S20 Sensor Chamber TP20 Temperature Sensor CP10 PT12-TI-9-403-US

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

S20 Sensor Chamber, TP20 Temperature Sensor,

CP10 Sensor and PT2 Plug-tail Cast Iron Valves


• Purpose designed sensor chamber for in-line mounting
• Simple removal of sensors for cleaning
• Allows continuous temperature compensated measurement

Description S20 sensor


The Spirax Sarco S20 sensor chamber is fitted with conductivity and chamber
temperature sensors used to determine the conductivity of liquids.
The use of a temperature sensor enables an accurate measurement
to be made when the temperature varies, as in the case of conden-
sate return monitoring systems.
The hexagonal in-line sensor chamber is screwed 1" NPT. CP10
conductivity
Adjacent radial female screwed connections are provided for:- sensor
CP10 conductivity sensor (3/8" BSP).
TP20 temperature sensor (1/4" BSP).
An additional 1/4 “ BSP (1/4" NPT) connection is provided on the
other side of the chamber for taking a sample if required. TP20
Spirax Sarco can supply a sample cooler for cooling hot samples, or temperature
sensor
a blanking plug if the connection is not required.

Limiting Conditions
S20 sensor chamber and TP20 Temperature sensor Installation
Maximum pressure 160 psig (11 barg)
Maximum media temperature 302°F (150°C) Caution: Do not install the sensor outdoors without additional
weather protection.
CP10 conductivity sensor Fit the sensor chamber in a vertical or horizontal pipeline with suit-
Maximum pressure 464 psig (32 barg) able isolation valves to allow inspection/cleaning of the sensors.
Maximum media temperature 462°F (239°C) Reducers may be fitted if required. Flow can be in either direction.
Torque rating 44 lb./f (60 Nm) The sensors themselves must be horizontal.
The TP20 has a taper thread, and may be installed using PTFE seal-
PT2 Plug-tail ing tape if required.
Cable temperature range -67 to 248°F (-55 to 120°C)
The CP10 sensor is provided with an S-type stainless steel gasket for
Supplied cable length 50" (1.25m)
sealing and does not require sealing tape.
Maximum voltage 20Vdc, 14Vac
Maximum current 10mA Fit the sensors to the chamber and tighten.
Fit PT2 plug tail (or other Spirax Sarco plug tail) to the CP10 sensor
and tighten.

Red wire to
CP10 PT2 sensor pin
Controls

Sensor Plug-tail
Boiler

To blow-
down
controller
Insulation
Sensor pin Screen
Boiler Blowdown

3/8" Blue wire ground- Junction box


24mm Nut 1 ed via PT2 and
BSP (For connection to conduit)
AF CP10 body.
& CCD

25mm
AF
The supply to the Sensor and Plug-tail must be provided by a low voltage limited power source.
Spirax Sarco Blowdown controllers provide this type of supply.
Flexible metal conduit (M16) can be connected to the PT12 by removing the nut (1)

Local regulation may restrict the use of this product below the conditions quoted. Limiting conditions refer to standard connections only.
In the interests of development and improvement of the product, we reserve the right to change the specification. TI-9-403-US 08.11
294

Valves_Instruments_Flowmeters 2016.indd 294 5/11/16 3:18 PM


S20 Sensor Chamber, TP20 Temperature Sensor,
CP10 Sensor and PT2 Plug-tail
Maintenance How to Specify
The equipment requires no specific maintenance other than periodic In line sensor chamber with conductivity and temperature sensors.
inspection and cleaning. CP10 Conductivity Sensor and PT2 Plug-tail.

Materials
S20 sensor chamber How to Order
Stainless steel ASTM 8582303 BS 970 303 S31 Spirax Sarco’s S20 sensor chamber complete with CP10 sensor,
PT2 plug tail, 1/4" BSP blanking plug, and TP 20 temperature
PT2 Plug-tail sensor.
Body Brass
Nuts Brass (2off)
Insulator Polyphenylene Sulphide Resin (glass filled)

CP10 conductivity sensor


Body Stainless steel BS 970 303 S31
Insulator PEEK plastic
Pin Stainless steel BS 970 303 S31

TP20 temperature sensor


Body and probe Stainless steel 316 ASTM A269 Gr. 316
Cable insulation PFA

Dimensions (approximately) in inches

4"
S-type stainless
2-1/4" steel gasket
CP10 sensor
(shown with PT2
Standard plug tail fitted).
length 50"
4-3/4"

1/4" BSP
50" sample
connection

TP20
temperature
sensor

1" NPT Controls


Boiler
Boiler Blowdown
& CCD

Spirax Sarco, Inc., 1150 Northpoint Blvd., Blythewood, SC 29016 • Phone: (803) 714-2000 • Fax: (803) 714-2222 TI-9-403-US 08.11
© Spirax Sarco, Inc. 2001 295

Valves_Instruments_Flowmeters 2016.indd 295 5/11/16 3:18 PM