Mammo QCManual
Mammo QCManual
Digital Mammography
Radiologist’s Section
COPYRIGHT NOTICE/PERMISSION
Copyright© 2018 American College of Radiology. All rights reserved.
No part of this document may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopy, recording or any information storage or retrieval system, without the prior
written consent of the American College of Radiology.
The is a registered trademark and service mark of the American College of Radiology.
Certain state laws do not allow limitations on implied warranties or the exclusion or limitation of certain
damages. If these laws apply to you, some or all of the above disclaimers, exclusions, or limitations may not apply
to you, and you might have additional rights.
DISCLAIMER ON FIGURES
The inclusion of photographs, illustrations or images of any manufacturer’s mammography or test equipment
(including mammography phantoms) in this manual does not imply endorsement of such mammography or
test equipment by the American College of Radiology. ACR includes these photographs, illustrations and images
only as educational aids to more clearly explain the steps that are needed to perform and evaluate the tests
outlined in the manual.
APPLICABLE LAW
By using any ACR product, you agree that the applicable federal law and the laws of the Commonwealth of
Virginia, without regard to principles of conflict of laws, will govern these Conditions of Use and any dispute of
any sort that might arise between you and ACR.
DISTRIBUTION
The Digital Mammography Quality Control Manual is provided free to all facilities accredited in the ACR
Mammography Accreditation Program and those applying for accreditation.
All others may purchase the manual from the ACR Education Catalog.
CITATION
Berns EA, Pfeiffer DE, Butler PF, et al. Digital Mammography Quality Control Manual. Reston, Va: American
College of Radiology; 2018.
Contents
PREFACE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
RADIOLOGIST’S SECTION
I. Revisions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
II. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
III. Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
IV. Responsibilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
V. QC Tests, Frequencies, and Timeframes for
Corrective Action . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
VI. Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
RADIOLOGIC TECHNOLOGIST’S SECTION
I. Revisions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
II. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
III. Technologist Quality Control. . . . . . . . . . . . . . . . . . . . . . . . 39
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
MEDICAL PHYSICIST’S SECTION
I. Revisions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
II. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
III. Mammography Equipment Evaluation and
Annual Survey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250
APPENDICES
I. Revisions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255
II. CAR Digital Mammography Phantom
Scoring Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .256
III. Artifact Evaluation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . 259
PREFACE, In February 2016, the FDA approved the ACR’s new Digital
Mammography Quality Control (QC) Manual and Digital
2 EDITION
ND
Mammography QC Phantom as an alternative standard for use in
routine QC of digital mammography equipment. This approval, as
an alternative standard, allowed mammography facilities, including
QC technologists and medical physicists, to use the new ACR
manual in lieu of manufacturers’ quality control manuals. However,
the approval specified that the new manual could not be used on
full-field digital mammography systems with advanced imaging
capabilities (e.g., tomosynthesis and contrast enhancement). In July
2018, the FDA approved the ACR’s QC procedures for digital breast
tomosynthesis (DBT) through an amendment to the 2016 alternative
standard. This updated 2018 Digital Mammography Quality Control
Manual integrates the new DBT QC procedures so that the entire
manual may be used for both full-field digital mammography and
DBT systems. [Author’s Note: Regarding digital mammography units
with contrast enhancement capability, the FDA has determined that
facilities may use this manual for QC of the 2D and DBT applications
of these units, and recommends that facilities follow manufacturer QC
procedures for contrast enhancement applications.]
I would again like to extend a sincere thank you to all who collaborated
on this 2nd Edition including the entire committee and ACR Staff. In
particular Doug Pfeiffer and Penny Butler dedicated countless hours to
bring this manual to fruition. Their efforts were invaluable.
September 2018
In February 2016, the FDA approved the ACR’s new Digital Mammography
Quality Control (QC) Manual and Digital Mammography QC Phantom
as an alternative standard for use in routine QC of digital mammography
equipment. Currently, the FDA requires digital mammography facilities to
perform QC for approved imaging systems, according to their respective
manufacturers’ quality control manuals. This approval, as an alternative
standard, allows mammography facilities, including QC technologists and
medical physicists, to use the new ACR manual in lieu of manufacturers’
quality control manuals. The FDA alternative standard specifies that the
new manual may be used only for full-field digital mammography systems
without advanced imaging capabilities (e.g., tomosynthesis and contrast
enhancement).
I would like to extend a personal thank you to Ms. Priscilla Butler who
kept the manual on track and did much of the work to bring this to fruition.
Her tireless work and perseverance are very much appreciated by me and
the entire committee. We could not have done this without her.
July 2016
Radiologist’s Section
Contents
I. REVISIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
II. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
III. DEFINITIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
A. Quality Assurance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
B. Quality Assurance Committee. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
C. Quality Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
D. QA/QC Procedures Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
IV. RESPONSIBILITIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
A. Radiologist’s Responsibilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1. Lead Interpreting Physician . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. Interpreting Physician. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Revisions
Date Page(s) Section Description of Revisions
RADIOLOGIST’S SECTION
November 2nd edition with digital breast
2018 tomosynthesis QC
RADIOLOGIST’S SECTION
removed from this manual. The ACR Mammography Quality Control
Manual for screen-film was the only ACR quality control manual
with a clinical image quality section; this change in the ACR Digital
Mammography Quality Control Manual makes it more consistent with
other ACR manuals and accreditation programs. The “Clinical Image
Quality Evaluation” section will be updated and made available on
the ACR website. Until a new document is provided, technologists and
radiologists can refer to the 1999 ACR Mammography Quality Control
Manual as well as training resources provided by the American Society
of Radiologic Technologists [5] for guidance on positioning and clinical
image quality.
• E
very imaging procedure is necessary and appropriate to the clinical
problem at hand
• Th
e images generated contain information critical to the solution of
that problem
• Th
e recorded information is correctly interpreted and results made
available in a timely fashion to the patient and her physician
• Th
e examination results in the lowest possible radiation exposure, cost,
and inconvenience to the patient consistent with imaging objectives
(3) Audit interpreting physician. Each facility shall designate at least one
RADIOLOGIST’S SECTION
interpreting physician to review the medical outcomes audit data at least once
every 12 months. This individual shall record the dates of the audit period(s) and
shall be responsible for analyzing results based on this audit. This individual
shall also be responsible for documenting the results and for notifying other
interpreting physicians of their results and the facility aggregate results. If
followup actions are taken, the audit interpreting physician shall also be
responsible for documenting the nature of the followup.
• O
ne or more radiologists, including the lead mammography
radiologist
• A medical physicist
• A facility manager
• O
ther radiology department personnel involved in caring for
mammography patients (this may include a nurse, desk attendant,
medical secretary, or others)
The QAC also may include medical and paramedical staff from outside
the radiology department, such as a surgeon, referring physician, nurse
educator, nurse from a comprehensive breast clinic, etc. Anyone who
helps provide care to the patient seeking breast cancer screening or
diagnosis should be considered as a member of the QAC since his or her
efforts affect the quality of care and the satisfaction of the patient.
C. Quality Control
QC is an integral part of QA and consists of a series of distinct technical
procedures that ensure the production of a satisfactory product, i.e., high-
quality screening or diagnostic images. Four steps are involved:
• A
cceptance testing to detect defects in equipment that is newly
installed or has undergone major repair
• D
etection and diagnosis of changes in equipment performance before
they become radiologically apparent
• V
erification of equipment performance after service has been
performed
• R
ecords of the QC tests performed by the QC technologist and
medical physicist
RADIOLOGIST’S SECTION
• Records of any corrective action as a result of the QA/QC testing
• R
ecords of routine and non-routine equipment service and
maintenance
• A
description of the orientation program for operators of
mammography equipment, including its duration and content
• M
ammographic techniques to be used, including pertinent
information on positioning, compression, appropriate image receptors,
imaging modes, and kVp-target-filter combinations if applicable
• P
recautions to protect the operator of the equipment, the patient,
and individuals in surrounding areas from unnecessary radiation
exposure
• P
olicies and employee responsibilities concerning personnel radiation
monitoring
A. Radiologist’s Responsibilities
1. Lead Interpreting Physician
RADIOLOGIST’S SECTION
oversight, and direction to all aspects of the QC program. One
mechanism the radiologist can use to demonstrate commitment to
QC is routine use of the new Optional System QC for Radiologist
procedure and form in the Radiologic Technologist’s Section to
quickly evaluate the entire mammographic imaging chain from
the radiologist workstation.)
11.
Oversee or designate a qualified individual to oversee the
radiation protection program for employees, patients, and other
individuals in the surrounding area.
12.
Ensure that records concerning employee qualifications,
mammography technique and procedures, infection control
procedures, QC, safety, and protection are properly maintained
and updated in the mammography QA/QC procedures manual.
2. Interpreting Physician
RADIOLOGIST’S SECTION
The medical physicist’s responsibilities relate to equipment performance
and include
1. Patient positioning
2. Compression
3. Image production
4. Image processing
5. Infection control
QC Tests, Frequencies, Before a facility QC technologist may start using the procedures in
the ACR Digital Mammography QC Manual for the first time on a
and Timeframes for
RADIOLOGIST’S SECTION
unit, the medical physicist must first conduct an annual survey of the
Corrective Action digital mammography unit and display devices using the manual and
the ACR Digital Mammography Phantom. This is important to provide
testing techniques and procedures for the QC technologist to use during
routine QC. After this is done, the QC technologist may start performing
routine QC using the ACR Digital Mammography QC Manual. For
current information and more details on transitioning to the ACR
Digital Mammography QC Manual, visit the Digital Mammography QC
Manual: Frequently Asked Questions on the ACR Digital Mammography
QC Manual Resources website.
Important: Before a facility may start using the procedures in the ACR
Digital Mammography QC Manual for the first time on a unit, the medical
physicist must first conduct an annual survey of the digital mammography
unit and display devices using the manual and the ACR Digital
Mammography Phantom.
The technologist and medical physicist will use the same forms they use
for most of the digital mammography tests to record the data and results
of the DBT tests. The previous digital mammography forms have been
revised to allow for this. New forms have been added for DBT-unique
tests (e.g., DBT Volume Coverage).
If any test fails, it is critical that the set up and techniques employed in the
test be checked and the test repeated to verify performance before initiating
corrective action. Upon confirmation of test failure, the MQSA Final Rule
requires that the source of the problem be identified and corrective action
be taken. In some cases, if test results fall outside of action limits, MQSA
requires that the source of the problem be identified and corrective action
taken before any further examinations are performed or any films are
processed using the component of the mammography system that failed
the test. Other test failures must be corrected within 30 days of the test
date (Table 1).
RADIOLOGIST’S SECTION
Test Minimum Frequency Corrective Action Timeframe
Technologist Tests
1. ACR DM Phantom Image Quality Weekly Before clinical use
2. C
omputed Radiography Cassette Weekly Before clinical use
Erasure (if applicable)
3. Compression Thickness Indicator Monthly Within 30 days
4. Visual Checklist Monthly Critical items: before clinical use; less
critical items: within 30 days
5. Acquisition Workstation Monitor QC Monthly Within 30 days; before clinical use for
severe defects
6. Radiologist Workstation Monitor QC Monthly Within 30 days; before clinical use for
severe defects
7. Film Printer QC (if applicable) Monthly Before clinical use
8. Viewbox Cleanliness (if applicable) Monthly Before clinical use
9. Facility QC Review Quarterly Not applicable
10. Compression Force Semiannual Before clinical use
11. Manufacturer Calibrations (if applicable) Mfr. Recommendation Before clinical use
Optional - Repeat Analysis As Needed Within 30 days after analysis
Optional - System QC for Radiologist As Needed Within 30 days; before clinical use for
severe artifacts
Optional - Radiologist Image Quality Feedback As Needed Not applicable
Medical Physicist Tests
1. M
ammography Equipment Evaluation (MEE) - MEE Before clinical use
MQSA Requirements
2. ACR DM Phantom Image Quality MEE and Annual Before clinical use
3. DBT Z Resolution MEE and Annual Within 30 days
4. Spatial Resolution MEE and Annual Within 30 days
5. DBT Volume Coverage MEE and Annual Before clinical use
6. Automatic Exposure Control System Performance MEE and Annual Within 30 days
7. Average Glandular Dose MEE and Annual Before clinical use
8. Unit Checklist MEE and Annual Critical items: before clinical use;
less critical items: within 30 days
9. Computed Radiography (if applicable) MEE and Annual Before clinical use
10. Acquisition Workstation Monitor QC MEE and Annual Within 30 days; before clinical
use for severe defects
11. Radiologist Workstation Monitor QC MEE and Annual Within 30 days; before clinical
use for severe defects
12. Film Printer QC (if applicable) MEE and Annual Before clinical use
13. Evaluation of Site’s Technologist QC Program Annual Within 30 days
14. E valuation of Display Device Technologist Annual Within 30 days
QC Program
15. Manufacturer Calibrations (if applicable) Mfr. Recommendation Before clinical use
16. Collimation Assessment MEE or Troubleshooting Within 30 days
Annual (DBT only)
MEE or Troubleshooting - Beam Quality MEE or Troubleshooting Before clinical use
(Half-Value Layer) Assessment
MEE or Troubleshooting - kVp Accuracy and MEE or Troubleshooting MEE: before clinical use;
Reproducibility troubleshooting: within 30 days
Troubleshooting - Ghost Image Evaluation Troubleshooting Before clinical use
Troubleshooting - Viewbox Luminance Troubleshooting Not applicable
Important: Corrective action for any test performed for MEEs must be
made before clinical use.
For detailed guidance on the FDA’s requirements for record retention, see
Quality Assurance Records and Retention of Personnel Records in the
FDA Policy Guidance Help System. All documentation must be made
available to MQSA inspectors during the annual inspection and the
facility’s accreditation body upon application and request.
Table 2 provides a complete list of all the digital mammography tests and
summarizes those tests that must be performed (as applicable) for the
modes that the facility uses clinically on its 2D and DBT systems.
Table 2. Required Tests for Imaging Modes Used on 2D and DBT Systems
RADIOLOGIST’S SECTION
Imaging Modes to Test
System Used for Both 2D and System Used for DBT
DBT Acquisition Acquisition Only
2D w/Add-on
Test 2D DBT Device DBT DBT
Technologist Tests
1. ACR DM Phantom Image Quality * & 2D*
2. C
omputed Radiography Cassette Erasure *
(if applicable)
3. Compression Thickness Indicator * * *
4. Visual Checklist *
5. Acquisition Workstation Monitor QC * *
6. Radiologist Workstation Monitor QC * *
7. Film Printer QC (if applicable) * *
8. Viewbox Cleanliness (if applicable) * *
9. Facility QC Review *
10. Compression Force * * *
11. Manufacturer Calibrations (if applicable) *
Medical Physicist Tests
1. Mammography Equipment Evaluation (MEE) * *
2. ACR DM Phantom Image Quality * & 2D*
3. DBT Z Resolution
4. Spatial Resolution *
5. DBT Volume Coverage
6. Automatic Exposure Control System Performance *
7. Average Glandular Dose *
8. Unit Checklist *
9. Computed Radiography (if applicable) *
10. Acquisition Workstation Monitor QC * *
11. Radiologist Workstation Monitor QC * *
12. Film Printer QC (if applicable) * *
13. Evaluation of Site’s Technologist QC Program *
14. Evaluation of Display Device Technologist QC Program * *
15. Manufacturer Calibrations (if applicable) *
16. Collimation Assessment * *
MEE or Troubleshooting - Beam Quality (Half-Value Layer * TF
*TF
[HVL]) Assessment
MEE or Troubleshooting - kVp Accuracy and Reproducibility *TF *TF
*Follow the procedures and frequency outlined for 2D QC
TF
HVL and kVp tests must include kVp, target, and filter combinations used for DBT
If a system is not used for 2D mammography and is only used for DBT
imaging, the technologist and the medical physicist are required to
evaluate phantom images for the DBT and 2D imaging modes. For the
Acquisition Workstation and Radiologist Workstation QC Tests, the 2D
procedures should be used to evaluate the 2D image.
RADIOLOGIST’S SECTION
commitment to QA by all parties involved in performing mammography
will validate that trust.
B. References
1. Taplin SH, Rutter CM, Finder C, Mandelson MT, Houn F, White E. Screening
mammography: clinical image quality and the risk of interval breast cancer.
AJR Am J Roentgenol. 2002;178:797-803.
2. McLelland R, Hendrick RE, Zinninger MD, Wilcox PA. The American
College of Radiology mammography accreditation program. AJR Am J
Roentgenol. 1991;157:473-479.
3. Destouet JM, Bassett LW, Yaffe MJ, Butler PF, Wilcox PA. The ACR’s
mammography accreditation program: ten years of experience since MQSA.
J Am Coll Radiol. 2005;2(7):585-594.
4. Department of Health and Human Services. FDA Mammography Quality
Standards, Final Rule. Fed Regist. 1997;62(208):55852-55994.
5. American Society of Radiologic Technologists. ASRT Continuing Education
Designed for Mammographers.
