This document provides an operations manual for the IAP COMP digital airgauge. It includes sections on keyboard description, specifications, measurement modes, program menu settings, factory setting mode, calibration mode, memory management, measurement with LVDT, wireless and serial interfaces, connector details, warranty, and test certificate. The manual provides detailed instructions on operating and setting up the various functions of the digital airgauge.

Uploaded by

byjurp
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
325 views42 pages

Iap Comp Digital Operational Manual v1.3

This document provides an operations manual for the IAP COMP digital airgauge. It includes sections on keyboard description, specifications, measurement modes, program menu settings, factory setting mode, calibration mode, memory management, measurement with LVDT, wireless and serial interfaces, connector details, warranty, and test certificate. The manual provides detailed instructions on operating and setting up the various functions of the digital airgauge.

Uploaded by

byjurp
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 42

OPERATION MANUAL

IGITAL IRGAUGE

MODEL:-

IAP COMP digital


VERSION: 1.3
INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

CONTENTS
A. INTRODUCTION 1
A-1. ADVANTAGES 1
A-2. APPLICATIONS 1
A-3. FEATURES 1

B. KEYBOARD DESCRIPTION 2

C. SPECIFICATION 3

D. FRONT PANNEL 4

E. FRONT AND BACK SIDE VIEW 4

F. POWER ON SETTINGS 5

G. PROGRAM MENU SETTINGS 6

G-1. New Value Entry Process 8

FANUC INTERFACE 12
G- 15. Fanuc-1 13
15.1. Reset 13
15.2. Max Average Count 14
15.3. Max Consecutive Count 15
15.4. Wear Limit 16
G- 16. Fanuc-2 16
G- 17. Air Saving Time 17
G- 18. IP Address 17
G- 19. Low Air Mode 19
G- 20. Control Limit 20
G- 21. Fanuc Delay 20

H. FACTORY SETTING MODE 21

I. CALLIBRATION MODE 22

J. MEMORY MANAGEMENT MODE 23

K. MEASUREMENT MODE 24

K-1. STATIC MODE 24


K-2. DYNAMIC MODE 26
K-3. AUTOSENSE MODE 28
INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
K-4. PROBE CENTER ADJUSTMENT MODE 28
K-5. CENTER OFFSET MODE 28

L. MEASUREMENT WITH LVDT 29

M. WIFI INTERFACE 30

N. SERIAL INTERFACE MODE 31

O. CONNECTOR DETAILS 36

P. WARRANTY 37

Q. TEST CERTIFICATE 38

List of Tables

B-1. Key function Description. 2

C-1. Airgauge specification. 3

G-1. program menu option settings. 8

G-2. Parameter settings menu. 9

M-1. Memory management function menu. 23

N-1. Static mode Key function Description. 24

N-2. Dynamic mode Key function Description. 26


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
A] INTRODUCTION :

As a result of extensive Research and Development at INNOVATIVE


AUTOMATION PRODUCTS, the ‘IAP COMP digital’ series of DRO is the most
comprehensive, reliable, economical & user friendly Airguage available in the market today.
This Airgauge is the latest model of Airgauge used for measurement of jobs, with pneumatic
gauges. Following are advantages & specifications.

A-1] ADVANTAGES :
1. Greater measurement accuracy.
2. Durability & reliability.
3. Easy assembly & functioning.
4. Non contact measurement.
5. Less operator fatigue.
6. Most economical Airgauge in the market.
7. High accuracy 0.1µm.
8. Automatic cleaning of measured surfaces.
9. Measurement in harsh workshop conditions.
10. RS 232 interface software for data exchange.
11. Airgauge type / LVDT type / Encoder type.
12. 16*2 LCD display and tricolor 7 segment Displays.
13. Compact in size.
14. Simple operation with menu Guidance.

A-2] APPLICATIONS :
The IAPCOMP digital can be used for the measurement of different jobs (such as ID, OD,
ovality, job center measurement) using air pressure.

A-3] FEATURES :

Piezo technology based sensor.


No moving diaphragm like parts.
Piezo channel operates in full differential mode of operation. This reduces errors caused
due to temperature electrical noise etc.
Compact in size.
16*2 LCD display for showing menu function.
Simple operations with menu program.
Up to 25 Measuring programs stored internally.
Tricolor 7 segments Display for Accept , Reject , Rework tolerance setting.
Auto measurement on plug insertion.
Foot-switch I/P for measurement .
Optional relay O/P’s for accept ,reject, rework.
RS232 interface for SPC analysis [SPCV3 software].
Support various expression mode.max ,min ,Avg, dynamic and static.
Optional wireless interface.
7500 job storage internally.
Display can set to Inch or mm.
Power supply 230V.
Resolution 0.001,0.0005, 0.0001,0.005.
Single Master , dual Master.
ID , OD selection.
Absolute or Increment.
Scale range mm - 4um, 40um,400um.4000um.
Inch - 0.2mil,2mil,20mil,200mil.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 1


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
B] KEYBOARD DESCRIPTION :

Table no: B-1: Key function Description.


