Kiwi Store SAB V001 Manual
Kiwi Store SAB V001 Manual
Kiwi Store SAB V001 Manual
for
Liconic
Kiwi
Stores
SAB_V001_MANUAL
Important Notice
2016 LiCONiC AG Liechtenstein, all rights reserved.
Contact Address
LiCONiC AG
Industriestrasse 8
FL - 9493 Mauren
Principality of Liechtenstein
Before operating the instrument, the user must read and understand this
manual.
LiCONiC regularly offers Operator Training Courses. We highly recommend the attendance
of such a course prior to working with a LiCONiC Instrument.
For future reference, please read the instrument Serial No. on the rear of your
instrument and write it in the provided space.
1.6 Definitions
Operator: Any person who uses the equipment for its intended
purpose.
Make sure that all Safety Instructions in this publication are strictly followed.
WARNING !
CAUTION !
NOTE !
DANGER !
Prior to any first time application, test runs should be made with a neutral liquid to
allow optimization of all liquid handling parameters.
Handling and disposing of waste has to be in accordance with all local, state and
federal environmental, health and safety laws and regulations.
Inside the chamber temperatures below zero degree Celsius are present.
Contact to any surfaces with bear skin should be avoided. Protective clothing has
to be worn when manual operations are performed inside the chamber.
WARNING !
Inert gases are asphyxiates and will displace oxygen to produce localized
oxygen deficient atmospheres
Inert gases are odorless, colorless and tasteless
Inert gases give no warning and may cause death in few seconds
The Exhaust of the CO2 or LN2 backup cooling option needs to be lead out
of the building otherwise danger of suffocation is immanent
In their liquid state inert gases are extremely cold fluids which can cause
serious burns when in contact with the skin
A cryogenic spill often gives a low lying cloud of vapor that creeps along the
ground
Due to very specific local rules and regulations in different countries it is the
customers / owners responsibility to install suitable oxygen (O2) and carbon
dioxide (CO2) monitor systems which trigger alarms to protect operators in
and around the SAB store.
It is not allowed to work inside the system alone. A second person has to be
outside surveying the work inside and to be in permanent contact with a radio
with the person inside.
1.11.4 Modifications
Modifications shall only be carried out by an authorized Engineer. LiCONiC AG
will not accept responsibility for any claim resulting from unauthorized
modification or alteration.
Storage
Dryer
Antechamber; User
Access to load and
unload material
Main Controller
cabinet
Cooling storage
Cooling antechamber
The Kiwi Stores SAB are Stand Alone for manual semi-automated operation.
These capacities are an excerpt of the most commonly requested sizes. Further
Storage sizes are available on request. All Kiwi Stores can handle and store a
wide range of types of plates and racks with SBS standard format. The
corresponding capacity is only depending of the height of the racks and if the
racks / plates are stored in separate levels or stacked on each other. Special
formats such as cryo boxes and larger sizes are available upon request.
For further systems dimensions please contact LiCONiC AG. These Dimensions
are examples and are not binding.
All Kiwi Store SAB units have compact dimensions. The Kiwi Store SAB has a
user front door which allows comfortable and easy access to the ante chamber.
The ante chamber is the operation space for the manual transfer of racks and
plates from any additional equipment or transport box into the actual storage
area. Exchangeable cassettes make the use of the Kiwi Store SAB extremely
flexible. Cassettes are available for all common plate types and Cryo Boxes.
Storage
Dryer
Antechamber; User
Access to load and
unload material
Main Controller
cabinet
Cooling storage
Cooling antechamber
4.1 Overview
The KiWi SAB DF Stores are designated for the use as compound storage,
whereas the ULT version is mainly for Biobanking use.
4.2 Cooling
Within all units the air in the DF (-20°C) part, is actively dried by forcing the air
through an evaporator. This allows cooling and actively drying the air. Since the
evaporator gets covered with ice after a certain time these units have an
automatic defrost function. The collected water will be exhausted trough a drain.
For DF Storage units this does apply for the complete storage area. In case of
the ULT systems the active drying is done in the user access chamber, the ante
chamber and the service area.
4.3 Temperature
All units are equipped with a Temperature-Sensor. This allows displaying the
temperature inside the storage chamber. A very high quality sensor is used for
this purpose. Quality and accuracy of the sensor is maintained by the Swiss
Federal Office of Metrology and the Swiss Accreditation Service (see
specifications in next chapter).
Due to the limited temperature range this sensor is used in the ULT storage
systems to monitor and control the climate in the service area and the ante
chamber. To monitor all other temperature, such as parameters of cooling and
storage chamber temperature, thermocouples are used.
The Humidity sensors used in the Kiwi Store series are suitable for extremely
high humidity as well as for extreme dryness. They incorporate a humidity and a
temperature sensor allowing a relative humidity (rH) readout.
Specifications
Measuring Range 0..100 % RH
Accuracy@23°C ±1.5 % RH
±1.3 °K Temp.
