Rotograph Evo 3d
Rotograph Evo 3d
Rotograph Evo 3d
Service Manual
Version 5 December 2011 (Rev. 1)
SERVICE MANUAL
Revision history
Revision history
Contents
1. INTRODUCTION 1-1
1.1. Description of the system .................................................................... 1-2
1.2. Icons appearing in the manual ............................................................ 1-3
1.3. How to contact VILLA SISTEMI MEDICALI technical service ................ 1-3
3. DESCRIPTION 3-1
3.1. Identification labels and laser labels .................................................... 3-1
3.1.1. Identification plates and laser labels "220-240V" version ...................3-2
3.1.2. Identification plates and laser labels "110-120V" version ...................3-3
3.2. Function, Models and Version ............................................................. 3-4
3.2.1. Basic version....................................................................................3-4
3.2.2. Version with cephalometric device ....................................................3-5
3.2.3. EVO XP (Extended Projection Package) - Optional .............................3-6
3.3. Block diagram ..................................................................................... 3-7
3.3.1. Power supply assembly.....................................................................3-8
3.3.2. CPU Board (A5 and A6) ....................................................................3-9
3.3.2.1. CPU Board jumper configuration .....................................3-10
3.3.3. Generator Board (A10) and Tubehead ............................................. 3-11
3.3.4. Keyboard (A4)................................................................................. 3-12
3.4. Control panel - Descriptions and functions........................................ 3-14
3.4.1. Key functions description ............................................................... 3-17
5. PRE-INSTALLATION 5-1
5.1. Electrical setting up ............................................................................. 5-2
5.2. Packaging ............................................................................................ 5-5
5.3. Space requirements ............................................................................. 5-6
5.3.1. Version without CEPH...................................................................... 5-6
5.3.2. Version with CEPH........................................................................... 5-7
6. INSTALLATION 6-1
6.1. Setting of the wall ................................................................................ 6-2
6.2. Column mounting................................................................................ 6-3
6.3. Mounting of the rotating arm assembly................................................ 6-6
6.4. Mounting of CEPH-arm (Optional)...................................................... 6-11
6.5. How to mount the coverings............................................................... 6-13
6.6. Inserting the CEPH sensor in the sensor holder ................................. 6-14
6.7. Final checks ...................................................................................... 6-15
6.7.1. PC set up ....................................................................................... 6-15
6.7.1.1. PC – USB protection keys................................................ 6-15
6.7.1.2. PC - Rotograph EVO 3D communication ......................... 6-15
6.7.1.3. DICOM set-up ................................................................ 6-16
6.7.1.4. Image treatment filters setup .......................................... 6-17
6.7.2. Verification of the PANORAMIC function ......................................... 6-18
6.7.3. Image verification for the 3D function ............................................. 6-21
6.7.4. Verification of the CEPH function ................................................... 6-22
6.7.5. Verification of exposure parameters................................................ 6-28
6.7.5.1. Invasive method.............................................................. 6-28
6.7.5.1.1. kVp ................................................................................6-31
6.7.5.1.2. mA Check......................................................................6-32
6.7.5.1.3. Time...............................................................................6-32
6.7.5.2. Non-invasive method ...................................................... 6-33
6.7.6. Storing of automatic exposure parameters ...................................... 6-35
6.7.6.1. Table of pre-set anatomic parameters.............................. 6-36
7. TROUBLESHOOTING 7-1
7.1. LEDs.................................................................................................... 7-1
7.1.1. Generator board A10 LED ................................................................ 7-1
7.1.2. CPU board A5 LED........................................................................... 7-2
7.2. Displayed messages ............................................................................. 7-3
7.2.1. Errors with code from E000 to E199................................................. 7-7
7.2.1.1. E108: Hardware key fault ................................................. 7-7
7.2.1.2. E110: Battery fault ........................................................... 7-7
7.2.2. Errors with code from E200 to E299................................................. 7-8
7.2.2.1. E200: Zero position optical sensor of rotation always active /
E201: Zero position optical sensor of rotation never active /
E204: Unexpected activation of zero position rotation
sensor .............................................................................. 7-8
1. INTRODUCTION
NOTE:
* The present manual is updated for the product it is sold with in order to
grant an adequate reference in performing diagnostics and repair
operations normally carried out by the service engineer.
The manual may not reflect changes to the product not impacting service
operations.
WARNING:
1. Rotograph EVO 3D is an electro-medical device and it can be used
only under the supervision of a physician or of highly qualified
personnel, with the necessary knowledge on X-ray protection.
2. The device must be used in compliance with the procedures
described, and never be used for purposes different from those
herewith indicated.
3. The user is liable as concerns the legal fulfilment related to the
installation and the operation of the device.
In this type of device, the linear digital sensor traditionally used in the
digital panoramic X-rays has been replaced by a wide range Digital
Sensor, also known as the Flat Panel.
These sensors allow to capture an area of the maxillofacial complex that
include the most important anatomical details, by acquiring a series of
two-dimensional images (imaging).
2. SAFETY INFORMATION
WARNING:
Please read this chapter thoroughly.
Villa Sistemi Medicali designs and builds its devices complying with the
related safety requirements; furthermore it supplies all information
necessary for a correct use and the warnings related to danger associated
with X-rays generating units.
Only the authorised personnel can remove the covers and/or have
access to the components under tension.
2.1. Warnings
Rotograph EVO 3D must be used in dental surgeries, radiology and
hospital settings.
Do not let water, or other liquids, into the device, as this could cause
short-circuits and corrosion.
Before cleaning the device, please disconnect it from the line voltage.
While performing the radiography, no-one, apart from the operator and
the patient, must remain in the room.
Clean and disinfect all parts that come into contact with the patient.
The centring bite or the bite protective sleeve, the head strip for 3D
examinations and the ear centring devices of the Cephalostat must
be replaced after each examination in which they were used.
Never try to rotate the moving arm manually when the unit is
switched on, to avoid permanent damage to the unit.
The manual movement is only allowed in case of error 206 (no power to
motors) to allow the exit of the patient.
Although the dose supplied by dental X-ray units is quite low and
distributed on a small surface, the operator must adopt the precautions
and/or suitable protection for the patient and himself, during the
execution of radiography. It is advisable to control the X-ray emission
from a protected area, by means of a remote control. If it is necessary to
operate near the patient, stay as far as the remote control cable allows,
or at least 2 m both from the X-ray source and from the patient, as
shown in Figure 2-1 and Figure 2-2.
Protected area
Protected area
WARNING:
The network connector on the base of the Rotograph EVO 3D column
must be connected to the dedicated image acquisition and 3D and
panoramic reconstruction PC via a shielded Ethernet cable "Cat.5e" or
greater.
Do not use this connector to connect the Rotograph Evo 3D to LAN
networks.
NOTE:
* The dimension of the "patient's environment" is defined as a distance of
at least 1.5 m from the actual patient.
If the PC is positioned inside the patient's environment, it must conform
to the requirements specified by the IEC 60601-1 standard for medical
devices; if located outside of the patient's environment, it must be
compliant with the IEC 60950 standard.
If the PC is supplied by Villa (optional), it must be placed outside the
patient's environment.
NOTE:
* Do not apply movements to the rotating arm or the tube-head when
the equipment is on.
1 0.025 0.03 1
C
0 0.03 0.004 0
-1 0.03 0.02 -1
For transmitters rated at the maximum output power not listed above,
the recommended separation distance"d" in meters (m), can be estimated
using the equation applicable to the frequency of the transmitter, where
"P" is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE:
* At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE:
* These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection of structures, objects
and people.
Symbols Description
∼ A.C.
N Connection point to the neutral conductor
L Connection point to the line conductor
Protection grounding
Operation grounding
Laser
Dangerous voltage
Filtration
X-Ray tube
3. DESCRIPTION
4 5
6, 7
8 (*)
1 2
Rotograph EVO 3D data plate Tube-head characteristics plate
3 4
EVO XP plate CEPHALOMETRIC device plate
(Additional projection package)
6
(N° 2) Spot Laser
indicator plate
5
PANCEPH digital
sensor data plate
7
(N° 2) Laser symbol
label
1 2
Rotograph EVO 3D data plate Tube-head characteristics plate
3 4
EVO XP plate CEPHALOMETRIC device plate
(Additional projection package)
5
PANCEPH digital 6
sensor data plate (N° 2) Spot Laser
indicator label
7
(N° 2) Laser
symbol label
8
WARNING label
The version with cephalometric device allows you to perform the following
examinations:
• Panoramic, Sinus and TMJ, Adult and Child, with the same
characteristics described for the base version.
• 3D volumetric examinations with the characteristics described for the
standard version.
• Digital Cephalometry with a choice between Adult and Child with 3
sizes each.The execution of the high resolution or standard
resolution can be selected within each combination, for a total of 12
combinations in Automatic selection.When examining at a Normal
Resolution, the examination is conducted with a reduced scan time,
allowing a further reduction of dose. In manual selection it is possible
to select high voltage between 60kV and 86kV, in 2kV steps and
anodic current from 6 mA to 12 mA in 1 mA steps. The positioning of
the primary slide collimator, the secondary collimator and the Digital
Sensor (in its sensor holder) is automatic, depending on the
size/projection selected.The Soft Tissues Filter is powered to attain
the best possible result of the facial profile.
• Children only Carpus exam with 3 selectable sizes, for a total of 3
combinations of exam in automatic. in manual selection it is possible
to select high voltage between 60kV and 86kV, in 2kV steps and
anodic current from 6 mA to 12 mA with 1 mA steps. The positioning
of the primary slide collimator is automatic.
The unit, both the base and the version with cephalometric device, is
prearranged to be fitted with the EVO XP (Additional projection package),
function, which enables to perform the following examinations:
• The right or left Emi-panoramic is used when the patient is known to
have a problem only on one side of the arch, in order to reduce the
radiation
• The reduced dose Panoramic reduces the dose radiated on the
dentition by excluding the ascending rami of the TMJ from the exams
• The frontal dentition enables to perform examinations of the front
part (roughly from canine to canine)
• The dentition with improved orthogonality reduces the overlap of the
teeth, thereby improving the diagnosis of interproximal decay.
NOTE:
* All these examinations can be added to Rotograph EVO 3D systems
already installed in the field.
NOTE:
* The code inserted into Rotograph EVO 3D to enable the optional
examinations is protected by a Unique Identification Code (UIC); in the
event the UIC is not present or is faulty, an error E107 will be shown.
Pressing the "Patient entrance" (7) will reset this condition, but
only the Panoramic, Sinus and TMJ will be active at the end of the start-
up procedure.
The UIC code can be displayed by pressing the "Right Arrow/Left Arrow"
The UIC code is simply an identifier of the single Rotograph EVO 3D unit;
in order to enable the optional functions it is necessary to request the
activation code from Villa Sistemi Medicali, which derives from the
Unique Identification Code or from the device serial number.
From the electrical point of view, the system can be divided into 4 main
blocks:
• Power supply assembly
• Main CPU board (A5)
• Generator board (A10), Generator CPU board (A9) and tubehead
• Column CPU board (A1)
• Keyboard with its control board (A4).
All control boards above listed are equipped with a local microcontroller
that shares information with the main CPU using a CANBUS
transmission line and protocol.
Each of the main blocks above listed is here after described.
It is located in the rear part of the column and is mainly composed by the
mains switch (S1), a 24Vdc 7A switching mode power supply which
supplies all circuits of the machine excluding the column motor, and a
further power supply which supplies the column motor and the enabling
circuit for X-ray emission.
The power supply assembly also acts as an interface with a number of
external signals and circuits like:
• Input for remote X-ray push button (S7) and output to the CPU of the
same signal.
• Outputs for the "Ready" and "X-ray ON" lamps (X3).
