Chroma Meter Chroma Meter
Chroma Meter Chroma Meter
Chroma Meter Chroma Meter
CR-400/410
CR-410
CR-400
Introducing the successor to the Konica Minolta CR-300/310,
our best-selling colorimeter globally accepted as the standard
in a wide range of industries.
CR-400
Measurement area Ø 8mm
CR-410
Measurement area Ø 50mm
CR-400 CR-410
Data Processor DP-400
The measuring head can perform Easy-to-understand the name on the buttons, ensure
smooth measurement and setting operations.
measurement alone.
The measuring head is detachable from the data processor. Achieves exceptional accuracy
Now, you can take measurements directly with the head Inter-instrument agreement : CR-400: ∆E*ab within 0.6
alone. What's more, you can connect the measuring head CR-410: ∆E*ab within 0.8
directly to a PC. Simply install our optional software, and Repeatability : within ∆E*ab 0.07
your PC can function as the data processor.
User calibration function ensures higher accuracy.
User-defined evaluation formulas (Settings can be configured with the data processor or via a PC with optional software
installed.)
When measuring With the varied accessories, you can measure targets with
diverse profiles.
powders or pastes
Note: The evaluation formula and grade indicated above are hypothetical examples used only to demonstrate the user index function.
When a compact The measuring head can be used independently of the data
colorimeter processor. This is advantageous when portability is required or
is needed in the field limited space is available.
When measurements
need to be printed The compact data processor features a built-in printer for
on-site for labeling superior mobility.
of samples
System Diagram
Optional Accessories
Granular-Materials Attachment Glass Light-Projection Tube Pivoting Base CR-A12 (For CR-400)
CR-A50 CR-A33f (For CR-400) and CR-A33e (For CR-410) Attaching the Pivoting Base CR-A12 to the
With the Granular-Materials Attachment Glass Light-Projection Tube CR-A33f and Measuring head of the CR-400 ensures
CR-A50, the color of powders, pastes, CR-A33e have a glass plate at the tip and greater stability and accuracy in measurements
grains, and other granular substances can can be used for measuring wet surfaces or Light-Projection Tube CR-A33c is also included.
be easily and accurately measured. for ensuring that materials such as textiles
are flat during measurements.
Color Data Software CR-S3w ChromaMagic Ver.1.1 or later CR-400 Utility Software CR-S4w
(Not Available in USA and Latin America)
Enables comprehensive color analysis from To take measurements or change the
incoming raw materials through all phases of measurement parameters of the CR-400/410
the manufacturing and production processes. Series, you can control the unit with a PC.
Supports the eight universally accepted color Measurement data can be transferred directly
spaces, and provides easy-to-understand to a Microsoft Excel® file by means of the OLE
displays of color control data. function. (Excel 97/2000/2002 is required to use the Excel transfer function.)
® ®
Improves color control efficiency and reduces Calibration data and color-difference
costs. reference color data can be uploaded or
• Enables accurate and quick display of pass/fail results and modified.
alarm levels.
• Displays trend graph and statistical compilation of measured
data.
• Easily exports data to spreadsheet applications.
• Various graphs and data display windows can be viewed at
one time.
• Built-in color space and color indices offer complete
solutions for a wide range of applications and industries. System requirements
CIE L*a*b* Tint(CIE) OS Windows® 95/98/2000/XP, WindowsNT®4.0
CIE L*C*h Tint(CIE) Delta CPU Pentium 166MHz or higher
CMC WI(ASTM E313-1973)
CIE L*u*v* WI(ASTM E313-1973) Delta Memory 32MB or more
HUNTER Lab WI(ASTM E313-1996) Hard disk 100MB or more free space
CIE94 WI(ASTM E313-1996) Delta Display resolution SVGA (800× 600) or higher
XYZ/Yxy WI(Berger)
FMC2 WI(Berger) Delta Parallel port or USB port to connect protecting key
Dominant WL WI(CIE)
Excitation Purity WI(CIE) Delta Notes for users of Windows®95 and Windows NT® 4.0
NBS 100 WI(Hunter)
NBS 200 WI(Hunter) Delta (1) The protect key for the USB port connection cannot be used.
Rx WI(Stensby) (2) When using Windows®95, you must use Internet Explorer Version4.0 or higher.
