LTQ Orbitrap Discovery Hardware
LTQ Orbitrap Discovery Hardware
LTQ Orbitrap Discovery Hardware
Hardware Manual
“PEEK” polymer is a trademark of Victrex plc. “PEEKSil” is a trademark of SGE International Pty. Ltd.
“SEQUEST” is a registered trademark of University of Washington. “Swagelok” is a registered trademark of the
Crawford Fitting Company. “Teflon” is a registered trademark of E. I. du Pont de Nemours & Co. “Tygon” is a
registered trademark of Saint-Gobain Performance Plastics Company.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document
is for reference purposes only. System configurations and specifications in this document supersede all previous
information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that
might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Printing History: Revision A printed in September 2007.
Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of
this Thermo Scientific instrument requires a team effort for lifting and/or moving the
instrument. This instrument is too heavy and/or bulky for one person alone to handle safely.
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU
Member State, and this product should be disposed of or recycled through them. Further information
on Thermo Fisher Scientific’s compliance with these Directives, the recyclers in your country, and
information on Thermo Fisher Scientific products which may assist the detection of substances
subject to the RoHS Directive are available at www.thermo.com/WEEERoHS.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/
EC erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état
membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage
d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre
pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des
substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.
Read This First
Welcome to the Thermo Scientific, LTQ Orbitrap Discovery™ system!
The LTQ Orbitrap Discovery is a member of the family of LTQ™ mass
spectrometer (MS) hybrid instruments.
Who uses this Guide This LTQ Orbitrap Discovery Hardware Manual is intended for all
personnel that need a thorough understanding of the instrument (to
perform maintenance or troubleshooting, for example). This manual
should be kept near the instrument to be available for quick reference.
Scope of this Guide This LTQ Orbitrap Discovery Hardware Manual contains a description
of the modes of operation and principle hardware components of your
LTQ Orbitrap Discovery instrument. In addition, this manual provides
step-by-step instructions for cleaning and maintaining your instrument.
Changes to the To suggest changes to this manual, please send your comments to:
Manual
Editors, Technical Documentation
Thermo Fisher Scientific
Hanna-Kunath-Str. 11
28199 Bremen
Germany
• Data input
• Admonitions
• Topic headings
Data Input Throughout this manual, the following conventions indicate data input
and output via the computer:
• For brevity, expressions such as “choose File > Directories” are used
rather than “pull down the File menu and choose Directories.”
• Any button that you click on the screen is represented in bold face
letters. For example, “click on Close”.
Admonitions Admonitions contain information that is important, but not part of the
main flow of text.
Topic Headings The following headings are used to show the organization of topics
within a chapter:
Chapter Name
Safety and EMC In accordance with our commitment to customer service and safety, this
Information instrument has satisfied the requirements for the European CE Mark
including the Low Voltage Directive.
Identifying Safety The LTQ Orbitrap Discovery Hardware Manual contains precautionary
Information statements that can prevent personal injury, instrument damage, and
loss of data if properly followed. Warning symbols alert the user to check
for hazardous conditions. These appear throughout the manual, where
applicable, and are defined in Table i.
Instrument-Specific Every instrument has specific hazards, so be sure to read and comply
Hazards with the following precautions. They will help ensure the safe,
long-term use of your system.
2. Only use fuses of the type and current rating specified. Do not use
repaired fuses and do not short-circuit the fuse holder.
3. The supplied power cord must be inserted into a power outlet with a
protective earth contact (ground). When using an extension cord,
make sure that the cord also has an earth contact.
5. Never run the system without the housing on. Permanent damage
can occur.
10. Never try to repair or replace any component of the system that is
not described in this manual without the assistance of your service
representative.
Safety Advice for Possible Hazardous Material Might Contaminate Certain Parts of Your
Contamination System During Analysis.
Figure 1-2 on page 1-3 shows the schematic view of the LTQ Orbitrap
Discovery. The LTQ Orbitrap Discovery consists of three main
components, which are described in the following topics:
• A linear ion trap (Thermo Scientific LTQ XL) for sample ionization,
selection, fragmentation, and AGC™.
Lens stack
1HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the instrument is not equipped with this option.
The mains inlet as well as a power outlet for peripheral devices (data
system, for example) are located at the right side of the instrument.
Rotary pumps are hidden under the linear trap and accessible from the
front. The left side panel and the front panel are mounted on hinges and
the right side panel is removable.
The top lid opens upwards to allow easy access for Field Engineers from
the top. See Figure 1-3.
Specifications The LTQ Orbitrap Discovery has the following measuring properties:
Control Elements The LTQ Orbitrap Discovery is mainly operated from the desktop
computer (data system). Some control elements for important system
functions are located directly on the instrument. They are described in
the following sections.
System Status LEDs Figure 1-4 shows the system status LEDs at the front of the instrument.
Five LEDs indicate the main functions of the system. (See also
Figure 1-5 on page 1-7.) While the Power LED is directly controlled by
the 3 × 230 V input, all other LEDs are controlled by the power
distribution board (Refer to topic “Power Distribution Board” on
page 1-38). Table 1-1 explains the function of the various LEDs.
The system status LEDs at the front panel of the linear ion trap are
described in the LTQ XL Hardware Manual.
Control Panels Figure 1-5 shows the right side of the LTQ Orbitrap Discovery. Located
here are the control panels, switches, and the ports for the external
connections (mains supply, gases, Ethernet communication, and cooling
water).
Switches and
control LEDs
Main power
switch
Power connector
Upper Control Panel The upper instrument control panel comprises the bakeout timer, the
bakeout control buttons, and three circuit breakers. To access the upper
control panel, swing open the small lid (opens from left to right). See
Figure 1-5 and Figure 1-6 on page 1-8.
The timer allows setting the duration for the bakeout of the system.
After the duration is set, the bakeout procedure is started by pressing the
green button on the right. A running bakeout procedure can be stopped
by pressing the orange button on the left side. For instructions about
performing a bakeout, refer to topic “Baking Out the System” on
page 3-4.
Bakeout control
buttons
Circuit breakers
Three circuit breakers are located at the bottom of this control panel.
Table 1-2 shows the parts of the LTQ Orbitrap Discovery that are
protected by the respective circuit breaker. The proper function of each
circuit breaker is signaled by a dedicated LED in the power control
panel (for example, F1 corresponds to L1).
Power Control Panel In addition to the system status LEDs at the front side (see Figure 1-4
on page 1-6), the LTQ Orbitrap Discovery has three power control
LEDs above the Vacuum Pumps switch at the right side. See Figure 1-7.
They indicate whether the corresponding circuit breaker is closed and
the respective parts of the instrument have power. (See Table 1-2 on
page 1-8.)
FT Electronics switch
Figure 1-7. Power control panel with power control LEDs and switches
The use of the switches below the power control LEDs changes the
working mode of the power distribution. (See topic “Working Modes of
the Power Distribution” on page 1-48.)
The Vacuum Pumps switch can be set into the positions ON or OFF.
When the switch is in the OFF position, everything but the multiple
socket outlet is switched off.
