UV:SPECTROPHOTOMETERS

TRUE SYMMETRICAL DOUBLE BEAM

POWERFUL INTEGRAL SOFTWARE

FULL REFERENCE BEAM SPACE

DISPLAY SCROLLING OF SCANS

CHOICE OF OPTICAL BANDWIDTH

SUPERB VERSATILITY

ACCURACY : FLEXIBILITY : FAST-OPERA

FULL VALIDATION PROCEDURES

WIDEST ACCESSORY RANGE


Variable or Fixed Bandwidth Powerful Integral Control
The Aquarius spectrophotometers offer very high Rapid operation with ease of use is provided by the
performance with fixed optical bandwidth of less powerful software built into the computing system of
than 2nm pharmacopoeia compliant, 1nm, or the Aquarius instruments. The standard software is
variable optical bandwidth of 0.5, 1, 2 and 4nm for remarkably comprehensive but may be expanded, at
high resolution applications. any time, by E-SEF to release additional software
modules when required.
The spectrophotometer's performance and powerful
software are only normally available with much Rapid Menu Operation
more expensive instruments.
The large backlit high resolution LCD screen displays
Double Beam Optical System menus for all applications, with clear prompts for all
entries and procedures. Scans, plots, curves and data
The fully symmetrical double beam system of the are all displayed on the screen and may be reprocessed
Aquarius instruments provides the highest accuracy or manipulated before storing or producing a printout.
of measurement coupled with very high stability
with time. Secure Integral Memory
Generous built in secure memory allows scans, curves,
Both sample and reference beams are provided with plots, kinetic curves, as well as methods to be stored
the same space facilities, thus catering for the tamper proof protected by personal security codes.
widest possible range of experimental requirements
and not limited to just a simple cell in the reference Any scans, plots or data may be recalled from memory,
beam. with assured integrity, for reprocessing, manipulation
and printing as required.
Scans Viewed By Scrolling
By scrolling information across the display screen,
Dynamic Scan Recall
the effective width of the screen is a massive The Aquarius instruments are provided with the ability
430mm. to store the last 25 or more scans or plots without the
need to carry out a scan storage procedure. The scans
Long scans, plots and other data may be viewed in or plots are automatically stored as they are made and
much more detail, than with other instruments. may be recalled from a list on the display.
2
TRUE SYMMETRICAL DOUBLE BEAM

POWERFUL INTEGRAL SOFTWARE


SOFTWARE

FULL REFERENCE BEAM SPACE


Software Enhancement By E-SEF Benefits
This unique design feature - Encoded Software Enhancement Facility Variable or Fixed Bandwidth
- provided by Cecil Instruments, enables any required additional
DISPLAY SCROLLING OF SCANS
software module to be released by an encoded entry. The encoded
entry required is obtained by telephone, fax or E-mail, enabling you
Large Backlit LCD Screen

to expand your library of methods, without delay, when a new Screen Scrolling
application arises within your laboratory.
CHOICE OF OPTICAL BANDWIDTH Scrolled Screen Width
Method Operation 430mm

SUPERB VERSATILITY
Any procedure may be stored as a method in the security code
protected integral memory of the Aquarius instruments. When a
Dynamic Scan Recall

method is recalled, all instrument parameters are automatically set, Full Reference Sample Space
without operator intervention, thus ensuring that the original
ACCURACY : FLEXIBILITY : FAST-OPERATION
conditions are exactly replicated. Method Storage

Scan Storage Spectral Storage


FULL VALIDATION PROCEDURES
Scans, plots and curves may all be stored, for future use, with Curve Storage
confidence in the integrity of the data. Integral memory is provided
for up to 100 scans, protected by the operators personal security Password Protected Storage
WIDEST ACCESSORY RANGE
codes.
Multi/Single Cell Kinetics
Reprocessing Operates in 6 Languages
Although scans may be in absorbance, transmittance or derivative
format, any scan may be transformed to an alternative format by the
very rapid reprocessing facility provided. Performance
Bandwidth < 2nm or 1nm
Fully Documented Scans
Variable 0.5, 1, 2, 4nm
As shown below all scans are fully documented on screen and in
hard copy. Stability ±0.0001A/H

