SMO1002ED

Download as pdf or txt
Download as pdf or txt
You are on page 1of 63

Digital Storage Oscilloscope

User Manual

i
Table of Contents
1. General Safety Requirements.......................................................................................... 1
2. Safety Terms and Symbols................................................................................................2
3. Quick Start........................................................................................................................4
Introduction to the Structure of the Oscilloscope......................................................................4
Front Panel............................................................................................................................................... 4
Rear Panel.................................................................................................................................................5
Control Area.............................................................................................................................................6

User Interface Introduction......................................................................................................... 7


How to Implement the General Inspection................................................................................ 8
How to Implement the Function Inspection...............................................................................9
How to Implement the Probe Compensation........................................................................... 10
How to Set the Probe Attenuation Coefficient.........................................................................11
How to Use the Probe Safely......................................................................................................12
How to Implement Self-calibration...........................................................................................12
Introduction to the Vertical System..........................................................................................12
Introduction to the Horizontal System..................................................................................... 14
Introduction to the Trigger System.......................................................................................... 14
4. Advanced User Guidebook............................................................................................. 16
How to Set the Vertical System................................................................................................. 17
Use Mathematical Manipulation Function...............................................................................18
The Waveform Calculation......................................................................................................................18
Using FFT function..................................................................................................................................19

Use Vertical Position and Scale Knobs..................................................................................... 22


How to Set the Horizontal System.............................................................................................23
Zoom the Waveform................................................................................................................................ 23

How to Set the Trigger System..................................................................................................24


Single Trigger.......................................................................................................................................... 25
Alternate Trigger (Trigger mode: Edge)..................................................................................................26

How to Operate the Function Menu......................................................................................... 27


How to Set the Sampling/Display............................................................................................................27
How to Save and Recall a Waveform...................................................................................................... 28
How to Implement the Auxiliary System Function Setting.....................................................................35
How to Update your Instrument Firmware..............................................................................................37
How to Measure Automatically............................................................................................................... 38
How to Measure with Cursors................................................................................................................. 42
How to Use Executive Buttons................................................................................................................ 44

ii
5. Communication with PC................................................................................................ 47
6. Demonstration................................................................................................................ 48
Example 1: Measurement a Simple Signal...............................................................................48
Example 2: Gain of a Amplifier in a Metering Circuit........................................................... 49
Example 3: Capturing a Single Signal......................................................................................50
Example 4: Analyze the Details of a Signal..............................................................................51
Example 5: Application of X-Y Function................................................................................. 53
Example 6: Video Signal Trigger.............................................................................................. 54
7. Troubleshooting..............................................................................................................56
8. Technical Specifications.................................................................................................57
General Technical Specifications.............................................................................................. 59
9. Appendix......................................................................................................................... 60
Appendix A: Enclosure.............................................................................................................. 60
Appendix B: General Care and Cleaning.................................................................................60

iii
1.General Safety Requirements

1. General Safety Requirements


Before use, please read the following safety precautions to avoid any possible bodily
injury and to prevent this product or any other connected products from damage. To
avoid any contingent danger, ensure this product is only used within the ranges
specified.
Only a qualified person should perform internal maintenance.
To avoid Fire or Personal Injury:
 Use Proper Power Cord. Use only the power cord supplied with the product and
certified to use in your country.
 Connect or Disconnect Correctly. When the probe or test lead is connected to a
voltage source, please do not connect and disconnect the probe or test lead.
 Product Grounded. This instrument is grounded through the power cord grounding
conductor. To avoid electric shock, the grounding conductor must be grounded. The
product must be grounded properly before any connection with its input or output
terminals.
When the instrument is powered by AC, do not measure AC power sources
directly otherwise it will cause a short circuit. This is because the testing ground
and power cord ground conductor are connected.
 Check all Terminal Ratings. To avoid fire or shock hazard, check all ratings and
markings on this product. Refer to the user manual for more information about ratings
before connecting to the instrument.
 Do not operate without covers. Do not operate the instrument with covers or panels
removed.
 Use the Proper Fuse. Use only the specified type and rating fuse for this instrument.
 Avoid exposed circuit. Be careful when working on exposed circuitry to avoid risk of
electric shock or other injury.
 Do not operate if any damage. If you suspect damage to the instrument, have it
inspected by qualified service personnel before further use.
 Use your Oscilloscope in a well-ventilated area. Make sure the instrument installed
with proper ventilation.
 Do not operate in damp conditions.
 Do not operate in an explosive atmosphere.
 Keep product surfaces clean and dry.

1
2.Safety Terms and Symbols

2. Safety Terms and Symbols


Safety Terms
Terms in this manual (The following terms may appear in this manual):

Warning: Warning indicates conditions or practices that could result in injury or


loss of life.

Caution: Caution indicates the conditions or practices that could result in


damage to this product or other property.
Terms on the product. The following terms may appear on this product:
Danger: Indicates an immediate hazard or injury possibility.
Warning: Indicates a possible hazard or injury.
Caution: Indicates potential damage to the instrument or other property.

Safety Symbols
Symbols on the product. The following symbol may appear on the product:

Hazardous Voltage Refer to Manual

Protective Earth Terminal Chassis Ground

Test Ground

2
2.Safety Terms and Symbols
To avoid body damage and prevent product and connected equipment damage, carefully
read the following safety information before using the test tool. This product can only be
used in the specified applications.

Warning:
The two channels of the oscilloscope are not electrically isolated. The channels should
adopt a common ground during measuring. To prevent short circuits, the 2 probe
grounds must not be connected to 2 different non-isolated DC levels.
The diagram of the oscilloscope ground wire connection:

It is not allowed to measure AC power when the AC powered oscilloscope is


connected to the AC-powered PC through the ports.

Warning:
To avoid fire or electrical shock, when the oscilloscope input signal
connected is more than 42V peak (30Vrms) or on circuits of more than
4800VA, please take note of below items:
 Only use accessory insulated voltage probes and test lead.
 Check the accessories such as probe before use and replace it if
there are any damages.
 Remove probes, test leads and other accessories immediately after
use.
 Remove USB cable which connects oscilloscope and computer.
 Do not apply input voltages above the rating of the instrument
because the probe tip voltage will directly transmit to the
oscilloscope. Use with caution when the probe is set as 1:1.
 Do not use exposed metal BNC or banana plug connectors.
 Do not insert metal objects into connectors.

3
3.Quick Start

3. Quick Start
Introduction to the Structure of the Oscilloscope
This chapter makes a simple description of the operation and function of the front panel of
the oscilloscope, enabling you to be familiar with the use of the oscilloscope in the
shortest time.

Front Panel
The front panel has knobs and function buttons. The 5 buttons in the column on the right
side of the display screen are menu selection buttons, through which, you can set the
different options for the current menu. The other buttons are function buttons, through
which, you can enter different function menus or obtain a specific function application
directly.

Figure 3- 1 Front panel


1. Display area
2. Menu selection buttons: Select the right menu item.
3. Control (button and knob) area
4. Probe Compensation: Measurement signal (5V/1kHz) output.
5. Signal Input Channel
6. USB Host port: It is used to transfer data when external USB equipment connects to
the oscilloscope regarded as "host device". For example: Saving the waveform to USB
flash disk needs to use this port.
7. Power on/off

4
3.Quick Start

Rear Panel

Figure 3- 2 Rear Panel


1. Handle
2. Air vents
3. AC power input jack
4. Foot stool: Adjust the tilt angle of the oscilloscope.
5. USB Device port: It is used to transfer data when external USB equipment connects
to the oscilloscope regarded as "slave device". For example: to use this port when
connect PC to the oscilloscope by USB.

5
3.Quick Start
Control Area

Figure 3- 3 Control Area Overview


1. Function button area: Total 6 buttons.
2. Horizontal control area with 1 button and 2 knobs.
"HOR" button refer to horizontal system setting menu, "Horizontal Position" knob
control trigger position, " Horizontal Scale" control time base.
3. Trigger control area with 2 buttons and 1 knob.
The Trigger Level knob is to adjust trigger voltage. Other 2 buttons refer to trigger
system setting.
4. Copy button: This button is the shortcut for Save function in the Utility function
menu. Pressing this button is equal to the Save option in the Save menu. The
waveform, configure or the display screen could be saved according to the chosen type
in the Save menu.
5. Vertical control area with 3 buttons and 4 knobs.
"CH1" and "CH2 " correspond to setting menu in CH1 and CH2, "Math" button refer
to math menu, the math menu consists of six kinds of operations, including CH1-CH2,
CH2-CH1, CH1+CH2, CH1*CH2, CH1/CH2 and FFT. Two "Vertical Position" knob
control the vertical position of CH1/CH2, and two "Scale" knob control voltage scale
of CH1, CH2.
6. M knob(Multipurpose knob): when a symbol appears in the menu, it indicates
you can turn the M knob to select the menu or set the value. You can push it to close
the menu on the left and right.

