Bulletin Endura & Pulsera
Bulletin Endura & Pulsera
Bulletin Endura & Pulsera
Release 2.2.4
1. INTRODUCTION .....................................................................................................................................1
1.1 How to read this bulletin ..........................................................................................................................1
1.2 Applicable documents..............................................................................................................................1
1.2.1 Mandatory documents..........................................................................................................................1
1.3 Abbreviations and definitions...................................................................................................................2
3 SW INSTALLATION................................................................................................................................5
3.1 Creating a backup of the examination settings database........................................................................6
3.2 Programming the Exam File ....................................................................................................................6
3.3 Restore the Exam File .............................................................................................................................6
3.4 Quick Check of the System .....................................................................................................................6
The intended reader is the operations and service engineer. This bulletin will provide information about the
major features of this release and the differences with respect to the previous version.
2.1 COMPATIBILITY
• This Release 2.2.4 release is compatible with release Release 2.1.6, although the examination settings
database is not. It is compatible with all the previous 2.2 releases. See chapter 3 and 3.3 how to handle
when coming from Release 2.1.6.
NOTE
In case of a software upgrade, e.g. coming from release 2.1.6, a backup must be made of the examination
settings database, for having the possibility to import your old settings with this new release or to go back to
the old situation. See 3.1.
The two ways are described below. For a drawing of the connections, see 0-Level documentation part3 R2.2
4522 981 32303.
NOTE
On a BV Endura & BV Pulsera system the system software has already been installed in the factory; for
installation of a new DFI skip steps 3, 4 and 5.
1. Connect a cable between the service PC and the BV Family System: a cross cable to the Service
Ethernet port of the BV Family System or a serial cable to the RS232 Service port of the DFI.
2. Turn on the MVS and wait until the system is operational.
3. In the Setup panel of the MVS: select enable service.
4. In BV-Scope select Install > Program Manual > Network Configuration.
5. Enable the Extended Service Access.
6. Put the CD 4522 210 46214 into the CD-ROM drive of the service PC that has been connected to
the system.
7. Start the SWDownload.exe application on the CD-ROM. The window below appears.
8. In case of connection with a cross cable: select Ethernet Connection. In case of connection with a
serial cable: select Dial-up Connection.
9. Enter the Service IP address shown on the Setup panel of the MVS. (In case of a new DFI enter the
default IP address: 192.168.1.61).
10. Click the Download Software button. The software will be downloaded to the DFI and the result will
be displayed in the output window.
11. After about two minutes (in case of an Ethernet connection), the message ‘Download completed
successfully’ appears: close the window and switch the BV Endura & BV Pulsera off and on.
12. Close all windows.
13. In case of a software upgrade from release 2.1.6, the flash prom will be programmed on the DFI;
follow the instructions on the LIVE monitor.
14. The DFI will download the software to the Midas board in the C-stand if the software version is
newer. It will display the download progress on the User Interface of the C-stand. In addition, the
SHA1 LEDs will flash. The MVS will display being in “stand-alone” mode.
15. Wait until download will be finished (progress is 100% will be visible on the User Interface on the C-
stand). The stand will reboot automatically.
NOTE
With this release new examination files are available. In case of a new installation or 3D-RX software it is
compulsory to do the following extra steps (7) to use these new examination files:
In case of a software upgrade, coming from release 2.1.6, do the following extra steps (7) to restore the
original examination settings:
The contents of the BV Endura & BV Pulsera software package, 4522 210 46204 (Software Packet BV
Endura & BV Pulsera Release 2.2.4):
Code-number Date Rel 2.2.4 Date Rel 2.2.3 Date Rel 2.2.2 Date Rel 2.2.1 Description
4522 210 46214 2007-07-23 Application CD BV family release 2.2.3
5.2.2 PROCEDURE
The preferred method to create FPR’s is by using the FPR form version as displayed on the Customer
Service website. Please use the English language so other engineers may benefit the knowledge and facts
from your FPR !! The resulting FPR file should be sent as an enclosed file to the helpdesk X-ray via E-mail.
Please use the latest FPR form version as displayed on the Customer Support website, InCentre.
It may be possible that your local service organization requires defining a local reference number to the FPR.
Consult locally.
5.2.3 MAILING
Please send the FPR file to the appropriate BL Helpdesk.
To facilitate the rapid processing of the problem report, send also the complete error-log-file.
Product BL Helpdesk
Cardio / Vascular / Remote Control / Helpdesk X-ray
Surgery
CT Helpdesk CT
MR Helpdesk MR
US Helpdesk US
MIT products Helpdesk MIT
Universal R/F / X-ray tubes / Mammo Helpdesk
Hamburg
5.2.4 FOLLOW UP
Maintenance teams operating within the BL will process the FPR.
The reply to an FPR as well as a monthly overview is sent to the SSD contact person.
All incoming FPR's may help in obtaining BETTER products.