SPR2-330 815 01 05 02
SPR2-330 815 01 05 02
SPR2-330 815 01 05 02
ro
A
s
n
m
ie
S
ARCADIS Avantic
SP
Startup
System
Start-up Instructions
© Siemens, 2008
10048590
10047680
© Siemens, 2008
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2008
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2008“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Measuring the local line voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Adapting the programmed system line voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Function check of the EMERGENCY STOP switches . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Functional checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Functional test of C-arm lift movement and parking brakes . . . . . . . . . . . . . . . . . . . . . . 27
System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Note regarding Hibernation Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Customer and country-specific configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Service login: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Customer address configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Configuration of the main system up to SW version VC10A, without service pack SP1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Configuration of the main system with SW version VC10A and service pack SP1
installed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Country-specific configuration for Australia or Switzerland. . . . . . . . . . . . . . . . . . . . 48
Fluoroscopy footswitch configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Country-specific adaptation of the maximum tube current . . . . . . . . . . . . . . . . . . . . 52
Configuring the network (optional) and remote service . . . . . . . . . . . . . . . . . . . . . . 54
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
IQ test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Completing protocols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Leakage current measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
SRS / MNP configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Completing the "ARCADIS Avantic Installation" protocol. . . . . . . . . . . . . . . . . . . . . . 65
5 _______ Protocols______________________________________________________ 66
Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Körperverletzung führen kann.
Please note:
¹ The product-specific safety information in these
instructions,
¹ The general safety information in TD00-000.860.01...
and
¹ The safety information in accordance with ARTD Part 2.
Systems equipped with an I.I. laser light localizer or a single tank laser light local-
izer 0
Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
The protective conductor resistance of 0.2 ohms must not be exceeded.
First measured value
In the case of systems thoroughly tested at the factory prior to delivery, the protective con-
ductor resistance test was already performed and the measured values were recorded in
the system test protocol.
The measurements were performed using the measuring procedures and equipment
recorded in the system test protocol.
The test protocol is included with the accompanying documentation.
If no covers were opened while installing the system and no additional components (e.g.
options) were installed or modified, the values recorded in the test protocol including the
measuring point (e.g. cover of the basic unit or cover of the image intensifier) can be trans-
ferred in the protective conductor resistance protocol as the first measured values.
If the power plug was changed during system installation, covers were removed, or addi-
tional components were installed or modified, the values provided in the test protocol are
invalid.
The values must be labeled invalid. Cross out the values, enter the comment "invalid value"
and confirm this with name, date, and signature.
The protective conductor test must be performed again after all work has been completed.
In the case of systems with no measured values for the protective conductor test recorded
in the system test protocol, the protective conductor test must be performed after all work
has been completed.
Measurement
The measurement must be performed in accordance with DIN VDE 0751, Part 1 (refer to
ARTD Part 2). The protective conductor resistance for all touchable conductive parts must
be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for protective conductor connections.
During the measurement, the power cable and additional connection cable with the inte-
grated protective conductor (e.g., monitor cable between the basic unit and monitor trolley)
must be moved section by section to detect cable breaks.
The protective conductor resistance must not exceed 0.2 ohms.
The values, including the measuring points, must be recorded as first measured values in
the protective conductor resistance protocol.
The measuring procedure and the measuring device used (designation and serial number)
must also be documented.
Separate the page with the protocol from these instructions and file it in the "Protocols" reg-
ister in the system binder or log book.
Fig. 1: Measuring circuit for measuring the protective conductor resistance for units that are
Repeat measurement
In the case of maintenance or repairs, perform the protective conductor resistance mea-
surement again.
Document and evaluate the values determined in the repeat measurement.
The measurement must be performed in accordance with DIN VDE 0751, Part 1 (refer to
ARTD Part 2). The protective conductor resistance for all touchable conductive parts must
be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for protective conductor connections.
During the measurement, the power cable and additional connection cables with the inte-
grated protective conductor (e.g. monitor cable between the basic unit and monitor trolley)
must be moved section by section to detect cable breaks.
The protective conductor resistance must not exceed 0.2 ohms.
The values determined in the repeat measurement, including the measuring points, must
be recorded and evaluated in the protective conductor resistance protocol.
The measuring procedure and the measuring device used (designation and serial number)
must also be documented.
