Blood Gas Control - Level 3 (BG Control 3)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

BLOOD GAS CONTROL - LEVEL 3 (BG CONTROL 3)

CAT. NO. BG5003 LOT NO. 252BG


SIZE: 30 x 1.8 ml EXPIRY: 2023-09-28

GTIN: 05055273227123

INTENDED USE
This product is intended for in vitro diagnostic use, in the quality control of Blood Gas analysis.

DEVICE DESCRIPTION
The Blood Gas Controls are supplied at 3 levels, 1, 2 and 3. Target values and ranges are supplied for the
following analytes: Calcium, Chloride, Glucose, Lactate, PCO2, pH, pO2, Potassium, Sodium and Total CO2.

SAFETY PRECAUTIONS AND WARNINGS


For in vitro diagnostic use only. Do not pipette by mouth. Exercise the normal precautions required for
handling laboratory reagents. Health and Safety Data Sheets are available on request.

STORAGE AND STABILITY


UNOPENED: The product is stable to expiration date when stored at +2ºC to +8ºC. Avoid
exposure to freezing and temperatures greater than +30ºC.

OPENED: For pH/blood gas values, the control should be analysed within 1 minute of opening. For
electrolyte measurements, the control should be analysed within 1 hour after opening.

PREPARATION FOR USE


The Blood Gas Control should be brought to +20ºC to +23ºC before use. Allow at least 4 hours for
ampoules to equilibrate to this temperature, prior to testing. Before use, hold the ampoule at the top and
bottom (with forefinger and thumb) and shake 15 - 20 times to mix the solution. Tap the ampoule to restore
the liquid to the bottom of the ampoule. Open the ampoule by snapping off the tip at the score. Use gauze,
tissue, gloves or an appropriate ampoule opener to protect fingers from cuts. Immediately introduce the
liquid from the ampoule to the analyser.

MATERIALS PROVIDED
Blood Gas Control - Level 3 30 x 1.8 ml

ASIGNED VALUES
Due to the variation caused by test equipment, test reagents and laboratory technique, the quoted ranges are
provided for guidance. It is recommended that these ranges are used until each laboratory has established its
own ranges, based on individual laboratory requirements.

Each batch of Blood Gas Control is submitted to a number of external laboratories and values are assigned
from a consensus of results obtained by these laboratories.

Randox Teoranta, Meenmore,


Dungloe, Donegal,
F94 TV06, Ireland

03 Mar ‘22 me

PAGE 1 OF 2
A
BLOOD GAS CONTROL - LEVEL 3 (BG CONTROL 3)
Cat. No. BG5003 Lot. No. 252BG Size 30 x 1.8ml Expiry 2023-09-28
Range
Analyte unit Target low high methods
Calcium mmol/l 0.698 0.628 0.768 Ion selective electrode
mg/dl 2.80 2.52 3.08
mmol/l 0.652 0.587 0.717 Colorimetric
mg/dl 2.61 2.35 2.87
Chloride mmol/l 127 117 137 Colorimetric
mmol/l 123 113 133 ISE indirect
Glucose mmol/l 14.7 12.5 16.9 Enzymatic Electrode
mg/dl 265 225 305
mmol/l 14.6 12.4 16.8 Glucose oxidase
mg/dl 263 223 303
Lactate mmol/l 1.03 0.845 1.22 Enzymatic Electrode
mg/dl 9.28 7.61 11.0
mmol/l 0.940 0.771 1.11 Colorimetric
mg/dl 8.47 6.95 9.99
pCO2 kPa 3.12 2.50 3.74 Ion selective electrode
kPa 3.08 2.46 3.70 Optical Fluorescence
pH pH units 7.521 7.468 7.574 Ion selective electrode
pO2 kPa 19.6 16.7 22.5 Ion selective electrode
kPa 19.5 16.6 22.4 Optical Fluorescence
Potassium mmol/l 6.15 5.66 6.64 ISE method - direct
mmol/l 6.10 5.61 6.59 Optical Fluorescence
Sodium mmol/l 165 157 173 ISE method - direct
mmol/l 161 153 169 Optical Fluorescence
Total CO2 mmol/l 19.8 15.8 23.8 Ion selective electrode
mmol/l 19.9 15.9 23.9 Calculated
mmol/l 20.0 16.0 24.0 Optical Fluorescence

__________________________________________________________________________________________________

Randox Laboratories Ltd., 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom PAGE 2 OF 2
Tel: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912
Email: applications@randox.com Website: www.randox.com