Clinical Risk Management: Its Application in The Manufacture of Health IT Systems - Implementation Guidance

The provision and deployment of Health IT Systems within the National Health Service (NHS) can deliver substantial benefits to NHS patients through the timely provision of complete and correct information to those healthcare professionals that are responsible for administering care. However, it must be recognised that failure, design flaws or incorrect use of such systems has the potential to cause harm to those patients that the system is intending to benefit.

Uploaded by

wekhjlghawe
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
136 views61 pages

Clinical Risk Management: Its Application in The Manufacture of Health IT Systems - Implementation Guidance

The provision and deployment of Health IT Systems within the National Health Service (NHS) can deliver substantial benefits to NHS patients through the timely provision of complete and correct information to those healthcare professionals that are responsible for administering care. However, it must be recognised that failure, design flaws or incorrect use of such systems has the potential to cause harm to those patients that the system is intending to benefit.

Uploaded by

wekhjlghawe
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 61

Document filename: DCB0129 Implementation Guidance v3.2.

docx
Directorate Solution Design Standards and Project Clinical Safety
Assurance
Document Reference NPFIT-FNT-TO-TOCLNSA-1300.06
Director Debbie Chinn Status Approved
Owner Stuart Harrison Version 3.2
Author Sean White Version issue date 02.05.2018

Clinical Risk Management: its


Application in the Manufacture
of Health IT Systems -
Implementation Guidance

Copyright © 2018 NHS Digital


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Document Management
Revision History
Version Date Summary of Changes
0.1 16.11.2012 First draft
0.2 10.12.2012 Revised following review by external appraisal. For approval
1.0 03.01.2013 First issue
1.1 17.01.2013 Incorporation of changes following review by Information Standards
Management Services. For approval
2.0 21.01.2013 Approved
2.1 09.05.2013 HSCIC rebranded, correction to Figure A and additional paragraph added to
section 4.3
3.0 08.12.2105 Amended and Approved to reflect changes made to Requirement 2.5
3.1 16.08.2016 NHS Digital Rebranded
3.2 02.05.2018 Changed to include Medical Device Software and Data Coordination Board
reference

Reviewers
This document must be reviewed by the following people:
Reviewer name Title / Responsibility Date Version
NHS Digital Clinical Safety Group 02.05.2018 3.2

Approved by
This document must be approved by the following people:
Name Title Date Version
Dr Manpreet Pujara Clinical Director for Patient Safety 02.05.2018 3.2
Debbie Chinn Director of Solution Assurance 02.05.2018 3.2
Publication copy 02.05.2018 3.2

Copyright © 2018 NHS Digital Page 2 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Data Coordination Board


This information standard (DCB0129) has been approved for publication by the Department
of Health and Social Care under section 250 of the Health and Social Care Act 2012.
Assurance that this information standard meets the requirements of the Act and is
appropriate for the use specified in the specification document has been provided by the
Data Coordination Board (DCB), a sub-group of the Digital Delivery Board.
This information standard comprises the following documents:
• Requirements Specification
• Implementation Guidance
• Change Specification.

An Information Standards Notice (DCB0129 Amd 24/2018) has been issued as a notification
of use and implementation timescales. Please read this alongside the documents for the
standard.
The controlled versions of these documents can be found on the NHS Digital website. Any
copies held outside of that area, in whatever format (e.g. paper, email attachment), are
considered to have passed out of control and should be checked for currency and validity.
Date of publication: 7 June 2018

Copyright © 2018 NHS Digital Page 3 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Related Documents:
These documents provide additional information and are specifically referenced within this
document.
Ref Doc Reference Number Title Version
1. DCB0160 Amd 25/2018 Clinical Risk Management: its Application in the 4.2
Deployment and Use of Health IT Systems:
http://www.digital.nhs.uk/isce/publication/dcb0160
2. DCB0129 Amd 24/2018 Clinical Risk Management: its Application in the 3.2
Manufacture of Health IT Systems:
http://www.digital.nhs.uk/isce/publication/dcb0129
3. 2017/745/EC The EU Regulation on Medical Devices 2017/745
4. ISO 14971:2012 Medical Devices: Application of Risk Management to 2012
Medical Devices
5. ALARP (HSE Website)
6. 2006 Annual Report of the Chief Medical Officer On 2006
the State of Public Health, Department of Health
7. 0555 Healthcare risk assessment made easy, NPSA 2007
8. Managing competence for safety-related systems, 2007
HSE

Copyright © 2018 NHS Digital Page 4 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Contents
1 Introduction 7
1.1 Background 7
1.2 Scope 7
1.3 Glossary of terms 8

2 General requirements for clinical risk management 11


2.1 Clinical risk management process 15
2.2 Top Management responsibilities 16
2.3 Clinical Safety Officer 17
2.4 Competencies of personnel 19
2.5 Third party products 19
2.6 Regular clinical risk management process review 19

3 Project Safety Documentation and Repositories 21


3.1 Clinical Risk Management File 21
3.2 Clinical Risk Management Plan 21
3.3 Hazard Log 23
3.4 Clinical Safety Case 26
3.5 Clinical Safety Case Report 27
3.6 Safety Incident Management Log 31

4 Clinical risk analysis 32


4.1 Clinical risk analysis process 32
4.2 Health IT System scope definition 33
4.3 Identification of hazards to patients 33
4.4 Estimation of the clinical risks 34

5 Clinical risk evaluation 39


5.1 Initial clinical risk evaluation 39

6 Clinical risk control 40


6.1 Clinical risk control option analysis 40
6.2 Clinical risk benefit analysis 41

Copyright © 2018 NHS Digital Page 5 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

6.3 Implementation of clinical risk control measures 44


6.4 Completeness of clinical risk control 44

7 Delivery, Monitoring and Modification 45


7.1 Delivery 45
7.2 Post-deployment monitoring 46
7.3 Modification 47

Appendix A Example Hazards 49


A.1 Introduction of a new Patient Administration System into an acute hospital 49
A.2 Introduction of a new electronic prescribing system in a GP surgery 50

Appendix B Example Hazard Identification Techniques 52


B.1 FFA (Functional Failure Analysis) 52
B.2 HAZID (Hazard Identification) 56
B.3 SWIFT (Structured What-IF Technique) 58
B.4 Fishbone Diagram 60

Copyright © 2018 NHS Digital Page 6 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

1 Introduction
1.1 Background
The provision and deployment of Health IT Systems within the National Health Service (NHS)
can deliver substantial benefits to NHS patients through the timely provision of complete and
correct information to those healthcare professionals that are responsible for administering
care. However, it must be recognised that failure, design flaws or incorrect use of such
systems has the potential to cause harm to those patients that the system is intending to
benefit.
This document provides guidance to support the interpretation of the requirements presented
in DCB0129 [Ref. 2]. It is aimed at those persons within a Manufacturer’s organisation who
are responsible for ensuring the safety of Health IT Systems through the application of
clinical risk management. To aid readability, the structure of this guidance document mirrors
that of the revised standard; where each requirement is presented and then supported by
additional information.
Whereas this document is restricted to Health IT Systems, the recommended risk analysis
should be conducted within the context of any overall risk management system in place in the
Manufacturer’s organisation and any wider health information governance processes. In this
document the term ‘clinical risk’ is used to make clear that its scope is concerned with risks to
patient safety as distinct from other types of risk such as financial.
Throughout this guidance the term ‘standard’ is used to specifically mean DCB0129 [Ref. 2].
Within this document the term ‘should’ does not infer any additional requirements to those
explicitly taken from the standard (as shown in capital letters).

1.2 Scope
This guidance document considers the risk management processes required to ensure
patient safety in respect to the development of a new Health IT System or in respect to the
modification of an existing system.
A Health IT System is defined as a product used to provide electronic information for health
or social care purposes where the product may include hardware, software. or a combination
of both. The scope of the Health IT System may extend beyond a Manufacturer’s
organisation and include hardware and or software procured or supplied from other
organisations and include infrastructure already in use at a Health Organisation. Further
guidance is presented at Section 2 and in Figure 2.
The scope of a specific Health IT System will therefore need to be defined by a Manufacturer
as part of the early lifecycle phases, possibly in conjunction with the Health Organisation.
This document is addressed to those persons in the Manufacturer’s organisation who are
responsible for ensuring the safety of Health IT Systems through the application of risk
management.

Copyright © 2018 NHS Digital Page 7 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

In the development and modification of a Health IT System, the scope of the standard and
this supporting guidance includes:
• all clinical functionality which could potentially cause harm to patients
• operational use and potential misuse of the clinical functionality and its potential to cause
harm to patients
This standard applies to all Health IT Systems including those that are also controlled by
medical devices regulations [Ref. 3], though the requirements are broadly consistent with the
requirements of ISO 14971 [Ref. 4].
A Manufacturer will need to identify which other standards are applicable to the development
of the Health IT System.

1.3 Glossary of terms


Term Definition
Clinical Safety Officer Person in a Manufacturer’s organisation responsible for ensuring the safety of a
(previously referred to Health IT System in that organisation through the application of clinical risk
as Responsible Person) management.
Clinical risk Combination of the severity of harm to a patient and the likelihood of occurrence
of that harm.
Clinical risk analysis Systematic use of available information to identify and estimate a risk.
Clinical risk control Process in which decisions are made and measures implemented by which
clinical risks are reduced to, or maintained within, specified levels.
Clinical risk estimation Process used to assign values to the severity of harm to a patient and the
likelihood of occurrence of that harm.
Clinical risk evaluation Process of comparing a clinical risk against given risk criteria to determine the
acceptability of the clinical risk.
Clinical risk Systematic application of management policies, procedures and practices to the
management tasks of analysing, evaluating and controlling clinical risk.
Clinical Risk Repository of all records and other documents that are produced by the clinical
Management File risk management process.
Clinical Risk A plan which documents how the Manufacturer will conduct clinical risk
Management Plan management of a Health IT System.
Clinical Risk A set of interrelated or interacting activities, defined by the Manufacturer, to meet
Management Process the requirements of this standard with the objective of ensuring clinical safety in
respect to the development and modification of a Health IT System.
Clinical safety Freedom from unacceptable clinical risk to patients.
Clinical Safety Case Accumulation and organisation of product and business process documentation
and supporting evidence, through the lifecycle of a Health IT System.
Clinical Safety Case A Report that presents the arguments and supporting evidence that provides a
Report compelling, comprehensible and valid case that a system is safe for a given
application in a given environment at a defined point in a Health IT System’s
lifecycle.

Copyright © 2018 NHS Digital Page 8 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Term Definition

Harm Death, physical injury, psychological trauma and/or damage to the health or well-
being of a patient.
Hazard Potential source of harm to a patient.
Hazard Log A mechanism for recording and communicating the on-going identification and
resolution of hazards associated with a Health IT System.
Health Organisation Organisation within which a Health IT System is deployed or used for a healthcare
purpose.
Health IT System Product used to provide electronic information for health or social care purposes.
The product may be hardware, software or a combination.
Initial clinical risk The clinical risk derived during clinical risk estimation taking into consideration any
retained risk control measures.
Intended use Use of a product, process or service in accordance with the specifications,
instructions and information provided by the manufacturer to customers.
Issue The process associated with the authoring of a document. This process includes:
reviewing, approval and configuration control.
Likelihood Measure of the occurrence of harm.
Lifecycle All phases in the life of a Health IT System, from the initial conception to final
decommissioning and disposal.
Manufacturer Person or organisation with responsibility for the design, manufacture, packaging
or labelling of a Health IT System, assembling a system, or adapting a Health IT
System before it is placed on the market and/or put into service, regardless of
whether these operations are carried out by that person or on that person's behalf
by a third party.