6. Sickles EA, D’Orsi CJ. ACR BI-RADS® Follow-up and Outcome Monitoring.
In: ACR BI-RADS® Atlas, Breast Imaging Reporting and Data System.
Reston, VA: American College of Radiology; 2013.
7. D’Orsi CJ, Sickles EA, Mendelson EB, et al. ACR BI-RADS® Atlas, Breast
Imaging Reporting and Data System. Reston, VA: American College of
Radiology; 2013.
I. REVISIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
II. INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
A. MQSA and Quality Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
B. Responsibilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
C. Q
C Tests, Frequencies, and Timeframes for
Corrective Action. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
III. Technologist Quality Control . . . . . . . . . . . . . . . . . . . . . . . . 39
A. Test Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
1. ACR Digital Mammography (DM) Phantom
Image Quality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2. Computed Radiography (CR) Cassette Erasure
(if applicable). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3. Compression Thickness Indicator . . . . . . . . . . . . . . . . . . . . 51
4. Visual Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
5. Acquisition Workstation (AW) Monitor QC. . . . . . . . . . . . 56
6. Radiologist Workstation (RW) Monitor QC. . . . . . . . . . . . 59
7. Film Printer QC (if applicable). . . . . . . . . . . . . . . . . . . . . . . . . 64
8. Viewbox Cleanliness (if applicable) . . . . . . . . . . . . . . . . . . . 67
9. Facility QC Review. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10. Compression Force. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
11. Manufacturer Calibrations (if applicable) . . . . . . . . . . . . 74
Optional – Repeat Analysis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Optional – System QC for Radiologist. . . . . . . . . . . . . . . . . . . 78
Optional – Radiologist Image Quality Feedback. . . . . . . . . 80
B. Quality Control Forms for 2D and DBT. . . . . . . . . . . . . . . . . . . . . 81
1. ACR Digital Mammography (DM) Phantom
Image Quality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
2. Computed Radiography (CR) Cassette
Erasure (if applicable). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
3. Compression Thickness Indicator . . . . . . . . . . . . . . . . . . . . 86
4. Visual Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
TECHNOLOGIST’S SECTION
Optional – Radiologist Image Quality Feedback. . . . . . . . 100
RADIOLOGIC
C. Management Forms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
1. ACR Technique and Procedure Summaries . . . . . . . . . . 103
2. Corrective Action Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
3. Facility Offsite Display Locations. . . . . . . . . . . . . . . . . . . . 107
4. QC Summary Checklists . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
D. Mobile Mammography. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
E. Infection Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
REFERENCES. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
A. Downloadable from the ACR Website
(www.acr.org). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
B. References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
C. Additional Resources. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
TECHNOLOGIST’S SECTION
RADIOLOGIC
TECHNOLOGIST’S SECTION
Note: Facilities may use the new ACR Digital Mammography QC Manual
RADIOLOGIC
for digital mammography systems with contrast enhancement, but only
for the 2D and DBT applications. Facilities should follow manufacturer QC
procedures for contrast enhancement applications.
Facilities may not use the new procedures with the small ACR
mammography phantom or use the newly developed ACR DM Phantom
with the old test procedures. The ACR DM Phantom was explicitly
designed as a tool for the ACR Digital Mammography Quality Control
Manual (for both 2D and DBT QC) and for the ACR Mammography
Accreditation Program to meet FDA MQSA phantom image quality and
dose requirements.
Note: The ACR Digital Mammography Phantom is required for the majority
of the QC tests in this manual. ACR-approved manufacturers of the
ACR Digital Mammography Phantom are listed on the ACR website. (For
more information, visit https://www.acraccreditation.org/Resources/
Digital-Mammography-QC-Manual-Resources.)
One of the major and most important aspects of the manual is a quarterly,
documented review of the QC program performed by the lead interpreting
radiologist, the facility manager, and the QC technologist. (See Facility
QC Review) Although many high-quality facilities have such a review
in place already, experience from the ACR Mammography Accreditation
Program has demonstrated that lack of communication among the
radiologists, managers, technologists, and medical physicists is a frequent
underlying cause of poor-quality mammography. A structured review
process can be beneficial to all facilities. These reviews may be done in
person or via teleconference or video conference.
TECHNOLOGIST’S SECTION
In addition, several optional procedures and forms are provided to aid in
quality improvement. The System QC for Radiologist test is a structured
RADIOLOGIC
tool allowing the radiologist to work with the technologist in evaluating
the entire system performance at the radiologist workstation using a
standard clinical image selected by the radiologist. The Radiologist Image
Quality Feedback procedure and form enables the interpreting radiologist
to notify the technical staff when asked to interpret sub-optimal cases
(and also provides a way to commend the technical staff for exceptional
quality). Using this procedure also helps facilities comply with the MQSA
Final Rule that “All interpreting physicians interpreting mammograms
for the facility shall follow the facility procedures for corrective action
when the images they are asked to interpret are of poor quality.”
The manual also includes several management forms to help organize and
reference important information. The ACR Techniques and Summaries
form provides a location to record techniques used for routine QC testing.
The manual also provides a central location to record all corrective action
for the digital mammography systems in the Corrective Action Log.
(This makes it easier to document these important actions and have it
available for review by the management, the lead interpreting radiologist,
the medical physicist, and MQSA inspectors.) A Facility Offsite Display
Locations form creates a convenient list of all locations of radiologist
workstations and film printers (if applicable) to help with the management
of all offsite QC.
B. Responsibilities
The MQSA Final Rule requires that the facility’s lead interpreting
physician (typically a radiologist) has the general responsibility of
ensuring that the quality assurance program meets all requirements.
Note: If the medical physicist determines that there is need for corrective
action, the facility should provide a copy of its medical physicist’s full report
to its equipment service engineer.
TECHNOLOGIST’S SECTION
physicist, should work together to ensure that all “display devices” are QC’d
and reviewed properly.
RADIOLOGIC
C. QC Tests, Frequencies, and Timeframes for
Corrective Action
Before a facility QC technologist may start using the procedures in
the ACR Digital Mammography QC Manual for the first time on a
unit, the medical physicist must first conduct an annual survey of the
digital mammography unit and display devices using the manual and
the ACR Digital Mammography Phantom. This is important to provide
testing techniques and procedures for the QC technologist to use during
routine QC. After this is done, the QC technologist may start performing
routine QC using the ACR Digital Mammography QC Manual. For
current information and more details on transitioning to the ACR
Digital Mammography QC Manual, visit the Digital Mammography QC
Manual: Frequently Asked Questions on the ACR Digital Mammography
QC Manual Resources website.
Important: Before a facility may start using the procedures in the ACR
Digital Mammography QC Manual for the first time on a unit, the medical
physicist must first conduct an annual survey of the digital mammography
unit and display devices using the manual and the ACR Digital Mammography
Phantom.
The technologist and medical physicist will use the same forms they use
for most of the digital mammography tests to record the data and results
of the DBT tests. The previous digital mammography forms have been
revised to allow for this. New forms have been added for DBT-unique
tests (e.g., DBT Volume Coverage).
If any test fails, it is critical that the set up and techniques employed in the
test be checked and the test repeated to verify performance before initiating
corrective action. Upon confirmation of test failure, the MQSA Final Rule
requires that the source of the problem be identified and corrective action
be taken. In some cases, if test results fall outside of action limits, MQSA
requires that the source of the problem be identified and corrective action
taken before any further examinations are performed or any films are
processed using the component of the mammography system that failed
the test. Other test failures must be corrected within 30 days of the test
date (Table 1).
TECHNOLOGIST’S SECTION
9. Facility QC Review Quarterly Not applicable
10. Compression Force Semiannual Before clinical use
11. Manufacturer Calibrations (if applicable) Mfr. Recommendation Before clinical use
RADIOLOGIC
Optional - Repeat Analysis As Needed Within 30 days after analysis
Optional - System QC for Radiologist As Needed Within 30 days; before clinical
use for severe artifacts
Optional - Radiologist Image Quality Feedback As Needed Not applicable
Medical Physicist Tests
1. Mammography Equipment Evaluation (MEE) - MEE Before clinical use
MQSA Requirements
2. ACR DM Phantom Image Quality MEE and Annual Before clinical use
3. DBT Z Resolution MEE and Annual Within 30 days
4. Spatial Resolution MEE and Annual Within 30 days
5. DBT Volume Coverage MEE and Annual Before clinical use
6. Automatic Exposure Control System MEE and Annual Within 30 days
Performance
7. Average Glandular Dose MEE and Annual Before clinical use
8. Unit Checklist MEE and Annual Critical items: before clinical use;
less critical items: within 30 days
9. Computed Radiography (if applicable) MEE and Annual Before clinical use
10. Acquisition Workstation Monitor QC MEE and Annual Within 30 days; before clinical
use for severe defects
11 Radiologist Workstation Monitor QC MEE and Annual Within 30 days; before clinical
use for severe defects
12. Film Printer QC (if applicable) MEE and Annual Before clinical use
13. Evaluation of Site’s Technologist QC Program Annual Within 30 days
14. Evaluation of Display Device Technologist QC Annual Within 30 days
Program
15. Manufacturer Calibrations (if applicable) Mfr. Recommendation Before clinical use
16. Collimation Assessment MEE or Troubleshooting Within 30 days
Annual (DBT only)
MEE or Troubleshooting - Beam Quality MEE or Troubleshooting Before clinical use
(Half-Value Layer) Assessment
MEE or Troubleshooting - kVp Accuracy and MEE or Troubleshooting MEE: before clinical use;
Reproducibility troubleshooting: within 30 days
Troubleshooting - Ghost Image Evaluation Troubleshooting Before clinical use
Troubleshooting - Viewbox Luminance Troubleshooting Not applicable
Important: Corrective action for any test performed for MEEs must be made
before clinical use.
Table 2 provides a complete list of all the digital mammography tests and
summarizes those tests that must be performed (as applicable) for the
modes that the facility uses clinically on its 2D and DBT systems.
Table 2. Required Tests for Imaging Modes Used on 2D and DBT Systems
Imaging Modes to Test
System Used for Both 2D and System Used for DBT
DBT Acquisition Acquisition Only
2D w/Add-On
Test 2D DBT Device DBT DBT
Technologist Tests
1. ACR DM Phantom Image Quality * & 2D*
2. omputed Radiography Cassette Erasure
C *
(if applicable)
3. Compression Thickness Indicator * * *
4. Visual Checklist *
5. Acquisition Workstation Monitor QC * *
TECHNOLOGIST’S SECTION
6. Radiologist Workstation Monitor QC * *
7. Film Printer QC (if applicable) * *
8. Viewbox Cleanliness (if applicable) * *
RADIOLOGIC
9. Facility QC Review *
10. Compression Force * * *
11. Manufacturer Calibrations (if applicable) *
Medical Physicist Tests
1. Mammography Equipment Evaluation (MEE) * *
2. ACR DM Phantom Image Quality * & 2D*
3. DBT Z Resolution
4. Spatial Resolution *
5. DBT Volume Coverage
6. Automatic Exposure Control System Performance *
7. Average Glandular Dose *
8. Unit Checklist *
9. Computed Radiography (if applicable) *
10. Acquisition Workstation Monitor QC * *
11. Radiologist Workstation Monitor QC * *
12. Film Printer QC (if applicable) * *
13. Evaluation of Site’s Technologist QC Program *
14. Evaluation of Display Device Technologist QC Program * *
15. Manufacturer Calibrations (if applicable) *
16. Collimation Assessment * *
MEE or Troubleshooting - Beam Quality (Half-Value Layer * TF
*TF
[HVL]) Assessment
MEE or Troubleshooting - kVp Accuracy and Reproducibility *TF *TF
*Follow the procedures and frequency outlined for 2D QC
TF
HVL and kVp tests must include kVp, target, and filter combinations used for DBT
If a system is not used for 2D mammography and is only used for DBT
imaging, the technologist and the medical physicist are required to
evaluate phantom images for the DBT and 2D imaging modes. For the
Acquisition Workstation and Radiologist Workstation QC Tests, the 2D
procedures should be used to evaluate the 2D image.
FREQUENCY Weekly, after relevant service, and upon installation of new equipment
(before clinical use).
TECHNOLOGIST’S SECTION
adipose tissue. If you have multiple phantoms, use the same phantom
each time on a given unit (Figure 1).
RADIOLOGIC
• ACR DM Phantom Image Quality form.
TEST PROCEDURE Important: Do not follow the phantom imaging instructions or technical
factors provided in the manufacturer’s QC manual. Be sure to follow the
instructions below. This technique must be the same as that used clinically
for a 4.2 cm thick compressed breast consisting of 50% glandular and 50%
adipose tissue, as defined by the FDA.
900.2 Definitions. (uu) Standard breast means a 4.2 centimeter (cm) thick com
pressed breast consisting of 50 percent glandular and 50 percent adipose tissue.
a. The pink wax insert is on the top side of the phantom and nearer
the chest wall
b. The phantom is centered left-to-right
c. The edge of the phantom is aligned with the chest wall edge of the
digital image receptor
6. At the acquisition workstation, select the imaging mode from the
ACR Technique and Procedure Summaries form. This is the imaging
mode and technique that would be used for a clinical screening
exam acquisition of a 4.2 cm thick compressed breast consisting of
50% glandular and 50% adipose tissue. If a combination exposure
mode (i.e. 2D plus DBT) is most commonly used clinically, use the
combination mode for this test, record the data from the 2D and DBT
a. Facility
b. Mammography Accreditation Program (MAP) ID number
c. Room ID
d. X-ray unit manufacturer and model
TECHNOLOGIST’S SECTION
8. For each weekly image acquisition, verify and use the following
technique parameters in the ACR Technique and Procedure
RADIOLOGIC
Summaries form:
a. AEC mode
b. Paddle and image receptor size
c. Paddle type (regular or flex)
d. Compression force
e. AEC cell position (if applicable)
f. Target/filter (if applicable)
g. kVp (if applicable)
h. Density setting (if applicable)
10. Record the following parameters that appear after the exposure on
the form:
a. The pink wax insert is on the top side of the phantom and nearer
the chest wall
b. The phantom is centered left-to-right
c. The edge of the phantom is aligned with the chest wall edge of the
digital image receptor
6. At the acquisition workstation, select the DBT imaging mode and
technique that would be used for a clinical exam acquisition of a
4.2 cm thick compressed breast consisting of 50% glandular and 50%
adipose tissue. If a combination exposure mode (i.e. 2D plus DBT)
is most commonly used clinically, use the combination mode for this
test, record the data from the 2D and DBT acquisitions, and use those
images for analysis. If a DBT-only mode is most commonly used
clinically for screening, use the DBT-only mode for this procedure
step. (If the system uses selectable AEC sensor positions, be sure to
use the same position each time the phantom is acquired.)
TECHNOLOGIST’S SECTION
7. Record or verify the following demographic information at the top of
RADIOLOGIC
the form:
a. Facility
b. MAP ID number
c. Room ID
d. X-ray unit manufacturer and model
8.
For each weekly image acquisition, verify and use the following
technique parameters in the ACR Technique and Procedure
Summaries form:
a. AEC mode
b. Paddle and image receptor size
c. Paddle type (regular or flex)
d. View or selected image
e. Compression force
f. AEC cell position (if applicable)
g. Target/filter (if applicable)
h. kVp (if applicable)
i. Density setting (if applicable)
10. Record the following parameters that appear after the exposure on
the form:
For DBT images, scroll to the one slice in which the test objects are
3.
best visualized. If “slices” are not available, then proceed to use the slab
in which the objects are best visualized. (See the ACR Technique and
Procedure Summaries form for guidance.) Record the best visualized
slice or slab.
4. Adjust the window width (WW) and window level (WL) settings to
optimize visualization of test objects (the test objects will be scored
in the next section). It is important not to use unreasonably narrow
WWs, which may enhance the appearance of artifacts. For guidance
on approximate values, refer to those used by your medical physicist
and recorded on the technologist’s ACR Technique and Procedure
Summaries form.
7.
Record the absence or presence of artifacts on the form as a pass or
fail (P or F).
TECHNOLOGIST’S SECTION
RADIOLOGIC
9. Do not deduct for artifacts. (Deducting for artifacts is no longer part
of the ACR DM Phantom scoring procedure.)
Figure 5. ACR DM Phantom wax insert map (test object sizes are not to scale).
b. Fibers
i. The fibers are manufactured to be 10 mm in length. If the
entire length of the fiber is not visible, measure it using the
display device’s electronic calipers.
ii. Count each fiber as 1 point if 8 mm or more of the fiber is
visible in the correct location and orientation.
iii. Count a fiber as ½ point if the fiber appears to be equal to or
greater than 5 mm and less than 8 mm in length and is in the
correct location and orientation.
iv. If a small gap in the fiber is visible, and it is less than the width
of the fiber, count the fiber as a full or half point depending
on the total visible length.
c. Speck groups
i. Count each speck group as 1 point if 4 to 6 specks are visible
in the proper locations in the group.
ii. Count a speck group as ½ point if 2 or 3 specks are visible in
the proper locations in the group.
d. Masses
i. Count each mass as 1 point if an object is visible in the correct
location and the mass appears to be generally circular against
the background (at least ¾ of the border is continuous and
generally round).
ii. Count a mass as ½ point if a mass-like object is visible in
the correct location but does not have a generally circular
TECHNOLOGIST’S SECTION
appearance (greater than ½ but less than ¾ of a circle).
e. Enter the final scoring result in each category (fibers, speck
RADIOLOGIC
groups, and masses) on the form.
