LabModule1 Micros
LabModule1 Micros
LABORATORY MODULE 1
MICROSCOPY
INTRODUCTION
In 1673, with the aid of a crude microscope consisting of a biconcave lens
enclosed in two metal plates, Antoni van Leeuwenhoek, the man known as the
Father of Microbiology, introduced the world to the existence of microbial forms
of life. Over the years, microscopes have evolved from the simple, single-lens
instrument of Leeuwenhoek, with a magnification of 300X, to the present-day
electron microscopes capable of magnifications greater than 250,000X.
OBJECTIVES
Once this module has been completed, the student should be able to:
Determine the diversity of microscopic instruments.
List the theoretical principles of brightfield microscopy.
Identify the components parts of a compound microscope and determine
the use of each.
Discuss the proper use and care of a compound microscope.
DISCUSSION
Microscopy is the most common method used both for the detection of
microorganisms directly in clinical specimens and for the characterization of
organisms grown in culture. It is defined as the use of a microscope to magnify
(i.e., visually enlarge) objects too small to be visualized with the naked eye so that
their characteristics are readily observable. Because most infectious agents cannot
be detected with the unaided eye, microscopy plays a pivotal role in the
laboratory.
1
Lab Module 1: Microscopy
2
Lab Module 1: Microscopy
electrons onto photographic film, thus making internal cellular structures visible.
Scanning electron microscopes are used for visualizing surface characteristics
rather than intracellular structures. A narrow beam of electrons scans back and
forth, producing a three – dimensional image as the electrons are reflected off the
specimen’s surface.
3
Lab Module 1: Microscopy
Abbé condenser – this component is found directly under the stage and
contains two sets of lenses that collect and concentrate light as it passes
upward from the light source into the lens systems. The condenser is
equipped with an iris diaphragm, a shutter controlled by a lever that is
used to regulate the amount of light entering the lens system.
Body tube – above the stage and attached to the arm of the microscope is
the body tube. This structure houses the lens system that magnifies the
specimen. The upper end of the tube contains the ocular lens or eyepiece
lens lens. The lower portion consists of a movable nosepiece containing
the objective lenses. Rotation of the nosepiece positions objectives above
the stage opening. The body tube may be raised or lowered with the aid of
coarse-adjustment and fine-adjustment knobs that are located above or
below the stage, depending on the type and make of the instrument.
4
Lab Module 1: Microscopy
Based on this formula, the shorter the wavelength, the greater the
resolving power of the lens. Thus, for the same numerical aperture, short
wavelengths of the electromagnetic spectrum are better suited for higher
resolution than are longer wavelengths. However, as with magnification,
resolving power also has limits. Decreasing the wavelength will not
automatically increase the resolving power of a lens, because the visible
portion of the electromagnetic spectrum is very narrow and borders on the
very short wavelengths found in the ultraviolet portion of the spectrum.
This relationship between wavelength and numerical aperture is valid only
for increased resolving power when light rays are parallel. Therefore, the
resolving power is also dependent on another factor, the refractive index.
This is the bending power of light passing through air from the glass slide
to the objective lens. The refractive index of air is lower than that of glass;
as light rays pass from the glass slide into the air, they are bent or refracted
so that they do not pass into the objective lens. This would cause a loss of
light, which would reduce the numerical aperture and diminish the
resolving power of the objective lens. We can compensate for loss of
refracted light by interposing mineral oil, which has the same refractive
index as glass, between the slide and the objective lens. In this way,
decreased light refraction occurs and more light rays enter directly into the
objective lens, producing a vivid image with high resolution.
5
Lab Module 1: Microscopy
6
Lab Module 1: Microscopy
Figure 1.c. Relationship between objective, working distance and diaphragm opening
Once the microscope is placed on the laboratory bench, observe the following
rules:
1. Remove all unnecessary materials (including books, papers, purses, and hats)
from the laboratory bench.
2. Uncoil the microscope’s electric cord and plug it into an electrical outlet.
3. Clean all lens systems; the smallest bit of dust, oil, lint, or eyelash will decrease
the efficiency of the microscope. The ocular, scanning, low power, and high
power lenses may be cleaned by wiping several times with acceptable lens tissue.
Never use a paper towel or cloth on a lens surface. If the oil immersion lens is
gummy or tacky, use a piece of lens paper moistened with xylol to wipe it clean.
Remove the xylol immediately with a tissue moistened with 95% alcohol, and
wipe the lens dry with lens paper. Note: This xylol cleansing procedure should be
performed only by the instructor and only if necessary; consistent use of xylol
may loosen the lens.
7
Lab Module 1: Microscopy
The following routine procedures must be followed to ensure correct and efficient
use of the microscope.
1. Place the microscope slide with the specimen within the stage clips on the fixed
stage. Move the slide to center the specimen over the opening in the stage directly
over the light source.
2. Raise the microscope stage up as far as it will go. Rotate the scanning lens or
low-power lens into position. Lower the body tube with the coarse-adjustment
knob to its lowest position. Note: Never lower the body tube while looking
through the ocular lens; this may allow for an impact with the slide and damage
to the slide or the microscope.
3. While looking through the ocular lens, use the fine-adjustment knob, rotating it
back and forth slightly, to bring the specimen into sharp focus.
4. Adjust the substage condenser to achieve optimal focus.
5. Routinely adjust the light source by means of the light-source transformer
setting, and/or the iris diaphragm, for optimum illumination for each new slide
and for each change in magnification.
6. Most microscopes are parfocal, which means that when one lens is in focus,
other lenses will also have the same focal length and can be rotated into position
without further major adjustment. In practice, however, usually a half-turn of the
fine-adjustment knob in one direction or the other is necessary for sharp focus.
7. Once you have brought the specimen into sharp focus with a low-powered lens,
prepare to visualize the specimen under oil immersion. Place a drop of oil on the
slide directly over the viewing area. Rotate the nosepiece until the oil-immersion
objective locks into position. Note: Care should be taken not to allow the high-
power objective to touch the drop of oil. Observe the slide from the side as the
objective is rotated slowly into position. This will ensure that the objective will be
properly immersed in the oil. Readjust the fine-adjustment knob to bring the
image into sharp focus.
8. During microscopic examination of microbial organisms, it is always necessary
to observe several areas of the preparation. To do so, scan the slide without the
application of additional immersion oil. Note: This will require continuous, very
fine adjustments by the slow, back-and-forth rotation of the fine adjustment knob
only.
8
Lab Module 1: Microscopy
REFERENCES
Cappucino, J. & Welsh, C. 2020. Microbiology: a laboratory manual, 12th ed.,
Pearson Education, Inc., New York
Granato, P., Morton, V. & Morello, J. 2019. Laboratory manual and workbook in
microbiology: applications to patient care, 12th ed., McGraw – Hill
Education, New York
Tille, P. 2014. Bailey & Scott’s diagnostic microbiology, 13th ed., Elsevier, Inc.,
Missouri