EBIT 50 SERVICE MANUAL

SERVICE MANUAL
FOR

DIAGNOSTIC ULTRASOUND SYSTEM

MODEL EBIT 50

Digitally signed by Ermicev Alexandr


Date: 2019.10.16 09:08:47 EEST
Reason: MoldSign Signature
Location: Moldova

CHISON MEDICAL IMAGING CO., LTD.


We reserve the right to make changes to this manual without prior notice.

-1-
EBIT 50 SERVICE MANUAL

CONTEXT

CHAPTER 1 GENERAL INTRODUCTION ..................................................................- 3 -

1.1 Introduction............................................................................................................- 3 -
1.2 Service policy ........................................................................................................- 3 -
1.3 Service Procedure .................................................................................................- 3 -

CHAPTER 2 DISASSEMBLE MACHINE ....................................................................- 6 -

2.1 SYSTEM OVERVIEW ...........................................................................................- 6 -


2.2 Disassemble the ultrasound module and power module .......................................- 8 -
2.3 Disassemble the monitor module ........................................................................- 13 -
2.4 Disassemble the keyboard module .....................................................................- 15 -

CHAPTER3 TROUBLE SHOOTING .........................................................................- 17 -

CHAPTER4 YSTEM AND SOFTWARE UPGRADE ..................................................- 18 -

4.1.1. SYSTEM UPGRADE .......................................................................................- 18 -


4.2.SOFTWARE AND HARDWARE UPGRADE .......................................................- 21 -

-2-
EBIT 50 SERVICE MANUAL

Chapter 1 GENERAL INTRODUCTION

1.1 Introduction

This guide covers the preparation and maintenance of the ultrasound system, transducers, and
accessories. Refer to the manufacturers’ instructions for specific information about peripherals.
Please read the information in this guide before providing service to EBIT 50 color ultrasound
system. It applies to the ultrasound system, and transducers.

1.2 Service policy

CHISON provides life-time service for all produced ultrasound machines. Meanwhile, according to
the warranty card enclosed in the packing box, the corresponding free warranty period is provided
for the main unit and probe (transducer) from the date of Ex-factory.
CHISON Service Dept. provides efficient service solutions to customers timely, including technical
support, spare parts support, field service, upgrade, warranty extension, and on demand service.
The engineer must have ultrasound training before provide service, CHISON can provide varieties of
training, including operation, application maintenance, and troubleshooting.

1.3 Service Procedure

If you encounter difficulty with proving service to CHISON products, please refer to the user manual
and service manual firstly to correct the problem.
If the problem is still not solved there, please contact with CHISON Service Dept at the following
numbers or addresses:

CHISON website: www.chison.com.


Service Support: Tel: 0086-510-85310012
Fax: 0086-510-85310726
E-mail: [email protected]

-3-
EBIT 50 SERVICE MANUAL

Caution :Important information

1. It is the customer’s responsibility to maintain and manage the system after


delivery.
2. The warranty does not cover the following items, even during the warranty
period:
a) Damage or loss due to misuse or abuse with system and probes, for
example, drop the probe, the liquid or the metal part fall into the
system.
b) Damage or loss caused by Acts of God such as fires, earthquakes,
floods, lightning, etc.
c) Damage or loss caused by failure to meet the specified conditions
for this system, such as inadequate power supply, improper
installation or environmental conditions.
d) Damage or loss caused by non approved transportation by CHISON.
e) Damage or loss due to use the system outside the region where the
system was originally sold.
f) Damage or loss involving the system purchased from a source other
than CHISON or its authorized agents.
3. Do not make changes or modifications to the software or hardware of this
system and probes.
4. During operate the system, if user has any doubt, difficulty or any unclear,
please contact CHISON's authorized service engineer immediately. Please
describe the situation clearly to solve the question in time. Before solve
the question, please don’t operate the system.
5. This system shall not be used by persons other than fully qualified and
certified medical personnel.
6. The device can only be sold to qualified medical institutions or doctors.
The users shall fully understand and master the devices before operating.
The users shall have got the qualification, and shall comply with the local
laws and regulations, the local religion and customs, etc.
7. The System modified or repaired by people other than CHISON’s qualified
service engineers, CHISON shall not be liable for the system.
8. The purpose of this system is to provide physicians with data for clinical
diagnosis. It is the physician’s responsibility for diagnostic procedures.
CHISON shall not be liable for the results of diagnostic procedures
9. This manual contains warnings regarding foreseeable potential dangers,
but user shall always be alert to dangers other than those indicated as well.
CHISON shall not be liable for damage or loss that results from negligence
-4-
EBIT 50 SERVICE MANUAL