6
3.Quick Start

User Interface Introduction

Figure 3- 4 Illustrative Drawing of Display Interfaces


1. Waveform Display Area.
2. Run/Stop
3. The state of trigger, including:
Auto: Automatic mode and acquire waveform without triggering.
Trig: Trigger detected and acquire waveform.
Ready: Pre-triggered data captured and ready for a trigger.
Scan: Capture and display the waveform continuously.
Stop: Data acquisition stopped.
4. The two blue dotted lines indicates the vertical position of cursor measurement.
5. The T pointer indicates the horizontal position for the trigger.
6. The pointer indicates the trigger position in the record length.
7. It shows present triggering value and displays the site of present window in
internal memory.
8. It indicates that there is a USB disk connecting with the oscilloscope.
9. Channel identifier of current menu.
10. The waveform of CH1.
11. Right Menu.
12. The waveform of CH2.
13. Current trigger type:
Rising edge triggering

7
3.Quick Start
Falling edge triggering
Video line synchronous triggering
Video field synchronous triggering
The reading shows the trigger level value of the corresponding channel.
14. It indicates the measured type and value of the corresponding channel. "T" means
period, "F" means frequency, "V" means the average value, "Vp" the peak-peak
value, "Vr" the root-mean-square value, "Ma" the maximum amplitude value,
"Mi" the minimum amplitude value, "Vt" the Voltage value of the waveform's
flat top value, "Vb" the Voltage value of the waveform's flat base, "Va" the
amplitude value, "Os" the overshoot value, "Ps" the Preshoot value, "RT" the rise
time value, "FT" the fall time value, "PW" the +width value, "NW" the -Width
value, "+D" the +Duty value, "-D" the -Duty value, "PD" the Delay A->B
value, "ND" the Delay A->B value, "TR" the Cycle RMS, "CR" the Cursor
RMS, "WP" the Screen Duty, "RP" the Phase, "+PC" the +Pulse count, "-PC"
the - Pulse count, "+E" the Rise edge count, "-E" the Fall edge count, "AR" the
Area, "CA" the Cycle area.
15. The readings show the record length.
16. The frequency of the trigger signal.
17. The readings show current sample rate.
18. The readings indicate the corresponding Voltage Division and the Zero Point
positions of the channels. "BW" indicates bandwidth limit.
The icon shows the coupling mode of the channel.
"—" indicates direct current coupling
"~" indicates AC coupling
" " indicates GND coupling
19. The reading shows the setting of main time base.
20. It is cursor measure window, showing the absolute values and the readings of the
cursors.
21. The blue pointer shows the grounding datum point (zero point position) of the
waveform of the CH2 channel. If the pointer is not displayed, it means that this
channel is not opened.
22. The two blue dotted lines indicate the horizontal position of cursor measurement.
23. The yellow pointer indicates the grounding datum point (zero point position) of
the waveform of the CH1 channel. If the pointer is not displayed, it means that
the channel is not opened.

How to Implement the General Inspection


After you get a new oscilloscope, it is recommended that you should make a check on
the instrument according to the following steps:

1. Check whether there is any damage caused by transportation.

8
3.Quick Start
If it is found that the packaging carton or the foamed plastic protection cushion has
suffered serious damage, do not throw it away first till the complete device and its
accessories succeed in the electrical and mechanical property tests.

2. Check the Accessories


The supplied accessories have been already described in the "Appendix A: Enclosure" of
this Manual. You can check whether there is any loss of accessories with reference to
this description. If it is found that there is any accessory lost or damaged, please get in
touch with our distributor responsible for this service or our local offices.

3. Check the Complete Instrument


If it is found that there is damage to the appearance of the instrument, or the
instrument can not work normally, or fails in the performance test, please get in touch
with our distributor responsible for this business or our local offices. If there is
damage to the instrument caused by the transportation, please keep the package. With
the transportation department or our distributor responsible for this business informed
about it, a repairing or replacement of the instrument will be arranged by us.

How to Implement the Function Inspection


Make a fast function check to verify the normal operation of the instrument, according
to the following steps:
1. Connect the power cord to a power source. Press the button on the bottom
left of the instrument.

The instrument carries out all self-check items and shows the Boot Logo. Push the
Utility button, select Function in the right menu. Select Adjust in the left menu,
select Default in the right menu. The default attenuation coefficient set value of the
probe in the menu is 10X.
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope
with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten
the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe
compensator.
3. Push the Autoset Button on the front panel.
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in
several seconds (see Figure 3- 5).

9
3.Quick Start

Figure 3- 5 Auto set

Check CH2 by repeating Step 2 and Step 3.

How to Implement the Probe Compensation


When connect the probe with any input channel for the first time, make this
adjustment to match the probe with the input channel. The probe which is not
compensated or presents a compensation deviation will result in the measuring error
or mistake. For adjusting the probe compensation, please carry out the following
steps:
1. Set the attenuation coefficient of the probe in the menu as 10X and that of the
switch in the probe as 10X (see "How to Set the Probe Attenuation Coefficient"
on P11), and connect the probe with the CH1 channel. If a probe hook tip is used,
ensure that it keeps in close touch with the probe. Connect the probe tip with the
signal connector of the probe compensator and connect the reference wire clamp
with the ground wire connector of the probe connector, and then push the
Autoset button on the front panel.
2. Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Figure 3- 6 and Figure 3- 7).

Overcompensated Compensated correctly Under compensated


Figure 3- 6 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if needed.

10
3.Quick Start

Figure 3- 7 Adjust Probe

How to Set the Probe Attenuation Coefficient


The probe has several attenuation coefficients, which will influence the vertical scale
factor of the oscilloscope.
To change or check the probe attenuation coefficient in the menu of oscilloscope:
(1) Push the function menu button of the used channels (CH1 or CH2 button).
(2) Select Probe in the right menu; turn the M knob to select the proper value in the
left menu corresponding to the probe.
This setting will be valid all the time before it is changed again.
Caution:
The default attenuation coefficient of the probe on the instrument is preset to
10X.
Make sure that the set value of the attenuation switch in the probe is the
same as the menu selection of the probe attenuation coefficient in the
oscilloscope.

The set values of the probe switch are 1X and 10X (see Figure 3- 8).

Figure 3- 8 Attenuation Switch

Caution:
When the attenuation switch is set to 1X, the probe will limit the bandwidth
of the oscilloscope in 5MHz. To use the full bandwidth of the oscilloscope,
the switch must be set to 10X.

11
3.Quick Start

How to Use the Probe Safely


The safety guard ring around the probe body protects your finger against any electric
shock, shown as Figure 3- 9.

Figure 3- 9 Finger Guard

Warning:
To avoid electric shock, always keep your finger behind the safety guard
ring of the probe during the operation.
To protect you from suffering from the electric shock, do not touch any
metal part of the probe tip when it is connected to the power supply.
Before making any measurements, always connect the probe to the
instrument and connect the ground terminal to the earth.

How to Implement Self-calibration


The self-calibration application can make the oscilloscope reach the optimum
condition rapidly to obtain the most accurate measurement value. You can carry out
this application program at any time. This program must be executed whenever the
change of ambient temperature is 5℃ or over.
Before performing a self-calibration, disconnect all probes or wires from the input
connector. Push the Utility button, select Function in the right menu, select Adjust.
in the left menu, select Self Cal in the right menu; run the program after everything is
ready.

Introduction to the Vertical System


As shown in Figure 3- 10, there are a few of buttons and knobs in Vertical Controls.
The following practices will gradually direct you to be familiar with the using of the
vertical setting.

12
3.Quick Start

Figure 3- 10 Vertical Control Zone

1. Use the Vertical Position knob to show the signal in the center of the waveform
window. The Vertical Position knob functions the regulating of the vertical display
position of the signal. Thus, when the Vertical Position knob is rotated, the pointer of
the earth datum point of the channel is directed to move up and down following the
waveform.
Measuring Skill
If the channel is under the DC coupling mode, you can rapidly measure the DC
component of the signal through the observation of the difference between the wave
form and the signal ground.
If the channel is under the AC mode, the DC component would be filtered out. This
mode helps you display the AC component of the signal with a higher sensitivity.
Vertical offset back to 0 shortcut key
Turn the Vertical Position knob to change the vertical display position of channel
and push the position knob to set the vertical display position back to 0 as a shortcut
key, this is especially helpful when the trace position is far out of the screen and want
it to get back to the screen center immediately.
2. Change the Vertical Setting and Observe the Consequent State Information Change.
With the information displayed in the status bar at the bottom of the waveform
window, you can determine any changes in the channel vertical scale factor.
 Turn the Vertical Scale knob and change the "Vertical Scale Factor (Voltage
Division)", it can be found that the scale factor of the channel corresponding to
the status bar has been changed accordingly.
 Push buttons of CH1, CH2 and Math, the operation menu, symbols, waveforms
and scale factor status information of the corresponding channel will be displayed
in the screen.

13
3.Quick Start

Introduction to the Horizontal System


Shown as Figure 3- 11, there are a button and two knobs in the Horizontal Controls.
The following practices will gradually direct you to be familiar with the setting of
horizontal time base.

Figure 3- 11 Horizontal Control Zone

1. Turn the Horizontal Scale knob to change the horizontal time base setting and
observe the consequent status information change. Turn the Horizontal Scale
knob to change the horizontal time base, and it can be found that the Horizontal
Time Base display in the status bar changes accordingly.
2. Use the Horizontal Position knob to adjust the horizontal position of the signal
in the waveform window. The Horizontal Position knob is used to control the
triggering displacement of the signal or for other special applications. If it is
applied to triggering the displacement, it can be observed that the waveform
moves horizontally with the knob when you rotate the Horizontal Position knob.
Triggering displacement back to 0 shortcut key
Turn the Horizontal Position knob to change the horizontal position of channel
and push the Horizontal Position knob to set the triggering displacement back to 0
as a shortcut key.

3. Push the Horizontal HOR button to switch between the normal mode and the
wave zoom mode.

Introduction to the Trigger System


As shown in Figure 3- 12, there are one knob and three buttons make up Trigger
Controls. The following practices will direct you to be familiar with the setting of the
trigger system gradually.

Figure 3- 12 Trigger Control Zone


14
3.Quick Start
1. Push the Trigger Menu button and call out the trigger menu. With the operations
of the menu selection buttons, the trigger setting can be changed.
2. Use the Trigger Level knob to change the trigger level setting.
By turning the Trigger Level knob, the trigger indicator in the screen will move
up and down. With the movement of the trigger indicator, it can be observed that
the trigger level value displayed in the screen changes accordingly.
Note: Turning the Trigger Level knob can change trigger level value and it is
also the hotkey to set trigger level as the vertical mid point values of the
amplitude of the trigger signal.
3. Push the Force button to force a trigger signal, which is mainly applied to the
"Normal" and "Single" trigger modes.

