oxy.IQ
Panametrics Oxygen Analyzer
User’s Manual
User’s Manual
910-296 Rev. D
September 2017
bhge.com
© 2017 Baker Hughes, a GE company – All rights reserved.
Baker Hughes reserves the right to make changes in specifications and features shown
herein, or discontinue the product described at any time without notice or obligation.
Contact your BHGE representative for the most current information. The Baker Hughes
logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of
the General Electric Company.
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
87
[no content intended for this page]
Product Registration
Thank you for purchasing a model oxy.IQ from GE. Please register your
product at www.gemeasurement.com/productregistration for product support
such as the latest software/firmware upgrades, product information and
special promotions.
Services
GE provides customers with an experienced staff of customer support
personnel ready to respond to technical inquiries, as well as other remote and
on-site support needs. To complement our broad portfolio of industry-leading
solutions, we offer several types of flexible and scalable support services
including: Training, Product Repairs, Extended Warranties, Service
Agreements and more. Please visit www.gemeasurement.com/services for
more details.
Typographical Conventions
Note: These paragraphs provide information that provides a deeper
understanding of the situation, but is not essential to the proper
completion of the instructions.
Important : These paragraphs provide information that emphasizes
instructions that are essential to proper setup of the equipment.
Failure to follow these instructions carefully may cause unreliable
performance.
Safety Issues
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in
accordance with local codes, standards, regulations, or laws applicable to
safety.
Working Area
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable
to the auxiliary equipment.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of
the equipment.
Regulatory Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE is an active participant in Europe’s Waste Electrical and Electronic
Equipment (WEEE) take-back initiative (Directive 2012/19/EU).
The equipment that you bought has required the extraction and use of natural
resources for its production. It may contain hazardous substances that could
impact health and the environment.
If you need more information on the collection, reuse and recycling systems,
please contact your local or regional waste administration. Visit
http://www.gemeasurement.com/environmental-health-safety-ehs for take-
back instructions and more information about this initiative.
RoHS
The oxy.IQ fully complies with RoHS regulations (Directive 2011/65/EU).
Figure 1: oxy.IQ
• Natural gas
1.4 Features
The oxy.IQ oxygen sensor uses an advanced galvanic fuel cell that
provides superior performance, accuracy, stability and long life. The
cell’s innovative design eliminates the potential for negative signal
output and reduces sources of contamination.
The cell is unaffected by other background gases or hydrocarbons and
is compatible with acidic gases (OX-2 and OX-4 cells). Recovery from
air at low ppm levels takes just a few minutes. Because the cell is
self-contained, minimal maintenance is required. There is no
electrolyte to change and no electrodes to clean.
The oxy.IQ offers the following features:
• Two-wire, loop-powered, 4 to 20 mA transmitter
• Intrinsically-safe option
Chapter 2. Installation
2.1 Mounting the oxy.IQ
To install the oxy.IQ into the process or sample system, refer to
Figure 9 on page 38 and/or Figure 2 below and proceed to the next
page.
CONNECTOR
2.05
1.58
Note: All dimensions are inches [mm]. [52.1]
[40.3]
B 2.75
[69.8] C
4.10
[104.2]
0.32
1.33 [8.25]
[33.9]
2X 2X 1.03 A
1/8-27NPT-2B 8-32 UNC-2B [26.27]
.27 [6.9] .51 [12.9] MAX
0.53
[13.57]
1.00
[25.40]
Ø1.75
[44.3]
Sensor Manifold
PTFE Tape
Blank Label
Std. Cable
Sensor Base
Knurled Nut
Sensor Manifold
Display
Down Arrow
Up Arrow
Cancel Enter
• and Keys - In the Main Menu, use these keys to move the
cursor between rows one row at a time in the direction
indicated.
To enter the Main Menu from normal display mode, simply press the
Enter key at any time. To leave the Main Menu and return to
measurement mode, press the Cancel key.
Note: Depending on how deep you are in the menu structure, it may be
necessary to press the Cancel key more than once to return all
the way back to measurement mode.
2. Press the key twice and then press the Enter key to enter
the Output menu.
5. After selecting the desired output range, press the Enter key to
save the selection. Then, press the Cancel key to return to the
Output menu.
