BP701T Instrumental Method of Analysis
Unit IIlI:
rm
(Dr. S. K. Patro, Asst. Professor, IPT Salipur)
A. Chromatography
1. Introduction to Chromatography
ha
2. Column Chromatography
3. Thin layer Chromatography
4. Paper Chromatography
B. Electrophoresis
1. Paper Electrophoresis
lP
2. Gel electrophoresis
3. Capillary electrophoresis
l
we
re
Ca
a
A. Chromatography
Introduction:
rm
Chromatography was invented by the Russian botanist Mikhail Tswet in the
year 1903. He employed the technique to separate various plant pigments (i.e.
Chlorophylls and Xanthophylls) by passing solutions of these substances (in
petroleum-ether extract) through a glass column packed with finely powdered CaCO3.
The separated species appeared as separate bands having colored bands; the various
pigments migrating through the column at different rates (because of differences in
ha
their distribution ratios). The separated species appearedasseparate bands having
colored bands: the various pigments migrating through the column at different rates
(because of differences in their distribution of ratios). The various solutes were
isolated by cutting and sectioning of the chalk packing. Twswett choose to designate
the name of such a process of separation as chromatography (Chroma-color,
lP
graphein-writing). Tswett's original experiments remained unnoticed in the literature
for several decades. It was not until 1931 when kuhan and Lederer investigated
polyene pigments that interest in such a technique was renewed.
But later on a diversify group of techniques which allow the separation of
closely related components of the complex mixtures. In this technique, the sample is
l
a
Classification of Chromatographic methods:
rm
Type of
General Type of Stationary Mobile
equilibration Name of the Technique
Classification method Phase Phase
Process
Paper Chromatography
(PC)
Liquid- Liquid Partition Thin layer
Liquid
ha
Liquid Supported between the Chromatography
Chromatography Liquid
Or on a solid immiscible (TLC)
(LC)
Partition surface liquids High Performance thin
layer Chromatography
(HPTLC)
lP
Liquid-Solid,
Adsorption Column
Or Solid Liquid Adsorption
Chromatography (ACC)
adsorption
Adsorption
(using very
Very finely
Liquid-Solid divided much higher
l
adsorption packed in a
phase)
column
re
Ca
a
exchange Liquid Partition/Sieving chromatography
Exchange
resign) (IEC)
Group
rm
Specific liquid
Affinity
usually uses bonded to a
solid surface (an Partition between
enzymesS Liquid Affinity
antibody, surface liquia chromatograpny.
Or (immobilized)
immobilized on
Antigen-
a stationary and mobile phase
Antibody
phase by
ha
highly covalentldy
specific
bindingtoit
interactions an affinity
ligand)
Gas-solid
lP
Gas -
Solid chromatography
Adsorption
(or Gas (GSC)
Solid
adsorption)
Gas
Gas-Liquid
Chromatography |Liquid adsorbed Partition between
Gas -liquid Chromatography
on a solid Gas and liquid
or Gas (GLC)
l
Partition)
we
Super-critical
fluid
Supercritical Bio specific Organic Species Partition between
reritica
Supercritical chromatography
fluid adsorption or bonded to a Super-critical fluid |
fluid (SFC) or
chromatography bio affinity solid surface and bonded species
Bioaffinity
(SFC)
re
chromatography
(BC)
Ca
a
Column Chromatography
When a column of stationary phase is used, the technique is called as column chromatography.
Based on the nature of the stationary phase i.e. whether it is solid or liquid, it is called as column
rm
adsorption chromatography or. Column partition chromatography is not widely used.
Principle:
. This technique is based on the principle of differential adsorption where different molecules
in a mixture have different affinities with the adsorbent present in the stationary phase.
2. The molecules having higher affinity remain adsorbed for a longer time decreasing their
speed of movement through the column.
ha
3. However, the molecules with lower affinity move with a faster movement, thus allowing the
molecules to be separated in different fractions.
4. Here, the stationary phase in the column chronmatography also termed the adsorbent is a
solid (mostly silica) and the mobile phase is a liquid that allows the molecules to move
through the column smoothly. The type of interaction between the stationary phase
lP
(adsorbent) & the solute is reversible in nature.
The rate of movement of a component (R) is given as follows
R
Rate of movement of component
=
Rate of movement of mobile phase
The equation can be simplified as follows:
a
Stationary Phase
Adsorbents are used in this technique may be organic and inorganic classes of compounds. The
ideal requirements of adsorbent are:
rm
It should produce only adsorption of the analyte over it
ii. The particles should have uniform size distribution and have spherical shape. Particle
size: 60-200u.
ii. It should have high mechanical stability
iv. It should be inert & should not react with the solute or other components.
V. Insoluble in the solvents or mobile phases used.
Vi. It should be colorless to facilitate observations of zones and recovery of components.
ha
lP
mobile
phase
loaded
sample sample
separation stronger
interactions
stationary
phase -resolved-
weaker bands
interactions
l
we
fractions eluted
collection molecules
re
Figure: 1
Column chromatography.
a
Calcium carbonate Acetone
Medium
Calcium phosphate Benzene
rm
Magnesium carbonate Toluene
Magnesium oxide Esters
Calcium hydroxide Chloroform
Activated magnesium silicate Acetonitrile
Activated alumina Alcohols
Activated charcoal Water
Strong
ha
Activated magnesia Pyridine
Activated silica Organic acids
Mixtures of acids or bases
Fuller's earth with ethanol or pyridine
lP
The most commonly used adsorbent is Silica gel of 80-100 mesh or 100 -
200 mesh size
which has a particle size of 60-200u.