NOTE For evaluation purposes, the first measured value and the
values documented during previous maintenance or safety
checks must be compared to the measured values. A sud-
den or unexpected increase in the measured values may indi-
cate a defect in the protective conductor connections
(protective conductor or contacts) - even if the limit value of
0.2 ohms is not exceeded.
Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
If the local line voltage or line frequency differs from the delivery state of the system, the
values listed in the test protocol are invalid.
The values must be labeled invalid. The reason for newly determining the first measured
value must be documented and confirmed with date, name, and signature.
Adjust/program the system to the local line voltage/line frequency before the measurement.
In the case of systems without a system leakage current measurement value recorded in
the system test protocol, the system leakage current measurement must be performed
after all work has been completed.
Measurement
Perform the measurement according to DIN VDE 0751, Part 1 (see ARTD-002.731.17....),
and record the determined value as the first measured value.
Measurement of the system leakage current according to the differential current method
(measurement setup according to (Fig. 2 / p. 15)) must be given preference, since this
method is not dangerous to the person performing the measurement and other persons.
However, please note the minimum resolution of the system leakage current measuring
device and any additional manufacturer information restricting the use of the measuring
device.
Fig. 2: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
If the direct measurement of the leakage current is used (measurement setup according to
(Fig. 3 / p. 15)), the system must be insulated during the measurement and must not be
touched.
Fig. 3: Measuring circuit for direct measurement of the system leakage current in compliance with
The system must be switched on during measurement. Measuring devices with automated
measuring sequences must therefore be set to manual measurement.
Enter the highest value as the first measured value in the system leakage current protocol.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1/2001-10, Table F.1, row "system leakage current for units according to remarks 1
and 3" of
2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from the
nominal voltage, correct the measured value to the value corresponding to a measurement
at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
Separate the protocol sheet from these instructions and attach it to the protocols already
in the system binder or log book.
Repeat measurement
When maintenance or repair work is performed on the primary power supply circuit (e.g.
repairs to the power-on circuit or replacement of the line filter), the system leakage current
measurement must be repeated.
The same measuring conditions as in the first measurement apply.
Record and evaluate the highest value determined in the repeat measurement in the exist-
ing leakage current protocol.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1/2001-10, Table F.1, row "system leakage current for units according to remarks 1
and 3" of
2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from the
nominal voltage, correct the measured value to the value corresponding to a measurement
at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
NOTE For evaluation purposes, the first measured value and the
values documented during previous maintenance or safety
checks must be compared to the measured values. A sudden
or unexpected increase in the measured values may indicate
that a fault has occurred in the primary power supply circuit
(insulation damage, damage from moisture, defective inter-
ference suppressor, etc.) - even if the limit value of 2.5 mA is
not exceeded.
The log book is located in the monitor cart behind the keyboard. It is accessible if the cover
on the back of the monitor cart is removed.
The customer can have the power plug on the supplied power cable replaced with an
appropriate local plug by an electrician, provided that:
• The power plug used can conduct the power required for the operation of the ARCADIS
Avantic (refer to current and voltage values stated on the line voltage label, back of
monitor cart).
• The required line internal resistance is reached (refer to ARCADIS Avantic Planning
Guide).
• The selectivity of the fuse protection is assured - according to the relevant national
standards.
It must also be ensured by the customer that connection to an unsuitable power supply is
prevented (e.g., by identification or design measures).
As concluding measures, perform and document protective conductor measurements (in
Germany, e.g., in the medical device book).
The system delivery state is provided by the power connection values on the the line volt-
age label (back of the monitor cart).
Necessary adjustments to local conditions are to be performed as required according to
the adjustment instructions of the ARCADIS Avantic system.
The system leakage current measurement guidelines must be observed.
• Measure the voltage and frequency at the intended outlet for the ARCADIS Avantic sys-
tem.
• Compare the measured line voltage with the line voltage label on the back of the moni-
tor cart.
¹ The measured voltage (+/- tolerance) has to correspond to the voltage read from
the label.
NOTE Perform this procedure only if the on-site line voltage does
not correspond to the programmed line voltage of the system
delivered.
NOTE After the line voltage is adapted, the leakage current values
recorded in the system test protocol are no longer valid.
After closing all covers and successfully performing a pro-
tective conductor test, repeat and document the device leak-
age current measurement.
Observe the leakage current measurement instructions in
the "General Information" chapter of these instructions.
• Remove the back and bottom covers (Fig. 4 / p. 24) from the monitor cart.
• Pull the power plug out of the UPS in the monitor cart.