Tables of all peaks and valleys are produced together with Scan rate 1 to 4000nm/min
instrument, sample and operator details. Scans recalled from the
instruments integral memory or by dynamic scan recall are also fully Wavelength Precision 0.1nm
documented.
Straylight <0.02%
The screens shown are not two separate displays but part of a
continuous display viewed by scrolling. Validation Software

3
POWERFUL INTEGRAL SOFTWARE

FULL REFERENCE BEAM SPACE


SPACE

Sampling SCROLLING
DISPLAY Flexibility OF SCANS
The design of the double beam optical sampling allows
almost identical space and facilities in both the sample and
CHOICE OF OPTICAL BANDWIDTH
reference beams. Multi-cell reference sampling, with
temperature control is therefore possible.

Operation is not limited to a single standard cell in the


SUPERB VERSATILITY
reference beam as in some other instruments.

Reflectance and Gel Scanning


Accessories include fixed angle and variable angle
ACCURACY : FLEXIBILITY
reflectance units and a gel scanning system for:gels
FAST-O
up to
100mm long.

DataStream
FULL VALIDATION PROCEDURES
Data may be transferred from the spectrophotometer to a
PC, or the software used to control the instrument from a
Multi-User Convenience WIDEST ACCESSORY RANGE
PC. Datastream software, developed by Cecil Instruments,
has a wide range of facilities including graphical and
Aquarius double beam instruments were designed
tabular presentation, normalisation of overlayed traces,
for the greatest convenience in a laboratory where
point and peak pick using a cursor, library facilities, data
multi user operation is required.
storage in files on disk, history logging and copy and paste
All methods, stored scans, plots, curves and data to other applications such as Excel, Word etc.
remain in secure integral memory, tamper
protected by passwords, readily accessible to all
operators without the possibility of mislaid disks The photograph below shows an Aquarius instrument in
or plug in card modules. use in Cecil Instruments applications laboratory.

4
FULL REFERENCE BEAM SPACE

DISPLAY
DISPLAY SCROLLING OF SCANS

CHOICE
Curve and LineOF OPTICAL BANDWIDTH
Fitting
A straight line, quadratic or cubic curve may automatically be fitted to a
set of up to 30 standards. Fits may be made with an intercept or forced
SUPERB VERSATILITY
through zero. Replicate standards may be used for any given
concentration. All entries are prompted for simplicity of operation. A
quadratic curve fit is shown here.

ACCURACY : FLEXIBILITY : FAST-OPERATION


Editing Suspect Standards and Curve Storage STANDARD SOFTWARE
FULL VALIDATION PROCEDURES
Suspect standards Wavelength Scan
may be observed
by inspection of the
screen display and
WIDEST ACCESSORY RANGE either deleted or
Peak and Valley Tabulation
substituted by a
replicate standard. Derivatives 1st-4th
Up to 30 curves
may be stored in Time/Rate Plot
integral memory,
security code
protected for future Dynamic Scan Recall
use.

Auto Sequential Scans

Derivative Spectra Scan Storage

Derivative spectroscopy is a powerful technique for revealing hidden


information or enhancing fine detail in a spectral scan. Bands masked Scan Overlay
by other overlapping bands or distorted by a sloping background may
be accurately located for quantification.
Scan Reprocessing
Derivative plots may be displayed on their own or superimposed on a
wavelength scan, with or without an offset. The degree of smoothing
used is operator entered and ranging is normally selected to be Method Storage
automatic.

Curve/Line Fit

Curve Storage

Editing of Standards

Absorbance Ratio

Absorbance Difference

Self Test and Calibration

Measurements in A, C, %T
5
DISPLAY SCROLLING OF SCANS

CHOICE OF OPTICAL BANDWIDTH

SUPERB VERSATILITY
Optical Bandwidth
The Aquarius range provides the choice of fixed
bandwidth of less than 2nm, 1nm, or variable
ACCURACY : FLEXIBILITY : FAST-OPE
bandwidth of 0.5, 1, 2 and 4nm. The variable
resolution performance of the CE 7500 is ideal for
samples with sharper features requiring high
FULL VALIDATION PROCEDURES
resolution - particularly vapours and some liquid
samples.

