Diq S 282 284
Diq S 282 284
Diq S 282 284
ba76188e01 03/2017
DIQ/S 282
DIQ/S 284
IQ SENSORNET SYSTEM 28X FOR 2 OR 4 DIGITAL IQ SENSORS
IQ SENSORNET System 28X
Contact YSI
1725 Brannum Lane
Yellow Springs, OH 45387 USA
Tel: +1 937-767-7241
800-765-4974
Email: [email protected]
Internet: www.ysi.com
2 ba76188e01 03/2017
DIQ/S 28X Contents
Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 Structure and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 Functions in the IQ SENSORNET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.3 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.3.1 USB interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.3.2 Ethernet interface (DIQ/S 28X-E[F]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.3.3 Fieldbus interface (DIQ/S 28X -MOD, -PR, -EF) . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.4 Behaviour of the system to a power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.1.1 Safety information in the operating manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.1.2 Safety signs on the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.1.3 Further documents providing safety information ....................... 16
2.2 Safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.2 Requirements for safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 User qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1 Scopes of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 Basic principles of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2.1 Requirements of the measurement location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3 Safety requirements of the electrical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4 Installation guidelines for lightning protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5 Drawing up the power rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6 Connecting the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7 On-site mounting of the DIQ/S 28X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7.2 Mounting on a mounting stand with the SSH/IQ sun shield . . . . . . . . . . . . . . . . . 29
3.7.3 Mounting under the SD/K 170 sun shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.7.4 Top hat rail mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.7.5 Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.8 Electrical connections: General instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.9 Connecting the voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.9.1 DIQ/S 28X (line power version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.9.2 DIQ/S 28X[-XX]/24V (24 V version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.10 Connections to the relay and current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.10.1 General installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.10.2 Usage of the auxiliary voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
ba76188e01 03/2017 3
Contents DIQ/S 28X
3.11 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.12 Installation examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.12.1 Connecting two sensors without compressed air cleaning . . . . . . . . . . . . . . . . . . 49
3.12.2 Connecting two sensors with compressed air cleaning . . . . . . . . . . . . . . . . . . . . 50
3.12.3 Connection of a sensor that is located at a distance
(without compressed air cleaning) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.13 Figures of the terminal strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.1 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.1.1 Overview of the operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.1.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.2 General operating principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.2.1 Navigating in menus, lists and tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4.2.2 Entering texts or numerals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
4.3 Access to the IQ SENSORNET with active access control . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.4 Display of current measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
4.4.1 Displaying a single measured value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
4.4.2 Displaying four measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
4.4.3 Displaying eight measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
4.4.4 Displaying recorded measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
4.4.5 Transmitting recorded measurement data to a PC . . . . . . . . . . . . . . . . . . . . . . . . 66
4.5 Messages and log book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.5.1 Message types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.5.2 Log book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
4.5.3 Viewing detailed message texts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4.5.4 Acknowledge all messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4.6 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.6.1 Calibration entries in the log book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.6.2 Calibration history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.7 Status info of sensors and outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.8 General course when calibrating, cleaning, servicing or repairing an IQ sensor .... 74
4.8.1 Maintenance condition of IQ sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.8.2 Switching on the maintenance condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.8.3 Switching off the maintenance condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.9 USB interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.9.1 Saving IQ SENSORNET data to a USB memory device . . . . . . . . . . . . . . . . . . . . . 77
4.9.2 Restore system configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.10 Info on software versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.11 Software-Update for DIQ/S 28X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
5 Settings/setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5.1 Selecting the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5.2 Terminal settings of the DIQ/S 28X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5.3 Access control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
5.3.1 Activating the access control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
4 ba76188d01 03/2017
DIQ/S 28X Contents
ba76188d01 03/2017 5
Contents DIQ/S 28X
7 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7.1 Outputs of the System 28X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7.2 Basic information on relay functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.2.1 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.2.2 Limit indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.2.3 Proportional output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
7.3 Entering / editing the name of an output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
7.4 Linking the output with a sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
7.5 Deleting a link with an output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
7.6 Setting the relay outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
7.6.1 Relay action . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
7.6.2 System monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
7.6.3 Sensor monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
7.6.4 Limit indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
7.6.5 Frequency controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
7.6.6 Pulse-width output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
7.6.7 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
7.6.8 Sensor-controlled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
7.6.9 Manual control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
7.6.10 Alarm contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
7.7 Setting the current outputs (variant -CR3, CR6) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
7.7.1 Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
7.7.2 PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
7.7.3 Fixed current value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
7.8 Checking the condition of the outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
7.9 Behavior of linked outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
7.9.1 Behavior in case of error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
7.9.2 Behavior in non-operative condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
7.10 Maintenance condition of the sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
7.10.1 Switching on the maintenance condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
7.10.2 Switching off the maintenance condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
6 ba76188d01 03/2017
DIQ/S 28X Contents
12 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
12.1 Explanation of the message codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
12.1.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
12.1.2 Informative messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
13 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
ba76188d01 03/2017 7
DIQ/S 28X
8 ba76188d01 03/2017
DIQ/S 28X Overview
1 Overview
1.1 Structure and function
The IQ SENSORNET System 28X is a modular measuring system for online anal-
ysis.
Webserver
email
Internet SPS
(Industrial) Ethernet-LAN
SPS
RS 485 RS 485
The functional units are connected with one another via a common line. The line
consists of two wires and a shield. It transports digital information between the
ba76188d01 03/2017 9
Overview DIQ/S 28X
Universal Transmitter and the other modules. At the same time it is used for the
power supply of all components.
Sensors All main and secondary measured parameters of the sensors connected can be
displayed and administrated. Each sensor is automatically recognized after
being connected and immediately starts measuring.
Universal The Universal Transmitter DIQ/S 28X[-XX] provides the core functions of the
Transmitter IQ SENSORNET system (power supply, controller, terminal, relays).
The variants of the Universal Transmitters DIQ/S 28X enhance the
IQ SENSORNET System 28X by additional functions and interfaces.
10 ba76188d01 03/2017
DIQ/S 28X Overview
Branches To be able to connect more sensors than there are SENSORNET connections
available at a Universal Transmitter DIQ/S 28X, use the branching module DIQ/
JB (accessory). The DIQ/JB branching module is a simple passive module for
the branching of IQ SENSOR NET lines.
Relay and current The relay and current outputs can be linked with sensors. Linked outputs can be
outputs used to monitor sensors and for the output of measured values.
A relay output can be programmed as:
Monitoring relay
Limit monitor
Proportional output of measured values (pulse width or frequency output)
Control unit of a sensor cleaning system driven by compressed air.
ba76188d01 03/2017 11
Overview DIQ/S 28X
For quick orientation, the states of all relay and current outputs are clearly indi-
cated on the display.
Compressed air- The DIQ/S 28X Universal Transmitter is prepared for the compressed air-driven,
driven cleaning time-controlled sensor cleaning function. For this, a DIQ/CHV valve module and
system if necessary a CH cleaning head is required per sensor (both available as acces-
sories). The cleaning procedure is controlled by the Universal Transmitter. The
Universal Transmitter provides the supply voltage and control relay for the
compressed air valve in the DIQ/CHV valve module. This enables a simple
setup and uncomplicated wiring.
Alternatively, the MIQ/CHV PLUS valve module can be installed in the system.
It combines the relay, valve and valve power supply in one MIQ enclosure. Thus,
no additional wiring is required, which makes installation easier especially if the
distance between the Universal Transmitter and sensor is great.
If necessary, an additional power pack can be added to supply sensors with high
power consumption (e.g. UV/VIS sensor).