Selection of Stationary Phase
The selection of stationary phase in column chromatography depends on the following:
1. Removal of impurities: When a small quantity of impurity is present and there is
sufficient.
2. No. of components to be separated: When few components are to be separated, weak
we
a
Different mobile phases used: It is used in increasing order of polarity or elution strength. The
solvents are given in the above Table 1. These solvents can be used in either pure form or as a
mixture of solvents of varying compositions
rm
ColumnCharacteristies
Column is mostly best quality of neutral glass since it should not be affected by solvents, acids or
alkalies. An ordinary burette can be used as column for separation.
Length/diameter ratio is 10-15:1.
For more efficiency, the length/diameter ratio is 100:1.
Column length
ha
a. Multi-component system long column
b. Components with similar affinities for adsorbent long column
C. Components with different affinities for adsorbent short column
Preparation of the Column
The column mostly consists of a glass tube packed with a suitable stationary phase.
lP
Glass wool/cotton wool or an asbestos pad is placed at the botton of the column before
packing the stationary phase.
After packing, a paper disc kept on the top, so that the stationary layer is not disturbed
during the introduction of sample or mobile phase.
There are two types of preparing the column, they are:
1. Dry packing/ dry filling
l
In this the required quantity of adsorbent is poured as fine dry powder in the column and the solvent
is allowed to flow through the column till equilibrium is reached.
we
a
Eg. Use of chloroform alone or Pet.ether: Benzene = 1:l only, etc.
Gradient elution technique: Solvents of gradually 1 (increasing) polarity or t (increasing)
elution strength are used during the process of separation.
rm
E.g. initially benzene, then chloroform, then ethyl acetate then chloroform
Other techniques like Frontal analysis and Displacement analysis where a graph of concentration of
eluate Vs. volume of eluate will give an idea of how compounds are eluted out from the column.
D. Detection of Components
1. If the compounds separated in a column chromatography procedure are colored, the progress
of the separation can simply be monitored visually.
ha
2. If the compounds to be isolated from column chromatography are colorless. Then the
technique depends upon the properties of the components. Different properties which can be
used are
3. Absorption of light (UV/Vis) Using UV-Visible Spectropotometer
-
.
lP
By using flame ionization flame detector
. Refractive index detector- based on the refractive index difference between the mobile
phase and mobile phase + component
7. Evaporation of the solvent and weighing the residue
8. Small fractions of the eluent are collected sequentially in labeled tubes and the composition
of each fraction is analyzed by TLC (thin layer chromatography).
l
Recovery of components: Earlier, recoveries of the components were done by cutting the column
into several distinct zones. Later, extrusions of the column into zones were done by using plunger.
we
The best technique is to recover the components by a process called aselution. The components are
called as eluate, the solvent called as eluent and the process of removing the components from the
column is called as elution. The different elution techniques like isocratic elution technique and
gradient elution technique.Recovery is done by collecting different fractions of mobile phase of
equal volume like 1Oml, 20ml etc or unequal volume. They can also be collected time wise i.e. a
fraction every 10 or 20 minutes etc. The recovered fractions are detected by using the techniques
re
discussed earlier. Similar fractions are mixed so that the bulk of the compound of each type is
obtained in a pure form. If a fraction still contains several components, it can be resolved by using
another column.
Applications:
1. Separation of mixture of compounds: Separation of glycosides, amino acids, plant extracts
Ca
2. Removal of impurities Isolation of the active constituents from the plant extract or from
formulations
3. Isolation of metabolites from the biological fluids: 17-ketosteroids from urine, cortisol
4. Estimation of drugs in formulations or crude extracts
i. Determination of % w/w of strychnine in syrup of ferrous phosphate
with quinine and strychnine
a
ii. Separation of diastereomers.
iii. Separation of tautomers and racemates
Factor affecting Column efficiency
rm
1. Dimensions of the column
2. Particle size of the adsorbent
3. Nature of the solvent
4. Temperature of the column
5. Pressure
ha
Advantages:
1. Any type of mixture can be separated by column chromatography.
2. Any quantity of the mixture can also be separated (pg to mg of substance).
3. Wider choice of mobile phase.
lP
4. In preparative type, the sample can be separated and reused.
5. Automation is possible.
Limitation or Disadvantages of Column chromatography
. Time consuming method
2. More amounts of solvents are required which may be expensive.
3. Automation makes the technique more complicated and costly.
l
we
re
Ca
10
a
Thin layer Chromatography (TLC)
Introduction:The history of thin layer chromatography dates back to 1938 when Izmailov and
rm
Shraiber separated plant extracts using 2mm thick and firm layer of alumina set on glass plate. In
1944, Consden, Goden and Martin used filter papers for separating amino acids. In 1950, Kirchner
identified terpens on filter paper and later glass fibre paper coated with alumina. Only in 1958,
Stahl developed standard equipment for analyzing by Thin layer chromatography.
Principle:
Thin Layer Chromatography can be defined as a method of separation or identification of a mixture
of components into individual components by using finely divided adsorbent coated or spread over
ha
a chromatographic plate. The mobile phase solvent flows through because of capillary action
(against gravitational force). The components move according to their affinities towards the
adsorbent. The component with more affinity towards the stationary phase travels slower. The
component with lesser affinity towards the stationary phase travels faster. Thus the components are
separated on a thin layer chromatographic plate based on the affinity of the components towards the
lP
stationary phase.
OH QH OH OH OH
l
Silica (Si02) is a solid with an extended structure of tetrahedral silica atoms bridged together by
bent oxygen atoms. On the surface of the silica particles, the solid terminates in very polar silanol
(Si-0-H) groups. The silica is the stationary phase because it remains adhered to the glass plate and
does not move during the chromatography process.
