Service Manual OMNI C Rev 1
Service Manual OMNI C Rev 1
T E S T I N G
Roche OMNI C
Service Manual
Roche Diagnostics GmbH
D-68298 Mannheim / Germany
www.roche.com
5IJTJOTUSVNFOUJTJOUFOEFEUPCFVTFEPOMZGPSUIFTQFDJBMJ[FEQVSQPTFEFTDSJCFEJOUIFJOTUSVDUJPOT5IF
NPTUJNQPSUBOUQSFSFRVJTJUFTGPSVTF
PQFSBUJPO
BOETBGFUZBSFFYQMBJOFEUPFOTVSFTNPPUIPQFSBUJPO/P
XBSSBOUZPSMJBCJMJUZDMBJNTXJMMCFDPWFSFEJGUIFJOTUSVNFOUJTVTFEJOXBZTPUIFSUIBOUIPTFEFTDSJCFEPSJG
UIFOFDFTTBSZQSFSFRVJTJUFTBOETBGFUZNFBTVSFTBSFOPUPCTFSWFE
5IFJOTUSVNFOUNBZCFPQFSBUFEPOMZCZQFSTPOTXIPTFRVBMJGJDBUJPOTFOBCMFUIFNUPDPNQMZXJUIUIFTBGFUZ
NFBTVSFTUIBUBSFOFDFTTBSZEVSJOHPQFSBUJPOPGUIFNBDIJOF
"EKVTUNFOUTBOENBJOUFOBODFQFSGPSNFEXJUISFNPWFEDPWFSTBOEDPOOFDUFEQPXFSNBZCFBUUFNQUFEPOMZ
CZBRVBMJGJFEUFDIOJDJBOXIPJTBXBSFPGUIFBTTPDJBUFEEBOHFST
*OTUSVNFOUSFQBJSTBSFPOMZUPCFQFSGPSNFECZUIFNBOVGBDUVSFSPSRVBMJGJFETFSWJDFQFSTPOOFM
0OMZBDDFTTPSJFTBOETVQQMJFTFJUIFSEFMJWFSFECZPSBQQSPWFECZ3PDIFBSFUPCFVTFEXJUIUIFJOTUSVNFOU
5IFTFJUFNTBSFNBOVGBDUVSFEFTQFDJBMMZGPSVTFXJUIUIJTJOTUSVNFOUBOENFFUUIFIJHIFTURVBMJUZSFRVJSF
NFOUT
0QFSBUJPOPGUIFJOTUSVNFOUXJUITPMVUJPOTXIPTFDPNQPTJUJPOJTOPUDPOTJTUFOUXJUIUIBUPGUIFPSJHJOBM
TPMVUJPOTDBOOFHBUJWFMZBGGFDU
BCPWFBMM
UIFMPOHUFSNNFBTVSFNFOUBDDVSBDZ%FWJBUJPOTJOUIFDPNQPTJ
UJPOPGUIFTPMVUJPOTDBOBMTPEFDSFBTFUIFTFSWJDFMJGFPGUIFFMFDUSPEFT
5IF RVBMJUZ DPOUSPM SFRVJSFNFOUT NVTU CF DPNQMFUFE BU MFBTU PODF EBJMZ GPS TBGFUZ SFBTPOT
#FDBVTFBDDVSBUFNFBTVSFNFOUSFTVMUTEFQFOEOPUPOMZPOUIFQSPQFSGVODUJPOJOHPGUIFJOTUSV
NFOU
CVUBMTPPOBOVNCFSPGPUIFSGBDUPST TVDIBTQSFBOBMZUJDT
UIFSFTVMUTQSPEVDFECZUIF
JOTUSVNFOU TIPVME CF FYBNJOFE CZ B USBJOFE FYQFSU CFGPSF TVCTFRVFOU EFDJTJPOT BSF SFBDIFE
UIBUBSFCBTFEPOUIFNFBTVSFNFOUWBMVFT
&YQMBOBUJPO
.FBOJOH$BVUJPO SFGFSUPBDDPNQBOZJOHEPDVNFOUTl
£*NQPSUBOUJOGPSNBUJPO£"MXBZTGPMMPX£
£0QFSBUJOHTBGFUZJOGPSNBUJPO£
• 5IFJOTUSVNFOUIBTCFFODPOTUSVDUFEBOEUFTUFEBDDPSEJOHUPUIFQSPUFDUJWFNFBTVSFTTUJQV
MBUFECZ&/*&$GPSFMFDUSJDBMNFBTVSFNFOU
DPOUSPM
*7%
BOEMBCP
SBUPSZJOTUSVNFOUTBOEXBTEFMJWFSFEGSPNUIFGBDUPS ZJOGMBXMFTTDPOEJUJPOXJUISFHBSETUP
TBGFUZGFBUVSFT*OPSEFSUPQSFTFSWFUIJTDPOEJUJPOBOEFOTVSFTBGFPQFSBUJPO
UIFVTFSNVTU
SFTQFDUUIFOPUJDFTBOEXBSOJOHTUIBUBSFDPOUBJOFEJOUIFTF*OTUSVDUJPOTGPS6TF
• 5IJT JOTUSVNFOU JT DMBTTJGJFE VOEFS UIF QSPUFDUJPO DMBTT * BDDPSEJOH UP &/ *&$
• 5IFJOTUSVNFOUNFFUTUIFDPOEJUJPOTGPSPWFSWPMUBHFDBUFHPSZ**
• 5IFJOTUSVNFOUNFFUTUIFDPOEJUJPOTGPSDPOUBNJOBUJPOMFWFM
• %P OPU PQFSBUF UIF JOTUSVNFOU JO BO FYQMPTJWF FOWJSPONFOU PS JO UIF WJDJOJUZ PG FYQMPTJWF
BOFTUIFUJDNJYUVSFTDPOUBJOJOHPYZHFOPSOJUSPVTPYJEF
• *G BO PCKFDU PS MJRVJE FOUFST UIF JOUFSOBM BSFBT PG UIF JOTUSVNFOU
SFNPWF UIF JOTUSVNFOU
GSPNJUTQPXFSTVQQMZBOEBMMPXBOFYQFSUUPDIFDLJUUIPSPVHIMZCFGPSFVTJOHJUBHBJO
• 5IFJOTUSVNFOUJTTVJUBCMFGPSMPOHUFSNPQFSBUJPOJOEPPST
$"65*0/
• 5IF QPXFS DPSE NBZ CF QMVHHFE JOUP B HSPVOEFE TPDLFU POMZ 8IFO VTJOH BO FYUFOTJPO
DPSE
NBLFTVSFJUJTQSPQFSMZHSPVOEFE
• "OZSVQUVSFPGUIFHSPVOEMFBEJOTJEFPSPVUTJEFUIFJOTUSVNFOUPSBMPPTFHSPVOEDPOOFD
UJPO NBZ SFTVMU JO IB[BSEPVT PQFSBUJOH DPOEJUJPOT *OUFOUJPOBM EJTDPOOFDUJPO PG UIF
HSPVOEJOHJTOPUQFSNJUUFE
• 5IFJOTUSVNFOUJTOPUTVJUBCMFGPSPQFSBUJPOXJUIBEJSFDUDVSSFOUQPXFSTVQQMZ
6TFPOMZUIFPSJHJOBMNBJOTQMVHEFMJWFSFEXJUIUIF3PDIF0./*$
£0QFSBUJOHTBGFUZJOGPSNBUJPO£
1 Introduction
2 Revisions
3 Components
4 AutoQC
5 Fluidic procedures
6 Service area
7 Annual maintenance
8 Troubleshooting
9 Index
Service Manual, Roche OMNI C
1 Introduction
1 Introduction.........................................................1-1
1.1 General notes ......................................................................................... 1-1
1 Introduction
1.2 Symbols
Sections marked with this symbol contain information that must be observed for the
prevention of personal injury to patients, users or third parties.
All sections or text passages that are marked with this symbol describe procedures
and/or indicate conditions or dangers that could damage or lead to malfunctions in the
Roche OMNI C.