The FT Electronics switch can be set into the Operating Position (ON)
or into the Service Position (OFF). When the switch is in the Service
position, all components are switched off with exception of the
following:
• Fans
• Heater control
• Vacuum control
The linear ion trap also remains on because it has a separate Service
switch.
Main Power Switch The main power switch must be turned 90° clockwise/anti-clockwise to
switch on/off the instrument (see Figure 1-8). Placing the main power
switch in the Off position turns off all power to the LTQ Orbitrap
Discovery (and linear ion trap as well, including the vacuum pumps).
On
Off
Note When the main power switch is in the Off position, you can secure
it with a padlock or a cable tie (to prevent unintended re-powering when
performing maintenance, for example). ▲
External Connections Figure 1-9 on page 1-11 shows the lower right side of the instrument
with the external connections for mains supply, gases, cooling water, and
Ethernet communication.
The power connector for the mains supply is located on the center. The
cooling water ports are located below the power connector. (See also
topic “Cooling Water Circuit” on page 1-27.) A Teflon® hose connects
the instrument to the nitrogen gas supply. An analogous port is used for
the HCD collision gas supply.1 Metal tubing connects the instrument
with the helium gas supply. (See also topic “Gas Supply” on page 1-24.)
Located at the top are two ports for Ethernet cables for connecting the
LTQ Orbitrap Discovery and the linear ion trap via an Ethernet hub
with the data system computer.
The exhaust hose from the rotary pumps is led backwards below the
instrument, comes out the back of the instrument, and connects the
pumps to the exhaust system in the laboratory.
1HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the
instrument is not equipped with this option.
Ethernet ports
Power connector
The power outlet for peripheral devices (data system, for example) is
located above the mains supply port. The outlet provides the mains
supply for the peripherals via a multiple socket outlet.
1
Not used in the standard version of the LTQ Orbitrap Discovery.
Linear Ion Trap The LTQ Orbitrap Discovery system can utilize a variety of ionization
techniques such as ESI, APCI, or APPI. Maintenance of the API source,
as well as switching between ionization methods, is vent-free. Ions are
transferred by octapole and “square” quadrupole lenses into an ion trap
that is optimized for axial ion ejection into the curved linear trap. (See
Figure 1-2 on page 1-3.)
All the ion handling, selection and excitation capabilities of the ion trap
can be used to prepare ions for analysis in the Orbitrap. These features
include storage and ejection of all ions, storage of selected masses or
mass ranges, as well as ion isolation. Isolated ions can be excited and
then fragmented as necessary for MS/MS and MSn experiments. The
patented Automatic Gain Control (AGC) provides extended dynamic
range and insures optimized overall performance of the ion trap and
Orbitrap MS.
The linear ion trap and the transfer chamber are mounted on a table.
See Figure 1-1 on page 1-2. The table also serves as a housing for the
forepumps. See Figure 1-18 on page 1-21. The LTQ Orbitrap
Discovery provides power for the linear ion trap – and for the data
system.
The linear ion trap is delivered with power connector, gas lines (He, N2,
and collision gas), and vacuum tube lines extending to the ESI source.
On the rear side of the LTQ XL ion trap is a flange with an O-ring seal.
When the flange is removed, the Orbitrap transfer chamber is mounted
to the flange of the linear ion trap. The transfer chamber is held with
supports on the table. The components of the ion optics and the
Orbitrap are fixed to the transfer chamber.
Orbitrap Analyzer This section describes the basic principle of the Orbitrap analyzer. The
heart of the system is an axially-symmetrical mass analyzer. It consists of
a spindle-shape central electrode surrounded by a pair of bell-shaped
outer electrodes. See Figure 1-10. The Orbitrap employs electric fields
to capture and confine ions.
Figure 1-10. Schematic view of the Orbitrap cell and example of a stable
ion trajectory
Measuring Principle In the mass analyzer shown in Figure 1-10, stable ion trajectories rotate
around an axial central electrode with harmonic oscillations along it.
The frequency ω of these harmonic oscillations along the z-axis depends
only on the ion mass-to-charge ratio m/q and the instrumental
constant k:
q
w = m- × k
---
Two split halves of the outer electrode of the Orbitrap detect the image
current produced by the oscillating ions. By Fast Fourier Transformation
(FFT) of the image current, the instrument obtains the frequencies of
these axial oscillations and therefore the mass-to-charge ratios of the
ions.
Curved Linear Trap On their way from the linear trap to the Orbitrap, ions move through
the gas-free RF octapole (Oct 1) into the gas-filled curved linear trap
(C-Trap). See Figure 1-11 on page 1-14. Ions in the C-Trap are returned
by the trap electrode. Upon their passage, the ions loose enough kinetic
energy to prevent them from leaving the C-Trap through the Gate. The
nitrogen collision gas (bath gas) is used for dissipating the kinetic energy
of ions injected from the LTQ XL and for cooling them down to the axis
of the curved linear trap.
Voltages on the end apertures of the curved trap (Trap and Gate
apertures) are elevated to provide a potential well along its axis. These
voltages may be later ramped up to squeeze ions into a shorter thread
along this axis. The RF to the C-Trap (“Main RF”) is provided by the
CLT RF main board. (See page 1-48.) Trap and gate DC voltages as well
as RF voltages to octapole 1 are all provided by the ion optic supply
board. (See page 1-45.) High voltages to the lenses are provided by the
high voltage power supply board. (See page 1-51.)
1
Figure 1-11. Layout of the instrument, also showing the applied voltages
Extraction of Ion Packets For ion extraction, the RF on the rods of the C-Trap is switched off and
extracting voltage pulses are applied to the electrodes, pushing ions
orthogonally to the curved axis through a slot in the inner hyperbolic
electrode. Because of the initial curvature of the curved trap and the
subsequent lenses, the ion beam converges on the entrance into the
Orbitrap. The lenses form also differential pumping slots and cause
spatial focusing of the ion beam into the entrance of the Orbitrap. Ions
are electrostatically deflected away from the gas jet, thereby eliminating
gas carryover into the Orbitrap.
1HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the
instrument is not equipped with this option.
Owing to the fast pulsing of ions from the curved trap, ions of each
mass-to-charge ratio arrive at the entrance of the Orbitrap as short
packets only a few millimeters long. For each mass/charge population,
this corresponds to a spread of flight times of only a few hundred
nanoseconds for mass-to-charge ratios of a few hundred Daltons/charge.
Such durations are considerably shorter than a half-period of axial ion
oscillation in the trap. When ions are injected into the Orbitrap at a
position offset from its equator (Figure 1-12), these packets start
coherent axial oscillations without the need for any additional excitation
cycle.
The evolution of an ion packet during the increase of the electric field is
shown schematically on Figure 1-12. When the injected ions approach
the opposite electrode for the first time, the increased electric field
(owing to the change of the voltage on the central electrode) contracts
the radius of the ion cloud by a few percent. The applied voltages are
adjusted to prevent collision of the ions with the electrode. A further
increase of the field continues to squeeze the trajectory closer to the axis,
meanwhile allowing for newly arriving ions (normally, with higher m/q)
to enter the trap as well. After ions of all m/q have entered the Orbitrap
and moved far enough from the outer electrodes, the voltage on the
central electrode is kept constant and image current detection takes
place.