The performance of the Aquarius CE 7500 is


WIDEST ACCESSORY RANGE
illustrated in the plots shown here of Benzene
Vapour at two different bandwidths, 0.5nm and
2nm overlayed.

Also shown here is the spectrum of iodine vapour


near 560nm.

The broad flat bands resulting from a scan at 2nm


bandwidth are well resolved when scanned at
0.5nm bandwidth.

A 1 nm optical bandwith is provided by the CE 7400S


version of the Aquarius, for applications requiring
this particular performance.

Micro-Sipette Sampling
Samples of 300µl or less, and multiple samples, may be
measured with very low cross contamination.

Samples are measured in the same cell enhancing


accuracy and speed of measurement. The pump, in the
sample compartment, is controlled by the instrument's
microprocessor.

Ultra-Micro Sampling
Valuable samples, such as biological samples prepared
with costly reagents, may be measured using a special
ultra-microcell and holder. With the system, samples of
50µl or even less may be accurately measured.

Tablet Dissolution
The CE 7460 and CE 7560 versions of Aquarius are fitted
with a special cell compartment and pipe port for tablet
dissolution assays.

The instrument may be used with either a six or eight


cell automatic changer.

6
CHOICE OF OPTICAL BANDWIDTH

SUPERB VERSATILITY
VERSATILITY

ACCURACY : FLEXIBILITY : FAST-OPERATION


Reaction Kinetics
Both single cell and multi cell Kinetics are available with reaction plots to screen and printer. Regression analysis
FULL VALIDATION PROCEDURES
is applied to the user-selected portion of reaction curves and results are reported in Units/Litre. Data may be
reprocessed as required and curves stored. Sample temperature control is available for single cells and four and
six cell changers.

WIDEST ACCESSORY RANGE


Advanced Kinetics
An example of advanced kinetics is shown below where time based reaction data generated by the
spectrophotometer has been transferred into Grafit, fitted to a curve and transformed into a more suitable format
for mathematical interpolation and calculation. In this particular example, the basic data plot can yield the limiting
rate (Vmax), the Michaelis constant (Km) and the initial linear portion (Vmax/Km).

This has been transformed into the Eadie-Hofstee format and plotted to give better precision in the calculation of
these important factors and constants.

Other transformations are available such as Lineweaver-Burk and the Hill plot for allosteric enzyme systems.

PC Data Manipulation

Data may be transferred to a PC for use with spread


sheets and other mathematical software packages.

More advanced reaction kinetics data manipulation


may be readily performed with Grafit software. Linear,
non-linear or polynomial curve fitting is easily
accomplished together with non-linear regression
analysis. A wide selection of standard equations are
provided, together with an equation editor for
customised equation fitting.

Data may also be transformed into other mathematical


formats for plotting and calculation.

Band Quantification
Accurate quantitative assays using band area or height,
corrected for background absorption, are exceptionally
easy to carry out with the easy to use Aquarius
instruments.

Analytical results are displayed and printed with


construction lines enabling the quality of the
parameters to be assessed.

The procedure may be stored as a method with security


code protection in the integral memory of the
instrument.

7
SUPERB VERSATILITY

ACCURACY : FLEXIBILITY

FULL VALIDATION PROCED


SCANNING QUANTITATION

WIDEST ACCESSORY RAN


Double Beam Wavelength Scans Powerful Quantification
Scans at up to 4000nm/min are presented on screen or Many quantitative methods are available including band
printed with all peaks annotated with wavelength and area computation with background correction, differencing
absorbance. Tables are also presented listing all peaks and and stripping of spectra, wine and colour assays,
troughs with values. mathematical calculations between several wavelengths,
protein and DNA assays.