11. If 2D add-on devices are used clinically with DBT, repeat the above
steps to score the resulting 2D image.
Test
Object Full Point Half Point
Fibers (6) • Full length visible (≥8 mm • At least half of length
long) visible (≥5 and <8 mm long)
• Correct location • Correct location
• Correct orientation • Correct orientation
• 1 break allowed (must be • 1 break allowed (must be
≤ width of fiber) ≤ width of fiber)
Speck • 4 to 6 specks visible • 2 to 3 specks visible
Groups (6) • Correct locations • Correct locations
Masses (6) • Density difference visible • Density difference visible
• Border is continuous and • Border is not continuous or
generally circular (≥ ¾ generally circular (≥ ½ and
border visible) < ¾ border visible)
• Correct location • Correct location
PRECAUTIONS AND If any of these quantities are outside of the action limits stated below, the test
CAVEATS should be repeated. Make sure the correct exposure mode has been used.
Note: If the facility replaces all of its PSP plates with new ones, the medical
physicist must perform a Mammography Equipment Evaluation consisting
of applicable tests.
TIMEFRAME FOR All failures of required items must be corrected before clinical use.
CORRECTIVE ACTION
FREQUENCY Weekly.
TECHNOLOGIST’S SECTION
• Computed Radiography Cassette Erasure form.
RADIOLOGIC
TEST PROCEDURE 1. U
sing the menu on the CR reader, select the option for CR cassette
erasure.
2. Place the CR cassette into the CR reader and initiate the erasure
process.
5. If the CR cassette fails again, enter the date, cassette number, reason
for failure, and action taken under “Action Taken on Cassette” at the
bottom of the form.
6. Date and initial the form to indicate the test was completed.
7. If failing cassette(s) are put back into service at any time, document this
on the form along with the date and description of service performed.
PERFORMANCE CRITERIA 1. E
ach cassette should successfully pass the CR cassette erasure
AND CORRECTIVE ACTIONS process.
3. If an error message occurs again, pull the cassette from service and
investigate the error message.
TIMEFRAME FOR Cassettes that do not successfully pass the erasure process must be taken
CORRECTIVE ACTION out of service immediately. After service or repair, the cassette should
undergo the erasure process before being put back in to service.
TECHNOLOGIST’S SECTION
compression paddle or bucky.
– If tape is used, cover the sides of the roll (by using a thin plastic
bag or paper) to prevent adhesive from sticking to the equipment.
RADIOLOGIC
– Be sure to set aside the compression thickness indicator phantom
and label the object for use only for this test.
• A ruler with a mm/cm scale.
2. Measure the thickness of the phantom using the same units provided
on the indicator (cm or mm) and record this value on the form
(Figure 6).
3. In contact mode, place the phantom on the breast support so that it
is centered laterally and aligned flush with the chest wall edge of the
support (Figure 6).
DATA ANALYSIS AND 1. Subtract the actual, measured thickness of the phantom from the
INTERPRETATION indicated thickness.
PRECAUTIONS AND Many systems use the indicated compression thickness to drive the
CAVEATS selection of initial kVp and filter under automatic exposure control.
Omitting such a test may have an impact on image quality and patient
dose, as suboptimal imaging techniques may be selected during imaging
if the compression thickness is not accurate.
2. If the test fails, recheck the measurement of the phantom with the
ruler and the setup of the equipment. Repeat the test.
TIMEFRAME FOR The source of the problem must be identified and corrective action taken
TECHNOLOGIST’S SECTION
CORRECTIVE ACTION within 30 days.
RADIOLOGIC
4. Visual Checklist
OBJECTIVES To ensure that digital mammographic x-ray system indicator lights,
displays, mechanical locks and detents are working properly and that the
mechanical rigidity and stability of the equipment is optimum.
FREQUENCY Monthly, after relevant service, and upon installation of new equipment
(before clinical use).
TEST PROCEDURE 1. Review all items listed on the visual checklist and indicate their status.
Be sure to rotate the C-arm the way you would for patient imaging.
2. If specific checks that are not on the form are recommended by your
medical physicist, add them to the checklist.
Items on the list that are not applicable to your system (e.g., “condition of
imaging plates and cassettes” if computed radiography [CR] is not used)
should be marked “not applicable” (NA).
PERFORMANCE CRITERIA The following tests listed in the Visual Checklist are critical, and failures
AND CORRECTIVE ACTIONS must be repaired or replaced before clinical use, as applicable:
TECHNOLOGIST’S SECTION
• Cables must be safely positioned.
• The C-arm motion must be smooth.
RADIOLOGIC
• The compression paddle motion must be smooth.
Items missing from the room should be replaced. Malfunctioning
equipment should be reported to the authorized service representative
for repair or replacement as soon as possible.
TIMEFRAME FOR Failures of critical tests must be corrected before clinical use; failures of
CORRECTIVE ACTION less critical tests must be corrected within 30 days.
FREQUENCY Monthly, after relevant service, and upon installation of new equipment
(before clinical use).
TEST EQUIPMENT • ry, soft, lint-free cloth or cleaning tissue recommended by your AW
D
manufacturer.
Note: Any other cleaning methods may lead to damage of the anti-reflective
screen coating. Please follow your AW manufacturer’s recommendations for
proper cleaning and cleaning material. If you don’t have them, ask the AW
manufacturer.
Figure 8. Test patterns: A. AAPM TG18-QC test pattern. B. SMPTE test pattern.
3. After drying, recheck the monitor surface to be sure the items noted
in step 1 have been eliminated. If they were not, clean the monitor
TECHNOLOGIST’S SECTION
again.
RADIOLOGIC
and Corrective Actions).
2. Display the test pattern on the monitor. (If an appropriate test pattern
is not available on the AW, skip this test.)
3. Evaluate the test pattern for the following visible targets and record
pass or fail on the form (see the ACR Technique and Procedure
Summaries form for further guidance on monitor test pattern
evaluation):
2. Review the results and verify that all tests have passed.
PRECAUTIONS AND CAVEATS Ideally, monitor screens should be free of dust, fingerprints, and other
marks. Similarly, there should be no “shiny” patches or obvious non-
uniformities on the surface. As described below, significant blemishes
that interfere with the interpretation or QC of images must be corrected.
OBJECTIVES • o ensure that radiologist workstation monitors are clean and free
T
from dust, fingerprints, and other marks that may interfere with
clinical information.
TECHNOLOGIST’S SECTION
Important: Monitor Manufacturer Automated Tests are required if such
RADIOLOGIC
tests are available in the manufacturer’s documentation.
FREQUENCY Monthly, after relevant service, and upon installation of new workstations
(before clinical use).
TEST EQUIPMENT • ry, soft, lint-free cloth or cleaning tissue recommended by your RW
D
manufacturer.
Note: : Any other cleaning methods may lead to damage of the anti-reflective
screen coating. Please follow your RW manufacturer’s recommendations for
instructions on proper cleaning and cleaning material. If you don’t have
them, ask the RW manufacturer.
3. After drying, recheck the monitor surface to be sure the items noted
in step 1 were eliminated. If they were not, clean the monitor again.
ACR DM Phantom
1. Display a phantom image that was acquired on one of the digital
mammography units per instructions in Test #1 ACR Digital
Mammography Phantom Image Quality.
2. Evaluate for artifacts and score the phantom test objects as in Test #1
ACR Digital Mammography Phantom Image Quality.
2. Evaluate the test pattern for the following visible targets and record
pass or fail on the form (see the ACR Technique and Procedure
Summaries form for further guidance on monitor test pattern
evaluation):
2. Review the results and verify that all tests have passed.
PRECAUTIONS AND Ideally, monitor screens should be free of dust, fingerprints, and other
CAVEATS marks. Similarly, there should be no “shiny” patches or obvious non-
uniformities on the surface. As described below, significant blemishes
TECHNOLOGIST’S SECTION
that interfere with the interpretation or QC of images must be corrected.
If there are questions regarding the significance of a monitor blemish, the
lead interpreting radiologist should be consulted.
RADIOLOGIC
Most problems can be corrected by cleaning according to the
manufacturer’s instructions. However, if cleaning does not correct the
problem, the manufacturer should be contacted to evaluate and correct
the problem.
ACR DM Phantom
1. Artifacts must not be clinically significant. This aspect of the test
fails if any artifacts are in a location that could impact clinical
interpretation and
TECHNOLOGIST’S SECTION
RADIOLOGIC
FREQUENCY • onthly (if printer is used less than monthly, before clinical films are
M
printed).
Note: Film printer QC is only required if they are used clinically for
mammography (i.e., for interpretation and to provide images to referring
physicians and patients). If such is the case, it is important that the facility
documents in its QC logs that that the film printer is not used clinically.
• Densitometer.
TEST PROCEDURE 1. Print the ACR DM Phantom image (acquired in Test #1 ACR Digital
Mammography Phantom Image Quality) without adjusting any
parameters (window width/window level, sizing, etc.) using the film
size used for the majority of clinical printing. (Refer to the ACR
Technique and Procedure Summaries form.) Print the digital images
without magnification or minification and as close to “true size” as
possible. The ACR recommends printing the phantom images so that
it is within 25% of the actual phantom size.
2. On the form, note the workstation used to print the image.
Contrast
1. Record the background OD from above in the “Contrast” section of
the form.
TECHNOLOGIST’S SECTION
3. Subtract the background OD from outside of the cavity from the OD
inside the cavity and record this on the form. This is the film contrast.
RADIOLOGIC
Contrast = Cavity OD–Background OD
2. See the ACR Technique and Procedure Summaries form for a detailed
schematic of where to measure the ODs.
PRECAUTIONS AND Many facilities never actually provide a final interpretation report
CAVEATS based on the hard copy images. However, even if a facility is using the
printer only to provide final interpretation-quality hardcopy images to
representatives, health-care providers, patients, or for retention purposes,
it still needs to perform all the required printer QC tests at appropriate
frequencies or prior to printing clinical images for patients and health-
care providers or for retention, whichever is less frequent. See the FDA
MQSA website for the most current requirements.
Contrast
Contrast (cavity OD – background OD) must be ≥0.1.
Dmax
The Dmax must be ≥3.1 (≥3.5 is recommended).
TIMEFRAME FOR All failures of required items must be corrected before clinical use.
CORRECTIVE ACTION
FREQUENCY Monthly, after relevant service, and upon installation of new equipment
(before clinical use)
TECHNOLOGIST’S SECTION
TEST PROCEDURE 1. Clean viewbox surfaces using cleaner recommended by viewbox
manufacturer and soft paper or cotton towels.
RADIOLOGIC
2. Ensure that all marks have been removed.
PRECAUTIONS AND High-quality viewboxes are still important in digital imaging because
CAVEATS comparison images may only be available on film. The accuracy of
the diagnosis and the efficiency of the radiologist are influenced by
the conditions under which the mammograms are viewed. Viewing
conditions may affect the diagnostic potential of even the best quality
mammograms. These conditions are determined by the luminance of the
viewboxes, the ambient room illumination or the amount of light falling
on the viewbox surface, and good masking of films on the viewbox.
PERFORMANCE CRITERIA AND Any marks that are not easily removed with window cleaner must be
CORRECTIVE ACTIONS removed with a safe and appropriate cleaner. If viewboxes appear non-
uniform, all of the fluorescent lamps must be replaced as soon as possible.
If viewbox masks are difficult to use, appropriate service or modifications
should be requested.
TIMEFRAME FOR Failures must be corrected before clinical images are viewed on the
CORRECTIVE ACTION viewbox.
9. Facility QC Review
OBJECTIVES To ensure the lead interpreting radiologist and facility manager are aware
that all QC tests are performed at the required frequencies, that data
are collected appropriately, that results are adequately documented, that
corrective action is taken, and that no patient exams are conducted when
tests requiring correction before clinical use failed.
FREQUENCY Quarterly.
TECHNOLOGIST’S SECTION
radiologist and facility manager. These reviews may be done in person or
remotely (e.g., via teleconference or video conference).
RADIOLOGIC
1. Enter QC results from the most recent Medical Physicist’s ACR Digital
Mammography (DM) QC Test Summary form into the Facility QC
Review form prior to the meeting.
3. Review the most recent quarter of QC data with both the lead
interpreting radiologist and facility manager.
5. If any tests fail, note that corrective action was documented.
PRECAUTIONS AND The MQSA Final Rule specifies that the lead interpreting physician has
CAVEATS “the general responsibility of ensuring that the quality assurance program
meets all MQSA QA [quality assurance] requirements.” This responsibility
cannot be delegated to individuals such as the medical physicist or the
QC technologist. Routine and appropriate conduct of the Facility QC
Review demonstrates that this responsibility is being properly exercised.
The lead interpreting radiologist, along with the facility manager, must
review the QC test results at least quarterly or more frequently if problems
are noted. It is particularly important for the lead interpreting radiologist
and the facility manager to note that all tests and necessary corrective
action are properly documented. Too often, investigations performed
by MQSA inspectors, MQSA certifying bodies, and mammography
accreditation bodies for non-compliance and quality issues demonstrate
that neither facility management nor lead interpreting radiologists were
aware that QC was not being performed at the required intervals or
documented at their facilities.
PERFORMANCE CRITERIA 1. If tests are not being done at an appropriate level or frequency, or if
AND CORRECTIVE ACTIONS corrective action for failures are not implemented and/or documented,
the facility needs to determine the reasons and allocate appropriate
training and/or time so the work is done properly.
2. If there are questions regarding the conduct of the test or data
interpretation, the medical physicist should be consulted for assistance
and additional training.
OBJECTIVES To ensure that the mammography system can provide adequate compression
in both manual fine-adjustment and hands-free, initial power-drive modes,
that the equipment does not allow too much compression to be applied when
used in initial power-drive mode, and that adequate compression can be
maintained throughout image acquisition.
TEST EQUIPMENT • alibrated bathroom scale (The scale should be a flat, conventional,
C
analog type. Digital scales sample the data and may not respond properly
as additional pressure is applied slowly to the scale. Digital scales designed
TECHNOLOGIST’S SECTION
specifically to measure compression force may be used.)
• Several towels.
RADIOLOGIC
• ther calibrated tools specifically designed to measure compression
O
force (such as digital gauges or compression force tools) may also be
used.
TEST PROCEDURE 1. Place a towel on the breast support surface (to protect the image
receptor), then place the bathroom scale on the towel with the dial
or read-out positioned for easy reading. Locate the center of the scale
directly under the compression device (Figure 9).
2. Place one or more towels on top of the scale to prevent damage to the
compression paddle.
Figure 9. Setup for the compression force test. Note the towel covering the image
receptor under the scale and the towel on top of the scale to protect the paddle.
TECHNOLOGIST’S SECTION
the maximum force. Failure to do so may lead the person conducting the
test to report an artificially low maximum compression force. This could
RADIOLOGIC
lead to an inappropriate failure of the initial power drive compression
device quality control test. See FDA Guidance.
Note: This is the minimum compression force that is available to use on the
unit. It is not the minimum force that must be used to compress the patient.
TIMEFRAME FOR The source of the problem must be identified and corrective action taken
CORRECTIVE ACTION before any examinations are performed.
TEST PROCEDURE 1. See manufacturer’s documentation for exact procedure steps. The
medical physicist should help the facility in locating and implementing
these procedures.
PRECAUTIONS AND Most manufacturers provide specific instructions for system calibrations
CAVEATS (e.g., detector calibration). See the manufacturer’s documentation for
precautions and caveats.
PERFORMANCE CRITERIA The unit must meet the prescribed periodic calibrations.
AND CORRECTIVE ACTIONS
FREQUENCY As needed. (In order for the repeat rates to be meaningful, a patient
volume of at least 250 patients is needed, if possible. Some facilities may
choose to conduct routine repeat analyses. Forms are provided for both
monthly and quarterly checks.)
TECHNOLOGIST’S SECTION
the following 2 key elements:
• Count of the total # of exposures made during the evaluation period
• % repeats during the same period: (# Repeat Exposures/Total #
RADIOLOGIC
Exposures) × 100%
TEST PROCEDURE 1. Count all repeated images for the length of time needed to examine at
least 250 patients (including those repeated images that were saved in
the patient’s medical record).
2. Also count the total number of patient exposures made during the
same time period.
3. Sort the repeated images into the categories listed in the Tally Sheet or
Daily Counting Sheet.
facility’s analysis and overall % repeats may be determined for the entire
facility or for each mammography unit at the facility. (Many facilities
choose to calculate both a facility-wide repeat rate and a repeat rate for
each unit in order to determine if equipment problems result in higher
repeat rates.) However, if a facility chooses to collect and analyze its repeat
data, it is important that the facility does this in a consistent manner so
that valid trends may be noted.
There is a real danger that technologists may alter their routine procedures
or criteria for accepting images if they know their repeated images will be
analyzed. This should be avoided.
