Vanquish Pumps C, Pumps F (VC-PXX, VF-PXX) - Operating Manual
Vanquish Pumps C, Pumps F (VC-PXX, VF-PXX) - Operating Manual
Vanquish Pumps C, Pumps F (VC-PXX, VF-PXX) - Operating Manual
Vanquish
Pumps C, Pumps F
VC-Pxx, VF-Pxx
Operating Manual
4820.4405-EN Revision 5.1 • May 2023
*4820.4405-EN-5.1*
Copyright © 2023 Thermo Fisher Scientific Inc. All rights reserved.
The hardware descriptions in this manual revision refer to devices: VC-P10-A-01, VC-P20-A-01,
VC-P21-A-01, VC-P32-A-01, VC-P33-A-01, VC-P40-A-01, VF-P10-A-01, VF-P20-A, VF-P32-A-01.
Trademarks
Acrobat, Adobe, and Adobe Reader are trademarks of Adobe Systems Incorporated.
Microsoft and Windows are trademarks of Microsoft Corporation.
MP35N is a trademark of SPS Technologies.
PharMed is a trademark of Saint-Gobain Performance Plastics.
Torx is a trademark of Acument Intellectual Properties, LLC.
All other trademarks are property of Thermo Fisher Scientific and its subsidiaries.
Disclaimer
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to
use in the product operation. The document is copyright protected; any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization
of Thermo Fisher Scientific Inc.
This manual is provided "as is." The contents of this manual are subject to being changed,
without notice, in future revisions.
Thermo Fisher Scientific Inc. makes no representations that this document is complete,
accurate, or error-free. Thermo Fisher Scientific Inc. assumes no responsibility and will not be
liable for any errors, omissions, damage, or loss that might result from any use of this document,
even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale.
The Terms and Conditions of Sale shall govern all conflicting information between the two
documents.
Manufacturer's address
Dionex Softron GmbH, Part of Thermo Fisher Scientific, Dornierstrasse 4, D-82110 Germering
Contacting Us
Contacting Us
There are several ways to contact us:
Ordering Information
For ordering information or sales support for HPLC products, contact
your local Thermo Fisher Scientific sales organization. For contact
information, go to Contact Us on http://www.thermofisher.com.
Technical Assistance
For technical support for HPLC products, contact your local Thermo
Fisher Scientific support organization. For contact information, go to
Contact Us on http://www.thermofisher.com.
Contents
1 Using this Manual....................................................................... 11
1.1 About this Manual ............................................................................................... 12
1.2 Conventions ......................................................................................................... 13
1.2.1 Conventions for Safety Messages ............................................................ 13
1.2.2 Special Notices ......................................................................................... 13
1.2.3 Typographical Conventions ...................................................................... 14
1.3 Reference Documentation................................................................................... 15
6 Operation................................................................................... 81
6.1 Introduction to this Chapter ................................................................................ 82
6.2 Safety Guidelines for Operation........................................................................... 83
6.3 Control Elements ................................................................................................. 84
6.3.1 Keypad...................................................................................................... 84
6.3.2 Status Indicators....................................................................................... 86
6.4 Power On/Off Control.......................................................................................... 89
6.5 Use of Solvents and Additives.............................................................................. 90
6.6 Preparing the Device for Operation..................................................................... 92
6.7 Important Operating Parameters ........................................................................ 94
6.8 Purging the Pump .............................................................................................. 100
6.8.1 Automatic Purging.................................................................................. 101
6.8.2 Manual Purging ...................................................................................... 102
6.9 Optimizing the Performance of the Device........................................................ 107
6.9.1 General Guidelines ................................................................................. 107
6.9.2 Installing Solvent Shut-Off Valves .......................................................... 107
6.10 Shutting Down the Device ................................................................................. 110
6.10.1 Short-Term Shutdown (Interruption of Operation) ............................... 110
6.10.2 Long-Term Shutdown............................................................................. 111
6.10.3 Restart after Long-Term Shutdown........................................................ 113
13 Appendix.................................................................................. 281
13.1 Compliance Information .................................................................................... 282
13.1.1 Declarations of Conformity .................................................................... 282
13.1.2 WEEE Compliance .................................................................................. 283
13.1.3 FCC Compliance...................................................................................... 283
13.1.4 Manual Release History.......................................................................... 284
13.2 Digital I/O ........................................................................................................... 285
1.2 Conventions
This section describes the conventions that are used throughout this
manual.
Make sure that you understand and follow all safety messages
presented in this manual.
NOTICE
Highlights information necessary to prevent damage to the device or
invalid test results.
• Input that you enter by the keyboard or that you select with the
mouse
Viewpoint
If not otherwise stated, the expressions left and right in this manual
always refer to the viewpoint of a person that is facing the device from
the front.
• Index entries
Hardware Documentation
Additional hardware documentation includes the following:
• Operating manuals for the other modules of the Vanquish system
• Vanquish System Operating Manual
• Instrument Installation Qualification Operating Instructions
Thermo Fisher Scientific provides up-to-date operating manuals as PDF
(Portable Document Format) files that you can access from our customer
manuals web site. To open and read the PDF files, Adobe™ Reader™ or
Adobe™ Acrobat™ is required.
Go to the following web site: www.thermofisher.com/HPLCmanuals
Software Documentation
Additional software documentation includes the following:
• Installation Guide
For basic information about device installation and configuration,
refer to the Installation Guide.
• Reference Card
For a concise overview of the most important workflows, refer to
the Reference Card.
TIP The Chromeleon Help and documents are included in the software
shipment.
Third-Party Documentation
Refer also to the user documentation provided by the manufacturers of
third-party components and materials, for example, Safety Data Sheets
(SDSs).
2 Safety
This chapter provides general and specific safety information and
informs about the intended use of the device.
The safety symbols and signal words in this manual include the
following:
CAUTION
Indicates a hazardous situation that, if not avoided, could result in minor
or moderate injury.
WARNING
Indicates a hazardous situation that, if not avoided, could result in
serious injury.
• Always keep the manual near the device for quick reference.
Symbol Description
— Power supply is on
Ο Power supply is off
Indicates alternating current.
TIP An additional type label on the leak tray of the device indicates the
module name, serial number, part number, and revision number (if any).
To facilitate device identification, have the information from this label
available when communicating with Thermo Fisher Scientific.
Laboratory Practice
Thermo Fisher Scientific recommends that the laboratory in which the
Vanquish system is used follow best practices for LC analyses. This
includes among others:
• Establishing shelf life limits and following them for all consumables
used with the system
Safety Standard
This device is a Safety Class I instrument (provided with terminal for
protective grounding). The device has been manufactured and tested
according to international safety standards.
Installation
General Operation
All users must know the hazards presented by the device and the
substances they are using. All users should observe the related Safety
Data Sheets (SDSs).
Protective Clothing
To protect you from chemical splashes, harmful liquids, or other
contamination, put on appropriate protective clothing, such as a lab
coat.
Protective Eyewear
To prevent liquids from striking your eyes, put on appropriate protective
eyewear, such as safety glasses with side shields. If there is a risk of
splashing liquids, put on goggles.
Gloves
To protect you from harmful liquids and avoid personal injury during
maintenance or service, put on appropriate protective gloves.
WARNING—Hazardous Substances
Solvents, mobile phases, samples, and reagents might contain toxic,
carcinogenic, mutagenic, infectious, or otherwise harmful substances.
The handling of these substances can pose health and safety risks.
• Be sure that you know the properties of all substances that you are
using. Avoid exposure to harmful substances. If you have any doubt
about a substance, handle the substance as if it is potentially
harmful.
• Wear personal protective equipment as required by the hazard and
follow good laboratory practice.
• Reduce the volume of substances to the minimum volume required
for sample analysis.
• Avoid handling of solvent reservoirs above head height.
• Do not operate the device in a potentially flammable environment.
• Avoid accumulation of harmful substances. Make sure that the
installation site is well ventilated.
• Dispose of hazardous waste in an environmentally safe manner that
is consistent with local regulations. Follow a regulated, approved
waste disposal program.
WARNING—Biohazard
Biohazardous material, for example microorganisms, cell cultures,
tissues, body fluids, and other biological agents can transmit infectious
diseases. To avoid infections with these agents:
• Assume that all biological substances are at least potentially
infectious.
• Wear personal protective equipment as required by the hazard and
follow good laboratory practice.
• Dispose of biohazardous waste in an environmentally safe manner
that is consistent with local regulations. Follow a regulated,
approved waste disposal program.
WARNING—Self-Ignition of Solvents
Solvents with a self-ignition temperature below 150 °C might ignite
when in contact with a hot surface (for example, due to leakage in the
chromatography system).
Avoid the use of these solvents.
WARNING—Hazardous Vapors
Mobile phases and samples might contain volatile or flammable
solvents. The handling of these substances can pose health and safety
risks.
• Avoid accumulation of these substances. Make sure that the
installation site is well ventilated.
• Avoid open flames and sparks.
• Do not operate the device in the presence of flammable gases or
fumes.
CAUTION—Allergic Reaction
Some capillaries in the system are made of MP35N™, a nickel/cobalt-
based alloy. Individuals with sensitivity to nickel/cobalt may show an
allergic reaction from skin contact.
WARNING—Safety Hazard
In case of emergency, disconnect the device from the power line.
• Use only solvents and additives that are compatible with all parts in
the flow path.
VC‑pumps except The pumps are shipped with UHMW-PE piston seals. Using
VC‑P21 and VC‑P33 tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
In rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with high pH,
ammonium hydroxide containing mobile phases and
prolonged exposure.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals.
Consider the following:
• With the piston seals, also replace the seal wash seals
(part no. 6040.0306, containing 2 seals for use as both,
piston seal and seal wash seal).
• Install new pistons. Exchanging the piston seals but
keeping the pistons that were used with the UHMW-PE
piston seals may impair the seal life of the PTFE seals.
• Note that slightly increased abrasion may be observed
with these seals, depending on the application. Consider
testing the permeability of the static mixer regularly at
short intervals.
VC‑P21 and VC‑P33 The pumps are shipped with carbon-fiber filled PTFE piston
pumps seals. Slightly increased abrasion may be observed with
these seals, depending on the application. Consider testing
the permeability of the static mixer regularly at short
intervals.
VF‑pumps The pumps are shipped with UHMW-PE piston seals. Using
tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
In rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with high pH,
ammonium hydroxide containing mobile phases and
prolonged exposure.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals.
Consider the following:
• With the piston seals, also replace the seal wash seals
(part no. 6040.0306, containing 2 seals for use as both,
piston seal and seal wash seal).
• Install new pistons. Exchanging the piston seals but
keeping the pistons that were used with the UHMW-PE
piston seals may impair the seal life of the PTFE seals.
• In addition, note the following:
¨ Slightly increased abrasion may be observed with the
PTFE seals, depending on the application. Consider
testing the permeability of the static mixer regularly at
short intervals.
¨ The pump should not be operated with pressures
higher than 70 MPa.
¨ The pump is no longer biocompatible.
• Refer also to the Operating Manuals for all modules in the Vanquish
system. They may provide additional guidelines and information.
NOTICE
If the system configuration includes a non-standard detector, for
example, a charged aerosol detector or refractive index detector, refer
to the Operating Manual for the detector for specific recommendations
regarding solvents and additives.
Changes that you make to the device may void compliance with one or
more of these EMC and safety standards. Changes to the device include
replacing a part or adding components, options, or peripherals not
specifically authorized and qualified for the product by Thermo Fisher
Scientific. To ensure continued compliance with EMC and safety
standards, replacement parts and additional components, options, and
peripherals must be ordered from Thermo Fisher Scientific or one of its
authorized representatives.
The device has been shipped from the manufacturing site in a safe
condition.
See also
2 Compliance Information (} page 282)
3 Device Overview
This chapter introduces you to the device and the main components.
Main Features
• Active piston seal wash system for high lifetime of pistons and piston
seals
The piston seal wash system keeps the pistons wet and rinses the
piston seals automatically. This prevents crystallization and particle
deposits on the piston surfaces and seals, which may damage the
components.
• Mixing system with capillary mixer and static mixer for smoother
baselines
With the isocratic pump, the mixing system serves as inline filter.
• Binary pump
Solvent selector "2 from 6" for increased flexibility
• Dual pump
The dual pump combines two ternary pumps in a single enclosure.
Pump-Specific Information
The picture shows the flow path through the seal wash system.
Figure 1: Flow path through the seal wash system (left: isocratic pump,
quaternary pump; right: binary pump, dual pump)
No. Description
No. Description
1a Isocratic pump, dual pump, quaternary pump: Capillary from the pump head
Binary pump: Capillary from the right pump head
1b Binary pump only: Capillary from the left pump head
2 System pressure sensor
3 Capillary mixer
4 Purge outlet
5 Purge valve knob
Binary Pump
With the binary pump, the purge/sensor unit also combines the solvent
streams from the pump heads.
• Capillary mixer
In the capillary mixer, the solvent streams delivered by the pump are
combined and premixed.
• Static mixer
The static mixer improves the mixing quality.
Mixing with capillary mixer and static mixer results in smoother
baselines.
TIP Mixing systems with other volumes are available as an option (see
Gradient Delay Volume, Dwell Volume and Ripple (} page 236)).
The leak sensor inside the device monitors the device for liquid leaks
from the flow connections. The liquid is collected in the leak tray and
guided to the drain port. From the drain port, the liquid is discharged to
waste through the drain system of the Vanquish system.
When the leak sensor detects leakage, the status indicators change to
red and beeping starts to alert you. Follow the instructions in this
manual to find and eliminate the source for the leakage.
If leakage is not resolved after 3 minutes, the pump stops the flow.
3.8 Operation
The device is designed to be operated from a computer configured with
the Chromeleon Chromatography Data System (CDS). The Chromeleon
software provides complete instrument control, data acquisition, and
data management.
TIP The device can be operated also with other data systems, such as
Thermo ScientificTM XcaliburTM. In this case, installation of additional
software is required in addition to the data system software. For details,
contact the Thermo Fisher Scientific sales organization.
4 Unpacking
This chapter provides information for unpacking the device and informs
you about the scope of delivery.
4.1 Unpacking
Damaged Packaging, Defective on Arrival
Inspect the shipping container for signs of external damage and, after
unpacking, inspect the device for any signs of mechanical damage that
might have occurred during shipment.
If you suspect that the device may have been damaged during shipment,
immediately notify the incoming carrier and Thermo Fisher Scientific
about the damage. Shipping insurance will compensate for the damage
only if reported immediately.
Tools required
Screwdriver, TorxTM T20
3. Remove the device from the shipping container: Grasp the device by
the carrying handles. Slowly and carefully, lift the device out of the
shipping container.
No. Component
1 Carrying handles
2 Attachment screw (one on each carrying handle)
5. If applicable:
Remove any additional packing material. Leave any protective films
attached to the surfaces of the device until it is properly positioned
in the system stack.
