2100N Manual
2100N Manual
2100N Manual
Model 2100N
Laboratory Turbidimeter
Instruction Manual
© Hach Company, 1993, 1996–2006, 2008. All rights reserved. 4/08 ed4
2
TABLE OF CONTENTS
CERTIFICATION ............................................................................................................................................................ 7
SAFETY PRECAUTIONS .............................................................................................................................................. 9
SPECIFICATIONS......................................................................................................................................................... 11
3
TABLE OF CONTENTS, continued
INSTALLATION/MAINTENANCE ......................................................................................................................47
4
TABLE OF CONTENTS, continued
5
6
CERTIFICATION
Hach Company certifies this instrument was tested thoroughly, inspected and
found to meet its published specifications when it was shipped from the factory.
The 2100N has been tested and is certified as indicated to the following
instrumentation standards:
Product Safety:
Listed by ETL to UL 1262 (Listing # H0492805390)
Certified by ETL to CSA C22.2 No. 1010.1 (Certification # H0492805390)
Certified by Hach to EN 61010-1 (IEC1010-1), supporting test records
by ETL
Immunity:
EN 50081-2: 97 (European Generic Immunity Standard) per 89/336/EEC
EMC: Supporting test records by Hach Company, certified compliance by
Hach Company.
Emissions:
Radiated Emissions per 89/336/EEC EMC: Supporting test records by
Amador Corp., Pinewood Oats (NVLAP #0271 01), certified compliance by
Hach Company.
7
CERTIFICATION, continued
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that
may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at
his own expense. The following techniques of reducing the interference problems
are applied easily.
1. Disconnect the Model 2100N Turbidimeter from it’s power source to verify
that it is or is not the source of the interference.
2. If the Model 2100N Turbidimeter is connected into the same outlet as the
device with which it is interfering, try another outlet.
3. Move the Model 2100N Turbidimeter away from the device receiving
the interference.
4. Reposition the receiving antenna for the device receiving the interference.
8
SAFETY PRECAUTIONS
Please read this entire manual before unpacking, setting up, or operating this
instrument. Pay particular attention to all danger and caution statements. Failure
to do so could result in serious injury to the operator or damage to the equipment.
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTE
Information that requires special emphasis.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to
the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for
operational and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12
August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
9
10
SPECIFICATIONS
(Specifications subject to change without notice)
Measurement Units:
Measurement Unit: NTUs, EBCs or Nephelos
Accuracy: ±2% of reading plus 0.01 NTU from 0-1000 NTU; ±5% of reading
from 1000 to 4000 NTU based on Formazin primary standards and with Ratio
ON; ±2% of reading plus 0.01 NTU from 0-40 NTU with Ratio OFF. Reference
Conditions: 0 to 40 °C, 0 to 90% RH Noncondensing @ 25 °C,
115/230 Vac ±17%, 50/60 Hz.* **
Response Time: 6.8 seconds with signal averaging off or 14 seconds with signal
averaging on.
Display: 5-character LED, 13.7 mm (0.54 in.) high digits with custom
annunciators
Light Source: Tungsten filament lamp. Lamp life 8,800 hours (typical).
* Turbidity specifications determined using USEPA Filter Assembly. Use of Flow Cell is required to achieve the cited
specifications.
** Intermittent electromagnetic radiation of 3 volts/meter or greater may cause slight accuracy shifts. Please see SUPPLEMENTAL
COMPLIANCE INFORMATION on page 80.
11
SPECIFICATIONS, continued
Temperature:
Storage Temperature: -40 to 60 °C (-40 to 140 °F)
Operating Temperature: 0 to 40 °C (32 to 104 °F)
Sample Temperature: 0 to 95 °C
Instrument Stabilization Time: 30 min. with ratio on, 60 min. with ratio off;
typical application leaves instrument on 24 hrs/day
Air Purge: 0.1 scfm at 69 kPa (10 psig: 0 hose barb connection for 1/8ð tubing,
Max 138 kPa (20 psig). Dry nitrogen or instrument grade air
(ANSI MC 11.1, 1975)
Serial I/O: RS232C serial interface via DB9 subminiature D shell connector for
data output to computer or printer, and data input (command). No handshaking.
Factory set for a 1200 baud rate, one stop bit, no parity, eight bit character length.
12
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary
Material Safety Data Sheets and become familiar with all safety procedures before handling
any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les
Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les
procédures de sécurité avant de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las
fichas de seguridad de materiales y familiarícese con los procedimientos de seguridad antes
de manipular productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden.
Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren
Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle
entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERICOLO
La manipolazione di campioni, standard e reattivi chimici può essere pericolosa. La
preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza
dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni
prodotto chimico.
13
14
SECTION 1 GENERAL DESCRIPTION
15
SECTION 1, continued
90° Detector
(Mandatory) Scattered
Light
Forward
Scatter
APERTURE Detector
IR Filter
Transmitted
Detector
Lamp Neutral
Sample Cell Density
Lens
Filter
16
SECTION 1, continued
• Instrument Manual
• Set of StablCal Primary Standards in sealed vials, one of each: <0.1 NTU, 20
NTU, 200 NTU, 1000 NTU, 4000 NTU
• Oiling Cloth
• Power Cord
• Dust Cover
If any of the items are missing or damaged, please contact the Customer Service
Department, Hach Company, Loveland, Colorado. Do not return the instrument
without prior authorization. In the United States, the toll-free number is
1-800-227-4224. Outside the United States, contact your nearest Hach dealer.
A power cord suitable for U.S. and Canadian 115 Vac line voltage is supplied with
the Model 2100N (Cat. No. 47000-00). If this model is to be configured for
230 Vac, an approved UL/CSA power cord with NEMA 6-15P type cord cap must
be used in place of the 115 Vac power cord supplied.
The Model 2100N (Cat. No. 47000-02) is factory configured for European
230 Vac line voltage. The power cord supplied with this model is VDE approved,
and has a Continental European type plug.
17
18
SECTION 2 TURBIDITY MEASUREMENT
Close the cell cover and press the I/O switch on the back instrument panel to turn
power on. Dark detector readings are taken immediately after the instrument is
switched on; error code ERR07 may be displayed if the cell cover is left open
during power up.
The EPA Filter Assembly is required for turbidity measurements reported for
United States Environmental Protection Agency (USEPA) National Primary
Drinking Water Regulations (NPDWR) or National Pollutant Discharge
Elimination System (NPDES) permits.
2. Hold the tab on the filter assembly, and insert the filter with the arrows
pointing toward the front of the turbidimeter.
3. Press the filter assembly all the way down into the housing. (See Figure 3.)
4. To remove the filter assembly from the instrument, grasp the tab and pull
straight up. Store the filter assembly in a clean environment.
