MU-N Series: User's Manual LR-W500 (C) Edition
MU-N Series: User's Manual LR-W500 (C) Edition
MU-N Series: User's Manual LR-W500 (C) Edition
KEYENCE Corporation has confirmed, on the basis of the following specifications, that this
product complies with the essential requirements of the applicable EU Directive. Be sure to
consider the following specifications when using this product in a member state of the
European Union.
MU-N Series These specifications do not give any guarantee that the end-product with this product
incorporated complies with the essential requirements of the EMC Directive. The manufacturer
of the end-product is solely responsible for confirming the compliance of the end-product itself
User's Manual according to the EMC Directive.
1-4 Specifications
This product can be used by connecting it to KEYENCE sensor models that support the MU-N
Series. MU-N11 MU-N12
Model
It can also be used with product series that feature models that support N-bus (the name of Main unit Expansion unit
Connected sensor LR-W500(C)
KEYENCE's wire-saving system).
Single output: 300 μs/1.1 ms/11 ms/100 ms/500 ms
For details on the these functions, see the "MU-N Series User's Manual". selectable
There is a separate "MU-N Series User's Manual" for each sensor model that can be Response time
Multiple output: 2 ms/3 ms/11 ms/100 ms/500 ms
connected. These manuals can be downloaded from the KEYENCE website selectable
(http://www.keyence.com). Please contact your nearest KEYENCE office, when it is not Mutual interference reduction function Up to 2 units with alternate frequencies set
possible to download these files over the internet. Timer function OFF/ON delay/OFF delay/One-shot
Power supply voltage 24 VDC, ripple (P-P) 10% or less, Class 2 or LPS
Power
135 mA or less 120 mA or less
Read this manual before using the product in order to achieve maximum performance. supply Current consumption
(without load)*1 (without load)*2
Keep this manual in a safe place after reading it for future reference.
4 max.,
NPN open collector/PNP open collector selectable,
The following symbols alert you to important messages. Be sure to read these sections
24 V or less,
carefully.
Control output Main unit: 50 mA or less *3 ,
Expansion unit: 20 mA or less,
Indicates a hazardous situation which, if not avoided, could result in death Residual voltage: 2 V or less,
WARNING N.O./N.C. selectable
or serious injury.
Indicates a hazardous situation which, if not avoided, could result in minor 5 max.,
CAUTION
or moderate injury. Short-circuit current: NPN: 1 mA or less,
Indicates a situation which, if not avoided, could result in product damage External input PNP: 2 mA or less,
NOTICE For the applied voltage, see wiring diagrams
as well as property damage.
*4 (page 3 of the user's manual)
I/O
• This product uses DC power. The product may explode or burn if an AC Ambient humidity 35 to 85%RH (no condensation)
voltage is applied. Environmental
resistance Shock resistance 1,000 m/s 2 in X, Y, Z axis directions respectively 6 times
• Do not wire this product along with power lines or high-tension lines.
Doing so may lead to product malfunctions or damage due to noise.
NOTICE 10 to 55 Hz Double amplitude 1.5 mm in the X, Y, Z axis
• Do not use this product outdoors or in a place where extraneous light can Vibration resistance
directions respectively, 2 hours
enter the light-receiving element directly.
Case and dust cover: Polycarbonate
Material Buttons: Polyacetal
Display panel: PMMA
1-2 Precautions on Regulations and Weight Approx. 70 g Approx. 70 g
-1-
1-5 Names of Main Unit Components 2-2 Connecting the Sensor and Power
Output Display Arrow
Cables
indicator buttons
• Turn the power off before connecting the sensor cable and power cable.
NOTICE • Insert the sensor cable and power cable securely into their respective
ports. If not properly inserted, the cables can be easily disconnected.
SELECT
Sensor cable
OUT
STB H A
MODE Cable type sensor
L MU-N
SET
Multi-Sensor Controller
Up to four expansion units can be connected to one main unit. aligning the pin numbers on the connector
Connector
* You can also connect up to four models that support N-bus (the name of body with the corresponding wire colors, lock
KEYENCE's wire-saving system) to one main unit. refer to the table below for more information.
Press fitting
Mount the product on a DIN rail and install on a metal surface when
CAUTION Connector pin
connecting multiple expansion units or mounting main units close together. Wire color
number Black/ Blue/ White/ Brown
NOTICE Turn the power off before connecting expansion units.
