Service Manual - DM0412S
Service Manual - DM0412S
DM0412S
VERSION20170118
CONTENTS
1.1 Introduction
DM0412S clinical centrifuge support applications with small sample volumes, the centrifuge is
widely used in medical and veterinary practices for the centrifugation of blood and urine samples,
in environmental analyses for the clarification of water and soil samples or in universities and
other research institutes.
1 2 3 4 5 6 7
FIG.1
Item 1 2 3 4 5 6 7
Master Speed Time Door Power Power view
Function
control board display display button switch interface window
When instrument failure occurs, first, you should conduct a failure analysis; if the failure is
caused by the damage of instrument hardware, the related component must be repaired or
replaced. Here are the relevant contents of the replacement and disassembly of instrument.
2.1 Removal
1
Step 1:
Step 2:
Step 3:
2
Step 4
Step 5:
3
2.2 Main parts illustration
4
2.3 Replacement of driver board
5
2.5 Replacement of brake resistor
Remove the screw, plug out the
connector marked by red circle,
then replace the brake resistor,
assembled in sequence.
6
2.7 Replacement of speed sensor
Step 1:
Step 2;
Step 1:
7
Step 2:
8
Driven board goes failure Replace driven board
E10 Unable to drive Connection between Reconnect motor and driven
driven board and motor board
is improper
Driven board is failure Replace driven board
Long time no running Rotate rotor by hand and run
again
E20 Over speed (Speed signal is Speed sensor goes failure Replace speed sensor
unstable.) Driven board goes failure Replace the driven board