19
SECTION 2, continued
DANGER DANGER
The 2100N Laboratory Le turbidimètre de laboratoire 2100N n’est pas prévu pour utilisation avec des liquides
Turbidimeter is not intended for inflammables ou contenant des hydrocarbures ou acides concentrés qui pourraient
use with flammable samples or attaquer les composants du 2100N. Effectuer des essais préalables en cas de doute sur
those containing hydrocarbons or la compatibilité de l’échantillon à contrôler.
concentrated acids that might
attack the 2100N components. PELIGRO
Conduct compatibility tests prior to El Turbidímetro de Laboratorio 2100N no está diseñado para usarse con muestras
analysis if the sample to be inflamables o que contengan hidrocarburos o ácidos concentrados que puedan atacar
monitored is in question. los componentes del 2100N. Ensaye antes del análisis si existe duda sobre la
compatibilidad de la muestra que se intenta analizar.
GEFAHR
Das Labortrübungsmessgerät 2100N darf nicht zur Analyse entflammbarer Proben
oder Proben, die Kohlenwasserstoffe oder konzentrierte Säuren enthalten, welche die
Teile des 2100N angreifen könnten, verwendet werden. Wenn die Verträglichkeit der zu
bestimmenden Probe fraglich ist, sollten vor der Analyse Tests durchgeführt werden.
PERIGO
O Turbidímetro de Laboratório 2100N não é feito com o fim de ser empregado com
amostras inflamáveis ou aquelas que contêm hidrocarbonetos ou ácidos concentrados
que possam atacar os componentes do 2100N. Os testes devem ser executados antes da
análise se existe alguma dúvida com respeito à compatibilidade da amostra a monitorar.
20
SECTION 2, continued
1. Collect a representative 2. Hold the sample cell by 3. Apply a thin bead of 4. Make sure the filter is in
sample in a clean container. the cap, and wipe to remove silicone oil from the top to place. Place the sample cell
Fill the sample cell to the water spots and finger prints. bottom of the cell—just in the instrument cell
line (approximately 30 mL). enough to coat the cell with compartment and close
Take care to handle the a thin layer of oil. Using the the lid.
sample cell by the top. Cap oiling cloth provided, spread
Note: For immediate update of the
the sample cell. the oil uniformly. Then, display, press ENTER.
wipe off the excess. The cell
Note: Instrument warm-up is not Note: When using the Flow-Cell
required. Optical and electronic
should appear nearly dry
with little or no visible oil. System, the Flow-Cell cover must
stabilization be in place for the LED light
is instantaneous. See Section 2.6.1. source to function.
5. Select manual or 6. Select the appropriate 7. Select the appropriate 8. Read and record
automatic range by pressing signal averaging setting (on Ratio setting (on or off) by the results.
the RANGE key. or off) by pressing the pressing the RATIO key.
Note: A measurement record can
SIGNAL AVG key. Select the appropriate unit be printed or transmitted via
(FNU or NTU) by pressing RS232 by pressing the PRINT key.
Note: See 3.1.3 for more
information.
the UNITS key.
Note: Values >40 NTU require
Ratio on.
21
SECTION 2, continued
• Do not leave a sample cell in the cell compartment for extended periods
of time.
• Empty the cell compartment and turn off the power if the instrument is stored
for extended periods of time.
22
SECTION 2, continued
Apply a thin bead of silicone oil from the top to bottom of the cell—just enough to
coat the cell with a thin layer of oil. Using the oiling cloth provided, spread the oil
uniformly. Then, wipe off the excess so that only a thin coat of oil is left. The cell
should appear nearly dry with little or no visible oil. Applying excess oil may
attract dirt and contaminate the sample compartment of the instrument.
1. Thoroughly clean and 2. Apply a thin bead of 3. Spread the oil uniformly
rinse the sample cell. silicone oil from the top to using the oiling cloth
bottom of the cell--just provided. Then, wipe off the
Note: See Section 2.5.1 on
page 22.
enough to coat the cell with excess so that only a thin
a thin layer of oil. coat of oil is left. The cell
should appear nearly dry
Note: See Section 2.6.1.
with little or no visible oil.
Note: Store the oiling cloth in a
plastic storage bag to keep the
cloth clean.
23
SECTION 2, continued
Clean all glassware with 1:1 hydrochloric acid and rinse several times with
dilution water when measuring low range turbidity samples. Cap the cells to
prevent small air-borne particles from contaminating the glassware if it is not
used immediately.
24
SECTION 2, continued
1. Fill the clean cell to the 2. Wipe the sample cell 3. Make sure the filter is in 4. Lift the cell
line with high-quality water, clean, and apply a film of place. Insert the sample cell compartment cover, rotate
and cap the sample cell silicone oil (refer to into the cell compartment, the sample cell (about 1/8 of
(refer to Section 2.6.2). Section 2.6.1). and close the cell cover. a turn). Close the lid, press
Record the reading. ENTER and record the
reading. Continue this
procedure until the smallest
NTU reading is obtained.
Note: “door” is displayed when
the sample compartment is
not covered.
5. Place an orientation
mark on the sample cell
marking band adjacent to the
index mark. Use this mark to
align the sample cell each
time a measurement is made.
25
SECTION 2, continued
1. Add portions of the high- 2. Make sure the filter is in 3. Insert the second cell
quality dilution water to place. Insert the first cell into the instrument, close the
multiple cells that are clean, into the instrument. Rotate lid, and note the value.
and coated with silicone oil. the cell slightly until the Rotate the cell
lowest reading is found. approximately 1/8 of a turn
Note: For more information, see:
Section 2.5.1 Cleaning
Note the cell orientation, and observe the reading.
Sample Cells record the reading, and add Repeat 1/8-turn rotations
Section 2.6.1 Applying an index mark to the until the reading matches the
Silicone Oil marking band of the cell. first cell reading within
Section 2.6.2 Preparing ±0.01 NTU. Add a permanent
Dilution Water.
orientation mark to the
marking band of the second
cell. Repeat this procedure
to match other cells.
Note: It may not be possible to
match all cells due to variability in
glass.
• Addition of a surfactant
• Application of heat
In some cases more than one method may be necessary for effective bubble
removal (e.g., some severe conditions may require combining use of heat with an
ultrasonic bath). Use care with these techniques; sample turbidity can be altered if
these methods are misused.
26
SECTION 2, continued
Letting the sample stand for a period of time to remove air bubbles is not
recommended. Particulates that cause turbidity may settle, and the sample
temperature may change. Both conditions may alter the turbidity of the sample
resulting in a measurement that is not representative of the original
sample turbidity.
To apply vacuum, use a sample degassing kit equivalent to Cat. No. 43975-00
(Sample Degassing Kit) as shown in Figure 4 or Cat. No.43975-10 (Sample
Degassing and Filtration Kit). These kits contain a syringe and stopper for
vacuum degassing. An electric or hand-operated pump equivalent to Cat. No.