1 Brow n
2 White
Point • Contact your nearest KEYENCE office when connecting units other than
N-bus (the name of KEYENCE's wire-saving system) compatible sensors. 3 Blue
• Turn the power off before connecting expansion units. 4 Black
• Do not touch the expansion unit connectors.
• When using the MU-N Series as the main unit, verify that the expansion 3 To make a connection, press together the body and the connector cover into
products used (other than MU-N Series expansion units) can operate which the cables were inserted. (Use a pair of pliers or a similar tool to press
within the power supply voltage range of the MU-N Series. them together.)
1 Remove the protection covers from If the sensor cable needs to be cut to a shorter length, install the
the main unit and expansion unit(s). connector as described above.
-2-
Power cable Analog output*1
Connect the power cable to the right side
of the controller. NPN/PNP
Brown*1(1*2)
24 V
White(2* )2
*1 MU-N11 only
* Connector cover
The sensor cable and power cable are both equipped with a connector cover. *2 Pin numbers when using an M12 connector cable
After the connectors are inserted into the main unit, place the connector covers
over the connectors. Finally, secure the connector covers with the connector
security rings. 2-4 Installing the Sensor
• Tightening torque for the mounting holes: 0.63 N·m (M3 screw)
2-3 Wiring • If the workpiece contains a glossy surface that could interfere with stable
detection, tilt the sensor approx. 15° to 20°. If tilting the sensor does not
improve detection, please attach the reflection canceling attachment
Power cable pin layout
(LR-WA1).
The tables below provide a breakdown of the power cables that can be used with
the MU-N Series. They also contain information on the associated wire colors and
15 to 20°
functions.
B3 Pink Input 4
B4 Violet Input 5
SET
FOCUS
FOCUS
Load (input device) Analog current or voltage input device • Turn the dial to the right to decrease the focal distance.
Control output
The focal point moves
NPN PNP closer to the sensor.
Brown*1(1*2) Brown*1(1*2)
24 V 24 V
Black(4*2)/White(2*2)/ Black(4*2)/White(2*2)/
Orange/Gray Orange/Gray • Turn the dial to the left to increase the focal distance.
(when output is (when output is
selected) selected)
Blue*1(3*2) Blue*1(3*2)
0V 0V
Blue*1(3*2) Blue*1(3*2)
0V 0V
-3-
3. Initial Settings *3 R (red) amount, G (green) amount, and B (blue) amount are values related to
that specific color elements portion of the full RGB reading. All three
components will total 1000.
When the MU-N Series and an LR-W500(C) are connected, the button For the selection method, see "8-2 Switching Banks through External Inputs"
NOTICE
operations for the LR-W500(C) main unit are disabled.
(page 11).
Depending on the setup statuses of the MU-N Series and the LR-W500(C), the
following initial settings are required when power is turned on. Selecting Control Output Setup
When using the control output(s), select from the following:
Condition Required initial settings • 1 output 16 banks [1out 16bank]: 1 output wire with 16 banks available
Both the MU-N Series and the LR-W500(C) 3-1 Settings when both • 4 outputs 2 banks [4out 2bank]: 4 output wires with 2 banks available
are in factory default. units are Initialized • Binary output (Up to 15 patterns) [Binary out]: 4 output wires with 1 to 15 signal
The settings on the MU-N Series and the 3-2 Selecting the Settings combinations possible
LR-W500(C) do not match. to Use (See "8-2 Switching Banks through External Inputs" (page 11).)
The settings on the MU-N Series and the Initial settings are not
LR-W500(C) match. required. Selecting I/O for white wire (2)
* When "6-2 Initialization" (page 6) is performed, "3-1 Settings when When "1out 16bank" is selected, the Input or Output functionality of the white wire
both units are Initialized" (page 4) is required. (2) can be selected.
• External input [Input]
• Control output [Output]
3-1 Settings when both units are * Disabled when [Off] is selected. See "8-1 Input/Output Wire Settings List"
(page 11).
Initialized
MODE
3-2 Selecting the Settings to Use
Press or to advance through the menus.
When the settings on the MU-N Series and the LR-W500(C) do not match at
Press to return to the previous menu.
power-on, the desired settings need to be chosen from either the sensor or the
Press or to select/change menu options. controller.
