1.1 Basic Measurement and Control Concepts: Quantity Unit Abbreviation
1.1 Basic Measurement and Control Concepts: Quantity Unit Abbreviation
Introduction
In a time of constant and rapid technological development, it would be quite ambitious to develop and
present a course that claimed to cover each and every industrial measuring type of equipment. This
course is not intended to be an encyclopedia of instrumentation and control valves, but rather a
training guide for gaining experience in this fast changing environment.
This course is aimed at providing engineers, technicians and any other personnel involved with
process measurement, more experience in that field. It is also designed to give students the
fundamentals on analysing the process requirements and selecting suitable solutions for their
applications.
The basic set of units used on this course is the SI unit system. This can be summarised in Table 1.1.
Table 1.1
SI units
Quantity Unit Abbreviation
Length metre m
Mass kilogram kg
Time second s
Current ampere A
Temperature degrees Kelvin °K
Voltage volt V
Resistance ohm Ω
Capacitance farad F
Inductance henry H
Energy joule J
Power watt W
Frequency hertz Hz
Charge coulomb C
Force newton N
Magnetic Flux weber Wb
Magnetic Field webers/metre2 Wb/m2
Density kilogram/metre3 kg/m3
2 Instrumentation for Automation and Process Control
There are a number of criteria that must be satisfied when specifying process measurement equipment.
Below is a list of the more important specifications.
1.2.1 Accuracy
The accuracy specified by a device is the amount of error that may occur when measurements are
taken. It determines how precise or correct the measurements are to the actual value and is used to
determine the suitability of the measuring equipment.
Accuracy generally contains the total error in the measurement and accounts for linearity, hysteresis
and repeatability. Figure 1.1 shows errors in measurement.
Reference accuracy is determined at reference conditions, i.e. constant ambient temperature, static
pressure, and supply voltage. There is also no allowance for drift over time.
Figure 1.1
Accuracy terminology
Introduction 3
1.2.3 Budget/Cost
Although not so much a specification, the cost of the equipment is certainly a selection consideration.
This is generally dictated by the budget allocated for the application. Even if all the other
specifications are met, this can prove an inhibiting factor.
More critical control applications may be affected by different response characteristics. In these
circumstances the following may need to be considered:
1.3.1 Hysteresis
Hysteresis is the difference in the output for given input when the input is increasing and output for
same input when input is decreasing. In other words, it is the difference in the way device works
when moving from 0% to 100%, compared to the way the device works when moving from 100% to
0%. When input of any instrument is slowly varied from zero to full scale and then back to zero, its
output varies. One example is shown in Figure 1.2. This is where the accuracy of the device is
dependent on the previous value and the direction of variation. Hysteresis causes a device to show an
inaccuracy from the correct value, as it is affected by the previous measurement.
Figure 1.2
Hysteresis
1.3.2 Linearity
Linearity expresses the deviation of the actual reading from a straight line. If all outputs are in the
same proportion to corresponding inputs over a span of values, then input output plot is straight line
else it will be non linear as shown in Figure 1.3. For continuous control applications, the problems
arise due to the changes in the rate the output differs from the instrument. The gain of a non-linear
device changes as the change in output over input varies. In a closed loop system changes in gain
affect the loop dynamics. In such an application, the linearity needs to be assessed. If a problem does
exist, then the signal needs to be linearised.
4 Instrumentation for Automation and Process Control
Figure 1.3
Linearity
1.3.3 Repeatability
Repeatability defines how close a second measurement is to the first under the same operating
conditions, and for the same input. Repeatability is generally within the accuracy range of a device and
is different from hysteresis in that the operating direction and conditions must be the same.
Continuous control applications can be affected by variations due to repeatability. When a control
system sees a change in the parameter it is controlling, it will adjust its output accordingly. However if
the change is due to the repeatability of the measuring device, then the controller will over-control.
This problem can be overcome by using the deadband in the controller as shown in Figure 1.4;
however repeatability becomes a problem when an accuracy of say, 0.1% is required, and a
repeatability of 0.5% is present.
Figure 1.4
Repeatability
Ripples or small oscillations can occur due to overcontrolling. This needs to be accounted for in the
initial specification of allowable values.