Ion Detection During ion detection, both the central electrode and deflector are
maintained at very stable voltages so that no mass drift can take place.
The outer electrode is split in half at z=0, allowing the ion image current
in the axial direction to be collected. The image current on each of half
of the outer electrode is differentially amplified and then undergoes
analog-to-digital conversion before processing using the fast Fourier
transform algorithm.
Peltier Cooling To allow stable operating conditions in the UHV chamber, it can be
cooled or heated (outgassing) by means of a Peltier element located on
the outside. A second Peltier element is located on the back of the
CE power supply board. See Figure 1-32 on page 1-44.
The Peltier cooling is based on the Peltier Effect, which describes the
effect by which the passage of an electric current through a junction of
two dissimilar materials (thermoelectric materials) causes temperature
differential (cooling effect). The voltage drives the heat to flow from one
side of the Peltier element to the other side, resulting in cooling effects
on one side and heating effects on the other side.
To remove the heat from the hot side of the Peltier elements, they are
connected to the cooling water circuit of the LTQ Orbitrap Discovery.
See topic “Cooling Water Circuit” on page 1-27 for further
information.
HCD Collision Cell HCD is an option for LTQ Orbitrap Discovery. The feature will be not
available if the instrument is not equipped with this option.
C-Trap
The fragment spectra generated in the collision cell and detected in the
Orbitrap show a fragmentation pattern comparable to the pattern of
typical collisional quadrupole spectra. See the LTQ Orbitrap Discovery
Getting Started manual for more information.
Vacuum System Figure 1-15 shows a schematic overview of the vacuum system. The
Orbitrap has the following vacuum compartments:
The forepumps of the linear trap provide the forevacuum for the
turbopumps.
LTQ XL
UHV Chamber
Vacuum Chamber
Turbopump TMH 071
(TMP 2) Orbitrap
Chamber
Pirani Gauge
Cold Ion Gauge
LTQ XL
Figure 1-15. Schematical view of vacuum system (C-Trap compartment not shown)*
* For an abridged version of the parts list, see Table 4-7 on page 4-3.
Turbopumps All parts of the system except for the Orbitrap analyzer are mounted in a
aluminum vacuum chamber evacuated by a 60 L/s turbopump (TMP 1,
see Figure 1-16). The rotary vane pumps of the linear trap (see below)
provide the forevacuum for this pump. This chamber is bolted to a
stainless steel welded UHV chamber housing the Orbitrap, lenses, and
corresponding electrical connections.
Preamplifier
HCD housing1
Cooling water
supply for Peltier
element
Cold Ion
Gauge
Turbopump
TMU 262 Turbopump TMH 071
(TMP 3) (TMP 2)
1HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the
instrument is not equipped with this option.
Linear Trap Turbopump A separate turbopump provides the high vacuum for the linear ion trap.
It it is mounted to the bottom of the vacuum manifold of the linear ion
trap. For more information, refer to the LTQ XL Hardware Manual.
Forevacuum Pumps The rotary vane pumps from the linear trap serve as forepumps for the
two smaller turbopumps (TMP 1 and TMP 2). The exhaust hose from
the forepumps is led to the back of the instrument and connects them to
the exhaust system in the laboratory. The forepumps are located on a
small cart in the forepumps cabinet below the linear trap. See
Figure 1-18.
Forepumps
To minimize the ingress of pump oil into the exhaust system, the outlets
of the forepumps are fitted to oil mist filters. See page 3-6 on
instructions about returning the collected oil to the forepumps.
Vacuum System Controls The power distribution board controls all turbopumps via voltage levels.
Refer to topic “Power Distribution Board” on page 1-38. An interface
for RS485 data via the instrument control board connects the
turbopumps with the linear ion trap. (Refer to topic “Instrument
Control Board” on page 1-36.) The linear ion trap has a separate
turbopump controller.
The analog values are digitized by the instrument control board. (Refer
to the topic “Instrument Control Board” on page 1-36.) They are then
sent as readout values to the data system.
Switching on the Vacuum System When the vacuum system is switched on, the following occurs:
1. After the Vacuum Pumps switch is switched On, the pumps of the
linear ion trap and the LTQ Orbitrap Discovery are run up. The
Pirani gauge (see above) controls the LTQ Orbitrap Discovery low
vacuum pressure as well as the pressure at the forevacuum pumps.
Within a short time, a significant pressure decrease must be
observed. The goodness of the vacuum can be estimated by means
of the rotation speed of the turbopumps (e.g. 80% after 15 min.).
2. If the working pressure is not reached after the preset time, the
complete system is switched off. At the status LED panel of the
power distribution board, an error message (Vacuum Failure) is put
out (see below).
3. The Cold Ion Gauge is only switched on after the low vacuum is
reached. It is then used to monitor the vacuum in the Orbitrap
region.
Vacuum Failure In case the pressure in the LTQ Orbitrap Discovery or the linear ion
trap exceeds a safety threshold, the complete system including linear ion
trap, electronics, and pumps is switched off. However, the power
distribution is kept under current and puts out an error message at the
LED panel. (Refer to the topic “Power Distribution Board” on
page 1-38.) It can be reset by switching the main power switch off and
on. (Refer to the topic “Main Power Switch” on page 1-10.)
Vacuum System Heating After the system has been open to the atmosphere (e.g. for maintenance
during a System Bakeout work), the vacuum deteriorates due to contaminations of the inner parts
of the vacuum system caused by moisture or a power outage. These
contaminations must be removed by heating the vacuum system: a
system bakeout. Refer to the topic “Baking Out the System” on
page 3-4.
Gas Supply Figure 1-19 shows a schematical view of the gas supply in the
instrument. The LTQ Orbitrap Discovery uses at least two gases for
operation:
• Nitrogen,
• Helium, and
The linear trap requires high-purity (99%) nitrogen for the API sheath
gas and auxiliary/sweep gas. The Orbitrap uses nitrogen as collision gas
(bath gas) for the curved linear trap. The linear trap requires ultra-high
1HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the
instrument is not equipped with this option.
The laboratory gas supply is connected to the inlets at the right side of
the instrument. See Figure 1-9 on page 1-11. Within the instrument,
the helium gas is led from the helium port through a stainless steel
capillary to the right rear side of the linear trap. Nitrogen gas and
HCD collision gas are both led via Teflon tubing to the right side of the
LTQ Orbitrap Discovery.
Part of the nitrogen gas flow is directed through Teflon tubing via a
pressure regulator to the vent valve of the linear trap. (See below for
further information.) Another part of the nitrogen flow is directed
through Teflon tubing to the vacuum chamber of the Orbitrap.
Nitrogen gas pressure to the C-Trap is kept constant by using an
“open-split” interface (gas flow divider, see Figure 1-21). It contains a
capillary line from the nitrogen line of the instrument to atmosphere
(flow rate: ~20 mL/min), with another capillary leading from the point
of atmospheric pressure into the C-Trap (flow rate: ~0.2 mL/min). For
the nitrogen gas to the C-Trap, black PEEKSil™ tubing is used
(75 μm ID silica capillary in 1/16 in PEEK tubing).
