SECURVIEW-DX INSTALLATION Man-00335 - 001 - 01
SECURVIEW-DX INSTALLATION Man-00335 - 001 - 01
SECURVIEW-DX INSTALLATION Man-00335 - 001 - 01
© Copyright Hologic 2005. All rights reserved. Printed in USA. Hologic and Hologic logos, LORAD, LORAD DSM, LORAD Elite, StereoLoc, SmartWindow, ScoutMarc,
StereoGuide, Multicare, Dataport, AutoFilm ID, HTC Imaging System, Affinity, Selenia, A Hologic Company, are trademarks or registered trademarks of Hologic, Inc., in the United
States and other countries.
Installation Manual
Table of Contents
List of Tables -v
Preface P-1
1.0 About This Manual ..............................................................................................................................P-1
2.0 Quality Control ...................................................................................................................................P-1
3.0 Training ...............................................................................................................................................P-1
4.0 Prerequisites ........................................................................................................................................P-1
5.0 Product Complaints .............................................................................................................................P-1
6.0 Terms and Definitions .........................................................................................................................P-2
7.0 Warnings, Cautions, and Notes ...........................................................................................................P-3
8.0 SecurViewDX Manual Set .....................................................................................................................P-4
9.0 Using the Installation Manual ..............................................................................................................P-4
Chapter 1—General Information 1-1
1.0 Introduction .........................................................................................................................................1-1
1.1 SecurViewDX Diagnostic Workstation ...........................................................................................1-1
1.2 SecurViewDX Manager ..................................................................................................................1-1
1.3 Multi-Workstation Configuration ...................................................................................................1-2
1.4 DICOM Information—Manager or Stand-Alone ............................................................................1-4
Chapter 2—Receiving and Unpacking 2-1
1.0 Receiving Instructions ..........................................................................................................................2-1
2.0 Unpacking ...........................................................................................................................................2-1
Chapter 3—Installing the Diagnostic Workstation 3-1
1.0 Installing a Stand-Alone or Client Workstation .....................................................................................3-1
1.1 Pre-installation Considerations ......................................................................................................3-1
1.2 Setting up the SecurViewDX Diagnostic Workstation .....................................................................3-2
2.0 Installing the SecurViewDX Manager ....................................................................................................3-5
2.1 Pre-Installation checklist ...............................................................................................................3-5
2.2 Starting the Manager .....................................................................................................................3-6
Chapter 4—Configuring the Diagnostic Workstation 4-1
1.0 System Startup .....................................................................................................................................4-1
2.0 Configuring the SecurViewDX Diagnostic Workstation .........................................................................4-1
2.1 Windows Setup .............................................................................................................................4-1
2.2 DICOM Settings—Manager and Stand-Alone ................................................................................4-1
2.3 Configuring a Multi-Workstation ...................................................................................................4-5
2.4 Calibrating the CRT Monitor .........................................................................................................4-7
2.5 Calibrating the LCD Display ..........................................................................................................4-9
3.0 Converting Stand-Alone Workstations to Clients ................................................................................4-14
3.1 Pre-Installation Checklist—Existing Stand-alone ..........................................................................4-14
3.2 Converting ..................................................................................................................................4-14
MAN-00335 i
Installation Manual
ii MAN-00335
Installation Manual
List of Figures
List of Figures
MAN-00335 iii
Installation Manual
List of Figures
iv MAN-00335
Installation Manual
List of Tables
List of Tables
MAN-00335 v
Installation Manual
List of Tables
vi MAN-00335
Installation Manual
Preface
About This Manual
Preface
3.0 Training
Hologic does not accept responsibility for injury or damage associated with improper or
unsafe system operation.
Service engineers must ensure that they receive training on SecurViewDX with Hologic
training programs prior to servicing the unit.
Service engineers should refer to the User/Administration Guide for directions on how to use
and service SecurViewDX.
4.0 Prerequisites
The procedures in this manual assume that you know how to do the following operations
related to the system’s graphical user interface:
• Perform mouse and trackball operations, such as click, drag and select
• Select from menus
• Click buttons
• Enter information in text fields
• Select options
• Select entries from drop-down lists
• Use scroll bars
Operating System (OS) The basic software control system which runs all functions
of a computer.