Introduction 5
1.3.4 Response
When the output of a device is expressed as a function of time (due to an applied input) the time taken
to respond can provide critical information about the suitability of the device. A slow responding
device may not be suitable for an application. This typically applies to continuous control applications
where the response of the device becomes a dynamic response characteristic of the overall control
loop. However in critical alarming applications where devices are used for point measurement, the
response may be just as important. Figure 1.5 shows response of the system to a step input.
Figure 1.5
Typical time response for a system with a step input.
Ambient
The surrounding or environment in reference to a particular point or object.
Attenuation
A decrease in signal magnitude over a period of time.
Calibration
The procedure of comparing and determining the performance accuracy is called calibration. To
configure a device so that the required output represents (to a defined degree of accuracy) the
respective input.
Closed loop
Relates to a control loop where the process variable is used to calculate the controller output. In a
closed loop system the control action is independent on desired output.
Coefficient( temperature)
A coefficient is typically a multiplying factor. The temperature coefficient defines how much change
in temperature there is for a given change in resistance (for a temperature dependent resistor).
Cold junction
The thermocouple junction, which is at a known reference temperature.
Compensation
A supplementary device used to correct errors due to variations in operating conditions.
Controller
A device, which operates automatically to regulate the control of a process with a control variable.
Elastic
The ability of an object to regain its original shape, when an applied force is removed. When a force is
applied that exceeds the elastic limit, then permanent deformation will occur.
Excitation
The energy supply required to power a device for its intended operation.
Gain
This is the ratio of the change of the output to the change in the applied input. Gain is a special case of
sensitivity, where the units for the input and output are identical and the gain is unitless.
Hunting
Generally an undesirable oscillation at or near the required setpoint is called hunting. Hunting
typically occurs when the demands on the system performance are high and possibly exceed the
system capabilities. The output of the controller can be overcontrolled due to the resolution of
accuracy limitations.
Ramp
Defines the delayed and accumulated response of the output for a sudden change in the input.
Range
The region between the specified upper and lower limits where a value or device is defined and
operated.
Reliability
The probability that a device will perform within its specifications for the number of operations or
time period specified.
Reproducibility
The similarity of one measurement to another over time, where the operating conditions have varied
within the time span, but the input is restored.
Resolution
The smallest interval that can be identified as a measurement varies.
Resonance
The frequency of oscillation is maintained due to the natural dynamics of the system.
Introduction 7
Self Heating
The internal heating caused within a device due to the electrical excitation. Self-heating is primarily
due to the current draw and not the voltage applied, and is typically shown by the voltage drop as a
result of power (I2R) losses.
Sensitivity
This defines how much the output changes, for a specified change in the input to the device.
Setpoint
Used in closed loop control, the setpoint is the ideal process variable. It is represented in the units of
the process variable and is used by the controller to determine the output to the process.
Span Adjustment
The difference between the maximum and minimum range values. When provided in an instrument,
this changes the slope of the input-output curve.
Steady state
Used in closed loop control where the process no longer oscillates or changes and settles at some
defined value.
Stiction
Shortened form of static friction, and defined as resistance to motion. More important is the force
required (electrical or mechanical) to overcome such a resistance.
Stiffness
This is a measure of the force required to cause a deflection of an elastic object.
Thermal shock
An abrupt temperature change applied to an object or device.
Time constant
The time constant of a first order system is defined as the time taken for the output to reach 63.2% of
the total change, when subjected to a step input change.
Transducer
An element or device that converts information from one form (usually physical, such as temperature
or pressure) and converts it to another (usually electrical, such as volts or millivolts or resistance
change). A transducer can be considered to comprise a sensor at the front end (at the process) and a
transmitter.
Transient
A sudden change in a variable, which is neither a controlled response, nor long lasting.
Transmitter
A device that converts one form of energy to another. Usually from mechanical to electrical for the
purpose of signal integrity for transmission over longer distances and for suitability with control
equipment.
Variable
Generally, this is some quantity of the system or process. The two main types of variables that exist in
the system are the measured variable and the controlled variable. The measured variable is the
measured quantity and is also referred to as the process variable as it measures process information.
The controlled variable is the controller output which controls the process.
8 Instrumentation for Automation and Process Control
Vibration
This is the periodic motion (mechanical) or oscillation of an object.
Zero adjustment
The zero in an instrument is the output provided when no, or zero input is applied. The zero
adjustment produces a parallel shift in the input-output curve.
