Doc316 53 01118
Doc316 53 01118
Doc316 53 01118
01118
(Orthophosphate)
USEPA 1, 2 PhosVer® 3 Method3 Method 8048
0.06 to 5.00 mg/L PO43– (0.02 to 1.60 mg/L P) Test ‘N Tube™ Vials
Scope and application: For drinking water, wastewater and seawater.
1 USEPA accepted for reporting wastewater analysis. Procedure is equivalent to USEPA and Standard Method 4500-P E for wastewater.
2 USEPA Accepted for reporting for drinking water analysis. Procedure is an acceptable version of EPA Method 365.1, approved at 40 CFR
part 141 NPDWR compliance monitoring.
3 Adapted from Standard Methods for the Examination of Water and Wastewater.
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows adapter and light shield requirements for the instruments that use them.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information for test tubes
Instrument Adapters Light shield
DR 6000, DR 5000 — —
DR 3900 — LZV849
DR 3800, DR 2800, DR 2700 — LZV646
DR 1900 9609900 (D1) —
DR 900 4846400 Cover supplied with the instrument
Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started.
For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water
in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results
automatically with the reagent blank adjust option.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
®
PhosVer 3 Reagent Powder Pillow 1
Reactive Phosphorus Test 'N Tube Vial 1
Funnel, micro 1
1
Items to collect (continued)
Description Quantity
Light shield or adapter (For information about sample cells, adapters or light shields, refer to
1
Instrument-specific information on page 1.)
®
Pipet, TenSette , 1.0- to 10.0-mL, with pipet tips 1
Test tube rack 1
Start
1. Start program 535 P 2. Add 5.0 mL of sample to 3. Put the cap on the vial. 4. Clean the vial.
React. PV TNT. For a Reactive Phosphorus Test Invert to mix.
information about sample 'N Tube Vial.
cells, adapters or light
shields, refer to Instrument-
specific information
on page 1.
Zero
5. Insert the vial into the 6. Push ZERO. The display 7. Add the contents of one 8. Put the cap on the vial.
16‑mm cell holder. shows 0.00 mg/L PO43–. PhosVer 3 Phosphate Shake for at least
Powder Pillow. 20 seconds. The powder will
not dissolve completely.
9. Start the instrument 10. When the timer expires, 11. Insert the vial into the 12. Push READ. Results
timer. A 2‑minute reaction clean the vial. 16‑mm cell holder. show in mg/L PO43–.
time starts.
Measure the sample
between two and eight
minutes after adding the
PhosVer 3 reagent.
Interferences
Interfering substance Interference level
Aluminum More than 200 mg/L
Arsenate Interferes at any level
Chromium More than 100 mg/L
Copper More than 10 mg/L
Sulfide More than 6 mg/L. Remove sulfide interference as follows:
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• Phosphate 2-mL Ampule Standard, 50-mg/L as PO43–
• Ampule breaker
• Pipet, TenSette®, 0.1–1.0 mL and tips
• Mixing cylinders, 25-mL (3)
1. Use the test procedure to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are small
variations in the reagents or instruments.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% confidence interval) Sensitivity
Concentration change per 0.010 Abs change
535 3.00 mg/L PO43– 2.94–3.06 mg/L PO43– 0.06 mg/L PO43–
Summary of method
Orthophosphate reacts with molybdate in an acid medium to produce a mixed
phosphate/molybdate complex. Ascorbic acid then reduces the complex, giving an
intense molybdenum blue color. The measurement wavelength is 880 nm (DR 1900:
710 nm) for spectrophotometers or 610 nm for colorimeters.
Consumables and replacement items
Required reagents
Optional standards
© Hach Company/Hach Lange GmbH, 1989–2014, 2017. All rights reserved. 08/2017, Edition 10