3. Press the key twice and then press the Enter key to enter
the Output menu.
4. Press the key and then press the Enter key to enter the Trim
menu.
1. Press the Enter key to enter the 4 mA Trim menu, and the
analog output is driven to about 4 mA.
3. Press the Enter key to save the trim adjustment and return to
the Trim menu.
1. Press the key and then press the Enter key to enter the 20
mA Trim menu, and the analog output is driven to about 20 mA.
3. Press the Enter key to save the trim adjustment and return to
the Trim menu.
1. Press the key and then press the Enter key to select the YES
menu option.
2. Press the Enter key to acknowledge that you are resetting the
sensor lifetime clock.
3. As instructed, remove the sensor manifold to expose the new
oxygen sensor to ambient air for about two minutes. Then, press
the Enter key to continue.
4. A message indicating that the calibration is in progress will be
displayed, and then the calibration data will be shown. At that
time, press the Enter key to save the calibration data and return
to measurement mode.
Note: A second calibration of the new sensor should be performed
within 1-2 days of the first calibration.
2. If you have not done so already, start the flow of span gas to the
sensor and allow both the 4-20 mA output reading and the display
reading to stabilize.
3. After the reading has stabilized, press the Enter key to enter the
Main Menu.
5. Press the key and then press the Enter key to select the
Span Gas menu option.
6. Press the and keys until the measurement agrees with the
span calibration gas value.
7. Confirm that the reading on the display has stabilized, and press
the
Enter key to save the calibration. Then, press the Cancel key
twice to return to measurement mode.
• Display Menu
• Output Menu
While programming these menus, refer to the menu map in Figure 16
on page 46.
Note: The menu options for initial setup are described in Chapter 3,
Initial Setup & Operation, and are only referenced in this chapter.
3. Press the key three times and then press the Enter key to
enter the Sensor Life menu.
4. The number of days your sensor has been in use is displayed.
When you have finished reading the information, press the
Enter key to return to the Calibration menu.
2. Press the key once and then press the Enter key to enter the
Display menu.
• % only
2. Press the key once and then press the Enter key to enter the
Display menu.
3. Press the key once and then press the Enter key to enter the
Display Range menu.
2. Press the key twice and then press the Enter key to enter
the Contrast menu.
3. Use the and keys to adjust the contrast to the desired value,
then press the Enter key to save the new value.
2. Press the key twice and then press the Enter key to enter
the Output menu.
3. Press the key twice and then press the Enter key to enter
the Error Type menu.
4. Use the and keys to select the desired option and then press
the Enter key to activate that error type. A check mark will
appear next to the selected option to indicate that it is activated.
The following options are available, and you may activate as many
of these options as you wish.
Note: Only the first four options are displayed on the screen upon
entering this menu. When you scroll down to the fourth option
(Low Temp), a down arrow to the right of this option indicates that
an additional screen of options is available.
• High O2
• Low O2 (programmable)
• High Temp
Note: Pressing the Enter key on an error type that has already been
activated, will deactivate that option and remove the check mark.
2. Press the key twice and then press the Enter key to enter
the Output menu.
3. Press the key three times and then press the Enter key to
enter the Error Output menu.
4. Use the and keys to select the desired option and then press
the Enter key to activate that error output. A check mark will
appear next to the selected option to indicate that it is activated.
The following options are available, and you may activate only one
option at a time.
Note: Only the first four options are displayed on the screen upon
entering this menu. When you scroll down to the fourth option
(NAMUR), a down arrow to the right of this option indicates that
an additional screen of options is available.
• None (no error output is generated)
2. Press the key three times and then press the Enter key to
select the Service menu.
5.2.1 Diagnostics
To enter the Diagnostics menu option from the Service Menu, complete
the following steps:
• Output mA
• Output %
When you have finished reading the information, press the Enter
key to move to Page 2 of the Diagnostics menu or press the Cancel
key to exit the Diagnostics menu.
4. Page 2 of the Diagnostics option displays the current values for the
following parameters:
• Temp °C
• Temp Res
• Gain
When you have finished reading the information, press the Enter
key to move to Page 1 of the Diagnostics menu or press the Cancel
key to exit the Diagnostics menu.
