800T Manual
800T Manual
Page 1
User’s Manual
tPrime Series
280T Ultrasonic Heat Meter
Created July 12, 2012
Version Man280T-12071403_v15
Page 2
tPrime 280T Ultrasonic Heat Meter - 1 -
WARNING!
(1) The 280T heat meter is not certified for use in hazardous environments. The
(2) The 280T heat meter contains Lithium batteries. Please check to see if they’re
working before using the heat meter. The batteries must be recycled or disposed of
properly.
Page 3
- 2 - tPrime 280T Ultrasonic Heat Meter
Contents
1. Introduction ..................................................................................................... 1
3. Operation ................................................................................................... 11
4. Installation ................................................................................................... 14
Page 4
tPrime 280T Ultrasonic Heat Meter - 3 -
5. Troubleshooting ............................................... 21
§5.1 Power-on Errors .............................................................................................................. 21
6. Interface/Communication ................................................................................................... 24
8. Appendix ................................................................................................... 36
§8.1 Battery Maintenance and Replacement ............................................................................ 36
Page 6
5 tPrime 280T Ultrasonic Heat Meter - 1 -
1. Introduction§1.1 Preface
A member of the tPrime Series, the 280T ultrasonic heat meter offers the most advanced BTU
measurement in the market by using state-of-the-art ultrasonic flow measurement technology. The
280T does not have moving parts to wear out, thus it literally requires no maintenance. It is also
very cost-effective, especially in the long run. This means both commercial and residential
installations can profit from the advantages of wear-free heat measurement, such as precision,
operating security, and long service life.
With its maximum operating temperature of 95°C (130°C version available upon request) and
nominal pressure of 1.6MPa, the technical specifications meet the standard for residential meters.
The high measurement dynamic allows a load of up to double the rating, thereby ensuring
high operating security.
The 280T heat meter uses transit-time ultrasonic flow measurement principle to reliably measure
the flow rate: sending and receiving transducers in opposition send ultrasonic signals alternately
with and against the flow direction. The flow rate can be precisely calculated from the difference
between the two transit times. There are no moving components in the sensor, and as a result, the
measurement is wear-free and stable over a very long time period.
Suitable for both commercial and residential applications, this compact meter fits in even the
smallest installation conditions and can be mounted separately from the electronics console. The
meter includes an order option for remote read-out (pulse, M-Bus, or RS485 output). The large
display can be set to display heat consumption, temperatures, flow total, working time, velocity, etc.
The 280T also offers tools for building energy management. When equipped with a M-Bus module,
the heat meters can be networked through a two-wire bus to a central location for integrated
resource management. An optional concentrator and data acquisition software make the whole
system installation and integration easy. Spire Metering provides a complete AMR (automatic meter
reading) solution as well.
Spire Metering’s 280T Ultrasonic BTU meter stands out among its competitors due to its unique
reflector-free sensor design and multipath technology. The 280T is able to work reliably even when
the water is dirty, which is useful because the HVAC circuit could be like this after years of
operation. Both commercial and residential installations can profit from the many advantages of
wear-free heating/cooling energy measurement, such as precision, operation security and long
service life.
§1.2 Features
• Wear-free ultrasonic measurement. No maintenance needed
• Proprietary robust sensor design. No reflector, thus works reliably for both clear water
Page 7 - 2 - tPrime 280T Ultrasonic Heat Meter
• Wider dynamic range allows for a load of up to double the rating, thereby ensuring
high operational security
• Maximum operating temperature of 95 °C (up to 130 °C optional)
• Nominal pressure of 1.6MPa
This compact meter fits into even the smallest installation locations and can be mounted separately
from the electronics console.
The large LCD screen can be set to display the heat consumption, temperature, flow total, working
time, flow velocity, etc. The meter also has a remote readout which can be configured as pulse,
M-Bus or RS485. An optional BACnet module is available upon request.
For more information on AMR and serial communication, see Section §6.
A time-keeper is integrated in the 280T heat meter. The time-keeper remains operating as long as
the battery is alive. In case of battery failure, the time-keeper will not keep running, and the time
data will be lost. The user must re-enter the proper time values after the battery failure is recovered.
The user can also set the date and time as desired.
Each set of the 280T series heat meter has a unique product identification number or ESN written
into the software that can only be modified with a special tool by the manufacturer. In case of any
hardware failure, please provide this number.