• Driving of the DC column motor (M1): this motor can be activated
either through its control board (A1), in case the movement is
requested by the operator acting on the keyboard, or through the
switch S2 located in the rear part of the column base. This switch
can be used to raise/lower the column during the installation phase
when the CPU has not been yet connected to the system.
• Input for the emergency column motor microswitches (S3 and S5):
these microswitches indicate the limit for the movement of the
column. If for any reason, the column goes beyond these
microswitches, the motor is de-activated by cutting the voltage.
Normally, the position of the column is also monitored by two other
microswitches (S4 and S6) that, as for all the other positioning
sensors, provide their signals to the Column CPU board (A1).
The unit does not include a voltage selector circuit for the mains voltage.
Therefore, the unit is manufactured in different versions, depending on
the line voltage of the installation place.
The power supply assembly module also includes the Column CPU board
(A1) that is dedicated to the control of the vertical column motion during
all phases (slow speed, ramp up to high speed, ramp down from high to
low speed, end run microswitches control, etc.).
In the CPU board are present some jumpers that define the system
configuration.
Wrong settings may affect the system functionality.
• XJ1 = Open
• XJ2 = Closed
• XJ6 = 2-3
• XJ8 = 2-3
• XJ11 = Open if CEPH is present / Closed if CEPH is not present
• XJ12 = Open
• XJ13 = Closed
XJ6 XJ8
XJ13
XJ12
XJ11
XJ2 XJ1
Figura 3-1
The Generator board and the tubehead are located on the rotating arm,
very close to each other. The power supply voltage is directly provided by
the Power supply assembly.
Dedicated switching circuits, directly located on the board, generate the
voltage used on the board itself.
The Generator board receives the signals concerning the X-ray dose to
provide (kV and mA), from the CPU board through CANBus messages; it
is the Generator CPU that generates the commands used for the X-ray
emission. The Generator board provides to the tubehead the voltages that
drive the high voltage transformers that then drive anode and filament of
the X-ray tube, also giving the relevant timing.
The keyboard is the interface with the operator, and is composed by the
following items:
• Matrix of keys, constituted by microswitches with short stroke,
necessary to activate the different available functions
• Signalling LED integrated in the touch panel
• LCD display equipped with back-light, composed by a matrix of 16
characters and 2 rows
• Keyboard PCB.
The keyboard PCB is directly connected to the main CPU board (A5)
which controls it. The language of the messages shown on the display
can be selected among 5 different options (English, Italian, French,
German and Spanish). The language selection is only available for the
messages dedicated to the user. The messages relative to the service
programs (Password) are always in English.
"Centring/ Patient
Entrance" button
Light signalling
"X-rays in progress"
"Test" button
"Adult/Child
selection" button
"Centring
"Size Selection" devices ON"
button button
Figure 3-2
Child
It is possible to select the anatomic
mode examinations (anatomic Adult
selection), using prefixed exposure
values.
This kind of selection enables to choose Small
between Adult/Child, each with three
different sizes (small, medium, large). Medium
Large
1 2
7 6
Display
8 14
12+13
9
11 4 15
10 16
3 5
LEGEND:
Messages Anatomic selection
Display: indicates operative messages, warnings and 8 - Patient selection key: Adult or Child
exposure parameters. (green LED)
9 - Size selection key: Small, Normal, or Large
Signal lights (green LED)
1 - Light indicating the machine is ready for X-ray 10 - Arch selection key: Protruded, Standard or
emission (green LED) Retracted (for panoramic execution) (green LED)
2 - Yellow LED indicating X-ray emission
Examination mode
Manual setting of exposure parameters 11 - Exam mode selection key
3 - Parameter selection key: kV, s or mA 12 +13 - Type of exam selection keys
4 - kV, mA or STF increase key
5 - kV, mA or STF decrease key Centring devices
14 - Sagittal and Frankfurt/Ala trago plane centring
Preparation functions device ON key
6 - Key to set Test function
(green LED) Column height adjustment
7 - Key for: 15 - Column up key
> Resetting and realigning the device's axes (in 16 - Column down key
case of collision with patient or in case of
release of rays button)
> Repositioning the rotation group (to bring the
group to the initial position after the examination
and to exit from the "making an exposure")
mode
> Confirmation
4. TECHNICAL DATA
General features
Device type Rotograph EVO 3D
Manufacturer VILLA SISTEMI MEDICALI
Buccinasco (MI) Italia
Class Class II B for European Directive
for Medical Devices 93/42
Class II for Canadian MDR
Class I with type B applied parts
according to IEC 60601-1
Class II according to 21CFR-
subchapter J
Operating mode Continuous with adaptive duty cycle
Protection degree IPX0
Rated line voltage 220-240 V∼ 110-120 V∼
Line frequency 50/60 Hz
Maximum line current 7 A @ 230 V∼ 15 A @ 115 V∼
50/60 Hz 50/60 Hz
Absorbed power 1.5 kVA @ 230 V∼ 1.6 kVA @ 115 V∼
50/60 Hz 50/60 Hz
Protection fuse (F1) 7AT 15 A T
Switching supply protection fuse (F2) 1.6 A T 3AT
Generator card protection fuses F1: 10 A F
F2: 5 A HF
F3: 2 A T
Line apparent resistance 0.5 Ω max --
Line voltage regulation -- < 3% @ 99 V∼
Rated output voltage (kVp) 60 ÷ 86 kVp, with 2 kVp steps
Anodic current 6 ÷ 10 mA, with 1 mA steps for PAN, TMJ
and Sinus
6 ÷ 12 mA, with 1 mA steps for Ceph
(up to 76 kVp)
6 ÷ 10 mA, with 1 mA steps for Ceph
(from 78 kVp to 86 kVp)
Additional filter for wide range sensor << 1.2 mm Al eq. @ 70 kVp
cover (for 3D exams)
Additional filtration for PANCEPH sensor 0.1mm Al eq. @ 70 kVp
cover
Exposure times
Panoramic (PAN) 13.8 s Adult/Child
RH/LH Emi-panoramic 7.4 s Adult / 7.3 s Child
Reduced dose Panoramic 11.4 s Adult/Child
Improved orthogonality dentition 11.9 s Adult/Child
Frontal dentition 4.4 s Adult/Child
TMJ mouth closed/open 2.44 s per image for left and right joint in open
and closed condition for a total of 9.7 s
SINUS 9.4 s
Volumetric 3D exams 11.2 s for Dentition and Sinus
10.1 s for TMJ Left and TMJ Right (each)
Cephalometry (Ceph) Variable exposure time depending on the type of
resolution and size selected.Minimum 4.5 s
(18x22nR), maximum 15 s (30x22hR)
Exposure time accuracy ± 10 %
Examination modes
Examination selection • Automatic selection for Adult and Child, 3
Sizes
• Dentition type selection (in Panoramic)
• Manual selection
• Collimator with automatic positioning
Panoramic • Standard Panoramic
NOTE: • RH and LH Emi-panoramic
Some of these exams are • Reduced dose Panoramic
optional and depend on the • Improved orthogonality dentition
system configuration. • Frontal dentition
TMJ (Temporo-Mandibular TMJ mouth closed/open
Joint)
SINUS Sinus P/A projection
Volumetric 3D exams Automatic selection for Adult and Child, 3 sizes
chosen between 4 types of exam: Dentition, TMJ
Left, TMJ Right, Sinus
Cephalometry and Carpus • Normal resolution in Latero-Lateral or Antero-
Posterior projections (different sizes)
• High resolution in Latero-Lateral or
Postero/Anterior projections (different sizes)
• High resolution Carpus exams
• Motorised soft tissue filter
Image magnification
Standard Panoramic 1 : 1.28 Adult and Child
TMJ open/close mouth, 4 images 1 : 1.25 (nominal)
Sinus 1 : 1.27 (nominal)
Ceph 1: 1.10 geometric magnification
1 : 1 magnification after software
correction (*)
Tube-head characteristics
Model MRE 05
Manufacturer Villa Sistemi Medicali S.p.A.
20090 Buccinasco (MI) Italia
Maximum tube voltage 86 kVp
kVp accuracy ±8%
Maximum anodic current 12 mA
Anodic current accuracy ± 10 %
Output radiation linearity < 0.2 according to standard
IEC 60601-2-7:1998 paragraph 50.102.2
Duty cycle Adaptive duty cycle depending on the
exposure parameters: from 1 : 8 (at 60
kV, 6 mA) up to 1 : 20 (at 76 kV, 12 mA).
Further reduction for three exposures in
close-up sequence: from 1: 3.6 (at 60 kV,
6 mA) up to 1 : 9 (at 76 kV, 12 mA).
Nominal power 1.032 kW (86 kVp - 12 mA -4s)
Total filtration 2.5mm Al eq. @ 70 kVp
HVL (Half value layer) > 2.0 mm Al eq. @ 60 kVp
> 2.7 mm Al eq. @ 74 kVp
> 3.2 mm Al eq. @ 86 kVp
Transformer insulation Oil bath
Cooling By convection
Leakage radiation at 1 m < 0.5 mGy/h @ 86 kVp - 12 mA - 3s duty
cycle 1/16
Tube-head maximum thermic capacity 310 kJ
(*) WARNING:
The declared image magnification value is valid after proper software
calibration.
NOTE:
* The Rotograph EVO 3D version without cephalometric arm, uses only the
wide range sensor both for 3D images as well as for Panoramic images
(narrow beam), while is also used the second PANCEPH sensor in the
version with Cephalometric arm.
NOTE:
* These removable parts are considered "type B applied parts", in
accordance with IEC 60601-1, 2nd edition.
Some of these parts do not carry identification codes due to their small
size. The use of these parts on other devices is not possible, since they
are parts designed specifically for the Rotograph EVO 3D.
• 21 CFR Subchapter J
• General safety:
IEC 60601-1:1988+A1:1991+A2:1995
IEC 60601-1-1:2000
IEC 60601-1-4:1996+A1:1999
IEC 60601-2-7:1998
IEC 60601-2-28:1993
IEC 60601-2-32:1994
IEC 60601-2-44:2001
UL 60601-1 (1st edition)
CAN/CSA C22.2 No. 601.1-M90 (2nd edition) +A1 + A2
• Electromagnetic compatibility:
IEC 60601-1-2:2001
• Usability:
IEC 60601-1-6:2004
Classifications
The EVO Rotograph 3D is a class I electro-medical device and Type B as
of IEC 60601-1 classification, foreseen for a continuous working at
intermittent load.
The classification according to EC Directive 93/42 and subsequent
amendments for medical devices is Class II B.
According to Canadian MDR, the equipment belongs to class II.
According to FDA 21 CFR, the equipment belongs to class II.
Load
400
350
300
250
E[kJ]
200
150
100
50
0
0 100 200 300 400 500 600
min
WARNING:
These measurements require the removal of the HF group covers; this
means to gain access to internal parts where high voltage are normally
present.
WARNING:
During the panoramic examination, the set value of kV and tube current
varies according to a pre-determined curve in order to compensate the
different absorption of X-ray beam due to different anatomical structures.
In this way, it is possible to obtain a good uniformity of the image
contrast. Particularly, the chosen value is lowered on the initial phase of
the panoramic and increased on the scissors/canine zone, in order to
compensate the effect of the cervical spine.
The value displayed during the panoramic examination corresponds to
the one chosen by the user, while the real value can be different; these
effects must be considered in case of measure of the exposure factors
using standard diagnostic mode. As an example, the variation follows the
curve hereafter:
Actual value kV
4.5. Dimensions
1020 (40,16")
Ø 1040 (40,95")
Ø 1140 (44,88")
with free standing base
Figure 4-1 - Base version
1020 (40,16")
927 (36,50")
905 (35,63")
1271 (50,04")
Ø 1040 (40,95")
Ø 1140 (44,88")
with free standing base
1910 (75,20")
1970 (77,56") with free standing base
5. PRE-INSTALLATION
The instructions indicated in this and in the following chapter enable to
perform a correct installation in order to grant a regular operation of
Rotograph EVO 3D.