Chapter 6. Specifications
6.1 Intrinsically Safe (IS) Installation
Intrinsically safe installations require an MTL7706 zener barrier,
one IS cable, and one non-IS cable.
Power Requirements
24 to 28 VDC at 50 mA
Cable
p/n 704-1318-02 (2 m length) or p/n 704-1317-10 (10 m length)
blue jacketed, twisted-pair with connector, 26 AWG conductors,
with connector
Output
Total load must equal 250 : ±5% when using a zener barrier
6.2 Non-Incendive (Div 2) and General Purpose Installation
No zener barrier or galvanic isolator is used.
Power Requirements
9 to 28 VDC, loop-powered, 0.7 W max
Cable
p/n 704-1317-02 (2 m length) or p/n 704-1317-10 (10 m length)
black jacketed, twisted-pair with connector, 26 AWG conductors,
with connector
• Percent sensors:
- 0% to 1% O2
- 0% to 2% O2
- 0% to 5% O2
- 0% to 10% O2
- 0% to 25% O2
- 0% to 50% O2
Accuracy
• ±1% of range at calibration point
Resolution
±0.1% of range
Linearity
• ±2% of range (OX-1, 2, 3, 5)
Sample Pressure
Vented to atmosphere during operation and calibration
Process Connections
1/8” NPT-F inlet and outlet
Dimensions
4.10 x 2.75 x 2.05 in. (104.1 x 69.9 x 52.1 mm)
Weight
1.35 lb (612 grams)
Electrical Classification
Intrinsically Safe package with zener barrier or galvanic isolator:
• USA/Canada:
IS for Class I, Div 1, Groups A, B, C, D; T4
IS for Class I, Zone 0, AEx ia IIC T4; Tamb -20 to +60°C
• EU ATEX:
II 1 G ia IIC Ga
IECEx Ex ia IIC T4; Tamb -20 to +60°C
Intrinsically safe package for non-incendive (Div 2) application without
use of zener barrier or galvanic isolator:
• USA/Canada:
Class I, Div. 2, Groups A, B, C, D; T4
• ATEX/IECEx:
Ex na IIC T4
European Compliance
See the EU Declaration of Conformity at the back of this manual
II 1 G
[13.57]
0.53
[26.27]
1.03
[25.40]
C
1.00
[52.1]
[33.9]
2.05
1.33
[69.8]
2.75
CONNECTOR
A B C
[104.2]
[40.3]
4.10
1.58
Ø1.75
[44.3]
B
2X
Ø .03 [.76]
C
C
URE RANGE -25°C TO 85°C
BLUE
BROWN
INDICATOR ARROW
39
Appendix A. Outline and Installation Drawings
40
.250±.020 NOTES:
BACK SHELL
END “A”
DETAIL C
PLUG INSULATOR BODY
SCALE 2 : 1
SECTION A-
SCALE 4 :
BLUE
135
BROWN
4.00 .25±.03
1.16
2.50
PART
ITEM NO. NUMBER DESCRIPTION
1 214-1060 PUSH-PULL CONNECTOR, PLUG, HIROSE
2 231-132 CABLE, BLUE JACKET, 4 X 1 INSULATED 26 AWG CONDUCT
3 410-064 SUP SLEEVING, SHRINK, 1/4” BLACK
4 222-036 FERRITE, CABLE CLAMP
INDICATOR ARROW
BACK SHELL
EXTENSION
136
COMPRESSION 2 METERS (6.56FT) 02 4=BLACK, PVC, 26 A
.195±.020
FITTING 10 METERS (32.81FT) 10 5=DRAIN, PVC, 26 A
6=EMPTY
.100±.020
END “A”
SECTION A
SCALE 4
BROWN RED
3 1 +24V
4 2
BLUE GREEN
24V RETURN
GREEN
BLACK
oxy.IQ Transmitter MTL7706
IS GROUND
IS Cable 4-20MA ANALOG
704-1318-xx INPUT DEVICE
(Blue Jacket) Non-IS Cable
R1
1. Equipment connected to barrier
inputs must not use or generate
BLACK
more than 250V.
2. Total load of R1 + R2 must equal R2
250 ohms ±5%.