Example of a
simple system Universal Transmitter
DIQ/S 28X-CR3
DIQ/JB
IQ Sensors, digital
12 ba76188d01 03/2017
DIQ/S 28X Overview
ba76188d01 03/2017 13
Overview DIQ/S 28X
1.3 Interfaces
The USB interface of the DIQ/S 28X provides the following functions:
Connection of a USB memory for the transfer of data (see section 4.9)
– Measured data
– Calibration data
– Configuration data
– Log book
– IQ LabLink data
Connection of a USB memory device to be used as an electronic key for sim-
ple access to the system when access control is switched on (see
section 5.3.3)
Connection of a USB hub for the reproduction of the USB interface.
NOTE:
The USB interface is designed for USB memory devices with a maximum
power consumption of 1 watt. USB memory devices with a higher power
consumption have to be supplied with power by an extra power supply. If
these instructions are not complied with, the availability of the system can
be adversely affected.
The Ethernet interface of the DIQ/S 28X -E[F] provides the following functions:
Integration into an Ethernet network (see section 6)
Monitoring and remote control via the Internet (IQ WEB CONNECT)
E-mail feature
14 ba76188d01 03/2017
DIQ/S 28X Overview
ba76188d01 03/2017 15
Safety DIQ/S 28X
2 Safety
2.1 Safety information
WARNING
indicates a possibly dangerous situation that can lead to seri-
ous (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to slight
(reversible) injury if the safety instruction is not followed.
NOTE
indicates a situation where goods might be damaged if the actions mentioned
are not taken.
Note all labels, information signs and safety symbols on the product. A warning
symbol (triangle) without text refers to safety information in this operating
manual.
16 ba76188d01 03/2017
DIQ/S 28X Safety
The authorized use of the DIQ/S 28X Universal Transmitter consists of its use
in online analysis. Only the operation and running of the sensor according to the
instructions and technical specifications given in this operating manual is autho-
rized (see chapter 10 TECHNICAL DATA). Any other use is considered unauthor-
ized.
ba76188d01 03/2017 17
Installation DIQ/S 28X
3 Installation
3.1 Scopes of delivery
The following parts are included in the scope of delivery:
DIQ/S 28X Universal Transmitter
Accessory kit with:
– Contact carrier with screws
– ISO cap nuts with screws and ring washers
– Cable glands with sealing gaskets
Operating manual.
18 ba76188d01 03/2017
DIQ/S 28X Installation
ba76188d01 03/2017 19
Installation DIQ/S 28X
DIQ modules are installed must be connected to the local potential equaliza-
tion system and the grounding system or must be individually sufficiently
grounded locally according to the codes of practice.
For the individual grounding of the measuring point the mounting construc-
tion must be solidly connected by means of a large-area auxiliary electrode
with the measuring medium.
Metallic control shafts/pipes and other large-area metallic bodies that reach
into the measuring medium are, for example, ideal for use in the grounding
of the mounting construction.
This creates a set path for the main surge. As a result it is possible to avoid
the surge being discharged via the cable and via the valuable sensor in the
measuring medium.
4 It is recommended to attach a metallic or nonmetallic sun shield to each out-
side location of the DIQ modules. Sun shields protect the electric field lines
in the area of the DIQ module due to an advantageous development of the
electrical field lines in the area of the MIQ module and promote the dissipa-
tion of the surge via the mounting construction.
5 The line voltage for the supply of the IQ SENSORNET must comply with over-
voltage category II. Generally this is ensured through the public operator of
the power supply networks. In company-owned networks, e.g. in all power
supply systems owned by wastewater treatment plants, this must be kept
separate by a potential equalization and a surge protection system for the
plant.
6 One part of the safety and lightning protection concept is based on high-
grade protective insulation of the IQ SENSORNET. It does not have or require
any protective ground conductor or earth terminal. Avoid any direct contact
of any SENSOR connections or the metallic sensor enclosures with the local
grounding or potential equalization system and with metallic construction el-
ements (see point 8).
7 Additional external lightning protection measures, e.g. the use of overvoltage
surge arresters, are not necessary for protection against the indirect effects
of lightning and could possibly result in malfunctions.
8 For the realization of the internal lightning protection of the system (e.g.
wastewater treatment plant control stands) and for the protection of external
resources, cable entry points into buildings or distributions coming from the
DIQ/S 28X must be carried out as follows:
– The shield of SNCIQ or SNCIQ-UG cables can be connected to the local
potential equalization with a gas overvoltage arrester. Use shield clamps
for the contacting of the shield. The shield of the cable must not be opened
under any circumstances.
– 0/4-20 mA interfaces must be realized with shielded cables. The cable
shield must be connected directly to the potential equalization(s) provided.
If plant potential equalization systems are provided on both sides, the
shield must also be connected on both sides. The inner conductors must
have no contact with the potential equalization.
– Relay lines should be connected to the local potential equalization in order
20 ba76188d01 03/2017
DIQ/S 28X Installation
Power rating - All components in the system require a specific level of electrical power for oper-
why? ation. Thus, it is necessary to draw up a power rating after selecting the desired
components. At the same time, this can determine whether the entire power
requirement of all components (consumers) is covered by the internal power
supply module of the Universal Transmitter. If this is not the case, the power
available in the system must be increased by further MIQ power supply
modules.
To operate the DIQ/S 28X safely, the power rating must meet the following
condition for continuous operation and for power peaks:
Maximum allowed The maximum allowed power delivery of the power supply components of the
power delivery of IQ SENSORNET is given in the following table:
IQ SENSORNET
components
ba76188d01 03/2017 21
Installation DIQ/S 28X
MIQ modules
MIQ/JB 0.1 0.1
MIQ/CR3 2.3 3.0
DIQ/CR3 2.3 3.0
MIQ/C6 2.0 3.0
MIQ/R6 1.2 1.5
22 ba76188d01 03/2017
DIQ/S 28X Installation
* The following text informs about allowing for the relative turn-on duration (TD)
Allowing for the Valves usually switch on periodically for a limited time and then require the
relative on-time in nominal power. Crucial for the load on the power unit of the DIQ/S 28X is the
valves time averaged (effective) power requirement that depends on the relative on-
time, OT:
Relative on-time OT = tOn / (tOn + tOff)
The effective power requirement is the product of the nominal power of the valve
component and the relative on-time:
P = Pnominal * OT
Since OT is always < 1, the effective power requirement is always smaller than
the nominal power of the valve component.
Determining the From the value determined for the power requirement, determine the number of
number of the MIQ power supply modules as follows:
additional
MIQ power supply Total power requirement P Number of additionally required
modules MIQ/PS power supply modules
P (continuous) P (peak)
P(c) ≤ 6.5 W P(p) ≤ 12 W -
P(c) ≤ 6.5 W P(p) > 12 W 1
P(c) > 6.5 W 1
ba76188d01 03/2017 23
Installation DIQ/S 28X
Calculation example:
Result:
Total power requirement P (continuous) < 6.5 W
Total power requirement P (peak) < 12 W
No additional power supply module is required.
Result:
Total power requirement P (continuous) > 6.5 W
Total power requirement P (peak) > 12 W
An additional power supply module is required.
24 ba76188d01 03/2017
DIQ/S 28X Installation
Temperature If the DIQ/S 28X is operated at an ambient temperature below 47 °C (117 °F),
dependency the operator has to make sure that the total power requirement P (continuous)
(degradation) is not more than 6.5 W.
of the maximum If the DIQ/S 28X is operated at an ambient temperature above 47 °C (117 °F),
allowed the allowed total power requirement P (continuous) is linearly reduced to 0.7 W
P(continuous) at 55 °C (131 °F).
7
6,5
6
1
0,7
0
Fig. 3-1 Degradation line of the line power variant DIQ/S 28X[-XX]
ba76188d01 03/2017 25
Installation DIQ/S 28X
6,5
5,5
4,5
NOTE
The sensor cable may only be connected to the SENSORNET connections. No
wire of the cable may be connected with an external electrical potential. Other-
wise, malfunctions could occur.
26 ba76188d01 03/2017
DIQ/S 28X Installation
1 On the left-hand side of the DIQ/S 28X, remove the two countersunk
screws and open the enclosure.