Power ON
Selecting Select Output
NPN Type mismatch Type match
NPN/PNP
NPN PNP
(page 4)
Sensor type
Sub unit Main unit is mismatch
Off
Selecting Analog Out 4-20mA MODE
Analog Off 0-10V
Output
(page 4) Off ≠Off Disp. Value
Copy From Controller
RGB Data
Controller Sensor
Selecting Analog Data
Analog Data Disp. Value MODE
(page 4) >3s
Selecting Control Output Mode 1out 16bank Return to the initial settings screen.
Output Setup 1 out 16 bank 4out 2bank
(page 4) Binary out Controller*1 (Default*2) Sends the MU-N settings to the LR-W500(C).
≠1out 16bank Sensor Sends the LR-W500(C) settings to the MU-N.
When 1out 16bank is selected
*1 If the settings stored in the MU-N cannot be sent to LR-W500(C), [Initialize] is
Selecting Wire2 Function Off
I/O for white Off
displayed instead of [Controller]. Selecting [Initialize] requires
Input
wire (2) "3-1 Settings when both units are Initialized" (page 4).
Output
(page 4) *2 If the connected MU-N is in the initial setting, [Sensor] is the default value.
4. Basic Settings
Return to the initial settings screen.
-4-
5. Sensitivity Adjustment < Precautions for master calibration>
• Continue calibration until the green light that indicates reference addition
does not turn on any more.
5-1 Auto/C+I/C Mode • If the master calibration is performed again, the registered contents from the
first master calibration will be overwritten. To add an allowable range after
the master calibration, perform the master addition calibration.
About the display value
• If the registration status is saturated and "---" is displayed. Perform the
• Conformity
master calibration again after lowering the "7-12 Master Calibration Set
The level of conformity of the current detected workpiece to the registered
Value" (page 10).
reference workpiece.
• Changing the master calibration set value after a master calibration has
Display range: 0 to 999 (The more the workpiece conform to reference workpiece,
been performed, does not affect the current setting value, only subsequent
the higher the value.)
calibrations.
• Setting value
The threshold of conformity at which a workpiece is judged to be the same as the Permitting color variations between different
registered workpiece. workpieces
To check or manually make fine adjustments to the setting value, Master addition calibration (when adding workpieces to be permitted)
see " Checking and adjusting the setting value" (page 5). Position a workpiece which is to be judged the same as the current
* The blinking numeric value that appears after calibration is the setting value. registered color. Then press and hold the [SET] button and the [] button.
When the added registration is successful, the "setting value" flashes three
Setting the sensitivity (apply one of the following three times, and the sensor returns to the normal screen (the setting value is not
changed at this point in time).
methods)
In this case, references are added to permit colors between "the current
1-point calibration (use to detect 1 specific color)
registered color" and "the additional registered color".
Register the color of the workpiece to be detected.
(When Auto mode is used, this function operates in C+I.)
1 OUT
SET
STB H
L
MU-N12
SELECT
MODE
Multi-Sensor Controller
1 MU-N12
FOCUS
SELECT
OUT
STB H
MODE
L
SET
SET
Multi-Sensor Controller
Press and
FOCUS
1 MU-N12
SELECT
2 MU-N12
SELECT
OK 750
STB H STB H
MODE MODE
L L
SET SET
Multi-Sensor Controller Multi-Sensor Controller
FOCUS
FOCUS
SET
[SET] [SET] When the In contrast, when the setting value is reduced, a wider detection tolerance is
button (for button difference is enabled. The [] and [] buttons can be used to increase or decrease the
1 second or again (for too small.
setting value.
less). 1 second
or less).
NG ---
MU-N12
SELECT RUN
OUT
STB H
Master calibration (use to permit color variations within the same SET
Multi-Sensor Controller
(for 1 second or less).
workpiece)
Press the [SET] button to register the reference color. Then, press and hold
Setting value
the [SET] button to perform sampling. During sampling, references are
OUT BANK16
845
added and are set to be judged as the same color. When a reference is
added, the indicator flashes (once) in green.
1 0 ... 999
When master calibration is executed, the setting value becomes 950
(default). To change this value, see "7-12 Master Calibration Set Value"
(page 10). <Note for sensitivity setting and setting value adjustment >
(When Auto mode is used, this function operates in C+I.) *1 When using the following settings, the master calibration and master addition
calibration cannot be used.
