Manual UltraTestLab V6 EN 2021 05
Manual UltraTestLab V6 EN 2021 05
User Manual
UltraTest GmbH
Am Schmiedeberg 6
28832 Achim
Germany
Tel: +49 4202 955 1390
Fax: +49 4202 955 1392
www.ultratest.de
[email protected]
-2-
3.3.1 Start of Measurements ................................................................................................... 28
3.3.2 Start of Measurements including Mixing Time .................................................................. 28
3.3.3 Additional Channels ........................................................................................................ 29
3.3.4 Stopping a Running Measurement ................................................................................... 29
3.4 Online Graphs ..................................................................................................................... 30
3.4.1 Graphic Marks ................................................................................................................ 31
3.4.2 Envelope ....................................................................................................................... 32
3.4.3 Cursor ........................................................................................................................... 32
3.4.4 Zoom ............................................................................................................................ 33
3.4.5 Graph Optimization ........................................................................................................ 34
3.4.6 Print Graphic.................................................................................................................. 34
3.4.7 Export Graphic ............................................................................................................... 34
3.4.8 Toolbox ......................................................................................................................... 35
3.4.9 Results .......................................................................................................................... 37
3.5 Online Tables ...................................................................................................................... 37
3.6 Evaluation .......................................................................................................................... 38
3.6.1 Spike Elimination............................................................................................................ 38
3.6.2 Data Export to MS-EXCEL ............................................................................................... 38
3.6.3 Parameter ..................................................................................................................... 39
3.6.4 Material Data Base (Material DB) .................................................................................... 40
3.6.5 Sub-Window Log (Log File) ............................................................................................. 42
3.6.6 Sub-Window „Read BP“ (for BP-700 Pro and BP-7 Tester only) ......................................... 43
3.7 Test ................................................................................................................................... 43
3.7.1 Dynamic E Modulus ........................................................................................................ 44
3.7.2 OK / NOK Testing (Traffic Light Function) ........................................................................ 45
3.7.3 Test Methods for Ultrasonic Sound Velocity Through Concrete .......................................... 45
3.8 Specimen ........................................................................................................................... 46
4 System Extensions ............................................................................................................ 49
4.1 Shrinkage Measurement ...................................................................................................... 49
4.1.1 Setup of Shrinkage Measurement .................................................................................... 49
4.1.2 Start of Shrinkage Measurement ..................................................................................... 50
4.2 External Temperature Measurement ..................................................................................... 50
4.2.1 Installation of External Temperature Measurement .......................................................... 51
4.2.2 Start of External Temperature Measurement .................................................................... 51
4.3 Differential Temperature / Room Temperature ...................................................................... 51
5 Troubleshooting ................................................................................................................ 52
6 EU Declaration of Conformity............................................................................................ 53
7 References ........................................................................................................................ 54
8 Application CD for Windows XP / 7 / 8 / 10 ..................................................................... 56
-3-
1 General
1.1 Cross References
This manual includes a variety of cross references. In the PDF version, you can easily follow these
references by clicking on the blue numbers with the left mouse button. This also applies for the table of
contents.
The PDF version of this manual can be opened via the help button directly from the program or it can also
be found on the enclosed CD.
1.2 Standards
The devices described in this manual comply with the following standards:
Concrete
EN 12504-4 (Europe)
EN 13791 (Europe)
ASTM C597-09 (North America)
IS13311 (Part 1) (India)
Natural Stone
EN 14146 (Europe)
EN 14579 (Europe)
-4-
1.3 Safety instructions
NOTE: Read the instructions carefully and completely. It is part of the product and contains important
notes for use. Keep this document!
Attention! Only the manufacturer may open the housing of any of the products. Opening by non-
authorized persons will cause the guarantee to be void!
Attention! Risk of electric shock! Avoid contacting mains voltages! Do not disconnect the transmitter
or receiver sensor during a running measurement! Do not start a measurement with the transmitter or
receiver sensor not connected!
Attention! Risk of short-circuit! Never bridge contacts! Do not put things into slots or ports! Do not
short-circuit transmitter and receiver outlets!
Attention! Not water-proof! Do not allow the BNC connectors of the transmitter and receiver sensors
to come in contact with water! The product is designed for indoor use only!
Attention! Danger of voltage pulses! Regularly check the function of the transmitter cable!
With badly damaged cables, noticeable voltage peaks can occur! At the first occurrence, immediately
replace the transmitter cable and destroy it!