Display Scrolling Curve and Line Fitting


Use of the display scrolling facility enhances the effective A straight line, quadratic or cubic concentration curve may
screen size and enables spectra and data up to 430mm wide be fitted to up to 30 standards, with or without an intercept at
to be viewed. the origin. Standards may be deleted and replaced and a
new curve fitted.

Dynamic Scan Recall Reaction Kinetics


The last 25 or more scans or plots are automatically retained Single cell kinetic measurements are possible with plots and
for recall by entering their scan sequence number and may data displayed on screen. Data may be reprocessed,
be reprocessed, overlaid or transferred to protected memory regression analysis applied and curves stored, security code
for future recall. protected against tamper or deletion.

Overlaid Spectral Scans Multi-Cell Reaction Kinetics


Up to eight scans, including stored scans, may be overlaid, Up to four or six cells are automatically measured with all
with or without offsets for convenient comparisons or reaction curves displayed on screen. Each curve may be
further manipulation. selected for processing with the results of regression
analysis printed.

Difference Spectra Multi Wavelength Assays


New wavelength scans or scans from store may be Assays are provided using either two or three wavelengths.
presented on screen and the difference spectra produced. A wine colour assay is provided which may use up to ten
The scans may be normalised enabling spectral stripping to different wavelengths.
reveal impurities or additives.

Derivative Spectra DNA and Protein Assays


Derivative spectra, up to fourth are readily produced, with DNA and Protein assays may be made using either the 2 or 3
selectable smoothing, either alone or superimposed on the wavelength methods of Warburg and Christian. Also
fundamental, with or without an offset. Scaling is automatic programmed are techniques and analytical procedures for
with all peaks and troughs quantified. assays by Lowry, Bradford, Biuret and BCA methods.

Scan Storage Band Area and Height Computation


Up to 100 spectra may be stored in safe memory for future Band areas with correction for sloping background
recall. Each spectrum is automatically allocated a recall absorption are readily quantified, with all details and
number, when stored, and may be security code protected construction displayed. Band areas may also be computed
against tamper or deletion. for derivative spectra to facilitate difficult quantification.

Programmed Scans Cell Program


Automatic scans, single or repeat may be made of a cell Using the various cell changers, measurements may be
program of up to 4, 6 or 8 cells. A programmed sequence of made on up to 4, 6 or 8 cells at a single wavelength with
scans may also be made of a single sample. cycle time and number of cycles entered by the operator. A
cell and wavelength program may be combined.

Time Course Plots Wavelength Program


Changing samples or flows may be examined and displayed Measurements may be programmed for 10 different
by making a time course plot. Data may be reprocessed and operator selected wavelengths with the cycle time and
stored, security code protected, with an automatically number of cycles selected by the operator. A wavelength
assigned recall number. and cell program may be combined.

8
:F
FAST
AST-OPERA
-OPERATION
TION

EDURES
SYSTEM FUNCTIONS SAMPLING

NGE
Integral Control System Variable or Fixed Bandwidth
All control software is within the instrument, tamper proof, Optical bandwidth selectable down to 0.5nm is provided by
providing easy menu driven operation with massive safe the CE7500 and allows the fine structure of narrow bands to
storage for scans, plots, curves etc. PC control and data be resolved which remain unresolved at just under 2nm
transfer to a PC are provided. bandwidth in instruments such as the CE7400.

Equal Sample and Reference Facilities Double Beam Optical System


Identical and generous sample and reference sampling The fully symmetrical optical system measures sample and
facilities widens the scope of measurement, for example, to reference simultaneously for ultimate accuracy, precision
enable multi-cell sampling in both beams and accommodate and baseline flatness. Long term stability is up to ten times
bulky accessories in the reference beam. better than single beam systems.

Method Operation and Storage Automatic Cell Changers


Up to 30 sets of instrument parameters, including Automatic changers are available for 4, 6 or 8 cells,
concentration curves may be stored, security code protected. providing rapid sample changing or the facility for full
When recalled, instrument parameters are automatically set, programming.
avoiding errors.

