Culber 2
Culber 2
Dissolved Oxygen
C. H. Culberson
College of Marine Studies
University of Delaware
Newark, Delaware 19716 U.S.A.
1. Introduction
This chapter describes procedures to be used for the calibration of volumetric glass-
ware, for the calculation of oxygen concentrations, for the preparation of reagents, and for
water sampling in the Winkler method for dissolved oxygen. These procedures are based
on the Chesapeake Bay Institute technique for the Winkler dissolved oxygen method (Car-
penter, 1965). All participants in WOCE are expected to closely follow Carpenter's method
for the measurement of dissolved oxygen in discrete water samples.
Carpenter's (1965) modication of the Winkler technique is reproducible (2 standard
deviations) to 0.010 mL/L (0.4 mol/kg-sw: approximately 0.2% of air saturated O2
concentrations). Achieving this level of precision, however, requires scrupulous attention to
detail.
The elimination of iodine loss by volatilization is the essential element of Carpenter's
technique. This is accomplished by optimizing the concentrations of the pickling reagents
(MnC12 , NaOH/NaI, H2 SO4 ), and by adopting a whole bottle titration which eliminates
iodine loss during sample transfer.
Institutions which titrate aliquots, rather than the entire oxygen sample, must
demonstrate that their technique does not exhibit systematic errors, when compared to
whole bottle titrations. All institutions muse use the pickling reagent concentrations spec-
ied by Carpenter (1965); no exceptions are allowed.
This chapter is not meant to be a cookbook or recipe which lists each step necessary
to determine dissolved oxygen. Rather, its purpose if (1) to specify algorithms for calibration
of volumetric glassware, and (2) to discuss those aspects of Carpenter's method which are
ambiguous or which require modication.
With the exception of item c, all volumetric apparatus must be calibrated to contain or
deliver. Since blanks are determined by dierence, the exact calibration of item c is not
essential, as long as the same volume is dispensed each time.
The volumetric apparatus should be calibrated by measuring the weight of pure
water contained or delivered at a reference temperature of 20.0 C. All weights must be
corrected for air buoyancy. The following equations should be used to calculate the true
volume from the measured weight of water in air.
where,
air = density of air at the temperature and pressure of the weighing oper-
ation ( 0.0012 g cm,3 at 20 C),
w = density of pure water ( 1.0 g cm,3 for water), and
weights = density of balance calibration weights ( 8.0 g cm,3 ) for stainless
steel, 8.4 g cm,3 for brass).
For weighing temperature between 10 and 30 C, use of the approximate buoyancy
factor,
fbuoy = 1:00105 (4)
will result in an error of less than 0.01%.
Table 1
Sample calibration of an oxygen
ask using WOCE equations. Calibration temperature,
21.7 C; buoyancy factor calculated using density of stainless steel weights, and the cubical
coecient of thermal expansion for borosilicate glass.
Table 2
Sample calibration of a 10 mL pipet using WOCE equations. Calibration temperature,
24.0 C; buoyancy factor calculated using density of stainless steel weights, and the cubical
coecient of thermal expansion for borosilicate glass.
,
(Vx Vblk;dw ) VIO3 NIO3 5598 , 1000 DO
,
O = (Vstd Vblk;dw ) reg
2
(Vbot , Vreg ) (8)
Terms in equation (8) have the following meaning. All volumes are expressed in
cubic centimeters.
WHP Operations and Methods { July 1991 7
= +AS +BS +C S
sw w
1:5 2
(10)
8 Dissolved Oxygen { July 1991
C = 4:8314 10, 7
(13)
Table 3
4. Preparation of Reagents
Reagents used in Carpenter's procedure include the following:
Substituting the pure water density for the density of dilute KIO3 in equation (18)
will not cause signicant error.
sample water through the drawing tube to remove air. Air bubbles frequently adhere to the
sides of the drawing tube. If this occurs, tap the drawing tube to remove them before lling
the oxygen
ask. Bubbles are very likely to form on the sides of new tygon tubing. This
problem can be avoided by soaking the tubing in seawater for several days prior to use.
Rinse the oxygen
ask several times with sample water ( 30 cm3 ) prior to lling;
then ll the
ask; and, nally, over
ow it with at least 3 bottle volumes of sample water
to minimize contamination with atmospheric oxygen. Over
owing the
ask with copious
amounts of sample water is the single most important factor in producing high-quality
oxygen results.
Prior to sampling, the MnC12 and NaOH/NaI reagents should be placed near where
the oxygen samples will be drawn. Add these two reagents immediately after each oxygen
sample is drawn. Do not stopper the oxygen
asks before adding the MnC12 and NaOH/NaI
reagents. The tip of the reagent dispensers should extend below the neck of the oxygen
asks, so that precipitate does not form in the excess seawater above the neck of the
asks.
The calculation of dissolved oxygen, equation (8), involves corrections for the vol-
umes of MnC12 and NaOH/NaI added, and for the oxygen that they contain. The reagent
dispensers must dispense exactly 1.00 cm3 . Their calibration should be checked, and ad-
justed if necessary, by dispensing ten 1 mL volumes into a 10 mL graduated cylinder. The
solubility of oxygen in the reagents is temperature dependent. Since the value used for
the concentration of oxygen in the reagents, DOreg , is the air solubility at 25.5 C, it is
important that the reagents be stored at a temperature as close to 25.5 C as possible. To
maintain their temperature as constant as possible, the MnC12 and NaOH/NaI reagents
should be stored in the laboratory, and only brought on deck when needed. These reagents
should never be stored in the cold, since this will signicantly increase their dissolved oxygen
content.
5.2 Standardization
Carpenter (1965) does not add 1 mL of MnC12 to standards, even though it is
added to blanks. Carpenter's standardization procedure should be modied to include the
addition of 1 mL of MnC12 after the addition of the NaOH/NaI reagent. Standards should
be determined in pure water, not in seawater.
6. Acknowledgements
Must of the discussion in section 2, Calibration of Volumetric Apparatus, is taken
from the unpublished notes of Andrew Dickson (1990). Some of the comments in section 5,
Water Sampling and Oxygen Analysis, are based on the Marine Technicians Handbook from
the Scripps Institution of Oceanography (Anonymous, 1971).
7. References
Anonymous, 1971. Marine Technicians Handbook: Oxygen Analysis. Scripps Institution
of Oceanography Reference No. 71-8.
Bigg, P. H., 1967. Density of water in S. I. units over the range 0-40 C. British Journal
of Applied Physics,18, 521-537.
Carpenter, J. H., 1965. The Chesapeake Bay Institute technique for the Winkler dissolved
oxygen method. Limnol. Oceanogr. 10, 141-143.
Commission on Atomic Weights and Isotopic Abundances, 1988. Atomic weights of the
elements 1987. Pure & Applied Chemistry 60(6):841-854.
Dickson, A., 1990. Calibration of oxygen bottles from the weight of the contained water
when weighed in air. Unpublished notes. 4 pages.
Hodgman, C. D., editor, 1958. Handbook of Chemistry and Physics, 39th edition. Chemical
Rubber Publishing Company, Cleveland, Ohio.
Millero F. J. and A. Poisson, 1981. International one-atmosphere equation of state of
seawater. Deep Sea Res., 28, 625-629.
WHP Operations and Methods { July 1991 15
Murray, C. N., J. P. Riley, and T. R. S. Wilson, 1968. The solubility of oxygen in Winkler
reagents used for the determination of dissolved oxygen. Deep Sea Res., 15, 237-238.
Weiss, R. F., 1981. Oxygen solubility in seawater. Unesco Technical Papers in Marine
Science. No. 36:22