Om Aa222new
Om Aa222new
Om Aa222new
Intended Use
The AA222 is a combined automatic impedance audiometer and
audiometer suited for clinics performing screening, as well as
diagnostic work. The test battery includes tympanometry, acoustic
reflex testing, ipsilateral and contralateral reflex decay testing,
Eustachian tube function test and AC audiometry.
Precautions
WARNING indicates a hazardous situation which,
if not avoided, could result in death or serious
injury.
2. Use only the disposable ear tips designed for use with this
instrument.
3. Use and store the instrument indoors only. Do not use this
instrument or its accessories in temperatures below 0°C/32°F
or above 50°C/122°F, or in relative humidity of more than
90%.
Explanation to Symbols
~ Alternating Current
This shall apply solely to the original purchaser. This warranty shall
not apply to any subsequent owner or holder of the product.
Furthermore, this warranty shall not apply to, and Interacoustics
shall not be responsible for, any loss arising in connection with the
purchase or use of any Interacoustics product that has been:
The equivalent volume of the middle ear, also called the compliance,
is easily derived by subtracting the two volume measurements above.
This is done automatically on the AA222 and the result is presented as
"Compliance", measured in ml.
Understanding Tympanograms
General Considerations:
A given curve drawn in a co-ordinate system will always have its
shape dictated by the vertical and horizontal graduations. The
printout of the AA222 complies with the international standards
in this respect, and therefore may not produce tympanogram
shapes directly comparable to other instruments if these do not
meet the standard requirements.
The Peak:
The peak of the tympanogram will horizontally be placed at the
air pressure of the middle ear, as equal pressures on both sides
of the tympanic membrane produces the highest compliance of
the system. A slight deviation of the peak in the direction of the
air pressure sweep may be experienced, due to an inherent
hysteresis of the middle ear and the test equipment. A slower
sweep speed may diminish the offset.
The Height:
The height of the tympanogram from its more or less horizontal
bottom line (measurements made from start pressure) to the top
shows the difference in compliance between stiffened tympanic
membrane and max. compliance. This difference is referred to
as "compliance" and is a measure for the equivalent volume of
the middle ear.
The Shape:
The shape of the tymp curve will change when the stiffness of
the system is changed (e.g. by ossicular chain disruption, otitis
media, etc.), and this is a primary reason for the diagnostic
value of this measurement. However, normal ears show a great
variety of tymp shapes so this should never be taken as the only
basis for making a diagnosis. Furthermore, two different
abnormalities may have opposing effects, resulting in a normal
shape of the tymp curve.
Classification of Tympanograms
Tympanograms can be classified according to compliance (height,
3
measured in ml. or cm ), pressure at compliance maximum
(measured in daPa), rate of compliance change (gradient in %), and
shape. Please refer to the chapter "Examples of Interpretations" in this
manual for illustrations of the classic curve categories, and the names
given to them by Liden and Jerger. On the following pages a more
detailed description of each category is presented.
Peak Pressure
Negative Normal Pressure: Positive Absence of
Pressure: Pressure: Pressure Peak:
Amplitude
Increased Decreased Amplitude: Unchanged Amplitude:
Amplitude:
1) Ossicular fixation 1) Blocked Eustachian
1) Eardrum bony or adhesive tube
abnormality 2) Serous otitis media 2) Early acute otitis media
2) Ossicular 3) Cholesteatoma,
discontinuity polyps, granuloma
4) Glomus tumours
Shape
........................Slopes....................... Smoothness:
Decreased / Flattened Increased slope: Altered smoothness:
slope:
1) Eardrum 1) Eardrum abnormality
1) Serous otitis abnormality 2) Ossicular
2) Ossicular fixation 2) Ossicular discontinuity
3) Tumours of middle discontinuity 3) Vascular tumours
ear 4) Patulous Eustachian
tube
Noise:
Acoustical signals showing up in the reflex recording, yet
irrelevant to the Stapedius reflex. As the reflex measurements
are based on observing the change in sound intensity of a 226
Hz tone, as explained in "Popular Introduction to Impedance", it
is possible that environmental noise of this frequency entering
the ear will show up as part of the test result. This is a problem
inherent to the measuring method and therefore common to all
normal impedance meters. Heart beat, talking and external
noises are common causes of noise peaks seen on the reflex
curve.
Negative Onset:
It is quite common to see reflexes start out with a small negative
deflecting dip. In ears with stapedial otosclerosis this dip with an
additional dip at the end of the stimulation can be the only
reaction left from the contraction of the Stapedius muscle. Some
tumour ears have been reported to give only the negative onset,
but no further reaction.
Reflex Threshold:
For a given stimulus the lowest level that elicits a detectable
reflex. This is not an absolute measurement as no exhausting
norm exits defining stimuli and related reflex characteristics.
Therefore, differences in test setups and reflex evaluation will
produce somewhat different results. It is not uncommon to
report the reflex threshold as the intensity which produces a 1%
or 2% change in equivalent volume (Test "A" with 2%
sensitivity).
Figure 3:
Air
Touch
Reflex Decay:
Reflex decay is calculated as the reduction of size of the
acoustic reflex during the first 10 seconds of muscular
contraction. Most normal ears will be able to maintain a
Stapedius contraction for 10 seconds or longer for frequencies
below 1000Hz, at a level 10dB above reflex threshold. A high
reflex decay score could indicate VIIIth nerve disorders.
Another way of stating the reflex decay is by finding the number
of seconds it takes before the Stapedius contraction falls to 50%
of its initial maximum.
