SYS-6010 Service Manul V2.0
SYS-6010 Service Manul V2.0
SYS-6010 Service Manul V2.0
Service Manual
CO., LTD.
Without prior approval from MEDCAPTAIN in writing, this Service Manual shall not be
photocopied, modified or translated, fully or partially, by any individual or
Version Information
content language version Publish time code
1 IMPORTANT INFORMATION.................................................................... 1
3 MAINTENANCE ...................................................................................... 21
1
Use the maintenance parts and accessories provided by MEDCAPTAIN to replace
and maintain the device.
Do not use the infusion pump in a flammable environment.
High-frequency surgical equipment, mobile phone, wireless device and defibrillator
may have interference on the infusion pump. Keep away from them while operating.
After maintenance, have a safety test and cleaning according to this manual.
2
1.7 List of symbols
HOME
1.8 Contacts
a) When the infusion pump is used and maintained normally and regularly, we provide
18 months free maintenance (replacement of battery and appearance parts are not
included). For failure caused by other factors, part replacement will be charged
accordingly.
b) If you have any questions when using the infusion pump, please contact local
distributor or directly contact us at any time.
The after-sales service contact details of Medcaptain Medical Technology Co., Ltd.
are as follows:
Address: 12th Floor, Baiwang Research Building, No.5158 Shahe West Road, Xili,
Nanshan District, Shenzhen, P.R.China
Telephone: 400-809-3369
Fax: 0755-26001651
Postal: 518055
Website:http://www.medcaptain.com
E-mail:MC.Service@medcaptain.com
3
System Introduction
2 System Introduction
2.1 Introduction
2.1.1 Purpose
This product is intended for hospitals to infuse liquid at constant speed or liquid medicine
continuously through the veins of patients.
4
drives the peristaltic sheet to infuse through the mechanical drive device, monitors the
sensors and infusion process, and provides sound and light alarms.
2.3 Appearance
Figure 1
5
1 –IV groove 2 –Peristaltic Pump Tablet 3 –Lighting lamp
Figure 2
10 – Shell
6
2.3.1 Drop Sensor
1
2
4
3
4 – Socket
Figure 4
Figure 5
7
2.4 Circuit Block Diagram
The whole circuit block diagram of SYS-6010 is shown as Figure 6. The parts are as
follows:
V2.0
J2
4PIN
Interface board Battery pack
8PIN 8PIN
CON2 CON1
CON2 CON1
8PIN 8PIN
1PIN
3PIN
14PIN 14PIN
J16
J3
J2 J5
socket
4PIN
J8
AC
AC/DC power board DC/DC power board Power control board Pin board
4PIN
J7
3PIN
J8 J8
J9
J2
4PIN
J7
14PIN 14PIN
小3PIN
J10
3PIN 3PIN 7PIN 24PIN 8PIN NA 8PIN 7PIN 6PIN 小4PIN 3PIN 4PIN 4PIN
J1 J1 J4 J2 J3 J9 J2 J1 J6 J3 J5 J4 J9
J7
J2 Air
control sensor
24PIN bu-
M
Pump tablet
Air-bubble
Pump door
Speed test
test board
Auto door
Pressure
bbl
sensor
sensor
sensor
board
30PIN
30PIN
JP1
Keyboard 24PINAlarm
J1 Lamp
cover lamp board
board
Figure 6
2.4.1 AC/DC Power Board
a. Circuit Introduction:
The AC/DC power board has adopted the medical level power supply with low
leakage current and high-insulation voltage. It satisfies the requirements of EMC. The
AC/DC power board has an input alternating voltage ranging from 90 to 240V, 50/60Hz,
direct output voltage of 13.6V and the maximum output current of 2A. The AC/DC power
board mainly consists of AC inlet, fuse, EMC filter circuit, rectifier, high voltage DC filter
capacitor, switching tube, pulse width modulator, transformer, rectifier and absorbing
circuit, DC filter circuit, stabilized light decoupling feedback control circuit, etc.