Orbitrap
CLT RF Main board housing
The HCD collision gas is also led to a gas flow divider. Part of the gas is
led through a capillary line to the atmosphere (flow rate: ~20 mL/min).
The other part of the gas (flow rate: ~0.5 mL/min) enters the IN port of
a three way valve. The gas leaves the valve through the OUT port and is
led to the collision octapole next to the curved linear trap.1 For the
HCD collision gas, red PEEKSil™ tubing is used (100 μm ID silica
capillary in 1/16 in PEEK tubing).
The valve is located at the left instrument side next to the temperature
controller board. See Figure 1-20 on page 1-25. It is switched by the
software via the power distribution board. (See page 1-38.)
Vent Valve of the Linear If the system and pumps are switched off, the system is vented. The vent
Ion Trap valve is controlled by the linear ion trap. The LTQ XL Hardware Manual
contains further information about the vent valve.
The instrument is vented with nitrogen from the same tubing that
supplies the LTQ XL sheath gas. See Figure 1-19 above. The vent valve
of the LTQ XL is attached to a pressure regulator that is set to a venting
pressure of 3–4 psi. The pressure regulator is located at the left side of
the LTQ Orbitrap Discovery. See Figure 1-21.
1
The third port of the valve is closed.
Cooling Water Circuit Figure 1-22 on page 1-27 shows a schematical view of the cooling water
circuit in the LTQ Orbitrap Discovery. Cooling water at a temperature
of 20 °C enters and leaves the instrument at the bottom of the right side.
See Figure 1-9 on page 1-11. First, the fresh water passes through the
turbopumps in the order TMP 3 → TMP 1 → TMP 2. Then it passes
through the heating element (Peltier element) that keeps constant
(±0.5 °C) the preset temperature of the analyzer. Before it leaves the
instrument, the water passes through the other Peltier element at the
back of the central electrode power supply board.
For instruction about performing maintenance for the chiller, see topic
“Maintenance of the Recirculating Chiller” on page 3-7. See also the
manufacturer’s manual for the chiller.
Properties of Cooling The water temperature is not critical, but should be in the range of 20 to
Water 25 °C (68 to 77 °F). Lower temperatures could lead to a condensation of
atmospheric water vapor. It is recommended to use distilled water rather
than de-ionized water due to lower concentration of bacteria and
residual organic matter.
Printed Circuit Boards The LTQ Orbitrap Discovery is controlled by a PC running the
Xcalibur software suite. The software controls all aspects of the
instrument. The main software elements are the communication with
the linear ion trap, the control of ion detection, and the control of the
Orbitrap mass analyzer.
Linear Ion Trap The linear ion trap is connected to the LTQ Orbitrap Discovery main
Electronics power switch. The linear ion trap has a sheet metal back cover.
Figure 1-23 shows the electronic connections at the rear side of the
linear trap.
The linear ion trap electronics has two connections with the
LTQ Orbitrap Discovery electronics:
For further information about the linear ion trap electronics, refer to the
LTQ XL Hardware Manual.
Electronic Boards at the Figure 1-24 shows the parts of the instrument when the right side panel
Right Side of the is opened. A transparent cover protects the lower part.
Instrument
Preamplifier
Computer
housing
Instrument Control
board housing
Preamplifier Figure 1-25 shows the preamplifier (P/N 207 8900)1. The preamplifier
is located in a housing next to the Cold Ion Gauge.
1
Part number of complete unit.
Internal Computer Figure 1-26 shows the components of the data acquisition unit
(P/N 206 4132). The unit is mounted in a housing located at the right
side of the instrument.
Location of
Data Acquisition Analog board
Data Acquisition
Digital PCI board
Computer housing
Data Acquisition Figure 1-27 shows the data acquisition digital PCI board
Digital PCI Board (P/N 206 0501). It is an add-on board to the internal computer (See
Figure 1-26 on page 1-33.).
This board is used to convert detected ion signals to digital form and to
interface to the computer mainboard. The board has two 16 bit parallel
connections to the DAC and the ADC on the data acquisition analog
board, which is used for controlling and reading-back signals. A
high-speed link port channel is also on the board that is used to
communicate with the electronics in the linear ion trap.
Communication takes place not only between the ion trap and the
internal computer of the LTQ Orbitrap Discovery system, but also
between the ion trap and the data system computer. For further
information about the data system, refer to the LTQ XL Hardware
Manual.
The diagnostic LEDs listed in Table 1-4 on page 1-35 show the status of
the board. The position of the LEDs on the board is indicated by a
white rectangle in Figure 1-27.
Table 1-4. Diagnostic LEDs of the Data Acquisition Digital PCI board
Name Color Description Normal Operating Condition
+5 V Green +5 V voltage present On
+3.3 V Green +3.3 V voltage present On
+2.5 V Green +2.5 V voltage present On
Data Acquisition Analog Board Figure 1-28 shows the data acquisition analog board (P/N 206 4150).
The diagnostic LEDs listed in Table 1-5 on page 1-36 show the status of
the voltages applied to the board. The position of the LEDs on the
board is indicated by a white rectangle in Figure 1-28.
Power Supply 2 Board The power supply 2 board (P/N 206 1440) provides the supply voltages
for the data acquisition analog board. It is mounted to the back inside
the housing of the internal computer. See Figure 1-26 on page 1-33.
The diagnostic LEDs listed in Table 1-6 show the status of the voltages
applied to the board.
Instrument Control Board Figure 1-29 shows the instrument control board (P/N 205 4221). The
instrument control board is located in a housing next to the internal
computer. It is connected to the LTQ Orbitrap Discovery main power.
The diagnostic LEDs listed in Table 1-7 show the status of applied
voltages to the board. The position of the diagnostic LEDs on the board
is indicated by a white rectangle in Figure 1-29.
Additionally, the board has four green LEDs that are directly connected
to the micro controller. They indicate the state of the micro controller
and possible error bits and can be used for software debugging. See
Table 1-8 on page 1-38. The position of the status LEDs on the board is
indicated by a white oval in Figure 1-29 on page 1-37.
Power Distribution Board Figure 1-30 on page 1-38 shows the power distribution board
(P/N 210 1320)1. It is located at the bottom of the right side of the
instrument.
1
Part number of complete unit.
The power distribution board controls the vacuum system and the
system power supplies, including the linear ion trap. Depending on the
quality of the vacuum and the status of the turbo molecular pumps, it
switches the vacuum gauges, the pumps, and the 230 V relays. It
controls external relays with 24 V dc connections. In case of a vacuum
failure, it initiates an automatic power down of the instrument. The
board also switches the valve that controls the flow of the HCD collision
gas.
The power distribution board indicates all system states and error
messages by status LEDs (see Table 1-9 on page 1-40) in the middle of
the left side of the board. A green LED indicates that the status is OK.