24V RETURN
oxy.IQ Transmitter
BROWN
BROWN
BLUE
+24V
24V RETURN
oxy.IQ Transmitter
BROWN
BROWN
Yes % only
Trim
Auto Select
Span Gas
4 mA Trim
Display Range
Flow Span Gas 20 mA Trim
On
Sensor Life Error Type
Off
140
Contrast
Low O2
Low Temp
Temp Comp
None
Low
High
Value
NAMUR
O2 uA
Output mA
Output %
Temp °C
141
Temp Res
Gain
OX-n (n=1-4)
OX-1
OX-2
OX-3
OX-4
oxy.IQ BCD-E
A Model Only
oxy.IQ Oxygen Transmitter; 4 to 20 mA output
B Sensor
0 No sensor
1 Standard ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
2 Acid ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
3 Standard percent sensor
4 Acid percent sensor
5 Standard ppm, 0 to 100, 200, 500 and 1000 ppm
C Package
1 Standard Package
3 Intrinsically safe (US/CAN Class 1 Div 1) or
Non-incendive (US/CAN Class 1 Div 2)
4 Ex flameproof
D- Cable Length
0 No cable
1 2 meter cable
2 10 meter cable
E Zener Barrier
0 None
1 Zener barrier
2 Galvanic isolator.
Note: For Class 1 Div 1 either zener barrier or
galvanic isolator must be selected. For
Class 1 Div 2, no barriers neeed. Please
refer to dwg 752-347 for installation
guidelines.
Example part number: oxy.IQ 132-1
Appendix D. Certifications
The following oxy.IQ certifications are included in this appendix:
• “ATEX EU-Type Examination Certificate” on page 52
8 FM Approvals Ltd, notified body number 1725 in accordance with Article 17 of Directive 2014/34/EU of
26 February 2014, certifies that this equipment has been found to comply with the Essential Health and
Safety Requirements relating to the design and construction of equipment intended for use in potentially
explosive atmospheres given in Annex II to the Directive.
The examination and test results are recorded in confidential report number:
9 Compliance with the Essential Health and Safety Requirements, with the exception of those identified in item
15 of the schedule to this certificate, has been assessed by compliance with the following documents:
10 If the sign ‘X’ is placed after the certificate number, it indicates that the equipment is subject to specific
conditions of use specified in the schedule to this certificate.
11 This EU-Type Examination certificate relates only to the design, examination and tests of the specified
equipment or protective system in accordance to the directive 2014/34/EU. Further requirements of the
Directive apply to the manufacturing process and supply of this equipment or protective system. These are
not covered by this certificate.
!"##$&$*;;
Mick Gower
Certification Manager, FM Approvals Ltd..
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
to EU-Type Examination Certificate No. FM14ATEX0032X
The Model oxy.IQ Transmitter’s electronics are contained on one printed circuit board. The circuit board is
located inside of a metallic housing. An oxygen cell holder is connected to the circuit board by way of pins.
The base of the oxygen cell holder contains a threaded joint for replacement of the oxygen cell. The
housing contains a polymeric window display. The polymeric window is approximately 1” by ¾”.The
housing, including the sensor assembly, is approximately 4.1” in height, by 2.75” in width, and 2.05” in
depth.
Whilst this certificate may be used in support of a manufacturer’s claim for CE Marking, FM Approvals Ltd
accepts no responsibility for the compliance of the equipment against all applicable Directives in all
applications.
This Certificate has been issued in accordance with FM Approvals Ltd’s ATEX Certification Scheme.
17 Schedule Drawings
A list of the significant parts of the technical documentation is annexed to this certificate and a copy has
been kept by the Notified Body.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
to EU-Type Examination Certificate No. FM14ATEX0032X
18 Certificate History
Details of the supplements to this certificate are described below:
Date Description
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail: [email protected] www.fmapprovals.com
Blueprint Report
GE Infrastructure Sensing Inc (1000000179)
Class No 3610
Original Project I.D. 3047174
Certificate I.D. FM14ATEX0032X
Drawing No. Revision Level Drawing Title Last Report Electronic Drawing
714-1344 C Oxy.IQ Safety Manual RR209822 Yes (pdf)
752-341 C MAIN PCB RR209822 Yes (pdf)
752-347 B Schematic Diagram - System Diagram oxy.IQ -FM Controlled Document RR209822 Yes (pdf)
15/06/2017 Page 1 of 1
CERTIFICATE OF CONFORMITY
6. The examination and test results are recorded in confidential report number:
8. If the sign ‘X’ is placed after the certificate number, it indicates that the equipment is subject to specific
conditions of use specified in the schedule to this certificate.