All sections or text passages marked with this symbol refer to situations and/or
potential dangers affecting personnel performing the servicing or repair work.
All sections or text passages marked with this symbol refer to components which
require special precautions related to electrostatic discharge. Packaging marked with
this symbol may only be opened by trained personnel (also see page 1-6, section 1, ESD
protection measures)
· Never use the analyzer near highly inflammable or explosive gasses (anesthetic
gasses, etc.).
· Use only 2-pin, grounded AC power sockets for the instrument.
· Ensure that the AC power cable and plug are in good condition.
Damaged AC cables or plugs must be replaced immediately.
· Switch the instrument off and unplug the AC power cable before opening its rear
panel.
· Be sure to replace blown fuses with new fuses of the same type.
· Always follow the instructions in this manual when servicing or repairing the
instrument.
· Always use suitable tools and testing equipment for servicing or repairs.
· To avoid damage to the instrument's electronic components, prevent liquids from
entering it.
· To clean the instrument, use lightly moistened tissue or cotton swabs.
1.4 Disinfectants
Do not attempt to decontaminate any part of the instrument before shutting it down and
unplugging it from the AC power socket.
Before plugging the instrument back in and turning it on, always wait 15 minutes to
allow the disinfectant to evaporate.
There is the risk of fire or explosion especially in the vicinity of the power supply!
Composition
Aqueous NaOCl solution with active chlorine (£ 2%)
Potential dangers
Due to the alkaline and oxidizing character of this preparation, we cannot rule out local irritation to
the skin, eyes, and mucous membranes.
Use a commercial alcohol-based (70%) disinfectant as a surface decontaminant. Please observe the
relevant product information!
1.5 Decontamination
Use the following decontamination procedures to minimize the risk of infection (including hepatitis
and HIV) when replacing parts that have come in contact with blood or other biological liquids.
1.5.1 Surfaces
Decontaminate all outside surfaces, including all covers (e.g. the measuring chamber cover, bottle
compartment cover, instrument cover) as well as the outside surfaces of the AutoQC module if
present, with a cloth moistened with disinfectant.
Press
"More functions – System – Wash and clean – Automatic routines – Decontaminate all
tubes" and follow the on-screen instructions.
For a detailed description of this procedure, please refer to section 6.1, Decontamination,
in the Roche OMNI C Instructions for Use.
ESD protection measures are designed to prevent electronic components from being
damaged or destroyed by uncontrolled electrostatic discharges.
Example:
· Shoes with rubber soles:
The friction of walking charges the person wearing the shoes in relation to the floor. A
discharge occurs when the person touches an object.
· Clothing made of synthetic fabrics:
Audible discharge; sparks visible in the dark.
· When transporting electronic assemblies, be sure to use ESD protection packaging 1 or other
suitable storage/shipping packaging (i.e. the original packaging material).
· Test and repair electronic assemblies only on tables with ESD mats1 .
· Wear a grounded ESD wrist strap.
· Handle electronic components/assemblies by the edges only. Avoid touching printed circuits or
component pins.
· Do not wear shoes with rubber soles or items of clothing containing synthetic fibres in facilities
in which electronic equipment is serviced.
· If necessary, use a humidifier to ensure optimal humidity (>20% rel.) in the room.
· Take care not to touch assemblies or individual components directly with your hand after
testing.
· When sending electronic components or assemblies in for repairs, always use ESD protection
packaging 1 to prevent further damage that may result in an incorrect interpretation of the cause
of the fault.
1.6.5 Conclusion
Naturally, not all printed circuit boards and electronic assemblies require such careful treatment. An
assembly that features simple plug connectors, for example, is insensitive to electrostatic discharge
and therefore does not need ESD protection packaging.
When in doubt, however, always take ESD precautions!
1
These are materials with a very low, defined conductivity (1012 ohms) which do not accumulate charges with friction and thus
cannot damage components.
2 Revisions ............................................................2-1
Revision 1.0 – First edition! ............................................................................................ 2-1
2 Revisions
3 Components ........................................................3-1
3.1 Important Instructions ............................................................................ 3-1
3.11 Micro switch unit (Bottle compartment cover switch) .......................... 3-37
3.11.1 General Information ........................................................................................ 3-37
3.11.2 Changing the micro switch unit ........................................................................ 3-38
3 Components
Components of the OMNI C (sample port, tubing, Waste container, etc.) contain
remnants of biological fluids after use, resulting in possible risk for infections.
After changing components, calibration parameters could have changed or some values
have to be entered again.
After finishing modifications, a system calibration must be performed to adapt the
calibration parameters to the new conditions.
3.2 Shutdown
Switch off the OMNI C before you begin with the modifications!
Determined by the time the analyzer was switched off, one of the following procedures
has to be performed:
· Perform the decontamination procedure according to sections 1.5.1 to 1.5.6 on page 1-4.
If the Docking mechanisms C1, C2 or C3 have been opened while the analyzer was
switched off, the respective solutions have to be prepared again under „More -
System – Tools– Fluid actions – Auto preparation routines“ to avoid improper
operation!
· Perform the decontamination procedure according to sections 1.5.1 to 1.5.5 on page 1-4.
· Press: „More - System – Tools- Shutdown“ and follow the instructions on the screen.
Only Roche Deproteinizer may be used for the decontamination of the tubing.
1. Switch off the OMNI C (please pay attention to section 3.2 Shutdown on page 3-2!).
2. Pull off the power cord from the power supply unit.
3. Remove the three mounting screws from the rear panel (see Fig. 3-1) and pull off the rear panel
from the analyzer (if needed, remove the power supply unit from the rear panel; a certain
amount of force is needed because of the mounting with Velcro tape).
4. Pull off the plug of the power supply cable from the PC tower (see Fig. 3-2).
When installing the rear panel, please pay attention to the correct routing of the power
supply cable inside the OMNI C (see Fig. 3-3)!
Principle:
The Mix is produced by the mixing valve V1 (two inputs, one output) in conjunction with the
peristaltic pump. For an optimal mixing process, the synchronous function of pump and the mixing
valve is essential.
The Mix consists of alternating fluid packages from solution C1 and solution C2. Homogenization is
accomplished by small packages of either solution in relation to the distance.
6. Open the four mounting screws of the FMS (see Fig. 3-5, 1 to 4).
(Do not remove the screws completely)
1 2
3 4
µC µC µC
Analog-Board AVR8535 AVR8515 AVR2313
SS1/SS2
V3 / V5
Temp. Fill Level
ADC Sinus
Regulation Baro
tHb-Electronic
Pre - Amplifier
V5
tHb / SO2
REF RCon Na Cl pH RCon Ca K PO2 PCO2 Preheater
SS1
Con
SS2
V3
FMS-Preheater
37 °C
Peltier 1 Peltier 2
Communication Power
Sample-
Sensor 1,2 MK cover sensor
Sinus generator Driver
Conductivity V3 u. V5 V3
Peltier 1
measurement
V5
NTC 1 AVR-8535
Fan 1
3xPWM Temp. Baro
bidir. over Regulation ADC
Peltier 2 Relais 11x12bit AVR 8515
3 Ampere 4x Temp
NTC 2
2 x S.S.
1x Filllevel opto coupler
Fan 2
thb/SO2
Cover DC/DC
I2C-Bus
measurement
3
ADC-Bus
K1...K4
5*
3+
MUX MUX
1:4 2 1:4 MUX
2
1:2
1. Switch off the OMNI C (please pay attention to section 3.2 Shutdown on page 3-2!).
2. Pull off the power cord from the power supply unit.
3. Remove the analyzer cover.
4. Dismount the sample port module (see section 3.6.1, Changing the sample port module on page
3-20).
5. Dismount the screen cover (see section 3.8.4, Changing the screen on page 3-30)
6. Open the measuring chamber cover.
7. Remove all electrodes from the measuring chamber.
8. Dismount the cover of the MC cover cable and pull the cable out (see Fig. 3-9)
9. Dismount the measuring chamber cover by pressing it to the left and then upwards
(see Fig. 3-10).