14697-00 or 14283-00, respectively, also may be used.
27
SECTION 2, continued
1. Fill a clean sample cell with sample. Leave the cell uncapped.
2. Immerse the cell (1/2 to 2/3 immersed) in an ultrasonic bath, and allow it to
stand until visible bubbles are expelled.
3. Remove the cell and install the cap. Thoroughly dry the cell, and apply a film
of silicone oil.
Bubble expulsion may take a few seconds to a minute or more. Follow this simple
procedure to avoid excessive application of ultrasound. First, apply ultrasound for
a short period of time, and again measure turbidity. Continue for several
repetitions, noting the treatment time and turbidity readings. If turbidity begins to
increase instead of decrease, the ultrasound waves probably have started to alter
the suspended particles. Note the treatment time before the turbidity increase, and
record it as the maximum time limit for ultrasonic treatment.
DANGER DANGER
Make sure the cap on the cell is Vérifier que le bouchon sur la cuvette est desserré. Le chauffage d’une cuvette bouchée
loose. Heating a tightly-capped cell hermétiquement peut provoquer une explosion.
may result in an explosion.
PELIGRO
Cerciórese de que la tapa de la célula esté suelta. Calentar una célula cerrada
ajustadamente puede originar una explosión.
GEFAHR
Der Verschluss muss lose auf der Küvette sitzten. Das Erhitzen einer fest verschlossenen
Küvette kann eine Explosion verursachen.
PERIGO
Tenha certeza de que a tampa na cela esteja solta. O aquecimento de uma cela tapada
apretada demais pode ocasionar uma explosão.
Avoid use of heat to accelerate degassing whenever possible. Heat may change the
characteristics of the suspended particles, and cause volatile components to come
out of solution. Gentle heating may be helpful in degassing very viscous samples
when combined with application of vacuum or ultrasound. If heating the sample is
necessary, do so only to the extent required to accomplish degassing. Cool the
sample to the original temperature before measurement.
28
SECTION 2, continued
Light absorbing particles, such as activated carbon and significant amounts of true
color, also may cause an instrument to “go blind.” Dilution may not be effective in
correcting for these interferences.
When too much light is absorbed by the sample matrix, sufficient light may not be
available for measurement. If this condition occurs, the lamp icon on the
instrument display flashes to warn the user.
29
SECTION 2, continued
Filter Holder
Tweezers
Filter
Stopper
Filter flask
Filter Pump
Hose
100
Dilution Factor = --------- = 5
20
For example, if the measured turbidity value is 1100 NTU, the final turbidity value is calculated as:
1100 × 5 = 5500
30
SECTION 2, continued
Small diameter sample cells are useful with the instrument when only a small
quantity of sample is available, the sample to be measured is in an ampule and
cannot be opened, or the sample is too turbid for use with the standard sample cell.
A shorter light path permits measurement of high-range samples without the need
for sample dilution.
Note: The 2100N Turbidimeter reads Adapters are available for test-tube diameters of 12- to 13-mm, 16-mm and
slightly different with cell adapters 19-mm O.D. The 12- to 13-mm adapter accommodates either 12-mm or 13-mm
installed because of the shorter path tubes. The minimum sample volumes that must be used are 2.5 mL for 12-mm
length associated with the smaller
diameter sample cells. Refer to the
tubes, 3.5 mL for 13-mm tubes, 5 mL for 16-mm tubes and 7 mL for
instruction sheet sent with the cell 19-mm tubes.
adapters for
additional information. The adapters come with a tall light shield supplied for test tubes taller than the
standard cover.
Carefully select sample-cell glassware used with the adapters to be clean and free
of significant scratches. The same handling and cleaning care applied to the
standard 2100N sample cells applies to the smaller cells (including the use of
silicone oil on the outside of the glass).
31
SECTION 2, continued
32
SECTION 3 OPERATION
5 6 7 8 9 10 11 12
2100N Turbidimeter
FNU
Manual Auto
NTU Range Range CAL?
NEPH
13
UNITS
CALIBRATE RANGE Exit 14
4
3 EDIT S0
15
CAL S1
2
SIGNAL
S2
AVG
RATIO 16
S3
1 ENTER S4 PRINT
19 18 17
20 21 22 23 24
33
SECTION 3, continued
6 NEPH Lights when the instrument is set for Nephelos unit of measure.
7 NTU Lights when the instrument is set for NTU unit of measure.
8 EBC Lights when the instrument is set for EBC unit of measure.
9 Lamp Lighted annunciator indicates when the instrument lamp is on. Flashes to indicate a low-level light condition.
10 Manual Range Lights when the instrument is in the manual ranging mode.
11 Auto Range Lighted annunciator indicates when the instrument is in the automatic ranging mode.
Lights to indicate that the calibration information recorded during the calibration process is outside of the
12 CAL? acceptable range. This may indicate an operator error during calibration, or possibly an instrument malfunction.
If the CAL? annunciator flashes, the instrument must be recalibrated.
13 RANGE Selects Auto Ranging or Manual Ranging. Pressing RANGE steps the instrument through the range options.
14 UNITS Exit Selects units of measure (i.e., NTUs, EBCs or Nephelos). Also, exits calibration without saving new values.
Turns ratio feature on or off. Lighted annunciator indicates ratio on. Flashes to indicate over-range of 40 NTU
15 RATIO
when ratio is off.
Transmits the result of measurement to a computer or printer. If the instrument is in the calibration review
16 PRINT mode, pressing PRINT transmits calibration data to a printer or computer. If the PRINT key is held during power
up, a full set of diagnostic results is transmitted to a computer or printer.
17 SIGNAL AVG Turns the signal averaging function on or off. Lighted annunciator indicates Signal Averaging mode is on.
34
SECTION 3, continued
35
SECTION 3, continued
To review calibration data currently in affect, press the CAL key followed by the
PRINT key. Press the UNITS/EXIT key to return to the operating mode.
Holding the PRINT key down while turning on instrument power transmits
diagnostic information to the device.
1. Press and hold the RIGHT ARROW key for 3 seconds. If the display does not
read "00," use the ARROW keys to edit the display until it reads “00.”
3. Use the UP or DOWN ARROW keys to display the desired operational mode.
Press ENTER. The instrument executes the selection and returns to the
measurement mode.
36
SECTION 3, continued
3.2 Calibration
The electronic and optical design of the 2100N Turbidimeter provides long-term
stability and minimizes the need for frequent calibration. The three-detector
ratioing optical system compensates for electronic and optical system variations
between calibrations. When data is used for USEPA reporting, recalibrate at least
every 90 days, or as stipulated by the regulating authority.
