Instruction Manual: 230 Series

Spellman's 230 Series power supplies provide regulated high voltage DC outputs from 1-30kV with low ripple and noise. They have a modular benchtop design, 3.5 digit front panel metering, and reversible polarity. Applications include high potential testing, electrostatics, and general laboratory use. The units have precision voltage and current regulation and monitoring, over voltage protection, and remote programming capability.

Uploaded by

alex lzg
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
373 views16 pages

Instruction Manual: 230 Series

Spellman's 230 Series power supplies provide regulated high voltage DC outputs from 1-30kV with low ripple and noise. They have a modular benchtop design, 3.5 digit front panel metering, and reversible polarity. Applications include high potential testing, electrostatics, and general laboratory use. The units have precision voltage and current regulation and monitoring, over voltage protection, and remote programming capability.

Uploaded by

alex lzg
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

R

Instruction Manual

230 SERIES

Bench Top High Voltage Power Supply

SPELLMAN
HIGH VOLTAGE ELECTRONICS
CORPORATION
One Commerce Park
Valhalla, New York, 10595

+1(914) 686-3600 * FAX: +1(914) 686-2870


E-mail: [email protected]
Website: http://www.spellmanhv.com

230 Series User’s Manual 118127-001 Rev A


230 HIGH VOLTAGE
POWER SUPPLY
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N

PAGE 1 OF 2

• MODULAR BENCH TOP DESIGN


• LOW RIPPLE AND NOISE
• 3.5 DIGIT FRONT PANEL DIGITAL METERING
• REVERSIBLE OUTPUT POLARITY
www.spellmanhv.com/manuals/230

Spellman’s Bertan brand of 230 Series high voltage power Accuracy:


supplies provide regulated high voltage outputs from 1 to Front panel control: ±(0.2% of setting + 0.2% of maximum)
30kV. The low noise, linear topology employed results in Front panel Meter: Voltage ±(0.5% of setting + 0.5% of
extremely low output ripple specifications. These 12 to 15 maximum), Current ±(2% of setting + 0.5% of maximum)
watt units are inherently reversible by design, providing Remote Programming: ±(0.1% of setting + 0.1% of maximum)
Voltage Monitor: ±(0.1% of reading + 0.1% of maximum)
either positive or negative output polarity. The 230 Series is
Current Monitor: ±(2% of reading + 1% of maximum)
fully arc and short circuit protected. Excellent regulation
specifications are featured along with outstanding stability Front Panel Metering and Controls:
performance. Power ON/OFF switch
3.5 digit metering for voltage and current, switch selectable
TYPICAL APPLICATIONS Polarity indicator
HiPot Testing 10 turn locking potentiometer to set output voltage
Electrostatics HV output connector
General Laboratory Usage Ground stud

Operating Temperature
SPECIFICATIONS 0°C to +50°C
Input Voltage: Storage Temperature:
115Vac, ±10%, 50/60 Hertz @ 0.5 amp -40°C to +85°C
230Vac, ±10%, 50/60 Hertz @ 0.25 amps
Input voltage is switch selectable Humidity:
20% to 85% RH, non-condensing
Output Voltage:
See “model selection” table Input Line Connector:
IEC320 EMI filter/input connector, a detachable
Output Polarity: line cord is provided
All units are reversible polarity by design
Interface Connector:
Output Current: 9 pin “D” connector, a mating connector is provided
See “model selection” table
Output Connector:
Voltage Regulation: A detachable 10 foot (3 meter) HV cable is provided
Line: ≤0.002% of rated output voltage over specified for units up to 5kV; 10kV through 20kV: 59” (1.5 meter);
input voltage range 30kV: 10 foot (3 meter)
Load:≤0.005% of rated output voltage for a full load
Cooling:
change
Convection cooled
Current Regulation:
Dimensions
Internally set to limit at less than 125% of rated current. 7.63˝ W X 5.03˝ H X 8.91˝ D
A rear panel switch allows limiting at 25% of rated full current. (194mm X 128mm X 226mm)
Ripple: Weight:
See “model selection” table ≤10 pounds (4.5kg)
Temperature Coefficient:
≤100ppm/°C Regulatory Approvals:
Stability: Compliant to 2004/108/EC, the EMC Directive and
≤0.01%/hour, 0.02% per 8 hours after a 1/2 hour warm up 2006/95/EC, the Low Voltage Directive.

