SERVICE MANUAL
FOR
MODEL EBIT 50
-1-
EBIT 50 SERVICE MANUAL
CONTEXT
1.1 Introduction............................................................................................................- 3 -
1.2 Service policy ........................................................................................................- 3 -
1.3 Service Procedure .................................................................................................- 3 -
-2-
EBIT 50 SERVICE MANUAL
1.1 Introduction
This guide covers the preparation and maintenance of the ultrasound system, transducers, and
accessories. Refer to the manufacturers’ instructions for specific information about peripherals.
Please read the information in this guide before providing service to EBIT 50 color ultrasound
system. It applies to the ultrasound system, and transducers.
CHISON provides life-time service for all produced ultrasound machines. Meanwhile, according to
the warranty card enclosed in the packing box, the corresponding free warranty period is provided
for the main unit and probe (transducer) from the date of Ex-factory.
CHISON Service Dept. provides efficient service solutions to customers timely, including technical
support, spare parts support, field service, upgrade, warranty extension, and on demand service.
The engineer must have ultrasound training before provide service, CHISON can provide varieties of
training, including operation, application maintenance, and troubleshooting.
If you encounter difficulty with proving service to CHISON products, please refer to the user manual
and service manual firstly to correct the problem.
If the problem is still not solved there, please contact with CHISON Service Dept at the following
numbers or addresses:
-3-
EBIT 50 SERVICE MANUAL
-5-
EBIT 50 SERVICE MANUAL
-6-
EBIT 50 SERVICE MANUAL
-7-
EBIT 50 SERVICE MANUAL
2.2.1
2.2.2
-8-
EBIT 50 SERVICE MANUAL
2.2.3
Remove the 4 screws marked in the picture and gently remove the side panels.
2.2.4
-9-
EBIT 50 SERVICE MANUAL
2.2.5
Remove the screws marked in the picture to disassemble the transducer board.
2.2.6
Remove the 13 screws marked in the picture to dismantle the metal baffle.
- 10 -
EBIT 50 SERVICE MANUAL
2.2.7
2.2.8
Remove the 4 screws marked in the picture to dismantle the OMAP board.
- 11 -
EBIT 50 SERVICE MANUAL
2.2.9
Remove the 8 screws marked in the picture to disassemble the cover and the power supply board.
2.2.10
- 12 -
EBIT 50 SERVICE MANUAL
2.3.1
Cut the ribbon and pull out the ribbon, then remove the ultrasound module to disassemble the
monitor module.
2.3.2
- 13 -
EBIT 50 SERVICE MANUAL
2.3.4
- 14 -
EBIT 50 SERVICE MANUAL
2.4.1
2.4.2
2.4.3
- 15 -
EBIT 50 SERVICE MANUAL
2.4.3
2.4.4
Remove the screws and you will get the keyboard PCB.
- 16 -
EBIT 50 SERVICE MANUAL
According to the most frequently occurred errors and system messages, the list of
possible causes and relevant solutions is attached as below:
1. Transmission
1. Adjust the transmission
frequency, gain or
frequency, gain or TGC
TGC control is not set
control.
Menu bar displays properly.
2.Ensure the probe is
3 on the screen but no 2. No probe is
connect correctly
scanning image connected or probe is
3. Defreeze the system
connected improperly
by pressing the FREEZE
The system is in
key.
frozen status
1.Examination mode 1. Set the exam mode
is not correct correctly.
Image quality is
4 2.Image processing 2.Adjust the setting of
abnormal
parameters are set image processing or set it
improperly to default setting
1. The probe is not
connected well with 1. Reconnect probe
Wrong probe type
5 probe connector. correctly.
show on screen.
2. Internal circuit 2. Restart the system.
protection
- 17 -
EBIT 50 SERVICE MANUAL
4.1.1
Find an empty USB disk, 2GB volume is easiest to be found by the main unit, change USB disk as
FAT32 and quick format.
4.1.2
Get the update file , the files are zipped and named as EBit system_update_ X.X.XXXX.XXXX-XX
Sometimes the name of the file will be marked EBit50. Don't worry, this file can be used normally.
- 18 -
EBIT 50 SERVICE MANUAL
4.1.3
Confirm when you open the folder filesys_update, there should be sub files in the folder.
4.1.4
Insert the U disk into the USB socket, Press SETUP key on the key on the keyboard, choose System
to get into System Settings interface.
Click the “update” item in system interface firstly and then click the “update” in the following
interface box .System will upgrade the software automatically.
- 19 -
EBIT 50 SERVICE MANUAL
4.1.5
If the upgrade is completed, please restart the machine, and the following dialog will pop up.
- 20 -
EBIT 50 SERVICE MANUAL
4.2.1
Please contact with CHISON Service Dept. Get the newest upgrade package. The package is
different with each unit, please provide the serial no. (SN) when you ask for the files. It includes two
files: fpga_update.rar and update_EBit-xxxxxxx (serial no.).zip
4.2.2
Unzip the files, we recommed “ WinRAR ” software tool to unzip. You will get two folders as below,
update_SN and fpga, copy them to the root directory of flash disk.
- 21 -
EBIT 50 SERVICE MANUAL
4.2.3
Please confirm when you open the folders, there should be subfile in the folder;
4.2.4
Insert the U disk into the USB socket, Press SETUP key on the key on the keyboard, choose System
to get into System Settings interface.
Caution:Never unplug the U disk or cut off the power supply during
upgrade, or the system will be breakdown, the software should be upgraded
firstly, hardware following.
4.2.4
Click the “update” item of software firstly and confirm “update”, the system will upgrade the
software automatically
- 22 -
EBIT 50 SERVICE MANUAL
4.2.5
Manually boot the machine, and you will see the following tips.
- 23 -
EBIT 50 SERVICE MANUAL
4.2.6
When the machine restart and get into the system interface, then click the “update” item of hardware
and confirm “update” , the system will upgrade the hardware automatically
4.2.7
Manually boot the machine, and you will see the following tips.
1) Please confirm the U disk is a good one, or you may try with another U disk.
2) The volume for U disk (2GB is the best), too small or too big volume is not
easy for machine to recognize.
3) Please confirm there will be subfiles once you open the upgrade folder or
fpga folder.
4) The upgrade package is damaged during the downloading or unzip, you can
try to download and unzip again.
5) Please confirm the USB disk is found.
- 24 -