8. Slide off the carrying handles from the rails towards the rear of the
device.
Figure 4: Sliding off the carrying handle from the left rail
TIP Keep the shipping container, the carrying handles with the
attachment screws, and all packing material. These items will be needed
if the device is transported to a new location or shipped.
• Pump
• Ship Kit
• Power cord
For information on contents of the ship kit or reordering parts, see
Accessories, Consumables and Replacement Parts (} page 259).
5 Installation
This chapter specifies the requirements for the installation site and
describes how to set up, install, and configure the device in the Vanquish
system and in the chromatography software.
NOTICE
The device is part of the Vanquish system. Therefore, follow the order
for installing the system modules as described in the Vanquish System
Operating Manual.
TIP
Before turning on the power to a Vanquish system module for the first
time, verify that the chromatography software is installed on the data
system computer. When the power is turned on, the required USB
drivers are automatically found and the Windows™ operating system
can detect the device.
5. Set up the device in the software. See Setting Up the Device in the
Software (} page 80).
6. Recommended:
Perform Instrument Installation Qualification.
In the Chromeleon software, a wizard is available to guide you
through the qualification process. On the Chromeleon 7 Console:
Click Tools > Instrument Qualification > Installation Qualification.
Follow the instructions in the Instruments Installation Qualification
Operating Instructions. The manual provides information about the
required materials and detailed instructions.
NOTICE
If the device is operated with another data system, refer to the
documentation for the software that you are using and/or perform the
qualification manually. The Instruments Installation Qualification
Operating Instructions provide information about the parameters to be
adapted and the required settings.
5.3.3 Condensation
No. Description
1 Solvent Rack
2 Detector
3 Autosampler
4 Pump
5 System Base
6 Column Compartment
For instructions on how to set up the system stack, refer to the Vanquish
System Operating Manual.
No. Description
1 Rating plate (example), indicating the serial number, part number, module
name, revision number (if any), and the line and fuse rating
2 Main power switch (on/off control)
3 Fuse holder
4 Power-inlet connector
5 System Interlink port
Allows power on/off control for the device from the Vanquish system base
and device communication
6 Digital I/O ports (Dig I/O)
Allow exchange of digital signals with external instruments
Each digital I/O port provides one input, one relay output, and one
bidirectional input/output. For connection and pin assignment information,
see Digital I/O (} page 285).
7 USB hub ("A" type connector)
Allows connection to other modules in the Vanquish system
8 USB (Universal Serial Bus) port ("B" type connector)
Allows connection to other modules in the Vanquish system or the
computer on which the data management system is installed, such as the
Chromeleon software
TIP Thermo Fisher Scientific recommends using the USB ports only as
described above. If the USB ports are used for any other purpose,
Thermo Fisher Scientific cannot ensure proper functionality.
NOTICE
• Never use defective communication cables. If you suspect that a
cable is defective, replace the cable.
• To ensure trouble-free operation, use only the cables provided by
Thermo Fisher Scientific for connecting the device.
NOTICE
Condensation in a device can damage the electronics.
• Before connecting the devices to the power line, be sure that no
condensation is present in the devices.
• If you suspect that condensation is present, allow the device to
warm up to room temperature slowly. Wait until the condensation is
completely gone before proceeding.
2. Connect the power cord to the power inlet connector on the device.
NOTICE
When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path. To flush
the Vanquish modules, follow the instructions in the Vanquish System
Operating Manual.
1. Connect the pump to the drain system (see Guiding Liquids to Waste
(} page 62)).
3. Set up the seal wash system (see Seal Wash System (} page 69)).
Figure 8: Tubing chase with tubing guides (left: view from inside, right:
view from top)
Tubing Brackets
Tubing brackets are available for holding the tubing in place. Slip the
bracket side onto the drain pipe.
• Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the product or additional or
spare capillaries and tubing as recommended by Thermo Fisher
Scientific.
• Install capillaries and fittings only at the positions for which they are
intended.
No. Description
NOTICE
• Tighten or loosen Viper capillaries only with your fingers and the
special torque tool. Do not use tools others than those provided by
Thermo Fisher Scientific for this purpose. Use the tool only for the
connections for which it is intended.
• To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries only when the system pressure is down to zero.
• Mind the correct orientation of the tool when tightening or
loosening connections (see pictures).
No. Description
1 Capillary
2 Torque toothing
4. Tighten the connection until the torque limitation has been reached,
which is indicated by a clicking sound.
NOTICE
• Tighten or loosen Viper capillaries only with your fingers. Do not use
tools other than the knurl that comes with the capillary.
• To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries only when the system pressure is down to zero.
No. Description
1 Knurl
2 Capillary
3 Slot
TIP Note the slot in the knurl. You can easily remove the knurl from the
capillaries through this slot if space is limited.
NOTICE
Overtightening may damage the connection (both, the ferrule and the
connection port). With a torque wrench, the torque should not exceed
3 Nm (for steel capillaries).
NOTICE
To avoid increased dead volume, leakage, or damage to the system,
reuse fittings and ferrules only for the same capillary connection.
• Waste liquid from the purge unit: When purging the pump, attach
silicone tubing to the purge outlet on the purge/sensor unit and
guide the free end of the tubing to the drain port outlet.
No. Description
1 Waste line from purge unit: routing path to drain port outlet (example)
2 Drain port
3 Drain port outlet (to Vanquish drain system)
4 Leak sensor
For information about how the liquid is discharged to waste through the
Vanquish drain system, refer to the Vanquish System Operating Manual.
The flow path determines the pump inlet (component at which the
solvent enters the pump) and the solvent flow through the pump.
• Solvent lines
• Tubing cutter
Preparations
3. On the pump inlet, remove the plug that protects the inlet during
shipment, if applicable.
4. For easier access to the lower degas chambers, remove the seal
wash detector assembly (funnel and detector).
No. Description
No. Description
TIP To avoid the formation of air bubbles in liquid lines, always place
reservoirs in the solvent rack on top of the system stack.
1. Connect each solvent line to the pump inlet port for which it is
intended (see the labeling on the solvent lines and in the pump).
2. Route the solvent lines from the pump to the solvent rack, through
the tubing guides in the pump and all modules above the pump in
the system stack. Observe the routing guidelines in Guiding
Capillaries and Tubing Through the System (} page 56).
3. Route the solvent lines through the guide hole in the solvent rack
and fix them in the appropriate tubing guides.
No. Description
No. Description
1 Solvent line
2 Retaining guide
3 Reservoir cap
4 Cap plugs
5 Solvent line filter
When
To shut off the solvent flow through the system at the solvent reservoir,
for example, to prevent solvent from flowing through the system when
you open a flow connection on the low-pressure side
• Tubing cutter
Preparations
Prepare the pump and follow the steps in Connecting the Solvent Lines
(} page 63).
No. Description
1 Solvent line
2 Fitting
3 Ferrule
4 Shut-off valve
1. Cut the solvent line about 10 cm from the reservoir cap with a
tubing cutter. Make sure that the cut is at a right angle to the length
of the line.
2. Onto the free end of each solvent line, slide a fitting and a ferrule.
Mind the correct orientation of the fitting and ferrule (see picture).
No. Description
1 Shut-off valve
2 Retaining guide
4. Check the solvent lines across the entire flow path. Make sure that
the lines are not bent, pinched, or squeezed at any point in the flow
path.
• In the pump: Seal wash pump (peristaltic pump), seal wash lines,
and seal wash detector (drop detector)
• The seal wash liquid is miscible with the solvent that is used in the
pump.
• The seal wash liquid is compatible with the seal wash tubing.
TIP To avoid the formation of air bubbles in liquid lines, always place
reservoirs in the solvent rack on top of the system stack.
Figure 22: Seal wash system in the Vanquish system (examples with a single split sampler,
left: with quaternary pump, right: with binary pump, as example for two-head pump)
No. Description
Preparations
Rinse the seal wash reservoir thoroughly. Use a high-purity solvent for
rinsing.
No. Description
1 Pump lever
2 Pump rotor
3 Tubing holder (peristaltic tubing)
4 Tubing connector
5 Seal wash line from autosampler (silicone tubing)
6 Peristaltic tubing
7 Seal wash tubing to pump head (silicone tubing)
1. Press down the lever on the seal wash pump and hold it in that
position with one hand.
2. With your other hand, insert the tubing between the lever and the
rotor and wind it around the rotor.
5. Check that the right tubing connector rests on the tubing holder. If it
does not, push the tubing connector onto the holder.
1. In the autosampler, locate the seal wash line (clear silicone tubing)
to the pump:
¨ Single split sampler: The seal wash inlet port is located on the
top enclosure frame (right tubing connector). The seal wash line
to the pump is interconnected to the seal wash inlet port.
¨ Dual split sampler: The seal wash inlet port is located on the
right enclosure frame (top tubing connector). The seal wash line
to the pump interconnects seal wash inlet and outlet port.
2. Disconnect the seal wash line from the seal wash inlet port.
3. Route the seal wash line from the seal wash outlet in the
autosampler through the tubing guides to the peristaltic tubing in
the pump.
¨ Single split sampler: The seal wash outlet is located on the
bottom end of the metering device.
¨ Dual split sampler: The seal wash outlet is located on the right
enclosure frame (bottom tubing connector).
No. Description
A Single split sampler: Seal wash connections on the top enclosure frame
B Dual split sampler: Seal wash connections on the right enclosure frame
1 Seal wash inlet port
1a Seal wash line to reservoir
No. Description
4. In the pump, connect the seal wash line to the tubing connector on
the free end of the peristaltic tubing.
1. Locate the seal wash line for connection from the autosampler to
the seal wash reservoir.
3. Route the seal wash line from the autosampler to the solvent rack
through the tubing guides in the autosampler and all modules above
the autosampler in the system stack.
4. Route the seal wash line through the guide hole in the solvent rack
and fix it in the appropriate tubing guide.
No. Description
No. Description
1 Wash line
2 Retaining guide
3 Reservoir cap
4 Cap plugs
2. Feed the wash line through an open hole in the cap of the wash
reservoir. The retaining guide keeps the tubing in place in the
reservoir.
3. Close any open holes in the reservoir cap with cap plugs.
5. Tighten the reservoir cap hand-tight. Press the retaining guide into
the hole in the reservoir cap so that the tubing is kept in place in the
cap.
During the seal wash cycle, the seal wash system is automatically filled
with the seal wash liquid. It will take 5 minutes for the liquid to travel
through the entire flow path.
No. Description
3. To allow the liquid to travel easily through the seal wash system,
press down the lever of the seal wash pump and hold it in that
position with one hand.
4. Pull out the syringe plunger to draw seal wash liquid into the tubing.
5. Release the lever of the seal wash pump when the liquid reaches the
syringe.
6. Remove the syringe and reconnect the tubing to the connection port
on the seal wash detector.
Mind the port labeling. To ensure correct functioning of the seal
wash detector, the tubing needs to be connected to the port labeled
in.
Figure 28: Guide holes in the autosampler and pump (left: view from VF‑/
VH-autosampler (example); right: view from pump)
No. Description
1. Check that the power button on the front left of the Vanquish
system base (system power button) is pressed in. If the power
button stands out, press the power button to turn on the power on
the system base.
• The pump starts operation with a seal wash cycle. Before turning on
the pump, check the liquid level in the seal wash reservoir. Consider
checking also the liquid level in the waste container.
See also
2 Power On/Off Control (} page 89)
The Help for the software that you are using provides detailed
information about the settings on each property page.
6 Operation
This chapter describes the elements for device control, provides
information for routine operation and for shutdown.
NOTICE
Pay attention also to the following guidelines:
• To prevent damage resulting from leakage or from running the
pump dry, always set the lower pressure limit for the pump.
• If there is evidence of leakage in the device, turn off the pump flow
and remedy the situation immediately.
• If the pump flow is interrupted, act appropriately to protect the
components in the detector. For details, refer to the Operating
Manual for the detector.
• Always verify that the autosampler is turned on before the pump
flow is on and pressure builds up. If the autosampler is turned off,
for example, after a power failure, stop the pump flow and wait until
the pressure is down to zero before turning on the autosampler or
other modules again.
• Keypad
The keypad buttons allow you to perform certain functions directly
from the device.
• Status indicators
The LEDs (Light Emitting Diodes) on the status indicator LED bar on
the front side of the device and the STATUS LED on the keypad
provide a quick visual check of the operational status of the device.
6.3.1 Keypad
The keypad inside the device allows you to perform certain functions
directly from the device. When you press a button, a short beep
confirms that the function is performed.
Figure 29: Keypad
STATUS
The STATUS LED provides a quick visual check of the operational status
of the device.
When the doors are closed, the LED bar on the front side indicates the
operational status.
MUTE ALARM
Beeping alerts you when the device detects a problem, for example,
leakage. To turn off the beep for the current alarm, press this button.
Eliminate the source for the alarm within 10 minutes. Otherwise,
beeping starts again. If the device detects a different problem, beeping
alerts you again immediately.
SELECT
Binary pump and dual pump
Pressing the button selects the pump head for you which you can
perform a DOCK or PURGE function from the keypad. The LEDs next to
the button indicate the selection:
LED Description
DOCK
Pressing the button moves the pistons of the selected pump head to the
appropriate position for maintenance procedures, such as pump head or
piston replacement, and undocks the pistons.
When undocking is complete, the LED next to the FLOW button starts
flashing green for the selected pump head. The LED continues flashing
while no pistons are docked. Pressing the button again docks the
pistons. Docking is complete when the LED stops flashing green.
• Before you press the button to undock the pistons, open the purge
valve by turning the valve knob counterclockwise.
PURGE
Pressing the button starts a purge cycle. Pressing the button while a
purge cycle is running stops the cycle. See Purging the Pump
(} page 100).
FLOW
Pressing the button starts or stops the flow, considering the selected
flow rate, solvent composition, and flow acceleration/deceleration.
LED Description
The status LED bar on the front side of the device and the STATUS LED
on the inside keypad provide information about the device status.
LED Bar
The LED bar colors provide the following information:
TIP—Dual pump only: With a dual pump, the LED bar is a two-part bar,
with the lights on the left referring to the left pump and the lights on the
right referring to the right pump.
STATUS LED
The STATUS LED on the keypad inside the device provides the following
information:
For information about the LEDs that are present next to a button on the
keypad, see Keypad (} page 84).
Red, flashing The pistons are not in the appropriate position for
normal operation (for example, after piston (seal)
replacement) or an error occurred during internal
maintenance (see Internal Maintenance (} page 120)).
In both cases, an accompanying message appears in the
Instrument Audit Trail when you try to start the flow,
for example, "The pump drive is still in undock
position; Please execute a dock command" (code
4127). Follow the remedial actions for the
accompanying message in Messages (} page 200).
For easier handling, you can use the power button on the front left of
the Vanquish system base (system power button) for power on/off.