USA +1-631-630-3000 FAX: +1-631-435-1620 e-mail: [email protected]


UK +44 (0)1798 877000 FAX: +44 (0)1798 872479 www.spellmanhv.com 128051-001 REV.D
JAPAN +81 (0)48-447-6500 FAX: +81 (0)48-447-6501
CHINA +86 (0)512-67630010 FAX: +86 (0)512-67630030 Spellman High Voltage is an ISO 9001:2000 and ISO 14001:2004 registered company
230 HIGH VOLTAGE
POWER SUPPLY
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N
PAGE 2 OF 2

DIMENSIONS: in.[mm]
MODEL SELECTION TABLE
230 Voltage Current Ripple
Series FRONT VIEW
230-01R 0 to 1kV 0 to 15mA 10mV
230-03R 0 to 3kV 0 to 5mA 30mV HIGH VOLTAGE
ADJUST PCS
230-05R 0 to 5kV 0 to 3mA 50mV METER

230-10R 0 to 10kV 0 to 1.5mA 500mV VF S

230-20R 0 to 20kV 0 to 0.5mA 2 volts


5,03 [128]
230-30R 0 to 30kV 0 to 0.4mA 5 volts POWER
ON
GROUND

OFF
HV CONN

7.63 194]
INTERFACE CONNECTOR
PIN SIGNAL PARAMETERS

1 Voltage Monitor 0 to 5Vdc = 0 to 100% rated voltage, Zout = 10KΩ SIDE VIEW
2 n/c none
3 Enable TTL “0” disables HV, TTL “1” or open enables HV
8.50 [216] 1.0 [25]
4 +5Vdc Reference +5.0Vdc @ 10mA, maximum
5 Current Monitor 0 to 5Vdc = 0 to 100% rated current, Zout = 10KΩ
6 Voltage Program Input 0 to 5Vdc = 0 to 100% rated voltage, Zin = 1MΩ
7 Analog Ground Ground
8 Digital Ground Ground
9 Polarity Indicator Open collector, 30V @ 25mA, positive = ON POS NEG

OPTIONS:
Isolated (Floating) Output-Option F
Units up to and including 5kV can be provided with differential outputs
.66 [17] 8.25 [210]
capable of floating up to ±2kV from ground. Voltage programming and
monitoring functions normally referenced to ground. Current monitoring
and metering is eliminated. Replace “R” suffix with “F” for this option.

BACK VIEW

USA +1-631-630-3000 FAX: +1-631-435-1620 e-mail: [email protected]


UK +44 (0)1798 877000 FAX: +44 (0)1798 872479 www.spellmanhv.com 128051-001 REV.D
JAPAN +81 (0)48-447-6500 FAX: +81 (0)48-447-6501
CHINA +86 (0)512-67630010 FAX: +86 (0)512-67630030 Spellman High Voltage is an ISO 9001:2000 and ISO 14001:2004 registered company
IMPORTANT SAFETY PRECAUTIONS

SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.

High voltage power supplies must always be grounded.

Do not touch connections unless the equipment is off and the


Capacitance of both the load and power supply is discharged.

Allow five minutes for discharge of internal capacitance of the power supply.

Do not ground yourself or work under wet or damp conditions.

SERVICING SAFETY
.

Maintenance may require removing the instrument cover with the power on.

Servicing should be done by qualified personnel aware of the electrical hazards.

WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.

CAUTION notes in the text indicate procedures to be followed to avoid possible


damage to equipment.

Copyright © 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.

118004-001 REV. B
WICHTIGE SICHERHEITSHINWEISE

SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.

Das Hochspannungsnetzteil muß immer geerdet sein.

Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.

Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.

Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.

SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.

Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.

“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.

“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.

118004-001 REV. B
PRECAUTIONS IMPORTANTES POUR VOTRE SECURITE

CONSIGNES DE SÉCURITÉ
CETTE ALIMENTATION GÉNÈRE DES TENSIONS QUI SONT DANGEUREUSES ET PEUVENT ÊTRE FATALES.
SOYEZ EXTRÊMENT VIGILANTS LORSQUE VOUS UTILISEZ CET ÉQUIPEMENT.

Les alimentations haute tension doivent toujours être mises à la masse.

Ne touchez pas les connectiques sans que l’équipement soit éteint et que la capacité à la fois de la charge et de
l’alimentation soient déchargées.

Prévoyez 5 minutes pour la décharge de la capacité interne de l’alimentation.

Ne vous mettez pas à la masse, ou ne travaillez pas sous conditions mouillées ou humides.

CONSIGNES DE SÉCURITÉ EN CAS DE REPARATION


La maintenance peut nécessiter l’enlèvement du couvercle lorsque l’alimentation est encore allumée.

Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques.

Dans le manuel, les notes marquées « WARNING » attire l’attention sur les risques lors de la manipulation de ces
équipements, qui peuvent entrainer de possibles blessures voire la mort.

Dans le manuel, les notes marquées « CAUTION » indiquent les procédures qui doivent être suivies afin d’éviter
d’éventuels dommages sur l’équipement.

118004-001 REV. B
IMPORTANTI PRECAUZIONI DI SICUREZZA
SICUREZZA
QUESTO ALIMENTATORE GENERA TENSIONI CHE SONO PERICOLOSE E
POTREBBERO ESSERE MORTALI.
PONI ESTREMA CAUTELA QUANDO OPERI CON QUESO APPARECCHIO.

Gli alimentatori ad alta tensione devono sempre essere collegati ad un impianto di terra.

Non toccare le connessioni a meno che l’apparecchio sia stato spento e la capacità interna
del carico e dell’alimentatore stesso siano scariche.

Attendere cinque minuti per permettere la scarica della capacità interna dell’alimentatore
ad alta tensione.

Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d’umidità.

SICUREZZA NELLA MANUTENZIONE.

Manutenzione potrebbe essere richiesta, rimuovendo la copertura con apparecchio


acceso.
La manutenzione deve essere svolta da personale qualificato, coscio dei rischi elettrici.

Attenzione alle AVVERTENZE contenute nel manuale, che richiamano all’attenzione ai


rischi quando si opera con tali unità e che potrebbero causare possibili ferite o morte.

Le note di CAUTELA contenute nel manuale, indicano le procedure da seguire per evitare
possibili danni all’apparecchio.

118004-001 REV. B
TABLE OF CONTENTS

Paragraph Title Page

SECTION 1 - GENERAL INFORMATION


1.1 Purpose of the Equipment 2
1.2 Description 2
1.3 Options 2
1.4 Safety Terms 2

SECTION 2 - OPERATION
2.1 Installation 2
2.2 Front Panel Controls and Displays 2/3
2.3 Rear Panel Controls, Connectors and Terminals 3
2.4 Polarity Reversing 3
2.5 Preparations for Use 3/4
2.6 Local Operation 4
2.7 Remote Operation 4/5
2.8 Computer Programming 5/6
2.9 Input Power 6
2.10 Current Limiting 6/7
2.11 High Voltage Output 7

SECTION 3 - THEORY OF OPERATION


3.1 Functional Description 7
3.2 Circuit Description 7
3.3 Current Capability 7/8

SECTION 4 - MAINTENANCE
4.1 General 8
4.2 Cleaning 8
4.3 Calibration 8

SECTION 1 GENERAL INFORMATION


Installation and Instruction Manual - Series 230 Page 1 118128-001 Rev A
SECTION 2 OPERATION
1.1 PURPOSE OF THE EQUIPMENT
The Series 230 is a family of regulated precision 2.1 INSTALLATION
laboratory high voltage power supplies. They WARNING! This unit produces hazardous
provide exceptional performance in critical voltage. Do not apply line voltage input unless
applications such as nuclear and electro-optical adequate ground is connected to the unit and
instrumentation, precision CRT and electron the high voltage output has been properly
beam applications. connected.

1.2 DESCRIPTION 2.2 FRONT PANEL CONTROLS AND


The Series 230 is a family of compact bench top DISPLAYS
power supplies with output voltages up to 30 Output Connector:
kV. The units consist of a DC power supply that The HV output connector on the 1kV through
converts the AC line power to a low DC voltage 5kV models mate with a shielded mating
and a DC to DC converter that generates the connector supplied with each unit. Refer to the
high DC output voltage. Low voltage electronic specifications on page 12 to identify the mating
solid-state circuitry is mounted on the PCB100, connector. Assembly procedures for mating
and the high voltage assembly is fully connectors are given at the end of this manual.
encapsulated for reliable, arc-free, operation. On 10kV through 30kV models, an unshielded,
preassembled high voltage cable connected to
These stable, low noise high voltage power the appropriate connector is provided with each
supplies feature front panel digital voltage and unit. Only the proper mating connector should
current metering, and calibrated direct-reading be used with the indicated power supply and the
front panel controls. The high voltage output is power supply should NEVER be energized
also obtainable from the front panel. The rear without a mating connector and suitable load
panel features a connector for remote analog connected.
programming and output voltage and output
current monitoring, line power plug, fuse, Power Switch:
and AC voltage selection switch. All units have A rocker switch turns the line power on or off to
arc and short circuit protection for safe, reliable, the entire instrument. The display panel will be
and arc-free operation. Although primarily lit when line power is applied to the unit.
designed for rack mounting, the unit may also be
used in benchtop applications. Output Meter:
The digital output meter can display the output
1.3 OPTIONS current or the output voltage. A switch below
Isolated Floating Output the meter allows the operator to select which
Units can be provided with the output capable of output parameter to monitor. The accuracy of
floating up to r2kV from ground. All controls, the meter is given in the specification (see page
programming and monitoring functions operate 12).
normally, referenced to ground. The high
voltage output polarity, with respect to the Polarity Indicator:
floating input terminal is reversible. An LED on the front panel display indicates the
polarity of the output. The appropriate LED is
1.4 SAFETY TERMS lit as soon as line power is applied regardless of
The WARNING used in this manual explains whether high voltage output is enabled or
dangers that could result in personal injury or disabled. To change the polarity setting, see
death. paragraph 2.4.