2. Technical Information
§2.1 Parts Identification
Integrator
Integrator Holder
Temperature Sensor
Mounting Port
Flow-cell Sensor
Temperature
Sensor Cable
Interface Cable
Temperature Sensor 2
Temperature Sensor 1
(Already Installed)
Gasket
Flow-cell Sensor
Removable Flange
Integrator
Temperature Sensor
Flow-cell Sensor
Interface Cable
Temperature Sensor
DN 15-40
DN 50+
§2.4 Specifications
(in.) (qpm) (qpm) (qpm) L (in.) W (in.) H (in.) (lb) Sensor Ext.
1/2” 6.6 0.1 13.2 4.33 4.72 3.38 1.54 G3/4B G1/2B
3/4” 11 0.2 22 5.11 4.72 3.38 1.76 G1B G3/4B
1” 15.4 0.3 30.8 6.29 5.03 3.50 1.98 G1 1/4B G1B
Electrical Data
Power Supply: Battery, 3.6V, Lithium
Battery type: Lithium, 3.6V
Battery life: 6 years. Optional 10 years
Backup Power Supply: Internal SuperCap
Communication Interface: M-Bus (default)
Optional: RS485 with MODBUS support, Pulse, optical
isolated, BACnet/MSTP, Radio, GSM
CE approval: EN61326-1:2006
Pressure: ≤1.6 MPa
Calculator (Integrator)
Display: LCD, 8 digits
Resolution: 999.99999 - 999999.99 - 99999999
Energy Unit: KWh – MWh – GJ
Communication Protocol: M-Bus (default). Optional: MODBUS, BACnet
Temperature Measurement
Sensor Type: PT1000, 2-wire
Measurement Range: 0…150°C C (32…302°F)
Difference Range: ΔΘ: 3K … 70K
Permissible Temperature: Θ: 2°C…60°C (35…140°F) for long term and up to 95°C
(203°F) for short term
High-temperature version: up to 130°C (266°F)
Mechanical Data
Metrological Class: 2 (according to OIML R75)
Environmental Class: B, optional C (outdoor)
Electromagnetic Class: E1
Environmental Temp: 0…55°C (32…131°F)
Pressure Loss
The pressure loss of a flow sensor is proportional to the square of the flow:
Δp = k x q2
Here Δp is pressure loss, q is volume flow rate and k is the coefficient.
All meters have Δp less than 0.25bar at qp.
Example:
280T-DN025-1-A-2-A stands for the 280T tPrime series BTU meter for pipe DN25mm with
M-Bus interface, BSP pipe joint, extension pieces and T-connector (thermal well), with
standard temperature rating.
3. Operation
§3.1 Built-in Battery
The instrument operates from the built-in Lithium battery, which is long-lasting with 6 to 10
years of operating time. The static current of the battery is less than 10uA, and the terminal
voltage reaches around 3.6V. If the 280T is connected to an outside power source, such as
M-Bus, the meter will not use the battery but instead draw power directly from the M-Bus.
Due to transport regulations, the battery might be deactivated by an insulating strip, which
must be removed completely in order to activate the meter. If a replacement battery is needed,
please contact Spire Metering. If the meter needs to be sent by air freight, then the battery
must be removed prior to shipping.
For safety precautions, the batteries should not be opened, come into contact with water, or be
exposed to temperatures above 80 °C. Batteries should be disposed of at proper collection
centers.
§3.2 Power On
First, make sure to activate the battery. Press and hold the red button until it turns on the
power.
The 280T screen should display E0 when it is not connected to a water pipe (There is no water
inside of the flow sensor). When the meter is connected to a water pipe, and the water is full
in the pipe, the device will display the quantity of heat. If there is any abnormality,
corresponding error messages will be displayed.
Generally, there should be no display of error messages, and the heat meter will go to the most
commonly used window, which displays the total energy in kW h.
The heat measurement program always operates in the background of the user interface. This
means that the heat measurement will keep running regardless of any user window browsing
or viewing.
When the red button on the front panel is pressed, it cycles through different display windows
such as supply water temperature (t1), return water temperature (t2), temperature difference
(∆t), working time (hr), accumulated water quantity, water velocity, etc.
Once it is turned on, the 280T will keep on running until its long-lasting battery runs out.
There is no need to turn off the device when not in use.
§3.3 Keypad
The keypad of the 280T heat meter has 1 key for all functions.
