Guidelines for
Troubleshooting
and Maintenance of
ICP-MS Systems
Presented by Eric Vanclay,
Atomic Spectroscopy
Consumables Product
Manager
ICP-MS Maintenance & Trouble Shooting Mar. 2014
Common ICP-MS Problems Reported by Customers
Sensitivity:
• Sensitivity is worse than it used to be
• I have a new application and I can’t get the sensitivity I need
• How come I can’t get the instrument to meet published detection limits?
Precision
• Sensitivity is acceptable but precision is terrible
High noise
• Can get the “right answers”, but very noisy signal – this is also giving bad precision.
Accuracy
• Instrument does not give the “right” results.
Poor Sample Throughput
• The instrument throughput needs to improve
• Nebulizer and/or interface cones blocks too quickly
ICP-MS Maintenance & Trouble Shooting Mar. 2014
2
Causes of Poor ICP-MS Sensitivity
Worn pump tubing
Sample Blocked nebulizer
introduction
system Blocked injector in torch
Worn or deformed torch
Blocked interface cones
Poor optimization – especially the neb. flow
Optimization Method setting – tuning?
Torch alignment
Wrong tubing type
Standard
High blank level
(& sample) Standards prepared correctly?
preparation
Samples prepared correctly? – internal standard selection
Reaction gas purity
ICP-MS Maintenance & Trouble Shooting Mar. 2014
3
Causes of Poor ICP-MS Precision
Worn pump tubing
Sample Beading in spray chamber
introduction
system Nebulizer condition and performance
Air leaks in transfer tubing
Sample line not grounded
Torch alignment
Interface cones not “conditioned” with your sample matrix
Optimization
Poor optimization – especially the neb. flow
Method setting – appropriate delay times?
Nebulizer choice for your samples
Standard
(& sample) Wash-out (memory effects)
preparation
ICP-MS Maintenance & Trouble Shooting Mar. 2014
4
Causes of Poor Accuracy in ICP-MS
Worn pump tubing
Sample
introduction Contamination from pump tubing
system
Blockage in nebulizer and/or torch
Blockage in interface cones
Interferences
Optimization Poor optimization – especially the neb. flow
Choice of internal standard
Insufficient stabilization time
Standard preparation
Standard Incomplete digestion – particles in solution
(& sample)
preparation No matrix matching
Wash-out (memory effects)
ICP-MS Maintenance & Trouble Shooting Mar. 2014
5
ICP-MS Sample Introduction System Tips
Do:
Check performance report each analysis
Don’t:
Assume system is still optimized
Check/monitor the nebulizer uptake Assume nebulizer flow rate is the same
Check/adjust the peri pump tubing Overtighten the pressure adj. screw
Check the blank reading Use a simple water blank
Rinse between samples & at the end of the
run
- Rinse should match sample matrix
Clean the torch/nebulizer/cones regularly Wait until you have blockage before
- Inspect condition of the nebulizer tip cleaning
- Inspect condition of the interface cones
Condition the interface cones with your
sample matrix before analysis
ICP-MS Maintenance & Trouble Shooting Mar. 2014
6
Drain tubing 1.52 mm id
Peri Pump Tubing Tips
Agilent P/N G1833-65570
- Tubing diameters
• Want tubing used for waste to be larger id than sample id
- Chemical compatibility
• Ensure tubing is resistant to the solvent being used
- Replace frequently
• Preclean new tubing to remove potential contamination
• Using “old” tubing can lead to problems with precision and stability
- Can also contribute to nebulizer blockage (if inside lining breaks down)
• Typical lifetime is ~5 days based on normal 8 hour working day Sample tubing 1.02 mm id
- Detach from tube holder after use – allows tube to “relax” Agilent P/N G1833-65569
- Maintaining tubes – What to check?