8
c. Circuit Socket Number and Definition
No. Socket Description Pin-out Range Explanation
No. Description
1 DC output 1: DC output 13.0~13.7V Connect J1 in the
voltage inlet 2:Voltage 2.78~2.94V DC/DC Power board
J1 sampling
3: Ground Output DC
ground
2 AC input 1: Live line AC L line input Connect AC power
inlet 2: NC / input inlet
J2
3: Neutral line AC N line
input
3 Ground 1: Ground / Connect AC power
J3 connecting input inlet
inlet
d. Caution:
The AC/DC power board can reach to a high voltage of 400V. The workbench is
required to be isolated and avoid touching the high voltage circuit parts. High voltage
may cause electric shock hurts.
While using a multimeter or oscilloscope to test, the pens cannot short out the live
parts in the circuit board, or the circuit board would be damaged.
While using an oscilloscope or multimeter powered by mains to test the circuit of
mains in the AC/DC power board (primary non-isolated side circuit), the isolated
differential voltage probe must be used to measure, or it will damage the board and
test instrument.
While using an oscilloscope multi-channel probe to test the circuit, all ground loops of
the probe in the oscilloscope must be connected to the same potential point, or it will
damage the board and test instrument.
9
clock circuit in the shutdown situation. The input is from battery.
(5)The stepper motor driving circuit: adopts the specialized stepper motor controlling
circuit. Under the control of CPU, the speed, subdivision, current ratio and damping ratio
can be adjusted, depending on requirements.
10
SYS02 board
2 J3 8 Socket to connect the Connect J2 in the pressure
pressure&speed detection &speed detection board of the
board of the infusion pump infusion pump
3 J4 7 Socket to connect the Connect J1 in the pressure
pressure&speed detection &speed detection board of the
board of the infusion pump infusion pump
4 J5 14 Socket to connect the Connect J2 in the DC/DC board
DC/DC board
5 J7 4 Socket to connect the Connect J7 in the
pressure pressure&speed pressure&speed detection board
detection board of the of the infusion pump
infusion pump
6 J8 14 Socket to connect the Connect J8 in the DC/DC board
DC/DC board
7 J9 3 Not used Apply in the infusion pump
8 J10 3 Socket to connect the Connect bubble sensor
bubble sensor
9 J16 3 Socket to connect the Connect the battery pack plug
power supply
10 CON1 8 Socket to connect the Connect in the CON1 interface
interface board board
11
No.
1 / 3 Socket to connect the Connect J16 in the power board
power board
12
No. Socket PIN No. Description Explanation
+V A B Drop GND
(2)USB2 socket (BUS2): USB3.0 socket interface. To connect the external DC power
supply, bar code scanner of RS485 serial port, drop sensor signal, nurse call, and
program update of SYS02 board CPU. The external power supply should be used with the
matched cables. The DC input power is 9-15V and the current is 1A.
+V A B Drop GND
Remark: When the voltage of external power supply connecting to the USB1 and USB2
socket is over 13.6V, the external battery in SYS-6010 can be fully charged. If the voltage
13
is under 13.6V, the battery may not be fully charged or not charged.
+V A B Drop GND
14
(3)2.4 inch TFT color display: with a resistance to touch. The screen resolution is
320X240 with FPC welded at the circuit board.
b. The main testing points: none
c. Number and definition of the sockets
No. Socket PIN No. Description Explanation
1 J1 24 Socket to connect the Connect J1 in the alarm light
alarm light board board
2 J2 24 Socket to connect the Connect J2 in the power board
power board
3 J3 8 Socket to connect the Connect the button cover
button cover
15
Red light flashing High level alarm
b. The main testing points: none
c. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 J1 24 Socket to connect SYS02 Connect J1 in the SYS02 board
board
16
open motor board
9 J9 4 Socket to connect the Connect the infusion motor
infusion motor board
2.4.11 The infusion motor
a. The motor introduction: adopts a stepper motor with 42mm diameter and 48 steps, 2
phases driving.The working voltage is 18V.
b. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 / 4 Socket to connect the Connect J8 in the infusion pump
infusion pump pressure pressure&speed detection board
&speed detection board
2.4.12 Stepper motor speed board
a. The circuit introduction
Motor speed function: two light couplings collect the guiding and shading light signals A
and B from the optical disc connecting on the stepper motor. You can judge the speed and
rotating direction of the stepper motor according to the order of A and B and if there are
A and B.
b. The testing points:none
c. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 J1 4 Socket to connect the Connect J4 in the infusion pump
infusion pump pressure pressure&speed detection board
&speed detection board
2.4.13 The pump tablets position testing board
a. The circuit introduction
The peristaltic pump position testing function: the light coupling collects the guiding
light signals from the optical disc connecting on the peristaltic pump and judge the position
of the peristaltic pump during the rotation.
b. The main testing points: none.
c. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 J1 3 Socket to connect the Connect J5 in the infusion pump
infusion pump pressure pressure&speed detection board
&speed detection board
2.4.14 The pressure sensor
a. The pressure sensor introduction
The pressure sensor adopts the resistive Wheatstone bridge strain gauge pressure
sensor. When strain gauge pressure changes, the bridge will be out of balance, the
17
voltage will output and the pressure signal is received.
Note: The pressure sensor is a very sensitive and precision device. Do not touch the
surrounding parts or pull the connecting lines while dismounting.
b. The main testing points: none.
c. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 / 4 Socket to connect the Connect J3 in the infusion pump
infusion pump pressure pressure&speed detection board
&speed detection board
2.4.15 Bubble sensor
a. Bubble sensor introduction
The bubble sensor adopts the ultrasonic testing method. Its testing is not affected by the
color of infusion line. It has high-sensitivity (greater than or equalto 30ul bubbles) and a
fast response speed.
18
phases driving. The working voltage is 24V.
d. Number and definition of the socket
No. Socket PIN Description Explanation
No.
1 / 4 Socket to connect the Connect J9 in the infusion pump
infusion pump pressure pressure &speed detection board
&speed detection board
2.6 Specifications
Power supply AC power supply: AC 100-240V,50/60 Hz, power,
consumption 25 VA
External DC power supply: DC 12 V
Internal battery: lithium battery 11.1 V 1500 mAh
Time of continuous use: no less than 5 hours (for infusion at 25
mL /h with a new battery)
Compatible IV sets All disposable infusion sets of 20d/mL and 60d/mL conform to the
standards.
Infusion mode 6 infusion modes: rate, drop, time, standard weight, weight,
multi-rate
Infusion setting 0.1-1200.0mL /h or (1-400d/min) (20d/mL IV set)
range 0.1-200.0mL /h or (1-200d/min) (60d/mL IV set)
Least increment of 0.1 mL /h(or 1d/min)
Total volume 0.0-99999.9mL
display
Accuracy ±5%
Purge operation 400.0mL /h (20d/mL IV set)
200.0mL /h (60d/mL IV set)
Bolus operation 35.0-1200.0mL /h(20d/mL IV set)
19
35.0-400.0mL /h(60d/mL IV set)
Auto-calculating the speed by the bolus volume
KVO rate 0.1-5.0mL /h
Occlusion level 11 levels are available
225mmHg~975mmHg,11 level
1:225mmHg
2:300mmHg
3:375mmHg
4:450mmHg
5:525mmHg
6:600mmHg
7:675mmHg
8:750mmHg
9:825mmHg
10:900mmHg
11:975mmHg
Alarm Near End, Infusion End, Occlusion Alarm, Low Battery, Battery
Empty, No Battery, No Power Supply, Door open, Air Bubble, No
Drop Sensor, No Drop Signal, Drop Error, Reminder Alarm, and
alarms during self-test and running.
Special function
Repeat alarming: If there is still alarm after mute alarm sound, it
scanning.