The fact that the light band of the probe goes off, and the test
starts is not an absolutely certain indication of a fit good enough
to produce valid decay test results. This is due to the prolonged
test time and the high sensitivity of the test.
Also great care should be taken to have the probe kept in a very
fixed position relative to the ear during testing.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : No hearing loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present or absent.
Audiogram : Sensory neural hearing loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : (Abnormal Decay.)
Audiogram : Sensory neural hearing loss
(May be unilateral).
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : Normal.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent.
Audiogram : Conductive loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent (Present by contra-
lateral stimulation).
Audiogram : Conductive loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent / Present.
Audiogram : Conductive loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Present.
Audiogram : Normal.
Volume : Normal.
Pressure : Peak not obtainable.
Ventilation : -
Reflex : Absent or elevated (rare).
Audiogram : Mild to moderate conductive
loss.
Volume : Low
Pressure : Normal.
Ventilation : -
Reflex : Absent.
Audiogram : Mild to moderate conductive
loss.
Volume : Normal.
Pressure : -100 daPa to +100 daPa.
Ventilation : Present.
Reflex : Absent or elevated (rare).
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative / moderate.
Ventilation : -
Reflex : Absent.
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative / negative.
Ventilation : -
Reflex : -
Audiogram : Mild conductive loss.
Volume : Normal.
Pressure : Not obtainable.
Ventilation : -
Reflex : Absent.
Audiogram : Moderate conductive loss.
Volume : Normal.
Pressure : Negative.
Ventilation : Absent or poor
Reflex : Absent or elevated (rare).
Audiogram : Mild to moderate
conductive loss.
Hearing threshold levels can be determined using test tones with fixed
frequencies or a test signal with frequency varying with time according
to predetermined rate of change.
Air Conduction
The purpose of air conduction audiometry is to establish the hearing
sensitivity at various frequencies. The test can specify the AC loss, but
it cannot distinguish between abnormality in the conductive
mechanism or sensor-neural mechanism.
Headset Placement:
Remove eyeglasses and earrings if possible and position the
headband directly over the top of the head. Place the rubber
cushions so that the diaphragms are aimed directly at the
opening into the ear canal. Pull down the yokes of the phones
and adjust for tight fit. If the cushions are not tight to the ears,
the test result will be false at lower frequencies.
Background noise:
Background noise can also give false test results, especially at
lower frequencies. If necessary, the audiometric headphone can
be equipped with our Audiocup Exclosures type 21925. Please
contact the distributor.
Instruction of subject:
Prior to hearing threshold level measurements, the following
instruction of the subject about the test tones and the response
Threshold Determination:
The test normally starts at 1000 Hz on the patient's best ear.
Familiarisation:
Present a tone of 40 dB to the test ear. Often this tone is
sufficient to evoke a clear response from the subject. Then
present a tone completely attenuated. If the subject does not
respond to a tone of 40 dB present tones that are successively
10 dB higher until a response occurs. Then reduce the tone
level by 20 dB.
Bone Conduction
The purpose of the BC test is to supply the test tone directly to the
inner ear around the middle ear mechanism via the skull to establish
the inner ear hearing threshold.
AC-BC gap:
The difference between the AC threshold and BC threshold, the
so called AC-BC gap, will be the loss of the middle ear. The
middle ear loss is of great diagnostic interest as it may indicate
the need for medical attention.
The bone conductor is placed behind the test ear directly on the
skull where the tone is best heard. Activate "Bone R L" (11) and
the test is performed in the same way as for the air conduction
test.
The output switch is set to Bone and the test is performed in the
same way as for the air conduction test.
When the masking and tone are connected to opposite ears, the tone
is damped by the cross damping of the head (40-50 dB by AC using
normal headphones and 0-10 dB by BC). The actual damping of the
tone from the test ear to the masked ear will therefore depend on the
actual test: AC or BC ( as well as the frequency tested).
Cross damping:
The cross damping of the head for AC is 40-50 dB.
The cross damping of the head for BC is 0-10 dB.
With AC-BC loss in the masked ear: The AC tone level minus
50 dB plus the AC-BC loss of the masked ear. (AC dB - 50 dB +
AB gap of masked ear.)
If the ear has a middle ear loss, the masking intensity should be
increased by this loss as the masking to the masked ear is AC
sound, which of course will be damped by the AC-BC loss in the
masked ear.
If the masked ear has an Air Bone loss of 20 dB, the minimum
masking will be: 50 dB - (0 to 5) dB + 20 dB = 65 to 70 dB.
Maximum masking:
Maximum masking is the highest level of masking intensity
which does not change the true threshold of the test ear.
When ears with large AC-BC gaps are tested, the audiologist
must be aware of the masking dilemma where minimum
masking becomes over-masking without finding the plateau
where the masking is correct.
A) Find the unmasked thresholds for air conduction and then for
the better ear by bone conduction - see air and bone
conduction.
B) Notify the subject that masking will be used. Tell him that he will
be hearing some noise, but to disregard it as best as he can and
respond only to the test signal, as he did previously.
C) Select Tone and output: Left or Right, Bone L R. Dial the “HL
dB” (30) to the right and the masking noise will automatically be
switched on.- Set the “HL dB” (26) dial to the test ear at the level
of the unmasked threshold by means of rotary switch.
F) When the subject indicates that he hears the noise, the test
begins.
Begin the threshold shift procedure from this starting point (the
“HL dB” (26) dial set to the unmasked threshold level for the test
ear, and the masking at the point it is first heard by the subject).