PACS Picture Archiving and Communications System. A
computer and network system for the transfer and
archiving of digital medical images
Pre-Fetch The retrieval of prior patient data in preparation for an
upcoming exam. Many PACS systems can be configured to
pull older images for a patient and forward them to the
SecurViewDX based on scheduling information that comes
from a PACS Broker/HIS/RIS. Pre-fetching is usually
performed during off-hours when traffic is lower.
WARNING! This alerts you to procedures that you must follow precisely to
avoid causing potentially serious or fatal injury to yourself or
others.
Caution: Cautions indicate procedures that you must follow precisely to avoid dam-
age to equipment, loss of data, or corruption of files in software applica-
tions
Note… Notes indicate important information that must be followed to ensure the
proper operation of the system.
1.0 Introduction
1.1 SecurViewDX Diagnostic Workstation
The SecurViewDX Diagnostic Workstation is used by a physician to review images.The
Workstation consists of:
• Two High Resolution Portrait displays
• Computer
• Keyboard and Mouse
• UPS
• Workflow Keypad with Trackball
DICOM
Network
Client Client
Manager
Database
Client Client
2.0 Unpacking
WARNING! To prevent injury to personnel and/or damage to equipment, care
must be taken when uncrating the equipment.
Unpack as follows:
1. If applicable, cut the retaining straps that secure the boxes to the shock-mounted
wooden pallet. Remove the boxes from the pallet.
2. Open all boxes and check their contents against the packing list and sales order.
Note… If there is a discrepancy between the contents and the packing list or sales
order, contact Hologic immediately. If it is necessary to repack any items for
future installation, use the original packaging materials.
3. Carefully remove all shipping materials (foam padding, tie-downs, straps, shipping wrap,
etc.).
4. Inspect each item for damage, then safely store them near the installation site.
MAN-00335 2-1
Installation Manual
Chapter 2—Receiving and Unpacking
Unpacking
2-2 MAN-00335
Installation Manual
Chapter 3—Installing the Diagnostic Workstation
Installing a Stand-Alone or Client Workstation
Note… The USB driver for the Photometer may be lost if the Photometer is
improperly disconnected. Hologic recommends that the Photometer not be
disconnected from the computer.
• Workflow keypad to a USB port on the rear of the computer.
• Network cable between the computer and the hospital network.
• Barcode Scanner (not shown) to a USB port.
Note… If you are installing LCD monitors, go to “Cable Connections for the LCD
Displays” on p. 3-4.
1. Set the monitors in position.
2. Referring to Figure 3-2, connect:
• The video cable for the left monitor to the top monitor adapter on the
computer, matching the colors on the adapter board to the colors on the
plugs.
• The video cable for the right monitor to the bottom monitor adapter on the
computer, being sure to match colors.
• A monitor serial cable between the COM1 port of the computer and the
Remote In port on the back of the left monitor.
• The second monitor serial cable between the Remote Out port on the back
of the left monitor and the Remote In port on the back of the right monitor.
protective
covers
6. Connect one end of the UPS serial cable into the UPS and the other into COM2
on the rear of the computer.
COM2
UPS Serial
Cable
Monitor
Power Cables
Computer
Power Cable
Figure 3-4: Monitor Power and Cable connections
Note… At the completion of this installation, the AWS, CAD server, and/or PACS
must be configured to send to the Manager (not to individual clients).
Consult the appropriate User’s Manuals for instructions on how to configure
these devices.
Note The Shift button must be held down during the entire sequence, until you
see the Logon prompt screen.
5. Log on to Windows as administrator.
6. Configure the following in Windows:
• User Interface Language
• Network Settings
• Regional Settings
• Host Name
7. Restart the computer.
Note… Only service users have access to DICOM settings.When entering DICOM
titles, always use capital letters.
1. Log into the SecurViewDX software using a service account.
2. Click the Administration tab on the right-hand side of the right monitor.
Default
Setting
Note… In box #2 of Figure 4-3, only PACS should be selected. In box #4 of Figure
4-3, only SecurViewDX or SecurViewRT Workstations and Managers for
SecurViewDX and SecurViewRT should be selected.