STATIC MODE DYNAMIC MODE
KEYS NAME
DESCRIPTION DESCRIPTION
STORE VALUE & SEND
FUNCTION - 1 SEND OVALITY VALUE OF CH1
CHANNEL-1 VALUE

STORE VALUE & SEND


FUNCTION -F2 SEND OVALITY VALUE OF CH2
CHANNEL-2 VALUE

RESET OVALITY CORRECTION


FUNCTION -F3 NO ACTION
FACTOR TO 1.0

CHARACTERIS PRESET MODE OF RESTART OVALITY TO ZERO OF CH1


TIC - 1 CHANNEL 1 OF CHANNEL 1

CHARACTERIS PRESET MODE OF RESTART OVALITY TO ZERO OF CH2


TIC - 2 CHANNEL 2 OF CHANNEL 2

CALIBRATION BY MASTER 1
CALIBRATION CALIBRATION BY OVALITY MASTER
& MASTER 2 I.E. STATIC
THIS KEY ONLY EFFECT
WHEN MEASUREMENT TYPE THIS KEY ONLY EFFECT WHEN
IS SET TO “AB-IN” IN MEASUREMENT TYPE IS SET TO
PARAMETER SETTING. IF “AB-IN” IN PARAMETER SETTING.IF
ABSOLUTE / LED IS ON, IT MEANS “ABS” LED IS ON, IT MEANS “ABS”
INCREMENT MEASUREMENT TYPE IS MEASUREMENT TYPE IS
SELECTED.ON OTHER HAND SELECTED.ON OTHER HAND IF LED
IF LED IS OFF IT MEANS IS OFF IT MEANS “INC”
“INC” MEASUREMENT TYPE MEASUREMENT TYPE IS SELECTED.
IS SELECTED.
THIS KEY ONLY EFFECT
WHEN MEASUREMENT TYPE THIS KEY ONLY EFFECT WHEN
IS SET TO “IN-MM” IN MEASUREMENT TYPE IS SET TO
PARAMETER SETTING. IF “IN-MM” IN PARAMETER SETTING.IF
LED IS ON IT MEANS “INCH” LED IS ON IT MEANS “INCH”
INCH/MM
MEASUREMENT UNIT IS MEASUREMENT UNIT IS
SELECTED.ON OTHER HAND SELECTED.ON OTHER HAND IF LED
IF LED IS OFF IT MEANS IS OFF IT MEANS “MM”
“MM” MEASUREMENT UNIT MEASUREMENT UNIT IS SELECTED.
IS SELECTED.

ENTER NO ACTION NO ACTION

ESCAPE ERASE LAST ENTRY ERASE LAST ENTRY

UP MEMORY MANAGEMENT
MEMORY MANAGEMENT

DOWN NO ACTION NO ACTION

LEFT PROBE CENTER PROBE CENTER

PROGRAM PROGRAM MODE PROGRAM MODE

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 2


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
SP1 Show Date & Time Show Date & Time

SP2 Display & Keyboard Testing Display & Keyboard Testing

STORE VALUE & SEND CH-1 RESTART OVALITY TO ZERO OF


Auto 1 VALUE CH-1 VALUE
STORE VALUE & SEND CH-2 RESTART OVALITY TO ZERO OF CH-2
Auto 2 VALUE VALUE
IF CH1 OVALITY ≥ OVALITY MIN.
Auto NO ACTION TOLERANCE THEN ENABLE RELAY
1-1 OF CH1, OTHERWISE NO ACTION
IF CH2 OVALITY ≥ OVALITY MIN.
Auto NO ACTION TOLERANCE THEN ENABLE RELAY
1-2 OF CH2, OTHERWISE NO ACTION

C] SPECIFICATION :

Input Voltage 80 TO 290VAC 50HZ


Power Consumption 5VA
Humidity 85 % at 45ºC Non-condensing.
Temperature 0ºC to 50ºC
Vibrations 1g,10Hz TO 50 Hz
Enclosure 250 x 280 x 110 mm
Weight 7kg
Resolution 0.1, 0.5, 1, 5 Micron (Micrometer)
Display 16*2 LCD display AND Tricolor 7 segments.
Table no:C-1: Airgauge specification.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 3


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

D] FRONT PANNEL :

E] FRONT AND BACK SIDE VIEW :

FRONT VIEW BACK VIEW

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 4


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

F] POWER ON SETTINGS :

At power ON Airguage LCD show “Memory Loading…” message

After that it will lead to Calibration screen. In Calibration screen it will ask you to select

channel 1 (press key “CH1” ) or channel 2 (press key “CH2” ) otherwise skip

this Calibration mode (press key “ESC” ).

After successful Calibration OR Exiting Calibration mode Airgauge lead to Home screen which
run default in Airguage.
Home screen shows characteristics name (like C1 and C2 ) , It’s measuring UNIT (like mm or
inch) and the channel selected for respective characteristics(like A, B, A+B, A-B ).

There are two rows of 8 7-segment in each rows . as shown bellow for Channel 1 and
Channel 2.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 5


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

G] PROGRAM MENU SETTINGS :

This mode is used to configure different parameters for each job measurement (25 programs
each for one job) to user’s need and requirement. This is a special mode and is generally
intended to use by person adverse to AIRGAUGE and its manual.

Airgauge will enter into this mode after pressing “PROG” key when Airgauge
showing Home screen. After it will show program menu screen. Here you can access different

program menus available under program menu. Use “UP” key and “DOWN”
key from accessing different options.
Following are options available under program menu:
Table no:G-1:program menu option settings.
Program
ACTION PERFORMED
menu
Total 25 programs can be set for
1]Prog no both channel with different
Parameters for different jobs.
It will give us access to different
parameter settings for
2]Char-1
characteristic 1 of program no set in
above setting(Prog no).
It will give us access to different
parameter settings for
3]Char-2
characteristic 2 of program no set in
above setting(Prog no).
It will save status number for every
4]Status measured job. assign status for
each status number.

5]WiFi Test WIFI testing functionality.