1.4 Usage
The measuring system is designed for a mains voltage of 100 - 240 V AC / 50-60 Hz.
Plug the power supply into a 100-240 V AC mains socket.
Connect the IP-8 system to the 12 V DC output of the power supply.
Connect the IP-8 system with a USB cable to a PC.
Disconnect all power connections after use.
Any use other than described in this user’s manual is not permitted and causes loss of warranty, loss of
guarantee and non-liability.
We reserve our rights for misprints and changes of the device, software, packing, or user’s manual.
-5-
2 IP-8 Setup, Connection and Startup
2.1 Placing
Installation Examples:
-6-
2.1.2 Setup of Measurement Cables and Thermo Sensors
Transmitter 1 ... 8
Receiver 1 ... 8
Power Supply 12V - After installing all thermo sensors and measurement cables, connect the power
supply plug to the 12V socket of the IP-8 tester. When connecting the ultrasonic tester, please ensure the
correct voltage (see label).
USB Connection to Laptop/PC - Plug the IP-8 tester into a free USB port.
Note - If your IP-8 system has been delivered with a laptop or a personal computer, please continue with
chapter 2 IP-8 Setup and Connection.
-7-
Note for administrators - All UltraTestLab users require read/write rights in the program directory
(standard: C:\Program Files\UltraTestLab) and in the subdirectory \data. Otherwise, the
configuration and log files will not be updated.
After the driver has been successfully installed, you will see “UltraTest: IP-8” under Control Panel
Hardware and Sound Devices and Printers.
-8-
When the IP-8 tester is connected and ready to use, you will see the green status message
in the bottom left corner of the window.
-9-
1. Select the desired language
2. Insert your licence key and press „Check“. You can find your licence key on the last page of your
manual. Licence keys for additional extensions and features (e.g. measurement of
shrinkage/swelling) can also be entered/activated over the same field.
The UltraTestLab licence is valid for one PC or laptop. In addition, the software may be installed
on up to five additional computers for the purposes of evaluating measured data. Entering a licence
key will permanently unlock the number of channels purchased. The activated channels will be
displayed in the field "Current License".
- 10 -
The UltraTestLab Setup can be recalled at any time under: Configuration Service Setup
3. Insert the path for the measuring date as well as standard measuring distance
- 11 -
5. Connect the corresponding measuring channels
7. Finished
- 12 -
2.2.4 Test Function (Slider “Test”)
Before you start your first measurement, all measuring channels should be tested.
The test function allows a quick check of all channels and their respective
measuring distance by measuring the ultrasonic sound velocity in air.
Procedure:
Click single for individual tests or interval for interval tests. Repeat this
for each channel you intend to test. Runtime, velocity, and temperature
will be displayed.
The velocity is calculated by means of the measuring distance (standard
40 mm).
In air, the system should measure a velocity of approx. 343,5 4 m/s
(at 20 °C air temperature).
If all connected measuring channels display correct ultrasonic sound velocities and
temperatures, the system is ready for measurement.
2. In order to improve the demoldability, you can apply silicone oil or petroleum jelly
(petrolatum) to the front surface of the transducers and the thermo sensor as a release agent.
Important: Do not use mineral-oil-based form release agents. They can cause silicone to
swell.
transmitter’s and receiver’s set collars fit tightly
against the white stop rings of the measuring
forms. Turning the transducers once by 30° may
improve the coupling to the material.
- 13 -
5. When removing the transducers from the cured
material, only pull and turn clock-wise. The transducer
pipes are very robust and can also be broken out by
slightly hitting the edge of a table.
Note: Only the front part of the transducers is waterproof. The BNC
connector is NOT waterproof! Penetrating water or cleaning agents
can destroy the transducers!
- 14 -
2.4 Configuration and Start of Measurement (Tab “set up measurement”)
This chapter enables a quick start of a measurement. The next chapters describe the operation of the
systems IP-8 and BP-700, respectively, more in detail. For a more detailed start description please refer to
3.3 Configuration and Start of Measurements.
2. Click on and enter a measuring file name. It is also possible to create new directories.
- 15 -
3 UltraTestLab Operation
3.1 Layout
The operating software UltraTestLab is divided in various main and sub windows. A main window can be
selected by clicking on the tabs at the top.