10. Pull the measuring chamber cover to the front, by doing this also the connection to the gas
spring to the housing is disconnected (see Fig. 3-11, 1).
2 4
1
13. Lift the measuring chamber carefully and pull off the two cables and the pressure sensor tube
from the analog board.
When installing the measuring chamber, the FMS pipe must slide exactly into the
holder of FMS connector sealing 1 (see Fig. 3-13, 1) and the two pipes at the
right side must slide through the two holes at the right side of the measuring
chamber (see Fig. 3-13, 2 and 3).
1
2
3
After installing the measuring chamber a Waste container fill check must be
performed:
1. Insert a full Waste container W into the analyzer and close the docking
mechanism W.
2. Press „More - System – Test – Waste Container sensor“ and measure
the Waste container level by pressing the button „W“ on the screen.
3. The measured fill level must be equivalent to the actual fill level in the Waste
container W.
4. If the fill levels are not equivalent, check whether the tube at the pressure
sensor (on the analog board) is connected properly.
1 2
4. Pull off all cable connections from the analog board (see Fig. 3-15)
5. Take out the analog board (by doing this also the electrode locking lever disengages).
6. Pull the tube out of the FMS connector sealing 2 (see Fig. 3-19, 1).
7. Open the tube connection and the barex tube connection (see Fig. 3-20, 1 and 2).
8. Push the tube connections to the side and take out the sample sensor board (see Fig. 3-21).
9. Pull off the cable from the sample sensor board (see Fig. 3-21, 1).
2
1
4
2
1 3
The tHb/SO 2 module is calibrated and sealed at the factory („Factory calibration“) and
can only be replaced as a unit.
Never try to open or disassemble the tHb/SO 2 module – the module would not be
calibrated any longer in this case!
1. Dismount the measuring chamber cover (Follow steps 1 to 10 of section 3.5.2 on page 3-9.)
2. Take out the TCon.
3. Open the two mounting screws at the tHb/SO2 – Module.
4. Take out the tHb/SO 2 module (pay attention to the tube connection at the right measuring
chamber side (see Fig. 3-27, 1).
5. Disconnect the cable connection to the analog board (see Fig. 3-27, 2).
5. Push the needle to the left, then upwards and take it out (see Fig. 3-29).
1
4
3
Pay attention to correct orientation of the needle sealing (see Fig. 3-31)!
3. Dismount the flap detection board by carefully disengaging the clips (see Fig. 3-33, 1 and 2).
1 2
4. Open the tension lever (translucent cover) of the peristaltic pump and push the linear clamp
(white plastic part) upwards (see Fig. 3-35).
5. Take out the pump tube.
1 4
2 3
6. Open the mounting screws of the peristaltic pump (see Fig. 3-35, 1 to 4) and take out the pump.
8. Place the pump tube around the pump head (the tubes must not be crossed!).
9. Close the tension lever (translucent cover).
10. Put the analyzer cover back onto the analyzer.
1. Follow the instructions in section 3.7.1 Changing the peristaltic pump on page 3-24.
2. Pull off the pump plate sealing from the plate.
The peristaltic pump can be tested at four speeds via this function.
The pump can be activated only counter-clockwise because a clockwise rotation would draw fluid
out of the Waste container.
3.8 PC tower
Modul
J34
Tower
BV2227
BP2464
(15)
LCD EL0290 (8)
Touch panel
Backlight J2 J3 EN0365
(2) Display
J5 Interface Actuator
DISPLAY PP
MODUL Board
BP2550 BB0979 MC
J1 J4 (20) BV2218
BV2259
BV2224 (40) (25) BV2217
(2) J2 J7 J6
FAN J8 BV2226
(2) J12 (10)
J9
BV2259 MBX Driver Modem
ZPA
(16) J3 Board BV2226
J13 (10)
BV2219 BB0978
J4
BV2228 RS 232
Power Supply J10
EN0364 P1
J1
(144) J5
J12
(7) BV2233
BV2220 (6)
Printer MBX
ST8 (7) Barcode
Interface Board
(26) Scanner
Board ST7 J13 EN0366
BV2225
J15
BV2229 (6)
DATA BUS
Thermal line
printer POWER BUS
10BASE-T
EN0369
1
2
1
2
3
3.8.3 Screen
1
2
3
6. Remove the two screws of the screen (see Fig. 3-42, 1 and 2).
7. Unplug the two connectors, pull off the cable and take out the screen.
Reinstall the two contact strips at the housing sides according to Fig. 3-43, 1 and 2.
1 2
1
2
Reinstall the two contact strips at the housing sides according to Fig. 3-43, 1 and 2.
3.8.6 Printer
5. Lift the printer carefully and unplug the two cables (see Fig. 3-46).
1 2
3 4
3. Lift the rear panel and unplug the cable connections (see Fig. 3-48, 1).
Pay attention to proper cable connections and correct orientation of the PCMCIA slots
to the MBX board!
3.9 Cables
3.10 Fan
3
2
5. Pull off the fan cable from the pump board (see Fig. 3-52, 1).
Test the micro switch unit by moving the bottle compartment cover!
4. Open the mounting screws of the micro switch unit in the bottle compartment
(see Fig. 3-1) and take out the micro switch unit.
Pin assignment:
15 Pin Sub-D, female
1. Dismount the rear panel according to section 3.3 Dismounting the rear panel on page 3-3.
2. Remove the power supply unit from the rear panel; a certain amount of force is needed because
of the mounting with Velcro tape
3.14 Valves
The 3 valve types used in the analyzer have different cable lengths (S, L, XL):
The needed valve head configuration has to be installed before mounting the valve
in the analyzer (see section 3.14.2 Single valve head and 3.14.3 Double valve head).
1. Dismount the rear panel according to section 3.3 Dismounting the rear panel on page 3-3.
2. Disconnect the valve cable from the respective board.
3. Take out the tube(s) - dismount the valve head and/or the bar if necessary.
4. Press out the valve from the backside.
Schematic drawings of the measuring chamber and the bottle compartment help to locate the
selected valve.
Fig. 3-63: Schematic drawing of the bottle compartment without AutoQC (left) and with AutoQC (right)
4. Open the mounting screw of the docking mechanism Waste W (see Fig. 3-66).
5. Pull off the tube from the top of the docking mechanism Waste W and close the flap in order to
protect the translucent plastic from being damaged when pulling out the docking mechanism
(see Fig. 3-67).
5. Pull out the respective docking mechanism C1 (or C2) to the front.
6. Pull off the tube from the backside of the docking mechanism (see Fig. 3-70).
1...Reference solution
2...Conditioning sol.
3...Zero point solution
4...Cleaning solution
Pay attention to correct connection of the tubes to the upper row of connectors at the
docking mechanism (see Fig. 3-71)!
3.18.1 Preparation
1. Open the bottle compartment cover.
2. Take out all bottles.
3. Dismount the rear panel according to section 3.3 Dismounting the rear panel on page 3-3.
4. Dismount the left side cover by opening the three mounting screws in the rear compartment and
in the bottle compartment (see Fig. 3-50, 1 to 3 on page 3-36).
5. Dismount the docking mechanisms (see section 3.17 on page 3-45).
6. Pull out the tubing cover plate behind the docking mechanism Pack C3 to the front
(see Fig. 3-72).
7. Change the tubing in the respective areas according to the following instructions and pictures.
Dismount the valve heads and/or the bars if necessary (see section 3.14 Valves on page 3-41).
Pay attention to the correct routing of the tubes according to the respective pictures,
especially at the valve heads.
Use the tubing diagrams in chapter 5 for checking.
Fig. 3-73 Tubing in the right bottle compartment area without AutoQC
Fig. 3-74 Tubing in the right bottle compartment area with AutoQC
1 2 3
After installing the analyzer tubing set, a FMS volume determination must be
performed (see chapter 6).
Connector:
· Code 39
· Full ASCII Code 39
· UPC/EAN
· Codabar
· UPC-E
· Code 11
· Interleaved 2 of 5
· Code 128
· MSI
· Code 32
The displayed number on the screen must be identical with the number
under the barcode (see Fig. 3-78).