The Silica extended structure and surface is shown in the Fig 2. The developed TLC plate is shown
in the Fig 3.
re
SOLVENT FRONT
.-CENTER OF SPOT
Ca
11
a
Advantages of TLC
1. It is a simple process with a short development time.
2. It helps with the visualization of separated compound spots easily.
rm
3. It helps in isolating of most of the compounds.
4. The separation process is faster and the selectivity for compounds is higher (even small
differences in chemistry is enough for clear separation).
5. The purity standards of the given sample can be assessed easily.
6. It is a cheaper chromatographic technique.
7. TLC offers a faster and more efficient separation than paper chromatography and the
majority of paper chromatographic separations have now been superseded by the TLC
ha
Procedures.
Practical Requirement:
1. Stationary Phase: There a several adsorbents which can be used as stationary phases. Some
of the stationary phases, their composition and the ratio in which they have to be mixed with
water or other solvents to form a slurry for preparing thin layer chromatographic plates are
lP
given in the below Table 2:
Name Composition Adsorbent: Water ratio
SilicagelH Silicagel without binder 1:1.5
Silicagel GF | Silicagel + Binder + Fluorescent indicator 1:2
Silicagel G Silicagel +CaSOa(gypsum) 1:2
l
Alumina 1:1.1
Neutral Al2O3 without binder
we
Basic
Acidic
AlLO G AlO3 +binder 1:2
powder 2:3)
12
a
Fluoroscent indicator Zine silicate
Silica gel and alumina are available with different specific surface areas and these grades are
identified by a number, e.g., silica gel 60 (or 40 or 150) which indicates the mean pore size in
rm
Angstroms (10-m).The particle size of silica gel for TLC is 10-40 um (average 15 um).
2. Preparation of the Glass Plates: The sizes of the glass plates for use with commercially
available spreaders are usually 20 X 20, 20 X 10 or 20 X 5 cm.
Microscopic slides can also be used for some applications like monitoring the progress of
chemical reaction.
In general, the glass plates should be of good quality and should be withstand temperatures
used for drying the plates.
ha
General method:
Mix 30gm of the adsorbent in a mortar to a smooth consistence with the requisite amount of
water or solvent specified in the manufacturer's instruction and transfer the slurry quickly to
the spreader. Spread the mixture over 4 to 5 plates (20 X 20cm) or a proportionately larger
number of smaller plates and allow the thin layers to set (about 4minutes when CaS04 is
lP
present). Transfer the plates carefully to a suitable holder and after a further 30minutes, dry
at 100-120 °C for 1 hour to activate the adsorbent. Cool and store the plates in a desiccator
over silica gel. The thickness of the moist thin layer should be about 0.25 mm.
Special methods:
a. Preparative thin layer. The layers are 0.5 -
2mm thick, prepared as described under the
general method, but using a smaller quantity of water and allowing a longer time for the
initial drying of the plate.
l
b. Mieroscopic slides are conveniently coated by a dipping technique in the following way:
we
using a capillary tube or micropipette. The spots can be placed at random or equidistant
from each other by using a template, with markings. The spot should be kept at least 2cm
above the base of the plate and the spotting area should not be immersed in the mobile phase
in the development tank.
13
a
4. Development Tank: For the purpose of development, a developing tank or chamber of
different sizes to hold TLC plates of standard dimensions are used. These require more
solvents for developing the chromatogram. When a new method is developed, it is better to
rm
develop in glass beakers or specimen jars, etc, to avoid more wastage of solvents. When
developed method or standard method is used, it is better to use development tank. In the
new type of development tanks have hump in the middlle, which require less solvent. The
development chamber or tank should be lined inside with filter paper moistened with the
mobile phase so as to saturate the atmosphere. If this kind of saturation of the atmosphere is
not done, "edge effect" occurs where the solvent front in the middle of the TLC plate moves
faster than that of the edge. The development tank examples are shown in the below Fig 3.
ha
Old type New type
Glass
TLC plates plates
lP
Mobile phase Hump
solvent
EES Mobile
Fig 3: Development tank
5. Mobile Phase:Selection of the mobile phase depends upon the below factors
. Nature of the substances to be separated
l
Pure solvents or mixture of solvents are used. The following gives a list of solvents (of
increasing polarity).
Petroleum ether, Carbon tetrachloride, Cyclohexane, Carbon di-sulphide, Ether, Acetone,
Benzene, Toluene, Ethyl acetate, Chloroform, Alcohols like methanol or ethanol, Water,
re
pyridine. The solvent composition is done by trial and error method only but with a review of
literature and other logical considerations like solubility of the substance, polar or non-polar
character of the samples, etec.
6. Development technique: Different development techniques are used for efficient
separations. They are
Ca
14
a
i. Vertical development (One dimensional): In this technique, the plates are
kept vertical and the solvent flows against gravity, because of capillary
action.
rm
Two dimensional techniques: For complex mixtures this technique is used.
First, the plates are developed in one axis and the plates after drying are
developed in the other axis. When large number compounds cannot be
separated by using one dimensional technique.
7. Detecting or Visualizing Agents
After the development of TLC plates, the spots should be visualized. Detecting colored
spots can be done visually. But for detecting colorless spots, any one of the following
ha
techniques can be used.
a. Specific methods: In this method particular detecting agents are used to find out the
nature of compounds or for identification purposes. Examples are
Ferric chloride- for phenolic compounds and tannins.
1. Ninhydrin in acetone- for amino acids
ii. Dragendroff's reagent- for alkaloids
lP
iv. 2,4- Dinitrophenyl hydrazine - for aldehydes and ketones
b. Nonspecific methods: Where the number of spots can be detected, but not the exact
nature or type of compound.
