HARK Acuitus Service Manual Rev C PDF
HARK Acuitus Service Manual Rev C PDF
Refractor/Keratometer
™
Acuitus Models
5000/5010/5015
Service Manual
CONFIDENTIAL & PROPRIETARY
Service Manual
CONFIDENTIAL & PROPRIETARY
Information contained in this document is copyright Carl Zeiss Meditec Incorporated and may not be
reproduced in full or in part by any person without prior written approval of Carl Zeiss Meditec. Its
purpose is to provide the User with adequately detailed information so as to repair, maintain, and
order spare parts for the instrument supplied. Every effort has been made to keep the information
contained in this document current and accurate as of the date of publication or revision. However,
no guarantee is given or implied that the document is error-free or that it is accurate with regard to
any specification.
All trademarks, registered trademarks, and trade names that appear in this manual are the
property of their respective holders.
ii
REVISION CONTROL LIST
Listed at the bottom of each page is the part number of the manual, along with the Rev letter
and date for that page (for example, nnnnnAmmyy). Subsequent revisions to a page will be
noted by a corresponding change to the Rev letter and date.
Pages in this document are at Revision B unless noted otherwise below.
iii
iv
Acuitus 5000/5010 Table of Contents
Table of Contents
Section 2 - Checkout
2.1 System Checkout Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
2.2 Optics Head Cleaning Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Section 4 - Adjustment/Calibration/Checks
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
4.2 Related Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
4.2.1 Calibration Software Setup For 17" VGA Display . . . . . . . . . . . . . . . . . 4 - 3
4.2.2 Write Enabled Service Calibration Disk . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
4.3 Camera Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
4.4 Source LED Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 9
4.5 Calibration Block Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 13
4.6 Chopper Motor Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
4.7 Refractor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
4.8 Keratometry Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 20
4.9 Optometer Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
4.10 Auto-Z Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
Section 5 - Troubleshooting
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
5.2 General Guidelines for Assembly Level Troubleshooting . . . . . . . . . . . . . . . . . . . 5 - 3
Section 6 - Diagrams
Figure 6-1. Acuitus Model 5000 System Interconnect Diagram
Figure 6-2. Acuitus Model 5010 System Interconnect Diagram
Figure 6-3. Acuitus Model 5015 System Interconnect Diagram
Section 7 - Parts
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
7.2 Replacement Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
7.3 Parts Orders — U.S. Domestic Service Operations . . . . . . . . . . . . . . . . . . . . . . . . 7 - 4
7.4 Parts Orders — International Service Operations . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
7.5 Returning Instruments / Defective Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
7.5.1 Equipment Return Authorization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
7.5.2 Packing the Instrument for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Appendices
Appendix A. Tools, Test Equipment and Service Supplies . . . . . . . . . . . . . . . . . . . . . . . A - 1
A.1 List of Tools, Test Equipment
and Service Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
A.2 The Special Tools - What They Are / What They Do . . . . . . . . . . . . . . . . . . A - 2
A.2.1 Service Calibration Diskette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
A.2.2 Carousel Mount Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
A.2.3 Tension Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
A.2.4 Level Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
A.2.5 Carousel Calibration Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 2
A.2.6 Customer Test Eye . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 3
A.2.7 LED Jumper Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 3
A.2.9 Static Protection Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 4
Appendix B. Restoring The Operating System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B - 1
Appendix C. Restoring The Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C - 1
Appendix D. Restore Calibration Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D - 1
Appendix E. Service Forms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E - 1
Notes:
The service manual is designed to support Level 1 field service, which employs modular
replacement of printed circuit boards and other assemblies that are most effectively repaired at a
central repair facility. This is the service strategy used in U.S. domestic field service, and
presented in Humphrey training classes. See section 1.3 for additional information regarding
Level 1 and Repair Center service.
The procedures in this manual assume that the reader is familiar with operation of the instru-
ment. Complete operating instructions are contained in the Acuitus Models
5000/5010/5015 User’s Guide. Information presented in the User’s Guide is not repeated
in this Service Manual. The User’s Guide can be ordered separately by standard Humphrey
parts order. Refer to section 7.7.1 for the User’s Guide part number information.
The general layout of the service manual is shown below. For greater detail, please refer to the
Table of Contents.
1.1.2 Conventions
The following conventions apply in this manual:
P Left and Right sides of the instrument are as viewed from the front (patient's
position) of the instrument, unless noted otherwise.
P Dimensions are given in inches unless noted otherwise.
P All tool sizes for screws and nuts mentioned in the instructions are metric unless
noted otherwise.
Field Service Bulletins are a vital element of service support. Bulletins are used to quickly convey
technical information on a variety of field service topics, including:
P instrument design changes P service manual revisions
P technical problems and P upgrade announcements/
corrections procedures
P software updates P system checkout – checklist
P new troubleshooting proce- P service disclaimer forms
dures
P problem alerts
Service bulletins are used to issue revised pages for service manual updates. Anyone on the
service bulletin distribution list will automatically receive updates to the Level 1 service manual.
Your service bulletins should be filed where easily accessible for quick reference.
NOTICE
Field Service Bulletins are Confidential and Proprietary, for the sole use of personnel
employed by Humphrey Systems, Humphrey affiliates, and authorized Humphrey
distributors.
Humphrey has a well-deserved reputation for high quality, reliable instruments,
unsurpassed in the industry. However, in the hands of an unethical competitor,
relatively minor issues discussed in service bulletins can be distorted to cause the
instrument owner unjustified concerns regarding the instrument.
As a Humphrey affiliate, employee, or distributor you are required to
handle your service bulletins as appropriate for proprietary and confi-
dential information.
A two-level service strategy is used for the Acuitus: 1) Level 1 for on-site service, and 2) Repair
Center service.
Level 1 service employs modular replacement, wherein faulty circuit boards and certain other
assemblies are replaced rather than repaired on-site. These faulty assemblies are shipped to a
Humphrey Repair Center for repairs. There are also certain procedures that require special
equipment available only at a Repair Center.
Designated Repair Centers (currently Dublin, CA and Oberkochen, Germany) are the second
level of service for the Acuitus. The Repair Centers perform major circuit board troubleshooting
and repair, plus any other service action that requires special equipment or procedures not
available in the field.
Several of the circuit boards in the Acuitus are multilayer boards and use Surface Mount
Technology (SMT) components. These boards require special equipment and techniques for
troubleshooting and repair.
The Acuitus service procedures listed below must be performed at a Humphrey
Repair Center. All other service procedures can be performed in the field.
Many of the major assemblies require special tooling and procedures for alignment. Before
loosening any part refer to table 3-1, in section 3 of this service manual, to determine the
required tooling and procedures.
Parts shown in boldface in the list below must not be loosened. These parts require alignment
with special equipment available only at the Repair Centers.
1.4 Precautions
The following precautions should be observed whenever the Acuitus is being installed or
serviced. Point out to the customer any potential hazard and the appropriate corrective action.
2. Ensure that the fuses installed in the instrument and the power table are of the proper
rating.
3. Use of an extension cord is not recommended. Doing so may compromise the safety of the
operator and/or patient.
5. If the power cord or plug on the instrument is damaged, a shock or fire hazard may result.
Do not allow continued operation of the instrument until the damaged cord or plug has
been replaced.
6. To prevent personal injury and damage to the instrument, do not place the instrument on
uneven or sloped surfaces.
Instrument Precautions
1. When the instrument is being unpacked, save the shipping materials for possible future use.
Whenever the instrument is shipped, it must be properly packed to prevent damage.
2. When spare parts are received, save the shipping materials for returning the defective
part(s), if appropriate.
3. To avoid possible damage to circuit board components, do not plug/unplug cables while
power is applied to the instrument.
4. Ensure that none of the ventilation openings in the instrument are blocked. Excessive heat
buildup within the instrument can cause instrument failures.
5. Do not flex the circuit boards. This instrument uses multilayer circuit boards. Multilayer
circuit boards are inherently susceptible to damage by excessive flexing.
6. Proper ElectroStatic Discharge (ESD) precautions must be observed whenever you are
disassembling or handling the instrument's circuitry. Many of the components are
extremely vulnerable to static discharge damage. A Field Service Static Protection Kit is
available for order and must be used for ESD protection during service of this instrument.
Refer to Appendix A for details regarding this kit.
7. Clean the instrument's exterior with a dry cloth. If necessary, a mild cleaning solution may
be used, but moisture must not be allowed inside the instrument.
The parts drawings in Section 7 of this Service Manual illustrate the internal physical layout of the
instrument. Diagrams in Section 6 illustrate the system interconnect arrangement and proper
cable routing.