Operating
Ambient temperature: 5℃-40℃ Relative humidity: 15%-95%,
conditions
nocondensation Atmospheric pressure: 70.0 kPa-106.0 kPa
Storage conditions
Ambient temperature: -20℃-60℃ Relative humidity:
10%-98%,
no condensation Atmospheric pressure: 22.0.0 kPa-107.4.0 kPa
Classification Class I, Type CF, IPX2
20
Maintenance
3 Maintenance
3.1 Cleaning and Sterilizing
Before cleaning the pump, be sure to turn off the power and disconnect the AC or DC
power cables.
If any solution spills on the pump or the pump gets heavily soiled, wipe it with wet soft
cloth dampened with cold or lukewarm water.
Use a piece of dry soft cloth to clean the AC power supply socket, USB socket or the
nurse call socket,ensure that the socket is dry before using it.
If the pusher and clutch need to be dismantled and cleaned, contact your distributors
or manufacturer.
Do not use organic solvent such as alcohol or thinner.
If disinfection is necessary, using the common disinfectors such as Chlorhexidine
gluconate and Benzalkonium chloride. After using the agent with a soft cloth,wipe off
it with a soft cloth dampened with water or warm water. When using the disinfecting
agent, follow the caution of each agent.
When sterilize the system,use EOG(Ethylene Oxide Gas) at the temperature of 50°C
or less and relative humidity of 60% or less. After sterilize,ventilate the room for over
24 hours or place system in aerator for 8 hours. The above is only a guideline,use
adequate method to check sterilization results.
The infusion pump cannot be autoclaved.
Never use a dryer or similar device to dry the infusion pump.
If liquid spills onto the pump, check whether the pump still functions normally. Test the
insulation and leakage current when necessary.
Do not soak the infusion pump into water.
3.2 Periodic Maintenance
Perform a periodic maintenance inspection to ensure safe operation and the longest
possible life of the infusion pump, and check the infusion pump once every six months.
You can maintain some items by yourself and contact your local distributor to maintain
some other items.
3.2.1 Checking the appearance (before every use)
Appearance checking: There are no cracks and damages.
Key operations: Keys response smoothly and effectively.
No abnormal noise during the peristaltic plate working.
3.2.2 Checking the power cable (before every use)
Check the appearance of the power cable. If the appearance is damaged and the
plug and the socket are in poor contact, contact the distributor for replacement in
time.
Connect the infusion pump to the AC/DC power and there is no indication of powering
on, contact the distributor for maintenance in time.
21
3.2.3 Checking the iInfusion rate (every 6 months)
Check the infusion flow once every 6 months by the graduate and timer.
Checking method: Use the preset brand IV set with a 60 mL /h rate, fill the infusion
bag with distilled water and access the graduate. Start, run 10 minutes at a 60mL /h rate
and observe the liquid volume in the graduate where 9.5-10.5mL is qualified.
22
3.3 Replacing the Battery
Remove an internal battery.
(1) Use a screwdriver to loosen the battery cover fixing screws at the bottom of the
pump.
(2) Remove the battery cover.
(3) Disconnect the battery cable connector.
(4) Remove the battery.
23
Software Upgrade
4 Software Upgrade
SYS-6010’s software upgrade involves two parts, being the Driver and UI upgrade.
24
In the menu bar, click on the "Configure" - > "Select Device", as shown in
figure choice chip PIC16F1527, click OK to exit;
In the menu bar, click on the "Programmer" - > "Select Programmer" - >
Select "PICkit 3" connection PICkit 3;
Click ok
25
PICkit 3 connection is successful, if not connected successfully check the
connection is correct or not;
In the menu bar, click on the "Programmer" => "Settings"=> "power", Click
the item "power target circuit from PICkit 3";then click “OK”
From the menu bar "File" - > "Import" hex File Import products. Burn MP - 30,
for example, find the corresponding version in the folder, and open;
26
Import success;
27
4.2 UI software Upgrade
Upgrade tool: please use long calbe to connect pump
28
Turn on the machine, then it would show “connect success”
In the Programer software, click "add", then open the upgrade file;
29
Then it would show “download complete”;
Warning!