Probe center adjustment mode for


6]Center-1
channel 1

Probe center adjustment mode for


7]Center-2
channel 2

8]Serial Serial number settings for channel


No-1 1

9]Serial Serial number setting for channel 2


No-2

10]Relay_ON_
Relay ON timing
Time

11]Operator
Operator number setting
Number

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 6


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

12] Machine
Machine number setting
Number

13] Custom
Custom number setting
Number

14] Tool
Tool number setting
Number

Fanuc settings for channel 1


15] Fanuc 1

Fanuc settings for channel 2


16] Fanuc 2

17] Air Saving


Air Saving time setting
Time

18] IP
IP Address Setting
Address

19] Low Air


Low Air Setting
Mode

20] Control
Control Limit Setting
Limit

21] Fanuc
Fanuc Delay Setting
Delay

In “char-1” Program menu Airgauge ask for password entry So Here Sequentially press key as
mentioned ahead “F1”, “F2”, “F3”, “CAL” AND “ENT”.

Enter password correctly so that it will lead to parameter setting menu.


Before going for parameter setting understand how to change different options within every
parameter menu and how to do new value entry.

For changing any parameter in parameter menu press “ENT” key.

Then for accessing different sub option within the parameter menu use “UP” OR

“DOWN” key and for confirming it press “ENT key.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 7


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

G-1] New Value Entry Process :


Left most digits starts Blinking. For changing Digit position press “LEFT” key and for

changing value on segment use “UP” OR “DOWN” key for incrementing or

decrementing value on segment.When you are at 7th digit position and press “LEFT”
key for moving 8th position it is blank.

This is Digit Sign position. Press “UP” or “DOWN” key for minus sign (positive
sign is not shown on display) When there is no any sign present then it will indicate positive

value. Press “ENT” key to store that value.

*Note : Total 25 programs you can set for 25 no. of different jobs.
Changes to these menus are applicable for only that job you have
selected. For every new job you have to set new parameters in a new
program.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 8


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

Table no: G-2: Parameter settings menu.


Parameter
ACTION PERFORMED
menu
Job diameter setting.for editing a value
DIAMETER assigned to it press “ENT” key.After entering
SETTING value press “ENT” key for confirming that
value.otherwise press “ESC” key.
Higher tolerance setting.for editing a value
HIGHER
assigned to it press “ENT” key.After entering
TOLERANCE
value press “ENT” key for confirming that
LIMIT
value.otherwise press “ESC” key.
Lower tolerance setting.for editing a value
LOWER
assigned to it press “ENT” key.After entering
TOLERANCE
value press “ENT” key for confirming that
LIMIT
value.otherwise press “ESC” key.
Higher master setting.for editing a value
HIGHER
assigned to it press “ENT” key.After entering
MASTER
value press “ENT” key for confirming that
SETTING
value.otherwise press “ESC” key.
Lower master setting.for editing a value
LOWER
assigned to it press “ENT” key.After entering
MASTER
value press “ENT” key for confirming that
SETTING
value.otherwise press “ESC” key.
ID OR OD measurement setting.
ID OR OD
For changing current setting press “ENT” key
TYPE
and for toggling settings available on display
MEASUREME
pressing “LEFT” key. After that press “ENT” or
NT
“ESC” key.
ABS OR INC measurement setting.For
changing current setting press “ENT” key and
MEASUREME
for toggling settings available on display
NT TYPE
pressing “LEFT” key. After that press “ENT” or
(ABS/INC)
“ESC” key. For changing these settings by
using “ABS/INC” key select “AB-IN” option.
0.001, 0.005, 0.0001 OR 0.0005 resolution
setting.For changing current setting press
DISPLAY
“ENT” key and for toggling settings available
RESOLUTION
on display pressing “LEFT” key. After that
press “ENT” or “ESC” key.
INCH OR MM measurement setting.For
changing current setting press “ENT” key and
UNIT OF
for toggling settings available on display
MEASUREME
pressing “LEFT” key. After that press “ENT” or
NT
“ESC” key. For changing these settings by
using “INCH/MM” key select “IN-MM” option.
A, B,A+B, A-B channel setting.
For changing current setting press “ENT” key
CHANNEL
and for toggling settings available on display
SELECTION
pressing “LEFT” key. After that press “ENT” or
“ESC” key.
In mm - 4um, 40um, 400um, 4000um.
In mil - 0.2mil,2mil,20mil,200mil range
DISPLAY
setting.For changing current setting press
RANGE
“ENT” key and for toggling settings available
SELECTION
on display pressing “LEFT” key. After that
press “ENT” or “ESC” key.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 9