Patient A person who is the recipient of healthcare.


Patient safety Freedom from harm to the patient.
Post-deployment That part of the lifecycle of a Health IT System after it has been manufactured,
released, deployed and is ready for use by the Health Organisation.
Procedure Specified way to carry out an activity or a process.
Process Set of interrelated or interacting activities which transform inputs into outputs.
Release A specific configuration of a Health IT System delivered to a Health Organisation
by the Manufacturer as a result of the introduction of new or modified functionality.
Residual clinical risk Clinical risk remaining after the application of risk control measures.
Safety incident Any unintended or unexpected incident which could have, or did, lead to harm for
one or more patients receiving healthcare.
Safety Incident Tool to record the reporting, management and resolution of safety incidents
Management Log associated with a Health IT System.
Severity Measure of the possible consequences of a hazard.

Copyright © 2018 NHS Digital Page 9 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Term Definition
Third party product A product that is produced by another organisation and not by the Health IT
System manufacturer. Examples include operating systems, library code,
database and application servers and network components.
Top Management Person or group of people who direct(s) and control(s) an organisation and has
overall accountability for a Health IT System.

Copyright © 2018 NHS Digital Page 10 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

2 General requirements for clinical risk management


In the manufacture of a Health IT System, clinical risk management is an essential activity in
ensuring the system does not compromise patient safety. The focus for a Manufacturer
should be on ensuring that the Health IT System they deliver does not introduce unnecessary
clinical risk and wherever possible reduces existing clinical risk.
General requirements for effective clinical risk management are:
• an understanding of the hardware/software and underlying development practices being
used to implement the Health IT System
• a thorough understanding of:
o the clinical functionality that the Health IT System is intending to provide or replicate
o the business processes that the Health IT System is intending to support
o usability issues which may result in unintended consequences to patients
• assessment of any known deficiencies in existing systems or business processes
• an appropriate awareness of clinical risk management
• an awareness of how clinical risk management aligns with any wider governance
processes
• a fully defined clinical risk assessment process which incorporates the application of
recognised and rigorous methodologies (for example, see Appendix B)
• a risk assessment to be carried out completely and competently
• the implementation and verification of any required clinical risk control measures
• any residual clinical risks are appropriately documented
• appropriate lifecycle management is in place.
Figure 1 presents a pictorial summary of the end-to-end clinical risk management process
including the activities and documentation as required by the Standard.

Copyright © 2018 NHS Digital Page 11 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3.1

Initiation of ISB 0129 Manufacture of a Health IT System Clinical Risk


Management File
4.1
2.1 2.1 Plan Risk Management
Activities
Establish and document
Procedures 3.2
a risk management
process Risk Management
4.2
Plan
Scope Definition

2.2
4.3
Appointment of trained
and experienced Clinical Hazard
personnel Identification
2.4
3.3
Competency 4.4
records Hazard Log
2.3 Clinical Risk Estimation

Appointment of Clinical
Safety Officer 3.4
5.1
Initial Clinical Risk
Evaluation

2.5 5.1

Risk Reduction
Non-health and COTS No Yes
Required?
software 6.1

Control Option Analysis

2.6 6.1

Clinical Risk
Yes Residual Clinical No
Management Process
Risk Acceptable?
Review

6.1

Further Measures
Yes
Practicable?
Clinical Safety
Case
No
6.2
Clinical Risk Benefit
Analysis

6.2

Yes Risk Acceptable No Re-appraisal

6.3
Control Measure
Implementation

6.4

Completeness
Evaluation

3.5
7.1
Clinical Safety Case
Delivery Report(s)

3.6
7.2
Back to Safety Incident Mgt
Post-deployment
appropriate place Log
Monitoring
in lifecycle

7.3

Maintenance

Figure 1 Clinical Risk Management Activities and Documentation

Copyright © 2018 NHS Digital Page 12 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

The term Manufacturer is used to cover all organisations which are involved in the
development of some aspect of a Health IT System. In this respect, Manufacturer is inclusive
of terms such as: provider, integrator, supplier, developer, vendor, etc.
Manufacturers which are including software or hardware procured from other Manufacturers,
must assure themselves of the suitability of these items. Ideally, such components should be
manufactured in accordance with DCB0129 and be accompanied by appropriate safety
documentation. It is, however, recognised that certain components such as third party
products will not fall under this definition; see section 2.5. Here the Manufacturer is to take
reasonable steps, based on perceived risks, to assure itself that the inclusion of such
products does not pose an unacceptable risk to patient safety. Where a Manufacturer uses
components that do not have the necessary safety documentation, then the Manufacturer will
also be responsible for undertaking and documenting an appropriate assessment of the
component concerned as part of his wider offering.
An example procurement chain for a new Health IT System is depicted in Figure 2. Within
this example, the Manufacturer responsible for the delivery of the Health IT System
(Manufacturer B) to the Health Organisation has integrated:
• third party product
• Health IT System A which implements specific functionality (purchased from
Manufacturer A)
• software internally developed.
This integrated software has then been installed on to a procured general hardware platform.
Note, none of the elements in this figure are considered to be a medical device.
The delivered third party products (software and hardware) from their respective vendors are
not accompanied by any supporting safety documentation. As a result, Manufacturer B is
required under requirement 2.5 of DCB0129 to assure itself that any clinical hazards arising
from these components are suitably addressed.
As Health IT System A from Manufacturer A has been specifically developed to implement
health functionality it meets the requirements of DCB0129. Its delivery is therefore
accompanied by a Clinical Safety Case Report. This Clinical Safety Case Report has then
been referenced by an overall Clinical Safety Case Report produced by Manufacturer B to
accompany the release of Health IT System B. Both of these Clinical Safety Case Reports
have been provided to the Health Organisation.
In the following sections the requirements as presented in the standard are reproduced to aid
readability. Each requirement is shown in a coloured table along with the original reference
number.

Copyright © 2018 NHS Digital Page 13 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Health IT System A
Internally developed
Third Party Software Third Party Hardare provided by
software
Manufacturer A

Clinical Safety Case


Report for Health IT
System A

Health IT System B produced by Manufacturer B

Clinical Safety Case


Report for Health IT
System B

Health Organisation

Figure 2 Health IT System procurement and documentation chain

Copyright © 2018 NHS Digital Page 14 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

2.1 Clinical risk management process


2.1.1 The Manufacturer MUST define and document a clinical risk management process
which recognises the risk management activities shown in Figure 1.
Note: the numbers shown in parentheses in this figure refer to sections later in this
document.

Figure 1 Clinical Risk Management Process

In order to ensure that the clinical risk analysis is performed in a structured and
comprehensive manner the Manufacturer needs to define and document its clinical risk
management process. The process is expected to cover the development and maintenance
phases of the lifecycle of any Health IT System impacted by this standard.
The clinical risk management process will recognise the activities shown in Figure 1. Further
detail on each activity and associated requirements can be found by referring to the
referenced section of this guidance document.

Copyright © 2018 NHS Digital Page 15 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

This process can be conducted within the context of the organisation’s wider risk
management policies, quality management system and any governance arrangements, e.g.
for information/health informatics.
The same clinical risk management process may not be appropriate for all releases;
definitions of risk acceptability criteria, hazard severity and likelihood may vary across Health
IT Systems. The Manufacturer is responsible for ensuring the clinical risk management
process is applicable to a given release. Prior to tailoring the clinical risk management
process, the Manufacturer will, as far as is reasonably practical, require an understanding of
the system to be deployed and how it will be used.
Strong co-operation between the Manufacturer and the Health Organisation throughout the
clinical risk activities is encouraged, where possible, if the clinical risk associated with the
deployment of a Health IT System is to be minimised.

2.2 Top Management responsibilities


2.2.1 In implementing the clinical risk management process for a given deployment, Top
Management MUST:
• make available sufficient resources
• assign competent personnel (see section 2.4) from each of the specialist areas that
are involved in developing and assuring the Health IT System
• nominate a Clinical Safety Officer.
2.2.2 Top Management MUST ensure that appropriate levels of authorisation for the Health
IT System and its safety documentation are defined in the Clinical Risk Management
Plan.

The extent of clinical risk analysis and hence the level of resources required to support the
process will need to be commensurate with the scale, complexity and level of clinical risk
associated with the release.
The assessment of the required resources may be guided by experience from previous
developments, including input from personnel involved in those developments. The level of
resource required would also be dependent on the Health IT System development
timescales.
The nominated resources need to have sufficient time available to allow them to apply a
suitable level of effort to ensure the clinical risk management process is completed in a
robust and competent manner. Such personnel will be able to contribute their specialist
knowledge to ensure the clinical risk management activities are executed as competently and
completely as possible. A list of typical specialist areas that may be required to support
clinical risk activities can be seen in section 4.1 of this guidance document.
The roles and responsibilities of personnel supporting the clinical risk management activities
will need to be documented in the Clinical Risk Management File.

Copyright © 2018 NHS Digital Page 16 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

A Clinical Safety Officer has to be nominated for a given release. The Clinical Safety Officer
will be responsible for ensuring the safe design of a Health IT System through the application
of the clinical risk management process. The nominated Clinical Safety Officer needs to
satisfy the requirements of section 2.3 of the standard.
Top Management remains responsible for identifying who in the Manufacturer organisation is
responsible for authorising the release of the Health IT System for subsequent use in live
service. This needs to be recorded in the Clinical Risk Management Plan.
Within the Clinical Risk Management Plan, Top Management will need to specify those
individuals who are able to approve the clinical risk management documentation. As a
minimum this will be the Clinical Safety Officer.
The Manufacturer needs to undertake a formal review of the clinical risk management
activities (see section 4.1) prior to the release of a Health IT System. The outcome of this
review will be a key input into this Top Management decision making process.
Top Management will need to satisfy itself that all foreseeable hazards have been identified
and that the risk of such hazards has been reduced to acceptable levels. By authorising the
release of the Health IT System, Top Management is accepting any residual clinical risk on
behalf of the Manufacturer.

2.3 Clinical Safety Officer


2.3.1 A Clinical Safety Officer MUST be a suitably qualified and experienced clinician.
2.3.2 A Clinical Safety Officer MUST hold a current registration with an appropriate
professional body relevant to their training and experience.
2.3.3 A Clinical Safety Officer MUST be knowledgeable in risk management and its
application to clinical domains.
2.3.4 A Clinical Safety Officer MUST make sure that the processes defined by the clinical risk
management process are followed.

