ErbaLyte ServiceManual
ErbaLyte ServiceManual
ErbaLyte ServiceManual
ERBALYTE
Company Contact Details:
ERBA MANNHEIM
Mallaustrasse 69-73
68219 Mannheim Germany
Tel : (49-6-21) 8799770
Fax : (49-6-21) 8799688
Preface
In order to operate and service the instrument properly, please read the
When receiving the goods, please check the product model and spare parts
against the packing list. If missing or mistake, please contact Erba immediately.
Never connect the plugs of the rear panel with electric metal materials.
before open it. Do not remove any symbols from the instrument.
The waste may contain materials harmful to human body. It should be discarded
-i
-
Contents
- ii
-
Chapter 1. Instrument installation
1. Power switch
4. Grounding pole
5. Machine label
-
2-
1.2 Environment requirements
1. The instrument should be installed on a stable and solid platform that is free of
2. The environment should be as free as possible from dust, corrosive gas, loud noises
3. Avoid placing the instrument in direct sunlight or in front of a source of heat or vent.
1. Power cable: should be connected to the power socket firmly and reliably
2. Grounding: the grounding pole should be connected to ground properly. The grounding
wire should be as short as possible and the grounding resistance must be less than 10Ω
3. If the power supply to the instrument is not stable, or there are high-power equipments
on the same line, then a power stabilizer (output ≥100W) is strongly suggested.
4. Avoid operating the instrument under high humidity environment for long
time, otherwise the instrument performance may be affected. Use desiccant or air
conditioner in case of necessary. Do not place the instrument near a water pool.
1. Pump tube installation: put the aspirating pump tube and the lactic pump tube
on the pump head, fix the tube connectors to the bracket, adjust the tube to the middle
-
2. Check the electrode assembly: take out the electrode assembly from the accessory kit.
Check if the setting screws are tightened, inlet and outlet port are clean, the filling
3. Installation of the electrode assembly: insert the whole assembly to the mounting base,
connect the
correctly. Connect the tube to the inlet and outlet port of the assembly.
4. Check the liquid positioner: replug the positioner and make sure it is well connected.
5. Check the tubing: make sure all the tubings are connected tightly and there is no leaking.
6. Connect to the reagent pack: connect the Std A tube to port A of the reagent pack.
Connect the Std B
tube to port B. Connect Std R tube to port R Connect the waste tube to port W.
7. Installation of printing paper: open the printer cover, insert the printing paper into the guide
slot (the thermal side face down), feed the paper out from the slot on the printer cover.
Note:
Make sure all tubings are connected firmly and there is no leakage.
Make sure all the lead wires of the electrodes are connected firmly.
The printing paper is using out when you see two red lines on it.
When switched on the power to the instrument, a system self-test will be performed
Note:
Try the other side of the printing paper or change the whole paper roll if there is no
should adjust or change the probe and make sure it can move into the wash block
correctly.
connections, reagent connections and electrode connections and make sure there
is no blockage or leakage.
Activate the electrode for 10-30 minutes before using if the electrode has
been stored or not used for a long time. Run calibration when electrode is stable.
-6-
2.2 Prompts during self-test and calibration
Error
Code message
1 The optical coupler of liquid distribution valve failed
#
2 Elevator switch failed
#
Error
Code message
Error 0# Liquid positioning failed
-7
-
Chapter3. Functional description and operation
Press Service in the main menu, press Time and enter “983”, press Exit, then the system
will enter into
(Attention: the extended programs are not open for the end-users to avoid wrong setting)
Press Pr.Vol , the button turns white. The voltages of the liquid positioner and all
electrodes will be printed out during calibration or sample measurement.
Press Pr.Vol again, white color disappear from the button. The voltages will not be
printed out
-8
3.3.2 Print PH value
Press Pr.PH , the button turns white. The pH value will be printed out when the sample
measurement is completed.
Press Pr.PH again, the button turns dark. The pH value will not be printed out when the
sample measurement is completed. This is the default setting.
Press Pr.TCa , the button turns white. The value of iCa and TCa will be printed out
when the sample measurement is completed.
Press Pr.TCa again, the button turns white. Only the value of iCa will be
printed out when the sample measurement is completed. This is the default setting.