In order to get into a certain menu window, the user can press the red button to change the
LCD display. The heat meter calculates the results and updates the display every 8 seconds.
The following values can be displayed:
Each short press of the red button switches the display screen to the next item. The menus are
displayed in loop-fashion. After pressing the red button, the display may stay at an item for a
short time and then return to the original display item. If you want the meter to display heat
output all the time, first press the red button to switch to the heat output screen, then press and
hold the button for 8 seconds. The display resolution is 0.001 kWh. If you want the meter to
display water flow total all the time, first press the red button to switch to the flow total screen,
then press and hold the button for 8 seconds. The display resolution is 0.00001m3.
These are the top layer (A1-layer) of the menu display structure. On the second layer
(A2-layer) of this structure, more information can be seen. To go from the A1-layer to the
A2-layer, press the red button until it displays the flow rate. Then, press and hold the button
for 8 seconds, and you should see “A2” on the LCD for a short time. This indicates that you
are ready to view the menus on the A2-layer. On the A2-layer, the following menus are
displayed:
To go back to the A1-layer, scroll to the Flowrate item of A2, then press and hold the button
for 8 seconds until “A1” appears on the LCD. Or, simply leave the button alone for a certain
time, and the meter will automatically switch to the top layer display.
Note: Factory Data 4 and Factory Data 5 are the calibration coefficients for Q1 (normal
flowrate), Q2 (0.2 times normal flowrate), Q3 (0.1 times normal flowrate), and Q4
(minimum flowrate).
Example: If Menu Factory Data 4 shows 626C 63CE, then the coefficient for Q1
is626C (Hex) or 25196 (Dec), and the coefficient for Q2 is 63CE (Hex) or
25550 (Dec). Similarly, this applies to Factory Data 5.
If an ‘E0’ flashes, it means there is no water flowing through the heat meter. Check the pipe’s
water supply and try to get it so that the pipe is full of water.
If an ‘E1’ is displayed, it means that the flow rate is over the supported range of the
instrument.
4. Installation
§4.1 Location selection
• Find a suitable location for connecting the flow cell (the “tube” part of the device
where water flows through) to the pipe line.
• Do not install the meter within 0.5m of an AC power line or a high-frequency
radiation source.
• It is recommended to have a 5D straight pipe run upstream and 2D straight pipe
run downstream, where D stands for pipe diameter.
• When two or more heat meters are installed closely, make sure they are distanced
by 0.3m or more.
• If the flow meter is installed on the common return of two heating circuits (e.g.
heating water and hot water), the mounting location must be at a sufficient
distance, at least 10D, from the Tee, to ensure that different temperatures
homogenize (refer to the figure on the bottom).
• For more information on optimal location, go to Appendix §8.4.
25 °C
>10D
85 °C
•
Tee with Temperature Sensor Port
•
•
• Return Return
•
•
PT1000 Temperature Sensor (Blue)
•
Thermal •
Exchange •
System • Valve
• Filter (optional)
Supply •
Supply
•
•
•
•
PT1000 Temperature Sensor (Red)
Plumbing • Gasket
Adapters
•
• Flange connection
(1) Check the flow-cell dimension and reserve sufficient clearance for the
flow-cell installation.
(2) Install flexible flanges if applicable (DN50-100), see next section for details
(3) Install the flow-cell between two shut-off valves.
Sealant
Transducer
Ultrasonic heat meters from sizes DN50 to DN100 normally use flexible flanges for pipe
connection. Here shows how to assemble the flexible flanges with the flow-cell sensor
and install them into a pipe line.
1. Each water meter or heat meter of size DN50-DN100 comes with a flow cell, a pair of
flexible flanges, and two pieces of metal rings.
2. First, put the 2 pieces of flanges into the flow cell.
Flow-cell Flange
Transducer
3. Put the metal ring into the metal ring slot. Do this for both sides of the flow cell.
Metal Ring
4. Connect the flow cell to the pipe:
Put the pipe sealant in place metal. Rotate the flange so that the two transducers are in
the horizontal plane. Note that if your pipe is vertical, you may skip this step. Just make
sure the flow in your vertical pipe is going upward, not downward. Put the bolts through
the flange holes as shown below. Screw in the nuts with caution. Leave the nuts loose
for this moment! Check whether the flow cell is in line with the pipe. If not, adjust the
flow cell position. Tighten the bolts and nuts.