15
4.Advanced User Guidebook

4. Advanced User Guidebook


This chapter will deal with the following topics mainly:

 How to Set the Vertical System


 How to Set the Horizontal System
 How to Set the Trigger System
 How to Set the Sampling/Display
 How to Save and Recall Waveform
 How to Implement the Auxiliary System Function Setting
 How to Update your Instrument Firmware
 How to Measure Automatically
 How to Measure with Cursors
 How to Use Executive Buttons
It is recommended that you read this chapter carefully to get acquainted the various
measurement functions and other operation methods of the oscilloscope.

16
4.Advanced User Guidebook

How to Set the Vertical System


The VERTICAL CONTROLS includes three menu buttons such as CH1, CH2 and
Math, and four knobs such as Vertical Position, Vertical Scale for each channel.
Setting of CH1 and CH2
Each channel has an independent vertical menu and each item is set respectively
based on the channel.
To turn waveforms on or off (channel, math)
Pushing the CH1, CH2, or Math buttons have the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off and its
menu goes away.
The description of the Channel Menu is shown as the following list:
Function Setting Description
Menu
DC Pass both AC and DC components of the input signal.
Coupling AC Block the DC component of the input signal.
Ground Disconnect the input signal.
ON Display inverted waveform.
Inverted
OFF Display original waveform.
1X
10X Match this to the probe attenuation factor to have an
Probe
100X accurate reading of vertical scale.
1000X
Yes If you are measuring current by probing the voltage
MeasCurr
No drop across a resistor, choose Yes.
Turn the M knob to set the Amps/Volts ratio. The
A/V V/A range is 100 mA/V - 1 KA/V.
or mA/V or mV/A Amps/Volts ratio = 1/Resistor value
Volts/Amp ratio is automatically calculated.
Full band Get full bandwidth.
Limit 20M Limit the channel bandwidth to 20MHz to reduce
display noise.

1. To set channel coupling


Taking the Channel 1 for example, the measured signal is a square wave signal
containing the direct current bias. The operation steps are shown as below:
(1) Push the CH1 button to show the CH1 SETUP menu.

17
4.Advanced User Guidebook
(2) In the right menu, select Coupling as DC. Both DC and AC components of
the signal are passed.
(3) In the right menu, select Coupling as AC. The direct current component of
the signal is blocked.
2. To invert a waveform
Waveform inverted: the displayed signal is turned 180 degrees against the phase
of the earth potential.
Taking the Channel 1 for example, the operation steps are shown as follows:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) In the right menu, select Inverted as ON, the waveform is inverted. Push
again to switch to OFF, the waveform goes back to its original one.
3. To adjust the probe attenuation
For correct measurements, the attenuation coefficient settings in the operating
menu of the Channel should always match what is on the probe (see "How to Set
the Probe Attenuation Coefficient" on P11). If the attenuation coefficient of the
probe is 1:1, the menu setting of the input channel should be set to1X.
Take the Channel 1 as an example, the attenuation coefficient of the probe is 10:1,
the operation steps are shown as follows:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) In the right menu, select Probe. In the left menu, turn the M knob to set it as
10X.
4. To measure current by probing the voltage drop across a resistor
Take the Channel 1 as an example, if you are measuring current by probing the
voltage drop across a 1Ω resistor, the operation steps are shown as follows:
(1) Push the CH1 button to show CH1 SETUP menu.
(2) In the right menu, set MeasCurr as Yes, the A/V radio menu will appear
below. Select it; turn the M knob to set the Amps/Volts ratio. Amps/Volts
ratio = 1/Resistor value. Here the A/V radio should be set to 1.

Use Mathematical Manipulation Function


The Mathematical Manipulation function is used to show the results of the addition,
multiplication, division and subtraction operations between two channels, or the FFT
operation for a channel. Press the Math button to display the menu on the right.

The Waveform Calculation


Press the Math button to display the menu on the right, select Type as Math.
Function Menu Setting Description

18
4.Advanced User Guidebook
Type Math Display the Math menu
CH1
Factor1 Select the signal source of the factor1
CH2
Sign +-*/ Select the sign of mathematical manipulation
CH1
Factor2 Select the signal source of the factor2
CH2
Next Page Enter next page
Vertical Turn the M knob to adjust the vertical
(div) position of the Math waveform.
Vertical Turn the M knob to adjust the voltage
(V/div) division of the Math waveform.
Prev Page Enter previous page

Taking the additive operation between Channel 1 and Channels 2 for example, the
operation steps are as follows:
1. Press the Math button to display the math menu in the right. The pink M
waveform appears on the screen.
2. In the right menu, select Type as Math.
3. In the right menu, select Factor1 as CH1.
4. In the right menu, select Sign as +.
5. In the right menu, select Factor2 as CH2.
6. Press Next Page in the right menu. Select Vertical (div), the symbol is in
front of div, turn the M knob to adjust the vertical position of Math waveform.
7. Select Vertical (V/div) in the right menu, the symbol is in front of the voltage,
turn the M knob to adjust the voltage division of Math waveform.

Using FFT function


The FFT (fast Fourier transform) math function mathematically converts a time-domain
waveform into its frequency components. It is very useful for analyzing the input signal on
Oscilloscope. You can match these frequencies with known system frequencies, such as
system clocks, oscillators, or power supplies.
FFT function in this oscilloscope transforms 2048 data points of the time-domain signal
into its frequency components mathematically (the record length should be 10K or above).
The final frequency contains 1024 points ranging from 0Hz to Nyquist frequency.
Press the Math button to display the menu on the right, select Type as FFT.
Function Menu Setting Description
Type FFT Display the FFT menu
CH1 Select CH1 as FFT source.
Source
CH2 Select CH2 as FFT source.

19
4.Advanced User Guidebook
Hamming
Rectangle
Blackman
Window Select window for FFT.
Hanning
Kaiser
Bartlett
Vrms Select Vrms for Format.
Format
dB Select dB for Format.
Next Page Enter next page
frequency Switch to select the horizontal position or
Hori (Hz) time base of the FFT waveform, turn the
frequency/div M knob to adjust it

div Switch to select the vertical position or


Vertical voltage division of the FFT waveform,
V or dBVrms turn the M knob to adjust it
Prev Page Enter previous page

Taking the FFT operation for example, the operation steps are as follows:
1. Press the Math button to display the math menu in the right.
2. In the right menu, select Type as FFT.
3. In the right menu, select Source as CH1.
4. In the right menu, select Window. Select the proper window type in the left menu.
5. In the right menu, select Format as Vrms or dB.
6. In the right menu, press Hori (Hz) to make the symbol in front of the frequency
value, turn the M knob to adjust the horizontal position of FFT waveform; then press
to make the symbol in front of the frequency/div below, turn the M knob to adjust
the time base of FFT waveform.
7. Select Vertical in the right menu; do the same operations as above to set the vertical
position and voltage division.
To select the FFT window
■ There are 6 FFT windows. Each one has trade-offs between frequency resolution and
magnitude accuracy. What you want to measure and your source signal characteristics
help you to determine which window to use. Use the following guidelines to select the
best window.
Type Characteristics Window

20
4.Advanced User Guidebook
Better solution for magnitude than Rectangle, and
good for frequency as well. It has slightly better
frequency resolution than Hanning.
Recommend to use for:
Hamming
 Sine, periodic and narrow band random noise.
 Transients or bursts where the signal levels
before and after the event are significantly
different.
Best solution for frequency, worst for magnitude.
Best type for measuring the frequency spectrum of
nonrepetitive signals and measuring frequency
components near DC.
Recommend to use for:
Rectangle  Transients or bursts, the signal level before and
after the event are nearly equal.
 Equal-amplitude sine waves with frequencies
those are very close.
 Broadband random noise with a relatively slow
varying spectrum.
Best solution for magnitude, worst for frequency.
Recommend to use for:
Blackman
 Single frequency waveforms, to find higher
order harmonics.
Good for magnitude, but poorer frequency
resolution than Hamming.
Recommend to use for:
Hanning  Sine, periodic and narrow band random noise.
 Transients or bursts where the signal levels
before and after the event are significantly
different.

21
4.Advanced User Guidebook
The frequency resolution when using the Kaiser
window is fair; the spectral leakage and amplitude
accuracy are both good.

Kaiser The Kaiser window is best used when frequencies


are very close to the same value but have widely
differing amplitudes (the side lobe level and shape
factor are closest to the traditional Gaussian RBW).
This window is also good for random signals.

The Bartlett window is a slightly narrower variant


Bartlett of the triangular window, with zero weight at both
ends.

Notes for using FFT


 Use the default dB scale for details of multiple frequencies, even if they have very
different amplitudes. Use the Vrms scale to compare frequencies.
 DC component or offset can cause incorrect magnitude values of FFT waveform. To
minimize the DC component, choose AC Coupling on the source signal.
 To reduce random noise and aliased components in repetitive or single-shot events, set
the oscilloscope acquisition mode to average.
What is Nyquist frequency?
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope
can acquire without aliasing. This frequency is half of the sample rate. Frequencies above
the Nyquist frequency will be under sampled, which causes aliasing. So pay more
attention to the relation between the frequency being sampled and measured.

Use Vertical Position and Scale Knobs

1. The Vertical Position knob is used to adjust the vertical positions of the
waveforms.
The analytic resolution of this control knob changes with the vertical division.
2. The Vertical Scale knob is used to regulate the vertical resolution of the wave
forms. The sensitivity of the vertical division steps as 1-2-5.
The vertical position and vertical resolution is displayed at the left bottom corner of
the screen (see Figure 4- 1).

