Manual Icanclave D Pro en

Download as pdf or txt
Download as pdf or txt
You are on page 1of 25

Instructions Manual

For Models: STE-18-D Pro


STE-23-D Pro
Thank you for choosing our steam sterilizer.

Prior to operating this instrument, please read the operations manual carefully and follow
all installation instructions.

Ningbo Ican Machines Co., Ltd. European Representative:


No. 77 Yunlin East Road, Gulin Town, Icanclave Europe S.L
Ningbo, China. Juan Ramon Jimenez 6
www.icanclave.com Quart de Poblet, Valencia, Spain.

Document: Version 00DP0000V2.9


Subject to technical changes

Save these instructions

2
3
Content
1. General 4 Scope of manual
4 Intended use
4 4 General safety instructions
4 Standards and directives
5 Symbols

2. Description of the sterilizer 5 Control panel


6 Sterilizer views
5 7 Technical specification
7 Packing content

3. Installation 7 General conditions


8 Power supply connection
7 8 Location requirements

4. Setup 8 Fill the distilled water tank


9 Preparation of the sterilization materials
8 10 Basic set
10 Advance set

5. Operation 13 Load the sterilization chamber


13 Select the program
13 13 Start the sterilization program
14 End of cycle
14 Manually interruption of the cycle
15 Delay start cycle
15 Test Program
16 Descale
16 Data
17 Print and save report (Optional)
18 Print Labels (Optional)

6. Maintenance 18 Clean the distilled water tank


18 Replacement of the bacteriological filter
18 19 Clean Chamber, trays and tray Rack
19 Door adjustment
19 Replacement of the door seal ring
19 The drain valve

7. Troubleshooting 20 Troubleshooting

20
8. Transportation and storage 21 Transportation and storage conditions

21
9. Safety devices 21 Safety device description

21
11. Appendix 22 Water properties/characteristics
23 Diagrams of the sterilization programs
22
1 General

4
Scope of Manual

This manual contains information concerning the installation, operation and


maintenance of the steam sterilizers. To ensure proper performance of the sterilizer, the
instructions given in this manual should be thoroughly understood and followed.
Keep the manual near the sterilizer in an accessible location for future reference.

Intended Use

The steam sterilizer described in this manual is intended for the sterilization in all
medical, dental, beauty, vet and tattoo fields of the following types of instrument loads:
solid, porous, hollow loads type A and hollow loads type B, un -wrapped, single wrapped
and double wrapped, liquid, that are compatible with steam sterilization.

General Safety Instructions

Read and understand this manual before attempting to install or operate th e


sterilizer.
Make sure that all the installation conditions are fully complied with.
Ensure that the voltage agrees with the supply voltage specified on the supply
on the type plate of the sterilizer.
This appliance must be grounded. Connect only to a pro perly grounded outlet.
Do not cover or block any openings on this appliance.
Use this appliance only for its intended use as described in this manual.
Do not exceed the maximum weight limit of the loads specified in this manual.
Do not operate this appliance if it has a damaged cord or plug if it is not working
properly or if it has been damaged or dropped.
Never put into the sterilizer inflammables or explosive products.
The sterilizer may not be operated in areas in which gas or any other explosive
volatile substance is present.
Installation and repair work should only be performed by authorized service
technicians. Work by unqualified persons could be dangerous and may void the
warranty.

Standards and directives

The steam sterilizers were designed and produced in conformity with the following
directives and standards:

Directives:

97/23/CE Pressure equipment.


93/42/EEC Medical devices (class II b).

Standards:

EN 13060 Relative to small steam sterilizers.


EN 61010-1 Safety regulations for laboratory devices - Part 1: General
regulations.
EN 61010-2-040 Safety regulations specific to sterilizers used in the processing
of medical material.
EN 61326-1 Electromagnetic compatibility regulations for laboratory devices.

5
Symbols

For safe operation, please pay close attention to the alert symbols below which can be
found on the sterilizer and throughout this manual.