For DN 10-40
The red-labeled Temperature Sensor should be already
PT1000
pre-plugged into the flow cell. In order to install the blue-labeled Temperature
Temperature Sensor on the return pipe, follow these steps: Sensor
(1) Insert the Tee into the desired location along the pipe line.
(2) Remove the Temperature Sensor Port plug from the Tee.
O-Ring
(3) Insert the O-Ring deep inside the Temperature Sensor Port.
(4) Gently insert the blue-labeled Temperature Sensor into the
Tee’s Temperature Sensor Port, being careful not to break the
For DN 50-120
(1) Install the red-labeled temperature sensor into the temperature sensor mounting pocket on
the flow-cell. Refer to the lower left figure for details.
(2) On the return pipe (assuming the heat meter is installed on the supply pipe), make a hole
with size slightly bigger than the mounting pocket.
(3) Insert the mounting pocket into the hole and weld it onto the return pipe.
(4) Install the blue-labeled temperature sensor into the mounting pocket:
- For DN80 and larger meters, refer to the lower center figure.
- For DN50 and DN65 meters, refer to the lower right figure.
PT1000
Temperature
Sensor
Mounting
Pocket
(Supply) Thermal-well
sleeve
PT1000
Temperature
Please wrap
Sensor
Teflon to seal
Thermal-well
Sleeve
(For sizes ≥DN80
only) Mounting
Pocket
Temperature sensor
Installation on the flow-cell For sizes ≥ DN80 For sizes DN50 and DN65
The electronic box can be mounted horizontally, vertically, or inclined on the meter mounting
base of the flow sensor.
(1) Make sure the surrounding temperature is less than 55°C. If not, remove the meter
mounting base and install it together with the electronic box on a wall that is at room
temperature.
(2) If the water temperature is higher than 90°C, remove the meter mounting base and
install it together with the electronic box on a wall.
(3) When used as a cold meter, remove the meter mounting base and install it together with
the electronic box on a wall. In addition, the box should be above the pipe so that no
condensed water will drop onto the box or run along the wire into the box.
For horizontal pipes, the flow-cell should be installed in a way so that the meter box faces upward and the
two ultrasonic transducers on the horizontal plane.
Pipe is Full
Pipe may not be Full
must go upward.
Flow Flow
instructions.
• Make sure the O-ring for the sealing is centered at the joining point. Otherwise, it
could generate disturbance to the flow, thus, degrading the meter accuracy.
• Use meter only under the specified operating conditions. Make sure flow rate
range is proper for the pipe.
• If the heating system is not running in winter, always empty the pipe. Otherwise,
the pipe may burst when water freezes.
• Important: The pipe must be full of liquids during operation!
Page 26 tPrime 280T Ultrasonic Heat Meter - 21 -
5. Troubleshooting
§5.1 Power-on Errors
When powered on, the 280T series ultrasonic heat meter automatically starts the
self-diagnosis process to see if there are any hardware and software problems. If a problem is
identified, an error message will be displayed. The following table shows the possible error
messages, the corresponding causes, and their solutions.
Whenever something goes wrong while the device is on, the 280T heat meter will show an
Error Code, just like with the power-on errors. When any abnormal Error Code shows,
counter-measures should be taken.
6. Interface/Communication
§6.1 General
The tPrime 280T series ultrasonic heat meter integrates a USB communication interface and a
complete set of serial communication protocol. By using this serial communication link, one
can configure the heat meter and acquire measurement results from a PC. Additionally, the
280T heat meter has multiple AMR/AMI integral options, including:
•Pulse
•M-Bus
•RS485 with Modbus
•RF
•TCP/IP
•BACnet modules
To facilitate the use of and to fully explore the potential of 280T heat meter, Spire Metering
has developed proprietary PC software, which is particularly useful for those who are not
familiar with computer communication technology.
Please visit the following technical support website for more information on available PC
software:
http://www.spiremt.com/support
These are the main methods to read data from tPrime 280T heat meters: Infrared Meter
Reading, RS485 Meter Reading, M-Bus Meter Reading, BACnet Meter Reading and RF
Wireless Reading.
Infrared reading is a local meter reading method. The others are remote meter reading
methods. The communication protocol discussed here is applicable to remote reading
methods.
MODBUS Protocol:
Master-Slave communication architecture. The 280T meter is the slave.
