AKTApure
AKTApure
AKTApure
User Manual
Table of Contents
Table of Contents
1 Introduction ...................................................................................................................................................... 6
1.1 Important user information ............................................................................................................................................ 7
1.2 ÄKTA pure overview ........................................................................................................................................................... 9
1.3 ÄKTA pure user documentation .................................................................................................................................... 11
1 Introduction
In this chapter
This chapter contains the following sections:
All users must read the entire ÄKTA pure Operating Instructions before installing, operating, or
maintaining the instrument. Always keep the ÄKTA pure Operating Instructions at hand when operating
ÄKTA pure.
Do not operate ÄKTA pure in any other way than described in the user documentation. If you do, you
may be exposed to hazards that can lead to personal injury and you may cause damage to the
equipment.
Intended use
ÄKTA pure is intended for purification of bio-molecules, in particular proteins, for research purposes
by trained laboratory staff members in research laboratories.
ÄKTA pure shall not be used in any clinical procedures, or for diagnostic purposes.
Prerequisites
In order to operate the system according to the intended purpose, it is important that:
• you have a general understanding of how the computer and the Microsoft® Windows® operating
system work.
• you understand the concepts of liquid chromatography.
• you have read and understood the Safety instructions chapter in ÄKTA pure Operating Instructions.
• a user account has been created according to UNICORN™ Administration and Technical Manual.
Safety notices
This user documentation contains safety notices (WARNING, CAUTION, and NOTICE) concerning the
safe use of the product. See definitions below.
WARNING
WARNING indicates a hazardous situation which, if not avoided, could result in death
or serious injury. It is important not to proceed until all stated conditions are met
and clearly understood.
CAUTION
CAUTION indicates a hazardous situation which, if not avoided, could result in minor
or moderate injury. It is important not to proceed until all stated conditions are met
and clearly understood.
NOTICE
NOTICE indicates instructions that must be followed to avoid damage to the product
or other equipment.
Main features
The main features of ÄKTA pure are listed below.
• ÄKTA pure is a flexible system that allows the user to configure both hardware and software to
meet many purification needs.
• The instrument can be easily extended with additional valves, detectors and fraction collectors.
• There are a large number of different hardware modules to choose from. The user can adjust, for
example, the number of columns, inlets, outlets and detectors and choose between different ways
to apply and collect samples.
• Purification and maintenance methods are easily created using predefined methods and method
phases. A method phase refers to a specific step/part in the method, such as column wash or elution.
Method phases are described in Phase Properties and are displayed graphically in a method outline.
This makes methods and phases easy to understand and edit.
• ÄKTA pure is controlled by the UNICORN software: a complete package for control, supervision and
evaluation of chromatography instruments and purification runs.
• UNICORN has different software licence options, such as Design of Experiments (DoE) and Column
logbook, to further add user support.
Administration Perform user and system setup, system log and database adminis-
tration.
Method Editor Create and edit methods using one or a combination of:
• Predefined methods with built-in application support
• Drag-and-drop function to build methods with relevant steps
• Line-by-line text editing
The interface provides easy viewing and editing of run properties.
System Control Start, monitor and control runs. The current flow path is illustrated
in the Process Picture, which allows manual interactions with the
system and provides feedback on run parameters.
When working with the modules Administration, Method Editor, System Control and Evaluation it is
possible to access descriptions of the active window by pressing the F1 key. This can be especially
helpful when editing methods
User documentation
The user documentation listed in the table below is delivered with ÄKTA pure. It is also available on the
user documentation CD.
ÄKTA pure Operating Instructions Instructions needed to install, operate and maintain ÄKTA
pure in a safe way.
Fraction collector F9-C and F9-R Instructions needed to install, operate and maintain Fraction
Operating Instructions collectors F9-C and F9-R in a safe way.
Sample pump S9 and S9H Operating Instructions needed to install, operate and maintain Sample
Instructions pump S9 and S9H in a safe way.
ÄKTA pure User Manual Detailed instrument and module descriptions and instruc-
tions on how to run, maintain and troubleshoot the system.
UNICORN Help Descriptions of UNICORN dialog boxes (available from the Help
menu).
Getting started with Evaluation • Video clips showing common workflows in the Evaluation
Note: module.
Available in UNICORN 7.0 and lat- • Overview of features of the Evaluation module.
er.
UNICORN Method Manual 1 • Overview and detailed descriptions of the method creation
features in UNICORN.
• Workflow descriptions for common operations.
UNICORN Administration and • Overview and detailed description of network setup and
Technical Manual1 complete software installation.
• Administration of UNICORN and the UNICORN database.
UNICORN Evaluation Manual1 • Overview and detailed descriptions of the Evaluation Classic
module in UNICORN.
• Description of the evaluation algorithms used in UNICORN.
UNICORN System Control Manual1 • Overview and detailed description of the system control
features in UNICORN.
• Includes general operation, system settings and instructions
on how to perform a run.
Additional literature
For practical tips on chromatography, refer to ÄKTA Laboratory-scale: Chromatography Systems Instru-
ment Management Handbook (product code 29010831).
In this chapter
This chapter contains the following sections:
2.6 Accessories 79
1 2 3
Part Function
1 Wet sides
2 Buffer tray
3 Holder rails
5 Power switch
6 Ventilation panel
11 12
10
9 1
8
2
7
6 5
Part Function
1 Multi-module panel
2 Inlet valve
4 System pump B
5 Pressure monitor
6 System pump A
7 Mixer
8 Outlet valve
9 Injection valve
10 Conductivity monitor
11 Column valve
12 UV monitor
Available modules
The modular design allows the user to customize ÄKTA pure in multiple ways. The system is always
delivered with the core modules of the selected configuration, but optional modules may be added to
the flow path.
The table below lists the available modules forÄKTA pure 25 and ÄKTA pure 150. Core modules are in-
dicated with an asterisk (*).
Note: The valves for ÄKTA pure 25 and ÄKTA pure 150 are compatible with both systems but for the
best performance the specific valve type should be used. The narrow channels in the valves
for ÄKTA pure 25 will give too high back pressure if used above 50 ml/min. The larger volumes
in the “H” valves for ÄKTA pure 150 may decrease resolution and increase peak broadening if
used in ÄKTA pure 25.
Module Label in
Pressure monitor* R9 R9
Mixer* M9 M9
V9-C V9H-C
V9-C2 V9H-C2
V9-Os V9H-Os
Module Label in
U9-M U9-M
Conductivity monitor C9 C9
L9-1.2 L9-1.2
F9-R F9-R
I/O-box E9 E9
Illustration convention
In the valve illustrations below, the following convention is used to point out the location of the ports
on the valve head. Loop valve V9-L is used as an example.
V9-L
2E
3E
4E • Ports located on the valve head rim are indicated outside the black
ring (e.g., 1E, 2E, etc.).
1E
5E
• Ports located on the pivot part of the valve head are indicated on
the inside of the black ring (e.g., 3E and 3F).
1F
5F
3F
4F 2F
(e.g., E and F).
Core modules
Core module Description
System pump P9 A or P9H A A high precision pump, which delivers buffer or sample in purification runs.
For further information, refer to Section 2.4.1 System pumps, on page 33.
System pump P9 B or P9H B A high precision pump, which delivers buffer in purification runs.
For further information, refer to Section 2.4.1 System pumps, on page 33.
Pressure monitor R9 Reads the system pressure after System pump A and System pump B.
For further information, refer to Section 2.4.12 Pressure monitors, on page 67.
Mixer M9 Mixes the buffers delivered from the system pumps to a homogeneous buffer
composition.
Three Mixer chambers are available for ÄKTA pure 25, their volumes are: 0.6
ml, 1.4 ml (mounted at delivery) and 5 ml.
Three Mixer chambers are available for ÄKTA pure 150. Their volumes are:
1.4 ml (mounted at delivery), 5 ml (included in delivery), and 15 ml.
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with
an ÄKTA pure 25 system configuration. The maximum
pressure for Mixer chamber 15 ml is 5 MPa.
Loo
W2
pF
SaP
Optional modules
Module Description
Inlet valve V9-IA or V9H-IA Inlet valve for System pump A with seven inlet ports and integrated air sensor.
For further information, refer to Section 2.4.4 Inlet valves, on page 39.
V9-IA
A3 A2
A4
A1
A5
Q
A7
A6
Out
Inlet valve V9-IB or V9H-IB Inlet valve for System pump B with seven inlet ports and integrated air sensor.
For further information, refer to Section 2.4.4 Inlet valves, on page 39.
V9-IB
B2 B3
B1
B4
Q
B5
B6
B7
Out
Inlet valve V9-IAB or V9H-IAB Inlet valve with two A inlet ports and two B inlet ports. No integrated air
sensor.
For further information, refer to Section 2.4.4 Inlet valves, on page 39.
A2 B2
A1 B1
A B
Sample inlet valve V9-IS or V9H-IS Inlet valve with eight inlet ports (seven sample inlets and one buffer inlet)
and integrated air sensor. The Sample inlet valve requires an external Sample
V9-IS pump module.
S2 S3
For further information, refer to Section 2.4.4 Inlet valves, on page 39.
S1
S4
Buff
S5
S7
S6
Out
Module Description
Inlet valve V9-IX or V9H-IX Inlet valve with eight inlet ports. No integrated air sensor.
For further information, refer to Section 2.4.4 Inlet valves, on page 39.
V9-IX
2 3
4
1
8
5
7
Out
Mixer valve V9-M or V9H-M Directs the flow to the Injection valve, bypassing the Mixer, or to the Injection
valve via the Mixer.
V9-M Re-Inj For further information, refer to Section 2.4.5 Mixer valve, on page 46.
Out
In
Mixer
Loop valve V9-L or V9H-L Enables the use of up to five loops connected to the instrument.
For further information, refer to Section 2.4.7 Loop valve, on page 52.
Column valve V9-C or V9H-C V9-C or V9H-C can connect up to five columns to the instrument. Up to ten
and V9-C2 or V9H-C2 columns can be connected by installing the optional column valves V9-C2
or V9H-C2.
The column valves direct the flow to one column at a time and feature two
integrated pressure sensors.
The valves allow the user to choose flow direction through the column, or
to bypass the column.
For further information, refer to Section 2.4.8 Column valves, on page 55.
Module Description
pH valve V9-pH or V9H-pH Enables the pH electrode to be included in the flow path or bypassed during
a run. The pH electrode may be calibrated when installed in the pH valve.
For further information, refer to Section 2.4.10 pH valve, on page 60.
Outlet valve V9-O or V9H-O Directs the flow to the Fraction collector, Fraction collector 2 (out 10), any
of the ten outlet ports, or waste.
V9-O
Fra
c Ou
t1 In
For further information, refer to Section 2.4.11 Outlet valves, on page 64.
W
t10
Ou
Ou
t2
Out9
Out3
Ou
t4
t8
Ou
Ou Out6 t5
t7 Ou
Outlet valve V9-Os or V9H-Os Directs the flow to the Fraction collector, Fraction collector 2, the outlet port,
or waste.
V9-Os In For further information, refer to Section 2.4.11 Outlet valves, on page 64.
Frac
Out
Module Description
Versatile valve V9-V or V9H-V A 4-port, 4-position valve, which can be used when adding extra features
to the flow path.
V9-V 1 For further information, refer to Section 2.4.9 Versatile valve, on page 59.
2
4
Module Description
External air sensor L9-1.5 or L9-1.2 Prevents air from being introduced into the flow path.
For further information, refer to Section 3.1 External air sensors, on page 87.
Fraction collector F9-C Flexible fraction collector that can collect up to 576 fractions.
Up to two fraction collectors can be connected at the same time, of which
only one (the primary) can be a Fraction collector F9-C.
For further information, refer to Section 3.2 Fraction collector F9-C, on page 89.
Fraction collector F9-R Round fraction collector that can collect up to 175 fractions.
Up to two fraction collectors can be connected at the same time.
For further information, refer to Section 3.3 Fraction collector F9-R , on
page 103.
I/O-box E9 Receives analog or digital signals from, or transfers analog or digital signals
to, external equipment that has been incorporated in the system.
For further information, refer to Section 3.5 I/O-box E9, on page 111.
Module Description
Sample pump S9 or S9H A high precision pump with an integrated pressure monitor. The sample
pump delivers buffer or sample in purification runs.
For further information, refer to Section 3.4 Sample pump S9 and S9H, on
page 107.
1 3
2
6
4 7 9
5 8
10
W1 W2
16
W
11 12 13 14 15
17
Part Description
1 Pressure monitor
2 Sample pump
Part Description
4 Inlet valve
5 System pump B
6 System pump A
7 Pressure monitor
8 Mixer
9 Injection valve
11 Column valve
12 Column
13 UV monitor
14 Conductivity monitor
15 Flow restrictor
16 Outlet valve
17 Fraction collector
W, W1, W2 Waste
Part Function
Lock/Unlock function
Follow the instruction below to lock or unlock the Pause and Continue buttons of the Instrument
control panel from UNICORN.
Step Action
Buttons
The Instrument control panel includes the following buttons:
Part Function
Status indications
The light indicators on the Instrument control panel indicate the current status of ÄKTA pure.
The table below describes the different states that can be displayed.
All light indicators are off. Off The instrument is turned off.
Off
The Power/Communication indicator Alarms and er- The system has been paused due to
displays a constant light. and the rors an alarm. To resume the run, acknowl-
Alarm and error indicator flashes. edge the alarm and continue the run
in UNICORN.
All indicators are lit in a wave pattern. Re-programming A module is being re-programmed to
be compatible with the current instru-
ment configuration.
In this section
This section contains the following subsections:
2.4.2 Mixer 37
2.4.10 pH valve 60
2.4.13 UV monitors 69
Introduction
This section describes the design and main functions of the system pumps, and also the pump piston
rinsing systems. The system can also be equipped with an external, optional sample pump, see Sec-
tion 3.4 Sample pump S9 and S9H, on page 107.
ÄKTA pure 150 P9H A and P9H B P9H 0.01 – 150 ml/min 5 MPa
Note:
When running the
Column packing
flow instruction, the
maximum flow rate
is 300 ml/min.
1 2 3
8 7 6 5 4
Part Description
3 Connections to pump piston rinsing system: Tubing is connected between the pumps
and the Pump piston rinsing system tube (6)
7 System pump B
8 System pump A
5 4
Part Description
4 Outlet tubing
5 Inlet tubing
1 2
Part Description
1 System pump A
2 System pump B
2.4.2 Mixer
Part Description
1 Outlet
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with
an ÄKTA pure 25 configuration. The maximum pressure
for Mixer chamber 15 ml is 5 MPa.
3 Inlet
Part Description
Note: Inlet and outlet ports are not visible in the picture. They are located on the opposite side of the
Valve connection block.
Inlet valve A V9-IA V9H-IA Enables automatic change between different buffers
and wash solutions (seven inlet ports) .
Can be used together with Inlet valve B to generate
gradients by mixing buffer from System pump A and
buffer from System pump B.
Inlet valve B V9-IB V9H-IB Enables automatic change between different buffers
and wash solutions (seven inlet ports).
Can be used together with Inlet valve A to generate
gradients by mixing buffer from System pump A and
buffer from System pump B.
Inlet valve AB V9-IAB V9H-IAB Enables automatic change between different buffers
and wash solutions (two A and two B inlet ports).
Can be used to generate gradients by mixing buffer
from System pump A and buffer from System pump
B.
Sample inlet valve V9-IS V9H-IS Enables automatic loading of up to seven samples
when used together with a sample pump.
Inlet valve X V9-IX V9H-IX Increases the total number of inlets to the system.
Can be used in two different configurations X1 and
X2. The configurations are called V9-X1 and V9-X2
for ÄKTA pure 25 and, V9H-X1 and V9H-X2 for ÄKTA
pure 150.
The modular design of ÄKTA pure allows the use of several combinations of inlet valves.
The possible combinations of Inlet valve A, Inlet valve B and Inlet valve AB are:
• one Inlet valve A
• one Inlet valve B
• Inlet valve A and Inlet valve B,
• Inlet valve AB together with Inlet valve A or Inlet valve B,
• one Inlet valve AB,
or
1 2
Part Description
Out Out
Port Description
1 2
4 3
Part Description
1 A inlet ports
2 B inlet ports
Note: Inlet valve AB does not have any integrated air sensor.
Port Description
A1-A2 Inlet ports A1 and A2 of Inlet valve AB are used when buffers or samples
should be delivered to System pump A
B1-B2 Inlet ports B1 and B2 of Inlet valve AB are used when buffers or samples
should be delivered to System pump B
Part Description
Port Description
Port Description
1-8 Inlets
Connect tubing
The table below shows the tubing and connectors that is delivered together with the optional inlet
valves.
A1-A2 Inlets to FEP, o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 1500
and Inlet valve AB i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
B1-B2 Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
A1-A7 Inlets to FEP, o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 1500
Inlet valve A i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
B1-B7 Inlets to FEP, o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 1500
Inlet valve B i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
InA From Inlet valve FEP, o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 300
A or i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
Inlet valve AB to Ferrule (yellow), Ferrule (blue),
System pump A 1/8" 3/16"
InB From Inlet valve FEP, o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 300
B or i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
Inlet valve AB to Ferrule (yellow), Ferrule (blue),
System pump B 1/8" 3/16"
S1-S7 Sample inlets to FEP o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 1000
Sample inlet i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
valve Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
Buff Buffer inlet to FEP o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 1000
Sample inlet i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
valve Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
InS Sample inlet FEP o.d. 1/8", FEP, o.d. 3/16", Tubing connec- Tubing connec- 580
valve to i.d. 1.6 mm i.d. 2.9 mm tor, 5/16" with tor, 5/16" with
Sample Pump Ferrule (yellow), Ferrule (blue),
1/8" 3/16"
Note: Narrow inlet tubing is available for S1-S7. Refer to Section 9.3 Tubing and connectors, on
page 420 for more information.
Port Description
In Port in which the flow enters the valve. Should be connected to the System pressure
monitor outlet.
Out Port from which the flow leaves the Mixer valve and bypasses the Mixer. Connect to the
injection valve SaP port.
Mixer Port from which the flow leaves the valve and is directed to the Mixer.
Connect tubing
The table below shows recommended tubing and connectors.
3-1 Pressure monitor R9 PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 160
to i.d. 0.75 mm i.d. 1.00 mm 1/16"
Mixer valve V9-M port
In
3-2 Mixer valve V9-M port PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 330
Mixer i.d. 0.75 mm i.d. 1.00 mm 1/16"
to Mixer M9
3-3 Mixer valve V9-M port PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 260
Out to i.d. 0.75 mm i.d. 1.00 mm 1/16"
Injection valve V9-Inj
port SaP
By-pass The flow from the system pumps bypasses the mixer.
Mixer The flow from the system pumps is directed to the Mixer. Mixer is the default flow path.
Port Description
LoopE Port for connection of a loop. Used to empty the loop into the flow path.
Manual load - Default posi- The system flow is directed onto the column or column valve. Sample
tion of the valve can be manually injected into the loop.
Excess sample leaves the valve through waste port W1.
Inject The system flow is directed through the loop and onto the column
or column valve. If a mixer valve or the sample pump is used, the
flow entering the SaP port is directed to waste port W2.
System pump waste The system flow is directed to waste port W1. If a mixer valve or the
sample pump is used, the flow entering the SaP port is directed to
the column or the column valve.
Direct inject The flow entering the SaP port is directed to the column or the col-
umn valve. This position is used with:
• the sample pump, or
• a mixer valve and System pump A, bypassing the mixer.
Flow entering the SyP port is directed to waste port W1.
Sample pump load The flow entering the SaP port is directed to the loop. This position
is used with:
• the sample pump, or
• a mixer valve and System pump A, bypassing the mixer.
Excess sample leaves the valve through waste port W1. The flow
entering the SyP port is directed to the column or the column valve.
Sample pump waste The flow entering the SaP port is directed to waste port W2. This
position is used with:
• the sample pump, or
• a mixer valve and System pump A.
The flow entering the SyP port is directed to the column via the loop.
Note: • In order to avoid sample carry-over when switching techniques for loading samples, wash
the injection valve with buffer between the loading of two different samples. For example,
when switching from loading sample onto the loop to loading sample directly onto the
column with the valve in Direct inject position.
• Make sure that the SaP port is plugged with a stop plug if neither the sample pump nor
the mixer is used.
Port Description
Note: Ports denoted by the letter F are used for filling the loop and ports denoted by the letter E are
used for emptying the loop.
Position 1-5 The flow direction depends on the Injection valve position.
By-pass The flow bypasses the loop(s). By-pass is the default flow path.
Connect tubing
The table below shows recommended tubing and connectors.
1 Connect port E on the Loop valve to port LoopE on the Injection valve.
2 Connect port F on the Loop valve to port LoopF on the Injection valve.
3 Connect one or many loops to the Loop valve. See Section 5.7 Sample application, on
page 179.
Note:
Always use the the first positions of the valve for the connected loops (e.g., if three loops
will be used, use port 1F-3F and the corresponding ports 1E-3E) to avoid cross-contamina-
tion.
Note: It is possible to place the Loop valve in other positions in the flow path than the one described
above. However, the volume used for washes will then be incorrect, just as the system config-
uration shown in the process picture.
Label in Function
V9-C V9H-C Connects columns to the system and allows the user
to choose column, flow direction through the column,
or to bypass the columns.
Up to five columns can be connected to the valve.
V9-Cs V9H-Cs Connects one column to the system and allows the
user to choose flow direction through the column,
or to bypass the column.
The inlet and outlet ports of Column valves V9-C, V9H-C and V9-C2, V9H-C2 have built in pressure
sensors that measure the actual pressure over the column. For further information on the pressure
sensors, see Function of pressure monitors integrated in Column valves V9-C or V9H-C, on page 68.
Column valves V9-Cs and V9H-Cs have no pressure sensors. See Section 5.5 Connect a column, on
page 172 for information on how to set the pressure alarm to protect the column when using V9-Cs or
V9H-Cs.
1 2
Part Function
By-pass
By-pass
Port Description
Port Description
By-pass The flow bypasses the column(s). By-pass is the default flow path.
Down flow The flow direction is from the top of the column to the bottom of the column.
Down flow is the default flow direction.
Up flow The flow direction is from the bottom of the column to the top of the column.
1A 1A 1A
1B 1B 1B
Port Description
Flow Description
path
By-pass The flow bypasses the column. By-pass is the default flow path.
Flow Description
path
Down The flow direction is from the top of the column to the bottom of the column. Down flow
flow is the default flow direction.
Up flow The flow direction is from the bottom of the column to the top of the column.
Connect tubing
The table below shows recommended tubing and connectors.
Tubing Tubing
Tubing
Connection Connector length
label ÄKTA pure 25 ÄKTA pure 150 (mm)
Injection valve to
5 100
Column valve
Column valve to
6 160
UV monitor
1 If no Column valve is installed, remove the Union F/F between tubing 5 and tubing 6
2 Connect tubing between Injection valve, Column valve and UV monitor according to the
table above.
Note: The built-in pressure sensors for column valve(s) V9-C, V9-C2,V9H-C and V9H-C2 have to be
re-calibrated after installation. See Calibrate the monitors, on page 305.
Note: When using two column valves, only the pre-column pressure sensor on the first valve and
the post-column pressure sensor on the second valve are used.
The Versatile valve has four available sets of flow paths; two where a single flow channel is used and
two where the flow can be directed through two different channels simultaneously.
Flow Description
path
1-3 A single flow channel where the flow is directed between port 1 and port 3.
2-4 A single flow channel where the flow is directed between port 2 and port 4.
1-4 and Two simultaneously used flow channels where the flow is directed between port 1 and
2-3 port 4 and between port 2 and port 3.
1-2 and Two simultaneously used flow channels where the flow is directed between port 1 and
3-4 port 2 and between port 3 and port 4.
2.4.10 pH valve
Part Description
1 pH valve
2 pH electrode
3 pH flow cell
4 Flow restrictor
By-pass
Port Description
W3 To Waste
Calibration Flow path used when calibrating the pH monitor and when filling the pH flow cell
with storage solution. The Cal port is used to inject solution into the flow cell using
a syringe.
Excess solution leaves the valve through port W3.
pH monitor
The pH monitor continuously measures the pH of the buffer and eluted proteins when the pH electrode
is inline. A pH electrode can be installed in the pH flow cell. For instruction on how to install the pH
electrode, see Section 7.5.2 Replace the pH electrode, on page 275. The pH electrode should not be ex-
posed to more than 0.5 MPa during a normal run, but can withstand pressure spikes of 0.8 MPa. It is
therefore important to place it after the column.
The illustration below shows the location of the pH flow cell and a pH electrode installed in the pH
valve.
Part Description
1 pH flow cell
2 pH electrode
Connect tubing
The table below shows recommended tubing.
8pH Conductivity monitor PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 180
to port In i.d. 0.50 mm i.d. 0.75 mm 1/16"
1R Flow restrictor to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 80
port ToR i.d. 0.50 mm i.d. 0.75 mm 1/16"
2R Flow restrictor to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 80
port FrR i.d. 0.50 mm i.d. 0.75 mm 1/16"
9pH Port Out to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 160
Outlet valve i.d. 0.50 mm i.d. 0.75 mm 1/16"
Note: The dimensions of the PEEK tubing depends on which tubing kit that is used. See Section 9.3
Tubing and connectors, on page 420 for more information.
Label in Description
Port Description
In Inlet port
Port Description
W Waste port
In
Frac Out1
W
V9-Os
Port Description
In Inlet port
W Waste port
Connect tubing
The table below shows recommended tubing and connectors.
9 Flow restrictor to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 135
Outlet valve i.d. 0.50 mm i.d. 0.75 mm 1/16"
9pH pH valve to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 160
Outlet valve i.d. 0.50 mm i.d. 0.75 mm 1/16"
Out1 - Outlets from the ETFE, o.d. 1/8", ETFE, o.d. 1/8", Fingertight connector, 1500
Out10 outlet valves i.d. 1 mm i.d. 1 mm 1/16"
Note: If ÄKTA pure previously has been configured without an Outlet valve, remove the Union F/F
between tubing 9 and tubing W before installation of the Outlet valve. Then connect tubing 9
to the In port on the Outlet valve and the waste tubing W to the W port on the Outlet valve.
Introduction
This section describes the location and function of the pressure monitors.
Up to four pressure monitors are included in ÄKTA pure.
Part Description
1 Pressure monitor
Pressure monitors are integrated in the inlet and outlet ports of the column
valve to measure pre-column pressure (A), and post-column pressure (B).
The Delta column pressure (C), or pressure drop, is the difference between
the pre- and the post-column pressure.
Pressure alarms can be set for both the pre-column pressure and the
Delta-column pressure. Pressure control of flow can use either the pre-
column pressure or the Delta-column pressure.
The table below shows the pressure monitor stettings when the instrument
is configured with one or two column valves. Set the pre-column and post-
column pressure monitors by turning the arrows on the switches on the
left and right sides of the module with a screwdriver. Set Node ID for the
column valve by turning the arrows in the two rotating switches at the
back of the module.
Note:
Column valves V9-Cs and V9H-Cs do not have pressure monitors.
Settings
Module Label Pressure
Node ID
monitor
2.4.13 UV monitors
Introduction
This section describes the design and function of UV monitors U9-M and U9-L.
The modules include a monitor unit and a detector with a UV flow cell.
1 2
4 3
Part Description
1 Inlet
3 UV detector
4 Outlet
1 2 3
5 4
Part Description
1 Inlet
2 UV monitor U9-L
4 Outlet
5 UV detector
Note: When using two UV monitors, the signal from the first UV monitor is by default used for peak
fractionation. This can be changed by editing the text instruction Fraction Collection:Peak
fractionation parameters:Signal source and choosing UV 2nd as Signal source.
Note: When using two UV monitors with different cell lengths to increase the UV absorption dynamic
range, the U9-L signal comes from the real cell length and has to be calibrated for exact calcu-
lations. The U9-M signal is automatically calibrated to nominal cell length.
UV monitor U9-L, 2nd can be located anywhere in the flow path and is therefore shown in the Process
Picture in UNICORN as a component without a fixed place. This means that it is possible to place U9-
L, 2nd before the other UV monitor in the flow path.
Note: If U9-L 2nd is placed on the high pressure side of the column, pressure limits have to be con-
sidered. See UV monitor options, on page 416
Connect tubing
The table below shows the tubing and connectors to be used with UV monitor U9-L.
Note: If a second UV monitor is used, the tubing for this has to be cut manually.
6 Tubing to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 160
UV monitor U9-L i.d. 0.50 mm i.d. 0.75 mm 1/16"
7 Tubing from PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 170
UV monitor U9-L i.d. 0.50 mm i.d. 0.75 mm 1/16"
Note: To perform a run with the flow in reverse direction through UV monitor U9-L, a longer tubing
7 is needed. Replace the 170 mm tubing from the UV monitor with tubing that is 210 mm long
and adjust the delay volume accordingly. For example, changing from 170 mm to 210 mm for
0.5 mm i.d. tubing increases the delay volume with 8 µl.
Part Description
1 Inlet
3 Conductivity monitor
4 Outlet
Connect tubing
The table below shows recommended tubing and connectors.
7 UV monitor U9-L to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 170
Conductivity monitor C9 i.d. 0.50 mm i.d. 0.75 mm 1/16"
8 Conductivity monitor C9 to PEEK, o.d. 1/16", PEEK, o.d. 1/16", Fingertight connector, 95
Flow restrictor i.d. 0.50 mm i.d. 0.75 mm 1/16"
4 3
Part Function
1 Flow restrictor
2 Holder
3 Inlet
4 Outlet
If ÄKTA pure is fitted with pH valve V9-pH or V9H-pH the Flow restrictor has to be moved from the
Conductivity monitor to the pH valve.
The illustration below shows Flow restrictor FR-902 fitted on the pH valve.
2
3
1
Part Function
1 Flow restrictor
Node ID
All of the available optional modules are preconfigured to give the desired function. However, the
function of a module or valve can be changed by changing its Node ID. Node ID is a unit number des-
ignation that is used by the instrument to distinguish between several units of the same type.
In a troubleshooting situation it may be useful to check a valve's or module's Node ID. Refer to Sec-
tion 9.16 Node IDs, on page 502 for a list of Node IDs for valves and modules.
Note: The function of a valve or module is defined by its Node ID, not by its physical position.
The Node ID is set by positioning the arrows of the one or two rotating switches at the back of the
valve. Use a screwdriver to set the arrows of the switches to the desired number.