The Acuitus uses a refractive method known as the “retinoscopic principle” to provide sphere
and cylinder readings. This principle can be explained in basic terms as follows: If somebody
would look at the retina of a patient from outside the eye, the portion of the retina the observer
would see would depend on the refractive error of the patient’s eye. For low myopic patients, he
would see a smaller area than for high myopic patients; for hyperopic patients, the image of the
retina would be rotated by 180 degrees compared to myopic patients. Therefore, once it is
known exactly what part of the retina can be seen from the outside, the refractive error can be
determined.
In the Acuitus, four photodiodes, shaped and arranged like four quarters of a donut (the “quad
detector”) are used to look at the retina of the patient. Once we know what part of the retina
the quad detector sees, we can calculate the spherical and cylindrical components of the patient’s
refractive error.
The chopper disk rotates at a rate of ten revolutions per second. Each revolution generates one
measurement set. Every set is checked against several criteria to ensure accurate data and to
discard bad measurements (e.g., due to patient movements, etc.). A rolling average is calculated,
consisting of the ten most recent measurements, and displayed on the screen. Measurements are
taken continuously; the system stops measuring once the joystick button is pressed and displays
the last average.
FIGURE 1-1.
The refractive error of the eye is defined for a completely relaxed crystalline lens, i.e., the patient
is looking at infinity. If the patient is looking at objects close by, the ciliary muscles increase the
refractive power of the crystalline lens in order to bring these objects into focus. A system which
shall measure the refractive error of the eye accurately, must, therefore, guarantee that the
patients’s accommodation is relaxed and the eye is looking at a distant object. The Acuitus
achieves this by means of a moving acuity target. An image of a balloon, which appears to be
coming from a distant point, is presented to the patient. In order to get the patient to look at this
image, we have to correct whatever refractive error may be present. For a nearsighted patient, a
far away image will look blurry - up to a point where the image cannot be recognized at all and
the patient doesn’t know where to look or how to accommodate. The Acuitus, therefore, uses
the first refractive measurement to move the acuity target into focus and relax the
accommodation of the patient. Once that has happened, all subsequent measurements should
be accurate.
The following descriptions are for each circuit board in the Acuitus 5000/5010 Automatic
Refractor/Keratometer. Refer to the appropriate sections in the parts section of this manual for
identification and location information for each assembly.
CPU PCB
The CPU PCB is the heart of the Acuitus system. The CPU contains the circuit elements listed in
Figure 1-2.
CPU PCB
SERIAL CONTROLLER
Keratometer Assembly
The Keratometer Assembly is mounted on the front of the model 5010 and 5015 optics head,
and performs the same functions as the LED Driver PCB does on the model 5000. In
addition to those functions, the Keratometer Assembly also contains the mirrors and LEDs
necessary to perform central and peripheral Keratometric measurements.
Detector PCB
The Detector PCB contains a custom four-segment photodiode array and the circuitry to
process the four resultant inputs created by the photodiode array. The Detector PCB
generates a signal corresponding to each of the four inputs and the signals are measured by
an analog-to-digital (A/D) convertor. The output of the A/D is read by the I/O Interface PCB.
LCD Display
The LCD display is a TFT (thin film transistor) active matrix color LCD (liquid crystal display).
The 5.5" (14 cm) LCD display area contains 320 x 240 pixels. The LCD display interfaces
with the Frame Grabber PCB via the Keypad PCB.
Backplane PCB
The Backplane PCB functions as a signal interconnect and to provide power for the CPU,
Frame Grabber, and I/O Interface PCBs. In addition, the backplane distributes power to the
floppy drive, fans, CCD camera, printer, LCD display/keypad, and Driver/Auto-Z PCB.
Power Supply
The Acuitus power supply is a 100-watt, switching type supply. The power supply is rated to
accept input voltages of 90 to 264 VAC, with input frequencies of 47 to 440 Hz.
1.8 Specifications
Dimensions
Sphere: -20D to +20D
Height: 20" 513 mm Cylinder: 0 to 7D
Width: 10" 256 mm Axis: 0E to 180E
Depth: 15" 385 mm Vertex Distance: 0.0, 10.5, 12.0, 13.5,
Weight: 40 lbs 18 Kg 15.0, 16.5 mm
Minimum Pupil Size: 3 mm
Operating Conditions
Keratometry Measurement
Operating Temperature: +10E to 40E C
Relative Humidity: 30% to 75% Range: 30 to 60D or
Atmospheric Pressure: 700 to 1060 hPa 5.60 to 11.20 mm
Axis: 0E to 180E
Peripheral K’s: 15E off visual axis
Central K’s: At visual axis
Electrical Contact Lens: 2.6 mm base leg for
43.50 dK cornea
Line Voltage: 100–120 V, 230 V
Line Frequency: 50/60 Hz
Current: 100–120 V, 2 A; 230 V, 1 A
Leakage Current: Less than 1 mA
Fuse Rating: T2.0 slow blow, 250 V,
5 x 20 mm
Output Devices
Input/Output Connections
Two RS-232 serial ports
Refraction Measurement
Notes:
Following is the System Checkout Checklist for the Acuitus 5000/5010. This checklist assumes
that all required service on the Acuitus has been completed. The checklist should be used to
perform a final operational checkout of the Acuitus prior to completion of the service call.
Many of the steps in the checklist assume that the technician is familiar with general operation of
the instrument. Complete operating instructions are contained in the Acuitus Owner's Guide.
1) Switch ON power to the unit and verify that the instrument powers up to the Main
Menu.
4) Check the functionality of the joystick. Move the optics head right/left and up/down.
Verify that the head can be secured in the down position.
6) Using the Customer Test Eye, perform a refraction and keratometry (keratometry on
model 5010 only). Verify that the measured values correspond with those marked on
the test eye.
7) Press the print button soft key. Verify that the printout is legible.
8) Switch OFF power to the unit. Install the Service Calibration Diskette and switch ON
power to the unit. Verify that the unit powers up to the calibration menu. Remove
the calibration diskette and switch OFF power to the unit.
9) Switch ON power to the unit and verify that the instrument powers up to the main
menu. Switch OFF power to the unit.
• The Eyepiece Lens has a special coating that is easily scratched, therefore careful cleaning is
imperative. The eyepiece lens should be checked regularly for dust and smudges because a
dirty eyepiece lens can cause erroneous readings.
1) Use a blower brush to gently brush off all visible dirt from the eyepiece lens.
2) Apply a small amount of lens cleaning solution on a lens tissue and gently wipe the
lens. Be careful to avoid scratching the lens.
3) Use a dry lens tissue to wipe off any excess solution from the surface of the lens.
Notes:
P Table 3-1 is the single source of reference for Removal / Adjustment requirements.
Whenever a removal procedure is performed, refer to Table 3-1 to determine which
follow-up checks/adjustments are required.
P In addition to the drawings in this section, the parts drawings in Section 7 provide
useful reference for parts identification and relative locations during removal/
replacement procedures.
P References from one procedure to another are shown in the form "(x.x)."
EXAMPLE: In the Optics Head Removal procedure, one of the steps is to "Remove
the optics head covers (3.4)." If you need further details for removing the optics head
covers, you can refer to section 3.4, and then return to the Optics Head Removal
procedure for the next step.
P Step-by-step procedures are provided when needed; but, when adequate, only
essential replacement notes are given. Simple removals that are clearly obvious are
not described in this manual.
P Unless noted otherwise, the steps for replacement (reassembly) are simply the reverse
order of the steps for removal, and are not listed. Replacement (reassembly) notes
are included as needed.
P All screws and nuts mentioned in the instructions are metric unless noted otherwise.
P In this manual, the terms left, right, front and rear of the instrument are as viewed
from the patient position, unless noted otherwise.
P Proper cable and wire routing is important. Before disassembly, observe the cable
routing and locations of the cable ties; then be sure to route the cables exactly the
same during reassembly. Be sure to replace all cable ties removed during
disassembly. Check that cables are positioned to avoid interference with moving
parts.
Any special tools or equipment required for a procedure are listed at the beginning of the
procedure.
When equipment needs to be returned to Humphrey Systems for repair, it is important that
it is properly packed for shipment, and that authorization for return is obtained before the
equipment is shipped.
When equipment is received, the original packing materials should be preserved for
possible later use. Costs to repair equipment damage caused by improper
packing for shipment to Humphrey become the responsibility of the sender.
Authorization must be obtained from Humphrey Systems before equipment is returned for
repair. A Return Goods Authorization (RGA) number is required on each return shipment
to Humphrey Systems. The procedure for obtaining an RGA number varies, depending
on your area of operation. Use the procedure that has been established by Humphrey
Systems for your operating organization.