Multi-Language Operation Sipette Sampling


The operator may select from the six operating languages Samples down to 300 microlitres may be measured in a
provided. Full Alpha Numeric entry is provided so that 10mm cell. Sampling is under control of the instrument’s
results may be fully documented for GLP. microprocessor system. Samples may be saved or pumped
to waste.
Formula Entry and Computation Ultra-Micro Sampling
Mathematical formulae may be entered enabling complex Samples of 50µl and less may be accurately measured
calculations involving many wavelengths and cells to be assisted by the very small beam size of only 7 x 1mm. All
carried out automatically. normal cells up to 100mm pathlength are accommodated.

Software Enhancement By E-SEF Sample Temperature Control


E-SEF - Encoded Software Enhancement Facility - allows Water and thermoelectric temperature control are available
software modules to be enabled using encoded numbers for single cells and automatic cell changers. Temperature is
provided by telephone, fax or E-mail. Software set from the control panel and displayed on the screen in the
requirements may be tailored, at any time, as required. latter case.

Data Storage and Protection Dissolution Testing


Data, including 100 spectra, concentration curves, kinetics Versions of the instruments fitted with a special cell
and time course plots may be stored in safe memory, compartment with piping port, are available for dissolution
security code protected against tamper or deletion using a testing using Cecil TD software and a PC. Six and eight cell
four digit password. changers are available.

Reprocessing Batch Sampling


Scans, time course plots, reaction curves and gel scans may Automatic batch sampling for up to 100 samples is available.
all be reprocessed to any required format on the display The pump fits within the sample compartment and volume
screen and then plotted. etc is set from the control panel and displayed on screen.
Batch number entry is provided.

Performance Validation Specular Reflectance


Software and certified standards enable wavelength Special accessories enable specular reflectance
accuracy, absorbance accuracy, bandwidth, stray light etc. to measurements to be made. Both fixed angle and variable
be validated. A timed and dated report includes the serial angle reflectance accessories are available. The latter is
number of the instrument. particularly useful for film thickness measurements.

9
ACCURACY : FLEXIBILITY : FAST-OP

FULL VALIDA
VALIDATION
TION PROCEDURES

Physical Sciences Applications


WIDEST ACCESSORY RANGE
Validation Using Filters
The near infrared region to 1100nm is covered by both Absorbance accuracy is most conveniently measured
instruments maintaining excellent performance. using certified filter sets calibrated by Cecil
Measurements beyond 900nm can be of value in Instruments against NPL certified standards.
physical science applications such as the Wavelength calibration is also carried out using
measurement of optical filters, heat absorbing filters
and semi-conductors. certified filters.

The transmission spectra of two different filters is


shown here overlayed, demonstrating their
characteristics beyond 900nm.
Straylight
Straylight is measured following the Pharmacopoeia
(EP/BP) test, using a 1.2% w/v solution of Potassium
Chloride in water measured against water in the
reference beam.
The more widely accepted ASTM tests at 220nm and
340nm are also available.

Instrument Validation
Assessment of instrument performance is readily
carried out using the validation software module built
into all Aquarius instruments and released by E-SEF,
‘Encoded software enhancement facility’.

Laboratories subject to regulatory control, or following


G.L.P. are provided with a range of Pharmacopoeia
(EP/BP/USP) tests.

As well as these tests, tests using filters traceable to


national standards with certificates of calibration
issued by Cecil Instruments are available.

Pharmacopoeia and other tests are shown here.

Optical Bandwidth Absorbance Accuracy

The optical bandwidth of the instrument may be The Pharmacopoeia (EP/BP) test for absorbance
validated by measuring the half height width of the accuracy, using a 60mg/L solution of Potassium
narrow deuterium emission line at 656.1nm. The Dichromate in a 10mm pathlength cuvette, is shown
measurement is carried out automatically. here.