22
4.Advanced User Guidebook

Figure 4- 1 Information about Vertical Position

How to Set the Horizontal System


The HORIZONTAL CONTROLS includes the Horizontal HOR button and such
knobs as Horizontal Position and Horizontal Scale.
1. Horizontal Position knob: this knob is used to adjust the horizontal positions of
all channels (include those obtained from the mathematical manipulation), the
analytic resolution of which changes with the time base.
2. Horizontal Scale knob: it is used to set the horizontal scale factor for setting the
main time base or the window.
3. Horizontal HOR button: push it to switch between the normal mode and the
wave zoom mode. For more detailed operations, see the introductions below.

Zoom the Waveform


Push the Horizontal HOR button to enter wave zoom mode. The top half of the
display shows the Main window and the bottom half displays the Zoom window. The
Zoom window is a magnified portion of the Main window.

23
4.Advanced User Guidebook

In normal mode, the Horizontal Position and Horizontal Scale knobs are used to
adjust the horizontal position and time base of the Main window.
In wave zoom mode, the Horizontal Position and Horizontal Scale knobs are used
to adjust the horizontal position and time base of the Zoom window.

How to Set the Trigger System


Trigger determines when DSO starts to acquire data and display waveform. Once
trigger is set correctly, it can convert the unstable display to meaningful waveform.
When DSO starts to acquire data, it will collect enough data to draw waveform on
left of trigger point. DSO continues to acquire data while waiting for trigger
condition to occur. Once it detects a trigger it will acquire enough data continuously
to draw the waveform on right of trigger point.
Trigger control area consists of 1 knob and 2 menu buttons.
Trigger Level: The knob that set the trigger level; push the knob and the level will be
set as the vertical mid point values of the amplitude of the trigger signal.
Force: Force to create a trigger signal and the function is mainly used in "Normal"
and "Single" mode.
Trigger Menu: The button that activates the trigger control menu.

Trigger Control
The oscilloscope provides two trigger types: single trigger, alternate trigger. Each type
of trigger has different sub menus.
Single trigger: Use a trigger level to capture stable waveforms in two channels
simultaneously.
Alternate trigger: Trigger on non-synchronized signals.
The Single Trigger, Alternate Trigger menus are described respectively as follows:

24
4.Advanced User Guidebook
Single Trigger
Single trigger has two types: edge trigger, video trigger.

Edge Trigger: It occurs when the trigger input passes through a specified voltage level
with the specified slope.
Video Trigger: Trigger on fields or lines for standard video signal.
The two trigger modes in Single Trigger are described respectively as follows:

1. Edge Trigger
An edge trigger occurs on trigger level value of the specified edge of input signal. Select
Edge trigger mode to trigger on rising edge or falling edge.

Push the Trigger Menu button to display the Trigger menu on the right. Select Type as
Single in the right menu. Select Single as Edge in the right menu.

In Edge Trigger mode, the trigger setting information is displayed on bottom right of the

screen, for example, , indicates that trigger type is edge, trigger


source is CH1, coupling is DC, and trigger level is 0.00mV.

Edge menu list:


Menu Settings Instruction
Type Single Set vertical channel trigger type as single trigger.
Single Edge Set vertical channel single trigger type as edge trigger.
CH1 Channel 1 as trigger signal.
Source
CH2 Channel 2 as trigger signal.
Auto Acquire waveform even no trigger occurs
Mode Normal Acquire waveform when trigger occurs
Single When trigger occurs, acquire one waveform then stop
Next Page Enter next page
AC Block the direct current component.
Coupling
DC Allow all component pass.
Trigger on rising edge
Slope
Trigger on falling edge
100 ns - 10 s, turn the M knob to set time interval
Holdoff
before another trigger occur.
Holdoff
Set Holdoff time as default value (100 ns).
Reset
Prev Page Enter previous page

Trigger Level: trigger level indicates vertical trig position of the channel, rotate trig level
knob to move trigger level, during setting, a dotted line displays to show trig position, and
the value of trigger level changes at the right corner, after setting, dotted line disappears.

2. Video Trigger
25
4.Advanced User Guidebook
Choose video trigger to trigger on fields or lines of NTSC, PAL or SECAM standard
video signals.
Push the Trigger Menu button to display the Trigger menu on the right. Select Type as
Single in the right menu. Select Single as Video in the right menu.
In Video Trigger mode, the trigger setting information is displayed on bottom right of the

screen, for example, ,indicates that trigger type is Video, trigger source is
CH1, and Sync type is Even.

Video Trigger menu list:


MENU SETTING INSTRUCTION
Type Single Set vertical channel trigger type as single trigger.
Single Video Set vertical channel single trigger type as video trigger.
CH1 Select CH1 as the trigger source
Source
CH2 Select CH2 as the trigger source
NTSC
Modu PAL Select video modulation
SECAM
Next Page Enter next page
Line Synchronic trigger in video line
Field Synchronic trigger in video field
Odd Synchronic trigger in video odd filed
Sync
Even Synchronic trigger in video even field
Line NO. Synchronic trigger in designed video line. Press Line
NO. menu item, turn the M knob to set the line number.
Prev Page Enter previous page

Alternate Trigger (Trigger mode: Edge)


Trigger signal comes from two vertical channels when alternate trigger is on. This mode is
used to observe two unrelated signals. Trigger mode is edge trigger.

Alternate trigger (Trigger Type: Edge) menu list:


Menu Settings Instruction
Type ALT Set vertical channel trigger type as alternate trigger.
CH1 Channel 1 as trigger signal.
Source
CH2 Channel 2 as trigger signal.
Next Page Enter next page
AC Block the direct current component.
Coupling
DC Allow all component pass.
Trigger on rising edge
Slope
Trigger on falling edge
100 ns - 10 s, turn the M knob to set time interval
Holdoff
before another trigger occur.

26
4.Advanced User Guidebook
Holdoff
Set Holdoff time as default value (100 ns).
Reset
Prev Page Enter previous page

How to Operate the Function Menu


The function menu control zone includes 4 function menu buttons: Utility, Measure,
Acquire, Cursor, and 2 immediate-execution buttons: Autoset, Run/Stop.

How to Set the Sampling/Display


Push the Acquire button, the Sampling and Display menu is shown in the right as follows:
Function Menu Setting Description
Sample Normal sampling mode.

Peak Detect Use to capture maximal and minimal samples.


Finding highest and lowest points over adjacent
intervals. It is used for the detection of the jamming
Acqu Mode
burr and the possibility of reducing the confusion.

Average It is used to reduce the random and don't-care


noises, with the optional number of averages. Turn
the M knob to select 4, 16, 64, 128 in the left menu.
Dots Only the sampling points are displayed.
Type Vect The space between the adjacent sampling points in
the display is filled with the vector form.
OFF
1 Second
Persist 2 Seconds Set the persistence time
5 Seconds
Infinity
ON
XY Mode Turn on/off XY display function
OFF
ON
Counter Turn on/off counter
OFF

Persist
When the Persist function is used, the persistence display effect of the picture tube
oscilloscope can be simulated. The reserved original data is displayed in fade color and the
new data is in bright color.
(1) Push the Acquire button.
(2) In the right menu, press Persist to select the persist time, including OFF, 1 Second, 2

27
4.Advanced User Guidebook
Seconds, 5 Seconds and Infinity. When the "Infinity" option is set for Persist Time,
the measuring points will be stored till the controlling value is changed. Select OFF
to turn off persistence and clear the display.

XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display format is
selected, Channel 1 is displayed in the horizontal axis and Channel 2 in the vertical axis;
the oscilloscope is set in the un-triggered sample mode: the data are displayed as bright
spots.
The operations of all control knobs are as follows:
 The Vertical Scale and the Vertical Position knobs of Channel 1 are used to set the
horizontal scale and position.
 The Vertical Scale and the Vertical Position knobs of Channel 2 are used to set the
vertical scale and position continuously.
The following functions can not work in the XY Format:
 Reference or digital wave form
 Cursor
 Trigger control
 FFT
Operation steps:
1. Push the Acquire button to show the right menu.
2. Select XY Mode as ON or OFF in the right menu.

Counter
It is a 6-digit single-channel counter. The counter can only measure the frequency of the
triggering channel. The frequency range is from 2Hz to the full bandwidth. Only if the
measured channel is in Edge mode of Single trigger type, the counter can be enabled. The
counter is displayed at the bottom of the screen.

Operation steps:
1. Push Trigger Menu button, set the trigger type to Single, set the trigger mode to
Edge, select the signal source.
2. Push the Acquire button to show the right menu.

3. Select Counter as ON or OFF in the right menu.

How to Save and Recall a Waveform


Push the Utility button, select Function in the right menu, select Save in the left menu.

28
4.Advanced User Guidebook
By selecting Type in the right menu, you can save the waveforms, configures or screen
images.
When the Type is selected as Wave, the menu is shown as the following table:
Function Menu Setting Description
Function Save Display the save function menu
Type Wave Choose the saving type as wave.
CH1 Choose the waveform to be saved.
CH2 (Choose All to save all the waveforms that
Source Math are turned on. You can save into the
All current internal object address, or into USB
storage as a single file.)
The object Wave0 –Wave15 are listed in
the left menu, turn the M knob to choose
the object which the waveform is saved to
or recall from.
Recall or close the waveform stored in the
ON current object address. When the show is
Object
OFF ON, if the current object address has been
used, the stored waveform will be shown,
the address number and relevant
information will be displayed at the top left
of the screen; if the address is empty, it
will prompt "None is saved".
Next Page Enter next page
Close all the waveforms stored in the
Close All
object address.
BIN For internal storage, only BIN can be
File Format TXT selected. For external storage, the format
CSV can be BIN, TXT or CSV.
Save the waveform of the source to the
Save
selected address.
Save to internal storage or USB storage.
When External is selected, the file name is
Internal
Storage editable. The BIN waveform file could be
External
open by waveform analysis software (on
the supplied CD).
Prev Page Enter previous page

When the Type is selected as Configure, the menu is shown as the following table:
Function Menu Setting Description
Function Save Display the save function menu
Type Configure Choose the saving type as configure.