• Adapt the voltage by changing the connections to transformers T1 and T2. Refer to the
ARCADIS Avantic wiring diagram.
X • Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce-
dure.
• Mechanical functions of the monitor cart
• Please clarify in advance with the customer whether the following system configura-
tions are required.
Service login: 0
Fig. 7: Authentication_
• Enter the 6-character password and click OK (see system folder for password).
Fig. 8: Configuration_
• Select <Configuration>.
Fig. 9: Configuration
• Click <Next>.
• Start the service program and log in as described under "Service Login".
Configuration of the main system with SW version VC10A and service pack SP1
installed 0
• Start the service program and log in as described under "Service Login".
Buzzer Settings
[BUZZER OFF]: The audible radiation warning signal is not sounded during "radiation
release"*1.
[BUZZER ON]: An audible, pulsed radiation warning signal is sounded during "radiation
release"*1.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
0 BUZZER Radiation not - Only during "radiation - Automatically after 2 sec-
OFF blocked release"*1, after the con- onds of additional "radiation
figured <Warning Time release time (cumula-
Limit> time is exceeded, a tive)"*1.
continuous audible radia-
tion warning signal is
sounded
0 BUZZER Radiation not - Only during "radiation - If the audible radiation
ON blocked release"*1, an audible, warning signal was acti-
pulsed radiation warning vated by reaching the
signal is sounded . <Warning Time Limit>, the
- Only during "radiation audible continuous radia-
release"*1 and after the tion warning signal auto-
configured <Warning matically stops after 2
Time Limit> time is seconds of additional "radi-
exceeded, an audible ation release time (cumula-
continuous radiation tive)"*1.
warning signal is
sounded.
1 to 10 BUZZER - Block after 5 - Only during "radiation - If the audible radiation
OFF min. "radiation release"*1 and after the warning signal was acti-
release time configured <Warning vated by reaching the
(uninter- Time Limit> time is <Warning Time Limit>, the
rupted)"*1 exceeded, an audible audible continuous radia-
continuous radiation tion warning signal auto-
warning signal is matically stops after 2
sounded. seconds of additional "radi-
- Only during "radiation ation release time (cumula-
release"*1 and after 4 tive)"*1.
min. 30 sec. "radiation - If the audible radiation
release time (uninter- warning signal was
rupted)"*1 (= 30 seconds released by reaching the
before reaching the con- <Blocking Time Limit>, it
figured <Blocking Time can only be stopped by
Limit>), an audible con- deactivating the radiation
tinuous radiation warning release.
signal is sounded.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
1 to 10 BUZZER - Block after 5 - Only during "radiation - If the audible radiation
ON min. "radiation release"*1, an audible, warning signal was acti-
release time pulsed radiation warning vated by reaching the
(uninter- signal is sounded . <Warning Time Limit>, the
rupted)"*1 - Only during "radiation audible continuous radia-
release"*1 and after the tion warning signal auto-
configured <Warning matically stops after 2
Time Limit> time is seconds of additional "radi-
exceeded, an audible ation release time (cumula-
continuous radiation tive)"*1.
warning signal is - If the audible radiation
sounded. warning signal was
- Only during "radiation released by reaching the
release"*1 and after 4 <Blocking Time Limit>, it
min. 30 sec. "radiation can only be stopped by
release time (uninter- deactivating the radiation
rupted)"*1 (= 30 seconds release.
before reaching the con-
figured <Blocking Time
Limit>), an audible con-
tinuous radiation warning
signal is sounded.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
11 to 20 BUZZER - Block after 5 - Only during "radiation - If the audible radiation
OFF min. "radiation release"*1 and after the warning signal was
release time configured <Warning released by reaching the
(uninter- Time Limit> time is <Warning Time Limit>, it
rupted)"*1 exceeded, an audible can only be deactivated by
continuous radiation pressing the "audible radia-
warning signal is tion warning signal reset
sounded. The 7-segment button"*1. This also stops
radiation time display at the 7-segment radiation
the control console time display from flashing.
begins to flash. - If the audible radiation
- Only during "radiation warning signal was
release"*1 and after 4 released by reaching the
min. 30 sec. "radiation <Blocking Time Limit>, it
release time (uninter- can only be stopped by
rupted)"*1 (= 30 seconds deactivating the radiation
before reaching the con- release. The 7-segment
figured <Blocking Time radiation time display con-
Limit>), an audible con- tinues to flash. If necessary,
tinuous radiation warning the flashing can be stopped
signal is sounded. The with the "audible radiation
7-segment radiation time warning signal reset but-
display at the control con- ton"*1.