Examples
1. lodine chamber method: Where brown or amber spots are observed when the
TLC plates are kept in a tank with few iodine crystals at the bottom.
l
1. Sulphuric acid spray reagent: 70-80% v/v of sulphuric acid with few mg of either
potassium dichromate or potassium permanganate or few ml of nitric acid as
we
oxidizing agent is used. This reagent after spraying on TLC plates is heated in an
oven. Black spots are seen due to charring of compounds.
11i. Using fluorescent stationary phase: When the compound s are not flourscent, a
fluorescent stationary phase is used. When the plates are viewed under UV
chamber, dark spots are seen on a fluorescent background. Examples of such
stationary phase is Silica gel GF
re
detecting techniques are used in TLC method development and in preparative TLC.
In densitometric method, Densitometer is used which measures quantitatively the density of the
spots. When the optical density of the spots for the standards and test solution are measured, the
quantity of the substance can be calculated. The plates are neither destroyed nor eluted with the
solvents to get the compounds. This method is also called as in-situmethod.
15
a
8. Qualitative analysis
The Rf value is calculated for identifying the spots in qualitative analysis. Rf value is the
ratio of distance travelled by the solute to the distance travelled by the solvent front.
rm
Distance travelled by solute
Rf =
Distance travelled by the solvent front
The Rf value ranges from 0 to 1. The ideal value is 0.3 to 0.8.
The Rf value is constant for every compound in a particular combination of stationary and
mobile phase. When the Rf value of a sample and reference compound is same, the
compound is identified by its standard. When the Rf value differs, the compound may be
different from its reference standard.
ha
Rx value is the ratio of distance travelled by the sample and the distance travelled by the
standard. Rx value is always closer to 1.
Rm value used to find out whether the compounds belong to a homologous series. If they
i
belong to a homologous series, the ARm values are constant. The ARm values for a pair of
adjacent member of a homologous series are determined by using the formula:
lP
=
Rm log
9. Quantitative Analysis
Indirect method: Quantitative analysis can be done after eluting the individual spotswith
solvent and filtering off the stationary phase. The solution can be concentrated and the exact
quantities of the conmpound determined by the methods like UV-Visible spectrophotometry,
fluorescence method, flame photometric method, electrochemical methods of analysis.
l
Direct method: It can be done after eluting the individual spots with solvent and filtering
off the stationary phase. The solution can be the concentrated and the exact amount of the
we
potassium iodide
Ampicillin for oral Cellulose M.N-300 Citric acid Butyl Starch iodide reagent
:
suspension alcohol(5:)
16
a
Test for impurities, decomposition & related substances in pharmaceutical products (as per
British Pharmacopoeia substances and preparations)
Substance Tested for Mobile Phase Detection
rm
Chlorpropamide p-Chlorobenzene Chloroform Sodium
sulphonamide and methanol:cyclohexane:13.5M hypochlorite
|
ha
Nitrazepam Tablets | Decomposition and Nitromethane ethyl 254nm radiation
related substances acetate(85:15)
0.5%
Desipramine Iminodibenzyl Toluene: ethyl acetate: Potassium
Hydrochloride (0.2%) ethanol: diethylamine dichromate
lP
(20:20:4:1) (0.5%) in
sulphuric
acid:water(4:4)
l
we
re
Ca
17
a
Paper Chromatography
Paper chromatography (PC) is a type of a planar chromatography whereby chromatography
procedures are run on a specialized paper.It is considered to be the simplest and most widely used
rm
of the chromatographic techniques because of its applicability to isolation, identification and
quantitative determination of organic and inorganic compounds.It was first introduced by German
scientist Christian Friedrich Schonbein (1865).
Types of Paper chromatography:
) Paper Adsorption Chromatography: Paper impregnated with silica or alumina acts as
adsorbent (stationary phase) and solvent as mobile phase.
ha
(ii) Paper Partition Chromatography: Moisture/ Water present in the pores of cellulose fibers
present in filter paper acts as stationary phase & another mobile phase is used as solvent In
general paper chromatography mostly refers to paper partition chromatography.
Principle of Separation
The principle of separation is mainly partition rather than adsorption. Substances are distributed
between a stationary phase and mobile phase. Cellulose layers in filter paper contain moisture
lP
which acts as stationary phase. Organic solvents/buffers are used as mobile phase. The developing
solution travels up the stationary phase carrying the sample with it. Components of the sample will
separate readily according to how strongly they adsorb onto the stationary phase versus how readily
they dissolve in the mobile phase.
Instrumentation of Paper chromatography
2. Stationary phase & papers used
l
3. Mobile phase
4. Application of sample
we
5. Developing Chamber
6. Detecting or Visualizing agents
1.STATIONARY PHASE AND PAPERS: Whatmann filter papers of different grades like No.1,
No.2, No.3, No.4, No.17, No.20 ete are used.
In general the paper contains 98-99% of a-cellulose, 0.3 1% B
-
-cellulose. These papers differ in
sizes, shapes, porosities and thickness.
re
Other modified papers like Acid or base washed filter paper, glass fiber type paper.
Hydrophilic Papers Papers modified with methanol, formamide, glycol, glycerol etc.