Rx Delta WI(Stensby) Delta (3) When using Windows NT® 4.0, you must use SP4 or higher, or Internet Explorer
Ry WI(Taube)
Ry Delta WI(Taube) Delta
Version 4.0 or higher.
System requirements
Rz YI(ASTM D1925)
Rz Delta YI(ASTM D1925) Delta OS Windows® 98/2000/XP
Strength:Tristimulus(%) YI(ASTM E313-1973) CPU Pentium 166MHz or higher
Strength:Tristimulus X(%) YI(ASTM E313-1973) Delta
Strength:Tristimulus Y(%) YI(ASTM E313-1996) Windows®, Windows NT®, Excel®, Internet Explorer are trademarks or registered Memory 32MB or higher
Strength:Tristimulus Z(%) YI(ASTM E313-1996) Delta trademarks of Microsoft Corporation of America or its subsidiaries. Hard disk 100MB or more free space
Tint(ASTM E313-1996) YI(DIN 6167) Pentium is a trademark or registered trademark of Intel Corporation of America or its
subsidiaries. Display resolution VGA (640× 480) or higher
Tint(ASTM E313-1996) Delta YI(DIN 6167) Delta
Specifications Dimensions Units : mm
Name Chroma Meter Measuring Head
Model CR-400 Head CR-410Head
Measuring Head CR-400 Measuring Head CR-410
Illuminating/viewing system d/0 (Diffuse illumination/0° viewing angle) Wide-area illumination/0° viewing angle 63 102 63 102
(Specular component included) (Specular component included)
Detector Silicone photo cells (6)
Display range Y: 0.01 to 160.00% (reflectance)
Light source Pulsed xenon lamp
Measurement time 1 seconds.
Minimum measurement interval 3 seconds.
Battery performance Approx. 800 measurements
(when using batteries under company testing Minolta's conditions)
217
Measurement/illumination area φ8/φ11 φ50/φ53
Within ∆E*ab0.07 standard deviation (when the white calibration plate
244
Repeatability
is measured 30 times at intervals of 10 seconds)
Inter instrument ∆E*ab: within 0.6 ∆E*ab: within 0.8
agreement Average of 12 BCRA series II colors
Observer 2 degrees Closely matches CIE 1931 Standard Observers: ( x2λ, yλ, zλ)
Illuminant ∗1 C, D65
40.2
Display ∗1 Chroma values, color difference values, PASS/WARN/FAIL display
67.5
Tolerance judgment ∗1 Color difference tolerance (box tolerance and elliptical tolerance)
Color space/ XYZ, Y x y, L∗a∗b∗, Hunter Lab, L∗C∗h, Munsell (only illuminant C), CMC(l:c), CIE1994, Lab99, 32
colorimetric data LCh99, CIE2000, CIE WI•Tw (only illuminant D 65), WI ASTM E313 (only illuminant C),
YI ASTM D1925 (only illuminant C), YI ASTM E313 (only illuminant C), 32
User index (up to six can be registered from computer) Data Processor DP-400
Languages Operating keys : English 255
LCD : English (default)
(LCD : German, French, Italian, Spanish, Japanese) *1
Storable data sets 1000 (measuring head and data processor save different data)
Color difference target colors 100
100
Calibration channels *1 20 channels (ch00 : white calibration, ch01 to ch19 : user calibration)
Display Dot-matrix LCD with back light (15 chars x 9 lines + 1 line for icon display)
Interface RS-232C compliant (for data processor/PC)
* Baud rate : 4800, 9600, 19200 (bps), set at 9600 bps when shipped from factory
Power source 4 AAA size alkaline or Ni-MH batteries,
AC adapter (AC-A17) AC120V 50-60Hz 0.4A (for N.America and Japan)
AC230V 50-60Hz 0.4A (for worldwide except N.America)
Size 102(W) x 217(H) x 63(D)mm 102(W) x 244(H) x 63(D)mm
73
Weight Approx. 550g Approx. 570g
(including 4 AAA size batteries and not including RS-232C cable)
Operating temperature/ 0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation
humidity range
Storage temperature/humidity range -20 to 40°C, relative humidity 85% or less (at 35°C) with no condensation
00 ead
10 ead
Other LCD back light ON/OFF function (when ON, back light stays ON for 30
00 r
o
DP o c e s s
seconds after last key or measurement operation)
CR u r i n g
CR ring
∗1 indicates when connected to the Data Processor or when not set using the Data Processor or the optional software, Standard/Optional
u
-4
-4
r
-4
as
as
ta P
