Guideline of Packaging Validation
Guideline of Packaging Validation
Guideline of Packaging Validation
2012 recommendations | 5
Assurance of the quality properties must 4.5 Process control and monitoring tial validation (IQ, OQ and PQ). That
be verified for each packaging system The routine tests that are established dur- no changes were made to materials,
(sample). ing the validation as being necessary must processes or sterilization compared to
Pursuant to standard ISO 11607-2, § 5.3.2 c be documented (e. g.in the standard op- the previous validation must be con-
the quality properties for sterilization erating procedure). This is intended as firmed in the revalidation report. If
sheets and reusable sterilization contain- a means of ensuring that changes in the changes are made to materials, proc-
ers are as follows: packaging process are detected on time esses or sterilization how such chang-
before they compromise the sterile bar- es will affect the packaging process
–– continuous closeness/integrity
rier systems and the requirements are no results must be elucidated. The re-
–– no punctures or tears (not applicable to
longer met. These include, e. g.: sults must be documented. Based on
reusable sterilization containers)
–– Visual inspection16 these, an individual revalidation plan
–– no other visible damage or material ir- must be drafted. Accordingly, in the
–– Peelability (e. g. peel test as per EN 868-
regularities15. event of material changes, for exam-
5, Annex E «Method for determination of
These quality properties must be veri- the peel characteristics of paper/plastic ple, operational qualification (OQ) and
fied and documented by means of suita- laminate products») performance qualification (PQ) must
ble processes or tests. The sterile barrier be partially or fully repeated, and if
–– Seal integrity test (e. g. dye penetration
systems or packaging systems are opened changes are made to the packaging
test/ink test as per ISO 11607-1, Annex
one step after the other, verified and docu- process or to the equipment used in-
B17)
mented (for photographic documentation stallation qualification (IQ) must also
see Annex B.8/C.7). –– Seal integrity indicator18 (e. g. Seal
be repeated. For revalidation it must
Check)
be ensured that the documents used
4.3 Drafting of a validation report –– Tensile strength of seal seam (e. g. de- meet the current requirements. The
The validation procedures and results termination of seal seam strength as per checklists must be updated if neces-
must be documented in a summary re- EN 868-5, Annex D «Method for deter- sary. An individual validation plan is
port. The checklists, protocols and any mination of the seal seam strength of required for each revalidation or per-
photographic documentation used serve pouches and reels» formance requalification. The «Vali-
as evidence and must be enclosed in an –– Stepwise opening of packaging (in the dation plan» checklists in Annex A.1,
annex to the report. case of sterilization sheets or reusable B.1 and C.1 can be used. ■
The report must contain, at least, the fol- sterilization containers).
lowing information:
Intervals (e. g. daily, weekly, monthly,
Validation plan yearly) and acceptance values must be
–– Evidence of implementation of the vali- defined for the routine tests needed, in-
dation plan (IQ, OQ and PQ checklists cluding the action to be taken if a test re-
completed as per Annex) sult is not satisfactory. The routine test
–– Evaluation of the results results must be documented. This proce-
–– Photographic documentation for manual dure must be set out in the quality man-
packing processes agement system.
–– Details and explanation of any devia- 15 The ISO 11607-2 standard uses «No mate-
4.6 Process changes and revalidation rial delamination or separation» here.
tions from validation plan
Processes must be revalidated: 16 For visual inspection standardized test
–– Formal approval of validation methods can be used (e. g. ISO 11607-1,
–– Unscheduled revalidation,
–– Process control and monitoring Annex B [ASTM F1886] for seal seams or EN
• for example in the event of changes 868-8 for reusable sterilization containers.
–– Process changes and revalidation to materials, processes, including 17 The basis for this test method is ASTM
4.4 Formal approval of validation changes to equipment or occurring F1929 «Standard test method for detecting
during sterilization (revalidation) seal leaks in porous medical packaging by dye
Validation, as documented and evaluated
penetration».
in the report, must be formally approved, –– Scheduled revalidation,
18 The seal integrity indicator must not un-
and duly documented, by the competent • at regular intervals, i. e. in general af- der any circumstances be cut since it must
person appointed by the operator. This can ter one year if no changes were made always be guaranteed that the entire pinch
be recorded, for example, in a field pro- to materials, sealing process or steri- roller of the sealing device is printed off.
vided to that effect in the validation plan. lization (performance requalification). Furthermore, the seal integrity indicator shall
always be made of the same type of material
If all validation results are not accepted, • provides evidence that the packag- as the porous part of the packaging (medical
this must be clearly documented, includ- ing process continues to be within grade paper as per EN 868-3 or HDPE as per
ing assessment of any remaining risks. the limits defined at the time of ini- EN 868-9/10)