9. This certificate relates to the design, examination and testing of the products specified herein. The FM
Approvals surveillance audit program has further determined that the manufacturing processes and quality
control procedures in place are satisfactory to manufacture the product as examined, tested and Approved.
Intrinsically Safe (Entity) for use in Class I, Division 1, Groups A, B, C and D; Temperature Class T4 Tamb =
-20˚C to +60˚C in accordance with Control Drawing No. 752-347; Ex ia IIC T4 Tamb = -20˚C to
+60˚C; in accordance with Control Drawing No. 752-347; Nonincendive for use in Class I, Division 2, Groups A,
B, C and D; Temperature Class T4 Tamb = -20˚C to +60˚C Hazardous Locations.
13 June 2017
J. E. Marquedant Date
VP, Manager, Electrical Systems
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
Canadian Certificate Of Conformity No: FM17CA0101X
IS CL I DIV 1, GPS A, B, C, D;
CL I, DIV 2, GPS ABCD
Class I, Zone 0, Ex ia IIC T4 Ga
Tamb = -20°C to +60°C
Install per Control Drawing 752-347
The Model oxy.IQ is a two-wire, loop-powered transmitter for measuring oxygen in ten ppm ranges and
seven percentage ranges. The Model oxy.IQ Transmitter contains one of five different oxygen cells. The
cells are specified to be part numbers OX-1, OX-2, OX-3 and OX-4 and OX-5. The type of oxygen cell
determines the range of measurement. The measured oxygen is converted to a mA signal for delivery to the
control equipment located in the non-hazardous area.
The Model oxy.IQ Transmitter’s electronics are contained on one printed circuit board. The circuit board is
located inside of a metallic housing. An oxygen cell holder is connected to the circuit board by way of pins.
The base of the oxygen cell holder contains a threaded joint for replacement of the oxygen cell. The
housing contains a polymeric window display. The polymeric window is approximately 1” by ¾”.The housing,
including the sensor assembly, is approximately 4.1” in height, by 2.75” in width, and 2.05” in depth.
1. The Model oxy.IQ Oxygen Transmitter will not pass the 500Vrms dielectric strength test. This must be
taken into account during installation.
2. When installed in a Division 2 location, the Model oxy.IQ Oxygen Transmitter shall be installed in a final
enclosure accepting a Division 2 wiring method per the CEC.
This Certificate has been issued in accordance with FM Approvals Canadian Certification Scheme.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
Canadian Certificate Of Conformity No: FM17CA0101X
Date Description
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
CERTIFICATE OF CONFORMITY
6. The examination and test results are recorded in confidential report number:
8. If the sign ‘X’ is placed after the certificate number, it indicates that the equipment is subject to specific
conditions of use specified in the schedule to this certificate.
9. This certificate relates to the design, examination and testing of the products specified herein. The FM
Approvals surveillance audit program has further determined that the manufacturing processes and quality
control procedures in place are satisfactory to manufacture the product as examined, tested and Approved.
Intrinsically Safe (Entity) for use in Class I, Division 1, Groups A, B, C and D; Temperature Class T4
Tamb = -20˚C to +60˚C in accordance with Control Drawing No.752-347; Intrinsically safe (Entity) for use in
Class I, Zone 0, AEx ia IIC T4 Tamb = -20˚C to +60˚C; in accordance with Control Drawing No.
752-347; Nonincendive for us in Class I, Division 2, Groups A, B, C and D; Temperature Class T4
Tamb = -20˚C to +60˚C; Hazardous (Classified) Location.