The CAUTION used in this manual explains Voltage Control:


hazards that could damage the instrument.

Installation and Instruction Manual - Series 230 Page 2 118128-001 Rev A


A continuous multi-turn digital dial is used to mechanism that is easily accessible upon
adjust the high voltage output. The resolution removal of the top cover. The polarity reversal
0.2% of maximum. module is a clear plastic assembly identified by
the exiting silicone high voltage cables. It is a
2.3 REAR PANEL CONTROLS, two-part assembly. To change the polarity, turn
CONNECTORS, AND TERMINALS off power supply, remove all cover screws
Gnd: holding the top cover on and:
Ground is connected to the case of the Series
230. a. Remove the two diagonally opposed screws
fastening the top portion of the module
Fuse: assembly to the bottom portion. NOTE: DO
The fuse is the ac line power fuse. It is rated for NOT DESOLDER WIRES OR PINS.
1A, 250Vac for 105Vac-125Vac operation and b. Carefully separate the module by pulling
0.5A, and 250Vac for 210Vac-250Vac the top portion from the bottom portion.
operation. Should a fuse ever need replacement, The module portions are fitted very snugly
only these values should be used unless and removal may be eased by slightly
otherwise advised by a qualified BERTAN rocking the assembly.
service technician. c. Rotate the top portion of the module
assembly 180q, taking care not to unduly
Line Voltage Selector: stress the high voltage cables.
The line voltage selector selects the appropriate d. Rejoin the 2 portions of the module
line voltage 105Vac-125Vac or 210Vac-250Vac assembly. Make sure that the top portion is
at 50-60 Hz. By default, power supplies are entirely seated to the bottom portion.
shipped from the factory in the 105-125V NOTE: An interlock automatically insures
position. Before energizing your power supply, that the high voltage cannot be applied until
verify that the line voltage selector switch is in the portions of the module are properly
the proper position for your mains input. mated.
e. Resecure the top portion to the bottom
AC Line Plug: portion of the Polarity Reversal Module
The IEC 320 line plug receptacle accepts a Assembly.
three-wire female line plug for ac line power. f. Re-cover the power supply.

WARNING! This unit is equipped with a 2.5 PREPARATION FOR USE


three-wire grounded line cord. This must be WARNING! Before energizing your power
used with a three-wire receptacle where the supply, thoroughly review and follow these
"third wire" is connected to earth ground; procedures. Failure to do so may result in
otherwise personal injury or death may occur. damage to equipment and injury or death to
personnel.
2.4 POLARITY REVERSING:
WARNING! Before attempting to reverse the To prepare the Series 230 for use, use the
power supply’s polarity, the power supply must following procedure:
be turned off and the output fully discharged.
Failure to follow these procedures may result Set the Series 230 for the appropriate line
in damage to the power supply, associated test voltage as specified in Section 2.3.
equipment and/or personnel.
Connect a ground strap from case ground (on the
For 1kV to 5kV output models, a screwdriver- rear panel) to a system common.
adjustable POLARITY SELECTOR SWITCH is
accessible at the rear panel of the unit, next to Select the appropriate HV output polarity for the
the HV output connector. For 10kV to 50kV application.
output models, the polarity of the HV output is
reversible by means of an internal switching