The Clinical Safety Officer needs to be suitably trained and qualified in risk management or
have an understanding in principles of risk and safety as applied to Health IT Systems. A
Clinical Safety Officer needs to have completed appropriate training. Whilst suitable training
is provided by NHS Digital in partnership with other bodies it is recognised that there are
other methods to acquire relevant skills, e.g. Masters modules in Patient Safety. It would be
beneficial for a Clinical Safety Officer to have experience of conducting clinical risk
management activities in an appropriate clinical setting.
Table 1 summarises competencies for a Clinical Safety Officer. The table details the relevant
experience, knowledge and skills that the candidate should have together with an explanation
of why these are required.

Copyright © 2018 NHS Digital Page 17 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Whilst some activities may be delegated, the Clinical Safety Officer will retain overall
responsibility for the following activities:
• approval of the Clinical Risk Management Plan to confirm that the plan is appropriate and
achievable in the context of the Health IT System development and modification
• ensure that clinical risk management activities are completed in accordance with the
Clinical Risk Management Plan
• review and approval of all safety documentation including Clinical Safety Case Reports
and Hazard Logs
• review of evidence in the Clinical Risk Management File to ensure it is complete and
supports the Clinical Safety Case Report
• provide recommendation to Top Management regarding whether the Health IT System is
safe to release
• raise any unacceptable safety risks to Top Management.
COMPETENCY RATIONALE
EXPERIENCE
Relevant clinical experience A thorough understanding of the practice of related healthcare, clinical
workflow and supporting business processes is required in order to
understand how and why adverse outcomes occur in patients and to pre-
empt potential hazards associated with the Health IT System.
KNOWLEDGE
In depth knowledge of the A thorough understanding of why and how errors occur in the
Health IT System being development, deployment and subsequent use of Health IT Systems and
developed, human factors and how these can result in patient harm. Knowledge is required of measures
their contribution and control in that can be effectively applied to control associated clinical risk.
the context of patient harm.
SKILLS
Critical appraisal and logical Must be able to independently recognise potential defects in the inherent
reasoning design of a system, how unintentional errors may occur and how the
system impacts existing business processes, etc.
Needs to be able to critically analyse recommendations made by other
representatives.
Will have to make calculated decisions on whether proposed solutions are
warranted and cost-effective.
Problem solving For reported defects, must be able to identify the root cause and propose
practical and effective solutions from a clinical perspective.
Ability to facilitate consensus Must be able to consider and review differing opinions and broker the
optimum solution between involved stakeholders.

Table 1 Key Competencies for a Clinical Safety Officer

Copyright © 2018 NHS Digital Page 18 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

2.4 Competencies of personnel


2.4.1 Personnel MUST have the knowledge, experience and competencies appropriate to
undertaking the clinical risk management tasks assigned to them.
2.4.2 Competency and experience records for the personnel involved in performing the clinical
risk tasks MUST be maintained.

The Manufacturer needs to use multi-discipline teams appropriate to the level of risk for the
project. Evidence that the personnel have the required skill set to perform the clinical risk
management tasks assigned to them can be ascertained from the individual’s competency
and experience records. Any shortfalls need to be identified and addressed using the
Manufacturer’s competency framework.
Competency and experience records for personnel involved in the clinical risk activities need
to be maintained in line with the Manufacturer’s Human Resource process. The records need
to be reviewed and updated on a regular basis. Any new skills, training or knowledge gained
as part of continued professional development needs to be added to the records to ensure
that the latest information is available.
Additional guidance can be found in “Managing competence for safety-related systems”
published by the HSE [Ref. 8].

2.5 Third party products


2.5.1 The Manufacturer MUST assess any third party product that is included within a release
as part of the clinical risk management process.
2.5.2 The nature of this assessment MUST be included in Clinical Safety Case Reports.

Many Health IT Systems are reliant on the use of third party products. Such products can
introduce a variety of risks particularly where a Health IT System is reliant upon it or
interoperates with it. Risks may also arise when software updates or patches are applied to
these products. Such products are, however, unlikely to have been risk assessed for health
applications by the original supplier.
Manufacturers are required as part of their clinical risk assessment activities to consider any
third party product incorporated into their Health IT System. The Manufacturer needs to:
• document the assessment of third party products in their Clinical Safety Case Report
• assess any changes in third party products. The Safety Case report will need to be re-
issued if a change in the level of risk results.

2.6 Regular clinical risk management process review


2.6.1 The Manufacturer MUST formally review its clinical risk management process at
planned, regular intervals.

Copyright © 2018 NHS Digital Page 19 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

The clinical risk management process needs to be formally reviewed, to ensure the process
remains effective. Such a review will need to encompass representatives of the key
stakeholder communities and especially the relevant clinical staff including the Clinical Safety
Officer. An annual review is recommended.
The review should examine whether any changes are needed as a result of greater
experience, best practice or lessons learnt from other releases to support a process of
continual improvement. Any changes to the clinical risk management process may also need
to be captured in any current Clinical Risk Management Plans.
Detailed records of the formal review and its findings should be retained by the Manufacturer
in accordance with their management processes.

Copyright © 2018 NHS Digital Page 20 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3 Project Safety Documentation and Repositories


All clinical risk documentation needs to be subject to configuration control so that any
subsequent changes can be tracked.

3.1 Clinical Risk Management File


3.1.1 The Manufacturer MUST establish at the start of a project a Clinical Risk Management
File for the Health IT System.
3.1.2 The Clinical Risk Management File MUST be maintained for the life of the Health IT
System.
3.1.3 All formal documents and evidence of compliance with the requirements of this standard
MUST be recorded in the Clinical Risk Management File.
3.1.4 Any decisions made that influence the clinical risk management activities undertaken
MUST be recorded in the Clinical Risk Management File.

The purpose of the Clinical Risk Management File is to provide a physical or logical
repository of all records and documents that are produced by the clinical risk management
process and required by this standard. If the documents are referenced from the Clinical Risk
Management File then they must be capable of being retrieved.
Consideration should be given to ensuring adequate back-up or archiving procedures are in
place to ensure that the Clinical Risk Management File, and the artefacts it contains or
references, remains preserved and recoverable throughout the life of the Health IT System.

3.2 Clinical Risk Management Plan


3.2.1 The Manufacturer MUST produce at the start of a project a Clinical Risk Management
Plan, which will include risk acceptability criteria, for the new Health IT System.
3.2.2 A Clinical Safety Officer MUST approve the Clinical Risk Management Plan.
3.2.3 If the nature of the project changes, or key people change, during the development and
modification of a Health IT System, then the Clinical Risk Management Plan MUST be
updated.
3.2.4 The Clinical Risk Management Plan MUST be maintained throughout the life of the
Health IT System.

The purpose of the Clinical Risk Management Plan is to document and schedule the clinical
risk management activities to support the safe development and maintenance of the Health
IT System.
The Clinical Risk Management Plan should:
• define and describe the Health IT System and the clinical context in which it will be used.
Where the clinical context is not known, e.g. a new Health IT System, then assumptions
made regarding its anticipated usage need to be recorded

Copyright © 2018 NHS Digital Page 21 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

• state the relevant procedures, policies and resources required to ensure effective and
efficient clinical risk management
• adhere to the Manufacturer’s quality and project management processes and
requirements
• define the Health IT System development lifecycle and its associated phases
• qualify which clinical risk activities are applicable at a particular lifecycle phase
• specify the criteria that are to be used to estimate the clinical risk (see section 4.4) and
evaluate the acceptability of the clinical risk (see section 5.1)
• identify key roles of responsibility and authority for each clinical risk activity. Additionally it
needs to also identify what other resources are required to support the activity, e.g.
reviewer, subject matter expert, test analyst, etc.
• define those members of staff who are able to approve the safety documentation
• record under what circumstance or periodicity the plan should be reviewed. Triggers
could be at a transition to the next phase in the Health IT System lifecycle, a change of
resource or in line with existing governance arrangements. The motivation for review is to
maintain an up to date and effective plan and to support a process of continual
improvement
• describe the process that will be used to monitor and react to safety incidents should they
arise following delivery of a release of a Health IT System.
Where some of the above aspects are addressed in other management documentation, e.g.
a Clinical Safety Management System, then the Clinical Risk Management Plan may
reference out to such documentation.
The extent of the Clinical Risk Management Plan needs to be commensurate with the scale
and clinical functionality of the Health IT System being developed whilst addressing the
clinical risk management activities specified within this standard.
Both the clinical risk management and overall programme level activities need to be
integrated correctly. Careful consideration needs to ensure that activities are scheduled in the
correct order as certain activities may have a dependency on another. For example, Clinical
Risk Control Option Analysis (section 6.1) should be completed before the Manufacturer
finalises its training material; the rationale being that specific training requirements may be
identified as clinical risk controls.
The Clinical Risk Management Plan needs to be approved by the Clinical Safety Officer prior
to use. The Clinical Safety Officer shall have the appropriate clinical experience and risk
management expertise to assess whether the plan is appropriate and achievable in the
context of the Health IT System being developed and released.
The Clinical Risk Management Plan forms part of the Clinical Risk Management File.

Copyright © 2018 NHS Digital Page 22 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3.3 Hazard Log


3.3.1 The Manufacturer MUST establish and maintain a Hazard Log.
3.3.2 A Clinical Safety Officer MUST approve each version of the Hazard Log.
3.3.3 An issued Hazard Log MUST accompany each Clinical Safety Case Report.

The Hazard Log is a mechanism for recording and communicating the on-going identification
and resolution of hazards associated with the Health IT System. It is organised so that it
enables a systematic approach to the management of hazards and supports the effective
collation of safety case evidence. Such on-going revisions will:
• incorporate new hazards, when identified
• record the mitigation of defined hazards through the implementation of clinical risk control
mechanisms
• reference supporting evidence
• record the status of actions.
Whilst the Hazard Log is a living document and continues to be updated during the lifecycle
of the Health IT System, a base-lined version is to be issued with each Clinical Safety Case
Report.
Each version of the Hazard Log has to be reviewed and approved by the Clinical Safety
Officer to signify that the clinical safety information recorded is accurate and appropriate.
An example Hazard Log template is presented at Table 2. It is not prescriptive or definitive
but illustrates how, reading from left to right, a well-structured Hazard Log supports effective
clinical risk management and promotes the collection of relevant evidence in a timely
manner. Where clinical risk control is transferred to the Health Organisation, this shall be
clearly identified in the Hazard Log. In this situation, the Health Organisation will be
responsible for implementing the necessary clinical risk control measures.
Additional examples are presented in Table 3 and Table 4 which illustrate how such a log
would be populated for hazards with single or multiple causes respectively. Table 5
summarises the entries that are recorded in each column of a Hazard Log.