Patient Communication
Talk forward
The operator may talk to the patient through the patient’s headset by
activating the “T.Fwd.” button (9). The goose neck microphone is
normally used even though an external microphone can be selected
for speech testing. The volume in the patient’s headset is adjusted by
turning the “HL dB” rotary dial (26) while holding down the (T.Fwd”
button (9).
Talk back
If the patient is equipped with a microphone connected to the talk back
input, then the operator may listen to the patient by pressing the
“Monitor TB” button (22) twice. To adjust the volume in the patient’s
headset hold down “Shift” (18) and the “Monitor TB” button (22) for 3
seconds. A volume bar will appear on the screen. Adjust the volume
by turning the “HL dB” rotary dial (26). Press “Store” (27) to save as
the default level.
Monitoring
By activating “Monitor“ (22), the presentation from e.g. tape can be
heard through monitor headset or built-in speaker connected to
“Monitor” on the connection panel in the accessory compartment.
To adjust the volume in the monitor headset or speakers, hold down
“Shift” (18) and the “Monitor TB” button (22) for 3 seconds. A volume
bar will appear on the screen. Adjust the volume by turning the “HL
dB” rotary dial (26). Press “Store” (27) to save as the default level.
Pause
The pause function operated from “Pause” (10) has been built in to
AA222 to prevent the instrument from automatically starting the test as
soon as an air tight fit has been obtained. The pause function is mainly
used when using the clinical headset or shoulder strap on the patient.
Patient Instruction:
Place the patient in a comfortable chair or on an examining
table if necessary. Small children may feel more comfortable
sitting on a parent's or nurse's lap. Show the probe to the patient
and then explain the following:
That the aim of the test is to test the mobility of the eardrum.
That the tip of the probe will be inserted into the ear canal,
and that it has to make a perfect seal.
That a small amount of air will flow through the probe to
move the eardrum; this will produce a sensation equal to
pressing a finger slightly into the ear canal.
That one or more tones will be heard during the test.
That no participation is expected from the patient.
That coughing, talking and swallowing will ruin test results.
Visual Inspection:
Check the external ear canal for wax with an otoscope and
remove excessive wax to prevent probe opening from clogging
which will inhibit testing. Excessive hairs may have to be cut.
Picture 1:
Pull
Push
The other probe is then pushed into the socket until a distinct
"click" is heard. Please note that the squared hole in the probe
connection must face in the same direction as the push-button.
D) Insert the probe tip into the ear canal until you have an
airtight fit. Pull the transducer house upwards again, until the
probe in the ear is hanging by the tubes from the transducer.
Figure 5:
Hair coming out of the ear canal may make an airtight fit difficult
to obtain. Therefore, ideally, excessive should be removed prior
to testing. Also, a clinical ear tip covered with Vaseline may be
helpful. Make sure the ear tip does not have its opening closed
by the wall of the ear canal or clogged by Vaseline or cerumen.
For decay test an absolutely airtight fit is needed to avoid test
results indicating decay, where the test result is actually just an
artefact caused by a poor fit.
Figure 6:
From the main menu it is possible to adjust the contrast of the display
by means of F5 “LCD-“ and F6 “LCD+”. The instrument will always
remember the last setting. When holding down “shift” (18) the writing
above F5 and F6 changes to “LED-“ and “(LED+)”. Now the brightness
of the LEDs (indication lights) of the instrument keys can be adjusted.
Also here the instrument will remember the last setting.
Test Procedures
Before every test is performed it is recommended to delete all stored
patient data from the instrument. Hold down “shift” (18) while pressing
“new subject” (10). The following message will appear on the screen:
Figure 7:
Printout 1: Printout 2:
Printout 1:
Printout 1 contains the following:
Tympanogram.
Reflexes (sequence method with individual reflexes starting at 80
dB at 4 different frequencies).
ETF 1 test (Eustachian Tube Function test).
Printout 3: Printout 4:
2. Place the new paper roll in the paper compartment in such a way
that the paper ascends from the lower part of the paper roll. With a
pair of scissors cut two triangles of paper away. This will make the
feeding of the paper easier – see picture 4.
Picture 4: Picture 5:
4. Gently insert the paper between the lower part of the black rubber
platen and the black plastic – see picture 6.
Picture 6: Picture 7:
5. When the paper appears between the upper part of the rubber
platen and the paper cutter, pull out an extra 10 to 15 centimetres
of paper and return the small blue lever to its original position.
6. Now guide the paper through the slot of the cover of the paper
compartment.
Figure 9:
The probe will have an airtight fit when the indication light on the
probe is green. When the status “Leaking” or “Blocked” is displayed
in the upper right hand corner of the screen under the Status Bar, the
indication light on the probe will be yellow. This indicates a bad fitting
or a blocked ear tip.
On the screen the tymp test can be followed on-line. When the tymp
curve has been drawn, the AA222 will automatically continue with the
selected reflex test “Test A” or “Test B”.
4. Remove the probe from the first ear when the test
sequence has finished. Select the other ear “Right” (27) or
“Left” (28) and move the probe to the second ear and
repeat the test sequence once again.
5. To print the test result, simply press “Print” (8).
Figure 10:
Figure 11:
Figure 12:
Figure 13 shows the Modify Menu for Reflex Test A – it is similar for
Reflex Test B:
Figure 13:
When the individual reflex for Test A or B has been modified to the
special personal needs, hold down “shift” (18) while pressing the F-
key underneath the highlighted reflex once again.