TABLE 3-1
Adjustment / Calibration Requirements Following Replacements
This table identifies checks/adjustments that must be made following removal/replacement procedures. The
numbers listed under Follow-ups Required are keyed to the Follow-up Actions listed in the second
column. For each Assembly Replaced, the follow-up numbers are listed in the order in which they should be
performed, not necessarily in numerical sequence.
Items marked with an asterisk (*) only need to be performed if the assembly is actually being replaced, not just
removed and reinstalled. All other items need to be performed whether the assembly is replaced, or simply
removed and reinstalled.
1) Remove the 2 socket-head screws from the rear pod cover behind the LCD display (Fig. 3-
1A).
9) Tilt the unit on its side and remove the socket-head screw that secures the rear pod cover at
the base of the unit (Fig. 3-1B).
10) Disconnect the 2 ribbon cables that connect the keypad PCB to the instrument base.
11) Remove the 3 socket-head screws that secure the LCD display bracket to the rear pod
cover.
12) Disconnect the LCD display cable from the LCD display.
The LCD display cable can be disconnected by carefully lifting the upper portion of the
cable connector on the LCD display (Fig. 3-2).
13) Remove the 4 socket-head screws that secure the LCD display bracket to the LCD display.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
REMOVE SCREW
1) Remove the 2 socket-head screws from the rear pod cover behind the LCD display.
2) Tilt the unit on its side and remove the socket-head screw that secures the rear pod cover at
the base of the unit (Fig. 3-1B).
3) Disconnect the 2 ribbon cables that connect the keypad PCB to the instrument base.
4) Remove the 3 socket-head screws that secure the LCD display bracket to the rear pod
cover.
6) Remove the 4 socket-head screws that secure the keypad PCB to the rear pod cover, and
remove the keypad PCB.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
1) Using a 2.5 mm hex wrench, disengage the 2 quick disconnect latches located at the front
and rear of the optics covers by turning left or right one quarter turn (Fig. 3-3).
3) Loosen the 3 captive-screws that secure the right optics head cover to the tower
(Fig. 3-4), and remove the right optics head cover.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure that all cut cable
ties are replaced following this procedure.
2) Remove the ribbon cables connected to the LED driver PCB and detector box assembly.
3) Disconnect all of the individual optics head cables from the main wiring harness.
4) Remove the 3 socket-head screws that secure the optics head to the tower (Fig. 3-5).
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure that all cut cable
ties are replaced following this procedure.
2) Loosen the 2 socket-head screws that secure the chopper disk motor to the chopper disk
assembly (Fig. 3-7).
Note - The 2 socket-head screws are loosened to facilitate access to the 3 screws
removed in the next step. To prevent damage to the optical disk within the chopper disk
assembly, do not completely remove these screws.
3) Remove the 3 socket-head screws that secure the chopper disk assembly to the source LED
housing (Fig. 3-7).
4) Disconnect the 2 cables that connect the chopper disk assembly to the main wiring harness.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Notes:
1. The required follow-up procedures (Section 3, Table 3-1) necessary after loosening the
Target LED Assembly utilizes tooling that is not part of the standard tool list. Refer to
Appendix A before loosening the Target LED Assembly.
2. One or more cable ties must be cut during this procedure. Ensure that all cut cable ties
are replaced following this procedure.
2) Remove the 2 socket-head screws the secure the target LED assembly to the optometer
guide bracket.
3) Disconnect the target LED assembly cable from the LED driver PCB.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure that all cut cable
ties are replaced following this procedure.
2) Remove the nut that secures the optometer motor shaft to the optometer guide bracket.
3) Remove the 2 socket-head screws the secure the optometer motor to the optometer motor
bracket.
4) Disconnect the optometer motor’s cable from the main wiring harness.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Notes:
1. The required follow-up procedures (Section 3, Table 3-1) necessary after loosening the
Target LED Assembly utilize tooling that is not part of the standard tool list. Refer to
Appendix A before loosening the Optometer Edge Detector Assembly.
2. One or more cable ties must be cut during this procedure. Ensure that all cut cable ties
are replaced following this procedure.
2) Remove the socket-head screw and nut that secure the optometer edge detector to the
optometer edge detector bracket.
3) Disconnect the optometer edge detector’s cable from the main wiring harness.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
2) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
5) Remove the 3 socket-head screws that secure the LED Driver PCB to the optics head
assembly.
6) Remove the 4 socket-head screws from the back of the LED Driver PCB that secure the
calibration block to the LED Driver PCB.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
2) Using the joystick, adjust the optics head to its uppermost position.
3) Working from the patient side of the instrument, unlatch the joystick covers by lifting both
covers up.
4) Separate the two joystick covers by pressing down on the front joystick cover while holding
the rear joystick cover up.
5) Remove the front joystick cover by sliding off to one side of the instrument.
6) Remove the rear joystick cover by lifting over the joystick and sliding off the rear of the
instrument.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure that all cut cable
ties are replaced following this procedure.
2) Remove the 2 socket-head screws that secure the joystick weight to the chassis.
3) Remove the 3 socket-head screws that secure the joystick assembly to the chassis.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
2) Using the joystick, adjust the optics head to its lowest position.
3) Remove the 2 retaining rings that secure the two joystick shafts and pulleys to the tower.
Refer to the Tower Assembly parts list and drawing in section 7 of this manual for parts
identification.
4) Remove the 2 joystick shafts and pulleys from the tower frame.
After removing the retaining rings (step 3 of this procedure), the joystick shafts and pulleys
can be pushed down, and then pulled out of the tower frame towards the center of the
instrument.
6) Slip the joystick belt under the large pulley, and out of the instrument.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure all cut cable ties
are replaced following this procedure.
2) Remove the socket-head screw that secures the edge detector to the bearing retainer block.
3) Disconnect the edge detector’s cable from the main wiring harness.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
Note - One or more cable ties must be cut during this procedure. Ensure that all cut cable
ties are replaced following this procedure.
1) Remove the 8 socket-head screws securing the base cover to the instrument chassis (Fig. 3-
9).
2) With one hand on the front of the unit and the other on the rear, lift off the base cover and
place the base cover with tower/optics head on its side next to the instrument chassis.
3) Cut the cable ties securing the excess cable lengths to the bottom of the base cover.
4) Lift up the base cover with tower/optics head, and place it upright next to the instrument
chassis.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
3) Loosen the 2 socket-head screws on the CPU retainer bracket and let the bracket hang
down below the CPU PCB (Fig. 3-10).
Note - It is not necessary to completely remove the CPU retainer bracket during this
procedure.
4) Remove the socket-head screw that secures the CPU PCB to the instrument chassis (Fig. 3-
10).
Replacement Notes:
P To ensure that the factory calibration is not lost, the flash disk (32 pin DIP on the CPU
PCB) from the original CPU PCB must be transferred to the new CPU.
P If the original flash disk is inoperative, refer to the procedures listed in the Appendix for
restoring the operating system, the application software, and the calibration constants on
the new flash disk.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
3) Remove the 3 socket-head screws that secure the Frame Grabber PCB to the instrument
chassis.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
2) Remove the 4 socket-head screws that secure the main fan to the instrument chassis.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
3) Remove the 2 socket-head screws that secure the I/O Interface PCB to the I/O interface
bracket (Fig. 3-11).
Follow-up Checks/Adjustments:
Refer to Table 3-1.
3) Remove the 4 socket-head screws that secure the interface panel to the instrument chassis
(Fig. 3-12).
4) Remove the 3 socket-head screws that secure the I/O interface bracket to the instrument
chassis.
5) Remove the 4 socket-head screws that secure the disk drive bracket to the instrument
chassis (Fig. 3-13).
6) Disconnect all cables from the floppy disk drive, and lift the floppy disk drive with disk drive
bracket out of the instrument.
7) Remove the 4 socket-head screws that secure the disk drive bracket to the floppy disk drive.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
2) Remove the 4 socket-head screws that secure the power supply to the instrument chassis.
3) Remove the 4 socket-head screws that secure the interface panel to the instrument chassis
(Fig. 3-12).
5) Disconnect the power supply’s fan cable from the Backplane PCB.
Follow-up Checks/Adjustments:
Refer to Table 3-1.
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
4.2 Related Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
4.2.1 Calibration Software Setup For 17" VGA Display . . . . . 4 - 3
4.2.2 Write Enabled Service Calibration Disk . . . . . . . . . . . . . 4 - 3
4.3 Camera Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
4.4 Source LED Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 9
4.5 Calibration Block Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 13
4.6 Chopper Motor Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
4.7 Refractor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
4.8 Keratometry Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 20
4.9 Optometer Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
4.10 Auto-Z Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
Notes:
4.1 Introduction
This section contains the procedures for adjustment and calibration of the Acuitus.