Copyright © 2018 NHS Digital Page 23 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Initial Hazard Risk Assessment Additional Controls Residual Hazard Risk Assessment Actions
Potential Hazard
Hazard Hazard Hazard Possible Existing
Clinical Business
Number Name Description Causes Controls Risk Risk Status
Impact Severity Likelihood Design Test Training Process Severity Likelihood Summary Owner
Rating Rating
Change

Table 2 Representative Hazard Log Template

Initial Hazard Risk Assessment Additional Controls Residual Hazard Risk Assessment Actions
Potential Hazard
Hazard Hazard Hazard Possible Existing
Clinical Business
Number Name Description Causes Controls Risk Risk Status
Impact Severity Likelihood Design Test Training Process Severity Likelihood Summary Owner
Rating Rating
Change

1 X X X X X X X X ? ? ? ? X X X X X X

Table 3 Representative Hazard log with Single Cause

Initial Hazard Risk Assessment Additional Controls Residual Hazard Risk Assessment Actions
Potential Hazard
Hazard Hazard Hazard Possible Existing
Clinical Business
Number Name Description Causes Controls Risk Risk Status
Impact Severity Likelihood Design Test Training Process Severity Likelihood Summary Owner
Rating Rating
Change

2 X X X X X X X X X X

1 X ? ? ? ? X X

2 X ? ? ? ? X X

3 X ? ? ? ? X X

Table 4 Representative Hazard Log with Multiple Possible Causes

X – Field to be populated
? – Field to be populated as applicable

Copyright © 2018 NHS Digital Page 24 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Field Description
Hazard number A unique number for the hazard
Hazard name A short descriptive name for the hazard
Hazard description A short description of the hazard
Potential Clinical Description of effect of hazard in the care setting and potential impact on the patient
Impact
Possible Causes Possible cause(s) that may result in the hazard. These may be technical, human
error, etc. Note: a hazard may have multiple causes
Existing Controls Identification of existing controls or measures that are currently in place and will
remain in place post implementation that provide mitigation against the hazard, i.e.
used as part of initial Hazard Risk Assessment
Initial Hazard Risk Assessment
• Severity The severity of the hazard as defined in Table 7
• Likelihood The likelihood of the hazard as defined in Table 8
• Risk Rating The derived risk rating from the combination of likelihood and severity according to
Table 9
Additional Controls
• Design Identification of design features or configurations implemented in the Health IT
System in order to provide mitigation against the hazard.
• Test Identification of testing to be completed in order to provide mitigation against the
hazard
• Training Identification of training to be implemented in order to provide mitigation against the
hazard.
• Business Process Identification of any Business Process Changes implemented in order to mitigate
Change against the hazard
Residual Hazard Risk Assessment
• Severity The severity of the mitigated hazard as defined by Table 7
• Likelihood The likelihood of the mitigated hazard as defined by Table 8
• Risk Rating The derived mitigated risk rating from the combination of likelihood and severity
according to Table 9
Actions
• Summary Summary of the action being taken with regard to mitigation of the hazard or
individual causes
• Owner The owner of the action
Hazard Status The status of the hazard:
• 'Open' not all clinical risk management actions, owned by the Manufacturer, in
respect of this hazard, have been completed.
• 'Transferred' all clinical risk management actions owned by the Manufacturer, in
respect of this hazard, have been completed but not all actions, owned by the
deploying Health Organisation, have been completed.
• 'Closed' all clinical risk management actions in respect of this hazard have been
completed.

Table 5 Hazard Log Entries

Copyright © 2018 NHS Digital Page 25 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3.4 Clinical Safety Case


3.4.1 The Manufacturer MUST develop and maintain a Clinical Safety Case for the Health IT
System.

The Clinical Safety Case is a structured argument which is supported by a body of relevant
evidence that provides a compelling, comprehensible and valid case that a system is safe for
release. The argument provides an explanation of how the supporting evidence can be
interpreted as indicating that the Health IT System exhibits an adequate degree of safety,
e.g. by demonstrating compliance with requirements or sufficient mitigation of identified
hazards.
The supporting evidence is the result of observation, analysis, testing or simulation that
provides information from which system safety can be claimed.
Parallels can be drawn between a Clinical Safety Case and legal proceedings:
• presentation of a defence (argument) without evidence is unfounded; how does the
defence hold?
• presentation of evidence without a legal argument is unexplained; what is the meaning of
the evidence?
The Clinical Safety Case should not be thought of as a physical issued document but rather
the intellectual planning that needs to be considered and undertaken in order to establish the
safety argument and generate the supporting evidence. Every effort should be made to
establish the safety argument as soon as practical in the lifecycle. This will ensure that
resource and effort is directed efficiently to generate relevant evidence. If consideration of the
safety argument is left until later in the lifecycle it may become difficult to explain how the
available evidence supports claims over the safety of the Health IT System. Such an
approach may result in gaps or lack of evidence which may result in additional work, delays
and increased costs.
The Clinical Safety Case will evolve during the lifecycle of the Health IT System and must be
reviewed to ensure that it continues to provide sufficient confidence in the safety of the
Health IT System.
The relationship between the Clinical Risk Management File, the Clinical Safety Case and
the Clinical Safety Case Report can be understood by considering a filing cabinet:
• the filing cabinet itself can be thought of as the Clinical Risk Management File, i.e. the
repository in which relevant information is stored
• the organisation, indexing and cross referencing of the information within the filing
cabinet can be thought of as the Clinical Safety Case, i.e. the planning and structure

Copyright © 2018 NHS Digital Page 26 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

• the retrieval of information from the filing cabinet can be thought of as the Clinical Safety
Case Report, i.e. presentation of information that has previously been organised to
support a safety position at any point in time. It is permissible for this presentation to be
incomplete at a particular point in time. For example, it may not be possible to present
particular evidence at an early iteration of the report but it should to be possible to
present the need for that evidence.

3.5 Clinical Safety Case Report


3.5.1 The Manufacturer MUST produce a Clinical Safety Case Report at each lifecycle phase
defined in the Clinical Risk Management Plan.
3.5.2 A Clinical Safety Officer MUST approve each Clinical Safety Case Report.
3.5.3 The Manufacturer MUST make available each Clinical Safety Case Report to a receiving
organisation, which may be a Health Organisation or another Manufacturer.

The Clinical Safety Case Report is the physical document that summarises all the key
elements of the Clinical Safety Case and references all supporting material in a clear,
comprehensible and concise format. It serves to communicate the Clinical Safety Case to the
end users and Top Management but also where appropriate to other bodies such as
regulators.
As the underlying Clinical Safety Case continues to evolve during the Health IT System
lifecycle, then there is a need to issue Clinical Safety Case Reports in support of key
milestones. The issuing of Clinical Safety Case Reports will be dictated by the Health IT
System development lifecycle being followed as defined in the Clinical Risk Management
Plan.
Typical development phases are detailed below. Generally, a Clinical Safety Case Report will
be issued during but before completion of a phase. The principal motivation for issuing a
Clinical Safety Case Report before completion of the phase is to support the decision to
transition into the next development phase, e.g. via a design review and to influence, from a
safety perspective the work conducted in the succeeding phase.
• Requirements Analysis: At this phase of the development the scope of the Clinical Safety
Case Report will be limited to defining the scope of the Health IT System (4.2),
identification of potential hazards (4.3) and estimation of the initial clinical risk (4.4). The
purpose of issuing a Clinical Safety Case Report at this phase is to demonstrate that all
perceivable hazards have been pre-empted and their clinical risk considered. The result
of the clinical risk assessment establishes the safety requirements which are fed forward
and addressed in the design phase.

Copyright © 2018 NHS Digital Page 27 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

• Design: At this phase of the development the scope of the Clinical Safety Case Report
will extend to include the initial clinical risk evaluation (5.1) and where necessary clinical
risk control option analysis (6.1). The purpose of issuing a Clinical Safety Case Report at
this phase is to demonstrate that the clinical risk of the hazards identified during the
requirements analysis phase has been evaluated and where necessary risk control
methods have been identified and justified as being appropriate are addressed in the
design. If the subsequent design activities establish new requirements then hazard
identification (4.3) and risk estimation (4.4) activities need to be conducted. The Clinical
Safety Case Report should present the residual clinical risk assuming successful
implementation of the identified controls. Where the residual risk remains unacceptable,
the Clinical Safety Case Report must present evidence to show that the manufacturer
has undertaken a clinical risk benefits analysis (6.2).
• Test: At this phase of the development the scope of the Clinical Safety Case Report will
extend to demonstrate implementation of the clinical risk control measures (6.3), where
the owner of those controls is the Manufacturer. The purpose of issuing a Clinical Safety
Case Report at this phase is to provide evidence that necessary risk control methods
identified in the design phase have been implemented and the implementation is
effective in reducing the clinical risk to the levels identified in the preceding Clinical
Safety Case Report. If any defects are raised during testing then the clinical safety
impact of any such defects needs to be considered and any identified risk managed in
accordance with this standard.
• Delivery: At this phase of the development the scope of the Clinical Safety Case Report
will extend to show completeness of clinical risk control (6.4). The purpose of issuing a
Clinical Safety Case Report at this phase is to document completion of all the clinical risk
management activities and to present the residual clinical risk at the point of the Health IT
System being delivered by the Manufacturer for subsequent live use. Where any element
of risk control is being transferred to a Health Organisation, for example the requirement
to train end users, then this must be clearly communicated. The specific Health IT
System release needs to be recorded.
• Modification: If during its lifetime, the Health IT System is modified, then the release of
any such modification needs to be supported by a Clinical Safety Case Report. The
Manufacturer will need to assess whether a single report can be issued at the delivery
phase or whether reports will need to be issued at preceding phases as discussed
above. The scale of change and its impact on clinical safety will dictate the approach
taken. Again the Clinical Safety Case Report will need to specify the release highlighting
the extent of the change introduced.
The specific time when a Clinical Safety Case Report is issued within a development
programme will be influenced by many factors, including the development lifecycle that is
being followed. It is important that the Manufacturer identifies these issue dates within the
Clinical Risk Management Plan.
A single Clinical Safety Case Report may be maintained; being re-issued in accordance with
local configuration control procedures, or individual standalone Clinical Safety Case Reports
may be issued.

Copyright © 2018 NHS Digital Page 28 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

The Clinical Safety Case Report is the primary vehicle for presenting a statement of the
clinical safety of the Health IT System. It therefore needs to be a readable document rather
than simply a listing of the Clinical Safety Case or the content of the Clinical Risk
Management File.
It needs to provide the reader with:
• a summary of all the relevant knowledge that has been acquired relating to the clinical
risks associated with the Health IT System at that point in the lifecycle
• a clear and concise record of the process that has been applied to determine the clinical
safety of the Health IT System
• a summary of the outcomes of the assessment procedures applied
• a clear listing of any residual clinical risks that have been identified and the related
operational constraints and limitations that are applicable
• a clear listing of any hazards and associated clinical risks that have been transferred,
together with any declared risk control measures, that are to be addressed as part of the
Health Organisation clinical risk management process
• a listing of outstanding test issues / defects associated with the Health IT System which
may have a clinical safety impact.
The structure of a Clinical Safety Case Report will reflect the organisation of the underlying
Clinical Safety Case, which in turn will be influenced by the requirements of this standard. An
example structure is provided in Table 6 but should not be considered to be prescriptive or
definitive.
The Manufacturer’s Clinical Safety Case Reports must be made available to the deploying
Health Organisation. This deliverable is a key input into the Health Organisations clinical risk
management activities in support of compliance with ISB 0160.
Where required and within the framework of any contract terms that exist between a
Manufacturer and a Health Organisation, the Manufacturer needs to make available to the
Health Organisation any documentation or evidence that is referenced within a Clinical Safety
Case Report or which supports the underlying Clinical Safety Case.
It is strongly recommended that the Manufacturer proactively works in close collaboration
with Health Organisations following delivery in order to ensure safe and effective deployment
of the Health IT System. Such relationships will minimise the likelihood of unanticipated
issues occurring and will ensure that any risk controls that the Manufacturer is dependent on
the Health Organisation to implement are communicated across organisational boundaries.