PERFORMANCE CRITERIA The overall repeat rate ideally should be 2% or less, but a rate of 5% is
AND CORRECTIVE ACTIONS probably adequate if the radiologist and medical physicist agree that this
is a reasonable level. Too low or no repeat rate may also indicate that
TECHNOLOGIST’S SECTION
poor image quality is being tolerated. These rates should be based on a
volume of at least 250 patients to be meaningful. A “Reason for Repeat”
RADIOLOGIC
that is significantly higher than the others indicates an area for potential
improvement.
If the repeat rate exceeds the selected acceptance level of either 2% or 5%,
or if the repeat rate changes from the previously measured rate by more
than 2%, the change should be investigated and corrective action taken, if
necessary. For example, if the previous repeat rate was 1.8% and the new
repeat rate is 4.2%, then the follow-up described above is required.
TIMEFRAME FOR The source of the problem should be identified and corrective action
CORRECTIVE ACTION taken within 30 days of the repeat analysis.
FREQUENCY As needed.
4.
Example: A radiologist can sit at one workstation and view the images
for all the digital mammography units or CR readers within a facility.
The radiologist does not need to evaluate every monitor at every
workstation.
3. Place the same MLO image from the same case and breast on each
monitor (or, for wide-screen monitors, on each side of the monitor, if
possible).
4. Evaluate the images for artifacts and check the appropriate boxes.
PRECAUTIONS AND This test can be performed by the radiologist for all digital mammography
CAVEATS or CR units on the same workstation.
PERFORMANCE CRITERIA If any box is checked “Yes”, take appropriate corrective action.
AND CORRECTIVE ACTIONS
TIMEFRAME FOR 1. If an image quality problem or artifact impedes clinical interpretation,
CORRECTIVE ACTION seek service before using the workstation to interpret images.
2. If the artifact does not impede clinical interpretation, seek service
within 30 days.
TECHNOLOGIST’S SECTION
RADIOLOGIC
FREQUENCY As needed.
TEST PROCEDURE 1. The radiologist should check image quality assessment observations
as the cases are interpreted.
2. The radiologist should also give positive feedback if the image quality
is excellent.
DATA ANALYSIS AND The facility management and staff should routinely evaluate this
INTERPRETATION information to examine possible areas for quality improvement.
PRECAUTIONS AND The MQSA Final Rule requires that “All interpreting physicians
CAVEATS interpreting mammograms for the facility shall follow the facility
procedures for corrective action when the images they are asked to
interpret are of poor quality.” Utilizing this procedure will help facilities
meet this requirement.
TECHNOLOGIST’S SECTION
For simplicity and uniformity, the DBT tests are intended to use the same
forms as the 2D tests. Document that the DBT tests were performed by
selecting the correct image “mode” on the form. (If you are using an
RADIOLOGIC
electronic version of the form, use the pull-down menu to select the
mode.) For example, on the ACR DM Phantom Image Quality form,
create a form for testing the 2D mode by entering “2D” in the heading
of the form. Then copy the form to create one to test the DBT mode and
enter “DBT” in the heading of the form. Note that for DBT you will need
to use a second or third page of the form for complete QC documentation.
The forms are also downloadable as Excel spreadsheets from the ACR
Digital Mammography QC Manual Resources website (go to Digital
Mammography Quality Control Test Forms). Although they have been
designed to help you record and analyze your QC results on a computer,
they may also be printed and completed manually.
2. C
omputed Radiography (CR) Cassette Erasure
(if applicable)
4. Visual Checklist
9. Facility QC Review
C. Management Forms
TECHNOLOGIST’S SECTION
QC events needing corrective action that occur within a mammography
facility. This includes problems detected during the Radiologic
Technologist QC Tests, Medical Physicist Equipment Evaluation and
RADIOLOGIC
Annual Survey, and any other miscellaneous events that may arise.
The form should be reviewed annually and updated when there are any
changes in locations or facilities that provide interpretations.
4. QC Summary Checklists
To assist with the oversight of the QC program, mammography checklists
are provided to record the Weekly, Monthly, Quarterly, and Semi-Annual
Tests for both the mammography system and the display devices. These
checklists provide quick reminders of when QC tasks are due and also
provide records indicating that the tasks have been completed in a timely
manner. In addition, the ACR will request copies of these checklists
during the accreditation process to document that all required QC tests
were performed at the required frequencies.
All dates should be filled in prior to use of the checklist. Each time a
task is completed, the individual carrying out the task should initial
the appropriate area on the checklist. If a test is not performed because
a system is not in use (for example, one of the RW monitors is being
repaired, is not being clinically used, or is not available to test), the QC
technologist should put an “X” in the box and include a note of why the
test is not being performed.
DIGITAL This checklist is for the digital mammography imaging unit located at the
MAMMOGRAPHY UNIT mammography facility.
QC SUMMARY CHECKLIST
FACILITY DISPLAY DEVICE This checklist is for the display devices (including the RW monitors, film
QC SUMMARY CHECKLIST printers, and viewboxes) that may or may not be at a different location
from the mammography imaging system.
D. Mobile Mammography
Mobile mammography systems are subject to all the QC requirements and
frequencies that apply to stationary mammography systems, including
DBT. In addition, the tests outlined in Table 5 must be performed after
each move of the mobile system at each examination location. For
example, if the mobile unit moves to 2 different locations within the same
day, all the applicable tests outlined below must be performed prior to
using the equipment at each location, so the tests would be performed
twice on that day.
Note: The tests outlined in Table 5 must be performed after each move of
the mobile system at each examination location.
TECHNOLOGIST’S SECTION
The evaluations should be made at the mobile unit’s site of operation
to verify that the mobile unit is performing adequately before any
patient examinations are conducted, before the mobile RW is used for
RADIOLOGIC
interpretation, or before the mobile film printer is used to generate
hardcopy of patient images. If the image quality is inadequate, then
immediate corrective action is required and the results of the corrective
action need to be verified via repeat testing before proceeding.
QC Tests
• ACR DM Phantom Image Quality - after each move and prior to
examining patients
• Compression Thickness Indicator - after each move and prior to
examining patients
• Radiologist Workstation (RW) Monitor QC (for mobile RW only) - after
each move and prior to interpretation
• Film Printer QC (for mobile film printers only) - after each move and prior
to printing patient images
900.7 Mobile Units. The facility shall verify that mammography units used to
produce mammograms at more than one location meet the requirements
in paragraphs (e)(1) through (e)(6) of this section. [Quality assurance—
equipment requirements] In addition, at each examination location, before any
examinations are conducted, the facility shall verify satisfactory performance
of such units using a test method that establishes the adequacy of the image
quality produced by the unit.
E. Infection Control
OBJECTIVE To prevent and control the spread of infection to employees, patients, and
visitors within the mammography facility.
• Disposable wipes
PROCEDURE STEPS 1. All surfaces in contact with the patient are to be wiped clean with a
facility-approved disinfectant at the end of each exam.
2. All linens are for single patient use. Following use, they are to be
deposited in appropriate bags for transport to laundry.
900.13 Infection control. Facilities shall establish and comply with a system
specifying procedures to be followed by the facility for cleaning and disinfecting
mammography equipment after contact with blood or other potentially
infectious materials. This system shall specify the methods for documenting
facility compliance with the infection control procedures established and shall:
(i) Comply with all applicable Federal, State, and local regulations pertaining to
infection control; and
(ii) Comply with the manufacturer’s recommended procedures for the cleaning
and disinfection of the mammography equipment used in the facility; or
(iii) If adequate manufacturer’s recommendations are not available,
comply with generally accepted guidance on infection control, until such
recommendations become available.
TECHNOLOGIST’S SECTION
RADIOLOGIC
B. References
1. Department of Health and Human Services. FDA Mammography Quality
Standards, Final Rule. Fed Reg., 1997;62(208):55852-55994.
2. U.S. Food and Drug Administration. Mammography Quality Standards Act
(MQSA) Policy Guidance Help System.
3. FDA Mammography Quality Standards Act and Program, Facility
Certification and Inspection (MQSA) - Digital Accreditation.
4. Hendrick RE, Bassett L, Botsco MA, et al. Mammography Quality Control
Manual. Reston, Va: American College of Radiology; 1999.
5. Samei E, Badano A, Chakraborty D, et al. Assessment of display performance
for medical imaging systems: executive summary of AAPM TG18 report.
Med Phys. 2005;32:1205-1225.
6. Gray JE. Use of the SMPTE test pattern in picture archiving and
communication systems. J Digit Imaging. 1992;5:54-58.
C. Additional Resources
– Berns EA, Hendrick RE, Cutter GR. Optimization of technique factors for a
silicon diode array full-field digital mammography system and comparison
to screen-film mammography with matched average glandular dose. Med
Phys. 2003;30:334-340.
– Bloomquist AK, Yaffe MJ, Pisano ED, et al. Quality control for digital
mammography in the ACRIN DMIST trial: part I. Med Phys 2006;33:719-736.
– Kruger RL, Schuler BA. A survey of clinical factors and patient dose in
mammography. Med Phys. 2001;28:1449-1454.
– Quality Control Manual Template for Manufacturers of Displays and
Workstations Devices Labeled for Final Interpretation in Full-Field Digital
Mammography. Rosslyn, Va: National Electrical Manufacturers Association;
XR 22-2006; 2006.
– Quality Control Manual Template for Manufacturers of Hardcopy Output
Devices Labeled for Final Interpretation in Full-Field Digital Mammography.
Rosslyn, Va: National Electrical Manufacturers Association; XR 23-2006; 2006.
I. REVISIONS.......................................................................... 119
II. INTRODUCTION............................................................... 120
A. MQSA and Quality Control............................................................... 120
B. Responsibilities..................................................................................... 123
C. Q
C Tests, Frequencies, and Timeframes for
Corrective Action................................................................................. 126
D. S urveys of Systems with Multiple Units and
Display Devices (Including Offsite Equipment)........................ 131
1. Mammography Equipment Evaluation (MEE) –
All New Digital Mammography Units and
Display Devices........................................................................... 132
2. Mammography Equipment Evaluation (MEE) –
New Digital Mammography Units
(with Existing Display Devices).............................................. 132
3. Mammography Equipment Evaluation (MEE) –
New Display Devices (with Existing Digital
Mammography Units)............................................................... 133
4. Annual Surveys............................................................................ 134
5. Major Component Service/Upgrade/Replacement/
Repair.............................................................................................. 135
E. Equipment Adjustments, Changes, or Repairs.......................... 137
III. MAMMOGRAPHY EQUIPMENT EVALUATION
AND ANNUAL SURVEY................................................... 139
A. Test Procedures.................................................................................... 139
1. Mammography Equipment Evaluation (MEE) –
MQSA Requirements for Equipment................................... 139
2. ACR Digital Mammography (DM) Phantom
Image Quality............................................................................... 143
3. DBT Z Resolution........................................................................ 157
4. Spatial Resolution....................................................................... 161
5. DBT Volume Coverage.............................................................. 164
6. Automatic Exposure Control System
Performance................................................................................. 167
7. Average Glandular Dose.......................................................... 171
PHYSICIST’S SECTION
MQSA Requirements................................................................. 220
MEDICAL
2. ACR DM Phantom Image Quality.......................................... 221
3. DBT Z Resolution........................................................................ 222
4. Spatial Resolution....................................................................... 223
5. DBT Volume Coverage.............................................................. 224
6. Automatic Exposure Control System Performance........ 225
7. Average Glandular Dose.......................................................... 227
8. Unit Checklist............................................................................... 228
9. Computed Radiography (If Applicable)............................... 229
10. Acquisition Workstation (AW) Monitor QC....................... 231
11. Radiologist Workstation (RW) Monitor QC....................... 232
12. Film Printer QC (If Applicable)................................................. 233
PHYSICIST’S SECTION
MEDICAL
Note: Facilities may use the new ACR Digital Mammography QC Manual
for digital mammography systems with contrast enhancement, but only
for the 2D and DBT applications. Facilities should follow manufacturer QC
procedures for contrast enhancement applications.
PHYSICIST’S SECTION
area and provide the same attenuation as the small ACR mammography
phantom used in the 1999 Mammography Quality Control Manual, which
approximates a 4.2 cm thick compressed breast consisting of 50% glandular
MEDICAL
and 50% adipose tissue. If you have multiple phantoms, use the same
phantom each time on a given unit (Figure 1).
Note: The ACR Digital Mammography Phantom is required for the majority
of the QC tests in this manual. ACR-approved manufacturers of the
ACR Digital Mammography Phantom are listed on the ACR website.
(For more information, visit https://www.acraccreditation.org/Resources/
Digital-Mammography-QC-Manual-Resources.)
B. Responsibilities
The MQSA Final Rule requires that the facility’s lead interpreting physician
(typically a radiologist) has the general responsibility of ensuring that the
quality assurance program meets all requirements.
In a facility where more than one technologist does mammography,
one technologist must be assigned the responsibilities of QC (the QC
technologist). Other qualified individuals may perform specific QC tests,
but they must be reviewed and evaluated by the primary QC technologist.
The primary QC technologist is responsible for ensuring that QC tasks
are done properly by standardizing test methodology, reviewing all data,
overseeing repeat testing before calling the medical physicist or service
personnel, and conferring with the radiologist and medical physicist.
Each facility must have the services of a medical physicist to survey
mammography equipment and oversee the equipment-related quality
assurance practices of the facility.
PHYSICIST’S SECTION
The medical physicist is required to conduct an MEE of new equipment
MEDICAL
and after major repairs. This survey must be done and all relevant tests
must be passed prior to use of digital mammography equipment on
patients.
Note: During the MEE, the medical physicist should complete the
Technologist’s ACR Technique and Procedure Summaries form (located in
the Technologist Section) to help the QC technologist use the appropriate
techniques during routine QC. This form should be reviewed and updated as
necessary during annual surveys.
Note: If there is need for corrective action, the medical physicist should
instruct the facility to provide a copy of its full report for the equipment
service engineer.
PHYSICIST’S SECTION
allows the medical physicist 30 days from the date of the survey to send
a report to the facility, a 30-day delay allows the facility no time to take
corrective actions. To help facilities comply with MQSA regulations, the
MEDICAL
medical physicist should immediately communicate any failures both
verbally and in writing.
Communication of test results and recommendations of corrective actions
are areas that can be improved in the practices of most medical physicists.
Corrective actions should not be limited to the repair of x-ray equipment
by qualified service personnel but should include recommendations that
will improve image quality, including recommendations concerning
detectors, technique factors, viewing conditions in the reading room, and
technologist QC. The medical physicist must, at a minimum, annually
review the results of technologist QC tests and make recommendations
regarding these tests, if needed. Furthermore, the medical physicist must
participate in annual reviews of the mammography QC program as a
whole to make sure that the program is meeting its objectives.
Mammography team members are strongly encouraged to review other
sections of the ACR Digital Mammography QC Manual that are not
directed towards them. For example, the radiologist should be familiar
with Technologist’s Test Facility QC Review and the Technologist’s
Optional Tests System QC for Radiologist and Radiologist Image
Quality Feedback. The radiologic technologist should review the Medical
Important: Before a facility may start using the procedures in the ACR Digital
Mammography QC Manual for the first time on a unit, the medical physicist
must first conduct an annual survey of the digital mammography unit
and display devices using the manual and the ACR Digital Mammography
Phantom.
The technologist and medical physicist will use the same forms they use
for most of the digital mammography tests to record the data and results
of the DBT tests. The previous digital mammography forms have been
revised to allow for this. New forms have been added for DBT-unique
tests (e.g., DBT Volume Coverage).
If problems are occurring or if equipment is unstable, it may be necessary
to carry out some or all tests more frequently to identify problems before
they affect clinical image quality or patient safety. If the QC program
is just being initiated, it may be valuable to carry out QC tests more
frequently for the first few months. This will provide the QC technologist
with more experience in a shorter period of time and also will provide
better baseline data regarding the reliability of imaging equipment. The
necessity of performing tests designated as “Optional” or “If applicable” is
left to discretion of the Quality Assurance Committee, especially the lead
interpreting radiologist, QC technologist, and medical physicist team,
who are most familiar with the facility’s equipment and the quality needs
of the mammography practice.
In addition to performing the mammography QC tests at the minimum
frequencies indicated, tests also should be carried out for new equipment,
both when problems are suspected and after any service or preventive
maintenance. For example, the compression test should be carried out
both when a new x-ray system is installed and after any service adjustment
of compression force.
If any test fails, it is critical that the setup and techniques employed in the
test be checked and the test repeated to verify performance before initiating
corrective action. Upon confirmation of test failure, the MQSA Final Rule
requires that the source of the problem be identified and corrective action
PHYSICIST’S SECTION
be taken. In some cases, if test results fall outside of action limits, MQSA
requires that the source of the problem be identified and corrective action
MEDICAL
taken before any further examinations are performed or any films are
processed using the component of the mammography system that failed
the test. Other test failures must be corrected within 30 days of the test
date (Table 1).
Important: Corrective action for any test performed for MEEs must be made
before clinical use.