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
SET Here we can set production target.For
PRODUCTION changing current setting press “ENT” key and
TARGET edit value using value entry method.
Here it will show you how many jobs get
SET ACCEPT
accepted using present program number and
VALUE
characteristic.
Here it will show you how many jobs get reject
SET REJECT
using present program number and
VALUE
characteristic.
Here it will show you how many jobs get
SET REWORK
rework using present program number and
VALUE
characteristic.
Auto-sense mode ON OR OFF setting. For
changing current setting press “ENT” key and
JOB AUTO
for toggling settings available on display
DETECTION
pressing “LEFT” key. After that press “ENT” or
“ESC” key.
Here we can set center master value. for
editing a value assigned to it press “ENT”
MASTER
key.After entering value press “ENT” key for
CENTER
confirming that value.otherwise press “ESC”
key.
Calibration mode ON OR OFF setting. For
ON/OFF changing current setting press “ENT” key and
STATUS OF for toggling settings available on display
‘CAL’ MODE pressing “LEFT” key. After that press “ENT” or
“ESC” key.
SP-Fun mode ON OR OFF setting.If status is
ON then, at power ON “SP” mode will be ON
after calibration mode. If status is OFF then,
ON/OFF at power ON Airgauge will directly enter in
STATUS OF main measurement mode, after “CAL” mode.
‘SP’ MODE For changing current setting press “ENT” key
and for toggling settings available on display
pressing “LEFT” key. After that press “ENT” or
“ESC” key.
Here we can set Ovality tolerance Higher limit
OVALITY value. for editing a value assigned to it press
TOLERANCE “ENT” key.After entering value press “ENT”
HIGHER LIMIT key for confirming that value.otherwise press
“ESC” key.
Here we can set Auto-sense range value. for
editing a value assigned to it press “ENT”
AUTOSENSE
key.After entering value press “ENT” key for
RANGE
confirming that value.otherwise press “ESC”
key.
Here we can set Auto-sense time value. e. g.
Auto-sense time = 6.00ms (you can set
auto-sense time from 0ms to 9999ms) .
AUTOSENSE
for editing a value assigned to it press “ENT”
TIME
key.After entering value press “ENT” key for
confirming that value.otherwise press “ESC”
key.
Here we can set Ovality checking time value.
When job is inserted to check Ovality,
OVALITY
value of job will be displayed in the range
CHECKING
then ovality value will reset. Here start the
TIME
timer.When Ovality checking time is set,
then job should be measured till checking

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 10


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
time completion. If job measurement
is stopped before time & if Ovality is below
minimum ovality, then error will be
displayed & color changed but relay will be
OFF. If job measurement completed
after time completion & if Ovality is above
minimum ovality, then relay will be ON.
e.g. Ovality checking time = 2000ms
(You can set Ovality checking time from
0ms to 9999ms)
for editing a value assigned to it press “ENT”
key.After entering value press “ENT” key for
confirming that value.otherwise press “ESC”
key.
Here we can set Ovality lower tolerance
value.
For correct ovality checking, minimum
ovality tolerance value should be entered.
Job measured in given time & if ovality
OVALITY
measured is below minimum ovality, then
TOLERANCE
error will be displayed & all relays will be
LOWER LIMIT
OFF.
for editing a value assigned to it press “ENT”
key.After entering value press “ENT” key for
confirming that value.otherwise press “ESC”
key.
Here we can set Ovality tolerance Higher
master value.
This function can be used only in
differential mode (i.e. when A+B or A-B is
used) for measurement of concentricity or
squareness.
Suppose you are measuring concentricity,
OVALITY if higher master value is 9µ & concentricity
TOLERANCE measured value is 8µ then press CAL to
HIGHER effect higher master value in that function.
MASTER At the time of job measurement enter into
Ovality mode, measure & rotate the job
then ovality value will be automatically
updated.
for editing a value assigned to it press “ENT”
key.After entering value press “ENT” key for
confirming that value.otherwise press “ESC”
key.
Here we can set Ovality Calibration on status
value.
Ovality Calibration can be done depending
on status is Yes or No.
In differential mode this calibration is
possible. Measure ovality, if value is
OVALITY wrong & if status is yes then press CAL
CALIBRATION key, ovality will be automatically
STATUS calculated & calibrated value will be
displayed.
If you want to see original measured
ovality values press “F3” key.
After Ovality calibration, again go to
Function No. 26 & make status NO to save
the calibration.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 11


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
For changing current setting press “ENT” key
and for toggling settings available on display
pressing “LEFT” key. After that press “ENT” or
“ESC” key.
NORMAL OR SIMPLE measurement
setting.
IN NORMAL mode two masters getting used
NORMAL for calibration.while in SIMPLE mode only
SIMPLE single master used for calibration.
SELECTION For changing current setting press “ENT” key
and for toggling settings available on display
pressing “LEFT” key. After that press “ENT” or
“ESC” key.
Here we can set measurement type from
STATIC, TIR, MAX, MIN, AVG OR TIR_SP.
MEASUREME
For changing current setting press “ENT” key
NT TYPE
and for toggling settings available on display
SELECTION
pressing “LEFT” key. After that press “ENT” or
“ESC” key

Fanuc interface:
Fanuc Interface is used for communication of Airgauge with Fanuc
controller for offset adjustment. Port Selects Fanuc to send data to the Serial
Port.
When asked the main offset value is send to the Fanuc controller as
per the protocol, Baud rate- 9600, 8 bit, Stop bit=1, parity=none.
Offset value is calculated by taking average of 20 values
(settable Max 20).
If the average of 20 values is greater than control limits UCL or LCL
for consecutive 7 times (settable Max20) then the new offset is calculated &
added to the main offset value. This main offset is then saved in memory &
send to the Fanuc controller when asked to send. If offset value calculated is
more than wear limit then change tool alarm is showed & this offset value is
not added in the main offset value. Average count & consecutive count is
cleared.
If values for averaging taken is less than 20, then this average values
are not taken for checking with control limit.
Reset function is used to clear main offset value, Average count,
consecutive count, Also main offset value=0 is saved in memory.
After calibration or preset or at power ON, Reset, or change in
program, Average count=0, Consecutive Count=0, Average=0.
Array is moving average array Ie. Last In First Out.
Wear limit:- If the Offset value is greater than the wear limit then show
the “Change Tool “ alarm & don’t take the offset value in the calculation.
For setting of Max Average Count, Max Consecutive count, Wear
limit & Reset press key & then key. Then select respective function
& by Key. Press key to modify. Press key to quit.
For selection of Fanuc Function for serial port press key. Then
Press key till ‘port’ is not coming. Then press key. Now press
Key to select the ‘Fanuc’ Function.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 12


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
15. Fanuc 1:

Fanuc Parameters are:-


1. Reset.
2. Max Average Count.
3. Max Consecutive Count.
4. Wear Limit.

15.1. Reset:- Fanuc Offset resets to 0.0. Reset function is used to clear main
offset value, Average count, consecutive count, Also main offset value=0 is
saved in memory.
Press key to toggle the mode

Press key to enter in reset then it will ask 'YES' to confirm. Press

key to reset or key to exit.