The multi-workstation configuration consists of three radio buttons that allow the
service user to select the workstation type:
• Stand-Alone Workstation
The workstation is a single SecurViewDX Diagnostic Workstation, without a
connection to a Multi-Workstation Manager.
• Client
The Diagnostic Workstation is a client within a multi-workstation system. If the
Multi-Workstation Client option is selected, the Multi-Workstation manager
address (either as DNS name or as IP address) can be entered in the “Manager
Address” text field. The Check button is also enabled. The Check button checks
for the availability of the entered network address.
• Multi-Workstation Manager
The machine is a manager in a multi-workstation system.
Launch MediCal 2.03 by clicking on the MediCal 2.03 shortcut on the desktop and log in as
an advanced user:
1. On the New Display detected window, click the Add Now button.
3. When the Configuration Setup Wizard, Figure 4-8, appears, click Next and follow the
Wizard instructions.
4. When asked, click Yes to remove existing configuration.
5. When asked, choose Barco Flat Panel.
7. Select Yes for White and Black Luminance and click Next
8. Set the White Luminance to 600 and the Black Luminance to 1 and click Next.
9. Select Calibrate all displays now and click Next. Calibration is performed automatically.
Calibration Setup is complete.
19. Ensure the Measure Quality Level is enabled for Monthly Q/C.
20. Right-click Measure Quality Level.
21. Select Properties.
22. Change the Warning Level to 5% and Error Tolerance to 10% for both displays.
23. Right-click Measure Quality Level.
24. Select Run Now and ensure that the test returns an OK status.
25. Ensure the View Pattern: SMPTE is disabled.
3.2 Converting
1. On a workstation that is to be configured as a Client, log in to the SecurViewDX
application as the service user and select Exit to Windows.
2. From the Start menu, select Log Off scr.
3. At the Windows login prompt, log into Windows as Administrator.
4. Ensure that the Windows System time matches the Manager’s Windows system time by
opening a Command prompt and typing:
net time \\<MANAGER-IP> /set
where the <MANAGER-IP> is the Manager IP recorded in a previous section.
5. Log out and log in as SCR.
6. Log in to the application as the service user.
7. Update the appropriate license keys for the Client Workstation.
Note… SecurViewDX requires new license keys. A master list of license keys is
included in the e:\mbc-scr-1\Hologic-Service\license_keys.txt file. For
each softcopy workstation, note the ID printed on the dongle (begins with
8-5E7…), then search the license_keys.txt file to locate the appropriate
keys. If the keys cannot be located, contact the Hologic Help Desk.
a. Log in to the application as the service user.
b. Select Administration
c. Select the Licensing tab.
d. Highlight any currently installed Generation 1.0 or 2.0 license keys and select
Delete.
e. If you are converting a Stand Alone workstation to a Client, highlight the SCW-
STANDALONE key and select Delete.
Note… If this is a new installation, then the Client license keys were entered at the
factory. If this is a conversion of existing stand-alone workstations, enter the
key for SCW-CLIENT.
8. Under the Feature field, enter the following:
• Feature: SCW-CLIENT
• Generation: 3.0
• Key: 20 character string obtained from license_keys file
Note… Make sure that the Key field is a 20 character string, append zeros to the
end if the key contains less then 20 characters.
9. Click the Check button. The system verifies the information with the dongle and adds the
key if successful.
10. If connectivity to a CAD processor is required, enter the CAD license information and
again press the Check button. If there is no CAD processor, do not enter this information.
• Feature: MBC-CAD
• Generation: 3.0
• Key: 20 character string obtained from license_keys file
11. The client workstation is initially set up as a stand-alone system. To convert from a stand-
alone system to a Client:
a. Click the “Multi-Workstation” tab.
b. Click on the “Cluster Client, Manager Address” button.
c. Enter the Manager IP (See Chapter 3, Installing the SecurViewDX Manager, p. 3-5)
and click Check. Confirm that it connects to the Manager.
d. Click Save. Confirm that you want to save the new system type.
e. The MBC Workstation Cluster Integration window appears. Check the appropriate
boxes:
• Migrate Image Data: Click only if you want to migrate all images from the Client
to the Manager.