CAUTION
Disconnect power. Always switch off power to the ÄKTA pure instrument before
replacing any of its components, unless stated otherwise in the user documentation.
Step Action
1 Disconnect power from the instrument by switching off the instrument power switch.
2 Loosen the connectors and remove the tubing from the existing module.
Note:
This step does not apply for a Module Panel.
Step Action
Note: A warning message is displayed at start up if a module has been installed in the instrument
but not added to the current system configuration in UNICORN.
2.6 Accessories
Introduction
This section describes the holders and other available accessories. These are used to attach and organize
columns, tubing and bottles to the ÄKTA pure instrument. The holders are attached to the instrument
using the holder rails on the left side and the front of the instrument.
Rail extension
The Rail extension rod can be used to attach accessories, eg., column holders or a Multi-purpose
holder. The rod has extra rails on both sides. Push the button of the rod to attach it to a holder rail.
The illustration below shows the Rail extension rod.
1 2
Part Function
1 Button
2 Extension rod
Multi-purpose holder
The Multi-purpose holder can be used to attach accessories, eg., a Loop holder or a cassette. Attach
the holder to a holder rail.
The illustration below shows the Multi-purpose holder.
1 2
Part Function
4 Tab
3 4 3
Loop holder
The Loop holder can be used to attach up to five 10 ml sample loops. Use two Multi-purpose holders
to attach the holder to a holder rail.
The illustration below shows the Loop holder.
Part Function
Column holder
The Column holder has one position for medium sized columns and one position for small sized columns.
The Column holder can also be used for bottles. Use two holders to attach long columns.
The illustration below shows the Column holder.
1 2 3
Part Description
3 Tab
Column clamp
The column clamp can be used to attach small sized columns. Use two clamps to attach long columns.
The illustration below shows the Column clamp.
Part Description
2 3
Part Description
1
1 Snap-in to holder rails
2 Button
Part Function
5 6
1 Lower tubing
2 Lower part
3 Snap-in-strips
7 4 Attachment part
4 5 Upper tubing
8 6 Lever
3 7 Upper part
8 Upper connector
9 Lower connector
9
1
2 3
Part Description
2 Tab
2
Part Description
3 1 Positions for tubing
1
2 Tab
Bottle holder
The Bottle holder is used for holding bottles. For example, the Bottle holder can be attached to the
holder rails to hold a sample bottle.
The illustration below shows the Bottle holder.
Part Description
2
3 Part Description
2 1 Air sensor
1
2 Air sensor adapter
3 Bottle holder
Module Panel
Description
All positions in ÄKTA pure must be occupied. Positions not used for core or optional modules must be
fitted with a Module Panel.
Module Panels are installed in the same way as the other optional modules and the cable inside must
be connected to the Module Panel, see Hardware installation of a module, on page 76.
When an optional module is replaced by a Module panel, the removed module has to be deselected
in the software configuration, see Edit system properties, on page 127.
Illustration
The illustration shows the Module Panels installed in the example configuration used in this manual.
Extension box
Description
The Extension box can be used to install extra modules on the ÄKTA pure instrument outside the system
chassis when the positions on the chassis are filled. It is possible to install up to six Extension boxes
with extra modules when using ÄKTA pure. See the Extension box instruction for more information.
Location
The illustration below shows an Extension box with a module mounted on the side of ÄKTA pure.
In this chapter
This chapter contains the following sections:
Tubing connections
Air sensor L9-1.5
L9-1.5 and inlet FEP, o.d. 1/8", i.d. FEP, o.d. 3/16", i.d. Tubing connector, Tubing connector, See note
valves 1.6 mm 2.9 mm 5/16" + Ferrule 5/16" + Ferrule
(yellow), 1/8" (blue), 3/16"
Note: When sample is loaded at high flow rate and the external air sensor is placed before the pump
that is used for loading the sample, it is necessary to use longer tubing to ensure that no air
reaches the pump. Use tubing with the minimum lengths given below between the valve located
before the pump and the external air sensor. The length applies for maximum flow rate, but
shorter tubing can be used at lower flow rates.
• Sample pump S9: 40 cm.
• Sample pump S9H: 20 cm.
Injection valve and L9-1.2 PEEK, o.d. 1/16" Fingertight connector, 100
1/16"
L9-1.2 and Column valve/the PEEK, o.d. 1/16" Fingertight connector, 100
connected column 1/16"
In this section
This section contains the following subsections:
3.2.1 Function 90
3.2.1 Function
Introduction
Fraction collector F9-C can collect fractions in deep well plates, tubes of different sizes or bottles. Up
to six cassettes for deep well plates and tubes can be used. The cassettes can be used in any combi-
nation and are placed on the Cassette tray. A rack for 50 ml tubes and a rack for 250 ml bottles are
also available.
Scanner functions detect which types of trays, racks, cassettes and deep well plates that are used in
each run.
The Cassette tray or one of the racks is placed inside the fraction collector. A height exclusion bar en-
sures that the bottles, tubes and deep well plates are correctly positioned and cannot damage the
Dispenser head. The Tray catch and positioning discs on the floor of the fraction collector guide the
Cassette tray or the rack into the correct position.
Scanning of Cassettes
When the door of the fraction collector is closed automatic scanning is performed. There are two types
of scanning procedures:
• Full scan: Scanning of Cassette type codes to determine which types of Cassettes are used, and
scanning of rows and columns in deep well plates to identify which types of plates are used (24,
48, or 96 wells). Full scan is performed only when the system is in state Ready.
• Quick scan: Scanning of Cassette type codes to determine which type of Cassettes are used. Quick
scan is performed during the run to ensure that correct Cassettes are placed in the Fraction collector.
Introduction
This section provides illustrations of Fraction collector F9-C. The main features and components are
indicated.
Front view
The illustration below shows the main parts of the exterior of Fraction collector F9-C.
1 2 3 4 5 6
Part Description
1 Fractionation indicator
Symbol indicating that fractionation is ongoing. Do not open the door while the indicator
is lit.
2 Door
3 Window
4 Door handle
6 Vents
Rear view
The illustration below shows the rear view of Fraction collector F9-C.
3 2
Part Description
1 Vents
3 Waste tube
Interior
The illustration below shows the main parts of the interior of Fraction collector F9-C.
3 4 5 6 7 8
2 9
10
11
12
14 13
Part Description
3 Lamp
4 Tubing guide
5 Tubing connection
6 Dispenser head
7 Fractionation arm
8 Tubing guide
10 Waste funnel
11 Waste tube
12 Tray catch
14 Tray guides
Dispenser head
The illustration below shows the Dispenser head of Fraction collector F9-C.
1 2 3
6 5
Part Description
1 Dispenser head
2 Nozzle
Introduction
Fractions can be collected in deep well plates and in tubes of different sizes. A number of cassettes
and racks for different tubes and deep well plates are available. The cassettes are placed on a rack
with six cassette positions. The Cassette type codes are scanned by the Cassette code reader to deter-
mine the type of Cassette.
Note: The tray and racks are inserted into the fraction collector with the GE-logotype facing outwards.
Note: Do not use the Cassette tray when a rack for tubes or bottles is placed in the fraction collector.
Part Description
1 Cassette tray
3 Cassette
QuickRelease function
The Cassettes for the smaller tube sizes (3, 8, and 15 ml) have a built-in QuickRelease function. The
QuickRelease function enables easy handling of tubes in the Cassettes. With the QuickRelease device
in lock position the tubes are fastened in the Cassette and can easily be emptied. With the QuickRelease
function in release position, the Cassette can easily be loaded with tubes and used tubes can easily
be discarded.
Step Action
3 After fractionation, pull the QuickRelease device to the release position and remove the
tubes containing the fractions of interest.
Step Action
5 10.5 12 70 76 VWR™
3 15
8 25
15 50
50 150
250 150
Property Specification
BD Biosciences 353966
Connect tubing
Fraction collector F9-C is delivered with all internal tubing in place. The tubing between the fraction
collector and purification instrument need to be installed.
Follow the instructions in the table below to connect the tubing from the ÄKTA pure instrument to the
fraction collector.
Step Action
1 Connect the tubing Frac to the Frac port on the outlet valve on the ÄKTA pure instrument.
2 Connect the other end of the Frac tubing to the inlet port on the fraction collector.
3 Adjust the delay volume setting in UNICORN, see Section 9.8.8 System settings - Tubing
and Delay volumes, on page 459 for more details.
Function
The fraction collector collects fractions from ÄKTA pure purification runs.
The fraction collector can be used for:
• Fixed volume fractionation
• Peak fractionation
• Combined fixed volume fractionation and peak fractionation
Fraction collector F9-R has the following function for reducing sample spill during fractionation:
• Drop Sync
3 4
5
2 6
1 7
Part Function
1 Lock knob
3 Delivery arm
4 Tubing connector
5 Tube sensor
6 Collection tubes
7 Tube rack
8 Base unit
1 22
Part Function
1 Node ID switch
Available tubes
For Fraction collector F9-R the fractions are collected in tubes of different sizes.
Tubes with the following diameter can be used with Fraction collector F9-R:
• 12 mm
• 18 mm
• 30 mm
The tubes can have a tube length between 50-180 mm.
1
2
6
5 3
4
Part Function
1 Single cutout
2 L-shaped cutout
3 Bowl
4 Tube support
5 Tube guide
6 Tube holder
Note: Note that the tube guide has both single and L-shaped cutouts, while the tube holder only has
single cutouts. See Single and L-shaped cutouts, on page 200 for more information.
1 Lift out the Tubing holder (4) from the Delivery arm (1).
3
5
4
2 Loosen the nut of the Tubing holder. Do not remove the Tubing holder nut (5) from the
Tubing holder.
4 Place the Tubing holder with the tubing over the Tube adjustment cavity (2) of the Delivery
arm. Push the tubing down against the bottom of the Tube adjustment cavity, and then
fingertighten the Tubing holder nut. This ensures the correct length of the exposed tubing
end (6).
6 For Fraction collector F9-R: Connect the tubing from the fraction collector to the port
Frac on the outlet valve.
For Fraction collector F9-R, 2nd: Connect the tubing from the second fraction collector
to:
• port Out 10 on Outlet valve V9-O or V9H-O
• port Out 1 on Outlet valve V9-Os or V9H-Os
7 Adjust the delay volume setting in UNICORN to the volume of the tubing, see Section 9.8.8
System settings - Tubing and Delay volumes, on page 459 for more details.
Front view
3 4 5
2 6
9 8
Part Function
7 Vents
9 Inlet manifold
Rear view
Part Function
3 Vents
4 5
Part Function
3 Inlet tubing, from the rinsing liquid tube to the left pump head (lower rinsing system
connection) .
4 Tubing between the left pump head (upper rinsing system connection) and the right
pump head (lower rinsing system connection).
5 Outlet tubing, from the right pump head (upper rinsing system connection) to the rinsing
liquid tube.
InS Sample inlet valve port FEP FEP Tubing connector Tubing connector 580
Out to sample pump o.d. 1/8", o.d. 3/16", 5/16" with Ferrule 5/16" with Ferrule
inlet manifold i.d. 1.6 i.d. 2.9 (yellow), 1/8" (blue), 3/16"
mm mm
3S Sample pump pressure PEEK, PEEK, Fingertight con- Fingertight con- 530
monitor to Inlet valve o.d. 1/16", o.d. 1/16", nector, 1/16" nector, 1/16"
port SaP i.d. 0.75 i.d. 1 mm
mm
Note: The recommended location of the sample pump is on the bench to the left of the ÄKTA instru-
ment. If the sample pump is placed elsewhere, longer tubing may be needed.
3.5 I/O-box E9
About this section
This section describes the design and the function of the I/O-box E9
In this section
This section contains the following subsections:
Location
The illustration below shows the I/O-box, its recommended location and connection.
Part Description
1 I/O-box
3 Multi-purpose holder
4 Clip
Connectors
Part Description
Analog in/out Signal connector for analog input and output signals.
UniNet-9 Connector used to connect the I/O-box to the ÄKTA pure instru-
ment.
Digital in/out Signal connector for digital input and output signals.
5 4 3 2 1
Part Function
1 Analog in signal 1 +
4 Analog in signal 2 +
Analog signals
All analog input and output signals are confined to the same Analog in/out connector.
Parameter Description
Parameter Description
Default output The user defines the default output level for the selected signal.
Full scale output The user sets the desired output signal value of, e.g., mAU, % or mS, which
will corresponds to the full scale output voltage 1000 mV.
Negative full scale out- The corresponding negative full scale output value is set automatically.
put For example, if the full scale output is set to 2000 mAU, a UV value of 500
mAU will give an output voltage of 250 mV, and -100 mAU will give -50
mV output voltage. A signal value of 0 mAU, 0% or 0 mS will always give
an output voltage of 0 mV.
Fix point Zero is always a fix point. A desired output signal of, i.e., 0 mAU, 0% or 0
mS corresponds to an output voltage of 0 mV. At power-on, an output
signal of 0 mV is transferred to the connected equipment, until the output
signal range values are set.
Note: No warning will be displayed in UNICORN if the analog output signal exceeds the set full scale
output value or is less than the set negative full scale output value.
5 4 3 2 1
Part Function
1 Digital in signal 1
2 Digital in signal 2
3 Digital in signal 3
4 Digital in signal 4
5 Signal ground
Digital signals
All digital input and output signals will be confined to the same D-sub connector and have a common
ground. The four input signals will be scanned synchronously, and the outputs will be set synchronously.
Parameter Value
Channels 2
Range ±2000 mV
Input impedance 1 MΩ
Analog output
Parameter Value
Channels 2
Range ±1000 mV
Digital input
Parameter Value
Channels 4
Digital output
Parameter Value
Channels 4
Required material
The following material is required:
• Flat-blade screwdriver, 2 mm
• Shielded cable with 9 conductors, 4 to 8 mm diameter
Instruction
Follow the instructions to connect one or two external cables to the supplied D-sub connectors.
Step Action
1 Open the connector housing by removing housing screw and unlatch the housing top
shell using a flat-blade screwdriver.
2 Locate the connection block mounted on the PCB board. The screw terminals has numbers
corresponding to the connector contacts.
1 2 3 4 5
6 7 8 9
4 Loosen the strain relief clamp and insert the cable with the shield under the strain relief
clamp. Fasten the strain relief clamp over the cable shield.
Tip:
The connection block can be rotated inside the housing in order to position the screw ter-
minals for left side or right side cable entry.
6 Close the housing top shell with the latch and screw the housing together.
System properties
Follow the instruction to update the system properties.
Step Action
3 Select the I/O-box (E9) or I/O-box 2 (E9) check box in the Component selection list. Then
click OK.
System settings
Default values for digital out ports, noise reduction and configuration of analog out ports can be set.
Digital out X Sets the value of the signal sent out by digital port number X to
either 0 or 1. The default value is 1.
Noise reduction analog in X Filters the noise in the analog signal in port number X.
Alarm analog in X Enables or disables the alarm for the analog signal in port number
X. When enabled, it sets the alarm limits for the analog signal. If
the alarm is enabled and the analog signal falls outside the set
limits, an alarm will be triggered and the method will be paused.
Alarm digital in X Enables or disables the alarm for the signal in digital port number
X. The alarm can be triggered by either of the signal values, 0 or
1. If the alarm is enabled and the condition set in ‘Value’ occurs,
an alarm will be triggered and the method will be paused.
Configure analog out X Enables the user to send one of the pre-defined signals (UV signal,
conductivity, temperature, pH or concentration of eluent B) to the
analog out port number X, and also to set the range of that signal.
Note: The delay volume has to be updated if an external component is added to the flow path.
F-type D-type
Constraints
The table below indicates usage constraints for the different external modules.
I/O-box E9, 2nd F-type I/O-box E9, 2nd requires I/O-box E9.
External air sensor L9, 2 F-type External air sensor L9, 2 requires External air sensor L9.
External air sensor L9, 3 F-type External air sensor L9, 3 requires External air sensor L9,
2.
External air sensor L9, 4 F-type External air sensor L9, 4 requires External air sensor L9,
3.
Fraction collector F9-C D-type Fraction collector F9-C requires an Outlet valve.
Fraction collector F9-R F-type Fraction collector F9-R requires an Outlet valve.
Fraction collector F9-R, 2nd F-type Fraction collector F9-R, 2nd requires Fraction collector
F9-R or Fraction collector F9-C.
Sample pump S9 and S9H D-type Cannot be used at the same time as Mixer valve V9-M
or V9H-M.
Note: To optimize signal quality, the total cable length connecting all external modules to the ÄKTA
pure instrument should not exceed 10 m.
NOTICE
Do not connect any module to the connector Test on the ÄKTA pure instrument.
Note: Plug all unused UniNet-9 ports (ports 1 to 8) on the ÄKTA pure instrument with jumpers.
Note: The connector Test should be protected by a plastic lid. Do not plug the connector with a
jumper.
4 System configuration
In this chapter
This chapter contains the following sections:
8 9 11 12 13
6 7 10
1 2 3 4 5
All positions in ÄKTA pure must be occupied. Positions not used for core or optional modules should
be fitted with a Module Panel.
Column valve V9-C or 8 (V9-C or V9H-C) These positions give the shortest possible
V9H-C with 9 (V9-C2 or V9H-C2) flow path.
Second column valve V9-
C2 or V9H-C2
Sample inlet valve V9-IS or 8 Cannot be used at the same time as Mixer
V9H-IS valve V9-M or V9H-M.
Inlet valve V9-X1 or V9H- None Inlet valve V9-X1 or V9H-X1 has no con-
X1 straints, and has no support in the
method phases.
Inlet valve V9-X2 or V9H- None Inlet valve V9-X2 or V9H-X2 requires Inlet
X2 valve V9-X1 or V9H-X1, respectively. It has
no support in the method phases.
Loop valve V9-L or V9H-L 8 Loop valve V9-L or V9H-L has no con-
straints, but is connected with the injec-
tion valve by replacing the loop.
4 Connect tubing.
Step Action
2 Select the component type of interest from the Component types list.
Result: All available components are shown in the Component selection list.
• Click the checkbox to select the added component.
• When applicable, choose the appropriate Property.
Note:
Instrument modules are referred to as Components in UNICORN.
There are five main types of modules (named components in UNICORN) to select from:
• Valves and pumps
• Monitors and sensors
• Fraction collectors
• Other (e.g., I/O-box)
• Core components (always present)
Multiple choices are not shown unless a component has been selected. The second component is only
available when the first has been selected. The selection made is reflected in which instructions and
phase properties that are available.
In this dialog it is possible to set the parameters for the available instructions.
All system settings available for ÄKTA pure are found in Section 9.8 System settings, on page 448.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
There are no system settings available for the inlet valves.
System properties
Follow the instruction below to update the system properties.
Step Action
3 Select Mixer valve (V9-M) or Mixer valve (V9H-M) in the Component selection list.
System settings
The flow rate for Mixer by-pass wash can be set.
Mixer by-pass wash settings Sets the flow rate used during Mixer by-pass wash and defines
wash volumes for mixer by-pass wash options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
System properties
Follow the instruction below to update the system properties.
Step Action
3 Select Loop valve (V9-L) or Loop valve (V9H-L) in the Component selection list.
System settings
The flow rate for Loop wash can be set.
Loop wash settings Sets the flow rate used during Loop wash.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
After selecting Column valve V9-Cs, V9H-Cs or no column valve, in UNICORN, the tube length has to
be defined.
The instructions Tubing: Injection valve to column and Tubing: Sample pump to injection valve are
used in calculations of pre-column pressure when Column valve V9-C or V9H-C (5-columns) is NOT
mounted onto the system. In such cases, there is no pre-column pressure sensor present. The estimated
pre-column pressure is calculated using the measured system pressure and the length of the tubing
between the system pressure sensor and the Column.
The instructions are available only when the Column valve V9-C or V9H-C (5-columns) is not selected
in the component list. The instruction Tubing: Sample pump to injection valve is in addition only
available if the sample pump is selected in the list.
Follow the instructions below to set the tube length between the Injection valve and the Column and
between the sample pump and the injection valve.
Step Action
1 • In the System Control module, choose System:Connect to Systems or click the Con-
nect to Systems icon.
Result: The Connect to Systems dialog opens.
Step Action
2 • Select a system.
• Select Control mode.
• Click OK.
Result: The selected instrument can now be controlled by the software.
3 • Select Tubing and Delay Volumes and select Tubing: Injection valve to column.
• Select the I. D. from the drop-down list.
• Type in the length of the tube in the Length field and click OK.
4 Perform step 3 for the instruction Tubing: Sample pump to injection valve.
Note: The built-in pressure sensors for Column valves V9-C and V9H-C have to be re-calibrated after
installation. See Calibrate the monitors, on page 305.
Note: A Pre-Column pressure alarm shall always be set to protect the column, see Section 5.6 Pressure
alarms, on page 176. Column valve V9-Cs and V9H-Cs does not contain pressure monitors.
See Section 5.6 Pressure alarms, on page176 for how to protect columns when not using Column
valves V9-C and V9H-C.
System properties
Follow the instruction below to update the system properties.
Step Action
3 Select Versatile valve (V9-V), Versatile valve 2 (V9-V), Versatile valve 3 (V9-V) or Versatile
valve 4 (V9-V) in the Component selection list.
System settings
If the valve is placed in the flow path between the UV monitor and the outlet valve, the delay volume
must be set. See Check/Set delay volume, on page 145.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
If the valve is placed in the flow path between the UV monitor and the outlet valve, the delay volume
must be set. See Check/Set delay volume, on page 145.
Note: It is recommended not to alter the default values for restrictor and pH cell delay volumes when
standard modules and standard tubing for flow restrictor are used.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
If the valve is not placed in the recommended position using the standard tubing kit, the delay volume
must be set. See Check/Set delay volume, on page 145.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
If the monitor is not placed in the recommended position using the standard tubing kit, the delay volume
must be set. See Check/Set delay volume, on page 145.
For UV monitor U9-L the flow cell length must be set. This is done as a calibration. See Perform the
calibration, on page 313.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
If the monitor is placed in the flow path between the UV monitor and the outlet valve, the delay volume
must be set. See Check/Set delay volume, on page 145.
System properties
Follow the instruction below to update the system properties.
Step Action
3 Select External air sensor (L9) in the Component selection list, or the appropriate com-
ponent if multiple external air sensors are used.
System settings
The sensitivity of the air sensor can be set.
Note: The sensitivity should be set to Normal when the air sensor is located before the System pump
(Air sensor ext - Before pump A or Air sensor ext - Before pump B). Due to higher pressure
and risk of small air bubbles, the sensitivity should be set to High when the air sensor is located
after the Injection valve (Air sensor ext - after Injection valve).
System properties
Follow the instruction below to update the system properties.
Step Action
Note: Fraction collector 2 is only available in the Component selection list if Fraction collector is
already selected.
System settings
If non-standard tubing is used between the outlet valve and the fraction collector, the delay volume
must be set. See Check/Set delay volume, on page 145.
Fraction settings and numbering mode can be set.
Fraction collector F9-C
Last tube filled Last tube filled: Action when last tube is filled (pause, direct the
flow to one of outlets or direct the flow to waste.
Fraction collector lamp Fraction collector lamp: Lamps in the fraction collector chamber
on or off.
Peak fractionation parameters The Peak fractionation parameters set the detection parameters
for peak collection, that is they decide when a peak starts and
ends. This information is used by the instructions Peak fractiona-
tion and Peak frac in outlet valve in order to start/end the peak
collection.
Fractionation settings Drop sync synchronises tube change to drop release. The available
settings are on or off. It is recommended to use Drop sync for
flow rates below 2 ml/min. Higher flow rates can however be used,
depending on the properties (for example viscosity) of the liquid.
Fractionation numbering mode Determines whether the fraction number is reset at the end of a
method or not.
Note:
The default setting is Reset.
Fractionation settings frac 2 Drop sync on or off. It is recommended to use this setting for flow
rates below 2 ml/min. Higher flow rates can be however be used,
depending on the properties (e.g. viscosity) of the liquid.
Fractionation numbering mode Determines whether fraction number for the second fraction col-
frac 2 lector is reset at the end of a method or not.
Note:
The default setting is Reset.
Peak fractionation parameters Peak fractionation parameters sets the detection parameters
for peak collection, i.e. it determines when a peak starts and ends.
This information is used by the instructions Peak fractionation,
Peak fractionation frac 2 and Peak frac in outlet valve in order
to start/end the peak collection.
System properties
Follow the instruction below to update the system properties.
Step Action
System settings
Default values for digital out ports, noise reduction and configuration of analog out ports can be set.
Digital out X Sets the value of the signal sent out by digital port number X to
either 0 or 1. The default value is 1.
Noise reduction analog in X Filters the noise in the analog signal in port number X.
Alarm analog in X Enables or disables the alarm for the analog signal in port number
X. When enabled, it sets the alarm limits for the analog signal. If
the alarm is enabled and the analog signal falls outside the set
limits, an alarm will be triggered and the method will be paused.
Alarm digital in X Enables or disables the alarm for the signal in digital port number
X. The alarm can be triggered by either of the signal values, 0 or
1. If the alarm is enabled and the condition set in ‘Value’ occurs,
an alarm will be triggered and the method will be paused.
Configure analog out X Enables the user to send one of the pre-defined signals (UV signal,
conductivity, temperature, pH or concentration of eluent B) to the
analog out port number X, and also to set the range of that signal.
Note: The delay volume has to be updated if an external component is added to the flow path.
1 • In the System Control module, choose System:Connect to Systems or click the Con-
nect to Systems icon.
Result: The Connect to Systems dialog opens.
2 • Select a system.
• Select Control mode.
• Click OK.
Result: The selected instrument can now be controlled by the software.
4 • Select Tubing and Delay Volumes and select the delay volume option of interest.
• Check the delay volume in the Volume field and enter a new value if necessary.
• Click OK.
Lock/Unlock function
Follow the instruction below to lock or unlock the Pause and Continue buttons of the Instrument
control panel from UNICORN.
Step Action
Power-save
ÄKTA pure has a power-save mode. The instrument enters Power-save after having been in the Ready
state for a set period of time. Power-save can be used both in room temperature and in cold room
temperature. The system enters the Ready state when a method run, a method queue or a manual
run ends.
To enable Power-save, a system must be connected and in state Ready.
2 • Select Advanced
and
• select Power-save
5 Operation
In this chapter
This chapter contains the following sections:
Checklist
Make sure the system is prepared in accordance with the settings in the method to be run. Depending
on configuration, remember to check:
• which valve ports to use for inlets and outlets
• which column type to use
• which column position to use
• which buffers and samples to prepare
• which sample application technique to use
• that the pH electrode is connected and calibrated
1 3
2
6
4 7 9
5 8
10
W1 W2
16
W
11 12 13 14 15
17
Part Description
1 Pressure monitor
2 Sample pump
Part Description
4 Inlet valve
5 System pump B
6 System pump A
7 Pressure monitor
8 Mixer
9 Injection valve
11 Column valve
12 Column
13 UV monitor
14 Conductivity monitor
15 Flow restrictor
16 Outlet valve
17 Fraction collector
W, W1, W2 Waste
5 2-25 2-50
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with an ÄKTA pure 25 configura-
tion. The maximum pressure for Mixer chamber 15 ml is 5 MPa.
Note: In ÄKTA pure 25 at low flow rates (below 0.2 ml/min) a gradient of sufficient quality may be
achieved with the mixer bypassed. Similarly, for low flow rates in ÄKTA pure 150 the 0.6 ml
mixer might be used.
Note: The 1.4 ml mixer in ÄKTA pure 25 or 5 ml mixer in ÄKTA pure 150 might work up to maximum
flow rate provided that the buffers are easily mixed.
For information on how to install a Mixer chamber, refer to Section 7.8.3 Replace the Mixer, on page 322.
Injection valve Waste ports W1 and W2 Front of the ÄKTA pure instrument.
Outlet valve (optional) Waste port W Front of the ÄKTA pure instrument.
Buffer tray (Rescue drainage) Drainage hole of the Buffer Rear of the ÄKTA pure instrument.
tray
Fraction collector F9-C Fraction collector waste outlet Rear of the fraction collector.
(optional)
CAUTION
Fasten the waste tubing. During operation at high pressure the ÄKTA pure instrument
may release bursts of liquid in the waste tubing. Securely fasten all waste tubing to
the ÄKTA pure instrument and to the waste vessel.
CAUTION
Make sure that the waste vessel will hold all the produced volume of the run. For
ÄKTA pure, a suitable waste vessel should typically have a volume of 2 to 10 liters.
NOTICE
The maximum level of the waste vessel must be lower than the bottom of the ÄKTA
pure instrument.
Step Action
1 Insert the waste tubing from all installed modules, in this example Injection valve, Outlet
valve (W, W1 and W2) and the fraction collector, in a vessel.
2 Make sure that the tubing is securely fastened to the ÄKTA pure instrument:
• Fasten waste tubing from the valves with the clips on the front of the system.
• Fasten waste tubing from the Buffer tray with the clips on the rear of the system.
Step Action
3 Cut the waste tubing to appropriate length. It is important that the tubing is not bent and
will not be submerged in liquid during the run.
Note:
If the tubing is too short, replace it with new tubing. Do not lengthen the tubing as this
might cause obstruction of the tubing.
Connect to system
Follow the instructions to connect the instrument to UNICORN.
Step Action
or
• click Connect to Systems on the System menu.
Result: The Connect to Systems dialog opens.
Step Action
In this section
This section contains the following subsections:
Introduction
This section describes how to prime inlets and purge the System pumps.
Overview
The procedure consists of the following stages:
Stage Description
Tip: The procedures for purging the pump heads and priming the inlets using the Process Picture,
are described below. It is also possible to perform the procedures from the Manual instructions
dialog.
1 Make sure that all inlet tubing that is to be used during the method run is placed in the
correct buffer.
Step Action
4 Connect a 25 to 30 ml syringe to the purge valve of one of the pump heads of the pump
that is being prepared. Make sure that the syringe fits tightly into the purge connector.
5 Open the purge valve by turning it counter-clockwise about three quarters of a turn.
Draw liquid slowly into the syringe until the liquid reaches the pump.
6 Close the purge valve by turning it clockwise. Disconnect the syringe and discard its
contents.
7 Repeat steps 3 to 6 for each piece of inlet tubing that is to be used during the run.
1 Make sure that the piece of waste tubing connected to the Injection valve port W1 is
placed in a waste vessel.
Step Action
• Click Set % B.
Result: Only System pump B is active.
Step Action
6 Connect a 25 to 30 ml syringe to the purge valve of the left pump head of System pump
B. Make sure that the syringe fits tightly into the purge connector.
7 Open the purge valve by turning it counter-clockwise about three quarters of a turn.
Draw a small volume of liquid slowly into the syringe (with a rate of about 1 ml per second).