Copyright © 2018 NHS Digital Page 29 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

1 Introduction
Purpose of the Clinical Safety Case Report and phase of lifecycle it relates to.
2 System Definition / Overview
Description of the Health IT System; identification of Health IT System part and version
number; description of the clinical environment it is to be used in; description of any existing
systems it replaces or interfaces with; number of users and patients.
3 Clinical Risk Management System
Description of the Manufacturer’s clinical risk management system; identification of key
personnel, their roles and responsibilities; identification of clinical risk management
governance structure.
4 Clinical Risk Analysis
Hazard identification; description of patient safety consequences; explanation of hazard
causes and contributory conditions; identification of existing mitigating controls; estimation of
clinical risk; identification of participating personnel.
5 Clinical Risk Evaluation
Evaluation of initial level of risk of each identified hazard using pre-defined criteria.
6 Clinical Risk Control
Identification, justification, implementation and verification of adequate risk controls; residual
clinical risk evaluation and completion of controls.
7 Hazard Log
Presentation of associated Hazard Log.
8 Test Issues
Summary of any outstanding test issues and the impact on clinical safety.
9 Summary Safety Statement
Statement from the Clinical Safety Officer summarising the safety position of the Health IT
System in the context of the intended deployment.
10 Quality Assurance and Document Approval
Evidence of appropriate quality, review and approval regimes.
11 Configuration Control / Management
Evidence of appropriate configuration control being used.

Table 6 Representative Content of a Clinical Safety Case Report

Copyright © 2018 NHS Digital Page 30 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3.6 Safety Incident Management Log


3.6.1 The Manufacturer MUST maintain a Safety Incident Management Log.

The Manufacturer needs to establish and use a Safety Incident Management Log to support
the effective communication, resolution and archiving of safety related incidents. The log
should be used throughout the life of a Health IT System.
The Safety Incident Management Log should serve to provide a common portal to
appropriate Manufacturer staff so that they have an up to date view of the status and
management of both current and historical safety incidents associated with a Health IT
System.
The requirement for a Safety Incident Log can be met through the use of an existing service
management process but it will be necessary to identify those incidents that result or have
the potential to result in a clinical risk.
The Safety Incident Management Log should record the following parameters:
• Reference Number: Unique identifier
• Reported by: Name and contact details of person reporting the incident
• Reported Date: Date on which the incident was reported
• Incident Summary: Narrative of incident including as much detail as is available including
any prevailing conditions, causes and observed effects including any harm that occurred
• Clinical Risk Assessment: Determination of clinical risk by considering the severity of the
incident, the likelihood of re-occurrence and known mitigation for relevant hazards
• System Release: Details of system affected
• Journal: Record of work conducted, including date and time, to resolve the incident. This
entry would also identify any permanent risk control measures introduced to prevent re-
occurrence. Should include “who” “when” and “what”
• Made Safe Date: Date at which the incident was made safe through the introduction of
short-term risk controls
• Closed Date: Date at which the incident was resolved through the introduction of
permanent risk control measures
• Cause: Summary of root cause analysis conducted.

Copyright © 2018 NHS Digital Page 31 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

4 Clinical risk analysis


4.1 Clinical risk analysis process
4.1.1 The Manufacturer MUST implement the clinical risk analysis activities defined in the
Clinical Risk Management Plan.
4.1.2 Clinical risk analysis SHOULD be carried out by a multi-disciplinary group including a
Clinical Safety Officer.
4.1.3 The extent of clinical risk analysis MUST be commensurate with the scale, complexity
and level of clinical risk associated with the Release.

It is the responsibility of the Clinical Safety Officer to ensure that the clinical risk management
activities are implemented as documented in the Clinical Risk Management Plan. Any
departures from the Clinical Risk Management Plan should be documented within the Clinical
Risk Management File. The Manufacturer may wish to update the Clinical Risk Management
Plan in light of any such changes.
In order for the clinical risk analysis to be completed in a robust and competent manner it is
essential that it is carried out by a group with representatives from all areas that are involved
in the development and maintenance of the Health IT System. Clinicians that will use Health
IT System after deployment should be included in the process. Involvement of a diverse set
of expertise is more likely to result in hazards being identified which may have otherwise
been missed.
Representatives from the following specialist areas will need to be involved in the clinical risk
analysis, as appropriate:
• Clinical Safety Officer
• Subject Matter Experts / Specialists
• Message Analysts
• Technical Architects
• Clinical Users from the Health Organisation deploying the system
• Clinical Safety Officer from the Health Organisation deploying the system
• Programme Manager
• Quality Assurance
• Safety Engineering
• Risk Management
• Test Management and execution
• Quality Management

Copyright © 2018 NHS Digital Page 32 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

4.2 Health IT System scope definition


4.2.1 The Manufacturer MUST define the clinical scope of the Health IT System which is to be
delivered.
4.2.2 The Manufacturer MUST define the intended use of the Health IT System which is to be
delivered.

Prior to commencing any hazard identification / risk assessment activity for a release it is
essential to define the scope of the assessment. Defining the boundary correctly will limit the
extent of the analysis as well as ensuring all areas impacted by the release are considered.
Clinical scope is the extent of the functionality that is provided within the Health IT System
that can be used to support or influence the administration of healthcare to a patient.
Intended use is the definition or explanation of who will use the Health IT System and how
they will use it, in terms of existing business processes or within new business process.
When defining the boundary of a clinical risk assessment the Manufacturer needs to consider
the following where applicable:
• third party product integrated within the Health IT System to be released
• clinical process / use: understand the revised operations with the new solution and how
the deploying Health IT System will impact on the current business processes and ways
of working
• human interface: the interaction of users with the Health IT System and their behaviours
using it
• IT Infrastructure: assess any infrastructure at the Health Organisation that is within the
Manufacturer’s scope of influence, required to support the deployment of the Health IT
System. This may be achieved by the Manufacturer specifying the minimum system
requirements
• data migration: where data migration is to be undertaken by the Manufacturer it should
be included in the scope of the clinical risk management activities. Any hazards
associated with the data migration should be analysed and suitably mitigated, working in
conjunction with the relevant Health Organisation as appropriate.

4.3 Identification of hazards to patients


4.3.1 The Manufacturer MUST identify and document known and foreseeable hazards to
patients with respect to the intended use of the Health IT System in both normal and
fault conditions.

The Manufacturer will need to undertake and document hazard identification activities in
order to reveal and document potential hazards to patients. Hazard identification needs to
consider both normal and abnormal operating conditions and usage scenarios.

Copyright © 2018 NHS Digital Page 33 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Many techniques exist for hazard identification and an appropriate technique will need to be
chosen depending on the application and the available expertise. Example hazard
identification techniques are presented in Appendix B.
In order to identify pertinent hazards the following three key areas must be considered:
• end to end clinical process, including functionality and how that functionality is used
• inter and intra Health IT System messaging
• Health IT System architecture and design.
It is strongly recommended that a hazard workshop is run to support complete hazard
identification. Attendees at the workshop will be drawn from the areas identified in section 4.1
depending on the nature of the Health IT System.
The identified hazards and their causes shall be recorded in the Hazard Log. Examples of
populated Hazard Logs are presented in Appendix A.
Details of the hazard workshop, including date, attendees and minutes should be recorded in
the Clinical Risk Management File and documented in the Clinical Safety Case Report.
Where no clinical hazards are raised through the hazard identification activities then this
judgement shall be recorded in the Clinical Risk Management File along with details of what
was done, who carried out the assessment and the date of the assessment. As a result, the
remaining clinical risk management activities defined in the clinical risk management process
do not need to be conducted. Should any changes be proposed to the original scope or
content of the Health IT System then the hazard identification shall be repeated to ensure
that no hazards have been introduced as a result of the change. If this assessment identifies
a new hazard then the clinical risk management process needs to be brought into play.

4.4 Estimation of the clinical risks


4.4.1 For each identified hazard the Manufacturer MUST estimate, using the criteria specified
in the Clinical Risk Management Plan:
• the severity of the hazard
• the likelihood of the hazard
• the resulting clinical risk.

The following classifications of likelihood, severity and resulting clinical risk are given for
illustrative purposes only. It is for the Manufacturer to decide on the classifications to use for
the release of a Health IT System. The assessment criteria which will be used shall be
documented in the Clinical Risk Management Plan.
Further guidance on healthcare risk assessment has been published by the NPSA in their
guide “Healthcare risk assessment made easy” [Ref. 7].

Copyright © 2018 NHS Digital Page 34 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

4.4.1 Assessment of Severity


The assessment of severity is made on a qualitative scale which should take account of the
harm that might be experienced by patients if the hazard was to arise and an adverse event
was to occur. The number of points on the qualitative scale for severity is a matter of choice
by the Manufacturer. A five point scale might be:
• catastrophic
• major
• considerable
• significant
• minor.
Each severity category should have an associated meaning assigned. These meanings
should be used to support the severity assessment made for each hazard. The provision of
meanings will allow consistency of applications across hazards.
An illustration of a possible severity classification scheme is given in Table 7.

Copyright © 2018 NHS Digital Page 35 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Severity Interpretation Number of


Classification Patients Affected
Catastrophic Death Multiple
Permanent life-changing incapacity and any condition for which the Multiple
prognosis is death or permanent life-changing incapacity; severe
injury or severe incapacity from which recovery is not expected in
the short term
Major Death Single
Permanent life-changing incapacity and any condition for which the Single
prognosis is death or permanent life-changing incapacity; severe
injury or severe incapacity from which recovery is not expected in
the short term
Severe injury or severe incapacity from which recovery is expected Multiple
in the short term
Severe psychological trauma Multiple
Considerable Severe injury or severe incapacity from which recovery is expected Single
in the short term
Severe psychological trauma Single
Minor injury or injuries from which recovery is not expected in the Multiple
short term.
Significant psychological trauma Multiple
Significant Minor injury or injuries from which recovery is not expected in the Single
short term.
Significant psychological trauma Single
Minor injury from which recovery is expected in the short term Multiple
Minor psychological upset; inconvenience Multiple
Minor Minor injury from which recovery is expected in the short term; Single
minor psychological upset; inconvenience; any negligible severity

Table 7 Example Severity Classification

This classification deals not only with physical injury but also with psychological trauma. It
also distinguishes between harm to single and to multiple patients.

4.4.2 Assessment of Likelihood


Systematic faults are characteristic of software and, unlike random faults, the likelihood of
their occurrence is not amenable to quantification. Thus their likelihood is subject to
judgement on a qualitative scale.
Where the likelihood of the identified hazard cannot be quantified, hazards should still be
listed and a reasonably pessimistic qualitative judgement should be used to allow a risk class
to be assigned.