• When the power is on, the system power button is pressed in. When
the power is off, the system power button stands out.
• If the main power switch on a device is off, you cannot turn on the
device with the system power button.
• To turn off a device completely, you have to turn it off with the main
power switch on the device. Pressing the system power button will
not be sufficient to turn off the power to the device completely.
• The pump starts operation with a seal wash cycle. Before turning on
the pump, check the liquid level in the seal wash reservoir. Consider
checking also the liquid level in the waste container.
Solvent Reservoirs
• Before you change solvents, make sure that the new solvent is
miscible with the previous solvent. If the solvents are not miscible,
flocculation may occur.
Mix immiscible solvents with an intermediate solvent to replace
them step by step. For example, use isopropanol.
NOTICE
Flush the system flow path thoroughly before operating the device for
the first time:
• When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path. To
flush the Vanquish modules, follow the instructions in the Vanquish
System Operating Manual.
• Some components of the device are filled with isopropanol when
the device is shipped from the manufacturing site. When operating
the device for the first time, use solvents that are miscible with
isopropanol. If they are not, use an appropriate intermediate
solvent.
• Verify that any air bubbles are completely flushed out of the system
flow path.
• Check the liquid level in the solvent reservoirs. Verify that the
amount of solvent is sufficient for the analysis.
• Close the doors of all modules in the Vanquish system, if not already
done.
NOTICE
During pumping cycle keep the pump doors closed. Otherwise,
functionality may be adversely impacted.
System Equilibration
System equilibration should include the following operations:
• Purging the pump (all channels, including those not used for the
application)
Flow Parameters
Parameter Description
Pressure Parameters
Parameter Description
Parameter Description
Solvent composition In the dialog box for the pump in the Instrument
Configuration Manager, set the number of the
solvents to be used with the pump. You can change
the standard solvent names as required. The solvent
names appear in the Chromeleon user interface.
In the Chromeleon Client, set the solvent composition.
For more information, see the related section in
Pump-Specific Information (} page 215).
Solvent consumption You can monitor the liquid level in the solvent
reservoir (solvent consumption) when you have
entered the required information for the related
solvent. Based on this information, the Chromeleon
software calculates the liquid level. To monitor the
liquid level in real time, based on real physical
measurement, install the Vanquish Solvent Monitor.
The solvent monitor is available as an option (see
Optional Accessories (} page 267)).
For monitoring from the Chromeleon software, enter
the following information:
• Volume of solvent in the reservoir (at the beginning
of a sequence)
• Lower limit for the solvent in the reservoir, which is
the minimum volume of liquid that must be
available in the reservoir
The Remain Time property for the solvent reports
the time left until the liquid level is expected to
reach the lower limit. The Chromeleon software
calculates the time from the current flow rate and
volume entered.
• Warning limit when you want to be informed about
the liquid level in the reservoir
When the liquid in the reservoir reaches the lower
limit, the Chromeleon software stops the pump as
specified in the Emergency Instrument Method if
available or aborts the Queue and stops the pump
flow.
Relevant for dual pump only
You can monitor the solvent consumption of the dual
pump as a whole if the two pumps share the same set
of solvent reservoirs. If the pumps are connected to
different sets of solvent reservoirs, you can monitor
the solvent consumption separately each pump (see
Monitoring the Solvent Consumption (} page 226)).
Parameter Description
Waste level You can monitor the liquid level in the waste
container when you have entered the required
information. Based on this information, the
Chromeleon software calculates the liquid level. To
monitor the liquid level in real time, based on real
physical measurement, install the Vanquish Solvent
Monitor. The solvent monitor is available as an option
(see Optional Accessories (} page 267)).
For monitoring from the Chromeleon software, enter
the following information:
• Volume of liquid in the waste container at the
beginning of a sequence
• Upper limit for the liquid level in the waste container
The Remain Time property for the waste reports the
time left until the liquid level is expected to reach
the upper limit. The Chromeleon software calculates
the time from the current flow rate and calculated
current liquid level.
• Warning limit when you want to be informed about
the liquid level in the waste container
Relevant for dual pump only
You can monitor the liquid level in the waste
container for the dual pump as a whole if the two
pumps share the same container. If the waste liquid is
guided to different waste containers, you can monitor
the liquid level for each pump. For details, see
Monitoring the Liquid Level in the Waste Container
(} page 227).
Other Parameters
Parameter Description
Parameter Description
Settings
The following settings are considered for the purge cycle:
Setting Description
Procedure Description
Prerequisites
Automatic purging is possible only when the pump is fluidically
connected to the Vanquish autosampler and when both modules have
an appropriate firmware installed.
When
Items required
Solvents suitable for purging
To flush out the isopropanol when operating the pump for the first time,
use solvents that are miscible with isopropanol. If they are not, use an
appropriate intermediate solvent.
1. In the Chromeleon software, set the channel that you want to purge
to 100%.
3. Start the purge cycle. You can start the purge cycle from the
Chromeleon software or by pressing the PURGE button on the
keypad.
The purge cycle ends automatically when the specified purge time
ends. If you want to stop purging before the purge time ends, turn
the purge off from the Chromeleon software or by pressing the
PURGE button on the keypad.
4. Repeat the above steps for all channels, including those that are not
used for the application.
• When the pump is operated for the first time or when solvent lines
are empty, follow the steps in Purging with First Time Operation or
Empty Solvent Lines (} page 102).
Items required
• Silicone tubing
• Tissue
• Syringe
TIP For first time-operation: Silicone tubing and a syringe are available in
the ship kit.
Preparations
3. Insert the syringe tip into the open end of the tubing.
1. In the Chromeleon software, set the channel that you want to purge
to 100%.
3. Binary pump and dual pump: For purging from the keypad, verify
that you have selected the pump head (pump) for which you want to
perform the procedure.
4. Start the purge cycle. You can start the purge cycle from the
Chromeleon software or by pressing the PURGE button on the
keypad.
5. Use the syringe to draw enough solvent for filling the solvent line to
the purge outlet.
6. When the liquid enters the silicone tubing, examine the solvent line.
If the solvent line is completely filled with liquid and no air bubbles
are visible, stop the purge cycle (from the Chromeleon software or
by pressing the PURGE button).
7. Remove the syringe from the silicone tubing, absorbing any liquid
that leaves the tubing with a tissue.
8. Route the free end of the tubing to the drain port outlet (see
Guiding Liquids to Waste (} page 62)).
9. Perform the above steps for all channels, including those that are
not used for the application.
If necessary, empty the syringe before starting the procedure for the
next channel.
10. When purging is complete, close the purge valve by turning the
valve knob clockwise.
NOTICE
Turn the valve knob only with your fingers. If the knob leaks, tighten a
little more. Be careful; overtightening may damage the purge valve
seals.
12. Dual pump only: If applicable, perform the purge procedure also for
the other pump.
NOTICE
Always empty the silicone tubing when purging is complete. When the
purge valve is closed, purge liquid remains in the tubing connected to
the purge outlet. Leaving the liquid in the tubing for a longer period
limits the tubing lifetime.
1. To absorb any liquid that might be leaving the purge outlet when the
tubing is removed, place the tissue under the outlet.
2. On the purge/sensor unit, remove the silicone tubing from the purge
outlet. The other end of the tubing remains in the purge drain port
outlet.
3. Hold the tubing so that any liquid that is present in the tubing flows
off through the drain port outlet.
6.8.2.2 Purging for Removing Air Bubbles or Changing to Different Solvent Type
• To remove air bubbles that might be trapped in the system (pump
head, solvent lines) because air bubbles in the system may lead to
the following problems:
¨ Pressure pulsation
¨ High noise level or pulsation during pump operation
¨ Non-reproducible analysis
TIP When the pump is operated for the first time or when solvent lines
are empty, follow the steps in Purging with First Time Operation or
Empty Solvent Lines (} page 102).
Items required
• Silicone tubing
• Tissue
Preparations
3. Route the free end of the tubing to the drain port outlet (see
Guiding Liquids to Waste (} page 62)).
1. In the Chromeleon software, set the channel that you want to purge
to 100%.
3. Binary pump and dual pump: For purging from the keypad, verify
that you have selected the pump head (pump) for which you want to
perform the procedure.
4. Start the purge cycle. You can start the purge cycle from the
Chromeleon software or by pressing the PURGE button on the
keypad.
5. Perform the above steps for all channels, including those that are
not used for the application.
NOTICE
Turn the valve knob only with your fingers. If the knob leaks, tighten a
little more. Be careful; overtightening may damage the purge valve
seals.
8. Dual pump only: If applicable, perform the purge procedure also for
the other pump.
NOTICE
Always empty the silicone tubing when purging is complete. When the
purge valve is closed, purge liquid remains in the tubing connected to
the purge outlet. Leaving the liquid in the tubing for a longer period
limits the tubing lifetime.
1. To absorb any liquid that might be leaving the purge outlet when the
tubing is removed, place the tissue under the outlet.
2. On the purge/sensor unit, remove the silicone tubing from the purge
outlet. The other end of the tubing remains in the purge drain port
outlet.
3. Hold the tubing so that any liquid that is present in the tubing flows
off through the drain port outlet.
TIP Shut-off valves for solvent lines are available as an option (see
Installing Solvent Shut-Off Valves (} page 107)). These valves allow you
to shut off the solvent flow through the system at the solvent reservoir,
for example, to prevent solvent from flowing through the system when
the pump flow is zero for a longer period.
The steps in this section apply only when installing a shut-off valve after
initial operation of the pump. To install a shut-off valve during initial
setup of the flow connections, follow the steps in Connecting the
Solvent Lines (} page 63).
• Tubing cutter
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Prepare the pump for purging (see Purging the Pump (} page 100)).
No. Description
1 Solvent line
2 Fitting
3 Ferrule
4 Shut-off valve
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
5. Cut the solvent line about 10 cm from the reservoir cap with a
tubing cutter. Make sure that the cut is at a right angle to the length
of the line.
6. Onto the free end of each solvent line, slide a fitting and a ferrule.
Mind the correct orientation of the fitting and ferrule (see picture).
8. Check the solvent lines across the entire flow path. Make sure that
the lines are not bent, pinched, or squeezed at any point in the flow
path.
9. To remove any air from the solvent lines, purge the pump using the
solvent of your application.
No. Description
1 Shut-off valve
2 Retaining guide
Charged aerosol Check that sufficient gas is available to continue gas flowing
detector through the detector. This is to prevent any build-up of
residue from solvents or analytes. Gas must be flowing when
pump flow is delivered to the detector.
UV/VIS detectors The lamp(s) in the detector can remain turned on.
Variable wavelength detector and VH-D10 diode array
detector only: The shutter can be moved to a closed position
for protection of the flow cell.
Fluorescence Turn off temperature control for the flow cell.
detector
• Make sure that the temperature of the column does not exceed
40 °C.
• When resuming operation, let the flow equilibrate and verify that
the operating parameters for the other system modules are set as
required before proceeding.
TIP Shutting down the device affects the operation of the system. When
shutting down the device, also observe the shutting down instructions
for the other Vanquish system modules and take appropriate action
(refer to the Operating Manuals for the modules).
TIP With a Vanquish Core system that has been modified for using
normal-phase compatible solvents and additives, refer to the
information about the flushing liquid in the Considerations with Normal-
Phase Compatible Solvents and Additives section in the Vanquish System
Operating Manual.
Device remains in the Flush the system, for Flush the system with
laboratory after shutdown example with methanol. several volumes of
100% acetonitrile should methanol and water
not be used. (50:50) (for example,
1.0 mL/min for
10 minutes with the
standard system) to
prevent salt buildup in the
fluidics. If the solvents in
the device are not
miscible with water, use
an appropriate
intermediate solvent.
Device shall be Flush the system with Flush the system first with
transported or shipped isopropanol. several volumes of
after shutdown methanol and water
(50:50) (for example,
1.0 mL/min for
10 minutes with the
standard system) to
prevent salt buildup in the
fluidics. If the solvents in
the device are not
miscible with water, use
an appropriate
intermediate solvent.
Afterward, flush the
system with isopropanol.
3. Turn off the pump flow and wait until the system pressure is down
to zero.
TIP Removing the solvent lines from the tubing guides is usually not
required (not even for replacing the slide-in module). However, when
you remove the solvent lines from the tubing guides in the pump and all
modules above the pump in the system stack, be careful not to pull on
other tubing in the guides.
5. Empty the seal wash system. See Emptying the Seal Wash System
(} page 113).
Situation Steps
Device and all other system modules Turn off the system with the system
remain in the system stack and are to be power button on the system base.
turned off
Device shall be transported or shipped If one of the modules shall be removed
after shutdown from the system stack, turn off all
system modules with their main power
switch. Pressing the system power
button will not be sufficient to turn off
the power to the devices completely.
Follow the instructions in Transporting or
Shipping the Device (} page 187).
2. Together with the caps, remove the solvent lines from the
reservoirs.
2. Together with the cap, remove the seal wash line from the seal wash
reservoir.
3. Remove the peristaltic tubing from the seal wash pump. When the
tubing is removed, any liquid that may be present in the seal wash
line, flows off to waste.
2. Reconnect the seal wash line and the solvent lines to the
appropriate reservoirs.
Situation Action
If the device remained in the system Turn on the system with the system
stack and all system modules were power button on the system base.
turned off.
If the device is restarted after transport. Turn on the device with the main power
switch.
6. Purge the pump and then have the pump deliver for a short time.
7 Maintenance and
Service
This chapter describes the routine maintenance and the service
procedures that the user may perform.
The device is designed for easy maintenance and service. The user-
serviceable parts of the device can be accessed from the front. If not
stated otherwise, the maintenance procedures do not require that you
remove the device from the system.
The maintenance procedures do not require that you remove the doors.
However, it is possible to remove a door if this should ever be required
for a specific reason or procedure. If you need to remove a door, follow
the related steps in Replacing the Doors (} page 185).
WARNING—High Voltage
High voltages are present inside the device that could cause an electric
shock.
Do not open the housing or remove protective panels unless specifically
instructed to do so in this manual.
CAUTION—Spraying Solvent
Solvents can spray when under high pressure.
• Stop the pump flow prior to opening the flow path.
• Wait until the system pressure is down to zero.
• When opening the flow path, wear appropriate protective
equipment.
CAUTION—Hydrostatic Pressure
Solvent may spill when you open the flow path. This is due to hydrostatic
pressure in the system when the solvent reservoirs are located above
the pump outlet. Before you loosen a connection in the flow path:
• Turn off the pump flow and wait until the system pressure is down
to zero.
• Unscrew the caps of the solvent reservoirs and remove the solvent
lines together with the caps from the reservoirs.
• Empty the solvent lines. See Emptying the Solvent Lines
(} page 169).
• Retighten the reservoir caps.
General Rules
• Follow all instructions step by step and use the tools recommended
for the procedure.