Note: Alternately, obtain optimum The calibration is based on a first order linear equation consisting of up to three
calibration accuracy using the independent variables. Unpredictable results may occur if standards other than the
StablCal™ Calibration Kit for the recommended calibration points are used. The factory suggested calibration
2100N Turbidimeter. This kit contains
prepared, stabilized formazin
points are those determined by Hach Company chemists and engineers to provide
suspensions of <0.1-, 20-, 200-, 1000-, the best calibration accuracy. Use of Formazin standards other than those
and 4000-NTU. Order Cat. No. 26621- specified in Section 3.2.3 on page 38 may result in less accurate calibrations.
00 (500 mL each) or Cat. No. 26621-05
(ampules). Periodically, as experience or regulating authorities indicate, verify the instrument
calibration using pre-calibrated Gelex Secondary Standards (refer to Section 3.2.5
on page 42). If the reading in the range of use is not within 5% of the standard’s
assigned value, recalibrate using Formazin primary standards (refer to Section
3.2.4 on page 39).
37
SECTION 3, continued
Transfer approximately 30 mL of
well mixed 400-NTU Formazin stock
4000-NTU
solution to a clean sample cell. No
dilution is required.
* A class A volumetric pipet may be used in place of a TenSette Pipet.
38
SECTION 3, continued
1. Fill a clean sample cell 2. Make sure the filter is in 3. Press the CAL key. 4. Press the ENTER key.
to the line (approx. 30 mL) place. Insert the sample cell
with dilution water. Wipe into the cell holder, and The S0 annunciator lights. The instrument display
the cell clean and apply a close the cell cover. The NTU value of the counts down from
thin film of silicone oil dilution water used in the 60 to 0, and then makes a
(refer to Section 2.6.1 on previous calibration measurement. This result is
page 23). is displayed. stored and used to calculate
a correction factor for
Note: Ratio on and Ratio off
Note: For best accuracy use the
calibration data are measured and
measurement of all
matched sample cells for all NTU standards.
recorded simultaneously.
measurements during calibration
(refer to Section 2.6.3.2 Matching Note: Calibration for EBC and Note: If reading of dilution is >0.5
Sample Cells). An alternative may NEPH units of measure is NTU, an E1 error
be to use the same cell for all set automatically. message is displayed at the end of
standards. step 11.
Note: Upon entering the
Note: A portion of the same Calibration Mode, Automatic Note: The turbidity of the dilution
dilution water used for Range, Signal Averaging On and water can be “forced” to zero
preparing standards must be used NTU units are automatically rather than reading the dilution
in this step. selected. Upon completion of water. Refer to Section 3.3.1 on
calibration, all operational modes page 44.
Note: To exit the calibration
are restored to precalibration
procedure at any time without
settings.
changing any stored value, press
the UNITS/EXIT key.
39
SECTION 3, continued
5. The instrument 6. Fill a clean sample cell 7. Press ENTER. The 8. Fill a clean sample cell
automatically increments to to the line with well-mixed, display counts down from to the line with well-mixed,
the next standard, displays 20-NTU formazin standard. 60 to 0, and displays the 200-NTU formazin
the expected NTU value Wipe the sample cell clean, turbidity (compensated for standard.
(e.g., 20.00 NTU), and the and apply a thin film of dilution water turbidity).
S1 annunciator flashes. silicone oil on its surface. Repeat steps 6 and 7.
Remove the sample cell Place it into the cell holder, The instrument
automatically increments to The instrument will display
from the cell holder. and close the cell cover. 1000 NTU, and the S3
the next standard, the
Note: If using StablCal Stabilized display shows 200.0 NTU, annunciator lights. Remove
Formazin, use the 20-NTU and the S2 annunciator the sample cell from
Standard.
flashes. the instrument.
Note: If using StablCal Stabilized
Remove the sample cell Formazin, use the 200-NTU
from the instrument. Standard.
40
SECTION 3, continued
9. Fill a clean sample cell 10. Fill a clean sample cell 11. Press the CAL key. The
to the line with well-mixed, to the line with well-mixed, instrument stores the new
1000-NTU Formazin 4000-NTU Formazin calibration and returns the
standard. standard. instrument to the
measurement mode.
Repeat steps 6 and 7. Repeat steps 6 and 7.
Note: If power is lost
The instrument will display The display automatically during calibration,
1000 NTU, and the S4 increments back to the new calibration data is lost,
annunciator lights. Remove dilution water standard. The and the old calibration remains in
the sample cell from S0 annunciator lights, and effect. To exit calibration without
saving new values, press the
the instrument. the previously measured UNITS/EXIT key.
value of the dilution water is
Note: If using StablCal Note: If error message E1 or E2
stabilized formazin, use the
displayed.
appear in the display, an error
1000 NTU standard. Note: If using StablCal occurred during calibration (refer
stabilized formazin, use the to Table 5 on page 69). You may
4000 NTU standard. clear the error message and
proceed with measurements by
pressing the ENTER key.
However, the Cal? annunciator is
lighted indicating a questionable
calibration. The Cal? annunciator
is turned off only by recalibration,
which removes erroneous data.
Prepare new standards, and
recalibrate the instrument. Make
sure the Formazin standards are
fresh and well mixed. Also check to
ensure dilution water is < 0.5
NTU.
41
SECTION 3, continued
2. Verify that the instrument is set for the NTU mode and Automatic Ranging.
Select either Ratio on or off for the 0–2 and 0–20 range standards. Choose the
Ratio function that the instrument is intended to operate in. If measurements
will be made with Ratio on and off, assign a separate value for each.
3. Thoroughly clean the outside of the Gelex vials, and apply a thin coating of
silicone oil.
4. Place the lowest NTU Gelex Standard in the sample compartment with the
triangle on the vial aligned with the index mark on the instrument sample
compartment. Close the sample-cell cover.
5. Press the ENTER key. Record the value displayed. Remove the standard from
the instrument, and mark this value on the vial.
42
SECTION 3, continued
GEFAHR
Es wird dringend empfohlen, die Sicherheitsdatenblätter vor der Handhabung von
Behältern, Tanks und Zufuhrsystemen, die chemische Reagenzien und
Standardsubstanzen enthalten, aufmerksam durchzulesen, damit Sie sich mit den beim
Umgang mit diesen Chemikalien notwendigen Vorsichtsmaßnahmen, Risiken und
Notfallschutzmaß-nahmen vertraut machen. Es wird empfohlen, in allen Situationen,
in denen mit einem Kontakt von Chemikalien zu rechnen ist, eine Schutzbrille
zu tragen.
PERIGO
Para familiarizar-se com as precauções de manipulação, riscos e procedimentos de
emergência, examine sempre o Folheto de Dados de Segurança antes de manipular os
recipientes, tanques e sistemas de distribuição que contenham reagentes químicos e
outros elementos padronizados. Se recomenda sempre o uso de protetores para olhos,
quando possa acontecer contato com os produtos químicos.