8 Close the purge valve by turning it clockwise. Disconnect the syringe and discard its
contents.
Step Action
9 Connect the syringe to the purge valve on the right pump head of System pump B, and
repeat steps 6 to 8. Keep the system flow running.
• Click Set % B
Result: Only System pump A is active.
Introduction
This section describes how to prime inlets and purge Sample pumps S9 and S9H.
Overview
The procedure consists of the following steps:
1 Prime all sample inlet tubing to be used during the run
2 Purge the Sample pump
3 Validate purge
4 End the run
1 Make sure that all sample inlet tubing that is to be used during the method run is immersed
in the correct buffers.
2 Make sure that the waste tubing connected to Injection valve port W2 is immersed in a
waste vessel.
Step Action
5 Connect a 25 to 30 ml syringe to one of the purge valves of the pump heads of the
Sample pump. Make sure that the syringe fits tightly into the purge connector.
7 Close the purge valve by turning it clockwise. Disconnect the syringe and discard its
contents.
8 Repeat steps 2-5 for each sample inlet that is to be used in the method run.
1 Make sure that all sample inlet tubing that is to be used during the method run is immersed
in the correct buffers.
2 Make sure that the waste tubing connected to Injection valve port W2 is immersed in a
waste vessel.
Result: The injection valve switches to waste position. This is necessary to achieve a
low back pressure during the purge procedure.
•
Click and close the pop-up window.
Result: a sample pump flow starts.
Step Action
6 Connect a 25 to 30 ml syringe to the left purge valve of the Sample pump. Make sure
that the syringe fits tightly into the purge connector.
8 Close the purge valve by turning it clockwise. Disconnect the syringe and discard its
contents.
9 Connect the syringe to the right purge valve on the Sample pump, and repeat step 6 to
step 8.
Validate purge
Follow the instructions below to check that there is no air left in the pump after performing a purge.
Step Action
WARNING
To avoid exposing the column to excessive pressure, make sure that the pressure
limit is set to the specified maximum pressure of the column. Before connecting a
column to the ÄKTA pure instrument, read the instructions for use of the column.
Methods automatically include a pressure alarm based on the specifications of the chosen column
type. However, when running manual runs you have to set the pressure limits yourself. Also, to protect
the column media, special settings are needed. See Section 5.6 Pressure alarms, on page 176 for more
information on pressure alarms.
Note: Do not overtighten when connecting columns. Overtightening might rupture the connectors
or squeeze the tubing and thereby result in high back pressure.
Note: If no column valve is used, remove the column from the system before running a system wash.
The pressure during a system wash may become too high for the column.
Step Action
3 Connect a suitable tubing to a Column valve port, in this example port 1A.
Step Action
7 When buffer leaves the tubing in a continuous mode and the top part of the column is
filled with buffer, connect the tubing to the top of the column.
Step Action
9 When buffer leaves the tubing at the bottom of the column in a continuous mode, connect
this piece of tubing to the Column valve. Use the port opposite to the one already con-
nected to the column, in this example port 1B. If no column valve is used, connect the
tubing to the next module in the flow path.
10 Click the End icon in the System Control toolbar to end the run.
Step Action
3 • Select the inner diameter of the tubing between the injection valve and the column
from the I.D. drop-down list.
• Type in the tubing Length.
5 Click OK.
Note:
The system now calculates the Pre-column pressure.
Step Action
2 • Select Alarms
and
• select Alarm pre column pressure.
Note: In order to avoid sample carry-over when switching techniques for loading samples, wash the
valve with buffer between the loading of two different samples. For example, when switching
from loading sample in the loop to loading sample directly onto the column with the valve in
Direct inject position.
When using a pump for sample application, it is important to prime inlets and purge the pump before
using the pump to load the sample:
• Sample pump, see further instructions in Section 5.4.2 Sample pump, on page 167
• System pump A, see further instructions in Section 5.4.1 System pumps, on page 161
When loading sample using System pump A and an external air sensor, the sensor should be installed
according to Adapter for air sensor, on page 83.
Note: When sample is loaded at high flow rate and the external air sensor is placed before the pump
that is used for loading the sample, it is necessary to use longer tubing to ensure that no air
reaches the pump. Use tubing with the minimum lengths given below between the valve located
before the pump and the external air sensor. The length applies for maximum flow rate, but
shorter tubing can be used at lower flow rates.
• Sample pump S9: 40 cm.
• Sample pump S9H: 20 cm.
Introduction
There are two ways to load sample directly onto a column:
• a fixed volume is loaded, or
• all the sample is loaded.
To inject all the sample, one of the following configurations is required:
• system pump, external air sensor, inlet valve A and the mixer valve, or
• sample pump and sample inlet valve.
a fixed volume of sample • manually prime the sample inlet tubing with sample, see Prime
sample inlets, on page 167.
• in the Method editor, make the following selections for the
Sample Application phase of the method to be run:
- select Inject sample directly onto column,
- select Inject fixed sample volume and set the volume to be
injected.
• make sure that the flow path from the sample inlet valve up to
the injection valve will be filled with an appropriate buffer:
- make sure that the buffer inlet tubing of the inlet valve is im-
mersed in buffer, and
- enable the function Wash sample flow path with buffer in
the Sample application phase.
all the sample • in the Method editor, make the following selections for the
Sample Application phase of the method to be run:
- select Inject sample directly onto column, and
- select Inject all sample using air sensor.
• select Inject fixed sample volume and set the volume to be injected,
• enable the function Prime sample inlet with and set the volume to be used for priming.
Result: the step Finalize sample injection is automatically deactivated in order to maximize precision
and accuracy. See Minimize sample loss, on page 180 for more information.
Note: If manual priming of the flow path up to the injection valve is preferred, enable Prime
sample inlet with but set the volume to 0 ml.
• If the system pump is used to load the sample: enable the function Wash sample flow path with
buffer after sample application in order to ensure that the correct sample volume is loaded onto
the column.
Introduction
A Superloop allows injection of large sample volumes onto the column. A Superloop can also be used
for multiple injections, for example in a scouting experiment when the same application conditions
are required. Superloop models are available in 10 ml, 50 ml and 150 ml sizes.
A superloop can be connected to either the Injection valve or the Loop valve. When using the Loop
valve, up to five loops can be connected simultaneously.
Note: After loading a Superloop, always plug the Syr port on the Injection valve with a Stop plug.
With a Superloop connected to the valve, an over-pressure may be created during injection.
Step Action
or
• the F port corresponding to the connected E port, eg., 1F, on the Loop valve
Step Action
7 Load sample into the Superloop by emptying the syringe into the Injection valve.
8 Disconnect the syringe and plug the Syr port with a Stop plug.
NOTICE
Glass tube splinter. Make sure to set the sample pressure below the max pressure
of the Superloop before executing a flow in the Manual instructions dialog when the
Superloop is connected.
Tip: The Superloop can also be filled as part of a method run, as set in the Sample Application
phase in the Method Editor. For multiple injections, it may be more convenient to fill the Super-
loop once, as described in the instruction below.
Step Action
2 Make sure that the sample inlet tubing from the sample vessel is connected to the Sample
inlet valve.
• Click .
Result: A sample flow starts, in this example of 10 ml/min.
5 When the Superloop is filled with as much volume as is needed, click the End icon in the
System Control toolbar to end the run.
6 Plug the Syr port on the Injection valve with a Stop plug.
NOTICE
Glass tube splinter. Make sure to set the system pressure below the max pressure
of the Superloop before executing a flow, when the Superloop is connected.
Tip: The Superloop can also be filled as part of a method run, as set in the Sample Application
phase in the Method Editor. For multiple injections, it may be more convenient to fill the Super-
loop once, as described in the instruction below.
Step Action
Step Action
Step Action
8 When the Superloop is filled with as much volume as is needed, click the End icon in the
System Control toolbar to end the run.
9 Plug the Syr port on the Injection valve with a Stop plug.
Introduction
A sample loop is recommended for injection of smaller sample volumes onto the column.
A sample loop can be connected to either the Injection valve or the Loop valve. When using the Loop
valve, up to five loops can be connected simultaneously.
Note: Sample loop is called capillary loop in UNICORN.
1 Connect a suitable sample loop to Injection valve ports LoopF (fill) and LoopE (empty).
Step Action
Step Action
7 Load sample into the sample loop. To avoid sample loss due to siphoning, leave the syringe
in the port until the sample has been injected onto the column during the run.
Tip:
It is recommended to overfill the loop to make sure that the loop is completely filled. Excess
of sample will leave the valve through port W1.
5.8 Fractionation
Introduction
Fraction collector F9-C and Fraction collector F9-R collect fractions from ÄKTA pure purification runs.
The fraction collectors are connected to ÄKTA pure and controlled by UNICORN. Control of the fraction
collector can be achieved automatically in a method run, or manually.
In this section
This section contains the following subsections:
CAUTION
Fire Hazard. Do not fractionate flammable liquids using Fraction collector F9-C.
When running RPC methods, or other procedures using organic solvents, collect
fractions through the outlet valve or Fraction collector F9-R.
Note: The tray and racks can tilt slightly when not fully inserted into the fraction collector and may
harm the fractionation arm. The tilt is due to the height difference between the door and the
floor of the fraction collector. The tendency to tilt is affected by the placement and weight of
the cassettes, tubes or bottles
1 If you are to use cassettes with the QuickRelease function, open the cassettes. For more
information on the QuickRelease function see QuickRelease function, on page 98.
2 Place the tubes and deep well plates in the cassettes. Make sure that the deep well plates
are rotated so that the well marked A1 is positioned above the A1 marking on the cassette.
Step Action
4 Place the cassettes on the Cassette tray. Make sure that the cassette type code (see illus-
tration below) faces the front of the tray marked with the GE logo.
Step Action
• make sure that the Tray catch snaps into closed position, as shown below.
Note:
• The tray can tilt slightly when not fully inserted into the fraction collector and may harm
the fractionation arm. The tilt is due to the height difference between the door and the
floor of the fraction collector. The tendency to tilt is affected by the placement and
weight of the cassettes.
• A height exclusion bar ensures that the tubes or deep well plates are correctly positioned
and cannot damage the Dispenser head.
• make sure that the Tray catch snaps into closed position, as shown below.
Note:
• The rack can tilt slightly when not fully inserted into the fraction collector and may harm
the fractionation arm. The tilt is due to the height difference between the door and the
floor of the fraction collector. The tendency to tilt is affected by the placement and
weight of the tubes or bottles.
• A height exclusion bar ensures that the tubes or bottles are correctly positioned and
cannot damage the Dispenser head.
• Do not use the Cassette tray when the Rack for 50 ml tubes or the Rack for 250 ml
bottles is placed in the fraction collector.
Step Action
Introduction
This chapter describes how to prepare and assemble Fraction collector F9-R before a run.
Fraction collector F9-R is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction
collector can be achieved automatically in a method run, or manually.
Part Function
1
1 Single cutout
2
2 L-shaped cutout
3 Bowl
4 Tube support
5 Tube guide
6
5 3
6 Tube holder
4
Note: Note that the tube guide has both single and L-shaped cutouts, while the tube holder only has
single cutouts.
1 For 30-50 mm tubes, first insert the tube guide from the 18 mm rack using the single cutout, before inserting the
tube support for the 30 mm rack.
Assembly instructions
Follow the instructions below to assemble the Tube rack.
Step Action
1 Insert the Tube support (4), if required, into the bowl (3). The circular marks on the Tube
support should face down.
1
2
6
5 3
4
Note:
When assembling a Tube rack, Single cutouts (1) and L-shaped cutouts (2), are used for
various inserts depending on the length of the collection tubes. See Single and L-shaped
cutouts, on page 200 for detailed information.
2 Insert the Tube guide (5) with the tube position numbers upwards. The Tube guide should
rest about 1 cm above the Tube support.
3 Insert the Tube holder (6) with the tube position numbers upwards:
Check that tube position 1 (7) is directly above tube position 1 (7) of the Tube guide.
Step Action
4 • Push the flexible bowl out at each rib and snap the Tube holder under the top lip of
the rib (8).
Note:
Do not force the tube holder into place as this may damage the lip.
• Check that the surface of the Tube holder is level.
5 Gently move the Delivery arm (10) out to the outer stop.
6 Place the Tube rack (9) over the Central spindle (11) and pull the spring loaded Drive sleeve
(12) out so the Tube rack comes to rest.
9
10
12
11
1 Lift and then lower the Delivery arm (1), and allow it to move in so the Tube sensor (2)
touches the collection tubes of the outer track.
1
Step Action
3 • Adjust the height so that the horizontal mark (5) on the Tube sensor (6) is at the same
level as the top of the flat collection tubes and approximately 2 mm over the top of
the flanged collection tubes (4).
• Lock the Delivery arm at this height with the lock knob.
4 Check that the Tube sensor (9) is in the correct position for the tubes used (10). The eluent
tubing (7) should be above the center of the collection tube.
7
8
10
5 Use the Sensor control (8) to position the Tube holder over the center of the collection
tube.
NOTICE
Never lift Fraction collector F9-R by the Delivery arm. This may damage the Fraction
collector.
Sensor control
The sensor control can be switched between the two positions "small tubes" and "large tubes", indicated
in the illustration below.
The position for large tubes is used for tubes of approximately 18 mm i.d. and larger. The position for
small tubes is used for tubes smaller than 18 mm i.d.
Note that this is a rough approximation. Always check that the eluent tubing is centered above the
collection tube.
Fractionation types
The table below lists the types of fractionation that the fraction collectors can be used for.
Type Description
Fixed volume fractiona- During fixed volume fractionation the fraction collector continuously
tion switches tubes according to the set volume throughout the entire fraction-
ation. This type of fractionation is also known as straight fractionation.
Peak fractionation Peak fractionation can be used to further increase the purity of the collect-
ed protein peaks and minimize the number of tubes used. The monitor
signal is used to determine when to switch the tubes. See Section 9.8.10
System settings - Watch parameters, on page 462 for information about
different watch options.
Combined fixed vol- The two fractionation types listed above can be used in combination.
ume fractionation and Combination of fixed volume and peak fractionation allows fractions col-
peak fractionation lected by fixed volume fractionation and fractions collected by peak frac-
tionation to be directed to different collection tubes.
To be able to analyze different parts of the peak, the fraction size during elution is usually set to a value
smaller than the expected peak volume.
Illustration
The illustration below shows examples of fractionation using fixed volume fractionation and fraction-
ation using peak fractionation.
A) Straight/fixed fractionation
Absorbance
Volume
B) Peak fractionation
Absorbance
Volume
Delay volume
Description
The delay volume settings are used to make sure that the fractions collected during fractionation, using
the outlet valve or the fraction collector, correspond to the fractions indicated in the chromatogram.
The delay volume is the volume between the UV monitor, and the fraction collector or outlet that is
used, see the illustration below.
Valve
UV/Vis
Delay volume
The time when a peak is detected The peak reaches the fraction
by the UV/Vis detector is called T0 collector at time T1
As the delay volume is affected by the length and diameter of the tubing, it should be set according
to the tubing and modules used.
Volume
When fixed volume fractionation is used the delay volume is collected at the beginning of fractionation.
Fraction collector F9-R saves the delay volume in the first fraction (fraction 1), see the example illustration
above. Delay volumes are not indicated with a numerical fraction mark on the chromatogram.
Fraction collector F9-C does not save the delay volume as a fraction, instead the delay volume is dis-
posed of in the waste funnel.
Note: If two fraction collectors are used, the fractions collected by Fraction collector F9-R, 2nd are
indicated by the prefix 2: (i.e. fractions 2:1, 2:2, 2:3 etc.).
When peak fractionation is used, Fraction collector F9-R collects the delay volumes in the fractions
preceeding each peak. In the chromatogram above, delay volumes are collected in fractions 1, 5 and
8. The numerical fraction marks for the delay volume fractions are not indicated on the chromatogram.
Fraction collector F9-C does not save the delay volumes as fractions, instead the delay volumes are
disposed of in the waste funnel.
Note: If two fraction collectors are used, the fractions collected by the second Fraction collector are
indicated by the prefix 2: (i.e. fractions 2:1, 2:2, 2:3 etc.).
DropSync on: When using DropSync the sensors in the Tube sensor detect when a drop is released.
The Tube rack moves and positions the next tube under the Tube sensor just after a drop is released.
Fractionation with DropSync can be used at flow rates up to 2 ml/min. For water and solutions with
higher surface tension, a higher flow rate can be used. Volatile solutions and solutions with low surface
tension may require a lower flow.
Peak broadening
The width of peaks at the fraction collector is influenced by the properties of the column and the di-
mensions of tubing connecting the components. Initial sample volume affects the peak width in gel
filtration (GF) chromatography. A sample zone is broadened during passage through a GF column so
that the sample is diluted and the resolution decreases with increasing sample volume. Sample volume
does not however affect the resolution in adsorption chromatography techniques such as affinity
chromatography (AC), ion exchange chromatography (IEX), and hydrophobic interaction chromatography
(HIC). The effect of peak broadening in the system from sample injection to peak detection (including
dilution on the column) is apparent in the chromatogram from the UV monitor, but broadening from
the UV monitor to fraction collection is not visible in the chromatogram. This "hidden" effect can
sometimes be dramatic, especially for high-resolution columns.
To minimize peak broadening, use narrow and short tubing connections as far as possible. Remember
that using narrow tubing will increase the back pressure in the system.
In this section
This section contains the following sub-sections:
Introduction
The predefined methods are built up using phases, where each phase corresponds to a step in a
chromatography run with a number of properties associated with that phase.
See UNICORN Method Manual for more information about method structure, definitions and concepts
of methods in UNICORN.
Predefined methods
There are several predefined methods to choose from. All the predefined methods are listed below.
The predefined methods available for each system are defined by the Instrument Configuration. Refer
to Section 9.7 Predefined methods and phases, on page 438 for more information about each method.
Purification methods
• Affinity Chromatography (AC)
• Affinity Chromatography (AC) with Tag Removal
• Anion Exchange Chromatography (AIEX)
• Cation Exchange Chromatography (CIEX)
• Chromatofocusing (CF)
• Desalting
• Gel filtration (GF)
• Hydrophobic Interaction Chromatography (HIC)
• Manual Loop Fill
• NHS-coupling
• Reversed Phase Chromatography (RPC)
Maintenance methods
• Column CIP
• Column Performance Test
• Column Preparation
• System CIP
• System Preparation
or
• Open an existing method that can be edited and saved with a new name or overwritten
2 Build/edit the Method Outline and/or edit the Phase Properties for the appropriate phases
3 Save the method
or
• select File:New Method...
Step Action
See UNICORN Method Manual for more information about methods and method creation in UNICORN.
Note: Sample loop is called capillary loop in UNICORN.
The following modules require that the method is created using the Text Instruction mode:
• Inlet valve IX
Note: When using Inlet valve IX (V9-IX or V9H-IX) connected to another inlet valve in a run, it is
necessary to use the instructions for both Inlet valve IX and the other inlet valve, respectively.
• Versatile valve
• I/O-box
• UV monitor, 2nd
Note: It is possible to reset the UV monitor U9-L, 2nd using Phase Properties. All other UV monitor
U9-L, 2nd instructions need to be edited in the Text Instruction mode.
Note: If installing more than one external air sensor, it is necessary to create the methods for the
additional air sensors using the Text Instruction mode.
Adapt the method will be adapted to the new set of components. All functions and positions
method that are still available will be unchanged. For example, valve positions present on
both the original and the new system will be unchanged in the adapted method.
Non-compatible settings will be adjusted in order to properly adapt the method for
new components. Settings that have been changed are described and saved in
Method Notes, which are displayed after the method has been adapted.
Keep text the text method will be kept and nothing will be changed. Non-compatible settings
method will still be present in the method but they will not be functional. All Phase Properties
of the method will be replaced by tables of Phase Variables.
Introduction
This section describes how to start a run using a previously created method and how to perform a
manual run. For further information about the capabilities of the System Control module, refer to
UNICORN System Control Manual.
1 Open the System Control module and click the icon Open Method Navigator.
3 Step through the displayed pages in the Start Protocol, add requested input and make
appropriate changes if necessary. Click Next.
Introduction
During the run, the System Control module will display the run progress of the method being executed.
This section describes how to interact with the run from the process picture or by executing manual
instructions.
To find an overview of the System Control user interface, see UNICORN System Control Manual.
Note: When ending a method run in advance, it is possible to save the partial result.
More information regarding UNICORN capabilities during the method run is available in the UNICORN
System Control Manual.
Introduction
This section describes how to clean the instrument and columns after a chromatographic run, and
how to prepare the system for storage.
The instrument and the columns should be cleaned between the runs. This will prevent, for example,
sample contamination, protein precipitation and column clogging. If the instrument is not going to be
used for a couple of days or longer, the instrument, columns and the pH flow cell should be filled with
storage solution. For further information about cleaning and maintenance procedures, see Chapter 7
Maintenance, on page 251.
Tip: To clean and fill the instrument and columns with storage solution, use System CIP and Column
CIP either as separate, predefined methods or as phases included in a chromatographic
method.
CAUTION
Hazardous chemicals and biological agents. Before maintenance, service and
decommissioning, wash the ÄKTA pure instrument with a neutral solution to make
sure that any hazardous solvents and biological agents have been flushed out from
the system.
System cleaning
After a method run is completed, perform the following:
• Rinse the instrument with one or several cleaning solution(s) (e.g., NaOH, buffer solution or distilled
water) using System CIP.
Note: If Column valve V9-C or V9H-C is mounted, the integrated pressure sensor of the valve allows
the system to monitor the post-column pressure. The limit for the pressure sensor in Column
valve V9-C or V9H-C is automatically set so that the UV monitor and the pH monitor are
protected from high pressure. If Column valve V9-C or V9H-C is not mounted, make sure
to keep the pressure in the system after the column below the pressure limits for the modules
in the flow path.
• If applicable, empty the fraction collector.
• Clean all spills on the instrument and on the bench using a moist tissue.
• Empty the waste vessel.
• Clean the manual injection port of the injection valve, see Section 7.6.2 Perform System CIP, on
page 278 for detailed instructions.
• If applicable, clean the pH electrode manually and make sure to leave it in an appropriate buffer.
See Section 7.6.7 Clean the pH electrode, on page 293 for detailed instructions.
System storage
If the instrument is not going to be used for a couple of days or longer, also perform the following:
• Fill the system and inlets with storage solution (e.g., 20% ethanol) using System CIP.
Note: If Column valve V9-C or V9H-C is mounted, the integrated pressure sensor of the valve allows
the system to monitor the post-column pressure. The limit for the pressure sensor in Column
valve V9-C or V9H-C is automatically set so that the UV monitor and the pH monitor are
protected from high pressure. If Column valve V9-C or V9H-C is not mounted, make sure
to keep the pressure in the system after the column below the pressure limits for the modules
in the flow path.
• If applicable, prepare the pH electrode for storage as described in Section 7.6.6 Storage of the pH
electrode, on page 291.
Column cleaning
After a method run is completed, perform the following:
• Clean the column with one or several cleaning solution(s) using Column CIP.
Column storage
If the column is not going to be used for a couple of days or longer, also perform the following:
• Fill the column with storage solution (e.g., 20% ethanol) using Column CIP.
pH electrode storage
If pH monitoring will not be used for a week or longer, perform one of the following actions:
• Inject new storage solution into the pH flow cell. Refer to Section 7.6.6 Storage of the pH electrode,
on page 291 for instructions.
• Replace the pH electrode with the dummy electrode that is installed in the pH valve on delivery.
In the following situations, in order to increase the lifetime of the pH electrode, use the By-pass position
and store the electrode in storage solution inside the pH flow cell:
• pH monitoring is not needed during the run.
• Organic solutions are used.
• Extremely acidic or extremely basic solutions are used.
For further information on how to prepare the pH electrode for storage, refer to Section 7.6.6 Storage
of the pH electrode, on page 291.
Result: All open UNICORN modules close and the Log On dialog box
opens.
Note: If an edited method or result is open and not saved when you try to exit or log off UNICORN,
you will see a warning. Click Yes to save, No to exit without saving, or Cancel to stay logged
on.
6 Performance tests
In this chapter
This chapter contains the following sections:
1 In the System Control module, on the System menu, click Performance Test and Report.
Result: The System Performance Test and Report dialog opens.
2 In the System Performance Test and Report dialog, click one of the following tests:
Air sensor A and inlet valve V9-IA test Air sensor A and inlet valve V9H-IA test
Air sensor B and inlet valve V9-IB test Air sensor B and inlet valve V9H-IB test
Air sensor S and Sample inlet valve Air sensor S and Sample inlet valve
V9-IS test V9H-IS test
4 Click Next in the Start Protocol dialog to open the next page. The pages are described
in the table Overview of the Start Protocol dialog, on page 226.
Page Description
Notes Displays the Method Notes of the method. The Method Notes con-
tains a method description and instructions on how to run the
method. This dialog box also allows the user to enter Start Notes.
Evaluation Procedures Allows the user to select to save the report to file (recommended)
and/or to print the report.
Result Name and Location Allows the user to change result name and result location.
Automatic evaluation
The system automatically generates a report when the test is finished. The report can be printed in
two ways:
• It is recommended to select Save the report to file in the Evaluation Procedures page of the Start
Protocol dialog box when starting the test. The report is saved in the folder Temp in your UNICORN
installation folder. For example C:\Program Files\GE Healthcare\UNICORN\UNICORN 7.0\Temp.
• If the option Print report was selected in the Evaluation Procedures page of the Start Protocol
dialog box when starting the test, the report is also automatically printed on the system printer.
Refer to UNICORN Administration and Technical Manual for information on how to install a printer.
Print the report and check the status of the tests. For each of the tests the report states "The test
passed" or "The test failed".
Note: The fraction collector test is evaluated manually and no report is generated.
Method description
The Air sensor A and inlet valve A tests switches inlet valve A, labeled V9-IA or V9H-IA, to an empty
position and checks if the air sensor detects air.
The method run takes approximately 1 minute.
Required configuration
A correctly installed Inlet valve A is required to run the test.
Required material
The following materials are required:
• Syringe, 25-30 ml
• Distilled water
3 Prime the A1 inlet and purge the pump. See Section 5.4.1 System pumps, on page 161.
4 Disconnect the inlet tubing connected to the A3 inlet valve position. During the test
method air is introduced into the inlet valve through that inlet port to test the function
of the air sensor.
Cause Action
Faulty air sensor For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Incorrect preparation of Make sure that the tubing was correctly prepared, see Prepare the
tubing test, on page 228.
Cause Action
The valve did not switch For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Method description
The Air sensor B and inlet valve B test switches the inlet valve B, labeled V9-IB or V9H-IB, to an empty
position and checks if the air sensor detects air.
The method run takes approximately 1 minute.
Required configuration
A correctly installed Inlet valve B is required to run the test.
Required material
The following materials are required:
• Syringe, 25-30 ml
• Distilled water
3 Prime the B1 inlet and purge the pump. See Section 5.4.1 System pumps, on page 161.
4 Disconnect the inlet tubing connected to inlet valve position B3. During the test method
air is introduced into the inlet valve through that inlet port to test the function of the air
sensor.
Cause Action
Faulty air sensor For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Incorrect preparation of Make sure that the tubing was correctly prepared, see Prepare the
tubing test, on page 230.
Cause Action
The valve did not switch For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Method description
The Air sensor S and Sample inlet valve test checks the functionality of the Sample inlet valve, labeled
V9-IS or V9H-IS, and the integrated air sensor in ÄKTA pure.
The method run takes approximately 1 minute.
Required configuration
Core components of the ÄKTA pure and correctly installed Sample Inlet valve and Sample pump are
required to run the test.
Required material
The following materials are required:
• Syringe, 25-30 ml
• Distilled water
1 Open System Control and update the system setting Tubing: Sample pump to injection
valve (found in System Settings:Tubing and Delay volumes).
Note:
The system setting Tubing: Sample pump to injection valve is available only if Column
valve C is not included in the system.
4 Prime inlet tubing A1 and sample inlet Buffer and purge the sample pump. See Section 5.4
Prime inlets and purge pump heads, on page160. No air shall be left in the pumps or tubings.
Step Action
5 Disconnect the inlet tubing connected to the sample inlet valve position S3. During the
test method air is introduced into the sample inlet valve through that inlet port to test
the function of the air sensor.
Cause Action
Faulty air sensor For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Incorrect preparation of Make sure that the tubing was correctly prepared, see Prepare the
tubing test, on page 230.
Cause Action
The valve did not switch For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
Method description
The Column valve C test checks the functionality of Column valve C, labeled V9-C or V9H-C, and of the
integrated pressure sensors that measure pre-column and post-column pressure. If two valves are
used, the test will check the pre-column pressure sensor in the first valve and the post-column pressure
sensor in the second valve.
The method run takes approximately 1 minute.
Required configuration
A correctly installed Column valve C is required to run the test.
Required material
The following materials are required:
• Syringe, 25-30 ml
• Distilled water
• Ref 1 tubing, see Reference capillary, on page 426.
1 Connect the Ref 1 tubing between Column valve C ports 1A and 1B.
5 Prime the inlet A1 and purge the pump. See Section 5.4.1 System pumps, on page 161.
Cause Action
Faulty Column valve For further information, see Section 8.4 Troubleshooting: Valves, on
page 375.
The integrated pressure Calibrate the pressure monitors, see Section 7.7.2 Calibrate the
monitors are not calibrated pressure monitors, on page 305.
Incorrect preparation of the Make sure that the reference tubing is connected between column
system ports 1A and 1B and that the flow restrictor is in-line.
Required configuration
The core components of ÄKTA pure, an outlet valve and a correctly installed Fraction collector F9-C
are required to run the test.
Required material
The following materials are required:
• Distilled water
• Syringe, 25-30 ml
• Two deep well cassettes
• Two 96 deep well plates for collecting the fractions
3 Prime inlet tubing A1 and purge System pump A. See Section 5.4.1 System pumps, on
page 161. No air shall be left in the pump or tubing.
4 Place two 96 deep well plates in positions 1 and 6 in the fraction collector. Make sure
that no other cassettes are present in the fraction collector during the run.
Step Action
Cause Action
Incorrect volumes collected Air in pumps: Make sure to prime inlet tubing A1 and purge System
in the tubes, and distur- pump A before method start, see Section 5.4.1 System pumps, on
bances of system pressure page 161.
curves: Faulty pump: See Section 8.6 Troubleshooting: Pumps, on page 387.