sole begins to flash.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
11 to 20 BUZZER - Block after 5 - Only during "radiation - If the audible radiation
ON min. "radiation release"*1, an audible, warning signal was
release time pulsed radiation warning released by reaching the
(uninter- signal is sounded . <Warning Time Limit>, it
rupted)"*1 - Only during "radiation can only be deactivated by
release"*1 and after the pressing the "audible radia-
configured <Warning tion warning signal reset
Time Limit> time is button"*1. This also stops
exceeded, an audible the 7-segment radiation
continuous radiation time display from flashing.
warning signal is - If the audible radiation
sounded. The 7-segment warning signal was
radiation time display at released by reaching the
the control console <Blocking Time Limit>, it
begins to flash. can only be stopped by
- Only during "radiation deactivating the radiation
release"*1 and after 4 release. The 7-segment
min. 30 sec. "radiation radiation time display con-
release time (uninter- tinues to flash. If necessary,
rupted)"*1 (= 30 seconds the flashing can be stopped
before reaching the con- with the "audible radiation
figured <Blocking Time warning signal reset but-
Limit>), an audible con- ton"*1.
tinuous radiation warning
signal is sounded. The
7-segment radiation time
display at the control con-
sole begins to flash.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
21 to 30 BUZZER - Block after - Only during "radiation - If the audible radiation
OFF 10 min. "radia- release"*1 and after the warning signal was acti-
tion release configured <Warning vated by reaching the
time (uninter- Time Limit> time is <Warning Time Limit>, the
rupted)"*1 exceeded, an audible audible continuous radia-
continuous radiation tion warning signal auto-
warning signal is matically stops after 2
sounded. seconds of additional "radi-
- Only during "radiation ation release time (cumula-
release"*1 and after 9 tive)"*1.
min. 30 sec. "radiation - If the audible radiation
release time (uninter- warning signal was
rupted)"*1 (= 30 seconds released by reaching the
before reaching the con- <Blocking Time Limit>, it
figured <Blocking Time can only be stopped by
Limit>), an audible contin- deactivating the radiation
uous radiation warning release.
signal is sounded.
21 to 30 BUZZER - Block after - Only during "radiation
ON 10 min. "radia- release"*1, an audible, - If the audible radiation
tion release pulsed radiation warning warning signal was acti-
time (uninter- signal is sounded . vated by reaching the
rupted)"*1 - Only during "radiation <Warning Time Limit>, the
release"*1 and after the audible continuous radia-
configured <Warning tion warning signal auto-
Time Limit> time is matically stops after 2
exceeded, an audible seconds of additional "radi-
continuous radiation ation release time (cumula-
warning signal is tive)"*1.
sounded. - If the audible radiation
- Only during "radiation warning signal was
release"*1 and after 9 released by reaching the
min. 30 sec. "radiation <Blocking Time Limit>, it
release time (uninter- can only be stopped by
rupted)"*1 (= 30 seconds deactivating the radiation
before reaching the con- release.
figured <Blocking Time
Limit>), an audible contin-
uous radiation warning
signal is sounded.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
31 to 50 BUZZER - Block after - Only during "radiation - If the audible radiation
OFF 10 min. "radia- release"*1 and after the warning signal was
tion release configured <Warning released by reaching the
time (uninter- Time Limit> time is <Warning Time Limit>, it
rupted)"*1 exceeded, an audible can only be deactivated by
continuous radiation pressing the "audible radia-
warning signal is tion warning signal reset
sounded. The 7-segment button"*1. This also stops
radiation time display at the 7-segment radiation
the control console time display from flashing.
begins to flash. - If the audible radiation
- Only during "radiation warning signal was
release"*1 and after 9 released by reaching the
min. 30 sec. "radiation <Blocking Time Limit>, it
release time (uninter- can only be stopped by
rupted)"*1 (= 30 seconds deactivating the radiation
before reaching the con- release. The 7-segment
figured <Blocking Time radiation time display con-
Limit>), an audible contin- tinues to flash. If necessary,
uous radiation warning the flashing can be stopped
signal is sounded. The with the "audible radiation
7-segment radiation time warning signal reset but-
display at the control con- ton"*1.
sole begins to flash.