If Fanuc is Reset then it will show Done as below window.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 13


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

15.2. Average Number:-

Enter Average Number, Maximum average count


is 20.Press key to toggle the mode.

Press key to enter in the mode & will display the count,

When you enter in the mode last digit starts blinking, & you can set the count
but it will accept the count less than 20.

To change the count, Press to change the digit place & press to
change the digit value to set the count.

Press key to save the changes & to exit.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 14


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

15.3. Consecutive Number:-

Enter Consecutive Number, Maximum Consecutive count


is 20.Press key to toggle the mode.

Press key to enter in the mode, & will display the count,

When you enter in the mode last digit starts blinking, & you can set the count
but it will accept the count less than 20.

To change the count, Press to change the digit place & press
to change the digit value to set the count.

Press key to save the changes & to exit.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 15


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
15.4. Max Wear Value:-

Enter Wear Limit, Press key to toggle the mode.


Press to enter in the mode,

Press key will display the count,& will display the count,

When you enter in the mode last digit starts blinking,

To set the Wear Limit Value, Press to change the digit place & press
to change the digit value to set the count.

Press key to save the changes & to exit.

16. Fanuc 2:

Fanuc 2 has same settings as per Fanuc 1 for


channel 2

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 16


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
17. Air Saving Time:
In Air Save Time, If no Action of measurement is detected for Air
saving time, then Air will be Automatically OFF
Enter Air Save Time, Press key to toggle the mode.
Press to enter in the mode,

As you Enter in the Mode Last digit starts blinking.

To set the Air save time, Press to change the digit place & press to
change the digit value to set the count.

Press key to save the changes & to exit.

18. IP Address:

Setting Commands- WiFi / Ethernet Command Mode

Press key to enter in IP Address mode, then Command Setting starts,

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 17


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

If no any device is connected then you will display Command error, “Restart
Device” window as shown above.
Press key to quit. & will start Rebooting.

IP Address setting:-
1. Local IP
2. Subnet mask.
3. Gateway
4. Remote Host IP

1. Local IP:-

When you enter in Local IP mode you will get Initial Local IP
“000.000.000.000” will show old IP.

To set the Local IP, Press to change the digit place & press to change
the digit value to set the IP.
Press key to save the changes & to exit.

2. Subnet Mask:-
To set the Subnet Mask, Press to change the digit place & press
to change the digit value to set the IP.
Press key to save the changes & to exit.

3. Gateway:-
To set the Gateway, Press to change the digit place & press
to change the digit value to set the IP.
Press key to save the changes & to exit.

4. Remote Host IP:-


To set the Remote Host IP, Press to change the digit place &
press to change the digit value to set the IP.

Press key to save the changes & to exit.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 18


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

19. Low Air Mode:


Press key and go to “19: Low Air Mode” menu present in
Program setting menu

Low Air Modes:-

1. Low Air Mode.


2. Low Air Cut off.

1. Low Air Mode:-


When you enter in Low Air Mode, you will get option for Low Air
Mode ON/ OFF.

Press key to toggle ON/OFF & Press key to select.

2. Low Air Cut off:-

In Low Air Cut off, you will get "4.0","3.8","3.6","3.4","3.2" Low Air
cut off selections. Press key to toggle & Press key to select.

Press Key to exit.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 19


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

20. Control Limit:-


Press key and go to parameter no. “20: Control Limit” menu

present in Program setting menu

If control limits are “ON” then display colors will be as follows,


1. Red: Job value will be between above high tolerance or below low
tolerance.
2. Orange: Job value will be between high tolerance & high control Limit
(HCL). OR Job value will be between low tolerance or low Control Limit (LCL).
3. Green: If job value is between HCL & LCL, HCL & LCL,HL & LL is settable.

21. Fanuc Delay:


Press key and go to parameter no. “21: Fanuc Delay” menu

present in Program setting menu

When Fanuc sends “LS” Command to Airguage then after this delay time data
will be send to Fanuc.
1 Count = 0.1 sec

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 20


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
H] FACTORY SETTING MODE :

This mode is a useful mode and is required when the system malfunctions and does
not function correctly. Use of this function is cautioned as this will erase all the settings and
corrections applied.

This function is used to set the Airgauge to factory defaults.

How to Factory set Airguage UNIT :

To factory set Airguage UNIT directly switch off the Airguage and keep pressing “CAL”
key & then Power ON the Airgauge.

Effect of Factory setting on Airguage :

This will set default parameters in home position.

All the programs in setup mode will set to default parameters.

Reset channel reading.

This will erase correction factor, probe center, all the data in memory, ovality measured &
center offset calculated.

This process will take some time; wait up to Program No. 25 After that Airgauge will enter

in calibration mode, if you want to enter in Home Position, then press “ESC” key.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 21


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

I] CALLIBRATION MODE :

Requirement before calibration:


Ensure that before calibration input air pressure should be grater than 4 bar. This air is
regulated to 4 bar and this air is given to the unit.

In Regulator unit this air is again regulated to the 2 bar.