The supplier can supply the assistance and the necessary technical
advice for pre-installation, all masonry works and the pre-installation
phase are at the customer’s charge and must be performed complying
with the indications given below.
WARNING:
In its standard versions, Rotograph EVO 3D must be fixed to the wall
with the two brackets supplied. Each dowel must support a max.
extraction force of 120 kg (264 lbs).
The fixing dowels to be employed, for any kind of wall, are the following:
• full or concrete bricks: screw anchors (provided with the
installation kit) in cast iron M6 or chemical screws WURTH (not
provided)
• wood mountings: self-threading screws (not provided)
• hollow bricks: chemical dowels (not provided).
NOTE:
* The device is supplied as a unit to be installed permanently.
Please DO NOT connect the unit to the power using a normal socket, to
avoid compromising the electrical safety.
NOTE:
* The electrical connection must be done on the terminal board X0 (see
operation circuit diagram – chapter 10).
NOTE:
* Rotograph EVO 3D, IS SET TO connect, at the entrance of the X-ray
room, the following control and warning devices:
• REMOTE X-RAYS BUTTON: "Dead man switch" remote control,
enables to perform the exam at a distance, the operator can stand
outside the X-ray emission area. This button must be suitable to
prevent unwanted emission. The standard X-ray button supplied with
the unit has the above characteristic.
• READY light: Green light (24V 40W max.), it signals that the machine
is ready to perform the exam. (contact N.O.).
• X-RAYS light: Yellow light (24V 40 W max.) it signals the entry in the
X-ray room is forbidden, since an exposure is on the run (contact
N.O.).
Figure 5-1
WARNING:
It is installer’s responsibility to check the characteristics of the remote
X-ray button
NOTE:
* The unit only provides the closing contacts relative to the above
mentioned functions. Power voltages for the relevant devices have to be
provide from outside, making sure not to exceed the indicated ratings.
5.2. Packaging
Rotograph EVO 3D in base version is delivered in two carton-board
boxes; the device equipped with Cephalometric unit will be delivered in
three boxes.
Weight
Contents
Net Gross
- Accessories
- Cephalometric device
57 kg 72 kg
- Covers
(126 lbs) (159 lbs)
- Accessories for Cephalometric device
NOTE:
* All boxes mount shock detectors.
At the receiving and before opening boxes, verify that those sensors have
not been activated.
WARNING:
Villa Sistemi Medicali will not bear any responsibility for damages
caused to the equipment due to improper unpackaging procedure,
and for the relevant costs.
(*) A = minimum 600 mm (23.6"), recommended 800 mm (31.5") for service purpose
Figure 5-1
(*) A = 600 mm (23.6") minimum, 800 mm (31.5") recommended for service purpose
(**) B = 1460 mm (57.52") minimum front space, 1610 mm (63.39") recommended front space
for service purpose
Figure 5-2
6. INSTALLATION
NOTE:
* Rotograph EVO 3D is delivered pre-mounted in groups; it is contained in
2 (PAN version) or 3 boxes (CEPH version).
The mechanical mounting consists exclusively in assembling the above
mentioned groups. Most of the adjustment are carried out in factory.
Two technicians will be necessary to perform some procedures; the
phases requiring the intervention of two men are identified in the related
chapters.
NOTE:
* This paragraph is valid only for wall mounted version.
NOTE:
* Rotograph EVO 3D has been designed for a wall fixing by two brackets,
each of which requires to be fixed by three dowels. In order to find the
right position of the brackets, it is necessary to use the quotes indicated
in the Figure 6-1.
It is very important the vertical alignment of the central holes and
the perpendicularity with the floor; it is strictly suggested to use a
plumb.
1. Mark the wall by a centre punch, at the level of the fixing holes; drill
the wall according to the type of dowels (see chapter 5).
70.5 70.5
(2.76") (2.76")
1040 (40.95")
99 99
(3,9") (3,9")
(4.5")
114
Figure 6-1
NOTE:
* In case of free standing base version, it is necessary to remove the four
feet grub screws before to raise up the column.
Retaining bracket
"B"
Fixing brackets
"A"
Figure 6-2
1. Remove the screws used to fix the column to the wooden pallet and
raising it (this operation requires the presence of two persons).
2. With the column in vertical position, remove only the two fixing
brackets "A".
NOTE:
* Leave the retaining bracket "B" in place, because it will be necessary for
the following operations of support arm mounting.
NOTE:
* The following procedure is not valid for free standing base version.
3. Mount on the column basement the lower fixing plate "D" and mount
on the column the upper fixing plate "E" (Figure 6-3).
4. Insert the external back cover on the upper fixing plates "E" without
fixing it. You will fix it at the end of the mechanical and electrical
installation.
5. Mount the upper wall bracket "C" on the fixing plates "E" (Figure
6-3), without tighten the screws.
6. Position the column close to the wall.
7. Fix the upper wall bracket "C" to the wall. Then tight the screws
between upper fixing plates "E" and wall bracket "C".
8. Fix the lower fixing plate "D" to the wall, without tighten the screws.
9. Verify the parallelism between column and wall positioning a bubble
level on the chin rest support.
In order to correct this position, it is possible to insert some spacers
(Figure 6-3 - provided with the column) between the lower bracket
and wall. Once the position is reached, tighten the screws.
10. Acting on the feet grub screws (Figure 6-2), adjust them in order to
level the feet on the floor. At the end, cover the grub screws with the
provided cups.
To easily reach the back side of the column leaving it hook to the wall,
remove the fixing screws located in the back side of the column basement
and remove the two screws "F" (Figure 6-3) from the upper bracket "C".
Rotate the column.
NOTE:
* Rotograph EVO 3D is equipped with a switch (S2) which allows to move
the column vertically for service use. It is located in the back lower side
of the column, protected by a metallic cover.
NOTE:
* Rotograph EVO 3D is shipped with the column pre-set at the minimum
height.
In case the room layout permit to reach higher position, it is possible to
increase the stroke adjusting the reference cams located in the column,
close to the end-run microswitches plate.
Remove the microswitches plate and lift properly the column in order to
access the reference cams holes.
Retaining bracket
"B"
Screws "F"
Spacers
Figure 6-3
2. Remove the screws fixing the rotating arm to the wooden frame.
Raise up the support/rotating arm using the designated handling
zones (Figure 6-4 - This operation requires the presence of two
persons).
Figure 6-4
3. Move the rotating arm near the column from lateral side in order to
avoid contacts between the Digital Sensor holder and the chin
support.
Retaining bracket
Figure 6-5
5. Remove the two posterior screws fixing the CPU board support plate
to the rotating arm and tilt it at about 75°; lock the support plate to
the lateral frame.
6. Insert the reference pins "B" (Figure 6-6) between the rotating arm
and column using a hammer. Insert the remaining two screws "C"
(Figure 6-6) and tighten all the fixing screws.
B C A
Retaining bracket
Figure 6-6
(Rev. 0) 6-7 Rotograph EVO 3D
SERVICE MANUAL
Installation
7. Remove all brackets and rotate the rotating arm about 90° in
clockwise direction in order to reach easily the covers screws.
8. Remove the tubehead cover; remove the two spacers "D" and the
lower fixing bracket "E" (Figure 6-7).
Spacers "D"
Lower fixing
bracket "E"
Figure 6-7
9. Open the sensor holder and mount the rotating arm lower cover
(Figure 6-8).
NOTE:
* In case of Ceph up-gradable version or single sensor version, in order to
mount the rotating arm lower cover, it is necessary to remove the sensor
holder internal covers.
Figure 6-8
10. Mount the temple support group inserting and rotating it in the
relevant shaft (Figure 6-9) until the group is locked.
Figure 6-9
11. Insert the cables between column and rotating arm through the holes
present in the rotating arm; reposition and fix the CPU board and
connect the cables to the related connectors.
NOTE:
* For 110-120V version, to access to the CPU board connectors, it is
necessary to remove the metallic cover.
At the end of the connections, position the cables in the relevant seating
and remount the cover.
Handling zone
Red handle
Figure 6-10
NOTE:
* Do NOT remove the handle before the complete installation of the Ceph
arm on the column.
1. Remove the column rotation fixing screw, located in the back side of
the column basement and remove the two screws "F" (Figure 6-3)
from the upper bracket "C".
2. Rotate the column in clockwise direction in order to reach the
column back side.
3. Lift the Ceph arm with the red handle and the handling zone and
remove it from the package (Figure 6-10).
4. Position the Ceph arm near the column and using the reference pins
and screws, fix the Ceph arm to the column (Figure 6-11).
5. Insert the cables in the hole between column and rotating arm and
connect them to the relevant connectors on the CPU and Digital
boards.
6. Remove the red handle.
Reference pin
Reference pin
Figure 6-11
NOTE:
* If necessary, to easily access to the covers back side screws, remove the
column rotation fixing screw, located in the back side of the column
basement and remove the two screws "F" (Figure 6-3) from the upper
bracket "C". Rotate the column in clockwise direction in order to reach
the column back side.
Figure 6-12
NOTE:
* Not valid for PAN only version.
In order to insert the sensor in the desired station, carry out the following
operations:
1. Grip the sensor by the appropriate handgrip; close your fingers to
form a fist, by engaging the control lever and bring it to the position
where the lever disappears inside the handgrip, so that the whole
mobile system retracts.
2. Keep the sensor with the relative handgrips vertical, so that the
upper plane is parallel to the horizontal part of the sensor holder,
bring the sensor close to the fixed station, by engaging the protruding
part of the mobile sensor into the relative casing.
3. Push the sensor mobile part to the very end, in order to engage the
mobile part onto the fixed hooking system.
4. Carry out a movement towards the lower part, ensuring that the
movement is complete.
5. Only at this point, release the hooking lever, checking that the
sensor is correctly engaged before releasing the handgrip.
WARNING:
During the lever releasing operation, hold the sensor firmly, to prevent
the sensor from falling during the insertion phase due to possible errors.
NOTE:
* All sensor types are equipped with a shock detection sensor; this sensor
is also visible from the outside to enable to operator to perform checks.
Possible shocks are displayed by a change in colour (from
transparent/white to red) of this sensor. The digital sensor can still
function correctly also when the colour changes, displaying a fall that
might also not have damaged the sensor.
NOTE:
* The fall sensor colour change interrupts the warranty on the sensor.
6.7.1. PC set up
Before to switch ON the PC, insert the two license USB keys.
WARNING:
Do not change the IP address of the network card used to communicate
with the Rotograph EVO 3D (usually 192.168.2.xxx). In order to connect
the PC in the Dicom network, it is necessary to use the second network
card.
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
Figure 6-13
4. Place the centering tool (P/N 6107900200) on the chin rest and the
sensor calibration tool (P/N 5607900800) in front of the sensor fixing
it with tape.
Figure 6-14
Centering tool P/N 6107900200 Figure 6-15
Sensor Calibration tool
P/N 5607900800
Figure 6-16
Figure 6-17
8. Measure also the two half of the image in order to check symmetry.
The difference has to be max. 1mm.
If there is not symmetry on the image, it is necessary to adjust the
rotation offset entering in password 118 (see paragraph 7.3.4.1).
Figure 6-18
5. In case the arc is not correct, it will be necessary to modify the lateral
offset as described in the paragraph 9.2.2.
NOTE:
* The CEPH arm has been adjusted on the unit during testing procedure,
and adjustments are not usually necessary in field. Before to proceed,
verify if the arm has been properly mounted during field installation.
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
Figure 6-19
Figure 6-20
4. Select "CEPH sensor zero" menu (see paragraph 9.6.3) to check the
secondary collimator position.
Figure 6-21
collimators and the ceph sensor go to the ceph central position. Place
the centering tool P/N 5209900900 on the secondary collimator.