There are 90-240V AC high-voltage current inside the MP-60A and 400V DC high-voltage
current produced by the switching power supply.A careless touching may cause serious
personal injury to the service engineer when they are conducting the checks and
troubleshooting.Please be strict in accordance with the requirements of the rules and
regulations to maintain the device by professional maintainers.
When using equipment to check power board or other parts, short-circuit or incorrect
measurement may cause serious personal injury or a severe damage to the device.
After every disassembling and maintenance, check the safety and function of the device.See
details in section 7.
The basic instruments of maintaining and checking
PC (to download software and fill in forms)
Software download connector
To be used infusion set
Digital multimeter
20-50mL graduate(accurate to 0.2mL )
Stopwatch or timing device
30
Note!
The graduate to inspect the infusion precision,is just used for general maintenance.
The special materials for disassembling should refer to the requirements in the
disassembling and installation chapter.
The safety project inspection needs specialized testing equipments.After disassembling
and installing the device, the safety of the pump must be tested.
31
Alarm Symptoms Alarm level Causes Troubleshooting
abnormally.
Near End Low The infusion will finish Wait until the infusion
within 3 minutes. finishes.
Occlusion Alarm High 1. The tubing is occluded. Click [STOP] to clear the
high viscosity.
Door Open High Door is open during Click [STOP] to clear the
closed.
Air Alarm High 1. Air bubble in the Click [STOP] to clear the
detector.
Infusion End High The VTBI or the infusion Click [STOP] to clear the
Reminder Alarm Middle Operation forgotten (no Click any key to clear the
loaded.
32
1 ERR-0003 Door position fault Replace the hall switch board
2 ERR-0004 Pressure sensor is falling or Reinstall the tube again or
abnormal pressed. Replace the pressure sensor and
the infusion pressure speed auto
driving board.
3 ERR-0005 Stepper motor speed error Replace the hall switch board or
Replace the stepper motor
4 ERR-0006 The 5V power supply Replace DC/DC power board
voltage or 18V motor drive
voltage is abnormal.
5 ERR-0007 Pump tablets fault Replace the peristaltic pump
tablets position test board
6 ERR-0008 SYS02 board and power SYS02 board damaged or the
board correspondence fault cable disconnection
7 ERR-0009 System fault Update software
8 ERR-0010 EEPROM fault Replace the SYS02 board
33
response when clicking The surface of screen is dirty Clean screen
Disassemble the device in a correct order. Incorrect order or forcible assembly might
cause irreversible damage to the device.
Before separating the components, please be sure that all connections are plugged
off. During separation, take care not to break the connection wire and damage the
connectors.
Please store the removed screws and other parts by category for easy reach during
34
reinstalling, and avoid dropping, contaminating oer losing them.
During disassembly and reinstallation, pay attention to protecting the waterproof seal
and do not pull the seal strip hard. Please do not miss or damage it during installation.
After assembling, check if there are extra screws or other components, If there are, a
reassembling is needed.
Warnings!
Before and after each assembling, be sure to cut off the external AC power
supply to avoid causing a human body injury.
Philips Screwdriver
Pliers
M3 nut spanner M3
6.3 Disassembly
6.3.1 Remove the battery
35
Figure 6.3-1 Note:
Code name
1 –Cover of battery
2 – PB2.3 tapping screw
3 –battery pack
1) Use phillips screwdriver to take out the 4 PB2.3 tapping screws (figure 6.3-1/code name
2) in the cover of battery (figure 6.3-1/code name 1).
2) Remove the cover of battery.
3) Take out the battery pack lightly (figure 6.3-1/code name 3)
Attention: pull out the battery pack lightly for there are connection wires between the
battery pack and the device.
4) The connection wires connect the battery and the device by a connector .Press the
block button in the connector and separate the plug and socket.