In calibration mode when the +20 µm ring is placed then adjust the front knob to display the
value between 6.000 & 7.000. And for -20 µm ring display should show the value between
3.000 & 4.000. Then start the calibration.

At power ON, Airgauge will enter into calibration mode or you can press “CAL” key
to enter in this mode. After Airgauge entered in calibration mode, You first select Channel No.,

( press “CH1” key for channel-1 or press “CH2” key for channel-2). Display shows

Insert Higher Master message, after inserting press “ENT” key. Then display shows

Insert Lower Master message, then press “ENT” key. Then displays Done & goes
back to calibration window.

Note: You should re-calibrate the Airguage for better


Performance .

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 22


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

J] MEMORY MANAGEMENT MODE :

Press “UP” key to enter into memory management Mode. Here you can see the stored
values in the memory. You can send the value to the RS232 (Serial port). You can delete the
stored jobs.
Table no: M-1: Memory management function menu.
Memory ACTION PERFORMED
managem
ent menu
It displays information about Job Measurement in
memory.
e.g: if we measure a job with Program No.1 &
Ch-1, then information shows its Memory No.,
View Program & Ch No., its Status as Accept/ reject/

rework & corresponding value.Press “ENT”


key to to send reading on rs232 port.
Send Memory window opens & it shows
sending… Data on the Serial Port (RS232). After
Send
sending all data it shows Done.press “ESC”
key for exit.
It asks for Clear Memory? Select Yes(press key

“ENT” ) to erase all memory, after clearing it


Clear

shows Done. Select No(press “ESC” key)


to go back.

Port Toggle between Printer Or PC with “LEFT”

key. Press “ENT” key to print data.


It will show the number of jobs saved in memory
Pending that you should need to clear before it will
overflow(limit is 10000 jobs).

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 23


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
K] MEASUREMENT MODE :
The two major types are: Static & Dynamic Modes. (see parameter No. 28 in set up mode).
K-1] STATIC MODE :
In static mode we measure ID/ OD of job. If current position is Static Mode then following
functions are assigned to each key:
Table no:N-1: Static mode Key function Description.
KEYS NAME DESCRIPTION
FUNCTION - 1 STORE VALUE & SEND CHANNEL-1 VALUE

FUNCTION -F2 STORE VALUE & SEND CHANNEL-2 VALUE

FUNCTION -F3 NO ACTION

CHARACTERI PRESET MODE OF


STIC - 1 CHANNEL 1

CHARACTERI PRESET MODE OF


STIC - 2 CHANNEL 2

CALIBRATION CALIBRATION BY MASTER 1 & MASTER 2 I.E. STATIC

THIS KEY ONLY EFFECT WHEN MEASUREMENT TYPE IS SET TO


ABSOLUTE / “AB-IN” IN PARAMETER SETTING. IF LED IS ON IT MEANS “ABS”
INCREMENT MEASUREMENT TYPE IS SELECTED.ON OTHER HAND IF LED IS
OFF IT MEANS “INC” MEASUREMENT TYPE IS SELECTED.
THIS KEY ONLY EFFECT WHEN MEASUREMENT TYPE IS SET TO
“IN-MM” IN PARAMETER SETTING. IF LED IS ON IT MEANS “INCH”
INCH/MM
MEASUREMENT UNIT IS SELECTED.ON OTHER HAND IF LED IS
OFF IT MEANS “MM” MEASUREMENT UNIT IS SELECTED.

ENTER NO ACTION

ESCAPE ERASE LAST ENTRY

UP MEMORY MANAGEMENT

DOWN NO ACTION

LEFT PROBE CENTER

PROGRAM PROGRAM MODE

SP1 NO ACTION

SP2 NO ACTION

Auto 1 STORE VALUE & SEND CH-1 VALUE


Auto 2 STORE VALUE & SEND CH-2 VALUE
Auto NO ACTION
1-1
Auto NO ACTION
1-2

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 24


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

7 SEGMENT Color indications:-


Tolerance Band:- Also there is a tolerance band. It is selected as per the tolerance set
(Tolerance band will show Higher & Lower Tolerance Limit).
In ID mode when the current job value is
1) Less than Lower master : 7-segment LCD of related Channel becomes Orange
(Rework mode).

2) Above the Higher master : 7-segment LCD of related Channel becomes Red (Reject
mode).

3) Between higher and lower master : 7-segment LCD of related Channel becomes Green
(Accept).

In OD mode when the current job value is


1) Less than Lower master : 7-segment LCD of related Channel becomes Red (Reject
mode).

2) Above the Higher master : 7-segment LCD of related Channel becomes Orange (Rework
mode).

3) Between higher and lower master : 7-segment LCD of related Channel becomes Green
(Accept).

AUTOSENCE MODE :
If Auto sense(Auto storing) mode is ON and if the current value does not change for 2 seconds
then this value will be automatically stored in the memory, also data will be sent on RS232
(serial port) for further PC based SPC analysis.

The job will be compared with the tolerance setting and according to that accept, reject or
rework count will be incremented. Also this status will be stored in the memory.

Also if “F1” key or foot switch is pressed then also the current value is stored in the
memory with accept, reject or rework status and count will be sent on RS232 (serial port).

