Philips - 3T - Protocol DKI

The document describes MRI scan parameters for a diffusion tensor imaging (DTI) sequence with the following properties: - Dual-coil head imaging with 32 channel coils and a field of view (FOV) of 240x240x120mm - Voxel size of 3x3x3mm, reconstructed to 3x3x3mm - Single-shot echo planar imaging (EPI) acquisition with 4 b-factors including 0, 1000, 1500, and 2000 - Other parameters include a repetition time (TR) of 4577ms, echo time (TE) of 80ms, and sensitivity encoding (SENSE) factor of 2

Uploaded by

jerry22445550
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
130 views3 pages

Philips - 3T - Protocol DKI

The document describes MRI scan parameters for a diffusion tensor imaging (DTI) sequence with the following properties: - Dual-coil head imaging with 32 channel coils and a field of view (FOV) of 240x240x120mm - Voxel size of 3x3x3mm, reconstructed to 3x3x3mm - Single-shot echo planar imaging (EPI) acquisition with 4 b-factors including 0, 1000, 1500, and 2000 - Other parameters include a repetition time (TR) of 4577ms, echo time (TE) of 80ms, and sensitivity encoding (SENSE) factor of 2

Uploaded by

jerry22445550
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Multi-transmit = "no";

Coil selection 1 = "SENSE-Head-32P";


element selection = "selection 1";
Coil selection 2 = "SENSE-Head-32AH";
element selection = "selection 1";
Dual coil = "yes";
CLEAR = "yes";
body tuned = "no";
FOV RL (mm) = 240;
AP (mm) = 240;
FH (mm) = 120;
Voxel size RL (mm) = 3;
AP (mm) = 3;
Slice thickness (mm) = 3;
Recon voxel size (mm) = 3;
RFOV (%) = 100;
Small FOV imaging = "no";
Fold-over suppression = "no";
Matrix scan = 80;
reconstruction = 80;
Scan percentage (%) = 100;
SENSE = "yes";
P reduction (AP) = 2;
P os factor = 1;
k-t BLAST = "no";
Stacks = 1;
type = "parallel";
slices = 40;
slice thickness (mm) = 3;
slice gap = "user defined";
gap (mm) = 0;
slice orientation = "transverse";
fold-over direction = "AP";
fat shift direction = "P";
Stack Offc. AP (P=+mm) = 1.75740623;
RL (L=+mm) = -0.521792829;
FH (H=+mm) = -6.37945843;
Ang. AP (deg) = 3.93005323;
RL (deg) = -16.8998699;
FH (deg) = 3.35096574;
Minimum number of packages = 1;
Slice scan order = "default";
Large table movement = "no";
PlanAlign = "no";
REST slabs = 0;
Interactive positioning = "no";
Patient position = "head first";
orientation = "supine";
Scan type = "Imaging";
Scan mode = "MS";
technique = "SE";
Modified SE = "no";
Acquisition mode = "cartesian";
Fast Imaging mode = "EPI";
shot mode = "single-shot";
Echoes = 1;
partial echo = "no";
TE = "user defined";
(ms) = 80;
Flip angle (deg) = 90;
TR = "shortest";
Halfscan = "yes";
factor = 0.600000024;
Water-fat shift = "minimum";
Shim = "auto";
Fat suppression = "SPIR";
strength = "strong";
frequency offset = "default";
Water suppression = "no";
BB pulse = "no";
MTC = "no";
Research prepulse = "no";
Diffusion mode = "DTI";
sequence = "SE";
gradient duration = "minimum";
gradient overplus = "yes";
directional resolution = "high";
nr of b-factors = 4;
b-factor order = "user defined";
b-factors = 0, 1000, 1500,
2000, (12) 0;
average high b = "no";
SAR mode = "high";
B1 mode = "default";
PNS mode = "moderate";
Gradient mode = "default";
SofTone mode = "no";
Cardiac synchronization = "no";
Respiratory compensation = "no";
Navigator respiratory comp = "no";
Flow compensation = "no";
Temporal slice spacing = "default";
NSA = 1;
Manual start = "no";
Dynamic study = "no";
dyn stabilization = "yes";
Arterial Spin labeling = "no";
Preparation phases = "auto";
Interactive F0 = "no";
B1 field map = "no";
MIP/MPR = "no";
Images = " M", (3) " no";
Autoview image = " M";
Calculated images = (4) " no";
Reference tissue = "White matter";
Preset window contrast = "soft";
Reconstruction mode = "immediate";
Save raw data = "no";
Hardcopy protocol = "no";
Ringing filtering = "default";
Geometry correction = "default";
IF_info_seperator = 1634755923;
Total scan duration = "10:48.1";
Rel. signal level (%) = 100;
Act. TR (ms) = "4577";
Act. TE (ms) = "80";
ACQ matrix M x P = "80 x 78";
ACQ voxel MPS (mm) = "3.00 / 3.07 / 3.00";
REC voxel MPS (mm) = "3.00 / 3.00 / 3.00";
Scan percentage (%) = 97.727272;
Packages = 1;
Min. slice gap (mm) = 0;
Diffusion gradient timing DELTA / delta (ms) =
"51.1 / 18.0";
EPI factor = 43;
WFS (pix) / BW (Hz) = "11.334 / 38.3";
BW in EPI freq. dir. (Hz) = "2668.1";
SAR / head = "< 33 %";
Whole body / level = "< 0.1 W/kg / normal";
B1 rms = "1.33 uT";
PNS / level = "65 % / normal";
Sound Pressure Level (dB) = 15.4141731;