Depot Repair
• If you need to return the device for depot repair, follow the
instructions in Transporting or Shipping the Device (} page 187).
A maintenance kit is available for the pump, which includes all parts that
are required for routine maintenance (see Consumables and
Replacement Parts (} page 270)).
1. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Install the backpressure capillary on the pump outlet. Guide the free
end of the capillary to the waste container.
3. Replace the frits in the solvent line filters with new frits.
4. Replace the installed waste container with the new waste container
to prevent any chemical reaction between the nitric acid and the
solvent in the waste container.
5. For 1 hour, apply the flow rate of your application and flush the
pump with 6 M nitric acid.
6. Flush the pump with fresh HPLC-grade water until the pH value is
neutral.
7. Remove the waste container with the nitric acid and re-install the
waste container that was installed previously.
8. For 2 hours, apply the flow rate of your application and flush the
pump with fresh acetonitrile.
10. Turn off the pump flow and wait until the system pressure is down
to zero.
NOTICE
Wipe up all liquids spilled onto the system immediately. If surfaces are
exposed for longer periods, these liquids can cause damage.
Decontamination
Decontamination is required, for example, when leakage or spillage has
occurred, or before service or transport of the device. Use a suitable
cleaning detergent or disinfectant to ensure that the treatment renders
the device safe to handle.
Parts required
• Purified water
NOTICE
Observe the following:
• Only use cleaning detergents that will not damage the surfaces of
the system.
• Never use sharp tools or brushes for cleaning any surfaces.
• Do not use sprays for cleaning.
• Prevent cleaning detergent from entering the flow path.
• Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the device.
Liquids can cause a short circuit when getting in contact with the
electronic components.
Preparations
1. Turn off the power to the device and disconnect the power cord
from the power source.
1. Wipe the surfaces clean with a clean, dry, soft, lint-free cloth or
wipe. If necessary, slightly dampen the cloth or wipe with a solution
of lukewarm water and a suitable cleaning detergent.
3. Wipe the cleaned surfaces with purified water to ensure that all
cleaning detergent residues have been removed.
Some counters can be reset to zero after the required action was
performed. To keep the Predictive Performance information up-to-date,
consider resetting the counter when a maintenance, service, or
qualification procedure has been performed.
• To test the seal wash system for leakage, follow the steps in Testing
the Seal Wash System for Leakage (} page 128).
• To replace the seal wash lines, follow the steps in Replacing the Seal
Wash Lines (} page 129).
• To replace the seal wash detector (drop detector), follow the steps
in Replacing the Seal Wash Detector (} page 132).
TIP To replace the seal wash seals, which are part of the pump head, see
Replacing the Seal Wash Seals (} page 151).
• At regular intervals
Items required
• Syringe (12 mL)
No. Description
3. To allow the liquid to travel easily through the seal wash system,
press down the lever of the seal wash pump and hold it in that
position with one hand.
4. Pull out the syringe plunger to draw seal wash liquid into the tubing.
5. When about 10 mL of liquid are present in the syringe, release the
lever of the seal wash pump.
6. Press the lever firmly onto the tubing and press the liquid from the
syringe into the seal wash system.
7. Check whether liquid escapes from the connections of the seal wash
system or under the pump heads.
Tools required
Tubing cutter (optional)
Preparations
1. Unscrew the cap of the seal wash reservoir and remove the seal
wash line together with the cap from the reservoir.
2. To empty the seal wash line, press down the lever of the seal wash
pump. Release the lever when the line is empty.
NOTICE
When you must remove a seal wash line from the tubing guides, be
careful not to pull on other tubing in the guides.
Follow the sequence of steps for the seal wash line that you want to
replace:
• For the seal wash lines that connect the seal wash reservoir to the
peristaltic tubing in the pump, follow the steps in the Operating
Manual for the Vanquish autosampler.
• For the seal wash lines from peristaltic tubing to the pump heads
and on to the seal wash detector, see Replacing the Seal Wash Lines
between the Peristaltic Tubing and the Seal Wash Detector
(} page 131).
No. Description
1 Pump lever
2 Pump rotor
3 Tubing holder (peristaltic tubing)
4 Tubing connector
5 Seal wash line from autosampler (silicone tubing)
6 Peristaltic tubing
7 Seal wash tubing to pump head (silicone tubing)
2. Press down the lever on the seal wash pump and remove the
peristaltic tubing from the pump and from the tubing holders.
3. Connect the new peristaltic tubing to the silicone tubing using the
tubing connectors. Cut tubing to required length, if applicable.
4. Insert the peristaltic tubing into the right tubing holder. Make sure
that the tubing connector rests on the tubing holder. Push the
tubing connector vertically down onto the holder if required.
5. Press down the lever on the seal wash pump and hold it in that
position with one hand.
6. With your other hand, insert the tubing between the lever and the
rotor and wind it around the rotor.
3. Flush the seal wash system, using fresh seal wash liquid (see
Flushing the Seal Wash System (} page 76)).
Replacing the Seal Wash Lines between the Peristaltic Tubing and the Seal Wash Detector
1. Disconnect the seal wash tubing from the tubing connectors. Do not
remove the tubing connectors.
5. Flush the seal wash system, using fresh seal wash liquid (see
Flushing the Seal Wash System (} page 76)).
Parts required
Seal wash detector assembly, consisting of the detector and funnel
Preparations
To prepare the new seal wash detector assembly push the seal wash
detector into the funnel. Mind the orientation of the funnel.
1b
1a
No. Description
No. Description
1. Disconnect the seal wash line from the seal wash detector (port
labeled in).
3. Grasp and pull the seal wash detector assembly upward by the
funnel.
4. Push the new seal wash detector assembly into the pump.
5. Reconnect the seal wash line to the seal wash detector (port labeled
in).
7. Flush the seal wash system, using fresh seal wash liquid (see
Flushing the Seal Wash System (} page 76)).
NOTICE
All pumps except VC-P21 and VC-P33: Whenever you remove the pump
head, you must replace the seal wash seals.
• To replace the piston seals or support ring, see Replacing the Piston
Seals or Support Ring (} page 145).
• To replace the seal wash seals, see Replacing the Seal Wash Seals
(} page 151).
• To replace the pump head seals, see Replacing the Pump Head Seals
(} page 154).
• To test the pump for piston seal leakage, see Testing the Piston Seals
for Leakage (} page 156).
For an overview of the pump head parts, see Overview of Pump Head
Parts (} page 135).
No. Description
Parts required
• Pump head
Preparations
2. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Install the pump head. See Installing the Pump Head (} page 138).
No. Description
Binary pump Select the pump head for which you want to
perform the procedure.
Dual pump Select the pump for which you want to perform
the procedure.
Isocratic pump, quaternary pump Continue with the next step.
3. Undock the pistons either from the keypad or from the Chromeleon
software. Wait until undocking is complete before proceeding with
the next step.
4. Remove the capillary that connects the pump head to the purge/
sensor unit. Disconnect the capillary on both the pump head outlet
and the purge/sensor unit.
6. Binary pump: Unplug the cable of the pressure sensor from the P-
WORK connector.
8. Hold the pump head with one hand, remove the pump head screw,
and then remove the pump head by pulling it toward you.
9. It may happen that the seal wash plate remains in the pump when
you remove the pump head. Follow these steps:
a) Insert the spacing tool between the pump head and pump block.
b) Carefully pull the seal wash plate away from the pump block.
2. Before you begin, note the information for reinstalling pump heads
and installing new pump heads (replacement pump heads).
Reinstallation of pump heads (binary pump or dual pump only): Be sure
not to interchange the pump heads.
• With binary pump and dual pump: New pump heads can be installed
in either position when installed for the first time. They have no
label on the pressure sensor cable or U-tube. For easier
identification during future maintenance, consider labeling them
according to the installation position (A for the left pump head or B
for the right pump head).
No. Description
1. Insert the pump head into the pump and tighten the pump head
screw (recommended torque: 10 Nm).
3. Reconnect the capillary that connects the pump head to the purge/
sensor unit. Observe the following:
Install the capillary in the direction of solvent flow (indicated by the
arrow on the capillary).
5. Dock the pistons either from the keypad or from the Chromeleon
software. Wait until docking is complete before proceeding with the
next step.
6. Flush the seal wash system with seal wash liquid (see Flushing the
Seal Wash System (} page 141)).
10. Test the seal wash system and pump for leakage (see Testing the
Seal Wash System for Leakage (} page 128) and Testing the Pump
for Leakage (General Test) (} page 177)).
To flush the seal wash system for any other reason, see Flushing the Seal
Wash System (} page 76).
No. Description
3. To allow the liquid to travel easily through the seal wash system,
press down the lever of the seal wash pump and hold it in that
position with one hand.
4. Pull out the syringe plunger to draw seal wash liquid into the tubing.
5. When about 10 mL of liquid are present in the syringe, release the
lever of the seal wash pump.
• Pistons
NOTICE
All pumps except VC-P21 and VC-P33: Whenever you remove the pump
head, you must replace the seal wash seals.
Preparations
Remove the pump head (see Replacing the Pump Head (} page 136)).
3. Replace the seal wash seals (see Replacing the Seal Wash Seals
(} page 151)).
No. Description
6. Grasping the pistons by the piston base, push in the pistons until
they contact the tool.
7. Remove the spacing tool. Be careful not to pull out the pistons any
further.
8. Install the pump head (see Installing the Pump Head (} page 138)).
9. Flush the seal wash system with seal wash liquid (see Flushing the
Seal Wash System (} page 141)).
11. Inspect all flow connections on the pump head for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.
13. Test the seal wash system and pump for leakage (see Testing the
Seal Wash System for Leakage (} page 128) and Testing the Pump
for Leakage (General Test) (} page 177)).
Items required
• Light source
• Isopropanol
TIP Hold the rear side of the piston into the light. The refraction of the
light makes dirt particles appear larger.
2. Even if the piston seems to be clean: Hold the piston and carefully
move your fingernail over the piston surface. You should not feel
inconsistent or rough areas.
Situation Action
Parts required
• Piston seals
• Support ring
Note that the support ring is not a wear part. You need not replace
the support ring when you replace a piston seal. Replace the support
ring only when applicable.
NOTICE
All pumps except VC-P21 and VC-P33: Whenever you remove the pump
head, you must replace the seal wash seals.
• Cleaning swab
• Lint-free tissue
• Seal-handling tool
No. Description
Preparations
1. Remove the seal wash plate from the pump head if applicable.
No. Description
1 Pistons
2 Seal wash plate
3 Seal wash bodies
4 Support ring
NOTICE
Do not use the seal removal pin to remove the support ring. The support
ring can be removed from the tool only with difficulties.
Figure 45: Pump head (left: no piston seal installed, right: piston seal
installed)
No. Description
7. Slide the support ring onto the insertion pin of the seal-handling
tool.
8. Slide the piston seal onto the insertion pin, with the open side of the
seal facing away from the tool.
No. Description
1 Support ring
2 Piston seal (open side facing away from the tool)
9. With the insertion pin, push the seal-handling tool into the pump
head as far as it goes in.
10. Remove the tool from the pump head. The piston seal and support
ring remain in the pump head.
11. Insert the seal wash bodies into the pump head. Mind the correct
orientation. The even side faces the pump head.
No. Description
12. Replace the seal wash seals. Follow the appropriate steps in
Replacing the Seal Wash Seals (} page 151).
13. Install the pistons. Follow the related steps in Replacing the Pistons
(} page 142).
14. Install the pump head. See Installing the Pump Head (} page 138).
16. Inspect all flow connections on the pump head for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.
18. Test the seal wash system and pump for leakage (see Testing the
Seal Wash System for Leakage (} page 128) and Testing the Pump
for Leakage (General Test) (} page 177)).
Items required
Recommendations
2. Route the free end of the tubing to the drain port outlet (see
Guiding Liquids to Waste (} page 62)).
9. Turn off the pump flow and wait until the system pressure is down
to zero.
10. Remove the capillary from the pump outlet and reconnect the pump
to the system.
11. Purge the pump using the solvent for your application.
12. When purging is complete, empty the silicon tubing.
13. If an increased leakage rate is observed with new piston seals after
several hours of operation, operate the pump for at least 2 hours at
35 MPa.
TIP If the piston seals are not sufficiently tight yet, have the pump build
up high pressure and hold the pressure for some minutes.
NOTICE
Always empty the silicone tubing when purging is complete. When the
purge valve is closed, purge liquid remains in the tubing connected to
the purge outlet. Leaving the liquid in the tubing for a longer period
limits the tubing lifetime.
1. To absorb any liquid that might be leaving the purge outlet when the
tubing is removed, place the tissue under the outlet.
2. On the purge/sensor unit, remove the silicone tubing from the purge
outlet. The other end of the tubing remains in the purge drain port
outlet.
3. Hold the tubing so that any liquid that is present in the tubing flows
off through the drain port outlet.
4. On the purge/sensor unit, consider reconnecting the silicone tubing.
• Once a year
Parts required
Seal wash seals
• Seal-handling tool
No. Description
Preparations
1. Remove the seal wash plate from the pump head if applicable.
No. Description
1 Pump head
2 Seal wash plate, channel side
3 Seal wash seals
4 Pistons
3. Slide the new seal onto the insertion pin of the seal-handling tool.
5. On the seal wash plate (no-channel side), set the insertion pin of the
seal-handling tool at a 90° angle to the plate and push the insertion
pin vertically into the plate as far as it goes in.
6. Remove the tool from the seal wash plate. The seal remains in the
plate.
No. Description
7. Place the seal wash plate onto the pump head, with the seals facing
the pump head.
8. Install the pistons. Follow the related steps in Replacing the Pistons
(} page 142).
9. Install the pump head. See Installing the Pump Head (} page 138).
10. Inspect all flow connections on the pump head for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.
11. Test the seal wash system and pump for leakage (see Testing the
Seal Wash System for Leakage (} page 128) and Testing the Pump
for Leakage (General Test) (} page 177)).
12. If you also replaced the piston seals, proceed with the specific steps
further down.
NOTICE
All pumps except VC-P21 and VC-P33: Whenever you remove the pump
head, you must replace the seal wash seals.
Preparations
1. Remove the seal wash plate from the pump head if applicable.
2. Remove the seal wash bodies from the pump head if applicable.
No. Description
5. Insert the seal wash bodies into the pump head. Mind the correct
orientation. The even side faces the pump head.
No. Description
6. Replace the seal wash seals (see Replacing the Seal Wash Seals
(} page 151)).
7. Install the pistons. Follow the related steps in Replacing the Pistons
(} page 142).
8. Install the pump head. See Installing the Pump Head (} page 138).
10. Test the seal wash system and pump for leakage (see Testing the
Seal Wash System for Leakage (} page 128) and Testing the Pump
for Leakage (General Test) (} page 177)).