SENSORNET
X3 X2 X1
2
GREEN
SHIELD
RED
1
SENSORNET 1
3
SNCIQ SACIQ
SNCIQ/UG
Fig. 3-3 Connecting the cable (example DIQ/S 28XCR3)
2 Screw the cable gland (no. 029 212, pos. 1 in Fig. 3-3) with the sealing
(pos. 2) into the housing at the mounting position for the SENSORNET
connection (see label on the bottom of the housing)
3 Loosen the cap nut (pos. 3 in Fig. 3-3).
4 Feed the sensor cable through the cable gland into the enclosure.
ba76188d01 03/2017 27
Installation DIQ/S 28X
X3 X2 X1
RED
SHIELD
GREEN
SENSORNET
schwarz
rot grün
Sensor
Fig. 3-4 SENSORNET connector (example DIQ/S 28X CR3)
5 Connect the cable ends to the terminal strip. At the same time, look out
for the designations of the terminals (red / shield / green).
6 Tighten the cap nut (pos. 3 in Fig. 3-3).
7 Close the enclosure.
The universal transmitter DIQ/S 28X is designed for stationary installation. With
the aid of mounting accessories, the setup can be adjusted to meet various
requirements.
28 ba76188d01 03/2017
DIQ/S 28X Installation
NOTE
Components installed outside must always be protected by a sun shield against
the effects of the weather (snow, ice and direct solar radiation). Otherwise, mal-
functions can result. Always mount the universal transmitter in an upright posi-
tion. Do not under any circumstances install the universal transmitter without
rain protection with the lid facing upwards (danger of retained humidity and pen-
etration of moisture).
NOTE
No contact base may be mounted on the back of the module (danger of short-
circuit!) if the module is mounted on a wall, a sun shield, or a top hat rail.
Installation The most important types of installation for the universal transmitter are
options described in the following chapters:
Mounting with sun shield SSH/IQ:
(see section 3.7.2).
Mounting to the sun shield SD/K 170
The sun shield SD/K 170 provides room for the universal transmitter. The sun
shield can be mounted to round or square section pipes (e.g. rails) with the
aid of the MR/SD 170 mounting kit (section 3.7.3).
Wall mounting:
The universal transmitter is permanently fixed to a wall. For wall mounting,
use the WMS/IQ mounting set (see chapter 11 ACCESSORIES AND OPTIONS).
Panel mounting:
The universal transmitter is mounted in the aperture of a switch panel. The
dimensions of the aperture are 138 mm x 138 mm. The panel may not be
thicker than 10 mm (section 3.7.5).
To install the interface (MOD, PR, E, EF) of the DIQ/S 28X [-MOD], [-PR], [-
E(F)] behind the panel, use the PMS/IQ-X accessory (see section 11).
Top hat rail mounting:
The universal transmitter is mounted on a 35 mm top hat rail with the aid of a
bracket, e.g. in a control cabinet. The connection can be released again with
one simple movement (section 3.7.4).
Materials and tools SSH/IQ sun shield (see chapter 11 ACCESSORIES AND OPTIONS).
required
4 mm set screw wrench
Phillips screwdriver.
ba76188d01 03/2017 29
Installation DIQ/S 28X
2
3
4
1 Screw the sun shield (pos. 1 in Fig. 3-5) with the four hexsocket head
screws (pos. 2), the washers (pos. 3) and the clamps (pos. 4) at the
required height on the mounting stand from the back.
Premounting the 7
ISO blind nuts
6
Fig. 3-6 Mounting the sun shield: Premounting the ISO blind nuts
2 Remove the two countersunk screws (pos. 5 in Fig. 3-6) and swing open
the lid.
3 Insert the cheese-head screws (pos. 6 in Fig. 3-6) with the plastic wash-
ers in the drilled mounting holes and loosely screw in the ISO blind nuts
(pos. 7).
30 ba76188d01 03/2017
DIQ/S 28X Installation
Fig. 3-7 Mounting the DIQ/S 28X on the sun shield SSH/IQ
4 Position the universal transmitter on the sun shield and fix it into place
with the two screws (pos. 6 in Fig. 3-6).
5 Close the lid and fix it with the two countersunk screws (pos. 5 in Fig. 3-
6).
The SD/K 170 sun shield can be mounted directly on a wall, on a mounting stand
or on a railing. The MR/SD 170 mounting kit is also required for mounting on a
mounting stand or railing.
Materials required SD/K 170 sun shield (see chapter 11 ACCESSORIES AND OPTIONS)
The MR/SD 170 mounting kit is also required for mounting the sun shield on
a mounting stand or railing (see chapter 11 ACCESSORIES AND OPTIONS).
ba76188d01 03/2017 31
Installation DIQ/S 28X
2
Fig. 3-8 Mounting the DIQ/S 28X with sun shield SD/K 170
1 Remove the two countersunk screws (pos. 1 in Fig. 3-8) and swing open
the module lid.
2 Position the universal transmitter on the sun shield and fix it into place
with the two screws (pos. 2 in Fig. 3-8).
3 Close the lid and fix it with the two countersunk screws (pos. 1 in Fig. 3-
8).
Materials required THS/IQ kit for top hat rail mounting (see chapter 11 ACCESSORIES AND
OPTIONS).
32 ba76188d01 03/2017
DIQ/S 28X Installation
1 Screw the clamping assembly (pos. 1 in Fig. 3-9) on the back of the uni-
versal transmitter with the two plastic tapping screws (pos. 2).
2 Attach the universal transmitter to the top hat rail from above using the
clamping assembly and press against the rail until the clamping assem-
bly clicks into place. The universal transmitter can be moved sideways
afterwards.
3 To unhook the universal transmitter, press it downward and pull it for-
ward at the bottom.
Variants of the DIQ/S 28X without fieldbus or Ethernet interface can be mounted
to a switch panel with the PMS/IQ mounting set.
For variants of the DIQ/S 28X with fieldbus or Ethernet interface, we recom-
mend using the PMS/IQ-X mounting set. It enables to mount the DIQ/S 28X in
the switch panel so that the fieldbus or Ethernet interface is accessible behind
the switch panel.
ba76188d01 03/2017 33
Installation DIQ/S 28X
Switch panel Details on panel mounting with the PMS/IQ-X mounting set are given in the
mounting with operating manual of the PMS/IQ-X.
PMS/IQ-X
Switch panel Panel mounting with the PMS/IQ mounting set is described below:
mounting with
PMS/IQ
Materials required PMS/IQ kit for panel mounting (see chapter 11 ACCESSORIES AND OPTIONS).
Tools 3 mm set screw wrench (contained in the panel installation kit).
Switch panel
aperture 138
34 ba76188d01 03/2017
DIQ/S 28X Installation
3
e A räg
tag ktt
on onta
elm d K n
leb
ufk er
er
el
2
tap un re lab
i S n tie ve r
Be erne mon o
en
tf rem rrie
ng ca
nti ct
m ou onta
k ll c
tac ta
r s ins
Fo and
5
1
3
2
1 Angle bracket
2 Bolts
3 Bolts
4 Impact protection plate
5 Bolts to fix the impact protection plate
1 Insert the universal transmitter in the panel aperture from the front.
2 Slightly unscrew the bolts (pos. 2 and 3) of the two angle brackets (pos.
1 in Fig. 3-11), but do not remove them.
3 Push in the two angle brackets - as shown in Fig. 3-11 - into the lateral
guides of the universal transmitter up to the stop.
4 Tighten the bolts (pos. 2).
5 Screw in the bolts (pos. 3) until the bolts rest snugly against the panel.
ba76188d01 03/2017 35
Installation DIQ/S 28X
6 Fix the impact protection plate (Pos. 4) to the back of the DIQ/S 28X
with the bolts (Pos. 5).
sealing ring 20 x 15 x 1 mm
blind plug
sealing ring 20 x 15 x 1 mm
sealing ring 24 x 19 x 2 mm
36 ba76188d01 03/2017
DIQ/S 28X Installation
General instal- Observe the following points when attaching connecting wires to the terminal
lation instructions strip
Shorten all wires to be used to the length required for the installation
Always fit all the ends of the wires with wire end sleeves before connecting
them to the terminal strip
Any wires that are not used and project into the enclosure must be cut off as
closely as possible to the cable gland.
