no load power consumption The TPP 65 Series of 65 Watt AC/DC power supplies feature a reinforced dou-
ble I/O isolation system according to latest medical safety standards (60601-1
• 5-year product warranty
3rd edition, 2 x MOPP). The leakage current is below 75 µA what makes the
units suitable for BF (body floating) applications.
The excellent efficiency of up to 92% allows a high power density for the stan-
dard 2.38” x 3.53” packaging format. The full load operating temperature
range is –40°C to +60°C while it goes up to 85°C with 50% load derating (for
single output models). The EMC characteristic is dedicated for applications in
industrial and domestic fields.
High reliability is provided by use of industrial quality grade components and an
excellent thermal management. It makes the products an ideal solution for
medical devices and for demanding safety and space critical applications.
Models
Order Code Output Output 1 Output 2 Output 3 Efficiency
Power Vnom Imax Vnom Imax Vnom Imax typ.
TPP 65-105 50 W 5 VDC 10'000 mA 90 %
TPP 65-112 12 VDC 5'420 mA 93 %
TPP 65-115 15 VDC 4'340 mA 94 %
TPP 65-124 24 VDC 2'710 mA 94 %
TPP 65-221 +12 VDC 5'420 mA +5 VDC 8'000 mA 90 %
TPP 65-231 65 W +15 VDC 4'340 mA +5 VDC 8'000 mA 91 %
TPP 65-251 +24 VDC 2'710 mA +5 VDC 8'000 mA 89 %
TPP 65-321M2 +12 VDC 5'420 mA +5 VDC 8'000 mA –12 VDC 600 mA 89 %
TPP 65-331M3 +15 VDC 4'340 mA +5 VDC 8'000 mA –15 VDC 600 mA 90 %
TPP 65-3512 +24 VDC 2'710 mA +5 VDC 8'000 mA +12 VDC 600 mA 89 %
Input Specifications
Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range)
Rated Range: 100 - 240 VAC (Full Range)
- DC Range Operational Range: 120 - 370 VDC (Designed for, no certification)
Polarity: +DC: L / –DC: N
Input Frequency Operational Range: 47 - 440 Hz
Certified: 50/60 Hz
Input Current - Full Load & Vin = 230 VAC 950 mA max.
- Full Load & Vin = 115 VAC 1'650 mA max.
Power Consumption - No load & Vin = 230 VAC 150 mW max. (Ready to meet ErP directive)
- No load & Vin = 115 VAC 150 mW max.
Input Inrush Current - At 230 VAC 60 A max.
- At 115 VAC 35 A max.
Input Protection T 3.15 A / 250 VAC (Internal Fuse in L & N)
Recommended Input Fuse (The need of an external fuse has to be assessed
in the final application.)
Output Specifications
Output Voltage Adjustment ±10% (only Output 1)
(By trim potentiometer)
Output power must not exceed rated power!
Voltage Set Accuracy ±1% max. (Output 1)
±2% max. (Output 2 and 3)
Regulation - Input Variation (Vmin - Vmax) single output models: 0.2% max.
dual output models: 0.2% max.
triple output models: 0.2% max.
- Load Variation (0 - 100%) single output models: 0.7% max. (5 VDC model)
0.5% max. (other output models)
dual output models: 0.5% max. (Output 1)
1.5% max. (Output 2)
triple output models: 0.5% max. (Output 1)
1.5% max. (Output 2)
0.7% max. (Output 3)
- Cross Regulation dual output models: 1.5% max.
(25% / 100% asym. load)
triple output models: 1.5% max.
Ripple and Noise - single output 5 VDC model: 75 mVp-p typ. (w/ 10 µF X7R)
(20 MHz Bandwidth) 12 VDC model: 75 mVp-p typ. (w/ 10 µF X7R)
15 VDC model: 75 mVp-p typ. (w/ 10 µF X7R)
24 VDC model: 75 mVp-p typ. (w/ 1 µF X7R)
- dual output 12 / 5 VDC model: 120 / 100 mVp-p typ. (w/ 10 µF X7R)
15 / 5 VDC model: 150 / 100 mVp-p typ. (w/ 10 µF X7R)
24 / 5 VDC model: 240 / 100 mVp-p typ. (w/ 10 µF X7R)
- triple output 12 / 5 / -12 VDC model: 120 / 100 / 120 mVp-p typ. (w/ 10 µF X7R)
15 / 5 / -15 VDC model: 150 / 100 / 150 mVp-p typ. (w/ 10 µF X7R)
24 / 5 / 12 VDC model: 240 / 100 / 120 mVp-p typ. (w/ 10 µF X7R)
Capacitive Load - single output 5 VDC model: 20'000 µF max.
12 VDC model: 4'520 µF max.
15 VDC model: 2'900 µF max.
24 VDC model: 1'130 µF max.
- dual output 12 / 5 VDC model: 2'500 / 3'000 µF max.