Note: Preparing Formazin from raw Synthesize a 4000-NTU Formazin stock solution for making the calibration
materials is not recommended. standard dilutions as follows (in place of using the prepared stock solution):
Preparation is temperature and
technique sensitive. Use prepared 4000 1. Dissolve 5.000 grams of reagent grade hydrazine sulfate (N2H4•H2SO4) in
NTU Formazin stock solution to avoid
approximately 400 mL of deionzed water.
handling raw materials, and for best
instrument performance and assured
analytical accuracy.
2. Dissolve 50.000 grams of reagent grade hexa-methylenetetramine in
approximately 400 mL of deionzed water.
3. Quantitatively, pour the two solutions into a 1-liter volumetric flask, and
dilute to volume with deionzed water. Mix well.
43
SECTION 3, continued
44
SECTION 3, continued
45
46
INSTALLATION/MAINTENANCE
Some of the following manual sections contain information in the form of warnings, cautions and
notes that require special attention. Read and follow these instructions carefully to avoid personal
injury and damage to the instrument. Only personnel qualified to do so, should conduct the
installation/maintenance tasks described in this portion of the manual.
Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous la forme
d’avertissements, messages de prudence et notes qui demandent une attention particulière. Lire et
suivre ces instructions attentivement pour éviter les risques de blessures des personnes et de
détérioration de l’appareil. Les tâches d’installation et d’entretien décrites dans cette partie du
mode d’emploi doivent être seulement effectuées par le personnel qualifié pour le faire.
Algunos de los capítulos del manual que presentamos contienen información muy importante en
forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a
fin de evitar accidentes personales y daños al instrumento. Las tareas de instalación y
mantenimiento descritas en la presente sección deberán ser efectuadas únicamente por personas
debidamente cualificadas.
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von
Warnungen, Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden müssen.
Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder
Schäden am Gerät zu vermeiden. In diesem Abschnitt beschriebene Installations- und
Wartungsaufgaben dürfen nur von qualifiziertem Personal durchgeführt werden.
47
48
SECTION 4 AIR PURGE SYSTEM
Dry nitrogen or instrument grade air (ANSI MC 11.1, 1975) up to 138 kPa
(20 psig) can be used to air purge the optical housing compartment. The
recommended air consumption rate is between 3 and 10 SCFH (standard cubic
feet/hour). The connection is made at the AIR PURGE fitting on the rear panel.
Note: Do not exceed 138 kPa When the sample temperature is expected to be near or below 2 °C (35 °F), use a
(20 psig). desiccant dryer and particle filter in the air line to ensure the dew point of the air
purge is below the sample temperature (see Figure 9).
If only shop air is available, use a coalescing filter with automatic drain in
conjunction with a dryer and particle filter to achieve instrument quality air. Life
expectancy of the coalescing filter should exceed 2000 hours. Change the particle
filter at the same time as the air dryer. Figure 9 illustrates methods of connecting
the two types of air supply to the instrument. The dryer and filter are not necessary
if dry nitrogen is used for the air purge.
INSTRUMENT
AIR
3-10 SCFH PARTICLE FILTER AIR DRYER PRESSURE
AIR FLOW (Balston DFU 9933- (Balston DAU 9933- REGULATOR
05-BQ or equivalent) 05-101 or equivalent)
COALESCING
FILTER/REGULATOR
0-30 PSIG
SHOP
AIR
30-10 SCHF PARTICLE AIR FILTER FILTER
AIR FLOW FILTER DRYER HOUSING (Balston 100-
(Balston FR- 12-BX or
920-30 or equivalent)
equivalent)
AUTO DRAIN
(Balston 20-105
STANDARD SHOP AIR or equivalent)
49
50
SECTION 5 USING FLOW-CELL SYSTEM APPARATUS
5.1 Description
Three optional Flow-Cell kits are available for the 2100N
Laboratory Turbidimeter. Two kits are available for low-pressure applications
[< 34 kPa (5psig)] and one kit for high-pressure applications
[< 414 kPa (60 psig)].
A constant optical path is the most important benefit of a Flow Cell. Variability in
sample cells and poor sample-cell quality (due to handling, inadequate cleaning,
etc.) are the greatest sources of error in turbidity measurement.
DANGER DANGER
Do not use the Hach Flow Cells Ne pas utiliser les cuves à circulation Hach avec des échantillons inflammables ou ceux
with flammable samples or those contenant des hydrocarbures, solvants, acides concentrés ou bases concentrées qui
containing hydrocarbons, solvents, peuvent attaquer les parties au contact du liquide. Effectuer des essais avant l'utilisation
concentrated acids or concentrated des cuves à circulation si la compatibi-lité de l'échantillon est douteuse.
bases that may attack wetted parts
of the cells. Conduct tests prior to PELIGRO
use of Flow Cells if sample No use las Células de Flujo Flow Cells de Hach con muestras inflamables o que
compatibility is questionable. contengan hidtocarburos, soventes, ácidos concentrados o bases concentradas que
puedan atacar las partes mojables de la célula. Experimente antes de usat las Células de
Flujo, si existe duda sobre la compatibilidad de la muestra.
GEFAHR
Durchflußküvetten von Hach düfen nicht in Verbindung mit brennbaren Proben oder
Proben, die Kohlenwasserstoffe, Lösemittel, konzentrierte Säuren oder konzentrierte
Basen, die benetzten Teile der Küvetten angreifen können, verwendet werden. Wenn die
Verträglichkeit fraglich ist, sollten vor der Verwendung der Durchflußküvetten Tests
durchgeführt werden.
PERIGO
Não se deverá usar Celas de Fluxo hach con amostras inflamáveis ou aquelas que
contêm hidrocarbonetos, solventes, ácidos concentrados ou bases concentradas que
podem atacar as partes molhadas das celas. Realize os testes antes do uso das Celas de
Fluxo se é questionável a compatibilidade das amostras.
51
SECTION 5, continued
Sample is introduced into the top of the cell. A baffle deflects the incoming
sample to the side wall of the cell minimizing turbulence in the light path.
Sample is discharged from top and bottom outlets. The top outlet collects and
expels air bubbles and particles that tend to float. The cone-shaped bottom outlet
collects settleable solids; water discharged from the bottom outlet carries the
settled solids out of the cell. This novel, dual-outlet design eliminates dead
volume in the cell to provide rapid, thorough flushing of the cell from one sample
to the next.
52
SECTION 5, continued
Inlet Reservoir
Light Shield
Support Rod
Flow Cell
Collection Drain Assembly
Thumb Screw
Base
The kit consists of a Flow-Cell Stand Assembly, a Flow-Cell Inlet Reservoir with
a capacity of 350 mL, a reservoir cover, a Collection Drain Assembly, the
Flow-Cell Assembly, interconnecting tubing, and a Flow-Cell Light Shield.