Pro- Pro- Time until Audible radiation warn- Audible radiation warning
grammed grammed radiation is ing signal sounds signal stops
value parameter blocked
<Blocking <Buzzer
Time Limit> mode dur-
ing x-ray>
31 to 50 BUZZER - Block after - Only during "radiation - If the audible radiation
ON 10 min. "radia- release"*1, an audible, warning signal was
tion release pulsed radiation warning released by reaching the
time (uninter- signal is sounded . <Warning Time Limit>, it
rupted)"*1 - Only during "radiation can only be deactivated by
release"*1 and after the pressing the "audible radia-
configured <Warning tion warning signal reset
Time Limit> time is button"*1. This also stops
exceeded, an audible the 7-segment radiation
continuous radiation time display from flashing.
warning signal is - If the audible radiation
sounded. The 7-segment warning signal was
radiation time display at released by reaching the
the control console <Blocking Time Limit>, it
begins to flash. can only be stopped by
- Only during "radiation deactivating the radiation
release"*1 and after 9 release. The 7-segment
min. 30 sec. "radiation radiation time display con-
release time (uninter- tinues to flash. If necessary,
rupted)"*1 (= 30 seconds the flashing can be stopped
before reaching the con- with the "audible radiation
figured <Blocking Time warning signal reset but-
Limit>), an audible contin- ton"*1.
uous radiation warning
signal is sounded. The
7-segment radiation time
display at the control con-
sole begins to flash.
Terms used *1
• Radiation release
¹ The hand or foot switch for radiation release is activated (pressed) by the user.
Radiation is active.
• Radiation release switch
¹ Hand or foot switch to release radiation
• Radiation release time (uninterrupted)
¹ The timer for the radiation release time (uninterrupted) starts with the radiation
release.
¹ The timer for the radiation release time (uninterrupted) stops with the deactiva-
tion of radiation release.
¹ In this case, the timer for the radiation release time (uninterrupted) is reset to 0.
• If necessary, press the combination of buttons to increase or decrease the code value
until the correct code value is displayed at the 7-segment display of the control console.
(4/Fig. 19 / p. 49).
¹ Australia: select code value 1.
¹ Switzerland: select code value 2.
¹ Other countries: do not start the country-specific configuration or select code
value 0.
¹ This value is saved automatically once selected.
• Wait a few seconds, then pull plug X10 from the base unit.
¹ The base unit is off.
• Reconnect plug X10 at the base unit and wait for it to boot.
¹ The configuration is valid once the reset of the base unit is complete.
NOTE Once the configuration has been changed, the label affixed
to the footswitch has to be removed and replaced by the new
label included in the delivery volume. In the case of an
"enhanced" footswitch, there are two labels on the pedals
that have to be replaced.
• Remove the label affixed to the footswitch. In the case of an "enhanced" footswitch,
remove both labels from the pedals.
• If necessary, remove any remaining adhesive where the labels were removed.
• Affix the new label to the footswitch. In the case of an "enhanced" footswitch, affix both
labels to the pedals.
• If the footswitch configuration was changed, check the footswitch pedal function.
¹ The footswitch pedal functions have to correspond to the functions indicated on
the newly affixed label(s).
Backup 3.1
• Click <Go> and wait until the message “Ready” appears in the footer.
NOTE If the printer was ordered together with the system, then the
printer cables have already been routed, and all mechanical
work steps and programming for the ARCADIS system have
already been performed at the factory.
Paper printers delivered after the system was delivered must
be installed and started up in accordance with the "Installa-
tion and Setting Instructions" SPR2-310.814.03...
Perform the following steps according to the user manual for the printer (CD-ROM):
• Unpack the printer.
• Set the switch on the back of the Sony printer as follows:
¹ Analog/Digital --> Digital
• Load the paper roll or film roll into the printer (follow instructions in user manual).
• Plug the power and USB cables into the printer.
• Switch on printer. After the self-test, print out a test page.
• If problems occur, use the printer's user manual for troubleshooting.
IQ test 0
Completing protocols 0
After completing all adjustment work steps and check measurements, the responsible
technician must sign and date the test protocols and country-specific protocols to confirm
that all values have been correctly determined and recorded.
Transferring the values from the test protocol to the "protective conductor" protocol
• Complete the "Protective Conductor Resistance" protocol found in the "Protocol" sec-
tion of this manual.