Important information (Caution) Hot surface Ground


connection

2 Description of the sterilizer


Screen and Control panel

1 Time
2 Date
3 Process curve
4 Internal temperature
5 Internal pressure
6 Cycle time
7 Program information
8 Clean water tank
level
9 Used water tank
level
10 Printers
11 Door status
12 USB memory
13 Up button
14 Down button
15 Select button
16 Start button

Door closed / Tap to open the


The door is open
door

Fill the tank if it is flashing Min level of freshwater tank 0.5Liters

Drain the tank if it is flashing The capacity is 1.5 liters

Grey = No connected / no paper


1 Report printer
Green =Ready

Grey = No connected
USB memory
Green =Ready

6
Sterilizer views

1. Door 9. Touch pad 17. Power socket


2. Door seal ring 10. Printer 18. Steam vent
3. Door lock 11. USB port 19. Condenser vent
4. Door handle 12. Main switch 20. Bacterial filter
5. Chamber 13. Distilled water tank 21. Used water tank
vent vent(optional)
6. Door interlock 14. Used water tank vent 22. Rating plate
7. Distilled water tank 15. Safety valve
8. Screen 16. Circuit breaker

7
Technical specifications

Model STE-18-D Pro STE-23-D Pro


Chamber (mm) 250 x 350 250 x 450
Overall dimensions (mm) 490(W)*455(H)*600(D) 490(W)*455(H)*690(D)
Net Weight (kg) 50 53
Nominal power (VA) 1750 1750
Rated Voltage 220-240V / 110-120V ; 50/60 Hz
Sterilization temperatures 121°C/134°C
Capacity of the distilled 2.5 L (Water at level Max.) Approx.
water tank 0.5L (Water at level Min.) Approx.
Circuit breaker F16A /400 V
Operation temperature 5°C ~ 40°C
Operation relative humidity Max. 80%, non-condensing
Max. Noise level <70 dB
Atmospheric pressure 76 kPa ~ 106 kPa

Packing content

Item Accessories Quantity

1 Instruments tray 3

Instruments tray
2 1
rack

3 Draining hose 2

4 Door Seal 1

5 Tray handle 1
Door adjustment
6 1
wrench
7 Instructions Manual 1

3 Installation
General conditions

Position the device on a plane surface with a minimum capacity of 60 kgs.


The sterilizer should be placed on a level worktable.
Leave at least 10 cm between the device rear part and the wall. The clearance required
to open the door is 40cm.
Position the sterilizer at such a height as to make it possible for the operator to check the
whole sterilization chamber and carry out the normal cleaning operations.
The room where the device is installed must be sufficiently ventilated.

8
Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
Do not lean on the door when it is opened.
Do not place trays, papers, fluid containers or other objects on the sterilizer.
Power supply connection

Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to
connect the sterilizer to an appropriate power supply could result in damage to the unit and
electrical shock to personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only
used for the sterilizer is recommended. Never connect the device pin to reductions of any type.

Location requirements (mm)

4 Setup
Switch On and open the door to remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD
and shows the door position, water level, working program, date, time, etc.

Fill the distilled water tank

Manual water filling

When the level of distilled water reaches a minimum level, the distilled water tank icon will flash
and beep three times.
Press the button on the tank lid and open it to the maximum position.
Fill it carefully with distilled water.
If exceeds the maximum level, an alarm will sound, and the distilled water tank icon will blink.

9
Drain the distilled water tank

Attach the drain hose on a fitting connection located on the front of the sterilizer. Pull the
connector to start the draining.
Attention: The capacity of the distilled water tank is approximately 2.5 liters.

Drain the used water tank

Attach the drain hose on a quick fitting connection located inside the service door at the left.
Attention: The capacity of the used water tank is approximately 2 liters.

Preparation of sterilization materials

For the most effective sterilization and to preserve the sample, please follow below:

Clean instruments immediately after use.