In the same way all the reflexes for Test A or B can be modified.
Finally, press “Back” (7) to return to the test screen.
The idea is that light indication in the “Ipsi” (12) and “Contra” (13)
keys indicates whether or not the ipsilateral and / or contralateral
reflexes will be performed in a preceding test session.
Figure 14:
3. Remove the probe from the first ear when the test
sequence has finished. Select the other ear “Right” (19) or
“Left” (20) and move the probe to the second ear and
repeat the test sequence once again.
Extend Function
In the Tympanometry mode it is possible to extend the
pressure range to go from +300 daPa to –600 daPa by
pressing F6. When the Extend Function is active “Extend” in
lower right hand corner of the screen is highlighted.
Figure 15:
The Tympanometry mode for High Probe Tone looks very similar to
the normal Tympanometry mode; however, the following differences
can be seen on the screen:
Figure 16:
Figure 17:
In the Reflex Test mode the right indication light only on the
“Tymp/Reflex” key (10) will be active.
The probe will have an airtight fit when the indication light on the
probe is green. When the status “Leaking” or “Blocked” is displayed
in the upper right hand side of the screen the indication light on the
probe will be yellow. This indicates a bad fitting or a blocked ear tip.
When a correct fit has been obtained the reflex test will begin.
Place the clinical headset over the patient's head or use the
shoulder strap and insert the probe tip into the ear canal.
When an airtight fit has been obtained the reflex test is
ready to be performed. To start the test, select "Pause" (9)
once again and the test will be carried out automatically.
3. Remove the probe from the first ear when the test sequence
has finished. Select the other ear “Right” (19) or “Left” (20),
move the probe to the second ear and repeat the test
sequence once again.
Figure 18:
When ready, press the “Tone Switch” (29) and the manual reflex will
be performed and appear on the display. The parameters for the
manual reflex can be changed once again if the result is not
satisfactory. When satisfied, press “Store” (24) to store the manual
reflex next to the automatic reflexes in the lower right hand corner of
the screen and AA222 will automatically prepare for performing
another manual reflex.
Figure 19:
Just above the F1 and F2 keys there are two small arrows pointing to
the left and the right hand respectively. By pressing the F1 and F2
keys it is possible to browse between the recorded reflexes.
Now an empty and highlighted reflex box will appear and below it the
default intensity and frequency will be highlighted like in figure 20.
Also “Decay” above the F5-key will be highlighted.
Note: To ensure the right basis for performing a reflex decay test it is
necessary to find the reflex threshold for the frequency to be tested.
When the reflex threshold has been found 10 dB must be added to
this threshold to get the correct starting point.
When ready, press the “Tone Switch” (29) and the manual reflex
decay will be performed and appear on the screen. The parameters
for the manual reflex can be changed by pressing the “Frequency
Decr / Incr” keys (27 and 28) and dialling the “HL dB” rotary dial (26).
Figure 22:
When a reflex decay test has been accepted and stored it will be put
into one of the small reflex boxes. A small “D” (see figure 23) will
indicate that a Decay test has been performed.
Figure 23:
To choose the Eustachian Tube Function test press the “ETF” key
(11).
In the Eustachian Tube Function test mode the indication light on the
“ETF” key (11) will be active and at the same time the indication light
on the “Pause” key (9) will start blinking in order to tell the operator
that his/hers action is now needed.
Figure 24:
To perform the test the operator will have to press the blinking
“Pause” key (9) to make a tympanometric curve. After the first sweep
an information window appears on the screen:
Figure 25:
Now press the blinking “Pause” (9) key once again and the test will
make its second sweep. When the second sweep has been carried
out another information window will appear on the screen:
Press the blinking “Pause” (9) key once again as told in the
information window and the test will make its final and third sweep.
If the three peak pressures are the same, it has not been possible for
the patient to operate his Eustachian tube to change the middle ear
pressure. This means that the Eustachian tube is temporarily or
permanently out of function.
Note: From the Common Setup Menu it is possible to set up the
“Pause” key (9) to be controlled from the Remote Switch of the
probe.
”Child” Function
AA222 has been equipped with a special Child Function. The
intention of this function is to help the operator keeping the child
concentrated for the short while it takes to perform the test.
Figure 27:
The train will continue moving until the selected reflex test session
has finished. When the test session has finished the train will
automatically disappear.
Figure 28:
When the audiogram for the first test ear has been completed
the test will automatically continue testing the other ear.
In the horizontal bar above the right and the left audiogram it is
possible to follow the test procedure see figure 25. To the left
there is a small circle, which is highlighted whenever a tone is
audible to the patient. Next to the circle there is a small
rectangle, which is highlighted whenever the patient presses
the patient response button. Finally, the present dB value and
frequency is expressed numerically.
Figure 28:
Manual Audiometry:
To perform manual audiometry first select “Audiometry” (11)
and then press “Man/Auto” (17). In the Manual Audiometry
Test mode the left indication light of the “Man/auto” key (17)
will be active. Now, manual audiometry can be performed.
Select frequency by pressing “Frequency Decr / Incr” (27 and
28) and intensity by dialling the “HL dB” rotary dial (26).
To store thresholds for the audiogram press “Store” (24).
Figure 30:
Insert Masking:
In cases where danger of over masking exists, masking by
insert phone is recommended. This will improve the cross
hearing of the masking sound from the approximately 40 dB of
a traditional headphone to the approximately 70 dB of the
CIR33 insert phone.