or from ignoring the precautions and operating instructions described in


this operation manual.
10. Due to negligence not following operation manual, CHISON shall not be
liable for the results.
11. Each time before and after ultrasound examination, please check the
probe surface, probe cable and sheath whether they are abnormal, such as
cracking, peeling and deformation. Also check whether the lens is strongly
fixed. Abnormal probes may cause electric shock and injure the patient.
Once any abnormal, user must stop using and contact CHISON's
authorized service engineer.
12. If the probe is dropped or scratched by hard part, please stop using the
probe immediately. And contact CHISON's authorized service engineer to
make sure the safety and effectiveness is in good condition before use.
13. If there is any liquid or metal to enter to the system, please power off the
system and stop using it immediately. Please first contact CHISON’s
authorized service engineer to make sure it’s safe before restart using it.
14. Please don't use solvents (such as paint thinner, benzine, or alcohol) or
abrasive cleansers for cleaning the system (including monitor and probes,
etc). It may corrode the system and probes.
15. The following environmental conditions are within system tolerances for
operation:
Temperature: 10ºC ~ 40ºC Relative Humidity: 30%~75%,
non-condensing

16. Important data must be backed up on external memory media. CHISON


shall not be liable for loss of data stored in the memory of this system
caused by operator error or accidents.
17. LCD display screen may have some dark or light dots, it is normal for the
LCD. It does not mean that LCD screen is defective.
18. Please put this service manual with the system to ensure operator and
manager can reach it at any time.
19. Please start to disassemble the system in 2 minutes after the system is
OFF.
20. While open the metal box, please note the sharp edge of metal part to
avoid to cut the finger.
21. Before disassembling any circuit board, the connection between the power
supply and the built-in battery must be disconnected.

-5-
EBIT 50 SERVICE MANUAL

CHAPTER 2 DISASSEMBLE MACHINE

2.1 SYSTEM OVERVIEW

-6-
EBIT 50 SERVICE MANUAL

-7-
EBIT 50 SERVICE MANUAL

2.2 Disassemble the ultrasound module and power module

2.2.1

Remove 2 screws from the battery cover.

2.2.2

Remove the 6 screws to remove the panel.

-8-
EBIT 50 SERVICE MANUAL

2.2.3

Remove the 4 screws marked in the picture and gently remove the side panels.

2.2.4

In order to disassemble back panel we need to remove the 4 screws.

-9-
EBIT 50 SERVICE MANUAL

2.2.5

Remove the screws marked in the picture to disassemble the transducer board.

2.2.6

Remove the 13 screws marked in the picture to dismantle the metal baffle.

- 10 -
EBIT 50 SERVICE MANUAL

2.2.7

To disassemble the board we need to remove the 3 screws.

2.2.8

Remove the 4 screws marked in the picture to dismantle the OMAP board.

- 11 -
EBIT 50 SERVICE MANUAL

2.2.9

Remove the 8 screws marked in the picture to disassemble the cover and the power supply board.

2.2.10

Remove the screws and the main board.

- 12 -
EBIT 50 SERVICE MANUAL

2.3 Disassemble the monitor module

2.3.1

Cut the ribbon and pull out the ribbon, then remove the ultrasound module to disassemble the
monitor module.

2.3.2

Remove the 8 screws.

- 13 -
EBIT 50 SERVICE MANUAL

2.3.4

Remove the screws to get the monitor itself.

- 14 -
EBIT 50 SERVICE MANUAL

2.4 Disassemble the keyboard module

2.4.1

Remove the screws on the keyboard.

2.4.2

Pull out the cables marked in the picture.

2.4.3

Pull out the cables from the speakers.

- 15 -
EBIT 50 SERVICE MANUAL

2.4.3

Remove the nocks marked in the picture.

2.4.4

Remove the screws and you will get the keyboard PCB.