An orange LED indicates a status that differs from normal. The position
of the LEDs on the board is indicated by a white oval in Figure 1-30 on
page 1-38.
The system status LEDs on the front side of the instrument (See
Figure 1-4 on page 1-6.) are controlled by the power distribution board.
The information partially comes from external boards (for example, the
Communication LED is controlled by the instrument control board).
(Refer to topic “Instrument Control Board” on page 1-36.)
Diagnostic LEDs show the status of voltages applied from the board to
other devices. The positions of the diagnostic LEDs on the board are
indicated by white rectangles in Figure 1-30 on page 1-38.
Table 1-11 shows the possible operating states of the power distribution.
4. After the system has started, the Pirani Switch on Penning gauge
gauge returns a vacuum < 10-2 mbar
and both turbopumps reach 80%
rotation speed
5. Vacuum and 80% rotation speed of Switch off system (including linear ion
turbopumps not reached after preset trap) and light error LED*; power
time (< 8 min, otherwise the pumps distribution remains switched on
automatically switch off)
6. One or more vacuum gauges defective Light error LED only, otherwise ignore
(control signal < 0.5 V)
Power Supply 1 Board Figure 1-31 on page 1-43 shows the power supply 1 board
(P/N 205 5810). This board is located next to the power distribution
board. It provides the power for the ion optic supply board (Refer to
topic “Ion Optic Supply Board” on page 1-45.) and the instrument
control board. (Refer to topic “Instrument Control Board” on
page 1-36.)
The diagnostic LEDs listed in Table 1-12 show the status of the voltages
applied to the board. The position of the LEDs on the board is
indicated by the white rectangles in Figure 1-31.
Electronic Boards on the Figure 1-32 shows the left side of the instrument with the panel opened.
Left Side of the Instrument This side of the instrument contains mostly boards that are part of the
Orbitrap control.
Central Electrode
Pulser board
Temperature
Controller board
Central Electrode
Power Supply board
High Voltage Power
Supply board
The main components on this side are described starting from the top.
Ion Optic Supply Board Figure 1-33 shows the ion optic supply board (P/N 209 9810)1. The
board is located in a housing on top of the left instrument side of the
instrument. This board supplies the voltages and the radio frequency for
the ion guides and interoctapole lenses of the LTQ Orbitrap Discovery.
It has an RF detector for the RF output control. The board also provides
the trap voltage, gate voltage, and reflector dc voltages as well as the
RF voltages to the octapole 1 of the Orbitrap. See topic “Orbitrap
Analyzer” on page 1-13 for further information. Furthermore, the board
provides the voltages for the HCD collision cell.2 See page 1-17.
The diagnostic LEDs listed in Table 1-13 on page 1-46 show the status
of applied voltages to the board. The position of the LEDs on the board
is indicated by white rectangles in Figure 1-33.
1
Part number of complete unit.
2HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the
instrument is not equipped with this option.
Central Electrode Pulser Board The central electrode pulser board (P/N 207 9640)1 is located in a
housing that is mounted to the flange of the UHV chamber.
1
Part number of complete unit.
The board switches the injection and measurement voltages for the
central electrode and the detection electrodes of the Orbitrap.
Resistor-capacitor circuits on the board convert the switching pulse into
a smooth transition between the voltages.
The diagnostic LEDs listed in Table 1-14 show the status of the voltages
applied to the board as well as some operating states. The position of the
LEDs on the board is indicated by the white rectangles in Figure 1-34
on page 1-46.
Temperature Controller Board The temperature controller board (P/N 207 8930) is located on the top
left side of the instrument, next to the CLT RF main board. See
Figure 1-32 on page 1-44. The temperature controller board keeps the
temperature of the analyzer chamber to a preset value. A Peltier element
that can be used for heating as well as for cooling is used as an actuator.
Activation is done via the serial SPI (Serial Peripheral Interface) bus.
The diagnostic LEDs listed in Table 1-15 show the status of the voltages
applied to the board as well as some operating states. The positions of
the LEDs on the board are indicated by the white rectangles in
Figure 1-35.
CLT RF Unit The CLT RF unit (P/N 207 9581) comprises the CLT RF main board
and the RF off & feedback board. The unit operates the curved linear
trap (C-Trap) with four phases RF voltage and three pulsed DC voltages
(PUSH, PULL, and OFFSET).
The CLT RF main board (P/N 207 9591) is located in a housing in the
center of the left side of the instrument. See Figure 1-32 on page 1-44.
This board provides an RF voltage (“Main RF”) for the curved linear
trap (C-Trap). It allows switching off the RF and simultaneous pulsing
of each C-Trap electrode. See topic “Orbitrap Analyzer” on page 1-13
for further information. The board communicates with the instrument
control board via an SPI bus.
The RF off & feedback board (P/N 208 2540) is an add-on board to the
CLT RF main board. It is located in the same housing. See Figure 1-36
on page 1-49.
The diagnostic LEDs listed in Table 1-16 show the status of the voltages
applied to the board as well as some operating states. The position of the
LEDs on the board is indicated by the white rectangles in Figure 1-36.
Central Electrode The central electrode power supply board (P/N 207 9611)1 is mounted
Power Supply Board in a housing on the bottom left side of the instrument. See Figure 1-37.
For positive ions, the CE voltages are negative and the DE voltages are
positive. The maximum CE voltage is 3 kV and the maximum
DE voltage is 1 kV. The board communicates via the SPI bus.
1
Part number of complete unit.
The diagnostic LEDs listed in Table 1-17 on page 1-51 show the status
of the voltages applied to the board as well as some operating states. The
position of the LEDs on the board is indicated by the red rectangles in
Figure 1-37.
Table 1-17. Diagnostic LEDs of the Central Electrode Power Supply board
No. Name Color Description Normal Operating
Condition
LD1 OVL DE HI- Yellow Negative side of Deflector High Off when HV is
Supply has been overloaded switched on
LD2 OVL DE HI+ Yellow Positive side of Deflector High Off when HV is
Supply has been overloaded switched on
LD3 No Term Red SPI bus termination board Off
missing
LD4 Send Yellow Interface has been addressed Flashing on SPI-bus
and sends/receives data data transfer
LD5 Sel Green Board has been addressed Flashing on SPI-bus
data transfer
LD6 Polarity Blue Positive/negative ion mode Off (positive mode)
LD7 OVL CE LO+ Yellow Positive side of Central Off when HV is
Electrode Low Supply has been switched on
overloaded
LD8 OVL CE LO- Yellow Negative side of Central Off when HV is
Electrode Low Supply has been switched on
overloaded
LD9 OVL CE HI+ Yellow Positive side of Central Off when HV is
Electrode High Supply has been switched on
overloaded.
