C2000 Owners Manual
C2000 Owners Manual
C2000 Owners Manual
owner’s manual
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 3
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 4
TABLE OF CONTENTS
2 C-2000 Features
3 Key Functions
3 Check Calibration
4 Set the Scale
4 To Change the Set-Point
5 Information About Your Readings
5 To Check the Accumulated Readings
6 To Reset Meter
6 Applications
9 Temperature Correction
9 Care Of Your Meter
10 Service for Your Meter
11 Warranty
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 5
DELMHORST C-2000
#
! Set-Point
Key
Read Key
@
Calibration
Check Key
$
Scale
Key
C-2000 FEATURES
hDesigned primarily to check the moisture content (%MC)
in lint cotton and seed cotton. Versions 2.0 and higher
include scales for viscose rayon and wool.
hResistance technology recognized around the world as the
most accurate method for measuring moisture.
h Proven microcontroller circuit
h Averages up to 100 accumulated readings
h MC ranges - lint cotton 4%-16%, seed cotton 6%-20%,
viscose rayon 6%-25%, wool 12%-30%.
h Includes (1) 9-volt battery
h One-year warranty
2
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 6
KEY FUNCTIONS
! Read:
Reads the % MC value.
@ Check:
Checks the meter calibration. Displays the
average of up to 100 accumulated readings;
displays the highest stored reading; erases the
readings from memory.
# Set-Point:
Displays the current set-point. Also acts as an
arrow (scroll) key to increase the set-point
value in 1% increments. A buzzer will alert
you if the meter reads higher than the selected
%MC value.
$ Scale:
Displays the meter scale as #1 (Lint cotton),
#2 (Seed cotton), #3 (Viscose Rayon), #4
(Wool). Also acts as an arrow key to decrease
the set-point value.
CHECK CALIBRATION
hPress the calibration check key @ and read key ! simul-
taneously. Meter is in calibration if it displays 10.8% ±
0.2, regardless of the scale setting.
When checking calibration, there is no need to disconnect the
electrode.
If you check the calibration and the display does not read
10.8, it is likely an indication of a low battery. If this occurs,
change the battery immediately. Continued use with a low
battery may cause the meter to go out of calibration.If you
have a fresh battery and the instrument still does not indi-
cate an acceptable calibration, return it to DELMHORST for
service. See “Service for Your Meter” section.
Note: When the battery is removed and then reconnected,
the meter displays its software version for one second and
then turns itself off. After replacing the battery, you must
reset the meter as described in the “Resetting the Meter”
section.
3
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 7
TO CHECK THE
ACCUMULATED READINGS
This feature displays the total number of all accumulated read-
ings for the given material you have chosen, the average of
those readings, and the highest stored reading.
hTo view the readings, press and release the calibration check
key @. First the meter displays the number of accumulated
readings for one second, then the average of those readings
for two seconds. Then it displays the highest stored reading
for two seconds. The total “cycle” time is five seconds.
5
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 9
TO RESET METER
hPress and release the calibration check key @.
hWithin one second, press and hold the scale key @. The
meter will display a reset sequence as follows: “110”,
“8”, “10.8”. The last number, “10.8” is a calibration
check.
hResetting the meter clears the memory and restores the
default settings of scale #1, set-point 10.
APPLICATIONS
Testing Lint Cotton and
Seed Cotton
hAttach the #52-E/C Sample Cup Electrode to the connector
on top of the meter.
hSet the scale for (#1) Lint cotton or (#2) Seed cotton and
select the set point.
hPlace the material in the sample cup. Press the cotton
firmly into the cup with your finger. The sample in the cup
should be overflowing during the test, even while you are
pressing it with your finger.
hPress the read key ! and read the moisture content on the
display. The meter displays the %MC for two seconds.
Readings below 4% lint cotton will be displayed as a negative
number. Readings above 16% lint will be displayed as a
flashing “16.2%”. Both the underrange and overrange read-
ings should be disregarded. They will not be added to the
accumulated readings or used in calculation of the average
or highest reading.
6
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 10
7
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 11
Testing Yarn
hConnect the #37-E/C multi-pin electrode to the connector
on top of the meter.
hSet the scale to #1 Lint cotton and select a set point.
hInsert the electrode pins into the yarn and press the read
key ! .
8
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 12
Temperature Correction
The basic calibration assumes the cotton temperature to be
70°F. For best accuracy, apply a temperature correction if
the cotton temperature is outside the range of 60°F to 80°F.
Make a correction of approximately 1.0% for every 20° F.
The meter will read higher than the actual moisture content
as cotton temperature increases, and will read lower as cot-
ton temperature decreases.
Example:
At 70°F, a reading of 7% indicates actual %MC of 7%. At
varied temperatures, meter readings will vary as follows:
Cotton Temperature Meter Reading Actual %MC
110°F 9.0% 7%
50°F 6.0% 7%
9
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 13
10
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 14
WARRANTY
Delmhorst Instrument Co., referred to hereafter as
Delmhorst, guarantees its C-2000 meter for one year from
date of purchase and any optional electrodes against defects
in material or workmanship for 90 days. If, within the war-
ranty period of the C-2000, you find any defect in material
or workmanship return the meter following the instructions
in the “Service for Your Meter” section. This limited war-
ranty does not cover abuse, alteration, misuse, damage dur-
ing shipment, improper service, unauthorized or unreasonable
use of the meter or electrodes. This warranty does not cover
batteries, pin assemblies, or pins. If the meter or any option-
al electrodes have been tampered with, the warranty shall be
void. At our option we may replace or repair the meter.
11
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 15
NOTES
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 16
NOTES
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 1
(877)-DELMHORST
(3 3 5 - 6 4 6 7 )
www.delmhorst.com
[email protected]