-
Hydrophobie papers acetylation of OH groups leads to hydrophobic nature, hence can be used
-
for reverse phase chromatography. Silicon pretreatment and organic non-polar polymers can also be
Ca
18
a
2.Application of sample: The sample to be applied is dissolved in the mobile phase and applied
using capillary tube or using micropipette. Very low concentration is used to avoid larger zone
3. PAPER CHROMATOGRAPHY MOBILE PHASE
rm
Pure solvents, buffer solutions or mixture of solvents can be used. Some of the Examples of
Hydrophilic mobile phases
Isopropanol: ammonia:water 9:1:2
Methanol: water 4:1 or 3:1
n-Butanol: glacial acetic acid: water 4:1:5
Hydrophobic mobile phases
ha
kerosene: 70% isopropanol
Dimethyl ether: cyclohexane
The commonly employed solvents are the polar solvents, but the choice depends on the nature of
the substance to be separated.
If pure solvents do not give satisfactory separation, a mixture of solvents of suitable polarity may be
lP
applied.
4.CHROMATOGRAPHIC CHAMBER: The chromatographic chambers are made up of many
materials like glass, plastic or stainless steel. Glass tanks are preferred most. They are available in
various dimensional sizes depending upon paper length and development type. The chamber
atmosphere should be saturated with solvent vapor.
Development technique:
l
Sample loaded filter paper is dipped carefully into the solvent not more than a height of I cm and
waited until the solvent front reaches near the edge of the paper.
we
Hangers
Solvent
front
Paper Spacers
Paper
stnp
solvent
19
a
b. DESCENDING TYPE: This is carried out in a special chamber where the solvent
holder is at the top.The spot is kept at the top and the solvent flows down the
paper.In this method solvent moves from top to bottom so it is called descending
rm
chromatography (Fig: 5).
Through for
solvent
s
Anti-siphon rod
NOZzle
Wires fixed to sides
9 Rubber
of tank on which bulb
trough can restmie Jno bat
ha
Paper Glass tank 8Side
Visualisation-
Serrated edge reagent
lP
Fig 5: Descending technique and sprayer to spray the visualizing agent
c. ASCENDING DESCENDING DEVELOPMENT: A hybrid of above two
-
Cirular Paper
l
we
Wick
Solvent front.
re
spreads in all directions uniformly. Hence the individual spots after development look like
concentric circles. By making perforations radially, number of quadrants can be created allowing
more number of samples to be spotted (Fig 6).
e. Two dimensional developments: This technique is very similar to 2-Dimensional
TLC. Here the chromatogram development occurs in two directions at right angles
(Fig 7). In this mode, the samples are spotted to one corner of rectangular paper and
20
a
allowed for first development. Then the paper is again immersed in the mobile phase
at a right angle to the previous development for the second chromatogram. In the
second direction, either the same solvent system or different solvent system can be
rm
used for development.
Original
Spot
ha D
a. Nonspecific methods: where brown or amber of spots can be detected, but not the
exact nature or type of the compound.
we
Examples
(i) lodine chamber method where brown or amber spots are observed when
the developed papers are kept in atank with few iodine crystals at the
bottom.
ii) UV chamber for fluorescent compounds: When compounds are viewed
under UV chamber, at 254nm (short ) or at 365nm (long ), fluorescent
re
compounds can be detected. Bright spots can are seen against a dark
background.
b. Specific methods: Specific spray reagents or detecting or visualizing agents are used
to find out the nature of compounds or identification purposes.
a. Ferric chloride- For phenolic compounds and tannins
Ca
21
a
The detecting techniques can also be categorized as
a.Destructive technique: Specific spray reagents ete where the samples are
destroyed before detection e.g. Ninhydrin reagent.
rm
b. Non-destructive technique: UV chamber method, lodine chamber method,
densitometric method, e.t.c, where the sample is not destroyed even after
detection.
For radioactive materials, detection is by using autoradiographyor Geiger muller
counter.
For antibiotics, the chromatogram is layed on nutrient agar inoculated with
appropriate strain and the zone of inhibition is compared.
ha
Quantitative Analysis
Direct technique: Densitometer is an instrument whichmeasures quantitatively the density of the
spots. When the optical densities of the spots for the standard and test solution are determined, the
quantity of the substance can be calculated. The papers are neither destroyed nor eluted with
solvents to get the compounds. The method is also known as in-situ method.
lP
Indirect techniques: In this technique, the spots are cut into portions and eluted with solvents. The
solution can be analyzedby any conventional techniques of analysis like spectrophotometry,
electrochemical methods, etc.
Qualitative Analysis: a. Rf value
Distance travelled by solute
Rf
Distance traveled by solvent front
l
The Rf value ranges from 0 to But the ideal values are from 0.3 to 0.8.
1.
Rm =
log(p1)
Application:
) To check the control of purity of pharmaceuticals,
ii) For detection of impurities
Ca
22
a
(ii) Detect the contaminants in foods and drinks,
(iv) For the detection of drugs
Drug Mobile phase Detecting agent
rm
Gentamycin Chloroform: Methanol:| Ninhydrin in pyridine-
Ammonia: Water acetone mixture
(10:5:3:2)
Vancomycin t-Amylalcohol: Nutrient agar containing
Acetone: water (2:1:2) Bacilus subtilis
ha
(v) In analysis of cosmetics
(vi) Analysis of the reaction mixtures in biochemical labs.
(vii) Identification of decomposition products
(vii) Analysis of metabolites of drugs in blood, urine etc.
(ix) In the study of ripening and fermentation
lP
Advantages of Paper Chromatography:
1. Simple and Rapid
5. Paper chromatography does not occupy much space compared to other analytical methods
or equipment's.
we
23
a
B. ELECTROPHORESIS
Electrophoresis is a physical method of analysis based on themigration of electrically charged
rm
proteins, colloids, moleculesor other particles dissolved or dispersed in an electrolytesolution in the
direction of the electrode bearing the oppositepolarity when an electric current is passed through it.