Resolution 12 Bit / 0.02%
12 Bit / 0.1°C
Sensor PT100 Temperature
C94 Humidity
Adjustment - E2-PROM
5.1 Overview
The Kiwi Store auxiliary refrigeration units mainly cover the case of a failure of
the main cooling unit whereas the gas backup solutions are made to cover power
outages.
This option includes the tubing and the control. This backup solution is fully
monitored to ensure the temperature to stay at –20°C. Also an exhaust is part of
this option. For ULT unit an adapted regulation is in use to ensure to stay below -
70°C.
If the CO2 backup cooling is coming into action large amounts of CO2 are
released through the options exhaust. Therefore the exhaust needs to be lead
out of the building to not endanger personnel working in the area of the system.
WARNING !
Inert gases are asphyxiates and will displace oxygen to produce localized
oxygen deficient atmospheres
Inert gases are odorless, colorless and tasteless
Inert gases give no warning and may cause death in few seconds
The Exhaust of the CO2 backup cooling option needs to be lead out of the
building otherwise danger of suffocation is immanent
In their liquid state inert gases are extremely cold fluids which can cause
serious burns when in contact with the skin
A cryogenic spill often gives a low lying cloud of vapor that creeps along the
ground
Due to very specific local rules and regulations in different countries it is the
customers / owners responsibility to install suitable oxygen (O2) and carbon
dioxide (CO2) monitor systems which trigger alarms to protect operators in
and around the SAB store.
This option includes the evaporator, the tubing and the control, the Dewar or the
connection to a storage tank is not included. The size of the Dewar is defined by
the desired backup time and the desired refill interval. This backup solution is
fully monitored to ensure the temperature is maintained. Also an exhaust is part
of this option.
If the LN2 backup cooling is coming into action large amounts of N2-Gas are
released through the options exhaust. Therefore the exhaust needs to be lead
out of the building to not endanger personnel working in the area of the system.
WARNING !
Inert gases are asphyxiates and will displace oxygen to produce localized
oxygen deficient atmospheres
Inert gases are odorless, colorless and tasteless
Inert gases give no warning and may cause death in few seconds
The Exhaust of the LN2 backup cooling option needs to be lead out of the
building otherwise danger of suffocation is immanent
In their liquid state inert gases are extremely cold fluids which can cause
serious burns when in contact with the skin
A cryogenic spill often gives a low lying cloud of vapor that creeps along the
ground
Due to very specific local rules and regulations in different countries it is the
customers / owners responsibility to install suitable oxygen (O2) and carbon
dioxide (CO2) monitor systems which trigger alarms to protect operators in
and around the SAB store.
EAN 128, ISBT128, Code 93, Codabar, pharmaceutical codes, MSI, Code 11,
Code 16K, Code 49 and GS1 DataBar
• DataMatrix
Ziath DataPaq Express Rack Reader (suitable for placement in low temperature climate)
WARNING!
LiCONiC AG is doing there utmost to make sure that real sample location
matches the stores database by using barcode readers / scanners and guiding
SAB users during store and retrieve process.
But due to the semiautomatic nature of the SAB store the user / customer is
finally responsible for sample location by moving them into a particular cassette
and associated cassette level manually.
The SAB cassettes are typically customized to optimize the storage space for the
given application. The cassettes are either designed to hold a specific plate / tube
rack type, cryo box type or the stacking of plates is applied to further optimize
storage capacity.
The Kiwi Store SAB can handle almost every plate in SBS format currently
available on the market. Specific formats can be realized upon request. For
optimum performance LiCONiC offers a variety of cassettes.
Dryer
Cold Store
Store
Antechmaber
Cooler
Antechamber
Controller
Logic
Main Backup
Cooler Cooler
Insulation
Panels
Cold Store
–80°C
Cooler
Antechamber
400VAC
Main
Cooler
400 VAC
Main
Cooler
400 VAC
Antechamber
Aux. Cooler
400VAC
3.2 Electrical
Connection Store
2.4 LN2
inlet and
exhaust
LN2
Store
Backup
Interface
Store Auxiliary
Main Cooling
Cooling
Requirement
Details
8. Refrigeration
Cold Store
–80°C
Backup Cooler
Aux. Cooler
LN2 Dewar 1
400 VAC
Water Cooler
Outlet
st
1 Stage R404a
nd
2 Stage R508b
External
water
Cold water in
typ. ~14°C
Room AC
STC cooler
unit
Liconic STC
Store STC Cold water
return typ. ~30°C
Auxiliary Cooler
Back plane
cold store
LN2 Inlet
Connections
Main Cooler
Connections
Dryer
Power In
Main Cooler
Main Cooler
Cold store
Power In
aux. Cooler
Aux. Cooler
Connections
auxiliary cooler
Cooling water
return line
Water shut-off
valve
Refrigeration
cooling water
IN ¾“
Refrigeration
cooling water
OUT ¾”
Drain Interface Compartment and Storage Dryer may perform regular defrost cycles
Interface Compartment and Storage Dryer are equipped with a drain tube
leading the gathered water out of them.