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 25


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
K-2] DYNAMIC MODE :
This mode is used to measure Ovality of job. it has sub types as TIR (Ovality), max., min.,
Avg.(Mean), TIR_SP. One of them is active at a time (see parameter No. 28 in set up mode).
Table no:N-2: Dynamic mode Key function Description.
KEYS NAME DESCRIPTION

FUNCTION - 1 SEND OVALITY VALUE OF CH1


& Foot 1

FUNCTION -F2 SEND OVALITY VALUE OF CH2


& Foot 1

FUNCTION -F3 RESET OVALITY CORRECTION FACTOR TO 1.0

CHARACTERISTIC RESTART OVALITY TO ZERO OF CH1


-1 ( CHANNEL 1 )

CHARACTERISTIC RESTART OVALITY TO ZERO OF CH2


-2 ( CHANNEL 2 )

CALIBRATION CALIBRATION BY OVALITY MASTER

MEASUREMENT TYPE SELECTION OPTION.IT IS EITHER “ABS”


ABSOLUTE /
OR “INC”. THIS KEY ONLY EFFECT WHEN MEASUREMENT TYPE
INCREMENT
IS SET TO “AB-IN” IN PARAMETER SETTING NO.:07
MEASUREMENT UNIT SELECTION OPTION.IT IS EITHER “INCH”
INCH/MM OR “MM” .THIS KEY ONLY EFFECT WHEN MEASUREMENT TYPE
IS SET TO “IN-MM” IN PARAMETER SETTING NO.:09

ENTER NO ACTION

ESCAPE ERASE LAST ENTRY

UP MEMORY MANAGEMENT

DOWN NO ACTION

LEFT PROBE CENTER

PROGRAM PROGRAM MODE

SP1 NO ACTION

SP2 NO ACTION

Auto 1 RESTART OVALITY TO ZERO OF CH-1 VALUE


Auto 2 RESTART OVALITY TO ZERO OF CH-2 VALUE
IF CH1 OVALITY ≥ OVALITY MIN. TOLERANCE THEN ENABLE
Auto 1-1 RELAY OF CH1, OTHERWISE NO ACTION
IF CH2 OVALITY ≥ OVALITY MIN. TOLERANCE THEN ENABLE
Auto 1-2 RELAY OF CH2, OTHERWISE NO ACTION
Note: Auto1, Auto2, Auto1-1 & Auto1-2 are set time for auto sense operation.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 26


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
A) TIR (Ovality) :
• Start measurement, after inserting plug in job current reading is shown.
(See Parameter No. 20 to 28)
• If reading is in the auto-sense range then after auto-sense time the TIR (Ovality)
value will reset to zero automatically.
• Then rotate the job to check Ovality.
• If Ovality tolerance increases above Ovality Higher tolerance value the Selected
Channel’s 7- Segment color change to RED (i.e. reject) otherwise it will be Green
(i.e. accept).
• We set Ovality lower tolerance is 1 micron means job has minimum Ovality of 1
micron.

B) MAX :
• Start measurement, after inserting plug in job current reading is shown.
(See Parameter No. 20 to 28)
• If reading is in the auto-sense range then after auto-sense time the MAX value will
reset to current job value automatically.
• Then rotate the job to check MAX value. MAX reading is calculated automatically &
displayed.

C) MIN :
• Start measurement, after inserting plug in job current reading is shown.
(See Parameter No. 20 to 28)
• If reading is in the auto-sense range then after auto-sense time the MIN value will
reset to current job value automatically.
• Then rotate the job to check MIN value. MIN reading is calculated automatically &
displayed.

D) AVG :
• Start measurement, after inserting plug in job current reading is shown.
(See Parameter No. 20 to 28)
• If reading is in the auto-sense range then after auto-sense time the AVG value will
reset to current job value automatically.
• Then rotate the job to check AVG. Then MAX, MIN, AVG reading is calculated
automatically & AVG value is displayed.

E) TIR_SP :
• In this mode you can check both Ovality of the job as well as job size.
• Start measurement, after inserting plug in job current reading is shown.
(See Parameter No. 20 to 28)
• If reading is in the auto-sense range then after auto-sense time the TIR value will
reset to current job value automatically.
• Then rotate the job to check O
• SP mode will check for value entered, if this value is greater than size tolerance (i.e.
HL & LL) then it will be in reject mode.
• SP mode will check for Ovality: If Ovality is greater than Ovality Tolerance, then also it
will enter in Reject mode.
• If job is removed & again inserted & if its value is in the settable range, then Ovality
measurement will reset.
• Checking of value is continuous in SP mode.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 27


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL
K-3] AUTOSENSE MODE:
If Auto-sense mode is ON & if job value will be in the auto-sense range then after
auto-sense time, the job value will be stored automatically in the memory & send to the
serial port.
Press “F1” / Press “F2” / Foot-switch key to manually send data in the
memory.

(See key function Table in corresponding to mode).

K-4] PROBE CENTER ADJUSTMENT MODE :

sensor value is shown in percentage (i.e. Air Pressure value in percentage). This is used to
adjust the probe.
Press “LEFT” key, It will lead to selection of channel 1 or channel 2.

For selecting channel-1 press “CH-1” key or press “CH-2” for channel-2.

EXAMPLE :

OR

K-5] PRESET MODE :

Press “CH1” or “CH2” key to activate preset mode (i.e offset of center master) for

channel-1 or channel-2 respectively. Here it will show you current center master value for

respective channel. In this mode you can enter center master value by pressing “LEFT”

or “UP” key. Then center offset value will be calculated internally by the software using

bellow formula

Center offset = Center Master - Current Value.

Otherwise press “F3” or “ENT” key to calculate center master offset value using

above formula just difference is that here it will take currently stored Center master value and
exit this mode.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 28


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

L] MEASUREMENT WITH LVDT :

Airguage can measure length with LVDT, for more accuracy.