The electrophoretic mobility is the rate of movement in metre per second of the charged particles
under the action of an electric field of I volt per metre and is expressed in square metres per volt
second. For practical reasons it is given in square centimetres per volt second, cmV "S". The
mobility is specific for a given electrolyte under precisely determined operational conditions.
ha
6Ttrn
Where = Electrophoretic mobility
Q-Net charge on the ion
r- lonic radius of the solute
n-viscosity of the medium
lP
The electrophoretic mobility is directly proportional to net charge and inversely proportional to
molecular size and viscosity of the electrophoresis medium.
The pH of the solution affects the mobility of the ion by
Depending on the method used, the electrophoretic mobility is either measured directly or
compared with that of a reference substance.
Based upon the type of apparatus used, electrophoretic methods may be divided into two categories,
l
one called free or moving boundary and the other called zone electrophoresis (using a supporting
medium). Zone electrophoresis include paper, gel such as agar, starch or poly acrylamide.
we
re
Ca
24
a
Paper electrophoresis:
It is separation technique, where ions of different charges are separated on a medium of paper
rm
(moistened with a buffer), by the application of a voltage between two electrodes, which are in
contact with the paper.
Principle of Separation: A mixture of ions or ionisable substances is applied on the centre of a
paper, previously immersed in a buffer of known ionic strength. This paper is placed across two
trays, filled with buffer, into which two electrodes are immersed. When a voltage is applied across
these electrodes, the ions or ionisable substances migrate towards anode or cathode, based on their
charge and other factors. Neutral or non-ionisable substances do not migrate. Anionic substances
ha
move towards anode and cations move towards cathode. Ultinmately, there is separation of anionic,
cationic and non-ionie / zwitterionic substances. The spots/bands which migrate can be detected bu
using appropriate spray reagents or visualizing agents as in paper chromatofraphy and can be
quantified by using densitometer. Both qualitative and quantitative analysis can be performed in
paper electrophoresis.
Components of paper electrophoresis
lP
a. Normally Whatman® filter paper (Grade 3 MM or No. 1) of suitable dimension (2.5cm to
Scm) with a length so that ends of the strip of paper touch the buffer solution, kept in the
electrode vessels. The paper to be used is washed with double distilled water followed by
0.1 M HCI or 0.01 M EDTA to remove impurities.
b. Electrodes and voltage to be applied
The electrode in the form of a thin wire is made up of carbon or platinum. A DC voltage of
l
about 8-15 V/cm length of paper is normally applied. In low voltage electrophoresis, the
voltage across two electrodes is about 100-300V, with a current of 0.4mAmp per cm width
we
or 1.5mAmp/strip.
In high voltage electrophoresis, a potential of about 50-215V/cm(Total 10,000V/strip) is
applied across the electrodes.
C. Buffers used: The pH of buffer to be used depends upon the types of compounds to be
separated.
The following are some of the buffers used:
re
25
a
Horizontal and vertical modes are used in analytical scale; whereas continuous
electrophoresis is used on a preparative scale (i.e. large amount of sample mixture is
used). The principles involved in all the modes are same, but the design of each
rm
instrument varies.
In Horizontal type / Vertical (Fig 1 and 2), buffer solution of known pH and ionic
strength is filled into two troughs. Appropriate grade of Whatmann filter paper and
suitable width and length of filter paper are immersed in buffer solution. 10-20ul of
sample solution is applied at the centre of the paper and fixed in position.
The transparent lid is closed for safety as well as to prevent evaporation of
buffer/solvent. A suitable potential (100-300V) is applied across two electrodes dipped
ha
in buffer solution.
Fig 1: Horizontal Paper electrophoresis
Sample application Paper strip
Transparent insulating
Lid
(for safely)
we
Sample application
Supporting rod or string
Paper
Electrodes-
re
BuferBuffersolution-
Ca
26
a
In this vertical mode, the migrations of ions are assisted by gravity and hence a typical
separation takes place in about 6-8 hours. After sufficient migration, the paper is taken
out and dried, to fix the spots bands. Then the compounds / bands /spots can be
rm
visualized by using the visualizing agent. The quantitation of spots can be done by
densitometer.
The horizontal mode is similar to the vertical mode, in principle. However, the paper
is placed on a flat bed, as shown in Fig 2. The procedure to be followed is same as that
of vertical type. In horizontal mode, it takes about 12-14 hours for separation.
Continuous electrophoresis (Fig. 3) is meant for preparative samples, where a
predetermined sample volume through a valve device is applied continuously on the
ha
centre of paper. The application of voltage causes migration of samples and hence
compounds are separated as bands. Thus each band is made to fall down and pure
compounds are collected in separate containers. The solvent is evaporated and pure
fractions are reused.
Various factors like charge of ions, size of the ions, viscosity of the medium, applied
voltage, pH of buffer and ionic strength affect the migration of ions in paper
lP
electrophoresis
aU WIl
DC Supply in Fig
Sample solution
or separation
l
Valve
we
Buffer solution
Buffer solution
re
Paper
Ca
Compound(s)
or Band
Collection vials
27
a
Fig 3: Continuous paper electrophoresis
rm
B. There are two types of paper electrophoresis based on the voltage applied, i.e. Low
voltage or Hligh voltage Paper electrophoresis
Voltage/em Votage/strip
Low voltage PE 8-15 V/ cm 100-300
Highvoltage PE 50-215V/ cm 10,000v
ha
High voltage paper Electrophoresis has the following advantages:
1. Separation is faster, hence less time is required for separation
2. Sharp bands are obtained, since there is less diffusion of bands.
3. As sharp bands are obtained, separation of closely related compounds can be
achieved.
lP
4. More number of samples can be analysed simultaneously.
Disadvantages of High Voltage Paper Electrophoresis
1. It is dangerous to the operator, since high voltage is applied.
2. More heating effects are seen because of high voltage and the paper becomes
dry
So in most of the laboratories, low voltage paper electrophoresis is used.
l
2. Less expenditure
3. Number of samples can be separated on a sample paper, at a time.
4. Wide variety of ionisable substances such as amino acids, proteins and peptides,
antibiotics, alkaloids etc., can be separated.