Solution to remove the water may be foreseen. In the most simple case a
evaporation pan could be used. (Water would evaporate into the room over
time)
More sophisticated solution would be to provide a drain to guide the water out
of the room.
Auxiliary Cooler
Chamber
cold store
Manual shut
off valve
Automatic
LN2 Inlet valve
9.1 Overview
Power 230VAC
Interface unit
Power 230VAC
UPS Store
Power 400VAC
Power 230VAC main cooler
non UPS Store
Power 400VAC
Power 400VAC auxiliary cooler
main cooler
Communication
Power 400VAC
cooler
Main Power
switch
IEC 60309
3P+N+PE 400V / 16A
¾”
9.5 Requirements
Requirement
Details
In their liquid form these gases are very cold, (the boiling point of liquid nitrogen
is -196° C) . Such cold liquids are known as cryogenic liquids, (from the Greek,
kryos meaning frost and gene meaning producing). The liquid gases therefore
can be used as backup cooling option in case of power failure.
Major hazards
There are some major hazards associated with these gases
WARNING !
Inert gases are asphyxiates and will displace oxygen to produce localized
oxygen deficient atmospheres
Inert gases are odorless, colorless and tasteless
Inert gases give no warning and may cause death in few seconds
Oxygen is the only gas, which supports life. The normal concentration in the air
that we breathe is approximately 21%. If the oxygen concentration in air
diminishes or, (what amounts to the same) if the concentration of inert gases
increases, a situation is rapidly reached where the hazards of asphyxia become
very great.
The danger occurs as soon as the oxygen levels are lower than 16%. Inert
gases are odorless, colorless and tasteless and are therefore a great deal more
dangerous than other gases such as chlorine, ammonia, or hydrogen sulfide,
which can be detected by their odor, at very low concentrations. The asphyxiating
effect of inert gases occurs without any preceding physiological signals which
could alert the victim. It takes place very rapidly: just a few dozen seconds with
very low oxygen content. Stated clearly "you do not notice that you are passing
out!"
The danger from the inert gases such as nitrogen, argon, helium, etc cannot be
stressed enough!! . Especially in this case of Asphyxia as there are no warning
signs. People also vary considerably in their reactions to oxygen deficiency.
When a cryogenic liquid is released into the atmosphere, it evaporates and forms
a dense cloud. As it is very cold it is heavy and hangs close to the ground. So it
will flow into any low lying areas, such as drains or pits. Therefore the first
warning of a cryogenic spill is often given by the presence of a low lying cloud of
vapor that creeps along the ground. These clouds can be dangerous, as their
temperature or composition will be unknown. Visibility within the cloud is very low
with the possibility of falling, perhaps into contact with the cryogenic liquid. If the
cloud is nitrogen or argon then it is very likely that the cloud does not contain
sufficient oxygen to support life. In this case unconsciousness could be
immediate and death could follow.
CAUTION!
When you see a cryogenic spill or a large vapor cloud notify qualified
personnel immediately
Shut off the source of liquid, by a REMOTELY controlled valve
Do not enter a vapor cloud caused by vaporizing cryogenic liquid
DO NOT SMOKE!
When you see a cryogenic spill or a large vapor cloud notify qualified personnel
immediately. Call the Fire Department in the case of a large liquid oxygen spill.
Have them stand by until the spillage is dispersed.
If possible shut off the source of liquid, by a remotely controlled valve if one is
fitted. Do not expose yourself to the hazard, such as high oxygen, combustible or
asphyxiating atmospheres or cryogenic liquid. Do not allow liquid oxygen to enter
drains or sewers. If so advised, shut down any air conditioning or ventilating
systems that may draw vapor or gas from the spill into other locations.
CAUTION!
Wear loose fitting gloves of impermeable material such as leather when
working with cold liquid, solid, or vapor
Safety glasses are recommended when handling high-pressure cylinders
and in areas where vapors are discharged
Set-up gas cylinder in a well ventilated area. Keep cylinder away from sources of
heat. Storage should not be in heavy traffic areas to prevent accidental knocking
over or damage from passing or falling objects. Valve caps should remain on
cylinders not connected for use. Segregate full and empty cylinders. Storage
areas should be free of combustible material. Avoid exposure to areas where salt
or other corrosive chemicals are present. Install carbon dioxide cylinders with the
valve end up. See Compressed Gas Association Pamphlet P-1 for additional
recommendations.