- 16 -
3.2 System configuration
Note: From UltraTestLab version 5 a setup routine is launched at first time installation. All required settings
for standard operations are made within this routine. The System configuration does not need to be
changed unless system extensions are added or new settings are required later.
In the System configuration window various application parameters can be set. All settings will be saved
in INI files when the application is closed and will automatically be loaded the next time the program is
started.
3.2.1 Language
Select your desired language: German, English, French, Spanish, Italian or Polish. Additional languages are
available on request.
Select language 8
- 17 -
for both systems, this button will allow you to select between the two systems. The active measuring
system will be displayed in the top line.
Options:
1. “BP-7 V4.0“
2. “BP-7 V4.1“
3. “BP-7 V4.2“ and higher
4. “IP-8 V1.0“
5. “IP-8 V2.0“ (since 05/2009)
6. “IP-8 V3/V4“ (since 09/2012)
7. “BP-700 V1.0“ (since 12/2013)
The temperature sensors have an internal unique digital address. If a new temperature sensor is connected
to the IP-8, prior to measurements this address has to be assigned to a dedicated channel (1 to 8).
Activate Assignment
Attention: During this assignment only one sensor may be connected to the IP-8 system. The green
keys indicate which channels a temperature sensor is already assigned to.
assigned not assigned
A click on a free (grey) channel (K1 ... K8) stores the assignment of the connected temperature sensor to
this channel. The key turns green.
Prior to assignment of a new sensor the previous sensor assignment must be canceled > Remove
assignment: cancels the assignment of a sensor to a channel. The green key turns grey > Connect
ONLY the new temperature sensor (no other sensor may be connected) and assign to a free (grey)
channel (K1 … K8). The key turns green again.
Deactivate Assignment
- 18 -
3.2.4 Hardware - Calibration Thermometer
The temperature sensors have a resolution of 0.1°C and an average variation of approximately 0.1°-0.2°C.
In order to minimize this variation, the sensors can additionally be calibrated.
Path for logfile - Select the directory for storing the logging data file “logile.xls”.
Default is the same path as for measuring files. Any available network drive and directory can be selected.
Path for Material Database - Select the directory for storing of file database “mat_db.utl”.
Default is the same path as for measuring files. Any available network drive and directory can be selected.
Evaluation - Tick box to remember a directory for each channel individually! Under evaluation, up to 8
saved measurement files can be displayed simultaneously. For each channel the system will separately
remember the last selected directory from which a file has been opened if this option is selected.
Save XLS Files - By default, all measuring files will be saved in an ASCII file ending in *.lvm. These files
can be opened with MS Excel. If the option write data additionally to .xls file is selected, the system
will create an additional MS Excel file into which all of the data of a measurement is exported.
Template Export Excel - file name of the Excel template, that is used for the export of measured data
(see 3.6.2 Data Export to MS-Excel). Default path is C:\Program files (x86) \UltraTestLab \data
\export1.xls. This template file and path can be changed at will.
Logo – Logo to be used for printouts The UltraTest logo is set as a default. The user can indicate the path
to an individual logo (in .bmp format), see 3.4.6 Print Graphic. The selected logo will be displayed right
from this path indication.
- 19 -
3.2.8 Measurement Parameters - Parameter Measurement
Note: Changing the default setting can affect the measuring data.
Default Values:
No. of values Number of individual measurements from which the average will
be calculated.
Possible settings: 1, 3 or 6
Default: 1
A higher sensitivity (> 5) can be useful if no signal is received due to highly attenuating materials and/or
large measuring distances (> 40 mm). On the other hand, an environment with acoustic or electric
interferences can result in erroneous measurements due to premature tripping. In this case, runtimes will
be measured that are too short, which can result in outliers with very high velocities (see also Ignore
Runtime Val. below [µs] below). A low sensitivity reduces the risk of erroneous measurements.
If number of values, sensitivity or test voltage are modified from the default, the window on the right pops
up at the next start of UltraTestLab. Clicking the button set to default resets these settings back to
default.
- 20 -
Distance [mm] - Setting of the default measuring
distance in mm. This option sets the default measuring
distance for all 8 channels after starting the application or
after clicking on the set button. If you are using other than
standard moulds 140-50 with 40 mm, you should change
the default measuring distance. However, when configuring
a particular measurement, the measuring distance can be set individually.