• When using 1 output 16 banks, banks 5 through 16 cannot use these
1 OUT
H
MU-N12
SELECT
2 OUT
H
MU-N12
SELECT
functions.
STB STB
MODE MODE
L L
SET SET
Multi-Sensor Controller Multi-Sensor Controller
FOCUS
FOCUS
SET
Press the Press and OK 950 • When using the binary output mode
[SET] hold the
button (for [SET]
1 second or button for *2 After master calibration or master addition calibration has been executed, the
less). 3 seconds setting value cannot be increased.
or more.
-5-
When the received light intensity is saturated or
5-2 Super I Mode
insufficient
When using the product with the 300 μs*1 or 1.1 ms*2 response time selected,
About the display value
stable operation may be reduced. In this situation, it may be possible to increase
• Received light intensity
stability by adjusting the light intensity to the optimal value using the steps below.
The current received light intensity is displayed.
Display range: 0 to 999 (The greater the received light intensity, the higher the
*1 2 ms during multiple outputs
value.)
*2 3 ms during multiple outputs
• Setting value
The threshold at which the received light intensity is judged to indicate that a 1 OUT
STB H
L
MU-N12
SELECT
MODE
workpiece is present.
SET
Multi-Sensor Controller
FOCUS
To check or manually make fine adjustments to the value, see " Checking
SET
Press the
and adjusting the setting value" (page 6).
[SET] and [] OK
* The blinking numeric value that appears after calibration is the setting value. buttons at the
same time (for
Setting the sensitivity (apply one of the following three 1 second or
less.)
methods)
2-point calibration (basic intensity differentiation)
1 OUT
SET
STB H
L
Multi-Sensor Controller
MU-N12
SELECT
MODE
2 OUT
SET
STB H
L
Multi-Sensor Controller
MU-N12
SELECT
FOCUS
SET
MU-N12
SELECT
RUN
OUT
1 OUT
SET
STB H
L
MU-N12
SELECT
MODE
If necessary, enter the password to release the lock.
Multi-Sensor Controller
1 OUT
SET
STB H
L
MU-N12
SELECT
MODE
OUT
SET
STB H
L
MU-N12
SELECT
MODE
FOCUS
6-2 Initialization
SET
SET
Press and
hold the
OK 500
[SET] button It is possible to reset the product to its factory default settings. After initialization,
for 3 seconds
the user must configure the settings again.
or more.
"3-1 Settings when both units are Initialized" (page 4)
When "SET" flashes, continue After passing the
holding the [SET] button and workpiece in front of the RUN
MU-N12
pass the workpiece in front of sensor, release the SELECT
Yes
Initialize? No
MU-N12
SELECT
RUN No
OUT
STB H
SET
L
MODE
Press the [] button
MODE
Multi-Sensor Controller
(for 1 second or less).
-6-
6-3 Switching the Display Screen
Basic screen
• 7 segment display
When tuning, the bank and channel of the
Current value output that is having its setting value
changed is displayed.
Press
MODE
+
MODE
(Press + to return to the
previous screen.)
• Display
1 845 12
34
845 15 845
[OUT]: Illuminates when the control output is ON.
[1 to 15]: The number of the channel whose control output is ON The setting value
illuminates.*2
[C+I]: The selected mode illuminates.*3
*1 When the analog output [RGB Data] is selected, 1output 4banks is set.
Therefore, the bank can be selected from BANK1 to 4. (only for MU-N11)
*2 When using the binary output, the channel No. with the output ON illuminates.
*3 When super I mode is selected, press + to display the light source
being used to make judgments.
OUT BANK16
Setting value
MODE
Graphic mode*1
1 P
B
923
17 Bottom value*3
MODE
-7-
7. Advanced settings
The value shown on the display screen is the
OUT BANK16 Off
845
initial value.
1 Pressing and holding the Mode key exits the
7-6 Timer
(page 9)
Bank1 Timer
Off
On delay
Off delay
setting screen. One shot*1
MODE
MODE Press or to advance through the menus. MODE
>3S Press + to switch the bank No.
Press to return to the previous menu. Pressing these buttons on the timer duration setting screen
moves to the timer selection screen.
Press or to select/change menu options.
=Off ≠Off
999
Auto
to
C+I
7-1 Detection Bank1 Timer 1
Detect All C
Mode Super I*1 10 ms
(page 9)
Auto
Separate Set
≠Separate Set
When Separate Set is selected,
MODE
press + to switch the bank or channel.