Ventilation
Depending on the existing ventilation, amount of gas used and size of room
where the unit is installed and operated, ventilation may be required. A local
exhaust may be useful at the point of sources of possible gas vapors. Where low
lying areas are not naturally ventilated vents should be situated to avoid higher
than normal concentration. Hazards to be considered in confined spaces are:
In the case of permanent presence of personnel in a room where there are risks
of oxygen depletion mixture, ventilation must be maintained with at least 30mJ
per hour per person. The volume of air introduced per hour must not be less than
twice the enclosed space in question. Ventilation should never be carried out with
pure oxygen, but solely with air.
Ideally the monitor should be placed so that it can be read inside, as well as from
outside, the room which has gas present. The selection of the type of apparatus
depends on the nature of the work in the place to be monitored (temperature,
multiple detectors, portable equipment, etc). The picture below shows such a
monitor for CO2.
Work Permit
For certain types of work, safety instructions and a special work procedure must
be set up in the form of a work permit
Personnel Protection
Depending on the type of work to be performed, and the layout of the premises,
the decision to provide additional protection for personnel may be taken on the
initiative of the person in charge. Such additional protection may involve:
Considering the hazards mentioned above, it is essential to provide all those who
handle or use inert gases with all the information and training necessary
regarding safety instructions, means of prevention and procedures to be followed
to avoid accidents, as well as rescue procedures to be implemented in the event
of an accident.
Place the cylinder in an easily accessible and visible place next to the system.
Remember that the cylinder will have to be replaced or refilled routinely. As the
operator will have to check the cylinder routinely it should be clearly visible to all
personnel during routine work.
The room should provide a glass door or a window that will allow a good view of
the instrument and gas connection from outside the room. Often major defects
can easily be detected by visible vapor clouds or icy tubes without the risk of
entering a contaminated room.
Before installation make sure that the maximum specified inlet pressure is never
exceeded. Always install a main shut-off valve that will cut gas flow as close to
the source as possible. Secure cylinder when in use.
WARNING!
Gas cylinders are under extremely high pressure that must be reduced
for use with a StoreX incubator
Use a pressure-reducing regulator when connecting to lower pressure
piping system
Never exceed maximum specified inlet pressure
Never connect a Kiwi gassing inlet directly to a pressurized cylinder or
wall outlet
Use a two stage pressure regulator with indicating valves. The high pressure
gauge should have an indicating range of 0 to ~200 bar (0..3000 psi) to monitor
the pressure or the gas source. For CO2 a scale of 0..100 bar (0..2000psi) is
sufficient. The low pressure gauge should have an indicating range of 0..4 bar
(0..60 psi) and will monitor the input pressure to the Kiwi Store. A flow meter is
recommend since it will show correct operation of the system.
Low Pressure
Regulator
High Pressure
Adjustment Screw
Low Pressure Gauge
The gas inlet of the Kiwi Store is located on the back of the instrument. Attach the
low pressure hose securely to the fitting. Always use the security nut. Never
operate the unit without the security nut being tightly fit.
Security nut
Use 4x6mm PUR, PVC, Neoprene or Nylon tube for the low pressure connection
or in case of liquid nitrogen backup option a cryogenic safety tubing. Do not use
Silicon or natural rubber tubing. Make the low pressure tube long enough to allow
a minimum of movement of the gas cylinder for cleaning and maintenance work.
For longer length of tubing, copper or other metal tubing should be considered.
Foresee flexible tubing next to the instrument.
Be prepared for the following gas consumption values. These values may vary
significantly over time since they depend on the instruments use.
WARNING !
Improper use of CO2 or improper installation of the CO2 operated equipment
may cause death
Read CO2 safety information and installation instructions (See CO2
Installation) carefully. If you are not familiar with the handling of CO2 and / or
safety precautions contact trained personnel.
WARNING !
THRESHOLD LIMIT VALUE TLV = 5’000 ppm
TLV-STEL (Short Term Exposure Limit) = 15’000 ppm (=1.5% vol.)
Carbon dioxide does not support life and may produce immediately hazardous
atmospheres. At a concentration in excess of 1.5%, carbon dioxide may produce
hyperventilation, headaches, visual disturbances, tremor, loss of consciousness
and death. Symptoms of exposure In the concentration ranges of 1.5-5% may be
highly variable, but typical symptoms of carbon dioxide intoxication include the
following:
Persons suffering from the toxic effect of carbon dioxide should be moved to
areas with normal atmosphere. Self contained breathing apparatus may be
necessary to prevent toxic exposure or asphyxiation of rescue workers. Assisted
respiration and supplemental oxygen should be given if the victim is not
breathing. Frozen tissues should be flooded or soaked with tepid water (105-
115F; 41-46C). Do not use hot water. Cryogenic burns which result in blistering
or deeper tissue freezing should be seen promptly by a physician.
WARNING !
Nitrogen may cause suffocation by displacing the oxygen in air. Ensure
levels of at least 19.5% Oxygen
On evaporation Nitrogen forms dry ice at extremely cold temperatures of
–196°C (-320°F)
Nitrogen may cause mild skin irritation, moderate eye irritation and possible
gastric distress. Cause of chronic illness is not known.