Screw a small cable gland with sealing ring into each remaining free opening
and close it with a blind plug.
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. Always cut
off any wires that are not in use as closely as possible to
the cable gland.
ba76188d01 03/2017 37
Installation DIQ/S 28X
WARNING
If the power supply is connected incorrectly, it may repre-
sent a danger to life from electric shock. Pay attention to
the following points during installation:
The DIQ/S 28X universal transmitter may only be con-
nected by a trained electrician.
The connection of the DIQ/S 28X universal transmitter
to the power supply may only be carried out when it is
not carrying any voltage.
The power supply must fulfill the specifications given
on the nameplate and in chapter 10 TECHNICAL DATA.
When installed in a building, a switch or power switch
must be provided as an interrupt facility for the DIQ/
S 28X.
The interrupt facility must
– be installed in the vicinity of the DIQ/S 28X universal
transmitter, easily accessible by the user, and
– be labeled as the interrupt facility for the DIQ/S 28X
universal transmitter.
After the DIQ/S 28X universal transmitter has been in-
stalled, it may only be opened if the line voltage has
been switched off beforehand.
Materials required Wire end sleeves, suitable for the power line, with suitable crimping tool
1 x cable gland with sealing ring (included in scope of delivery of the DIQ/
S 28X).
Preparing the
1 Cut off the cable to the required length.
power cable
2 Strip the cable insulation for approx. 45 mm.
3 Bare the wires of phases L and N and fit them with wire end sleeves.
4 If present, cut off the ground wire at the end of the cable sheath.
38 ba76188d01 03/2017
DIQ/S 28X Installation
approx. 45 mm
NOTE
The ground wire must not project into the enclosure. Otherwise, malfunctions
could occur.
Connecting the
5 On the left-hand side of the DIQ/S 28X, remove the two countersunk
power line
screws and open the enclosure.
2
1
3 N
L
6 Screw a cable gland (pos. 1 in Fig. 3-13) with sealing ring (pos. 2) into
the enclosure below the power supply connection.
7 Loosen the coupling ring (pos. 3).
ba76188d01 03/2017 39
Installation DIQ/S 28X
8 Feed the power line through the cable gland into the enclosure. When
doing so bend the flexible divider (pos. 4) to the right.
X19 X18
L1 N
100...
240V AC
MAINS
L N
9 Connect phases L and N to the terminal strip. Make sure that the cable
assignment agrees with the specification on the terminal label under the
terminal strip.
10 Tighten the cap nut (pos. 3 in Fig. 3-13).
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. Always cut
off any wires that are not in use as closely as possible to
the cable gland.
40 ba76188d01 03/2017
DIQ/S 28X Installation
WARNING
If the 24 V AC/DC supply is connected incorrectly, it may
represent a danger to life from electric shock. Pay atten-
tion to the following points during installation:
The DIQ/S 28X universal transmitter may only be con-
nected by a trained electrician.
The 24 V AC/DC supply must meet the specifications
quoted on the name plate and in chapter 10 TECHNICAL
DATA (protective low voltage SELV).
The connection of the DIQ/S 28X universal transmitter
to the power supply may only be carried out when it is
not carrying any voltage.
When installed in a building, a switch or power switch
must be provided as an interrupt facility for the univer-
sal transmitter DIQ/S 28X.
The interrupt facility must
– be installed in the vicinity of the DIQ/S 28X universal
transmitter, easily accessible by the user, and
– be labeled as the interrupt facility for the DIQ/S 28X uni-
versal transmitter.
Materials required Wire end sleeves, suitable for the 24 V AC/DC feed line, with suitable crimp-
ing tool
1 x cable gland with sealing ring (included in scope of delivery of the DIQ/
S 28X).
Preparing the 24 V
1 Cut off the cable to the required length.
AC/DC line
2 Strip the cable insulation for approx. 45 mm.
3 Bare the wires 1 and 2 and fit them with wire end sleeves.
ba76188d01 03/2017 41
Installation DIQ/S 28X
ca. 45 mm
wire 1
wire 2
Fig. 3-15 Prepared 24 V AC/DC line.
Connecting the
4 On the left-hand side of the DIQ/S 28X, remove the two countersunk
24 V AC/DC line
screws and open the enclosure.
2
1
3 N
L
5 Screw a cable gland (pos. 1 in Fig. 3-16) with sealing ring (pos. 2) into
the enclosure below the 24 V AC/DC connection.
6 Loosen the coupling ring (pos. 3).
7 Feed the 24 V AC/DC line through the cable gland into the enclosure.
When doing so bend the flexible divider (pos. 4) to the right.
42 ba76188d01 03/2017
DIQ/S 28X Installation
X19 X18
24V AC DC
INPUT
POWER
L N
8 Connect wires 1 and 2 to the terminal strip. Make sure that the cable
assignment agrees with the specification on the terminal label under the
terminal strip.
9 Tighten the cap nut (pos. 3 in Fig. 3-16).
NOTE
No free wires are allowed to project into the housing. Otherwise there is the dan-
ger of short circuits that can cause a fire. Always cut off any wires that are not in
use as closely as possible to the cable gland.
ba76188d01 03/2017 43
Installation DIQ/S 28X
WARNING
If external electrical circuits that are subject to the danger
of physical contact are incorrectly connected to the relay
contacts, there may be a danger of life threatening electric
shock. Electrical circuits are regarded to be subject to the
danger of physical contact when there are voltages higher
than the Safety Extra Low Voltage (SELV).
Pay attention to the following points during installation:
Electrical circuits subject to the danger of physical
contact may only be connected by a qualified electri-
cian.
Electrical circuits subject to the danger of physical
contact may only be connected when they are voltage-
free.
If electrical circuits subject to the danger of physical
contact are switched with a relay, no circuit that is not
subject to this danger (e. g. the DIQ/CHV module) may
be operated on the further relays.
Switching voltages and switching currents on the relay
contacts must not exceed the values specified in
chapter 10 TECHNICAL DATA. Protect electrical circuits
against currents that are too high with an electrical
fuse.
Only single-phase consumers can be switched with the
relays. Under no circumstances may multiphase con-
sumers be switched with the aid of several relays (ex-
ample: three-phase current driven pumps). Always
switch multiphase consumers via a protective relay.
After the Universal Transmitter DIQ/S 28X has been in-
stalled, it may only be opened if all external voltages
have been switched off beforehand.
Materials required Wire end sleeves, suitable for the connecting wires, with suitable crimping
tool
4 x cable gland with sealing ring (scope of delivery of the DIQ/S 28X)
44 ba76188d01 03/2017
DIQ/S 28X Installation
Connecting lines
1 On the left-hand side of the DIQ/S 28X, remove the two countersunk
to the terminal
screws and open the enclosure.
strip
2
1
3 N
L
2 Screw a cable gland (pos. 1 in Fig. 3-18) with the sealing ring (pos. 2)
into the enclosure below the respective connections.
3 Loosen the coupling ring (pos. 3).
4 Feed the line through the cable gland in the enclosure.
5 Connect the wires to the terminal strip. While doing so, pay attention to
the specifications on the label located under the terminal strip.
6 Tighten the coupling ring (pos. 3).
ba76188d01 03/2017 45
Installation DIQ/S 28X
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. This could re-
sult in life threatening electric shock when working with
the universal transmitter DIQ/S 28X. Always cut off any
wires that are not in use as closely as possible to the cable
gland.
Connection Klemmleiste R3 R2 R1
compressed air
cleaning
Ventil-
Steuerleitung
Klemmleiste
DIQ/CHV X6 X5 X4 X3 X2 X1
DISTRUBUTION VALVE
NOTE
Run the bridge below the divider so the bridge does not bump against the circuit
board in the lid when the enclosure is closed.