36
Figure 6.3-2 Notes:
Code Name
1 –Rear Cover 2 – PB2.3 PB2.3 tapping screw
3 – PT3.0 tapping screw
1) Use phillips screwdriver to take out the 6 PB2.3 tapping screws (figure 6.3-2/code name
2) in the rear cover (figure 6.3-2/code name 1).
2)Use phillips screwdriver to take out the 6 PT3.0tapping screws (figure 6.3-2/code name
3) in the rear cover.
3)Balance the rear cover with left hand, then use screwdriver with right hand to stick the
inner side of the AC power socket. Use a uniform force to inner cover and separate the
rear cover carefully.
Attention: be careful to remove the rear cover, avoid pulling out other components inside
the device. After removing the rear cover, please take a picture first to know well the circuit
connections and wire bindings.
37
Figure 6.3-3 Notes
Code Name
1 –PCBA stand component 2 –PCBA stand board
3 –PB2.3 tapping screw 4–FRC flat cable
5 –Middle plate
Attention: please connect the power supply, press [OPEN] to open the pump door, then
remove the battery.
1)Use the short edge of 2# inner hexagon spanner to push-out a certain space between
baffle-axis (figure 6.3-3/code name 2) and round decorative cover (figure 6.3-3/code
name 3), take out the round decorative cover.
2)Use pilers tip to clamp the pushed-out baffle-axis, pull it out completely.
3)Remove the baffle (figure 6.3-3/code name 1).
38
Figure 6.3-4 Notes:
Code Name
1 –Round Decorative Cover
2 – PB2.3 Tapping Screw
3 –Square Decorative Cover
4 –Cover Board
5 –Panel
6 – FPC Flat Cable
7 – Medium plate
8 –Flat Cable Sealing Groove
9 – PCB Bracket for PCB
1)Use pilers to take down the 9 round decorative covers (figure 6.3-4/code name 1) and 2
square decorative covers (figure 6.3-4/code name 3) in the cover board (figure 6.3-4/code
name 4).
Attention: use pilers tip to insert between the decorative covers and stick them up, avoid
breaking the cover board.
2)Use phillips screwdriver to remove the 7 PB2.3 tapping screws.
3)Take out the cover board lightly.
Attention: Do not miss the magnets in the cover board.
4)Remove the FRC flat cable (figure 6.3-4/code name 6) inside the cover board (figure
6.3-4/code name 5).
5)Remove the flat cable sealing groove (figure 6.3-4/code name 8) in the middle board
(figure 6.3-4/code name 7).
6)Pull out the bracket for PCB (figure 6.3-4/code name9) lightly, and pull out the FPC flat
cable at the same time.
Attention: take care the FPC flat cable not to be hung.
39
6.3.5 Remove the pump door components
40
Figure 6.3-6 Notes:
Code Name
1 –Pressure sensor 2 –M4 Machine Screw
3 – M4 Spring Washer 4 – M4 Flat Washer
5 –Middle board Label 6 –M3.0 Machine Screw
7 – M3.0 Spring Washer 8 –Components of the Peristaltic pump
1) Use phillips screwdriver to take out the M4 machine screws (figure 6.3-6/code name 2)
in the pressure sensor (figure 6.3-6/code name 3).
2)Remove the pressure sensor.
Attention: Do not break off the connection wire of the pressure sensor and take down
the pressure from the board.
3)Use pilers to tear down the middle board label (figure 6.3-6/code name 5).
Attention: Do not break the surface.
4) Use phillips screwdriver to take out the 6 M3.0 machine screws (figure 6.3-6/code name
6).Remove the M3.0 spring washer (figure 6.3-6/code name 7).
5) Remove components of the peristaltic pump (figure 6.3-6/code name 8).