Disadvantages
1. The time required for separation is more, i.e. 6-8 hours in vertical mode and 12-
re
28
a
3. Separation of lipoproteins in serum (in case of hyperlipidemia)
4. Separations of enzymes in blood.
5. Separation of alkaloids and antibiotics in different samples can be carried out.
rm
GEL ELECTROPHORESIS
It is a separation technique. Gel is used as medium. The gel may be agar or agarose gel or
polyacrylamide gel.
The device consists essentially of a glass plate over the whole surface ofwhich is deposited a
firmly adhering layer of gel of uniform thickness. The connection between the gel and the
ha
conducting solution is effected in various ways according to the type of apparatus used. Precautions
are to be takento avoid condensation of moisture or drying of the solid layer.
Vertical Gel Instrument- The schematic diagram of a vertical gel electrophoresis apparatus is
given in Figure 4. It has two buffer chambers, upper chamber and a lower chamber. Both chambers
are fitted with the platinum electrodes connected to the external power supply from a power pack
which supplies a direct current or DC voltage. The upper and lower tank filled with the running
lP
buffer is connected by the electrophoresis gel casted in between two glass plates (rectangular and
notched). 'There are additional accessories needed for casting the polyacrylamide gel such as comb
(to prepare different well), spacer, gel caster etc.
l
we
Power supply
Gel cassette
Electrode
chamber
re
Comb
Tank
Ca
29
a
Casting of the gel: The acrylamide solution (a mixture of monomeric acrylamide anda bi-
functional eross linker bis-acrylamide) is mixed with the TEMED and APS and poured in between
the glass plate fitted into the gel caster. What is the mechanism of acrylamide polymerization?
rm
Ammoinum persulfate in the presence of TEMED forms oxygen free radicals and induces the
polymerization of acrylamide monomer to form a linear polymer. These linear monomers are
interconnected by the cross linking with bis-acrylamide monomer to form a 3-D mesh with pores.
The size of pore is controlled by the concentration of acrylamide and amount of bis-acrylamidein
the gel. n a vertical gel electrophoresis system, we cast two types of gels, stacking gel and
resolving gel. First the resolving gel solution is prepared and poured into the gel cassette for
polymerization. A thin layer of organie solvent (such as butanol or isopropanol) is layered to stop
ha
the entry of oxygen (oxygen neutralizes the free radical and slow down the polymerization) and make
the top layer smooth. After polymerization of the resolving gel, a stacking gel is poured and comb is
fitted intothe gel for construction of different lanes for the samples. Different steps involves the vertical
gel electrophoresis is shown in the below Fig 5.
lP
cathode buffer Anode buffer
Gel cassette
l
we
Electrode
chamber
sample
re
Tank
Separate protein
bands
Ca
Running of the gel: The sample is prepared in the loading dye containing SDS, B-mercaptoethanol
in glycerol to denature the sample and presence of glycerol facilitates the loading of sample in the
well. As the samples are filled vertically there is a distance drift between the molecules at the top
30
a
Vs at the bottom in a lane. This problem is taken care once the samples run through the stacking
gel. The pH of the stacking gel is 6.8 and at this pH, glycine is moving slowly in the front whereas
Tris-HCl is moving fast. As a result, the sample gets sandwiched between glycine-Tris and get
rm
stacked in the form of thin band. As the sample enters into the resolving gel with a pH 8.8, the
glycine is now charged, it moves fast and now sample runs as per their molecular weight (dueto
SDS they have equal negative charge). After tracking dye reaches to the bottom of the gel, gel is
taken out from the glass plate with the help of a spatula. Gel is stained with coomassie brilliant blue
R250 dye. The dye stains protein present on the gel. A typical SDS-PAGE pattern is given in the
Fig. 6.
ha
Potentials of discontinuous PAGE:
1. Number of disulfide bonds: Comparison of reducing and non-reducing denaturing gels can be used to
provide information related to the number of disulfide bonds present in the protein.
2. Seperating Proteins based on size alone: In the presence of SDS and reducing environment, PAGE
gel resolves two proteins of on the basis of molecular masses and the concentration of gel concentration.
lP
In SDS-PAGE, the relative mobility and the log molecular weight as given by
v=Vo4logM
Molecular weight of a protein can be determined by plotting relative migration Rf with the log
molecular weight of standard protein.
migration of protein from the lane
Rf= migration of tracking dye
l
we
.8
re
DIstance
avelled
y dye
U.
02 0.3 04
0s 0.8 O8
leg moleular weight(kD)
Ca
31
a
Calculation of molecular weight of the unknown protein sample is a 5 step process:
1. Resolve the protein sample on the SDS-PAGE along with the molecular weight markers.
rm
2. Calculate the relative mobility (Rf) using the following formula:
Rf-migrationofproteinfromthelane/migrationoftrackingdye
3. Plotlog molecular mass (y-axis) versus relative mobility (x-axis) of the standards.
4. Perform a linear regression using a calculator or using regression software such as
Microsoft Excel.
ha
5. Use the linear regression equation (Y = mx + c) to estimate the mass of the unknown
protein.