CAUTION PRUDENCE
The manual and automated, Low- La cuve à circulation basse pression, manuelle ou automatisée, est conçue pour
Pressure Flow Cell setup is utilisation sous faible pression uniquement [<0,34 bar (34 kPa - 5 psig)].
designed for low-pressure use only
[<34 kPa (5 psig)]. PRECAUCION
La Célula de Flujo de faja presión, tanto manual como automática, ha sido diseñada
para baja presión solamente [34 kPa (5 psig - lbs/pulg.2 sobre la presión atmosférica)].
VORSICHT
Die manuelle und die automatisierte Niederdruck-Durchflußküvette ist nur für
Niederdruckanwen-dungen geeignet [<34 kPa (ca. 0,3 bar)].
PRECAUÇÃO
A Cela de Fluxo à baixa pressão manual e automatizada é projectada apenas para uso à
baixa pressão [<34 kPa (5 libras/polegada quadrada manômetro)].
53
SECTION 5, continued
1. Verify the sample compartment is empty, and turn the instrument off.
2. Turn the instrument on its top (place it on a soft cloth to protect the instrument
from marring), and install the base plate of the stand as illustrated in
Figure 12. Do not over tighten the screws.
6. Place the instrument right-side up, and install the support rod into the base.
Base
54
SECTION 5, continued
3. Cut a 2.5-cm (1 in.) piece of clear, 1/8-in. I.D. Tygon tubing, and install it
between the “Y” connector and the Collection Drain Assembly.
4. Cut a 50-cm (20 in.) length of clear, 3/8-in. I.D. Tygon tubing for the drain
line. Connect one end to the drain barb on the Collection Drain Assembly, and
run the other end to a suitable drain.
The discharge end of the drain tube must be unrestricted and lower than the
instrument for proper drain flow and prevention of air locking. Locate the
instrument as close to the drain as is practical using the shortest length of drain
tubing possible.
The kit is supplied with 152 cm (5 ft.) of 3/8-in. tubing. The system will not drain
properly if this length is exceeded. If the entire 152-cm length is used, the end of
the drain tubing must discharge at a point at least 46 cm (15 in.) below the center
line of the instrument to ensure proper flow.
55
SECTION 5, continued
Refer to Section 5.2.1.1 and Section 5.2.1.2 for assembly instructions. Omit step 5
in Section 5.2.1.1; the Collection Drain Assembly is not provided with the
Automated Flow-Cell Kit.
56
SECTION 5, continued
5. Cut a 50-cm (20 in.) piece of clear, ¼-in. I.D. Tygon tubing for the drain line.
Connect one end to the 1/8 x ¼-in reducer, and run the other end to a
suitable drain.
The discharge end of the drain tube must be unrestricted and lower than the
instrument to prevent air locking and ensure proper drain flow. Locate the
instrument as close to the drain as is practical using the shortest length of
drain tubing possible.
The kit is supplied with 152 cm (5 ft.) of ¼-in. tubing. The system will not
drain properly if this length is exceeded. If the entire 152 cm is used, the end
of the drain tubing must discharge at a point at least 46 cm (15 in.) below the
center line of the instrument to ensure proper flow.
6. Connect the power supply to the Power Jack on the Flow Valve Module. Plug
the power supply into an appropriate wall receptacle.
57
SECTION 5, continued
• Keep all parts of the system clean. Air bubbles tend to form around areas that
are not cleaned thoroughly.
• Do not attempt to use the Flow Cell for samples containing large particles that
may clog the system.
• Install the reservoir cover when the system is not in use to prevent
contamination of the system by air-borne particles.
• Always carefully pour sample down the inside edge of the inlet reservoir to
minimize agitation of the sample, which can entrain air bubbles.
• If bubbles tend to accumulate in the Flow Cell, gently tap the cell on a soft
surface to dislodge the bubbles.
• Do not use the system for monitoring flammable solutions, solvents, strong
acids or strong bases.
• Fill the system with distilled or deionized water when it is not used for short
periods of time (a few hours). This minimizes air locks and build up of
residue on the components.
58
SECTION 5, continued
1. Connect ¼-in. O.D. tubing (of appropriate lengths for sample-in and sample-
out connections) to the inlet and outlet fittings on the cap assembly as
described in steps 1, 2 and 3 of Figure 15. The Flow-Cell cap fittings are
compression fittings suitable for ¼-in. O.D. polyethylene, metal, glass or
clear vinyl tubing. Clear vinyl tubing requires an internal tube support. Step 4
of Figure 15 describes the method for removing tubing from this type of
compression fitting (shown in the cross-section drawing in Figure 17).
Note: Hand-tighten the compression Figure 16 illustrates a suggested hookup using a pressure regulator in the inlet line
fitting nuts. Excessive force may and a flow meter or flow control valve on the outlet line. The control valve is
damage the fitting or cap assembly (see installed on the outlet side because it may introduce bubbles into the sample
Figure 17).
causing positive interference in the turbidity reading. This hookup also maintains
sufficient pressure to minimize outgassing in a carbonated stream.
Flow-Through Cell
Flow Meter
(optional)
Return
Sample Or
In To Drain
Pressure Control
Regulator Valve
59
SECTION 5, continued
Nut o-ring
Tubing
Grab Ring
Spacer
CAUTION PRUDENCE
Provide back-flow protection to Installer une protection anti-retour pour éviter les variations brusques de pression au-
prevent pressure surges from dessus de 4 bar (400 kPa, 60 psig) dans les applications où l’échantillon retourne dans
exceeding 60 psig in applications une ligne sous pression.
where sample is returned to a
pressurized line. PRECAUCION
Tome medidas para eliminar el reflujo para evitar golpes de presión que excedan 60
lbs/pulg. cuadrada en el manómetro (psig) en aplicaciones en las que la muestra
retorna a una línea bajo presión.
VORSICHT
Es ist für einen Rücklaufschutz zu sorgen, um zu verhindern, daß Druckstöße bei
Anwendungen, bei denen die Probe in eine unter Druck stehende Leitung
zurückgeführt wird, 4,2 bar überschreiten.
PRECAUÇÃO
Dever-se-á fornecer protecção de contra-fluxo para impedir aumentos repentinos de
pressão que excederem de 60 libras por polegada quadrada em aplicações em que a
amostra regressar a um cano sob pressão.
2. Thoroughly clean the glass sample cell, and screw it into the cap assembly
(see Figure 14). Make sure the O-ring is in place in the cap assembly to
ensure a good seal (hand tighten). Wipe the glass surface free of fingerprints
and smudges. Apply a thin, even film of silicone oil to the outside surface.