• Air trapped in System
pump A
• Faulty System pump A
Incorrect preparation of Make sure that the system was correctly prepared, see Prepare the
buffer and tubing test, on page 236.
Required configuration
A correctly installed Outlet valve, and a correctly installed Fraction collector F9-R are required to run
the test.
Required material
The following materials are required:
• Distilled water
• Syringe, 25-30 ml
• 7 tubes for collecting the fractions.
3 Prime inlet A1 and purge System pump A. See Section 5.4.1 System pumps, on page 161.
Cause Action
Incorrect volumes collected Air in pumps: Make sure to prime inlet tubing A1 and purge System
in the tubes, and distur- pump A before method start, see Section 5.4.1 System pumps, on
bances of system pressure page 161.
curves: Faulty pump: See Section 8.6 Troubleshooting: Pumps, on page 387.
• Air trapped in System
pump A
• Faulty System pump A
Liquid collected in wrong Make sure the fraction collector delivery arm is positioned above
tubes: tube number 1 before starting the test.
Incorrect preparation of Make sure that the system was correctly prepared, see Prepare the
buffer and tubing test, on page 239.
Method description
The Sample Pump test checks the functionality of the sample pump, labeled S9 or S9H.
The method run takes approximately 12 minutes.
Required configuration
The core components of ÄKTA pure and a correctly installed Sample pump are required to run the test.
Required material
The following materials are required:
• Distilled water
• 1% Acetone and 1.00 M NaCl in distilled water
• Syringe, 25-30 ml
• Ref 1 tubing, see Reference capillary, on page 426.
1 Connect a tubing between Injection valve port column and the UV inlet, preferably the
Ref 1 tubing included with the system at delivery.
4 Immerse sample inlet tubing Buffer in 1% Acetone and 1.00 M NaCl in distilled water.
5 Prime inlet tubing A1 and sample inlet Buffer and purge the pumps. See Section 5.4.2
Sample pump, on page 167 and. No air shall be left in the pumps or tubings.
Cause Action
The two injected peaks of liq- Air in pump: Make sure to prime inlet tubing Buffer and purge sample
uid are not delivered as pro- pump before method start, see Section 5.4.2 Sample pump, on
portional peaks: page 167.
• Air trapped in Sample Faulty pump: See Section 8.6 Troubleshooting: Pumps, on page 387.
pump
• Faulty Sample pump
The delivery of liquid is pulsat- Air in pump: Make sure to prime inlet tubing Buffer and purge sample
ing (unstable pump pressure) pump before method start, see Section 5.4.2 Sample pump, on
page 167.
Faulty pump: See Section 8.6 Troubleshooting: Pumps, on page 387.
The pump pressure is too Folded, twisted or blocked tubing: Check the tubing.
high (at a flow rate of 25 The Sample pump pressure monitor is not calibrated: Calibrate the
ml/min or 150 ml/min for pressure monitor, see Section 7.7.2 Calibrate the pressure monitors,
ÄKTA pure 25 and ÄKTA pure on page 305.
150, respectively)
Faulty pump: See Section 8.6 Troubleshooting: Pumps, on page 387.
Incorrect preparation of Make sure that the system was correctly prepared, see Prepare the
buffer and tubing test, on page 241.
Required configuration
A correctly installed UV monitor U9-L and a correctly installed Conductivity monitor are required to
run the test.
Required material
The following materials are required:
• Distilled water
• 1% acetone and 1.00 M NaCl in distilled water
• Ref 1 tubing, see Reference capillary, on page 426.
• Syringe, 25-30 ml
• Mixer, 1.4 ml
3 Place inlet tubing B1 into a solution of 1% acetone and 1.00 M NaCl in distilled water.
4 Connect the Ref 1 tubing between Injection valve port Col and the UV monitor U9-L inlet,
to generate a back pressure.
5 Make sure that the Mixer with a chamber volume of 1.4 ml is installed.
For further information, see Section 7.8.3 Replace the Mixer, on page 322.
6 Prime the buffer inlets and purge System pump A and System pump B. See Section 5.4.1
System pumps, on page 161.
Note: If the test is performed at cold room temperature, select Conductivity:Cond temp compensa-
tion in the System Settings dialog, and set the Compensation factor to 2.1.
Illustration of chromatogram
The illustration below shows a chromatogram from a System test with UV monitor U9-L.
Disturbances caused by air Make sure to prime the buffer inlets and to purge the System pumps
trapped in any of the pumps before method start. See Section 5.4.1 System pumps, on page 161.
Disturbances caused by Replace piston seals. See Chapter 7 Maintenance, on page 251.
damaged pump piston seals
Unstable or incorrect UV sig- See Section 8.3 Troubleshooting: Monitors, on page 360.
nal, or drifting base line:
• Faulty UV monitor
Wrong Mixer chamber size Replace the Mixer. See Chapter 7 Maintenance, on page 251.
or faulty Mixer
If all values are faulty: Air in pumps: Make sure to prime the buffer inlets and to purge the
System pumps before method start. See Section 5.4.1 System pumps,
• Air in the pump or a faulty
on page 161.
pump
Faulty pump: See Chapter 7 Maintenance, on page 251.
Faulty values at 5%: Replace piston seals. See Chapter 7 Maintenance, on page 251.
• Damaged pump piston
seal in System pump B
Faulty values at 95%: Replace piston seals. See Chapter 7 Maintenance, on page 251.
• Damaged pump piston
seal in System pump A
Incorrectly prepared acetone Make sure that the acetone solution is 1% and that no solution has
solution evaporated.
Wrong UV cell path length See, Calibration of the UV monitor U9-L flow cell length, on page 312.
set in UNICORN
Air trapped in the pumps Make sure to prime and purge the system pumps before starting
the test, see Section 5.4.1 System pumps, on page 161.
The value set for the Cond If the test is performed at cold room temperature, open the System
temp compensation factor is Settings dialog, select Conductivity:Cond temp compensation and
not optimal set the Compensation factor to 2.1
Air or dirt in the UV flow cell Flush or clean the UV cell, see Section 7.5.1 Clean the UV flow cell,
on page 273.
The inline filter is dirty Replace the inline filter, see Section 7.3.2 Replace the inline filter, on
page 266.
The System pressure monitor Calibrate the pressure monitor, see Section 7.7.2 Calibrate the pres-
is not calibrated sure monitors, on page 305.
Required configuration
A correctly installed UV monitor U9-M and a correctly installed Conductivity monitor are required to
run the test.
Required material
The following materials are required:
• Distilled water
• 1% acetone and 1.00 M NaCl in distilled water
• Ref 1 tubing, see Reference capillary, on page 426.
• Syringe, 25-30 ml
• Mixer, 1.4 ml
3 Place inlet tubing B1 into a solution of 1% acetone and 1.00 M NaCl in distilled water.
4 Connect the Ref 1 tubing between Injection valve port Col and the UV monitor U9-M inlet,
to generate a back pressure.
5 Make sure that the Mixer with a chamber volume of 1.4 ml is installed.
For further information, see Section 7.8.3 Replace the Mixer, on page 322.
6 Prime the buffer inlets and purge System pump A and System pump B. See Section 5.4.1
System pumps, on page 161.
Note: If the test is performed at cold room temperature, select Conductivity:Cond temp compensa-
tion in the System Settings dialog, and set the Compensation factor to 2.1.
Illustration of chromatogram
The illustration below shows a chromatogram from a System test with UV monitor U9-M.
Disturbances caused by air Make sure to prime the buffer inlets and to purge the System pumps
trapped in any of the pumps before method start. See Section 5.4.1 System pumps, on page 161.
Disturbances caused by Replace piston seals. See Chapter 7 Maintenance, on page 251.
damaged pump piston seals
Unstable or incorrect UV sig- See Section 8.3 Troubleshooting: Monitors, on page 360.
nal, or drifting base line:
• Faulty UV monitor
Wrong Mixer chamber size Replace the Mixer. See Chapter 7 Maintenance, on page 251.
or faulty Mixer
If all values are faulty: Air in pumps: Make sure to prime the buffer inlets and to purge the
System pumps before method start. See Section 5.4.1 System pumps,
• Air in the pump or a faulty
on page 161.
pump
Faulty pump: See Chapter 7 Maintenance, on page 251
Faulty values at 5%: Replace piston seals. See Chapter 7 Maintenance, on page 251.
• Damaged pump piston
seal in System pump B
Faulty values at 95%: Replace piston seals. See Chapter 7 Maintenance, on page 251.
• Damaged pump piston
seal in System pump A
Incorrectly prepared acetone Make sure that the acetone solution is 1% and that no solution has
solution evaporated.
Air trapped in the pumps Make sure to prime and purge the system pumps before starting
the test, see Section 5.4.1 System pumps, on page 161.
Incorrectly prepared NaCl Make sure that the NaCl solution is 1.00 M.
solution
The value set for the Cond If the test is performed at cold room temperature, open the System
temp compensation factor is Settings dialog, select Conductivity:Cond temp compensation and
not optimal set the Compensation factor to 2.1
Air or dirt in the UV flow cell Flush or clean the UV cell, see Section 7.5.1 Clean the UV flow cell,
on page 273.
Dirt in inline filter Replace the inline filter, see Section 7.3.2 Replace the inline filter, on
page 266.
System pressure monitor not Calibrate the pressure monitor, see Section 7.7.2 Calibrate the pres-
calibrated sure monitors, on page 305.
7 Maintenance
In this chapter
This chapter contains the following sections:
Note: Modules with no plus symbol (+) have no related maintenance notifications.
1 In the left pane of the Maintenance Manager dialog box, select a maintenance notification.
Result: Details of the selected maintenance notification are displayed in the dialog box.
Step Action
3 Click Apply to save the changes and apply the notification settings.
Click... to...
Note: The predefined maintenance notifcation periods use average values. The actual service interval
for a specific module can be shorter or longer.
WARNING
Hazardous biological agents during run. When using hazardous biological agents,
run System CIP and Column CIP to flush the entire system tubing with bacteriostatic
solution (e.g., NaOH) followed by a neutral buffer and finally distilled water, before
service and maintenance.
CAUTION
Hazardous chemicals during run. When using hazardous chemicals, run System
CIP and Column CIP to flush the entire system tubing with distilled water, before
service and maintenance.
CAUTION
Always use appropriate personal protective equipment when decommissioning the
equipment.
Weekly Change pump rinsing solution Section 7.3.1 Change pump rinsing so-
lution, on page 262
Weekly Clean fraction collector diodes Section 7.3.3 Clean the Fraction collec-
tor F9-C sensors, on page 267
Semiannually Clean the UV flow cell Section 7.5.1 Clean the UV flow cell, on
page 273
Clean the instrument externally Section 7.6.1 Clean the instrument externally, on page 277
Perform System CIP Section 7.6.2 Perform System CIP, on page 278
Perform Column CIP Section 7.6.3 Perform Column CIP, on page 284
Clean Fraction collector F9-C Section 7.6.4 Clean Fraction collector F9-C, on page 288
Clean Fraction collector F9-R Section 7.6.5 Clean Fraction collector F9-R, on page 290
Replace tubing and connectors Section 7.8.1 Replace tubing and connectors, on page 317
Clean the pH electrode Section 7.6.7 Clean the pH electrode, on page 293
Clean the Conductivity flow cell Section 7.6.10 Clean the Conductivity flow cell, on page300
Calibrate the Conductivity monitor Section 7.7.3 Calibrate the Conductivity monitor, on
page 308
Calibrate the UV monitor Section 7.7.4 Calibrate the UV monitors, on page 312
Replace inline filter Section 7.3.2 Replace the inline filter, on page 266
Replace O-ring in Mixer Section 7.8.4 Replace the O-ring inside the Mixer, on
page 323
Replace UV flow cell Section 7.8.5 Replace the UV monitor U9-M flow cell, on
page 325 and Section 7.8.6 Replace the UV monitor U9-L
flow cell, on page 328
Replace the Flow restrictor Section 7.8.7 Replace Flow restrictor, on page 330
Replace inlet filters Section 7.8.8 Replace the inlet filters, on page 331
Wipe off excess oil from the pump head Section 7.6.9 Wipe off excess oil from the pump head, on
page 299
Clean the check valves Section 7.6.8 Clean the pump head check valves, on
page 296
Replace check valves Section 7.8.9 Replace the pump head check valves, on
page 332
Replace pump piston seals, pump P9 and Section 7.8.10 Replace pump piston seals of Pump P9 or
P9H P9H, on page 335
Replace pump piston seals, pump P9-S Section 7.8.11 Replace pump piston seals of Pump P9-S,
on page 344
Replace pump pistons Section 7.8.12 Replace pump pistons, on page 351
Replace pump rinsing system tubing Section 7.8.13 Replace pump rinsing system tubing, on
page 352
In this section
This section contains the following subsections:
Maintenance interval
Replace the pump rinsing solution in the system pumps and the sample pump (optional module) every
week to prevent bacterial growth.
Required material
The following material are required:
• 20% ethanol
• Syringe, 25-30 ml
5 4
Part Description
4 Outlet tubing
5 Inlet tubing
4 Insert the inlet tubing to the System pump piston rinsing system into the fluid in the
rinsing solution tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing solution tube.
5 Connect a 25 to 30 ml syringe to the outlet tubing of the System pump piston rinsing
system. Draw liquid slowly into the syringe.
7 Insert the outlet tubing into the fluid in the rinsing solution tube.
8 Fill the rinsing solution tube so that the tube contains 50 ml of 20% ethanol.
4 5
Part Function
3 Inlet tubing, from the rinsing liquid tube to the left pump head (lower rinsing system
connection) .
4 Tubing between the left pump head (upper rinsing system connection) and the right
pump head (lower rinsing system connection).
5 Outlet tubing, from the right pump head (upper rinsing system connection) to the rinsing
liquid tube.
1 Unscrew and remove the rinsing liquid tube from the holder.
3 Put the rinsing liquid tube back in the holder and fasten it by screwing it into place.
4 Insert the inlet tubing to the piston rinsing system into the fluid in the rinsing liquid tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing liquid tube.
5 Connect a 25 to 30 ml syringe to the outlet tubing of the System pump piston rinsing
system. Draw liquid slowly into the syringe.
Note:
The outlet tubing is colored orange for clarity in the above illustration.
7 Insert the outlet tubing into the fluid in the rinsing liquid tube.
8 Fill the rinsing liquid tube so that the tube contains 50 ml of 20% ethanol.
Maintenance interval
Replace the inline filter that is located in the top section of the Mixer every week, or when required, for
example when the filter becomes clogged.
Required material
The following materials are required:
• Online filter kit
• Forceps
• Gloves
Instruction
Follow the instructions below to replace the inline filter that is located in the top of the Mixer.
Tip: Use forceps and gloves during the replacement procedure to avoid contaminating the Mixer
components.
Step Action
2 Remove the old filter using forceps. Replace the support net if this is damaged. Fit the
new filter.
3 Check the O-ring of the Mixer. If the O-ring is damaged, replace it according to Section 7.8.4
Replace the O-ring inside the Mixer, on page 323.
4 While holding the Mixer upright, screw the top section back onto the Mixer.
Maintenance interval
Clean the fraction collector sensors every week, or when required, for example if the fraction collector
fails to read the tray ID or do not collect fractions correctly.
Required material
The following material is required:
• Wash bottle
• Water or 20% ethanol
• Cloth
Instructions
Follow the instructions below to clean the fraction collector diodes. See Section 3.2.2 Fraction collector
F9-C illustrations, on page 92 for the locations of the components of the fraction collector.
Step Action
Result: The Dispenser head moves to cleaning position, and the Instrument display states
System pause.
Step Action
2 Wipe off the Dispenser head and the Drop sync and Type code reader sensor windows
using a wash bottle with water or 20% ethanol and a cloth.
5 In the System Control module, press the End icon in the toolbar.
Result: The Dispenser head moves to home position, and the Instrument display states
Ready.
Maintenance interval
Check the back pressure for the Flow restrictor every month.
1 Immerse the piece of inlet tubing marked A1 in distilled water, and insert the piece of
tubing from Outlet valve port W into a waste container.
Step Action
10 Check that the pressure difference is within the range 0.2 ± 0.05 MPa.
If this is not the case, the Flow restrictor should be replaced, see Section 7.8.7 Replace
Flow restrictor, on page 330.
1 Immerse the piece of inlet tubing marked A1 in distilled water, and insert the piece of
tubing from Outlet valve port W into a waste container.
Step Action
7 Calculate the difference between the two pressure values noted in step 4 and step 6.
8 Check that the pressure difference is within the range 0.2 ± 0.05 MPa.
If this is not the case, the Flow restrictor should be replaced, see Section 7.8.7 Replace
Flow restrictor, on page 330.
In this section
This section contains the following subsections:
Maintenance interval
Clean the UV flow cell every six months, or when required.
NOTICE
Keep UV flow cell clean. Do not allow solutions containing dissolved salts, proteins
or other solid solutes to dry out in the flow cell. Do not allow particles to enter the
flow cell, as damage to the flow cell may occur.
Required material
The following materials are required:
• Luer connector
• Waste container
• Syringe, 25-30 ml
• 10% surfactant detergent solution (e.g., Decon™ 90, Deconex 11, or RBS 25)
• Distilled water
Instruction
Follow the instructions below to clean the UV flow cell for UV monitor U9-M or UV monitor U9-L. The
UV flow cell can be mounted or not mounted on the instrument during the cleaning procedure.
The illustrations in the instructions below show UV monitor U9-M. UV monitor U9-L is cleaned in a
corresponding way.
Step Action
1 Disconnect the tubing from the top of the UV flow cell, and replace the fingertight con-
nector with a Luer connector.
2 Disconnect the tubing from the bottom of the UV flow cell, and connect a piece of waste
tubing to the UV flow cell. Insert the waste tubing into a waste container.
Step Action
3 Fill a syringe with distilled water, and connect the syringe to the Luer connector.
4 Squirt the distilled water through the UV flow cell in small amounts. Disconnect the syringe.
5 Fill a syringe with a 10% surfactant detergent solution, such as Decon 90, Deconex 11,
RBS 25 or equivalent, and connect the syringe to the Luer connector.
Tip:
Heat the 10% surfactant detergent solution to 40°C to increase the cleaning effect.
6 Squirt the detergent solution through the UV flow cell about five times.
7 Leave the detergent solution in the flow cell for at least 20 minutes.
8 Inject the detergent solution remaining in the syringe into the flow cell. Disconnect the
syringe.
9 Fill a syringe with distilled water. Connect the syringe to the Luer connector.
10 Inject the distilled water into the UV flow cell to rinse the flow cell. Disconnect the syringe.
11 Disconnect the Luer connector from the top of the UV flow cell. Reconnect the piece of
tubing from the Column valve to the top of the UV flow cell.
12 Disconnect the waste tubing from the bottom of the UV flow cell. Reconnect the piece
of tubing from the Conductivity monitor to the bottom of the UV flow cell.
Maintenance interval
Replace the pH electrode every six months, or when required.
Required material
The following materials are required:
• pH electrode
• Deionized water
• Standard buffer pH 4
Instruction
CAUTION
pH-electrode. Handle the pH-electrode with care. The glass tip may break and cause
injury.
1 Disconnect the pH electrode cable of the used pH electrode from the connection on the
front of the pH valve.
2 Unscrew the nut of the pH electrode by hand, and pull the used electrode away.
3 Unpack the new pH electrode. Remove the cover from the tip of the new pH electrode.
Make sure that the electrode is not broken or dry.
4 Prior to first use of the electrode, immerse the glass bulb in deionized water for 30 minutes
and then in a standard buffer, pH 4, for 30 minutes.
5 Carefully insert the new pH electrode into the pH flow cell. Tighten the nut by hand to
secure the electrode.
6 Connect the pH electrode cable of the new electrode to the connection on the front of
the pH valve.
7 Calibrate the new pH electrode, see Section 7.7.1 Calibrate the pH monitor, on page 303.
In this section
This section contains the following subsections:
7.6.9 Wipe off excess oil from the pump head 299
Maintenance interval
Clean the the instrument externally when required. Do not allow spilled liquid to dry on the instrument.
Required material
The following materials are required:
• Cloth
• Mild cleaning agent or 20% ethanol
Instruction
Follow the instructions below to clean the instrument externally.
Step Action
3 Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning agent or 20%
ethanol. Wipe off any excess.
Maintenance interval
Perform a System cleaning in place (System CIP) when required, for example between runs where dif-
ferent samples and buffers are used. This is important to prevent cross-contamination and bacterial
growth in the instrument.
Required material
The following materials are required:
• Appropriate cleaning solutions (e.g., NaOH, buffer solution or distilled water).
• Syringe, 25-30 ml
Introduction
The System CIP method is used to fill the instrument and the selected inlets and outlets with cleaning
solution.
WARNING
Hazardous biological agents during run. When using hazardous biological agents,
run System CIP and Column CIP to flush the entire system tubing with bacteriostatic
solution (e.g., NaOH) followed by a neutral buffer and finally distilled water, before
service and maintenance.
CAUTION
Hazardous substances. When using hazardous chemical and biological agents,
take all suitable protective measures, such as wearing protective glasses and gloves
resistant to the substances used. Follow local and/or national regulations for safe
operation, maintenance and decommissioning of the equipment.
CAUTION
Explosion hazard if flammable liquid leaks during cleaning of the flow path. When
cleaning the flow path of Fraction collector F9-C with a flammable liquid like ethanol,
carefully inspect the flow path, including the waste tubing, to make sure there will
be no leakage.
Tip: If hazardous chemicals are used for system or column cleaning, wash the system or columns
with a neutral solution in the last phase or step.
or
• select File:New Method.
Result: The New Method dialog opens.
2 In the New Method dialog, select System and then System CIP in the Predefined Method
drop-down list. Click OK.
Result: One Method Settings phase and three System CIP phases show in the Method
Outline pane. Each System CIP phase uses one cleaning solution.
3 If desired, add additional System CIP phases to the method using the Phase Library.
Step Action
Note:
• The pH electrode and the pH valve are not included in the system CIP. Refer to Sec-
tion 7.6.7 Clean the pH electrode, on page 293 for instructions on how to clean the pH
electrode.
• If the fraction collector is included, a lower flow rate might need to be set in the instruc-
tion Fraction collector wash settings.
• Make sure not to exceed the pressure limits for any of the modules that are part of the
flow path during the CIP.
Step Action
or
• select File:Save As
Result: The Save As dialog opens.
1 In the Method Editor module, create a System CIP method according to the instructions
above.
2 Connect bypass tubing to all selected column positions and loop positions if a loop valve
is used.
3 Prepare cleaning solutions and immerse the selected inlet tubing in the solutions.
Note:
Note that each phase uses one solution. All inlets selected in one phase should be immersed
in the same cleaning solution.
4 In the System Control module, select the created method and start the run.
5 For complete cleaning of the flow path, clean the manual injection port of the Injection
valve and the pH valve manually, see the instructions below.
2 In the Manual instructions dialog, select Flowpath:Injection valve, and select Manual
Load. Click Execute.
3 Connect a suitable sample loop to Injection valve ports LoopF (fill) and LoopE (empty).
Note:
Do not use a Superloop when cleaning the Injection valve.
4 Connect tubing to Injection valve port W1, and direct this tubing to a waste container.
6 Fill a syringe with distilled water. Connect the syringe to Injection valve port Syr, and inject
the distilled water.
1 Connect tubing to pH valve port W3, and direct the other end of this tubing to a waste
container.
2 Unscrew the pH electrode from the pH valve, and replace it with the dummy electrode.
4 In the Calibration dialog, select pH from the Monitor to calibrate drop-down list.
7 Fill a syringe with approximately 10 ml of 1 M NaOH. Connect the syringe to the pH valve
port Cal, and inject the solution.
8 Fill a syringe with distilled water. Connect the syringe to the pH valve port Cal, and inject
the distilled water.
Result: The pH valve switches back to the default position and the Calibration dialog
closes. No calibration is performed.
Maintenance interval
Perform a Column cleaning in place (Column CIP) when required, for example between runs where
different samples are used.
Required material
The following materials are required:
• Appropriate cleaning solutions. Please refer to the instructions for use of the column.
Introduction
The Column CIP method is used to clean the column after purification runs, to remove non-specifically
bound proteins and to minimize the risk for carry-over between different purification runs.
WARNING
Hazardous biological agents during run. When using hazardous biological agents,
run System CIP and Column CIP to flush the entire system tubing with bacteriostatic
solution (e.g., NaOH) followed by a neutral buffer and finally distilled water, before
service and maintenance.
CAUTION
Hazardous substances. When using hazardous chemical and biological agents,
take all suitable protective measures, such as wearing protective glasses and gloves
resistant to the substances used. Follow local and/or national regulations for safe
operation, maintenance and decommissioning of the equipment.
Tip: If hazardous chemicals are used for system or column cleaning, wash the system or columns
with a neutral solution in the last phase or step.
or
• select File:New Method.
Result: The New Method dialog opens.
2 In the New Method dialog, select System and then Predefined Method and Column CIP.
Click OK.
Result: One Method Settings phase and one Column CIP phase will be displayed in the
Method Outline pane.
3 In the Phase Properties tab of the Method Settings phase, select Column type and Col-
umn position.
Step Action
5 In the Phase Properties tab of the Column CIP phase, click the Get Suggested Steps
button to get a suggested procedure for the selected column type. Note that this function
is not available for all column types.
Result: Suggested cleaning steps for the selected column type are displayed.
6 If several cleaning solutions are used, change settings for Inlet A and/or Inlet B . Select
one inlet for each solution. If Inlet B is used, remember to edit the values in the %B column.
Step Action
or
• select File:Save As
Result: The Save As dialog opens.
1 In the Method Editor module, create a Column CIP method according to the instruction
above.
2 Prepare cleaning solutions and immerse the selected inlets in the solutions.
4 In the System Control module, select the created method and start the run.
Maintenance interval
Clean the Fraction collector when required, for example if liquid has been spilled in the Fraction collector
chamber. The internal tubing of the fraction collector may need to be replaced for maintenance or for
process purposes. Information on how and when to replace the internal tubing can be found in the
ÄKTA pure User manual.
Required material
The following materials are required:
• Wash bottle
• Water or 20% ethanol
• Cloth
Instruction
Follow the instruction below to clean the interior of the Fraction collector. The locations of the compo-
nents of the fraction collector are shown in Section 3.2.2 Fraction collector F9-C illustrations, on page 92.
Step Action
NOTICE
• If no column valve is used, make sure to replace any columns in
the flow path with tubing before a fraction collector wash is per-
formed.
• Lower the flow for the fraction collector wash in the instruction
Fraction collector wash settings if the system back pressure is el-
evated during the wash.
• Make sure not to exceed the pressure limits for any of the modules
that are part of the flow path.
3 Open the door of the fraction collector and remove the rack
Step Action
4 Wash the Cassette tray or rack and the Cassettes (if applicable), with water and a mild
cleaning agent.
5 Lift off the Waste funnel and wash it with water and a mild cleaning agent.
Refit the Waste funnel.
6 Wipe off the interior of the fraction collector using a damp cloth. Wipe off stains using a
mild cleaning agent or 20% ethanol.
7 Wipe off the Dispenser head and its diode windows (the Drop sync sensor and the Type
code reader) using a wash bottle with water or 20% ethanol and a cloth.
10 In the System Control module, press the End icon in the toolbar.
Result: The Dispenser head moves to home position.
Maintenance interval
Clean the Fraction collector when required, for example in case of liquid spill.
Required material
The following materials are required:
• Water or 20% ethanol
• Cloth
3 Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning agent or 20%
ethanol. Wipe off any excess.
Maintenance interval
When pH monitoring is not used, the pH electrode can be stored in storage solution inside the pH flow
cell. If pH monitoring is not used for a week or longer, inject new storage solution into the pH flow cell
or replace the pH electrode with the dummy electrode that was installed in the pH valve on delivery.
Required material
The following materials are required:
• Syringe, 25 to 30 ml
• Storage solution (1:1 mixture of standard buffer, pH 4, and 1 M KNO3)
Instruction
Follow the instructions below to fill the pH flow cell with storage solution. The calibration function is
used to switch the position of the pH valve. However, no calibration is performed.
Step Action
Step Action
5 Fill a syringe with approximately 10 ml of the storage solution. Connect the syringe to
the pH valve port Cal, and inject the storage solution.
6 Click Close.
Result: The pH valve switches back to the default position and the Calibration dialog
closes. No calibration is performed.
Maintenance interval
Clean the pH electrode when required. The pH electrode can be cleaned either when it is installed in
the pH valve or when it has been removed. The pH electrode has a limited longevity and should be
replaced every six months or when the response time is slow, see Section 7.5.2 Replace the pH electrode,
on page 275. After cleaning has been performed, re-calibrate the pH monitor, see Section 7.7.1 Calibrate
the pH monitor, on page 303.
Required material
The following materials are required:
• Syringe, 25-30 ml
• Distilled water
• 0.1 M HCl and 0.1 M NaOH
or
• Liquid detergent
or
• 1% pepsin solution in 0.1 M HCl
or
• 1 M KNO3
Cleaning agents
Clean the pH electrode using one of the following procedures:
Salt deposits
Dissolve the deposits by immersing the electrode for a five minute period in each of the following solu-
tions:
• 0.1 M HCl
• 0.1 M NaOH
• 0.1 M HCl
Rinse the electrode tip in distilled water between each solution.
Oil or grease films
Wash the electrode tip in liquid detergent and water. If the films are known to be soluble in a particular
organic solvent, wash with this solvent. Rinse the electrode tip in distilled water.
Protein deposits
Dissolve the deposit by immersing the electrode in a solution of 1% pepsin in 0.1 M HCl for five minutes,
followed by thorough rinsing with distilled water.
If these procedures fail to rejuvenate the electrode, try the following procedure.
Note: This procedure can be performed only when the pH electrode is not installed in the pH valve.
Step Action
3 Allow the electrode to cool while immersed in the KNO3 solution before re-testing.
If these steps fail to improve the electrode, replace the electrode, see Section 7.5.2 Replace the pH
electrode, on page 275.
Step Action
4 Fill a syringe with approximately 10 ml of chosen cleaning solution. Connect the syringe
to the pH valve port Cal. Inject the liquid and wait for 5 minutes. Disconnect the syringe.
5 If several cleaning solutions are to be used, repeat step 4 with distilled water and then
with the next solution.
Result: The pH valve switches back to the default position and the Calibration dialog
closes. No calibration is performed.