15 / 5 VDC model: 1'200 / 3'000 µF max.
24 / 5 VDC model: 625 / 3'000 µF max.
- triple output 12 / 5 / -12 VDC model: 2'500 / 3'000 / 500 µF max.
15 / 5 / -15 VDC model: 1'200 / 3'000 / 500 µF max.
24 / 5 / 12 VDC model: 625 / 3'000 / 500 µF max.
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Safety Specifications
Safety Standards - IT / Multimedia Equipment EN 62368-1
IEC 62368-1
UL 62368-1
- Medical Equipment EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents www.tracopower.com/overview/tpp65
Protection Class Class I & II (Prepared): Reinforced Insulation
Pollution Degree PD 2
Over Voltage Category OVC II
EMC Specifications
EMI Emissions EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 18 class B (internal filter)
- Radiated Emissions EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 18 class B (internal filter)
- Harmonic Current Emissions EN 61000-3-2, class A
- Voltage Fluctuations & Flicker EN 61000-3-3
EMS Immunity EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A
Contact: EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A
- EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A
L to L: EN 61000-4-5, ±1 kV, perf. criteria A
- Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A
- PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A
- Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
General Specifications
Relative Humidity 95% max. (non condensing)
Temperature Ranges - Operating Temperature –40°C to +85°C
- Storage Temperature –40°C to +85°C
Power Derating - High Temperature Depending on model
- Low Input Voltage 4 %/V below 90 VAC
See application note: www.tracopower.com/overview/tpp65
Cooling System Natural convection (20 LFM)
Altitude During Operation 5'000 m max.
Switching Frequency 50 - 140 kHz (PWM) (Output 1)
750 kHz typ. (PWM) (Output 2)
510 kHz typ. (PWM) (Output 3)
Insulation System Reinforced Insulation
Working Voltage (rated) 258 VAC
Isolation Test Voltage - Input to Output, 60 s 4'000 VAC
- Input to Case or PE, 60 s 2'500 VAC
- Output to Case or PE, 60 s 2'500 VAC
Creepage - Input to Output 8 mm min.
Clearance - Input to Output 8 mm min.
Isolation Resistance - Input to Output, 500 VDC 100 MΩ min.
Leakage Current - Touch Current 75 µA max.
(at 264 VAC)
Reliability - Calculated MTBF 1'500'000 h (for single output models)
1'000'000 h (for multi output models)
(MIL-HDBK-217F, ground benign)
Environment - Vibration IEC 60068-2-6
5 g, 3 axis, sine sweep, 5-500 Hz, 1 oct/min
- Mechanical Shock IEC 60068-2-27
50 g, 3 axis, half sine, 11 ms
Housing Material Aluminum
Housing Type Metal Case
Mounting Type Chassis Mount
Connection Type Screw Terminal
Weight - single output 172 g
- dual output 221 g
- triple output 221 g
Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf
REACH SVHC list compliant
REACH Annex XVII compliant
- RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf
Exemptions: 7a, 7c-I
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
- SCIP Reference Number b3f3b644-bdd9-4fcd-8836-8805a668dc5e
Supporting Documents
Overview Link (for additional Documents) www.tracopower.com/overview/tpp65
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Outline Dimensions
Voltage Adj.
Screw Terminal
44.45 (1.75)
40.8 (1.606)
1
Input (CON1) Output (CON2)
CON1
60.5 (2.38)
3
4
Pin Function Pin* Function
CON2
1 4-M3x0.5 1 Line 1,2 –Vout
3 Neutral 3,4 +Vout
10.0
(0.31)
8.0
78.7 (3.10) (0.39) 69.8 (2.748)
84.24 (3.317) *Terminal rated for 10 A max.
89.7 (3.53) (at higher current connection has to be split)
Max. corner screw penetration: 2.3 (0.09) CON1: Terminal Block
Max. center screw penetration: 2.0 (0.08)
mates with Screw locked torque max. 0.2 Nm / 2 kgfcm
All dimensions in mm (inch) Wire dimension range: 26 - 16 AWG
Tolerance: X.X ±0.5 (X.XX ±0.02)
X.XX ±0.25 (X.XXX ±0.010) CON2: Terminal Block
mates with Screw locked torque max. 0.2 Nm / 2 kgfcm
33.3 (1.31)
Screw Terminal
Voltage Adj.
6
40.8 (1.606)
1
CON1
60.5 (2.38)
CON2
3
Pin Function Pin* Dual Triple
1
4-M3x0.5 1 Line 1 NC Vout 3
3 Neutral 2,3 COM COM
10.0
(0.31)
8.0
The screw locked torque: mates with Screw locked torque max. 0.2 Nm / 2 kgfcm
1.2 (0.047)