Speech Audiometry
When the instrument is switched on it automatically powers-up in
Figure 31:
Figure 32:
While presenting live speech into the built-in microphone adjust the
VU meter so that maximum deflection reaches “0” indication on its
scale on the points of maximum intensity of the spoken words (see
arrow A). The adjustment is carried out by dialing the “HL dB” rotary
dial (26). To store the VU setting press “Store” (27).
Figure 33:
B
A
In case two CD inputs are used (e.g. one for recorded speech and
one for recorded noise) both inputs must be calibrated. When being
in the message box this is simply done by selecting CD1 or CD2 (see
arrow B) by pressing “CD1/CD2” (8) and store the adjusted VU
setting by pressing “Store” (27).
In order to see the two stored VU settings hold down “shift” while
pressing “CD1/CD2” (8).
Figure 34
Number of Words
The number of words to be tested must be entered in the
Speech Setup.
Number of Curves
In addition to the permanent benchmark reference curve it is
possible to record up to 4 speech curves. In order to select and
browse between speech curves 1 to 4 simply press F5.
Man/Rev
You may choose between manual or reverse speech
presentations simply by pressing “Man/Rev” (F3).
Pulse
You may choose pulsing speech for the speech presentation
simply by pressing “Pulsing” (F4).
Figure 35:
The patient must be informed that it is only the loudness of the tone
loudness and not the character of the tone that should be considered.
Stenger Test
The Stenger test is a test for malingering based on the auditory
phenomenon of referral to the ear in which the sound appears
loudest (the Stenger effect).
SISI Test
The Short Increment Sensitivity Index is designed to test the
ability to recognise one decibel increase in intensity during a
series of pure tones bursts presented 20 dB above threshold.
Figure 37:
If more frequencies have been tested during the SISI test procedure
it is possible to see them all in the display at the same time by
pressing F6 “Table”.
Figure 38 shows tables of the SISI scores for right and left ear.
Figure 38:
We recommend using a new ear tip for each patient. The ear tip is
not designed for reuse and an appropriate level of cleanliness cannot
be guaranteed even if recognized disinfecting procedures are used.
There is also a risk of material decomposition caused by the cleaning
agent. Only with new and original Interacoustics supplies is the
measuring accuracy guaranteed. Supplies were designed specifically
to obtain optimal measuring results and non-original supplies may
compromise results.
To clean the small acoustic and air pressure channels of probe tip
unscrew the small ribbed plastic nut that holds the probe tip:
Picture 8:
To clean the small acoustic and air pressure channels of probe tip
unscrew the probe cap that holds the probe tip:
After unscrewing the probe cap it is possible to detach the probe tip
with the small acoustic and air pressure channels from the
transducer house. Out of the probe tip you can take temporarily
remove the sealing gasket.
Picture 10:
The cleaning of the acoustic and air pressure channels of the probe
tip must be performed by means of the cleaning wire (nylon wire)
which is part of the cleaning tool that can be found in the Ear tips
Assortment provided with the AA222.
When cleaning the acoustic and air pressure channels of the probe
tip the cleaning wire must be inserted from the back of the probe tip
according to picture 11:
After cleaning all the acoustic and air pressure channels of the probe
tip can be reassembled. Place the sealing gasket onto the transducer
house like shown in picture 12 and make sure that holes of the
gasket are aligned with the holes in the holes of the transducer
housing.
Picture 12:
Picture 13:
9) Tymp/Reflex Selects:
1. Tympanometry.
2. Reflex and Tympanometry.
3. Reflexometry.
Medical CE-mark
The CE-mark indicates that Interacoustics A/S meets the
requirements of Annex II of the Medical Device Directive
93/42/EEC. TÜV Product Service, Identification No. 0123,
has approved the quality system.
Power:
Consumption: 15VA
Mains voltage/fuses: 100-240V~, 50-60Hz, 0.5A max.
Operation Environment:
Temperature: 15 – 35 C / 65 – 95 F
Rel. Humidity: 30 – 90%
Printer (Optional)
Type: Thermal printer with recording paper
in rolls.
Printing Time: Depending on the test printed.
Compliance
Range: 0.1 to 6.0 ml (Ear volume: 0.1 to 8.0
ml). No difference is measured in the
recorded values between static and
dynamic measuring mode.
Types
Tympanometry: Automatic, where the start and stop
pressure can be user-programmed
from the setup menu.
Eustachian Tube
Function: Williams test (automatic
function).
Memory
Tympanometry: 1 curve per ear.
Eustachian Tube
Function: 3 curves per ear.
Calibrated transducers
Audiometry headset: Telephonics TDH/DD45 with a static
force of 4.5N 0.5N
Bone conductor: Radioear B71 with a static force of
5.4N0.5N
Definition of units
1 ml at 226 Hz = 1 acoustic mmho,
-8 3
1 acoustics mho = 10 m /Pa·s
1 daPa = 10 Pascal
Inputs
CD1/2: Connection for CD player or tape
recorder.
Mic: Connection for external microphone
for live speech.
Talk Back: Connection for talk back
microphone.
Patient Response : Patient response switch connection.
Output
Audiometry Earphones: Audiometric headphone, left and
right.
Bone Conduction : B71 bone conductor.
Free Field 1 and 2: Electrical output for external 2
channel power amplifier (AP12 and
AP70).
Monitor: Monitor earphone disconnects the
internal monitor loudspeaker when
inserted. Operator can monitor signal
when presented to the patient as well
as the talk back signal from the
patient’s microphone.
Contra Earphone: Audiometric headphones for reflex
and audiometry measurements.