that some of the function are not available when the measuring head is not connected. accessories
Me
Me
Da
Name Data Processor Color Data Software
Model DP-400 ChromaMagic CR-S3w
CR-400 Utility Software
Display range Y : 0.01 to 160.00% (reflectance) CR-S4w
Measurement time ∗2 1 Seconds. White Calibration Plate
Minimum measurement interval ∗2 3 Seconds. CR-A43
Battery performance Approx. 800 measurements (when using batteries under company testing Minolta's conditions) White Calibration Plate
Illuminants C, D65 CR-A44
Display Chroma values, color difference values, color difference graphs, PASS/WARN/FAIL display Protective Cap
CR-A72
Tolerance judgment ∗2 Color difference tolerance (box tolerance and elliptical tolerance) Only for the display function Protective Cap
Color space/ XYZ, Y x y, L∗a∗b∗, Hunter Lab, L∗C∗h, Munsell (only illuminant C), CMC (l:c), CIE1994, Lab99, CR-A104
colorimetric data LCh99, CIE2000, CIE WI-Tw (only illuminant D65), WI ASTM E313 (only illuminant C), RS-232C Cable
YI ASTM D1925 (only illuminant C), YI ASTM E313 (only illuminant C), CR-A101(Head-DP)
User index (up to six registered in the Measuring Head can be used) RS-232C Cable
Languages Operating keys : English, LCD : English (default), German, French, Italian, Spanish, Japanese CR-A102(for PC)
AC Adapter
Storable data sets Max. 2000 pieces of data (divisible into 100 pages) AC-A17
Deletion and Undoing selected stored data (one piece of data or all data) are possible Wrist Strap
Color difference target colors ∗2 Only for the operating function (100 pieces of data when the measuring head is connected; input of CR-A73
measurement values or numeric) (independent of page function) Shoulder Strap
Calibration channels ∗2 Only for the operating function (20 channels when the measuring head is connected) SS-01
(ch00: white calibration; ch01 to ch19: user calibration) Hard Case
CR-A103
Page function 100 pages
Display Dot-matrix LCD with back light (16 chars x 9 lines + 1 line for icon display) Contrast adjustment Roll Paper (one roll)
Printer 384 dot line thermal printer (can also print graphs) Automatically prints out all measurement results (can be set not to print) Roll Paper
Statistical function Maximum, minimum, average, and standard deviation DP-A22(five rolls)
Automatic measurement ∗2 Date and time display: year, month, day, hour, minute 4 AA Size Batteries
Timer: 3seconds. to 99 minutes.
(Some measurement modes require more than 3 seconds.) 4 AAA Size Batteries
Interface RS-232C compliant Baud rate (bps) : 19200 fixed (when connected to PC) Glass Light-Projection Tube
When measuring head is connected baud rate is automatically set to that of the measurement head CR-A33a/f
Power source 4 AA size alkaline or Ni-MH batteries, Light-Projection Tube
AC adapter (AC-A17) AC120V 50-60Hz 0.4A (for N.America and Japan) CR-A33c/d
AC230V 50-60Hz 0.4A (for worldwide except N.America) Glass Light-Projection Tube
CR-A33e
Size 100(W) x 73(H) x 255(D)mm Granular-Materials Attachment
Weight Approx. 600g (not including batteries and paper) CR-A50
Operating temperature/ 0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation Pivoting Base
humidity range CR-A12
Storage temperature/humidity range -20 to 40°C, relative humidity 85% or less (at 35°C) with no condensation Color Tiles
Other User calibration function (multi-calibration/manual calibration) ∗2, Measurements for automatic average
function, Print ON/OFF function. CR-400 measurement data import function ∗2, All color space print ON/OFF Standard accessory
function, Data protection ON/OFF function. Back light ON/OFF function. Buzzer ON/OFF function. Display Optional accessory
color limit function, Remote mode (stored data output), Character input function (alphanumeric)
∗2 indicates that part of or all functions are not available when the measurement head is not connected. Specifications are subject to change without notice.