In case of an accident evacuate all personnel from the affected area. Persons
suffering from lack of oxygen should be removed to fresh air. If victim is not
Ventilate area and attempt to shut off the release of gas by tightening the main
valve. If it is not possible to stop the release of gas, allow the gas to be released
there. If leaking from cylinder or its valve, contact your supplier. Monitor area for
a minimum of 19.5% oxygen content before allowing re-entry by personnel.
A pressure limiter and a flow limiter protect the system from inlet pressure
fluctuations and make it relatively insensitive to the amount of input pressure.
This solution is adequate for very low demanding applications and for non toxic
gases. Note that a significant amount of gas may be emitted to the instrument’s
environment.
Pressure
Exhaust
Limiter
Flow Limiter
Flow Regulator
An exhaust channels the surplus amount of gas to a defined point. The exhaust
is usually connected to the slightly under-pressurized exhaust system of the
building. The flow regulator is combined with a visual flow indicator and a scale
calibrated in l/min.
The picture below explains the functioning of this system. At the gas inlet a
pressure limiter and a flow limiter protect the system from inlet pressure
fluctuations. This also makes the system relatively insensitive to the amount of
input pressure.
Flow Regulator
Chamber
Gas Inlet
Pressure
Limiter
Flow Limiter
Exhaust
Main CPU
Gate Switch
Glass Door Switch
The actual gas flow is controlled by a solenoid. The solenoid is controlled by the
main CPU. By sensing both, glass door switch and Gate switch, the CPU knows
when there is a user access or when the Gate is opened and reacts accordingly.
In fact the CPU holds three values that represent the gas needed to add. The
standby value maintains the gas concentration when there is no access and
compensates for minor leakages of the incubation chamber. The Gate Access
Value compensates the gas loss caused by a Gate access while the User Access
Constant allows recovery after opening the glass door. The CPU internally
calculates the actual amount of gas needed in a specific situation. Since this
system limits the gas consumption to a fairly low level the exhaust does not have
to be connected in most cases.
Right at the inlet a pressure limiter protects the successive components against
overpressure, up to ~10 bar. The pressure limiter also makes the gassing option
relatively insensitive to the amount of inlet pressure. A flow limiter also limits the
gas flow to ~10 l/min. A solenoid controls the gas flow to the incubation chamber
in accordance with the gas concentration inside the chamber.
A second tubing system measures the gas concentration inside the incubation
chamber. A pump removes a small amount of gas from the incubation chamber
and feeds it to the gas sensor. A particle filter protects the pump and sensor from
contamination.
Flow Limiter
Solenoid
Pressure
Limiter
Chamber
Gas Inlet
Gas Sensor
Exhaust
Gas Pump
Gate Switch
Measurement
Tube
Glass Door
Display Switch
Particle
Main CPU Filter
The main CPU monitors the opening of the incubation chamber preventing
unwanted gas flow to the surrounding environment. The solenoid is therefore
only opened when the glass door is shut and the Gate has not been opened over
an excessive period of time.
The continuous flow system has a minimal of safety functions. The exhaust limits
gas leaking into the surrounding environment. A slight under pressure on the
exhaust will also lower uncontrolled gas flow into the surrounding environment.
The safety elements of the continuous flow system simply limit the gas flow and
protect against overpressure to some extent. This system may not be used in an
higher risk environment.
The above flowchart lists the steps that are implemented to avoid unwanted gas
emissions. As a first step the gas is always shut-off when the glass door is open.
This will eliminate the risk of large amounts of gas exiting when the user door is
opened. The same applies to the Gate. However, because of its smaller size
some gas flow is allowed even when the Gate is open. A Gate timeout is
implemented cutting gas flow when the Gate has been opened for an excessive
period of time. An additional general safety step is the gassing timeout. A gassing
timeout will occur when the solenoid is opened over a limited period of time.
Feature Comment F1 O2 C3
Pressure Limiter Limits internal pressure to some extent + + +
Gas Flow Limiter Limits maximum flow avoiding full gas flow + + +
Exhaust Channels exhaust gas + + -4
Gas Switch Turns gas flow on and off - + +
Door Monitoring Cuts gas flow when door opens - + +
Gate Monitoring Cuts gas flow when Gate is open over a - + +
certain period of time
Gas Flow Timeout Cuts gas when a continuous gas flow occurs - + +
over a long period of time
Gas Flow Lookup Controls gas flow according to system status - + +
Gas Control Controls gas flow according to measured gas - - +
concentration values
Gas Measuring Gas sensor measures actual gas - - +
concentration
Gas Display Gas concentration is displayed to user - - +
Gas Transmission Remote gas concentration readout - - +
WARNING!
Because of the high safety standards maintained within the Kiwi the highest
risk of failure or unwanted gas emission lies in the external installations
1
Continuous Flow System
2
Open Loop System
3
Closed Loop System
4
Available on request
Place the cylinder in an easily accessible and visible place next to the system.