Treat the instruments by ultrasound cleaner.
Residual chemicals left over after cleaning and disinfecting process may damage and
corrode parts of the sterilizer, always rinse off the instruments using distilled water.
Follow instrument manufacturer’s guidelines and recommendations for handing and
cleaning instruments prior to sterilization.
Check the manufacturer’s instructions as to proper procedure for sterilizing of each item.
Arrange the samples of different materials on different trays or with at least 3cm of space
between them.
Clean and dry instruments thoroughly before placing them into tray.
Always insert a sterilization paper or cloth between the tray and sample to avoid direct
contact.
Arrange the containers (glasses, cups, test-tubes, etc.) on one side or inverted position,
avoiding possible water stagnation.
Don’t stack the trays one above the other or put them in direct contact with the walls of
the sterilization chamber.
Always use the instrument tray handle.
Wrap the samples one by one or, if more tools have to be set in the same bag, verify that
these are made of the same material.
Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the
sterilizer.
Don’t overload the trays over the stated limit (see appendix 2).

10
Basic Set

The “Basic Set” menu permits to set the following


options:
*Date *Time *Language
Select the “Basic Set” from the main menu by
pressing button.
Select the item by pressing button. The unit
you selected will be lighted.
Adjust the value by pressing buttons. Press
button to select the next item.
Press button to save and back.

Note: The Counter (cycle No) cannot be set by the


operator.

Advance Set

The “Setup” menu permits to set the following


options:
*Parameter *Unit *Preheat *Expire date (labels)
*Water quality (sensor) *Preheat *Factory reset
and see the information of: *Last error
Select “Setup” from the main menu by pressing
button.
Input the password digit to digit by pressing
and button to go next.
Password: 1111

Parameter

The “Parameter” menu permits to set the following


options:
*Holding time *Dry time

Select “Parameter” from the menu by pressing


button.
Select the program by pressing then press
button.
Select the parameter by pressing button. Adjust
it by pressing .
Press button to save and back.

Unit

11
Select “Unit” from the menu by pressing button.
Select the parameter by pressing button. Adjust
it by pressing .
Press button to save and back.

Preheat

When this mode is activated, the chamber and


steam generator start to warm until it reaches the
minimum temperature to begin a sterilization cycle,
this helps to reduce the total cycle time and the
drying efficiency. The “Preheat” mode will be
deactivated after one hour of inactivity.

Select “Preheat” from the menu by pressing


button.
Adjust it by pressing .
Press button to save and back.

Expiry date (Optional)

To modify the expiration date of the labels, select


“Expire date” from the menu by pressing button.
Adjust it by pressing .
Press button to save and back.

12
Water quality (optional)

If your sterilizer is equipped with a water


quality sensor and you want to deactivate it,
select “Water quality” from the menu by pressing
button.
Adjust it by pressing .
Press button to save and back.

Last Error

In order to help the technical assistance


process, the most relevant information
corresponding to the last error can be
displayed on the screen.

Select “Last error” from the menu by pressing


button.
Press button to back.

Factory reset

Select “Factory reset”


Press “Yes” to reset all settings
Press button to back.

5 Operation

13
Load

After selecting program, the materials to be sterilized can now be placed


on the tray and placed inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.

Select the program

Enter to the main menu by pressing button,


Select “Program”. Select the program by
pressing then press to confirm program,
in the screen will appear the information of
selected program as the temperature and
sterilization time (holding time), also th e date,
time, current pressure and current
temperature.

Start the sterilization program

Press to start the cycle. The stage, conditions


and the status of the cycle will appear on the
display. The sterilizers will perform the program
automatically. (see appendix 2).

14
End of the cycle

Once the cycle is completed, “End” will appear at the end of the graphic,
the printer will print out and the digital report saved in the USB memory is
these are connected.

Caution: Always use the tray handle to load or unload the tray
into the sterilizer. Failure to do so can result in burning.

Manually interruption of the cycle

To interrupt a started cycle prematurely, hold for 3 seconds.


If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer
may be considered sterile and considering that the cycle has been interrupted during the drying
phase the materials and instruments inside the chamber may be wet.
Note: If the cycle is manually interrupted before it reaches the drying phase, the items inside the
sterilizer must be considered not sterile. N20 will appear on the screen. (see Error code
description).

Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.