4 AutoQC module...................................................4-1
4.1 General Information ................................................................................ 4-1
4 AutoQC module
When transporting the OMNI C with the AutoQC module installed, do not use the
AutoQC module to carry the instrument.
The AutoQC module is a unit which, in conjunction with the OMNI C, performs automatic quality
control measurements at times pre-programmed by the user.
The module consists of the ampoule holder for 120 ampoules maximally, and a steel tube which,
after being positioned by motors (in Y-, Y-, and Z-direction), breaks the bottom of the ampoule and
withdraws QC fluid. The aspirated QC fluid is transported to the measuring chamber and then
measured.
3 identical DC motors control the unit in all 3 axes. The distance covered and the rotational speed
are measured by an incremental position transducer integrated in the motor.
The absolute position is determined by light barriers. A transmissive photo interrupter is used in all
3 axes. The light barrier for the X-direction is directly on the AQC board.
Sample detection is performed via a light barrier located on the Z-distributor board.
The cover of the AutoQC module is monitored by a hall sensor. When the cover is opened, all
mechanical actions are stopped by the software (except actions in the Component Test).
The ampoule ambient temperature is measured by the temperature sensor and used for correction of
QC measurement values (PO2 and PCO2 only).
The measured value of the temperature sensor can be read under “More - System - Test –
Control sensors – Temperature control”.
Moving the unit into the service position as described in the following may in some
cases only be partly or not at all possible due to defective components. lt should,
however, be performed at any rate to facilitate service work.
If the service positions cannot be approached at all, the vertical and the horizontal
slide can be moved manually, by removing the toothed belt from the carrier pins of the
vertical and/or horizontal slide..
6. Pull off the AutoQC cable from the backside of the OMNI C (see Fig. 4-2).
7. Unlock the AutoQC module by pulling the lever on the backside of the OMNI C (see Fig. 4-2).
9. Open the four cover mounting screws (on the lever) and remove the cover
(see Fig. 4-3, 1 and 2).
10. Remove the two screws at the front of the AutoQC top cover (see Fig. 4-3, 3 and 4)
1 2
3 4
11. Remove the two screws at the back of the AutoQC top cover (see Fig. 4-4, 1 and 2).
12. Remove the AutoQC top cover.
1 2
3 4
13. Remove the two screws at the back of the housing (see Fig. 4-4, 3 and 4).
14. Remove the screws of the side covers (see Fig. 4-5, 1 and 2).
1 2
15. Remove the two screws from the bottom of the AutoQC module (see Fig. 4-6, 1 and 2).
1 2
16. Pull off the AutoQC cable from the AQC board (see Fig. 4-7, 1).
17. Remove the Barex tube security clamp from the docking part by pressing it upwards
(see Fig. 4-8, 1).
1 2
5. Remove the two screws from the AQC magnetic valve (see Fig. 4-11).
6. Disconnect the valve cable.
1
2
The connectors of the flex cable (short) are not coded, therefore make sure that the
contacts are installed pointing toward the OMNI C!
The connectors of the flex cable (short) are not coded, therefore make sure that the
contacts are installed pointing toward the OMNI C!
The connectors of the flex cable (long) are not coded, therefore make sure that the
contacts are installed pointing toward the edge of the board respective the light
barrier.
8. Remove the fastening screws from the motor and remove the motor.
9. Remove the toothed disk from the motor and attach it to the new motor.
9. Remove the toothed wheel and the guiding sleeve from the motor and connect it to the new
motor.
Do not install or remove the AQC temperature sensor when the system is connected to
power!
4. Disconnect the cable from the back of the AQC temperature sensor.
5. Remove the AQC temperature sensor and clean the vertical slide cover from all glue rests.
6. Paste the new AQC temperature sensor on the vertical slide cover
5. Dismount the wash port holder by pressing the two snap locks together.
6. Insert a new wash port into the wash port holder.
1. Press „More – System – Test – Valves and aggregates – AutoQC position test –
Service position“ to move the motors into the service position (Remove the ampoule mats, the
ampoule holder and the AQC tray).
2. Open the cover of the AutoQC module.
3. Dismount the AutoQC top cover and the AutoQC cover (Follow the steps 9 to 12 in section 4.2
Dismounting the housing on page 4-2).
4. Open the screw (see Fig. 4-17, 1) and remove the wash port angle.
5. Disconnect the wash tube at the wash port and at the docking part.
6. Remove the wash tube from the magnetic valve.
Disconnect the cable from the back of the AQC temperature sensor.
Pull out the vertical slide cover to the front.
Remove the screw at the carrying wheel and remove the old toothed belt.
Stretch the new toothed belt over the toothed disk of the motor and over the carrying wheel.
Fasten the carrying wheel.
Insert the toothed belt into the carrier pins of the vertical slide.
5.5 Aspiration of C1, C2 or Mix / washing the measuring chamber ............. 5-5
5.1
Peristaltic pump
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Pa
V4
Tubing diagram OMNI C BG/ISE
V14
Air
FMS V1 V2 V11
C3 V9 Air
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8
solution
Cleaning
solution
Waste Solution C1 Solution C2 Reference
solution
5 Fluidic procedures
5-1
5-2
5.2
Peristaltic pump
5 Fluidic procedures
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Ref MCon MCon V3
Na Cl pH Ca K O2 CO2 SS1
Pa
V4
V14
Air
Tubing diagram OMNI C BG/ISE/AQC
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Pa Pressure Sensor
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Ref MCon MCon V3
Na Cl pH Ca K O2 CO2 SS1
Pa
V4
V14
Air
Aspiration of sample / QC material
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Pa
V4
V14
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Aspiration of C1, C2 or Mix / washing the measuring chamber
5 Fluidic procedures
5-5
5-6
5.6
Peristaltic pump
5 Fluidic procedures
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Ref MCon MCon V3
Na Cl pH Ca K O2 CO2 SS1
Pa
V4
V14
Air
V1 V2 V11 V13
Washing the needle (over the bypass)
FMS
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Pa
V4
V14
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
5 Fluidic procedures
5-7
5-8
5.8
Peristaltic pump
5 Fluidic procedures
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Ref MCon MCon V3
Na Cl pH Ca K O2 CO2 SS1
Pa
V4
V14
Air
Aspiration of the cleaning solution
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Pa
V4
V14
Filling of the reference electrode
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
5 Fluidic procedures
5-9
5-10
2
1
5 Fluidic procedures
Peristaltic pump
Air
V5
V6 V7
5.10 Conditioning cycle
Pa
V4
V14
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Pa
V4
V14
5.11 Aspiration of the AutoQC material
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
5 Fluidic procedures
5-11
5-12
Peristaltic pump
5 Fluidic procedures
Air
V5
V6 V7
Measuring chamber SS2 Needle
TCon tHb/sO2
Ref MCon MCon V3
Na Cl pH Ca K O2 CO2 SS1
5.12 Washing the AutoQC
Pa
V4
V14
Air
C3 V9 Air V12
O2 Zero point
solution
W C1 C2 Air
Conditioning V10 V8 AutoQC
solution
Cleaning V17
solution
Waste Solution C1 Solution C2 Reference
solution
C1 : C2
2 : 1 = Mix1
1 : 2 = Mix2
5.14 Calibrations
Calibration for ready
Automatic selection of calibration type!
System calibration
Conductivity C1 calibration + Conductivity C2 calibration + Conditioning cycle + O2 zero point
calibration + 2P (Mix2) calibration + O2 air calibration + 1P (Mix1) calibration
Conductivity calibration
Conductivity C1 calibration + Conductivity C2 calibration
1 Point calibration
O2 air calibration + 1P (Mix1) calibration
2 Point O2 calibration
O2 zero point calibration + O2 air calibration + 1P (Mix1) calibration
6 Service area
6.1.1 Password
Several software functions of the OMNI C are only accessible for the
service technician for safety reasons.
Upon opening this functions a password will be requested.