Procedure
1) Make a duplicate copy of the Service Calibration Disk.
2) Using the Windows Explorer program in Windows 95, rename the file LCRCAL.INI to
LCRCALF.INI
The LCRCAL.INI file can be found in the A:\LCRCAL directory of the Service Calibration
Disk copy.
3) Using the Windows Explorer program in Windows 95, rename the file LCRCALM.INI to
LCRCAL.INI
The LCRCALM.INI file can be found in the A:\LCRCAL directory of the Service Calibration
Disk copy.
4) Label the copied disk “Service Calibration Disk/17" VGA Display Copy”.
The Service Calibration Disk must be write-enabled to successfully complete the calibration
procedures listed in this section. Ensure that the tab on the floppy disk is correctly positioned on
the Service Calibration Disk prior to inserting into the floppy disk drive.
Purpose
To align the CCD camera to the quad detector.
Procedure
1) Open the base unit (3.15).
3) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
5) Replace the Acuitus Video/Frame Grabber PCB with the Atlas Video Frame Grabber PCB
(3.17). Leave the CPU PCB removed to facilitate installation of the Atlas Video Cable
(next step).
6) Plug the Atlas Video Cable into connector J16 on the Atlas Video/Frame Grabber PCB.
Note - Connector J16 is not keyed. Ensure that pin 1 on the cable is aligned with pin 1
on J16.
7) Reinstall the CPU PCB. Leave the base cover with tower/optics head removed and side-
by-side with the instrument chassis.
10) Screw the Camera Focus Tool into the eyepiece on the front of the optics head (Fig. 4-1).
FIGURE 4-1.
11) Mount the Camera Focus LED Tool on the Camera Focus Tool (Fig. 4-1).
12) Disconnect the source LED cable from connector JSRC on the LED Driver PCB.
13) Disconnect the visual acuity target LED cable from connector JVAT on the LED Driver
PCB, and connect the Camera Focus LED Tool cable to connector JVAT.
14) Disconnect the CCD camera’s video cable from the instrument’s wiring harness, and
connect it to the Monitor Video Cable Adapter.
15) Using the Coax Cable, connect the Monitor Video Cable Adapter to the color monitor.
16) Insert the “17" VGA Display Copy” of the Service Calibration Disk into the floppy disk
drive, and then switch ON the instrument.
The Calibration Startup Screen will be displayed on the VGA monitor after the system
boots from the calibration disk.
18) Observe the cross-hairs image on the color monitor, and adjust the position of the camera
mount until the best possible focus of the image is achieved.
(continued)
20) Remove the Camera Focus and Camera Focus LED Tools from the instrument, leaving
both the visual acuity target LED and source LED cables disconnected.
21) Remove the Monitor Video Cable Adapter and reconnect the CCD camera’s video to the
instrument’s wiring harness.
22) Slide the foam cover forward to expose the slot in the detector tube (Fig. 4-3).
24) Screw the Mirror Tool on the front of the optics head (Fig. 4-4).
26) Apply power to the fiber optic probe, and adjust the position of the mounting bracket on
the detector box as required to obtain the best illumination of the quad detector in the
VGA display.
28) Adjust the position of the camera PCB until the image of the 4 quad detector segments are
centered around the small green circle in the VGA display (Fig. 4-5).
30) Remove the Detector Illumination Tool, and slide the foam cover over the slot in the
detector tube (Fig. 4-3).
32) Reconnect the source LED and visual acuity cables to connectors JSRC and JVAT on the
LED Driver PCB.
33) Slide the foam cover back to its original position on the detector tube.
FIGURE 4-2.
FIGURE 4-3.
FIGURE 4-4.
FIGURE 4-5.
Purpose
To align the source LED output to the quad detector and camera.
Procedure
1) Open the base unit (3.15).
3) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
5) Replace the Acuitus Video/Frame Grabber PCB with the Atlas Video Frame Grabber PCB
(3.17). Leave the CPU PCB removed to facilitate installation of the Atlas Video Cable
(next step).
6) Plug the Atlas Video Cable into connector J16 on the Atlas Video/Frame Grabber PCB.
Note - Connector J16 is not keyed. Ensure that pin 1 on the cable is aligned with pin 1
on J16.
7) Reinstall the CPU PCB. Leave the base cover with tower/optics head removed and side-
by-side with the instrument chassis.
9) Screw the Mirror Tool on the front of the optics head (Fig. 4-4).
(continued)
10) Loosen the set screw that secures the source LED to the source LED mounting bracket,
and slide the source LED out of the mounting bracket (Fig. 4-6).
11) Disconnect the chopper disk motor from the main wiring harness.
12) Disconnect the CCD camera’s video cable from the instrument’s wiring harness, and
connect it to the Monitor Video Cable Adapter.
13) Using the Coax Cable, connect the Monitor Video Cable Adapter to the color monitor.
14) Insert the “17" VGA Display Copy” of the Service Calibration Disk into the floppy disk
drive, and then switch ON the instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
16) Look in the source LED mounting hole and make sure that the chopper disk is not blocking
the light path.
If necessary, temporarily reconnect the chopper disk motor to the main wiring harness to
change the position of the chopper disk. This may need to be performed several times
until a clear portion of the disk is positioned in front of the source LED mounting hole.
17) Insert the source LED into the source LED mounting hole, leaving the set-screw loose.
18) Disconnect the source LED and visual acuity LED cables from connectors JSRC and JVAT
on the LED Driver PCB.
19) Connect the source LED cable to the Source LED Dimmer Tool.
20) Connect the Source LED Dimmer Tool to connector JSRC on the LED Driver PCB.
21) Observe the image of the source LED on the color monitor, and adjust the brightness of
the source LED with the control on the Source LED Dimmer Tool.
Note - The brightness of the source LED must be dimmed to an adequate level to
determine proper focus in the next step of this procedure.
22) Observe the image of the source LED on the color monitor, and adjust the position of the
source LED in the source LED mount hole until the best possible focus of the image is
achieved.
24) Remove the Monitor Video Cable Adapter and reconnect the CCD camera’s video to the
instrument’s wiring harness.
25) Loosen the 2 socket-head screws that secure the source LED mount to the optics head (Fig.
4-7).
26) Adjust the position of the source LED mount until the image of the source LED is centered
within the cross-hairs on the VGA monitor.
27) Tighten the 2 socket-head screws on the source LED mount (Fig. 4-7).
28) Reconnect the chopper disk motor to the main wiring harness.
30) Reconnect the source LED and visual acuity LED cables to connectors JSRC and JVAT on
the LED Driver PCB.
FIGURE 4-6.
FIGURE 4-7.
Purpose
To align the calibration block to the optics path. The calibration block is used as a mount for the
Carousel Mount Tool.
Note - This procedure must be performed on a level surface. Verify that the surface, which
supports the instrument, is level using a standard bubble level.
Procedure
1) Remove the optics head covers (3.4).
2) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
4) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-8).
5) Mount the Tension Tool inside the Carousel Mount Tool by screwing the tool into the
instrument’s eyepiece (Fig. 4-8).
6) Mount the Level Tool on the Carousel Mount Tool (Fig. 4-8).
7) Loosen the 3 screws that secure the calibration block to the front of the optics head (Fig. 4-
9).
8) Disconnect the source LED cable from the connector JSRC on the LED Driver PCB.
9) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
(continued)
11) Adjust the calibration block up/down and right/left as required to center the image in the
camera display (Fig. 4-10).
12) Rotate the calibration block clockwise or counterclockwise as required to ensure that the
bubble in the level tool is centered between the black lines.
Note - It may be necessary to repeat steps 9 and 10 as required to ensure the calibration
block is both centered and level.
13) Tighten the 3 screws that secure the calibration block to the front of the optics head (Fig. 4-
9).
14) Reconnect the source LED cable to connector JSRC on the LED Driver PCB.
Purpose
To determine necessary software offsets to compensate for timing differences between the four
detector channels.
Procedure
1) Remove the optics head covers (3.4).
2) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
4) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-11).
5) Mount the Carousel Calibration Tool on the Carousel Mount Tool (Fig. 4-11).
6) Using the jumper cable, connect the Flashing LED (part of the Carousel Calibration Tool)
to the test connection (JLED) provided on the main wiring harness.
7) Align the Flashing LED portion of the Carousel Calibration Tool with the instrument
eyepiece.
8) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
9) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
10) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit,and switch OFF the instrument when the DOS prompt appears.