If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer
may be considered sterile.
Note: Considering that the cycle has been interrupted during the drying phase the materials and
instruments inside the chamber may be wet.

If the cycle is manually interrupted before it reaches the drying phase, the items inside the
sterilizer must be considered not sterile.

Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.

Delay start cycle

15
Select “Delay start”. Select the program by pressing then press to confirm.
Select the parameter by pressing button. Adjust it by pressing .
Press button to save and back.

Test Programs

Helix Test

Put the Helix test device into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu,
select “Helix test”; in the screen will appear the information as temperature and
sterilization time (holding time), also the date, time, current pressure and current
temperature.

Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).

After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.

B&D test

Put the B&D test package into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu,
select B&D test, in the screen will appear the information as the temperature and
sterilization time (holding time), also the date, time, current pressure and current
temperature.

Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).

After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.

Vacuum Test

16
Select “Program” from the main menu by
pressing then to enter in the menu,
select “Vacuum test”.

After closing the door, press to start the cycle.


The stage, conditions and the status of the cycle
will appear on the display. The sterilizers will
perform the program automatically. (see appendix
2).

In compliance with EN 13060, the test requires that


the air leakage rate less than or equal to 0.13
kPa/min during 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will
show void. That means the result of the test is fail. You need run the vacuum test again after the
chamber has cooled down.

Descale

Mix the descaling powder with distilled water to make descaler according with the
instruction of the descaling powder manufacturer.
From the Programs menu select descale.
After closing the door, press to start the cycle. The stage, conditions and the status of the
cycle will appear on the display. The sterilizers will perform the program automatically. (see
appendix 2).
Follow the instructions in the screen.

Remove all instruments inside the chamber a nd close the door.


Empty the freshwater tank and fill with descaler which prepared as above, to 2/3 of the water
tank.
Press , start descaling, it will take around 25 minutes.
After descaling, empty the freshwater tank and fill with distilled water.
Press , start flushing, it will take around 8 minutes.
After flushing, the cycle “Descale” is end.

Data

The internal memory will store the information of the last 9999 cycles.

17
Print report (Optional)

Internal Memory
In this menu you can get the information of all the cycles stored in the internal memory of the
sterilizer.
Select “Report” from the main menu and press button, you will see the list of records.
Select the record by pressing button.
Press button to print and save the report.
Press button to exit.
E.g.
If the serial number of the steam sterilizer is
E00001 and the cycle number is 0012.

The file name of the file will be 01001200.log


The first two numbers represent machine
number.
The middle four numbers represent cycle
number.
The last two numbers represent error code.
00: no error; 01: error E01

Sample of a printer report

==============================
==
Program: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying Time: 08Min
Holding Time: 4.0Min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -075.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -075.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
Cycle No.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v 2B00V2.5
==============================
==

18
Print labels (Optional)

Select “labels”, it will show cycle numbers.


Select the cycle number, press button to
print the labels.

6 Maintenance
To ensure proper operation and maximum steam sterilizer life, carefully follow all
recommendations for periodic maintenance.
One of the most important steps you can take to prevent problems with your sterilizer is to use
only distilled water.

Frequency Number of cycles Maintenance operation


Clean the door seal
Clean the filter inside the chamber and in the
Monthly 50 clean water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months 200 Clean the distilled water tank
Every year 800 Replace the door seal

Select Settings from the main menu and tap “Maintenance”, to enter in the menu. Select
the item to check the information.

Clean the distilled water tank

Disconnect the main cable.


Drain the tank completely using the drain connector at the back of the sterilizer and leave it
connected into the connector in an open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas difficult to reach
using and a distilled water.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water,
and put it back in to the position.

Replacement of the bacteriological filter.

The bacteriological filter is in the back of the sterilizer. Unscrew


the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.

19
Clean chamber, door seal ring, trays, and tray rack.

Remove the trays and tray rack from the chamber.


Clean trays, rack and inside of the chamber with mild soap.
Rinse the trays, rack and inside of the chamber with a smooth cloth and distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber.
Failure to comply may cause damage to the chamber and/or other components.

Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the
surface.