Installation and Instruction Manual - Series 230 Page 3 118128-001 Rev A


Set the front panel controls to: is reached. Apply power to the load by
a. Power Switch - OFF switching the High Voltage - ON. The output
b. Output Voltage Switch(s) – 0 (205B will quickly reach the value set by the controls.
only). Full stability will be achieved after
c. Multi Turn Digital Dial - 000 (fully approximately 30 minutes.
counterclockwise)
2.7 REMOTE OPERATION
Set the LOCAL/REMOTE (ANALOG/ PROGRAM CONTROL SWITCH:
DIGITAL) rear panel switches (205B only) to: Before the Series 230 can be remotely
programmed, the instrument must be configured
a. LOCAL for local front panel by setting the rear panel PROGRAM
operation. CONTROL SWITCH (S102) to the REMOTE
ANALOG position (206B is only programmable
b. REMOTE/ANALOG for remote through the rear panel and therefore S101 is
analog operation. Note: Selecting the omitted). All monitoring and enable functions
REMOTE control will override all local are active, independent of the S102 switch, as
front panel controls of the output. are the front panel meters. When in remote
mode the front panel controls are inactive.
c. REMOTE/DIGITAL for Computer
Programming if equipped with a CBNY REMOTE PROGRAMMING:
option. Reference Section 2.7. The high voltage output can be remotely
programmed from either an external voltage
Plug the line cord into the power line with a source or with an external potentiometer using
three-wire IEC receptacle to maintain proper the internal reference voltage source (Pin 4). A
case ground. 0 to +5Vdc programming voltage applied to Pin
6 of J107 (PROGRAMMING/MONITOR)
WARNING! This unit is equipped with a connector jack on the rear panel will remotely
three-wire grounded line cord. This must be program the high voltage output from zero to
used with a three-wire receptacle where the maximum output. Programming can also be
"third wire" is connected to earth ground; accomplished using a potentiometer connected
otherwise personal injury or death may occur. between Pin 4 (+5Vdc), Pin 7 (GND) and with
Connect the output of the Series 230 to the the wiper connected to Pin 6 (PRGM INPUT).
circuit. Use a properly rated shielded cable with The potentiometer should be a low temperature
the supplied HV output connector to insure good coefficient wirewound or cermet type, 5k: to
circuit connections and safe operation. Refer to 20k: resistance values. The power supply
Section 2.10. output will be proportional to the programming
input. The programming input impedance is
WARNING! Prior to connecting or removing greater than 1M:. TABLE 2.1 below lists the
any equipment from the High Voltage power PROGRAMMING / MONITOR connector pin
supply, always return the Output Voltage designations. The accuracy of the remote
Control(s) to 0V prior to applying or removing programming is detailed in the specifications.
power. External circuits may retain voltage
after controls are set to zero. Discharge any
residual voltage before connecting or removing
any equipment.

TABLE 2.1 - J107 PIN DESIGNATIONS


2.6 LOCAL OPERATION PIN # FUNCTION
Turn POWER - ON to the instrument. Slowly 1 Output voltage monitor, buffered, 0 to +5Vdc
increase the output voltage using the appropriate (output impedance 10k:)
2 No connection
Voltage Control(s) until the desired output level