For detailed guidance on the FDA’s requirements for record retention, see
Quality Assurance Records and Retention of Personnel Records in the
FDA’s MQSA Policy Guidance Help System. All documentation must be
made available to MQSA inspectors during the annual inspection and the
facility’s accreditation body upon application and request.
PHYSICIST’S SECTION
900.12(d) Quality assurance—general. (2) Quality assurance records. The lead
interpreting physician, quality control technologist, and medical physicist shall
MEDICAL
ensure that records concerning mammography technique and procedures,
quality control (including monitoring data, problems detected by analysis of
that data, corrective actions, and the effectiveness of the corrective actions),
safety, protection and employee qualifications to meet assigned quality
assurance tasks are properly maintained and updated. These quality control
records shall be kept for each test specified in paragraphs (e) and (f) of this
section until the next annual inspection has been completed and FDA has
determined that the facility is in compliance with the quality assurance
requirements or until the test has been performed two additional times at the
required frequency, whichever is longer.
Table 2 provides a complete list of all the digital mammography tests and
summarizes those tests that must be performed (as applicable) for the
modes that the facility uses clinically on its 2D and DBT systems.
Table 2. Required Tests for Imaging Modes Used on 2D and DBT Systems
Imaging Modes to Test
System Used for Both 2D and System Used for DBT
DBT Acquisition Acquisition Only
2D w/Add-On
Test 2D DBT Device DBT DBT
Technologist Tests
1. ACR DM Phantom Image Quality * & 2D*
2. Computed Radiography Cassette *
Erasure (if applicable)
3. Compression Thickness Indicator * * *
4. Visual Checklist *
5. Acquisition Workstation Monitor QC * *
6. Radiologist Workstation Monitor QC * *
7. Film Printer QC (if applicable) * *
8. Viewbox Cleanliness (if applicable) * *
9. Facility QC Review *
10. Compression Force * * *
11. Manufacturer Calibrations (if applicable) *
Medical Physicist Tests
1. Mammography Equipment Evaluation (MEE) * *
2. ACR DM Phantom Image Quality * & 2D*
3. DBT Z Resolution
4. Spatial Resolution *
5. DBT Volume Coverage
6. Automatic Exposure Control System Performance *
7. Average Glandular Dose *
8. Unit Checklist *
9. Computed Radiography (if applicable) *
10. Acquisition Workstation Monitor QC * *
11. Radiologist Workstation Monitor QC * *
12. Film Printer QC (if applicable) * *
13. Evaluation of Site’s Technologist QC Program *
14. Evaluation of Display Device Technologist QC * *
Program
15. Manufacturer Calibrations (if applicable) *
16. Collimation Assessment * *
MEE or Troubleshooting - Beam Quality *TF
*TF
(Half-Value Layer [HVL]) Assessment
MEE or Troubleshooting - kVp Accuracy and *TF *TF
Reproducibility
*Follow the procedures and frequency outlined for 2D QC
TF
HVL and kVp tests must include kVp, target, and filter combinations used for DBT
PHYSICIST’S SECTION
medical physicist for the combination of testing that must be performed
for MEEs and annual surveys.
MEDICAL
For purposes of these examples, “display devices” refers to acquisition
workstations (AWs), radiologist workstations (RWs), or film printers (if
applicable). The solid and dashed arrows in the figures indicate that the
ACR DM Phantom image must be sent along this pathway and evaluated
(see the ACR DM Phantom Image Quality test) on the designated display
device. The gray shaded box indicates that all applicable testing from this
manual must be done for the devices included in the box.
MEEs and annual surveys of radiologist workstations must be conducted
on site by the medical physicist since the quality of the image displayed
on the monitor itself must be evaluated. This cannot be done remotely.
However, if the workstation is located at a great distance from the
mammography facility (e.g., another part of the country) the facility
may use the services of a medical physicist closer to the location of
the radiologist workstation. It is essential that MEE and annual survey
reports of offsite radiologist workstations be available at the facility where
the mammography unit is located in order to satisfy inspection and
accreditation requirements.
MEEs and annual surveys of film printers (if applicable) may be
conducted remotely by the medical physicist since the quality of the
Figure 2. MEE – All New Digital Mammography Units and Display Devices.
Figure 3. MEE – New Digital Mammography Units (with Existing Display Devices).
PHYSICIST’S SECTION
may be new, previously owned, or relocated from another facility under
the same ownership. The phantom images used for evaluation should
have been acquired from any of the facility’s digital mammography units
MEDICAL
within the past month.
The following combination of testing must be performed:
Figure 4. MEE – New Display Devices (with Existing Digital Mammography Units).
4. Annual Surveys
Medical physicists must conduct annual surveys on all digital mammography
units, acquisition workstations, radiologist workstations, and film printers
(if applicable). At some facilities, it is possible for the medical physicist to
do this at one time. However, many medical physicists choose to separate
the annual survey for the digital mammography units from that of the
display devices because the workstations or printers are located at different
physical locations from the digital mammography units or equipment repair
has shifted the timing of the survey(s). For this reason, the annual survey
scenarios in Figure 5 are presented separately for the digital mammography
units and the display devices. However, it is important to note that the
annual survey of the digital mammography units and the display devices
may be done at one time.
If the annual surveys for the display devices are done separately from
the digital mammography unit, a phantom image stored on the facility’s
picture archiving and communication system (PACS) may be used.
This image should have been acquired from any of the facility’s digital
mammography units within the past month.
The following combination of testing must be performed:
PHYSICIST’S SECTION
MEDICAL
PHYSICIST’S SECTION
equipment meets standards) can be performed by other qualified
personnel (e.g., radiologic technologist or service representative
with appropriate training/experience) without involving the medical
MEDICAL
physicist. However, the facility can consult its medical physicist in
these situations if it wishes.
Table 3 lists typical component adjustments, changes, or repairs that may
occur in a digital mammography system along with medical physicist
involvement (i.e., on-site, oversight, or optional) recommended by the
FDA and the ACR. Be sure to check the FDA’s MQSA Policy Guidance
Help System for current guidance on testing.
Note: For the kVp Accuracy and Reproducibility and Beam Quality (Half-
Value Layer) Assessment tests, see the procedures and forms in the MEE or
Troubleshooting Test section.
PHYSICIST’S SECTION
TEST EQUIPMENT General [Section 900.12(b)]
MEDICAL
Mammography Equipment Evaluation - MQSA Requirements checklist
(required)
APPLICABLE MQSA
REQUIREMENTS 900.12(b) Equipment.
(3) Motion of tube-image receptor assembly.
(i) The assembly shall be capable of being fixed in any position where it is
designed to operate. Once fixed in any such position, it shall not undergo
unintended motion.
(ii) The mechanism ensuring compliance with paragraph (b)(3)(i) of this section
shall not fail in the event of power interruption.
(4) Image receptor sizes.
(iii) Systems used for magnification procedures shall be capable of operation
with the grid removed from between the source and image receptor.
(5) Light fields. For any mammography system with a light beam that passes
through the X-ray beam-limiting device, the light shall provide an average
illumination of not less than 160 lux (15 foot-candles) at 100 cm or the
maximum source-image receptor distance (SID), whichever is less.
(6) Magnification.
(i) Systems used to perform non-interventional problem solving procedures shall
have radiographic magnification capability available for use by the operator.
(ii) Systems used for magnification procedures shall provide, at a minimum, at
least one magnification value within the range of 1.4 to 2.0.
(7) Focal spot selection.
(i) When more than one focal spot is provided, the system shall indicate, prior to
exposure, which focal spot is selected.
(ii) When more than one target material is provided, the system shall indicate,
prior to exposure, the preselected target material.
(iii) When the target material and/or focal spot is selected by a system algorithm
that is based on the exposure or on a test exposure, the system shall display,
PHYSICIST’S SECTION
after the exposure, the target material and/or focal spot actually used during
the exposure.
MEDICAL
(8) Compression. All mammography systems shall incorporate a compression
device.
(i) Application of compression. Effective October 28, 2002, each system shall
provide:
(A) An initial power-driven compression activated by hands-free controls
operable from both sides of the patient; and
(B) Fine adjustment compression controls operable from both sides of the
patient.
(ii) Compression paddle.
(A) Systems shall be equipped with different sized compression paddles
that match the sizes of all full-field image receptors provided for the system.
Compression paddles for special purposes, including those smaller than the
full size of the image receptor (for “spot compression”) may be provided. Such
compression paddles for special purposes are not subject to the requirements
of paragraphs (b)(8)(ii)(D) and (b)(8)(ii)(E) of this section.
(B) Except as provided in paragraph (b)(8)(ii)(C) of this section, the compression
paddle shall be flat and parallel to the breast support table and shall not
deflect from parallel by more than 1.0 cm at any point on the surface of the
compression paddle when compression is applied.
(C) Equipment intended by the manufacturer’s design to not be flat and parallel
to the breast support table during compression shall meet the manufacturer’s
design specifications and maintenance requirements.
(D) The chest wall edge of the compression paddle shall be straight and parallel
to the edge of the image receptor.
(E) The chest wall edge may be bent upward to allow for patient comfort but
shall not appear on the image.
(9) Technique factor selection and display.
(i) Manual selection of milliampere seconds (mAs) or at least one of its
component parts (milliampere (mA) and/or time) shall be available.
(ii) The technique factors (peak tube potential in kilovolt (kV) and either tube
current in mA and exposure time in seconds or the product of tube current and
exposure time in mAs) to be used during an exposure shall be indicated before
the exposure begins, except when automatic exposure controls (AEC) are used,
in which case the technique factors that are set prior to the exposure shall be
indicated.
(iii) Following AEC mode use, the system shall indicate the actual kilovoltage
peak (kVp) and mAs used during the exposure. The mAs may be displayed as
mA and time.
(14) Lighting. The facility shall make special lights for film illumination, i.e., hot-
lights, capable of producing light levels greater than that provided by the view
box, available to the interpreting physicians.
(15) Film masking devices. Facilities shall ensure that film masking devices that
can limit the illuminated area to a region equal to or smaller than the exposed
portion of the film are available to all interpreting physicians interpreting for
the facility.
PHYSICIST’S SECTION
900.2 Definitions. (uu) Standard breast means a 4.2 centimeter (cm) thick
MEDICAL
compressed breast consisting of 50 percent glandular and 50 percent adipose
tissue.
the type of paddle chosen (e.g., flex or fixed) is the one used for the
majority of clinical imaging.
5.
Manually compress the paddle to approximately 5 decanewtons
(daN) or 12 pounds of compression force. It is important to use
the same compression force each time for this test. Note that at this
compression force, the compressed breast thickness indicator may
not read 4.2 cm.
6. At the acquisition workstation, select the imaging mode and technique
that would be used for a clinical screening exam acquisition of a 4.2
cm thick compressed breast consisting of 50% glandular and 50%
adipose tissue. If a combination exposure mode (i.e. 2D plus DBT)
is most commonly used clinically, use the combination mode for this
test, record the data from the 2D and DBT acquisitions, and use
those images for analysis. If a 2D-only mode is most commonly used
clinically for screening, use the 2D-only mode for this procedure
step. (If the system uses selectable AEC sensor positions, be sure to
use the same position each time the phantom is acquired.)
Note: This technique must match the technique used by the technologist
for his or her weekly QC. This technique should be listed in the ACR
Technique and Procedure Summaries form that the technologist maintains.
PHYSICIST’S SECTION
e. Compression force
MEDICAL
f. AEC cell position (if applicable)
g. Target/filter (if applicable)
h. kVp (if applicable)
i. Density setting (if applicable)
used clinically for screening, use the DBT-only mode for this
procedure step. (If the system uses selectable AEC sensor positions,
be sure to use the same position each time the phantom is acquired.)
Note: This technique must match the technique used by the technologist
for his or her weekly QC. This technique should be listed in the ACR Technique
PHYSICIST’S SECTION
and Procedure Summaries form that the technologist maintains.
MEDICAL
7. Record or verify the following demographic information at the top of
the form:
a. Facility
b. MAP ID number
c. ACR DM phantom manufacturer and serial number
d. Room ID
e. X-ray unit manufacturer and model
f. Unit SID in cm
8.
Record or verify and use the following technique parameters in the
“Phantom Setup” box at the top of the form. These parameters should
match the techniques found on the Technologist’s ACR Technique
and Procedure Summaries form, which should be located in the
phantom section of the technologist’s QC notebook.
a. AEC mode
b. Paddle and IR size
9.
Acquire an image of the phantom.
10. Record the following parameters that appear after the exposure on
the form:
a. Target/filter material (e.g., Rh/Rh, W/Ag, etc.)
b. kVp (e.g., 32)
c. mAs (e.g., 78.5)
d. The unit-indicated AGD in mGy for the phantom exposure, if
available (This will be compared to the measured AGD in the
Average Glandular Dose test.)
11.
If the system uses an add-on device for DBT, repeat steps 3-11 in 2D
mode with the DBT fixture in place.
For DBT images, scroll to the one slice in which the test objects are
3.
best visualized. If “slices” are not available, then proceed to use the
slab in which the objects are best visualized.
4. Adjust the window width (WW) and window level (WL) settings to
optimize visualization of test objects (the test objects will be scored
in the next section). It is important not to use unreasonably narrow
WWs that may enhance the appearance of artifacts.
5. Record the approximate WW and WL settings (and best visualized
slice or slab # for DBT scoring) on the Technologist’s ACR Technique
and Procedure Summaries form.
6.
Using approximately the same WW and WL settings used to
evaluate the test objects, examine the entire phantom for both broad
area artifacts and detailed artifacts. (See Figure 9 for examples of a
properly windowed ACR digital phantom images without artifacts.)
a. Broad area artifacts (e.g., non-uniformities, blotches, and streaks)
usually are best seen while observing the phantom image as a
whole and not in pieces (i.e., not magnified or at full resolution).
b. Detailed artifacts (e.g., black or white pixels, clusters of pixels,
lines, or dust particles) usually are best seen while observing the
phantom image at full spatial resolution, where one pixel on the
display matches one pixel in the image, or with magnification,
using a zoom factor greater than 1.0.
7. See the Artifact Evaluation Guide in Appendix III for examples of
“good” or “artifact free” images and some common digital artifacts.
8. For phantom images from each mode and target/filter combination,
PHYSICIST’S SECTION
record the absence or presence of artifacts on the form as a Pass or
Fail (P or F).
MEDICAL
9. To score the phantom image, adjust the WW and WL settings to
optimize visualization of test objects. You may need to slightly adjust
the WW and WL to obtain optimum visualization of each test object.
The zoom or magnification tool should also be used. Use a WW
and WL that permit the best visualization of fibers, speck groups,
and masses. Using the scoring methods described below, score the
number of fibers, speck groups, and masses seen in the phantom and
record the scores on the form. (See Figure 10.)
10. Do not deduct for artifacts. (Deducting for artifacts is no longer part
of the ACR DM Phantom scoring procedure.)
11. Scoring method (see Figure 11 and Table 4):
a. Count the number of visible objects, from the largest object of a
given type (fiber, speck group, or mass) downward, until a score
of 0 or ½ is reached, then stop counting for that object type.
(This step is the same as used in the 1999 Mammography Quality
Control Manual [5].) For each test object type, the minimum
possible score is 0 objects and the maximum possible score is
6 objects.
b. Fibers
i. The fibers are manufactured to be 10 mm in length. If the
entire length of the fiber is not visible, measure it using the
display device’s electronic calipers.
ii. Count each fiber as 1 point if 8 mm or more of the fiber is
visible in the correct location and orientation.
iii. Count a fiber as ½ point if the fiber appears to be equal to or
greater than 5 mm and less than 8 mm in length and is in the
correct location and orientation.
Figure 10. DBT image of a magnified and properly windowed ACR DM Phantom.
PHYSICIST’S SECTION
MEDICAL
Figure 11. ACR DM Phantom wax insert map (test object sizes are not to scale).
iv. If a small gap in the fiber is visible, and it is less than the width
of the fiber, count the fiber as a full or half point depending on
the total visible length.
c. Speck groups
i. Count each speck group as 1 point if 4 to 6 specks are visible
in the proper locations in the group.
ii. Count a speck group as ½ point if 2 or 3 specks are visible in
the proper locations in the group.
d. Masses
i. Count each mass as 1 point if an object is visible in the correct
location and the mass appears to be generally circular against
the background (at least ¾ of the border is continuous and
generally round).
ii. Count a mass as ½ point if a mass-like object is visible in
the correct location but does not have a generally circular
appearance (greater than ½ but less than ¾ of a circle).
12. Enter the final scoring result in each category (fibers, speck groups,
and masses) on the form.
13. If the system uses a separate breast support system for DBT, repeat
steps 1-12 for the 2D mode image acquired with the DBT breast
support system in place.
14. If 2D add-on devices are used clinically, repeat the above steps to
score the resulting 2D image.
15. See the ACR DM Phantom Scoring Guide in Appendix II for
examples on scoring.
(Omit the DC offset if this does not apply for the DM unit being
tested.)