The default password for this functions is „$ad“.
Delete databases
This function erases all databases and log files in the internal flash memory.
Reset settings
This function restores all analyzer settings to the default values (a new setup file with the default
values is created).
A calibration of SS1 and SS2 with alternating fluid and air packets can be started here:
Calibrate
Read the actual air pressure from a precision barometer and enter the barometric pressure value.
This value will be stored together with the measured mV value of the baro sensor as a
calibration point.
(Together with the slope, air pressure can be calculated)
The procedure for determining the FMS volume can be started here:
Proceed as follows:
1. Connect the filled syringe (air bubble free!) to the FMS cross piece by using the tube.
(It is helpful to use a coloured fluid, e.g. Combitrol).
5. To be sure that no drops are in front of the fluid, inject until the fluid reaches the tube of V5 and
pull slowly back until the fluid reaches the edge of the sample sensor cover again.
6. The difference between the two readings is the FMS tubing volume (in µl).
Designed for well trained service technicians only, this function offers the opportunity to customize
the test routines for servicing and fault finding.
The consequence action delay avoids the execution of consequence actions by the analyzer when
changing between test functions.
This allows e.g. switching the valves, then changing the test function and perform a manually
controlled aspiration via activation of the peristaltic pump.
When this mode is activated, no automatic test preparations are done by the analyzer (e.g. opening
the air valve when measuring the barometer value).
This function should be activated only by well trained service technicians because the
risk of damaging the analyzer by improper handling is very high in this mode!
Destination
The storage location for the log data (analyzer actions, etc.) can be entered here:
Days
Determines the period of time (in days), for which log data will be stored.
Bytes
Determines the maximum size for the log file.
Debug Level
Boot device
Determines the boot location:
Sample counter
Reset the sample counter by pressing this button.
QC counter
Reset the QC counter by pressing this button.
7 Annual maintenance
7.1 OMNI C
· Decontamination / Cleaning (see chapter 1, Introduction)
· Replace the needle sealing (see chapter 3, Components)
· Replace the fill port (see chapter 3, Components)
· Replace the pump head (see chapter 3, Components)
· Replace the pump tube (see chapter 3, Components)
· Replace the measuring chamber tubing set (see chapter 3, Components)
· Replace the (analyzer) tubing set (see chapter 3, Components)
· Check the barometer value and calibrate if necessary (see chapter 6, Service area)
All necessary components are included in the AQC maintenance kit OMNI C.
8 Troubleshooting..................................................8-1
8.1 System Stops .......................................................................................... 8-1
8.1.1 General Information........................................................................................... 8-1
8 Troubleshooting
A stop of all running actions will occur if danger for the user (e.g. by an overflow of the waste
while opening the docking mechanism of the W waste bottle) or for the analyzer exists or a proper
operation of the analyzer is not possible because of technical problems.
If a proper operation is not possible, but the complete control of the functions of the analyzer is
available (e.g. temperature alarm, measuring chamber cover open, flap open, fill level alarm) all
running actions will be finished and Mix 1 will be aspirated into the measuring chamber; the
resulting measuring data will be marked resp. discarded.
Unrecoverable system stops remain on the screen and fulfil an emergency program if possible, so
that the analyzer stays operating (wetting of the electrodes / keeping the tubing free).
It is possible to access the System menu if no automatic actions are running currently (except in the
warm-up period).When changing to the System menu during a System Stop, the System Stop will be
stopped and reactivated when closing the System Menu. If e.g. there is a change from a temperature
System Stop to the System Menu and thereafter the System Menu is closed, the analyzer will show
the temperature System Stop again. If e.g. a system calibration has been started in the System Menu,
it can not be conducted immediately because the System Stop has to be removed first.
After all System Stops have been removed, consequence actions will be executed. There is a
common list of consequences, which impedes double actions. This means that if e.g. „washing“ was
activated several times, it is then performed only once.
Warm-up!
Cause
Temperature in the measuring chamber, in the measuring chamber cover and/or in the tHb/SO 2
module is too low.
Appears after switching on the analyzer and after opening the measuring chamber cover for longer
than 5 seconds.
Removal
Close measuring chamber cover.
Wait! Will be removed automatically as soon as the proper temperatures are reached!
Out of operation!
Cause
The instrument has been taken out of operation.
Removal
Perform the installation procedure.
Removal
Close the AutoQC cover!
Check the AutoQC cover switch („More – System – Test – Control sensors – Monitoring
Sensors“),change components if necessary.
Fatal error!
Cause
Fundamental software functions can not be performed.
Removal
Press the "Reboot" button on the screen.
Switch the analyzer off /on.
Removal
Close bottle compartment cover.
Check to be sure that the bottles are inserted completely.
Check the bottle compartment cover micro switch („More – System – Test – Control sensors
– Monitoring sensors“), change if necessary.
Removal
Delete data records in order to free up additional memory
Removal
Close docking mechanism Pack C3
Check C3 micro switch („More - System – Test – Control sensors – Monitoring sensors“),
change if necessary.
Consequence actions
Auto-preparing of the Pack C3 solutions
Removal
Exchange solutions C1, C2 and Pack C3!
Exchange or empty W Waste bottle!
Consequences
If solutions C1 and C2 have been changed: => prepare solutions C1 and C2 => conductivity
calibration
If Pack C3 has been changed resp. the C3 docking mechanism was opened:
=> aspirate all Pack solutions.
If the waste bottle was changed resp. emptied: => Waste bottle fill level measurement
Flap open!
Cause
Flap was opened:
- during a measurement
- during a calibration
Flap detection is defective
Removal
Close flap.
Check function of the flap detection board („More - System – Test – Control sensors –
Monitoring sensors“), change if necessary.
Removal
Start a System calibration.
Check the tubing of solutions C1 and C2
Consequences
Last measurement/calibration could be invalid.
Removal
Close measuring chamber cover.
Check the measuring chamber cover sensor („More - System – Test – Control sensors –
Monitoring sensors“), check cable, change components if necessary.
Consequences
After opening without changing the electrodes: warm-up
After changing (an) electrode(s): warm-up and system calibration
Removal
Delete data from the PCMCIA card to have extra memory space.
Put PCMCIA card in a PC with a suitable place, export data and delete data from the card
afterwards.
Format new the PCMCIA card in the Roche OMNI C („More - System - Test - PC Components
– PCMCIA card – More“).
Removal
Sample sensors calibration
Check tubing
Change sample sensor board
Removal
Check under „More - Test – Valves & Aggregates – Peristaltic pump“. if the following
values are displayed:
Pump volume: 40 - 70µl
FMS volume: 920 - 1200µl
If the displayed values are out of the limits, perform a FMS volume determination and correct the
FMS volume value.
System error!
Cause
Program crashed
Removal
Switch the analyzer off/on.
In case of recurrence the electronic is defective, change components if necessary.
Temperature error!
Cause
Temperature is outside of the specified range.
Measuring chamber: 37.00 °C ± 0.2 °C
Measuring chamber cover left and right: 37.00 °C ± 0.2 °C
tHb-/SO 2 module: 37.00 °C ± 0.2 °C
Heating is defective.
Measuring chamber cover cable is defective.
Removal
Reduce / raise the room temperature.
Change measuring chamber cover or measuring chamber.
Removal
Reinsert the waste container
Check Waste Container micro switch („More - System – Test – Control sensors –
Monitoring sensors“), change if necessary.
Consequences
Waste fill level measurement
Removal
Change or empty the W Waste Container according to the instructions
Consequences
Waste fill level measurement
Info
Info text Description Removal
No.
Reproducibility couldn’t be
501 Reprod. check. pend. checked because of lack of Perform „Calibration for ready“
calibration cycles.