Remember that the cylinder will have to be replaced or refilled routinely. As the
operator will have to check the cylinder routinely it should be clearly visible to all
personnel during routine work.
The room should provide a glass door or a window that will allow a good view of
the instrument and gas connection from outside the room. Often major defects
can easily be detected by visible vapor clouds or icy tubes without the risk of
entering a contaminated room. If this is not possible a oxygen level meter with a
remote alarm needs to be installed creating a acoustic and optical alarm in case
of low oxygen level.
Before installation make sure that the maximum specified inlet pressure is never
exceeded. Always install a main shut-off valve that will cut gas flow as close to
the source as possible. Secure cylinder when in use.
WARNING!
Never operate a damaged liquid gas cylinder
Special pressure-reducing regulator may be required to operate a liquid
gas backup.
Never exceed maximum specified cylinder pressure
Always ensure tight fit of all connections.
Supplied connection for liquid gas backup is a ½” SAE male (outside thread). For
liquid gas cylinder such as Cryogenic cylinders, standard connection hose are
available.
Over Pressure
Shut-off Valve Outlet Release
Pressure build up
circuit shut off valve
WARNING !
Nitrogen may cause suffocation by displacing the oxygen in air. Ensure
levels of at least 19.5% Oxygen
Liquid Nitrogen has extremely cold temperatures of –196°C (-320°F)
Contact with liquid Nitrogen may cause skin frost burns and severe eye
damages.
In case of an accident evacuate all personnel from the affected area. Persons
suffering from lack of oxygen should be removed to fresh air. If victim is not
breathing, use trained personnel to administer a supplemental oxygen supply and
Ventilate area and attempt to shut off the release of gas by tightening the main
valve. If it is not possible to stop the release of gas, allow the gas to be released
there. If leaking from cylinder or its valve, contact your supplier. Monitor area for
a minimum of 19.5% oxygen content before allowing re-entry by personnel.
Refill interval of LN2 cylinder is depending on the required minimum backup time
and the available LN2 cylinder volume. All cryogenic cylinder have a certain
leakage rate which will define the time after which the minimum filling level of the
tank is reached. The filling level is shown by a gauge mounted next to the
connection points on the cylinder. The actual design may vary depending on
cylinder manufacturer.
Before disconnecting the connection hose close the pressure builder valve, the
shut off valve on the cylinder and the shut off valve on the connection point on
the actual storage chamber. Use two suitable spanner to loosen the connection
on the chamber (or on the cylinder depending on refill procedure). Usually the
cylinder is then moved to a refill station and refilled with LN2. Depending on
cylinder manufacturer the process may vary. Please refer to the manufacturers
cylinder manual for further details.
To reconnect the cylinder with the system, proceed in the reverse way to the
disconnecting procedure. Connect back the hose. Open the main shut off valve
on the cylinder, open the shut off valve on the storage chamber and open the
valve on the pressure builder. The pressure builder valve has to stay opened in
order to ensure proper operation of the LN2 backup option.
WARNING!
Wear personnel protection when disconnecting, refilling and reconnecting
LN2 cylinder.
Ensure the shut off valves are closed when disconnecting the cylinder.
Refer to cylinder manufacture manual for exact refill procedure
CAUTION !
Shut off valve of pressure builder has to stay open as long as the cylinder
is connected to the storage.
When replacing a cylinder, make sure the shut off valve on the CO2 cylinder and
on the unit are closed. Use a spanner to disconnect the pressure regulator from
the bottle. Put the new bottle in place and reconnect the pressure regulator.
Open the shut off valve of the cylinder and of the storage.
WARNING!
Ensure the shut off valves are closed when disconnecting the cylinder.
11. Doors
In order to prevent large amount of humidity introduced into the system, the user
door is heated. Note: Leaving the user door open for extended periods will cause
frost and ice inside the system.
User door
Springloaded Mechanism
The user door is opened by pulling on the left edge or on the handlebar and may
be opened to an angle of almost 180 degrees.
CAUTION !
Use the door handle to open user door
Keep user door closed when it is not being accessed
Keep user door access to a minimum
The seal must firmly close user door
Keep the seal clean
WARNING !
It is dangerous to leave the user door open over long periods of time.
Avoid excessive condensation on the glass door
12. Cassettes
Customized SAB cassettes allow simple and comfortable loading and unloading
of storage samples. Once prepared, the samples can be quickly transferred into
the climate chamber of the Kiwi SAB unit by means of the cassette lifts installed.
There is a wide variety of cassette types available. Make sure that you work with
the proper cassette size that matches the type of plates or Cryo Boxes you are
using.
CAUTION !
Do not add, bend or remove support rails.
If cassettes with different pitch height are used within the same system,
make sure a prior removed cassette is put back in exactly the same position.
Failures being generated like that, will result in destinations not found or
crashes.