Maintenance interval
Clean the check valves when required, for example if solids in the check valve cause irregular or low
flow. The cleaning procedure is the same for the system pumps and the sample pump.
Required material
The following materials are required:
• Adjustable wrench
• Methanol
• Distilled water
• Ultrasonic bath
Instruction
Follow the instructions below to remove and clean the pump head check valves.
Step Action
2 Disconnect the tubing from the pump head and disconnect the pump inlet tubing. Dis-
connect the tubing of the pump rinsing system.
3 Unscrew the purge valve by turning it counter-clockwise, and lift off the metal ring.
Step Action
4 Unscrew the plastic nut of the upper check valve using an adjustable wrench, and gently
lift off the upper check valve.
5 Unscrew the two white plastic screws located below each pump head. Pull the plastic
connectors to the sides to release the inlet manifold.
Step Action
7
CAUTION
Hazardous substances. When using hazardous chemical and biolog-
ical agents, take all suitable protective measures, such as wearing
protective glasses and gloves resistant to the substances used. Follow
local and/or national regulations for safe operation, maintenance and
decommissioning of the equipment.
Immerse the valves completely in methanol and place them in an ultrasonic bath for a
few minutes. Repeat the ultrasonic bath with deionized water.
9 Tighten the nut until fully finger-tight and then use the adjustable wrench to tighten a
further 90 degrees.
10 Refit the inlet manifold and reconnect the tubing to the pump head.
Maintenance interval
During the first months of use it is normal that excess oil leaks out of the drain hole below the System
pump. The function of the pump is not in any way affected by this.
Required material
The following materials are required:
• Cloth
• Mild cleaning agent or 20% ethanol
Instruction
Follow the instructions below to clean the System pumps externally.
Step Action
3 Wipe off the excess oil from the pump head with a damp cloth. Wipe off stains using a
mild cleaning agent or 20% ethanol.
Maintenance interval
Clean the Conductivity flow cell when required.
Required material
The following materials are required:
• Luer connector
• Waste container
• Syringe, 25-30 ml
• 1 M NaOH
• Distilled water
Instruction
Follow the instruction below to clean the flow cell of the Conductivity monitor.
Step Action
1 Disconnect the fingertight connector and the piece of tubing from the top of the Conduc-
tivity monitor, and attach a Luer connector.
2 Disconnect the piece of tubing from the bottom of the Conductivity monitor, and connect
a piece of waste tubing to the Conductivity monitor. Place the waste tubing in a waste
container.
Step Action
3 Fill a syringe with distilled water, and connect the syringe to the Luer connector.
4 Squirt distilled water through the Conductivity flow cell in small amounts. Disconnect the
syringe.
5 Fill a syringe with 1 M NaOH, and connect the syringe to the Luer connector.
6 Squirt 1 M NaOH through the Conductivity flow cell about five times.
8 Fill a syringe with distilled water. Connect the syringe to the Luer connector.
9 Inject the distilled water into the Conductivity flow cell to rinse the flow cell. Disconnect
the syringe.
10 Disconnect the Luer connector from the top of the Conductivity flow cell, and reconnect
the fingertight connector with tubing.
In this section
This section contains the following subsections:
Maintenance interval
Calibrate the pH monitor once a day, when the pH electrode has been replaced, or if the ambient
temperature has changed by more than ± 5°C.
Required material
Use two pH calibration buffers with a difference of at least one pH unit. Preferably use a pH standard
buffer, pH 4 or pH 7, as the first calibration point, and a pH standard buffer close to the lowest or
highest pH you need to measure as your second point. Allow the buffers to equilibrate to ambient
temperature before use.
Instruction
CAUTION
pH-electrode. Handle the pH-electrode with care. The glass tip may break and cause
injury.
1 In the Calibration dialog, select pH from the Monitor to calibrate drop-down list.
3 Enter the pH of the first pH standard buffer in the pH for buffer 1 field.
Step Action
4 Fill a syringe with approximately 10 ml of the first pH standard buffer. Connect the syringe
to the Luer connector of pH valve port Cal, and inject the buffer.
6 Wash the pH flow cell by injecting water into pH valve port Cal using a new syringe.
7 Enter the pH of the second pH standard buffer in the pH for buffer 2 field.
9 Is the Calibrated electrode slope ≥ 80% and the Asymmetry potential at pH 7 inside
the interval ± 60 mV?
• If Yes: Click the Close button to switch the pH valve back to the default position and
to close the Calibration dialog.
• If No: Clean the pH electrode and repeat the calibration procedure. If this does not
help, replace the electrode.
Maintenance interval
ÄKTA pure can have up to four pressure monitors: System pump pressure monitor, Sample pump
pressure monitor, Pre-column pressure monitor and Post-column pressure monitor. Check the pressure
monitors every week, or when the the ambient temperature has changed by more than ± 5°C. Calibrate
the monitor if the zero pressure reading is outside the range ± 0.02 MPa.
1 Disconnect the relevant tubing from the pressure monitor to obtain zero-pressure, see
table Tubing and pressures, on page 306.
2 Click the Customize icon to open the Customize dialog. In the Customize dialog, under
the Run Data Groups tab, select the relevant pressure to display, see table Tubing and
pressures, on page 306. Click OK to close the Customize dialog.
3 In the Run Data pane in the System Control module, check what pressure is displayed.
4 If the zero pressure reading is outside the range ± 0.02 MPa, calibrate the pressure
monitor according to the instruction below.
1 Disconnect the relevant tubing from the pressure monitor, see table Tubing and pressures,
on page 306.
2 In the Calibration dialog, select the pressure monitor to calibrate from the Monitor to
calibrate drop-down list.
System pump pressure Tubing from the System pump pressure monitor System pressure
monitor
Sample pump pressure Tubing from the Sample pump pressure monitor Sample pressure
monitor
Pre-column pressure Tubing to Column valve port In (V9-C, V9H-C or PreC pressure
monitor V9-C2, V9H-C2).
Post-column pressure Tubing to Column valve port Out (V9-C, V9H-C PostC pressure
monitor or V9-C2, V9H-C2).
Introduction
Two types of calibrations can be performed:
• Conductivity monitor - factory calibration: Restores the conductivity cell constant to the factory
default value.
• Conductivity monitor - user calibration: Calibrates the conductivity cell constant.
Maintenance interval
Recommended maintenance intervals for the two types of calibrations:
• Conductivity monitor - factory calibration: Perform calibration to override an incorrect user cali-
bration.
• Conductivity monitor - user calibration: The conductivity cell is factory calibrated, and should not
require recalibration under normal usage. Perform calibration when the signal is unstable or you
suspect that it is incorrect. It is also recommended to recalibrate the Conductivity monitor after
cleaning.
1 In the Calibration dialog, select Conductivity monitor - factory calibration from the
Monitor to calibrate drop-down list.
Result: The time for the new calibration and the current value are displayed.
2 Click Restore.
Result: The conductivity cell constant is restored to the factory default value. The conduc-
tivity cell constant is written on the packaging of the Conductivity monitor.
1 Make sure that the instrument has been switched on for at least one hour.
Step Action
3 In the System Settings dialog, select Conductivity and Cond temp compensation. Set
the Compensation factor to 0%, and click OK.
4 Wash the whole flow path and fill it with deionized water using a suitable inlet and the
System pump, until the conductivity value reaches 0.00 mS/cm.
5 Connect the tubing (labelled 3) from the system pump pressure monitor directly to the
conductivity cell inlet, by-passing the Mixer, Injection valve, Column valve and UV-monitor,
see image below. Ensure that the Flow restrictor is inline.
6 Prime and purge the inlet used in step 4 with the 1.00 M NaCl calibration solution. Fill the
conductivity cell with the calibration solution at 1 ml/min. Pump in at least 15 ml of the
calibration solution and wait until the conductivity signal and the temperature have
stabilized before continuing the calibration. See Section 5.4.1 System pumps, on page 161.
7 Under continuous flow, read the current Conductivity temperature from Run Data with
an accuracy of one decimal place.
8 In the Calibration dialog, select Conductivity monitor - user calibration from the Mon-
itor to calibrate drop-down list.
Step Action
9 In the Run Data pane of System Control, read the current Cond temp.
Tip:
If Cond temp is not showing, click the Customize icon. In the Customize dialog, under the
Run Data Groups tab, select to view Cond temp.
10 In the Calibration dialog, enter the theoretical conductivity value at the current conduc-
tivity temperature in the Enter theoretical conductivity value input field.
• If a certified conductivity standard solution is used, use the supplied theoretical con-
ductivity value.
• If a manually prepared 1.00 M NaCl calibration solution is used, see the graph for
conductivity value at the current temperature on page 311.
12 In the System Control toolbar, click the End icon to end the run.
13 In the System Settings dialog, select Conductivity:Cond temp compensation and set
the Compensation factor back to desired value, default 2.0%. Click OK.
Note:
When using NaCl, the optimal compensation factor is 2.1%.
97
95
90
Conductivity (mS/cm)
85
80
77
20 25 30
Temperature (°C)
Equipment needed
To perform the calibration, a calibration kit containing test solutions, syringes and accessories is
needed. A specified kit is available for each cell length.
To calculate the real path length of the UV flow cell, use the following software:
• UV-900 cell calibration Excel-file (Product Code 18-6324-06)
1 Ensure that the flow restrictor is inline in the flow path after the UV flow cell.
2 Mount the union Luer female/1/16” male, included in the test kit, in the upper inlet of the
UV flow cell.
Step Action
4 The solution bottles are labelled with the concentration value and the reference ab-
sorbance value for each solution. Enter the concentrations of the solutions in ascending
order into the column UV Test kit Concentration (mg/l). Enter the corresponding ab-
sorbance values into the column UV Test kit Absorbance (AU/cm).
2 Fill one of the supplied syringes with 1.5-2 ml of the first solution (0 mg/l). Ensure that
there are no air bubbles in the syringe.
3 Fit the syringe in the union Luer connector and inject the solution. DO NOT remove the
syringe.
Note:
Air trapped in the UV cell causes inaccurate measurements. To avoid introducing air into
the UV cell, gently fill the union Luer up to the edge with test solution that is to be introduced,
using the syringe. Then insert the syringe into the union Luer.
Step Action
7 Repeat the injections with the remaining four test solutions in increasing concentration
order. Use a new syringe for each solution.
8 After each injection, wait for a stable absorbance value. Note the measured absorbance
values for each solution.
9 Enter the measured absorbance values into the table in the column UV-900 Absorbance
(AU).
Note:
The values should be converted from mAU to AU.
10 When all absorbance values have been entered into the table, the real UV flow cell path
length is shown at the bottom of the table.
Note:
The regression coefficient R2 should be larger than 0.999. If this is not the case, one or
more measured values are faulty.
3 Enter the nominal flow cell path length in the Nominal length input field and click Set.
Step Action
4 • If a calibration has been performed: enter the calculated flow cell path length, obtained
in the calibration procedure, in the Real length input field and click Set.
• If no calibration has been performed: enter the nominal flow cell path length in the
Real length input field and click Set.
Result: The UV flow cell path length is updated.
WARNING
Disconnect power. Always disconnect power from the instrument before replacing
any component on the instrument, unless stated otherwise in the user documenta-
tion.
In this section
This section contains the following subsections:
Maintenance interval
Replace tubing and connectors when required, for example when a tubing has clogged or has been
bent so that the flow is stopped.
Required material
The following material are required:
• Tubing and connectors
• Tubing cutter
• Fingertight wrench
Instruction
Follow the instruction below to replace tubing and connectors.
Step Action
3 If the tubing has labels, remove the labels to be used with the new tubing later. Discard
the tubing and connectors.
Step Action
4 Cut the new tubing to the same length as the old tubing. Use a tubing cutter to get a
straight angle cut.
CAUTION
Cut injuries. The tubing cutter is very sharp and must be handled with
care to avoid injuries.
Note:
When replacing system tubing, use the original inner diameter and length to ensure that
the correct delay volumes are maintained. Inlet and outlet tubing may be shortened if re-
quired.
7 Insert the tubing with connector into the port. Make sure to insert the tubing all the way
into the bottom of the port.
8 Tighten the connector fully. For areas difficult to access, use the fingertight wrench in-
cluded in the accessory kit.
Maintenance interval
Replace tubing and connectors when required, for example when the tubing has clogged or has been
bent so that the flow is stopped.
Required material
The following material are required:
• Tubing and connectors
• Tubing cutter
• Fingertight wrench
Instructions
Follow the instructions below to replace tubing and connectors. See Section 3.2.2 Fraction collector
F9-C illustrations, on page 92 for the locations of the components of the fraction collector.
Step Action
1 Switch off the power to the fraction collector by switching off power to the ÄKTA pure
instrument.
2 Remove the connector for the internal tubing that is attached to the inlet port:
• Unscrew the connector from the inlet port.
Step Action
4 Cut the new tubing to the same length as the old tubing. Use a tubing cutter to get a
straight angle cut.
CAUTION
Cut injuries. The tubing cutter is very sharp and must be handled with
care to avoid injuries.
Note:
Use tubing with the original inner diameter and length to ensure that the correct delay
volume is maintained.
Step Action
6 Thread the tubing into the fraction collector and through the tubing guide in the ceiling
of the fractionation chamber.
9 Make sure that the tubing not is twisted or bent and that it does not dip downwards.
Maintenance interval
Replace the Mixer when a different Mixer chamber is desired, or when the Mixer is damaged.
Required material
The following is required:
• Mixer
Instruction
Follow the instruction below to change the Mixer.
Step Action
1 Disconnect the tubing from the top and bottom of the Mixer.
4 Reconnect the tubing to the top and bottom of the new Mixer.
Maintenance interval
Replace the O-ring inside the Mixer if it is damaged.
Required material
One of the following O-rings are required:
• O-ring 13.1 x 1.6 mm (for Mixer chambers 0.6, 1.4, and 5 ml)
• O-ring 13.1 x 1.6 mm (highly resistant, for use when the system is exposed to organic solvents or
high concentrations of organic acids such as acetic acid or formic acid for longer periods of time).
• O-ring 22.1 × 1.6 mm (for Mixer chamber 15 ml)
Instruction
Follow the instruction below to replace the O-ring inside the Mixer.
Tip: Use a forceps and gloves during the replacement procedure to avoid contaminating the Mixer
components.
Step Action
2 Unscrew the top section of the Mixer and pull apart the Mixer in two halves.
Step Action
4 Lift up the top section of the Mixer and pull away the old O-ring inside.
5 Wet the new O-ring with 20% ethanol and fit it in position. Make sure that the inline filter
is still in position.
6 Reassemble the Mixer components and, while holding the Mixer upright, screw the top
section back onto the Mixer.
Maintenance interval
Replace the UV flow cell when it is desired to use a flow cell with a different path length, or if the cell
is damaged. Clean the optical fiber connectors if they have accidentally been touched.
Required materials
The following materials are required:
For replacement of flow cell
• UV flow cell
For cleaning of the optical fiber connectors
• Lens paper
• Isopropanol
CAUTION
Hazardous chemicals or biological agents in UV flow cell. Make sure that the entire
flow cell has been flushed thoroughly with bacteriostatic solution (e.g., NaOH) and
distilled water, before service and maintenance.
Note:
While the UV detector is disconnected, the UV lamp becomes inoperable so no UV light
can be emitted from the instrument.
Step Action
4 Pull off the detector and the flow cell from the monochromator. Be careful not to damage
the UV flow cell.
Note:
Make sure that the flow cell does not come into contact with any liquid, and that no liquid
enters the UV detector or monochromator.
Note:
While the UV detector is disconnected, protect the fiber connectors from dust or other im-
purities by mounting the rubber protective caps onto them.
Note:
Do not touch the optical fiber connectors as this will result in poor monitor performance.
If you accidentally touch the optical fiber connectors, clean them according to Clean the
optical fiber connectors, on page 327.
6 Pull off the black protective caps from the new UV flow cell, and connect the new UV flow
cell to the detector.
7 Connect the detector, with the new flow cell connected, to the monochromator. Pull the
latch upwards to fasten the detector.
WARNING
Hazardous substances. When using hazardous chemicals, take all suitable protective
measures, such as wearing protective clothing, glasses and gloves resistant to the
substances used. Follow local and/or national regulations for safe operation and
maintenance of the product.
Step Action
Maintenance interval
Replace the UV flow cell when it is desired to use a flow cell with a different path length, or if the cell
is damaged.
Required material
• UV flow cell
CAUTION
Hazardous chemicals or biological agents in UV flow cell. Make sure that the entire
flow cell has been flushed thoroughly with bacteriostatic solution (e.g., NaOH) and
distilled water, before service and maintenance.
3 Unscrew the knurled wheel at the bottom of the UV monitor (1). Press the wheel upwards
to release the flow cell (2).
1 2
Step Action
4 Pull the flow cell upwards out of the monitor. Hold the flow cell by the top part with the
O-ring: do not touch the optical surfaces of the flow cell.
Note:
Make sure that the flow cell does not come into contact with any liquid, and that no liquid
enters the monitor.
9 Update the UV flow cell path length in the Calibrate dialog, in System Control.
Note:
The flow path cell length can be:
• updated with the nominal value, see Update the cell path length , on page 314 to set
the nominal flow cell path length
or
• calibrated and updated with the calculated value, see Calibration of the UV monitor
U9-L flow cell length, on page 312 to calibrate the flow cell path length and to update
the software with the calculated value.
Maintenance interval
Replace the Flow restrictor when required, for example when the back pressure of the Flow restrictor
is outside the range 0.2 ± 0.05 MPa.
Required material
The following material is required:
• Flow restrictor FR-902
Instruction
Follow the instruction to replace the Flow restrictor.
Step Action
1 Disconnect the tubing connected from the used Flow restrictor, and discard the used
Flow restrictor.
2 Connect the tubing to the new Flow restrictor. Make sure that the Flow restrictor connector
marked IN is connected to the pH valve port ToR (To Restrictor), and that the Flow restrictor
connector marked OUT is connected to the pH valve port FrR (From Restrictor).
3 Check the back-pressure of the new Flow restrictor, see Section 7.4 Monthly maintenance,
on page 269.
Maintenance interval
Replace the inlet filter when required, for example when the filters are clogged.
Required materials
The following material is required:
• Inlet filter set
Instruction
Follow the instruction to replace an inlet filter and a support net.
Step Action
1 Pull off the inlet filter and the support net from the inlet filter holder.
2 Fit the new support net and inlet filter, and press the filter into position.
Maintenance interval
Replace a check valve when required, for example if the check valve is damaged or clogged. The fol-
lowing instructions are valid for the system pumps and the sample pump.
Required materials
The following materials are required:
• Check valve kit
• Adjustable wrench
Instruction
NOTICE
Handle the check valves with care when they have been removed from the pump
heads, to prevent loss of any internal components.
1 Disconnect the tubing from the pump head and disconnect the pump inlet tubing.
2 Unscrew the purge valve by turning it counter-clockwise, and lift off the metal ring.
Step Action
3 Unscrew the plastic nut of the upper check valve using an adjustable wrench, and gently
lift off the upper check valve.
5 Tighten the nut until fully finger-tight and then use the adjustable wrench to tighten a
further 180 degrees.
6 Place the new metal ring onto the new upper check valve, and screw the new purge
valve.
7 Unscrew the two white plastic screws located below each pump head. Pull the plastic
connectors to the sides to release the inlet manifold.
Step Action
10 Tighten the nut until fully finger-tight and then use the adjustable wrench to tighten a
further 180 degrees.
11 Refit the inlet manifold and reconnect the tubing to the pump head.
Maintenance interval
Replace the O-rings and piston seals and rinse the membranes of the pumps if they are damaged.
After replacement, perform a run to break in the new piston seals.
The instructions in this section apply to the following pumps.
NOTICE
Advanced maintenance. Read the instruction carefully before disassembly of the
pump head.
Required material
The following materials are required:
• Adjustable wrench
• For Pump P9: Star screwdriver, T20
• For Pump P9H: Star screwdriver, T10 and T20
• Ultrasonic bath
• Ethanol, 20%
• For Pump P9: Tubing giving a back pressure of 6 to 8 MPa (60 to 80 bar).
• For Pump P9H:Tubing giving a back pressure of 2 to 3 MPa (20 to 30 bar)
• For Pump P9: P9 Seal kit, 25 ml
• For Pump P9H: P9H Seal kit, 150 ml
Illustrations
The illustrations below show the parts of the pump heads of the pumps P9 and P9H.
Pump P9
8 9 10
1 2 3 4 5 6 7
11
12
13
Pump P9H
8 9 10
1 2 3 4 5 6 7
11
12
13
13 Star screws
Introduction
Follow the instructions below to replace the O-rings, piston seal, and pump membrane housing of
pumps P9 and P9H.
Note: Always replace the O-rings, piston seals, and pump membrane housing of both pump heads
of a pump at the same time.
2 Disconnect the tubing from the pump head, and disconnect the pump inlet tubing.
3 Unscrew the two white plastic screws located below each pump head by hand. Pull the
plastic connectors to the sides to release the inlet manifold.
Step Action
5 Unscrew the two screws of the front section of the pump head using a star screwdriver
T20, and pull off the front section.
6 Place the front section of the pump head face down on the bench. For Pump P9, unscrew
the two screws of the support washer using a star screwdriver, T20. For Pump P9H, un-
screw the four screws of the support washer using a star screwdriver, T10. Discard the
O-ring (1) on the support washer, and the discard the piston seal (2) located in the front
section of the pump head.
Note:
Be careful not to scratch the metal surfaces.
2
1
2 1
Step Action
7 Unscrew one of the two screws securing the pump membrane housing using a star
screwdriver, T20. Unscrew the second screw, and at the same time push firmly on the
front of the pump membrane housing to compensate for the pressure of the piston return
spring.
8 Carefully pull off the pump membrane housing together with the piston and return spring.
9 Inspect the piston and return spring for signs of damage. If damaged, discard the piston
and return spring and use a new piston and return spring when assembling the pump
head.
10 Clean the pump head and pump membrane housing in an ultrasonic bath. If there are
particles on any surfaces, the check valves should be removed and cleaned separately,
see Section 7.6.8 Clean the pump head check valves, on page 296.
1 Unscrew the purge valve of the pump head. Replace the O-ring on the purge valve with
a new O-ring, and screw the purge valve back into the pump head.
Note:
Always use Lubricant 56686700 when exchanging the O-ring 3 x 1 mm.
2 Wet a new seal with 20% ethanol. Place the new seal in the hole in the front section of
the pump head and press it into position.
Step Action
3 Place the support washer on top of the new seal in the front section of the pump head.
Screw the two or four screws of the support washer. Make sure to tighten the screws
fully.
4 Wet a new O-ring, 21.4 x 1.6 mm, with 20% ethanol. Fit the O-ring around the support
washer.
Step Action
1 Insert the piston into the return spring. Insert piston and return spring into hole in the
pump module.
Note:
Do no touch the ceramic or glass part of the pump piston.
2 Wet the membrane in the hole with 20% ethanol before mounting.
3 Place the pump membrane housing onto the locating pins on the front of the pump
module.
4 Screw one of the two screws securing the pump membrane housing using a star
screwdriver, T20. Push firmly on the front of the pump membrane housing to compensate
for the pressure of the piston and then screw the second screw.
5 Make sure that the new seal is wetted with 20% ethanol and then tighten both screws
fully.
8 Reconnect the tubing to the pump head, and reconnect the pump inlet tubing.
1 Fill a buffer vessel with 20% ethanol in water. Immerse the inlet tubing, for example A1
for System pump A, or B1 for System pump B, in the buffer vessel. Place the buffer vessel
on the Buffer tray.
2 Prime the inlets and purge the pump, see Section 5.4.1 System pumps, on page 161.
3 • For pump P9: Connect tubing that gives a back pressure of 6 to 8 MPa to one of the
column positions of the Column valve or between the Injection valve and the UV
monitor.
• For pump P9H: Connect tubing that gives a back pressure of 2 to 3 MPa to one of the
column positions of the Column valve or between the Injection valve and the UV
monitor.
3 Immerse the waste tubing in the buffer vessel to recirculate the liquid.
Step Action
Maintenance interval
Replace the O-ring and the piston seal, and rinse the membrane of the pump if they are damaged.
After replacement, perform a run to break in the new piston seals.
The instructions in this section apply to the following pump.
NOTICE
Advanced maintenance. Read the instruction carefully before disassembly of the
pump head.
Required material
The following materials are required:
• Adjustable wrench
• Star screwdriver, T20
• Cross-headed screwdriver
• Hex wrench
• Ultrasonic bath
• Ethanol, 20%
• Reference capillary Ref 1
• P9-S Seal kit, 65 ml
Illustration
The illustration below shows the parts of the pump heads of Pump P9-S.
7 8 9
1 2 3 4 5 6
10
Introduction
Follow the instructions below to replace the o-ring, piston seal, and rinse membrane of pump P9-S.
Note: Always replace the o-rings, piston seals, and rinse membranes of both pump heads of a pump
at the same time.
2 Disconnect the tubing from the pump head, and disconnect the pump inlet tubing.
Step Action
3 Unscrew the two white plastic screws located below each pump head by hand. Pull the
plastic connectors to the sides to release the inlet manifold.
5 Unscrew one of the two screws of the pump head using a hex wrench. Unscrew the
second screw, and at the same time push firmly on the front of the rinse chamber to
compensate for the pressure of the piston return spring.
6 Place the pump head face down on the bench. Pull out the piston together with the return
spring.
Step Action
7 Inspect the piston and return spring for sign of damage. If damaged, discard the piston
and return spring and use a new piston and return spring when assembling the pump
head.
8 Unscrew the two screws securing the drain plate and the rinse chamber. Lift off the drain
plate, and discard the membrane located between the drain plate and the rinse chamber.
9 Lift off the rinse chamber. Gently pull off the piston seal. Discard the used seal.
10 Clean the pump head, rinse chamber and drain plate in an ultrasonic bath. If there are
particles on any surfaces, the check valves should be removed and cleaned separately,
see Section 7.6.8 Clean the pump head check valves, on page 296.
1 Unscrew the purge valve of the pump head. Replace the O-ring on the purge valve with
a new O-ring, 3x1 mm, and screw the purge valve back into the pump head.
Note:
Always use Lubricant 56686700 when exchanging the O-ring 3 x 1 mm.
2 Wet a new seal with 20% ethanol. Place the new seal in the hole in the front section of
the pump head and press it into position.
3 With the pump head facing downwards on the bench, place the rinse chamber onto the
front section of the pump head with the rinse ports in line with the check valves. The
conical depression in the rinse chamber shall be facing upwards. Wet a new membrane
with 20% ethanol, and place the membrane into the rinse chamber with the conical face
upwards.
Step Action
1 Place the drain plate on top of the assembly. Screw the two screws through the drain
plate and the rinse chamber using a cross-headed screwdriver.
2 Wipe clean the piston and remove all finger prints. Wet the piston with 20% ethanol, and
insert the piston into the return spring. With the pump head facing downwards on the
bench, insert the piston into the pump head by pushing it gently but firmly vertically
downwards into the seal.
3 Place the complete pump head over the locating pins on the front panel of the sample
pump module. Turn the pump head so that the text UP on the drain plate is facing up-
wards. Push firmly on the front of the pump head, and at the same time screw one of
the screws to fasten the pump head onto the front of the module using a hex wrench.
Screw the second screw of the pump head. Make sure to tighten both screws fully.
6 Reconnect the tubing to the pump head, and reconnect the pump inlet tubing.
1 Fill a buffer vessel with 20% ethanol in water. Immerse a piece of sample inlet tubing, for
example S1, in the buffer vessel. Place the buffer vessel on the Buffer tray.
2 Prime the inlets and purge the pump, see Section 5.4.2 Sample pump, on page 167
3 Connect the reference capillary Ref 1 (or equivalent tubing that gives a backpressure of
2-3 MPa) to one of the column positions of the Column valve (e.g., ports 1A and 1B).
4 Immerse the waste tubing in the buffer vessel to recirculate the liquid.
Step Action
Maintenance interval
Replace the pump pistons if they are damaged.
Required material
The following materials are required:
• Adjustable wrench
• Star screwdriver, T20
• Piston kit
Maintenance interval
Replace the pump rinsing system tubing when required , for example if the tubing is clogged or damaged.
Replacement instructions for the system pumps and the sample pump (external module) are given
below.
Required material
The following material is required:
• Rinsing system tubing
5 4
Part Description
4 Outlet tubing
5 Inlet tubing
4 5
Part Function
3 Inlet tubing, from the rinsing liquid tube to the left pump head (lower rinsing system
connection) .
4 Tubing between the left pump head (upper rinsing system connection) and the right
pump head (lower rinsing system connection).
5 Outlet tubing, from the right pump head (upper rinsing system connection) to the rinsing
liquid tube.
4 Fit all pieces of tubing into the tubing holders on the pump modules.
8 Troubleshooting
In this chapter
This chapter contains the following sections:
Text in the Process Picture The operating system of the computer does not have the text font
pane in the System control Calibri True Type installed. For example, Windows XP does not
module looks strange include this font by deafault.
Install the font Calibri True Type or switch to an operating system
that includes the font.
Troubleshooting procedure
To troubleshoot ÄKTA pure:
Step Action
1 Always start by checking the General checklist. See Section 8.2 Troubleshooting: General
Checklist, on page 358.
2 In this document, search for solutions in the section corresponding to the problem.
4 If problems remain after corrective actions, generate a System error report and contact
your local GE representative.
Step Action
1 In the System Control module, on the System menu, click Create System Error Report.
Result: The first page of a wizard is displayed.
2 • Click Next and start to enter information about the problem, click Next.
• Choose to enclose methods, logs or result files.
• Select location for the report and click Finish to generate the report. The filename of
the zip file will be Report_YYYYMMDD.zip and the default folder location is: C:Program
Files\GE Healthcare\UNICORN.
System checks
• Is the correct system selected in UNICORN System Control? For more details, see Section 5.3 Start
UNICORN and connect to system, on page 157.
• Are the fans blowing at the back and at the right side of the system?
Monitor checks
• Is the UV monitor set to the correct wavelength? Check the wavelengths that are used in the method.
For the predefined methods the wavelengths are set in the Method Settings phase. For more details,
see UNICORN Method Manual.
• Is the air sensor sensitivity set to normal in UNICORN System Control:System Settings to avoid
unnecessary stops due to minor air bubbles?
Instrument communication
• Have Node IDs been set correctly for all instrument modules? To check and change Node IDs, see
Node ID, on page 76.
• Does the positioning of the columns correspond to the selections made in the method? For more
details, see Section 5.5 Connect a column, on page 172.