12) From the Calibration Startup Screen, select Manual, Refract, Chopper, Start.
The system will now collect data. When the readings stabilize, the message “Timing OK”
should appear.
13) When the message “Timing OK” appears, select Next, Exit, Save LCR.
An updated copy of the calibration file is copied to the Acuitus C: (disk-on-chip) drive.
14) When the system indicates that the calibration data has been saved, select Cancel to return
to the Calibration Startup Screen.
Purpose
To determine necessary software offsets to provide valid sphere and cylinder readings.
Procedure
1) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
3) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-11).
4) Mount the Carousel Calibration Tool on the Carousel Mount Tool (Fig. 4-11).
5) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
6) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
7) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit,and switch OFF the instrument when the DOS prompt appears.
10) At the Adjust Calibration Lenses screen, adjust each of the nominal values to the actual
value labeled for the lens on the Carousel Calibration Tool (Figs. 4-12 and 4-13).
Use the up/down arrow keys to select the value to be adjusted. Use the plus or minus
softkeys to change the selected value.
Note - The cylinder lens (“CYL”) is always set to 90 degrees.
11) After adjusting the lens values, press the joystick button.
12) Align each lens as requested with the instrument eyepiece by rotating the carousel drum,
and then press the joystick button.
13) The system will collect data for the current lens and then ask for the next lens to be read.
14) When the system requests the “axis correction lens”, position the lens labeled CYL, and
press the joystick button.
15) When the message “Press Joystick To See Calibration Constants” appears, press the
joystick button.
17) When the system indicates that the calibration data has been saved, select Cancel to return
to the Calibration Startup Screen.
Purpose
To calibrate the keratometry function of the instrument.
Procedure
1) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
3) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-11).
4) Mount the Carousel Calibration Tool on the Carousel Mount Tool (Fig. 4-11).
5) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
6) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
7) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit,and switch OFF the instrument when the DOS prompt appears.
10) Align the R8.00 lens as requested with the instrument eyepiece by rotating the carousel
drum, and then select LED Bal.
The instrument will take several readings on the R8.00 lens.
12) When requested by the system, align the CYL lens with the instrument eyepiece by rotating
the carousel drum, and then select Axis.
The system will take several readings on the CYL lens.
14) The next screen will show the current setup for the three Keratometry lenses (Fig. 4-14).
Adjust each of the nominal values to the actual value labeled for the lens on the Carousel
Calibration Tool. Figure 4-14 is shown with the nominal value for each lens.
Use the up/down arrow keys to select the value to be adjusted. Use the plus or minus
softkeys to change the selected value.
15) After adjusting the lens values, press the joystick button.
16) When requested by the system, align the R9.65 lens with the instrument eyepiece by
rotating the carousel drum, and then select Start QA.
The system will take a reading on the R9.65 lens.
18) When requested by the system, align the R8.00 lens with the instrument eyepiece by
rotating the carousel drum, and then select Start QA.
The system will take a reading on the R8.00 lens.
20) When requested by the system, align the R6.15 lens with the instrument eyepiece by
rotating the carousel drum, and then select Start QA.
The system will take a reading on the R6.15 lens.
22) When requested by the system, align the CYL lens with the instrument eyepiece by rotating
the carousel drum, and then select Start QA.
The system will take several readings on the CYL lens.
23) When the system indicates that CYL lens passed, select Next, Continue, Cancel, Save
LCR.
An updated copy of the calibration data is copied to the Acuitus C: (disk-on-chip) drive.
24) When the system indicates that the calibration data has been saved, select Cancel to return
to the Calibration Startup Screen.
Purpose
To determine and store the necessary calibration offset value to present a focused Visual Acuity
Target to the patient during refractions.
Note - The NTSC Camera, the 110 V Power Supply, and the NTSC Monitor must be used
together as a set. The PAL Camera, the 220 V Power Supply, and the PAL Monitor must be
used together as a set.
Procedure
1) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
3) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-15).
4) Mount the Optometer Alignment Camera on the Carousel Mount Tool (Fig. 4-15).
5) Connect the Camera Power Supply to the camera and apply power to the power supply
(Fig. 4-16).
6) Using the Coax Cable, connect the video output of the camera to the video input on the
monitor (Fig. 4-16).
7) Disconnect the source LED cable from the connector JSRC on the LED Driver PCB.
8) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
(continued)
9) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
10) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit, and switch OFF the instrument.
12) From the Calibration Startup Screen, select Manual, Refract, Optom.
13) Observe the balloon target image on the monitor display, and using the Up/Down keys,
adjust the position of the optometer until the best possible focus of the image is achieved.
15) When the system indicates that the calibration data has been saved, select Cancel to return
to the Calibration Startup Screen.
16) Reconnect the source LED cable to connector JSRC on the LED Driver PCB.
FIGURE 4-15.
FIGURE 4-16.
Purpose
To calibrate the Auto-Z function of the instrument.
Procedure
1) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
3) Mount the Carousel Mount Tool to the calibration block on the front of the optics head
(Fig. 4-11).
4) Mount the Carousel Calibration Tool on the Carousel Mount Tool (Fig. 4-11).
5) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
6) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
7) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit,and switch OFF the instrument when the DOS prompt appears.
10) At the Calib Z screen, align each lens as requested with the instrument eyepiece by rotating
the carousel drum, and then press the joystick button.
The system will collect data for the current lens and then ask for the next lens to be read.
11) At the Test Z screen, align each lens as requested with the instrument eyepiece by rotating
the carousel drum, and then press the joystick button.
The system will collect data for the current lens and then ask for the next lens to be read.
12) When the message “Press Continue” appears, select Continue, Exit, Save LCR.
An updated copy of the calibration file is copied to the Acuitus C: (disk-on-chip) drive.
13) When the system indicates that the calibration data has been saved, select Cancel to return
to the Calibration Startup Screen.
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
5.2 General Guidelines for Assembly Level Troubleshooting . . . . . . . . 5 - 3
Notes:
5.1 Introduction
The troubleshooting aids in this section will help you to diagnose operating difficulties and system
malfunctions in the Acuitus 5000/5010. As Field and Repair Center experience identifies
recurrent problems and appropriate diagnostic processes, this section will be expanded to include
additional troubleshooting aids.
On-site assembly level troubleshooting poses unique and complex challenges to the Field Service
Representative. In addition to identifying and replacing the faulty assembly, the Field Service
Representative must often travel great distances to the customer location, resolve operator errors,
and identify power and environmental causes of problems while satisfying the customer's
perception of quality service.
A structured approach to resolving field problems can greatly increase customer satisfaction and
the Field Service Representative's effectiveness. There are seven major tasks in resolving field
equipment problems:
Prior to Going On-Site While On-Site
1. Obtain a Service History for the 4. Have the Operator Demonstrate the
Instrument Problem
2. Query the Operator of the 5. Resolve the Obvious
Instrument
6. Substitute the Failing Assembly
3. Check for Field Service Bulletins
7. Confirm the Repair
that Address the Problem
Each of these tasks is explained below.
While On-Site:
4) Have the Operator Demonstrate the Problem.
This is very important. Before doing anything, always have the operator attempt to
demonstrate the problem to you. By having the operator demonstrate the problem,
you may solve the problem immediately, or avoid some incorrect assumptions. Having
the operator demonstrate the problem will ALWAYS result in one of the following:
P Reveal Operator Error - You may immediately see that the cause of the
problem is operator error. You then have the opportunity to train the
customer in the proper use of the instrument, instead of spending time
troubleshooting a nonexistent instrument problem.
P Demonstrate specific problem sequence - The operator will show you
the specific sequence of events needed to reproduce the problem, reducing
your diagnostic time.
P Help Form Ideas - The operator may not be able to duplicate the
problem. However, the sequence of steps used to demonstrate the problem
will give you ideas on how to pursue the problem in other ways. In this case
you have the opportunity to explain to the customer the difficulty in resolving
a nonduplicable problem.
5) Resolve the Obvious.
Obvious mechanical or physical defects should be resolved when discovered. Often,
resolving the obvious will lead you to the real problem.
The frequency of the problem helps you determine how long it will take to verify a repair. A
good rule of thumb when troubleshooting intermittent problems is: 2 times the frequency of
the symptom gives an 80% confidence level of the repair.
EXAMPLE:
Problem
The problem appears only about once an hour. The sequence of events needed
to reproduce the problem leads you to believe a circuit board is the likely
cause.