Door adjustment

Under normal circumstances, the chamber door does not require


adjustments. However, if the seal fails (resulting in steam leaking from the
front of the chamber), you may adjust it.

Open the door, insert your finger into the bottom of the door, and pull the
ring to unlock the mechanism. Turn the lid counterclockwise to tighten it. If it
is too tight, you may also turn the lid clockwise to loosen it.

Replacement of the door seal ring

Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:

Press in the top and bottom Press in the left and right Press the remaining sections
of the door seal. sides of the door seal. of the seal.

Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.

The drain valves

To drain the water from the used water tank or to fill the tank from an external container, insert the
connector attached to the silicone tube until a click is heard. To disconnect the connector, press
the metal button.

20
1. Press the included 2. Turn the drain valve 3. Push and turn the drain
hose on to the drain valve clockwise to drain the valve anticlockwise to be
firmly. tank. back after draining the tank.

7 Troubleshooting

Code Description Proposed solution


Steam generator temperature Power off & run a new cycle
E1
sensor error. Contact your supplier if error persists.
Power off & run a new cycle
E2 Inner temperature sensor error.
Contact your supplier if error persists.
Temperature sensor of the Carefully ensure that the chamber wall is heated
E3
chamber wall error. and contact your supplier.
Power off & run a new cycle
E5 Fail to release the pressure.
Contact your supplier if error persists.
Make sure you had closed the door properly.
E6 Door lock problem during the cycle.
check the door switch.
Error between temperature and Power off & run a new cycle
E7
pressure correlation. Contact your supplier if error persists.
Error between temperature and Power off & run a new cycle
E8
pressure correlation. Contact your supplier if error persists.
Ensure the distilled tank isn't empty. Check the
E9 Failure to hold temperature. inner temperature sensor. Check somewhere for
leaking.
The electromagnet of locking system doesn’t
The door locking system doesn’t
E10 work.
work.
The switch of locking system doesn’t work.
Failure to preheat the steam Power off & run a new cycle
E11
generator. Contact your supplier if error persists.
Power off & run a new cycle
E12 Failure to preheat the chamber.
Contact your supplier if error persists.
Power off & run a new cycle
E13 Vacuum failed.
Contact your supplier if error persists.
Power off & run a new cycle
E15 Inner temperature sensor error #2*
Contact your supplier if error persists.
Replace the air filter
E16 Pressure error Power off & run a new cycle
Contact your supplier if error persists.
N20 Program manually interrupted Reset the error from main screen.
Somewhere is leaking. Check the door seal.
E22 Vacuum test failure
Or contact your supplier if error persists.
The temperature of the chamber is high.
N23 Result of vacuum test is void
Try again after the chamber has cooled down.
It takes too long time to enter the Check somewhere leaking.
E24
next status. Or contact your supplier if error persists.
Switch off. Then switch on after the chamber cool
N27 The vacuum test fails.
down and try again.

21
Power off and contact your supplier if error
E28 The pressure is overload.
persists.
Power off & run a new cycle
E30 Vacuum failed.
Contact your supplier if error persists.
Power off & run a new cycle
E31 Inner temperature sensor error #2*
Contact your supplier if error persists.

8 Transportation and storage

Switch off the sterilizer before transportation or storage.


Pull out the plug. Let the machine cool down.
Drain the distilled water tank and the used water tank.

Conditions for transport and storage


Temperature: -20°C ~ +50°C
Relative humidity: ≤ 85%
Atmospheric pressure: 50kPa~ 106kPa.

9 Safety devices

1. Main breaker: Protection of the instrument against possible failures of the heating resistor.
Action: Interruption of the electric power supply.

2.Thermal cutouts on the main transformer winding: protection against possible short circuit and
main transformer primary winding overheating
Action: Temporary interruption of winding.

3.Safety valve: Protection against possible sterilization chamber over-pressure.


Action: Release of the steam and restoration of the safety pressure.

4.Safety micro-switch for the door status: Comparison for the correct closing position of the door.
Action: Signal of the wrong position of the door

5.Thermostat on chamber heating resistors: Protection for possible overheating of the chamber
heating resistors.
Action: Interruption of the power supply of the chamber resistors.