46 ba76188d01 03/2017
DIQ/S 28X Installation
Separating plate
Relais contact Auxiliary voltage output
Bridge
3.11 Commissioning
Start checklist and Before starting the system, carry out the system check using the following
system start checklist. Always carry out the check:
before the initial commissioning
before any further commissioning if the system has been previously extended
or modified.
Start checklist:
1 Are all components correctly connected with each other (see
section 3.6, if necessary, section 3.7)?
2 Is the universal transmitter correctly connected to the power
supply (see section 3.9)?
3 Do the line voltage and line frequency agree with the data on
the name plate of the universal transmitter?
4 Is the sensor ready for measuring, e.g. a D.O. sensor filled
with electrolyte solution?
Starting the Switch on the power supply of the DIQ/S 28X. As soon as the system is suc-
system cessfully initialized, the measured value display appears. With a sensor not yet
providing any measurement values, "Init" appears temporarily:
Select your system language during the first system start. The
selected system language can be changed in the Einstellungen/
Settings menu at any time.
ba76188d01 03/2017 47
Installation DIQ/S 28X
Initial start phase The Universal Transmitter is initialized during the first start phase. All
IQ SENSORNET components are automatically registered with the Universal
Transmitter DIQ/S 28X. The system then performs a self test. This process can
take several seconds. During this period, the following display appears:
Second start As soon as the Universal Transmitter is successfully initialized, the measured
phase value display appears (fourfold display). In the case of IQ sensors that are not
yet providing measured values, "Init" appears temporarily
48 ba76188d01 03/2017
DIQ/S 28X Installation
USB
DIQ/S 28x(/24V)
SNCIQ(/UG) SNCIQ(/UG)
DIQ/JB
1 2 3 4 5 6 7
1 2 3 4 5 6 7
SACIQ SACIQ
SACIQ SACIQ
Sensor 2 Sensor 1
Sensor 2 Sensor 1
X1 X2 X3 X4 X5 X6 X7
red green
red green
SACIQ SACIQ
black black
Sensor 2 Sensor 1
ba76188d01 03/2017 49
Installation DIQ/S 28X
ESC
SNCIQ(/UG)
Valve control line 1
Valve control line 2
SACIQ SACIQ
Air Air
Sensor 2 Sensor 1
Connection Example:
scheme of valve Relay 1 controls the cleaning of sensor 1.
control Relay 2 controls the cleaning of sensor 2.
Variant: Relay 1 controls the cleaning of both sensors. Cleaning of both sensors
is carried out with the same settings.
NOTE
In this configuration, the free switching contact (here: R3) may be used to switch
SELV voltages only.
50 ba76188d01 03/2017
DIQ/S 28X Installation
Variante
Klemmleiste R3 R2 R1 AUXILIARY
VOLTAGE R3 R2 R1 AUXILIARY
VOLTAGE
DIQ/S 28X
X17 X16 X15 X14 X13 X12 X11 X10 X17 X16 X15 X14 X13 X12 X11 X10
a b c
a b c
Ventil-
Steuerleitung 1
Klemmleiste
DIQ/CHV 1 X6 X5 X4 X3 X2 X1
DISTRUBUTION VALVE
Ventil-
Steuerleitung 2
Klemmleiste
X6 X5 X4 X3
DIQ/CHV 2 X2 X1
DISTRUBUTION
VALVE
X1 X2 X3 X4 X5 X6 X7
red green
red green
SACIQ SACIQ
black black
Sensor 2 Sensor 1
ba76188d01 03/2017 51
Installation DIQ/S 28X
DIQ/S 28x(/24V)
M C S
DIQ/S 284
UNIVERSAL TRANSMITTER ESC OK
SNCIQ(/UG)
DIQ/S 28x
DIQ/JB
SNCIQ(/UG)
1 2 3 4 5 6 7
SACIQ
1 2 3 4 5 6 7
SACIQ
Sensor
red green
SACIQ
black
52 ba76188d01 03/2017
DIQ/S 28X Installation
DIQ/S 28X X19 X18 X17 X16 X15 X14 X13 X12
≤ 240V AC ≤ 240V AC ≤ 240V AC
X11 X10 X9
0/4...20mA
X8 X7
0/4...20mA
X6 X5
0/4...20mA
X4 X3 X2 X1
SHIELD
GREEN
RED
-CRx[-E] L1 N ≤ 2A AC ≤ 2A AC ≤ 2A AC AUXILIARY
VOLTAGE
100...
240V AC
+ REC - + REC - + REC -
MAINS R3 R2 R1 AUXILIARY C3 C2 C1 SENSORNET 1
DIQ/S 28X X19 X18 X17 X16 X15 X14 X13 X12 X11 X10 X9 X8 X7 X6 X5 X4 X3 X2 X1
-CRx[-E] ≤ 240V AC ≤ 240V AC ≤ 240V AC 0/4...20mA 0/4...20mA 0/4...20mA
SHIELD
GREEN
RED
≤ 2A AC ≤ 2A AC ≤ 2A AC AUXILIARY
/24V 24V AC DC
VOLTAGE
INPUT
+ REC - + REC - + REC -
POWER R3 R2 R1 AUXILIARY C3 C2 C1 SENSORNET 1
DIQ/S 28X X19 X18 X17 X16 X15 X14 X13 X12
≤ 240V AC ≤ 240V AC ≤ 240V AC
X11 X10 X6 X5 X4 X3 X2 X1
GREEN
SHIELD
-MOD, -PR, -EF
RED
GREEN
RED
SHIELD
L1 N ≤ 2A AC ≤ 2A AC ≤ 2A AC AUXILIARY
VOLTAGE
100...
240V AC
DIQ/S 28X X19 X18 X17 X16 X15 X14 X13 X12 X11 X10 X6 X5 X4 X3 X2 X1
-MOD, -PR, -EF ≤ 240V AC ≤ 240V AC ≤ 240V AC
SHIELD
GREEN
RED
SHIELD
RED
GREEN
≤ 2A AC ≤ 2A AC ≤ 2A AC AUXILIARY
/24V 24V AC DC
VOLTAGE
INPUT
GREEN
RED
SHIELD
GREEN
RED
SHIELD
(component of the
£ 2A AC £ 2A AC £ 2A AC
DIQ/S 284-CR6
+ REC - + REC - + REC -
[-XX]) R3 R2 R1 C3 C2 C1 SENSORNET 2 SENSORNET 1
1 2 3 4 5 6 7
DIQ/JB (7 passive, potential-free terminals
for line extension)
1 2 3 4 5 6 7
X6 X5 X4 X3 X2 X1
ba76188d01 03/2017 53
Operation DIQ/S 28X
4 Operation
4.1 Operating elements
The Universal Transmitter is equipped with a large display for the clear presen-
tation of current measured values, the graph of measured values, status
displays and message texts.
The 5 keys <M>, <C>, <S>, <ESC>, <OK> and the arrow keys <> are
used to operate the IQ SENSORNET system.
M C S
USB
54 ba76188d01 03/2017
DIQ/S 28X Operation
4.1.2 Display
2 3 4 5 6
ba76188d01 03/2017 55
Operation DIQ/S 28X
Measured value The measured value display contains the following information for each IQ sen-
display sor:
1 2 3 4 5
56 ba76188d01 03/2017
DIQ/S 28X Operation
4.1.3 Keys
Key Function
<M> Display measured values
<C> Start calibration
of the IQ sensor selected in the measured value display
<S> Open the Einstellungen/Settings menu
<ESC> Change to the higher menu levels or
abort entries without storing them
<OK> Confirm selection
<> Highlight and select:
Menu items
(arrow keys) Entries
Columns or fields
Letters or numerals
ba76188d01 03/2017 57
Operation DIQ/S 28X
Short operating instructions are given in the help lines on the dis-
play.
58 ba76188d01 03/2017
DIQ/S 28X Operation
You can assign names to IQ sensors and MIQ output modules. Example: Enter-
ing a sensor name:
ba76188d01 03/2017 59
Operation DIQ/S 28X
60 ba76188d01 03/2017
DIQ/S 28X Operation
You can break off entering the name with <ESC>. The old name is
retained.
ba76188d01 03/2017 61
Operation DIQ/S 28X
62 ba76188d01 03/2017
DIQ/S 28X Operation
The measured value is shown numerically and as a bar graph on the Measured
values (1 sensor) display.