10
FULL VALIDATION PROCEDURES

WIDEST ACCESSORY
ACCESSORY RANGE

Cell and Wavelength Programming


All Aquarius instruments are fully programmable. Cell programmes
may use four, six or eight cell changers, with a single autozero or
individual zeros for each cell.
Wavelength programmes are provided for up to ten different
wavelengths and cell and wavelength programmes may be
combined.

Sampling and Temperature Control

A complete range of sampling accessories is available. A selection


of auto-sample changers, some with water temperature control, is
shown on the left. Thermoelectric single, four, four plus four and
six cell changers are shown above. Control is through the
instrument microprocessor system with screen display of
temperature.

ORDERING ACCESSORIES
Cell Holders Temperature Control - Water Circulation
Single cell holders - sample or reference 505 26 00 Single or micro sipette cell holder 503 36 00
Micro sipette 10mm cell holder 503 26 00 4 cell holder 202 36 16
Adjustable holder for micro cell 6600 57 00 2 x 4 cell holder, sample or reference 202 36 15
50µl ultra microcell holder 8020 39 00 6 cell holder 5500 36 15
Holder for rectangular cells up to 100mm 595 27 00 Temperature Control - Thermoelectric
Holder for cylindrical cells up to 100mm 595 26 00 Thermoelectric Control System 7400 31 00
Cell Changers Holder for single or sipette cell 8020 56 00
Auto changer for 4 or 4 + 4 cells, requires holder 9070 31 00 4 cell holder 9070 57 00
Holder for 4 x 10mm cells sample or reference 594 54 00 2 x 4 cell holder 9070 58 00
Auto changer for 6 cells, requires holder 9070 34 00 6 cell holder 9070 59 00
Holder for 6 x 10mm cells 5500 33 00 Sample Stirring
Holder for 6 x 20mm cells 5500 34 00 Electronic stirrer and sample cell holder 7200 31 00
Automatic 8 cell changer with holder 9070 32 00 Electronic stirrer and sample and reference holder 7200 32 00
4 position microcell holder 505 38 00 Gel Scanning
Sipette and Batch Sampling Gel scanner - requires trough or film holder 9070 39 00
Return sipette system, requires holder and cell 9070 21 00 Gel trough - 5mm x 100mm silica 570 07 02
10mm sipette cell 202 07 02 Printers and Cables
20mm sipette cell 202 07 51 RS232 cable for 25 pin PC with protocol manual 2021 26 00
40mm sipette cell 202 07 54 RS232 cable for 9 pin PC with protocol manual 2021 83 00
Sipette 10mm cell holder 503 26 00 Laser Printer 8000 73 01
Batch sampler for 40 samples with pump 2020 82 00 Dot matrix printer 8000 70 01
Batch sampler for 100 samples with pump 2020 82 02 Colour ink jet printer 8000 72 01
Reflectance Calibration Filters
Fixed angle reflectance accessory - pair CE 575 Holmium filter in holder 202 01 44
Variable angle reflectance accessory CE 574 Didymium filter in holder 303 01 30
Lamps Set of 2 certified Wavelength filters in holders 303 40 00
Deuterium lamp with hours indicator 2202 01 42 Set of 4 certified Absorbance filters in holders 594 44 00
Tungsten halide lamp - in pairs 2303 01 40 Set of 6 certified Absorbance filters in holders 594 66 00
11
ORDERING
ORDERING
CE 7400 Spectrophotometer
< 2nm pharma-compliant Optical
bandpass. Wavelength range 190-1100nm.
SPECIFICATIONS
SPECIFICATION
Optical System Fully symmetrical double beam system CE 7400S Spectrophotometer
1 nm Optical bandpass.
All mirrors coated for long life SPECIFICATIONS
Wavelength range 190-1100nm.
Monochromator Modified Czerny-Turner design using 1200
CE 7500 Spectrophotometer
L/mm holographic diffraction grating 0.5, 1, 2 and 4nm Optical bandpass.
Detectors Two silicon diode detectors Wavelength range 190-1100nm.
Self Test and Calibration Automatic at switch on CE 7200 Spectrophotometer