Ipsi Earphone: Probe earphone incorporated in
Attenuator
Range: 0 to 130 dB in 1 or 5 dB steps.
Typical range is –10 to 120 dB HL.
Range is individual for different
modes – see table 1.
Test Types
Manual Audiometry: Manual control of all functions.
Automatic Audiometry: Auto threshold according to ISO
8253-1 (Patient controlled Hughson
Westlake). Threshold is determined
by the activation of the patient
response.
Properties of stimuli
General
Specifications for stimulus and audiometer signals are made to
follow IEC 60645-1
Contralateral Earphone
Pure tone: ISO 389-1 for audiometric headphone
Wide Band noise (WB): Interacoustics standard
- spectral properties: As “Broad band noise” specified in IEC
60645-5, but with 500 Hz as lower cut-
off frequency.
Low pass noise (LP): Interacoustics standard
- spectral properties: Uniform from 500 Hz to 1600 Hz, ±5
dB re. 1000 Hz level
Hign pass noise (HP): Uniform from 1600 Hz to 10 kHz, ±5
dB re. 1000 Hz level
Ipsilateral Earphone
Pure tone: Interacoustics standard
Wide Band noise (WB): Interacoustics standard
- spectral properties: As “Broad band noise” specified in IEC
60645-5, but with 500 Hz as lower cut-
off frequency.
Low pass noise (LP): Interacoustics standard
- spectral properties: Uniform from 500 Hz to 1600 Hz, ±10
dB re. 1000 Hz level
Frequency Reflex
Contralateral Ipsilateral
Audiometric EAR-Tone 3A Insert/CIR33
headphone
Min Max* Min Max* Min Max* Min Max*
Hz dB HL dB HL dB HL dB HL dB HL dB HL dB HL DB HL
250 -10 110 -10 105 0 100 - -
500 -10 120 -10 110 0 105 10 105
1000 -10 120 -10 120 0 110 10 110
2000 -10 120 -10 120 0 105 10 105
3000 -10 120 -10 120 0 100 10 100
4000 -10 120 -10 115 0 95 10 100
6000 -10 120 -10 100 - - - -
8000 -10 110 -10 95 - - - -
WB noise -10 120 -10 120 0 100 10 105
LP noise -10 120 -10 120 0 100 10 105
HP noise -10 120 -10 120 0 100 10 105
Minimum level step size 5 dB
Maximum difference error between tow steps is less than ±3 dB
Residual noise is less than 25 dB (A)
Signal to noise ratio is larger than 80 dB (A)
ON-OFF ration is larger than 80 dB (A)
Rise-/fall time: 35 ms (±10 ms)
Outputs:
Phones, Left/ Jack, 6.3mm Voltage: Up to 5.5V rms. by 10 load
Right mono Min. load impedance: 5
Bone conductor Jack, 6.3mm As for Phones,
mono Left/Right
Free Field 1, 2 Phono Voltage: Up to 8.0V rms. by 100 load
Min. load impedance: 100
Freq. response: 75-12000 Hz +/-3dB
Output impedance: 500 ohm
Monitor Jack, 6.3mm Voltage: Up to 2.0V rms. by 8 load
mono Min. load impedance: 0
Phones, Jack, 6.3mm Voltage: Up to 5.5V rms. by 10 load
Contralateral mono Min. load impedance: 5
(Insert masking)
Parts
Included Parts:
Audiometric headphone Single contralateral headset
CIR33 Insert earphones
B71 Bone conductor
APS2 Patient Signal
ATP-AT235U Universal probe system with shoulder strap and
wrist strap (tubing 360mm Ø1.3xØ3.3 transparent silicone)
TPR26 3 rolls of recording paper
Power cable
Dust cover
Operation Manual.
Multilingual CE instructions for use
Additional Parts:
EARtone 5A Audiometric Insert Phones
50250 Peltor noise exclosures
21925 Amplivox noise exclosures
ACC400 carrying case
CAT50 calibration unit 0.2-0.5-2.0-5.0 ml
IES impedance ear simulator
GSE10 RS232 Galbanic isolation adapter
USB cable 2m black
Connection for
audiometric Contra Phone
Connection to headphone connection
mains Phones
Connection for
Connection for
FF amplifiers
Probe System
Please note that the built-in test cavities are intended for a daily
check only and not for calibration purposes.
Contents of Shipment
When AA222 is delivered as a standard unit the case contains the
following:
1 Instrument AA222
1 Audiometric Headset Audiometric
headphone
1 Bone Conductor B71
1 Patient Signal APS2
1 Universal Probe System ATP-AT235U
3 Rolls of Recording Paper TPR26
1 Power Cable -
1 Box of 65 Assorted Eartips BET55
1 Operation manual
1 CE-manual
Reporting Imperfections
Inspect before connection:
Prior to connecting AA222 to mains it should once more be
inspected for damage. All of the cabinet and the accessories
should be checked visually for scratches and missing parts.
Annual calibration:
The AA222 has been designed to provide many years of
reliable service, but annual calibration is recommended due to
possible impact on transducers.
Normally, only the indication light of the “Ipsi” key (12) will be
active informing that only the ipsilateral reflex test will be
performed during the subsequent test procedure. Press the
“Contra” key (13) to activate also the contralateral reflex test
for the subsequent.
What is Gradient?
Gradient is explained in the chapter “Tympanometry Setup
Menu” of this manual.
