Doc316 53 01118

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Phosphorus, Reactive DOC316.53.

01118

(Orthophosphate)
USEPA 1, 2 PhosVer® 3 Method3 Method 8048
0.06 to 5.00 mg/L PO43– (0.02 to 1.60 mg/L P) Test ‘N Tube™ Vials
Scope and application: For drinking water, wastewater and seawater.
1 USEPA accepted for reporting wastewater analysis. Procedure is equivalent to USEPA and Standard Method 4500-P E for wastewater.
2 USEPA Accepted for reporting for drinking water analysis. Procedure is an acceptable version of EPA Method 365.1, approved at 40 CFR
part 141 NPDWR compliance monitoring.
3 Adapted from Standard Methods for the Examination of Water and Wastewater.

Test preparation

Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows adapter and light shield requirements for the instruments that use them.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information for test tubes
Instrument Adapters Light shield
DR 6000, DR 5000 — —
DR 3900 — LZV849
DR 3800, DR 2800, DR 2700 — LZV646
DR 1900 9609900 (D1) —
DR 900 4846400 Cover supplied with the instrument

Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started.
For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water
in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results
automatically with the reagent blank adjust option.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.

Items to collect
Description Quantity
®
PhosVer 3 Reagent Powder Pillow 1
Reactive Phosphorus Test 'N Tube Vial 1
Funnel, micro 1

1 The D adapter is not available with all instrument versions.

1
Items to collect (continued)
Description Quantity

Light shield or adapter (For information about sample cells, adapters or light shields, refer to
1
Instrument-specific information on page 1.)
®
Pipet, TenSette , 1.0- to 10.0-mL, with pipet tips 1
Test tube rack 1

Refer to Consumables and replacement items on page 4 for order information.

Sample collection and storage


• Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1)
hydrochloric acid and rinsed with deionized water.
• Do not use a detergent that contains phosphate to clean the sample bottles. The
phosphate in the detergent will contaminate the sample.
• Analyze the samples as soon as possible for best results.
• If immediate analysis is not possible, immediately filter and keep the samples at or
below 6 °C (43 °F) for a maximum of 48 hours.
• Let the sample temperature increase to room temperature before analysis.

Test 'N Tube procedure

Start

1. Start program 535 P 2. Add 5.0 mL of sample to 3. Put the cap on the vial. 4. Clean the vial.
React. PV TNT. For a Reactive Phosphorus Test Invert to mix.
information about sample 'N Tube Vial.
cells, adapters or light
shields, refer to Instrument-
specific information
on page 1.

Zero

5. Insert the vial into the 6. Push ZERO. The display 7. Add the contents of one 8. Put the cap on the vial.
16‑mm cell holder. shows 0.00 mg/L PO43–. PhosVer 3 Phosphate Shake for at least
Powder Pillow. 20 seconds. The powder will
not dissolve completely.

2 Phosphorus, Reactive, PhosVer 3 TNT Method (5.00 mg/L)


Read

9. Start the instrument 10. When the timer expires, 11. Insert the vial into the 12. Push READ. Results
timer. A 2‑minute reaction clean the vial. 16‑mm cell holder. show in mg/L PO43–.
time starts.
Measure the sample
between two and eight
minutes after adding the
PhosVer 3 reagent.

Interferences
Interfering substance Interference level
Aluminum More than 200 mg/L
Arsenate Interferes at any level
Chromium More than 100 mg/L
Copper More than 10 mg/L
Sulfide More than 6 mg/L. Remove sulfide interference as follows:

1. Measure 25 mL of sample into a 50-mL beaker.


2. Swirl continuously and add bromine water by drops until a permanent yellow color is seen.
3. Swirl continuously and add phenol solution by drops only until the yellow color is removed.
Use this treated sample in the test procedure.

Iron More than 100 mg/L


Nickel More than 300 mg/L
Highly buffered samples or Can prevent the correct pH adjustment (of the sample) by the reagents. Sample pretreatment
extreme sample pH may be necessary.
Silica More than 50 mg/L
Silicate More than 10 mg/L
Turbidity or color Samples with a high amount of turbidity can give inconsistent results. The acid in the reagents
can dissolve some of the suspended particles and variable desorption of orthophosphate from
the particles can occur.
Zinc More than 80 mg/L

Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• Phosphate 2-mL Ampule Standard, 50-mg/L as PO43–
• Ampule breaker
• Pipet, TenSette®, 0.1–1.0 mL and tips
• Mixing cylinders, 25-mL (3)

Phosphorus, Reactive, PhosVer 3 TNT Method (5.00 mg/L) 3


1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 25-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.