LD10 OVL CE HI- Yellow Negative side of Central Off when HV is
Electrode High Supply has been switched on
overloaded
LD11 OVL DE LO+ Yellow Positive side of Deflector Low Off when HV is
Supply has been overloaded. switched on
LD12 OVL DE LO- Yellow Negative side of Deflector Low Off when HV is
Supply has been overloaded. switched on
LD13 HV ON Green High voltage switched on On when HV is
switched on
High Voltage The high voltage power supply board (P/N 207 7991)1 is mounted in a
Power Supply Board housing on the bottom left side of the instrument. See Figure 1-32 on
page 1-44. This board provides five dc voltages for the ion optics of the
LTQ Orbitrap Discovery. Two voltages supply the lenses of the
instrument. Three voltages are applied to the RF CLT main board to be
1
Part number of complete unit.
used as focusing potentials for the curved linear trap. See topic
“Orbitrap Analyzer” on page 1-13 for further information. The board
communicates via the SPI bus.
The diagnostic LEDs listed in Table 1-18 on page 1-53 show the
operating states of the board. The position of the LEDs on the board is
indicated by the white rectangles in Figure 1-38.
Table 1-18. Diagnostic LEDs of the High Voltage Power Supply board
No. Name Color Description Normal Operating
Condition
LD1 NO TERM Red SPI bus termination board Off
missing
LD2 SEND Yellow Interface has been addressed Flashing on SPI bus
and sends/receives data data transfer
LD3 SEL Green Board has been addressed Flashing on SPI bus
data transfer
LD4 HV ON Green High voltage is switched on On
LD5 POLARITY Green Positive/negative ion mode Off (positive mode)
SPI Bus Termination Board Various boards communicate via the SPI bus, a serial RS485-based bus
system. The SPI bus termination board reduces unwanted signal
reflections. The boards indicate a missing termination (after
maintenance, for example) by LEDs.
The SPI bus termination board (P/N 208 1480) is located at the bottom
left side of the instrument, below the high voltage power supply board.
See Figure 1-39.
Figure 1-39. High Voltage Power Supply board with SPI Bus Termination
board
Shutting Down the If you need to turn off the MS detector in an emergency, place the main
System in an power switch (located on the power panel at the right side of the
LTQ Orbitrap Discovery) in the Off (O) position. This turns off all
Emergency power to the instrument, including the linear ion trap, multiple socket
outlets for the data system, and the vacuum pumps. The main power
switch must be turned 90° anti-clockwise to switch off the instrument.
See Figure 2-1.
On
Off
Behavior of the System in A main power failure has the same consequence as switching off via the
Case of a Main Failure main power switch. If the power is available again, the system is started
up automatically i.e. the pumps are switched on and the vacuum is
created. If the system has been vented during the mains failure, it is
necessary to bakeout the system to obtain the operating vacuum. Refer
to the topic “Baking Out the System” on page 3-4.
It is not possible to check whether the system was vented. The log file of
the data system indicates a reboot of the system. In case of frequent but
short power failures we recommend installing an uninterruptible power
supply (UPS). If main power failures occur frequently while the system
is not attended (e.g. in the night), we recommend installing a power fail
detector.
Note The intentional venting of the system is performed with the vent
valve of the linear ion trap. ▲
Placing the System in The LTQ Orbitrap Discovery system should not be shut down
Standby Condition completely if you are not going to use it for a short period of time, such
as overnight or over the weekend. When you are not going to operate
the system for 12 hours or more, you can leave the system in Standby
condition.
2. Turn off the flow of sample solution from the LC (or other sample
introduction device).
Note For instructions on how to operate the LC from the front panel,
refer to the manual that came with the LC. ▲
3. From the Tune Plus window, choose Control > Standby (or click on
the On/Standby button to toggle it to Standby) to put the
instrument in Standby condition. The consequences of this user
action are described in the LTQ XL Hardware Manual. The System
LED on the front panel of the LTQ XL is illuminated yellow when
the system is in Standby condition.
Shutting Down the The LTQ Orbitrap Discovery system does not need to be shut down
System completely if you are not going to use it for a short period of time, such
as overnight or over a weekend. Refer to the topic “Placing the System in
Standby Condition” on page 2-4. This section describes how to shut
down the system for a maintenance or service procedure.
Use the following procedure to shut down the LTQ Orbitrap Discovery
system:
2. Turn off the flow of sample solution from the LC (or other sample
introduction device).
Note For instructions on how to operate the LC from the front panel,
refer to the manual that came with the LC. ▲
3. From the Tune Plus window, choose Control > Off to put the
instrument in Off condition. When you choose Control > Off, all
high voltages are shut off, as are the sheath and auxiliary gas.
4. Put the FT Electronics switch to the OFF position. See Figure 1-7
on page 1-9.
5. Put the Vacuum Pumps switch to the OFF position. See Figure 1-7.
When you place the switch in the OFF position, the following
occurs:
6. Leave the main power switch of the LTQ Orbitrap Discovery in the
On position.
Starting Up the To start up the LTQ Orbitrap Discovery after it has been shut down,
System after a you need to do the following:
Shutdown 1. Start up the instrument
Starting Up the Instrument Note The recirculating chiller and data system must be running before
you start up the instrument. The instrument will not operate until
software is received from the data system. ▲
4. Make sure that the main power switch of the LTQ XL is in the On
position and the electronics service switch of the LTQ XL is in the
Operating position.
5. Place the main power switch at the right side of the LTQ Orbitrap
Discovery in the On position.
6. Put the Vacuum Pumps switch to the ON position. See Figure 1-7
on page 1-9. The rotary-vane pumps and the turbomolecular pumps
are started.
Note Pumping the system after a complete shut down takes hours and
requires overnight baking of the system. ▲
Setting Up Conditions for Set up your LTQ Orbitrap Discovery for operation, as follows:
Operation
1. Before you begin data acquisition with your LTQ Orbitrap
Discovery system, you need to allow the system to pump down for
at least 8 hours. Operation of the system with excessive air and water
in the vacuum manifold can cause reduced sensitivity, tuning
problems, and a reduced lifetime of the electron multiplier.
2. Ensure that the gas pressures are within the operational limits:
• Argon if present: 690 ± 140 kPa (6.9 ± 1.4 bar, 100 ± 20 psi)
Note Air in the helium line must be purged or given sufficient time to
be purged for normal performance. ▲
3. Select the Display Status View button in the Tune Plus window.
Check to see whether the pressure measured by the ion gauge is
≤5 × 10-9 mbar, and the pressure measured by the Pirani gauge is
around 1 mbar. Compare the values of the other parameters in the
status panel with values that you recorded previously.
Resetting the System If communication between the LTQ Orbitrap Discovery and data
system computer is lost, it may be necessary to reset the system using the
Reset button of the LTQ XL.
The procedure given here assumes that the LTQ Orbitrap Discovery
and data system computer are both powered on and are operational. If
the instrument, data system computer, or both are off, refer to topic
“Starting Up the System after a Shutdown” on page 2-7.
To reset the LTQ Orbitrap Discovery, press the Reset button of the
linear ion trap. See the LTQ XL Hardware Manual for the location of the
Reset button. When you press the Reset button, the following occurs:
Resetting the Tune You can reset the LTQ Orbitrap Discovery system tune and calibration
and Calibration parameters to their default values at any time. This feature may be useful
if you have manually set some parameters that have resulted in less than
Parameters to their optimum performance. To reset the LTQ Orbitrap Discovery tune and
Default Values calibration parameters to their default values, proceed as follows:
• Choose File > Restore Factory Tune Method to restore the default
tune parameters.