Biswas, Anirban:
Clinical Audiovestibulometry, (Bhalani Medical Book House,
th
Bombay, India 4 edition 2009)
Brask, T.:
Extratympanic Manometry in Man. (Scandinavian Audiology,
supp. 7. 1978)
Fiellau-Nikolajsen, Mogens:
Tympanometry and Secretary Otitis Media. (Acta Oto-L. 1983)
Jerger, J.:
Clinical Experience with Impedance Audiometry. (1970)
Katz:
Handbook of Clinical Audiology, Fourth Edition 1994 (Williams
& Wilkins 1985)
Liden, G. et al.:
International Symposium on Impedance Audiometry and
Pediatric Audiology Göteborg 1982. (Scandinavian Audiology
supp. 17. 1983)
Liden, G.:
Audiology (Almqvist & Wiksell. 1985) (Swedish language)
Popelka, G. R. et al.:
Hearing Assessment with the Acoustic Reflex. (Grune &
Stratton 1981)
Acoustic Compliance:
Another term for Acoustic Admittance.
Acoustic Immittance:
Refers collectively to acoustic impedance and / or acoustic
admittance.
Compliance:
1) Ease with which air moves (e.g. influenced by the
eardrum and middle ear mechanism).
2) Often used to indicate the equivalent volume of air in
the middle ear.
Ear Tip: A cuff which is used to seal the probe into the external
auditory canal.
Myringoplasty:
Surgical repair of the eardrum membrane.
Myringotomy:
(tympanotomy) A small incision made in the eardrum
membrane to remove fluid from the middle ear.
Pascal (Pa):
2
A unit of pressure or stress, equal to one Newton per m .
Probe:
A coupling device that is inserted into the external auditory
canal, to connect it to the acoustic immittance meter.
Probe Ear:
The ear into which the probe is inserted.
Probe Signal:
An acoustic signal that is emitted into the external auditory
canal by means of a probe. The signal is used to measure
acoustic immittance.
Stimulus Ear:
The ear to which the reflex activating stimulus is presented in
order to elicit a middle ear muscle reflex. Note: If a bone
vibrator or a loudspeaker is used to deliver an acoustic reflex it
may not be possible to define the stimulus ear.
Toynbee Test:
Test designed to determine the function of the Eustachian tube
in ears with perforated eardrums.
Toynbee's Manoeuvre:
See Valsalvation.
Tympanogram:
A chart of the results of tympanometry - compliance
measurements at the eardrum.
Tympanometry:
The measurement of the ability of the eardrum and ossicular
chain to transmit sound pressure waves. An intact eardrum is
subjected to air pressure changes to determine its stiffness
(impedance) and compliance (admittance).
Valsalva's Manoeuvre:
Blowing forcibly to open Eustachian tube by holding nose and
closing mouth. Named for its originator, Antonio Valsalva.
Sometimes called Valsalva's experiment.
Williams Test:
Test designed to determine the function of the Eustachian tube
in ears with non-perforated eardrums.
Figure 39:
From the Main Setup Menu it is possible to enter the following Setup
Menus by pressing F1 to F5:
Tympanometry Setup
Setup for Reflex Test A and B
Common Setup
Clinic Setup
In the individual setup menus navigation through the setup points and
changing parameters in the individual setup points is very simple.
With the F-keys (F2 to F5) representing arrow keys “, , , ” it is
possible to browse through the different setup points. Parameters
can be changed with “Change” (F1) when it is highlighted.
To leave the individual setup menus press “Back” (7). This key will
always take you back one level. If one or more parameters have
been changed the message in figure 36 will appear on the screen.
To confirm the changes press F1 for “Yes” or F6 for “No”.
Figure 41:
Start Pressure:
Indicates the starting point for the pressure sweep for the
tympanometric curve. It goes from 25 daPa to 300 daPa.
Stop Pressure:
Indicates the pressure where the sweep for the tympanometric
curve will stop.
A slow speed will be more time consuming, but may give more
detailed information. The horizontal displacement of the
tympanometric curve’s peak in the sweeping direction caused
by inherent hysteresis of the system and the middle ear itself
will be smaller with lower speed.
Compensated Mode:
Figure 42: Compensated Figure 43: Non-Compensated
Gradient Unit:
The Gradient unit is an expression of the shape of the
tympanometric curve (narrow or wide).
If ml is selected the program will go 50 daPa to each
side from compliance value and save the two ml
values. The gradient is the average of the two ml
values subtracted from the compliance value.
If daPa is selected the gradient will be calculated the
following way: The program searches from each end
of the tymp curve and stores the pressure at the point
where the tymp curve is equal to compliance value
divided by 2. Gradient is equal to the two stored
pressures subtracted from one another.
Tympanometry Scale:
For the tympanic tests is defined which is the default scaling. If
needed, you can set the scaling differently in the test screen by
pressing button F5 (5).
Figure 44:
Figure 45:
The four arrow keys (F2 to F5) browses between the parameters of
the individual reflex.
Output:
It is possible to select between ipsilateral and contralateral
reflexes as output.
Stimuli Frequency:
For ipsilateral reflexes the stimuli frequency options are: 500,
1000, 2000, 3000 and 4000 Hz, including also the noise stimuli
WN, LP and HP.
Level:
The intensity level for the reflexes goes from 60 dB to
maximum. Maximum depends on the selected frequency and
the selected transducer. In the final line of the table it is
possible to switch on or off the reflexes according to the
number of reflexes needed.
Reflex Methods
There are four different reflex methods for Reflex Test A and B:
Reflex Sensitivity:
To fully understand the reflex search procedure of the
following three Reflex Methods it is important to know about
the Reflex Sensitivity idea. Reflex Sensitivity appears under
the Reflex Method line when Screening, Auto or Sequence
Method are selected.