Wavelength Range and Accuracy 190 - 1100nm; 0.5nm <2 nm Optical bandpass.
Wavelength range 190-900nm.
Wavelength Reproducibility 0.1nm
Wavelength Readout Digital display on screen CE 7460/CE 7560 Dissolution Spectros
Wavelength Selection By GOTO Key and keypad entry Suitable for tablet dissolution monitoring
using a choice of dissolution software and
Optical Bandwidth < 2nm CE7400; 0.5, 1, 2, 4nm CE7500 an automatic six or eight cell changer.
Stray Light Less than 0.02% at 220nm, NaI; and 0.02%
8000 70 01 Dot Matrix Printer and Cable
at 340nm, NaNO2
Display Screen Large backlit LCD displays spectra, menus, 8000 72 01 Colour Ink Jet Printer and Cable
data, prompts and instrument status. 8000 73 01 Laser Printer and Cable
Display Scrolling Provides viewing width of 430mm
Derivative Spectra 1st to 4th derivatives SOFTWARE MODULES
Autozero Automatic by press button
Program C
Photometric Readout Digital display of A, %T and C on screen Cell program for up to 4, 6 or 8 cells
Photometric Accuracy ±.004A at 1A
Program W
Photometric Noise Less than ±0.0001A at 550nm, 2nm b/width Wavelength program for up to 10
Baseline Stored in non-volatile memory wavelengths and time interval
measurements at a fixed wavelength
Baseline Flatness ±0.001A (300nm/min, 200-1100nm) 2nm
bandwidth Quant M
Baseline Stability Better than ±0.0001A/hour Multi-wavelength assays at 2 or 3
wavelengths, Warburg and Christian
Computer Interface Bi-directional RS 232 c assays at 2 or 3 wavelengths, wine/colour
Scan Speeds 1-4000nm/min assay at up to 10 wavelengths
Wavelength Scale Expansions 1-100nm/cm Quant P
Method Storage 30 methods may be stored and retrieved Protein assays using Lowry, Bradford,
Method Protection Each protected by 4 digit passcode Biuret, BCA and direct UV
Reaction Kinetics With plots, reprocessing of data as Quant S
required, for single cell or up to 6 cells Quantification of corrected bands,
difference spectra, spectral stripping
Overlaid Scans Up to 8 scans may be overlaid
Cell Programming Up to 4, 6 or 8 cells Kinetics
Programme Cycles From 1 to 9999 Kinetic measurements with plots and
reprocessing of data
Programme Delay (cycle time) From 0 seconds to 4 days (1 sec resolution)
Wavelength Programming Up to 10 wavelengths. May be combined Kinetics M
Multi-cell Kinetics linked with
with a cell programme autochanging of cells and plotting of
Dynamic Scan Recall Last 25 scans automatically stored on board reaction curves
Storage of Spectra 100 spectra may be stored code protected
Formulae
Concentration Curve Fitting Linear, quadratic or cubic with editing of Construction and entry of mathematical
standards. 30 curves may be stored formulae with multiple wavelengths and
cells; automatic computation of results
Multi-language Operation Six languages provided
Formula Entry and Computation Complex formulae may be computed Validation
Batch Sampling Up to 100 samples under internal control Validation of absorbance, wavelength,
bandwidth, stray light etc. using liquid
Date and Time Information Printed from real time clock samples or certified standards
Power Requirements 110-250V, 50/60Hz, 170 watts.
DataStream
Dimensions and Weight 635 x 410 x 210mm; 33kg Data transfer to, or control by a PC.
Graphical presentation, transfer of data to
Cecil Instruments policy is one of continuous development. We therefore reserve the right to change specification without notice. Excel, Word etc.

CECIL INSTRUMENTS LIMITED MILTON TECHNICAL CENTRE CAMBRIDGE CB24 6AZ ENGLAND ISO 9001
TEL: 01223 420821 FAX: 01223 420475 E-MAIL: [email protected] WEB SITE: www.cecilinstruments.com CERTIFIED COMPANY

12