29
4.Advanced User Guidebook
Setting1
Configure ….. The setting address
Setting8
Save the current oscilloscope configure to
Save
the internal storage
Recall the configure from the selected
Load
address

When the Type is selected as Image, the menu is shown as the following table:
Function Menu Setting Description
Function Save Display the save function menu
Type Image Choose the saving type as image.
Save the current display screen. The file
can be only stored in a USB storage, so a
Save USB storage must be connected first. The
file name is editable. The file is stored in
BMP format.

Save and Recall the Waveform


The oscilloscope can store 16 waveforms, which can be displayed with the current
waveform at the same time. The stored waveform called out can not be adjusted.
In order to save the waveform of CH1, CH2 and Math into the object Wave0, the
operation steps should be followed:
1. Turn on CH1, CH2 and Math channels.
2. Push the Utility button, select Function in the right menu, select Save in the left
menu. In the right menu, select Type as Wave.
3. Saving: In the right menu, select Source as All.
4. In the right menu, press Object. Select Wave0 as object address in the left menu.
5. In the right menu, press Next Page, and select Storage as Internal.
6. In the right menu, press Save to save the waveform.
7. Recalling: In the right menu, press Prev Page, and press Object, select Wave0 in the
left menu. In the right menu, select Object as ON, the waveform stored in the address
will be shown, the address number and relevant information will be displayed at the
top left of the screen.

In order to save the waveform of CH1 and CH2 into the USB storage as a BIN file, the
operation steps should be followed:
1. Turn on CH1 and CH2 channels, turn off the Math channel.
2. Push the Utility button, select Function in the right menu, select Save in the left

30
4.Advanced User Guidebook
menu. In the right menu, select Type as Wave.
3. Saving: In the right menu, select Source as All.
4. In the right menu, press Next Page, and select File Format as BIN.
5. In the right menu, select Storage as External.
6. In the right menu, select Storage, an input keyboard used to edit the file name will
pop up. The default name is current system date and time. Turn the M knob to choose
the keys; press the M knob to input the chosen key. The length of file name is up to
25 characters. Select the key in the keyboard to confirm.
7. Recalling: The BIN waveform file could be open by waveform analysis software (on
the supplied CD).
Shortcut for Save function:
The Copy button on the bottom right of the front panel is the shortcut for Save function in
the Utility function menu. Pressing this button is equal to the Save option in the Save
menu. The waveform, configure or the display screen could be saved according to the
chosen type in the Save menu.
Save the current screen image:
The screen image can only be stored in USB disk, so you should connect a USB disk with
the instrument.
1. Install the USB disk: Insert the USB disk into the "7. USB Host port" of "Figure
3- 1 Front panel". If an icon appears on the top right of the screen, the USB disk
is installed successfully. If the USB disk cannot be recognized, format the USB disk
according to the methods in "USB disk Requirements" on P31.
2. After the USB disk is installed, push the Utility button, select Function in the right
menu, select Save in the left menu. In the right menu, select Type as Image.
3. Select Save in the right menu, an input keyboard used to edit the file name will pop
up. The default name is current system date and time. Turn the M knob to choose the
keys; press the M knob to input the chosen key. The length of file name is up to 25
characters. Select the key in the keyboard to confirm.

USB disk Requirements


Support USB disk format: USB 2.0 or below, FAT16 or FAT32, allocation unit size no
exceed 4k, max capacity 64G. If the USB disk doesn't work properly, format your USB
disk and then try again. There are two methods in format the USB disk, first by using
computer system to format, the other one is through formatting software to format. (8G or
above USB disk can only use the second method to format, that is through formatting
software to format.)

Use system-provided function to format the USB disk


1. Connect the USB disk to the computer.

31
4.Advanced User Guidebook
2. Right click Computer- Manage to enter Computer Management interface.
3. Click Disk Management menu, and information about the USB disk will display on
the right side with red mark 1 and 2.

Figure 4- 2: Disk Management of computer


4. Right click 1 or 2 red mark area, choose Format. And system will pop up a warning
message, click Yes.

Figure 4- 3: Format the USB disk warning


5. Set File System as FAT32, Allocation unit size 4096. Check "Perform a quick
format" to execute a quick format. Click OK, and then click Yes on the warning
message.

32
4.Advanced User Guidebook

Figure 4- 4: Formatting the USB disk setting


6. Formatting process.

Figure 4- 5: Formatting the USB disk


7. Check whether the USB disk is FAT32 with allocation unit size 4096 after formatting.

Use Minitool Partition Wizard to format


Download URL: http://www.partitionwizard.com/free-partition-manager.html
Tip: There are many tools for the USB disk formatting on the market, just take Minitool
Partition Wizard for example here.
1. Connect the USB disk to the computer.
2. Open the software Minitool Partition Wizard.
3. Click Reload Disk on the pull-down menu at the top left or push keyboard F5, and
information about the USB disk will display on the right side with red mark 1 and 2.

33
4.Advanced User Guidebook

Figure 4- 6: Reload Disk


4. Right click 1 or 2 red mark area, choose Format.

Figure 4- 7: Choose format


5. Set File System FAT32, Cluster size 4096. Click OK.

Figure 4- 8: Format setting


6. Click Apply at the top left of the menu. Then click Yes on the pop-up warning to
begin formatting.

34
4.Advanced User Guidebook

Figure 4- 9: Apply setting


7. Formatting process

Figure 4- 10: Format process


8. Format the USB disk successfully

Figure 4- 11: Format successfully

How to Implement the Auxiliary System Function Setting


●Config
Push the Utility button, select Function in the right menu, select Configure in the left
menu.
The description of Configure Menu is shown as the follows:

35
4.Advanced User Guidebook

Function Menu Setting Description


Function Configure Show the configure menu
Lock all keys. Unlock method: push Trigger
KeyLock Menu button in trigger control area, then push
Force button, repeat 3 times.
About Show the version and serial number

●Display
Push the Utility button, select Function in the right menu, select Display in the left
menu.
The description of Display Menu is shown as the follows:

Function Menu Setting Description


Function Display Show the display menu
BackLight 0% - 100% Turn the M knob to adjust the backlight.

Graticule Select the grid type

Turn the M knob to set the disappear time


Menu Time OFF, 5S – 30S
of menu

●Adjust
Push the Utility button, select Function in the right menu, select Adjust in the left
menu.
The description of Adjust Menu is shown as the follows:
Function Menu Description
Self Cal Carry out the self-calibration procedure.
Default Call out the factory settings.
ProbeCh. Check whether probe attenuation is good.
Do Self Cal (Self-Calibration)
The self-calibration procedure can improve the accuracy of the oscilloscope under the
ambient temperature to the greatest extent. If the change of the ambient temperature is
up to or exceeds 5℃, the self-calibration procedure should be executed to obtain the
highest level of accuracy.
Before executing the self-calibration procedure, disconnect all probes or wires from the
input connector. Push the Utility button, select Function in the right menu, the function
menu will display at the left, select Adjust. If everything is ready, select Self Cal in the

36
4.Advanced User Guidebook
right menu to enter the self-calibration procedure of the instrument.
Probe checking
To check whether probe attenuation is good. The results contain three circumstances:
Overflow compensation, Good compensation, Inadequate compensation. According to the
checking result, users can adjust probe attenuation to the best. Operation steps are as
follows:
1. Connect the probe to CH1, adjust the probe attenuation to the maximum.
2. Push the Utility button, select Function in the right menu, select Adjust in the left
menu.
3. Select ProbeCh. in the right menu, tips about probe checking shows on the screen.
4. Select ProbeCh. again to begin probe checking and the checking result will occur
after 3s; push any other key to quit.

● Save
You can save the waveforms, configures or screen images. Refer to "How to Save and
Recall a Waveform" on page 28.

● Update
Use the front-panel USB port to update your instrument firmware using a USB memory
device. Refer to "How to Update your Instrument Firmware" on page 37.

How to Update your Instrument Firmware


Use the front-panel USB port to update your instrument firmware using a USB memory
device.
USB memory device requirements: Insert a USB memory device into the USB port on

the front panel. If the icon appears on the top right of the screen, the USB memory
device is installed successfully. If the USB memory device cannot be detected, format the
USB memory device according to the methods in "USB disk Requirements" on P31.
Caution: Updating your instrument firmware is a sensitive operation, to prevent damage
to the instrument, do not power off the instrument or remove the USB memory device
during the update process.
To update your instrument firmware, do the following:
1. Push the Utility button, select Function in the right menu, select Configure in the left
menu, select About in the right menu. View the model and the currently installed
firmware version.
2. Check if the website offers a newer firmware version. Download the firmware file. The
file name must be Scope.update. Copy the firmware file onto the root directory of your
USB memory device.
3. Insert the USB memory device into the front-panel USB port on your instrument.
4. Push the Utility button, select Function in the right menu, select Update in the left
menu.
5. In the right menu, select Start, the messages below will be shown.

37
4.Advanced User Guidebook

6. In the right menu, select Start again, the interfaces below will be displayed in
sequence. The update process will take up to three minutes. After completion, the
instrument will be shut down automatically.

