Bulletin Endura & Pulsera

Download as pdf or txt
Download as pdf or txt
You are on page 1of 15

Release Bulletin

BV Endura & BV Pulsera


Release 2.2.4

CSIP level 1 Documentation

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Proprietary Notice CSIP Level 1:
This document and the information contained in it is strictly reserved for current Philips Medical Systems
(“Philips”) personnel, Philips licensed representatives and Philips customers who have purchased a valid
service agreement for use by the customer’s designated in-house service employee on equipment located at
the customer’s designated site. Use of this document by unauthorized persons is strictly prohibited. This
document must be returned to Philips when the user is no longer licensed and in any event upon Philips’ first
written request.
Liability / Warranty Disclaimer:
Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not
limited to, the implied warranties of merchantability and fitness for a particular purpose. Philips has taken
care to ensure the accuracy of this document. However, Philips assumes no liability for errors or omissions
and reserves the right to make changes without further notice to any products herein to improve reliability,
function, or design. Philips may make improvements or changes in the product(s) or program(s) described in
this document at any time.

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin
BV Endura & BV Pulsera

Release 2.2.4

Author: Philips Medical Systems Netherlands


BL GXR – Software Management

Date: July 23rd , 2007

Doc.: 4522 210 46234

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
CSIP Level 1 doc. no: 4522 210 46234
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Contents

1. INTRODUCTION .....................................................................................................................................1
1.1 How to read this bulletin ..........................................................................................................................1
1.2 Applicable documents..............................................................................................................................1
1.2.1 Mandatory documents..........................................................................................................................1
1.3 Abbreviations and definitions...................................................................................................................2

2 CHARACTERISTICS OF THE RELEASE ..............................................................................................3


2.1 Compatibility ............................................................................................................................................3
2.2 Operational restrictions............................................................................................................................3
2.3 New Features ..........................................................................................................................................3
2.4 Solved Problems......................................................................................................................................4
2.5 Known Problems......................................................................................................................................4

3 SW INSTALLATION................................................................................................................................5
3.1 Creating a backup of the examination settings database........................................................................6
3.2 Programming the Exam File ....................................................................................................................6
3.3 Restore the Exam File .............................................................................................................................6
3.4 Quick Check of the System .....................................................................................................................6

4 RELEASE CONTENTS ...........................................................................................................................7

5 COMMUNICATION WITH PHILIPS MEDICAL SYSTEMS ....................................................................8


5.1 Ordering procedure..................................................................................................................................8
5.2 Problem reporting ....................................................................................................................................8
5.2.1 Purpose ................................................................................................................................................8
5.2.2 Procedure.............................................................................................................................................8
5.2.3 Mailing ..................................................................................................................................................8
5.2.4 Follow up ..............................................................................................................................................8
5.3 X-RAY Helpdesk......................................................................................................................................9

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
CSIP Level 1 doc. no: 4522 210 46234
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
1. INTRODUCTION
This release bulletin contains information about the Release 2.2.4 Software of the BV Endura & BV Pulsera
System for the Surgery systems. The information can be used for:
• Initial installation of the software on a new system
• Upgrade installation of the software to this release
• Re-installing the software after an error

The intended reader is the operations and service engineer. This bulletin will provide information about the
major features of this release and the differences with respect to the previous version.

1.1 HOW TO READ THIS BULLETIN


A brief outline of the chapters is given below:
Chapter 1: Introduction
Chapter 2: Describes the contents of the software delivery, the functionality of the software and its
resolved and known problems
Chapter 3: Describes the software installation

1.2 APPLICABLE DOCUMENTS


1.2.1 MANDATORY DOCUMENTS
1. Service Manuals BV Family system R2 4522 981 32293, 4522 981 32303,
4522 981 32313, 4522 981 32324
2. Instructions for use BV Pulsera 9896 001 93131
3. Instructions for use BV Endura 9896 001 93121
4. Instructions for use BV Pulsera R2.2 addendum 3D-RX 4522 165 14751
5. Instructions for use BV Pulsera R2.2 addendum 15kW 4522 165 15161

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
1
1.3 ABBREVIATIONS AND DEFINITIONS
3D-RX 3D rotational X-ray
DFI Digital Fluoroscopy Imaging
FPR Field Problem Report
GXR General X-Ray
HW Hardware
MIT Medical IT
NA Not Applicable
PEI Product End Item
SW Software
SuCo Surgery Coordinator

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
2
2 CHARACTERISTICS OF THE RELEASE
This release contains all the required functionality of the BV Endura & BV Pulsera.
The software is identified by the following id: Release name: 2.2.4

2.1 COMPATIBILITY
• This Release 2.2.4 release is compatible with release Release 2.1.6, although the examination settings
database is not. It is compatible with all the previous 2.2 releases. See chapter 3 and 3.3 how to handle
when coming from Release 2.1.6.