• Isopropanol
2. When the seal wash cycle is complete, disconnect the tubing from
the seal wash detector (port labeled in).
No. Description
3. Remove some liquid from the tubing, for example, by shaking the
tubing.
Liquid level in the tubing remains The piston seals seal tightly. Reconnect the
unchanged seal wash tubing to the seal wash detector
(port labeled in). This completes the test.
Liquid level in the tubing is The seal wash seals are leaky:
decreasing 1. Replace the seal wash seals.
2. Reconnect the seal wash tubing to the seal
wash detector (port labeled in).
3. Repeat the test starting with step 1.
Liquid level in the tubing is One or more piston seals are leaky.
increasing Isocratic pump and quaternary pump:
1. Replace the piston seals in the pump head.
2. Reconnect the seal wash tubing to the seal
wash detector (port labeled in).
3. Repeat the test starting with step 1.
Binary pump and dual pump:
Continue with the next steps.
8. On the right pump head, remove the seal wash tubing that connects
the left pump head to the right pump head.
10. For about 5 minutes, observe the liquid level in the tubing. Finish the
observation before a new seal wash cycle starts.
Liquid level in the tubing is 1. Replace the piston seals in the left pump
increasing again head.
2. Reconnect the seal wash tubing to the right
pump head.
3. Reconnect the seal wash tubing to the seal
wash detector (port labeled in).
4. Repeat the test starting with step 1.
Liquid level in the tubing does not 1. Replace the piston seals in the right pump
increase again head.
2. Reconnect the seal wash tubing to the right
pump head.
3. Reconnect the seal wash tubing to the seal
wash detector (port labeled in).
4. Repeat the test starting with step 1.
TIP If you observe leakage from a pump head, verify that the seal wash
tubing is properly connected to the pump head. If the tubing is not
properly connected or if the tubing is crimped or kinked, seal wash liquid
may leak into the pump.
• To replace the outlet check valves, see Replacing the Outlet Check
Valves (} page 161).
No. Description
When
• Issue with the check valve that could not be solved by cleaning (for
example, persistent leakage)
Parts required
• Valve cartridge
Preparations
2. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Use the open-end wrench to loosen and remove the valve nut from
the pump head.
Valve cartridge removed together with Hold the nut over your hand and turn it
valve nut upside down to allow the cartridge to drop
into your hand.
Valve cartridge remained in pump Pull out the cartridge using the tweezers.
4. Insert the new cartridge into the valve nut. Insert the cartridge in
the direction of flow (indicated by the arrow on the cartridge).
5. Screw the valve nut into the pump head and tighten with the open-
end wrench.
NOTICE
Overtightening may damage the valve cartridge. The torque should not
exceed 10 Nm.
8. Inspect the inlet check valve and fitting connection for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.
9. Optional: Test the pump for leakage (see Testing the Pump for
Leakage (General Test) (} page 177)).
No. Description
1 U-tube
2 Outlet check valve, including
2a Valve nut
2b Valve cartridge
3 Pump head
When
• Issue with the check valve that could not be solved by cleaning (for
example, persistent leakage)
Parts required
• Valve cartridge
Preparations
2. Turn off the pump flow and wait until the system pressure is down
to zero.
1. Remove the capillary that connects the pump head to the purge/
sensor unit. Disconnect the capillary on both the pump head outlet
and purge/sensor unit.
3. Use the open-end wrench to loosen and remove the valve nut from
the pump head.
Valve cartridge removed together with Hold the nut over your hand and turn it
valve nut upside down to allow the cartridge to drop
into your hand.
Valve cartridge remained in pump Pull out the cartridge using the tweezers.
5. Insert the new cartridge into the valve nut. Insert the cartridge in
the direction of flow (indicated by the arrow on the cartridge).
6. Screw the valve nut into the pump head and tighten with the open-
end wrench.
NOTICE
Overtightening may damage the valve cartridge. The torque should not
exceed 10 Nm.
8. Reconnect the capillary that connects the pump head to the purge/
sensor unit. Observe the following:
Install the capillary in the direction of solvent flow. Note that one
end of the capillary is longer than the other end. Connect the longer
end to the pump head.
10. Inspect all flow connections on the pump head for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.
11. Optional: Test the pump for leakage (see Testing the Pump for
Leakage (General Test) (} page 177)).
When
• Once a year
• Ultrasonic bath
• Backpressure capillary
You can use any Viper capillary that can generate the backpressure
indicated in the table with a flow rate between 1 mL/min and
4 mL/min. For example, you can connect the two Viper capillaries
(length: 15 m each) from the Diagnostics tool kit with the connection
union to obtain a 30 m capillary with 180 µm inner diameter.
Preparations
Remove the check valves that you want to clean.
Recommended: Clean the inlet check valves and outlet check valves of all
pump heads at the same time. To remove the check valves, follow the
steps in Replacing the Inlet Check Valves (} page 159) and Replacing the
Outlet Check Valves (} page 161).
1. In the direction of flow, fill the check valves with isopropanol, using
the syringe.
2. Place the check valves in the beaker. Clean them in the ultrasonic
bath for 10 minutes.
4. Reinstall the check valves. Follow the steps in Replacing the Inlet
Check Valves (} page 159) and Replacing the Outlet Check Valves
(} page 161).
5. Install the backpressure capillary on the pump outlet. Guide the free
end of the capillary to the waste container.
6. Have the pump deliver the solvents suitable for the run-in
procedure: Start with 50% B and a flow rate of 1 mL/min. Increase
the flow rate until the system pressure is well above 60 MPa
(VC-pump) or 90 MPa (VF-pump) and run the pump for 20 minutes.
7. Inspect the pump heads and check valves for indications for leakage.
Tighten leaky connections only when the system pressure is down to
zero.
10. Test the pump for leakage (see Testing the Pump for Leakage
(General Test) (} page 177)).
Follow the sequence of steps for the maintenance procedure that you
want to perform:
• To test the static mixer for permeability, see Testing the Static Mixer
for Permeability (} page 166).
TIP With the dual pump, test each static mixer separately.
When
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
• Water as solvent
• Tissue
NOTICE
Turning the static mixer in the mounting bracket when loosening or
tightening capillaries can damage the mounting bracket.
To prevent the mixer from turning in the bracket, hold the mixer with an
open-end wrench when you loosen or tighten capillaries on the mixer.
2. To absorb liquid leaving the static mixer, place the tissue on the
mixer outlet.
7. Test the pump for leakage. See Testing the Pump for Leakage
(General Test) (} page 177).
• As applicable
¨ Static mixer
¨ Capillary mixer
Preparations
3. Turn off the pump flow and wait until the system pressure is down
to zero.
NOTICE
Turning the static mixer in the mounting bracket when loosening or
tightening capillaries can damage the mounting bracket.
To prevent the mixer from turning in the bracket, hold the mixer with an
open-end wrench when you loosen or tighten capillaries on the mixer.
1. On the static mixer outlet, disconnect the capillary that connects the
static mixer to the autosampler injection valve.
4. Remove the assembly of static mixer and capillary mixer from the
pump.
7. Insert the assembly of static mixer and capillary mixer into the
pump.
10. On the static mixer outlet, connect the capillary to the autosampler
injection valve.
11. Apply the flow rate of your application and have the pump deliver
for a short time.
12. Inspect the flow connections on the mixer for indications of leakage
before resuming operation.
13. When the static mixer has been replaced: In the Chromeleon
software, consider updating the Predictive Performance information
for the static mixer (MixerChanged command).
• To replace the solvent line filter, see Replacing the Solvent Line Filter
(} page 173).
Parts required
• Silicone tubing
• Syringe
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Prepare the pump for purging (see Purging the Pump (} page 100)).
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
The flow path determines the pump inlet (component at which the
solvent enters the pump) and the solvent flow through the pump.
• Solvent line from the solvent reservoir to the pump inlet: Follow the
steps in Solvent Lines From Reservoir To Pump Inlet (} page 171).
Preparations
3. Remove the solvent line from the pump and all modules above the
pump in the system stack. When removing the solvent line from the
tubing guides, do not pull on other tubing in the guides.
4. To install the new solvent line, follow the related steps in Connecting
the Solvent Lines (} page 63).
5. To remove any air from the solvent lines, purge the pump using the
solvent of your application.
When
Damage or blockage of the solvent line
Preparations
2. Remove the solvent line on the pump inlet and on the degasser.
4. To remove any air from the solvent lines, purge the pump using the
solvent of your application.
When
Damage or blockage of the solvent line
Preparations
3. To remove any air from the solvent lines, purge the pump using the
solvent of your application.
No. Description
• As applicable:
¨ Filter holder
¨ Filter frit
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
3. As applicable: Open the filter holder and remove the filter frit.
7. To remove any air from the solvent lines, purge the pump using the
solvent of your application.
• Leakage is observed around the valve knob when the valve is open.
Parts required
Purge valve knob (with integrated seals)
No. Description
NOTICE
Avoid holding the new valve knob by the seal end. Scratches on the seals
will prevent a proper seal and cause leakage.
• Cleaning swab
Preparations
2. Turn off the pump flow and wait until the system pressure is down
to zero.
1. To remove the purge valve knob from the purge/sensor unit, turn
the valve knob counterclockwise all the way and pull it straight off
the purge/sensor unit.
2. Before installing the new purge valve knob, clean the opening in the
purge/sensor unit with a cleaning swab.
3. Insert the new purge valve knob into the purge/sensor unit.
4. To tighten, turn the valve knob clockwise with your fingers until it is
finger-tight.
NOTICE
Turn the valve knob only with your fingers. Overtightening may damage
the purge valve seals.
5. Resume operation.
6. Inspect the purge outlet. If liquid leaves the outlet, the valve knob
may not be tightened enough. Tighten a little more.
Test Description
Prerequisite
A compatible chromatography data system that supports basic tightness
testing for the configuration of your Vanquish system
2. If the test fails, check the Instrument Audit Trail for a message.
VC‑pump 60 MPa
VF‑pump 80 MPa
4. Apply a flow of 50 µL/min and use the solvent for the application.
Pump Pressure
6. Have the pump deliver until the following pressure has built up:
Pump Pressure
VC‑pump 55 MPa
VF‑pump 75 MPa
7. When the pressure has built up, apply the following flow:
Pump Flow
Standard pump
VC‑P10, VC‑P20, VC‑P32, VC‑P40 1 µL/min
VC‑P21, VC‑P33 4 µL/min
VF‑pump 1 µL/min
Pump modified for NP applications
VC-P20, VC-P21, 4 µL/min
VC-P32, VC-P32, VC‑P40
VC-P10 6 µL/min
4. Tighten the connection that you want to test a little more. The
pressure will suddenly increase a little.
6. When testing the pump for leakage is complete, reset the upper
pressure limit to the value used before the leakage test.
Parts required
Fuses (2 fuses, 3.15 AT, 250 V AC, slow-blow, 5 x 20 mm) from Fuses Kit
Tools required
Slotted screwdriver, any size between 3.3 mm and 5.5 mm is
appropriate
Preparations
WARNING—Electric Shock
High voltages are present inside the device that could cause an electric
shock or damage to the device.
• Turn off the device with its main power switch. Disconnect the
power cord from both the power source and the device.
• Use only the fuses of the type and current rating specified for the
device by Thermo Fisher Scientific.
• Do not use repaired fuses and do not short-circuit the fuse holders.
No. Description
2. Replace the two fuses with new fuses of the specified type and
current rating. Always replace both fuses.
4. Reconnect the power cord to the power source and to the device.
TIP When a new firmware version is released, the new version will be
included in the next available Chromeleon version. The new firmware
will not be transferred automatically to the device when you install the
Chromeleon version.
Preparations
1. Read the release notes provided with the firmware and/or
Chromeleon version.
2. Connect the device in the Chromeleon software.
3. Stop all operations on the Instrument that includes the device.
4. Wait until the Instrument is idle.
5. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
NOTICE
A firmware downgrade or incomplete firmware update may result in loss
of functionality or malfunctioning of the device.
• Do not interrupt communication between the Chromeleon software
and the device at any time during the procedure.
• At the beginning of the update process, a message appears showing
the firmware version currently installed in the device and the
version that will be transferred from the Chromeleon software. If
the firmware installed in the device is a later version than the
version in the Chromeleon software, cancel the download.
Situation Action
TIP The maintenance procedures do not require that you remove the
doors. If this should ever be required for a specific reason or procedure,
follow the steps in this section.
Parts required
Replacement door
Preparations
If the door to be replaced is located directly below the solvent rack,
remove all reservoirs from the solvent rack.
If the display is installed next to the door to be replaced, consider
loosening the fastening block of the display to push the display toward
the system rear.
Follow these steps
NOTICE
To avoid damage to the door hinges, be careful when performing the
following sequence of steps and do not apply force.
1. If the door is located directly below the solvent rack, lift the solvent
rack slightly on the front edge.
2. To remove a door, push the door upward while opening.
Open the door to a position in which the two hinges on the housing
are aligned in the grooves on the door. You can remove the door
only when the hinges are in the grooves.
No. Description
3. Slightly tilt the door to the outside, away from the housing, and
remove the door.
4. To install the door, align the door with the hinges on the housing. Do
not clamp tubing or capillaries between the door and the enclosure.
1. Prepare the device for transport. See Preparing the Device for
Transport (} page 187).
2. Turn off the device with its main power switch and disconnect the
power cord.
5. Check the leak tray and drain port. If liquid is present, absorb the
liquid with a tissue.
1. Observe the notes for handling and lifting the device safely.
3. Install and set up the device in the system stack. Follow the
instructions on mounting the system stack in the Vanquish System
Operating Manual.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
• Decontaminate all parts of the device that you want to return for
repair.
• Fill in and sign the Health and Safety Form. Thermo Fisher Scientific
refuses to accept devices for repair if the Health and Safety Form is
missing, incompletely filled in, or unsigned.
2. If you need to return the device to Thermo Fisher Scientific for depot
repair, contact your local Thermo Fisher Scientific support
organization for the appropriate procedure.
Tools required
Screwdriver, Torx T20
Preparations
1. Loosen the four captive screws on the front left and front right of
the device.
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the leak tray and pull the module out
of the enclosure by approximately 10 cm.
NOTICE
The slide-in module can fall down when pulling it out of the enclosure
too far.
Pull out the slide-in module just far enough so that you can grasp it on
both sides from below.
4. Remove the slide-in module from the enclosure. The following steps
require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
• Decontaminate all parts of the device that you want to return for
repair.
• Fill in and sign the Health and Safety Form. Thermo Fisher Scientific
refuses to accept devices for repair if the Health and Safety Form is
missing, incompletely filled in, or unsigned.
Preparations
Remove the slide-in module from the enclosure. See Removing the Slide-
In Module (} page 190).
NOTICE
Shipping the slide-in module improperly leads to damage to the device.
Always ship the slide-in module as described in this operating manual.