• If the system test protocol only contains the confirmation of the protective conductor
resistance threshold, enter this value in the "Protective conductor resistance" table, row
"Measurement point 1", column "First measured value" (< 0.18 Ohm). Enter "n.a." in
rows "measurement point 2" through "measurement point 22" of column "First mea-
sured value."
• If the system test protocol specifies the measured protective conductor resistance val-
ues and their measurement points, transfer the values to rows "Measurement point 1"
through "Measurement point 22" of column "First measured value". Mark empty fields
with "n.a." as "not applicable".
• In row "(1) Meas. circuit", column "First measured value", enter the text "VDE0751-1
Image C2".
• If specified in the system test protocol, read off the measuring equipment used at the
factory for the protective conductor test.
¹ Enter this measuring equipment in the "Protective conductor resistance" proto-
col, in the "Measuring device, type:" row.
• If specified in the system test protocol, read off the identification of the measuring
equipment specified in the test protocol.
¹ Enter the ID of the measuring equipment in the "Protective conductor resistance"
protocol, in the "Measuring device, Ser. No.:" row.
• Mark the following rows in the "Protective conductor resistance" protocol with "n.a." as
"not applicable":
¹ Row "Meas. inst. calibrated up to:"
¹ Row "Evaluation:"
• Complete the following rows in the "Protective conductor resistance" protocol:
¹ In the row "Date:" enter the current date.
¹ In the "Name" row, enter your name in block letters.
¹ Sign the protocol in the "Signature" row.
• If you want to complete the protocol with additional remarks, enter them in the
"Remarks" table in the protocol addendum.
Measurement points
Measurement from the protective conductor connection of the power plug to the:
Base unit, handle attachment screw for I.I. cover
Base unit, one attachment screw at I.I. grid
Base unit, one attachment screw at I.I. flange
Base unit, one attachment screw at I.I. cap
Base unit, one attachment screw at POWERPHOS
Base unit, one screw at cover of horizontal carriage
Base unit, one lateral screw at brake pedal (left or right)
Base unit, back cover (electronics box)
Base unit, equipotential bonding connection
Base unit, near footswitch cable outlet (electronics box)
Base unit, one lifting column screw at the base plate of the electronics box
Base unit, base plate of electronics box (at unpainted location)
Base unit, one screw at the back of each monitor
Monitor cart, one screw at monitor support
Monitor cart, one screw at the small, top back cover
Monitor cart, one screw at the large, lower back cover
Monitor cart, one screw at back cover of switch-on assembly
Monitor cart, one screw at the front cover
Monitor cart, one screw at the right side cover
Monitor cart, one screw at the left side cover
Monitor cart, one screw at the right handle
Monitor cart, one screw at the left handle
Transferring the values from the test protocol to the "device leakage current" proto-
col
NOTE If the programmed line voltage has not been changed from its
status at delivery, and there have been no changes of parts
in the system's primary circuit, the device leakage current
value specified in the test protocol delivered with the system
is valid.
The device leakage current test need not be performed.
The value recorded in the test protocol must be used as the
"first measured value" in the "Device Leakage Current" pro-
tocol.
Perform the device leakage current measurement if the con-
ditions specified above do not apply.
Refer to the section "Device leakage current measurement
during start-up (if necessary)".
• Complete the "Device leakage current" protocol found in the "Protocol" section of this
manual.
• Enter the system serial number on the "Serial Number" row of the "Device leakage cur-
rent" protocol.
• If applicable, enter the customer-specific system ID number in the "Customer-specific
ID Number" row of the "Device leakage current" protocol.
• In test protocol 1, table "Alternative leakage current measurement (DIN VDE 0751)" or
"Device leakage current measurement (DIN VDE 0751-1), read off the higher recorded
value between rows "Leakage current under normal operation and with nominal volt-
age" and "Leakage current with power reversion of polarity and nominal voltage."
¹ Enter the higher value from test protocol 1 in the "Device leakage current" proto-
col, row "Device leakage current (highest measured value) [mA]" and in row
"Device leakage current corrected value [mA]", column "First measured value".
• Read off the line voltage in test protocol 1 under "Revision Level".
¹ Enter the value from test protocol 1 in the "Device leakage current" protocol, row
"Line voltage during measurement [V~]", column "First measured value".
• Read off the production resources entered in test protocol 1, table "Production
resources used".
¹ Enter the production resources from test protocol 1 in the "Device leakage cur-
rent" protocol, row "Measuring device: type", column "First measured value".