13 June 2017
J. E. Marquedant Date
VP, Manager, Electrical Systems
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
US Certificate Of Conformity No: FM17US0190X
IS CL I DIV 1, GPS A, B, C, D;
CL I, DIV 2, GPS ABCD
Class I, Zone 0, AEx ia IIC T4 Ga
Tamb = -20°C to +60°C
Install per Control Drawing 752-347
The Model oxy.IQ is a two-wire, loop-powered transmitter for measuring oxygen in ten ppm ranges and
seven percentage ranges. The Model oxy.IQ Transmitter contains one of five different oxygen cells. The
cells are specified to be part numbers OX-1, OX-2, OX-3 and OX-4 and OX-5. The type of oxygen cell
determines the range of measurement. The measured oxygen is converted to a mA signal for delivery to the
control equipment located in the non-hazardous area.
The Model oxy.IQ Transmitter’s electronics are contained on one printed circuit board. The circuit board is
located inside of a metallic housing. An oxygen cell holder is connected to the circuit board by way of pins.
The base of the oxygen cell holder contains a threaded joint for replacement of the oxygen cell. The
housing contains a polymeric window display. The polymeric window is approximately 1” by ¾”.The housing,
including the sensor assembly, is approximately 4.1” in height, by 2.75” in width, and 2.05” in depth.
1. The Model oxy.IQ Oxygen Transmitter will not pass the 500Vrms dielectric strength test. This must be
taken into account during installation.
2. When installed in a Division 2 location, the Model oxy.IQ Oxygen Transmitter shall be installed in a final
enclosure accepting a Division 2 wiring method per the NEC.
This Certificate has been issued in accordance with FM Approvals US Certification Requirements.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
SCHEDULE
US Certificate Of Conformity No: FM17US0190X
Date Description
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: [email protected] www.fmapprovals.com
,(&([&HUWLILFDWH
RI&RQIRUPLW\
,17(51$7,21$/(/(&7527(&+1,&$/&200,66,21
,(&&HUWLILFDWLRQ6FKHPHIRU([SORVLYH$WPRVSKHUHV
IRUUXOHVDQGGHWDLOVRIWKH,(&([6FKHPHYLVLWZZZLHFH[FRP
$SSOLFDQW *(,QIUDVWUXFWXUH6HQVLQJ
7HFKQRORJ\3DUN'ULYH
%LOOHULFD0$
8QLWHG6WDWHVRI$PHULFD
(TXLSPHQW 0RGHOR[\,42[\JHQ7UDQVPLWWHU
2SWLRQDODFFHVVRU\
7\SHRI3URWHFWLRQ ,QWULQVLF6DIHW\
0DUNLQJ
([LD,,&7*Dr&ื7Dืr&
(QHUJ\/LPLWDWLRQ3DUDPHWHUVb
$SSURYHGIRULVVXHRQEHKDOIRIWKH,(&([ -DPHV0DUTXHGDQW
&HUWLILFDWLRQ%RG\
3RVLWLRQ 930DQDJHU(OHFWULFDO6\VWHPV
6LJQDWXUH
IRUSULQWHGYHUVLRQ
'DWH
7KLVFHUWLILFDWHDQGVFKHGXOHPD\RQO\EHUHSURGXFHGLQIXOO
7KLVFHUWLILFDWHLVQRWWUDQVIHUDEOHDQGUHPDLQVWKHSURSHUW\RIWKHLVVXLQJERG\
7KH6WDWXVDQGDXWKHQWLFLW\RIWKLVFHUWLILFDWHPD\EHYHULILHGE\YLVLWLQJWKH2IILFLDO,(&([:HEVLWH
&HUWLILFDWHLVVXHGE\
)0$SSURYDOV//&
%RVWRQ3URYLGHQFH7XUQSLNH
1RUZRRG0$
8QLWHG6WDWHVRI$PHULFD
,(&([&HUWLILFDWH
RI&RQIRUPLW\
&HUWLILFDWH1R ,(&([)0*; ,VVXH1R
'DWHRI,VVXH
3DJHRI
0DQXIDFWXUHU *(6HQVLQJ(0($
)5((=21(($676+$1121
&2&/$5(99
,UHODQG
$GGLWLRQDO0DQXIDFWXULQJORFDWLRQV
7KLVFHUWLILFDWHLVLVVXHGDVYHULILFDWLRQWKDWDVDPSOHVUHSUHVHQWDWLYHRISURGXFWLRQZDVDVVHVVHGDQGWHVWHGDQGIRXQGWRFRPSO\ZLWKWKH
,(&6WDQGDUGOLVWEHORZDQGWKDWWKHPDQXIDFWXUHU
VTXDOLW\V\VWHPUHODWLQJWRWKH([SURGXFWVFRYHUHGE\WKLVFHUWLILFDWHZDVDVVHVVHGDQG
IRXQGWRFRPSO\ZLWKWKH,(&([4XDOLW\V\VWHPUHTXLUHPHQWV7KLVFHUWLILFDWHLVJUDQWHGVXEMHFWWRWKHFRQGLWLRQVDVVHWRXWLQ,(&([6FKHPH
5XOHV,(&([DQG2SHUDWLRQDO'RFXPHQWVDVDPHQGHG
67$1'$5'6
7KHHOHFWULFDODSSDUDWXVDQGDQ\DFFHSWDEOHYDULDWLRQVWRLWVSHFLILHGLQWKHVFKHGXOHRIWKLVFHUWLILFDWHDQGWKHLGHQWLILHGGRFXPHQWVZDVIRXQG
WRFRPSO\ZLWKWKHIROORZLQJVWDQGDUGV
,(& ([SORVLYHDWPRVSKHUHV3DUW*HQHUDOUHTXLUHPHQWV
(GLWLRQ
,(& ([SORVLYHDWPRVSKHUHV3DUW(TXLSPHQWSURWHFWLRQE\LQWULQVLFVDIHW\L
(GLWLRQ
7KLV&HUWLILFDWHGRHVQRWLQGLFDWHFRPSOLDQFHZLWKHOHFWULFDOVDIHW\DQGSHUIRUPDQFHUHTXLUHPHQWVRWKHUWKDQWKRVHH[SUHVVO\LQFOXGHGLQWKH
6WDQGDUGVOLVWHGDERYH
7(67 $66(660(175(32576
$VDPSOHVRIWKHHTXLSPHQWOLVWHGKDVVXFFHVVIXOO\PHWWKHH[DPLQDWLRQDQGWHVWUHTXLUHPHQWVDVUHFRUGHGLQ
7HVW5HSRUW
4XDOLW\$VVHVVPHQW5HSRUW
*%%$64$5
A
Adjusting, oxy.IQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Air Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Analog Output
see Output
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
B
Buttons, Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
C
Cable
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Installed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Longer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Calculating Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calibration
Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Existing Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
New Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
oxy.IQ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Span Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Contrast, Adjusting Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
D
Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Display