9. Record the SNR on the form.
10. Calculate the CNR as:
PHYSICIST’S SECTION
(Mean Cavity Signal – Mean Bkgd Signal)
CNR =
Std Dev of Bkgd
MEDICAL
11. Record the CNR on the form.
PHYSICIST’S SECTION
a. Artifacts are as prominent as (or more prominent than) the
visible test objects in the phantom image, or
MEDICAL
b. Artifacts obscure test objects in the phantom, or
c. Artifacts could affect clinical interpretation.
The cause of the artifact should be identified and isolated to
determine if it originates from the x-ray system, the detector,
or the monitor. If the artifact is confirmed to originate from the
detector, a recalibration or flat-fielding of the detector may be
needed. Artifacts isolated to other components of the imaging
chain should be investigated.
After the artifact is resolved, repeat the phantom artifact test. If
a clinically significant artifact persists, contact your authorized
service representative. If the clinically significant artifact
originated from the x-ray/detector system, do not image patients
until it is corrected. If the clinically significant artifact originated
from the monitor, do not use the monitor until it is corrected.
2. The fiber score must be ≥2.0.
3. The speck group score must be ≥3.0.
4. The mass score must be ≥2.0.
Note: During annual surveys, the CNR is compared to the value measured
during the most recent MEE. The deviation of the CNR from this original
value is used to monitor changes in the imaging chain over time. A deviation
more than -15% from the value measured at the time of the last MEE would
result in a failure. Corrective actions would be required prior to clinical use.
During a MEE, the value measured would be used for future comparisons.
General
1. If any of these quantities are not within action limits, the test should
be repeated, making sure the correct AEC mode has been used. If
phantom scores are below the stated minimum score, the facility
should contact its authorized service representative.
2. Record and date any comments and required corrective action in the
Technologist’s Corrective Action Log form.
TIMEFRAME FOR All failures of required items must be corrected before clinical use.
CORRECTIVE ACTION
3. DBT Z Resolution
OBJECTIVES To ensure that blurring in the z-direction is not excessive.
FREQUENCY As part of the mammography equipment evaluation (MEE) of new units,
annually, and after relevant service.
TEST EQUIPMENT • DBT image from ACR Digital Mammography (DM) Phantom Image
Quality test.
• DBT Z Resolution form.
TEST PROCEDURE 1. If annual testing, record the baseline full width at half maximum
(FWHM) value from the MEE (or the oldest annual survey if the
MEE is not available). Otherwise, perform the following steps to
create a baseline Z-resolution value.
2. Using the phantom image acquired from the ACR DM Phantom Image
Quality test (Figure 14), obtain signal data over the specks as follows:
PHYSICIST’S SECTION
MEDICAL
Figure 14. ACR DM Phantom Image.
Note: Use the reconstructed DBT slices (for presentation) for Z-resolution
measurements.
a. Scroll to the slice (slab) where the center speck in the largest
speck group is most in focus and brightest. This is slice 0.
b. Zoom in on the largest speck group on the ACR DM Phantom
image (Figure 15).
c. Place a region of interest (ROI) over the center speck (Figure 16).
d. Record the maximum signal value on the form at the Slice
Location 0.
3.
Repeat step 2 for the other 5 specks in the group.
4.
Determine mean background signal as follows:
a. Place an ROI over the background adjacent to the center speck
(Figure 17).
Figure 16. ROI placement over center speck in largest speck group.
b. Record the mean background signal value on the form at the Slice
Location 0.
5. Repeat steps 2-4 for the following slices and record the signal values
on the form at the appropriate slice locations:
a. 2 slices below slice 0 (−2)
Figure 18. Illustration of the blurring of the largest speck through the five DBT slices.
PHYSICIST’S SECTION
MEDICAL
Figure 19. Images of the 5 different slice locations for Z-resolution signal
measurements.
Figure 20. Images of the 5 different slice locations for Z-resolution measurements
with ROI placement locations.
DATA ANALYSIS AND 1. For each DBT slice (−2, −1, 0, +1, +2), subtract the Mean Background
INTERPRETATION Signal from the Average Maximum Speck Signal, yielding
Z-Resolution Signal Difference (Z-Res Diff):
Z-Res Diff = Average Max Speck Signal – Mean Background Signal
2. For DBT slices −2, −1, +1, and +2, calculate the Δ Z-Res Diff, the
fraction of Z-Res Diff of that slice relative to DBT slice 0:
Δ Z-Res Diffi = Z-Res Diffi / Z-Res Diff0
3. If Δ Z-Res Diff2 is greater than or equal to 0.5, calculate Δ Z-Res
Diff3.
4. If Δ Z-Res Diff-2 is greater than or equal to 0.5, calculate Δ Z-Res
Diff-3.
5.
From the DBT Z Resolution form, determine the FWHM.
Note: FWHM is the distance between points on the curve at which the
function reaches half its maximum value. In this test, the curve represents
ratios of signal in each slice compared to Slice 0, so the maximum is 1.0
and half maximum is 0.5. The x-axis represents distance in the depth (z)
direction. Therefore, FWHM is in units of mm and is the distance that the
speck is spread in the z-direction. For a more complete description of the
graphical plot measurement, see the American College of Radiology Digital
Mammography QC Manual: Frequently Asked Questions.
PRECAUTIONS AND It may be helpful to review the DBT Z-Resolution form, which illustrates
CAVEATS the Δ Z-Res Diffi calculation.
PERFORMANCE MEEs
CRITERIA AND
CORRECTIVE ACTIONS The initial (MEE) Z-Resolution measurement becomes the baseline value
for future Z-Resolution measurements. There is no action limit for the
initial Z-Resolution measurement.
Annual Surveys
The FWHM value of the current image set must be within ±30% of the
baseline value. If the Performance Criteria are not met, a qualified service
engineer must be contacted to correct the problem.
TIMEFRAME FOR Failures must be corrected within 30 days.
CORRECTIVE ACTION
4. Spatial Resolution
OBJECTIVES To measure the limiting spatial resolution as an indicator of detector
performance.
FREQUENCY As part of the mammography equipment evaluation of new units,
annually, and after relevant service.
TEST EQUIPMENT • ACR Digital Mammography (DM) Phantom.
• Line pair (lp) pattern with frequencies up to 10 lp/mm.
• Spatial Resolution form.
TEST PROCEDURE – 2D 1. Create a test patient.
AND DBT
2. Turn off all image processing to acquire a “for processing” (e.g.,
“raw”) image.
3. Install a compression paddle.
4. Place the ACR DM Phantom on the breast support with the wax
insert facing up and away from the chest-wall edge (rotated 180°
from the normal orientation of the phantom).
5. Place the line pair pattern on the phantom at a 45° angle (see Figure 21).
6. Lightly compress the line pair pattern to ensure that it remains secure
during the exposure.
7. Make one exposure using a manual technique as close to the ACR
DM Phantom technique as possible.
8. Repeat steps 3-7 for any other targets used clinically.
PHYSICIST’S SECTION
9. Install the magnification stand and paddle and enter the unit’s most
frequently used clinical magnification mode. Repeat steps 3-7 for
MEDICAL
clinically used targets.
10. Repeat steps 3-8 in the DBT imaging mode.
11. If the system uses an add-on device for DBT, repeat steps 3-8 in 2D
mode with the DBT fixture in place.
A B
C
Figure 21. A. Bar pattern placement for the Spatial Resolution test on a 2D system.
The bar pattern is at a 45° angle to the chest-wall edge. B. Magnification setup. C. DBT
setup.
DATA ANALYSIS AND 1. At the acquisition station, view each image using full resolution and
INTERPRETATION the greatest zoom available. (See Figures 22 and 23.)
2. Record the highest frequency for which at least half the length of the
lines can be continuously resolved in each image. (See Figure 23.)
3. Ensure that the polarity of the lines does not reverse. If reversal
occurs, the limiting resolution has been surpassed.
Figure 22. Image of bar pattern properly visualized for the Spatial Resolution Test.
Figure 23. Bar pattern magnified for analysis. In this pattern, 4.0 lp/mm are
distinguishable.
4. If the system uses an add-on device for DBT, repeat steps 1-3 for the
2D image acquired with the DBT fixture in place.
PRECAUTIONS AND It is recognized that limiting spatial resolution is an imperfect substitute
CAVEATS for a detailed determination of modulation transfer function. However,
limiting spatial resolution is more easily measured in the field and serves
as an acceptable analog for purposes of detector performance consistency.
PHYSICIST’S SECTION
PERFORMANCE 1. Spatial resolution of the 2D image(s) must be ≥4.0 lp/mm.
CRITERIA AND
MEDICAL
CORRECTIVE ACTIONS 2. Spatial resolution of the 2D magnification mode image(s) must be
≥6.0 lp/mm.
3. Spatial resolution of the DBT image(s) must be ≥2.0 lp/mm.
4. If limiting spatial resolution does not meet these criteria, service
must be scheduled.
TIMEFRAME FOR Failures must be corrected within 30 days.
CORRECTIVE ACTION
PHYSICIST’S SECTION
MEDICAL
B
Figure 24. A. ACR DM Phantom with 0.1 mm aluminum sheets positioned for image
acquisition. B. Zoomed photograph of DBT Volume Coverage setup.
DATA ANALYSIS AND At the acquisition workstation, scroll through the image set using
INTERPRETATION the thinnest slices available. Determine if both the top and bottom
aluminum sheets are well defined and in focus in their respective planes.
(See Figure 25.)
A B
Figure 25. A. Image of ACR DM Phantom with aluminum sheets in focus at bottom
of phantom. B. Image of ACR DM Phantom with aluminum sheets in focus at top of
phantom.
PRECAUTIONS AND If the system fails the test, verify that the aluminum sheets are properly
CAVEATS positioned and repeat the test.
PERFORMANCE Each aluminum sheet must be well defined within one slice, or the system
CRITERIA AND is not imaging the entire breast volume.
CORRECTIVE ACTIONS
PHYSICIST’S SECTION
compensation step) to 0.
6. If applicable, set the AEC sensor to the center of the phantom.
MEDICAL
7. Select the large focal spot.
8. Acquire an image using the AEC mode used clinically.
9. Make an exposure.
A B
Figure 26. Position of attenuator for Automatic Exposure Control System Performance
test in contact mode. A. 2D. B. DBT.
Figure 27. Position of attenuator for Automatic Exposure Control System Performance
test in magnification mode.
DATA ANALYSIS AND 1. If the manufacturer provides a DC offset, record this value on the
INTERPRETATION form.
2. Record the signal-to-noise ratio (SNR) results from the last MEE on
the form. (This does not apply to MEEs.)
3.
Calculate the lower limit and the upper limit for each mode/
attenuator SNR as ±15% of the last MEE’s SNR, and record.
4. Use only “for processing” (e.g., “raw”) images for analysis.
5. For DBT, find the center slice or slab in the image set and use the
thinnest available slice.
6. On the radiologist workstation, use a circular or rectangular region
of interest (ROI; approximately 3 cm from chest wall and centered
left to right) to measure the mean signal value (or mean analog to
digital units) in the middle of the phantom. (See Figure 28.)
7. Record the mean signal value as Mean Bkgd Signal on the form;
record the standard deviation as Std Dev of Bkgd.
A B
Figure 28. Position of ROI to measure pixel value (approximately 3 cm from chest wall
and centered left to right). A. 2D. B. DBT.
Note: If the acquisition workstation does not have ROI capability, the
medical physicist should use one of the following alternatives to complete
the test:
PHYSICIST’S SECTION
• Make the ROI measurement on the radiologist workstation.
• Using image analysis software, make the ROI measurement on an external
MEDICAL
computer system.
• If neither of the above alternatives are available, use the manufacturer’s
AEC evaluation procedure, equipment, and form.
It is recognized that SNR is not strictly defined for DBT images. However,
these calculated values should remain consistent year-to-year if the AEC
is performing consistently.
Note that if components in the imaging chain are replaced (i.e., detector,
x-ray tube, etc.) SNR may not meet the annual survey performance
criteria of ±15% of the last MEE’s SNR. If this is the case, the reason
should be noted, a new baseline should be established, and the test should
be passed.
Annual Surveys
The SNR must be within ±15% of the last MEE’s SNR for each thickness
and mode tested. (This component of the test does not apply to MEEs.)
General
1. If any of these quantities are not within action limits, the test should
be repeated. If the results remain below the performance criteria, the
facility should contact its authorized service representative.
2. Record and date any comments and required corrective action in the
Technologist’s Corrective Action Log form.
TIMEFRAME FOR Failures must be corrected within 30 days; for MEEs, before clinical use.
CORRECTIVE ACTION
Note: This procedure uses the technique factors that result from the ACR
Digital Mammography (DM) Phantom Image Quality test. This technique
must be the same as that used clinically for a 4.2-cm thick compressed
breast consisting of 50% glandular and 50% adipose tissue, as defined
by the FDA. Exposure measurements made at a corresponding manual
PHYSICIST’S SECTION
technique in 2D mode (or fixed mode) can be used to calculate average
glandular dose for DBT images.
MEDICAL
1. Place a lead sheet or other protective device on the image receptor.
This is intended to protect the detector from repeated exposures.
2. Position the center of the dosimeter at a height of 4.2 cm above the
breast support and just under the paddle. Center the dosimeter
approximately 4 cm in from the chest-wall edge of the image receptor.
Make sure that the entire dosimeter is exposed (see Figure 29).
3. Secure the dosimeter in position, and do not change its position
during the following measurements.
A B
Figure 29. Exposure setup. Compression device positioned so that it is just in contact
with the dosimeter. A. Side view diagram. B. Top view diagram. C. Photo.
5. Set the target material, filtration, and kVp at the values at which the
ACR DM Phantom image was acquired in the 2D mode, and record
the settings on the form. Also, record the HVL (previously measured
during the MEE) for those same parameters on the form. Finally,
record on the form the automatic exposure control (AEC) mode and
mAs at which the ACR DM Phantom image was acquired in the 2D
mode.
6. Manually set mAs as close as possible to that obtained under the
AEC exposure of the ACR DM Phantom and record the value on the
form.
7. Make a manual exposure and record the measured exposure on the
form.
8.
Repeat step 7 until three exposures have been recorded.
9. If available, record the unit-indicated average glandular dose from
the ACR DM Phantom Image Quality test form.
10. Set the target material, filtration, and kVp at the values at which the
ACR DM Phantom image was acquired in the DBT mode, and record
the settings on the form. Also, record the HVL (previously measured
during the MEE) for those same parameters on the form. Finally,
record on the form the AEC mode and mAs at which the ACR DM
Phantom image was acquired in the DBT mode. Repeat steps 6-9 for
the DBT mode.
Note: If the phantom image was acquired in a combination mode, only use
the technical factors associated with the DBT portion of the exposure for the
DBT average glandular dose assessment.
Note: Use the acrylic g-factor * c-factor * 8.76 mGy/R values when
determining the average glandular dose from the ACR DM Phantom
techniques for a 4.2-cm thick compressed breast consisting of 50% glandular
and 50% adipose tissue. Both acrylic and BR-12 values are provided for
breast thicknesses of 2 to 8 cm if the medical physicist wishes to determine
average glandular doses for a broader range of breast thicknesses.
4.
Compute the average glandular dose using the following equation:
AGD = Kgcs
AGD = Average Glandular Dose (mGy)
K = Entrance Exposure (mR)
PHYSICIST’S SECTION
g = g-factor for breast simulated with acrylic or BR-12
MEDICAL
c = c-factor for breasts simulated with acrylic or BR-12
s = s-factor for clinically used spectra
a. K is the exposure in the absence of backscatter at the entrance
surface of the breast. (The amount of backscatter in the described
setup is small enough to be ignored while maintaining adequate
accuracy for this test.)
b. The factor g corresponds to a glandularity of 50% and is derived
from values calculated from Dance [6].
c. The factor c corrects for any difference in breast composition
from 50% glandularity. (See Table 5 and Table 6; additional
phantom thicknesses are provided if the medical physicist would
like to determine dose for other breast thicknesses.)
d. The factor s corrects for differences due to the choice of x-ray
spectrum (see Table 7) [6, 7, 8, 9].
PRECAUTIONS AND Current and accurate calibration of dosimeters is essential and required
CAVEATS under the FDA MQSA Final Rule [1]. See the FDA’s MQSA Policy Guidance
Help System [2] frequently asked questions on air kerma calibration. It is
well known that the energy response for ionization chamber-based air
kerma measuring instruments is typically flat in the 20-40 kVp range.
For solid-state air kerma measuring instruments, however, the energy
response is not flat, and because of this, the air kerma readings from these
instruments may need to be adjusted by an appropriate correction factor.
The correction may already be handled internally by the instrument, or
you may need to contact the instrument manufacturer for the correction
factor [10].
The medical physicist should ensure that the HVL has been determined
for the target/filter and kVp combination used for DBT imaging during
the MEE.
If the facility has changed its target/filter and kVp combination for an
average patient (and the corresponding HVL was not determined during
the MEE), or if the medical physicist suspects that the HVL may have
significantly changed since the MEE, the medical physicist should re-
evaluate the HVL for use in the current Average Glandular Dose test.