Up to four measured values of IQ sensors are shown on the display at the same
time.
Up to eight measured values of IQ sensors are shown on the display at the same
time.
ba76188d01 03/2017 63
Operation DIQ/S 28X
If the measured value recording has been activated for an IQ sensor (see
section 5.10), the temporal course of the recorded measured values can be
displayed numerically and graphically.
The following display options are possible:
Monthly load of selected sensor (graphic display)
Weekly load of selected sensor (graphic display)
Daily load of selected sensor (graphic display)
Measured value list of selected sensor (numerical display)
64 ba76188d01 03/2017
DIQ/S 28X Operation
X 3
4
5
1 Display
2 Range of the graphical display
The time range (x-axis) corresponds to the selected type of display
(1 day, 1 week, 1 month).
The value range of the y-axis is automatically adjusted to min and
max values
3 Position of the cursor (X)
4 Data on the position of the cursor (X)
(date, time, main and secondary measured value)
5 Navigation instructions
Fig. 4-12 Daily load of selected sensor (example)
Displaying
1 Switch to the measured value display with <M>.
recorded
measured values 2 Select an IQ sensor with <>.
3 Open the Display/Options menu with <OK>.
4 Select one of the display types with <>
Monthly load of selected sensor
Weekly load of selected sensor
Daily load of selected sensor
Measured value list of selected sensor
and confirm with <OK>.
The selected option is displayed.
ba76188d01 03/2017 65
Operation DIQ/S 28X
5 Move the cursor (X) along the measured value curve with <>.
The cursor (X) marks the selected measured value.
6 Switch to the next display option with a shorter display period with
<OK>.
or
Switch to the next display option with a longer display period with
<ESC>.
66 ba76188d01 03/2017
DIQ/S 28X Operation
In the case of new error messages, the error symbol flashes on the display.
Information i
Information that does not require immediate action.
In the case of new information, the info symbol flashes on the display.
The log book is a list with all the messages from all modules. The log book
contains up to 1000 entries. If there are more than 1000 messages, the oldest
entries are deleted.
New messages can be recognized by the flashing information symbol or error
symbol. The log book always contains the latest message in the first position.
They do not yet have a checkmark in the status field.
The flashing of the info or error symbol only stops after all detailed message
texts in the log book have been opened and marked with a checkmark () (see
section 4.5.3).
ba76188d01 03/2017 67
Operation DIQ/S 28X
1 2 3 4 5
Fig. 4-14 Log book of entire system
There is a detailed message text for each message of a module that is ready for
operation. The detailed message text on each message is given in the log book
(see section 4.5.3) and in the operating manual of the registering module.
Structure of the The message code consists of 6 characters and can contain numerals and let-
message code ters, e.g.: II21171.
68 ba76188d01 03/2017
DIQ/S 28X Operation
Type number
Type
Component code
Category
II2 171
Example: The "171" module (Universal Transmitter DIQ/S 284) sends a message with the
Message code short message form "II2".
II2171 This is an info message (I) of the type Installation (I) with the type number (2).
The detailed message text of the short form message (II2) can be found in the
log book and in the operating manual of the component that sent it.
ba76188d01 03/2017 69
Operation DIQ/S 28X
The log book shows the current status at the point of time it was
opened. If new messages arrive while a log book is open, these
do not appear in the log book. As usual, you are informed of new
messages by flashing info or error symbols. The current log book
with the new messages can be viewed by closing and reopening
the log book.
70 ba76188d01 03/2017
DIQ/S 28X Operation
With the Acknowledge all messages function you can acknowledge all
messages at the same time (see section 4.5.4).
Message texts are stored in the modules that caused them. There-
fore, further information on a log book entry for an IQ sensor such
as, e.g. calibration messages, instructions and help texts are only
accessible in the case of connected components that are ready for
operation.
If a message text is not available because, e.g. a module is not connected with
the system, you can look up the detailed text of the message as follows:
ba76188d01 03/2017 71
Operation DIQ/S 28X
The detailed calibration data of the last calibrations are integrated in the Calibra-
tion history of selected sensor overview.
All calibration data are stored in the IQ sensor. In order to view the
calibration data of an IQ sensor, the IQ sensor must be connected
to the IQ SENSORNET and must be ready for operation.
The date and time of a calibration are entered in the log book. The respective
message text contains an indication as to whether a calibration was successful
or not. The values determined with calibration can be viewed in the calibration
history (see section 4.6.2).
The calibration history contains the detailed calibration data of the last calibra-
tions.
72 ba76188d01 03/2017
DIQ/S 28X Operation
ba76188d01 03/2017 73
Operation DIQ/S 28X
Therefore, keep to the following course when you calibrate, clean, service or
repair an IQ sensor.
Outlet
1 Switch on the maintenance condition for the IQ sensor (see
section 4.8.2).
The display of the sensor in the measured value display flashes.
2 Pull the sensor out of the sample.
3 Perform the calibration in the laboratory, cleaning, maintenance or
repair (removing and replacing) of the sensor (about these topics, see
the component operating manual of the corresponding sensor).
4 Submerse the sensor in the sample again.
5 Wait until the measured value no longer changes.
6 Switch off the maintenance condition for the sensor (see section 4.8.3).
The display of the sensor in the measured value display no longer
flashes.
74 ba76188d01 03/2017
DIQ/S 28X Operation
Maintenance
Sensor
condition
cleaning
Measured value Measured value display ON manual Measured value
active
display display
The measured value
Clean flashes stops flashing. The measured value
Maintenance
Sensor flashes.
condition
cleaning
OFF c e/m OFF manual
(automatic sequence)
c
Display mesage
Switch on the
maintenace
condition
Continue / Cancel
Calibration
Step 1
m
g e
Calibration
Step n
end g/m/e
Switch on the maintenance condition manually when you want to clean, service
or repair (remove and replace) an IQ sensor.
ba76188d01 03/2017 75
Operation DIQ/S 28X
76 ba76188d01 03/2017
DIQ/S 28X Operation
The configuration data can be transferred from the USB memory device back to
the IQ SENSORNET. Thus, you can very easily create systems that are config-
ured identically.
Data backup
1 Switch to the measured value display with <M>.
2 Open the Einstellungen/Settings menu with <S>.
3 Use <> and <OK> to open the Data transfer to USB memory
menu.
4 Select the data to be saved with <>.
Save configuration
Measured data storage
Log book
Calibration history
and confirm with <OK>.
ba76188d01 03/2017 77
Operation DIQ/S 28X
You can cancel the data saving process by highlighting and confirm-
ing the Cancel menu item with <> and <OK>. In this case
the data will not be stored to the USB memory device.
6 Press <> to highlight the Save menu item and confirm with
<OK>.
The selected data will be stored to the USB memory device.
78 ba76188d01 03/2017
DIQ/S 28X Operation
5 Exit the List of all components dialog box with <M> or <ESC>.
You can view the software versions of all components in the dialog
box, List of all components (see section 4.11).
ba76188d01 03/2017 79
Settings/setup DIQ/S 28X
5 Settings/setup
5.1 Selecting the language
A list shows all the available system languages.
3 Select a language from the list with <> and confirm with <OK>.
The active language is marked with a checkmark.
4 Switch to the higher menu level with <ESC>.
or:
Switch to the measured value display with <M>.
80 ba76188d01 03/2017
DIQ/S 28X Settings/setup
Illumination brightness
Illumination brightness (standby)
ba76188d01 03/2017 81
Settings/setup DIQ/S 28X
Access to the system with active access control (see section 4.3).
Access control provides three preconfigured user rights in the system. Each
user type can be protected with its own password.
User right The current user right is shown on the display with the following symbols.