Solution
The suspected circuit board is replaced. To provide an 80% confidence level
that the circuit board resolved the problem, the instrument will need to be
tested for 2 hours without failing.
important that before replacing an assembly, good solid rationale for its replacement be
developed and a backup plan devised in case replacement of the assembly does not affect the
symptom. It is best NOT to replace any hardware until solid rationale for replacement is
clearly evident.
1) Analyze the service history in detail. Look for trends or the possibility of a previous
repair inducing the problem.
2) Query the operator(s) at length for any other clues such as:
P Times of failures;
P More failures with certain operators.
9) Document the customer service report. Record what you feel should happen if the
problem returns. This will help a different Field Service Representative in the
resolution of the problem.
11) Educate the customer on logging pertinent information that may benefit a Field
Service Representative should the problem return. Often, engaging the customer
in the pursuit of a solution emphasizes the difficulty of the situation. In some
instances, the customer may welcome the feeling of being involved in resolution of
the problem.
Notes:
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
7.2 Replacement Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
7.3 Parts Orders — U.S. Domestic Service Operations . . . . . . . . . . . 7 - 4
7.4 Parts Orders — International Service Operations . . . . . . . . . . . . . 7 - 5
7.5 Returning Instruments / Defective Parts . . . . . . . . . . . . . . . . . . . . . 7 - 5
7.5.1 Equipment Return Authorization . . . . . . . . . . . . . . . . . . 7 - 5
7.5.2 Packing the Instrument for Shipment . . . . . . . . . . . . . . . 7 - 5
7.5.3 Returning Defective Parts . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
7.6 Recommended Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 6
7.7 Level 1 Parts Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 7
7.7.1 Accessories/Supplies/Consumables . . . . . . . . . . . . . . . . . 7 - 7
7.8 Illustrated Parts Breakdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 9
7.8.1 Final Assembly – 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
7.8.2 Final Assembly – 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 12
7.8.3 Model 5015 Chinrest Assembly -1 . . . . . . . . . . . . . . . . . 7 - 14
7.8.4 Model 5015 Chinrest Assembly -2 . . . . . . . . . . . . . . . . . 7 - 16
7.8.5 Tower Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 18
7.8.6 Tower Support Assembly . . . . . . . . . . . . . . . . . . . . . . . 7 - 20
7.8.7 POD Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 22
7.8.8 Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 24
7.8.9 Optics Head – 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 26
7.8.10 Optics Head – 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 28
7.8.11 Optics Head – 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 30
Notes:
7.1 Introduction
This section contains Level 1 parts lists and associated information for the Acuitus Models 5000
and 5010. It contains instructions for ordering parts and returning defective parts.
Ú Ú Ú Ú Ú Parts Notice Ú Ú Ú Ú Ú
The parts in the following list are kept in stock and are generally always available for immediate
shipment as replacement parts. Any part not in this list is subject to delays in delivery. These
parts, as well as all other parts in the instrument, are listed in the Level 1 Parts Lists and the
Illustrated Parts Breakdown sections of this manual.
Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú Ú
Part No. Description
14449 Cord, Power, Hospital Grade (100 – 120 V)
21455 Fuse, Metric, 2A / 250 V, SB
22581 Cord, Power, Euro (220 – 240 V)
29244 Floppy Drive, 3½", 2 MB
43042-320 Belt, Timing, 320 Teeth
43836 SIMM, RAM, 2M x 36, 70ns
46370 PCB, I/O Interface (models 5000/5010)
46480 PCB Assy, LED Driver (model 5000)
46510 Motor, Optometer
46511 Chopper Disk Assembly
46513 CCD Camera Assembly (models 5000/5010)
46570 PCB Assy, Backplane
46610-4 PCB Assy, CPU (Includes 4MB Flash Disk)
46611 LCD Display, 5.5", Color
49756 Power Supply
46730 Cable Assy, Keypad/LCD Display
46810 Tube, Eyepiece
46811 Window, Red, Front
(continued)
Spare parts may be ordered monthly or weekly by using the standard parts order form available
from Service Parts Stores. Parts needed overnight can be ordered by phone from the Parts
Department. The cost of shipping parts for next day delivery is very high and should be used
only in emergencies.
The Parts Department phone number is 877-HUMPHREY (877-486-7473).
There is a $25.00 minimum for customers ordering parts or supplies. Customers are also billed
for shipping charges (most items are shipped UPS).
For International Service Operations, please use the ordering procedures that have been
established for your area of operations, and which meet the requirements of the Humphrey
International Parts Department.
When equipment needs to be returned to Humphrey Systems for repair, it is important that it is
properly packed for shipment, and that authorization for return is obtained before the equipment
is shipped.
When you return parts or assemblies to the Repair Center, attach a paper tag to each assembly.
On the tag, list the following information:
P the serial number of the instrument from which the part was removed
P the date of removal
P the problem (if you could not find the problem, describe what the customer reported,
and add "CND" - Can Not Duplicate)
P the part number of the part/assembly being returned
P your territory number (U.S. Domestic Reps)
This information will enable the Repair Center to make repairs as efficiently as possible. (For U.S.
Domestic Reps, tags are available from the Parts Department.)
When you receive spare parts from the Parts Department, save the packaging material for
returning the defective part for repair. Many of the parts are delicate and expensive; extra
attention to packaging will pay off in less damage to parts during shipment.
Parts that are recommended as spares for supporting field repairs on the Acuitus are not
indicated in this manual. Instead, please refer to service bulletin ACT-002.
Sparing of small hardware (screws, washers, etc.) is necessary, but is left to the discretion of the
individual service technician.
7.7.1 Accessories/Supplies/Consumables
Part No. Description
14449 Cord, Power, Hospital Grade (100 – 120 V)
21455 Fuse, Metric, 2A / 250 V, SB
22581 Cord, Power, Euro (220 – 240 V)
43836 SIMM, RAM, 2M x 36, 70ns
46822 Paper, Thermal Printer, 60 mm x 30 m, RW, 1 roll
47511 Dust Cover, Acuitus
47546 Eye Piece Assy
47344 Contact Lens Holder
47603 Lens Cloth, 5 x 7, Microfiber
47629-4 IC, Flash Disk, 4MB
47661 Card, Reorder
47976 User Guide, English
48005 Paper, Thermal Printer, Adhesive, 60 mm x 10.7m, RW, 1 roll
The drawings on the following pages provide illustrated parts breakdowns of the instrument.
The parts drawings are keyed by item numbers to the associated parts lists.
Parts listed without a part number are not available separately — order the next higher assembly.
For part numbers of individual cables, refer to the System Interconnect diagram
in Section 6 (Figure 6-1). Cables that do not have a part number listed in
Figure 6-1 are supplied only as part of an associated assembly shown in this
parts section.
7.8.8 Computer
Item Part No. Description
( 7.8.8 Computer )
Notes:
The tools and test equipment required for servicing the Acuitus 5000/5010 are listed below.
Additional detail regarding the special tools and test equipment is provided in section A.2.
This is a 3.5" floppy diskette that contains calibration software and is required when performing
specific Acuitus calibrations. With the instrument power switched Off, the diskette is inserted into
the instrument’s floppy drive and the power is switched On. After a delay, the Calibration
Startup Screen will be displayed on the LCD.
The Carousel Mount Tool is a cylinder with open sides, open ends and protruding IR windows.
After removal of the head covers, retaining ring, and IR window, the Carousel Mount Tool
attaches at three points to the front of the optics head. The tool acts as a stand, supporting either
the Level Tool or the Carousel Calibration Tool. Figure 4-1 illustrates the correct positioning of
the Carousel Mount Tool.
The Tension Tool is a cylinder with open ends, machined threads and four plunger springs. The
Tension Tool is used with the Level Tool and Carousel Mount Tool when centering the
Calibration Block to the optics path. The Tension Tool is screwed into the instrument’s eyepiece
after the Carousel Mount Tool is mounted. Figure 4-1 illustrates the correct positioning of the
Tension Tool.
The Level Tool is a disk with a small bubble level at the top and five pinhole openings at its
center. The Level Tool is used with the Tension Tool and Carousel Mount Tool when centering
the Calibration Block to the optics path. The Level Tool attaches to the Carousel Mount Tool
after the Tension Tool has been mounted. Figure 4-1 illustrates the correct positioning of the
Level Tool.
The Carousel Calibration Tool is a drum with several test lenses positioned along its outside edge.
Each lens can be positioned in front of the instrument’s eyepiece by simply rotating the drum.
The test lenses are used during both calibration and calibration verification. The flashing LED
required for the Chopper Motor Offset Calibration is also part of the Carousel Calibration Tool.
The Carousel Calibration Tool attaches to the Carousel Mount Tool. Figure 4-4 illustrates the
correct positioning of the Carousel Calibration Tool.