The Piping & Instrumentation Diagram (&ID) is a multi-disciple drawing (typically controlled by
the process or chemical engineers), which may also be referred to as the Process &
Instrumentation Diagram. It gives a graphical representation of the process including hardware
(Piping, Equipment) and software (Control systems); this information is used for the design
construction and operation of the facility. Other synonyms for the P&ID are:
• MFD – Mechanical Flow Diagram: Used where materials handling is predominant
• EFD – Engineering Flow Diagram
• UFD – Utility Flow Diagram
The P&ID also provides important information needed by the constructor and manufacturer to
develop the other construction input documents (the isometric drawings, or orthographic physical
layout drawings, etc.). The P&ID provides direct input to the field for the physical design and
installation of field-run piping. For clarity, it is usual to use the same general layout of flow paths
on the P&ID as used in the flow diagram.
The P&ID ties together the system description, the flow diagram, the electrical control schematic,
and the control logic diagram. It accomplishes this by showing ALL of the piping, equipment,
principal instruments, instrument loops, and control interlocks. The P&ID contains a minimum
amount if text in the form of notes (the system description minimizes the need for text on the
P&ID). The first P&ID in the set for the job should contain a legend defining all symbols used; if
some symbols are defined elsewhere, it may be appropriate only to reference their source. The
P&ID is also used by the start-up organizations for preparing flushing, testing, and blow-out
procedures for the piping system and by the plant operators to operate the system. The correctness
and completeness of the P&ID drawings are critical to the success of a plant start-up program.
There are three parts to process measurement and control functions as shown on a P&ID. These are:
• Letter (which describe WHAT the device is)
• Symbols or instrumentation bubbles (which describe WHERE the device is)
• Lines connecting the bubbles (which show HOW the data is moved)
First consider the letters. The first letter ALWAYS corresponds to the parameter being measured.
The subsequent letters describe what is being done with the parameter
Identifying letters for Process measurement and control functions are listed below:
Introduction 9
A Analysis Alarm
B Burner
C Conductivity Control
D Density
E Voltage Primary element
F Flow
G Glass (sight tube)
H Hand
I Current (electric) Indicate
J Power
K Time Control station
L Level Light
M Moisture1
O Orifice
P Pressure Point
Q Quantity
R Radioactivity Record, Relief
S Speed Switch, Safety
T Temperature Transmit
U Multivariable Multifunction
V Viscosity Valve
W Weight Well
Y Relay (transformation)
Z Position Drive2
1
Moisture is not the normal use of the letter “M”, and is not conform to ISO-5, but it is one of the
most common uses of the letter “M”.
2
In many industries, “Z” as the second letter is used to show a safety device. This is often done when
the control system is completely separate from the safety system.
Some examples …
Letters Meaning
PT Pressure Transmitter
PI Pressure Indicator
PIT Pressure Indicating Transmitter
LV Level Valve
FQI Flow Quantity Indicator (shows the total
quantity)
The difference between a PT and a PIT is simple … you will not get an indication of the pressure with
a PT
The second portion of the way instrumentation logic is represented on a P&ID is the “instrumentation
bubble”. This shows where the device is located. In addition, a series of horizontal lines is used to
further define the location, and to show who has access to the information. The letters are placed in the
top portion, and the device number is placed in the bottom portion.
10 Instrumentation for Automation and Process Control
Discrete Instruments
LSH
A level switch high (tag number 855) in an auxiliary PLC. Access to the set
855
point is limited (for example, only the control systems engineer has access to
the setpoint)
TIC A temperature indicating controller (tag number 642) with a shared display (for
642
example, a control room with multiple control displays). Everyone has access
to the reading and the setpoint.
The third portion of the instrumentation represented on the P&ID is the way the data is moved from
process to device or device to device. This is done with a set of lines connecting either the process
device itself (for example, a pipe) to the instrument bubble, or connecting instrument bubbles. The
following figure shows some of the ISA standard line symbols.
Introduction 11
Figure 1.6
Instrument representation on flow diagrams (b)
Some examples:
LIC LAH
The information from Level Indicating Controller 442 is
442 442 transferred via a data link (software) within the DCS to Level
Alarm High 442
Flow elements have some special requirements. Sometimes we need only a primary device to measure
the flow. Other times we need a primary and a secondary device. For example, consider measuring
glow with an orifice meter. The primary element is the orifice plate (this is a flow element, so it
would be labelled “FE”). The secondary element could be a differential pressure transmitter, but
because it is in flow service, it would be a Flow transmitter (FT). We can use the instrument symbol
to better define what type of flow meter is being used.