Only a trained System Integrator may alter the instrument configuration and
reset the lift accordingly.
The SAB cassettes (see picture below) are open at their front for loading.
Cassettes must be treated with great care. They are aligned within 1mm when
leaving the factory. Misaligned cassettes must be returned for re-alignment.
Misaligned, tilted or bent cassettes may cause handling failures. Special care
must be taken when cassettes are fully loaded with plates.
Never apply force to the cassettes. Do not drop a cassette as mechanical shock
will misalign it. Never lift a cassette by its side walls. This may bend the side walls
of the cassette. Bent side walls cause the plates to block when being loaded or
unloaded.
The plates / tube racks or Cryo Boxes are presented to the operator by means of
the cassette lifts shown in the pictures below :
CAUTION !
Make sure that ALL plates are pushed to the back of the cassette. Plates
jutting out the front side of the cassette may cause collision with the cassette
lift.
Lift cassettes by their cassette handles only
Do not drop cassettes. Misaligned cassettes can cause handling errors
Do not autoclave cassettes with plastic rails. Use disinfecting solution for
sterilization. All stainless steel cassettes may be autoclaved.
Keep cassette clean. Particles and ice could prevent reliable handling.
The Barcodes must be placed on the shorter side of the plates and must face
towards the open side of the cassette. The barcodes should be parallel to the
plate’s lower edge. The barcode label should be centered both vertically and
horizontally. On higher plates (e.g. deep well plates) place the barcode further
down. Preferably barcode labels should be at the same height independent of the
plate height. Barcode reading reliability is improved when the barcode is not
covered by the plate’s lid.
WARNING !
DANGER !
Use barcode testing device to verify barcode quality. Quality Class A, B or C are
required. Identification of Class D is not guaranteed. For sufficient print quality
use barcodes printed by thermal transfer or photographic methods. The barcode
label surface should be mat and clean. Do not use yellowed, dirty or damaged
barcodes.
The 2D codes must be placed on the bottom of the tube. The 2D reader operates
either on a scanner or camera based principle. If a camera is used it requires
lighting by short flashes to take pictures. These flashes may be a potential trigger
for people with photosensitive epilepsy.
WARNING !
A few people may experience epileptic seizures when viewing flashing lights
or patterns. Avoid watching the 2D code reading for longer time.
Use barcode testing device to verify barcode quality. The code label surface
should to be mat and clean. Do not use dirty or damaged barcodes.
Service COM-Port
Power Switch
Indicator Light
Display Temperature;
Dryer Evaporator
Display Temperature;
Service Area
The alarm indicates that the handling is not activated. In order to activate the
handling, this has to be triggered out of the controlling software.
To switch on the cooling unit the main power switch needs to be turned into the
position stating ON. The cooling unit will immediately start on the auxiliary cooling
as the handling is not switched on. As soon as the SAB handling (and therefore
the PLC controlling the climate) is switched on the main cooling will take over.
USB communication
RS232 communication
Hardware communication
Thermocouple Terminals
CAUTION !
Do not disconnect any cables, except a LiCONiC engineer did advise to
do so.
All cooling units have to be permanently switched on.
The key switch has to stay in automatic position.
Ensure normal conditions (p.e.: Power and Water) before switch on the
cooling units.
To switch on the cooling unit the main power switch needs to be turned into the
position stating ON. The key switch has to be in automatic position (see Picture
3). This has to be performed on both cooling units main and auxiliary.
USB communication
RS232 communication
Hardware communication
Thermocouple Terminals
CAUTION !
Do not disconnect any cables, except a LiCONiC engineer did advise to
do so.
All cooling units have to be permanently switched on.
The key switch has to stay in automatic position, for normal operation.
Ensure normal conditions (p.e.: Power and Water) before switch on the
cooling units.
CAUTION !
Ensure normal conditions being restored to operate the main cooling
Ask for assistance from a LiCONiC engineer in case of insecurity about
normal conditions and performing the recovery procedure.
Observe cooling after recovery. Contact immediately a LiCONiC engineer
in case of repeated stop of cooling after recovery.
16.3.1 Procedure
SAB ULT System running on auxiliary cooling
o Ensure normal operation conditions being restored. Optimal reset point would
be system returned to target operation Temperature.
o Toggle the key switch of main cooling (move key switch out of normal position
and back again)
o Auxiliary cooling should stop now and the red alarm light of main cooling
should switch off
o The Error Flag 1506 has to be manually set back with “RS 1506” sent through
the service terminal.
o Toggle the key switch of aux and main cooling (move key switch out of normal
position and back again)
o Either Main only or both cooling units will operated. When Target operation
Temperature of the Storage is reached toggle the key switch of the main
cooling again.
o As soon as normal operation conditions are restored the gas backup should
stop operation an electric cooling unit takes over.
CAUTION !
Ensure normal conditions being restored to operate the main cooling
Ask for assistance from a LiCONiC engineer in case of insecurity about
normal conditions and performing the recovery procedure.