Copyright © 2018 NHS Digital Page 36 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

An example five point scale for likelihood is:


• very high
• high
• medium
• low
• very low.
Each likelihood category should have an associated meaning assigned. These meanings
should be used to support the likelihood assessment made for each hazard within the Clinical
Safety Case Report. The provision of meanings, see Table 8, will allow consistency of
applications across hazards
Likelihood Interpretation
Category

Very high Certain or almost certain; highly likely to occur


High Not certain but very possible; reasonably expected to occur in the majority of cases
Medium Possible
Low Could occur but in the great majority of occasions will not
Very low Negligible or nearly negligible possibility of occurring

Table 8 Example Likelihood Classification

For each hazard the severity and likelihood assessment together with the resulting clinical
risk shall be recorded in the Hazard Log. Any information required to support the assessment
should be also be recorded in the Hazard Log.

4.4.3 Assessment of Risk


Clinical risk is defined as the combination of the severity of harm to a patient and the
likelihood of occurrence of that harm. A two-dimensional clinical risk matrix, Table 9, is used
to combine hazard likelihood and severity to yield a measure of risk.
Very High 3 4 4 5 5
High 2 3 3 4 5
Likelihood

Medium 2 2 3 3 4
Low 1 2 2 3 4
Very Low 1 1 2 2 3
Minor Significant Considerable Major Catastrophic

Severity

Table 9 Example Clinical Risk Matrix

Copyright © 2018 NHS Digital Page 37 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Controls that are in place prior to the deployment of the Health IT System and will still remain
in place post deployment should be factored into the assessment.
The assessment of the hazard should consider the most realistic and typical scenario that
could result in patient harm. It is possible that a hazard consequence may have more than
one credible severity rating. For example the consequence of a patient being inadvertently
prescribed an overdose of one particular drug could be more severe and immediate, giving
little or no time for intervention, than the overdose of a different drug which may simply
induce inconvenience after an elapsed period of time. The likelihood of the former occurring
may be significantly lower than the latter if it is used to treat a very rare condition. The
assessment needs to consider all scenarios and establish which is most credible.
It should be noted that a single hazard may have multiple possible causes. Where there are
multiple possible causes the hazard risk assessment should be reported at the hazard level.
In deriving the overall risk for a particular hazard, it is permissible to complete the Severity,
Likelihood and Risk Rating cells for each individual cause. In this situation the most
pessimistic Risk Rating for the related causes may be selected for the hazard.

Copyright © 2018 NHS Digital Page 38 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

5 Clinical risk evaluation


5.1 Initial clinical risk evaluation
5.1.1 For each identified hazard, the Manufacturer MUST evaluate whether the initial clinical
risk is acceptable. This evaluation MUST use the risk acceptability criteria defined in the
Clinical Risk Management Plan.
5.1.2 If the initial clinical risk is acceptable, then the risk control requirements defined in
sections 6.1 to 6.3 do not apply to this hazard.

A key element of the clinical risk evaluation process is to gain:


• an understanding of the specific risk levels
• an understanding of where significant risks lie that may or may not subsequently be
found capable of risk reduction to acceptable levels.
In order to establish risk acceptability, each clinical risk rating obtained from the risk matrix
has to be compared against the risk acceptability criteria defined in the Clinical Risk
Management Plan.
Controls that are in place prior to the deployment of the Health IT System and will remain in
place post deployment should be factored into the assessment.
Definition of acceptable clinical risk is a decision for the Manufacturer, taking into account, as
far as practical, the current values of society perhaps expressed in local, national or regional
regulations. However, the general acceptability of an estimated risk, as defined in the risk
matrix, (Table 9), is shown in Table 10.
5 Unacceptable level of risk
4 Mandatory elimination of hazard or addition of control measure to reduce risk to an
acceptable level
3 Undesirable level of risk
Attempts should be made to eliminate the hazard or implement control measures to
reduce risk to an acceptable level. Shall only be acceptable when further risk reduction is
impractical
2 Acceptable where cost of further reduction outweighs benefits gained or where further
risk reduction is impractical
1 Acceptable, no further action required

Table 10 Example Risk Acceptability Definitions

Risk control options will need to be identified for those risks that are considered unacceptable
or undesirable, see section 6.
The results of the clinical risk evaluation and the rationale on which it is based are to be
recorded in the Clinical Risk Management File.

Copyright © 2018 NHS Digital Page 39 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

6 Clinical risk control


6.1 Clinical risk control option analysis
6.1.1 The Manufacturer MUST identify appropriate clinical risk control measures to remove
any unacceptable clinical risk.
6.1.2 Proposed clinical risk control measures MUST be assessed by the Manufacturer to
determine whether:
• new hazards will be introduced as a result of the measures
• the clinical risks for previously identified hazards will be affected.
6.1.3 The Manufacturer MUST manage any new hazards or increased clinical risks in
accordance with sections 4.4 to 6.4.
6.1.4 The Manufacturer MUST evaluate the residual clinical risk. This evaluation MUST use
the risk acceptability criteria defined in the Clinical Risk Management Plan.
6.1.5 Where a residual clinical risk is judged unacceptable, the Manufacturer MUST identify
additional clinical risk control measures in order to reduce the clinical risk.
6.1.6 If the Manufacturer determines that no suitable risk control measures are possible then
the Manufacturer MUST conduct a clinical risk benefit analysis of the clinical risk
(section 6.2).

Risk control options are to be investigated for each risk that has been evaluated as
unacceptable. Whilst risk reduction might not be practicable in all cases, it should be
considered.
Risk control measures can reduce the likelihood of the hazard occurring or reducing the
severity if the hazard arises. For example, patient safety risks associated with provision of
unintended medications may be reduced by using TallMan lettering to reduce the likelihood
of errors being made in drug name perception or using native dm+d (dictionary of medicines
and devices) terms to prevent communication errors between systems.
A reduction in risk can be achieved through the application of one or more of the following
mechanisms, listed in order of preference:
• changes to the design or the inclusion of protective measures in the Health IT System
• product verification and validation (for example, testing). A testing programme should
address each of the hazards and thus provide a practicable demonstration that the
claimed risk reduction has been achieved
• administrative and implementation procedures
• user, operator and other stakeholder training and briefing
• information for patient safety, including warnings.

Copyright © 2018 NHS Digital Page 40 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Where a Manufacturer is unable to implement a suitable clinical risk control measure, e.g.
training required by Health Organisation staff, then the Manufacturer will ensure this need is
clearly documented and communicated to the Health Organisation.
Once the risk control options have been identified the residual clinical risk needs to be
evaluated using the criteria defined in the Clinical Risk Management Plan.
For each hazard, there are two possible results:
• the residual clinical risk is acceptable
• the residual clinical risk is unacceptable.
There is an important distinction to be made between residual clinical risks that are so low
that there is no need to consider them and residual clinical risks that are greater than that but
which are accepted because of the associated benefits and the impracticability of reducing
the risks. If the residual clinical risk meets the Manufacturer’s risk acceptability criteria then
no further risk reduction is necessary.
An impact assessment of a proposed risk control measure should also be undertaken to
determine if the risk control measure:
• creates a new hazard
• increases the estimated risk associated with another identified hazard.
Where this is the case, then the suitability of the proposed risk control measure should be re-
evaluated. If no other option is available, then the new hazards or changed risks will
themselves need to be managed in accordance with the risk management process.
The clinical risk control measures selected will need to be recorded in the Clinical Risk
Management File.

6.2 Clinical risk benefit analysis


6.2.1 Where a residual clinical risk is deemed unacceptable and further clinical risk control is
not practicable, the Manufacturer MUST determine if the clinical benefits of the intended
use outweigh the residual clinical risk.
6.2.2 If the clinical benefits do not outweigh the residual clinical risk, then the clinical risk
remains unacceptable and the project SHOULD be re-appraised.

Clinical risk benefit analysis is not required for every hazard.


Clinical risk benefit analysis is used to justify the residual clinical risk associated with a
hazard once all practicable measures to reduce the clinical risk have been applied. If, after
applying these measures, the clinical risk is still judged not acceptable, a clinical risk benefit
analysis is needed to establish whether the Health IT System is likely to provide more clinical
benefit than harm.

Copyright © 2018 NHS Digital Page 41 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

The decision as to whether the residual clinical risk associated with a hazard is outweighed
by the benefit the Health IT System provides is essentially a matter of judgement by
experienced and knowledgeable individuals, which would normally include the Clinical Safety
Officer. Unfortunately, there can be no standardised approach to estimate clinical benefit and
a greater degree of variation will be the inevitable result of using different approaches.
Those involved in making clinical risk benefit judgements have a responsibility to understand
and take into account the technical, clinical, regulatory, economic, sociological and political
context of their risk management decisions.
If the analysis does not support the conclusion that the clinical benefits outweigh the residual
clinical risk, then the clinical risk remains unacceptable.
However, the Health Organisation may be able to put in place additional control measures to
those identified by the Manufacturer to further reduce the residual clinical risk. These
additional external risk controls may then result in the acceptability criteria for a specific
deployment being met. In such circumstances the Manufacturer needs to make sure that the
Health Organisation understands that they are accepting certain residual risks that may arise
in the use of the Health IT System. In such circumstances, the Health Organisation may
decide that the clinical benefits outweigh the risk and the Health IT System can be delivered
and deployed.
Generally, if all practicable clinical risk control measures, including those of an external
Health Organisation are insufficient to satisfy the Manufacturer’s clinical risk acceptability
criteria, then the Health IT System should not be delivered. This decision should be
communicated to the deploying Health Organisation.
It is possible (probable) that the risk acceptance criteria for a Manufacturer and a Health
Organisation will be different. Inevitably some alignment will be required when assessing risk
acceptability across organisational boundaries.
The clinical risk benefit analysis needs to be documented in the Clinical Safety Case Report
and whether the residual clinical risk is now acceptable needs to be documented.
6.2.1 ALARP
The concept of ALARP (As Low As Reasonably Practicable) [Ref. 5] may be used to
establish risk acceptance and justify the residual clinical risk associated with any identified
hazard. Practicable has two elements:
• technical practicability; the ability to achieve further risk reduction
• economic practicability; the cost of achieving further risk reduction.
Any ALARP justification of residual clinical risk should consider these two elements.

Copyright © 2018 NHS Digital Page 42 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Establishing whether a residual clinical risk is ALARP involves considering the level of risk
remaining against the efforts required to reduce it further. The assessment is one of
proportionality. Whilst it may be feasible to reduce the level of residual clinical risk through
further mitigation or control the cost of doing so may be so great that it far outweighs the
benefits to be gained in doing so. Conversely, there may be situations where for modest
additional effort significant benefits in risk reduction could be realised.
ALARP is not prescriptive and requires expert judgement to be expressed in order to
substantiate a claim of ALARP. Consequently any such assessment is always subjective.
The ALARP triangle, shown in Figure 3, depicts the concept.