3.3.4 Adjust pH
The instrument will enter into Standby mode if there is no operation in 2 minutes. In standby
mode, the operator should press Wake Up to activate the instrument to the working mode. If
the instrument stays in standby mode for 4 hours, it will maintain the electrodes
automatically. When wake up the instrument, it will carry out a calibration before returning to
The operator can change the waiting time for entering into standby mode. The suggested
To change the waiting time, press Standby ,input the value, then press Yes to confirm.
The liquid distribution valve may be blocked or leaking after using for a long time. The
measuring results could be affected. Therefore it is necessary to check the valve periodically.
Touch Valve, the system will aspirate Std A and Std B automatically. Observe the
aspiration volume, flow speed, the switch of Std.A/Std.Bin the valve and the bubbles to
Press Showmv, the button turns white. The voltages of the liquid positioner and all
electrodes will be displayed on the screen during calibration or sample measurement.
Press Show mv again, the button turns white. The voltages of the liquid
positioner and all electrodes will not be displayed on the screen during calibration or
sample measurement.
K 1 xxx 2 xxx
Na 3 xxx 4 xxx
Cl 5 xxx 6 xxx
ica 7 xxx 8 xxx
Tca 9 xxx 10 xxx
PH
Set Exit
Press Pr.Clea to print out the last date that the de-protein program been performed.
body. Do not open the instrument without disconnecting it from the power supply.
2. Because the samples may have pathogenesis bacterium or viruses, all replaced
3. The reagents are irritating to eyes, skin and diaphragm. Wear proper personal protective
equipment (e.g.gloves, lab coat, etc.) and follow safe laboratory procedures when
handling them in the laboratory The reagents are harmful to human body. If the
reagents accidentally spill on your skin, wash them off with plenty of water and if necessary,
go to see a doctor.
5. If the reagents accidentally spill into your eyes, wash them off with plenty of water and
Always avoid the hematolysis. In addition, the following points should be noted:
1. The serum or plasma can be stored in the refrigerator, but they must be warmed up
2. When preparing the blood serum samples, do not add any materials like the surface
active agent that may interfere in the measurement or even damage the sensor.
3. Hemolysis can affect K result significantly as well as other parameters. Special remark
should be made for hemolysis sample, and the results are for reference to the lab only.
4. If the serum has not been separated completely, the fibrins in the sample may affect the
aspiration, and the proteins may deposit inside the tubing thus causing contamination or
blockage. In such a case, the results will be affected, so the operator should clean the
tubing on time.
5. The sample should be tested within 2 hours, otherwise the result may be not accurate.
1. The analyzer is designed to work continuously for 24 hours a day. No need to shut
strong acid and other supplements that may damage the electrodes.
3. Not all commercial controls are suitable for ISE measurement. Some of them contain
4. The bubbles should not be aspirated during sample aspiration, otherwise the
5. When the sample reaches the checking point, make sure there are no bubbles
because it is very easy to block the tubing system. Always move the serum to
a small and clean tube or cup with a pipette before testing on the instrument.
7. If the ambient temperature fluctuates for more than 10 , the instrument should be
calibrated again
8. The pH value of the standard solutions and the samples should be within 6-9,
11. Every electrode has a serial number. Keep it on the electrode. The warranty of the
12. When removing the sample tube after aspiration, be careful not to hit the sample
probe, otherwise the probe will bend or get misaligned and may not move into the
13. Do not open the front door during calibration or sample measurement, otherwise
14. Check and make sure the grounding is good if static interference is found.
1. Check the electrodes and make sure the filling solutions are sufficient. Add filling
2. Check if there is salty crystal on the internal electrode, if so, clean it with wet tampon.
3. Check and clean the rubber in the wash block. Clean the sample probe.
4. Run deprotein program in service menu. (If less than 5 samples per day,
then just run deprotein program every 2 weeks) Check and clean the tubing system.
5. Run “Na adjust” program if the slope of Na electrode is less than 50.
2. Check the internal electrode and make sure it doesn’t blanch or its coating doesn’t flake
off. Change the internal electrode if necessary. Check the filling solution, change it if it
3. Observe the voltage of the electrode. If the voltage of K,、Na,、Cl are all less than 20 mV,
then change the reference filling solution first. If it doesn’t improve, change the reference
membrane.
found. Observe the flow speed, liquid position and bubble status during sample test or
calibration.