Figure 6-22
Centering tool P/N 5209900900
and exposure time and use keys increase (4) or decrease (5)
to set values for the exposure (suggested values: 60kV, 6mA and
0.5s).
8. Take X-ray pressing the X-ray button and check if in the obtained
narrow image the projection of the slit of the centering tool is vertical
and in the middle of the impressed area, which is defined by the arc
identified in the figure.
Figure 6-23
Figure 6-24
10. Enter in "STF zero" (see paragraph 9.6.4) to verify the functionality of
the Soft Tissue Filter.
12. Make a exposure with the same parameters (60kV – 6mA – 0.5s).
Figure 6-25
13. Measure the distance between the center of the ear rings and the
edge of the soft tissue filter. The value has to be around 50mm
(considering that the measure scale is not 1:1).
kVp and time can be measured directly on the Generator board (A10);
this method has higher accuracy than the so called non-invasive mode.
The system accuracy is guaranteed by this measuring method.
NOTE:
* If the following actions are performed during maintenance some covers
have to be removed:
a. Remove the outer covering of the H.F generator placed on the rotating
arm.
b. Remove the protection grid of the Generator board.
WARNING:
The Generator board has a working voltage of about 400V.
1. Power off and on the unit. The following message will be displayed:
R O T O G R A P H E V O
R E L E A S E * . * * *
> T E S T <
When this message appears, press the column up and column down
S E R V I C E M E N U
C O N F I G U R A T I O N
2. Using the scroll right and scroll left keys to change the
S E R V I C E M E N U
T R O U B L E S T O O T I N G
displayed:
S E T U P
P A S S W O R D ? 1 0 0
3. Using the increase/decrease keys, set the password equal to 112 and
confirm with the key "Patient Entrance" (7).
4. Using the scroll right and scroll left keys go to the "RX
EMISSION" menu and confirm with the key "Patient Entrance" (7).
At the selection of the X-ray emission test, the display will show the
following message:
X - R A Y E M I S S I O N
7 2 k V 0 6 m A 1 4 . 0 S
WARNING:
From now, the emission is enabled; it starts with the X-ray button press,
so take care of this situation.
> 7 2 k V 0 6 m A 1 4 . 0 S
In this situation, the increase key (4) and the decrease key (5)
NOTE:
* This action allows you to carry out the measuring of the exposure
parameters having the tubehead-arm in a fix position (not rotating).
WARNING:
X-ray will be emitted during the performance of the following operations.
Authorised technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
6.7.5.1.1. kVp
Use a multimeter (R input >= 10MΩ) in working conditions VDC and end
of scale 5/10V (maximum value 4.3 V); position the cold pole on TP40
(GND) and the hot pole on TP44 (kV) on the Generator board (A10). Set
the following parameters: 60kV-6mA-3s.
Perform an exposure and considering that the ratio between the value on
the voltmeter and high voltage is 1V= 20kV, verify that the value
indicated by the multimeter ranges from 2.85÷3.15 V (3V±5%).
Should this last value be outside the specified limit, detect the "set kV"
voltage supplied by CPU, connecting the voltmeter between TP40 GND
test points and TP46 kV (set nearby X57). The value on these points
must range from 2.7 and 3.3 V (3V+3%). If it is out of tolerance replace
the Generator CPU board (A9), otherwise replace the Generator board.
Measured values for different set values are contained in the following
table.
If all measures are within the specified range, and there is an evidence of
a performance loss, measure the high voltage supplied by the tubehead,
and the exposure time using a non-invasive kilovoltmeter with < +3kVp
tolerance. The high voltage value must be within 8% of the set value,
while the time value must be within 10% of the set value. If the voltage is
not within the expected range, replace the tubehead; otherwise replace
Generator CPU board.
6.7.5.1.2. mA Check
Use a multimeter (R input >= 10MΩ) in working conditions VDC and end
of scale 5/10 V (maximum value 4VDC), position the cold pole on TP40
(GND) and the hot pole on TP29 (mA). Set the following parameters:
60kV-6mA-3s.
Perform an exposure and considering that a ratio 1V DC ≈ 3 mA, verify
that the value indicated by the multimeter ranges from 1.86÷2.138V
(6mA±8%).
Set 80kV - 12mA - 3s. The value read must range from 3.643 to 4.276 V
(12mA±8%).
Measured valued for different set values are contained in the following
table.
In case the detected values are outside these ranges, check that the
voltage between TP40 (GND) and TP34 (ma) is contained on the above
ranges. If not, replace the Generator CPU board (A9).
6.7.5.1.3. Time
The exposure parameters (kV, time and dose) can also be verified using
the so called "non-invasive method".
WARNING:
The machine collimator gives a narrow X-ray beam.
Measurement by a non invasive kV meter on a very narrow beam can be
difficult and/or unreliable and special probes with reduced sensitive area
must be used.
It may result helpful to use a fluorescent screen to locate the X-ray beam
and consequently position the probe of the kV meter.
1. With the unit switched on, select the Panoramic Examination mode
the same time, the LEDs relating to "Patient type", "Patient size" and
"Biting" switch off, and the display shows:
x x k V x x m A x x . x s
E M I S S I O N P R O G R A M
WARNING:
The following operations involve the emission of X-rays, so the
Authorised Technician must pay the greatest attention and respect the
protection regulations in force in that country.
NOTE:
* This program allows you to carry out the measuring of the exposure
parameters with the tube-head arm in a fixed position (not rotating)
without variations due to spine compensation.
4. To set the exposure parameters, press key (3) , the display will
> x x k V x x m A x x . x s
x x k V > x x m A x x . x s
x x k V x x m A > x x . x s
The parameters can vary within the limits shown in the following
table:
NOTE:
* The performances are guaranteed only if the measurement of kV and
time is done with the invasive method, due to the fact that the
non-invasive method may introduce errors for instruments tolerance or
wrong measurement condition.
6. To end the control program, press key "Test" (6) ; the display
will indicate:
x x k V x x m A x x . x s
P A N O R A M I C - S T D
on the left of the value, the reference character ">" will appear.
3. Press increase (4) or decrease (5) keys to set the value to store.
After pressing the keys, the display will show the following message:
S T O R E ?
> 0 < = S , T = N
and the LEDs relating to the keys for patient and size (that are being
modified) will light up.
Press the key "Patient Entrance" (7) to confirm or the key "Test"
68 kV 64 kV 68 kV 62 kV
Small Small
6 mA 6 mA 6 mA 6 mA
72 kV 66 kV 72 kV 64 kV
Medium Medium
6 mA 6 mA 6 mA 6 mA
74 kV 68 kV 76 kV 66 kV
Large Large
6 mA 6 mA 6 mA 6 mA
SINUS
Adult Child
66 kV 62 kV
Small
6 mA 6 mA
70 kV 64 kV
Medium
6 mA 6 mA
72 kV 66 kV
Large
6 mA 6 mA
3D Dentition
Adult Child
68 kV 64 kV
Small
6 mA 6 mA
72 kV 66 kV
Medium
6 mA 6 mA
74 kV 68 kV
Large
6 mA 6 mA
68 kV 62 kV
Small
6 mA 6 mA
72 kV 64 kV
Medium
6 mA 6 mA
76 kV 66 kV
Large
6 mA 6 mA
3D Sinus
Adult Child
66 kV 62 kV
Small
6 mA 6 mA
70 kV 64 kV
Medium
6 mA 6 mA
72 kV 66 kV
Large
6 mA 6 mA
66 kV 62 kV 74 kV 70 kV
Small Small
6 mA 6 mA 12 mA 10 mA
70 kV 64 kV 76 kV 72 kV
Medium Medium
6 mA 6 mA 12 mA 10 mA
72 kV 66 kV 80 kV 74 kV
Large Large
6 mA 6 mA 10 mA 10 mA
7. TROUBLESHOOTING
7.1. LEDs
The following table shows the LED that are present on the Generator
board, their significance and the recommended corrective actions in case
of defects. To locate the LEDs, refer to the layout of the Generator board
illustrated in chapter 10, drawing 13, of this manual:
The following table shows the LED that are present on the CPU board A5,
their significance and the recommended corrective actions in case of
defects. To locate the LEDs, refer to the layout of the CPU board A5
illustrated in chapter 10, drawing 4, of this manual:
NOTE:
* If all the LEDs are OFF, verify the column fuse F3; if it is OK, verify cable
X11 between column CPU board A1 and CPU board A5, on both sides.
Verify the column CPU board A1.
E x x x
P R E S S R E S E T
xxx code number of the error message
E x x x
C A L L T E C H A S S .
xxx code number of the error message
E x x x
S W I T C H O F F S Y S T .
xxx code number of the error message
The error messages are divided into different areas that can be distinguished by the
error number; the following table contains the different errors with meanings.
Rotation motor
Y motor
Sensor holder
X-ray Controls
CanBus
Vertical motor
Keyboard
All these are errors related to the main CPU board (A5) and its internal
peripheral.
Power off the unit and, after 1 minute delay, power it ON again; if the
error is displayed again, replace the Microprocessor CPU board (A6).
This error is shown when the firmware of the Rotograph EVO 3D does
not sense the presence of the U.I.C. (Unique Identification Code).
The unique code is read, with is check byte, from the control system at
the start-up; if the check byte is incorrect, the system displays the above
error number.
In case this error is displayed, a pressure of key " Patient entrance " (7)
allows the system to continue its functioning, but only standard base
examination will be possible.
In case there is a fault on the hardware key itself, it must be replaced. All
the optional features must be re-enabled with proper codes. Before
requesting a new hardware key, the S/N of the equipment and/or the
U.I.C. itself must be recorded and reported to Villa Sistemi Medicali.
This message means that the clock battery on main CPU board (A5) is
low or fault.
If after the power ON, a 90 second cooling time starts, wait until the end
of the time; then the display will show "E110 – Press >0<".
Follow the message shown on the display and perform an examination.
At the end of the examination, power OFF the machine and wait a couple
of minutes before powering ON again.
1. If the message is not yet present, it means that the battery is low.
Leave the machine powered ON to recharge it.
These errors are related to the various axis movements and require the
replacement of some parts; only the following error message E206 can be
generated by a special condition and can be reset.
T Z E R O [ f f ] a
Z E R O O F F S ± e e e e
1. If there is no variation of the above signals, if the arm does not move
or moves with difficulty or jumps:
• check the belt and verify that it is not broken; if the belt is loose,
adjust its tension
• check cable of stepper motor; there can be a short circuit or a
broken wire; check also for a loosen contact. In case of short
circuit, replace the cable, verifying also that no damage has been
caused to the motor driver on the CPU.
2. If the arm moves, verify the cable up to the CPU board (A5). If it still
continues to have no variation of the signals when there is an
activation of the switches, change the CPU board.
Y Z E R O [ f f ] a
Z E R O O F F S ± e e e e
Using the left and right arrows it is possible to rotate the arm, checking if
the status of the sensor changes accordingly.
1. If there is no variation of the above signal, if the arm does not move
or moves with difficulty or jumps.
• Check the belt and verify that it is not broken; if the belt is loose,
adjust its tension
• Check cables of the Y axes motor; there can be a short circuit or
a broken wire; check also for a loosen contact. In case of short
circuit, replace the cable, verifying also that no damage has been
caused to the motor driver (D27) on the CPU.
2. If there is no variation of the above signal and the arm moves, verify
the cable from the optical sensor up to the CPU board (A5). If still
there is no variation of the signals when there is an activation of the
switches, change the CPU board.
All the messages refer to the optical barrier that is sensing the zero
position of the Digital Ceph sensor holder; differences between their
meanings are that the first one (E280) may occur during the
examination, meaning that the senor activation has occurred before the
expected time. This can be caused by the loss of motion of the stepper
motor due to a collision with the patient or a fault of the motor stepper
driver.