Tools required
Screwdriver, Torx T20
Preparations
2. Verify that the device enclosure into which the slide-in module shall
be installed is clean. If required, clean the inner and outer surfaces
of the enclosure. See Cleaning or Decontaminating the Device
(} page 124).
1. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to insert the slide-in module properly into the
enclosure in the next step.
2. On the slide-in module, check that you can move the captive screws
back and forth with your fingers. If you cannot, screw them in until
you can.
5. With the screwdriver, move each captive screw back and forth while
pressing inward until the screw slips into the thread.
6. Push the slide-in module back into the enclosure as far as it goes in.
7. Gradually and evenly, tighten the four captive screws on the slide-in
module hand-tight.
NOTICE
• Verify that the screws are tightened. Pull the slide-in module by the
leak tray towards the front and check whether the screws move. If
they do not move, the slide-in module is installed properly.
• If the screws move, tighten the screws further. With a torque
wrench, the recommended torque is 1.2 Nm.
2. Prepare all other modules of the Vanquish system for operation and
restart them. Refer to the Operating Manuals for the modules.
8 Troubleshooting
This chapter is a guide to troubleshooting issues that may arise during
operation of the device.
NOTICE
When the Vanquish system includes a Vanquish charged aerosol
detector, the detector may affect the operation and/or performance of
the pump. For example, errors in the detector may affect the pump by
automatically stopping the pump flow. For more information, refer to
the Operating Manual for the detector.
TIP For information about operating issues that might occur during the
operation of a Vanquish system, refer to the Vanquish System Operating
Manual.
Status Indicators
The status indicator LED bar on the front side of the device and the
STATUS LED on the keypad inside provide quick visual feedback on the
operational status of the device. If the device firmware detects a
problem, the status indicators are red.
Alarms
Leaks are a potential safety issue. Therefore, if a leak sensor detects
leakage, beeping starts to alert you in addition to the message in the
Instrument Audit Trail and the status indicators changing to red. Follow
the instructions in this manual to find and eliminate the source for the
leakage.
The data system logs information about all events related to instrument
operation for the current day in an Instrument Audit Trail. The
Instrument Audit Trail is named with the current date, using the format
yyyymmdd. For example, the Instrument Audit Trail for May 15, 2019, is
named 20190515.
The Instrument Audit Trails can be found on the ePanel Set (Audit
ePanel). In addition, Audit Trails for each instrument are available in the
Chromeleon 7 Console Data view, in the folder of the Instrument.
Messages in the Instrument Audit Trail are preceded by an icon. The icon
identifies the seriousness of the problem. For possible causes and
remedial actions, see Messages (} page 200).
Firmware Failure
If a firmware failure occurred during operation of the module, an
exception log has been created about the processes during the firmware
failure. The firmware sends the exception log to the Instrument Audit
Trail when the module is connected in the chromatography data system.
8.2 Messages
This section provides information about the messages that may appear
in the Instrument Audit Trail during operation of the pump. For easier
identification of the source for a problem, note the Device information
that may appear in front of the message:
[Main Device Name] With messages relating to the entire pump, for example, to
the leak detection, main board or with a communication
error
[Pump Device Name] With messages relating to the entire flow unit, for example,
to the system pressure
[Pump Device With messages relating to the right pump block or
Name]_ components in the right pump head, for example, the
Wellness_RightBlock pistons or check valves.
[Pump Device With messages relating to the left pump block or
Name]_ components in the left pump head, for example, the pistons
Wellness_LeftBlock or check valves.
The Device Names are specified in the dialog box for the pump in the
Instrument Configuration Manager. The default names are:
The table lists the most frequently observed messages for the device
and provides troubleshooting assistance.
Each message consists of a code number and a text. The code number is
the unique identifier for the problem while the wording may change.
Note the following:
• If you cannot find the code you are looking for, check the message
text. The two messages "Unexpected module behavior" and
"Module malfunction detected" can be assigned to different codes.
See the beginning of the table for more information.
TIP If you are unable to resolve the problem following the instructions in
this manual, or if you encounter a message not listed in the table, write
down the code and wording of the message and contact us. For details,
see the Contacting Us section at the beginning of this manual.
Code 92 Data verification failed. • This may be a temporary error. Turn off the module. Wait for 5
seconds and turn on the module again.
• The firmware may be defective. Update the firmware.
Code 118 USB Buffer Overflow. This is a software problem. The module produces data faster than
the computer on which the chromatography data system is
running reads the data.
1. In the chromatography data system, disconnect and reconnect
the module.
2. If this does not solve the problem, update the firmware or the
chromatography data system version.
3. If the problem persists: Also, third-party software on the
computer, for example, virus scanners or poor computer
performance can cause the problem. Contact the onsite IT
department.
Code 120 System interlink request timed Communication with the module failed. The module did not
out. respond in time.
For the module for which the message appears:
1. Turn on the module if it is not yet turned on.
2. Check the system interlink connections to the module. Verify
that all system interlink cables are connected at both ends.
3. If the message persists, replace the system interlink cables.
Code 126 Device lost system interlink A system interlink cable may be disconnected from the module or
connection. defective.
1. Check the system interlink connections to the module. Verify
that all system interlink cables are connected at both ends.
2. If the message persists, replace the system interlink cables.
Code 128 System interlink data errors. A system interlink cable may be disconnected from the module or
defective.
1. Check the system interlink connections to the module. Verify
that all system interlink cables are connected at both ends.
2. If the message persists, replace the system interlink cables.
Code 129 Device X not found on system X = serial number
interlink. The module with the given serial number may be disconnected,
turned off or a firmware update is running.
1. Check the system interlink connections to the module.
2. If the message persists, check the module operational status.
Code 131 System interlink bus not A system interlink cable may be disconnected from the module or
operable. defective.
1. Check the system interlink connections to the module. Verify
that all system interlink cables are connected at both ends.
2. If the message persists, replace the system interlink cables.
Code 136 Lock request rejected – already X = lock holder ID, with USB address referring to the
locked by X. chromatography data system or system interlink address referring
to the system controller or a module
The module is already locked by another software (system
controller or chromatography data system).
Wait until the module is released from the locked state.
Code 137 Lock by X expired. X = lock holder ID, with USB address referring to the
chromatography data system or system interlink address referring
to the system controller or a module
Inform Thermo Fisher Scientific about the occurrence. No further
action required.
Code 145 Lock holder X lost. X = lock holder ID, with USB address referring to the
chromatography data system or system interlink address referring
to the system controller or a module
The lock was released automatically since its holder disappeared
from system interlink.
Check the system interlink connections to the module.
Code 152 Assignment or command Module is not ready to accept command or property assignment
rejected – device locked. since locked by another software (system controller or
chromatography data system).
Wait until module is released by the current software (system
controller or chromatography data system).
Code 4098 Upper pressure limit Check the flow path for blockage.
exceeded If the flow path is blocked:
• The column may be contaminated. Rinse or replace the column.
If the problem occurs due to column ageing, it may be sufficient
to increase the setting for the upper pressure limit.
• The autosampler may be blocked. Check the autosampler
components and eliminate the source for the blockage.
If the flow path is not blocked:
Test the static mixer for permeability (see Testing the Static Mixer
for Permeability (} page 166)).
Code 4099 Pressure fallen below lower Check the following causes:
limit. • One or more solvent reservoirs are empty. Fill the reservoirs and
purge the pump. If you are monitoring the solvent consumption
from the Chromeleon software, the message "Code 4158 Out of
eluent X" alerts you when a reservoir is empty.
• Air bubbles may be trapped in the solvent lines. Check the
solvent line filters. Purge the pump.
• Leakage may be present in the system. Resolve the leakage. For
the pump, see Resolving Liquid Leaks (} page 210). For the other
system modules, refer to the Operating Manuals for the
modules.
• A check valve may be contaminated or defective. Clean or
replace the check valves as appropriate (see Check Valves
(} page 159)).
• With the applied flow, the pump cannot build up enough
pressure to reach the lower pressure limit. Adapt the lower
pressure limit or increase the flow.
Code 4100 Purge pressure limit exceeded A pressure of more than 5 MPa built up after a purge cycle has
started.
Perform the following sequence of steps:
1. Open the purge valve if it is not yet open.
2. Check the flow path for indications of blockage. See also
Code 4098 Upper pressure limit exceeded.
3. Take appropriate remedial action as necessary.
4. Mute the alarm by pressing the MUTE ALARM button on the
pump keypad.
5. Start a new purge cycle.
Code 4101 Not ready. Please run self- The pump is not ready because the self-test was not successful.
test. Perform the following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4112 Self-test failed because of Changes in pressure have occurred during the self-test. Perform
fluctuating pressure. the following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4113 Self-test failed because The pressure has not been down to zero during the self-test.
pressure is out of range. Verify that the pressure sensor cables are properly connected.
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
If the message appears again, the system pressure sensor may be
defective. Contact Technical Support.
Code 4114 The device is busy. Please The pump is not ready because, for example, a self-test is running.
retry after 1 minute. Retry after 1 minute. If the message appears again, perform the
following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4115 Working piston pressure zero The message may appear for the pressure sensor in a pump head
out of range. or for the system pressure sensor:
• If the message appears for only the right pump block or the left
pump block, the cable for the pressure sensor in the pump head
may not seat properly in the P-WORK connector.
• If the message appears for both pump blocks at the system time,
the cable for the system pressure sensor may not seat properly
in the P-SYS connector.
Verify that the pressure sensor cable is properly plugged into the
connection port and retry. If the problem persists, contact
Technical Support.
Code 4116 The pressure signals are not The pressure calibration factor for the working piston is out of
consistent. There may be air in the range. Perform these steps:
working cylinder. 1. Verify that the degasser is working.
2. Purge the pump.
Code 4119 Can’t start pump while alarm An alarm is present, for example, because leakage has been
is on. detected. You can restart the pump flow only when the alarm is
no longer present. Mute the alarm by pressing the MUTE ALARM
button on the pump keypad. Find and eliminate the source for the
alarm as suggested for the accompanying message.
Code 4120 The rear seal wash system has This message alerts you only when the drop counter functionality
run out of wash solution. of the seal wash detector is enabled (standard setting).
• Verify that wash liquid is present in the seal wash reservoir.
• Inspect the seal wash lines across the entire flow path for
indications of blockage or leakage. Verify that all seal wash lines
are properly connected and routed. Replace the seal wash lines
as necessary (see Replacing the Seal Wash Lines (} page 129)).
• Inspect the peristaltic tubing for indications of blockage or
damage. Replace the tubing as necessary (see Replacing the Seal
Wash Lines (} page 129)).
• Verify that the peristaltic tubing is properly inserted in the seal
wash pump and that the lever of the pump is not blocked.
• Inspect the electrodes of the seal wash detector for
contamination or damage. Replace the seal wash detector as
necessary (see Replacing the Seal Wash Detector (} page 132)).
• Make sure that the seal wash liquid is sufficiently conductive
(see Choosing the Seal Wash Liquid (} page 69)).
• Start an additional seal wash cycle from the Chromeleon
software. While the wash cycle is running, inspect the pump
heads. If droplets leave the pump head at the rear, the seal
wash seal is defective. Replace the seal wash seals (see
Replacing the Seal Wash Seals (} page 151)).
Code 4121 Piston seal leakage has This message alerts you only when the drop counter functionality
exceeded the recommended limit. of the seal wash detector is enabled (standard setting).
Test the pump for piston seal leakage and take the recommended
action (see Testing the Piston Seals for Leakage (} page 156)).
Code 4122 The rear seal leak sensor This message alerts you only when the drop counter functionality
detects drops constantly. of the seal wash detector is enabled (standard setting).
Inspect the electrodes of the seal wash detector for
contamination or damage. Replace the seal wash detector as
necessary (see Replacing the Seal Wash Detector (} page 132)).
The message appears also when the percentage of water in the
seal wash liquid is too high. Observe the information about the
composition of the seal wash liquid in Choosing the Seal Wash
Liquid (} page 69).
Code 4125 Degasser malfunction The vacuum level monitoring function detected insufficient
degasser vacuum.
Turn the pump off and on again. In the chromatography data
system, check the degasser vacuum. After about 1 minute, the
setting should change from NotOk to Ok. If the degasser vacuum
is still insufficient, the degasser may be leaky. Locate the
DegasserPressure property in the Command window, write down
the pressure reading, which can provide helpful information to
identify the source for the problem, and contact Technical
Support.
Code 4127 The pump drive is still in The message alerts you during pump maintenance:
undock position. Please execute a dock This message alerts you during pump maintenance (for example,
command. when replacing the piston seals or pistons) when you try to start
the pump while the pistons are not yet in the correct position for
normal operation. To return the pistons into the correct position,
perform the following sequence of steps:
1. Open the purge valve.
2. Dock the pistons.
3. Close the purge valve.
The message alerts you during internal pump maintenance:
If the message appears during the internal maintenance
procedure, perform an additional purge cycle.
Code 4130 Pressure buildup during A pressure of more than 0.5 MPa built up during piston
maintenance operation. maintenance or piston seal maintenance. To reduce the pressure,
perform the following sequence of steps:
1. Open the purge valve.
2. Dock or undock the pistons, depending on when the message
appeared.
3. Close the purge valve.
If the message appears during the internal maintenance
procedure, perform an additional purge cycle.
Code 4148 Can’t perform this command You tried to run a self-test while the flow rate was not yet down
while the flow is on. to zero. Turn off the flow and repeat the command.
Code 4152 Pressure sensor malfunction. Verify that the cable for the pressure sensor is properly plugged
Check the cable connection and retry. into the connection port and retry. If the message appears again,
follow these steps:
1. Open the purge valve.
2. Perform a self-test.
3. Close the purge valve.
4. Restart the flow.
5. If this message appears again or if another message relating to
the pressure or a pressure sensor appears, contact Technical
Support.
Code 4156 Compression limit reached. The compression value was 100% during the last stroke. If the
compression value is lower than 100% for several strokes
afterward, the message "Compression back to normal" (code
4157) appears. If this message does not appear, inspect the flow
connections for indications of air bubbles. Purge the pump if
necessary.
If the message "Compression limit reached" persists, see
Checking the Compression Values (} page 211).
Code 4158 Out of eluent X This message alerts you only if you are monitoring the solvent
consumption from the Chromeleon software. The message
indicates for which solvent the reservoir is empty.
Code 4159 The waste bottle is full. This message alerts you only if you are monitoring the liquid level
in the waste container. Empty the waste container when needed.
Code 4161 Pressure recalibration The pump pressure was not down to zero during the self-test.
deviates by xx bar. Verify that the pressure sensor cable is properly connected to the
P-SYS connector. Wait until the pressure is down and repeat the
test.
To speed up the pressure drop, open the purge valve and then,
start the self-test.