The previously published 1999 ACR Mammography Quality Control
Manual [5] used a method adapted from Barnes and Wu to calculate
average glandular dose. This method was limited to four target/filter
combinations, a limited range of kVp and HVL settings, one breast
thickness (4.2 cm), and one tissue glandularity value (50% adipose-50%
glandular). This manual is using a different method published by
Dance [6], which utilizes a single formula (D = Kgcs), where D is the
average glandular dose, K is the entrance exposure at the upper surface
of the breast without backscatter, g is the entrance exposure to mean
glandular dose conversion factor, c corrects for differences in breast
composition from 50% glandularity, and s corrects for differences in x-ray
spectra. This method accommodates most target/filter combinations
used in modern systems, nearly all clinically relevant breast thicknesses,
a wider range of beam qualities, and a wider range of breast glandularity.
There is a minor difference between the two methodologies: an average
of 1.0% difference in AGD values using the Barnes and Wu method [5]
PHYSICIST’S SECTION
compared to the Dance method [6] for a cross-section of target/filter
combinations and kVps used in the 1999 manual.
MEDICAL
PERFORMANCE CRITERIA 1. The average glandular dose for a single cranio-caudal view of the
AND CORRECTIVE ACTIONS ACR DM Phantom in either 2D or DBT mode must not exceed 3.0
mGy. (Dose from a mode that combines a 2D view with a DBT view,
such as “Combo” mode, is not subject to the 3.0 mGy performance
criteria. Each view within the mode should be compared separately
against the performance criteria.)
If the calculated value exceeds these levels, action must be taken to
2.
evaluate and eliminate the cause of excessive dose.
3. If the unit-indicated average glandular dose is available, it should be
within ±25% of the calculated average glandular dose.
TIMEFRAME FOR • Doses exceeding 3 mGy must be corrected before clinical use.
CORRECTIVE ACTION
• ailures of the unit-indicated average glandular dose must be
F
corrected within 30 days; for MEEs, before clinical use.
APPLICABLE MQSA
REQUIREMENTS 900.12(e)(6) Quality control tests—other modalities.
For systems with image receptor modalities other than screen-film, the quality
assurance program shall be substantially the same as the quality assurance
program recommended by the image receptor manufacturer, except that the
maximum allowable dose shall not exceed the maximum allowable dose for
screen-film systems in paragraph (e)(5)(vi) of this section.
900.12(e)(5)(vi) Dosimetry.
The average glandular dose delivered during a single cranio-caudal view
of an FDA-accepted phantom simulating a standard breast shall not exceed
3.0 milligray (mGy) (0.3 rad) per exposure. The dose shall be determined with
technique factors and conditions used clinically for a standard breast.
8. Unit Checklist
OBJECTIVES To ensure that all locks, detents, angulation indicators, and mechanical
support devices for the x-ray tube and breast support assembly are
operating properly and that the DICOM image file headers are correctly
populated.
FREQUENCY As part of the mammography equipment evaluation (MEE) of new units,
annually, and after relevant service.
TEST EQUIPMENT Unit Checklist form.
TEST PROCEDURE 1.
Verify that the freestanding dedicated mammography unit is
mechanically stable under normal operating conditions (*critical test).
2. Verify that all moving parts move smoothly, without undue friction;
that cushions or bumpers limit the range of available motions; and
that no obstructions hinder the full range of motions within these
limits.
3. Set and test each lock and detent independently to ensure that
mechanical motion is prevented when the lock or detent is set
(*critical test).
4. Verify that the detector or image receptor holder assembly is free
from wobble or vibration during normal operation (*critical test).
5. If cassette-based, verify that the image receptor slides smoothly into
the proper position in the image receptor holder assembly.
6. If cassette-based, verify that the cassette is held securely for any
orientation of the image receptor holder assembly (*critical test).
PHYSICIST’S SECTION
7. Verify that in normal operation the patient and operator are not
exposed to sharp or rough edges or other hazards including electrical
MEDICAL
hazards (*critical test).
8. Verify that the compression paddles are all intact with no cracks or
sharp edges (*critical test).
9. Verify that the mammography area is clean and free from significant
dust and debris that may cause artifacts.
10. Verify that the operator is protected by adequate radiation shielding
during exposure (*critical test).
11. Verify that all indicators work properly.
12. Verify that automatic decompression can be overridden to maintain
compression (for procedures such as needle localizations) and
its status displayed continuously (if automatic decompression is
available) (*critical test).
13. Verify that compression can be manually released in the event
of a power or automatic release failure by turning power off to
the equipment. This can be verified by placing a phantom under
compression and using manual controls to release the compression
(*critical test).
PERFORMANCE CRITERIA Critical items that are hazardous, inoperative, or operate improperly must
AND CORRECTIVE ACTIONS be repaired by appropriate service personnel or replaced.
TIMEFRAME FOR Failures of critical tests (*) must be corrected before clinical use;
CORRECTIVE ACTION less critical tests must be corrected within 30 days; for MEEs, before
clinical use.
APPLICABLE MQSA
REQUIREMENTS 900.12(e)(5)(xi) Decompression.
If the system is equipped with a provision for automatic decompression after
completion of an exposure or interruption of power to the system, the system
shall be tested to confirm that it provides:
(A) An override capability to allow maintenance of compression;
(B) A continuous display of the override status; and
(C) A manual emergency compression release that can be activated in the event
of power or automatic release failure.
PHYSICIST’S SECTION
5. Record the mAs on the form.
MEDICAL
6. Process the acquired CR image using no image processing if possible.
7.
At a workstation, place a region of interest having an area of
approximately 5 cm2 in the center of each image. Record the
measured mean signal values and standard deviation on the form.
Calculate the signal-to-noise ratio (SNR) for each image using the
procedure described in the ACR Digital Mammography Phantom
Image Quality test.
8. Repeat steps 1-7 for all plates at the facility (i.e., large and small).
2. Place a pair of thin steel rulers or another thin metal object on top of
the cassette in a T shape. (See Figure 30.)
3. Make a manual exposure at a technique close to 25 kVp and 4 mAs.
4. Process the cassette.
Note: If the facility replaces all of its PSP plates with new ones, the medical
physicist must perform an MEE consisting of applicable tests.
PHYSICIST’S SECTION
is likely in the x-ray unit or in the CR reader. The source should be
isolated and appropriate service called.
MEDICAL
CR Reader Scanner Performance
It is recommended that the edges of the steel ruler in both directions be
crisp and linear. There should be no jagged or non-linear edges visible at
the ruler’s edge.
TIMEFRAME FOR Failures of required items must be corrected before clinical use.
CORRECTIVE ACTION
A B
Figure 31. Test patterns: A. AAPM TG18-QC test pattern. B. SMPTE test pattern.
A B
Figure 32. Test patterns: A. AAPM TG18 LN8-01 test pattern. B. AAPM TG18 LN8-18 test
pattern.
PHYSICIST’S SECTION
MEDICAL
Figure 33. AAPM TG18 UNL80 test pattern.
3. Evaluate the test pattern for the following visible targets and record
pass or fail on the form:
a. Is the test pattern centered appropriately?
b. Are the 0%-5% contrast boxes visible?
c. Are the 95%-100% contrast boxes visible?
d. Are the alphanumerics sharp and legible?
e. Are the line-pair images at the center and four corners visible
and clearly distinguishable?
f. Are the grayscale ramps smooth and continuous?
PHYSICIST’S SECTION
a monitor blemish, the lead interpreting radiologist should be consulted.)
MEDICAL
1. The test pattern must be centered appropriately.
2. The 0%-5% and 95%-100% contrast boxes must be visible.
3. The alphanumerics must be sharp and legible.
4. The high-contrast line-pair patterns must be distinguishable at the
center and corners.
5. The grayscale ramps must be smooth and continuous.
PHYSICIST’S SECTION
check, or other patterns that allow for measurement of Lmin and Lmax.
(See Figure 32.)
MEDICAL
• APM TG18 UNL80 test pattern for luminance uniformity, or
A
other patterns that allow for measurement of luminance uniformity.
(See Figure 33.)
• Luminance meter.
• Radiologist Workstation Monitor QC form.
Monitor Condition
1. Visually inspect the surface of the monitor for the presence of dust,
scratches, defects, fingerprints, shiny patches (from grease or gel),
and other foreign material (e.g., pen marks, etc.).
2. Record significant findings on the form (see Performance Criteria
and Corrective Actions).
ACR DM Phantom
1.
Display a phantom image that was acquired on one of the
digital mammography units per instructions in the ACR Digital
Mammography Phantom Image Quality test.
2. Evaluate for artifacts and score the phantom test objects as in the
ACR Digital Mammography Phantom Image Quality test.
Distance Measurement
1.
Using a measurement tool within the manufacturer software,
measure the distance across the wax insert as shown in Figure 34.
2. Record results on the form.
Luminance Check
1.
Using the TG18 LN8-01 and LN8-18 test patterns (or other
appropriate phantom as indicated above), measure Lmin and Lmax with
a luminance meter in the center of each box. (See Figure 32.)
2. Record these values on the form.
3. If possible, obtain the manufacturer specifications for Lmin and Lmax
and enter them on the form.
Luminance Uniformity
1. Using the TG18 UNL80 test pattern, measure the luminance at
each of the four corners of the monitor and at the center of monitor.
(See Figure 33.)
2. Record these values on the form.
PHYSICIST’S SECTION
1. Open monitor manufacturer automated test program.
MEDICAL
2. For initial setup, review the monitor manufacturer’s frequencies,
action limits, and other test parameters to verify if appropriate for
mammography.
3. Review the results and verify that all tests have passed.
4. Record an overall pass or fail on the form.
Luminance Uniformity
1. Calculate the percent difference of the luminance values measured in
the image display area using the following equation:
200 * (Lmax – Lmin)
% difference =
(Lmax + Lmin)
where Lmax and Lmin are the maximum and minimum measured
luminance values respectively.
2. Compare the center luminance measurements on both monitors.
The % difference should be ≤20%.
PRECAUTIONS AND Ideally, monitor screens should be free of dust, fingerprints, and other
CAVEATS marks. Similarly, there should be no “shiny” patches or obvious non-
uniformities on the surface. As described below, significant blemishes
that interfere with the interpretation or QC of images must be corrected.
Most problems can be corrected by cleaning according to the
manufacturer’s instructions. However, if cleaning does not correct the
problem, the manufacturer should be contacted to evaluate and correct
the problem.
In most cases, Monitor Manufacturer Automated Tests and action
limits are available in manufacturer manuals or documents published
by the manufacturer. These tests are extremely valuable in maintaining
quality and are specific to each manufacturer. If a Monitor Manufacturer
Automated Test is available, the medical physicist should assist the facility
in verifying that the automated system is set up and functioning properly.
Monitor Condition
Any large, significant blemish that interferes with the interpretation or QC
of images is a failure. (If there are questions regarding the significance of a
monitor blemish, the lead interpreting radiologist should be consulted.)
ACR DM Phantom
1. Artifacts must not be clinically significant. This aspect of the test
fails if any artifacts are in a location that could impact clinical
interpretation and
a. They are as prominent as (or more prominent than) the visible
test objects in the phantom image, or
Distance Measurement
The distance measured across the wax insert parallel to the anode-cathode
axis must be 70.0 mm ±14.0 mm.
PHYSICIST’S SECTION
1. The test pattern must be centered appropriately.
2. The 0%-5% and 95%-100% contrast boxes must be visible.
MEDICAL
3. The alphanumerics must be sharp and legible.
4. The high-contrast line-pair patterns must be distinguishable at the
center and corners.
5. The grayscale ramps must be smooth and continuous.
Luminance Check
Lmin must be within ±30% of the values specified by the manufacturer,
if available. Lmax must be within ±10% of the values specified by the
manufacturer, if available. If no specified values are available, luminance
values must be Lmin ≤1.5 cd/m2 and Lmax ≥420 cd/m2.
Luminance Uniformity
The calculated % difference for an individual monitor must be ≤30%.
The calculated % difference between a pair of monitors for the center
luminance value must be ≤20%.
PHYSICIST’S SECTION
2. On the form, note the workstation used to print the image.
MEDICAL
3. Optional – print images on other size film.
4. View the image.
Contrast
1. Record the background OD from above in the “Contrast” section of
the form.
2. Measure the OD inside the cavity and record the measurement on
the form.
3. Subtract the background OD from outside of the cavity from the OD
inside the cavity and record it on the form. This is the film contrast.
Contrast = Cavity OD – Background OD
Distance Measurement
1. With a ruler, measure the distance across the wax insert in the anode-
cathode (A-C) direction and record on the form. (See Figure 35.)
PHYSICIST’S SECTION
Background optical density must be ≥1.6. (Background optical densities
between 1.7 and 2.2 are recommended; approximately 2.0 is optimal.)
MEDICAL
Contrast
Contrast (cavity OD – background OD) must be ≥0.1.
Dmax
The Dmax must be ≥3.1 (≥3.5 is recommended).
Distance Measurement
The distance measured across the wax insert parallel to the A-C axis must
be 70.0 mm ±14.0 mm.
TIMEFRAME FOR All failures of required items must be corrected before clinical use.
CORRECTIVE ACTION
PHYSICIST’S SECTION
QC records the facility manager and lead interpreting radiologist
noted that they are aware of this occurrence, why it occurred,
MEDICAL
and that corrective action was taken so it would not occur again.
The site’s technologist QC program would PASS since corrective
action was taken to prevent future occurrences. The medical
physicist should make a note of the circumstances and corrective
action on the form. For example, “the QC technologist was on
temporary disability; since that time a backup QC technologist
was trained to assume QC responsibilities in case of future
unexpected absences.”
d. The QC technologist was routinely subtracting for artifacts when
scoring the ACR DM Phantom. Although the medical physicist
should note this on the form, the overall evaluation for the site’s
technologist QC program should PASS. (However, if the medical
physicist notes during the following year’s annual survey that
the same QC technologist continues to subtract for artifacts, the
overall evaluation for the site’s technologist QC program should
FAIL.)
4. Be sure to note on the Evaluation of Site’s Technologist QC
Program form if the QC technologist is performing his or her QC
responsibilities well.
PRECAUTIONS AND The medical physicist should conduct this review so that he or she
CAVEATS is confident that all required tests have been correctly completed at
the required intervals and appropriate corrective action was taken if
necessary. For example, when reviewing the tests of a QC technologist
or facility that is new to the medical physicist, he or she should review
all the tests and corrective action for the previous year. Similarly, if the
equipment is new to the facility, a thorough review of QC should be done.
If the medical physicist establishes confidence that QC is being conducted
as required, he or she may only need to review samples of the QC during
future surveys. If the samples reveal missing data, irregular results, or
missing corrective action, a complete review should be conducted.
This important evaluation helps the facility prepare for its annual MQSA
inspection by identifying missing tests or inadequate documentation. If
tests are found missing, however, it is imperative that the QC technologist
does not perform and back-date any tests to compensate for the omissions.
The medical physicist should advise the facility to make a note that they are
aware of the missing tests and document efforts to prevent occurrence in
the future. See the FDA’s topic on Radiologic Technologists: Falsification
of Documentation for information on QC records.
Medical physicists should provide the QC technologist with one-on-one
training or recommend sources of appropriate training if deficiencies in
the conduct of the tests are noted. Although it is the medical physicist’s
responsibility to perform this review annually and to bring any
deficiencies to the attention of the facility and lead interpreting physician,
it is the responsibility of the facility and lead interpreting physician to see
that measures are taken to ensure that all required tests are performed as
specified by the medical physicist and at the required frequencies.
PERFORMANCE CRITERIA 1. The Site’s Technologist QC Program must receive an overall PASS
AND CORRECTIVE ACTIONS evaluation.
2. If significant deficiencies are found, they must be documented in the
form, and action should be taken to correct the problems. Depending
on the nature of the deficiency, correction may entail actions such as
obtaining appropriate test equipment or providing additional training
or coursework for the QC technologist, or facility management could
decide to provide sufficient time for the QC technologist to perform the
required duties. QC technologists should be assisted by both the medical
physicist and facility management to resolve any identified issues.
TIMEFRAME FOR Failures must be corrected within 30 days.
CORRECTIVE ACTION
APPLICABLE MQSA
REQUIREMENTS 900.12(e) Quality assurance—equipment. (9) Surveys. (ii) The results of all tests
conducted by the facility in accordance with paragraphs (e)(1) through (e)(7)
[mostly technologist quality control tests] of this section, as well as written
documentation of any corrective actions taken and their results, shall be
evaluated for adequacy by the medical physicist performing the survey.
Note: This evaluation may be done by the primary site’s medical physicist
PHYSICIST’S SECTION
or the medical physicist providing services for the offsite equipment. In any
case, the evaluation must be performed and documented at least annually
MEDICAL
for each display device.
APPLICABLE MQSA
REQUIREMENTS 900.12(e) Quality assurance—equipment. (9) Surveys. (ii) The results of all tests
conducted by the facility in accordance with paragraphs (e)(1) through (e)(7)
[mostly technologist’s quality control tests] of this section, as well as written
documentation of any corrective actions taken and their results, shall be
evaluated for adequacy by the medical physicist performing the survey.