Default: 40 mm
Ignore Runtime Val. below [µs] - Displays the runtime threshold in µs. With this function, erroneous
measurements due to premature tripping can be avoided. This type of erroneous measurements is usually
caused by acoustic or electric interferences. Runtimes measured below this threshold will not be shown
(see also Sensitivity above).
Default: 5 µs
Threshold Dyn E modulus [m/s] – Displays the threshold in m/s for the display of the dynamic E
modulus curve. Background: The dynamic E modulus only makes sense for solid compounds, it does not
exist in a fluid compound.
Start time – Measuring time (in minutes) before the search for a maximum of the acceleration is started.
Default: 10
Stop time – Measuring time (in minutes) after which the search for a maximum of the acceleration ends.
Default: 999999
Initial velocity – Velocity (in m/s) at which the search for the maximum of the acceleration is started.
Default: 10
% decr. – required minimum decrease (in %) of the values that follow a potential maximum in order to
define and display this maximum as such.
Default: 30
Acceler.1 Limit – Threshold for the first derivative (= gradient) of the measuring curve.
Default: 0.00
Acceler.2 Limit – Threshold for the second derivative (= curvature) of the measuring curve.
Default: 0.10
- 21 -
3.2.10 Measurement Parameters - Standard Measurements
At the right side of the window
measurement parameters up to 8 default
parameter sets can be individually defined
both for measuring period and interval (e.g.
for particular products or product groups) and
be selected when configuring the
measurement. Each default parameter set has
a long and a short description field.
The term entered in the abstract field will be
displayed in the set up measurement
window instead of the term mentioned in the
standard field (default standard, standard 1 to
8).
Procedure
After clicking on the button, a new
window opens on the right side. Apply a thin layer of contact
agent (e.g. oil) to the contact areas of the transducers and
press them tightly to both sides of the calibration bar.
- 22 -
The calibration will start 10 seconds after clicking on one of
the buttons C1 to C8. The first measured value will be
applied for the calibration and displayed in the right
window. Further below in that window, subsequently
measured values are displayed. Clicking Accept will apply
the measured value for the calibration. Repeat restarts the
measurement. Factory preset restores the factory-set
value for the selected channel. Clicking Stop aborts the
running measurement, a possibly measured value will not
be applied. The offset data thus determined will be assigned
to the channel. If possible, every transducer pair should
therefore always be used with the same measuring channel.
Note: At IP-8 systems version 3 equipped with firmware version 2.00 onwards, factory calibration has set
all channels to a value of 0.00 µs. In order to obtain the most accurate measuring results, the calibration
process should be repeated after replacing the transducers.
3.2.12 Extensions
Chapter 4 System Extensions describes program components that require their individual licence keys:
Preparation
- 23 -
Start performance check
4. Check connected hardware on functionality. During this process 6 measurements per channel will
take place in order to measure the noise level
After the performance check the results will be stored as a log file:
Default: C:\UltraTestLab\Diagnose\diag_ddmmyyyy_hhmm.xls
The file can be read by any kind of editor or by MS Excel. By clicking Send results to ultratest.de the
protocol file will be emailed in excel format to [email protected]. The button View results opens the
Excel file on the PC / laptop. By clicking EXIT the diagnosis will be quit.
- 24 -
Procedure
1. Exit all active measurements. (At channels with active measurement the noise level measurement
will be interrupted at the next measuring step.)
2. Select a free measuring channel and connect transmitter and receiver cable.
3. Select the time period (60 seconds) and the integration time (1-2 seconds)
4. Click start
5. Disturbances can be turned on and off and the influence on the measurement can be uncovered
(Course of the noise level at different locations within a climate chamber caused by turning the compressor
on and off)
Note: The rate of the noise level does not depend on the sensitivity setting (1 … 10, see Fehler! V
erweisquelle konnte nicht gefunden werden. Measurement Parameters). However, a constant noise
level at a high sensitivity setting (e.g. 9) will have a stronger influence on the measuring values than a low
sensitivity setting.
If advanced mode and noise detection are active (see 3.2.15 Service), another button
will appear in the online and evaluation graph window (see 3.4 Online Graphs). After the
activation of noise an additional graph will be displayed indicating the noise level on a scale from 0 to 10.
The noise level will be measured as follows: After the actual measurement the receiver will be turned on
for approx. 4 seconds and the highest amplitude will be measured. This way all recurrent background
noises are measured. Particular incidents such as e.g. banging doors will only be measured if they occur
during this measuring interval.