Contrast, Adjusting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21, 23
O2 Parameter, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Range, Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Document Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Drawings, Outline & Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
E
Error
Output Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Type, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
F
Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
I
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation
Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Mounting the oxy.IQ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Sensor Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Interference Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intrinsically Safe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
K
Keypad, Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
L
Label, Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
M
Main Menu
Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Menu
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21, 23
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 21, 25
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Trim. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14, 21
Menu Maps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Main Menu, Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Main Menu, User’s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Mounting the Oxy.IQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
O
O2 Display Parameter, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Order String . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
O-Ring, Sensor Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Outline & Installation Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 37
Output
Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Error Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Error Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
High (20 mA), Trimming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Low (4 mA), Trimming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 25
Range, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Trimming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
oxy.IQ
Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Calibrating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Calibration, Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calibration, Span Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Output Range, Selecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Oxygen Sensor
Calibration, Existing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration, New . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Sensor Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
P
Passcode, Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Pressure Rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Programming, User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Publication Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
R
Range
Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
RoHS Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
S
Safety
Auxiliary Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
General Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Personal Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 7
Sensor Manifold
Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O-Ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Service Menu
Diagnostics Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Passcode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Span Gas Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Support
Customer Support Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
T
Terms and Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Trimming
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Output, High (20 mA). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Output, Low (4 mA) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Typographical Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
W
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
WEEE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Wiring the oxy.IQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
U.S.A Ireland
The Boston Center GE Sensing EMEA
1100 Technology Park Drive Free Zone East, Shannon,
Billerica, MA 01821 Co. Clare, V14 V992, Ireland
U.S.A Tel: +35 361 470200
Tel: 800 833 9438 (toll-free), 978 437 1000
E-mail:[email protected]
Page 2 of 6
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
oxy.IQ User’s Manual 166 69
Addendum A. oxy.IQ Safety Manual
GE Sensing
oxy.IQ 1100 Technology Park Drive Oxy.IQ Safety Manual
Billerica, MA 01821
MODEL DWG NO. 714-1344 REV. C TITLE
• Only trained, competent personnel may install, operate and maintain the
equipment.