12 Instrumentation for Automation and Process Control
Figure 1.7
Some P& ID symbols for flow measurements
Introduction 13
Finally, the control valve itself can have different actuators. ISA has developed some symbols to show
the different actuators. Some examples are as follows:
Figure 1.8
1.6.1 Advantages
The range of operation not only determines the suitability of the device for a particular application but
also can be chosen for a range of applications. This can reduce the inventory in a plant as the number
of sensors and models decrease. This also increases system reliability as sensing equipment can be
interchanged as the need arises.
An increased operating range also gives greater over and under-range protection, should the process
perform outside of specifications.
Widening the operating range of the sensing equipment may be at the expense of resolution.
Precautions also need to be made when changing the range of existing equipment. In the case of
control systems, the dynamics of the control loop can be affected.
14 Instrumentation for Automation and Process Control
Fast Response
With a fast response, delays are not added into the system. In the case of continuous control, lags can
accumulate with the various control components and result in poor or slow control of the process. In a
point or alarming application, a fast speed of response can assist in triggering safety or shutdown
procedures that can reduce the amount of equipment failure or product lost.
Often a fast response is achieved by sacrificing the mechanical protection of the transducer element.
Good Sensitivity
Improved sensitivity of a device means that more accurate measurements are possible. The sensitivity
also defines the magnitude of change that occurs. High sensitivity in the measuring equipment means
that the signal can be read easily by a controller or other equipment.
High Accuracy
This is probably one of the most important selection criteria. The accuracy determines the suitability
of the measuring equipment to the application, and is often a trade off with cost.
High accuracy means reduced errors in measurement; this also can improve the integrity and
performance of a system.
High over range protection is different to having a wide operating range in that it does not measure
when out of range. The range is kept small to allow sufficient resolution, with the overrange protection
ensuring a longer operating life.
Maintenance is reduced with fewer pieces to wear, replace or assemble. There are also savings in the
time it takes to service, repair and replace, with the associated procedures being simplified.
Cost
Any application that requires a control solution or the interrogation of process information is driven by
a budget. It therefore is no surprise that cost is an important selection criterion when choosing
measurement equipment.
Repeatability
The ability of the measurement system to give same output for same input repeatedly. Good
repeatability ensures measurements vary according to process changes and not due to the limitations of
the sensing equipment.
Introduction 15
Size
This mainly applies to applications requiring specifically sized devices and has a bearing on the cost.
Stable
If a device drifts or loses calibration over time then it is considered to be unstable. Drifting can occur
over time, or on repeated operation of the device. In the case of thermocouples, it has been proven that
drift is more extreme when the thermocouple is varied over a wide range quite often, typically in
furnaces that are repeatedly heated to high temperatures from the ambient temperature.
Even though a device can be recalibrated, there are a number of factors that make it undesirable:
• Labour required
• Possible shutdown of process for access
• Accessibility
Resolution
The resolution is the smallest measurable difference between two consecutive measurements.
Robust
This has the obvious advantage of being able to handle adverse conditions. However this can have the
added limitation of bulk.
Temperature Corrected
Ambient temperature variations often affect measuring devices. Temperature correction eliminates the
problems associated with these changes.
Intrinsic Safety
Required for specific service applications. This requirement is typically used in environments where
electrical or thermal energy can ignite the atmospheric mixture.
Simple to Adjust
This relates to the accessibility of the device. Helpful if the application is not proven and constant
adjustments and alterations are required.
A typical application may be the transducer for ultrasonic level measurement. It is not uncommon to
weld in brackets for mounting, only to find the transducer needs to be relocated.
Non Contact
This is usually a requirement based on the type of material being sensed. Non-contact sensing is used
in applications where the material causes build-up on the probe or sensing devices. Other applications
are where the conditions are hazardous to the operation of the equipment. Such conditions may be
high temperature, pressure or acidity.
Reliable Performance
This is an obvious advantage with any sensing device, but generally is at the expense of cost for very
reliable and proven equipment. More expensive and reliable devices need to be weighed up against the
cost of repair or replacement, and also the cost of loss of production should the device fail. The costs
incurred should a device fail, are not only the loss of production (if applicable), but also the labour
required to replace the equipment. This also may include travel costs or appropriately certified
personnel for hazardous equipment or areas.