- 25 -
3.2.15 Service - Service
This service mode is for firmware updates, remote support, troubleshooting and guest access.
Advanced Mode
Various service features require to activate the advanced
mode by ticking the appropriate box.
Noise Detection
Beside advanced mode, in the Service window, it is possible to activate noise detection. If both boxes
are ticked, there will be an additional button noise in the online and evaluation graphs (see 3.4 Online
Graphs) that enables measurement of the noise level in addition to the other measured values.
Toolbox
The use of the toolbox functions requires activation of the advanced mode (see 3.4.8 Toolbox).
Printing Background
In the advanced mode it is possible to alter the online and evaluation graph
background color (e.g. for printing purposes) (see 3.4.6 Print Graphic). On
the right, background graph determines the background color of the online
or evaluation graph. Just click on the field and easily select any color.
- 26 -
3.2.16 Service - Simulation
Simple simulation mode for a demonstration of
measurements with no measuring system connected. In
the simulation mode, one or various measurements will
proceed in time-lapse mode.
When clicking the off button, it will change its color and
its name (on). Moreover, at the lower right end of the
window, the word Simulation shows up.
To start the simulation, the window set up measurement must be opened. Here, the already
preconfigured measurements can be started. After clicking the green start button of a channel, the
following window pops up:
The simulation will be stored as described in 3.3 Configuration and Start of Measurements as a file
named default Cx_uvw.lvm, where x stands for the selected channel number and uvw is a consecutive
numbering.
IP-8: 1 – 8 channels
BP-700 Pro: 1 channel
When you reach the maximum number of allowed channels, the Start buttons of the remaining channels
will automatically be disabled.
- 27 -
3.3.1 Start of Measurements
1. Select a free channel (green start button). Do not press the start button yet.
2. Select the parameters for the measurement using a default (8 default settings possible) for certain
products/product groups or manually set the measuring period and measuring interval. As a
standard, at the IP-8 the lowest settable measuring interval is 10 seconds. In the service mode,
however, it is possible to reduce the interval down to 1 second. Due to the measuring time of the
thermo sensors, below an interval time of approx. 3 seconds, the number of available channels will
be reduced.
If no valid file name has been entered yet, the alert window
shown on the right will appear.
While the measurement is running, the measured values will be stored in the selected file after every 10
measurements. After the pre-set measuring period has expired, the measurement will automatically be
terminated and the results will be saved in the specified file.
- 28 -
The mixing time can be added any time later, during or after
the measurement (see section 3.6.3 Parameter).
A maximum of 2 additional channels can be activated for each measuring channel. The measuring values
from the additional channels will measured in the same interval as the ultrasonic measurements.
Continue measuring
Abort the measurement and save the measured
values.
Abort the measurement without saving the
measured values.
Tip: Each measurement saved will be recorded in a measuring protocol (see 3.6.4 Sub-Window Log).
- 29 -
3.4 Online Graphs
In the Online Graphs window, ongoing measurements are graphically displayed.
Depending on the available extensions up to two of the following graphs can additionally be displayed for
each channel:
In general, the ultrasonic velocity correlates very well with the strength properties of a specimen. For solid
parts, a direct physical correlation exists between velocity and dynamic e-modulus.
Edyn = k * v² * ρ
k = (1 + μ) * (1 - 2μ) / (1 - μ)
µ= Poisson's ratio
v= velocity in [m/s]
ρ= density of product sample in [kg/dm³]
- 30 -
In order to display the dynamic e-modulus the value of density and Poisson's ratio need to be entered.
(see 3.6.3 Parameter).
Sets of graphic marks - Up to 20 sets of 4 graphic marks each can be defined and individually named.
Just click on the field Set 1 … 20, enter the Name of the set and the parameters in the Text field and
click on Save new values. Options: velocity [m/s] or time [min]
The figures belonging to the graphic marks can easily be moved to any other place by means of the
mouse.
- 31 -
3.4.2 Envelope
Envelopes let you define allowed deviations from reference measurements. Two additional curves above
and below the reference curve create a corridor that limits the allowed area. For this purpose, a percentage
velocity offset, an absolute velocity offset in m/s, and a time offset in minutes are provided in relation to
the reference curve. The reference curve can be selected via the additional channel reference (on the
right side).
Offset 1 - 2 time offset (shifting the reference curve on the time axis)
Enter permissible positive and negative time offset in minutes.
Envelope sets - up to 20 sets of offset values for different products can be defined.