• For product training please contact GE Customer Support Center:
U.S.A Ireland
The Boston Center GE Sensing EMEA
1100 Technology Park Drive Free Zone East, Shannon,
Billerica, MA 01821 Co. Clare, V14 V992, Ireland
U.S.A Tel: +35 361 470200
Tel: 800 833 9438 (toll-free), 978 437 1000
E-mail:[email protected]
• The product is an electrical apparatus and must be installed in the hazardous area
in accordance with the requirements of the EC Type Examination Certificate. The
installation must be carried out in accordance with all the appropriate
international, national and local standard codes and practices and site
regulations for Intrinsically Safe apparatus and in accordance with the
instructions contained in the manual. Access to the circuitry must not be made
during operation.
• The maximum operating pressure for the oxy.IQ is 10psi. Be sure the sample
conditioning system is designed to maintain the oxy.IQ pressure below these
limits, and that the oxy.IQ outlet is vented to atmosphere during operation and
calibration.
• WARNING - Substitution of components may impair Intrinsic Safety.
• WARNING - To prevent ignition of flammable or combustible atmospheres,
disconnect power before servicing.
• AVERTISSEMENT - Pour éviter l'inflammation d'atmosphères inflammables ou
combustibles, débrancher l'alimentation avant l'entretien.
• AVERTISSEMENT - Remplacement des composants peut compromettre la
sécurité intrinsèque.
• Equipment is not intended for the measurement of oxygen in fluid of liquid phase.
Page 3 of 6
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
70 167 oxy.IQ User’s Manual
Addendum A. oxy.IQ Safety Manual
GE Sensing
oxy.IQ 1100 Technology Park Drive Oxy.IQ Safety Manual
Billerica, MA 01821
MODEL DWG NO. 714-1344 REV. C TITLE
Markings
Markings shall appear on the oxy.IQ as shown below for the Intrinsically Safe version
of the product.
II 1 G
Page 4 of 6
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
oxy.IQ User’s Manual 168 71
Addendum A. oxy.IQ Safety Manual
GE Sensing
oxy.IQ 1100 Technology Park Drive Oxy.IQ Safety Manual
Billerica, MA 01821
MODEL DWG NO. 714-1344 REV. C TITLE
Page 5 of 6
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
72 169 oxy.IQ User’s Manual
Addendum A. oxy.IQ Safety Manual
GE Sensing
oxy.IQ 1100 Technology Park Drive Oxy.IQ Safety Manual
Billerica, MA 01821
MODEL DWG NO. 714-1344 REV. C TITLE
Power Requirements:
Nominal Operating Parameters: 28VDC at 50mA
Blue (+)
Brown (-)
Power Connector
Page 6 of 6
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
oxy.IQ User’s Manual 170 73
Addendum A. oxy.IQ Safety Manual
We, GE Sensing
1100 Technology Park Drive
Billerica, MA 01821
USA
U.S.A.
The Boston Center
1100 Technology Park Drive
Billerica, MA 01821
U.S.A.
Tel: 800 833 9438 (toll-free)
978 437 1000
E-mail: [email protected]
Ireland
Sensing House
Shannon Free Zone East
Shannon, County Clare
Ireland
Tel: +35 361 470200
E-mail: [email protected]
bhge.com
© 2017 Baker Hughes, a GE company – All rights reserved.
Baker Hughes reserves the right to make changes in specifications and features shown
herein, or discontinue the product described at any time without notice or obligation.
Contact your BHGE representative for the most current information. The Baker Hughes
logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of
the General Electric Company.
910-296 Rev. D
PET-MSM-MAN-809439 - 01 - Reviewed - 02-05-2019 - 5 - For information only
175