When the Analog Output is used (only for main unit)
Other
Auto 999
7-1 Detection 7-7 Analog Analog Scaling
Bank1 Detect C+I to
Mode Lower and
Auto C 4 mA = 0 0
(page 9) Upper Limits
Super I*1
(page 10)
=Off
Other When the Timer of the selected bank is set to Off
OUT BANK16
1 845
End
7-13 Graph Mode On Peak
Graphic On Trigger
Mode On Peak
Return to the basic screen. (page 10)
-8-
7-5 Output Modes
Off
7-15 Password Password Lock
On Stability [Stability]
(page 11)
Off
This function can indicate when the amount of light received is not stable.
=On This output will turn ON when the amount of light is unstable and the stability
indicators do not illuminate.
=Off
9999
Enter Password
to
Error [Error]
0000 0000 This output turns ON when an error occurs. As soon as the error is cleared, the
output is automatically reset. For the cause of the error, see "9-1 Error
Input from the top digit. Display" (page 12).
Press to move between digits.
7-6 Timer
OUT BANK16
1 845
End
This function can be used to delay the timing of the sensor output switching.
• On delay
• Off delay
Return to the basic screen.
• One shot
ON
7-2 Response Time N.C.
OFF
The longer the response time, the more reliable and stable the detection will be. ON
N.O.
On delay
When detection is unstable due to the workpieces moving at a high speed, set the OFF
response time to a smaller value. ON
N.C.
OFF
OFF
(N.O./N.C. Selection)
ON
N.C.
Set the output logic to N.O. or N.C.. OFF
• N.O.: Turns the output on when the registered condition is met (turns the output
ON
on when light is received) *. N.O.
One shot
• N.C.: Turns the output on when a condition other than the registered condition OFF
is met (turns the output on when light is not received) *. ON
N.C.
OFF
* The condition within parentheses indicates the condition when super I mode is Timer duration
selected.
ON
Input
OFF
> 60 ms > 60 ms
ON
Input
OFF >100 ms
< 60 ms < 60 ms
-9-
7-7 Analog Lower and Upper Limits 7-9 Screen Brightness
Configure scaling settings, if necessary. Screen brightness after no operation has been done for a certain period of time
can be configured.
Current output (4 to 20 mA) Item Description
Initial settings When settings are changed 100% The display brightness is always kept at 100%.
(example) After a certain period of time, the display brightness
30% (default value)
will be set to 30%.
Analog 4 mA 0 Analog 4 mA 400
After a certain period of time, the display will be turned
Analog 20 mA 999 Analog 20 mA 800 Display Off
OFF.
20.5 mA
20 mA
• After using LR-W for a long period of time, the brightness will gradually
20 mA
decrease.
• When screen brightness is set to [100%], the brightness will decrease at a
faster rate.
4 mA 4 mA • When screen brightness is set to [Display Off], the 7 segment display will show
3.8 mA
as follows.
0 999 0 400 800
Analog 4 mA 800
Analog 20 mA 400
* Display will be back to normal state after button operation.
20.5 mA
20 mA
0 400 800 The effect of mutual interference can be reduced by changing the light emission
For details on the analog output when an error occurs, period. When using multiple LR-W Series units in close proximity, set each unit to
see "9-2 Displays Other than Numeric Values" (page 12). a different light emission period. When selecting frequency B (alternate
frequency), the response time becomes approximately 20% slower.
Voltage output (0 to 10 V)
Initial settings When settings are changed 7-11 Detection Light Source
(example)
Analog 0 V 0 Analog 0 V 400 When using Super I mode, the light source used for detection is automatically
Analog 10 V 999 Analog 10 V 800 selected to provide optimal performance. To require the sensor to use a specific
light source, adjust this setting to Red, Green, Blue, or RGB.
10 V 10 V
This graph displays the peak values that occur during each output cycle.
7-8 Analog Output Test The peak value is updated each time the output turns off*.
It is possible to output arbitrary current or voltage values to ensure the analog • Vertical axis: Peak value while the output was ON
signal is properly received by the analog input device. • Horizontal axis: The number of times the output turned ON. The maximum
value is selectable from the following :
50 (default), 500, 2000, 10000
- 10 -
Trigger at ON [On Trigger]
8-2 Switching Banks through External
999
Inputs
941 0.1s To switch banks through external inputs, the unit needs to be in a key locked
status.