3.4.3 Cursor
The cursor is activated by clicking on the respective button
at the bottom of the diagram. When the cursor is activated,
an additional button to select the channel shows up and the
exact values of the x- and y-position of the cursor of the selected channel will be displayed (e.g. 10 min,
1556.42 m/s).
- 32 -
Eliminate noise spikes (evaluation graphs only)
The curve in the left window includes two noise spikes, that do not refer to the actual behavior of the
tested material. In the middle window the right spike has been selected by the cursor. By clicking interpol.
& resave the spike will be removed and a new dot will be interpolated that fits to the adjacent dots of the
curve as displayed in the right window. Clicking delete & resave would have deleted this dot only, a little
gap would have occurred in the curve. The curve will be automatically stored in the new condition.
The automatic suppression of spikes can be set – even during an active measurement - in the toolbox
under Spikes, see 3.4.8 Toolbox.
3.4.4 Zoom
After clicking on the Zoom button, an area can be selected by the mouse that will be displayed in enlarged
mode. A second click on the Zoom button deactivates the function and the display will return to normal.
- 33 -
3.4.5 Graph Optimization
Adjust Scale
The scale of x-axis and y-axis can easily be adjusted by double-clicking on the highest (or lowest) value
(e.g. “0 m/s” or “1200 m/s”). Subsequently you can enter your desired new value via keyboard and confirm
by pressing Enter on the keyboard. The y-axis default value can be set in the software menu, see 3.2.6
Software > Scale Standard Y-axis.
Show Legend
Clicking the Legend button once or twice activates the legend field and it will be displayed. This feature
displays the description or assignment of the graphs, e.g. for print editions. By left-clicking on the frame
and holding down the mouse button, you can move the legend to any place in the graphic.
Type of diagram
Color
Line style and width
Smoothing and interpolation
Axis assignment
opens the window on the right. The top line will be printed left-aligned (logo), centered (first
text field), and right-aligned (second text field) with the entered text.
- 34 -
1. Right-click into the diagram.
2. Select first line of the pop-up
menu (export graphic)
3. Select the desired graphic
format (selecting enhanced
ctrl+ v
metafile .emf format usually
leads to a better resolution).
4. Export to the clipboard or file.
5. Copy into the desired
application.
3.4.8 Toolbox
The content of the toolbox varies in the main menus online graph and evaluation. Use of the toolbox
features requires activation of the advanced mode (see 3.2.15 Service > Service).
The boundary points of the two tangents with the curve are marked with white cursors. These cursors can
easily be moved on the curve by clicking them with the mouse. The belonging tangents are displayed as
dotted white lines. At the intersection of the two tangents, a dotted orange line indicated the belonging
time value. The colors white and orange, resp., are default values that can easily be changed to any other
value in the toolbox window.
- 35 -
Filters upper left side (evaluation only) - Filter for
smoothing the speed graphs.
Auto fitting: A defined range can be activated by setting a start time and stop time and subsequently
clicking the range button.
Width of spike (evaluation only) - enables the automatic suppression of spikes that extend over more
than one measured value.
- 36 -
3.4.9 Results
Copy Data - The values of a table can be copied to the clipboard by right-clicking them.
- 37 -
3.6 Evaluation
The evaluation window allows for the evaluation of completed measurements. The structure corresponds
to the window online graphs, please refer to 3.4 Online Graphs. Up to 8 saved measurements
(file 1 … 8) can be displayed and edited. BP-7 only: Subordinate windows are available as follows: file,
log, import and read BP-7 (BP-7 only).
Using the buttons file 1 to file 8, you can load and display saved measurements via a browse prompt. For
every channel, the last opened file is pre-set so it can be re-opened by pressing Enter. Clicking the file x
button again will delete the present file.
The channels can be set to either display the velocity [m/s] or the runtime [µs]. The conversion takes place
using the measuring distance set before starting the measurement. Moreover, you have the option of
displaying either the temperature or the first derivative (acceleration).
The features cursor, print, zoom, envelopes, graphic marks, x-axis, and grid are identical with the
features in the online graphs window (see 3.4 Online Graphs).
Caution: File 8 is predefined as the reference curve for the envelope feature here. For the cursor
feature, the respective channel can be selected by means of the arrow keys in the file display.
- 38 -
3.6.3 Parameter
By clicking on a new window will open where data and description of the selected
measurement can be displayed and changed.