• Check 2 key things on pump tubing
- Roundness of tube – should not be any “flat” spots
- Tubing should still be elastic – replace if obviously stretched
• Don’t over tighten – just need smooth and even sample flow
- Remember to check other tubing for wear, leaks and crimps
ICP-MS Maintenance & Trouble Shooting Mar. 2014
7
Peri Pump Tubing Tips
Symptoms:
- Peri pump tubing that looks/feels worn or has a strange color
• IF IN DOUBT, CHANGE IT
– Erratic liquid flow
• Check tension from clamps
– Bubbles in the liquid stream
• Check all gas fittings, tubing and connectors – deposits, burrs, damage
– Spurting Nebulizer or disconnecting tubing segments
• Plugging in the transfer line. Requires cleaning or replacement.
ICP-MS Maintenance & Trouble Shooting Mar. 2014
8
Sample Introduction
– Nebulizer Tips
Think “PREVENTION”
- Micro-flow nebulizers
- Zero tolerance to undissolved solids
• Plugging of annulus and/or capillary
• Filter/Centrifuge/gravitational settling – sampler positioning
• Use only lintless wipes
- Autosampler enclosures
- Rinse at least 10 minutes with a reagent blank before
extinguishing plasma
1. Image modified from “Pneumatic Nebulisers and Spray Chambers for Inductively Coupled Plasma Spectrometry. A Review, Part 1. Nebulisers“
by Barry Sharp, JAAS, vol.2, p. 613-652, 1988
2. Image provided by Meinhard Glassblowing Products
ICP-MS Maintenance & Trouble Shooting Mar. 2014
9
Cleaning the Nebulizer
Never sonicate or attempt to clean with wire!
For normal cleaning:
• Soak in 5% nitric acid for ~10 mins.
To remove a nebulizer blockage:
Suction
• Reverse pump the nebulizer with the tip in solvent; OR
• Apply suction from the wide end of the capillary using a vacuum
aspirator; OR
• Apply high pressure clean air via a tubing snugly fitted over the
nebulizer tip (use with caution); OR
• Use a dedicated nebulizer cleaning tool to force methanol solution Pressure
through the tip
For salt deposits:
• Soak the nebulizer overnight in a beaker of 25% Fluka RBS-25
detergent. Rinse with pure water
For “stubborn” deposits: Nebulizer cleaning tool
• Soak the nebulizer overnight in conc. nitric acid. Use a pipette to P/N G3266-80020
ensure there are no air bubbles in capillary. Rinse with pure water
ICP-MS Maintenance & Trouble Shooting Mar. 2014
10
Cleaning the Spray Chamber
For routine cleaning, soak the end
cap and spray chamber in 5% nitric
acid for ~30 mins
• Rinse, allow to dry and refit
If you see precision problems or can
see “beading” or droplet formation on
the walls of the spray chamber:
• Soak overnight in a 25% detergent
solution
- Best to leave it soaking for 24 hours
- Use any laboratory detergent
e.g. Fluka RBS25, Triton X-100, Decon
90 etc
• Rinse, allow to dry and refit
ICP-MS Maintenance & Trouble Shooting Mar. 2014
11
Cleaning the Torch
Visually check the torch, bonnet and
shield when removing the torch
• Replace if deformed or chipped
Do not sonicate!
Soak in >5% nitric acid for ~30 mins.