7. Press the button to power on the instrument.

How to Measure Automatically


Push the Measure button to display the menu for the settings of the Automatic
Measurements. At most 8 types of measurements could be displayed on the bottom left
of the screen.
The oscilloscopes provide 30 parameters for auto measurement, including Period,
Frequency, Mean, PK-PK, RMS, Max, Min, Top, Base, Amplitude, Overshoot, Preshoot,
Rise Time, Fall Time, +PulseWidth, -PulseWidth, +Duty Cycle, -Duty Cycle, Delay
A→B , Delay A→B , Cycle RMS, Cursor RMS, Screen Duty, Phase, +PulseCount,
-PulseCount, RiseEdgeCnt, FallEdgeCnt, Area, and Cycle Area.
The "Automatic Measurements" menu is described as the following table:
Function Menu Setting Description
Press to show the left menu, turn the M knob to
Meas Type
AddCH1 select the measure type, press AddCH1 again
(left menu)
to add the selected measure type of CH1.

38
4.Advanced User Guidebook
Press to show the left menu, turn the M knob to
Meas Type
AddCH2 select the measure type, press AddCH2 again
(left menu)
to add the selected measure type of CH2.
OFF Hide the window of measures
Show CH1 Show all the measures of CH1 on the screen
CH2 Show all the measures of CH2 on the screen
Press to show the left menu, turn the M knob to
Meas Type select the type need to be deleted, press
Remove
(left menu) Remove again to remove the selected measure
type.
Remove All Remove all the measures

Measure
Only if the waveform channel is in the ON state, the measurement can be performed.
The automatic measurement can not be performed in the following situation: 1) On the
saved waveform. 2) On the Dual Wfm Math waveform. 3) On the Video trigger mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:

1. Push the Measure button to show the right menu.


2. Select AddCH1 in the right menu.
3. In the left Type menu, turn the M knob to select Period.
4. In the right menu, select AddCH1. The period type is added.
5. In the left Type menu, turn the M knob to select Frequency.
6. In the right menu, select AddCH1. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically (see
Figure 4- 12).

Figure 4- 12 Automatic measurement

39
4.Advanced User Guidebook

The automatic measurement of voltage parameters


The oscilloscopes provide automatic voltage measurements including Mean, PK-PK,
RMS, Max, Min, Vtop, Vbase, Vamp, OverShoot, PreShoot, Cycle RMS, and Cursor
RMS. Figure 4- 13 below shows a pulse with some of the voltage measurement points.

Figure 4- 13
Mean: The arithmetic mean over the entire waveform.
PK-PK: Peak-to-Peak Voltage.
RMS: The true Root Mean Square voltage over the entire waveform.
Max: The maximum amplitude. The most positive peak voltage measured over the
entire waveform.
Min: The minimum amplitude. The most negative peak voltage measured over the
entire waveform.
Vtop: Voltage of the waveform's flat top, useful for square/pulse waveforms.
Vbase: Voltage of the waveform's flat base, useful for square/pulse waveforms.
Vamp: Voltage between Vtop and Vbase of a waveform.
OverShoot: Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms.
PreShoot: Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms.
Cycle RMS: The true Root Mean Square voltage over the first entire period of the
waveform.
Cursor RMS: The true Root Mean Square voltage over the range of two cursors.
The automatic measurement of time parameters
The oscilloscopes provide time parameters auto-measurements include Period,
Frequency, Rise Time, Fall Time, +D width, -D width, +Duty, -Duty, Delay A→B ,
Delay A→B , and Duty cycle.
Figure 4- 14 shows a pulse with some of the time measurement points.

40
4.Advanced User Guidebook

Figure 4- 14

Rise Time: Time that the leading edge of the first pulse in the waveform takes to rise
from 10% to 90% of its amplitude.
Fall Time: Time that the falling edge of the first pulse in the waveform takes to fall
from 90% to 10% of its amplitude.
+D width: The width of the first positive pulse in 50% amplitude points.
-D width: The width of the first negative pulse in the 50% amplitude points.
+Duty: +Duty Cycle, defined as +Width/Period.
-Duty:-Duty Cycle, defined as -Width/Period.
Delay A→B : The delay between the two channels at the rising edge.
Delay A→B : The delay between the two channels at the falling edge.
Screen Duty: Defines as (the width of the positive pulse)/(Entire period)
Phase: Compare the rising edge of CH1 and CH2, calculate phase difference of two
channels.
Phase difference=(Delay between channels at the rising edge÷Period)×360°.

Other measurements
+PulseCount : The number of positive pulses that rise above the mid
reference crossing in the waveform.
-PulseCount : The number of negative pulses that fall below the mid
reference crossing in the waveform.
RiseEdgeCnt : The number of positive transitions from the low reference
value to the high reference value in the waveform.
FallEdgeCnt : The number of negative transitions from the high reference
value to the low reference value in the waveform.

41
4.Advanced User Guidebook
Area : The area of the whole waveform within the screen and the unit is
voltage-second. The area measured above the zero reference (namely the vertical
offset) is positive; the area measured below the zero reference is negative. The area
measured is the algebraic sum of the area of the whole waveform within the screen.
Cycle Area : The area of the first period of waveform on the screen and the
unit is voltage-second. The area above the zero reference (namely the vertical offset)
is positive and the area below the zero reference is negative. The area measured is the
algebraic sum of the area of the whole period waveform.
Note: When the waveform on the screen is less than a period, the period area
measured is 0.

How to Measure with Cursors


Push the Cursor button to turn cursors on and display the cursor menu. Push it again to
turn cursors off.
The Cursor Measurement for normal mode:
The description of the cursor menu is shown as the following table:
Function
Setting Description
Menu
Voltage Display the voltage measurement cursor and menu.
Time Display the time measurement cursor and menu.
Time&Voltage Display the time and voltage measurement cursor
Type
and menu.
AutoCursr The horizontal cursors are set as the intersections of
the vertical cursors and the waveform
Line Type Time Makes the vertical cursors active.
(Time&Vol
tage type)
Voltage Makes the horizontal cursors active.
Window
(Wave Main Measure in the main window.
zoom Extension Measure in the extension window.
mode)
a Turn the M knob to move line a.
b Turn the M knob to move line b.
Line
ab Two cursors are linked. Turn the M knob to move
the pair of cursors.
CH1 Display the channel to which the cursor
Source
CH2 measurement will be applied.
Perform the following operation steps for the time and voltage cursor measurement of
the channel CH1:
1. Push Cursor to display the cursor menu.
2. In the right menu, select Source as CH1.
3. Press the first menu item in the right menu, select Time&Voltage for Type, two
blue dotted lines displayed along the horizontal direction of the screen, two blue

42
4.Advanced User Guidebook
dotted lines displayed along the vertical direction of the screen. Cursor measure
window at the left bottom of the screen shows the cursor readout.
4. In the right menu, select Line Type as Time to make the vertical cursors active.
If the Line in the right menu is select as a, turn the M knob to move line a to the
right or left. If b is selected, turn the M knob to move line b.
5. In the right menu, select Line Type as Voltage to make the horizontal cursors
active. Select Line in the right menu as a or b, turn the M knob to move it.
6. Push the horizontal HOR button to enter wave zoom mode. Push Cursor to show
the right menu, select Window as Main or Extension to make the cursors shown
in the main window or zoom window.

Figure 4- 15 Time&Voltage Cursor Measurement

Auto Cursor
For the AutoCursr type, the horizontal cursors are set as the intersections of the vertical
cursors and the waveform.

43
4.Advanced User Guidebook
The Cursor Measurement for FFT mode
In FFT mode, push the Cursor button to turn cursors on and display the cursor menu.
The description of the cursor menu in FFT mode is shown as the following table:
Function
Setting Description
Menu
Vamp Display the Vamp measurement cursor and menu.
Freq Display the Freq measurement cursor and menu.
Freq&Vamp Display the Freq and Vamp measurement cursor
Type
and menu.
AutoCursr The horizontal cursors are set as the intersections of
the vertical cursors and the waveform
Line Type Freq Makes the vertical cursors active.
(Freq&Vamp
type)
Vamp Makes the horizontal cursors active.
Window Main Measure in the main window.
(Wave zoom
mode)
Extension Measure in the FFT extension window.
a Turn the M knob to move line a.
b Turn the M knob to move line b.
Line
ab Two cursors are linked. Turn the M knob to move
the pair of cursors.
Display the channel to which the cursor
Source Math FFT
measurement will be applied.
Perform the following operation steps for the amplitude and frequency cursor
measurement of math FFT:
1. Press the Math button to display the right menu. Select Type as FFT.
2. Push Cursor to display the cursor menu.
3. In the right menu, select Window as Extension.
4. Press the first menu item in the right menu, select Freq&Vamp for Type, two
blue dotted lines displayed along the horizontal direction of the screen, two blue
dotted lines displayed along the vertical direction of the screen. Cursor measure
window at the left bottom of the screen shows the cursor readout.
5. In the right menu, select Line Type as Freq to make the vertical cursors active. If
the Line in the right menu is select as a, turn the M knob to move line a to the
right or left. If b is selected, turn the M knob to move line b.
6. In the right menu, select Line Type as Vamp to make the horizontal cursors
active. Select Line in the right menu as a or b, turn the M knob to move it.
7. In the right cursor menu, you can select Window as Main to make the cursors
shown in the main window.

How to Use Executive Buttons


Executive Buttons include Autoset, Run/Stop, Copy.