82 ba76188d01 03/2017
DIQ/S 28X Settings/setup
Switching on the
1 Open the Settings menu with <S>.
access control
2 Using <> and <OK>, select and confirm the menu item, Access
control.
The Simple access control dialog window opens.
3 Using <>, select the function, Activate extended access control
and confirm with <OK>.
User right and instrument block can be selected.
ba76188d01 03/2017 83
Settings/setup DIQ/S 28X
Fig. 5-3 Settings -> Access control -> Extended access control
The instrument automatically generates a password for each user right. This
password can be accepted or changed.
84 ba76188d01 03/2017
DIQ/S 28X Settings/setup
With the function Device lock you can protect the Universal Transmitter no only
against unauthorized operating bus also against unauthorized reading of the
current measured values. By activating the instrument block the password for
viewing the measured values is activated.
After a selected interval without user entries, the system is automatically
blocked. The display shows only the IQ SENSORNET logo.
Switching on the
1 Call up the access control (see section 5.3.1).
instrument block
2 Press <> to select the Device lock function and confirm with
<OK>.
A checkmark appears next to the function.
3 Press <> to select the Apply settings function and confirm with
<OK>.
A security prompt opens.
4 Using <>, select OK and confirm with <OK>.
The settings are adopted.
The Access control dialog window is still open.
The current passwords are visible.
The administrator can simplify access to the IQ SENSORNET by saving the pass-
word to USB memory. The USB memory thereby becomes an electronic key.
When the electronic key is connected to the System 28X, the user right stored
there with the relevant password is automatically read out. The user of the elec-
tronic key is logged on to the system with his user right without any further pass-
word prompt.
When the electronic key is disconnected, the IQ SENSORNET automatically
switches to the lowest user right.
ba76188d01 03/2017 85
Settings/setup DIQ/S 28X
Saving a password
1 Connect the USB memory to the USB-A interface.
to an electronic
key 2 Open the Settings menu with <S>.
3 Using <> and <OK>, select and confirm the menu item, Access
control.
The Access control dialog window opens.
4 Press <> to select an authorization level and confirm with
<OK>.
5 Using <>, select the menu item, Save access code on USB
memory and confirm with <OK>.
The password for the system settings is saved to the USB memory.
86 ba76188d01 03/2017
DIQ/S 28X Settings/setup
For the easier identification of the IQ sensors you can assign an individual name
to each IQ sensor.
5 Enter the name with <> and <OK> and confirm with <OK> (see
section 5.4.1).
ba76188d01 03/2017 87
Settings/setup DIQ/S 28X
The numbering of the sensors is generated by the system. The order of the sen-
sors in the measured value display and in the Edit list of sensors overview can
be individually determined.
5 Using <>, select the menu item, Set display position and con-
firm with <OK>. A dialog window opens.
6 Press <> to select the required number for the display position
and confirm with <OK>.
The sensor is displayed at the new position in the list of sensors. The
other sensors are moved accordingly.
88 ba76188d01 03/2017
DIQ/S 28X Settings/setup
5 Using <>, select Erase inactive sensor and confirm with <OK>.
The dialog window for the security prompt appears.
6 Using <>, select Erase inactive sensor and confirm with <OK>.
The inactive sensor is erased.
ba76188d01 03/2017 89
Settings/setup DIQ/S 28X
Sensor settings The following diagram shows the operating steps to call up the sensor setting
menu and enhanced sensor functions:
(Sensor - plug-in)
Settings Sensor
90 ba76188d01 03/2017
DIQ/S 28X Settings/setup
Requirements for a The following requirements have to be met for a sensor-sensor link:
sensor-sensor link Hardware
– A sensor for which an influence quantity can be corrected (e. g. FDO® 700
IQ, TriOxmatic® 700 IQ) is in the IQ SENSORNET system.
– A sensor that measures the influence quantity (e.g. TetraCon® 700 IQ) is
in the IQ SENSORNET system.
Software
– The software of the sensor with correction function (e.g. FDO® 700 IQ)
supports the Sensor-sensor link function.
– The software of the sensor measuring the influence quantity (e.g.Tetra-
Con® 700 IQ) supports the Sensor-sensor link function.
Sensor settings
– The correction function is enabled in the setting menu for the sensor with
ba76188d01 03/2017 91
Settings/setup DIQ/S 28X
Establishing the
1 Using <S>, open the Settings menu.
link
2 Using <><> and <OK>, select and confirm the menu item, System
settings -> Sensor-sensor link.
The Sensor-sensor link overview opens.
The displayed list includes all sensors with which the correction of an
influence quantity is possible (e.g. FDO® 700 IQ).
92 ba76188d01 03/2017
DIQ/S 28X Settings/setup
5 Using <><>, highlight a sensor that can be linked, and confirm with
<OK>.
The sensors are linked.
The influence quantity to be linked and the linked sensor
are displayed.
6 Leave the system settings with <M>.
In the measured value display, the corrected measured value is marked
by an asterisk (*).
ba76188d01 03/2017 93
Settings/setup DIQ/S 28X
94 ba76188d01 03/2017
DIQ/S 28X Settings/setup
For the easier identification of the outputs, you can assign an individual name to
each output in the Edit list of outputs display.
4 Enter the name with <> and <OK> and confirm with <OK> (see
also section 4.2.2).
ba76188d01 03/2017 95
Settings/setup DIQ/S 28X
An inactive dataset for an MIQ output module arises if the system receives no
signals from a registered MIQ output module. Inactive datasets can be recog-
nized by a question mark, e.g. "?01" in the Edit list of outputs overview.
An inactive dataset can be reactivated by assigning it, e.g. to an MIQ output
module of the same type (see section 9.3.3). All settings are retained. If you no
longer require the stored data, you can erase them.
96 ba76188d01 03/2017
DIQ/S 28X Settings/setup
5 Using <>, select Erase output module and confirm with <OK>.
The dialog window for the security prompt appears.
6 Using <>, select Erase output module and confirm with <OK>.
The output is erased.
The procedure and possible settings for linking outputs with sensors
are given in the operating manual of the respective output module.
Under this menu item you can specify reactions on certain alarm events.
An alarm event is when a certain measured value (limiting value) of a sensor is
exceeded or undercut.
Alarm events can be relayed as follows:
As a message on the display
As a relay action
As an Email (see section 5.9.5)
ba76188d01 03/2017 97
Settings/setup DIQ/S 28X
98 ba76188d01 03/2017
DIQ/S 28X Settings/setup
4 To set up a new alarm, select a sensor from the list with <> and
confirm with <OK>. The Set alarm link display opens.
5 Edit the setting table. The required operating steps are described in
detail in section 4.2 GENERAL OPERATING PRINCIPLES.
ba76188d01 03/2017 99
Settings/setup DIQ/S 28X
1
2
3
4
5
1 Alarm no. Axx and date and time of the alarm event
2 User-defined designation
3 Sensor number and model name of the sensor that triggered the
alarm event
4 Series number and name of the sensor that triggered the alarm event
5 Description of the event with specification of the limiting value:
"<" = Undercut
">" = Exceeded
Operating If there are several alarm messages indicated on the display, you can scroll
instructions through the messages with <>. This can be recognized from the page
number in the right lower corner. The latest message is always on the first posi-
tion.
Pressing <M> hides the alarm messages and switches to the measured value
display. After one minute the alarm messages appear again if what caused them
is still present.
All alarm messages can be sent to an Email address. In this case, the same
information that is output on the display is transmitted. For details, see section
5.9.5.
For critical alarm events take into account that Email messages may
possibly be received delayed.
3 Using <> and <OK>, select and confirm the menu item,
Change password. The dialog window for entering the password opens.
4 Press <> and <OK> to enter the new password and press
<OK> to confirm. The setting is changed.
The real time clock is used for the display of date and time in the measured value
display and in log book entries.
Air pressure values can be set within a range of 500 to 1100 mbar.
The TCP/IP settings menu includes functions and settings for the operation of
the DIQ/S 28X in an Ethernet network.
4 Enter texts with <> and <OK>, or select and confirm an option
from the list.
