Example of Well Organized Report
Example of Well Organized Report
Example of Well Organized Report
Mechanical Measurements
Partner’s Name
Instructor’s Initials
Table of Contents
Procedure page 2
Data/Calculations page 4
i
Purpose
Background Theory
device has both static and stagnation holes, which are connected to either end of a
differential manometer.
Figure 1
In the figure 1, point a is the location of the stagnation hole, while point b if the
location of the static hole. The value h is the manometer fluid height.
1
Procedure
2) Next, the pitot tube was aligned in the center of the duct, and the fan was
3) Step two was then repeated using approximately half of the fan speed.
Equipment Description
Figure 1
Figure 1 shows a pitot tube arrangement that was very similar to the type used in the
experiment. The pitot tube was attached to a manometer. The pitot tube was housed in a
round duct, and was moved from one wall of the duct to the other. The position of the
pitot tube was recorded with a scale that was rigidly attached to the outside of the duct. A
fan on one end provided the air flow for this experiment.
2
Sample Calculations
u = (2*g*h*ρm/ρ).5 g = gravity
u=
(2*32.2*.0028(62.4*.862/.0715))^.5 h = manometer fluid height
u = 25.62 ft/s ρm = manometer fluid density
ρ = air density
3
Data/Calculations
1390 rpm 3
r h u-
r*u
No. Static Velocity r/R u/U
in ft in ft (ft2/sec)
(ft/sec)
0 0 0.000 0 0.000 1.36 0 0 0 0
1 0.33 0.028 0.17 0.014 1.36 25.65 0.71 0.05 0.71
2 1.04 0.087 0.26 0.022 1.36 31.72 2.75 0.17 0.88
3 1.86 0.155 0.31 0.026 1.36 34.63 5.37 0.30 0.96
4 2.88 0.240 0.33 0.028 1.36 35.73 8.58 0.46 0.99
5 4.85 0.404 0.33 0.028 1.36 35.73 14.44 0.78 0.99
6 6.25 0.521 0.32 0.027 1.36 35.18 18.33 1.00 0.98
7 7.65 0.638 0.31 0.026 1.36 34.63 22.08 1.22 0.96
8 9.62 0.802 0.34 0.028 1.36 36.27 29.07 1.54 1.01
9 10.64 0.887 0.35 0.029 1.34 36.80 32.63 1.70 1.02
10 11.46 0.955 0.3 0.025 1.34 34.07 32.53 1.83 0.95
11 12.15 1.013 0.26 0.022 1.34 31.72 32.11 1.94 0.88
12 12.5 1.042 0 0.000 1.34 0 0 2 0
808 rpm 2
r h u-
r*u
No. Static Velocity r/R u/U
in ft in ft (ft2/sec)
(ft/sec)
0 0 0.000 0 0.000 0.42 0 0 0 0
1 0.33 0.028 0.05 0.004 0.42 13.91 0.38 0.05 0.70
2 1.04 0.087 0.07 0.006 0.42 16.46 1.43 0.17 0.82
3 1.86 0.155 0.09 0.008 0.42 18.66 2.89 0.30 0.93
4 2.88 0.240 0.1 0.008 0.42 19.67 4.72 0.46 0.98
5 4.85 0.404 0.1 0.008 0.42 19.67 7.95 0.78 0.98
6 6.25 0.521 0.1 0.008 0.42 19.67 10.24 1.00 0.98
7 7.65 0.638 0.1 0.008 0.42 19.67 12.54 1.22 0.98
8 9.62 0.802 0.11 0.009 0.42 20.63 16.54 1.54 1.03
9 10.64 0.887 0.11 0.009 0.42 20.63 18.29 1.70 1.03
10 11.46 0.955 0.09 0.008 0.42 18.66 17.82 1.83 0.93
11 12.15 1.013 0.06 0.005 0.42 15.24 15.43 1.94 0.76
12 12.5 1.042 0 0.000 0.42 0 0 2 0
4
Figure 2
5
Figure 3
6
Figure 4
7
Figure 5
8
Results/Conclusion
The velocity profiles of figure 4 and 5 were not perfectly parabolic, but they did
follow a parabolic form. There is a strange deviation that occurs on both plots at the same
place that has the same discrepancy (between 1.5 and 2 r/R). This suggests to me some
sort of systematic error in the experiment. There is a slight dip in the center of the
velocity profile (see conclusion for discussion). The overall profile, steep slope giving
way to zero slope, then back to a steep negative slope indicates that the fluid has a very
low viscosity, since the center section has virtually the same velocity profile.
There are a few area that could have imparted error into this experiment. The pitot
tube may not have been perfectly perpendicular to the duct and thus the flow direction.
The pitot tube may not have been exactly in the position indicated, due to human error.
Ways to improve the above stated would be to have an indicator that would show
the direction in which the pitot tube was pointed, also to be careful when taking
measurements.
the velocity profile in a circular duct. I was surprised that the profile was not more
parabolic. Perhaps if the duct was smaller in diameter, of the fluid was more viscous, then
the result would have been a more parabolic shape. Another way to improve the
measurements would be to use a lower specific gravity fluid in the manometer. That
would allow for a larger height of fluid for the same pressure differential, and thus a more
9
I feel that there may be some systematic error in this experiment, dealing with
figure 4 and 5. The pitot tube is attached to an arm; this arm may be disrupting the air
flow when the pitot tube is all of the way in the air duct. This could be the reason why
there was a strange dip in figures 4 and 5. This could also explain the behavior of the
Overall, the results obtained with the current procedures and measurement deices
10