Sample, text and user can also be amended after the measurement has been finished.
Distance [mm] - If the measuring distance is amended later, the velocity and derivation will be
recalculated.
Mix. Time - Increasing the mixing time (e.g. of 5 minutes) will move the curve of the setting process on
the time axis by 5 minutes to the right. The UltraTestLab version V3.00.43 and higher does allow
amendments of the mixing time during the active measurement (online graphs).
Dyn. E-Module - The density of the material and the Poisson's ratio are necessary to calculate the dynamic
e-modulus. If the density is not known, it can be calculated by weighting the test specimen. For some
measurement forms the volume is already stored. For standard shapes 140-50 (D50), the volume is 95 ml.
The Poisson's ratio µ is often not known for all products. Therefore for many cement bonded materials a
value of µ=0.20 ... 0.22 is assumed.
Info File - For each stored measurement, an individual info file can be stored. This file may include any
content. This option is helpful to store all data relevant to a particular measurement, see pic at the next
page:
- 39 -
3.6.4 Material Data Base (Material DB)
The material database can store measurement parameters as well as material data and descriptions for
any number of products or materials. By selecting an already generated material all parameters and
descriptions for new measurements are automatically adopted.
- 40 -
Click on „Details“ or „Material DB“ and the start screen opens:
The material data base allows storage of parameter sets of measuring data settings.
Description
Duration: Measurement duration in days and hours, measuring interval in minutes and
seconds
Distance: The sensor distance can be entered. This is helpful if various different measuring
forms are in use (due to varying aggregate size). The value for standard measuring
forms is 40 mm.
Reference file: Storage location for a reference curve for this measurement
Graphic marks set: for the evaluation of times and speeds (see 3.4.1 Graphic Marks)
Envelope set: for monitoring permissible deviations and envelopes (for error monitoring, see
3.4.2 Envelope).
The “Strength” button at the top can only be activated if the optional license strength measurement is
included in the UltraTestLab.
- 41 -
Storage Location for Material Database
Die gemeinsame Nutzung in einer Netzwerkumgebung ist möglich. Schreibrechte ggf. überprüfen.
Sorting options are: No. / Date / Time / File / Sample / Text / User
Delete will remove the highlighted rows from the log file. The log file itself will
not be deleted.
- 42 -
3.6.6 Sub-Window „Read BP“ (for BP-700 Pro and BP-7 Tester only)
This feature reads the measurements saved in the BP-7 and
saves them in the *.lvm format. For this feature you need
to set the BP-7 to the “PC” and “read” mode.
After the files have been read, the data can be displayed in
the “evaluation” window, printed, and exported as *.xls file.
The imported files will be included in the log file. Double-
clicking the row will display the imported file in the
subordinate window “file”.
3.7 Test
The testing feature allows for quick testing of all channels and checking the measuring distance through
measuring the sonic velocity in air.
Individual tests or interval operation for all free channels. Displays runtime, velocity, and
temperature.
In air, the system should detect a velocity of about 343 +-3 m/s (at 20°C).
If additional measurements are activated, their measurement values will be displayed as well.
(Shrinkage and Quelling will be measured with the Mitutoyo measuring gauge here).
- 43 -
BP-7 Version 4.0 and 4.1 only - After restarting the BP-7, you need to set the BP-7 to the “PC” measuring
mode prior to the first measurement. Then you need to send a start signal by clicking the Start button
ONCE. You must NOT send a start signal for subsequent measurements - click on “skip”.
Edyn = ρ * v² * k
k = (1 + μ) * (1 - 2μ) / (1 - μ)
µ= Poisson's ratio
v= sonic velocity in [m/s]
ρ= density of the sample in [kg/dm³]
The dynamic E modulus is defined only for firm materials. Calculations for green mortar do not yield any
useful values.
- 44 -
3.7.2 OK / NOK Testing (Traffic Light Function)
When testing similar parts, quality assurance often uses reference values. In the example below, parts with
a runtime ≤ 50 µs are classified as OK. For runtimes over 70 µs, the part is to be considered defective.
Ultrasonic runtime
≤ 50 µs: part is OK
> 50 µs and ≤ 70 µs: ? / review
> 70 µs: part is not OK / NOK
The devices described in this manual comply with the following standards:
EN 12504-4 (Europe)
ASTM C597-09 (North America)
IS13311 (Part 1) (India)
- 45 -
3.8 Specimen
The chapter Specimens deals with the measurement and storage of solid specimens such as e.g. prisms.