Note Make sure that any problems you might be experiencing are not
due to improper API source settings (spray voltage, sheath and auxiliary
gas flow, ion transfer capillary temperature, etc.) before resetting the
system parameters to their default values. ▲
General Remarks Preventive maintenance must commence with installation, and must
continue during the warranty period to maintain the warranty. Thermo
Fisher Scientific offers maintenance and service contracts. Contact your
local Thermo Fisher Scientific office for more information. Routine and
infrequent maintenance procedures are listed in Table 3-1.
• Proceed methodically
• Always cover the opening in the top of the vacuum manifold with a
large, lint-free tissue whenever you remove the top cover plate of the
vacuum manifold.
• Dirty tools can contaminate your system. Keep the tools clean and
use them exclusively for maintenance and service work at the
LTQ Orbitrap Discovery.
Returning Parts In order to protect our employees, we ask you for some special
precautions when returning parts for exchange or repair to the factory.
Your signature on the Repair Covering letter confirms that the returned
Cleaning the Surface of Clean the outside of the instrument with a dry cloth. For removing
the Instrument stains or fingerprints on the surface of the instrument (panels, for
example), slightly dampen the cloth (preferably made of microfiber)
with distilled water.
Baking Out the This section provides information and help concerning the system
System bakeout of the LTQ Orbitrap Discovery. The bakeout procedure
removes unwanted gases or molecules (collected or remaining) from the
high vacuum region of the instrument. Ions can collide with those gases
or molecules resulting in lower overall sensitivity. Therefore, we
recommend to bake out the instrument if the high vacuum decreases
noticeable during routine operation.
Note Pumping down the system after venting takes at least 8 hours, and
usually requires overnight baking of the system. ▲
In case the system has been vented during a power failure, it is necessary
to bake out the system to obtain the operating vacuum. Refer to the
topic ”Behavior of the System in Case of a Main Failure” on page 2-2.
2. Put the FT Electronics switch at the power control panel into the
ON position.
3. Set the bakeout time by entering the desired time (hh:mm) with the
up/down keys of the bakeout timer. See Figure 3-1.
Up keys
Down keys
Maintaining the The turbopumps1 need maintenance work that is briefly outlined below.
Vacuum System
Note The manuals of the pump manufacturers give detailed advice
regarding safety, operation, maintenance, and installation. Please note
the warnings and precautions contained in these manuals! ▲
Exchanging the Lubricant For all manipulations at the pumps, note the advice, warnings, and
Reservoir of the cautions contained in the pump manuals!
Turbopumps For the turbopumps, we recommend exchanging the lubricant reservoir
in a one-year cycle. At each exchange procedure, the complete lubricant
reservoir must be exchanged!
1For
maintenance of the forepumps, refer to the LTQ XL Hardware Manual or the pump
manufacturer’s manual.
Maintenance of the For the NESLAB ThermoFlex 900 recirculating chiller, the following
Recirculating Chiller checks should be carried out on a regular basis.
Reservoir Periodically inspect the fluid inside the reservoir. If cleaning is necessary,
flush the reservoir with a cleaning fluid compatible with the circulating
system and the cooling fluid.
Fluid Bag Filter The ThermoFlex 900 recirculating chiller installed in the cooling circuit
of the instrument is equipped with a fluid bag filter, which needs to be
replaced on a regular basis. Replacement bags are available from Thermo
Fisher Scientific.
Condenser Filter To prevent a loss of cooling capacity and a premature failure of the
cooling system, clean the condenser filter regularly. If necessary, replace
it.
† Optional equipment.
† Some systems are delivered with NESLAB Merlin M33 recirculating chiller.
Parts Basic System This section contains parts lists for the mechanical components of the
LTQ Orbitrap Discovery.
Table 4-4. Parts Upgrade kit Orbitrap HCD option (P/N 122 4800)*
Pos. No. Qty. Designation Part No.
0010 1 Collision cell, with mount 122 1320
0020 1 Flange, for options; complete 122 1250
0030 1 Cable kit, Upgrade LTQ Orbitrap Discovery 122 5100
0040 1 Upgrade kit gas supply LTQ Orbitrap XL† 122 4820
0050 1 Cable, power distribution/valve 210 0310
* HCD
is an option for LTQ Orbitrap Discovery. The feature will be not available if the instrument is not
equipped with this option.
†
Module, for parts list see below.
Table 4-5. Parts Orbitrap Installation Kit (P/N 118 8120), continued
Pos. No. Qty. Designation Part No.
0100 1 Cap nut, stainless steel; 1/8“ 052 0890
0110 1 Connector; C19 to C14 082-2528
0120 10 Miniature fuse; 5 x 20 mm, 8 A 201 1850
*
Specify desired length.
Parts Orbitrap Analyzer Table 4-6. Parts Orbitrap-2 chamber; complete (P/N 122 4780)
Pos. No. Qty. Designation Part No.
0010 1 UHV chamber; complete 116 6110
0020 1 Vacuum chamber, complete 116 6131
0030 1 Octapole Orbitrap, complete 116 6120
0040 4 Washer; 8.4×15×1.5, stainless steel 047 0860
0050 4 Screw; M8X25, silver plated 119 1210
0060 4 Pan head screw; M 2 x 8 DIN 84 A4 045 0720
0070 2 Screw cyl. M 4 x 12 DIN912 045 2800
Parts Pumping System Table 4-7. Parts pumping system Orbitrap (P/N 118 4490)*
Orbitrap Pos. No. Qty. Designation Part No.
0010 1 Turbomolecular pump TMU262; modified 118 4340
0020 2 Turbomolecular pump; TMH 071 P 114 1500
0040 1 Water cooling for TMH 262 114 9140
0050 5 Water cooling for TMH 071 P 079 4742
0080 1 UHV gauge IKR 270; short 118 1380
0130 1 Compact Pirani Gauge TPR280 115 6400
*
For a schematic overview of the vacuum system, refer to Figure 1-15 on page 1-19.
Two kits are available for the pumping system, comprising either
pumps, gauges, and cooling (pump kit Orbitrap: P/N 117 5000) or
hoses, clamps, and gaskets (pump system Orbitrap: P/N 117 5010):
Table 4-8. Parts pump kit Orbitrap (P/N 117 5000), continued
Pos. No. Qty. Designation Part No.
0050 1 Water cooling for TMH 262 114 9140
0060 5 Water cooling for TMH 071 P 079 4742
0070 1 PVC hose, with steel helix; ID=45 mm, L=1.3 m 118 4330
0080 2 Hose nipple, DN 40, ISO-KF-45 115 9230
0090 1 Venting flange; DN 10, KF-G1/8" 118 4400
0100 1 Splinter shield for turbopumps, with DN 100 CF-F 119 8590
flange
0110 2 Centering ring, with integrated splinter shield; 119 8600
DN 63 ISO
Parts Gas and Water This topic contains parts lists for the gas and water supply of the
Supply LTQ Orbitrap Discovery. For a schematical overview of the gas supply,
refer to Figure 1-19 on page 1-24.