Purification checks
• Have all columns been cleaned and prepared according to the column recommendations?
• Have the samples been adjusted to binding buffer conditions?
• Have the samples been clarified by centrifugation and/or filtration prior to sample loading?
• Are the correct buffers used for the chosen columns and proteins?
• Check buffers for precipitations. Adjust to room temperature.
• Are the chosen columns suitable for the chosen target proteins?
• Do the buffers have correct pH? The pH of some buffers changes with the temperature.
• Are the UV-wavelengths used by the method appropriate with respect to used buffers and proteins?
For more details see the method handbooks available from GE.
The internal temperature of • The air intake on the rear or on the left side of the instrument
the UV monitor is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The UV curve shows a gradi- • Large difference in refractive index between buffer A and buffer
ent that is inverted com- B
pared to the expected gradi-
Due to light spreading effects in the UV cell, the buffer with the
ent
highest UV absorption shows the lowest UV absorption in the
chromatogram, and the buffer with the lowest UV absorption
shows the highest UV absorption. This can occur if there is a large
difference in refractive index between buffer A and buffer B and
the UV is run at high sensitivity.
UV monitor U9-L
Problem Possible cause and action
The internal temperature of • The air intake on the rear or on the right side of the instrument
the UV monitor is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The UV cell is not correctly • Check that the UV flow cell is fitted correctly, see Section 7.8.6
installed Replace the UV monitor U9-L flow cell, on page 328.
Conductivity monitor
Problem Possible cause and action
The conductivity part of the Test was performed at cold room temperature and the conductiv-
system performance test ity temperature compensation factor was not optimized.
failed In the System Settings dialog, select Conductivity:Cond temp
compensation and set the Compensation factor to 2.1%.
Temperature out of range • This error can only occur when the temperature compensation
for calibration is turned on. The error will occur when the temperature is
outside the range 2°C to 40°C.
Make sure the temperature of the calibration solution is within
2°C and 40°C.
Incorrect or unstable read- • The temperature compensation factor is not properly set
ing
Use a temperature compensation factor. The temperature com-
pensation factor is found in System Control:System Settings:Con-
ductivity. Instruction regarding the factor is also found in Sec-
tion 7.7.3 Calibrate the Conductivity monitor, on page 308.
• The column is not equilibrated
Equilibrate the column. Use the method phase Equilibration.
If necessary, clean the column. Use the predefined method Col-
umn CIP. See Section 7.6.3 Perform Column CIP, on page 284.
• Poor mixing function
Check that the correct Mixer chamber size is used. See Select
Mixer chamber, on page 151 for recommendations. To change the
Mixer, see Section 7.8.3 Replace the Mixer, on page 322.
Check the motor operation of the mixer. Place a magnet close
to the Mixer chamber during run. The magnet should vibrate.
Check that the Mixer chamber is free from solids. To open the
mixer, see Section 7.3.2 Replace the inline filter, on page 266.
The pH module is not found • The cable between the pH valve and the ICU is not connected
by the instrument
Remove the pH valve and make sure that the cable is connected.
See Hardware installation of a module, on page 76.
The internal valve tempera- • The air intake on the rear or on the left side of the instrument
ture is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
Drift of pH signal when the • Decreasing salt concentration in the electrode membrane due
pH electrode has been re- to osmosis to buffer
moved from storage solu-
Regenerate the pH electrode. Place the electrode in deionized
tion
water for 30 minutes followed by 30 minutes in a buffer with pH
4.
Pressure monitors
Problem Possible cause and action
The pressure monitors are • The cable between the Pressure monitors and the ICU is not
not found by the instrument connected.
Remove the monitor and make sure that the cable is connected.
The internal temperature of • The air intake on the rear or on the left side of the chromatog-
the pressure monitor is too raphy instrument or the sample pump is covered
high
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
Irregular pressure curves • Air bubbles are passing through or are trapped in the pump
Check that there is a sufficient volume of buffer present in the
flasks.
Check all connections for leaks.
Check pump pressure curves. See Examples of pump pressure
curves, on page 390 for examples of pump pressure curves.
Purge the pumps, see Section 5.4.1 System pumps, on page 161
or Section 5.4.2 Sample pump, on page 167
• The check valve does not function correctly
Remove any solids in the valves by cleaning the check valves
according to the instructions in Section 7.6.8 Clean the pump
head check valves, on page 296.
• Piston seal is leaking
Replace the piston seal of a system pump according to the in-
structions in Section 7.8.10 Replace pump piston seals of Pump
P9 or P9H, on page 335. Replace the piston seal of the sample
pump according to the instructions in Section 7.8.11 Replace
pump piston seals of Pump P9-S, on page 344
• Blockage of flow path
Use the predefined method Prepare System to flush through to
clear blockage.
If necessary, replace tubing. See Section 7.8.1 Replace tubing and
connectors, on page 317.
Check the mixer inline filter. It can be clogged if unfiltered buffers
or samples are applied. See Section 7.3.2 Replace the inline filter,
on page 266 for instructions how to replace the mixer inline filter.
Check the inlet filters. They can be clogged if unfiltered buffers
or samples are applied. To replace the filters, see Section 7.8.8
Replace the inlet filters, on page 331.
Check the column. It can be clogged if unfiltered buffers or
samples are applied. To clean a column, see Section 7.6.3 Perform
Column CIP, on page 284.
The valve is not found by • The cable between the valve and the ICU is not connected
the instrument
Remove the valve and make sure that the cable is connected.
See Hardware installation of a module, on page 76.
• Wrong Node ID
Check the valve Node ID. If necessary, change the Node ID. See
Node ID, on page 76.
The internal valve tempera- • The air intake on the rear or on the left side of the instrument
ture is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
Inlet valves
The Inlet valves include Inlet valve V9-IA, V9H-IA, V9-IB, V9H-IB, V9-IAB, V9H-IAB, V9-IX and V9H-IX, as
well as Sample inlet valve V9-IS and V9H-IS.
pH valve
Problem Possible cause and action
Leaking pH valve • The dummy electrode was dry when it was installed in the
valve.
1 Remove the dummy electrode.
2 Wet the dummy electrode properly with distilled water.
3 Insert the dummy electrode into the pH valve.
4 Rotate the dummy electrode before securing it with the nut.
Module Panel
Problem Possible cause and action
The instrument is unable to • A Module Panel is missing and the position is left empty
find some of the modules
Install the missing Module panel.
• The cable between the Module Panel and the ICU is not con-
nected
Remove the Module Panel and make sure that the cable is con-
nected.
The fraction collector can- • The cable between the fraction collector and the ÄKTA pure
not be found by the instru- instrument is not connected
ment
Generate a System error report and contact Service.
• A fuse in the instrument ICU is broken
The ICU needs to be changed. Generate a System error report
and contact Service.
The accumulator is jammed • Salt crystals or protein residuals block the accumulator
or there is an internal error
Restart the instrument and perform an accumulator wash.
in the instrument
• Mechanical error
If this error is recurrent, generate a System error report and
contact Service.
Fraction collection tube or • Too many commands are pending in the fraction collector
well is overfilled and frac-
The reason could be that too many fraction collector instructions
tionation movements are
have been sent. Wait for a while and try again.
lost.
The fraction collector failed • Cassette tray or tube rack is inserted in the wrong direction
to detect the code on the
Take it out and insert it in the correct direction.
Cassette
• Unclean Cassette type code reader
Clean the dispenser head and its four diode windows using a
cloth and a mild cleaning agent or 20% ethanol. See Section 7.6.4
Clean Fraction collector F9-C, on page 288 for more information.
• Unclean Cassette type codes
Clean the Cassette type codes. See Section 7.6.4 Clean Fraction
collector F9-C, on page 288 for more information.
If this error is recurrent, set the Fraction collector configuration
manually in UNICORN. In System Control, select System:Settings.
Navigate to Fraction collector:Cassette configuration and select
Manual.
The Cassette tray is loose in • The Tray guides in the fractionation chamber are missing
the fraction collector
Replace the Tray guides. See Section 7.6.4 Clean Fraction collector
F9-C, on page 288.
• The Tray catch is broken
Contact service.
Calibration of the Drop sync • Unclean Drop sync sensor diode windows
sensor failed
Clean the Drop sync sensor diode windows. See Section 7.6.4
Clean Fraction collector F9-C, on page 288 for location of the Drop
sync sensor diode windows and cleaning instructions.
If this error is recurrent, generate a System error report and
contact Service.
Drop sync does not work • Liquid spatter in the fraction collector
Change to another type of tubes or deepwell plates or collect
fraction volumes of maximum two thirds of the volume of the
tubes. Clean the diode windows of the drop sync sensor more
frequently than once a week, using a cloth and a mild cleaning
agent or 20% ethanol.
Tubes do not fit in the Cas- • Wrong tube dimensions are used
sette
Check that the used tubes have the right dimensions. See Fraction
collector tubes and bottles, on page 99 for information about
tubes and Cassettes.
• QuickRelease function is worn out
Order a new Cassette. See Chapter 10 for ordering information.
Deep well plate does not fit • Unsupported deep well plate model
in the Cassette
Check that the deep well plates are supported. See Deep well
plates, on page 100.
The Cassette does not fit • The Cassette is turned in the wrong direction
the Cassette tray
See Prepare and insert the Cassette tray, on page 194 for informa-
tion of how to place the Cassettes.
• Objects or dirt under the Cassette
Remove the object or dirt.
Cassette tray in wrong posi- • The Tray guides in the fractionation chamber are missing
tion
Replace the Tray guides. See Section 7.6.4 Clean Fraction collector
F9-C, on page 288.
• The Cassette tray is facing the wrong direction
Make sure that the front of the tray (marked with the GE-logotype)
is facing outwards. See Prepare and insert the Cassette tray, on
page 194.
• Dirt under the Cassette tray
Remove the dirt.
The loaded Cassette tray • The Cassette tray is facing the wrong direction
cannot be inserted into the
Check that the GE-logotype is facing outwards when the tray is
fractionation collector
inserted into the fraction collector.
• Some of the tubes or plates are incorrectly placed in the Cas-
settes
Check that all tubes and plates are correctly inserted in the Cas-
settes. See Prepare and insert the Cassette tray, on page 194 .
• Some of the tubes or plates have the wrong dimensions
Check that the deep well plates and the tubes used are of the
right type. See Fraction collector tubes and bottles, on page 99
and Deep well plates, on page100 for information about supported
tubes and plates.
Quick scan or Full scan does • The Tray guides in the fractionation chamber are missing
not work
Replace the Tray guides. See Section 7.6.4 Clean Fraction collector
F9-C, on page 288.
• The Cassette tray is facing the wrong direction
Check that the GE-logotype is facing outwards.
• The Cassette type codes are unclean
Clean the the Cassette type codes.
• The QuickRelease of a Cassette is in open position
Close the Cassette. See Prepare and insert the Cassette tray, on
page 194.
• The Cassette code reader diode window is unclean
Clean the Cassette code reader diode window. See Section 7.6.4
Clean Fraction collector F9-C, on page 288.
• The automatic scanning is turned off in UNICORN
Make sure that the automatic scanning is turned on in UNICORN
software. In System Control, select System:Settings. Navigate
to Fraction collector:Cassette configuration and select Auto-
matic.
• Wells in deep well plates are prefilled to a volume above 25%
of the total well volume
Full scan will not work with prefilled wells during these conditions.
• Hardware error
Generate a System error report and contact Service.
The interior of the fraction • The light has been turned off in UNICORN
collector is dark
Turn on the light in UNICORN. In System Control, select Sys-
tem:Settings. Navigate to Fraction collector:Fraction collector
lamp and select Lamps On.
• The lamp is broken
Contact Service.
The waste or the interior of • The waste tubing is positioned so that the flow is obstructed
the fraction collector is
Untwist the waste tubing.
flooded
• The waste container is placed in a position higher than the
waste outlet
Place the waste container in a position lower than the waste
outlet.
• The waste tubing is blocked.
Clean or replace the waste tubing.
The liquid leaving the nozzle • Check the position of the waste funnel
does not strike the waste
Refit the waste funnel.
funnel
The fraction volume found • Leakage on the wet side of the instrument
in the tubes or wells are
Localize the leakage and take care of the leakage, for example
smaller than expected
by tightening connectors.
• Air in pumps
Purge pumps. See Section 5.4.1 System pumps, on page 161.
• Bad pump function
See troubleshooting of pumps in Section 8.6.
• Leakage inside the Frac chamber
Contact Service.
Liquid on the floor of the • The fingertight connector on the Dispenser head is not tight
fraction collector enough
Tighten the connector. Replace the connector if the leakage is
persistent.
• The waste tubing or the waste funnel is blocked.
Make sure that there is no blockage or clogging and that the
waste tubing not is bent.
• Leakage in the Frac arm
Contact service.
Spillage by the fraction col- • Unclean diode windows in the Drop sync sensor or unclean
lector during fractionation nozzle
Clean the Drop sync diode windows and the nozzle. See Sec-
tion 7.6.4 Clean Fraction collector F9-C, on page 288. Use the
Fraction collector cleaning position.
• Too high flow rate during usage of Drop sync
Use a flow rate below 2 ml/min.
• One or more Cassettes have empty positions
Make sure that all Cassette positions contain tubes or plates.
• The tubes are flooded
Make sure that the fraction volume is adapted to the tube volume.
Fraction tubes have not been replaced by empty tubes when the
Fraction collector was opened.
It is not possible to fraction- • The door of the fraction collector is not properly closed
ate
Close the door.
The fraction collector frac- • Quick scan has not detected the correct Cassette
tionates in the wrong well
Clean the Cassette type code of the Cassette. See Section 7.6.4
or tube
Clean Fraction collector F9-C, on page 288.
• Unsupported deep well plate is used.
Make sure that approved deep well plates are used. See Deep
well plates, on page 100.
• Wells in deep well plates are prefilled to a volume above 25%
of the total well volume
Full scan will not work with prefilled wells during these conditions.
• The deep well plate is not correctly positioned in the Cassette
See Prepare and insert the Cassette tray, on page 194 for informa-
tion of how to place the deep well plate.
• Dirt on the nozzle or Drop sync sensor diode windows. Dirt
may effect where the drops fall.
Clean nozzle and Drop sync sensor diode windows. See Sec-
tion 7.6.4 Clean Fraction collector F9-C, on page 288.
Error message when the • Some parts in a method require that the door of the fraction
door of the fraction collec- collector is closed.
tor is opened during a run
Do not open the door when the Fractionation indicator is lit.
Fraction collector wash is The Fraction collector wash instruction is aborted when the fraction
reported as Completed collector door is opened during state pause. When selecting to
when it has been aborted. continue the run the instruction is reported as Completed, it should
state Cancelled.
Action: Do not open the door during the Fraction collector wash
instruction. If the door has been opened, restart the Fraction collec-
tor wash instruction to make sure that the Fraction collector wash
has been performed as recommended.
The instrument does not • The Detector - Outlet valve delay volume has been set to zero.
fractionate via the Outlet
1 In System Control, select System:Settings:Fraction collec-
valve ports
tions:Delay volumes.
2 Correct the Detector - Outlet valve delay volume.
When starting a method run Action: End a manual run before starting a method run.
during a manual run, the
fraction collector generates
an error message indicating
that the wrong Cassettes
are in place, even though
the correct Cassettes are
present in the Fraction col-
lector.
Spillage in Fraction collector Action: If possible, place the Cassettes close to waste, i.e., position
when the Frac arm is mov- 5 or 6 during fractionation.
ing from Cassette placed in
position 1 and 2 to waste
The fractionation starts in • The door of the fraction collector has been opened between
the first row again two runs. When the instrument is in state Ready and the door
is opened and closed, the Fraction collector content is reset.
If fractionation will continue after the last fraction of the first run,
do not open the door of the fraction collector between two runs.
Spillage between fractions • The combination of high flow rates and liquids with low surface
tension might lead to spillage in the Fraction collector.
If possible use Cassette positions 5 and 6. Lower the flow rate
when using liquids with low surface tension.
The fraction collector can- • The cable between the fraction collector and the ÄKTA pure
not be found by the instru- instrument is not connected
ment
Make sure that the cable is connected.
• Wrong node ID
Check the Node ID of the fraction collector. If necessary, change
the Node ID. See Section 9.16 Node IDs, on page 502.
The fraction collector failed • To high flow rate for drop sync
to detect a drop and has
Decrease the flow rate or disable drop sync.
changed tube without drop
sync. • Air in the flow path
Check the flow path for air. Fill system and purge pumps. If this
error is recurrent, generate a System error report and contact
service.
• Unclean drop sync sensor
Clean the drop sensor photocell located above the tube sensor
with a damp cloth.
• The tubing is not correctly mounted in the tubing holder nut
Check that the tubing is correctly mounted, see Connect tubing
to ÄKTA pure, on page 106.
The tube sensor has not de- • The fraction collector movement is blocked
tected a new tube.
Make sure that the fraction collector can move and is free from
obstructions.
Fraction numbering does Fraction numbering continues from where it was at the time of
not start at 1 when the the No tube error
fractionation is restared af- Manually reset the fraction number in the System settings menu.
ter a No tube error
Pumps
Problem Possible cause and action
Liquid is leaking between • Piston seal or rinsing membrane incorrectly fitted or worn
the pump head and the side
Replace or reinstall the seal or the membrane. For System pumps
panel
and Sample pump S9H, see Section 7.8.10 Replace pump piston
seals of Pump P9 or P9H, on page 335. For Sample pump S9, see
Section 7.8.11 Replace pump piston seals of Pump P9-S, on
page 344
The pump is not found by • The cable between the system pump and the ICU is not con-
the instrument nected.
Contact Service.
• The cable between the sample pump and the ÄKTA pure instru-
ment is not connected.
Connect the cable.
• A fuse in the instrument ICU is broken.
Contact Service.
The internal temperature of • The air intake on the rear or on the left side of the instrument
the pump is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
High pressure alarm • The pressure has increased due to increased viscosity
The viscosity increases in cold room. Lower the flow when per-
forming runs in a cold room.
High pressure alarm when • The parameter selected for pressure control is not the most
pressure control is activat- appropriate one
ed
The pressure control is based on either the Pre column pressure
or the Delta column pressure. To change the parameter for
pressure control, select Pre column pressure or Delta column
pressure from the Pressure control drop-down list in the Instruc-
tion box of the instruction of interest.
• The flow is too high
Lower the flow.
Too slow pressure build up • Too low I factor in the Pressure control parameters instruction
when pressure control is
In the Manual Instructions dialog, increase the I factor of the
active
Pressure control parameters instruction.
Too slow pressure build up • Too low I factor in the Constant pressure flow parameters in-
when using constant pres- struction
sure flow
In the Manual Instructions dialog, increase the I factor of the
Constant pressure flow parameters instruction.
Pressure overshoot or oscil- • Too high I factor in the Pressure control parameters instruction
lating pressure when pres-
In the Manual Instructions dialog, decrease the I factor of the
sure control is active
Pressure control parameters instruction.
Pressure overshoot or oscil- • Too high I factor in the Constant pressure flow parameters in-
lating pressure when using struction
constant pressure flow
In the Manual Instructions dialog, decrease the I factor of the
Constant pressure flow parameters instruction.
Modules cannot be found by • The cable between a module and the ICU is not connected
the instrument
Remove the module and make sure that the cable is connected.
• Two similar modules have been added to the instrument, for
example two Inlet valve A
Remove one of the modules with the same Node ID.
• The Node ID for one or more of the modules is incorrect, for
example an Inlet valve A2 has the same Node ID as Inlet valve
A. The instrument then considers them to have the same
identity.
Remove Inlet valve A2 and change the Node ID according to
Section 9.16 Node IDs, on page 502.
An unknown instrument • The Node ID for one or more of the modules is incorrect
module is connected to the
Check Node ID and change the Node ID according to Section 9.16
system
Node IDs, on page 502.
Mixer
Problem Possible cause and action
The Mixer chamber was not • The Mixer chamber is not correctly installed
recognized
Verify that the Mixer chamber is correctly installed. See Sec-
tion 7.8.3 Replace the Mixer, on page 322. If the error is recurrent,
replace the Mixer chamber. See Chapter 10 Ordering information,
on page 505.
The internal Mixer tempera- • The air intake on the rear or on the left side of the instrument
ture is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
Superloop
Problem Possible cause and action
Overpressure during filling • The Superloop is filled to the max. Pressure is not released
anywhere.
Manually turn Injection valve to Manual load position.
Cabinet
Problem Possible cause and action
The temperature of the in- • The air intake on the back or on the left side of the instrument
strument or an instrument is covered
component is too high
Make sure that none of the air intakes on the instrument are
covered.
• Broken fans
Contact Service.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
Liquid from the Buffer tray • The waste tubing from the Buffer tray is loose
is flowing onto the table
Contact Service.
One or more modules are • One or some of the minor modules use too much current. Minor
automatically turned off modules include all modules except the Fraction collector, the
UV monitor and the pumps.
The current is cut off by a temperature sensitive component. The
instrument can be restarted when the temperature has de-
creased. If persistent, generate a System error report and contact
Service.
One or more module(s) is • The cable between a module and the ICU is not connected.
not found by the instrument
Remove valves and check that cables are connected.
• One or more jumpers are loose or missing.
Check the connections on the back of the instrument. Empty
positions for UniNet-9 connectors 1 to 8 must have connected
jumpers. See the illustration below. The connector 9 should be
protected with a plastic lid and must never have a jumper con-
nected to it.
The internal instrument • The air intake at the rear or on the right side of the instrument
temperature is too high is covered
Make sure that none of the air intakes on the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The external air sensor is • The cable between the external air sensor and the ICU is not
not found by the instrument connected
Check the back of the instrument and make sure that the cable
is connected.
• The external air sensor has not been selected in UNICORN
Select the external air sensor in UNICORN. Refer to the installation
instructions delivered together with the external air sensor.
Air is introduced into the • One of the connections is not tight enough
flow path
Tighten the connectors.
Liquid is leaking from the • One of the connections is not tight enough
external air sensor
Tighten the connectors.
I/O-box E9
Problem Possible cause and action
Noisy analog signal • Too long cable between the external equipment and E9
Use as short cable as possible. Use shielded cable. Connect the
cable shield to the D-sub connector’s shield.
The internal temperature of • The air intake on the rear or on the right side of the instrument
the instrument control pan- is covered.
el is too high
Make sure that none of the air intakes of the instrument are
covered.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35 C.
• Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
Multiple error messages in • A cable to a module (including Module Panels) is not connected
UNICORN: Lost modules
Connect the cable:
1 Switch off the instrument.
2 Check all modules and connections.
3 Switch on the instrument.
4 In the displayed dialog in UNICORN, select the option Restart
the system only and click OK.
• A UniNet-9 connector is not plugged
Check that all UniNet-9 connectors that are not in use are
plugged.
UNICORN has lost communi- • The UNICORN client has lost connection to the instrument
cation with the instrument server during a temporary overload of the processor
server
Restart the UNICORN client to regain control. The run continues
and no data will be lost.
Phase properties or text in- • The module required for the instruction has not been enabled
structions are not available in the component selection in the administration module
in the method editor as ex-
Review the component list.
pected
• The wrong system has been selected when creating the new
method
Make sure that the right system has been selected in the "new
method" dialog.
All modules
Error code Description Action
Valve
Error code Description Action
20, 24 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
22 Valve not finding position Restart the instrument. If recurrent please con-
tact Service.
Pressure monitor
Error code Description Action
20, 21, 24- Internal instrument error Restart the instrument. If recurrent please con-
27 tact Service.
Air sensor
Error code Description Action
Mixer
Error code Description Action
pH monitor
Error code Description Action
20, 21 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
Conductivity monitor
Error code Description Action
20-27 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
Pump
Error code Description Action
51-53 Internal pump error Check that there is no blockage of the pump
outlet. Restart the instrument. If recurrent please
contact Service.
UV monitor U9-M
Error code Description Action
21, 25, 26, Grating error Restart the instrument. If recurrent please con-
31 tact Service.
22, 23, 32 Block filter error Restart the instrument. If recurrent please con-
tact Service.
UV detector U9-D
Error code Description Action
24, 26, 28, Internal instrument error Restart the instrument. If recurrent please con-
29, 31, 33 tact Service.
25 Too low light intensity Check that the detector and flow cell are fitted
correctly. If warning reappears, contact Service.
27 Too high light intensity Check that the detector and flow cell are fitted
correctly. If warning reappears, contact Service.
30 Too high light intensity, R channel Check that the detector is fitted correctly. If
warning reappears, contact Service.
32 Too high S light intensity, S chan- Check that the flow cell is fitted correctly. If
nel warning reappears, contact Service.
35 Too low light intensity, R channel Check that the detector is fitted correctly. If
warning reappears, contact Service.
36 Too low light intensity, S channel No light through flow cell. Check solution absorp-
tion and that the cell is fitted correctly.
UV monitor U9-L
Error code Description Action
58 Too low light intensity, S channel No light through flow cell. Check solution absorp-
tion and that the cell is fitted correctly.
59, 60 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
20 - 22 Internal instrument error. Restart the fraction collector with the power
switch on the ÄKTA pure instrument. If recurrent,
generate a system error report and contact ser-
vice.
32 The fraction collector failed to Restart the fraction collector with the power
operate the accumulator. switch on the ÄKTA pure instrument. If recurrent,
generate a system error report and contact ser-
vice.
33 - 35, 37 - Fraction collector configuration The instrument restarts. If the error is recurrent,
40, 55 error. generate a system error report and contact ser-
vice.
47 - 48 The fraction collector failed to Clean the cassette code reader, the cassette type
detect the code on the cassette code on the cassette and the plate.
or the type of plate.
53 Calibration of the dropsync sen- Clean the Drop sync sensor, see user documen-
sor failed. tation for the location of the sensor. If this error
is recurrent, please contact service.
57 - 59 Internal fraction collector error. The instrument restarts. If the error is recurrent,
generate a system error report and contact ser-
vice.
63 Tube change is too fast. Lower the flow rate or increase the fractionation
volume.
68, 72 Fraction collector arm movement The instrument restarts. If the error is recurrent,
is blocked. generate a system error report and contact ser-
vice.
21 Drop sync warning Clean the sensor and remove air bubbles in the
flow path.
22 Tube sensor error Check that the tube sensor is adjusted properly.
23 Delay queue full Increase the fraction size. Fraction size must be
greater than 1/10 of the delay volume.
24, 26 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
25, 28 Too fast tube change Increase the fraction size or lower the flow rate.
I/O-box
Error code Description Action
21 Analog in signal below -2V Check the external equipment connected to the
I/O-box.
23-28 Internal instrument error Restart the instrument. If recurrent please con-
tact Service.
9 Reference information
In this chapter
This chapter contains the following sections:
Parameter Data
Environmental ranges
Parameter Data
Operating ranges
Parameter Data
System pumps
Parameter Data
Sample pump
Parameter Data
Weight 11 kg
Valves
Parameter Data
Number of valves Up to 12
Inlet options
Parameter Data
Inlet A 1, 2 or 7 inlets
Inlet B 1, 2 or 7 inlets
Outlet options
Parameter Data
Number of outlets 1 or 10
Mixer
Parameter Data
Gradient formation
Parameter Data
Pressure monitors
Parameter Data
Number of sensors Up to 4
Placement of sensors Standard: The System pressure monitor is located after the
System pump
Options:
• The Pre-column pressure monitor and the Post-column
pressure monitor are integrated in Column valve V9-C
or V9H-C.
• The Sample pressure monitor is located after the Sample
pump.
Number of sensors Up to 7
UV monitor options
Parameter Data
Number of monitors Up to 2
Absorbance range -6 to 6 AU
Parameter Data
pH monitor option
Parameter Data
pH reading range 0 to 14
Number of outlets 10
Parameter Data
Weight • F9-C: 21 kg
• F9-R: 5 kg
I/O box
Parameter Data
PEEK, o.d. 1/16", i.d. 0.25 Blue High pressure tubing 0.5 μl
mm Reference capillary 1
Tubing Kit 0.25
PEEK, o.d. 1/16", i.d. 0.50 Orange High pressure tubing 2.0 μl
mm Tubing kit 0.5 (stan-
dard)
PEEK, o.d. 1/16", i.d. 0.75 Green High pressure tubing 4.4 μl
mm Tubing kit 0.75
PEEK, o.d. 1/16", i.d. 1.0 mm Beige High pressure tubing 7.8 μl
Tubing kit 1.0
FEP, o.d. 3/16", i.d. 2.9 mm Transparent Inlet tubing for high 66.0 μl
flow rate and high vis-
cosity
ETFE, o.d. 1/16", i.d. 0.75 Transparent Narrow inlet tubing 4.4 μl
mm (optional)
ETFE, o.d. 1/16", i.d. 1.0 mm Transparent Outlet and waste tub- 7.8 μl
ing
Note: • Many different sizes/types of tubing can be connected to a chromatography system. Tubing
with a smaller inner diameter (i.d.) holds less delay volume and will therefore generate less
dilution of the protein peak. Narrow tubing, however, increases the system pressure, espe-
cially when running at high flow rates. The tubing used should match the application needs.
See Section 9.4 Recommended tubing kits for prepacked columns, on page 428 for more
information.
• When using the high pressure tubing kit with i.d. 1.0 mm to allow high flow rates in combi-
nation with high viscosities in the pumps, inlet tubing with a larger i.d. than standard might
be needed to avoid outgassing.
Tubing connectors
The table below shows the tubing connectors used in ÄKTA pure.
Tubing connector 5/16" + ferrule (blue) FEP o.d. 3/16" i.d. 2.9 mm
3/16"
Other connectors
The table below shows other connectors used in ÄKTA pure.
Tubing labels
The illustration below shows the tubing labels for for a typical system configuration.
1A1 1S2
4
1S1
InA 1A2 2B
3S
1B1
1B2 2A 2S
A2 B2 InS S1
A1 B1 3
W1 W2
InB
6 7 8 9 W
Out1 Frac
Inlet tubing
The table below shows the labels, standard diameters, and standard lengths of the inlet tubing.