Installation and Instruction Manual - Series 230 Page 4 118128-001 Rev A


3 Enable/Disable. Input logic zero disables high All Series 230 instruments can be provided with
voltage generation. Open circuit or input a factory-installed option for remote digital
logic one enables high voltage generation.
programming of the high voltage output. The
4 Precision +5Vdc reference output referenced
to analog ground. programming inputs are TTL compatible and the
5 Output current monitor, buffered, 0 to +5Vdc data is positive logic (all data bits low yield 0
(output impedance 10k:) high voltage output). The addition of this option
6 Remote analog voltage programming input, 0 allows the unit to be easily interfaced to any
to +5Vdc computer or microprocessor utilizing one of its
7 Analog Ground three user selectable modes of operation.
8 Digital Ground
9 Polarity Indicator
16 Bit Transparent: The 16 bit data is passed
from the inputs directly to the DAC. This is the
REMOTE ANALOG MONITORING:
default mode (is 100% compatible with all
Buffered, analog output monitors, 0 to + 5Vdc,
previous CBNY digital programming boards).
linearly proportional to the power supply's
voltage and current output are provided. To
16 Bit Register: The 16 bit data is latched into an
monitor the output voltage, connect a high
internal 16-bit register in one write cycle.
impedance meter to pin 1 and pin 7 (ground). To
monitor the output current, connect a high
8 Bit Register: Two 8-bit bytes (Most
impedance meter to pin 5 and pin 7 (ground).
Significant Byte and Least Significant Byte) are
The accuracy of the voltage and current
latched into two 8-bit registers. The MSB and
monitors is given in the specifications (see page
LSB registers are individually addressed and
12). The monitor output impedance is
written. This allows an 8 bit data bus system to
approximately 10k:.
provide 16 bit programming in two write cycles.
ENABLE/DISABLE:
The register modes utilize standard Chip Select
A TTL level logic TRIP input signal can be used
and Write Enable protocol allowing the CBNY
to enable or disable the power supply output
to act as a memory mapped register or an I/O
remotely. Input logic zero or grounding pin 3
port attached directly to an 8 or 16 bit
disables high voltage generation. Open circuit or
microprocessor system bus. In addition, any
input logic one on pin 3 enables high voltage
standard unit in the Series 230 can be computer
generation.
programmed and monitored using the Bertan B-
HiVE minicomputer controlled enclosure with
+5Vdc REFERENCE OUTPUT:
the appropriate interface module. Remote TTY
A precision +5Vdc reference output is provided
or RS-232C control at selectable 110 to 9600-
on pin 4 for the user’s convenience. This fixed
baud rate is possible. A separate IEEE-488
output can be used for remote resistance
interface is also available for use with the Series
programming (see REMOTE
205B high voltage power supplies. See the
PROGRAMMING, above) or various control
Model 200-C488 data sheet for complete
functions. This output is referenced to analog
information.
ground (pin 7).
FUNCTIONAL DESCRIPTION:
POLARITY INDICATOR:
Connector J3 (Amphenol 57-40240) accepts the
A TTL polarity indicator output signal is
remote binary coded inputs and is located on the
available at pin 9. An NPN open collector
rear panel. When the REMOTE/LOCAL switch
connection with respect to digital ground
is in the REMOTE DIGITAL position, control
indicates the high voltage output polarity. NPN
of the unit is dependent upon the digital signals
saturation denotes positive polarity.
present at Pin 2 through Pin 17 of J3. Pin 2 is
for the most significant bit, with increasing pin
2.8 COMPUTER PROGRAMMING numbers having lower significant bits. Pin 17 is
(Optional) the least significant bit input. Positive logic is
GENERAL:

Installation and Instruction Manual - Series 230 Page 5 118128-001 Rev A


used, with logic 0 on all 16 data lines needed for Pin 23: Enable. If strapped to "0" or open the
0 volts output. output of the latch (input of the DAC) is at high
impedance and the input of the DAC set to zero.
The incremental resolution is a function of the If strapped to "1" the output of the latch is
power supply's maximum output. The enabled and the data at the input of the DAC is
resolution for different models in the Series is dependant on the CS WR and F/L.
listed in TABLE 2.2 below.
Pin 24: F/L-Latch/Follow. If strapped to "0" the
TABLE 2.2 – SERIES 230 DIGITAL device operates in a latching mode where data
PROGRAMMING RESOLUTION on the data bus is latched intoº on board latches
-01R 0.02V by a write pulse. If strapped to "1" or left open,
-03R 0.05V the data at the input to the D to A follows the
-05R 0.08V data on the data bus regardless of the levels on
-10R 0.16V CS, WR or BS. The 16/8 must be open or
-20R 0.31V strapped to "1" when in the follow mode or the
-30R 0.50V most significant byte to the DAC will be
-50R 0.80V identical to the least significant byte input.

OPERATION: Byte select must be valid at least 25ns before


In addition to the 16 data lines, J3 contains WR goes low and remain valid at least 40ns
ground and +5Vdc output on Pins 21 and 22 after the first of CS or WR to go high.
respectively, and five control lines described
below. The +5Vdc and ground allow remote All timing is from the latter of CS or WR going
control via an external digitally coded switch. low and from earlier of CS or WR going high.

The digital data at the input can continuously The minimum common low time of WR and CS
control the power supply output high voltage. is 40ns. Data must be valid at least 125ns before
The detailed functions of the control inputs are the earlier of CS or WR goes high and remain
described below. valid at least 20ns after that time.
Pin 01: 16/8. If strapped to "1" or left open the
device operates as a 16 bit parallel input. If 2.9 INPUT POWER
strapped to "0" the device operates as an 8 bit Input AC line voltage required is
parallel bus with the BS input selecting the byte 115Vac/230Vac r10%, 50-60Hz, single phase.
being written. The recessed LINE VOLTAGE selector switch
on the rear panel selects either 115 Vac or 230
Pin 18: CS-Chip Select (device select). When Vac operation. By default, power supplies are
"0" the data on the bus may be written into the shipped from the factory in the 115Vac position.
latches by the WR signal. Before energizing your power supply, verify that
this switch is in the proper position for your
Pin 19: WR-Write. When "0" with CS, "0" the mains input.
input to the DAC the data on the bus. When the
first CS or WR is returned to "1" the data 2.10 CURRENT LIMITING
present on the bus is latched into the DAC. The Series 230 includes a current limiting circuit
that drops the output voltage to a safe level
Pin 20: BS-Byte Select. If the 16/8 input is when the rated output current is exceeded by
strapped to "0", a "1" on the BS selects the least approximately 5%. (See specification on
significant byte to write data to. If BS is "0" Current Capability when operating the unit at
data is written to the most significant byte. If reduced output voltages or when operating in a
16/8 input is "1" or open the BS lead has no current limit mode for capacitor charging).
effect.
2.11 HIGH VOLTAGE OUTPUT