Use RS485 serial communication. Default parameters are:
• Baud Rate: 9600 bps
• Checksum: None
• Data bit: 8 bits
Notes:
*Unit code:
05– KWH; 14 – Watt; 32 – Litre/H; 2C – Cubic Meter;
02– WH; 08 – GWH;
** Use factory software to change the communication protocol. If you set the Communication
Mode to 1, and set the duration to 6556, then the meter will switch to M-Bus protocol.
Resetting the external power will switch the mode back to MODBUS protocol.
M-Bus Protocol:
Master-Slave communication architecture. The 280T meter is the slave.
Use M-Bus serial communication. Default parameters are:
• Baud Rate: 2400 bps
• Checksum: None
• Data bit: 8 bits
• Stop bit: 1 bit
The 280T adapts the EN1434 / EN13757 M-Bus protocol. This allows it to communicate with
any standard M-Bus master device.
Please note that only the major functions of the protocol have been implemented.
Pulse Interface:
Vcc
Specifications:
Optocoupler
Vcc: 12~24 VDC power supply Isolation Red/Blue Rx
Rx: 2kOhm
Pulse width: 15ms
Pulse frequency: Pulse
Counter
1 kWh/p (≤DN32),
10kWh/p (DN40~DN80),
100kWh/p (DN100~DN150),
Black/Yellow
1000kWh/p (≥DN200)
M-Bus Interface:
RS485/MODBUS Interface:
Blue (Minus)
Black (Ground)
Yellow (B)
BACnet Interface
Yellow (B)
Interface: Upper Link: BACnet MSTP
Downlink: RS485/MODBUS
Power Supply: 5VDC/0.5A
The pulse system is the easiest way to connect the 280T meter to an output source for serial
communication. As shown in the wiring diagram in §6.5, be sure to include a black output
wire for the ground.
The pulse output interface can be connected to a pulse wireless transmitter to send the meter
data to a remote data center through a fixed wireless AMI infrastructure (refer to section $6.8
for details). It may also send data to a handheld meter reading device for mobile reading when
AMI infrastructure is not available. Please contact [email protected] for more
information.
M-Bus Cable
The M-Bus uses two wire cables which are going from the M-Bus Master / Repeater to each
M-Bus device (bus structure). The M-Bus is polarity independent and needs no line
termination resistors at the end of the cables.
Any cable type may be used as long as the cable is suitable for 42 V / 500 mA. Shielding is
not necessary and not recommended since the capacity of the cable should be minimized.
In most cases a standard telephone cable is used which is a twisted-pair wire with a diameter
of 0.8 mm each (2 x 0.8 mm). This type of cable should be used for the main wiring. For the
wiring to the meters from the main wiring (last 1 … 5 m to the meter) a cable with smaller
diameter may be used.
The M-Bus system is an European instrument “bus” standard designed for domestic
metering devices, such as water meters, heat/water meters, gas meters, etc., to communicate
with data centers. The “bus” simply uses two non-polarized wires to achieve a variety of
options for reliable meter reading, remote diagnosis, remote control, incremental pricing,
time-based pricing, batch service, prepaid billing, and more. This ‘bus’ system is both
simple and economical to wire and implement.
A typical M-Bus AMR system consists of a number of M-Bus utility meters, several M-Bus
concentrators, a GSM/GPRS Data Transmitter Unit (DTU) for each M-Bus concentrator, and
a data center. The M-Bus Concentrator communicates with the data center computer through a
GSM/GPRS network. The data center first issues a meter reading command and sends it to the
network. The DTU receives the command and forwards it to the M-Bus concentrator. Then,
the concentrator either replies to the command with requested data or passes the command to
its submeters transparently.
Please note that you may not need the DTU unit if you can connect the M-Bus
concentrator(s) to your computer directly. Alternatively, you may connect the
concentrator(s) to your computer through TCP-IP network by using Ethernet-232 adapters.
Similarly, you may connect the concentrator(s) to your BACnet or MODBUS network by
using proper adapters.
The 280C Concentrators are used for an AMR system to facilitate the communication between
the data center and the M-Bus utility meters of the AMR system. These concentrators support
not only the 280T BTU meters, but also the 280W water meters and STE101C electricity
energy meters. A wireless M-Bus concentrator is also available, where the M-Bus
concentrator is affixed with a GSM/GPRS data transmitter unit (DTU).