1. Blockage will cause insufficient aspiration volume for sample, Std A and Std B, or no
between Std A and Std B. When aspirating Std A, if small bubbles found at the
outlet of distribution valve, then the flow path inside the distribution valve for Std A
is leaking. If bubbles always stay at the inlet of Std A and can not be excluded, then
the flow path for Std A is blocked slightly. Check flow path for Std B by the same way.
5. The bubbles will appear near the connector if it is leaking. Tighten the tube or connect
6. If leakage found between the electrodes, then the O ring between the electrodes may
not be present or maybe out of position, or the setting screws might not be not
tightened.
7. If the pump tube is distorted, leaking or aged, it will cause no aspiration, insufficient
aspiration volume or incorrect liquid position. Change the pump tube if necessary.
8. Wash the tubing system regularly to prevent the blockage caused by protein
deposition. Clean the sample probe externally. Clean the wash block.
4.5 Check the sample probe
1. Adjust the sample probe if it is misaligned or bend. Make sure the sample probe can
2. If the sample probe is blocked, move it up to top position, take the probe out, and
3. If the wash block is blocked, lift up the sample probe, remove the cover of the wash
Follow the steps below to replace the internal electrode or filling solution:
1. Pull out all electrode wires from the plugs. Remove the tubes from the inlet and
6. Assemble the electrodes in correct order. Make sure that the rubber gaskets are in
right position.
7. Tighten the fixing nuts of the electrode assembly.
8. Connect the tubing to inlet and outlet. Insert the electrode wires back to the plugs.
If the internal electrode blanches or its coating flakes off, then remove the old filling solution,
add fresh filling solution and install a new internal electrode.
Change the filling solution if it is insufficient or the slope of the electrode is too low. Take out
the internal electrode (do not touch any metal materials), remove the old filling solution, add
fresh solution to 85% of total volume, thenput the internal electrode back and tighten it.
K filling solution is for K electrode only, Ca filling solution is for Ca electrode only, and reference
filling solution is for reference electrode only. Na/Cl filling solution is for Na, Cl and pH electrodes.
o-ring
internal part sleeve nut
of electrode
tool for adjustment
of o-ring
4.7 Maintenance of pump, valve and lifting unit
After long time of using, the pump tube will get aged and may cause
inaccurate aspiration volume and result in bad results. So, it is necessary to change the pump
tube.
If there is big noise when the pump head rotating, disconnect the pump tube, pull
out the pump head, loose three setting screws, take out the plate, add a little lubrication oil
to the ball bearing, then assemble the pump head back. If the pump head is damaged, then
The function of the distribution valve is to switch the flow path to Std A or Std B according to
the control program. The valve status is detected by an optical coupler. The performance of
the optical coupler can be affected by dust or foreign objects, and may cause
message will be shown on the screen and will be printed out. Clean the optical coupler
periodically.
Loosen the four setting screws on the valve plate, pull the whole valve out a little, loosen four
nuts, take out the valve plate and four copper standings, loose the setting screws for the
optical coupler. Take out the optical coupler. Turn the valve head with a tweezers and take it
out. Loose two screws on the valve cap and take the cap out. Clean five O rings on the valve
Install the rotor and cap to the valve body, tighten two screws. Pay attention to the five O rings
on the cap. Install the optical coupler to center position. Put on the valve head, tighten the
screws on the valve motor axis. Put the copper standings back, tighten the valve
Note:
Pay attention to the position of the O-rings, copper standings, optical coupler and
Maintain the valve once every year. Keep the optical coupler clean and check it
periodically.
The maintenance of the lifting unit is by removing the dust, foreign objects and oil dirt from it.
Remove the cover of the sample probe, remove the sample probe, clean the axis and add a
little grease. Let the sample probe move up and down for several times, make sure it moves
Note:
The electrode should be sealed and stored under dry environment. Do not add filling
solution to it.
The tubing and O-rings should be sealed and stored under dry environment.
The internal electrode should be sealed and stored under dry environment. Do not
Chapter 5. Hardware
5.1 Connector and Cable list
loose connections. Disconnect the power cable, open the rear cover, and check the cables
and IC chips.