Standard solution method


Use the standard solution method to validate the test procedure, the reagents and the
instrument.
Items to collect:
• 3.0-mg/L phosphate standard solution

1. Use the test procedure to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are small
variations in the reagents or instruments.

Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% confidence interval) Sensitivity
Concentration change per 0.010 Abs change
535 3.00 mg/L PO43– 2.94–3.06 mg/L PO43– 0.06 mg/L PO43–

Summary of method
Orthophosphate reacts with molybdate in an acid medium to produce a mixed
phosphate/molybdate complex. Ascorbic acid then reduces the complex, giving an
intense molybdenum blue color. The measurement wavelength is 880 nm (DR 1900:
710 nm) for spectrophotometers or 610 nm for colorimeters.
Consumables and replacement items
Required reagents

Description Quantity/test Unit Item no.

Reactive Phosphorus Test ’N Tube Reagent Set — 50 tests 2742545


Includes:
PhosVer 3 Phosphate Reagent Powder Pillow, 10 mL 1 50/pkg 2106046
Reactive Phosphorus Test ‘N Tube Dilution Vials (not sold
1 50/pkg —
separately)

4 Phosphorus, Reactive, PhosVer 3 TNT Method (5.00 mg/L)


Required apparatus

Description Quantity/test Unit Item no.

Funnel, micro, poly 1 each 2584335


Pipet, TenSette, 1.0–10.0 mL 1 each 1970010
Pipet tips, for TenSette Pipet, 1.0–10.0 mL varies 50/pkg 2199796
Test tube rack 1 each 1864100

Recommended standards and apparatus

Description Unit Item no.

Phosphate Standard Solution, PourRite Ampule, 50-mg/L as PO43–, 2 mL 20/pkg 17120H


Phosphate Standard Solution, 50 mg/L as PO43– 500 mL 17149
Phosphate Standard Solution, 1 mg/L as PO4 3– 500 mL 256949
Phosphate Standard Solution, 3-mg/L as PO4 3– 946 mL 2059716
Drinking Water Standard, Mixed Parameter, Inorganic for F-, NO3–N, PO43–, SO42– 500 mL 2833049
Wastewater Effluent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO43–,
500 mL 2833249
COD, SO42–, TOC
Ampule Breaker, 2-mL PourRite Ampules each 2484600

Optional reagents and apparatus

Description Unit Item no.

Bromine Water, 30-g/L 29 mL 221120


Hydrochloric Acid Solution, 6.0 N (1:1) 500 mL 88449
Phenol Solution, 30-g/L 29 mL 211220
Pipet, TenSette, 0.1–1.0 mL each 1970001
Pipet tips for TenSette Pipet, 0.1–1.0 mL 50/pkg 2185696
Pipet tips for TenSette Pipet, 0.1–1.0 mL 1000/pkg 2185628
Filter paper, folded, 3–5-micron, 12.5-cm 100/pkg 69257
Funnel, poly, 65 mm each 108367
Thermometer, non-mercury, –10 to +225 °C each 2635700
Bottle, sampling, with cap, low density polyethylene, 250 mL 12/pkg 2087076
Beaker, 50 mL each 50041H

Optional standards

Description Unit Item no.

Ampule Breaker, 10-mL Voluette Ampules each 2196800


Phosphate Standard Solution, 10-mg/L as PO43– 946 mL 1420416
Phosphate Standard Solution, 15-mg/L as PO4 3– 100 mL 1424342
Phosphate Standard Solution, 50-mg/L, 10-mL Voluette Ampules 16/pkg 17110
Phosphate Standard Solution, 100-mg/L as PO4 100 mL 1436832
Phosphate Standard Solution, 10-mL ampule, 500 mg/L as PO4 16/pkg 1424210
Phosphate Standard Solution, 500-mg/L as PO4 100 mL 1424232

Phosphorus, Reactive, PhosVer 3 TNT Method (5.00 mg/L) 5


FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY
In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – [email protected] FAX: (970) 669-2932

© Hach Company/Hach Lange GmbH, 1989–2014, 2017. All rights reserved. 08/2017, Edition 10

You might also like