0% found this document useful (0 votes)
90 views7 pages

Comprehensive Infant Care Centre

The document summarizes the features of the CIC 101 infant care system from Phoenix, which is designed to provide sophisticated care for preterm infants. It has dual temperature monitoring, electronic CPAP and suction, and a pulse oximeter to monitor vital signs. Additional features include a tilting bed, examination lights, alarms and a touchscreen display. The system is designed to be user-friendly for caregivers while keeping infants warm, comfortable and healthy.

Uploaded by

Vignesh Guru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
90 views7 pages

Comprehensive Infant Care Centre

The document summarizes the features of the CIC 101 infant care system from Phoenix, which is designed to provide sophisticated care for preterm infants. It has dual temperature monitoring, electronic CPAP and suction, and a pulse oximeter to monitor vital signs. Additional features include a tilting bed, examination lights, alarms and a touchscreen display. The system is designed to be user-friendly for caregivers while keeping infants warm, comfortable and healthy.

Uploaded by

Vignesh Guru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

COMPREHENSIVE INFANT CARE CENTRE

CIC 101
The CIC 101 from PHOENIX is simply the most sophisticated neonatal open care system available today. it has
been designed to meet the requirements involved in in providing care for the preterm infant and to meet the
critical situations that may arise in providing such care. CIC 101 is remarkably convenient and user friendly,
the space around the infant is organized efficiently for the convenience of care-givers, the mother and visitors.

Dual temperature monitoring for Electronic blending resuscitation Pulse oximeter


accurate thermal environment or CPAP + suction unit

The CIC helps keep infants warm, The ow-based operation of the The ow-based operation of the
comfortable and healthy by continuously CPAP unit is infant- friendly and CPAP unit is infant- friendly and
and accurately maintaining a constant eases breathing. During an eases breathing. During an
temperature. emergency, the infant can be emergency, the infant can be
resuscitated. resuscitated.

Weighing scale Foot Operated Tilting and


Height adjustments rotation of bed

The built-in weighing scale allows The height of the bed (along with the The gas-asssisted mechanism tilts
hassle-free measurement of the infant’s control display panel and the warmer the bed 12° smoothly to both
weight without moving or disturbing the unit) may be adjusted to the Tr e n d e l e n b u r g a n d r e v e r s e
infant. convenience of the caregiver. Foot Trendelenburg positions when
switches are provided on both sides required. The bed can also be
for easy access. rotated 60° degree for easy access to
the infant.

Non touch Quick infant access Intelligent


examination lamp & Minimal risk infection safety alarm
The two soft-start examination lamps The oval-shaped, single-piece, The visual and audible alarms alert
help examine with minimal disturbance crevice-free bed is easy to clean. the caregiver if the set measured
to the sleeping infant. The lamps can be Thus bacterial growth is reduced, the value crosses the set limits. Apilot
controlled using an innovative non-touch oval shape gives easy access to the lamp attached to the warmer gives a
system, ensuring reduced risk or spread infant from anywhere around. visual indication.
of infections.

Removable X-ray sheet 7 Inch Display


X-ray trays can be slid in and out of the 7-inch touch screen colour display
X-ray compartment without disturbing clearly displays the set and measured
the infant while taking X-rays. vital parameters in a single screen.
CDSCO Registration No. : 000379 / 2020
TECHNICAL SPECIFICATIONS

Self-checking microcontroller based baby control system

Physical dimensions:

Height+ 1840 mm - 2140 mm


Length 1425 mm
Width 940 mm

LCD graphical display with trend facilities 7" colour TFT

Skin Temperature
Heater set value
Baby Weight
Heart rate (BPM)
SpO2

Mechanical
Castors 5" with brakes
Infant bed Oval shaped bed for greater accessibility
Bed tilt ± 12°
Bed rotation ± 60°
Mattress-padding foam density 25 kg/m3.
X-ray provision Provision to take X-rays with the help of portable machines
without shifting the infant.
Mattress cover Memory Foam Mattress, waterproof, washable.