In case that no collision has been observed or the message is frequently
displayed, follow the procedure.
The last 3 messages are displayed during the power on phase, when the
system is controlling the functionality of the whole system.
E281 and E282 mean that the zero position sensor is never activated;
due to a motor not running or to a fault of the optical barrier (the sensor
or its harness); E283 means that this sensor is always activated. Also in
this case the fault can be the motor or the optical barrier.
• Check that during the power on phase there are movements on these
axes; the normal motion is going to engage the sensor and a reverse
motion to disengage it. If no motion is observed, the fault can be the
stepper motor or the cable connecting it to the main CPU. Check that
the cables between the motor M8 and X6 of board A12 is well
inserted, as for the cable X1-X2 of A12 to X23, X24 and X34 of the
main CPU A5
• Check continuity of the above cables, from M8 to board A12 and from
X1-2 of board A12 up to X23, X24 and X34 of the main CPU (A5)
• Repeat the power on sequence, verifying the motion.
If still there is no motion, replace the stepper motor and, if not fixed,
replace the CPU.
If a motion has been observed, but messages are displayed, this can be
the sensor or its wiring:
• Power off the system and try to manually disengage the sensor,
taking care of a smooth motion.
• Power on the system and check the functioning of the sensor,
measuring the DC voltage on the terminal of the sensor B10, with
and without an obstacle manually inserted. A variation of about
3.5 V minimum should be observed. If not, replace the sensor.
• If the variation is observed and the message still continues to be
displayed, check the correct insertion of cable X5 on A12 board and
from X1-2 of the board up to X23, X24 and X34 of the main CPU;
check the continuity of the interested wires and replace the faulty
one (if any).
• If all cables are OK, the fault is on the CPU, so replace it.
(Rev. 0) 7-11 Rotograph EVO 3D
SERVICE MANUAL
Troubleshooting
The X-ray beam coming out from the tube head assembly is collimated to
the area under exam by a secondary collimator that is moving
synchronously with the CEPH sensor holder; this collimator is moved by
a stepper motor M7.
All the above messages refer to this mechanism; all the considerations
above described for E280 to E283 for patient’s collision.
In case of absence of collision between the secondary collimator and the
patient, the cause can be a fault on motor M7 and/or optical sensor
barrier B11.
• Check that during the power on phase there are movements on these
axes; the normal motion is going to engage the sensor and a reverse
motion to disengage it. If no motion is observed, the fault can be the
stepper motor or the cable connecting it to the main CPU (A5). Check
that the cables between the motor M7 and X4 of board A12 is well
inserted, as for the cable X1-X2 of A12 to X23, X24 and X34 of the
main CPU.
• Check continuity of the above cables, from M7 to X4 board A12 and
from X1-2 of A12 up to X23, X24 and X34 of the main CPU A5.
• Repeat the power on sequence, verifying the motion.
If still there is no motion, replace the stepper motor and, if not fixed,
replace the CPU.
If a motion has been observed, but messages are displayed, this can be
the sensor or its wiring.
• Power off the system and try to manually disengage the sensor,
taking care of a smooth motion.
• Power on the system and check the functioning of the sensor,
measuring the DC voltage on the terminal of the sensor B11, with
and without an obstacle manually inserted. A variation of about
3.5 V minimum should be observed. If not, replace the sensor.
• If the variation is observed and the message still continues to be
displayed, check the correct insertion of cable X3 on A12 board and
from X1-2 of the board up to X23, X24 and X34 of the main CPU;
check continuity of the interested wires and replace the faulty one (if
any).
• If all cables are OK, the fault is on the CPU, so replace it.
Rotograph EVO 3D 7-12 (Rev. 0)
SERVICE MANUAL
Troubleshooting
All the messages refer to the optical barrier that is sensing the zero
position of the primary beam collimator, that is moved by a stepper
motor. This mechanism is located in front of the tube head assembly. The
slit collimator does not move during the PAN examination, while it has a
scanning motion during Ceph exams, synchronously with the secondary
collimator and CEPH sensor holder.
Also in this case, the first one (E320) may occur at the end of an
examination, while the system is checking the correctness of the motion,
meaning that the senor activation has occurred before the expected time.
This can be caused by the loss of motion of the stepper motor.
The last 3 messages are displayed during the power on phase, when the
system is controlling the functionality of the whole system.
E321 and E322 mean that the zero sensor position is never activated;
due to a motor not running or to a fault of the optical barrier (the sensor
or its harness); E323 means that this sensor position is always activated.
Also in this case the fault can be the motor or the optical barrier.
• Power off the system and remove the frontal cover of the tubehead
assembly.
• Check that during the power on phase there are movements on the
primary slit collimator; the normal motion is going to engage the
sensor and a reverse motion to disengage it. If no motion is observed,
the fault can be the stepper motor M5 or the cable connecting it to
the main CPU. Check that the cables between the motor M5 and X64
is well inserted, as for the cable X64 to X29 and X30 of the main CPU
A5.
• Check continuity of the above cables, from M5 to X64 and from X64
up to X29 and X30 of the main CPU A5.
• Repeat the power on sequence, verifying the motion.
If still there is no motion, replace the stepper motor and, if not fixed,
replace the CPU.
If a motion has been observed, but messages are displayed, this can be
the sensor or its wiring.
• Power off the system and try to manually disengage the sensor, taking
care of a smooth motion.
• Power on the system and check the functioning of the sensor,
measuring the DC voltage on the terminal of the sensor B8, with and
without an obstacle manually inserted. A variation of about 3.5 V
minimum should be observed. If not, replace the sensor.
Close the PAN sensor holder to its position and press the Patient
centering key to restart the power on procedure.
This message is displayed if, during the power on procedure, the X-ray
button, local (E360) or remote (E361) have been sensed as pressed.
Release the button if it was pressed: the error condition is reset powering
off the unit and on again.
If the error is still present, check the continuity of cables of the interested
button. For the local one, S8, it is connected to X18 of the keyboard and
it is transferred to the main CPU using the CANBus cable.
• Check that the cable X18-X18 to the main CPU is well inserted and
its continuity.
• From the main CPU, the signal is routed to the Generator board
(A10), using the CANBus cable X20-X20; check also that this cable is
well inserted and its continuity. Replace it if damaged.
• If the error is still present, check the correct insertion of the flat cable
between the Generator CPU board (A9) to the Generator board (A10).
The cable is labelled X53-X53; replace it if damaged.
WARNING:
On the Generator board (A10) there are dangerous high voltages,
230 VAC and 400 VDC.
Before accessing the Generator CPU and power board (A9 and A10),
it is mandatory to switch off the mains and wait up to 2 minutes in
order to allow the discharge of the capacitor.
7.2.3.6. E380: Invalid CANBus message (from Generator CPU board A9)
This error is displayed when the main CPU board (A5) receives an invalid
message from the Generator board (A10); that is a message which
contents is not listed on the valid data dictionary of the unit.
This can be generated by a bad connection or by a fault of the local
Generator CPU board (A9).
Check the CANBus cable X20-X20 and replace it if faulty, otherwise
replace the Generator CPU board or then the Generator board.
These two messages have the same meaning, that is a no answer to main
CPU from CAN messages, but their occurring is in different times.
E381 is generated during power on, while E382 is displayed after a
regular power on sequence.
In both cases, the error can be generated by a bad connection of the
CANBus or a fault on the local Generator CPU board or a foult on the
Generator board, so the steps above described for E380 are applied.
• Power off the system and remove the frontal cover of the tubehead
assembly.
• Check that during the power on phase there are movements on the
primary slit collimator; the normal motion is going to engage the
sensor and a reverse motion to disengage it. If no motion is observed,
the fault can be the stepper motor M6 or the cable connecting it to
the main CPU. Check that the cables between the motor M6 and X65
is well inserted, as for the cable X65 to X29 and X30 of the main CPU
A5.
• Check continuity of the above cables, from M6 to X65 and from X65
up to X29 and X30 of the main CPU A5.
• Repeat the power on sequence, verifying the motion.
If still there is no motion, replace the stepper motor and, if not fixed,
replace the CPU.
If a motion has been observed, but messages are displayed, this can be
the sensor or its wiring.
• Power off the system and try to manually disengage the sensor,
taking care of a smooth motion.
• Power on the system and check the functioning of the sensor,
measuring the DC voltage on the terminal of the sensor B9, with and
without an obstacle manually inserted. A variation of about 3.5 V
minimum should be observed. If not, replace the sensor.
• If the variation is observed and the message still continues to be
displayed, check the correct insertion of the cable from B9 in X33 of
the main CPU and the continuity of the connection.
• If all cables are OK, the fault is on the CPU, so replace it.
WARNING:
Those errors are related to the X-ray generator, so they can be safety
related.
In case of Error messages E759 and E755, the system must be
immediately powered off, because there is an unexpected emission
(E759) or the emission has not been terminated into the expected
time.
WARNING:
On the Generator board (A10) there are dangerous high voltage,
230 V AC and 400 VDC.
Before accessing the Generator CPU and power board (A9 and A10),
it is mandatory to switch off the mains and wait up to 2 minutes in
order to allow the discharge of the capacitor.
This message is signalling that the Generator board (A10) is not powered.
This message is displayed when a value higher than expected has been
detected on the Generator board (A10).
Check that connector X57 is well inserted. With the connector inserted,
measure the resistance between pins 2 and 4 of connector (harness to
tubehead) that should be 13.3k ± 2% while is 14.3k ± 2% between pins 4
and 3 of the same connector; if correct values are measured, replace
before the tubehead and, eventually, the Generator board.
If incorrect values are measured, replace the tubehead and then the
Generator board.
This message is signalling that the PFC (Power Factor Control) circuit is
not correctly functioning.
When the CPU command for the emission, the X-ray on signal is set to on
condition when it reaches the 75% of its set value at the beginning and is
higher than 25% at the end of emission.
If the first condition is not met, the exam is halted and the system has to
be reset, and the message E758 "No X-ray" is displayed.
In this case, check the correct insertions of X56 and X57 between the
Generator board to the tubehead and X53-X53 (flat cable) between
Generator board (A10) and Generator CPU board (A9).
Check the CANBus connection between the Generator CPU board (A9),
and the main CPU A5 (cable X20-X20).
If all connections are OK and the fault is repetitive, same as for Led H1 in
chapter 7.1.1.
If the second condition is not met, that is, the X-ray emission do not fall
under the value in a pre set time, the message E759 "Unexpected
emission" is displayed. This message can be also displayed if the signal
X-ray on is sensed during the idle state.
In this case, check the main CPU board (A5) and the Generator CPU
board (A9) and the cable X20-X20. If all OK, replace the Generator board.
This message is displayed when the Generator CPU board (A9) is not
detecting the RX button during the emission start-up.
In this case, check the correct insertions of X53-X53 (flat cable) between
Generator board and Generator CPU board (A9).
Check the CANBus connection between the Generator CPU board (A9)
and the main CPU A5 (cable X20-X20).
If the error is still present, check the cable of the RX button. It is
connected to X18 of the keyboard and it is transferred to the main CPU.
If all connections are OK and the fault is repetitive, replace the Generator
CPU board (A9).
This message is displayed when the Generator CPU board (A9) is not
detecting the X-ray emission feedback signal.
In this case, check the correct insertions of X53-X53 (flat cable) between
Generator board and Generator CPU board (A9).
Check the CANBus connection between the Generator CPU board (A9)
and the main CPU A5 (cable X20-X20).
Check the correct insertion of X56 and X57 between the Generator board
to the tubehead.
If all connections are OK and the fault is repetitive, replace the Generator
CPU board (A9).
This error message is displayed when the Generator CPU board (A9) is
not detecting the X-ray button pressed also if the main CPU (A5) has
commanded the emission with the corresponding CANBus message.
In this case a possible interruption on cable X20-X20 from main CPU A5
to control CPU A9 is possible.