In this area, measuring processes for different test specimens can be created and managed. The strength
development can be displayed as a graphic. When entering the corresponding parameters (density and
Poisson number), the dynamic modulus of elasticity can also be displayed with absolute values (in GPa).
Under Control, there is an overview of all created test specimens with active or finished test series. For
the test series that have not been completed, the time for the next measurement is displayed in the "Next
test" column. By pressing on the column header, the measurement series can be sorted according to the
selected parameter. When clicking on the column header "Next test", the measurement series can be, for
example, sorted in chronological order for the next test to be performed. The red line mark may indicate
the measurement with the longest overdue. The default path for storing measurements is entered at the
bottom left at ‘Path for measured values’.
The ‘New test block’ key allows the creation of a new test specimen. After a specimen is created it can
be edited any time later. However, the following initial input options only exist at this point:
- 46 -
The following setting points are also possible in the edit mode:
- 47 -
Under the Measurement tab, further measured values are assigned to a test specimen.
In addition to the selection key for the file to be edited, the Edit button can be
used to open a window for editing the measuring process. Here the number of
measurement points used for the measurement as well as the test intervals for
which measurement values are to be acquired can be defined. Description see
„New test block“
The Graph key enables display in one diagram of the measured values for
ultrasonic velocity and dynamic E modulus.
- 48 -
4 System Extensions
The elongation can be displayed as an additional graph above the ultrasonic velocity and the temperature.
It is also possible to measure the elongation independent of the ultrasonic velocity. With some kinds of
measuring gauges, the accuracy can be increased from 0.01 mm to 0.001 mm.
3. Let the system search for the USB driver automatically and follow the prompts.
Ignore the Windows compatibility warning and install the driver
6. Select interface:
- 49 -
7. Select type e.g. IMB-mi4 and click on “setup.”
Installation completed!
In the area for the additional channel, you can select the
corresponding measuring gauge Length C1 to Length
C8.
After starting the measurement, you can now choose to display or to hide the
additional channels (here length) as well as the temperature and acceleration
graphs in the Online Graphs window.
The elongation will also be displayed in the online tables. Under evaluation, the
additional channels are displayed as Add. Ch.1 and Add. Ch.2. The length values
can be exported to MS Excel by clicking on to.xls (see 3.6.2 Data Export to MS
Excel).
- 50 -
4.2.1 Installation of External Temperature Measurement
1. Install the current version of UltraTestLab (version
1.4.45 or higher).
2. Insert the included software CD in your CD-ROM
drive and follow the instructions on the screen.
3. Let the system search for the USB driver
automatically and follow the prompts.
Ignore the Windows compatibility warning and
install the driver.
4. The external temperature measurement option
can be unlocked with a licence key:
Info Licence Options
5. Activate the external temperature measurement:
Configuration Additional Measurements
External Thermometers on
This Feature is switch on with and is therefore activated as an additional channel for each
measurement (see 3.3.3 Additional Channels).
- 51 -
5 Troubleshooting
The majority of known problems is caused by defective measuring cables or USB cables.
Check the device settings according to 3.2.2 “Configuration Interface Connected device.”
default: “IP-8 V3.0” since 09/2012
Close the application and restart Windows. Restart UltraTestLab, select interface USB-IP-8, and
run “Test” (see 3.7 Test). The characteristic clicking of the transmitters should be audible.
1. Disconnect the thermometer from the faulty channel (see 3.2.3 configuration -> delete
assignation)
2. Re-assign the thermometer
3. Run “test” and check if the room temperature is displayed
Check power supply: the voltage should be between 11.8 V and 12.2 V.
(The positive pole is inside the plug.)
In the case of individual extremely high velocity values, the threshold value is usually reached too
early due to interferences. Possible reasons for such interferences are:
All IP-8 systems from version V3.0 on include an interference level measurement feature that is factory
activated (see 3.2.14 Diagnosis - Noise Level).
If a device has to be sent in for checking / repair, please contact us before taking any further action.
UltraTest GmbH
Am Schmiedeberg 6
28832 Achim
Germany
- 52 -
6 EU Declaration of Conformity
- 53 -
7 References
Mortar, Plaster, Thermal Composited Systems
- 54 -
Refractories, Castables
- 55 -
8 Application CD for Windows XP / 7 / 8 / 10
- 56 -