44
4.Advanced User Guidebook
 [Autoset] button
It's a very useful and quick way to apply a set of pre-set functions to the incoming
signal, and display the best possible viewing waveform of the signal and also works
out some measurements for user as well.
The details of functions applied to the signal when using Autoset are shown as the
following table:
Function Items Setting
Vertical Coupling Current
Channel Coupling Current
Vertical Scale Adjust to the proper division.
Horizontal Level Middle or ±2 div
Horizontal Sale Adjust to the proper division
Trigger Type Slope or Video
Trigger Source CH1 or CH2
Trigger Coupling DC
Trigger Slope Current
Trigger Level 3/5 of the waveform
Trigger Mode Auto
Display Format YT
Force Stop
Inverted Off
Zoom Mode Exit

Judge waveform type by Autoset


Five kinds of types: Sine, Square, video signal, DC level, Unknown signal.
Menu as follow:
Waveform Menu
Sine Multi-period, Single-period, FFT, Cancel Autoset
Square Multi-period, Single-period, Rising Edge, Falling Edge,
Cancel Autoset
Video signal Type (line, field), Odd, Even, Line NO., Cancel
Autoset
DC level/Unknown signal Cancel Autoset
Description for some icons:
Multi-period:To display multiple periods
Single-period: To display single period
FFT: Switch to FFT mode
Rising Edge: Display the rising edge of square waveform
Falling Edge: Display the falling edge of square waveform
Cancel Autoset:Go back to display the upper menu and waveform information
Note: The Autoset function requires that the frequency of signal should be no lower

45
4.Advanced User Guidebook

than 20Hz, and the amplitude should be no less than 5mv. Otherwise, the Autoset
function may be invalid.

 [Run/Stop] button
Enable or disable sampling on input signals.
Notice: When there is no sampling at STOP state, the vertical division and the
horizontal time base of the waveform still can be adjusted within a certain range, in
other words, the signal can be expanded in the horizontal or vertical direction.
When the horizontal time base is ≤50ms, the horizontal time base can be expanded
for 4 divisions downwards.

 [Copy] button
This button is the shortcut for Save function in the Utility function menu. Pressing
this button is equal to the Save option in the Save menu. The waveform, configure or
the display screen could be saved according to the chosen type in the Save menu. For
more details, please see "How to Save and Recall a Waveform" on P28.

46
5.Communication with PC

5. Communication with PC
The oscilloscope supports communications with a PC through USB. You can use the
Oscilloscope communication software to store, analyze, display the data and remote
control.
To learn about how to operate the software, you can push F1 in the software to open the
help document.
Here is how to connect with PC via USB port.
(1) Install the software: Install the Oscilloscope communication software on the
supplied CD.
(2) Connection: Use a USB data cable to connect the USB Device port in the right panel
of the Oscilloscope to the USB port of a PC.
(3) Install the driver: Run the Oscilloscope communication software on PC, push F1 to
open the help document. Follow the steps of title "I. Device connection" in the
document to install the driver.
(4) Port setting of the software: Run the Oscilloscope software; click
"Communications" in the menu bar, choose "Ports-Settings", in the setting dialog,
choose "Connect using" as "USB". After connect successfully, the connection
information in the bottom right corner of the software will turn green.

47
6.Demonstration

6. Demonstration
Example 1: Measurement a Simple Signal
The purpose of this example is to display an unknown signal in the circuit, and
measure the frequency and peak-to-peak voltage of the signal.
1. Carry out the following operation steps for the rapid display of this signal:
(1) Set the probe menu attenuation coefficient as 10X and that of the switch in the
probe switch as 10X (see "How to Set the Probe Attenuation Coefficient" on P11).
(2) Connect the probe of Channel 1 to the measured point of the circuit.
(3) Push the Autoset button.
The oscilloscope will implement the Autoset to make the waveform optimized, based
on which, you can further regulate the vertical and horizontal divisions till the
waveform meets your requirement.
2. Perform Automatic Measurement
The oscilloscope can measure most of the displayed signals automatically. To
measure the period, the frequency of the CH1, following the steps below:
(1) Push the Measure button to show the right menu.
(2) Select AddCH1 in the right menu.
(3) In the left Type menu, turn the M knob to select Period.
(4) In the right menu, select AddCH1. The period type is added.
(5) In the left Type menu, turn the M knob to select Frequency.
(6) In the right menu, select AddCH1. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically
(see Figure 6- 1).

48
6.Demonstration

Figure 6- 1 Measure period and frequency value for a given signal

Example 2: Gain of a Amplifier in a Metering Circuit


The purpose of this example is to work out the Gain of an Amplifier in a Metering
Circuit. First we use Oscilloscope to measure the amplitude of input signal and output
signal from the circuit, then to work out the Gain by using given formulas.
Set the probe menu attenuation coefficient as 10X and that of the switch in the probe
as 10X (see "How to Set the Probe Attenuation Coefficient" on P11).
Connect the oscilloscope CH1 channel with the circuit signal input end and the CH2
channel to the output end.
Operation Steps:
(1) Push the Autoset button and the oscilloscope will automatically adjust the
waveforms of the two channels into the proper display state.
(2) Push the Measure button to show the right menu.
(3) Select AddCH1 in the right menu.
(4) In the left Type menu, turn the M knob to select PK-PK.
(5) In the right menu, select AddCH1. The peak-to-peak type of CH1 is added.
(6) In the right menu, select AddCH2. The peak-to-peak type of CH2 is added.
(7) Read the peak-to-peak voltages of Channel 1 and Channel 2 from the bottom left
of the screen (see Figure 6- 2).
(8) Calculate the amplifier gain with the following formulas.
Gain = Output Signal / Input signal
Gain (db) = 20×log (gain)

49
6.Demonstration

Figure 6- 2 Waveform of Gain Measurement

Example 3: Capturing a Single Signal


It's quite easy to use Digital Oscilloscope to capture non-periodic signal, such as a
pulse and burr etc. But the common problem is how to set up a trigger if you have no
knowledge of the signal? For example, if the pulse is the logic signal of a TTL level,
the trigger level should be set to 2 volts and the trigger edge be set as the rising edge
trigger. With various functions supported by our Oscilloscope, user can solve this
problem by taking an easy approach. First to run your test using auto trigger to find
out the closest trigger level and trigger type, this helps user to make few small
adjustments to achieve a proper trigger level and mode. Here is how we achieve this.
The operation steps are as follows:
(1) Set the probe menu attenuation coefficient to 10X and that of the switch in the
probe to 10X (see "How to Set the Probe Attenuation Coefficient" on P11).
(2) Adjust the Vertical Scale and Horizontal Scale knobs to set up a proper vertical
and horizontal ranges for the signal to be observed.
(3) Push the Acquire button to display the right menu.
(4) In the right menu, select Acqu Mode as Peak Detect.
(5) Push the Trigger Menu button to display the right menu.
(6) In the right menu, select Type as Single.
(7) In the right menu, select Single as Edge.
(8) In the right menu, select Source as CH1.
(9) In the right menu, press Next Page, select Coupling as DC.
(10)In the right menu, select Slope as (rising).

50
6.Demonstration
(11)Turn the Trigger Level knob and adjust the trigger level to the roughly 50% of
the signal to be measured.
(12)Check the Trigger State Indicator on the top of the screen, if it is not Ready, push
down the Run/Stop button and start acquiring, wait for trigger to happen. If a
signal reaches to the set trigger level, one sampling will be made and then
displayed in the screen. By using this approach, a random pulse can be captured
easily. For instance, if we want to find a burst burr of high amplitude, set the
trigger level to a slightly higher value of the average signal level, push the
Run/Stop button and wait a trigger. Once there is a burr occurring, the instrument
will trigger automatically and record the waveform during the period around the
trigger time. By turning the Horizontal Position knob in the horizontal control
area in the panel, you can change the horizontal triggering position to obtain the
negative delay, making an easy observation of the waveform before the burr
occurs (see Figure 6- 3).

Figure 6- 3 Capturing a Single Signal

Example 4: Analyze the Details of a Signal


Noise is very common inside most of the electronic signal. To find out what's inside
the noise and reduce the level of noise is very important function our oscilloscope is
capable to offer.
Noise Analysis
The level of noise sometime indicates a failure of electronic circuit. The Peak Detect
functions acts an important role to help you to find out the details of these noise. Here
is how we do it:
(1) Push the Acquire button to display the right menu.
(2) In the right menu, select Acqu Mode as Peak Detect.

51
6.Demonstration
The signal displayed on the screen containing some noise, by turning on Peak Detect
function and changing time base to slow down the incoming signal, any peaks or burr
would be detected by the function (see Figure 6- 4).

Figure 6- 4 Signal with Noises

Separate Noises from the Signal


When focusing on signal itself, the important thing is to reduce the noise level as
lower as possible, this would enable user to have more details about the signal. The
Average function offered by our Oscilloscope can help you to achieve this.
Here are the steps for how to enable Average function.
(1) Push the Acquire button to display the right menu.
(2) In the right menu, select Acqu Mode as Average.
(3) Turn the M knob and observe the waveform obtained from averaging the
waveforms of different average number.
User would see a much reduced random noise level and make it easy to see more
details of the signal itself. After applying Average, user can easily identify the burrs
on the rising and falling edges of some part of the signal (see Figure 6- 5).

52
6.Demonstration

Figure 6- 5 Reduce Noise level by using Average function

Example 5: Application of X-Y Function


Examine the Phase Difference between Signals of two Channels
Example: Test the phase change of the signal after it passes through a circuit network.
X-Y mode is a very useful when examining the Phase shift of two related signals.
This example takes you step by step to check out the phase change of the signal after
it passes a specified circuit. Input signal to the circuit and output signal from circuit
are used as source signals.
For the examination of the input and output of the circuit in the form of X-Y
coordinate graph, please operate according to the following steps:
(1) Set the probe menu attenuation coefficient for 10X and that of the switch in the
probe for 10X (see "How to Set the Probe Attenuation Coefficient" on P11).
(2) Connect the probe of channel 1 to the input of the network and that of Channel 2
to the output of the network.
(3) Push the Autoset button, with the oscilloscope turning on the signals of the two
channels and displaying them in the screen.
(4) Turn the Vertical Scale knob, making the amplitudes of two signals equal in the
rough.
(5) Push the Acquire button to display the right menu.
(6) In the right menu, select XY Mode as ON. The oscilloscope will display the
input and terminal characteristics of the network in the Lissajous graph form.
(7) Turn the Vertical Scale and Vertical Position knobs, optimizing the waveform.
(8) With the elliptical oscillogram method adopted, observe and calculate the phase
difference (see Figure 6- 6).