Code 4176 The pump pressure exceeded Follow these steps:
the absolute limit. Check pressure sensor 1. Verify that the pressure sensor cables are properly connected.
cables for proper connection and flow Reconnect the cables if necessary.
path for clogging. Then, perform a self-
2. Check the flow path for clogging and remedy the situation if
test.
necessary.
3. Perform a self-test.
If the self-test is not successful, contact Technical Support.
Code 4234 The flow was stopped due to The pump flow was stopped because of a problem with another
a request from another module. module. Check the messages from other modules.
Code 4237 Out of solvent. The solvent reservoir is empty.
Refill the solvent reservoir and purge the pump.
Code 4238 Mainboard initialization Main board initialization has not been successful.
failed. 1. This may be a temporary error. Turn off the module. Wait for 5
seconds and turn on the module again.
2. If the message appears again, update the firmware.
3. If the message persists, the main board may be defective.
Contact Technical Support.
Code 4239 Could not build up enough • Air bubbles may be trapped in the system. Purge the pump.
pressure. • Leakage may be present. As leakage usually occurs at a
connection, visually inspect all components and connections in
the flow path. For tests that can help you to identify the source
for the leakage, see Testing the Pump for Leakage (} page 209).
Code 4240 Column temperature is too The procedure that you want to perform will stop the pump flow.
high. This may damage the column if the column temperature is too
high. Reduce the column oven temperature. Repeat the
procedure.
When the leak sensor reports leakage, see Resolving Liquid Leaks
(} page 210).
Tests
The following tests assist you in finding out whether leakage is present
in the pump and they help you to locate the source of a leak:
When
The leak sensor is wet. The leak sensor reports leakage.
If leakage is not resolved after 3 minutes, the pump stops the flow.
• Cloth or tissue
Preparations
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in Maintenance and Service
(} page 115).
2. With a cloth or tissue, thoroughly absorb all liquid that has collected
in the leak tray and under the leak sensor. Be careful not to bend the
sensor.
• The pressure is not down to zero when the purge valve is open
See Pressure Not Zero with Open Purge Valve (} page 211).
• Pressure pulsation
See Checking the Compression Values (} page 211) and Testing the
Static Mixer for Permeability (} page 166).
1. For the pressure sensor for which the pressure is not down to zero,
check the pressure sensor cable and connector for proper contact.
a) Unplug the pressure sensor cable.
b) Reconnect the cable. To ensure proper contact push the cable
firmly into the connector.
2. Perform a self-test.
When
You observe pressure pulsation or shifts in retention time.
Solvent Compression in %
Water 25 % ± 10 %
Methanol 50 % ± 10 %
Acetonitrile 45 % ± 10 %
Lower than indicated in the table The outlet check valve may be defective.
Replace the valve cartridge. See Replacing the
Outlet Check Valves (} page 161).
Higher than indicated in the table • The piston seal may be defective. Replace
the piston seal. See Replacing the Piston
Seals or Support Ring (} page 145).
• The inlet check valve may be defective.
Replace the valve cartridge. See Replacing
the Inlet Check Valves (} page 159).
Very high (message "Compression • Air bubbles may be present in the pump.
limit reached" and low pressure) Make sure that no air bubbles are present in
the solvent lines and purge the pump.
• If solvent shut-off valves are installed: Verify
that the valves are open.
• Excessive leakage may be present on the
inlet check valve. Remove the solvent line
from the solvent reservoir and the solvent
line filter from the solvent line. While the
pump is running, observe the liquid in the
solvent line. If the liquid moves also against
the direction of flow during the stroke, the
inlet check valve may be defective. Replace
the valve cartridge. See Replacing the Inlet
Check Valves (} page 159).
TIP The information and references to the static mixer in this section
apply equally to the inline filter if installed.
When
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
• Water as solvent
• Tissue
9 Pump-Specific
Information
This chapter provides specific information about the pump, for example,
the interior components and operating principle.
TIP The dual pump combines two ternary pumps in a single enclosure.
The two pumps are referred to as left pump and right pump. If not
otherwise stated, the information in this section applies equally to the
quaternary pump and the two pumps of the dual pump.
No. Description
No. Description
No. Description
No. Description
The pump contributes to the system gradient delay volume by the dwell
volume, which is the calculated volume between the point where the
gradient is formed and the pump outlet.
No. Description
Examples
• To deliver 20% of solvent %A, 40% of solvent %B, 30% of solvent %C,
and 10% of solvent %D:
Set %B to 40%, set %C to 30%, and set %D to 10%. %A is
automatically set to 20%.
• SyncWithPump
To enable synchronization, set the property to On. To disable
synchronization, for example, for a specific application, set
SyncWithPump to Off.
• Pump
Shows the name of the pump to which the autosampler is linked.
• The two pumps are referred to as left pump and right pump.
¨ On the keypad, L refers to the left pump and R refers to the right
pump.
¨ In the Chromeleon user interface, Left refers to the left pump
and Right refers to the right pump, for example, as in
PumpLeft_Pressure and PumpRight_Pressure.
• You can monitor the liquid level in the waste container of the dual
pump as a whole or separately for each pump (see Monitoring the
Liquid Level in the Waste Container (} page 227)).
The default assignment for the two pumps is the instrument in which
the dual pump is installed.
TIP Notice that the pumps now appear in the instruments to which they
have been assigned.
TIP When monitoring the liquid level in the solvent reservoirs (solvent
consumption) as described in this section, the Chromeleon software
calculates the liquid level. To monitor the liquid level in real time, based
on real physical measurement, install the Vanquish Solvent Monitor. The
solvent monitor is available as an option (see Optional Accessories
(} page 267)).
• The two pumps can share the same set of solvent reservoirs or they
can be connected to different sets of reservoirs.
TIP It is possible to share solvent reservoirs even if the two pumps are
assigned to different instruments.
2. On the Bottles tab page, select the Share Eluent Bottles check box.
2. On the Bottles tab page, clear the Share Eluent Bottles check box.
TIP When monitoring the liquid level in the waste container as described
in this section, the Chromeleon software calculates the liquid level. To
monitor the liquid level in real time, based on real physical
measurement, install the Vanquish Solvent Monitor. The solvent
monitor is available as an option (see Optional Accessories (} page 267)).
For monitoring the liquid level in the waste container, note the
following:
• The waste liquids from the two pumps can be guided into the same
waste container or into different waste containers.
• Depending on the setup, you can monitor the liquid level in the
waste container for the dual pump as a whole or separately for each
pump.
TIP It is possible to share the waste container even if the two pumps are
assigned to different instruments.
2. On the Bottles tab page, select the Share Waste Bottles check box.
2. On the Bottles tab page, clear the Share Waste Bottles check box.
No. Description
No. Description
No. Description
The pump contributes to the system gradient delay volume by the dwell
volume, which is the calculated volume between the point where the
gradient is formed and the pump outlet.
No. Description
1 Solvent selectors
2 Solvent degasser
3+4 Pump heads
5 Purge/sensor unit
6 Mixing system with capillary mixer and static mixer
7 Pump outlet
8 Dwell volume of the pump
Examples
No. Description
No. Description
A Solvent
1 Pump head, with
1a Working cylinder
1b Equilibration cylinder
2 Purge/sensor unit, with
2a Purge valve knob
2b Purge outlet
2c System pressure sensor
3 Mixing system, serving as inline filter, with
3a Capillary mixer
3b Static mixer
4 Pump outlet (outlet of the static mixer)
No. Description
1 Pump with
1a Point where the gradient is formed:
Quaternary pump: The gradient is formed after the proportioning valve.
Dual pump: The gradient is formed after the proportioning valve (with each
of the two pumps).
Binary pump: The gradient is formed in the purge/sensor unit.
2 Autosampler
3 Column
4 Detector
5 Gradient delay volume of the system
The pump contributes to the system gradient delay volume by the dwell
volume (see Pump-Specific Information (} page 215)). For the dwell
volume of the pump, see Performance Specifications (} page 252).
In the standard configuration, the pump has been optimized for the
dwell volume and ripple. You can adapt the dwell volume or ripple to
your requirements by replacing the mixing system with which the pump
is shipped with a mixing system with a different volume:
Description Remarks
Description Remarks
1. Remove the mixing system with which the pump is shipped (see the
next section).
2. Install the new mixing system. Follow the sequence of steps for the
mixing system that you want to install. See:
¨ Installing a Mixing System with a Volume higher than 100 µL
(} page 240)
¨ Installing the 100 µL Mixing System (} page 243)
¨ Installing the 35 µL Inline Filter Set (} page 245)
Preparations
3. Turn off the pump flow and wait until the system pressure is down
to zero.
NOTICE
Turning the static mixer in the mounting bracket when loosening or
tightening capillaries can damage the mounting bracket.
To prevent the mixer from turning in the bracket, hold the mixer with an
open-end wrench when you loosen or tighten capillaries on the mixer.
1. On the static mixer outlet, disconnect the capillary that connects the
static mixer to the autosampler injection valve.
2. If you want to install the 100 µL mixing system or 35 µL inline filter
set: Disconnect the other end of the capillary on the autosampler
injection valve and remove the capillary.
5. Remove the assembly of static mixer and capillary mixer from the
pump.
Parts required
Static mixer as required for the mixing system (see Available Mixing
Systems (} page 237)).
• Screwdriver, Torx T10 (required only for static mixers with a volume
of 750 µL or 1500 µL)
Preparations
1. If not yet done: Remove the mixing system with which the pump is
shipped (see Removing the Mixing System (} page 239)).
No. Description
2. Using the screws shipped with the mounting bracket, attach the new
bracket inside the pump (same position and orientation as the one
you removed).
Binary pump: Reinsert the spacer sleeves.
No. Description
NOTICE
Turning the static mixer in the mounting bracket when loosening or
tightening capillaries can damage the mounting bracket.
To prevent the mixer from turning in the bracket, hold the mixer with an
open-end wrench when you loosen or tighten capillaries on the mixer.
2. Insert the assembly of static mixer and capillary mixer into the
pump.
7. Apply the flow rate of your application and have the pump deliver
for a short time.
• Operating the pump with a mixing system with a volume lower than
the volume installed in the pump upon shipment reduces the
gradient delay volume but increases the ripple.
• Operating the pump with a mixing system with a volume higher than
the volume installed in the pump upon shipment reduces the ripple
but increases the gradient delay volume.
• The maintenance procedures are the same as for the mixing system
with which the pump was shipped (see Mixing System (} page 166)).
Parts required
Preparations
1. If not yet done: Remove the mixing system with which the pump is
shipped (see Removing the Mixing System (} page 239)).
No. Description
2. Connect the static mixer to the capillary mixer. Mind the direction of
flow through the static mixer (indicated by the arrow on the mixer).
3. Turn and insert the assembly of static mixer and capillary mixer into
the pump so that the knurl of the capillary mixer rests on the edge
of the purge/sensor unit.
No. Description
1 Static mixer
2 Capillary mixer
3 Knurl resting on the purge/sensor unit
9. Apply the flow rate of your application and have the pump deliver
for a short time.
10. Inspect the flow connections on the mixer for indications of leakage
before resuming operation.
• Operating the pump with a mixing system with a volume lower than
the volume installed in the pump upon shipment reduces the
gradient delay volume but increases the ripple.
• The maintenance procedures are the same as for the mixing system
with which the pump was shipped (see Mixing System (} page 166)).
Parts required
Preparations
1. If not yet done: Remove the mixing system with which the pump is
shipped (see Removing the Mixing System (} page 239)).
2. The inline filer requires a different mounting bracket. Install the new
mounting bracket. The mounting bracket is included in the inline
filter ship kit.
1. Unplug the system pressure sensor cable from the P-SYS connector.
2. Remove the seal wash detector assembly (funnel and detector).
Grasp and pull the assembly upward by the funnel. The detector
including the tubing and cable can remain connected to the funnel.
No. Description
3. To attach the new mounting bracket inside the pump (above the
purge/sensor unit):
a) Remove the push rivets that cover the mounting holes while no
mounting bracket is installed.
b) Using the screws shipped with the mounting bracket, attach the
new bracket inside the pump.
No. Description
2. Connect the inline filter to the capillary mixer. Mind the direction of
flow through the inline filter (indicated by the arrow on the filter).
3. Turn and insert the assembly of inline filter and capillary mixer into
the mounting bracket.
No. Description
1 Inline filter
2 Capillary mixer
10. Apply the flow rate of your application and have the pump deliver
for a short time.
11. Inspect the flow connections on the mixer for indications of leakage
before resuming operation.
• Operating the pump with the inline filter reduces the gradient delay
volume but increases the ripple.
11 Specifications
This chapter provides the physical and performance specifications,
including information about the materials used in the flow path of the
device.
Type Specification*
Type Specification*
Type Specification*
Type Specification*
Communication USB:
1 USB port (USB 2.0, "B" type connector)
1 USB hub with 3 ports (USB 2.0, "A" type connectors)
I/O Interface:
2 Dig I/O ports (mini-DIN), each providing one input, one relay output,
and one bidirectional input/output
System Interlink:
2 system interlink ports (RJ45-8 connectors)
Control Chromeleon 7
The pump can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Keypad with 5 buttons for performing certain functions directly from
the pump
Materials in the analytical flow path VC-pumps: VF-pumps:
Stainless steel, titanium, ceramics, MP35N, titanium, ceramics,
sapphire, PEEK, UHMW PE, sapphire, PEEK, UHMW PE,
fluoropolymers fluoropolymers
For information about the chemical resistance of materials refer to the
technical literature.
Solvent and additive information See Solvent and Additive Information (} page 26).
Safety features Leak detection and safe leak handling, excess pressure monitoring
Good Laboratory Practice (GLP) Predictive Performance functions for scheduling maintenance
features procedures based on the actual operating and usage conditions of the
pump.
All system parameters logged in the Chromeleon Audit Trail.
* Typical operating conditions for measurable specifications:
Flow accuracy, flow precision, pulsation: 1 mL/min at 20 MPa or 2 mL/min at 40 MPa, water
Proportioning accuracy, proportioning precision: 1 mL/min at 20 MPa or 2 mL/min at 40 MPa, 0-100%, water/
spiked water
Type Specification*
Type Specification*
Good Laboratory Practice (GLP) Predictive Performance functions for scheduling maintenance
features procedures based on the actual operating and usage conditions of the
pump.
All system parameters logged in the Chromeleon Audit Trail.
* Typical operating conditions for measurable specifications:
Flow accuracy, flow precision, pulsation: 1 mL/min at 20 MPa or 2 mL/min at 40 MPa, water
Type Specification
12 Accessories,
Consumables and
Replacement Parts
This chapter describes the standard accessories that are shipped with
the device and the accessories that are available as an option. This
chapter also provides information for reordering consumables and
replacement parts.