• Read off the measuring device ID entered in test protocol 1, table "Production
resources used".
¹ Enter the measuring device ID from test protocol 1 in the "Device leakage cur-
rent" protocol, row "Measuring device, ser. no.", column "First measured value".
• In "Device leakage current" protocol, row "(1) Meas. circuit", column "First measured
value", enter the text "VDE0751-1 Image C5".
• Mark the following rows in the "Device leakage current" protocol, column "First mea-
sured value", with "n.a." as "not applicable":
¹ Row "Meas. inst. calibrated up to:"
¹ Row "Evaluation:"
• Complete the following rows in the "Device leakage current" protocol, column "First
measured value":
¹ In the row "Date:" enter the current date.
¹ In the "Name" row, enter your name in block letters.
¹ Sign the protocol in the "Signature" row.
• If you want to complete the protocol with additional remarks, enter them in the
"Remarks" table in the protocol addendum.
NOTE If the programmed line voltage has been changed from its
status at delivery, and/or there have been changes of parts in
the system's primary circuit, the device leakage current value
specified in test protocol 1 delivered with the system is no
longer valid.
The device leakage current measurement has to be per-
formed.
• Perform the device leakage current measurement after successfully completing the
protective conductor test.
• Complete the "First measured value" column of the "Device leakage current" protocol.
• If you want to complete the protocol with additional remarks, enter them in the
"Remarks" table in the protocol addendum.
On the ARCADIS systems, SRS should be configured. In addition to SRS, MNP should be
installed so that RUH can be enabled.
Remote connection
SRS configured Yes No
MNP installed Yes No
The manufacturer of this product requires information and the legislator demands proofs
that a product delivered free of defects from the factory continues to possess the required
and certified quality properties on installation and start-up.
It is therefore absolutely necessary that the installation report with the installation and
start-up data is sent without delay after completion of the work to the address stated on the
installation report.
You will find the report in the system folder.
Tab. 8
Measuring circuit 0
Fig. 27: Measuring circuit for measuring the protective conductor resistance for units that are
disconnected from power, in compliance with DIN VDE 0751-1/2001-10, Fig. C2.
Pos. 1 Measurement setup (measuring device)
Pos. 2 System
Pos. 3 Application part (if applicable)
Remarks: 0
Tab. 9
Tab. 10
Measuring circuit 0
Direct measurement
Fig. 28: Measuring circuit for direct measurement of the system leakage current in compliance with
DIN VDE 0751-1/2001-10, Fig. C5 for protection class I.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
Fig. 29: Measuring circuit for direct measurement of the system leakage current in compliance with
DIN VDE 0751-1/2001-10, Fig. C5 for protection class II.
Pos. 1 = System
Pos. 2 = Application part (if applicable)
Pos. 3 = Measurement setup (integrated into measuring device)
Differential measurement
Fig. 30: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
Fig. 31: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class II.
Pos. 1 = System
Pos. 2 = Application part (if applicable)
Pos. 3 = Measurement setup (integrated into measuring device)
Remarks: 0
Tab. 11
Name:
Phone/fax:
E-mail:
Not networked
Identification:
Computer name * Configuration/Local Host/TCP IP address
IP address:
TCP/IP address * Configuration/local host/TCP IP address
Print * Configuration/DICOM/general
Comments
Network data for HIS/RIS network partner (only with HIS/RIS option)
Tab. 14 Network data for HIS/RIS
Not available
Device: Version: Manufacturer:
Host properties:
Host name * Configuration/DICOM/HIS RIS nodes
Application entity
AE title * Configuration/DICOM/HIS RIS nodes
Comment:
Application entity:
AE title * Configuration/DICOM/network nodes
Comment:
Network data for diagnostic console/archive system (only with DICOM Basic option)
(2)
Tab. 16 Network data for reporting console/archiving system
Application entity:
AE title * Configuration/DICOM/network nodes
Comment:
Application entity:
AE title * Configuration/DICOM/Print devices
Comment:
Network data for a DICOM camera in the network (only with the DICOM Print option)
(2)
Tab. 18 Network data for DICOM camera in the network
Application entity:
AE title * Configuration/DICOM/Print devices
Comment:
Chapter 3 Change
Main system configuration Completely revised and additions regarding SW SP1
Country-specific configura- incorporated.
tions Country-specific configurations for Australia and Switzer-
land added.