A1-A2 Inlets to Inlet valve FEP, o.d. 1/8", FEP, o.d. 3/16", 1500 3.0 9.9
and AB i.d. 1.6 mm i.d. 2.9 mm
B1-B2
A1-A7 Inlets to Inlet valve A FEP, o.d. 1/8", FEP, o.d. 3/16", 1500 3.0 9.9
i.d. 1.6 mm i.d. 2.9 mm
B1-B7 Inlets to Inlet valve B FEP, o.d. 1/8", FEP, o.d. 3/16", 1500 3.0 9.9
i.d. 1.6 mm i.d. 2.9 mm
InA From Inlet valve A or FEP, o.d. 1/8", FEP, o.d. 3/16", 300 0.6 2.0
Inlet valve AB to Sys- i.d. 1.6 mm i.d. 2.9 mm
tem pump A
InB From Inlet valve B or FEP, o.d. 1/8", FEP, o.d. 3/16", 300 0.6 2.0
Inlet valve AB to Sys- i.d. 1.6 mm i.d. 2.9 mm
tem pump B
S1-S7 Inlets to Sample inlet FEP, o.d. 1/8", FEP, o.d. 3/16", 1000 2.0 6.6
valve V9-IS or V9H-IS i.d. 1.6 mm i.d. 2.9 mm
Buff Inlet to Sample inlet FEP, o.d. 1/8", FEP, o.d. 3/16", 1000 2.0 6.6
valve V9-IS or V9H-IS i.d. 1.6 mm i.d. 2.9 mm
InS From Sample inlet FEP, o.d. 1/8", FEP, o.d. 3/16", 580 1.2 3.8
valve V9-IS or V9H-IS i.d. 1.6 mm i.d. 2.9 mm
to Sample pump S9
or S9H, respectively
Note: Narrow inlet tubing, ETFE, o.d. 1/16", i.d. 0.75 mm, is available for S1-S7.
1A1 System pump A left PEEK, o.d. PEEK, o.d. 340 150 270
to Restrictor A 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
1A2 System pump A right PEEK, o.d. PEEK, o.d. 340 150 270
to Restrictor A 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
1B1 System pump B left PEEK, o.d. PEEK, o.d. 340 150 270
to Restrictor B 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
1B2 System pump B right PEEK, o.d. PEEK, o.d. 340 150 270
to Restrictor B 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
1S1 Sample pump left to PEEK, o.d. PEEK, o.d. 340 150 270
Restrictor S 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
1S2 Sample pump right PEEK, o.d. PEEK, o.d. 340 150 270
to Restrictor S 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
Frac Outlet valve to Frac- PEEK, o.d. PEEK, o.d. 400 180 310
tion collector F9-R 1/16", i.d. 0.50 1/16", i.d. 0.75
mm mm
Note: Tubing of different inner diameter can be used for tubing Frac and tubing 5 to 9. An inner di-
ameter (i.d.) of 0.5 mm is standard, but tubing kits of 0.25 mm i.d., 0.75 mm i.d. and 1.0 mm
i.d. are also available. Tubing 3S is not included in the kits, but should be changed to 1.0 mm
i.d. if the Tubing kit 1.0 is used.
3-2 Mixer valve to Mixer PEEK, o.d. PEEK, o.d. 330 150 260
1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
3-3 Mixer valve to Injec- PEEK, o.d. PEEK, o.d. 260 120 200
tion valve 1/16", i.d. 0.75 1/16", i.d. 1.0
mm mm
Note: Tubing of different inner diameter can be used for tubing 8pH, 9pH, 1R and 2R. An inner diam-
eter (i.d.) of 0.5 mm is standard, but tubing kits of 0.25 mm i.d., 0.75 mm i.d. and 1.0 mm i.d.
are also available.
Reference capillary
The table below shows the label, diameter, and standard length of the reference capillary. The capillary
is used during the System performance tests.
Outlet tubing
The table below shows the labels, diameters, and standard lengths of the outlet tubing. The tubing is
not mounted on delivery.
Out Outlets from Outlet valve ETFE, o.d. 1/16", i.d. 1.0 mm 1500 1.2
V9-Os or V9H-Os
Out1 - Outlets from Outlet valve V9-O: ETFE, o.d. 1/16", i.d. 1.0 mm 1500 1.2
Out10 V9-O or V9H-O V9H-O: FEP o.d.1/8", i.d. 1.6 mm 3.0
Waste tubing
The table below shows the labels, diameters, and standard lengths of the waste tubing. The waste
tubing is mounted on delivery.
W1 System pump waste ETFE, o.d. 1/16", i.d. 1.0 mm 1500 1.2
W2 Sample pump waste ETFE, o.d. 1/16", i.d. 1.0 mm 1500 1.2
All HiTrap 5 ml 0.5 or 0.75 0.75 0.75 with Pressure control for
(including HisTrap, GSTrap except maximum flow rate combined
4B, StrepTrap) with high viscosity.
Mini Q™, Mini S™ 4.6/50 PE 0.25 or 0.5 0.25 or 0.5 Use 0.25 for maximum resolution.
Avoid column valve V9-C and pH
position in V9-pH.
Mono Q™, Mono S™, Mono P™ HR 0.5 0.5 Avoid combination of column
5/5 valve V9-C or V9H-C and pH posi-
tion in valve V9-pH or V9H-pH.
Resolution may be improved with
0.25.
Superdex 200 5/150 GL 0.25 0.25 Avoid column valve V9-C and pH
position in V9-pH.
HiLoad 16/60 Superdex 30 pg, 75 pg, 0.5 or 0.75 0.5 with Pressure
200 pg control or 0.75
HiLoad 26/60 Superdex 30 pg, 75 pg, 0.5 or 0.75 0.5 with Pressure
200 pg control or 0.75
In this section
Biocompatibility
The ÄKTA pure instrument is designed for maximum biocompatibility, with biochemically inert flow
paths constructed mainly from titanium, PEEK and highly resistant fluoropolymers and fluoroelastomers.
Titanium is used as far as possible to minimize contribution of potentially deactivating metal ions such
as iron, nickel and chromium. There is no standard stainless steel in the flow path. Plastics and rubber
materials are selected to avoid leakage of monomers, plasticizers or other additives.
Cleaning chemicals
Strong cleaning works well with 2 M sodium hydroxide, 70% acetic acid or the alcohols methanol,
ethanol and isopropyl alcohol. Complete system cleaning using 1 M hydrochloric acid should be
avoided in order to not damage the pressure sensors. If you are cleaning separation media using 1 M
hydrochloric acid, use loop injections of the acid and make sure that the column is not mounted on
the Column Valve V9-C. The Column Valve V9-C contains a pressure sensor which can be damaged
by 1 M hydrochloric acid.
If sodium hypochlorite is used as sanitizing agent instead of 2 M sodium hydroxide, use a concentration
up to 10%.
Organic solvents
Reversed phase chromatography of proteins works well with 100% acetonitrile and additives trifluo-
roacetic acid (TFA) up to 0.2% or formic acid up to 5%.
Strong organic solvents like ethyl acetate, 100% acetone or chlorinated organic solvents should be
avoided. These might cause swelling of plastic material and reduce the pressure tolerance of PEEK
tubing. For this reason, flash chromatography and straight (“normal”) phase chromatography is gen-
erally not recommended on the system
Assumptions made
The ratings are based on the following assumptions:
• Synergy effects of chemical mixtures have not been taken into account.
• Room temperature and limited overpressure is assumed.
Note: Chemical influences are time and pressure dependent. Unless otherwise stated, all concentra-
tions are 100%.
Introduction
This section provides detailed information about chemical resistance of the ÄKTA pure instrument to
some of the most commonly used chemicals in liquid chromatography. Regarding exposure to solutions
not covered by this information, contact your GE representative for recommendations.
Note: A user can be exposed to large volumes of chemical substances over a long time period. Ma-
terial Safety Data Sheets (MSDS) provide the user with information regarding characteristics,
human and environmental risks and preventive measures. Make sure that you have the MSDS
available from your chemical distributor and/or databases on the internet.
Aqueous buffers
The specified aqueous buffers are suitable for continuous use.
1 If hydrochloric acid, HCl, is used as a cleaning agent when columns are connected to the system, the HCl concen-
tration should not exceed 0.1 M in the pressure sensors. Remember that the ÄKTA pure system has pressure sensors
in the column valve V9-C.
For other parts of the system up to 1 M HCl is acceptable for short periods of use. See Cleaning chemicals, on page 432
1 Organic solvents can penetrate weaknesses in PEEK tubing walls more easily than water based buffers. Special
care should therefore be taken with prolonged use of organic solvents close to pressure limits.
Depending on pressure, tubing between pump head and pressure monitor needs to be changed. See Section 7.8.1
Replace tubing and connectors, on page 317 for more information.
2 Mobile phase system.
Note: It is recommended to replace the mixer sealing ring with the highly resistant O-ring (product
code 29-0113-26) if the system is to be exposed to organic solvents or high concentrations of
organic acids, such as acetic acid and formic acid, for a longer period of time.
Other substances
The following chemicals are suitable for continuous use.
PolyChloroTriFluoroEthylene PCTFE
PolyEtherEtherKetone PEEK
PolyPropylene PP
PolyTetraFluoroEthylene PTFE
UltraHighMolecularWeightPolyEthylene UHMWPE
Aluminum oxide
Elgiloy™
Hastelloy™ C-276
Quartz glass
Ruby
Sapphire
Titanium grade 2
Titanium grade 5
PolyEtherEtherKetone PEEK
PolyPropylene PP
Silicone
Manual Loop Fill The sample application loops of the loop valve
are manually filled with samples. The method
will guide the user through the process by
pausing and displaying on-screen instructions.
Up to five loops can be filled with different sam-
ples. The loops are filled in descending order.
Partial or complete loop fill can be chosen. Auto-
matic washing of loops and flow path are inte-
grated in the method.
The required components for this phase are a
column valve and a loop valve.
Method Settings The first, and mandatory, phase in any method. Defines common parameters
used in the subsequent phases.
The Method Settings phase defines:
1 Column type
Note:
The Column type list can be filtered in two steps:
• Select the chromatography technique to be used in the list Show by
technique
• Select Show only suggested columns to show the columns that are
suggested for the selected chromatography technique.
2 Pressure limit
3 Flow rate
4 Option to control the flow to avoid overpressure
Note:
Default values for pressure limits and flow rate are given for the selected column
type.
• Column position
• Flow restrictor use
• Unit selection for Method base and Flow rate
• Monitor settings:
- pH monitor
- Air sensor alarm settings
- UV monitor
Note:
The first wavelength of U9-M and the fixed wavelength for U9-L or U9-
L 2nd is always turned on. The second and third wavelengths for UV
monitor U9-M can be set on or off.
• Settings for Column Logbook
• Start Protocol
• Result name and location
Note:
Some of these options may not be required by certain methods.
Equilibration Equilibrates the column before purification, or re-equilibrates the column after
purification.
Sample Application Applies sample to the column. Defines the sample application technique, the
sample volume, and the handling of flowthrough.
Column Wash Washes out unbound sample after sample application or removes strongly
bound proteins after elution.
Elution Elutes the sample from the column. Defines parameters for the elution and
fractionation settings.
Column Prepara- Prepares the column before use by removing the storage solution and equili-
tion brating the column. By adding steps, several preparation solutions can be
used sequentially.
Column CIP Cleans the column after purification runs by rinsing the column with a
cleaning solution to remove nonspecifically bound proteins. By adding steps,
several cleaning solutions can be used sequentially.
System Prepara- Prepares the system before a run by removing storage solution and filling
tion the system and inlets with buffer solution. One preparation solution is used
per phase.
System CIP Cleans the system after purification runs by rinsing the system with a cleaning
solution. One cleaning solution is used per phase.
Column perfor- Tests the efficiency of a packed column in terms of height equivalent to a
mance test theoretical plate (HETP), and the peak asymmetry factor (As).
Manual Loop Fill Is used to manually fill the additional sample application loops mounted on
the loop valve. The filling options are:
• Partial loop fill
• Complete loop fill
Miscellaneous Can be added to any method at suitable places. The instructions can help
the user to better organize the graphical output of the results or introduce a
controlled delay in the method run.
The following subsections list the system settings available for ÄKTA pure.
Alarm UV Alarm UV enables or disables the alarm for the UV signal. When
enabled, it sets the alarm limits for the UV signal from UV monitor
U9-L. When the UV signal falls outside the set limits, an alarm will
be triggered and the method will be paused.
Alarm UV1 Alarm UV 1 enables or disables the alarm for the UV 1 signal from
UV monitor U9-M. When enabled, it sets the alarm limits for the
UV 1 signal. When the signal falls outside the set limits, an alarm
is issued and the method will be paused.
Note:
It is not possible to set an alarm signal for the UV 2 or UV 3 signals
from UV monitor U9-M.
Alarm UV 2nd Alarm UV 2nd enables or disables the alarm for the UV signal from
U9-L, 2nd. When enabled, it sets the alarm limits for the UV signal
from UV monitor U9-L, 2nd. When the signal falls outside the set
limits, an alarm will be triggered and the method will be paused.
Wavelength Wavelength sets the wavelengths for UV monitor U9-M. The cor-
responding curves in the chromatogram are marked with the set
wavelengths.
Note:
• The instruction is available when UV monitor U9-M (variable) is
selected in the component list. For best performance, do not
use more wavelengths than necessary for the application.
• At low wavelengths, the eluent can have significant absorbance
of its own.
Noise reduction UV Noise reduction UV filters the noise in the UV signal from U9-M
or U9-L (depending on the configuration). A column-specific aver-
aging time is set automatically when a column is defined in a
method run and Averaging time is set as a variable.
Noise reduction UV 2nd Noise reduction UV 2nd filters the noise in the UV signal from U9-
L, 2nd. A column-specific averaging time is set automatically when
a column is defined in a method run and Averaging time is set
as a variable.
Relative scale cond Relative scale cond facilitates monitoring of a gradient, for which
the user sets the conductivity values for 0% and 100%. The Rela-
tive scale cond can be set in ascending manner (0% for low and
100% for high conductivity) or in descending manner (0% for high
and 100% for low conductivity).
Note:
The Relative scale cond in descending manner is especially useful
for conductivity visualization in RPC and HIC, where the conductiv-
ity curve is reversed compared to the concentration curve (i.e., high
conductivity at 0% B and low conductivity at 100% B).
Cond temp condensation Cond temp compensation is used to adjust the conductivity values
to a reference temperature in order to compare conductivity val-
ues between runs that have been performed at different temper-
atures.
Setting the compensation factor to 0% turns this function off.
Alarm system pressure Alarm system pressure sets the alarm limits for the system
pressure. When enabled and the system pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm system pressure is set as a variable. Low alarm is
only triggered if the pressure first exceeds the Low alarm limit for
ten seconds continuously and then falls below the Low alarm
limit.
Note:
Setting the Low alarm or the system flow rate to 0 deactivates the
low pressure alarm.
Alarm sample pressure Alarm sample pressure sets the alarm limits for the sample
pressure. When enabled and the sample pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm sample pressure is set as a variable. Low alarm is
only triggered if the pressure first exceeds the Low alarm limit for
ten seconds continuously and then falls below the Low alarm
limit.
Note:
Setting the Low alarm or the system flow rate to 0 deactivates the
low pressure alarm.
Alarm delta column pressure Alarm delta column pressure sets the alarm limits for the delta
column pressure (pre-column pressure minus post-column pres-
sure). When enabled and the delta column pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Low alarm is only triggered if the pressure first
exceeds the Low alarm limit for ten seconds continuously then
falls below the Low alarm limit.
Note:
• Setting the Low alarm to 0 deactivates the low pressure alarm.
• Instruction Alarm delta column pressure is available only when
Column valve V9-C or V9H-C (5-columns) is selected in the
component list.
Alarm pre column pressure Alarm pre column pressure sets the alarm limits for the pre col-
umn pressure. When enabled and the pre column pressure falls
outside the set pressure limits, an alarm is issued and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm pre column pressure is set as a variable. Low alarm
is only triggered if the pressure first exceeds the Low alarm limit
for ten seconds continuously then falls below the Low alarm limit.
Note:
Setting the Low alarm to 0 deactivates the low pressure alarm.
Alarm inlet A air sensor Alarm inlet A air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet A. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm inlet B air sensor Alarm inlet B air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet B. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm inlet S air sensor Alarm inlet S air sensor enables or disables the air sensor alarm
for the built-in air sensor at the sample inlet. If the alarm is enabled
and air is detected, an alarm will be triggered and the method will
be paused.
Alarm external air sensor X Alarm external air sensor X enables or disables the alarm for the
optional air sensor number X. If the alarm is enabled and air is
detected, an alarm will be triggered and the method will be
paused.
Sensitivity inlet A air sensor Sensitivity inlet A air sensor is used together with the Alarm inlet
A air sensor instruction and sets the sensitivity of the built-in air
sensors at inlet A.
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
Sensitivity inlet B air sensor Sensitivity inlet B air sensor is used together with the Alarm inlet
B air sensor instruction and sets the sensitivity of the built-in air
sensors at inlet B.
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
Sensitivity inlet S air sensor Sensitivity inlet S air sensor is used together with the Alarm inlet
S air sensor instruction and sets the sensitivity of the built-in air
sensors at the sample inlet.
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
Sensitivity external air sensor X Sensitivity external air sensor X is used together with the Alarm
external air sensor X instruction and sets the sensitivity of the
optional air sensor number X. The optional air sensor can be locat-
ed either before any of the inlets A or B or after the injection valve.
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
- When located before an inlet, the default sensitivity is Normal.
- When located after the injection valve, the default sensitivity is
High and the pump currently pumping onto the column is used
for calculating the air volume for the external air sensor.
Note:
Using an air sensor after the injection valve is only useful when
running at lower pressures. High pressure dissolves any small air
bubbles present.
Digital out X Digital out X sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Noise reduction analog in X Noise reduction analog in X filters the noise in the analog signal
in port number X.
Alarm analog in X Alarm analog in X enables or disables the alarm for the analog
signal in port number X. When enabled, it sets the alarm limits for
the analog signal. If the alarm is enabled and the analog signal
falls outside the set limits, an alarm will be triggered and the
method will be paused.
Alarm digital in X Alarm digital in X enables or disables the alarm for the signal in
digital port number X. The alarm can be triggered by either of the
signal values, 0 or 1. If the alarm is enabled and the condition set
in ‘Value’ occurs, an alarm will be triggered and the method will
be paused.
Configure analog out X Configure analog out X enables the user to send one of the pre-
defined signals (UV signal, conductivity, temperature, pH or con-
centration of eluent B) to the analog out port number X, and also
to set the range of that signal.
Last tube filled Last tube filled: Action when last tube is filled (pause, direct the
flow to one of outlets or direct the flow to waste.
Fraction collector lamp Fraction collector lamp: Lamps in the fraction collector chamber
on or off.
Peak fractionation parameters The Peak fractionation parameters set the detection parameters
for peak collection, that is they decide when a peak starts and
ends. This information is used by the instructions Peak fractiona-
tion and Peak frac in outlet valve in order to start/end the peak
collection.
Fractionation settings Drop sync synchronises tube change to drop release. The available
settings are on or off. It is recommended to use Drop sync for
flow rates below 2 ml/min. Higher flow rates can however be used,
depending on the properties (for example viscosity) of the liquid.
Fractionation numbering mode Determines whether the fraction number is reset at the end of a
method or not.
Note:
The default setting is Reset.
Fractionation settings frac 2 Drop sync on or off. It is recommended to use this setting for flow
rates below 2 ml/min. Higher flow rates can be however be used,
depending on the properties (e.g. viscosity) of the liquid.
Fractionation numbering mode Determines whether fraction number for the second fraction col-
frac 2 lector is reset at the end of a method or not.
Note:
The default setting is Reset.
Peak fractionation parameters Peak fractionation parameters sets the detection parameters
for peak collection, i.e. it determines when a peak starts and ends.
This information is used by the instructions Peak fractionation,
Peak fractionation frac 2 and Peak frac in outlet valve in order
to start/end the peak collection.
Tubing: Injection valve to col- The instruction is used in calculations of pre-column pressure
umn when Column valve V9-C or V9H-C (5-columns) is not mounted
onto the system. In such cases, there is no pre-column pressure
sensor present. The estimated pre-column pressure is either cal-
culated by using the measured system pressure and the length
of the tubing between the system pressure sensor and the column
valve or by using the measured sample pressure and the length
of the tubing between the sample pressure sensor and the column
valve depending on the position of the injection valve.
The instruction is available only when Column valve V9-C or V9H-
C (5-columns) is not selected in the component list.
Tubing: Sample pump to injec- The instruction is used in calculations of pre-column pressure
tion valve when Column valve V9-C or V9H-C (5-columns) is not mounted
onto the system. In such cases, there is no pre-column pressure
sensor present. The estimated pre-column pressure is either cal-
culated by using the measured system pressure and the length
of the tubing between the system pressure sensor and the column
valve or by using the measured sample pressure and the length
of the tubing between the sample pressure sensor and the column
valve depending on the position of the injection valve.
The instruction is available only when Column valve V9-C or V9H-
C (5-columns) is not selected and a sample pump (S9 or S9H) is
selected in the component list.
Delay volume: pH valve Delay volume: pH valve is used to calculate the delay volume
between the monitor and the Outlet valve. The instruction is used
to make sure that the collected fractions correspond to the frac-
tions indicated in the chromatogram. It is recommended not to
alter the default values for restrictor and pH cell delay volumes
when standard modules and standard tubing for flow restrictor
are used.
Delay volume: Monitor to outlet Delay volume: Monitor to outlet valve is used to define the delay
valve volume between the monitor and the Outlet valve. The instruction
is used to make sure that the collected fractions correspond to
the fractions indicated in the chromatogram. The delay volume
must be changed when changing tubing to another I.D. or length
or when removing or adding components.
Delay volume: Monitor to frac Delay volume: Monitor to frac is used to define the delay volume
between the monitor and the Fraction collector. The instruction
is used to make sure that the collected fractions correspond to
the fractions indicated in the chromatogram. The instruction is
available only when the Fraction collector is selected in the com-
ponent list.
Delay volume: Monitor to frac 2 Delay volume: Monitor to frac 2 is used to define the delay volume
between the monitor and the second Fraction collector. The in-
struction is used to make sure that the collected fractions corre-
spond to the fractions indicated in the chromatogram. The instruc-
tion is available only when the second Fraction collector is selected
in the component list.
System wash settings System wash settings sets the flow rate used for System wash.
Note:
• The volume for system wash is set in the System wash instruc-
tion.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
• Adjust the flow rate during the system wash so that the system
pressure does not exceed 2 MPa.
System pump wash settings System pump wash settings sets the flow rate and the wash
volume used during system pump washes.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Sample pump wash settings Sample pump wash settings sets the flow rate and the wash
volume used during sample pump washes.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Fraction collector wash settings Fraction collector wash settings sets the flow rate during Frac-
tion collector wash.
Mixer by-pass wash settings Mixer by-pass wash settings sets the flow rate used during Mixer
by-pass wash and defines wash volumes for mixer by-pass wash
options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Loop wash settings Loop wash settings sets the flow rate used during Loop wash.
Note:
• The volume for system wash is set in the Loop wash instruction.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Watch UV parameters Watch UV parameters sets the accepted signal fluctuation and
Delta peak limit of the UV signal for some of the tests in the Watch
and Hold until instructions.
Watch UV 2nd parameters Watch UV 2nd parameters sets the accepted signal fluctuation
and Delta peak limit of the UV 2nd signal for some of the tests in
the Watch and Hold until instructions.
Watch cond parameters Watch cond parameters sets the accepted fluctuation and Delta
peak limit of the conductivity signal for some of the tests in the
Watch and Hold until instructions.
Watch pH parameters Watch pH parameters sets the value for the accepted fluctuation
of the pH signal used for the test Stable signal in the instructions
Watch and Hold until.
Watch flow parameters Watch flow parameters sets the value for the accepted fluctua-
tion of the flow rate signal used for the test Stable signal in the
instructions Watch and Hold until with signal System flow.
Watch pressure parameters Watch pressure parameters sets the value for the accepted
fluctuation of the pressure signals used for the test Stable signal
in the instructions Watch and Hold until.
Watch analog in parameters Watch analog in parameters sets the accepted signal fluctuation
and Delta peak limit of the analog signal for some of the tests in
the Watch and Hold until instructions.
Power-save Power-save sets the instrument into power saving mode. When
the function is enabled, the instrument enters power-saving mode
after having been in state Ready for a certain time period. The
instrument turns into state Ready when a method run, a method
queue or a manual run ends. The time interval before the instru-
ment enters power-saving mode is defined by the user.
Instrument control panel Instrument control panel locks/unlocks the control panel located
on the front side of the instrument. When unlocked, the buttons
on the Instrument control panel are active and can be used to
control a few basic functions of the instrument. When the Instru-
ment control panel is locked, no functions are available.
Pressure control parameters By using Pressure control the method can be run with the set
flow rate without the risk of method stop due to pressure alarm.
Pressure control is enabled in the instruction System flow or
Sample flow. Pressure control parameters provides the P and I
factors used in the regulator and can be adjusted for different
columns.
Constant pressure flow param- Constant pressure flow parameters sets the values for the P and
eters I factors needed to keep a constant pressure by varying the flow
rate. The signal used for pressure control is set in the instruction
Constant pressure flow.
Max flow during valve turn Max flow during valve turn sets the maximum flow rate used
during the turning of the injection and outlet valve in order to
avoid high pressure alarms.
Method progressing flow Method progressing flow sets which flow (automatic, system
flow, sample flow) the progress of the method is calculated after.
In automatic mode, the position of the injection valve determines
if the system flow or the sample flow is used.
Sample pump setting Sample pump setting enables sample pump flow while the injec-
tion valve is in manual load position.
Data collection The Data collection settings determine the maximum number of
data points collected for a given curve. Data reduction occurs if
the maximum number of data points is exceeded. To avoid data
reduction, set the maximum number of data points to be collected
to 180000 or insert a New Chromatogram instruction in the
method.
Note:
The default setting is 54000 data points, which corresponds to 1.5
h for a signal of 10 Hz.
3 To execute several instructions at the same breakpoint, select and edit an instruction
and click Insert. Repeat for several instructions.
4 To update parameter fields during method run, check the Auto update... box.
Sample flow Sample flow defines the system flow rate. Flow rate can be set
either as volumetric or as linear flow. A column type must be se-
lected before using linear flow.
Pump A wash Pump A wash is used to change buffers in the inlet tubing, pump
and mixer.
Note:
• Pressing End during Pump wash will terminate both the wash
and the run immediately.
Pressing Continue during Pump wash will terminate the wash
and the run will continue from the point at which the pump
wash instruction was executed.
• An instruction issued when a Pump wash is in progress will not
be executed until the wash is completely finished and all valves
have turned back to the previous positions.
• Pump wash cannot be executed when the system is in state
HOLD.
Pump B wash Pump B wash is used to change buffers in the inlet tubing, pump
and mixer.
Note:
• Pressing End during Pump wash will terminate both the wash
and the run immediately.
Pressing Continue during Pump wash will terminate the wash
and the run will continue from the point at which the Pump
wash instruction was executed.
• An instruction issued when a Pump wash is in progress will not
be executed until the wash is completely finished and all valves
have turned back to the previous positions.
• Pump wash cannot be executed when the system is in state
HOLD.
Sample pump wash Sample pump wash is used to change buffers in the sample inlet
tubing and sample pump.
Note:
• Pressing End during Sample pump wash will terminate both
the wash and the run immediately.
Pressing Continue during Sample pump wash will terminate
the wash and the run will continue from the point at which the
Sample pump wash instruction was executed.
• An instruction issued when a Sample pump wash is in progress
will not be executed until the wash is completely finished and
all valves have turned back to the previous positions.
• Sample pump wash cannot be executed when the system is
in state HOLD.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Mixer by-pass wash Mixer by-pass wash is used to wash or fill the flow path (mixer
valve and tubing) that is used for sample application with system
pump. It is possible to choose a more extensive or a less extensive
wash option.
Note:
• Pressing End during Mixer by-pass wash will terminate both
the wash and the run immediately.
Pressing Continue during Mixer by-pass wash will terminate
the wash and the run will continue from the point at which the
Mixer by-pass wash instruction was executed.
• An instruction issued during a Mixer by-pass wash will not be
executed until the wash is completely finished and the valves
have turned back to the previous positions.
• Mixer by-pass wash cannot be executed when the system is
in state HOLD.
Loop wash Loop wash is used to wash the Loop valve. It is possible to wash
a single sample application loop, all loops or only the by-pass
position. Loop wash instruction is not available, if a Column valve
is not mounted on the system.
Note:
• Pressing End during Loop wash will terminate both the wash
and the run immediately.
Pressing Continue during Loop wash will terminate the wash
and the run will continue from the point at which the Loop wash
instruction was executed.
• An instruction issued when a Loop wash is in progress will not
be executed until the wash is completely finished and the valves
have turned back to the previous positions.
• Loop wash cannot be executed when the system is in state
HOLD.
System wash System wash is used to fill the system with the selected buffer
composition. The flow can be directed to the waste position of
either the injection valve or the outlet valve. The flow is directed
to the end of the flow path if outlet valve is not present.
Note:
• Pressing End during System wash will terminate both the wash
and the run immediately.
Pressing Continue during System wash will terminate the wash
and the run will continue from the point at which the System
wash instruction was executed.
• If System wash is performed during a Gradient operation, the
current component B concentration is maintained during the
wash.
• An instruction issued during a system wash operation cannot
be executed until the wash is completely finished and all valves
have turned back to the previous positions.
• System wash cannot be executed when the system is in state
HOLD.
• Adjust the flow rate during the system wash so that the system
pressure does not exceed 2 MPa.
Sample inlet Sample inlet turns the Sample inlet valve to the selected position.
Injection valve Injection valve sets the Injection valve to the selected position.
The instruction gives an injection mark in the chromatogram when
the inlet valve switches to Inject or Direct Inject.
Note:
Sample flow refers to the flow that enters the injection valve via
SaP port.
Column position Column position turns the Column valve to the position specified
in the parameter Position.
pH valve pH valve sets the pH cell and the 0.2 MPa restrictor in positions
inline or offline.
The pH valve also has a calibration position. This position is only
available when performing calibration of the pH monitor (In Sys-
tem control select System:Calibrate).The calibration position can
also be used to fill the pH cell with storage solution since the pH
valve is in open position.
Note:
• It is not possible to turn the pH valve during any type of fraction-
ation as it affects the delay volume.
• The pH valve instruction can be given during the delay volume
of the different stop fractionation instructions, but it is executed
only after the set delay volume has been collected.
Outlet valve Outlet valve turns the Outlet valve to the selected position. The
instruction gives a mark in the chromatogram when the valve is
switched to the selected position.
Loop valve Loop valve turns the Loop valve to the selected position.
Mixer valve Mixer valve turns the Mixer valve to the selected position.
Note:
Setting Mixer valve in by-pass position makes it possible to use
System pump A for direct loading of the sample onto the column.