Perform „Internal cleaning of sample
path“; Perform „Calibration for ready“;
Check environment conditions
The signal noise of an electrode is
1000 Signal noise (vibrations, electrical fields), Check
outside the specified limits.
electrode for air bubbles in the
electrolyte, Change electrode if
necessary
Perform „Internal cleaning of sample
path“; Perform „Calibration for ready“;
Check environment conditions
The sensor signal of an electrode
1001 Signal distorted (vibrations, electrical fields), Check
is distorted.
electrode for air bubbles in the
electrolyte, Change electrode if
necessary
While measurement: The sample is not
suitable for the selected sample type or
The result is outside the specified
1004 Out of range the measurement result is outside the
measuring range.
specified rangesWhile calibration:
Check electrode
The barometer value is outside the Check the barometric pressure value,
1009 Baro out of range
specified range. Enter barometric pressure value
When calculating a result 2
samples are taken into
2nd measurement not
1010 consideration, but in this case one Perform the missing measurement
available
was missing e.g. calculating the
Shunt values.
The sample was identified as
1012 Interferences (2) Unsuitable sample
unsuitable
The sample was inhomogeneous
1013 Sensor signal unstable Homogenize the sample by mixing.
(sampling criteria II was violated)
The tHb/SO2 module/algorithm
has detected an interference
1022 Interferences (3) Unsuitable sample
substance, e.g. COHb is higher
than 15%
A calculation error for calculating
1023 Interferences (4) Unsuitable sample
the SO2 result was detected.
Info
Info text Description Removal
No.
The tHb result is higher than The measurement result is outside the
1024 Out of range (+)
25g/dL. specified ranges
The measurement result is outside the
1025 Out of range (-) The tHb result is lower than 3 g/dL.
specified ranges
The SO2 result is between 0 and The measurement result is outside the
1026 Out of range (-)
50% specified ranges
SO2 error because the tHb result The measurement result is outside the
1054 tHb not ok
is higher than 25 g/dL specified ranges
SO2 error because the tHb result The measurement result is outside the
1055 tHb not ok
is lower than 3 g/dL. specified ranges
Measurement not
1065 Improper tHb sample scan. Switch analyzer off/on
called
Electrode is deactivated for
1070 Not activated measurement.(Electrode status Activate electrode for measurement
MS_DUMMY)
Electrode is deactivated by
Remote lockout
1071 Remote lock Remove Remote lockout(OMNILINK)
(OMNILINK)(Electrode status
MS_RMLOCK)
If measurement result is needed, do
Electrode was temporarily
1072 Not activated not temporarily deactivate electrode for
deactivated for measurement.
measurement.
Perform „Calibration for ready“, Check
1073 Calibration pending 2 point calibration is missing. electrode, Change Electrode if
necessary
Electrode is in 2 point calibration
1074 Slope nOK Perform „Calibration for ready“
alarm.
1075 Calibration pending 1 point calibration is missing. Perform „Calibration for ready“
1 point calibration failed; Info No.
1076 1P Error Perform „Calibration for ready“
in calibration report
Parameter is locked because a QC
1077 QC lock status Repeat QC-measurement
measurement failed.
Check solution C2 fill level, Check
Conductivity calibration with tubing for leaks, Check (and clean) C2
1078 Conductivity C2 nOk
solution C2 failed. docking mechanism, Check sample
port and needle
Check solution C2 fill level, Check
Conductivity calibration with tubing for leaks, Check (and clean) C1
1079 Conductivity C1 nOk
solution C1 failed. docking mechanism, Check sample
port and needle
1501 fill FMS FMS not ready. Switch analyzer off/on
1502 fill FMS FMS input parameters invalid. Switch analyzer off/on
1503 fill FMS Wrong valve position(s). Switch analyzer off/on
While measurement: Avoid air bubbles
in the sample, Check for sufficient
sample volume, Wet sample path
1511 fill check Fill error, measuring chamber left
(Perform blood measurement) While
calibration: Check C1 and C2 tubing for
leaks, Check sample port and needle
While measurement: Avoid air bubbles
in the sample, Check for sufficient
sample volume, Wet sample path
1512 fill check Fill error, measuring chamber right.
(Perform blood measurement) While
calibration: Check C1 and C2 tubing for
leaks, Check sample port and needle
Info
Info text Description Removal
No.
While measurement: Avoid air bubbles
in the sample, Check for sufficient
sample volume, Wet sample path
1513 fill check Fill error, location unknown.
(Perform blood measurement) While
calibration: Check C1 and C2 tubing for
leaks, Check sample port and needle
Check sample volume according to
1515 Sample too small Insufficient sample volume.
specifications.
Proper sampling (avoid air bubbles in
1516 Sample fragmented The sample contains air bubbles.
the sample)
1521 fill FMS FMS not ready. Switch analyzer off/on
1522 fill FMS FMS input parameters invalid. Switch analyzer off/on
1523 fill FMS Peristaltic pump does not move. Switch analyzer off/on
Problem with FMS. Improper air /
wash Packages (no 0,6s air before
Check pump tube, Check FMS volume,
1525 fill FMS first wash Package, after
Determine FMS volume
production of CON/CAL).
timeout=15s.
Problem with FMS. Missing Check solution C1/C2 fill level, Check
1526 fill FMS unfragmented wash Package, tubing for leaks, Check (and clean)
timeout=10s. C1/C2 docking mechanisms
Problem with FMS. Missing
Check solution C1/C2 fill level, Check
unfragmented wash Package (after
1527 fill FMS tubing for leaks, Check (and clean)
CON/CAL production)
C1/C2 docking mechanisms
Timeout=15s,
1541 fill FMS FMS not ready. Switch analyzer off/on
1542 wash MC Wrong valve position(s). Switch analyzer off/on
Problem in the washing procedure.
Check solution C1/C2 fill level, Check
SS2 has been empty too early.
1545 wash MC tubing for leaks, Check (and clean)
(FMS is still working), CON before
C1/C2 docking mechanisms
MK
Problem in the washing procedure. Check solution C1/C2 fill level, Check
1546 wash MC CON Package not detected, tubing for leaks, Check (and clean)
timeout=10s, C1/C2 docking mechanisms
Problem in the washing procedure. Check solution C1/C2 fill level, Check
1547 wash MC SS2 has been empty too early. tubing for leaks, Check (and clean)
(FMS is still working), CON in MK C1/C2 docking mechanisms
Problem in the washing procedure. Perform valve test, Check FMS tubing,
1548 wash MC End of CON filling couldn’t be Check FMS volume, Determine FMS
reached, timeout=10s. volume
No reference contact before Check sample port and needle, Check
1554 pos Ref.
reference solution aspiration. measuring chamber for clots
Check tube and plug of Reference
Reference solution aspiration electrode, Check fill level of Reference
1556 pos Ref. problem (hub is too small after 3s solution (by auto-preparing), Check
of aspiration). (and clean) C3 docking mechanism,
Check sample port and needle.
1561 asp. Mix1 FMS not ready. Switch analyzer off/on
1562 asp. Mix1 Wrong valve position(s). Switch analyzer off/on
Proper filling of reference electrode
1564 check Ref couldn’t be checked because of air Check sample port and needle
bubble in the measuring chamber.
Info
Info text Description Removal
No.
Perform “Fill reference electrode”,
Properly positioning of reference Check tube and plug of Reference
solution not possible, air bubbles in electrode, Check fill level of Reference
1565 fill Ref
the reference tubing, (Reference solution (by auto-preparing), Check
difference >30) (and clean) C3 docking mechanism,
Check sample port and needle
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early. tubing for leaks, Check (and clean)
1566 Aspirate Mix1
Separating air bubble is before C1/C2 docking mechanisms,
MCC-MCO) Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early. tubing for leaks, Check (and clean)
1567 Aspirate Mix1
Separating air bubble is at MCC- C1/C2 docking mechanisms,
MCO). Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early. tubing for leaks, Check (and clean)
1568 Aspirate Mix1
Measuring chamber is filled but air C1/C2 docking mechanisms,
separation bubble before V6). Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (timeout=9s, filling of tubing for leaks, Check (and clean)
1569 Aspirate Mix1
measuring chamber terminated C1/C2 docking mechanisms,
because of invalid CAL Package). Determine FMS volume
1581 Aspirate Mix2 FMS not ready. Switch analyzer off/on
1582 Aspirate Mix2 Wrong valve position(s). Switch analyzer off/on
Improper filling of measuring
Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early.