DANGER DANGER
Use only glass sample cells that are Utiliser seulement les cuves en verre marquées PRESSURE TESTED, 60 PSIG
screened PRESSURE TESTED, (éprouvée à la pression, 60 psig # 4 bar). L’utilisation de toute autre cuve présente des
60 PSIG. Use of any other cells risques de blessures de l’opérateur et de dommages pour l’appareil. En cas de fuite ou
may result in injury to the operator de bris de cuve, dépressuriser immédiatement le système et débrancher
and damage to the instrument. In l’alimentation électrique.
the event of leakage or breakage,
immediately depressurize the PELIGRO
system and disconnect power. Use solamente células de vidrio que estén clasificadas PRESSURE TESTED, 60 PSIG.
El empleo de cualquier otro tipo de célula puede resultar en lesiones al operador y
daños al equipo. Si ocurrieran escapes o roturas, inmediatamente alivie la presión del
sistema y desconecte la energía.
60
SECTION 5, continued
GEFAHR
Es dürfen nur Glasküvetten benutzt werden, die bis 4,2 bar druckgeschützt sind. Die
Verwendung anderer Küvetten kann zu Verletzungen des Benutzers oder zu Schäden
am Gerät führen. Sollten Küvetten auslaufen oder zerbrechen, muß das System sofort
entlüftet und die Stromzuführung unterbrochen werden.
PERIGO
Dever-se-á usar celas de amostras em vidro que sejam pré-examinadas com PRESSÃO
DE TESTE DE 60 LIBRAS/POLEGADA. O uso de quaisquer outras celas pode
ocasionar lesão ao operário e estragar o instrumento. Em caso de esvazamento o
ruptura, dever-se-á imediatamente soltar a pressão do sistema e desligar a
corrente eléctrica.
3. Start sample flow through the system, and watch for leaks at the top of the
sample cell. Adjust the pressure regulator to maintain pressure below
414 kPa (60 psig). Adjust the flow-control valve to a suitable flow rate below
500 mL/minute. Generally, low flow rates minimize signal noise caused by
bubbles and particulate matter.
4. Insert the Flow Cell into the instrument cell holder.
DANGER DANGER
Use eye protection and handle with Porter des lunettes de protection et manipuler avec une extrême prudence lors de la
extreme care when pressuring the pressurisation du système avec la cuve à circulation hors de l’appareil. Ne pas tenir
system with the Flow Cell out of l’ensemble par la cuve en verre. Utiliser un écran de protection entre l’opérateur
the instrument. Do not hold the et la cuve.
unit by the glass cell. Use a
protective shield between the PELIGRO
operator and the cell. Use protección de los ojos y manipule con extrema precaución al iniciar la presión en el
sistema mientras la célula de Flujo Continuo (Flow-Thru) esté retirada del
instrumento. No sostenga la unidad tomándola por la célula de vidrio. Interponga un
escudo protector entre el operador y la célula.
GEFAHR
Wenn das System unter Druck gesetzt wird, während sich die Durchflußküvette nicht im
Gerät befindet, ist mit äußerster Vorsicht zu arbeiten; ein Augenschutz sollte getragen
werden. Die Einheit darf nicht an der Glasküvette festgehalten werden. Zwischen
Küvette und Benutzer sollte ein Schutzschild aufgestellt werden.
PERIGO
Usar protecção para os olhos e manusear com muito cuidado ao pressurizar o sistema
con a cela Flow-Thru fora do instrumento. Não sostenha com a mão a unidade pela
cela de vidro. Usar um protector entre o usuário e a cela.
61
SECTION 5, continued
Coat the outside surface of the glass, Flow-Cell parts with a thin film of silicon oil
before installation in the instrument (see Section 2.6.1 on page 23).
Note: Always test the system for leaks Fill the system with distilled or deionized water when it is not used for short
before inserting the Flow Cell into the periods of time (a few hours). This minimizes air locks and build up of residue on
turbidimeter. the components. Always disassemble, thoroughly clean, and air dry all
components before long-term storage.
62
SECTION 6 DATA OUTPUT
Data is transmitted to the printer as a 39-character string plus the line feed and
carriage return.
1 5
6 9
63
SECTION 6, continued
Instrument Printer
(DB9 Female) (DB 25 Male)
3 3
5 7
6 20
SHELL 1 (SHELL)
Instrument Computer
(DB9 Female) (DB9 Female)
2 3
3 2
5 5
SHELL SHELL
64
SECTION 6, continued
The Citizen printer requires some configuring for compatibility with the 2100N
Turbidimeter. Set the switches and jumper positions on the Citizen Model iDP-
562RSL II printer board as follows (see Figure 23 and Table 4):
65
SECTION 6, continued
4. Press the ENTER key to activate the print speed selection mode.
5. Use the UP and DOWN ARROW keys to select the flashing SL Pr for slow or
FS Pr for fast print speed.
6. Press the ENTER key to accept the desired setting, and exit the print speed
configuration mode.
• Key in VAL (for value) and press enter on the computer keyboard. This action
recalls the current 2100N measurement with the measurement units.
• Key in LST (for list) and press enter on the computer keyboard. This action
lists the calibration standards and coefficients.
66
SECTION 7 MAINTENANCE
7.1 Cleaning
Keep the turbidimeter and accessories clean. Use a cloth dampened with mild
detergent and water when the enclosure and keypad require cleaning. Wipe up
spills promptly. Wash sample cells with nonabrasive laboratory detergent, rinse
with distilled or demineralized water, and air dry. Avoid scratching the glass cells,
and wipe all moisture and fingerprints off of the cells before inserting them into
the instrument (refer to Section 2.5.1 Cleaning Sample Cells).
CAUTION PRUDENCE
Turn the 2100N Turbidimeter off Eteindre le turbidimètre 2100N et débrancher l'alimentation électrique avant de
and disconnect the power before nettoyer l'appareil.
cleaning the instrument.
PRECAUCION
Apague el Turbidímetro 2100N antes de limpiar el instrumento.
VORSICHT
Vor der Reinigung muß das Trübungsmeßgerät 2100N abgestellt und der Netzstecker
gezogenwerden.
PRECAUÇÃO
Apague o Turbidímetro 2100N e desligue a corrente eléctrica antes
de limpar o instrumento.
1. Turn the instrument off, and disconnect the power cord from the back
panel receptacle.
CAUTION PRUDENCE
The lamp must be cool before La lampe doit être froide avant de la retirer del’appareil.
removal from the instrument.
PRECAUCION
La lámpara debe dejarse enfriar antes de intentar quitarla del instrumento.
VORSICHT
Die Lampe muß vor der Entnahme aus dem Gerät abgekühlt sein.
PRECAUÇÃO
A lâmpada deverá estar esfriada antes de tirá-la do instrumento.
CAUTION PRUDENCE
Wear protective eye wear if the Porter des lunettes de protection si la lampe est allumée, alors que le capot de la lampe
lamp is turned on while the lamp est retiré.
cover is removed.