A DTU (Data Transmitter Unit) is a GSM/GPRS modem which provides a wireless interface
for the concentrator. Spire Metering’s EP280 GSM modem and EP260 GPRS modem are
specially designed for this application. Both EP280 and EP260 have the following features:
For remote locations where wired metering networks or RF networks are not suitable, Spire
Metering offers GSM-enabled utility meters. This allows the utility meter to communicate
with the data center directly.
One example is a 280T heat meter that has a built-in GSM modem (EP280). When the EP280
GSM modem is powered up, it starts to establish a wireless link with the GSM wireless
network. On the data center side, another GSM modem (EP260) is connected to the computer.
When this modem is powered up, it will also start to establish a wireless link with the GSM
wireless network.
The data center software sends a query command together with the destination’s phone
number and address to its modem, EP260. EP260 then packs the command in SMS format
and sends the SMS package to the GSM/GPRS cellular wireless network. The SMS service
routes this package all the way to the destination modem EP280, whose phone number
matches that of the command.
When the EP280 receives the SMS package, it strips off the SMS packing information and
extracts the original command. Consequently, it sends the command to the meter whose
address matches the one in the command. In this way, a number of utility meters, such as
water meters, electricity meters, gas meters and valves, can be connected to the same EP280
GSM modem, as long as they have different addresses.
When the heat meter receives the command from its EP280 modem, it executes the command
and responds with the requested data. The data will be sent to the GSM/GPRS cellular
wireless network via SMS service. The SMS service routes this data all the way back to the
data center computer. After receiving the response from the water meter, the Data Center
Manager software will process the data, saving the data to SQL database.
This fixed network Advanced Metering Infrastructure (AMI) system utilizes a Remote Data
Logger (RDLR) to receive and transfer meter data from transmitters located at each utility
meter on to the utility computer. The SpireCapture data management software suite allows
easy access to system information and a variety of customer service tools. In addition to
providing meter data to the utility’s billing system in a flexible and compatible file format, the
meter data management software also monitors and controls system performance, manages
data, and remotely reconfigures the RDLR as needed.
Repeaters
amplify
messages Wireless Utility Meter:
280W, 280T, STE 101
Data Center
Clubhouse
The data link layer options include BACnet IP (Annex J), MS/TP master (Clause 9), with a
baud rate of up to 76.8 Kbps, and MS/TP slave (Clause 9). The networking options are
BACnet/IP Broadcast Management Device (BBMD), as well as registrations by foreign
devices.
The standard object types supported by BACnet interface are device object, analog input,
analog output, analog value, binary input, binary output, binary value, multi state input, multi
state output, multi state value, and notification class object.
GPRS network
Wireless M-Bus Concentrator (cellular data network)
Gateway
Data Center
The products manufactured by Spire Metering Technology are warranted to be free from defects in
materials and workmanship for a period of two years from the date of shipment to the original
purchaser. Spire Metering’s obligation should be limited to restoring the meter to normal operating
conditions or replacing the meter, at Spire Metering’s choice, and shall be conditioned upon
receiving written notice of any alleged defect within 10 days after its discovery. Spire Metering will
determine if the return of the meter is necessary. If it is, the user should be responsible for the
one-way shipping fee from the customer to the manufacturer.
Spire Metering is not liable to any defects or damage attributable to misusage, improper installation,
out-of-spec operating conditions, replacement of unauthorized parts, and acts of nature.
Additionally, fuses and batteries are not part of this warranty.
§7.2 Service
The manufacturer provides instrument installation services for its customers, and the charge will
depend on the complexity of the installation.
For operational problems, please contact the technical support department by telephone, fax, or
email. In most cases, the problem can be resolved immediately.
For any hardware failure of the instrument, we recommend our customers to send back the
instrument for service. Please contact the technical support department with the model number and
serial number of the unit before sending the unit back to us. Both numbers can be found on the
product label. For each service or calibration request, we will issue a Return Materials
Authorization (RMA) number.
Take notice that the cost for repairing can only be determined after receipt and inspection of the
instrument. A quotation will be sent to the customer before proceeding with the service.
Before returning the instrument for warranty repair or service, please read the following carefully:
1. If the return item has been exposed to nuclear or other radioactive environments, or has been in
contact with hazardous material which could pose any danger to our personnel, the unit cannot be
serviced.