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Item Quantity in
shipment
Tubing bracket 1
Solvent lines (pack of 4), from solvent reservoirs to pump degasser 1
System interlink cable (RJ45), 0.5 m 1
USB cable, type A to type B, high-speed, USB 2.0, cable length: 5 m 1
Item Quantity in
shipment
Solvent monitor
Static mixers
Capillary mixers
TIP Maintenance kits can include parts that are not required for your
specific pump. These parts are marked accordingly in the contents lists
for the kits.
VC-P10
VC-P20
VC-P32
VC-P40
VF-P10
VF-P20
VF-P32
VC-P10
Pump head seals (PTFE O-rings, sizes: 32x1.5 and 9x1.5; 5 each) 6040.2208
Capillary kit, pump head, including: 6040.3001
• U-tube (pack of 2)
• Capillary from pump head to purge/sensor unit (pack of 2)
Seal-handling tool 6040.7158
* With pump modified for NP applications: Seals (pack of 2) for 6040.0306
use as both piston seals and seal wash seals
VF-P10
Pump head seals (PTFE O-rings, sizes: 32x1.5 and 9x1.5; 5 each) 6040.2208
Capillary kit, pump head, including: 6044.5016
• U-tube (pack of 2)
• Capillary from right pump head to purge/sensor unit
• Capillary from left pump head to purge/sensor unit
Seal-handling tool 6040.7158
Installation tool for Viper capillaries with torque toothing 6040.2314
Solvent line filter, filter holder (pack of 6) (no filter frits included) 6268.0115
Solvent line filter, filter frits, 10 µm (pack of 10) for
VF‑pumps (biocompatible frits) 6268.0111
VC‑pumps (stainless steel frits) 6268.0110
Solvent lines to connect the solvent reservoirs to the pump inlet
Quaternary pump: Solvent lines to connect the solvent reservoirs 6036.1703
to the degasser inlet ports
Dual pump: Solvent lines to connect the solvent reservoirs to the 6036.1704
degasser inlet ports
Binary pump: Solvent lines to connect the solvent reservoirs to the 6036.1701
solvent selectors
Isocratic pump: Solvent line to connect the solvent reservoir to 6036.1705
the pump head
Solvent lines to connect the pump inlet to the pump heads
Quaternary pump: Tubing kit, including: 6044.8017
• Solvent lines from degasser to proportioning valve
• Solvent line from proportioning valve to pump head
Dual pump: Tubing kit, including: 6044.8005
• Solvent lines from degasser to proportioning valve
• Solvent line from proportioning valve to pump head
Binary pump: Tubing kit, including: 6044.8006
• Solvent lines from solvent selectors to degasser
• Solvent lines from degasser to pump heads
The tables list the reordering part numbers for the components with
which the pump is shipped from the factory. Mixing systems with other
volumes are available as an option. For ordering information, see
Optional Accessories (} page 267).
Static mixers
Capillary mixers
Power cords
13 Appendix
This chapter provides additional information about compliance and the
use of the digital I/O ports.
RoHS Compliance
This product complies with the RoHS (Restrictions of Hazardous
Substances) directives:
Logo Description
The green logo marks items that do not contain the hazardous
substances identified by the regulations.
Revision Covering
Pin Assignment
1 Bidirectional input/output
2 Relay output — Relay_NC (Normally Closed contact)
3 Ground — GND
4 Digital input — Input
5 Relay output — Relay_COM
COM is the common contact for NO and NC. If the relay is not activated or if
the device is turned off, the connection is between COM and NC. If the relay
is activated, the connection is between COM and NO.
6 Relay output — Relay_NO (Normally Open contact)
The next table lists the functions assigned to the connector pins and the
color of the cable wire connected to each pin.
Prerequisites
To use the digital I/O functionality, the following must be fulfilled:
• The digital I/O port is connected to the external device with the
digital I/O signaling cable.
• The inputs and outputs that you want to use are selected in the
Instrument Configuration Manager.
1. Plug the 6-pin connector of the cable into the digital I/O port that
you want to use.
1. In the dialog box for the pump, on the Inputs and Outputs pages,
select the inputs (Pump_Input_X) and outputs (Pump_Relay_X) that
you want to use. The numbering in the dialog box corresponds to
the port numbers on the pump.
Index
A cleaning ......................................................... 124
compression .................................................... 98
accessories ..................................................... 259 compression values ....................................... 211
optional ..................................................... 267 condensation ............................................. 50, 54
ship kit....................................................... 261 connection
additives ........................................................... 26 power cord ................................................. 54
information ................................................. 26 connectors ....................................................... 53
use............................................................... 90 consumables.......................................... 259, 270
air bubbles (remove) .............................. 101, 105 cTUVus mark.................................................. 283
algae ................................................................. 90 curve ................................................................ 98
audit trail........................................................ 199
Audit Trail messages ...................................... 200 D
B decontamination ........................................... 124
degasser mode ................................................ 98
Basic Tightness Test ....................................... 177 delivery ............................................................ 44
buffer use ......................................................... 90 device
buffers .............................................................. 26 optimization ............................................. 107
concentration.............................................. 26 preparing for operation.............................. 92
information ................................................. 26 restart after long-term shutdown ............ 113
use............................................................... 90 shutdown.................................................. 110
Dig I/O...................................................... 53, 285
C Digital I/O................................................. 53, 285
capillaries Dock (button)................................................... 85
conventional (non-Viper) ............................ 61 door ................................................................. 51
guide ........................................................... 56 open............................................................ 51
install........................................................... 57 remove ..................................................... 116
Viper............................................................ 58 replace...................................................... 185
capillary mixer .................................................. 37 unhinge............................................. 116, 185
replace....................................................... 167 drain system .................................................... 62
CE mark .......................................................... 282 drainage........................................................... 62
charged aerosol detector ............................... 198 drop detector.......................... 218, 219, 228, 233
check valve drop detector (replace) ................................. 132
clean.......................................................... 164 dual pump...................................................... 217
maintenance ............................................. 159 specifics .................................................... 225
replace....................................................... 159 dwell volume .................................. 222, 231, 236
CheckValvesServiceDone ............................... 126
chloride concentration ..................................... 28 E
Chromeleon...................................................... 39 EAC mark ....................................................... 282
audit trail................................................... 199 equilibration .................................................... 93
device setup ................................................ 80 ExceptionLogClear (Chromeleon) .................. 199
module setup .............................................. 80
Predictive Performance............................. 125 F
Smart Shutdown........................................ 110
Smart Standby........................................... 110 FCC................................................................. 283
Smart Startup .............................................. 93 filter permeability.......................................... 213
firmware failure.............................................. 199 L
firmware update ............................................ 183
flow .................................................................. 94 leak detection ............................................ 38, 98
Flow (button).................................................... 86 leak sensor..................................................... 210
flow acceleration.............................................. 94 leak sensor mode............................................. 98
flow connections .............................................. 55 leak test
flow deceleration ............................................. 94 automatic testing ..................................... 177
Flow LEDs ......................................................... 86 Basic Tightness Test.................................. 177
flow ramp ......................................................... 94 manual testing.......................................... 178
flush piston seals ....................................... 156, 210
pump ................................................. 122, 123 pump ................................................ 177, 210
fuses ............................................................... 181 purge valve knob ...................................... 210
seal wash system .............................. 128, 210
G leakage........................................................... 210
LED bar................................................ 84, 86, 198
gloves ............................................................... 23 liquid leaks..................................................... 210
gradient delay volume.................... 222, 231, 236 liquid level monitoring..................................... 96
guide hole......................................................... 56 long-term shutdown...................................... 111
guidelines
installation................................................... 46 M
maintenance ............................................. 117
operation..................................................... 83 main power fuses .......................................... 181
maintenance.......................................... 115, 120
I cleaning .................................................... 124
decontamination ...................................... 124
inlet check valve doors......................................................... 185
clean.......................................................... 164 firmware update....................................... 183
replace....................................................... 159 fuses ......................................................... 181
inline filter .............................................. 237, 268 general rules............................................. 119
permeability ...................................... 166, 213 internal ..................................................... 120
replace............................................... 166, 213 interval ..................................................... 120
installation........................................................ 45 introduction.............................................. 116
capillaries and tubing .................................. 56 predictive performance............................ 125
safety guidelines ......................................... 46 safety guidelines....................................... 117
site requirements........................................ 49 messages ....................................................... 200
system ......................................................... 47 MixerChanged ............................................... 126
Instrument Audit Trail .................................... 199 mixing system .................................................. 37
intended use..................................................... 20 available volumes ..................................... 237
interior view ..................................................... 34 exchange (change to different volume) ... 239
Interlink ............................................................ 53 permeability ............................................. 166
internal maintenance ..................................... 120 replace...................................................... 166
isocratic pump................................................ 233 move................................................................ 48
Mute Alarm...................................................... 85
K
keypad.............................................................. 84 N
non-Viper capillaries........................................ 61
normal-phase compatible solvents and additives
.................................................................. 249
O pump head
maintenance............................................. 134
operating parameters ...................................... 94 parts overview.......................................... 135
operating principle ........................................... 33 pump head seals....................................... 154
operation.................................................... 39, 81 replace...................................................... 136
control elements ......................................... 84 pump head bushing ....................................... 135
interrupt.................................................... 110 pump head seals
power on/off ............................................... 89 replace...................................................... 154
safety guidelines ......................................... 83 pump inlet ............................................. 221, 230
optimization ................................................... 107 pump outlet.... 218, 220, 221, 229, 230, 233, 234
guidelines .................................................. 107 pump pressure (record)................................... 95
outlet check valve purge ............................................................... 98
clean.......................................................... 164 automatic ................................................. 101
replace....................................................... 161 changing solvents ..................................... 102
overview (functional) ....................................... 31 empty solvent lines .................................. 102
first time operation .................................. 102
P manual...................................................... 102
peristaltic tubing remove air bubbles................................... 102
replace....................................................... 130 Purge (button) ................................................. 85
permeability purge settings ................................................ 100
static mixer................................................ 166 purge valve ............................... 36, 218, 219, 233
permeability test ............................................ 213 purge valve knob (replace) ............................ 175
pH range........................................................... 27 purge/sensor unit ............................................ 36
piston purging........................................................... 100
clean.......................................................... 144
replace....................................................... 142 Q
piston seal QualificationDone.......................................... 126
leak test..................................................... 156 quaternary pump........................................... 217
replace....................................................... 145
run-in......................................................... 149 R
piston seal wash cycle ...................................... 99
piston seal wash system..................... 35, 69, 127 rating plate ...................................................... 19
PistonsChanged.............................................. 126 rear seal wash cycle......................................... 99
power considerations................................. 49, 50 rear seal wash system......................... 35, 69, 127
power cord ................................................. 50, 54 regulatory compliance..................................... 29
power on/off .................................................... 89 remove slide-in module................................. 190
power up .......................................................... 79 reordering...................................................... 259
Predictive Performance.................................. 125 replacement parts ................................. 259, 270
preparation requirements
remove module......................................... 190 power cord ................................................. 50
pressure limits.................................................. 95 restart device after long-term shutdown ...... 113
pressure range ................................................. 95 return
protective clothing ........................................... 22 slide-in module ......................................... 192
Pump .............................................................. 224 RoHS mark ..................................................... 282
pump (flush) ........................................... 122, 123
pump block LED.................................. 84, 87, 198
S shutdown....................................................... 110
long-term.................................................. 111
safety class ....................................................... 21 short-term ................................................ 110
safety glasses.................................................... 23 signal words..................................................... 18
safety guidelines site requirements ............................................ 49
case of emergency ...................................... 25 condensation ........................................ 50, 54
electrical safety ........................................... 23 power considerations ........................... 49, 50
general ........................................................ 21 slide-in module .............................................. 190
general hazards ........................................... 24 install ........................................................ 192
installation................................................... 46 remove ..................................................... 190
maintenance ............................................. 117 return........................................................ 192
operation..................................................... 83 Smart Shutdown ............................................ 110
power cord .................................................. 50 Smart Standby ............................................... 110
protective equipment ................................. 22 Smart Startup .................................................. 93
qualification of personnel ........................... 22 solvent
servicing .................................................... 117 chloride concentration ............................... 28
safety symbols............................................ 18, 19 information................................................. 26
scope of delivery .............................................. 44 pH range ..................................................... 27
seal wash body ............................................... 135 use .............................................................. 90
seal wash cycle ..................................... 79, 89, 99 solvent composition ........................................ 96
seal wash detector (replace) .......................... 132 solvent consumption ............................... 96, 226
seal wash seal (replace).................................. 151 solvent line filter...................................... 63, 169
seal wash system........................................ 35, 69 assemble............................................. 64, 174
flush............................................................. 76 replace...................................................... 173
maintenance ............................................. 127 solvent lines............................................. 63, 169
replace....................................................... 129 connect....................................................... 63
seal wash detector .................................... 132 empty ....................................................... 169
seal wash liquid ........................................... 69 replace...................................................... 170
setup ........................................................... 70 shut-off valves .......................................... 107
test for leakage ......................................... 128 solvent monitoring .......................................... 96
SealsChanged ................................................. 126 solvent name ................................................... 96
Select (button).................................................. 85 solvent reservoir (share)................................ 226
Select LEDs ....................................................... 85 solvent selection valve..................................... 32
self-test ...................................................... 79, 89 solvent selector ....................................... 32, 232
service ............................................................ 115 specifications ................................................. 251
ServiceDone ................................................... 126 performance............................................. 252
setup physical..................................................... 258
hardware..................................................... 52 startup ............................................................. 93
software ...................................................... 80 static filter............................... 166, 213, 237, 268
system arrangement ................................... 52 static mixer ...................................................... 37
ShareEluentBottles......................................... 226 permeability ..................................... 166, 213
ShareWasteBottles......................................... 227 replace...................................................... 167
ship................................................................. 187 status indicator
slide-in module.......................................... 192 LED bar .......................................... 84, 86, 198
ship kit ............................................................ 261 Status LED...................................... 84, 86, 198
short-term shutdown ..................................... 110 Status LED ........................................... 84, 86, 198
support ring.................................................... 145
SyncWithPump ......................................... 99, 224
system arrangement ........................................ 52
system equilibration......................................... 93
System Interlink................................................ 53
system pressure sensor.................................... 36
T
transport ........................................................ 187
troubleshooting.............................................. 197
check compression values......................... 211
general information .................................. 198
leakage ...................................................... 209
messages................................................... 200
pressure issue ........................................... 211
retention time shift ................................... 211
tubing ............................................................... 57
tubing bracket .................................................. 56
tubing chase ..................................................... 56
tubing guides.................................................... 56
type label.......................................................... 19
U
UKCA mark ..................................................... 282
UL/CSA compliance ........................................ 283
Universal Serial Bus .......................................... 53
unpack.............................................................. 42
USB ................................................................... 53
V
Viper
knurl ............................................................ 58
torque toothing........................................... 58
W
waste ................................................................ 62
waste container (share) ................................. 227
waste level ..................................................... 227
waste level monitoring..................................... 97
WEEE .............................................................. 283