1585 Aspirate Mix2 tubing for leaks, Check (and clean)
Separating air bubble is before
C1/C2 docking mechanisms
MCC-MCO)
Improper filling of measuring
Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early.
1586 Aspirate Mix2 tubing for leaks, Check (and clean)
Separating air bubble is at MCC-
C1/C2 docking mechanisms
MCO)
Improper filling of measuring
Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early.
1587 Aspirate Mix2 tubing for leaks, Check (and clean)
Measuring chamber is filled but air
C1/C2 docking mechanisms
separation bubble before V6).
Improper filling of measuring
Check solution C1/C2 fill level, Check
chamber (timeout=9s, filling of
1588 fill MC tubing for leaks, Check (and clean)
measuring chamber terminated
C1/C2 docking mechanisms
because of invalid Mix2 Package)
Perform proper sample aspiration
Improper filling of measuring
within 15 seconds, Check tube to air at
1601 Aspirate sample chamber (timeout=15s, no sample
V9, Wet sample path (Perform blood
at SS1)
measurement)
Aspiration procedure incorrect.
Wet sample path (Perform blood
1602 Aspirate sample (SS1 sees fluidic residues before
measurement)
measurement has been started)
Perform proper sample aspiration
Aspiration procedure incorrect. within 15 seconds, Check tube to air at
1603 Aspirate sample
(time out=15s, no sample at SS1) V9, Wet sample path (Perform blood
measurement)
Check sample sensor calibration,
1605 SS1 inactive Sample sensor 1 is not active.
Perform sample sensor calibration
Check sample sensor calibration,
1606 SS2 inactive Sample sensor 2 is not active.
Perform sample sensor calibration
Info
Info text Description Removal
No.
Improper filling of tHb-/SO2
1622 pos. sample module (Timeout=10s, no air Perform proper sample aspiration
separation bubble)
Improper filling of tHb-/SO2
1623 pos. sample Perform proper sample aspiration
module (Timeout=10s, no sample)
Improper filling of tHb-/SO2
Perform proper sample aspiration
1624 pos. sample module (Timeout=10s, no
(Avoid air bubbles in the sample)
reproducibility with sample)
Perform proper sample aspiration,
Improper filling of measuring
Perform „Internal cleaning of sample
1642 pos. sample chamber (Timeout=15s, no air
path“, Wet sample path (Perform blood
separation bubble) at MCI-MCM)
measurement)
Perform proper sample aspiration,
Improper filling of measuring
Perform „Internal cleaning of sample
1643 pos. sample chamber (Timeout=15s, no sample
path“, Wet sample path (Perform blood
at MCI-MCM)
measurement)
Perform proper sample aspiration,
Improper filling of measuring
Perform „Internal cleaning of sample
1644 pos. sample chamber (Timeout=15s, no air
path“, Wet sample path (Perform blood
separation bubble at MCC-MCO)
measurement)
Perform proper sample aspiration,
Improper filling of measuring
Perform „Internal cleaning of sample
1645 pos. sample chamber (Timeout=15s, no sample
path“, Wet sample path (Perform blood
at MCM-MCO)
measurement)
Improper AutoQC sampling (SS3
1668 asp. AQC Perform „Wash AutoQC“
detects no sample)
Improper AutoQC sampling (SS3
1669 asp. AQC Perform „Wash AutoQC“
detects no air).
Improper aspiration of calibration
Check tube to air at V2, Check sample
1704 asp. C1 solution C1 (time out=30s, no air
sensor calibration
Package at SS2).
Improper aspiration of calibration Check solution C2 fill level, Check
solution C1 (time out=30s, no tubing for leaks, Check (and clean) C1
1705 asp. C1
calibration solution C1 Package at docking mechanism, Check sample
SS2 detected) port and needle
Improper aspiration of calibration Check solution C2 fill level, Check
solution C1 (time out=30s, no tubing for leaks, Check (and clean) C1
1707 asp. C1
calibration solution C1 Package at docking mechanism, Check sample
SS2 detected port and needle
Improper filling of measuring Check solution C2 fill level, Check
chamber (timeout=10s, no tubing for leaks, Check (and clean) C1
1722 C1 pos.
calibration solution C1 at MCI- docking mechanism, Check sample
MCC detected). port and needle
Improper aspiration of calibration
Check tube to air at V2, Perform
1744 asp. C2 solution C2 (time out=30s, no air
sample sensor calibration
Package at SS2 detected).
Improper aspiration of calibration Check solution C2 fill level, Check
solution C2 (time out=30s, no tubing for leaks, Check (and clean) C2
1745 asp. C2
calibration solution C2 pre- docking mechanism, Check sample
Package at SS2 detected) port and needle
Improper aspiration of calibration Check solution C2 fill level, Check
solution C2 (time out=30s, no tubing for leaks, Check (and clean) C2
1747 asp. C2
calibration solution C2 Package at docking mechanism, Check sample
SS2 detected). port and needle
Info
Info text Description Removal
No.
Improper filling of measuring Check solution C2 fill level, Check
chamber (timeout=10s, no tubing for leaks, Check (and clean) C2
1762 C2 pos.
calibration solution C2 at MCI- docking mechanism, Check sample
MCC detected). port and needle
Improper filling of measuring
1782 asp. air chamber (timeout=15s, measuring Check tube to air at V9, perform V6 test
chamber is not filled with air)
Timeout=25s, O2 electrode scan
1784 asp. air Switch analyzer off/on
not completed(no GrpCompl)
Improper filling of measuring
chamber (air separation bubble Check tubing for leaks, Check tube to
1802 pos. Cond.
has not been detected by MCC- air at V9, perform V9 test
MCO)
Improper filling of measuring
Check fill level of Conditioning solution
chamber (Conditioning solution
1803 pos. Cond. (by auto-preparing), Check (and clean)
has not been detected at MCC-
C3 docking mechanism
MCO)
Improper filling of measuring
Check fill level of Conditioning solution
chamber (Conditioning solution
1804 Cond. pos. (by auto-preparing), Check (and clean)
has not been detected at MCM-
C3 docking mechanism
MCO)
Check measuring chamber for leaks,
End of Mix 1 was not detected at
1808 Cond. pos. Wet sample path (Perform blood
SS1 when refilling
measurement)
1861 fill FMS FMS not ready Switch analyzer off/on
1862 fill FMS Wrong valve position(s). Switch analyzer off/on
Perform “Fill reference electrode”,
Proper positioning of reference Check tube and plug of Reference
solution not possible, air bubbles in electrode, Check fill level of Reference
1865 fill Ref
the reference tubing (reference solution (by auto-preparing), Check
difference >30). (and clean) C3 docking mechanism,
Check AutoQC steel tube
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early, tubing for leaks (including AutoQC),
1866 aspirate Mix1
air separation bubble before MCC- Check (and clean) C1/C2 docking
MCO) mechanisms, Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early. tubing for leaks (including AutoQC),
1867 aspirate Mix1
Separating air bubble is at MCC- Check (and clean) C1/C2 docking
MCO). mechanisms, Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (SS2 is empty too early. tubing for leaks (including AutoQC),
1868 aspirate Mix1
Measuring chamber is filled but air Check (and clean) C1/C2 docking
separation bubble before V6). mechanisms, Determine FMS volume
Improper filling of measuring Check solution C1/C2 fill level, Check
chamber (timeout=9s, filling of tubing for leaks (including AutoQC),
1869 aspirate Mix1
measuring chamber terminated Check (and clean) C1/C2 docking
because of invalid CAL Package). mechanisms, Determine FMS volume
Check fill level of PO2 zero-point
Improper aspiration (PO2 zero- solution (by auto-preparing), Check
1900 asp. O2zero point solution has not been (and clean) C3 docking mechanism,
detected at SS1) Check tubing for leaks, Check sample
port and needle
Info
Info text Description Removal
No.