6.Thermostat on steam generator heating resistors: Protection for possible overheating of the
steam generator heating resistors.
Action: Interruption of the power supply of the steam generator resistors.

7.Door safety lock: Protection against accidental opening of the door.


Action: Impediment of the accidental opening if the door during the program.

8.Self-leveling hydraulic system: Hydraulic system for the natural pressure leveling in case of
manual cycle interruption, alarm or blackout.
Action: Automatic restoration of the atmospheric pressure inside the chamber.

Appendix 1

22
Water properties / Characteristics

Description Feed water Condensate

Evaporate residue ≤ 10mg/ I ≤ 1.0mg/kg

Silicium oxide sio2 ≤ 1mg/ I ≤ 1.0mg/kg

Iron ≤ 0.2mg/ I ≤ 0.1mg/kg

Cadmiun ≤ 0.005mg/ I ≤ 0.05mg/kg

Lead ≤ 0.05mg/ I ≤ 0.1mg/kg

Rest of heavy metals ≤ 0.1mg/ I ≤ 0.1mg/kg

Chloride ≤ 2mg/ I ≤ 0.1mg/kg

Phosphates ≤ 0.5 mg/ I ≤ 0.1mg/kg

Conductivity ≤ 15μs /cm ≤ 3 μs /cm

PH Value 5 – 7.5 5-7

Appearance Colorless, clean Colorless, clean

Hardness 0.02 mmol/ I 0.02 mmol/ I

23
Appendix 2

Diagrams of the sterilization programs STE-18-D Pro


Temperature (°C)

Max load per tray


Total time (min)
Pressure (kPa)

Max load (kg)


Holding time
(min)

(kg)
Programs Type

Solid 134 134 210 3.5 12-30 Unwrapped solid material 4.50 1.20

Wrapped Unwrapped solid material 4.50 1.20


134 210 4 30-49 Single-wrapped solid or hollow
134 material 4.00 1.10

Wrapped Unwrapped solid material 4.50 1.20


121 110 20 38~54 Single-wrapped solid or hollow
121 material 4.00 1.10

Unwrapped porous material 1.00 0.30

Single-wrapped porous
0.80 0.25
material

Prion 134 134 210 18 37-52 Dual-wrapped porous material 0.60 0.20

Single-wrapped hollow
3.50 1.00
material
Dual-wrapped solid and hollow
material 1.50 0.50
Drying
(optional) ― ― ― 1-20 ― ― ―

B&D test 134 210 3.5 28-35 ― ― ―

Helix test 134 210 3.5 28-35 ― ― ―

Vacuum test ― ― ― 20-25 ― ― ―

The time required for sterilizer to be ready for routine use after the power is switched is less than
5 minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C

24
Appendix 3

Temperature (°C) Diagrams of the sterilization programs STE-23-D Pro

Max load per tray


Total time (min)
Pressure (kPa)

Max load (kg)


Holding time

(min)

(kg)
Programs Type

Solid 134 134 210 4 15-30 Unwrapped solid material 5.00 1.50

Wrapped Unwrapped solid material 5.00 1.50


134 210 4 38-57 Single-wrapped solid or hollow
134 material 4.50 1.20

Wrapped Unwrapped solid material 5.00 1.50


121 110 20 46~62 Single-wrapped solid or hollow
121 material 4.50 1.20

Unwrapped porous material 1.25 0.40

Single-wrapped porous
1.10 0.30
material

Prion134 134 210 18 45-60 Dual-wrapped porous material 0.75 0.25

Single-wrapped hollow
4.00 1.25
material
Dual-wrapped solid and hollow
material 2.00 0.60
Drying
(optional) ― ― ― 1-20 ― ― ―

B&D test 134 210 3.5 28-35 ― ― ―

Helix test 134 210 3.5 28-35 ― ― ―

Vacuum test ― ― ― 20-25 ― ― ―

The time required for sterilizer to be ready for routine use after the power is switched is less than
5 minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C

25

You might also like