A Customer Test Eye is shipped with every Acuitus. The test eye is used to verify the Sphere and
Cylinder readings. The Customer Test Eye is also used in the System Checkout Checklist
Procedure (2.1). Refer to Figure A-1 for proper positioning of the test eye on the instrument’s
forehead rest.
The LED Jumper Cable is used to connect the flashing LED on the Carousel Calibration Tool to
the test connection (JLED) provided on the main wiring harness.
The Optometer Alignment Camera, Camera Power Suppy, Color Monitor and Coaxial Cable
are used to view the visual acuity target during the Optometer Alignment procedure. Figures 4-7
and 4-8 illustrate the correct positioning and connection of these tools.
The Static Protection Kit is essential for protection of the Acuitis against electrostatic discharge
whenever you are disassembling or handling Acuitus circuitry. The kit contains the following
items:
The static protection kit must be properly grounded, and you must be wearing the wrist cuff
whenever working on the Acuitus with the covers removed.
1) Connect the bench mat alligator clip to a known good electrical ground point.
Any unpainted screw or chassis surface on the Acuitus is an acceptable ground point,
provided that the Acuitus has its power cord connected and is plugged
into a grounded power receptacle.
3) Connect the wrist cuff to the snap clip on the banana plug lead.
4) Adjust the wrist cuff to fit snugly around your wrist. The wrist cuff should always be
worn when working on the Acuitus with the cover removed.
Purpose
The operating system software must be restored after replacement of the flash disk.
Procedure
1) Insert the operating system software installation diskette into the floppy drive.
The system will boot from the operating system installation diskette and load the operating
system software.
3) When the system indicates that the installation is complete, remove the operating system
software installation diskette, and switch the unit OFF.
Purpose
This procedure provides a method of installing or updating the System Software on the flash
disk.
Note - If the flash disk was replaced. The Restoring The Operating System Procedure
(Appendix B) must be performed prior to this procedure.
Procedure
1) Insert the application software installation diskette into the floppy drive.
The system will boot from the application software installation diskette and load the
application software.
3) When the system indicates that the installation is complete, remove the application software
installation diskette, and switch the unit OFF.
The factory calibration constants must be restored to the flash disk after replacement of the flash
disk or optics head.
Notes:
1. If the flash disk has been replaced, the operating software and system software must be
restored to the new flash disk prior to this procedure. Refer to appendix B.
2. A Service Calibration Diskette with the appropriate factory calibration constants is
provided with the replacement optics head.
3. To obtain a copy of the Service Calibration Diskette with the appropriate factory
calibration constants, contact Humphrey Systems Technical Support, and provide the
serial number of the optics head.
Procedure
1) Insert the appropriate Service Calibration Diskette into the floppy drive.
2) Switch ON the power to the unit. Wait for the calibration menu to be displayed.
3) Select Manual, then Save LCR. The calibration data will be read from the floppy drive
and will be written to the flash disk.
The forms in this section should be used as masters for making additional copies.
# Disclaimer Form
Owner: Date:
Location:
Acuitus Serial #:
At your request, Humphrey Instruments has agreed to service the Acuitus 5000/5010 described
above. The possibility exists that, during the course of the servicing, data which has previously been
entered into the Acuitus may be erased or otherwise adversely affected. Therefore, Humphrey
Instruments cannot guarantee that your data will not be lost or altered and has no obligation or
responsibility to reconstruct data which may have to be re-entered into the Acuitus for any reason.
Please sign below to indicate that you have read this disclaimer, agree to be bound by it, and
authorized the Humphrey Instruments service technician to provide the requested service.
Signature
Print Name
Date:
Revision B 0999
System Checkout Checklist
Following is the System Checkout Checklist for the Acuitus 5000/5010. This checklist assumes
that all required service on the Acuitus has been completed. The checklist should be used to
perform a final operational checkout of the Acuitus prior to completion of the service call.
Many of the steps in the checklist assume that the technician is familiar with general operation of
the instrument. Complete operating instructions are contained in the Acuitus Owner's Guide.
1) Switch ON power to the unit and verify that the instrument powers up to the Main
Menu.
4) Check the functionality of the joystick. Move the optics head right/left and up/down.
Verify that the head can be secured in the down position.
6) Using the Customer Test Eye, perform a refraction and keratometry (keratometry on
model 5010 only). Verify that the measured values correspond with those marked on
the test eye.
7) Press the print button soft key. Verify that the printout is legible.
8) Switch OFF power to the unit. Install the Service Calibration Diskette and switch ON
power to the unit. Verify that the unit powers up to the calibration menu. Remove
the calibration diskette and switch OFF power to the unit.
9) Switch ON power to the unit and verify that the instrument powers up to the main
menu. Switch OFF power to the unit.
Revision B 0999
SERVICE
BULLETIN
Instrument: Acuitus 5000/5010/5015 No.: ACT-ID01
Subject: Service Bulletin Index Date: 28 Jun 00
This bulletin indexes all Acuitus and General (GEN) service bulletins issued to date. The bulletins
are indexed in two ways: by number and by instrument subassembly.
The index will be updated periodically. Meanwhile, we recommend that you manually list in the index
each new bulletin you receive.
INDEX BY NUMBER
NUMBER SUBJECT
INDEX BY SUBASSEMBLY
DOCUMENTS
The Acuitus 5000/5010/5015 Service Manual - Rev. C (P/N 47881) is now available.
Distribution
# One copy of the new manual is being sent to each Zeiss Humphrey Systems Repair Center.
# Additional copies of the service manual are available by standard parts order.
This bulletin provides revision and date of release information for software used on the Acuitus. The part
numbers for the current revision of software are listed in each section. It is recommended that only the
most current version of software be used when servicing the instrument.
Application Software
! Transmit measured data using the Packet Exchange Protocol ( PEP ) after the operator
presses print
! The instrument will display a dialog while the Acuitus is attempting to transmit data.
rd
! Allows operator to set parameters for Serial Port #1 via the 3 system set up menu.
! If the data transmission is unsuccessful, the Acuitus will now display a message.
! Changes Video Offsets to properly center video image for both old and new frame grabber
boards.
! Fixed a bug that caused the 5000 Model to display international characters incorrectly
! Screen Saver activation time was reduced from 15 to 5 min.
An update package of revised pages for the Acuitus Model 5000/5010/5015 Service Manual has
been produced to make a variety of corrections and additions to the manual. The update package
updates the Revision B manual (P/N 47881) to Revision C. Please remove the corresponding
pages within the manual and replace them with these updated pages.
Rev C changes to the Acuitus Service Manual cover a broad Replace selected pages in your Acuitus Service Manual with
range, including: the corresponding Rev C update pages enclosed.
The latest Acuitus calibration software (revision A14 or later) allows calibration of instruments equipped
with both the original and aperture enhanced optics heads. This bulletin provides the updated Refractor
Calibration procedure required. For more information concerning the Aperture enhancement, refer to
service bulletin ACT-005.
Important changes are noted with a vertical bar in the left margin of the procedure. These changes will
be implemented in the next version of the Acuitus Service Manual.
Purpose
To determine necessary software offsets to provide valid sphere and cylinder readings.
Procedure
1) Unscrew and remove the retaining ring that secures the IR window to the front of the optics
head assembly.
3) Mount the Carousel Mount Tool to the calibration block on the front of the optics head (Fig.
4-11).
4) Mount the Carousel Calibration Tool on the Carousel Mount Tool (Fig. 4-11).
5) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the
instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration
disk.
6) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
7) When the system indicates that the calibration data has been saved on the calibration disk,
select Cancel, Exit, and switch OFF the instrument when the DOS prompt appears.
9) At the Calibration Startup Screen, press the help button one or more times to select the
appropriate aperture setting (refer to service bulletin ACT-005).
Note - If the correct aperture setting is already indicated, press the help button twice to
reselect it. It is necessary to press the help button one or more times to force the software to
update the calibration data.
10) From the Calibration Startup Screen, select Manual, Refract, Refract.
11) At the Adjust Calibration Lenses screen, adjust each of the nominal values to the actual value
labeled for the lens on the Carousel Calibration Tool (Figs. 4-12 and 4-13).
Use the up/down arrow keys to select the value to be adjusted. Use the plus or minus softkeys
to change the selected value.
Notes:
1. Each lens on the drum tool includes two values. For aperture enhanced units, the values
with the “A” designation are used. For non-aperture units, the values without the “A”
designation are used.
2. For aperture enhanced units, the cylinder value with the “A” designation is used. For non-
aperture units, the cylinder value is always set to 90 degrees.
