ACCURACY–PRESSURE MEASUREMENT
0 to 110% of Range: ±(0.05% of Reading + Floor Term)
69.8
(2.75)
DIMENSIONS ARE IN MM (IN)
Includes all effects of linearity, hysteresis, repeatability,
temperature, and stability for one year.
Floor Term: 0.005% of FS or display resolution,
whichever is greater All models indicate vacuum, but vacuum specification applies
to 16 psi, 36 psi, and 300 psi models only.
Vacuum: ±(0.25% of Reading
+ Vacuum Floor Term) Not recommended for continuous use below -14.5 psi. Refer to
30 Series Pressure Calibrator MICRO-USB
XP2i-DP data sheet for gauges that are intended for continuous CONNECTION
Vacuum Floor Term: 0.004 psi for 16 and 36 psi high vacuum use. 2 LINE
0.01 psi for 300 psi
137.2 8 CHARACTER
(5.40) LCD DISPLAY
Connection: 4 mm jacks, 19 mm (0.75") spacing Jacks are compatible with safety sheathed banana plugs. SENSOR
B DIAPHRAGM
LOCATIONS
DISPLAY
DO NOT EXCEED MAXIMUMS
POWER
The 30 Series is Intrinsically Safe only if powered by one of the following battery types.
CERTIFICATIONS
II 2G Ex ia IIC T4
ATEX/IECEx: Approved Battery Type Ta= Marking FTZU 06 ATEX 0010X
ATEX
Duracell Alkaline 9V, MN 1604
Energizer Alkaline 9V, 522 0 to 50° C II 2G Ex ia IIC T4 Gb
Ex ia IIC T4 Gb IECEx FTZU 10.0018X
Energizer Lithium 9V, L522
Varta High Energy Alkaline 9V, 4922 0 to 45° C
Exia Intrinsically Safe and Non-incendive for Hazardous
Locations: Class I, Division 1, Groups A, B, C and D, Temperature
CSA: Approved Battery Type Marking
Code T4 or T3C. For hazardous location product warnings, refer
Duracell Alkaline 9V, MN 1604 to the operation manual.
Energizer Alkaline 9V, 522 Class I, Division 1, A, B, C, D T4
Energizer Lithium 9V, L522 30 Series complies with the Electromagnetic Compatibility and
Varta High Energy Alkaline 9V, 4922 Class I, Division 1, A, B, C, D T3C the Pressure Equipment Directives.
1 x 9V: 90 hours, typical Uses one alkaline 9V battery. Use of USB interface reduces
operating time. 30 Series complies with the Australian Radiocommunications
For hazardous location product warnings, refer to the
(Electromagnetic Compatibility) Standard 2008
operation manual.
ENCLOSURE
Model 31≤300 psi: 342 g (12.0 oz) Weight includes batteries.
Anodized aluminum penetrated All welded with a permanently filled diaphragm seal. ACCESSORIES
with PTFE (polytetrafluoroethylene) Magnetic Hanging Strap P/N 5203
Diaphragm Seal Fluid: Silicone Oil Test Lead Kit P/N 1351
Connection: Bottom, 1⁄8 " female NPT USB A to Micro USB Cable P/N 4431
Waterproof Carrying Case P/N 2888
STORAGE TEMPERATURE
Temperature Range: -20 to 70° C (-4 to 158° F) Batteries should be removed if stored for more than
one month. COMPLEMENTARY PRODUCTS
TSPECIAL
FEATURES
he following requires the use of ConfigM30 software
Crystal Engineering offers a wide range of products that
work with the 30 Series:
User Defined Unit: Allows the user to define and display any pressure units not included,
• Fittings that connect without tools, safely and without leaks
or to use the gauge to display force, level, or other pressure related parameters. • Lightweight, super flexible high pressure hoses
Reduce or increase displayed resolution.
• Pneumatic hand pumps
Remove: Unwanted pressure units from Units button. • Hydraulic hand pumps
Password Protect: Changes to configuration or userspan calibration factor(s). • Portable pressure comparators
Save: Configuration to a file. • Software, for the quickest way to calibrate pressure transmitters
Copy: Configuration to other calibrators. and gauges
ORDERING INFORMATION
Intrinsic 1st Pressure 2nd Pressure Pressure Pressure AMETEK offers a variety of solutions for pressure genera-
Model — / Pump System* Carrying Case~
Safety Range P/N Range P/N Unit Manifold
tion and measurement. Our line of products for pressure
IS PSI
generation includes everything from small pneumatic
Single Sensor. . . 31 Standard. . . . . . . . (omit) No Pump. . . . (omit)
hand pumps to a precision, hydraulic pressure comparator
Dual Sensor. . . . 33 Stainless Steel. . . -SPSS System A. . . . -AXX Aluminum . . . . (omit)
capable of generating up to 15 000 psi / 1000 bar / 100 MPa.
System A. . . . -AHX Waterproof. . . . . . . -W
System B. . . . . -BXX All of our pumps may be ordered as part of a Pump System,
SAMPLE PART NUMBERS System B. . . . . -BHX complete with a 30 Series and delivered in a sturdy carry-
~ The Waterproof Case is
IS31-36PSI. . . . . . . . . . . . . . . . . . . . . . . . . . Calibrator with one sensor (36 psi). System C. . . . . -CXX an option for Systems A, B,
ing case with custom insert.
IS33-16/1500PSI-SPSS . . . . . . . . . . . . . Calibrator with two sensors (16 psi and 1500 psi) and C only.
System C. . . . . -CHX
The Waterproof Case is
with two stainless steel manifolds. System D. . . . -DOX * Refer to the following page for a more detailed description
the only option for Systems
IS33-36/3000PSI-SPSS-HOX-W. . . . . Calibrator with two sensors (36 psi and 3000 psi) System D. . . . -DWX G and H. of each pump system.
and two stainless steel manifolds with a System H pump system System E. . . . . -EOX
and a waterproof carrying case. System F. . . . . -FOV
System F. . . . . -FWV
XXOrdering a Pump System Only
System G. . . . -GOX
Any pump system, carrying case, and connection fittings for a 30 Series calibrator may be ordered
System G. . . . -GWX
separately from the calibrator. Enter IS30-NONE followed by the Pump System part number and the
System H . . . . -HOX
Carrying Case option code.
P014-CPF
System E EOX 0 to 10 000 psi / 700 bar (Oil)