Table 4-11. Upgrade kit gas supply LTQ Orbitrap XL (P/N 122 4820)
Pos. No. Qty. Designation Part No.
0050 0.10 m* Capillary 1/16" ID-SS 060 5470
0070 1 Stainless reducing union; Swagelok, 6 mm-1/16” 117 8070
0080 1 T-piece; 1/16" (SS-100-3) 052 3550
0090 1 Coupling; 1/16”, SS-100-6 052 4340
0100 2 PEEKSIL capillary; 1/16", 500 × 0.075 mm 118 6970
0130 6 Ferrule; 1/16" GVF/16 067 4800
0190 2 Capillary; PEEKsil, 1/16", 0.1 × 500 mm 122 3420
0200 1 Valve, with angle 122 4070
* Specify required length.
Electronic Parts This section contains parts list for the electronic components of the
LTQ Orbitrap Discovery.
Electronics - Right Panel The following replaceable parts are available for the electronics at the
right panel of the LTQ Orbitrap Discovery.
Table 4-15. Parts Orbitrap-2 electronics; right panel (P/N 208 1010)
Pos. No. Qty. Designation Part No.
0010 1 Power supply 1 205 5810
0020 1 Power supply 29V/2A 206 4040
0030 1 Power distribution 206 2130*
0040 1 Unit PS 24V/20A 208 1130
0050 1 Unit PS +/-15V/2.75A 208 1140
0060 1 KIT E_RIGHT PANEL ASSEMBLY 206 9790
0070 1 Unit data acquisition† 206 4132
0080 1 Instrument control board 205 4221
0085 1 CLT RF TRIGGER board 208 5880
0090 1 LT ANALOG INTERFACE board Orbitrap-2 208 1940
0100 1 LT DIGITAL INTERFACE board 208 7180
0110 1 UNIT PS-BASIC LOAD 208 5900
* This item is replaced by P/N 210 1320.
Electronics - Left Panel The following replaceable parts are available for the electronics at the left
panel of the LTQ Orbitrap Discovery.
Table 4-17. Parts Orbitrap-2 electronics; left panel (P/N 208 1020)
Pos. No. Qty. Designation Part No.
0010 1 UNIT HIGH VOLTAGE PS 207 7991
0020 1 UNIT CENTRAL ELECTRODE PS 207 9611
0030 1 LP SPI-BUS-TERMINATION 208 1480
Electronics Analyzer The following replaceable parts are available for the analyzer electronics
of the LTQ Orbitrap Discovery.
Electronics Main Supply The following replaceable parts are available for the electronics main
supply of the LTQ Orbitrap Discovery.
Table 4-19. Parts electronics main supply Orbitrap (P/N 208 1040)
Pos. No. Qty. Designation Part No.
0010 1 UNIT SWITCH PANEL 208 1120
0020 1 Mains cable; IEC-CONNECTOR-16A 209 6110
0030 1 Mains cable; MAIN INPUT/LINE FILTER 208 1420
0040 1 KIT BAKEOUT-SWITCHES 207 9040
0050 1 Cable loom, mains supply 208 1110
0070 1 Mains cable; LINE FILTER/MAIN SWITCH 208 1280
0080 1 Mains cable; terminal/LTQ-relay 208 1290
0090 1 Mains cable; POWER DIS./SWITCH PANEL 1 208 1300
0100 1 Mains cable; POWER DIS./SWITCH PANEL 2 208 1310
0110 1 Cable; power distribution/LTQ-relay 208 1320
0120 1 Cable loom, power distribution/bakeout timer 207 9010
0130 1 Unit bakeout timer Orbitrap 208 0960
0140 1 KIT MAINS SUP.ASSEMBLY MAT. 208 1100
0150 1 Cable; BAKEOUT-TIMER/POWER DIS. 208 1470
0160 1 Mains cable; POWER DIS./SOCKET R-PUMP 208 1350
0170 1 Mains cable; LTQ-relay/LTQ-filter 208 9230
cm centimeter ft foot
FTMS Fourier Transformation Mass Spectroscopy KEGG Kyoto Encyclopedia of Genes and
Genomes
FTP file transfer protocol
kg kilogram
FWHM Full Width at Half Maximum
l length
g gram
L liter
9
G Gauss; giga (10 )
LAN local area network
GC gas chromatograph; gas chromatography
lb pound
GC/MS gas chromatograph / mass spectrometer
LC liquid chromatograph; liquid chromatography
GUI graphical user interface
LC/MS liquid chromatograph / mass spectrometer
h hour
LED light-emitting diode
h height
μ micro (10-6)
HCD Higher Energy Collision Induced
Dissociation m meter
NCBI National Center for Biotechnology RS-232 industry standard for serial
Information (USA) communications
S T
safety interlock switch 2-8
temperature
SEM detector 1-12 analyzer chamber 1-47
control 1-17 V
differential 1-17
vacuum
temperature controller board 1-17, 1-25, 1-44, 1-47
analyzer system 2-8
diagnostic LEDs 1-48
chamber 1-19–1-20
location 1-47
compartments 1-19
temperature, of chiller operating fluid 3-7 deterioration 1-23
thermoelectric elements 1-17 failure 1-23, 1-39
TMP 1 1-19–1-20 gauges 1-22, 1-36
TMP 2 1-19–1-20 quality 1-39
TMP 3 1-19–1-21 safety threshold 1-22–1-23
tools 3-2 system 1-10, 1-19
top lid, of instrument 1-4 vacuum components
transfer chamber 1-12 left instrument side 1-20
trap voltage 1-45 UHV chamber 1-20
tune parameters 2-11 Vacuum LED
Tune Plus window 1-29, 2-5, 2-9, 2-11 linear trap 2-8
buttons 2-4, 2-9 LTQ Orbitrap Discovery 1-6
turbopumps 1-20, 1-36 Vacuum Pumps switch 1-6, 1-9, 1-41
controller 1-21–1-22 vacuum system
error 1-40 controls 1-22
exchanging the lubricant reservoir 3-6 heating 1-23
linear trap 1-21 maintenance 3-6
maintenance intervals 3-6 moisture 1-23
TMP 2 1-20 valve, for HCD collision gas 1-26, 1-39
TMP 1 1-20 vent valve 1-26, 2-2–2-3, 2-5
TMP 3 1-20 linear trap 1-25, 2-3
vent valves 1-23 venting pressure 1-26
turning off, the peripherals 2-2, 2-6 venting, the system 2-3
U W
UHV chamber 1-17, 1-19–1-20 water 1-10
temperature control 1-17 chiller 1-4
uninterruptible power supply 2-2 hoses 1-28
up keys 3-4 ports 1-11
upper control panel 1-7–1-8 temperature 1-28
UPS wheels, of instrument 1-4
See uninterruptible power supply working modes, of the power distribution 1-41
user maintenance 3-1–3-2 working principle, of the Orbitrap analyzer 1-3
procedures 3-2
X
Xcalibur 1-29
www.thermo.com