Check fill level of Cleaning solution (by
auto-preparing), Check (and clean) C3
Improper aspiration of Cleaning
1901 asp. clean. sol. docking mechanism, Check tubing for
solution (not detected at SS1)
leaks, Check tubing for leaks, Check
sample port and needle
Check tube and plug of Reference
Improper aspiration of Reference electrode, Check fill level of Reference
1902 asp. Ref solution (not detected at MCC- solution (by auto-preparing), Check
MCO) tubing for leaks, Check (and clean) C3
docking mechanism
Improper aspiration of calibration Check solution C2 fill level, Check
1903 aspirate C1 solution C1 (no calibration solution tubing for leaks, Check (and clean) C1
C1 at SS2 detected) docking mechanism
Improper aspiration of calibration Check solution C2 fill level, Check
1904 aspirate C2 solution C2 (no calibration solution tubing for leaks, Check (and clean) C2
C2 at SS2 detected) docking mechanism
1998 Undefined Undefined error Perform Software update
The mixing ratio of the solutions Check solution C1 and C2 fill levels
2002 FMS out of range C1 and C2 is outside the specified (also for equal levels), Perform valve
range. test, Check FMS tubing
2 point calibration failed, because
2004 1P Error no valid 1 point calibration was Perform „Calibration for ready“
available.
The standard deviation is outside
2006 Reproducibility nOK the specified range for checking Perform „Calibration for ready“
the reproducibility (including tHb)
Perform „Internal cleaning of sample
2007 Signal noise tHb module is polluted.
path“
tHb error because of improper
calibration values caused by
Perform „Internal cleaning of sample
2009 Pollution error polluted tHb module. A cleaning
path“
will be performed during the next
system calibration.
2011 Invalid EEPROM data tHb EEPROM values are incorrect. Change tHb/SO2 module
tHb factory calibration is
2012 Invalid factory cal. Change tHb/SO2 module
improperly stored in the EEPROM.
2014 Cal. time expired Timeout during calibration. Perform „Calibration for ready“
The electrode drift is outside the
2015 Ready alarm Perform „Calibration for ready“
specified limits.
Check fill level of Conditioning solution
Na electrode is in 2 point
(by auto-preparing), Check (and clean)
2016 conditioning nOK calibration alarm, because a
C3 docking mechanism, Perform
conditioning is missing.
“Conditioning cycle”
3030 asp. AQC Initialisation not o.k. Switch analyzer off/on
3031 asp. AQC SS3 calibration not o.k. Perform „Wash AutoQC“
The AutoQC steel tube has not
3032 asp. AQC Perform „AutoQC position test“
reached the ampoule position.
The AutoQC steel tube has not
3033 asp. AQC Perform „AutoQC position test“
reached the start position Z.
The AutoQC steel tube has not
3034 asp. AQC Perform „AutoQC position test“
reached the end position Z.
The AutoQC steel tube has not
3035 wash AQC Perform „AutoQC position test“
reached the wash position.
The AutoQC steel tube has not
3036 wash AQC Perform „AutoQC position test“
reached the ampoule XY
Info
Info text Description Removal
No.
The AutoQC steel tube has not
3037 wash AQC reached the ampoule position Z Perform „AutoQC position test“
(down).
The AutoQC steel tube has not
3038 wash AQC Perform „AutoQC position test“
reached the ampoule Z (up).
The AutoQC steel tube has not
3039 wash AQC reached the ampoule position Z Perform „AutoQC position test“
(up)
The AutoQC steel tube has not
3040 wash AQC reached the wash position after Perform „AutoQC position test“
back-wash procedure.
Check:
· the sample port for damage (by glass splinters) and soiling
· the wash plate for pollution
· the needle for bends, damage and soiling
· the needle sealing for damage and soiling
Prepare solutions C1 and C2 under „More - System – Tools – Fluid actions – Auto
preparation routines“ and observe the tubing in the right measuring chamber window and at the
peristaltic pump.
The solutions must be visible as homogeneous fluid packages without air bubbles.
The reference solution must be visible in the tube of the reference as homogeneous fluid package
without air bubbles when filling the reference electrode.
9 Index
Numerics
1 Point calibration 5-13
2 Point calibration excl. O2 5-13
2 Point calibration incl. O2 5-13
2 Point O2 calibration 5-13
A
Actuator board 3-44
Air tube at V2 is blocked by crystallization 8-15
Air tube at V9 is blocked by crystallization 8-16
Analyzer tubing 3-49
Annual maintenance 7-1
AQC cover open! 8-2
Aspiration of C1, C2 or Mix / washing the measuring chamber 5-5
Aspiration of sample / QC material 5-4
Aspiration of the AutoQC material 5-11
Aspiration of the cleaning solution 5-8
Aspiration of the O2 zero point solution 5-7
AutoQC module 4-1
B
Barcode scanner 3-53
Baro sensor calibration 6-6
Boot parameters 6-15
Bottle compartment cover open! 8-3
C
Cables 3-35
Calibration for ready 5-13
Calibration solutions C1 and C2 have flown back 8-15
Calibrations 5-13
Changing the actuator board 3-44
Changing the analog board 3-12
Changing the AQC board 4-6
Changing the AQC cable 4-16
Changing the AQC docking part 4-16
Changing the AQC magnetic valve 4-6
D
Decontamination 1-4
Determination of the FMS volume 6-8
Disinfectants 1-3
Dismounting the horizontal slide cover 4-5
Dismounting the housing 4-2
Dismounting the rear panel 3-3
Docking mechanisms 3-45
Double valve head 3-42
E
Error messages 8-7
Error sources 8-15
ESD protection measures 1-6
F
Fan 3-36
Fatal error! 8-2
Fill level alarm! 8-3
Filling of the reference electrode 5-9
Flap open! 8-4
Flash memory full! 8-3
Fluid Pack switch! 8-3
Fluidic procedures 5-1
FMS (Fluid Mixing System) 3-5
I
Important notes and warnings 1-2
Introduction 1-1
L
Leak at the tube under the V1 bar 8-16
Leaks at the docking mechanisms 8-17
Leaks at the tube or plug of the reference electrode 8-17
Leaks in the Sample port / needle area 8-17
List of error messages (sorted by Info No.) 8-7
List of System Stops 8-2
Log unit 6-13
M
Measurement procedure 5-14
Measuring chamber 3-7
Measuring chamber cover open! 8-4
Micro switch unit (Bottle compartment cover switch) 3-37
Micro switches 3-39
O
Out of operation! 8-2
P
Password 6-1
Password protected software functions 6-1
PC tower 3-27
PCMCIA memory full! 8-5
Peristaltic pump 3-24
Power supply unit 3-40
Printer 3-33
Pump board 3-45
Pump cal. failed! 8-5
R
Reset counters 6-16
Roche Deproteinizer 1-3
Roche Service 1 6-2
S
Sample detection failed! 8-5
Sample port module 3-20
Sample sensor calibration 6-5
Screen 3-30
Serial number 6-14
Service area 6-1
Shutdown 3-2
Single valve head 3-41
Symbols 1-1
System calibration 5-13
System calibration procedure 5-16
System error! 8-5
System Stops 8-1
T
Temperature error! 8-6
Testing the barcode scanner 3-54
Testing the micro switches 3-40
Testing the peristaltic pump 3-27
Testing the valves 3-43
The FMS principle 5-13
The OMNI C stays switched off longer than 24 hours 3-2
The OMNI C stays switched off shorter than 24 hours 3-2
Troubleshooting 8-1
Tubing diagram OMNI C BG/ISE 5-1
Tubing diagram OMNI C BG/ISE/AQC 5-2
Tubing in the area of docking mechanism Pack C3 3-49
Tubing in the FMS area (Fluid Mixing System) 3-51
Tubing in the pump area 3-51
Tubing in the right bottle compartment area with AutoQC 3-50
Tubing in the right bottle compartment area without AutoQC 3-50
Tubing of the reference electrode connector 3-52
V
Valves 3-41
Valves and sensors 5-3
W
Warm-up! 8-2
Washing the AutoQC 5-12