Risk cannot be tolerated

Unacceptable

Risk tolerable only if


reduction impractical or cost
is disproportionate to benefit
gained
Tolerable
Increasing Risk

Risk tolerable if cost of reduction


exceeds improvement gained

Broadly Acceptable No additional measures


required

Figure 3 ALARP Triangle

Copyright © 2018 NHS Digital Page 43 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

6.3 Implementation of clinical risk control measures


6.3.1 The Manufacturer MUST implement the clinical risk control measures identified in
section 6.1.1, except where these are to be implemented by another organisation.
6.3.2 The Manufacturer MUST verify each clinical risk control measure implemented under
6.3.1.
6.3.3 The Manufacturer MUST verify the effectiveness of each clinical risk control measure
implemented under 6.3.1.

The Manufacturer must verify the implementation and effectiveness of the risk control
measures for which they are responsible. Where a Manufacturer is dependent upon another
organisation to implement further risk control measures in order to reduce the residual clinical
risk to acceptable levels then they remain responsible for informing such organisations of
their need to conduct the same activities.
Verification of a clinical risk control measure should consider both the implementation of the
measure and the effectiveness of the measure.
The implementation verification activity, which should happen before delivery, is to confirm
that the risk control measure is in place. The means and responsibility by which verification is
carried out will depend on the nature of the measure.
Verifying the effectiveness of a clinical risk control measure demonstrates that the measure
provides the intended result. Hence, effectiveness of a clinical risk control measure can only
be determined with a clear understanding of the required effect of that clinical risk control
measure.
The results of these activities will need to be recorded in the Clinical Safety Case Report.

6.4 Completeness of clinical risk control


6.4.1 The Manufacturer MUST ensure that the clinical risks from all identified hazards have
been considered and accepted.

The Manufacturer of a Health IT System needs to be able to demonstrate that the clinical
risk(s) from all identified hazards have been identified and considered and that the
acceptability of all residual clinical risks is suitably justified.
Prior to delivery of any release a formal review will need to be conducted to ensure that the
clinical risk management process as defined in the Clinical Risk Management Plan has been
completed (see section 7.1).
The results of this review will need to be recorded in the Clinical Safety Case Report.

Copyright © 2018 NHS Digital Page 44 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

7 Delivery, Monitoring and Modification


7.1 Delivery
7.1.1 The Manufacturer MUST undertake a formal review of the Health IT System prior to its
delivery to ensure that all of the requirements of this standard have been addressed.
7.1.2 The results of this review MUST be recorded in the Clinical Safety Case Report.
7.1.3 The Health IT System configuration for the release MUST be recorded in the Clinical
Safety Case Report.

Prior to delivery of the Health IT System, the Manufacturer will need to undertake a formal
review of the clinical risk management activities conducted to ensure that the requirements of
this standard have been addressed.
The scale of this review should be commensurate with the scale of the overarching clinical
risk management process but needs to be sufficient to ensure that Top Management are
adequately appraised of the work conducted.
In essence the review needs to show that:
• the Manufacturer’s Clinical Risk Management Plan has been implemented and the
outcomes recorded
• the residual clinical risk for each hazard is acceptable
• appropriate methods are in place to obtain relevant post-deployment information and to
feed these into the Manufacturer’s clinical risk management system.
Top Management need to be satisfied that all foreseeable hazards have been identified and
that the clinical risk of each hazard has been reduced to acceptable levels. In all
circumstances, Top management remains responsible for the release of the Health IT
System.
In addition, Top Management needs to be satisfied that any outstanding defects that remain
unresolved at release have been reviewed by the Clinical Safety Officer and the potential
impact on clinical safety has been assessed. If the risk of any such defects remains
unacceptable then the Manufacturer must consider implementing additional risk control
measures. Where further risk control measures are considered impractical then clinical risk
benefits analysis needs to be conducted to establish whether the benefits of the release
outweigh the residual clinical risk.
It may not be practical in all circumstances for a Manufacturer to implement adequate risk
controls to reduce the risk to acceptable levels. In such circumstances a Manufacturer may
identify the requirement for a Health Organisation to implement additional controls. Such
requirements must be clearly documented in the Clinical Safety Case Report to ensure that a
deploying Health Organisation is aware that an element of clinical risk is being transferred to
it and that the Health Organisation will be responsible for implementing appropriate risk
control measures.

Copyright © 2018 NHS Digital Page 45 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

This review and its findings must be recorded and summarised in the supporting Clinical
Safety Case Report.

7.2 Post-deployment monitoring


7.2.1 The Manufacturer MUST establish, document and maintain a process to collect and
review reported safety concerns and safety incidents for the Health IT System following
its deployment.
7.2.2 The Manufacturer MUST assess the impact of any such information on the on-going
validity of the Clinical Safety Case.
7.2.3 Where any such evidence is assessed to undermine the safety case the Manufacturer
MUST take appropriate corrective action in accordance with the Clinical Risk
Management Plan and document it in the Clinical Safety Case Report.
7.2.4 The Manufacturer MUST ensure safety related incidents are reported and resolved in a
timely manner.
7.2.5 A record of safety incidents, including their resolution, MUST be maintained by the
Manufacturer in a Safety Incident Management Log.

Once deployed, there is a need for the Manufacturer to proactively monitor and review the
achieved safety characteristics of the Health IT System. This monitoring needs to extend
beyond the Health IT System itself to include the impact on users and related healthcare
processes.
The validity of any assumptions and the effectiveness of any controls made in the Clinical
Safety Case Report need to be monitored to ensure the perceived level of clinical risk
remains representative and acceptable. If it is found that the Clinical Safety Case does not
hold in live system use, then the Manufacturer will need to undertake the clinical risk activities
described in Sections 4 to 6 of the standard. This may result in additional or modified clinical
risk control mechanisms being introduced to manage the clinical risk. Any such changes
need to be recorded in a re-issued Clinical Safety Case Report.
To support continued safe use of a Health IT System within a Health Organisation the
Manufacturer needs to establish, implement and follow a safety incident management
process.
The purpose of the safety incident management process is to:
• enable users of the Health IT System to report incidents, to the Manufacturer, they have
had or they consider may have an impact on patient safety
• enable personnel within the Manufacturer’s organisation to report incidents they have
had or they consider may have an impact on patient safety
• provide a communication mechanism within the Manufacturer’s organisation and where
appropriate the Health Organisations that are using the Health IT System
• ensure appropriate and sufficient resources are allocated by the Manufacturer to manage
and resolve the reported incident

Copyright © 2018 NHS Digital Page 46 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

• enable the Manufacturer to issue safety alerts or bulletins to Health Organisations to


advise users of potential safety incidents and associated mitigation.
To achieve this, the safety incident management process needs to provide:
• a central point of contact (helpdesk) where the incident is logged
• a mechanism by which a clinical risk assessment can be made. In practice this will be the
same criteria defined in the clinical risk management plan. As part of this assessment the
existing safety case should be reviewed:
o does the reported incident constitute a new hazard?
o is the incident a realisation of a recorded hazard?
o have clinical risk controls failed?
• a mechanism by which the user community can be advised of the safety incident
• a mechanism through which effective root cause analysis can be conducted. This
mechanism will need to consider:
o the provision of appropriate resource
o collaboration with appropriate organisations
• corrective changes have been approved by the appropriate authority as identified in the
Clinical Risk Management Plan
• appropriate key point indicators to ensure effective management
• a mechanism by which the user community can be advised that the safety incident has
been resolved
If, as a consequence of an incident, the Clinical Safety Case is impacted then the Clinical
Safety Case will need to be updated to capture this. In such circumstances the Manufacturer
should consider issuing an updated Clinical Safety Case Report.
The reporting, management and resolution of each issue shall be recorded in a Safety
Incident Management Log which will be referenced in the Clinical Risk Management File.
The requirement to record and manage safety incidents can be addressed by a Manufacturer
in any existing service management framework, e.g. ITIL (IT Infrastructure Library), If this
approach is followed then it will be necessary to be able to distinguish between those
incidents that have a potential safety impact and any other incident recorded within that
management system.

7.3 Modification
7.3.1 The Manufacturer MUST apply their clinical risk management process to any
modifications or updates of the deployed Health IT System.
7.3.2 The application of this process MUST be commensurate with the scale and extent of the
change and the introduction of any new clinical risks.

Copyright © 2018 NHS Digital Page 47 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

7.3.3 The Manufacturer MUST issue a Clinical Safety Case Report to support any
modifications to the Health IT System that changes its clinical risk.
7.3.4 The Manufacturer MUST maintain an audit trail of all versions and patches released for
deployment.

It is expected that a Health IT System will be modified or updated during its service. The
motivation for any such changes could be defect fixes or introduction of new functionality.
The Manufacturer will need to review its Clinical Safety Case to see if the modification or
updates impact on existing hazards or introduces new hazards. In the event that the safety
case is no longer valid the Manufacturer’s Clinical Safety Case Report will need to be up-
issued.
It is important that a direct correlation is maintained between a particular issue of a Clinical
Safety Case Report and the corresponding Health IT System release. The Clinical Safety
Case Report must record the specific Health IT System release it relates to. This record is in
addition to any other configuration control processes used within a Manufacturer’s
organisation. This will need to be reviewed by the Health Organisation as part of their own
clinical safety management activities to assess the impact on their own Clinical Safety Case.
The Manufacturer needs to create a log of all patches and modifications delivered into the
NHS. This log should identify the release deployed at each Health Organisation. If an incident
is reported which relates to a particular release of the Health IT System this information can
be used to identify the extent of the issue and contact affected Health Organisations.

Copyright © 2018 NHS Digital Page 48 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Appendix A Example Hazards


Two scenarios are presented below to illustrate example hazards, the potential safety consequence, the possible causes and associated controls(s) which all need to be considered when conducting the
related risk assessment. It is important to record the correlation between the identified cause and associated control(s) so that the appropriateness and sufficiency of the control(s) can be evaluated. No risk
assessment has been conducted and the associated columns have been omitted from the Hazard Log.
The key areas of Clinical Process (CP), Messaging (M) and Design (D) that are to be considered in hazard identification are also indicated in the examples.

A.1 Introduction of a new Patient Administration System into an acute hospital


This Hazard Assessment is set in the context of the introduction of a new Patient Administration System (PAS), replacing an existing manual administration and covers one particular scenario of a patient
attending who is unable to provide their identity. It is assumed that existing manual processes will be retained to support the introduction of the new system to mitigate any inherent risk with the new system
not being able to support care.

Additional Controls
Hazard Hazard Potential Clinical
Hazard Name Possible Causes Existing Controls
Number Description Impact Business Process
Design Test Training
Change

1 Patient Incorrect Presenting patient is


misidentification identification of incorrectly identified
presenting within the Patient
patient on their Administration System
admission to and is mapped to
hospital. another patient’s
medical record.
Possibility that the
patient subsequently
receives inappropriate
care or a delay in care.
Patient unknown to hospital None, introduction of a new PAS design should PAS to be tested to assure Hospital staff to be In emergency
and is unable to confirm their system support patient correct implementation trained how to admit situations hospital
own identity (CP). administration in patient administration. unidentified patients procedures must allow
emergency situations in emergency temporary admission
when their identity cannot situations. of patients even when
be established. their identity cannot be
initially established.
Patient’s identity is similar to None, introduction of a new PAS to implement NHS PAS to be tested to assure Hospital staff to be Hospital procedures to
an existing patient held in system number correct implementation of trained in use of NHS accommodate use of
PAS (CP). NHS number. number as prime NHS number as prime
identifier. identifier.