Note… If multiple workstations contain the same patient data, then you only need
to perform the image migration for one of them.
• Migrate Annotations: Click only if you want to migrate all annotations &
markings from the Client to the Manager
• Migrate User Data: Click only if you want to migrate any users from the Client to
the Manager
f. Follow the on screen instructions to migrate desired users.
Note… If this is a new installation, then there is no data to transfer and you should
leave the check boxes unchecked. However, you must still perform the
Migrate step.
g. Press Migrate.
• After the migration is complete, both the Client and the Manager restart and
synchronize.
• When the system restarts, the Client synchronizes with the Manager. The length
of time it takes to synchronize depends on how much data is on the system.
h. Repeat this section for all clients in the multi-workstation system.
Note… You should use the Network Time Protocol to synchronize manager and
client times.
2.0 Configuration
The scanner has to be configured for the same country code as the keyboard (American for
example).
MAN-00335 5-1
Installation Manual
Chapter 5—Configuring the Barcode Scanner
Configuration
RECEIVER sets which module gets the result of the scan. Only LIGHTBOX is currently
supported.
5-2 MAN-00335
Installation Manual
Appendix A—Technical Characteristics
This section lists the relevant technical characteristics for the stand-alone or client
workstation.
Operating System Windows 2000 Professional with Service Pack
3 or greater.
Computer/Memory Dual Pentium Processors with clock rate 2
GHz or higher; 2 GBytes of 800 MHz ECC
RAM or greater.
Disk capacity >140 GB Image Storage
Archival Media CD-RW with a read rate of at least 24X.
Graphic Controller Card(s) 5 MegaPixel 10-bit data output.
Monitor 5 MegaPixel, grayscale, Portrait, 2048 x 2560
pixels
Network Interface (NIC) 10/100Base-T Ethernet
DICOM Functionality Provides the following DICOM functionality
Verification SCU
Store SCP
Query and Retrieve SCU
Print SCU
Accessories Workflow keypad
Printing User may select an image and send it to a
validated DICOM printer. The system prints
the image with all the current Window/Level
settings, zoom factor, and displayed
annotations or a diagnostic image (as
selected).
MAN-00335 A-1
Installation Manual
Appendix A—Technical Characteristics
A-2 MAN-00335
Installation Manual
Appendix B—Troubleshooting
Network Problems
Appendix B—Troubleshooting
If the network error has occurred due to a misconfiguration of the manager address on the
client side:
• The Microsoft Windows “Administrator” user can change the address using the separate
tool setmwserver.exe.
• This tool allows you to reconfigure the system type and to set the manager address.
Index
B F
bar code final checks 4-18
beep volume and sound 5-1
emulate keypad 5-1 H
barcode scanner 5-1
hardware setup 3-2
beep volume and sound 5-1
I
C
image export 4-6
calibrating
Image Processing Configuration 4-6
CRT monitor 4-7
installing
LCD display 4-9
client workstation 3-1
client 3-1
manager 3-5, 3-6
workstation 3-1
stand-alone workstation 3-1
complaints, product P-1
configuring L
barcode scanner 5-1
image processing LCD display, calibrating 4-9
deleting images 4-6 licensing 4-6
image export 4-6
multi-workstation 4-5
M
client 4-5 manager 1-1
manager 4-5 manual set P-4
standalone 4-5 MediCal LCD Setup 4-7, 4-9
connections multi-workstation
CRT displays 3-3 configuration 1-4
converting stand-alone workstations to clients 4-14 multi-workstation configuration 1-2
CRT cable connections 3-3, 3-5
CRT monitor, calibrating 4-7 N
D network A-1
node types and tasks 1-3
definitions P-2
deleting images 4-6 P
destination tab 4-4
diagnostic workstation 1-1 printer tab 4-3
DICOM product complaints P-1
destination tab 1-4, 4-1, 4-4 Q
printer tab 1-4, 4-1, 4-3
settings 4-1 Q/A setup 4-8, 4-12
setup tab 1-4, 4-1, 4-2 quality control P-1, 4-8, 4-12
E R
emulate keypad 5-1 receiving 2-1
receiving and unpacking 2-1
MAN-00335 i
Installation Manual
Index
ii MAN-00335