Calibration of Centrifuge Apparatus - Pharmaguideline
Calibration of Centrifuge Apparatus - Pharmaguideline
Calibration of Centrifuge Apparatus - Pharmaguideline
-->
Set the time for five minutes using set parameters. Record the results by using a digital stopwatch. Repeat the test for ten minutes.
Calibration of revolutions per minutes (RPM)
Set the RPM of rotor to 12000. Record the results by using tachometers.
Set RPM of rotor to 6000. Record the results by using tachometers
Set the RPM of rotor to 4000. Record the results by using tachometers.
Set RPM of rotor to 2000. Record the results by using tachometers.
Calibration of temperature
Fixed angle rotor
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 5°C, RPM: 12000
and time 45 minutes.
Record the temperature of the sample solution by using digital thermometer.
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 10°C, RPM: 12000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 20°C, RPM: 12000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 5°C, RPM: 4000 and
time 45 minutes.
Record the temperature of the sample solution by using digital thermometer.
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 10°C, RPM: 4000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Fill the two sample tubes with ethylene glycol and place it in the sample holder. Set the parameters like temperature 20°C, RPM: 4000
and time 45 minutes. Record the temperature of the sample solution by using digital thermometer
Frequency
Perform the calibration after every six months. Record the results in the calibration record format.
Maintenance and cleaning
Only neutral agents may be used for cleaning and disinfection (70% isopropanol/ water mixture or an alcohol -based disinfectant)
After cleaning with detergent, the rubber seals in the rotor chamber should be rinsed well in water and lubricated with glycerin in order to
prevent the seals from becoming brittle.
If condensation water forms in the rotor chamber after freezing occurs, dry with a soft absorbent cloth.
The rotor and bucket must be cleaned regularly to prevent contamination caused by residue.
When instrument does not comply with the specified limits the instrument should be labeled as “out of calibration” and should get repaired
/serviced.
After servicing /repairing the instrument must be calibrated as per above-mentioned procedure.
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 1/3
9/19/23, 10:56 AM Calibration of Centrifuge Apparatus : Pharmaguideline
Make : Model :
A] Speed Measurement
Sr. No. RPM Set RPM on Display RPM on Tachometer Tolerance RPM
1 4000 ±100
2 2000 ±100
B] Temperature Measurement
Sr. No. Temp Set (°C) Temp on Display (°C) Temp on Thermometer (°C) Tolerance (°C)
1 5 ±2
2 10 ±2
C] Time Measurement :
Sr. No. Time Set (Min.) Time on Display (Min.) Time on Stop watch (Min.) Tolerance (Min.)
1 5 ±2
2 10 ±2
Make : Model :
https://www.pharmaguideline.com/2011/02/calibration-of-eppendorf-centrifuge.html 2/3
9/19/23, 10:56 AM Calibration of Centrifuge Apparatus : Pharmaguideline
2 10 ±2
3 20 ±2
Remark : Satisfactory / Not satisfactory.
Performed by : Checked by:
Date : Date :
C] Time Measurement :
Instrument used : Stop Watch
Sr. No. Time Set (Min.) Time on Display (Min.) Time on Stop watch (Min.) Tolerance (Min.)
1 5 ±2
2 10 ±2
Remark : Satisfactory / Not satisfactory.
Performed by : Checked by:
Date : Date :
Conclusion : Instrument working is Satisfactory / Not satisfactory.
Next calibration due on :
Performed by : Checked by :
Date : Date :