From the main CPU, the signal is routed to the Generator board, using
the CANBus cable X20-X20; check also that this cable is well inserted
and its continuity. Replace it if damaged.
If the error is still present, check the correct insertion of the flat cable
between the Generator CPU board to the Generator board. The cable is
labelled X53-X53; replace it if damaged.
This message has the same meaning as the corresponding error E362,
but it is generated by the Generator CPU board (A9), that is signalling a
possible broken connection with the main CPU (A5).
Repeat the test as per error E774 above.
This error is displayed when there is no answer to main CPU board (A5)
from CAN messages during power on.
The error can be generated by a bad connection of the CANBus or a fault
on the Column CPU board (A1).
Check that the column movement power supply fuse F4 is not blown.
Check the safety column movement microswitches work correctly,
otherwise replace the microswitches assy (microswitch + cable).
CANBus cable X11-X11 and replace it if faulty, otherwise replace the
column CPU board.
Check on the Power HUB board (A21) the LEDs H1, H2, H3 and H4
status; they should be all ON.
If LED H1 is OFF, disconnect the cable X84-X75 (X84 side) and verify if
the LED switch ON; if it doesn't light change the cable (there is likely a
short circuit between the cable pins), otherwise replace the Power HUB
board.
If one of the other LEDs is OFF, replace the Power HUB board.
If all the LEDs of the Power HUB board are ON, check the connections on
the VDSIB board, mainly the cable X84-X75 (X75 side); if necessary
replace the faulty cable.
Otherwise, the problem can be related to a bad syncronization of the
VDSIB board, in this case:
1. Switch OFF the unit
2. Disconnect all the VDSIB cables
3. Turn ON the unit
4. Reconnect the VDSIB cables
5. Switch OFF the unit and ON again to reboot the system and restore
communication.
During the power on phase, the local control board of the keyboard is
controlling that during power on no more than 1 button is pressed.
Only the case of column up/column down is allowed and used to enter
the set up procedure, so if this procedure is started without an explicit
request, it means that those buttons are pressed.
In this case, replace the keyboard membrane and after the control board.
E 8 5 0 ( x x x x x x x x )
C A L L T E C H A S S .
xxxxxxxx code number for message 850
This message is signalling that only one of the two buttons is pressed.
Release the pressed button and power off the system and power on it
again. If the message is displayed again, replace first the keyboard
membrane and after the control board.
When this message appears, press the column up and column down
NOTE:
* SYSTEM and BURN-IN operations are used only during factory set-up;
there are special passwords not accessible to Service Engineers. For this
reason, they are not explained on the manual.
In all menus of the service programs, the keyboard keys provide the
following usage.
S E R V I C E M E N U
C O N F I G U R A T I O N
Pressing the Patient centering key (>0<) the next message is displayed:
S E T U P
P A S S W O R D ? 1 0 0
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
7.3.3.1. Language
In this menu, it is possible to adjust the internal Real Time Clock (RTC);
this RTC is used to check the correctness of the Cooling Down time of the
tubehead.
NOTE:
* It will take care of standard operation (hour and date change, leap year
calculation, etc.) but does not consider Summer Time Hour Changes.
The RTC clock is set at the Factory hour/date (Central Europe date/time)
during the final test of the equipment. It is the installer’s responsibility to
set, if wanted, the local date/time.
D A T E - T I M E S E T
D A T E : D D / M M / Y E A R
A D J U S T D A T E
D > X X M Y Y Y Y Y
A D J U S T D A T E
D X X M > Y Y Y Y Y
A D J U S T D A T E
D X X M Y Y Y > Y Y
NOTE:
* If the emission is OFF, a warning message will be displayed just after the
initial message one and before the TEST message. The display is cleared
by pressing the Patient centering (>0<) key.
These selection allow the set of brightness and contrast of the display in
order to set those as a convenient level for the operators.
The display brightness will decrease with higher value; the factory set
value is set to the maximum level.
The "clicking" sound of the pressed key can be set ON/OFF; once set to
ON, also its volume can be adjusted at the appropriate volume.
F F F F F F F F F F F F F F F F
V A L U E [ 0 ] = F F
Where VALUE[x] is the value assigned to the couple of digits, with "x"
ranging from 0 to 7, and "FF" is the hexadecimal value assigned to the
specific couple. Using the standard buttons to move into the selected
field, set its value equal to the one received for the appropriate position.
Values are displayed and changed using a couple of digits, with a
variation of one unit for each pressure of the parameter
increase/decrease key; the auto repeat function for fast changes is
enabled.
In this menu, it is possible to set the system for EUROPEAN tooth style
or AMERICAN tooth style.
After the selection, the display will show the following message:
T O O T H S T Y L E O P T
E U R O P E A N S T Y L E
or
T O O T H S T Y L E O P T
A M E R I C A N S T Y L E
keys .
In this menu, it is possible to set the system for digital or film base; the
WARNING:
Changing the set mode will affect the complete system
functionality.
This parameter has to be set as "3D SYSTEM" for PAN CEPH units and
"WITHOUT SYSTEM" for PAN only units.
NOTE:
* For PAN only system, changes on this menu do not have any effect on
system functionality.
This menu allows to set the system for 1 mobile sensor or 2 fixed sensors
configuration.
After the selection, the display will show the following message:
S E N S O R H A N D L I N G
S E N S O R I S M O B I L E
or
S E N S O R H A N D L I N G
S E N S O R I S F I X E D
WARNING:
Wrong setting provide operative messages not consistent with the
equipment configuration.
The DSPU has its own static IP address, that is used for the connection
with the external PC. The default set value for IP address is
192.168.002.215.
The DSPU has also a specific net mask, default set value of
255.255.255.0000.
Using the standard method to change the value, set both according to
values defined by the Network manager of the installation.
The QSD has its own static IP address, that is used for the connection
with the PC. The set value for this IP address is 192.168.002.045.
The QSD has also a specific net mask, default set value of
255.255.255.000.
Using the standard method to change the value, set both according to
values above mentioned.
S E R V I C E M E N U
C O N F I G U R A T I O N
Using the left/right scroll key is possible to change the message, use the
keys until to the following message is visualized:
S E R V I C E M E N U
A X I S A L I G N E M E N T
Pressing the Patient centering (>0<) key the next message asking for the
correct password is shown.
S E T U P
P A S S W O R D ? 1 0 0
and confirm with the Patient centering key . Before entering the
NOTE:
* Interruption of this operation must be performed only in case those
functions are used to check some hardware faults. If the system needs to
be centered, the operation must be completed.
The exit from each menus is performed using the Test (T) button; the
value is updated or not by pressing the appropriate button as an answer
to the following message:
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
WARNING:
Changes to one or more offset values affect the system’s
functionality, so take care to not alter those values if not needed.
Normally only the replacement of faulty parts (motors, belts, zero
optical sensors, etc.) will require to act on those data.
In this menu, it is possible to set the offset of rotation axis in order to set
the correct starting value of emission. Once the Patient centering (>0<)
key is pressed, the following message is displayed:
T Z E R O [ Y Y ] a
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
• "xxxx" is the current value of the offset, that can be changed with the
left/right keys.
This function allows the set of the offset of Y axis, the motion along mid
sagittal plane of the patient, in order to place the correct position of the
central path. Once the Patient centering (>0<) key is pressed, the
following message is displayed:
Y Z E R O [ Y Y ] a
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
• "xxxx" is the current value of the offset, that can be changed with the
left/right keys.
This function is not active on the digital mode; if selected it will generate
an error message.
WARNING:
In this menu the emission is enabled; it starts with the X-ray button
press, so take care of this situation.
This function allows to center the offsets of the linear slit collimator.
Once the Patient centering (>0<) key is pressed, the following message is
displayed:
P C C Z E R O [ Y Y ] a
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
• "xxxx" is the current value of the offset, that can be changed with the
left/right keys.
be set by pressing the key ; when it has been pressed, the normal
NOTE:
* The position of the primary slit is set to the correct value during the
factory final test; changes on the stored value must be done only in case
of:
• Replacement of one part of the linear slit collimator (motor, zero
barrier sensor, etc.); in this case the beam must be aligned to the
sensor entrance acting on the offset value.
• Replacement of a defective tubehead; in this case, the beam aligned
moving the tubehead leaving the offset value unchanged.
Set this parameter with the same value used to set "Primary collimator
PAN".
NOTE:
* The complete procedure to enter in password 124 is described on
paragraph 7.3.1 and following.
S E R V I C E M E N U
C E P H S E T T I N G S
Pressing the Patient centering (>0<) key the next message is displayed,
asking for the correct password.
S E T U P
P A S S W O R D ? 1 0 0
and confirm with the Patient centering key. The system will begin
D I G I T A L C E P H
E N A B L E S E N S O R
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
D I G I T A L C E P H
C E P H I S E N A B L E D
or
D I G I T A L C E P H
C E P H I S D I S A B L E D
increase/decrease keys.
Pressing the key will quit the menu and the disabled/enabled
7.3.5.2. Y Offset
WARNING:
Do NOT modify this parameter: functionality can be severely
impaired.
This function allows the set of the offset of Y axis, the motion along the
mid sagittal plane of the patient, but its function is to center the X-ray
beam on the ceph sensor. Once the Patient centering (>0<) key is
pressed, the following message is displayed:
Y Z E R O [ Y Y ] a
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
• "xxxx" is the current value of the offset, that can be changed with the
left/right keys.
In this menu, it is possible to set the offset of rotation axis in order to set
the correct starting value of emission. Once the Patient centering (>0<)
key is pressed, the following message is displayed:
T Z E R O [ Y Y ] a
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
• "xxxx" is the current value of the offset, that can be changed with the
left/right keys.
This menu allows to adjust the zero offset for the secondary collimator
(the collimator that is used to limit the X-ray beam before patient on
CEPH).
When the display is showing the following messages, using the left/right
scroll keys selects the message
D I G I T A L C E P H
E N A B L E C E P H
D I G I T A L C E P H
C E P H S . O L L . Z E R O
C Z E R O [ Y Y ] a b c d
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "xxxx" is the zero offset actual value, that can be changed with the
left/right keys.
This menu allows to adjust the zero offset for the digital sensor.
When the display is showing the following messages, using the left/right
scroll keys selects the message
D I G I T A L C E P H
E N A B L E C E P H
D I G I T A L C E P H
C E P H S E N S O R Z E R O
S Z E R O [ Y Y ] a b c d
Z E R O O F F S ± x x x x x
where:
• "YY" is the stepper motor speed, that can be changed using the
• "xxxx" is the zero offset actual value, that can be changed with the
left/right keys.
This program is used to adjust the offset of the Soft Tissue motorized
filter.
S E R V I C E M E N U
T R O U B L E S T O O T I N G
Pressing the Patient centering (>0<) key the next message is displayed:
S E T U P
P A S S W O R D ? 1 0 0
NOTE:
* Of these submenus, actually only the Test Column, Test Input Ports,
PAN RX emission, CEPH RX emission, Pre-heating time and Preheating
level are active.
C O L U M N T E S T
I D L E 0 0 0 a b
Using the column up/column down keys verify that the last two
This function is used to carry out diagnostics at low level of the various
input signals of the CPU board A5. It may be used, for example, to verify
the limit switches without necessarily moving the axes but manually
activating the optical sensor / microswitches.
T E S T I N P U T P O R T S
R y 7 6 5 4 3 2 1 0
where:
• Ry= selected port code (R4, R5, R6, R7)
• 76543210= logic status of the port inputs.
NOTE:
* The following table describes the correspondence between the displayed
figures and the corresponding input signal. The digits are numbered from
7 to 0. The 7 digit corresponds to the character furthest to the left on the
display, while digit 0 corresponds to the character furthest to the right.