• For more stubborn stains, soak in aqua regia
(1:3 HNO3:HCl)
• For salt deposits:
- Rinse with water to remove deposits
- Soak the torch overnight in a beaker of 25% Fluka
RBS-25 detergent
Rinse and allow to dry
Caution! Reinstall only when dry
Torch damage due to incomplete drying
ICP-MS Maintenance & Trouble Shooting Mar. 2014
12
Re-installing the Torch
Refit the torch shield & torch bonnet
Replace the torch into the torch
holder
Ensure the torch projection fits into
the slot on the torch holder
Can check the alignment of the RF
coil when re-installing the torch
Reconnect gas fittings and transfer
tube from spray chamber
Test plasma ignites and instrument
switches to “analysis” mode
• If plasma fails to ignite, check all
connections for possible air leaks
ICP-MS Maintenance & Trouble Shooting Mar. 2014
13
Interface Cone Choices
Sampler and skimmer cones extract a portion of the plasma ion
population and transfer this into the higher vacuum mass
spectrometer for detection
- Nickel cones are standard
• Inexpensive
• Good thermal and chemical resistance
• Uses standard brass skimmer base
• Typically use 3-5/year (based on ~350 samples/day)
- Ni plated sampling cone (optional)
• Used when samples containing > 0.5% HCl are analyzed
• Also used for routine operation with HMI aerosol dilution
• Requires stainless steel skimmer base
- Pt tipped cones are optional
• Essential for users wanting to analyze aggressive acids
(esp. HF digests)
• Also used when O2/Ar option gas is used to analyze organic
solvents
- Pt skimmer with the Ni base used for organics analysis –
requires brass skimmer base
- Pt skimmer with the Cu base used for best DLs and high
matrix samples (incl. HF) - requires brass skimmer base
- Typically use 1-2/year (based on ~350 samples/year)
ICP-MS Maintenance & Trouble Shooting Mar. 2014
14
Interface Cone Cleaning –Why?
The necessity to clean the cones depends on your
(in)tolerance limits for:
• Sensitivity – are orifices getting plugged?
• Long term precision – are cones adequately conditioned?
• High background or distorted peak shapes
• Interface vacuum changing – deposition?
Close up of sampler
Other reasons to clean the cones? cone after measuring
75 undiluted sea
• If there is a build up of deposits on the orifice (should be circular and water samples
free of deposits) using HMI
• If the orifice of the cone is discolored (due to excess heat)
A conditioned cone has a uniform coating that leads to long term stability
If analyzing different sample types where a major element in the first sample type
is a trace element in the second, more than one cleaning step is required
If analyzing the same type of samples, clean to remove only superficial deposits.
Step 1 may be all that is required
ICP-MS Maintenance & Trouble Shooting Mar. 2014
15
Interface Cone Cleaning – Step 1
For normal contamination:
Just need a simple clean with
pure water:
• Dip a cotton swab in pure water and
clean both sides of the cone
• Rinse with pure water
• Ultrasonic clean the cones in pure
water for more than 5 minutes
Agilent P/N 9300-2574
ICP-MS Maintenance & Trouble Shooting Mar. 2014
16
Agilent P/N 5188-5359
Interface Cone Cleaning – Step 2
For normal contamination:
If performance is still not
satisfactory, clean with a 2%
Citranox solution
(NOT MORE THAN 2%!)
• Ultrasonic clean the cones in a 2%
Citranox solution for 2 to 3 minutes
• Rinse with pure water
• Ultrasonic clean in pure water for an
additional 2 to 3 minutes
ICP-MS Maintenance & Trouble Shooting Mar. 2014
17
Agilent P/N 5188-5359
Interface Cone Cleaning – Step 3
ONLY For more severe
contamination:
Clean with a 2% nitric acid
solution
• Dip a cotton swab in 2% nitric acid and
clean both sides of the cone
(DO NOT SOAK IN ACID!)
• Rinse with pure water
• Ultrasonic clean in pure water for
2 to 3 minutes
• Rinse with pure water
• Ultrasonic clean in pure water
Pitted nickel cone from effect
for an additional 2 to 3 minutes of HNO soak (left side) and
3
clean machined metal on right.