PRECAUTIONS AND Most manufacturers provide specific instructions for system calibrations
CAVEATS (e.g., detector calibration). See the manufacturer’s documentation for
precautions and caveats.
PHYSICIST’S SECTION
PERFORMANCE CRITERIA Manufacturers’ recommendations should be followed.
AND CORRECTIVE ACTIONS
MEDICAL
TIMEFRAME FOR Manufacturers’ recommendations should be followed.
CORRECTIVE ACTION
Note: The Collimation test for DBT units is performed in 2D mode, not DBT
mode.
PHYSICIST’S SECTION
of the deviation: if the radiation field extends beyond the edge of the
light field, record the value as positive; if the radiation field does not
extend fully to the edge of the light field, record the value as negative.
MEDICAL
2. From the image in the acquisition workstation, determine the
difference between each light field edge and the border of the visible
image, maintaining the sign: if the coin is fully visible, the difference
will be positive; if the coin is only partially visible, the difference will
be negative.
3. For each edge, sum the values determined in steps 1 and 2 of this
section; this is the deviation between the radiation field and the
visible image.
4. For the larger coin, measure the distance between the diameter of
the coin in the image on the acquisition workstation and the dimension
of the coin perpendicular to the chest-wall edge as the distance
from the paddle edge to the image receptor edge, maintaining the sign.
(See Figure 36.)
a. If the coin is not fully visible, record the value as positive.
b. If the coin is fully visible, record the value as negative.
PRECAUTIONS AND It may be necessary to correct for magnification effects in the images. As
CAVEATS necessary corrections will depend on the exact methodology used and
the system being tested, only this general warning can be provided.
Note: If a large exposure is required for the external detector (such as self-
developing film), care must be taken to protect the unit’s detector. Either
use an appropriate attenuator between the external detector and unit’s
detector or make one exposure to capture the coins on the image, then
place adequate lead over the unit’s detector to complete the light field/
radiation field measurement with the external detector. A 3 mm (1/8”)
aluminum plate, large enough to cover the entire detector, also may be laid
directly on the breast support surface.
PERFORMANCE CRITERIA AND 1. If a light localizer is used, congruence of the light field with the
CORRECTIVE ACTIONS radiation field should be such that the total misalignment (sum
of misalignments on opposite sides) is within 2% of source-image
receptor distance (SID) as required by 21 CFR §1020.31(f)(3) [13].
The x-ray field should not extend beyond any of the four sides
of the image receptor by more than +2% of the SID. At the chest
wall side, the radiation field must extend to the edge of the image
to minimize excluding breast tissue adjacent to the chest wall.
Ideally, the entire image receptor should be exposed. However,
since most manufacturers have designed their collimators to meet
the FDA manufacturer’s requirement for general x-ray equipment,
which specifies that the x-ray field not exceed the size of the image
receptor, a tolerance of −2% of the SID on the right and left sides is
acceptable. (For example, an image with a 1.1-cm “white gap” on its
left side would be acceptable on a unit with a SID of 60 cm). A greater
tolerance of −4% of SID is acceptable at the anterior side of the film
to allow for appropriate image marking.
PHYSICIST’S SECTION
MEDICAL
Note: If the facility changes technique factors (that impact beam quality and
dose) for a 4.2-cm thick compressed breast consisting of 50% glandular and
50% adipose tissue, and the HVL was not previously determined for those
factors, the medical physicist should measure the HVL for the appropriate
factors during the annual survey. Also, if there is any concern regarding a
significant dose change, the HVL should be re-evaluated.
Note: The use of type 1100 aluminum alloy for HVL measurement can give
(depending on specific samples) HVL values up to 7.5% lower than those
measured using high-purity aluminum. If type 1100 aluminum is used,
results should be corrected to agree with those obtained using high-purity
aluminum.
TEST PROCEDURE 1. Raise the breast compression paddle as close as possible to the x-ray
tube.
2. Cover the detector with a protective device (e.g., a lead sheet or lead
apron).
3. Place the dosimeter 4.2 cm above the image receptor holder assembly,
centered left to right and 4 cm in from the chest-wall edge of the
image receptor. The dosimeter should be fully within the x-ray field.
(See Figure 37.)
A B
Figure 37. HVL setup. Compression device positioned so that it is as close to the x-ray
tube as possible. A. Side view diagram. B. Top view diagram. C. Photo.
4. Select the target/filter and kVp used for the ACR DM Phantom
PHYSICIST’S SECTION
acquisition, and record the information on the data form.
5. Set the unit to manual mAs, with a time setting long enough to
MEDICAL
provide an exposure sufficiently close to the ACR DM Phantom
technique.
6. Use the collimator, if possible, to collimate the x-ray beam so that the
dosimeter is just fully exposed (to minimize backscatter).
7. Make an exposure without aluminum sheets between the x-ray tube
and the dosimeter.
8. Only 2 additional exposures are necessary to complete the test. Select
two thicknesses that will most likely result in exposures just above
and just below one-half the original exposure reading (taken without
any added aluminum sheets between the x-ray tube and dosimeter).
a.
Add sufficient aluminum (e.g., 0.3 mm) between the x-ray
tube and the dosimeter by placing the aluminum on top of the
compression paddle. Use the light field (if available) to verify
that the x-ray path to the dosimeter is fully blocked by the
aluminum sheet(s). Make an exposure and record the dosimeter
reading. This reading should be greater than one-half the original
exposure reading.
The HVL shall meet the specifications of FDA’s Performance Standards for
Ionizing Radiation Emitting Products (Part 1020.30) for the minimum HVL.
These values, extrapolated to the mammographic range, are shown in the
table below. Values not shown may be determined by linear interpolation or
extrapolation.
PHYSICIST’S SECTION
MEDICAL
TEST PROCEDURE 1. In manual timing mode, select the most commonly used clinical kVp
setting (within the specified accuracy range of the meter) and record
on the data form. Also record nominal focal spot size, exposure time,
and mA (or mAs) setting.
2. Set up the test device following the manufacturer’s instructions.
3. Make three exposures in the same manual mode settings, and record
the measured kVp values.
4. Repeat the procedure at the lowest clinically used kVp that can be
measured by the kVp test device and the highest available clinically
used kVp, but make only one exposure at each setting. (Reproducibility
needs to be checked only at the most commonly used clinical kVp
unless variability is suspected at other settings.)
DATA ANALYSIS AND 1.
To determine kVp accuracy, average the readings for each kVp setting
INTERPRETATION tested and compare this average value with the value of the preset
nominal kVp.
2. To determine kVp reproducibility, compute the standard deviation
of the kVp values for each kVp setting and calculate the coefficient of
variation of kVp (standard deviation divided by the mean).
PERFORMANCE CRITERIA 1. If the average measured kVp differs by more than ±5% (e.g., ±1.5 kVp
AND CORRECTIVE ACTIONS at 30 kVp) from the nominal kVp setting, the unit should be checked
by appropriate service personnel.
2. If the coefficient of variation exceeds 0.02 for any kVp setting, the
unit should be checked by appropriate service personnel.
TIMEFRAME FOR • When the test is performed for an MEE, all failures must be corrected
CORRECTIVE ACTION before clinical use.
• hen the test is performed for troubleshooting, all failures must be
W
corrected within 30 days.
PHYSICIST’S SECTION
MEDICAL
Figure 39. Phantom placed on the image receptor with the aluminum sheet placed
correctly.
10. One minute after the first image was acquired, or within as short
PHYSICIST’S SECTION
a time as the system will allow if longer than 1 minute, acquire a
second image using the same techniques as the first image.
MEDICAL
11. Use a circular or rectangular region of interest tool (area ~1 cm2) to
measure the mean signal value in the following 3 regions in the raw
version of the second image acquired (see Figure 40):
a. The mean signal (S1) value over the uniform attenuator on the
side where the attenuator was present in the first image
b. The mean signal (S2) value over the uniform attenuator plus the
aluminum square on the side where the attenuator was present in
the first image
c. The mean signal (S3) value over the uniform attenuator plus the
aluminum square on the side where no uniform attenuator was
present in the first image
DATA ANALYSIS AND Calculate the Ghosting Index using the formula:
INTERPRETATION S –S
Ghosting Index = 3 2
S1 – S2
PERFORMANCE If the ghosting index is more than ±0.3, repeat the test. If the ghosting
CRITERIA AND index still exceeds ±0.3, service personnel should be contacted.
CORRECTIVE ACTIONS
TEST PROCEDURE
Note: The measurement procedures described below follow, as closely as
possible, those recommended by the National Association of Photographic
Manufacturers (NAPM). Additional measurements are described in the
NAPM standard that the medical physicist may wish to include, e.g.,
luminance uniformity.
1.
Reproduce the typical ambient lighting conditions for the reading
room, including overhead and task lighting, that is used when
mammograms are interpreted.
2.
For each viewbox used for mammographic interpretation, turn
on the lights in the viewbox at least 30 minutes before making the
following measurements.
3. Place the luminance meter with its detector parallel to and facing the
viewbox surface.
4.
Make the measurement and record the result as the viewbox
luminance.
5. Visually inspect each viewbox for uniformity of luminance and
PHYSICIST’S SECTION
for uniformity of color of the lighting. Note nonuniformities. Also
evaluate viewboxes for proper function of masking devices and the
presence of dirt or marks.
MEDICAL
6. Repeat the tests for all viewboxes used for interpreting mammograms
on film.
DATA ANALYSIS AND Compare the measured luminance to the action limit and determine pass
INTERPRETATION or fail.
PRECAUTIONS AND Many photometric units of measurement exist but are seldom used. For
CAVEATS simplicity we will consider only the SI photometric unit of luminance, the
candela per square meter.
Luminance is the amount of light either scattered or emitted by a surface,
measured in cd/m2 (formerly “nit”). Other units often used for luminance
are foot-lamberts. To convert foot-lamberts to cd/m2, multiply the
numerical value by 10.764/π (3.426).
B. Test Forms
Routine and complete documentation of QC and corrective actions are
a critical part of quality mammography and are required by the FDA
mammography regulations. The QC forms in this section have been
developed to help meet this goal and to comply with FDA requirements.
The QC form number corresponds to the test procedure number provided
in the previous section. Each form has been set up with a brief summary
of the test procedure, an area to record the test conditions and techniques
(to ensure that test conditions that could cause variability in the results
are not changed), and reminders of action limits and timeframes for
corrective action. Be sure to consult the quality control test procedure for
the complete instructions if there are questions about performing tests or
analyzing data.
For simplicity and uniformity, the DBT tests are intended to use the same
forms as the 2D tests. Document that the DBT tests were performed by
selecting the correct image “mode” on the form. (If you are using an
electronic version of the form, use the pull-down menu to select the
mode.) Note that for some DBT tests you will need to use a second or
third page of the form for complete QC documentation.
The forms are also downloadable as Excel spreadsheets from the ACR
Digital Mammography QC Manual Resources website (go to Digital
Mammography Quality Control Test Forms). Although they have been
designed to help you record and analyze your QC results on a computer,
they may also be printed and completed manually.
PHYSICIST’S SECTION
Requirements
MEDICAL
2.
ACR DM Phantom Image Quality
3.
DBT Z Resolution
4.
Spatial Resolution
5.
DBT Volume Coverage
6.
Automatic Exposure Control System Performance
7.
Average Glandular Dose
8.
Unit Checklist
9.
Computed Radiography (if applicable)
10.
Acquisition Workstation (AW) Monitor QC
11.
Radiologist Workstation (RW) Monitor QC
12.
Film Printer QC (if applicable)
13.
Evaluation of Site’s Technologist QC Program
14.
Evaluation of Display Device Technologist QC
15.
Manufacturer Calibrations (if applicable)
16.
Collimation Assessment
2.
MEE or Troubleshooting – kVp Accuracy and
Reproducibility
3.
Troubleshooting – Ghost Image Evaluation
4.
Troubleshooting – Viewbox Luminance
1.
Medical Physicist’s ACR DM QC Test Summary
2.
Mammography Technique Chart
3.
Medical Physicist QC Letter for the Radiologist
PHYSICIST’S SECTION
MEDICAL
Utilize Corrective
Action Log Form
Initials
Required/ Date
Recommended Time Frame Description Completed
E. Supplemental Forms
1.
Facility, Unit, and Test Equipment Data
PHYSICIST’S SECTION
MEDICAL
B. References
1. Department of Health and Human Services. FDA Mammography Quality
Standards, Final Rule. Fed Reg. 1997;62(208):55852-55994.
2. U.S. Food and Drug Administration. Mammography Quality Standards Act
(MQSA) Policy Guidance Help System.
3.
FDA Mammography Quality Standards Act and Program, Facility
Certification and Inspection (MQSA) - Digital Accreditation.
4. Williams MB, Goodale PJ, Butler PF. The current status of full-field digital
mammography quality control. J Am Coll Radiol. 2004;1:936-951.
PHYSICIST’S SECTION
5. Hendrick RE, Bassett L, Botsco MA, et al. Mammography Quality Control
Manual. Reston, Va: American College of Radiology; 1999.
6.
Dance DR, Skinner CL, Young KC, Beckett JR, Kotre CJ. Additional
MEDICAL
factors for the estimation of mean glandular breast dose using the UK
mammography dosimetry protocol. Phys Med Biol. 2000;45:3225-3240.
7. Dance DR, Young KC, van Engen RE. Further factors for the estimation
of mean glandular dose using the United Kingdom, European and IAEA
breast dosimetry protocols. Phys Med Biol. 2009;54(14):4361-4372.
8. Dance DR, Young KC, van Engen RE. Estimation of mean glandular dose
for breast tomosynthesis: factors for use with the UK, European and IAEA
breast dosimetry protocols. Phys Med Biol. 2011;56:453-471.
9. van Engen RE, Bosmans H, Bouwman RW, et al. Protocol for the Quality
Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis
Systems, Version 1.01. Nijmegen, The Netherlands: European Reference
Organisation for Quality Assured Breast Screening and Diagnostic Services
(EUREF); June 2016.
10.
Brateman LF, Heintz PH. Solid-state dosimeters: a new approach for
mammography measurements. Med Phys. 2015;42:542-557.
11. Samei E, Badano A, Chakraborty D, et al. Assessment of display performance
for medical imaging systems: executive summary of AAPM TG18 report.
Med Phys. 2005;32:1205-1225.
12.
Gray JE. Use of the SMPTE test pattern in picture archiving and
communication systems. J Digit Imaging. 1992;5:54-58.
13. Electronic products; performance standard for diagnostic x-ray systems and
their major components. Fed Regist. 2005;70:33997-34042.
C. Additional Resources
– Berns EA, Hendrick RE, Cutter GR. Optimization of technique factors for a
silicon diode array full-field digital mammography system and comparison
to screen-film mammography with matched average glandular dose. Med
Phys. 2003;30:334-340.
– Bloomquist AK, Yaffe MJ, Pisano ED, et al. Quality control for digital
mammography in the ACRIN DMIST trial: part I. Med Phys. 2006;33:719-
736.
– Kruger RL, Schuler BA. A survey of clinical factors and patient dose in
mammography. Med Phys. 2001;28:1449-1454.
– Pollard BJ, Samei E, Chawla AS, et al. The influence of increased ambient
lighting on mass detection in mammograms. Acad Radiol. 2009;16:299-304.
– Quality Control Manual Template for Manufacturers of Displays and
Workstations Devices Labeled for Final Interpretation in Full-Field Digital
Mammography. Rosslyn, Va: National Electrical Manufacturers Association;
XR 22-2006; 2006.
– Quality Control Manual Template for Manufacturers of Hardcopy Output
Devices Labeled for Final Interpretation in Full-Field Digital Mammography.
Rosslyn, Va: National Electrical Manufacturers Association; XR 23-2006;
2006.
Appendices
Contents
I. REVISIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255
II. A
CR DIGITAL MAMMOGRAPHY (DM) PHANTOM
SCORING GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
A. Phantom. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
B. Scoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258
III. ARTIFACT EVALUATION GUIDE. . . . . . . . . . . . . . . . . . . . . 259
A. Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
B. Examples of Clinically Significant Artifacts. . . . . . . . . . . . . . . . 260
C. E xamples of Minor, Clinically Insignificant
Artifacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 262
APPENDICES
Figure 2. ACR DM Phantom wax insert map (test object sizes are not to scale).
APPENDICES
B. Scoring
1. For digital breast tomosynthesis images, scroll to the slice or slab in
which the test objects are best visualized.
2.
Adjust window width and window level to optimize visualization of
each set of test objects.
Artifact A. Procedure
Evaluation Guide 1. Zoom image to full resolution and pan.
2.
Using approximately the same window width and window level
settings used to evaluate the test objects, examine the entire phantom
for both broad area artifacts and detailed artifacts. Entire field of view
should be free of artifacts that may impede clinical interpretation.
3.
Artifact evaluations are assigned a “Fail” if artifacts are present that
impede clinical interpretation.
4.
Artifact evaluation passes if the image is free of artifacts (see Figure 4) or
artifacts are clinically insignificant.
APPENDICES
2. Ghosting.
3. Collimator cut-off.
Figure 7. White line along chest-wall edge of image due to collimator misalignment.
APPENDICES