PAS human interface design None, introduction of a new PAS to present PAS interface design to be Hospital staff trained Hospital procedures to
is such that transcription or system demographic details in a assessed by Manufacturer’s to remain vigilant accommodate use of
mis-selection errors occur; format that clearly CSO prior to release. whilst identifying new PAS.
administrator believes that identifies differences. PAS interface to be patient.
they have entered/selected Design of PAS interface evaluated by hospital staff Hospital staff trained
correct demographic details to comply with national prior to release to ensure in PAS use before
for the presenting patient (D). standards suitability. deployment.

Copyright © 2018 NHS Digital Page 49 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

A.2 Introduction of a new electronic prescribing system in a GP surgery


This Hazard Assessment is focusing on the introduction of a new prescribing system and design and messaging flaws in that system that result in some error in the electronic prescription. There is an
assumption that the dispensing system is working correctly and that existing competencies and procedures that already exist in the dispensing organisation would provide mitigation in the circumstance of
the identified cause.

Additional Controls
Hazard Hazard
Hazard Name Potential Clinical Impact Possible Causes Existing Controls
Number Description Business Process
Design Test Training
Change

2 Incorrect Electronic On receipt at the


electronic prescription sent dispensing pharmacy the
prescription to dispensing electronic prescription
pharmacy is contains different
different to what information to what the
the prescriber prescriber had intended or
intended. thought they had
prescribed. Possibility that
the patient receives the
incorrect medications,
dosage or quantity or
suffers a delay in the
administration of
medications.
Faults in prescribing Existing dispensary Clinical safety design Clinical safety testing by
system that result in competencies and requirements to minimise Manufacturer to assure
unintended but credible procedures may alert likelihood. system
changes to the electronic dispenser to a problem and
message (D, M) prevent unintended
medications to be
dispensed. In this case the
patient would experience
some delay.
If prescriber suspects the
system is not working
correctly they can revert to
paper prescribing.

Copyright © 2018 NHS Digital Page 50 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Additional Controls
Hazard Hazard Hazard
Potential Clinical Impact Possible Causes Existing Controls
Number Name Description Business Process
Design Test Training
Change

Incorrect mapping of Existing dispensary Prescribing system to use Prescribing system to be Surgery staff to remain
medications within competencies and native dm+d terms tested to assure correct vigilant whilst
electronic prescribing procedures may alert use of native dm+d prescribing medication
system translates dispenser to a problem and
intended prescription into prevent unintended
something different (D). medications to be
dispensed. In this case the
patient would experience
some delay.
If prescriber suspects the
system is not working
correctly they can revert to
paper prescribing.
Electronic prescription is No existing control to Clinical safety design Clinical safety testing by Surgery procedures
routed to the incorrect prevent immediate effect of requirements to minimise Manufacturer to assure to ensure an
dispensary (D, M) patient experiencing a likelihood. system. alternative
delay in the provision of Prescribing system to prescription can be
their medication. provide a prescription provided in such
If prescriber suspects the cancellation or recall circumstances.
system is not working capability.
correctly they can revert to
paper prescribing.

Copyright © 2018 NHS Digital Page 51 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Appendix B Example Hazard Identification


Techniques
B.1 FFA (Functional Failure Analysis)
B.1.1 Description
A hazard identification technique that takes a functional view of the system and for each
function considers what the potential “safety consequences” may be if:
• the function is lost, i.e. not available when it is required
• the function is wrong, i.e. is available but performs an unintended action
• the function is provided when not required, i.e. function performs as intended but not at
the correct time or out of sequence
Safety consequences document the potential consequence(s) the functional failure may have
on a patient from which a hazard can be identified.
It considers “contributing factors” that may affect the safety outcome. Contributing factors
include environmental conditions or other influences but excludes other functional causes
within the system.
The analysis is captured and documented in a simple table.

B.1.2 Advantages
Simple analytical principles
Systematic and methodical technique which ensures all functionality is considered
Relatively efficient and can be conducted by a small team
Readily identifies those functions that are safety impacting.

B.1.3 Disadvantages
Can yield a massive amount of output if analysis is undertaken at too low a level of functional
decomposition
Can be difficult to apply to systems where information is more important than function.

Copyright © 2018 NHS Digital Page 52 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.1.4 Example Analysis


This example considers the scenario where a Health IT System is being introduced into a social care organisation to provide Service Users
with a facility to retrieve electronic demographic data. The availability or non-availability of paper based records within the social care
organisation will have a direct impact on the safety consequences. Numbers are used to map the contributing causes with the respective
safety consequence.
Function: Retrieve Personal Details
Failure Mode Contributing Factors Safety Consequences
No Function: 1 No Paper records 1 Delay in identifying Service User and completing contact assessment if
Service user demographics details are not 2 Paper records are available demographic details are not available.
available electronically to the Care Professional and up to date. Subsequent delay in assessing care needs and providing care services.
3 Paper records are available Care services not provided.
but are out of date. Hazard: Delay in provision of care services.
Hazard: No provision of care services
2 Care professional would refer to paper records and conduct assessment
on that basis.
Little or no Service User harm
3 Care professional would refer to paper records but would conduct
assessment on out of date data.
Potential for provision of wrong or inappropriate care services.
Hazard: Inappropriate care services provided.
Incorrect Function: 1 No Paper records. 1 Care Professional conducts contact assessment using wrong
Incorrect Service user demographic details 2 Paper records are available demographic information. Subsequent care services could be based on
presented electronically to the Care Professional and up to date. another Service User’s care history.
3 Paper records are available Hazard: Inappropriate care services provided.
but are out of date.
2 Care professional cross-checks with paper records (assumption).
Little or no Service User harm

Copyright © 2018 NHS Digital Page 53 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

3 Care professional cross-checks with paper records (assumption) and


conducts contact assessment using out of date data. Potential for provision
of wrong or inappropriate care services.
Hazard: Inappropriate care services provided
Function Provided when not required None Distraction whilst using IT system
No Service User harm - usability issue.

This table can be readily expanded to support hazard risk assessment.

Copyright © 2018 NHS Digital Page 54 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

Assessing the hazard “Delay in provision of care services” and considering the failure mode “No function” it’s reasonable to establish a
cause as being a loss of connection to Spine within the system. The contributing factors can now be thought of as risk controls which in turn
influence (in this example) the “Severity” assessment:
Function: Retrieve Personal Details
Hazard Failure Mode Cause Existing Controls Severity Likelihood Risk
Delay in No Function: No connection None Significant – delay in provision of Medium – 2 Acceptable where cost
provision of Service user to Spine care services could result in unproven new of further reduction
care services demographics details discomfort or short term harm. Not system outweighs benefits
are not available considered credible that care gained.
electronically to the services would not be provided.
Care Professional No connection Availability of Minor - Contact assessment can Medium – 2 Acceptable where cost
to Spine paper records continue using existing paper unproven new of further reduction
based details (Assumption: paper system outweighs benefits
records are maintained and up to gained.
date).
Significant – severity of outcome High – 3 Undesirable Attempts
increases in circumstances where introduction of should be made to
paper records are not maintained. IT will reduce eliminate the hazards or
Delay in provision of care services paper implement control
could result in discomfort or short housekeeping measures to reduce risk
term harm. Not considered to an acceptable level
credible that care services would
not be provided.

Copyright © 2018 NHS Digital Page 55 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.2 HAZID (Hazard Identification)


B.2.1 Description
A hazard identification technique that focuses on the characteristics of information flow within
a system. It is a structured brain storming technique which uses characteristic keywords (e.g.
None, Wrong, Late, Incomplete Duplicate) to consider what the potential safety
consequences may be. It can also be used to document the initial risk assessment.
The analysis is captured and documented in a simple table.

B.2.2 Advantages
Simple analytical principles
Systematic and methodical technique which ensures complete coverage
Readily identifies those functions that are safety impacting
Highly suited to IT systems where focus is the messaging and display of information.

B.2.3 Disadvantages
Can yield a massive amount of output if analysis is undertaken at too low a level of data
definition
Generally requires a larger team and an experienced facilitator to keep task on track.

Copyright © 2018 NHS Digital Page 56 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.2.4 Example Analysis


This example considers the same scenario as B1 but the focus is the information (Demographic Details) rather than a particular function.
Deviations in the intended characteristics of that data can then be considered to identify the potential safety consequences.
Contact Assessment
Demographic Details (this is the information flow being analysed)
Deviation Keyword Cause Existing Controls Consequences & Severity Likelihood Risk
None
Wrong
Late
Incomplete Spine PDS record Outside scope of Health IT N/A N/A N/A
incomplete System influence
Incomplete record Availability of paper based Minor - Missing information Medium – unproven 2 Acceptable where cost of
retrieved by Health records to yield missing would be apparent and Contact new system further reduction outweighs
IT System data Assessment can continue using benefits gained.
paper based information.

Duplicate

Copyright © 2018 NHS Digital Page 57 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.3 SWIFT (Structured What-IF Technique)


B.3.1 Description
A hazard identification technique that uses pertinent questions to explore the consequences
of unintentional actions. It can include functions, information and users. It takes the form of a
brain storming exercise with a strong cross section of relevant expertise.
The analysis is captured and documented in a simple table.

B.3.2 Advantages
Simple process that is easily followed.
Can identify significant issues quite quickly.

B.3.3 Disadvantages
Need relevant and appropriate set of questions if all issues are to be uncovered.
Can be difficult to do for the first time although generic themes (User Interface, Design
Features, System Integration, Equipment Failures, Training) can be explored.
Can involve a large number of people.
Needs an experienced facilitator to ensure analysis stays focused.

Copyright © 2018 NHS Digital Page 58 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.3.4 Example Analysis


Task Initial Contact Assessment
Description of Task (input / output activity resource, equipment)
Care Professional conducts contact assessment using local IT system to retrieve personal demographic details from NHS Spine to establish and confirm identity of
Service User.
Current Risk
What If Question Cause Consequence Safeguards
S L R
Presenting Service User is unable to
provide identity
Care Professional is unable to enter right
data
IT system doesn’t retrieve correct data from
Spine

Copyright © 2018 NHS Digital Page 59 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.4 Fishbone Diagram


B.4.1 Description
A root cause analysis technique which uses the concept of a fishbone to capture the causes
to a pre-defined hazard.
A diagrammatical technique that enables the causes and the relationships between them to
be conveyed.

B.4.2 Advantages
Intuitive technique that is easy to use and findings are readily conveyed
Generic themes can be used to prompt consideration of potential causes.

B.4.3 Disadvantages
Identification of hazard is a pre-requisite.
Limited expression of logical relationship between identified causes.

Copyright © 2018 NHS Digital Page 60 of 61


Clinical Risk Management: its Application in the Manufacture of Health IT Systems -
Implementation Guidance v3.2 02.05.2018

B.4.4 Example Analysis

User problem/ Human Factors Terminology Team and Social Factors

Design Features Factor Resource Factors

Hazard

Equipment and Failure Factors Safety Factors

Education and Training Factors Communication Factors System Integration Factors

Copyright © 2018 NHS Digital Page 61 of 61