Permits rotation of the heater without disturbing the infant during radiography

Electrical power requirements


Operating voltage 230 V AC ~
Power rating 750W
Fuse rating 4A
Frequency 50 Hz
Leakage current 0.5 mA
Observation lamp with dimmer controller 12v, 20w (touch free)

Skin temperature probe


Range 6°C to 50° C
Accuracy ±0.2° C
Resolution 0.1° C
Probe interchangeability ±0.2° C

Servo Mode
Set temperature range (Servo) 32° C to 37° C,
Override able up to 38° C

Heater
Single element 600 W quartz infrared heating element placed in a parabolic reflector
Heater Life Time 12months
Manual Mode
Heater output control range (Manual) 0% to 100% (5% increments)

Pre-Warm Mode
Heater output control 60% after 3 minutes
30% after 12 minutes
+ Height adjustment is optional
Audio and visual alarms

· High / Low infant temperature


· Probe failure
· Heater failure
· Power failure ( up to 10 minutes)
· SpO2, BPM (High / Low)
· System failure with automatic heater cut off
· Power down alarm in manual mode (after 10 minutes, heater power reduce to 60%)
· Manual mode alert alarm (every 15 minutes)
· Over temperature
· High, low oxygen concentration (±5% from the set value) for CPAP function
· High, low pressure (±2 cm H2O from the set value) for CPAP function

Visual alarms (Screen message)


·Peripheral probe failure
·Skin temperature difference (1 deg. C between two probes)

Battery backup
For nCPAP and for temperature monitoring1 hour (max)

APGAR Timer (Optional) Works in Pre-Warm mode only


Beep sound at 1, 3, 5 & 10th minutes
Stop at 20 minutes
Integrated timer 1 to 59 min, with count-up and count-down feature

TIMER (Optional) Works in Manual and Baby servo mode


UP / Down

Internal Weighing Scale (Optional)


Accuracy +/-10 g
Load capacity 10 kg (max)

RESUSCITATION: (Optional)
Breathing aid
Airway ventilation outlet (hand operated port)
Oxygen flow rate control 0 _ 8 Lit/min.
Breathing rate control Manual
Maximum airway pressure 35 cm H2O
Adjustable airway pressure 10 _ 35 mm H2O

Suction control
Maximum suction 250 mm Hg
Vacuum gauge range 0 - 760 mm Hg
(Electrical option on request)

Auxiliary oxygen outlet


Oxygen flow control 0 _ 8 Lit/min
Maximum airway pressure 55 cm H2O

MASIMO PULSE OXIMETER (Optional)


Oxygen saturation range 0% to 100%
Pulse rate range 25 to 240 BPM
Perfusion index 0.00% to 20.00%
CPAP (Optional)

Oxygen flow rate control 0- 8 Lit/min.


O2 concentration 21% to 100 %
Oxygen inlet pressure 3 bar to 6 bar
Air inlet pressure 3 bar to 6 bar
Suction pressure 0 to 250 mm Hg
Manometer (digital) 0 to 99 cm H2O
Bar display digital 0 to 20 cm H2O
CPAP pressure 0 to 20 cm H2O

Air & oxygen blender

The patient should always be open to the atmosphere, and the resistance (PEEP) should be generated by flow.

Accessories (Optional)

· Supplied with bonnet and cap for correct positioning of nasal prong
· Supplied with prongs of 3 sizes (small, medium and large)
· Provided with PEEP generator near baby to ease respiratory effort by the baby and reduce dead space.

Standard accessories

· Height-adjustable IV stand
· Mayo trays
· IV pipe for fixing accessories

Coating Epoxy/powder coating for scratch resistance and rust protection

Electrical protection
Type of protection against electric shock Class 1
Degree of protection against electric shock Type B
Mode of operation Continuous
Protection against hazards of explosion Not-protected

Manufactured to standards IEC 60601 - 1, IEC 60601 - 1- 2, IEC 60601 - 2- 21

Environmental specifications
Operation temperature range 18°C to 30°C
Storage temperature range 10°C to 60°C
Operating humidity range 0% to 100% RH

CDSCO Registration No. : 000379 / 2020