The transition of the display value before and after the output turned ON for the Set the key lock beforehand. "6-1 Key Lock" (page 6)
most recent output cycle will be plotted on the graph*.
• Vertical axis: Display value at ON
Bank inputs for 1 output 16 banks
• Horizontal axis: Time and display range selectable from the following :
0.1 sec. (default), 1 sec., 10 sec. Wire color Orange Gray Pink Violet
I/O No. Input 2 Input 3 Input 4 Input 5
Wiring
* Operation for N.O. ON and OFF are reversed for N.C. Bank Bank Bank Bank
External input
input A input B input C input D
BANK1 OFF OFF OFF OFF
7-14 Sensor Indicator Blinking Function BANK2 ON OFF OFF OFF
BANK3 OFF ON OFF OFF
While [Yes] is selected, the sensor indicator blinks in green. This function notifies BANK4 ON ON OFF OFF
users which sensor is being set. Exiting this screen returns the setting to [No]. BANK5 OFF OFF ON OFF
BANK6 ON OFF ON OFF
BANK7 OFF ON ON OFF
BANK8 ON ON ON OFF
7-15 Password Bank No.
BANK9 OFF OFF OFF ON
BANK10 ON OFF OFF ON
If you set this to [ON], you can set a personal identification number that must be BANK11 OFF ON OFF ON
entered to release the "6-1 Key Lock" (page 6). You can set the personal BANK12 ON ON OFF ON
identification number to a value from 0 to 9999. BANK13 OFF OFF ON ON
BANK14 ON OFF ON ON
8. Input/Output Wire Settings BANK15 OFF ON ON ON
BANK16 ON ON ON ON
ON
Input 2
OFF >100 ms
< 60 ms < 60 ms
- 11 -
8-3 Control Outputs during Binary Output 9-3 WARRANTIES AND DISCLAIMERS
Wire color Black (4) White (2) Orange Gray
Wiring
(1) KEYENCE warrants the Products to be free of defects in materials and
I/O No. 1 2 3 4
Control output Binary output A Binary output B Binary output C Binary output D workmanship for a period of one (1) year from the date of shipment. If any
No match OFF OFF OFF OFF models or samples were shown to Buyer, such models or samples were used
Matches ch1 ON OFF OFF OFF
merely to illustrate the general type and quality of the Products and not to
Matches ch2 OFF ON OFF OFF
Matches ch3 ON ON OFF OFF represent that the Products would necessarily conform to said models or
Matches ch4 OFF OFF ON OFF samples. Any Products found to be defective must be shipped to KEYENCE
Matches ch5 ON OFF ON OFF with all shipping costs paid by Buyer or offered to KEYENCE for inspection
Signal Matches ch6 OFF ON ON OFF
Combination Matches ch7 ON ON ON OFF and examination. Upon examination by KEYENCE, KEYENCE, at its sole
When Match Matches ch8 OFF OFF OFF ON option, will refund the purchase price of, or repair or replace at no charge any
Occurs Matches ch9 ON OFF OFF ON Products found to be defective. This warranty does not apply to any defects
Matches ch10 OFF ON OFF ON
Matches ch11 ON ON OFF ON resulting from any action of Buyer, including but not limited to improper
Matches ch12 OFF OFF ON ON installation, improper interfacing, improper repair, unauthorized modification,
Matches ch13 ON OFF ON ON misapplication and mishandling, such as exposure to excessive current, heat,
Matches ch14 OFF ON ON ON
Matches ch15 ON ON ON ON
coldness, moisture, vibration or outdoors air. Components which wear are not
warranted.
(2) KEYENCE is pleased to offer suggestions on the use of its various Products.
• I f the unit matches on multiple channels, it outputs the channel with the highest They are only suggestions, and it is Buyer's responsibility to ascertain the
matching degree. fitness of the Products for Buyer's intended use. KEYENCE will not be
• I f N.C. is set, ON and OFF output logic is reversed in the above chart. responsible for any damages that may result from the use of the Products.
(3) The Products and any samples ("Products/Samples") supplied to Buyer are not
(Bar mov es along Normal Normal Normal
the display )
* When Output modes is set to [Error], MU-N would turn ON with N.O. and turn
OFF with N.C.
2015
243006E 1125-1 541GB
- 12 -