2.2 OPERATIONAL RESTRICTIONS

With this release, there are no operational restrictions.

2.3 NEW FEATURES

• Configurable auto push export of 3D runs.

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
3
2.4 SOLVED PROBLEMS
This is the list of the solved problems since release 2.2.3:
• FPR SURG00008619: C-arm restarts itself when remote control is used.
• FPR SURG00008281: DICOM MPPS does not work with GE PACS.
• FPR SURG00008406: Fluoro time not correct in dose report.

This is the list of the solved problems since release 2.2.2:


• System blockage after set caps-lock and select WS button.
• Change default-setting for all Vaculair-exams Endura
• All translations have been validated.

This is the list of the solved problems since release 2.2.1:


• Error in Pulsera examination settings for USA.

This is the list of the solved problems since release 2.1.6:


• Worklist has been increased to 100 patients.
• Improvements in the user interface region.
• More languages.

2.5 KNOWN PROBLEMS


This is an overview of the known software problems within the BV Endura & BV Pulsera Release 2.2.4:

These are the known problems of this release:


• Black image in DVD title menu after pulsed or continues HDF with auto recording to DVD.
• If another language is selected and a scheduled examination is selected for acquisition then warning
“Examination type set to default” is displayed very short.

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
4
3 SW INSTALLATION
The system software for the BV Endura & BV Pulsera system has to be downloaded from CD-Rom to the
system in case of a software update, a new release or reinstallation.

NOTE
In case of a software upgrade, e.g. coming from release 2.1.6, a backup must be made of the examination
settings database, for having the possibility to import your old settings with this new release or to go back to
the old situation. See 3.1.

The software can be downloaded in two ways:


• With an Ethernet connection (with a network cross cable). This takes about 2 minutes.
• With a dial-up connection (with a serial cable). This takes about 25 minutes.

The two ways are described below. For a drawing of the connections, see 0-Level documentation part3 R2.2
4522 981 32303.

NOTE
On a BV Endura & BV Pulsera system the system software has already been installed in the factory; for
installation of a new DFI skip steps 3, 4 and 5.

1. Connect a cable between the service PC and the BV Family System: a cross cable to the Service
Ethernet port of the BV Family System or a serial cable to the RS232 Service port of the DFI.
2. Turn on the MVS and wait until the system is operational.
3. In the Setup panel of the MVS: select enable service.
4. In BV-Scope select Install > Program Manual > Network Configuration.
5. Enable the Extended Service Access.
6. Put the CD 4522 210 46214 into the CD-ROM drive of the service PC that has been connected to
the system.
7. Start the SWDownload.exe application on the CD-ROM. The window below appears.

8. In case of connection with a cross cable: select Ethernet Connection. In case of connection with a
serial cable: select Dial-up Connection.
9. Enter the Service IP address shown on the Setup panel of the MVS. (In case of a new DFI enter the
default IP address: 192.168.1.61).
10. Click the Download Software button. The software will be downloaded to the DFI and the result will
be displayed in the output window.
11. After about two minutes (in case of an Ethernet connection), the message ‘Download completed
successfully’ appears: close the window and switch the BV Endura & BV Pulsera off and on.
12. Close all windows.
13. In case of a software upgrade from release 2.1.6, the flash prom will be programmed on the DFI;
follow the instructions on the LIVE monitor.
14. The DFI will download the software to the Midas board in the C-stand if the software version is
newer. It will display the download progress on the User Interface of the C-stand. In addition, the
SHA1 LEDs will flash. The MVS will display being in “stand-alone” mode.
15. Wait until download will be finished (progress is 100% will be visible on the User Interface on the C-
stand). The stand will reboot automatically.

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
5
16. After reboot, the complete system will be operational and standby.

3.1 CREATING A BACKUP OF THE EXAMINATION SETTINGS DATABASE


1. In the Setup panel of the MVS: select enable service.
2. In BV- Scope select Install > Export > Examination.
3. Make a backup of the examination file: EXAMyyyy.END or EXAMyyyy.PUL.
4. Store the examination file on the hard disk of the service PC.