Installation and Instruction Manual - Series 230 Page 6 118128-001 Rev A


The high voltage output connector is located on voltage power supplies. The B+ supply is a
the rear panel. An appropriate shielded mating filtered full wave rectifier circuit located on the
connector is supplied with each unit. These chassis. The regulated low voltage power
connectors are as listed in Table 2.3. Refer to supply circuit (+12Vdc) consists of a rectifier
pages 13-16 for the mating connector assembly circuit located on T1 and output regulators
instructions. Only the proper mating connector located on the PCB 100.
should be used with the indicated power supply The output of the oscillator circuit is amplified
and the power supply should never be energized in the AGC amplifier. The gain of the AGC
without a mating connector and suitable load amplifier is a function of the control voltage
connected. developed at the output of the error amplifier.

TABLE 2.3: SERIES 230 HIGH VOLTAGE The encapsulated high voltage assembly
CONNECTORS includes a high voltage power transformer,
MODEL OUTPUT MATING rectifier or multiplier circuits, ripple filter and
-01R JDK PDB sensing circuits. These are all critical custom
-03R JDK PDB designed and encapsulated components.
-05R JDK PDB
-10R JJA 405787 A sample of the high voltage DC output is fed to
-20R JJA 405787 the output voltage sensing circuit and is
compared to a command voltage. Output
-30R JJA 405787
voltage control is obtained by varying the
-50R JJB 405786
command voltage fed to the error amplifier. The
error amplifier compares the command voltage
SECTION 3 and the signal from the output voltage sense
THEORY OF OPERATION circuit. Any difference causes a correction in
the gain control of the AGC amplifier. The
3.1 FUNCTIONAL DESCRIPTION command voltage is controlled by the front
The circuit uses a DC to DC converter that panel controls when the rear panel program
converts low voltage DC power to a high switch is in the LOCAL position.
voltage DC output. This output voltage is
highly regulated and filtered and can be varied The reference and reference control and buffer
either by the front panel controls or through the provide a stable +5Vdc to the front panel output
REMOTE PROGRAM input on the rear panel. voltage controls.
The input to the DC to DC converter is obtained
from internal low voltage power supplies The current sensing circuit monitors the output
powered by the AC line input. current. The buffered output of this circuit is
employed for both internal and remote current
An oscillator determines the frequency monitoring.
(approximately 20kHz) at which all
amplification, high voltage transformation, 3.3 CURRENT CAPABILITY
rectification and filtering occurs. The The maximum current rating for each model, as
amplification is a function of a control voltage shown in the table to the left, is applicable when
that performs the function of control and the unit is operated at maximum output voltage.
regulation. A sample of the output voltage is When operated at reduced output voltage levels,
compared against a reference voltage in the or when operating in a current limit mode for
sensing circuit. The sensing circuit generates charging capacitors, the output current must be
the control voltage to set and maintain a fixed limited to reduced levels. This is required to
high voltage output. protect against excessive power dissipation of
the driver transistors.
3.2 CIRCUIT DESCRIPTION
The input AC line is converted to the B+ The maximum output current must be linearly
(36Vdc) supply and regulated +12Vdc low derated from maximum output voltage to 30% of

Installation and Instruction Manual - Series 230 Page 7 118128-001 Rev A


maximum current at zero output voltage. 4.3 CALIBRATION SERVICES
Maximum output current available at any Your BERTAN high voltage power supply is
desired voltage can be calculated by applying designed to provide many years of reliable
the following formula: service. For a nominal charge it can be returned
to the factory for calibration and certification to
Imax = Vset(.7Irated/Vrated)+.3Irated its original specification. For traceability, a
certificate will be issued, identifying the serial
When operating the power supply as a capacitor number of the unit calibrated and all test
charger (i.e., a capacitor is being continuously equipment used to perform the calibration. All
discharged then recharged from zero voltage by measurements are traceable to the National
the power supply) use a charging resistor in Institute of Standards and Technology (NIST).
series with the power supply output. The Calibration is guaranteed from 1 year of
resistance should be equal to the power supply’s issuance. Contact the factory at (516) 433-3110
maximum rated output voltage divided by the or your local sales representative for additional
maximum rated output current. details.