The software of AA222 uses an algorithm to determine if a
reflex is acceptable or not. There is a selection of three
different algorithms to choose from, each representing its
individual level of Reflex Sensitivity. These three Reflex
Sensitivities are referred to as Sensitive, Normal and Robust
and they are expressed in ml.
Note: The pass box is not shown in the screen of AA222 but
serves as an example of what sensitivity is and how it is used.
Figure 47: Sensitive (0.03 ml) Figure 48: Normal (0.05 ml)
Figure 49: Robust (0.08 ml) Figure 50: Robust (0.08 ml)
The first three reflexes are all examples of reflexes that have
been accepted according to the different algorithms (sensitive,
normal and robust). They are accepted because they do not
touch the pass box. However, figure 50 shows an example of
a reflex that is not accepted according to the chosen algorithm.
The reflex is not accepted because it goes through the pass
box.
Figure 51:
Auto (5 dB steps):
Figure 52 shows the screen for the Auto (5 dB steps) Reflex
Method. The Auto test is used for testing where only one reflex
is needed per frequency, but a reflex of a certain size is
preferred (see “Reflex sensitivity” above). The Auto test then
searches for suitable stimulus intensity before each reflex is
tested.
Figure 52:
4) If the reflex found the second time does not meet the pass
criteria either, a new reflex is recorded at 5dB higher
intensity. Consecutively, higher intensities are used until a
reflex is accepted (if a pass cannot be obtained, the reflex
recorded at maximum intensity is displayed as the test
result for the test).
Sequence:
Figure 53 shows the screen for the Sequence Reflex Method:
Figure 53:
Figure 54:
Level Increase:
Selects how many dB the intensity will be increased between
each single reflex. Choose between 5, 6, 8, 9, or 10.
Figure 55:
Default Intensity:
Default Intensity is used to select the step-down value of the
attenuator when the frequency is changed and can be set from
–10 dB to 50 dB in steps of 5 dB and Off. If the value is set to
“Off”, the intensity will not change when changing transducer.
Masking Intensity:
It is possible to set the default masking intensity between 0 and
50 dB in steps of 5 dB. The masking intensity can also be set
to “Turn off”. In this situation when changing test ear the
masking will automatically turn off. The masking intensity can
also be set to “unchanged”. In this situation when changing test
ear the masking will automatically stay the same.
Bone at 8 kHz:
Selects whether or not it should possible have 8 kHz as output
when using the Bone Conductor.
Microphone Selection:
Selects between the built-in swan neck microphone and an
external microphone.
Output Ch2:
Selects between Narrow Band and White Noise as masking for
output in channel 2.
Figure 59:
Audiometer Frequencies:
It is possible to disable/enable one or more of the following
frequencies: 125 Hz, 250 Hz, 750 Hz, 1500 Hz, 3000 Hz, 6000
Hz and 8000 Hz. To browse between the frequencies press
“Decr Incr Frequency” (27 and 28). The individual frequency is
set to “On” or “Off” by pressing F1 “Change”.
Pulse Length:
It is possible to select the pulse length. The pulse length can
be changed from 250 mS to 5000 mS in steps of 50 mS. When
multipulse is selected in the above Pulse Key Function the
tone presentation and intermitted pause is equal.
Level Decr/Incr:
Selects 1 or 5 dB increments or decrements when dialling the
“HL dB” rotary key (26) between 1 dB and 5 dB.
Figure 60:
Number of words:
Here you can assign the number of words in your word list.
This will make the basis of the result obtained via the
automatic speech score counter.
Score Method:
It is possible to select between Default or German method.
Percent 100
Words selected in Item 21
L/R Standard Curves and FF1/FF2 Standard Curves:
The speech audiogram holds standard curves indicating
normal hearing of one-syllable words and two-syllable words.
Standard curves for headphone testing as well as Free Field
testing are available.
Figure 61:
Pause:
If “On” is selected the “Pause” key (9) is pre-selected as
default for 1) the Tympanometry Test, 2) the Reflex Test and
3) the Reflex and Tympanometry Test.
Communication:
The setting is USB. It cannot be changed.
Remote Switch:
Selects the function of the Remote Switch of the probe. Select
between “Off”, “L/R”, “Pause” and “L/R or Pause”.
1. When “Off” is selected the Remote Switch is out of function.
Printer:
There are two different “Printer” settings; “Off” and “Internal”.
Language:
It is possible to select between “English” and “German”.
Insert the CD, if the PC is not connected to the internet, click Next.
Click Finish, the Found New Hardware Wizard now starts over again
because a driver for Serial Converter B needs to be installed, follow
the directions above.
The Driver for the USB Serial Converter is now installed, to find out
which COM port to use when communicating with the instrument start
the Device Manager (Click Start, My Computer -> properties,
Hardware -> Device Manager). The Port is recognized as “USB
Serial Port” use the one with the lowest number.
Before the connection to the mains network, be sure that the local
mains voltage corresponds to the voltage labelled on the
instrument. Always disconnect the power cord if the instrument is
opened or by control / replacement of the mains fuses.
For maximum electrical safety, turn off the power from a mains
powered instrument when it is left unused.
Do not site the instrument next to a heat source of any kind, and
allow sufficient space around the instrument to ensure proper
ventilation.
1 2 3 4 5 6 7
8 9 10 11 12 13 14 15 16 17
18 19 20 21 22 23 24 25
27 28 29 30
26