ICP-MS Maintenance & Trouble Shooting Mar. 2014
18
Re-installing the Cleaned Cones
Check the condition of the graphite
gasket and replace if necessary
Refit the skimmer cone using the
removal tool
Refit the sample cone and tighten by
hand
Check the vacuum levels to confirm
correct installation
• Interface pressure: 500 Pa (~4 torr,
0.005 atm)
• Analyzer pressure: 0.002 Pa (~1.5 x10-5 torr,
2 x10-8 atm)
If necessary, retension the skimmer and
sample cones
ICP-MS Maintenance & Trouble Shooting Mar. 2014
19
Recommended Procedures at End of the Day
1. Aspirate acid rinse solution for a few minutes before shutting off
the plasma
- Helps to prevent sample deposition inside the nebulizer after the run
2. Extinguish the plasma and switch off the chiller
3. Remove the sample capillary from the rinse, start the pump again and
pump any remaining rinse solution from the spray chamber
4. Release the pressure bars on the pump tubing and remove the bridges
from the securing slot
- Ensure the tubes are no longer stretched over the pump rollers
5. Empty waste vessel
6. a) Close the current worksheet – leave Mass Hunter S/W running
b) Leave main power and argon on
- Keeps instrument in stand-by mode (ensures fastest start-up)
ICP-MS Maintenance & Trouble Shooting Mar. 2014
20
Agilent ICP-MS Performance - Benefits
Best matrix tolerance & less interferences
• Handles tough sample matrices better Relative signal (suppression) in undiluted seawater
than any other ICP-MS
- Highest plasma temperature (lowest
CeO/Ce) as standard
- HMI (high matrix introduction) allows
direct analysis of difficult samples
e.g. sea water
• ORS3 operates in He collision mode, to
remove ALL polyatomic interferences Long-term stability in undiluted seawater
- Enables reliable removal of all polyatomic
interferences using a single set of
conditions
- Gives lower levels of interference and HMI gives lower suppression and
better long-term stability better long-term stability
ICP-MS Maintenance & Trouble Shooting Mar. 2014
21
ICP-MS – Potential Autosampler Issues
- More customers use autosamplers for automation
- Issues to consider:
• Longer transfer tube between sampler and ICP-MS
- May need to program a longer sample uptake delay
- May exacerbate problems with memory effects
• Ensure probe diameter is appropriate for sample matrix
- Use wider bore for high % TDS or viscous samples
• Sample stability - potential for sample changes while uncovered in racks – impacts accuracy
- Dust ingress can introduce contamination
- Sample evaporation may occur during long unattended runs
- Sediment in the sample may settle out, esp. with wear metals or suspensions
• Ensure transfer line to ICP-MS is in good condition
- Kinks in the line may cause poor uptake, or pulsing in the sample
- Impacts on precision and accuracy
ICP-MS Maintenance & Trouble Shooting Mar. 2014
22
ICP-MS – Recommended Maintenance Schedule
Daily:
Argon gas pressure
Check peristaltic pump tubing for damage/deterioration
Visual check of glassware
(connections OK, no filling of spray chamber or connector)
Visual inspection of sample cone exterior (orifice shape & deposition)
Frequently, as needed -- perform these operations:
Empty the drain reservoirs
Thorough visual inspection of interface cones
Check nebulization
Replace peristaltic pump tubing
Clean/replace torch
Check recirculation water level
Frequency and extent of maintenance depends on the usage of the instrument:
this overview assumes daily use, 8 hours/day.
For systems run 24/7, more frequent maintenance is required.