NOTE
With this release new examination files are available. In case of a new installation or 3D-RX software it is
compulsory to do the following extra steps (7) to use these new examination files:

3.2 PROGRAMMING THE EXAM FILE


1. In the Setup panel of the MVS: select enable service.
2. In BV- Scope select Install > Program Automatic > Examination.
3. In the instruction window select Intern.
4. Click [ok] to continue installation of the examination data into the system.
5. Select in the next instruction window the configuration, which is applicable.
6. Click [ok] to finish.
7. Restart the system.

In case of a software upgrade, coming from release 2.1.6, do the following extra steps (7) to restore the
original examination settings:

3.3 RESTORE THE EXAM FILE


1. In the Setup panel of the MVS: select enable service.
2. In BV- Scope select Install > Program Automatic > Examination.
3. In the instruction window select Extern, browse for the previous (e.g. EXAMyyyy.PUL) made
backup file.
4. Click [ok] to continue installation of the examination data into the system.
5. Select in the next instruction window the configuration, which is applicable.
6. Click [ok] to finish.
7. Restart the system.

3.4 QUICK CHECK OF THE SYSTEM


1. Check if in the start-up panel Release 2.2.4 is mentioned.
2. Check the examination names.
3. Check during the following steps (if applicable) if the image on the monitor screen looks normal.
4. Select an examination and start LDF fluoroscopy for a few seconds.
5. Select a vascular examination and make a subtraction run. (Only if available on the system)
6. Select roadmap and make a roadmap run. (Only if available on the system)

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
6
4 RELEASE CONTENTS

The contents of the BV Endura & BV Pulsera software package, 4522 210 46204 (Software Packet BV
Endura & BV Pulsera Release 2.2.4):

Code-number Date Rel 2.2.4 Date Rel 2.2.3 Date Rel 2.2.2 Date Rel 2.2.1 Description
4522 210 46214 2007-07-23 Application CD BV family release 2.2.3

4522 210 46234 Release Bulletin


4522 210 46212 2006-11-27 Application CD BV family release 2.2.3
No 12nc 2006-10-18 Application CD BV family release 2.2.2
4522 210 46211 2006-10-03 Application CD BV family release 2.2.1
4522 210 46146 Application CD BV family release 2.1.6

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
7
5 COMMUNICATION WITH PHILIPS MEDICAL SYSTEMS

5.1 ORDERING PROCEDURE


The software packages can be ordered with the code numbers given in chapter 4 on page 7 of this
document.

5.2 PROBLEM REPORTING


5.2.1 PURPOSE
Field Problem Report (FPR) describes a product problem with respect to the performance, installability,
serviceability or maintainability, which cannot be solved by the local service organization.

5.2.2 PROCEDURE
The preferred method to create FPR’s is by using the FPR form version as displayed on the Customer
Service website. Please use the English language so other engineers may benefit the knowledge and facts
from your FPR !! The resulting FPR file should be sent as an enclosed file to the helpdesk X-ray via E-mail.
Please use the latest FPR form version as displayed on the Customer Support website, InCentre.
It may be possible that your local service organization requires defining a local reference number to the FPR.
Consult locally.

5.2.3 MAILING
Please send the FPR file to the appropriate BL Helpdesk.
To facilitate the rapid processing of the problem report, send also the complete error-log-file.

Product BL Helpdesk
Cardio / Vascular / Remote Control / Helpdesk X-ray
Surgery
CT Helpdesk CT
MR Helpdesk MR
US Helpdesk US
MIT products Helpdesk MIT
Universal R/F / X-ray tubes / Mammo Helpdesk
Hamburg

5.2.4 FOLLOW UP
Maintenance teams operating within the BL will process the FPR.
The reply to an FPR as well as a monthly overview is sent to the SSD contact person.
All incoming FPR's may help in obtaining BETTER products.

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
8
5.3 X-RAY HELPDESK
If there are any questions/problems that need assistance, then contact:

Philips Medical Systems


Helpdesk X-Ray
Address: Building QG150
P.O. Box 10000
5680 DA Best The Netherlands
Telephone +31 40-2764777
FAX +31 40-2762628
E-mail [email protected]

CSIP Level 1 doc. no: 4522 210 46234


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2007 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
9

You might also like