53
6.Demonstration

The signal must be


centered and kept in the
horizontal direction.

Figure 6- 6 Lissajous Graph

Based on the expression sin (q) =A/B or C/D, thereinto, q is the phase difference
angle, and the definitions of A, B, C, and D are shown as the graph above. As a result,
the phase difference angle can be obtained, namely, q =± arcsin (A/B) or ± arcsin
(C/D). If the principal axis of the ellipse is in the I and III quadrants, the determined
phase difference angel should be in the I and IV quadrants, that is, in the range of
(0 - π /2) or (3π / 2 - 2π). If the principal axis of the ellipse is in the II and IV
quadrants, the determined phase difference angle is in the II and III quadrants, that is,
within the range of (π / 2 - π) or (π - 3π /2).

Example 6: Video Signal Trigger


Observe the video circuit of a television, apply the video trigger and obtain the stable
video output signal display.
Video Field Trigger
For the trigger in the video field, carry out operations according to the following
steps:
(1) Push the Trigger Menu button to display the right menu.
(2) In the right menu, select Type as Single.
(3) In the right menu, select Single as Video.
(4) In the right menu, select Source as CH1.

54
6.Demonstration
(5) In the right menu, select Modu as NTSC.
(6) In the right menu, press Next Page, select Sync as Field.
(7) Turn the Vertical Scale, Vertical Position and Horizontal Scale knobs to obtain
a proper waveform display (see Figure 6- 7).

Figure 6- 7 Waveform Captured from Video Field Trigger

55
7.Troubleshooting

7. Troubleshooting
1. Oscilloscope is powered on but no Display.
 Check whether the power connection is connected properly.
 Restart the instrument after completing the checks above.
 If the problem persists, please contact us and we will be under your service.
2. After acquiring the signal, the waveform of the signal is not displayed in the
screen.
 Check whether the probe is properly connected to the signal connecting wire.
 Check whether the signal connecting wire is correctly connected to the BNC
(namely, the channel connector).
 Check whether the probe is properly connected with the object to be measured.
 Check whether there is any signal generated from the object to be measured (the
trouble can be shot by the connection of the channel from which there is a signal
generated with the channel in fault).
 Make the signal acquisition operation again.
3. The measured voltage amplitude value is 10 times or 1/10 of the actual value.
Look at the attenuation coefficient for the input channel and the attenuation ration of
the probe, to make sure they are match (see "How to Set the Probe Attenuation
Coefficient" on P11).
4. There is a waveform displayed, but it is not stable.
 Check whether the Source item in the TRIG MODE menu is in conformity with
the signal channel used in the practical application.
 Check on the trigger Type item: The common signal chooses the Edge trigger
mode for Type and the video signal the Video. If Alternate trigger is selected,
both of the channel 1 and channel 2 trigger levels should be adjusted to the
proper position. Only if a proper trigger mode is applied, the waveform can be
displayed steadily.
5. No Display Responses to the Push-down of Run/Stop.
Check whether Normal or Signal is chosen for Polarity in the TRIG MODE menu and
the trigger level exceeds the waveform range.
If it is, make the trigger level is centered in the screen or set the trigger mode as Auto.
In addition, with the Autoset button pressed, the setting above can be completed
automatically.
6. The displaying of waveform seems getting slow after increasing AVERAGE value
in Acqu Mode (see "How to Set the Sampling/Display" on P27 ), or a longer duration
is set in the Persist in Display (see "Persist" on P27).
It's normal as the Oscilloscope is working hard on many more data points.

56
8.Technical Specifications

8. Technical Specifications
Unless otherwise specified, the technical specifications applied are for the
oscilloscope only, and Probes attenuation set as 10X. Only if the oscilloscope fulfills
the following two conditions at first, these specification standards can be reached.
 This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
 If change of the operating temperature is up to or exceeds 5℃, do a
"Self-calibration" procedure (see "How to Implement Self-calibration" on P12).
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Performance Characteristics Instruction
Bandwidth 100 MHz
Channel 2 channels
Mode Normal, Peak detect, Averaging
Acquisition Sample rate
1 GS/s
(real time)
Input coupling DC, AC , Ground
Input impedance 1 MΩ±2%, in parallel with 20 pF±5 pF
Input coupling 1X,10X,100X,1000X
Max. input voltage 400V (DC+AC, PK - PK)
Input Channel –channel 50Hz: 100 : 1
isolation 10MHz: 40 : 1
Time delay between
150ps
channel(typical)
Bandwidth limit 20 MHz, full bandwidth
Sampling rate range 0.5 S/s~1 GS/s
Interpolation (Sinx)/x
Max Record length 10K
2 ns/div – 1000 s/div,
Scanning speed (S/div)
Horizontal step by 1 – 2 - 5
System Sampling rate / relay
±100 ppm
time accuracy
Single:
Interval(△T) accuracy ±(1 interval time+100 ppm×reading+0.6 ns);
(DC - 100MHz) Average>16:
±(1 interval time +100 ppm×reading+0.4 ns)
Vertical Resolution
8 bits (2 channels simultaneously)
(A/D)
Vertical system Sensitivity 5 mV/div~5 V/div
±2 V (5 mV/div – 100 mV/div)
Displacement
±50 V (200 mV/div – 5 V/div)

57
8.Technical Specifications
Performance Characteristics Instruction
Analog bandwidth 100 MHz
Single bandwidth Full bandwidth
Low Frequency ≥10 Hz (at input, AC coupling, -3 dB)
Rise time (at input,
≤ 3.5 ns
Typical)
DC gain accuracy ±3%
Delta Volts between any two averages of
≥16 waveforms acquired with the same
DC accuracy (average)
scope setup and ambient conditions (△V):
±(3% reading + 0.05 div)
Waveform inverted ON/OFF
△V, △T, △T&△V between cursors,
Cursor
auto cursor
Period, Frequency, Mean, PK-PK, RMS, Max,
Min, Top, Base, Amplitude, Overshoot,
Preshoot, Rise Time, Fall Time, +Pulse
Width, -Pulse Width, +Duty Cycle, -Duty
Automatic
Cycle, Delay A→B , Delay A→B , Cycle
RMS, Cursor RMS, Screen Duty, Phase,
Measurement
+Pulse Count, -Pulse Count, Rise Edge Count,
Fall Edge Count, Area, and Cycle Area.
Waveform Math +, -, *, / ,FFT
Waveform storage 16 waveforms
Bandwidth Full bandwidth
Lissajous
Phase
figure ±3 degrees
difference
Communication
USB 2.0 (USB storage)
port
Counter Support

Trigger:

Performance Characteristics Instruction


Trigger level range Internal ±5 div from the screen center
Trigger level
Internal ±0.3 div
Accuracy (typical)
Trigger
According to Record length and time base
displacement
Trigger Holdoff
100 ns – 10 s
range

58
8.Technical Specifications
Performance Characteristics Instruction
50% level setting
Input signal frequency ≥ 50 Hz
(typical)
Edge trigger slope Rising, Falling
Support standard NTSC, PAL and SECAM
Modulation
Video Trigger broadcast systems
Line number range 1-525 (NTSC) and 1-625 (PAL/SECAM)

General Technical Specifications


Display
Display Type 7" Colored LCD (Liquid Crystal Display)
Display Resolution 800 (Horizontal) × 480 (Vertical) Pixels
Display Colors 65536 colors, TFT screen
Output of the Probe Compensator
Output Voltage
About 5 V, with the Peak-to-Peak voltage ≥1 MΩ.
(Typical)
Frequency (Typical) Square wave of 1 KHz
Power
Mains Voltage 100 - 240 VACRMS, 50/60 Hz, CAT Ⅱ
Power Consumption < 15 W
Fuse 2 A, T class, 250 V
Environment
Working temperature: 0 ℃ - 40 ℃
Temperature
Storage temperature: -20 ℃ - 60 ℃
Relative Humidity ≤ 90%
Operating: 3,000 m
Height
Non-operating: 15,000 m
Cooling Method Natural cooling

Mechanical Specifications
Dimension 301 mm× 152 mm×70 mm (L*H*W)
Weight About 1.1 kg

Interval Period of Adjustment:


One year is recommended for the calibration interval period.

59
9.Appendix

9. Appendix
Appendix A: Enclosure
(The accessories subject to final delivery.)

Standard Accessories:

Power Cord CD Rom USB Cable Probe


Probe Adjust

Appendix B: General Care and Cleaning


General Care
Do not store or leave the instrument where the liquid crystal display will be exposed
to direct sunlight for long periods of time.
Caution: To avoid any damage to the instrument or probe, do not exposed it to any
sprays, liquids, or solvents.
Cleaning
Inspect the instrument and probes as often as operating conditions require.
To clean the instrument exterior, perform the following steps:
1. Wipe the dust from the instrument and probe surface with a soft cloth. Do not
make any scuffing on the transparent LCD protection screen when clean the LCD
screen.
2. Disconnect power before cleaning your Oscilloscope. Clean the instrument with a
wet soft cloth not dripping water. It is recommended to scrub with soft detergent
or fresh water. To avoid damage to the instrument or probe, do not use any
corrosive chemical cleaning agent.

Warning: Before power on again for operation, it is required to confirm that


the instrument has already been dried completely, avoiding any
electrical short circuit or bodily injury resulting form the moisture.

60

You might also like