12) After adjusting the lens values, press the joystick button.
13) Align each lens as requested with the instrument eyepiece by rotating the carousel drum, and
then press the joystick button.
14) The system will collect data for the current lens and then ask for the next lens to be read.
15) When the system requests the “axis correction lens”, position the lens labeled CYL, and press
the joystick button.
16) When the message “Press Joystick To See Calibration Constants” appears, press the joystick
button.
18) When the system indicates that the calibration data has been saved, select Cancel to return to
the Calibration Startup Screen.
ACT-004 Page 3
A new source aperture is being installed in the optics head of all new units. The new aperture enhances
the instrument’s refraction results for accuracy and repeatability. The new aperture is mounted on the
detector path assembly (see Figures 1 and 2). This bulletin provides the necessary information to
upgrade existing units with the new aperture.
Notes:
1. Several special tools are required for this upgrade. It is not possible to successfully
complete this upgrade without all of the special tools listed on page five of this
bulletin.
2. An updated version of the Refractor Calibration Procedure must be followed when
calibrating aperture enhanced units. Refer to service bulletin ACT-004.
3. An updated Carousel Calibration Tool with Aperture values must be used when
performing the Refractor Calibration Procedure on aperture enhanced units.
Carousel Calibration Tools without aperture values must be sent back to the factory
for recalibration.
4. The Customer Test Eye must be recalibrated after the aperture upgrade. The
Customer Test Eye Calibration Procedure is listed on page five of this bulletin.
Figure 3.
ACT-005 Page 3
Upgrade Procedure
1) Remove the optics head covers. Refer to the Optics Head Covers Removal Procedure in the Acuitus
service manual.
2) Remove the two screws that secure the source LED housing to the front plate of the optics, and then
carefully lower the source LED housing and chopper disk assembly to the side of the instrument (see
Figure 4).
Note – In Figure 4, the bottom screw is hidden by the source LED housing.
3) Apply 4 small drops of adhesive to the points shown in Figure 5.
4) Install the Source Aperture. Carefully press the aperture in place by hand for 2 minutes.
6) Modify the optics head serial number by marking the letter “A” at the end of the optics head serial
number. See Figure 3.
7) Perform the Source LED Alignment procedure as listed in section 4 of the Acuitus service manual.
8) Insert the Service Calibration Disk into the floppy disk drive, and then switch ON the instrument.
The Calibration Startup Screen will be displayed after the system boots from the calibration disk.
9) From the Calibration Startup Screen, select Manual, Save A, OK, Save A.
A copy of the instrument’s calibration file is saved on the calibration disk.
10) When the system indicates that the calibration data has been saved on the calibration disk, select
Cancel, Exit, and switch OFF the instrument when the DOS prompt appears.
Note – The instrument must reboot from the calibration disk to load the new calibration data.
12) From the Calibration Startup Screen, press the help button on the Acuitus keypad to select “YES” for
the Aperture setting.
Note – On some older instruments, the aperture setting may already indicate “YES” before pressing
the help button. In these cases, the help button must be pressed twice selecting “NO” and then
“YES” to properly setup the calibration data.
14) When the system indicates that the calibration data has been saved, select Cancel, Exit, and switch
OFF the instrument when the DOS prompt appears.
15) Perform the Chopper Motor Offset Calibration Procedure as listed in section 4 of the Acuitus service
manual.
16) Perform the Refractor Calibration Procedure as listed in service bulletin ACT-004.
Note – It is not necessary to reinstall the A12 or later Application software if it is already installed.
18) Perform the Customer Test Eye Calibration Procedure listed on page five of this bulletin.
Page 4 ACT-005
REMOVE 2 SCREWS
Figure 4.
GLUE (4 LOCATIONS)
Figure 5.
ACT-005 Page 5
5) Take three readings on the test eye, selecting print after each reading is taken.
6) Verify that for all three readings the following acceptance criteria is achieved:
Sph Eq is between 4.00 and 6.00
Cyl is between 0.00 and –0.50
Avg K is between 41.50 and 43.00
7) Average the three Sph Eq readings to the nearest .12D, and write this value on the test eye label.
8) Write –0.25 on the test eye label for the CYL value.
9) Average the three Avg K readings to the nearest .12D, and write this value on the test eye label.
Required Tools
Note - Refer to the Acuitus Service Manual, Appendix A for tool descriptions and part numbers.
Two versions of the I/O Interface PCB are currently available for the Acuitus. The original I/O Interface
PCB (P/N 46370) is only compatible with instruments equipped with stepper motor chopper disk
assemblies. The latest I/O Interface PCB (P/N 49750) is compatible with instruments equipped with
stepper motor or DC motor chopper disk assemblies.
Jumper Configuration
The figure below illustrates the correct jumper configuration for the latest I/O Interface PCB. There are no
jumpers to configure on the original I/O Interface PCB.
Four different Video/Frame Grabber PCBs have been used with Acuitus instruments. This bulletin
provides the necessary service information concerning each PCB.
Four different CPU PCBs have shipped with Acuitus instruments. This bulletin provides the necessary
service information concerning each PCB.
P/N 55063 CPU PCB, Pentium 266 MHz Processor, no Flash Disk
This is the latest CPU PCB used on the Acuitus. This PCB has changes to the BIOS that allow operation
with the latest camera assembly (P/N 54608-1). For details on the camera, refer to service bulletin
ACT-009. This PCB is not compatible with the older camera assemblies.
With the exception of the BIOS change, the features and capabilities of this PCB are identical to the
52192 PCB. Unlike earlier versions, this PCB does not include RAM or a Flash Disk. One 64 MB DIMM
RAM Card and one Flash Disk are required when using this PCB. When exchanging PCBs, the RAM and
Flash Disk from the defective PCB can be transferred to the new PCB. This PCB will accept 4 or 8 MB
Flash Disks.
P/N 52192 CPU PCB, Pentium 266 MHz Processor, no Flash Disk
This is the third generation CPU PCB used on the Acuitus. Unlike the previous versions, this PCB does
not include RAM or a Flash Disk. One 64 MB DIMM RAM Card and one Flash Disk is required when
using this PCB. When exchanging PCBs, the RAM and Flash Disk from the defective PCB can be
transferred to the new PCB. This PCB will accept 4 or 8 MB Flash Disks.
When installing this Pentium PCB on a system previously equipped with a 486 PCB, in addition to the
RAM, three cables must also be replaced. Refer to the parts list below for more information.
Note — This PCB is no longer available. The 46610-4 or the 52192 CPU PCBs can be used as a
replacement part. Additional parts are required if the 52192 PCB is used as a replacement. Refer to the
parts list on page 2 of this bulletin.
This is the second generation CPU PCB used on the Acuitus. This PCB is identical to the 46610-4 PCB,
except that the flash disk is now 8 MB. This PCB does not include RAM. One 8 MB SIMM RAM Card is
required when using this PCB. When exchanging PCBs, the RAM and Flash Disk from the defective PCB
can be transferred to the new PCB. This PCB will accept 4 or 8 MB Flash Disks.
A new camera assembly (P/N 54608-1) is being installed on the Acuitus. The two original cameras used
on the Acuitus are no longer manufactured by the vendor. The new camera was introduced to continue
production of new Acuitus instruments, and to ensure adequate quantities of the original cameras are
reserved for service spares.
The new camera has been installed on new instruments starting with the following serial numbers:
5000-1345 5015-2897
Notes:
1. A supply of the original cameras have been reserved for spare parts. The new camera offers no
performance or reliability improvements. Upgrades from the original camera to the new camera
are not supported by Carl Zeiss Meditec.
2. The new camera was never installed on new model 5010 instruments.
Parts
Several new parts are required for the new camera.
Part No. Description
54608-1...................................... Camera Assembly
53460-1 or 53460-2 .................. Video/Frame Grabber PCB1
55063......................................... CPU PCB2
54720......................................... Cable, Main Cable Harness
52239......................................... Cable, CPU/Serial (COM1)
52239......................................... Cable, CPU/Serial (COM2)
53281......................................... Cable, CPU/Keyboard
52428......................................... Cable, CPU/Printer
52762......................................... Cable, Keypad/IO PCB
1 – Refer to service bulletin ACT-007A for more information concerning the Video/Frame Grabber
PCBs.
2 – Refer to service bulletin ACT-008A for more information concerning the CPU PCB.
Cable Connections
An updated system interconnect diagram is shown on page 2 of this document. This diagram is only valid
for instruments equipped with the new camera assembly. For older instruments, refer to the system
interconnect diagrams shown in section 6 of the Acuitus Service Manual.
Belts, Cables
Other Miscellaneous