Versatile valve X Versatile valve X turns the Versatile valve number X to the selected
position.
Note:
Four sets of positions are available. In positions 1-3 and 2-4 only a
single flow channel can be used. In positions 1-4 & 2-3 and 1-2 &
3-4 the flow can be directed through two channels simultaneously.
Inlet valve (X1) Instruction Inlet valve X1 turns the extra valve to the selected
position. The extra valve is a basic 8 port valve, without air sensor,
to be used for general applications.
Inlet valve (X2) Instruction Inlet valve X2 turns the extra valve to the selected
position. The extra valve is a basic 8 port valve, without air sensor,
to be used for general applications.
Auto zero UV Auto zero UV sets the UV signals from U9-M or U9-L to 0 mAU.
Auto zero UV 2nd Auto zero UV 2nd sets the UV signal from UL-9, 2nd to 0 mAU.
Wavelength Wavelength sets the wavelengths for UV monitor U9-M. The cor-
responding curves in the chromatogram are marked with the set
wavelengths.
Note:
• The instruction is available when UV monitor U9-M (variable) is
selected in the component list. For best performance, do not
use more wavelengths than necessary for the application.
• At low wavelengths, the eluent can have absorbance of its own.
Noise reduction UV Noise reduction UV filters the noise in the UV signal from UV
monitor U9-M or U9-L. A column-specific averaging time is set
automatically when a column is defined in a method run and Av-
eraging time is set as a variable.
Noise reduction UV 2nd Noise reduction UV 2nd filters the noise in the UV monitor signal
from U9-L, 2nd. A column-specific averaging time is set automat-
ically when a column is defined in a method run and Averaging
time is set as a variable.
UV lamp Sets the UV lamp ON or OFF. Default is ON. The UV lamp is turned
ON when the system changes state to RUN, HOLD, or WASH.
UV lamp 2nd Sets the UV lamp 2nd ON or OFF. Default is ON. The UV lamp 2nd
is turned ON when the system changes state to RUN, HOLD, or
WASH.
Relative scale cond Relative scale cond facilitates monitoring of a gradient, for which
the user sets the conductivity values for 0% and 100%. The Rela-
tive scale cond can be set in ascending manner (0% for low and
100% for high conductivity) or in descending manner (0% for high
and 100% for low conductivity).
Note:
The Relative scale cond in descending manner is especially useful
for conductivity visualization in RPC and HIC, where the conductiv-
ity curve is reversed compared to the concentration curve (i.e., high
conductivity at 0% B and low conductivity at 100% B).
Stop fractionation Stop fractionation ends the fractionation after the set delay vol-
ume (specified in System Settings:Tubing and Delay volumes)
has been collected. The outlet valve is then turned to position
Waste.
Note:
If Stop fractionation is issued when both Fractionation and Peak
fractionation are active, fractionation is stopped after the set delay
volume has been collected. The outlet valve remains in position
Frac and peak fractionation continues.
Peak fractionation Peak fractionation enables collection of only those peaks that
fulfill the conditions set in the Peak fractionation parameters in-
struction.
Stop peak fractionation Stop peak fractionation ends the peak fractionation after the set
delay volume (specified in System Settings:Tubing and Delay
volumes) has been collected. The outlet valve is then turned to
position Waste.
Last tube filled Only for Fraction collector F9-C. Last tube filled sets the action
to perform after the fraction collector fills the last tube of the run:
pause the fractionation, direct the flow to one of the outlet ports
of the outlet valve or direct the flow to waste.
Reset frac number Only for Fraction collector F9-R. Sets fraction numbers to restart
from 1. The restart occurs when the instruction is issued. The in-
struction overrides the continuous numbering mode if Fractiona-
tion numbering mode is set to Continue in System Settings.
Feed tube Fraction collector F9-C: Feed tube moves the fractionation arm
to the position specified by the parameter Start position, after
the set delay volume has been collected. A fraction mark is given
in the chromatogram.
Fraction collector F9-R: Feed tube moves the tube rack forward
one tube after the set delay volume has been collected and a
fraction mark is set. When fractionation or peak fractionation is
not ongoing, Feed tube moves the rack instantly and no fraction
mark is set.
Fractionation numbering mode Only for Fraction collector F9-R. Fractionation numbering mode
determines whether the fraction number is reset at the end of a
method or not.
Note:
The default setting is Reset.
Fraction collector wash Only for Fraction collector F9-C. Fraction collector wash is used
to wash the fraction collector with the current solution present in
the system. The wash flow rate is set in the instruction Wash
settings:Fraction collector wash settings and the current inlet
positions are used. After the wash, the flow rate and the valve
positions automatically go back to their previous settings.
Note:
Fraction collector wash cannot be executed during any type of
fractionation.
Reset tube type Only for Fraction collector F9-C. Reset tube type resets all the
tube types in the fraction collector.
Note:
It is not allowed to execute the Reset tube type instruction during
fractionation.
Frac cleaning position Only for Fraction collector F9-C. Frac cleaning position enables
manual cleaning of the dispenser head. The system is paused and
the fractionation arm is moved to the middle front of the interior
of the fraction collector. It is then possible to open the door of the
fraction colelctor and manually clean the dispenser head.
Note:
The Frac cleaning position instruction cannot be executed during
fractionation.
Fraction collector lamp Only for Fraction collector F9-C. Fraction collector lamp turns the
light in the fraction collector on or off.
Cassette configuration Only for Fraction collector F9-C. Cassette configuration is set to
either automatic or manual:
• Automatic: the fraction collector automatically detects the
cassette types present in the fraction collector.
• Manual: The fraction collector content is manually set.
Fractionation frac 2 Fractionation frac 2 is used when collecting fractions with the
second Fraction collector.
Stop fractionation frac 2 Stop fractionation frac 2 ends the fractionation after the set delay
volume for the second Fraction collector (specified in System
Settings:Tubing and Delay volumes) has been collected. The
outlet valve is then turned to position Waste.
Note:
If Stop fractionation frac 2 is issued when both Fractionation
frac 2 and Peak fractionation frac 2 are active, fractionation is
stopped after the set delay volume has been collected. The Outlet
valve V9-O or V9H-O remains in position Outlet 10 / Frac 2 (or
Outlet 1 if Outlet valve V9-Os is mounted on the instrument) and
peak fractionation in the second Fraction collector continues.
Peak fractionation frac 2 Peak fractionation frac 2 enables collection of only those peaks
that fulfill the conditions set in the Peak fractionation parameters
instruction.
Stop peak fractionation frac 2 Stop peak fractionation frac 2 ends the peak fractionation in
second Fraction collector after the set delay volume (specified in
System Settings:Tubing and Delay volumes) has been collected.
The outlet valve is then turned to position Waste.
Reset frac number frac 2 Sets fraction numbers to restart from 1 for the second Fraction
collector. The restart occurs when the instruction is issued. The
instruction overrides the continuous numbering mode if Fraction-
ation numbering mode frac 2 is set to Continue in System Set-
tings.
Feed tube frac 2 During fractionation or peak fractionation the instruction Feed
tube frac 2 moves the second Fraction collector tube rack forward
one tube after the set delay volume has been collected and a
fraction mark is set. When fractionation or peak fractionation is
not ongoing, Feed tube frac 2 moves the rack instantly and no
fraction mark is set.
Stop frac in outlet valve Applicable if no fraction collector is used. Stop frac in outlet valve
ends the fractionation in outlet valve after the set delay volume
(specified in System Settings:Tubing and Delay volumes) has
been collected. The outlet valve is then turned to position Waste.
Peak frac in outlet valve Applicable if no fraction collector is used. Peak frac in outlet valve
enables collection of only those peaks that fulfill the conditions
set in Peak fractionation parameters. When the set fraction
size/outlet has been collected, the outlet valve turns to the next
position. A fraction mark is set in the chromatogram for each new
outlet position. If Outlet valve V9-Os or V9H-Os (1-outlet) is used,
only one peak can be collected via Outlet 1 position.
Stop peak frac in outlet valve Applicable if no fraction collector is used. Stop peak frac in outlet
valve ends the peak fractionation in outlet valve after the set delay
volume (specified in System Settings:Tubing and Delay volumes)
has been collected. The outlet valve is then turned to position
Waste.
Peak fractionation parameters Peak fractionation parameters sets the detection parameters
for peak collection, i.e. it determines when a peak starts and ends.
This information is used by the instructions Peak fractionation,
Peak fractionation frac 2 and Peak frac in outlet valve in order
to start/end the peak collection.
Auto zero analog in X Auto zero analog in X sets the value of the analog signal in the
analog port number X to 0 mV.
Reset auto zero analog in X Reset auto zero analog in X sets the signal in analog port number
X to its current value, i.e. the actual voltage in the analog port
number X.
Noise reduction analog in X Noise reduction analog in X filters the noise in the analog signal
in port number X.
Digital out X Digital out X sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Pulse digital out X Pulse digital out X generates a pulsed signal in digital port number
X. The signal changes from the initial state (0 or 1) to the opposite
state and returns to the initial state after the defined length of
time.
Configure analog out X Configure analog out X enables the user to send one of the pre-
defined signals (UV signal, conductivity, temperature, pH or con-
centration of eluent B) to the analog out port number X, and also
to set the range of that signal.
Alarm system pressure Alarm system pressure sets the alarm limits for the system
pressure. When enabled and the system pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm system pressure is set as a variable. Low alarm is
only triggered if the pressure first exceeds the Low alarm limit for
ten seconds continuously and then falls below the Low alarm
limit.
Note:
Setting the Low alarm or the system flow rate to 0 deactivates the
low pressure alarm.
Alarm sample pressure Alarm sample pressure sets the alarm limits for the sample
pressure. When enabled and the pressure falls outside the set
pressure limits, an alarm will be triggered and the method will be
paused. Low alarm is only triggered if the pressure first exceeds
the Low alarm limit for ten seconds continuously then falls below
the Low alarm limit.
Note:
Setting the Low alarm to 0 deactivates the low pressure alarm.
Alarm delta column pressure Alarm delta column pressure sets the alarm limits for the delta
column pressure (pre-column pressure minus post-column pres-
sure). When enabled and the delta column pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Low alarm is only triggered if the pressure first
exceeds the Low alarm limit for ten seconds continuously then
falls below the Low alarm limit.
Note:
• Setting the Low alarm to 0 deactivates the low pressure alarm.
• Instruction Alarm delta column pressure is available only when
Column valve V9-C or V9H-C (5-columns) is selected in the
component list.
Alarm pre column pressure Alarm pre column pressure sets the alarm limits for the pre col-
umn pressure. When enabled and the pre column pressure falls
outside the set pressure limits, an alarm is issued and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm pre column pressure is set as a variable. Low alarm
is only triggered if the pressure first exceeds the Low alarm limit
for ten seconds continuously then falls below the Low alarm limit.
Note:
Setting the Low alarm to 0 deactivates the low pressure alarm.
Alarm UV Alarm UV enables or disables the alarm for the UV signal. When
enabled, it sets the alarm limits for the UV signal from UV monitor
U9-L. When the UV signal falls outside the set limits, an alarm will
be triggered and the method will be paused.
Alarm UV1 Alarm UV 1 enables or disables the alarm for the UV 1 signal from
UV monitor U9-M. When enabled, it sets the alarm limits for the
UV 1 signal from UV monitor U9-M. When the UV signal falls out-
side the set limits, an alarm is issued and the method will be
paused.
Note:
It is not possible to set an alarm signal for the UV 2 or UV 3 signals
from UV monitor U9-M.
Alarm UV 2nd Alarm UV 2nd enables or disables the alarm for the UV signal from
UV monitor U9-L, 2nd. When enabled, it sets the alarm limits for
the UV signal from U9-L, 2nd. When the UV signal falls outside the
set limits, an alarm will be triggered and the method will be
paused.
Alarm inlet A air sensor Alarm inlet A air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet A. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm inlet B air sensor Alarm inlet B air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet B. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm sample inlet air sensor Alarm sample inlet air sensor enables or disables the air sensor
alarm for the built-in air sensor at the sample inlet. If the alarm is
enabled and air is detected, an alarm will be triggered and the
method will be paused.
Alarm external air sensor X Alarm external air sensor X enables or disables the alarm for the
optional air sensor number X. If the alarm is enabled and air is
detected, an alarm will be triggered and the method will be
paused.
Alarm analog in X Alarm analog in X enables or disables the alarm for the analog
signal in port number X. When enabled, it sets the alarm limits for
the analog signal. If the alarm is enabled and the analog signal
falls outside the set limits, an alarm will be triggered and the
method will be paused.
Alarm digital in X Alarm digital in X enables or disables the alarm for the signal in
digital port number X. The alarm can be triggered by either of the
signal values, 0 or 1. If the alarm is enabled and the condition set
in ‘Value’ occurs, an alarm will be triggered and the method will
be paused.
System wash settings System wash settings sets the flow rate used for System wash.
Note:
• The volume for system wash is set in the System wash instruc-
tion.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
• Adjust the flow rate during the system wash so that the system
pressure does not exceed 2 MPa.
System pump wash settings System pump wash settings sets the flow rate and wash volume
used during system pump washes.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Sample pump wash settings Sample pump wash settings sets the flow rate and wash volume
used during sample pump washes.
Fraction collector wash settings Fraction collector wash settings sets the flow rate during Frac-
tion collector wash.
Mixer by-pass wash settings Mixer by-pass wash settings sets the flow rate used during Mixer
by-pass wash and defines wash volumes for mixer by-pass wash
options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Loop wash settings Loop wash settings sets the flow rate used during Loop wash.
Note:
• The volume for system wash is set in the Loop wash instruction.
• The flow rate should not exceed 10 ml/min if narrow inlet tubing
(I.D. 0.75 mm) is used.
Watch UV parameters Watch UV parameters sets the accepted signal fluctuation and
Delta peak limit of the UV signal for some of the tests in the Watch
and Hold until instructions.
Watch UV 2nd parameters Watch UV 2nd parameters sets the accepted signal fluctuation
and Delta peak limit of the UV 2nd signal for some of the tests in
the Watch and Hold until instructions.
Watch cond parameters Watch cond parameters sets the accepted fluctuation and Delta
peak limit of the conductivity signal for some of the tests in the
Watch and Hold until instructions.
Watch pH parameters Watch pH parameters sets the value for the accepted fluctuation
of the pH signal used for the test Stable signal in the instructions
Watch and Hold until.
Watch flow parameters Watch flow parameters sets the value for the accepted fluctua-
tion of the flow rate signal used for the test Stable signal in the
instructions Watch and Hold until with signal System flow.
Watch pressure parameters Watch pressure parameters sets the value for the accepted
fluctuation of the pressure signals used for the test Stable signal
in the instructions Watch and Hold until.
Watch analog in parameters Watch analog in parameters sets the accepted signal fluctuation
and Delta peak limit of the analog signal for some of the tests in
the Watch and Hold until instructions.
Pressure control parameters By using Pressure control the method can be run with the set
flow rate without the risk of method stop due to pressure alarm.
Pressure control is enabled in the instruction System flow or
Sample flow. Pressure control parameters provides the P and I
factors used in the regulator and can be adjusted for different
columns.
Pressure control min flow rate can be set either as volumetric or
as linear flow. A column type must be selected before using linear
flow.
Constant pressure flow Constant pressure flow enables column packing at constant
pressure. The system pump automatically adjusts the flow rate
within the specified Minimum allowed flow rate – Maximum al-
lowed flow rate range. The goal is to reach and keep the set
Pressure at the selected Pressure sensor using the P and I factors
set in the Constant pressure flow parameters instruction. The
total volume is continuously updated using the actual flow rate.
Both pressure control flow rates can be set either as volumetric
or as linear flow. A column type must be selected before using
linear flow.
Note:
• When Constant pressure flow is used, the P and I factors set
in the Constant pressure flow parameters instruction are used
to control the pressure, instead of the P and I values set in the
Pressure control parameters instruction.
• Pressure sensor Delta column pressure is available only when
column valve V9-C or V9H-C (5-columns) is selected in the
component list.
Constant pressure flow param- Constant pressure flow parameters sets the values for the P and
eters I factors needed to keep a constant pressure by varying the flow
rate. The signal used for pressure control is set in the instruction
Constant pressure flow.
Column packing flow Column packing flow is used to set flow rates over 25 ml/min and
150 ml/min for ÄKTA pure 25 and ÄKTA pure 150, respectively.
Both A and B pumps are used to generate the flow, making it
possible to set flow rates up to 50 ml/min and 300 ml/min for
ÄKTA pure 25 and ÄKTA pure 150, respectively. Flow rate can be
set either as volumetric or as linear flow. A column type must be
selected before using linear flow.
Before executing the Column packing flow instruction it is impor-
tant to:
• Immerse inlet tubing A1 and B1 in the same buffer
• Disconnect the column outlet tubing from the Column valve
and place the tubing in a waste vessel
Note:
When running Column packing flow only isocratic runs can be
performed, gradients cannot be generated.
Delay volume: Monitor to outlet Delay volume: Monitor to outlet valve is used to define the delay
valve volume between the monitor and the Outlet valve. The instruction
is used to make sure that the collected fractions correspond to
the fractions indicated in the chromatogram. The delay volume
must be changed when changing tubing to another i.d. or length
or when removing or adding components.
Delay volume: Monitor to frac Delay volume: Monitor to frac is used to define the delay volume
between the monitor and the Fraction collector. The instruction
is used to make sure that the collected fractions correspond to
the fractions indicated in the chromatogram. The instruction is
available only when the Fraction collector is selected in the com-
ponent list.
Delay volume: Monitor to frac 2 Delay volume: Monitor to frac 2 is used to define the delay volume
between the monitor and the second Fraction collector. The in-
struction is used to make sure that the collected fractions corre-
spond to the fractions indicated in the chromatogram. The instruc-
tion is available only when the second Fraction collector is selected
in the component list.
Start volume count Start volume count starts the volume counter function. The
counted volume is saved into a memory.
This instruction is best used in combination with Watch instruc-
tions.
Stop volume count Stop volume count stops the volume counter function. The
counted volume is stored in the memory and can be recalled with
the instruction Hold counted volume. The counted volume can
also be recalled in following runs and is stored until a new Stop
volume count instruction is issued.
This instruction is best used in combination with Watch instruc-
tions.
Hold counted volume Hold counted volume sets the system to Hold. The system will
remain in the state Hold until the accumulated volume reaches
the volume stored by the instructions Start volume count / Stop
volume count.
Set mark Set mark inserts a mark into the current chromatogram with the
text entered for the parameter Mark text .
Timer Timer sets the system to pause or end after a set volume or time
has passed. Select base sets the base to either accumulated time
or accumulated volume. Timeout sets the volume or time. Action
sets the action to perform (pause or end)
System flow 0.001 – 50.000 ml/min The set flow rate of the system pumps.
(ÄKTA pure 25)
0.01 – 300.00 ml/min (ÄKTA
pure 150)
System linear flow cm/h The set flow velocity of the system pumps.
Only available if a column is selected.
Sample flow 0.001 – 50.000 ml/min The set flow rate of the sample pump.
(ÄKTA pure 25)
0.01 – 300.00 ml/min (ÄKTA
pure 150)
Sample linear flow cm/h The set flow velocity of the sample pump.
Only available if a column is selected.
Inlet S S1 - S7, buff The set position of the sample inlet valve.
Column flow direction N/A The set flow direction position of the Column
valve V9-C, Column valve V9H-C, Column
valve V9-Cs and V9H-Cs.
System pressure -1.00 – 20.00 MPa (ÄKTA The system pressure signal (at the system
pure 25) pumps).
-1.00 – 5.00 MPa (ÄKTA
pure 150)
Sample pressure -1.00 – 10.00 MPa (ÄKTA The sample pressure signal (at the sample
pure 25) pump).
-1.00 – 5.00 MPa (ÄKTA
pure 150)
PreC pressure -1.00 – 20.00 MPa (ÄKTA The pre-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA
pure 150)
DeltaC pressure -1.00 – 20.00 MPa (ÄKTA The delta-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA
pure 150)
PostC pressure -1.00 – 20.00 MPa (ÄKTA The post-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA
pure 150)
UV 1 -6000.000 – 6000.000 mAU The first UV/Vis absorbance signal of the U9-
M monitor.
UV 2 -6000.000 – 6000.000 mAU The 2nd UV/Vis absorbance signal of the U9-
M monitor.
UV 3 -6000.000 – 6000.000 mAU The 3rd UV/Vis absorbance signal of the U9-
M monitor.
UV 2nd -6000.000 – 6000.000 mAU The UV absorbance signal of the second U9-
L monitor.
Cond temp 0.0 – 99.0 °C The temperature signal (in the conductivity
flow cell).
Frac position N/A The current tube position of the fraction col-
lector.
Frac 2 position N/A The current tube position of the fraction col-
lector 2.
Air inlet A No air, Air The current state of the air alarm for the inte-
grated air sensor in inlet valve A.
Air inlet B No air, Air The current state of the air alarm for the inte-
grated air sensor in inlet valve B.
Air inlet S No air, Air The current state of the air alarm for the inte-
grated air sensor in inlet valve IS.
Ext. air sensor, to No air, Air The current state of the air alarm for the ex-
Ext. air sensor 4 ternal air sensors.
System flow 0.001 – 50.000 ml/min (ÄKTA 1 Hz The set flow rate of the system
pure 25) pumps.
0.01 – 300.00 ml/min (ÄKTA
pure 150)
Sample flow 0.001 – 50.000 ml/min (ÄKTA 1 Hz The set flow rate of the sample
pure 25) pump.
0.01 – 300.00 ml/min (ÄKTA
pure 150)
PreC pressure -1.00 – 20.00 MPa (ÄKTA 1 Hz The pre-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA pure
150)
DeltaC pressure -1.00 – 20.00 MPa (ÄKTA 1 Hz The delta-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA pure
150)
System pressure -1.00 – 20.00 MPa (ÄKTA 10 Hz The system pressure signal (at the
pure 25) system pumps).
-1.00 – 5.00 MPa (ÄKTA pure
150)
Sample pressure -1.00 – 10.00 MPa (ÄKTA 10 Hz The sample pressure signal (at the
pure 25) sample pump).
-1.00 – 5.00 MPa (ÄKTA pure
150)
PostC pressure -1.00 – 20.00 MPa (ÄKTA 1 Hz The post-column pressure signal.
pure 25)
-1.00 – 5.00 MPa (ÄKTA pure
150)
Cond temp 0.0 – 99.0 °C 0.5 Hz The temperature signal (in the con-
ductivity flow cell).
UV cell path 0.2, 0.5 cm 1 Hz The nominal cell path length of the
length U9-L monitor.
UV cell path 0.05, 0.2, 1.0 cm 1 Hz The nominal cell path length of the
length U9-M monitor.
UV cell path 0.2, 0.5 cm 1 Hz The nominal cell path length of the
length 2nd 2nd U9-L monitor.
Valve
UV/Vis
Delay volume
The time when a peak is detected The peak reaches the fraction
by the UV/Vis detector is called T0 collector at time T1
1 Identify all components in the system flow path that contribute to the delay volume of
interest.
2 Determine the internal volumes of all hardware modules and tubing, see Section 9.14
Component volumes, on page 498 for information about theoretical module volumes and
Section 9.3 Tubing and connectors, on page 420 for information about tubing lengths and
dimensions.
3 To obtain the total delay volume, sum up half of the flow cell volume of the monitor used
(that is, the UV or UV/Vis monitor) with all volumes of tubing and modules that are located
after the monitor in the flow path.
Note:
For pH-valve V9-pH and V9H-pH always use the volume for the valve in bypass position.
The system automatically adds the volumes for the flow restrictor and the pH flow cell
when if they are part of the system.
2 • Select Tubing and Delay Volumes and select Delay volume: Monitor to outlet valve.
• Type in the volume in the Volume field and click OK.
Tubing kit pH Delay vol- Delay volume UV monitor to point of collection of eluent
valve ume pH
V9-pH valve re- Delay volume Delay volume Delay volume
strictor Monitor to outlet Monitor to frac Monitor to frac
valve F9-R F9-C
Tubing kit pH Delay vol- Delay volume UV monitor to point of collection of eluent
valve ume pH
V9H-pH valve re- Delay volume Delay volume Delay volume
strictor Monitor to outlet Monitor to frac Monitor to frac
valve F9-R F9-C
Note: • Always set the delay volume from the UV monitor to the point of collection of eluent. De-
pending on the system configuration, this will be Delay volume: Monitor to outlet valve
or Delay volume: Monitor to frac in the system settings.
• If the pH valve is mounted, make sure to set the value for the delay volume of the pH valve
restrictor. The system automatically adds the volumes for the flow restrictor and the pH
flow cell if they are part of the flow path.
2 Fill a small sample loop, for example 20 μl, with a 1% to 5% acetone solution.
3 Set all components between the monitor and the fraction collector to by-pass mode.
4 Fill the system with water. Run the pump at a flow rate of 1 ml/min and inject the acetone
solution as a sample.
5 Note the volume from the point of injection to the maximum of the peak in the chro-
matogram. This volume is volume measurement V1.
6 Reconfigure the system as follows to include the fraction collector in the flow path before
the monitor:
• Disconnect the two tubing segments from the UV monitor flow cell and connect these
with a low-dead-volume connector, a 1/16'' female to a 1/16'' female union connector.
• Mount the fractionation collector tubing tip on the top of the UV monitor flow cell.
Connect tubing to the bottom of the UV monitor flow cell and lead this tubing to waste.
It is preferable if the waste tubing gives some back pressure as the flow restrictor
now is placed before the UV flow cell.
7 Set the Frac size to a large volume, for example 100 ml, so that the Outlet valve is in the
Frac position during the entire run.
Step Action
8 Start the pump at a flow rate of 1 ml/min and inject the acetone solution.
9 Note the volume from the point of injection to the appearance of the peak in the chro-
matogram. This volume is volume measurement V2.
Mixer, 5 ml 5000
Mixer, 15 ml 15000
Conductivity cell 22
pH flow cell 76
Note: The given values for the component volumes of the valves are average values. Depending on
the chosen flow path the actual component volume may differ somewhat.
Parameter Description
Target value for Sets the target value for the PI pressure regulation as a percentage of the
pressure control pressure limit. If the target pressure is too close to the pressure limit there is
a risk that a short pressure spike will trigger the pressure alarm. The pressure
limit is set in the Alarm pressure instruction. The Alarm pressure used for
pressure control depends on the settings in the System flow instruction.
Pressure control If the flow rate is reduced below the value set in Pressure control min flow
min flow rate rate, the method is paused and the system is set to state ALARMS AND ER-
RORS.
Pressure control min flow rate can be set either as volumetric or as linear
flow. A column type must be selected before using linear flow.
Default 8 40 N/A
Large soft 1 media 8 300 - 600 A higher I value than the default val-
columns ue is needed to speed up pressure
ramp-up times.
1 Soft media is defined as all GE separation media, except silica and MonoBeads.
2 Rigid media is defined as GE separation media that is based on silica and MonoBeads.
Back pressure
Using narrow tubing between components will improve resolution but will lead to increased back
pressure in the system. Narrow tubing after the column will increase the pressure in the column at a
given flow rate. Make sure that the pressure sensor limits in the system are set so that the maximum
pressure for the column used is not exceeded.
Pressure monitor R9 0
Mixer M9 0
UV detector U9-D 0
Conductivity monitor C9 0
I/O-box E9 0
I/O-box, 2nd E9 1
Check/Change Node ID
Step Action
1 Remove the module according to the instruction in Section 2.5 Installation of internal
modules, on page 76.
Step Action
2 The Node ID of a module is set by the position of an arrow on a rotating switch at the
back of the module.
Valve modules have two rotating switches, as shown in the image below:
• the first switch, labeled A, sets the tens and
• the second switch, labeled B, sets the units.
For example, to set the Node ID to 6 for a valve module,
switch A is set to 0 and switch B is set to 6.
A B
3 Check the Node ID and compare it with the listed Node IDs in the tables above.
4 To change the Node ID, use a screw driver to set the arrows of the switches to the desired
number.
10 Ordering information
Introduction
This chapter lists accessories and user replaceable spare parts available for ÄKTA pure.
Mixer
Item Code no.
Tubing
Item Code no.
Holders
Item Code no.
UV monitor
Item Code no.
I/O box
Item Code no.
Bowl 18305103
Valves
Item Code no.
Superloop 10 ml 19758501
Superloop 50 ml 18111382
pH monitor
Item Code no.
pH electrode 28954215
Conductivity monitor
Item Code no.
Flow restrictor
Item Code no.
Module components
Item Code no.
Cables
Item Code no.
Sample Pump S9
Item Code no.
UNICORN
There are different UNICORN products and licenses available for different purposes, for example licenses
for use with a workstation or for working remotely. Contact your local GE salesperson for more infor-
mation on UNICORN products and licenses and how to order.
Index
A storage, 221
Affinity chromatography Column CIP
predefined method, 439–440 maintenance method, 445
Air sensors predefined phase, 447
air sensor adapter, 83 Column performance test
A test, 228 maintenance method, 445
B test, 230, 232 predefined phase, 215, 447
error codes, 405 Column preparation
external air sensors, 87 maintenance method, 445
inlet valves, 40 predefined phase, 447
troubleshooting, 398 Column valve
ÄKTA pure description, 55
description, 13, 86 flow paths V9-C, 56
illustrations, 14 flow paths V9-Cs, 57
main features, 9 V9-C test, 234
maintenance, 251 Column wash
overview, 9 predefined phase, 447
troubleshooting, 356 Component volumes, 498
Anion exchange Conductivity monitor
predefined method, 441 calibration, 308
clean the conductivity flow cell, 300
C description, 72
Cabinet error codes, 405
troubleshooting, 396 troubleshooting, 367
Cassettes, 96 Connectors
quickrelease function, 98 for ÄKTA pure, 421
Cation exchange replace connectors, 317
predefined method, 441 D
Chromatofocusing
predefined method, 441 Delay volume
CIP reference information, 493
column CIP, 284 settings, 207
column maintenance method, 445 Delta column pressure
column predefined phase, 447 description, 68
system CIP, 278 Desalting
system maintenance method, 445 predefined method, 442
system predefined phase, 215, 447 Documentation, 11–12
Cleaning E
column, 221
Elution
system, 220
predefined phase, 447
Column
Equilibration
attach column holder, 172
predefined phase, 447
cleaning, 221
Extension box, 85
column CIP, 284
column holder, 80 F
column valve, 55 Flow cells
connect column, 172 conductivity, 72
pressure alarm, 176