µSwitch /
Port Bit Association Logic status
Optical sensor
R4 7 not used -
6 not used -
0 for Analog
5 DSPU board initialization
1 for Digital
4 Motor +5V power supply (0 = alarm status) 1
3 Column CPU board A1 initialization 1
2 Generator board A10 initialization 1
1 not used -
0 not used -
R5 7 (*) Primary collimator position optical sensor B8 0
6 (**) Soft tissue filter 0 position B8 1
5 not used -
4 not used -
3 not used -
2 (**) Digital Sensor ready 0
1 not used -
0 not used -
R6 7 (*) Panoramic cassette start position B5 1
6 (*) Panoramic cassette end position B4 1
5 (*) CEPHALOMETRIC cassette present S10 0
4 (*) PANORAMIC cassette present S9 0
3 Cassette / Sensor holder in B7 1
CEPHALOMETRIC position
2 Cassette / Sensor holder in PANORAMIC B6 1
position
1 Remote control X-rays button X71-1 / X71-2 1
0 X-rays button on column S8 1
R7 7 not used 1
6 (**) Ceph sensor holder start position B11 1
5 not used 1
4 (**) Ceph secondary collimator start position B10 1
3 (**) Slit primary collimator 0 position B9 1
2 Y axis start position B1 1
1 not used -
0 Rotation arm 0 position B3 1
(*) Functionality only on Analog version
(**) Functionality only on Digital version
NOTE:
* The logic status of these signals depends on the physical position of the
relevant optical sensor / microswitch with respect to the "0" position.
WARNING:
Do NOT modify this parameter: functionality can be severely
impaired.
This function allows the set of the Tubehead pre-heating time. After the
selection, the display will show the following message:
P R E - H E A T I N G T I M E
V A L U E = 2 0 0 0
keys .
I D L E F I L L E V E L
L E V E L = 3 0
NOTE:
* DO NOT modify the "LEVEL" value in this step.
F I L L E V E L @ X X m A
L E V E L = Y Y Y
where:
• "XX" is the Tubehead filament current value, that can be changed
These menus is used to test the X-ray emission excluding all interlocks
that are active during examination. It is used to test the function of the
X-ray generator and tube head assembly.
At the selection of the X-ray emission test, the display will show the
following message
7 2 k V 0 6 m A 1 . 0 0 s
X - R A Y P A R A M E T E R S
WARNING:
In this menu the emission is enabled; it starts with the X-ray button
press, so take care of this situation.
> 7 2 k V 0 6 m A 1 . 0 0 s
case the kV. To move to mA and exposure time, press key once or
twice again.
In this situation, the keys increase and decrease will change the
parameter.
Pressing the X-ray button will start the emission.
Both those menus are used only for factory set and testing, so they are
not available for field service.
In any case, accession to those is protected by a special password, to
avoid the risk of dangerous operations.
This menu allows to show all parameters with its own set value; there is
no possibility to change the parameters. This menu is useful to record on
a paper sheet the condition of the machine.
Its usage is suggested at the end of installation to review and record the
parameters.
Using the right/left scroll keys, it is possible to view all parameters.
NOTE:
* The last displayed value is the Hardware Protection Key code; this is an
alternate method to view it.
8. PERIODIC MAINTENANCE
As with all electrical appliances, this unit must be used correctly and
maintenance and inspections must be made at regular intervals. Such
precautions shall guarantee the safe and efficient function of the
appliance.
Periodic maintenance consists in inspections made directly by the
operator and/or Technical Service Department.
The inspections made directly by the operator are the following:
• ensure that the rating labels are intact and correctly fastened
• check whether there are any traces of oil on the tubehead
• check to ensure that the X-rays push-button switch cable is not split
or damaged
• check to ensure that there is no external damage to the appliance
which could jeopardise protection from radiation.
WARNING:
The operator is recommended to perform these inspections before each
session of operations.
If the operator detects irregularities or damage, he should immediately
inform the Technical Service Department.
WARNING:
Components may only be replaced by original spare parts.
NOTE:
* The Service Engineer has to take special care for all what concerns
electrical safety of the device and must make sure of restoring all
provisions for electrical safety which may be affected during a service
intervention and to solicit the customer to have the electrical safety tests
repeated every time the intervention has caused the replacement of
important parts or the intervention has significantly affected safety
provisions of the device.
5209900900 Digital sensor centering tool Used for Sensor calibration and
5607900800 Copper filter for Digital sensor Cephalometric arm checks
5807304100 Calibration programs CD
5. From the task bar menu, click with the right button of the
mouse on Villa icon and select "Close Control Panel".
Figure 8-1
6. From desktop open the software ViVA that permit the direct control
of the 3D digital sensor.
Figure 8-2
Figure 8-3
10. Flag the parameter "User Sync" in order to manage manually the
exposure.
Figure 8-4
Figure 8-5
12. Set Rotograph EVO 3D in 3D dentition with 76kV 10mA. Position the
Cu filter of 1.5mm thickness on the primary collimator. Start X-ray
and when the emission starts, press on "Continue" to start
acquisition for Gain calibration.
Figura 8-6
Figure 8-7
14. At the end it will be possible to check the Gain median value that
should be between 1600 and 4000. In case the value is not within the
range, it will be necessary to repeat the calibration selecting the
exposure parameters that provide the correct median value.
Figure 8-8
15. Repeat the calibration selecting the other 3D modalities which are
"2x2 4pF G4 30fps" and "1x1 0,5pF G4 10fps".
Figure 8-9
Figure 8-10
Figure 8-11
Figure 8-12
19. Set Rotograph EVO 3D in PAN with 60kV – 6mA removing the Cu
filter used to calibrate the 3D modality. Start X-ray and when the
emission starts, press on "Continue" to start acquisition for Gain
calibration.
20. At the end it will be possible to check the Gain median value that
should be between 1600 and 4000. In case the value is not within the
range, it will be necessary to repeat the calibration selecting the
exposure parameters that provide the correct median value.
Figure 8-13
21. At the end of calibration verify that the calibration files present in the
directory "C:\IMAGERs\S/N of the detector\working modality" have
been updated.
Figura 8-14
9. CORRECTIVE MAINTENANCE
In this chapter are included the operation necessary after the
replacement of the system main components.
The Pleora MAC address is indicated in the test report of the digital panel
supplied as a spare part.
It is necessary to write the correct address in the file "HcpConfig.ini"
(using notepad) included in the directory "C:\IMAGERs\sensor s/n":
[ReceptorControl]
FgDll=HcpFgExPleora
RpDll=HcpRp1313
[HcpConfigFiles]
HcpFile_01 = *.dat
HcpFile_02 = vivacy.xml
[ImageAcquisition]
FrameStartAvailable=0
[HcpFgPleora]
PleoraSerialBaudRate = 9600
PleoraMacAddress=00-11-1C-00-8C-CC
[VirtCp]
FileRev=3
In order to set the new digital sensor 3D address, once is opened the
virtual keyboard (by pressing TAC), select the icon "Utilities".
Figure 9-1
Figure 9-2
Set the correct S/N of the 3D digital sensor using the button "1" and
verify that the relevant "MAC Address" is the same you received on the
test report.
1. Enter in "Y Axis zero" menu; the primary collimator will move to the
panoramic position.
rotating arm will move to 90° position and will be ready to start X-ray
emission.
4. Run the ViVa program and verify if the S/N of the sensor is the
correct one.
Figure 9-3
In case of digital sensor replacement, select "No" and set the correct
sensor S/N selecting the function "Receptor Setup" in the
"Acquisition" drop-down menu.
Figure 9-4
It will be possible to select the S/N of the detector (if it has been
previously installed in the directory "C:\IMAGERs") in the "Receptor
Serial #" window.
Figure 9-5
Figure 9-6
Figure 9-7
6. Move the small black square present in the upper window of the
screen to the lower one and stop acquisition with the red square
present in the commands on the right side of the screen. Confirm
with "Yes".
Figure 9-8
NOTE:
* When you replace digital sensor, in order to center the X-ray beam move
ONLY the sensor. In case you replace the tubehead, move ONLY the
tubehead. It is very important to have a fixed point for the beam
centering
Figure 9-9
• In case the reference pins are aligned and centered on the window,
but the X-ray beam is not centered, it is necessary to move the
tubehead acting on the screws "4" (Figure 9-10).
• In case the reference pins and the X-ray beam are centered in the
window but there is a vertical misalignment of the image (left right
side), it is necessary to move the primary collimator acting on the
screws 1, 2 and 3 (Figure 9-10).
Figure 9-10
Figure 9-11
At the end of the adjustment, the image must be well centred in the
green window.
Figure 9-12
In case the 3D image doesn’t provide a continuous arc, but the edge line
has some interruptions, may be necessary to check the setting of the
Lateral Offset parameter.
Decrease the parameter Lateral Offset Increase the parameter Lateral Offset
Figure 9-13
Figure 9-14
NOTE:
* The only parameter allowed to be modified is the Lateral Offset. Changes
to one of the other parameters may effect the system’s functionality.
The Lateral Offset parameter has to be set within the range of +/- 2
maximum.
WARNING:
The board shipped as replacement carry the Hardware key and the
EEPROM not configured.
To make the system working, the Hardware key A13 (code 5807302700)
must be retrieved from the failed board and positioned on the new board.
This component includes the U.I.C. (Inique Identification Code) witch
determines the enabling codes for the radiological exams.
NOTE:
* At the end of the replacement, restore the metallic cover and the ground
connection. Both parts has to be recovered from the failed board.
D19
A13
Figure 9-15
9.5. PC replacement
In case of PC replacement, it will be necessary to install the SW and set
properly the communication IP addresses in order to permit the system
communication.
Figure 9-16
At the first use of the SW, it will be required the name of the user. This
name will be always stored in the USB key.
In case you need to make some tests, without installing the user name,
you can select "Close" without writing any name. The Dicom images will
have the warning of "Unregistred license. Only for evaluation" but it will
be possible to set the user’s data at the next initialization (on customer
site).
Figure 9-17
NOTE:
* Application Data is an hidden folder, so be sure that Microsoft Explorer
has the option "Display Hidden File and Folders" active.
• WINDOWS XP
C:\Documents and Settings\ All Users\
Application Data\OWANDY\OSP - LAN PANORAMIC\
Example:
In the folder:
C:\Documents and Settings\All Users\Application Data\OWANDY\
OSP - LAN PANORAMIC\
you must have the following folder:
- 3_KAS21126.
Figure 9-18
Figure 9-19
For the CEPH digital sensor, it is necessary to set the address in the
following mask selecting as device "Rotograph Evo".
Figure 9-20
You have to set the IP address in the computer according to the one
stored in Rotograph EVO 3D selecting in "My network", with the right
button of the mouse, the board Intel Pro 1000 and, selecting with the
right button "Properties" and changing the IP address.
Example:
In Rotograph EVO 3D the default address is 192.168.002.215 (subnet
mask 255.255.255.0), you have to set the address in the PC to
192.168.002.16 and the same subnet mask 255.255.255.0.
NOTE:
* When these operations are performed during Maintenance it will be
necessary to remove some of the covers.
NOTE:
* This adjustment needs a personal computer directly connected to
Rotograph EVO 3D where the "DentalStudio" program is installed and
the use of the CD (P/N 5807304100) supplied with the centering tools
kit.
Using the increase key (4) and the decrease key (5) set the
password equal to 124 and confirm with the "Patient entrance" (7)
M A C H I N E S E T T I N G
P R E S S > 0 <
2. Press key (7) . The machine will move and the following
W A I T F O R
followed by
C E P H P O S I T I O N I N G
P L E A S E W A I T
D I G I T A L C E P H
E N A B L E S E N S O R
D I G I T A L C E P H
C E P H I S D I S A B L E D
4. Using increase key (4) and the decrease key (5) select the
D I G I T A L C E P H
C E P H I S E N A B L E D
5. Press key "Test" (6) and if the setting has been changed the
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
will be displayed:
D I G I T A L C E P H
E N A B L E S E N S O R
WARNING:
Do NOT modify the "Digital CEPH Y offset" of Password 124:
functionality can be severely impaired.