PRECAUCION
Use protección de ojos cuando la lámpara esté encendida y su cubierta protectora esté
fuera de posición.
VORSICHT
Wenn die Lampe bei abgenommener Lampenabdeckung brennt, ist ein Augenschutz
zu tragen.
67
SECTION 7, continued
PRECAUÇÃO
Quando a Lâmpada estiver acesa e a proteção de lâmpada estiver removida, use
protetor ocular.
2. Remove the cover over the lamp compartment (see Figure 24). Squeeze the
latch tab together, and pull out on the cover.
3. Use a small screwdriver to loosen the two screws securing the lamp leads in
the lamp terminal block. Pull the leads free.
4. Remove the two Phillips-head screws securing the lamp holder in the
instrument (upper left-hand corner and lower right-hand corner). Remove the
entire lamp assembly.
Note: Do not touch the lamp; oil from 5. Mount the new lamp assembly in the instrument.
skin will cause damage. Clean the lamp
with alcohol if necessary. 6. Insert the lamp leads under the clamping plates in the lamp terminal block.
Lamp leads are not polarized and therefore may be placed in either position
on the terminal block. Tighten the terminal block screws.
7. Replace the lamp compartment cover, and connect the power cord to the back
panel receptacle.
68
SECTION 8 TROUBLESHOOTING
8.1 Introduction
The Model 2100N Laboratory Turbidimeter has several self-diagnostic functions
and Error Codes to permit convenient and effective system troubleshooting.
69
SECTION 8, continued
Access diagnostic information by pressing and holding the RIGHT ARROW key for
3 seconds. Use the ARROW keys to edit the display to read the diagnostic code
number of interest. Press the ENTER key to display the diagnostic value.
70
SECTION 8, continued
8.3.2.3 Flashing 9s
If the display flashes all 9s, the sample being measured is over-range (for the
selected range of measurement). The instrument also indicates over-range if the
sample is >40 NTU (268 Nephelos or 9.8 EBCs) with the Ratio off (refer to
Section 2.8 on page 29).
8.3.2.4 Flashing 0s
If the display flashes 0s, the measurement indicates a negative turbidity value.
Recalibrate the instrument (see Section 3.2.4 on page 39).
71
72
General Information
73
74
REPLACEMENT PARTS AND ACCESSORIES
Description Cat. No.
2100N Laboratory Turbidimeter, 110/220 V with UL and CSA approved power cord and fuse ........................47000-00
2100N Laboratory Turbidimeter, 110/220 V with European power cord and electrical fuse..............................47000-02
REPLACEMENT PARTS
Cover, lamp compartment ....................................................................................................................................47032-00
Cover assembly, sample cell compartment ..........................................................................................................47713-00
Dust Cover............................................................................................................................................................47030-00
Formazin Primary Turbidity Stock Solution, 4000 NTU, 100 mL ........................................................................2461-42
Gelex® Secondary Turbidity Standardization Kit ................................................................................................25890-00
Includes:
Empty Vial (for stray light measurement..........................................................................................................25891-00
0–2 NTU ...........................................................................................................................................................25891-01
0–20 NTU .........................................................................................................................................................25891-02
0–200 NTU .......................................................................................................................................................25891-03
200–4000 NTU .................................................................................................................................................25891-04
Instrument Manual ............................................................................................................................................47000-88
Lamp Kit, Replacement ....................................................................................................................................47089-00
Oiling Cloth ......................................................................................................................................................47076-00
Power Cord, 18/3 SVT, 10 A, 125 V, North American, 110 V, UL/CSA approved Power Cord .....................18010-00
or
Power Cord, 0.75 mm SQX3 conductor, European, 220 V, VDE approved Power Cord ................................46836-00
Quick Reference Card.......................................................................................................................................47000-44
Sample Cells, 6/pkg, 2100 N ............................................................................................................................20849-00
Silicone Oil, 15 mL dropper bottle .....................................................................................................................1269-36
75
REPLACEMENT PARTS AND ACCESSORIES, continued
76
How To Order
By Telephone: By Mail:
6:30 a.m. to 5:00 p.m. MST Hach Company
Monday through Friday P.O. Box 389
(800) 227-HACH Loveland, CO 80539-0389
(800-227-4224) U.S.A.
By FAX: (970) 669-2932
Information Required
International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the
representative nearest you, send an e-mail to: [email protected] or contact:
77
Repair Service
Authorization must be obtained from Hach Company before sending any
items for repair. Please contact the HACH Service Center serving
your location.
In Canada:
Hach Sales & Service Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only)
Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: [email protected]
78
Limited Warranty
Hach Company warrants its products to the original purchaser against any defects
that are due to faulty material or workmanship for a period of one year from date
of shipment unless otherwise noted in the product manual.
In the event that a defect is discovered during the warranty period, Hach Company
agrees that, at its option, it will repair or replace the defective product or refund
the purchase price excluding original shipping and handling charges. Any product
repaired or replaced under this warranty will be warranted only for the remainder
of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents;
or consumable components of a product, such as, but not limited to, lamps
and tubing.
Limitations
This warranty does not cover:
• Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or
undeclared), terrorism, civil strife or acts of any governmental jurisdiction
• Damage caused by any repair or attempted repair not authorized by Hach Company
This warranty contains the sole express warranty made by Hach Company in
connection with its products. All implied warranties, including without limitation,
the warranties of merchantability and fitness for a particular purpose, are
expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied
warranties and if this is true in your state the above limitation may not apply to
you. This warranty gives you specific rights, and you may also have other rights
that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty
terms and no person is authorized to make any other warranties or representations
on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above
are the exclusive remedies for the breach of this warranty. On the basis of strict
liability or under any other legal theory, in no event shall Hach Company be liable
for any incidental or consequential damages of any kind for breach of warranty
or negligence.
79
SUPPLEMENTAL COMPLIANCE INFORMATION
The 2100N Turbidimeter shows an accuracy shift when exposed to radio
frequency interference of 3 volts per meter. If the displayed data are not stable
(within the accuracy and repeatability specifications) and other interferences are
not suspected as the cause, inspect the area for portable phones, paging service
tower/antennas, or any other transmitting communication device. For example, if
the user of the instrument detects a measurement deviation while using a hand-
held radio communication device to relay information to a base station, the
interference is clearly caused by the hand-held radio communication device. To
resolve the interference problem, move the hand-held radio communication
device at least 3 to 4 meters from the instrument to provide isolation.
As shown below, the instrument reading may deviate slightly win the presence of
specific frequencies. Any electronic instrument is sensitive to RF fields if the
power is great enough. However, typical transmitting power is limited by
regulatory controls, and field strengths of greater than 3 volts/meter are unusual.
80