Tango FT Nir Spectrometer
Tango FT Nir Spectrometer
Specifications
Rugged NIR analyzer with sealed and desiccated optics for routine quality control. The analyzer
is equipped with a gold coated integrating sphere for the reflection measurement of samples.
Sample spinners for inhomogeneous samples are optionally available.
The analyzer is either equipped with a built-in data system with a touch screen or optionally
available with an external PC-based data system. The easy-to-use and GMP compliant TANGO
software package allows qualitative and quantitative analysis.
Design
< Housing: rugged, sealed and desiccated housing
< NIR source: air cooled Tungsten source (5 V/ 7W), average lifetime
9,000 h, pre-aligned, easily exchangeable by operator
< Interferometer: RockSolidTM, frictionless bearing, permanently aligned,
Quartz substrate beamsplitter with proprietary coating
< Laser diode: average lifetime 10+ years
< Detector: InGaAs diode, TE-cooled and temperature-stabilized
< Sample presentation: reflection measurements in cups, Petri dishes or disposable vials
F T-NIR
Innovation with Integrity
Performance Software
< Spectral range: 11,500 - 4,000 cm -1
< Resolution: better than 4 cm -1
< Wavenumber reproducibility: better than 0.04 cm -1, at 7,181 cm -1
< GMP compliant TANGO
< Wavenumber accuracy: better than 0.1 cm -1, at 7,181 cm -1 software with user-friendly
< Photometric accuracy: better than 0.1% T interface
< Product based operation
Electronics (Barcode reader friendly)
< Qualitative and/or quan-
< Data acquisition: 24-bit A/D converter
< Automation: microprocessor controlled optical bench, digital titative evaluations (no
speed control, automatic gain selection, internal method development)
< Customizable analysis
background, advanced system check
< Performance check: permanent on-line diagnostics of all optical com- reports
< Data exchange via USB
ponents, automation units and sampling accessories
< Connectivity: Ethernet interface, 10/100 Mbps stick and/or LAN
Dimensions
< Spectrometer (w x d x h): 292 mm x 438 mm x 175 mm
Operating Environment
< Protection: NEMA 13 / IP54
< Operating temperature: 5° C to 35° C
< Storage temperature: -20° C to 70° C
< Operating relative humidity: < 80% non-condensing Product selection by pressing
< Power requirements: 100 to 240 VAC, 50/60 Hz, 75 W (typical 40 W) the respective icon
< Regulatory certification and
Data System
< Embedded PC with 10.5” touch display; optional external PC data system
Validation
< TANGO software: the software package includes functions that facilitate
the use of the analyzer in a 21 CFR Part 11 compliant
environment
< Instrument qualification: TANGO software supports Operational Qualification (OQ)
PQ tests
< Validation manual: complete hardware and software validation
Bruker Optics
is ISO 9001 certified.
Covered by one or more of the following patents: DE102004025448; DE19940981.
Additional patents pending. Laser class 1 product.
www.bruker.com/optics Bruker Optics Inc. Bruker Optik GmbH Bruker Hong Kong Ltd.
Billerica, MA · USA Ettlingen · Germany Hong Kong
Phone +1 (978) 439-9899 Phone +49 (7243) 504-2000 Phone +852 2796-6100
Fax +1 (978) 663-9177 Fax +49 (7243) 504-2050 Fax +852 2796-6109
[email protected] [email protected] [email protected]
Bruker Optics is continually improving its products and reserves the right to change specifications without notice.
© 2013 Bruker Optics BOPT-4000085-02