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
D I G I T A L C E P H
R O T A T I O N O F F S E T
displayed:
P L E A S E W A I T . . .
while the slit primary collimator moves to the DIGITAL CEPH central
position.
Then the following message will be displayed:
T Z E R O [ f f ] a
Z E R O O F F S ± e e e e e
decrease (5) set values (kV and mA) for the exposure
machine keyboard; the system gets ready to take the Rotation Arm
alignment test, moving the sensor to a correct position.
NOTE:
* If the following message will be displayed:
D I G I T A L S E N S O R
I S N O T R E A D Y
Press key (7) to reset the message and press key (11) again.
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
13. Press the X-ray button and keep it depressed until the end of the
exposure.
Figure 9-21
15. If it will be necessary to change the height of the CEPH arm, loose the
two screws "A" (Figure 9-22), and acting on the screw "B" adjust the
height of the Ceph group. Repeat the exposure until the vertical
alignment is reached. Tighten the loosened screws "A".
Figure 9-22
16. Check on the image if the Ear Centering Circles are concentric (small
circle inside the big circle – i.e. as shown in Figure 9-21).
17. To adjust the vertical alignment between the two rings, loose the two
screws "F" (Figure 9-22) and adjust the position of the arm acting on
screw "G" (Figure 9-23). Once the aligned position has been reached
(test exposures are required), tighten bolt "H" (Figure 9-23) and
tighten screws "F".
The performance of this adjustment could require the reiterate of
step 15.
G
H
Figure 9-23
With keys increase (4) and decrease (5) set the speed value
turn the tubehead arm, increasing the displayed offset value if the
left distance is lower than the right one or viceversa.
19. In order to reset the unit to the new parameter, exit from
will be displayed:
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
Press key (7) to store the changes and press key "Test"
21. Take a new image and repeat the test from point 18 until the circle is
well centered.
22. To adjust the horizontally alignment between the two rings, loose the
two screws "C" (Errore. L'origine riferimento non è stata trovata.)
and rotate the ear rings support group acting on screw "D". Once the
aligned position has been reached (test exposures are required),
tighten screws "C".
Figure 9-24
23. Measure the distance between the borders of the image and the
borders of the central vertical stripe (
Figure 9-25
The right and left distances must not differ more than ±3mm.
24. In case it is not centered, exit from Rotation Offset menu pressing
25. Using keys "Arrow right" (13) or "Arrow left" (12) select
D I G I T A L C E P H
C E P H S E N S O R Z E R O
26. Press key "Patient entrance" (7) . The display will show:
S Z E R O [ f f ] a
Z E R O O F F S ± e e e e e
27. With keys increase (4) and decrease (5) set the speed value
move the CEPH Sensor, increasing the displayed offset value if the
left distance is higher than the right one or viceversa.
28. Press key "Test" (6) to exit from the Ceph Sensor Zero
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
T Z E R O [ f f ] a
Z E R O O F F S ± e e e e e
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
30. Press the X-ray button to take a new image and repeat the test from
point 20 until the image is well centered.
31. Once the aligned position has been reached, press key "Test" (6)
to exit from the Rotation offset menu and return in the main menu:
D I G I T A L C E P H
R O T A T I O N O F F S E T
33. Press keys "Arrow right" (13) or "Arrow left" (12) until
34. Press key "Patient entrance" (7) and the following message will
be displayed:
C Z E R O [ f f ] a b c d
Z E R O O F F S ± e e e e e
35. Press key " Patient entrance " (7) on the machine keyboard.
Figure 9-26
(Rev. 1) 9-29 Rotograph EVO 3D
SERVICE MANUAL
Corrective maintenance
Figure 9-27
Centering tool P/N 5209900900
and exposure time and use keys increase (4) or decrease (5)
to set values for the exposure (suggested values: 60kV, 6mA and
0.5s).
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
39. Take X-ray pressing the X-ray button and check if in the obtained
narrow image the projection of the slit of the centering tool is vertical
and centered with the arc (Figure 9-28 and Figure 9-29).
Figure 9-28
Figure 9-29
40. If it is not the case, with keys increase (4) and decrease (5) set
displayed offset value if the left distance is lower than the right one or
vice versa; repeat the test from point 30.
NOTE:
* This is a fine setting; if the zero offset (±eeeee) is lower than –1000 steps
or greater than 1000 steps repeat the first step of ceph arm centering
described in 9.6.1.
41. When the right setting is reached press key "Test" (6) to exit
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
cancel them.
NOTE:
* This adjustment needs a personal computer directly connected to
Rotograph EVO 3D where the "DentalStudio" program is installed.
1. When in password 124 scroll the menu items pressing keys "Arrow
D I G I T A L C E P H
S T F Z E R O O F F S E T
P L E A S E W A I T . . .
NOTE:
* If the following message will be displayed:
D I G I T A L S E N S O R
I S N O T R E A D Y
Press key (7) to reset the message and press key (11) again.
WARNING:
X-rays will be emitted during the performance of the following operations.
Authorised Technicians are therefore recommended to use the greatest
caution and to comply with the safety regulations and laws of their
country.
7. Press the X-ray button and keep it depressed until the end of the
exposure; the following message will be displayed:
E N T E R I M G O F F S E T
O F F S E T ( m m ) 5 0
8. On the image, measure the distance "A" (Figure 9-30) between the
Soft Tissue Filter (STF) edge and the center of the rings; perform the
measure in "mm".
Figure 9-30
9. If the distance "A" is different from 50±2mm, press keys increase (4)
The system will calculate automatically the new Soft Tissue Filter
offset.
S T F Z E R O [ x x ] x
Z E R O O F S ± x x x x x x
10. Check proper alignment of the Soft Tissue Filter by performing a new
exposure (repeat the actions from point 3 to point 8).
If the alignment is correct, store it into the non-volatile memory of the
U P D A T E C H A N G E S ?
> 0 < = Y , T = N
1. General diagram
15. Lay-out and Circuit diagram CEPH arm connection PCB A12
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 1 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 2 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 3 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 4 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 5 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 6 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 7 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 8 of 9
ROTOGRAPH EVO 3D
General diagram 1
Code 39079003 - Rev. 1 Page 9 of 9
ROTOGRAPH EVO 3D
Lay-out 2
Column CPU PCB A1
Code 58071010 - Rev. P Page 1 of 1
ROTOGRAPH EVO 3D
Circuit Diagram 3
Column CPU PCB A1
Code 39079005 - Rev. P Page 1 of 3
ROTOGRAPH EVO 3D
Circuit Diagram 3
Column CPU PCB A1
Code 39079005 - Rev. P Page 2 of 3
ROTOGRAPH EVO 3D
Circuit Diagram 3
Column CPU PCB A1
Code 39079005 - Rev. P Page 3 of 3
ROTOGRAPH EVO 3D
Lay-out 4
Keyboard PCB A4
Code 58071012 - Rev. 1 Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 5
Keyboard PCB A4
Code 39079023 - Rev. P Page 1 of 4
ROTOGRAPH EVO 3D
Circuit diagram 5
Keyboard PCB A4
Code 39079023 - Rev. P Page 2 of 4
ROTOGRAPH EVO 3D
Circuit diagram 5
Keyboard PCB A4
Code 39079023 - Rev. P Page 3 of 4
ROTOGRAPH EVO 3D
Circuit diagram 5
Keyboard PCB A4
Code 39079023 - Rev. P Page 4 of 4
ROTOGRAPH EVO 3D
Lay-Out 6
CPU PCB A5
Code 58073033 - Rev. P Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 1 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 2 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 3 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 4 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 5 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 6 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 7 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 8 of 9
ROTOGRAPH EVO 3D
Circuit diagram 7
CPU PCB A5
Code 39079036 - Rev. P Page 9 of 9
ROTOGRAPH EVO 3D
Lay-out 8
Microprocessor PCB A6
Code 58073042 - Rev. 1 Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 9
Microprocessor PCB A6
Code 39079009 - Rev. 1 Page 1 of 4
ROTOGRAPH EVO 3D
Circuit diagram 9
Microprocessor PCB A6
Code 39079009 - Rev. 1 Page 2 of 4
ROTOGRAPH EVO 3D
Circuit diagram 9
Microprocessor PCB A6
Code 39079009 - Rev. 1 Page 3 of 4
ROTOGRAPH EVO 3D
Circuit diagram 9
Microprocessor PCB A6
Code 39079009 - Rev. 1 Page 4 of 4
ROTOGRAPH EVO 3D
Lay-out and circuit diagram 10
Rotation group PCB A7
Code 58073025 - Rev. P Page 1 of 1
ROTOGRAPH EVO 3D
Lay-out 11
Generator CPU PCB A9
Code 58074037 - Rev. P Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 12
Generator CPU PCB A9
Code 39079040 - Rev. P Page 1 of 2
ROTOGRAPH EVO 3D
Circuit diagram 12
Generator CPU PCB A9
Code 39079040 - Rev. P Page 2 of 2
ROTOGRAPH EVO 3D
Lay-out 13
Generator PCB A10
Code 58074021 - Rev. 8 Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 14
Generator PCB A10
Code 39079013 - Rev. 7 Page 1 of 3
ROTOGRAPH EVO 3D
Circuit diagram 14
Generator PCB A10
Code 39079013 - Rev. 7 Page 2 of 3
ROTOGRAPH EVO 3D
Circuit diagram 14
Generator PCB A10
Code 39079013 - Rev. 7 Page 3 of 3
ROTOGRAPH EVO 3D
Lay-out and circuit 15
diagram PCB A12
Code 58098150 - Rev. 0 Page 1 of 1
ROTOGRAPH EVO 3D
Lay-out 16
Power HUB PCB A21
Code 58073035 - Rev. 1 Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 17
Power HUB PCB A21
Code 39079038 - Rev. 1 Page 1 of 1
58073031/
ROTOGRAPH EVO 3D
Lay-out 18
VDSIB PCB A22
Code 58073031 - Rev. P Page 1 of 1
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 1 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 2 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 3 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 4 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 5 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 6 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 7 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 8 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 9 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 10 of 11
ROTOGRAPH EVO 3D
Circuit diagram 19
VDSIB PCB A22
Code 39079042 - Rev. P Page 11 of 11
ROTOGRAPH EVO 3D
Lay-out 20
DSPU PCB A14
Page 1 of 1
1 - COLUMN
Electrical and mechanical parts
Cables
3 - ROTATION ARM
4 - COVERS
5 - CEPH DEVICE
1 - COLUMN
7
2
13
4
3
8
16 12
11
14
15
10
Cables
Cables
1
6 (*)
8 (B3)
7
3
2
8 (B1)
8 (B2)
3 - ROTATION ARM
1 6607000300 3D HF Tubehead
11
14
1
13
12
5 (B6 / B7)
5 (B8 / B9)
6
4 – COVERS
5 – CEPH DEVICE
6 5207084000 Pulley
1 (B11) 5
3 7
6
2 1 (B10)
7 4
11
10
12. APPENDIX
NOTE:
* The information listed in the table are the technical parameters set
during factory testing. Preferences set by the user (e.g. exposure
parameters different than the default ones) are not listed.
The table also has columns with blank cells. These cells must be filled in
when, during installation or during the life on the unit, any of the listed
parameters will be modified (e.g. after replacing a motor or a positioning
sensor).
Rotograph EVO 3D
Unit code:
Unit S/N:
U.I.C.:
Pleora MAC Address:
Date
Language (English,
Italian, French,
German, Spanish)
Digital Ceph status
(OFF=disabled;
ON=enabled)
COLL setup type
3D sensor
(OFF=disabled;
ON=enabled)
STF setup type