Observe cooling after recovery. Contact immediately a LiCONiC engineer
in case of repeated stop of cooling after recovery.
16.4.1 Procedure
STC ULT backup cooling switch to manual operation
o Turn key switch into horizontal position (Manual mode)
o The first stage will switch on immediately. The second stage will kick in after
about 15 minutes (yellow and green light lit)
o When main cooling is back into normal operation (yellow and green light lit)
switch back key switch of backup cooling into horizontal position.
Contact:
Liconic AG Main: +423 373 63 39
The operator should regularly read the display readings and check whether they
match the desired values. Readings like “0.00” or “9.99”, readings that are far
out, instable or that differ over long periods of time could indicate a malfunction of
the instrument or could be caused by faulty settings of the system.
The electrical cables should be routinely inspected for wear, defects or contusion.
There should be no mechanical tension on any cable. Broken connectors should
be replaced immediately. All electrical installations should be kept clean and well
protected from water and humidity.
The same applies to cables and gas tubing. They should be free of mechanical
tension and should show no signs of wear or contusion. Sharp bends should be
avoided under all circumstances. All fittings should be secured. Missing security
nuts must be reinstalled immediately. Any broken pneumatics parts must also be
replaced immediately.
The mechanical appearance of the instrument should be clean and “straight”. All
movements should be smooth and precise. Watch out for unusual noises as
these may indicate misalignments or mechanical wear.
Seals should close the chamber firmly. Check seals by placing a piece of paper
between the seal and door. When trying to pull on the paper, there should be an
even resistance. Check the seal on the glass door, gate and front door.
Condensation, mainly on the front glass door, presents a possible severe danger
as it may lead to water flowing into the electrical parts of the instrument.
Excessive condensation on the glass door will occur when the front door is open
over a long period of time. Furthermore it may also indicate a faulty heater or
high environmental humidity.
WARNING !
Icing on the gas cylinder or other parts of the gassing system indicate
excessive gas flow.
In the case of such icing leave the room immediately and call for help
Do not try to locate the source of the problem or try to fix it
- LIMS integration: A LIMS system does send or provide data of plates going to
be loaded into the system.
- Reading of content: Reading the 1D code and the the 2D codes present on
the plate going to be imported, prior to the import.
LIMS integration
This is customer specific. In general the data are provided in the background to
the system. The individual plates will be identified with a 1D barcode scanner to
identify which plates are available to be imported at the given time
Scanning of content
With a scanning application the individual plates can be scanned prior to the
import process. The content is read and shown to the user to identify if the read
content does comply with the physical appearance of the plate. Samples
presence, readable and unreadable 2D codes are clearly marked in a visual
display of the plate.
In some cases the scanning has to be performed multiple times with some
cleaning steps in between to get all of the samples correctly identified.
18.2.2 Import
The import of plates is executed within PlateExplorer. Plates which have been
preloaded to the system for being imported can be stored within the system. The
import is triggered by pressing the import button in Plate Explorer:
After pressing the import button the below window is shown. It does display the
preloaded plates waiting for import. The plates are shown on the left.
18.2.3 Export
The export of plates and samples is executed within PlateExplorer. Plates which
have samples on them or are requested as a whole can be exported from the
system. The export is triggered by pressing the export button in Plate Explorer:
After pressing the export button the export is started. Depending on the
integration the plate overview is immediately displayed or the job selection is
opened. In the job selection the job is searched on the computer and sent to the
system. After this also the plate overview is shown.
Suffocation / burning by indirect contact with • When LN2 has been introduced into
LN2 the system and the DEWAR needs to
be exchanged, wear protective
clothing.
• In case no exhaust leading the N2
gas out of the room, only enter the
room wearing a breathing mask with
oxygen supply.
• During operation of LN2 backup, do
not touch any piping showing frost
without protective clothing.
Bodily harm while entering the area by • Use suitable equipment for access to
slipping off or falling down service area (save and robust ladder)
Contaminated or destroyed samples or • To avoid a contamination of or by the
probes by the user (product safety and samples, no sharp edge tools should
personnel safety) be used within the service area.
Sample location / Database integrity • By using barcode readers / scanners
and guiding the SAB user Liconic AG
is doing utmost to make sure that real
sample location matches the stores
database but due to the semi -
automatic nature of the SAB store the
user / customer is finally responsible
for sample location by moving them
into a particular cassette and
associated cassette level manually.
General Rules:
1) DO NEVER WORK ALONE WITHIN THE DEVICE. ALWAYS ORGANIZE A
SECOND PERSON AS SAFTEY GUARD ON THE OUTSIDE OF THE DEVICE
THAT IS CONNECTED TO YOU VIA RADIO. FURTHER AN ADMINISTRATIVE
PERSON MUST BE REACHABLE IN A SHORT TIME. BOTH MUST BE 100%
PHYSICALLY FIT.