Unaffected by Density
Many applications measure process materials that may have variations in density. Large variations in
the density can cause measurement problems unless accounted for. Measuring equipment that is
unaffected by density provides a higher accuracy and is more versatile
The effect of moisture content can cause problems in both cases, i.e. when a product goes from a dry
state to wet, or when drying out from a wet state.
Unaffected by Conductivity
The conductivity of a process material can change due to a number of factors, and if not checked can
cause erroneous measurements. Some of the factors affecting conductivity are:
• pH
• salinity
• temperature
Mounting External to the Vessel
This has the same advantages as non-contact sensing. However it is also possible to sense through the
container housing, allowing for pressurised sensing. This permits maintenance and installation without
affecting the operation of the process.
Another useful advantage with this form of measurement is that the detection obstructions in chutes or
product in boxes can be performed unobtrusively.
This is more a criteria that determines the suitability of the device for the application.
Polarity Insensitive
Sensing equipment that is polarity insensitive generally protects against failure from incorrect
installation.
Generally, spot sensing is done with smaller transducers, with area or average sensing being
performed with large transducers.
Remote Sensing
Sensing from afar has the advantage of being non-intrusive and allowing higher temperature and
pressure ratings. It can also avoid the problem of mounting and accessibility by locating sensing
equipment at a more convenient location.
No Calibration Required
Pre-calibrated equipment reduces the labour costs associated with installing new equipment and also
the need for expensive calibration equipment.
No Moving Parts
The advantages are:
• Long operating life
• Reliable operation with no wear or blockages
If the instrument does not have any moving or wearing components, then this provides improved
reliability and reduced maintenance.
Maintenance can be further reduced if there are no valves or manifolds to cause leakage problems. The
absence of manifolds and valves results in a particularly safe installation, an important consideration
when the process fluid is hazardous or toxic.
FLOW APPLICATIONS
Selecting equipment with low pressure losses results in safer operating pressures with a lower
operating cost.
Good Rangeability
In cases where the process has considerable variations (in flow for example), and accuracy is
important across the entire range of operation, the selecting of equipment with good rangeability is
vital.
Unaffected by Viscosity
The viscosity generally changes with temperature, and even though the equipment may be rated for the
range of temperature, problems may occur with the fluidity of the process material.
No Obstructions
This primarily means no pressure loss. It is also a useful criterion when avoiding equipment that
requires maintenance due to wear, or when using abrasive process fluids.
1.6.2 Disadvantages
The following is a discussion of effects of the disadvantages and reasons for the associated limitations.
Hysteresis
Hysteresis can cause significant errors. The errors are dependent on the magnitude of change and the
direction of variation in the measurement. One common cause of hysteresis is thermoelastic strain.
NOTE: Sometimes Hysteresis is desirable. For a pressure relief valve, hysteresis helps ensure the
valve does not “chatter”.
Linearity
This affects the resolution over the range of operation. For a unit change in the process conditions,
there may be a 2% change at one end of the scale, with a 10% change at the other end of the scale.
This change is effectively a change in the sensitivity or gain of the measuring device.
In point measuring applications this can affect the resolution and accuracy over the range. In
continuous control applications where the device is included in the control loop, it can affect the
dynamic performance of the system.
Indication Only
Devices that only perform indication are not suited for automated control systems as the information is
not readily accessible. Errors are also more likely and less predictable as they are subject to operator
interpretation.
Poor over range protection in the device may not be a problem if the process is physically incapable of
exceeding the operating conditions, even under extreme fault conditions.
Unstable
This generally relates to the accuracy of the device over time. However the accuracy can also change
due to large variations in the operation of the device due to the process variations. Subsequently,
unstable devices require repeated calibration over time or when operated frequently.
20 Instrumentation for Automation and Process Control
Size
Often the bulkiness of the equipment is a limitation. In applications requiring area or average
measurements then too small a sensing device can be a disadvantage in that it does not “see” the full
process value.
Special Cabling
Measurement equipment requiring special cabling bears directly on the cost of the application.
Another concern with cabling is that of noise and cable routing. Special conditions may also apply to
the location of the cable in reference to high voltage, high current, high temperature, and other low
power or signal cabling.
