Sys Master - Pdfs - hc4 - h21 - 10538525327390 - COA - DRE C10569000 - ST WB CERT 4861948 1 1 1
Sys Master - Pdfs - hc4 - h21 - 10538525327390 - COA - DRE C10569000 - ST WB CERT 4861948 1 1 1
Sys Master - Pdfs - hc4 - h21 - 10538525327390 - COA - DRE C10569000 - ST WB CERT 4861948 1 1 1
Reference Material
This certificate is designed in accordance with ISO 17034 and ISO Guide 31. This reference material (RM) was designed,
produced and verified in accordance with ISO/IEC 17025, ISO 17034 and a registered quality management system ISO
9001.
Product Name
Benzylalcohol
CERTIFIED CERTIFIED
Purity Expanded Uncertainty (U)
98.61% (g/g) 0.30% (g/g)
Uncertainty
The certified value(s) and uncertainty(ies) are determined in accordance with ISO 17034 with an 95% confidence level (k=2).
Uncertainty is based on the Total Combined Uncertainty, including uncertainties of characterisation, homogeneity and
stability testing. Stability values are based on real evidence opposed to simulation.
The producer certifies that this reference material meets the specification stated in this certificate until the expiry date,
provided it is stored unopened at the recommended temperature herein. Product warranties for this reference material are
set out in the terms and conditions of purchase.
CERTIFIED BY CERTIFIED ON
RM Release
N. Müller 24 Jan 2023
Instrument
CHROMATOGRAM
UHPLC/DAD
Detection
DAD
Column
LUNA Omega C18 1.6 µm 100 x 2.1
mm
Method Details
see Batch Information
Inj.-Vol.
2.0 µL
Flow
0.5 mL/min
Method of Characterisation
Purity = 100% - Assay impurities – Water content (KF)
Method of Identification
EA, NMR, RT, IR, MS, UV
Batch Information
Water Content: 0.72% (g/g) by Karl-Fischer-Titration (U(exp) = 0.06% (g/g)).
Method Details:
Eluent A: 100% Acetonitrile
Eluent B: Water + 0.5% H3PO4
Time [min] Eluent A [%] Eluent B [%]
0 10 90
0.3 10 90
8 100 0
9.5 100 0
Intended Use the balances is verified daily internally Instructions for use
This RM is intended for use in a and annually by an external accredited It is recommended to use 1 mg as the
laboratory as a calibration and quality calibration service. Chromatographic minimum sample size and if less
control standard or in method methods are traceable to the material is used, to increase the
development for analytical International System of Units (SI). certified uncertainty by a factor of two
techniques. for half sample and four for a quarter
Homogeneity of sample. If storage after opening is
Safety Random replicate samples of the final necessary, the RM should be tightly
Proper precautions should be packaged RM have been analysed to closed and kept from light and
observed while handling. See Safety prove homogeneity compliant with moisture. If the RM was in a sealed
Data Sheet. ISO 17034. ampoule, it should be transferred to a
vial with minimum head space. Visit
Traceability Storage the support section of our website
The balances used for gravimetric The RM should be stored in the lgcstandards.com for a series of
measurements are calibrated with original sealed container at the Dr. Ehrenstorfer Tech Tip videos and
weights traceable to the national indicated temperature. frequently asked questions.
standards (DKD). The calibration of