Connect LVDT probe to the Airguage & probe stand as shown in below fig.
For length measurement with LVDT do the following settings in Airguage.

• In Parameter settings menu : (Refer Chapter G for settings)

1) Set the Nominal value (i.e. Diameter setting ) Parameter No.1:- e.g. 20.000

2) Set the higher tolerance limit (i.e. HL) Parameter No.2:- e.g. 20.0100

3) Set the lower tolerance limit (i.e. LL) Parameter No.3:- e.g. 19.9900

4) Set the higher master (i.e. HM) Parameter No.4:- e.g. 1.0000

5) Set the lower master (i.e. LM) Parameter No.5:- e.g. 0.0000
Above HM & LM settings for 1mm measuring range, set these parameters as per required
range.

6) Set the Outer Diameter Mode Parameter No.6:-

7) Set the ABS mode Parameter No.7.

8) Set the Resolution & range as per requirement Parameter No. 8 :-


e.g. Resolution = 0.0001.

9) Set SIM - NOR mode from Parameter No.8.

Select SIM mode, for calibration by only one master i.e. HM (Higher Master).

[You can select NOR mode, for calibration by two masters i.e. HM & LM (Higher Master &
Lower Master).]

Calibration of LVDT Probe :


Here select SIM mode for calibration.

1) Now Press key to enter in calibration mode. Here (Big) 7

Segment display shows Actual probe value (when probe not pressed).

2) Adjust the LVDT probe to 0.0000 (i.e. Mechanical zero position.)

3) Place the Higher Master of 1.000 value & press key to do the calibration.

Calibration is done. Now, you can start the measurement.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 29


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

M] WIFI INTERFACE :

This is used for wireless communication to another device with Wi-Fi.

Start Wi-Fi communication as follows:

Press key and go to “WiFi testing” menu present in program setting menu

Press key then display will show you following message as “WiFi testing”,

then configure the WiFi module and after configuration press key for escaping from
“WiFi testing mode”.

Status No. entry mode:

Press key and go to “Status” menu present in program setting menu

Press key .Here you can change Status Number just by pressing key, changed

no. will be displayed on the LCD. The Status No. will be automatically saved.

(Status No. can be changed from 0 to 9.)

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 30


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

N] SERIAL INTERFACE MODE :

This software can be used for saving different jobs data (readings) from Airgauge, through
serial interfacing.

Open Serial interface software (from CD) (i.e. SPC v3.5), Install the software in PC.
Open the SPC Setup software

Following screen will appear

Select COM port (communication port of your computer)

On Airgauge, if key is pressed then data will be sent to the Serial Software or if

foot-switch is pressed then last saved memory data will be automatically sent to the serial port.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 31


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

Sent data is shown in software.

If you want to store data click on New from file menu & press Add button. You can see added
values in List as below:

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 32


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

When values are stored in the list you can see the Line Graph & Histogram.

Line Graph:-

Histogram:-

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 33


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

Setup mode:-

In this mode you can set different parameters

1) Data Setup: Here you can select COM port, fractional digits & multiplier for entered
value.

2) Target: This is to enter target value.

3) Statistical setup: LSL (Lower Specific Limit), USL (Upper Specific Limit) & No. of
groups can be changed.

4) Print setup: Company Name, Part Name, Part no., Operation Desc, Operator Name,
Machine & Gauge Name are settable.

5) Language selection: You can change English to German & German to English.

6) Color: Color combination of Line graph & Histogram can be changed.

7) Unit: Unit of measurement can be toggled between INCH (i.e. Imperial & MM (i.e.
Metric).

8) Select Range: Here you can select all data or can select specific range.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 34


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

Print from the software:-

You can print whatever you want to print by selection.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 35


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

O] CONNECTOR DETAILS :

INNOVATIVE AIRGAUGE 9 PIN D’ TYPE SOLD. MALE (SERIAL OUTPUT) CONNECTOR


DETAILS

PIN NO. SIGNAL


1 _
2 RX
3 TX
4 GND
5 _
6 _
7 _
8 _
9 _

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 36


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

P] INNOVATIVE AUTOMATION PRODUCTS LIMITED WARRANTY :


This product is warranted against defects for 12 months from the date of purchase from
authorized Innovative franchisees and dealers.

Within this period, we will repair it without charge for parts and labor. Warranty does not cover
transportation cost, nor does it cover a product subjected to misuse or accidental damage, nor
does it support damages due to lightning or very high power transients.

Except as provided herein, Innovative Automation Products make no warranties, express or


implied, including warranties of merchant ability and fitness for a particular purpose.

This warranty is limited to IAP COMP digital manufactured by us. The warranty on
scale/sensor is subjected to the warranty given by its respective manufacturer.

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 37


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

* TEST CERTIFICATE *

MODEL:

SR. NO.:

DATE:

TESTED AS PER OUR COMPANY STANDARDS AND ALL TESTS


FOUND TO BE SATISFACTORY.

FINAL INSPECTION DEPT.


INNOVATIVE AUTOMATION PRODUCTS

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 38


INNOVATIVE AUTOMATION PRODUCT
IAP COMP digital OPERATIONAL MANUAL

USER NOTES
Specifications and the design are subject to possible modifications
without notice due to constant improvements.

INNOVATIVE AUTOMATION PRODUCTS


B-24, MIDC, INDUSTRIAL ESTATE,
SATARA 415 004 (MAHARASHTRA)
INDIA
TELEPHONE : 91 2162 246509
TELEFAX : 91 2162 248012

EMAIL : [email protected]

Innovative Automation Product , B/24, MIDC, Satara - 415004,MH.,India. 39