2. When after switch on the power, the LCD doesn’t light on, the lifting unit doesn’t work, the
pump motor doesn’t work: check 9-pin DC power cable and transformer output cable J107.
3. When switch on the power, the LCD doesn’t light on, but the lifting unit work normally:
Check U102 (LM7812)、9-pin DC power cable, NBQ connector and 12V voltage.
The LCD inverter may be damaged.
4. The lifting unit can not move, the valve doesn’t work: Check U103 (LM7812)、
9-pin DC power cable、U106、U107.
5. The pump motor doesn’t work: Check U110、LM7812、9-pin DC power cable、U108、
U109
6. The signals of calibration or measurement are confused or all zero: check cable J204.
7. If the program works incorrect, clean the dust accumulated around U301 IC chip and
check. It is recommended to clean PCB with pure alcohol and blow to dry.
10.)No print. Check J108, J207(PRT) connectors and the connectors between printer head
Note:
5.3.1 Features
Electric feature: comply with EIA RS-232C
4 NC none
6 NC none
7 NC none
8 NC none
9 NC none
The instrument communicates with external PC through RS-232 serial port. After each test
finish, the data will be automatically sent out during the time of printing. The content of the
transmitting data is almost same as the printing data. The operator can also send
Example:
Data format:
Sample number + 2 blank + ID + 2 blank + symbol byte + blank + k result + blank + Na result +
blank +
Note:
Chapter 6 Trouble-shooting
Display prompts
Code Erro Solutio
1 r
Optical coupler error Check J101, clean optical n coupler, adjust the
# position of optical coupler
2 Elevator switch error Check J102, J114, elevator switch,
# U103(LM7812)、
U107(ULN2803)、U106(74HC373) ,9-pin DC
Printing prompts
Code Erro Solutio
r n
Error 0# Liquid Connect the positioner again.
positionin
Error 3# Too much or less Check the distribution valve and positioner
sample aspirated
May cause the instrument to work abnormally; the slopes become abnormal during
calibration; the testing result not stable.
May cause: instrument works unstable, the deviation of Na, Cl is too big
Positioner:
First, make sure the connection of the positioner is good. Then check the
positioner’s voltage with “Voltage” program in “Service” menu. Rotate the pump
manually and let the liquid pass through the positioner. The positioner’s voltage
should become small when there is liquid or big when there is no liquid. The
difference should be more than 500.
Electrode:
Check the voltage of the electrode by “Voltage” program in “Service” menu. If the voltage is
If aspiration alarm appeared or no aspiration at all, replace a new reagent pack (Std A, Std
B)
Cause Solution
Unreliable grounding Check the connection of
the grounding wire
Cause Solution
Too many organic deposits on Wash it with deproteinised Solution.
the electrode membrane
Reagent contaminated Replace the reagent
Insufficient filling solution Refill the filling solution
Electrode does not work Replace the failed electrode
Dust or moisture around the electrode or Clean and dry the electrode and plug
plug
tubing loose reconnected
Pump tube stuck recover the tube
Pump tube broken replace the tube
Tubing blockage clean the tubing
Gasket between the electrodes Re assemble it
wrong placed or missing
Leakage between the electrode Reinstalled firmly
Optical coupler too dirty or broken Clean or replace it
Mixing Chamber Blocked Clean Mixing Chamber
Cause Solution
condenser condensing Liquid outlet blocked
AB slope is zero No Std R, pressure sensor cable loose or
Mixer don’t move damaged.
Reconnect J111 , replace mixing motor
Liquid outlet valve don’t work Reconnect J111 , replace liquid outlet valve
Air outlet don’t work, leakage, AB slope is Reconnect J109, replace air outlet valve
low
AB slope abnormal Std R invalid, leaked or blocked
Cause Solution
- 27
-
Auto sampler on incorrect position Check J104 connector and cable, clean the
dust on the counting optical coupler,
replace the optical coupler
Cause Solution
Aspirating tube loose or broken Connect again or replace it
Pump tube sticks Restore the tube
Pump tube broken Replace the tube
Pump tube blocked Clear the blockage
The gasket between the electrodes Place the gasket properly
does not placed properly or
The electrode assembly leaks Tighten the assembly again
Dust in the positioner, Clean or replace the positioner
or positioner