2. If the return item has been exposed to or in contact with dangerous materials, but has been
certified as hazard-free device by a recognized organization, you are required to supply the
certification for the service.
3. If the return item does not have a RMA# associated, it will be sent back without any service
conducted.
Page 41- 36 - tPrime 280T Ultrasonic Heat Meter
8. Appendix
§8.1 Battery Maintenance and Replacement
The battery is a Lithium non-rechargeable battery. Therefore, one needs to replace the batteries
once they run out of power (usually lasts 6 or 10 years). Please contact the manufacturer for
purchasing new batteries. Dead batteries should be disposed of at proper collection centers.
The tPrime series 280T heat meter consists of an ultrasonic flow sensor, a pair of PT1000
temperature sensors, and a BTU unit. The microprocessor-based BTU unit controls the ultrasonic
sensor to transmit and receive ultrasound in an orderly fashion in order to conduct precise flow
measurement. The BTU unit also has electronics dedicated to measure the temperature in the supply
pipe as well as the return pipe via PT1000 sensors. The BTU unit calculates the heat energy based
on the flowrate and the temperature difference between the supply and the return.
The figure on the right illustrates how the
ultrasonic flow sensor works. Two ultrasonic
transducers (A and B) are mounted on a
spool piece face-to-face; one is on the
upstream and the other on the downstream.
The electronic console (e.g., the BTU unit)
operates by alternately transmitting and
receiving a burst of sound energy between
the two transducers and measuring the transit
time it takes for sound to travel between the
two transducers. The difference in the transit
time measured corresponds directly to the
velocity of the liquid in the pipe
where:
Q is the quantity of heat given up or absorbed
V is the volume of liquid passed
k is called the heat coefficient, a function of the properties of the energy-conveying liquid at the
relevant temperatures and pressure
ΔΘ is the temperature difference between the flow and return of the heat exchange circuit
The true conventional value of the heat coefficient (k) for water, if it is used as the system’s heat
conveying liquid, can be obtained from a standard heat coefficient table.
Another method is to use the specific enthalpy table. Spire Metering Technology’s BTU meter
utilizes the second method to calculate the heat energy of a heat exchange circuit. The advantage of
this method is that the flow sensor can be installed in either the supply or return line, as long as the
Red-tagged temperature sensor is installed on the flow sensor side. The equation is below:
where:
Q is the total heat given out or absorbed, from time τ0 to τ1, [J] or [Wh]
ρ(Ts, Tr) is the density of the liquid passed[kg/m3]
∆h(Ts, Tr) is the specific enthalpy difference, calculated from the supply and the return line
temperatures, in [J/kg].
To calculate heat (Q) accurately, both flow and temperature measurements must be accurate. Of the
two, measuring flow rate is more challenging, as all of the values in the integral on the right side of
the equation below must be correctly measured, determined, or pre-calibrated.
L ∆T
V= × and qv = k S V
cos2 θ TT
up
• down
where:
ϴ is the angle between the flow direction and the sound path
L is the transducer separation
Tup (Tdown) is the transit time upstream (downstream), and ΔT = Tup – Tdown
k is a factory-calibrated scale factor, a function of the sensor structure and the fluid properties
S is the cross-sectional area of the flow at the sensor
Temperature
Sensor (blue)
Return
A2 When pressure is higher than 1.0MPa but less than or equal to 2.5MPa:
The first step in the installation process is to select an optimal location for installing the device in
order to make the measurement reliable and accurate. A basic knowledge about the piping and its
plumbing system would be advisable.
An optimal location would be defined as a long straight pipe line full of liquid that is to be
measured; it can be in a vertical or horizontal position. Principles to select an optimal location:
(1) The straight pipe should be long enough to eliminate irregular flow induced error. Typically,
the length of the straight pipe should be 5 times of the pipe diameter (5D) straight pipe run
upstream and 2D straight pipe run downstream, the longer the better.
(2) Make sure that the temperature of the mounting location does not exceed the range for the
heat meter. When the meter is used as cold meter, remove the meter’s main box and its
mounting base and install the main box on a wall or other objects where the temperature is
within 0~55°C. In addition, the meter box needs to be higher than the cold pipe, so that the
condensed water on the cold pipe will not flow into the meter box along the wires. If the
heating system is not running in winter, always empty the pipe. Otherwise, the pipe may burst
when water freezes.
(3) Select a relatively new straight pipe line if possible. Old pipe tends to have corrosions and