- 16 -
EBIT 50 SERVICE MANUAL

CHAPTER3 TROUBLE SHOOTING

According to the most frequently occurred errors and system messages, the list of
possible causes and relevant solutions is attached as below:

No. Errors & Messages Possible Cause Solution


When turn on the
Battery failure, the Check the power cable
system,
1 adapter is not working connection to make sure
power-indicating
properly. it’s good.
lamp is not lit.
1. Wait for 1 minute to
1. The system is
restart the system after
When turn on the restarted too shortly
turning off.
system, power after it is turned off.
2. Check the LVDS
2 indicating lamp is lit, 2. the connection
cable connection to make
but no images on between the monitor
sure it’s connection is
the monitor. and DSC board
good.

1. Transmission
1. Adjust the transmission
frequency, gain or
frequency, gain or TGC
TGC control is not set
control.
Menu bar displays properly.
2.Ensure the probe is
3 on the screen but no 2. No probe is
connect correctly
scanning image connected or probe is
3. Defreeze the system
connected improperly
by pressing the FREEZE
The system is in
key.
frozen status
1.Examination mode 1. Set the exam mode
is not correct correctly.
Image quality is
4 2.Image processing 2.Adjust the setting of
abnormal
parameters are set image processing or set it
improperly to default setting
1. The probe is not
connected well with 1. Reconnect probe
Wrong probe type
5 probe connector. correctly.
show on screen.
2. Internal circuit 2. Restart the system.
protection

- 17 -
EBIT 50 SERVICE MANUAL

CHAPTER4 YSTEM AND SOFTWARE UPGRADE

4.1.1. SYSTEM UPGRADE

4.1.1

Find an empty USB disk, 2GB volume is easiest to be found by the main unit, change USB disk as
FAT32 and quick format.

4.1.2

Get the update file , the files are zipped and named as EBit system_update_ X.X.XXXX.XXXX-XX
Sometimes the name of the file will be marked EBit50. Don't worry, this file can be used normally.

- 18 -
EBIT 50 SERVICE MANUAL

4.1.3

Confirm when you open the folder filesys_update, there should be sub files in the folder.

Note: Confirm there should be uncovered sub files in the filesys_update


folder. Otherwise, system interface will prompt:No files can be found!!

4.1.4

Insert the U disk into the USB socket, Press SETUP key on the key on the keyboard, choose System
to get into System Settings interface.

Click the “update” item in system interface firstly and then click the “update” in the following
interface box .System will upgrade the software automatically.

- 19 -
EBIT 50 SERVICE MANUAL

4.1.5

If the upgrade is completed, please restart the machine, and the following dialog will pop up.

- 20 -
EBIT 50 SERVICE MANUAL

4.2.SOFTWARE AND HARDWARE UPGRADE

4.2.1

Please contact with CHISON Service Dept. Get the newest upgrade package. The package is
different with each unit, please provide the serial no. (SN) when you ask for the files. It includes two
files: fpga_update.rar and update_EBit-xxxxxxx (serial no.).zip

4.2.2

Unzip the files, we recommed “ WinRAR ” software tool to unzip. You will get two folders as below,
update_SN and fpga, copy them to the root directory of flash disk.

- 21 -
EBIT 50 SERVICE MANUAL

4.2.3

Please confirm when you open the folders, there should be subfile in the folder;

4.2.4

Insert the U disk into the USB socket, Press SETUP key on the key on the keyboard, choose System
to get into System Settings interface.

Caution:Never unplug the U disk or cut off the power supply during
upgrade, or the system will be breakdown, the software should be upgraded
firstly, hardware following.

4.2.4

Click the “update” item of software firstly and confirm “update”, the system will upgrade the
software automatically

- 22 -
EBIT 50 SERVICE MANUAL

4.2.5

Manually boot the machine, and you will see the following tips.

- 23 -
EBIT 50 SERVICE MANUAL

4.2.6

When the machine restart and get into the system interface, then click the “update” item of hardware
and confirm “update” , the system will upgrade the hardware automatically

4.2.7

Manually boot the machine, and you will see the following tips.

Caution:Solution for upgrade failure:

1) Please confirm the U disk is a good one, or you may try with another U disk.
2) The volume for U disk (2GB is the best), too small or too big volume is not
easy for machine to recognize.
3) Please confirm there will be subfiles once you open the upgrade folder or
fpga folder.
4) The upgrade package is damaged during the downloading or unzip, you can
try to download and unzip again.
5) Please confirm the USB disk is found.

- 24 -