Signal Conditioning
Primarily used when transmitting signals over longer distances, particularly when the transducer signal
requires amplification. This is also a requirement in noisy environments. As with cabling, this bears
directly on the cost and also may require extra space for mounting.
Maintenance
High maintenance equipment increases the labour, which become a periodic expense. Some typical
maintenance requirements may include the following:
• Cleaning
• Removal/replacement
• Calibration
If the equipment is fragile then there is the risk of it being easily damaged due to repeated handling.
Sampled measurement equipment is mainly used for quality control applications where specific
samples are required and the quality does not change rapidly.
Pressure Applications
This applies to applications where the measuring equipment is mounted in a pressurised environment
and accessibility is impaired. There are obvious limitations in installing and servicing such equipment.
In addition are the procedures and experience required for personnel working in such environments.
Introduction 21
Access
Access to the process and measuring equipment needs to be assessed for the purpose of:
• The initial installation
• Routine maintenance
The initial mounting of the measuring equipment may be remote from the final installation; as such the
accessibility of the final location also needs to be considered. This may also have a bearing on the
orientation required when mounting equipment.
The cost of the installation is greatly increased if no compressed air is available for such a purpose.
More common is the requirement to tap into the existing supply but this still requires the installation of
air lines.
Material Build-up
Material build-up is primarily related to the type of process material being measured. This can cause
significant errors, or degrade the operating efficiency of a device over time. There are a number of
ways to avoid or rectify the problems associated with material build-up:
• Regular maintenance
• Location (or relocation) of sensing equipment
• Automated or self cleaning (water sprays)
Radiation
The use of radioactive materials such as Cobalt or Cesium often gives accurate measurements.
However, problems arise from the hazards of using radioactive materials, which require special safety
measures. Precautions are required when housing such equipment, to ensure that it is suitably enclosed
and installation safety requirements are also required for personal safety.
Electrolytic Corrosion
The application of a voltage to measuring equipment can cause chemical corrosion to the sensing
transducer, typically a probe. Matching of the process materials and metals used for the housing and
sensor can limit the effects; however in extreme mismatches, corrosion is quite rapid.
High Resistance
Devices that have a high resistance can pick up noise quite easily. Generally high resistance devices
require good practice in terms of cable selection and grounding to minimise noise pickup.
Orientation Dependent
Depending on the technology used, requirements may be imposed on the orientation when mounting
the sensing transducer. This may involve extra work, labour and materials in the initial installation. A
typical application for mounting an instrument vertically would be a variable area flowmeter.
For those technologies that sense through the process material, the phase change can result in
reflections and possibly make the application unmeasurable.
Mechanical Failure
Failure of mechanical equipment cannot be avoided; however the effects and consequences can be
assessed in determining the suitable technology for the application. Flow is probably the best example
of illustrating the problems caused if a measurement transducer should fail. If the device fails, and it is
of such a construction that debris may block the line or a valve downstream, then this can make the
process inoperative until shutdown and repaired.
Filters
There are two main disadvantages with filters:
• Maintenance and cleaning
Introduction 23
The pressure loss can be a process limitation, but from a control point of view can indicate that the
filter is in need of cleaning or replacement.
Flow Profile
The flow profile may need to be of a significant form for selected measuring equipment. Note that the
flow profile is dependent on viscosity and turbulence.
Acoustically Transparent
Measuring transducers requiring the reflection of acoustic energy are not suitable where the process
material is acoustically transparent. These applications would generally require some contact means of
measurement.
Figure 1.9 shows how instrument and control valves fit into the overall control system structure. The
topic of controllers and tuning forms part of a separate workshop.
Figure 1.9
Instruments and control valves in the overall control system
24 Instrumentation for Automation and Process Control
Petrochemical Applications
• Co-generation
• Light oils
• Petroleum products
• Steam
• Hydrocarbon vapours
• Flare lines, stacks
Natural Gas
• Gas leak detection
• Compressor efficiency
• Fuel gas systems
• Bi-directional flows
• Mainline measurement
• Distribution lines measurement
• Jacket water systems
• Station yard piping
Power Industry
• Feed water
• Circulating water
• High pressure heaters
• Fuel oil
• Stacks
• Auxiliary steam lines
• Cooling tower measurement
• Low pressure heaters
• Reheat lines
• Combustion air
Emissions Monitoring
• Chemical incinerators
• Trash incinerators
• Refineries
• Stacks and rectangular ducts
• Flare lines