The above derating factors are safe for all


conditions and all models. Consult BERTAN
for special cases before exceeding these factors.

SECTION 4 MAINTENANCE

4.1 GENERAL
The Series 230 instrument should not require
any maintenance. It is designed for reliable,
trouble free operation. If any question should
arise, contact the Bertan Customer Service
Department for assistance or return
authorization. It is suggested that the unit be
returned to the factory if service should become
necessary.

4.2 CLEANING
Cleaning of the power supply should only be
performed with the supply disconnected from
the ac power source. A soft cloth moistened
with conventional ammonia-based cleaning
agents will suffice for all exposed surfaces. The
metal shell of the HV connector should be
cleaned with isopropyl alcohol.
If the supply is operated in a dusty environment,
an accumulation of dust/debris may build-up
inside the unit which may cause noisy operation
(i.e., “ticking” or minor crackling) in the area of
the HV cabling on the –10R through –50R. The
safest way to remove such debris is with
compressed air. Ensure that no dust/debris is
left behind in the insulative medium of the HV
output connector after this cleaning operation.
Such dust may be removed with a cotton swab
moistened with isopropyl alcohol.

Installation and Instruction Manual - Series 230 Page 8 118128-001 Rev A


SPELLMAN HIGH VOLTAGE ELECTRONICS
WARRANTY
Spellman High Voltage Electronics (“Spellman”) warrants that all power supplies it manufactures will be
free from defects in materials and factory workmanship, and agrees to repair or replace, without charge, any
power supply that under normal use, operating conditions and maintenance reveals during the warranty
period a defect in materials or factory workmanship. The warranty period is twelve (12) months from the
date of shipment of the power supply. With respect to standard SL power supplies (not customized) the
warranty period is thirty-six (36) months from the date of shipment of the power supply.

This warranty does not apply to any power supply that has been:
• Disassembled, altered, tampered, repaired or worked on by persons unauthorized by Spellman;
• subjected to misuse, negligent handling, or accident not caused by the power supply;
• installed, connected, adjusted, or used other than in accordance with the original intended application and/or
instructions furnished by Spellman.

THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

The buyer’s sole remedy for a claimed breach of this warranty, and Spellman’s sole liability is limited, at
Spellman’s discretion, to a refund of the purchase price or the repair or replacement of the power supply at
Spellman’s cost. The buyer will be responsible for shipping charges to and from Spellman’s plant. The
buyer will not be entitled to make claim for, or recover, any anticipatory profits, or incidental, special or
consequential damages resulting from, or in any way relating to, an alleged breach of this warranty.

No modification, amendment, supplement, addition, or other variation of this warranty will be binding unless
it is set forth in a written instrument signed by an authorized officer of Spellman.

Factory Service Procedures


For an authorization to ship contact Spellman’s Customer Service Department. Please state the model and
serial numbers, which are on the plate on the rear panel of the power supply and the reason for return. A
Return Material Authorization Code Number (RMA number) is needed from Spellman for all returns. The
RMA number should be marked clearly on the outside of the shipping container. Packages received without
an RMA Number may delay return of the product. The buyer shall pay shipping costs to and from Spellman.
Customer Service will provide the Standard Cost for out-of-warranty repairs. A purchase order for this
amount is requested upon issuance of the RMA Number (in-warranty returns must also be accompanied by
a “zero-value” purchase order). A more detailed estimate may be made when the power supply is received
at Spellman. In the event that the cost of the actual repair exceeds the estimate, Spellman will contact the
customer to authorize the repair.

Factory Service Warranty

Spellman will warrant for three (3) months or balance of product warranty, whichever is longer, the repaired
assembly/part/unit. If the same problem shall occur within this warranty period Spellman shall undertake all
the work to rectify the problem with no charge and/or cost to the buyer. Should the cause of the problem be
proven to have a source different from the one that has caused the previous problem and/or negligence of
the buyer, Spellman will be entitled to be paid for the repair.

Spellman Worldwide Service Centers

For a complete listing of Spellman’s Global Service facilities please go to:


http://www.spellmanhv.com/customerservice/service.asp

101520-007 REV D