ICP-MS Maintenance & Trouble Shooting Mar. 2014
23
EMF (Early Maintenance Feedback)
EMF window shows usage of various components and
predicts when to perform maintenance (user definable)
ICP-MS Maintenance & Trouble Shooting Mar. 2014
24
ICP-MS System Tips – User Log
• Use the “Maintenance Log” to record
routine and non routine maintenance
activities
• Maintenance log can track:
- When the maintenance activity was
completed
- Operator who completed the maintenance
- Type of maintenance activity
- Any operator comments
ICP-MS Maintenance & Trouble Shooting Mar. 2014
25
Summary – Key Consumables for ICP-MS
Sample preparation/presentation:
- Peristaltic pump tubing
- Transfer and drain tubing
- ICP-MS standard solutions
- Internal Standard solutions
- Torches
- Spray chambers
- Nebulizers
Ion Extraction:
- Sampler and skimmer cones
Autosampling:
- Sample tubes, racks, probes and transfer tubing
ISIS:
- Peristaltic pump tubing, ferrules & fittings
ICP-MS Maintenance & Trouble Shooting Mar. 2014
26
Agilent ICP-MS Consumable Kits
Part Number Description Content
Peristaltic pump tubing for sample intro Graphite gasket for sampling cone
G3280-67003 Basic Consumables Kit for 7700x Sample tubing (id 0.5 mm) Rotary pump oil (I L)
Torch (2.5 mm)
Peristaltic pump tubing Long life shield plate
Sample tubing (id 0.5 mm) Bonnet
Online ISTD addition kit plasma gas tubing and gas line connector
Drain tube assembly Graphite gasket for sampling cone
Spray Chamber (G3280-80008) 2 Ni Sampling and Skimmer cones
G3280-67004 Comprehensive Spares Kit for Ball joint connector Screw and spacer kit for x-lens
7700x/e Gas connector for dilution gas port O-ring for cell
Carrier gas connector for Micro Mist Octopole assembly
Plug for dilution gas port Polishing paper for ion lens
Torch (G3280-80001) Oil mist filter element
Clamp Rotary pump oil
Peristaltic Pump tubing for sample intro
Spray chamber (G3280-80008) Graphite gasket for sampling cone
Ball joint connector PT sampling and sklimmer cone (one each)
Gas connector for dilution gas port Screw and spacer kit for x-lens
G3280-67007 Clamp O-ring for cell
Comprehensive Spares kit for 7700s Torch (G3280-80001) Octopole assembly
Long life shield plate Polishing paper for ion lens
Bonnet Oil mist filter element
Plasma / aux gas line tubing Rotary pump oil
Work coil
Tee, Cross and Union joints
PFA sample tubing (0.3 mm and 0.5 mm id) Teflon nuts for 1.6 mm and 3.0 mm od tubing
PTFE sample tubing (0.8 mm and 2.0 mm id) Front and back ferrules for 1.6 mm and 3.0 mm od
G4911-67001 Comprehensive Spares kit for ISIS Peristaltic Pump tubing for sample intro tubing
Tubing identification tags Tubing clamp
Spiral tubing
There’s also Speciation kits – basic LC connection kit (G1820-65541) & comp. LC connection kit (G1833-65200)
ICP-MS Maintenance & Trouble Shooting Mar. 2014
27
Where to Find the Right Consumable?
Analytical Consumables: Agilent Assist:
Consumables & Supplies Instrument Sales &Services
1-800-227-9770 (Option 1,1) 1-800-227-9770 (Option 1,3)
www.agilent.com/chem/contactus www.agilent.com/chem/contactus
On-Line resources:
ICP-MS Parts & Supplies
2nd Edition Speciation Handbook
ICP-MS Journal Archives
ICP-OES Parts & Supplies
Atomic Absorption Supplies
AA FAQs
Instrument Parts & Supplies
Atomic Spectroscopy Application Notes
Recorded Agilent e-Seminars
Agilent Quick Reference Guide for ICP-MS (pub. # 5990-8182EN)
Agilent Spectroscopy Supplies Catalogue (pub # 5991-1060EN)
Instrument User Manuals
ICP-MS Maintenance & Trouble Shooting Mar. 2014
28
Summary – To Achieve Quality Data
• Most “instrument” failures occur in the sample introduction area
- Includes
• Interface cones
• Peristaltic pump tubing
• Drain Assembly
• Torch
• Spray chamber
• Nebulizer
• Improper maintenance of this area can result in poor data quality
• Frequently less experienced analysts can fail to recognize problems resulting in
productivity losses
• Establishing maintenance procedures can prevent problems
ICP-MS Maintenance & Trouble Shooting Mar. 2014
29