6.3.7 Remove the Door Drive and Components of the Middle board
41
Figure 6.3-7 Notes:
Code Name
1 –Alarm lamp board 2 –PB2.3 tapping screw
3 –SYS02 module and display screen 4 – Lamp shade
5 –Lamp board 6 –Baffle
7 –FPC flat cable 8 –Panel
9 –M3 machine screw 10 –M3 spring washer
11 –Door catch 12 –Baffle spring
13 –Baffle block 14 –Membrane
15 –Baffle shaft
1)Use phillips screwdriver to remove 2 M3 machine screws (figure 6.3-7/code name 9) in
the baffle block (figure 6.3-7/code name 13), take down the 2 M3 spring washer (figure
6.3-7/code name 10).
Attention: hand press on the baffle, avoid the baffle spring pushes out it.
2)Take down the baffle block.
3)Turn up the baffle (figure 6.3-7/code name 6),take down the 2 baffle springs(figure
6.3-7/code name 12).
4)Use pilers to take down the baffle shaft (figure 6.3-7/code name 15).
Attention: use the pilers tip to push out it backward for a length and pull put it.
5)Take down the baffle.
6)Use phillips screwdriver to remove the 2 PB2.3 tapping screws (figure 6.3-7/code name
2) in the lamp shade(figure 6.3-7/code name 4).
7)Take down the lamp shadeand lamp board (figure 6.3-7/code name 5).
8)Take down the FPC flat cable (figure 6.3-7/code name 7).
9)Remove the connection wire in the alarm lamp board (figure 6.3-7/code name 1)from
42
SYS02 module board (figure 6.3-7/code name 3) lightly.
10)Use phillips to remove the 2 PB2.3 tapping screws in the alarm lamp board and take
down the alarm lamp board and lamp shadelightly.
11)Take out the SYS02 module board and display screen lightly.
Attention: put the display screen upward, avoid scratching or pollute it.
12) Use phillips to remove the 4 machine screws in the 2 door catches (figure 6.3-7/code
name 11) and take down the 4 M3 spring washers.
Attention: do not miss the plastic chips under the door catch,2 pieces of each door
catch.
13) Take down the door catch.
14) Tear down the membrane (figure 6.3-7/code name 14) from the panel (figure
6.3-7/code name 8).
Note: the installation operation method: Reverse operation according to the steps above.
6.4 Installation
The Safety Test
The electric and safety test is used to check the abnormal electric hazards.
If the hazards can not be discovered timely, it might cause injury to the
patient and operator.
The equipment of the electric and safety test adopts the safety analyzer,
maintainers should master the usage.
The electric and safety test completely complies with the standard of
IEC/EN60601-1.
The electric and safety test in the chapter takes the example of 601 safety
analyzer. Different brands of the safety analyzer might have a different
analysis.
If the electric and safety test fails, please contract our after-sales service
engineers.
43
3. Via power cord, connect the tested equipment to the auxiliary power output port
on 601 safety analyzer.
4. Connect one end of the red testing wire to the “Red input terminal” on the safety
analyzer, and clip another end tightly onto the metal foil on the surface of tested
equipment enclosure.
5. Turn on the power source of 601 safety analyzer and press “5-Enclosure
leakage” on the panel of 601 safety analyzer to enter the enclosure leakage
current testing interface.
6. The enclosure leakage current shall not exceed 100μA under normal conditions
and shall not exceed 300μA in event of a single fault.
44
Maintenance
6. Spare Parts List
Code Name
No. Part number Description No. Part number Description
1 60Z0020261 Cover of Battery a 7901000004 Interface Board
2 7404000006 Battery b 9114002321 Pressure&speed detecting board
3 9114002441 Rear Cover c 9114002252 Power Controll Board
45
MADCAPTAIN TECHNOLOGY CO., LTD
The after-sales service address: 12th Floor, Baiwang Research Building, No.5158
Shahe West Road, Xili, Nanshan District, Shenzhen, P.R.China
Telephone: 0755-26953369
Fax: 0755-26001651
Postal: 518055
Website:http://www.medcaptain.com
E-mail:info@medcaptain.com
2014 All rights reserved. SHENZHEN MEDCAPTAIN MEDICAL
NOLOGY CO., LTD.
46