Log Molecular Weight = (slope) (mobility of the unknown) + Y intercept
Buffer and reagent for electrophoresis- The different buffer and reagents with their purpose for
lP
vertical gel electrophoresis is as follows-
1. N, N, N', N'-tetramethylethylenediamine (TEMED)-it catalyzes the acrylamide
polymerization.
2. Ammonium Persulfate (APS)-it is an initiator for the acrylamide polymerization.
3. Tris-HCl- it is the component of running and gel casting buffer.
4. Glycine-it is the component of running buffer
l
5. Bromophenol blue- it is the tracking dye to monitor the progress of gel electrophoresis.
6. Coomassie brilliant blue R250-it is used to stain the polyacrylamide gel.
we
7. Sodium dodecyl sulphate-it is used to denature and providenegative charge to the protein.
8. Acrylamide- monomeric unit used to prepare the gel.
9. Bis-acrylamide- eross linker for polymerization of acrylamide monomer to form gel.
re
Ca
32
a
Capillary Electrophoresis
Capillary electrophoresis employing a narrow bore fused quartz silica capillary tube usually
rm
50-75cm long with an i.d. of 25-100um (and an o.d. of 400 um) containing an appropriate
electrolyte using a direct current (DC) high voltage source, capable of producing a current of 250
uA at voltage ranging from 1000 to 30,000volts and on-line detector that similar to those HPLC are
involved (high voltage electrophoresis).
A cross -sectional view of such a capillary is shown in the Fig 7. The capillary is protected with an
outer layer of a polyimide (polymer of imide monomer)
ha O -Capillary
opening
lP
Polyimide
Coating Fused silica
Fig 7: A cross -
sectional view of such a capillary is shown in the
Integrator or
l
Capillary
Computer
we
Detector
Anode- Cathode
re
Buffer Buffer
Source Vial Sample Vial Destination Vial
High Voltage
Power Supply
components are a sample vial, source and destination vials, a capillary, electrodes, a high voltage
power supply, a detector, and a data output and handling device. The source vial, destination vial
and capillary are filled with an electrolyte such as an aqueous buffer solution. To introduce the
sample, the capillary inlet is placed into a vial containing the sample. Sample is introduced into the
capillary via capillary action, pressure, siphoning, or electrokinetically, and the capillary is then
33
a
returned to the source vial. The migration of the analytes is initiated by an electric field that is
applied between the source and destination vials and is supplied to the electrodes by the high-
voltage power supply. In the most common mode of CE, all ions, positive or negative, are pulled
rm
through the capillary in the same direction by electroosmotic flow (EOF) (Fig 9). The analytes
separate as they migrate due to their electrophoretie mobility, and are detected near the outlet end of
the capillary. The output of the detector is sent to a data output and handling device such as an
integrator or computer. The data is then displayed as an electropherogram, which reports detector
response as a function of time. Separated chemical compounds appear as peaks with different
retention times in an electropherogram.
ha
Electrophoretic
migration
Double layer
S®%0®
Sio Sio Sio Sio Sio Sio Sio
Sio Sio
Capillary
surtace
l
a
utilizes the small volumes. One example of representative Electropherogram which was obtained
from the seized heroin sample is showing in the Fig 10.
rm
80
2)
60- ISTD
40
N
20-
ha
20
0.0 0.2 0.4 0.6 0.8 1.0 2 1.4
Time (min)
Fig. 10:Representative Electropherogram showing the separation of a seized heroin sample byusing
lP
MEKC with short end injection (1) morphine, (2) heroin, (3) acetylcodeine, ISTD = internal
standard (N, N-dimethyl-5-methoxytryptamine). UV absorbance at 210 nm, uncoated fused silica
capillary 50 cm x 50 um LD. x360um O.D., effective separation length 8 cm, back- ground
electrolyte: 15 mM sodium borate, 25 mM sodium dodecylsulfate, 15% (v/v) acetonitrile, pH 9.5,
25 °C.-25 kv, hydrodynamicinjection.
Application:
l
1. Capillary electrophoresis (CE) is the primary methodology used for separating and
detecting short tandem epeat (STR) alleles in forensic DNA laboratories.
we
2. Capillary electrophoresis may be used for the simultaneous determination of the ions NH",
Na", K", Mg" and Ca" in saliva.
3. llicit Drug Analysis: Applications of capillary electrophoresis to illicit drugs in seizures and
toxicology samples
Analytes Matrix Method
Coca alkaloids and sugars Illicit cocaine MEKC (Micellar electrokinetic
re
35
a
The analytical separations by capillary electrophoresis are done in many ways which are
called modes. There are four modes are: i) capillary zone electrophoresis (CZE) (i) capillary gel
rm
electrophoresis (CGE) (iii) capillary isoelectrie focusing(CIEF) and (iv) capillary
isotachophoresis(CITP).
ha
l lP
we
re
Ca
36
a
References:
1. P.C. Kamboj Pharmaceutical Analysis - II (Instrumental methods). First edition,
Vallabh Publications; 2010.
rm
2. A. H. Beckett, J. B. Stenlake, Practical Pharmaceutical Chemistry -Part Two Fourth
edition CBS Publishers and Distributors, New Delhi (India); 2001
3. F. Tagliaro, ... S. W. Lewis, in Encyclopedia of Forensic Sciences (Second Edition);
2013 (Science direct)
4. B. K. Sharma, Instrumental Methods of Chemical Analysis-Twenty Seventh Edition;
2011
5. Indian Pharmacopoeia, Volume 1, 'The Pharmacopoeial commission Ghaziabad 2007.