5 Complete the entries on the display TCP/IP settings.
The email menu contains functions and settings for the sending of alarm mes-
sages via Email.
4 Enter texts with <> and <OK>, or select and confirm an option
from the list.
5 Complete the entries on the display email.
6 Using <> and <OK>, select and confirm the menu item, Save
and quit.
The settings are stored. The menu ist closed.
The Settings bus interfaces menu contains functions and settings for the
communication of the IQ SENSORNET with a fieldbus. Prerequisite: An interface
to a fieldbus is available (hardware, optional).
For details, see operating manual ba77010 "IQ SENSORNET Fieldbus linking",
for download at www.YSI.com.
The system allocates memory blocks to an IQ sensor for the Measured value
logging. With a recording interval of 1 measurement dataset per minute, a
recording duration of 1 day per memory block is possible. 360 memory blocks
are available and can be distributed to the sensors.
The recording duration directly depends on the recording interval. If a recording
duration of 0 d is set for a sensor, there is no memory block allocated to the
IQ sensor. The Dur. setting also contains the number of the allocated memory
blocks (see table).
Recording interval Possible recording durations with the given recording interval
1 min 0d 1d 2d 3d ... 360 d
5 min 0d 5d 10 d 15 d ... 1800 d
10 min 0d 10 d 20 d 30 d ... 3600 d
Recording interval Possible recording durations with the given recording interval
15 min 0d 15 d 30 d 45 d ... 5400 d
30 min 0d 30 d 60 d 90 d ... 10800 d
60 min 0d 60 d 120 d 180 d ... 21600 d
5.10.1 Setting the recording interval (dt) and recording duration (Dur.)
When the number of memory blocks for a sensor is reduced, the memory block
with the oldest data is erased. All other data is retained.
PC SmartPhone
DIQ/S 28X-E
Fig. 6-1 DIQ/S 28X-E[F] in a local network and on the Internet (example)
Network address of the Instrument name (is specified in the terminal settings of the Universal
Universal Transmitter in Transmitters (siehe section 5.2).
the LAN When delivered, the combination of instrument name + serial number
is entered as the name (e.g. S284-16160001).
or
Fixed local IP address of the Universal Transmitter (defined in the set-
tings of the Universal Transmitter (see section 6.1.2).
6.3 Establishing the connection with the DIQ/S 28X via a network
The web server of the DIQ/S 28X provides functions for the (remote) operation
of the DIQ/S 28X and for the data exchange (e.g. with a PC) via a network
address.
The IP address and the other network settings for the Universal
Transmitter DIQ/S 28X are configured in the System settings ->
TCP/IP settings menu.
Procedure Enter the network address of the DIQ/S 28X in the address line in the web
browser
1 Enter the network address of the Web browser DIQ/S 28X-E[F] in the
address line of the web browser.
in the LAN
e.g. name or IP address of the DIQ/S 28X-E[F]
on the Internet,
e.g. fixed IP address of the router or DynDNS name
The network connection to the DIQ/S 28X-E[F] is established.
The start page of the IQ SENSOR NET is displayed.
The IQ WEB CONNECT start page of the DIQ/S 28X-E[F] provides links to
the following functions:
"IQ WEB CONNECT Terminal"
(operating the DIQ/S 28X-E[F])
"IQ WEB CONNECT DatalogTransfer"
(transmitting measurement data)
"IQ WEB CONNECT ConfigSaveLoad"
(saving and loading the system configuration of the DIQ/S 28X-E[F])
With the "IQ WEB CONNECT Terminal", you can operate your DIQ/S 28X-E[F]
just like you are used to on the instrument.
You can save data to a USB memory device connected to the instru-
ment (similar to operation of the instrument). To save data to a PC,
select the function "IQ WEB CONNECT DatalogTransfer" (see
section 6.3.3).
Via the website "IQ WEB CONNECT DatalogTransfer", you can save measure-
ment data of the DIQ/S 28X-E[F] directly to a PC.
The transmitted measurement data are stored in a separate file in CSV data
format for each sensor. The file name (e. g. S03_TriOxmatic700IQ_1.csv) is
automatically assigned and includes:
sensor number (e. g. S03)
sensor model (e. g. TriOxmatic700IQ)
consecutive number.
For processing you can open the csv file, e.g. with Microsoft Excel.
2 3 4 5 6 7 8 9
Fig. 6-4 Example of a CSV file after being imported into Microsoft Excel
Via the website IQ WEB CONNECT ConfigSaveLoad, you can display or save the
system configuration of the DIQ/S 28X-E[F] or load it to a DIQ/S 28X-E[F].
6.5 Troubleshooting
The Universal Transmitter DIQ/S 28X provides an Ethernet interface for the
connection of the IQ SENSORNET to private networks, company networks and
public networks.
To establish accessibilty of the IQ SENSORNET in a public network (e.g. Internet),
external services (e.g. Internet access, fixed public IP address etc.) are
required.
A detailed error analysis for connection errors is not possible within the scope of
this operating manual because of the multitude of service providers, network
architectures and connection options involved.
7 Outputs
7.1 Outputs of the System 28X
Functioning of the outputs
Relay outputs work as openers or closers.
Current outputs provide a current depending on the measured value.
Current outputs are available in the variants DIQ/S 28X-CRX and when an
extension module with current outputs is used).
Recorder
PID controller
Fixed current value
7.2.1 Monitoring
When using a relay for monitoring, a relay action (Open, Close) occurs when
certain states occur. This function is suitable, for example, for the monitoring of
errors in the system.
Monitoring limit-
ing values using Measured t1
one or two relays value
Relay 1 t1
UL
Hysteresis UL
2
1 4 6
3
5
Hysteresis LL
LL Relay 2
t2 t2
Ti
Fig. 7-1 Switching points for relays with the function of a limit indicator
A switching delay (t) can be set up for each relay for switching processes. This
is the time period for which a limiting value must be exceeded before the relay
switches. This prevents frequent switching if the measured values are close to
the limiting value.
100
90
Proportional
band
10
0 Measured value
1 2
Fig. 7-2 Output with one relay
10
0 Measured value
2 1 1 2
Relay 1 Relay 2
1 Start value
Measured value with a minimum pulse width or switching frequency
2 End value
Measured value with a maximum pulse width or switching frequency
Fig. 7-3 Output with two relays
Pulse width output The output of the pulse width is used, e.g. for controlling valves.
Pulse-width regulation changes the duration of operation (ton) of the output sig-
nal. Depending on the position of the measured value in the proportional range,
the relay is operated for a longer or shorter period.
Relay
tOn t Off
On
Off
Time [s]
T
Fig. 7-4 Relay output of the pulse-width output
The cycle duration (T) is made up of the turn-on and turn-off switching
duration (ton, toff) of the relay together. While the selected cycle duration
(T) remains constant, the turn-on duration (ton) changes depending on
the measured value and, with it, the pulse width (v). The pulse width can
be set from 0 % to 100 %.
v = (ton / T) * 100 %
v = pulse width, T = cycle duration, ton = turn-on duration
If the measured value is at the end of the proportional range (End value), the
turn-on duration (ton) is long, the turn-off duration is short. This means the
relay operates for a longer period.
If the measured value is at the beginning of the proportional range (Start
value), the turn-on duration (ton) is short, and the relay operates for a corre-
spondingly shorter period.
Frequency output Switching frequency output is used, e.g. for controlling dosing pumps.
In contrast to the pulse-width output, the pulse width is not modulated with the
frequency output but the switching frequency of the output signal. Depending on
the position of the measured value in the proportional range, the relay is
switched more often or less often.
Relay
tOn = 0.3 s
On
Off
T Time [s]
Fig. 7-5 Relay output of frequency output
While the selected switching duration (ton = 0.3 s) always remains con-
stant, the switching frequency at which the relay switches changes
depending on the measured value.
If the measured value is at the end of the proportional range (End
value), the switching frequency is higher.
If the measured value is at the beginning of the proportional range
(Start value), the switching frequency is low.
Characteristic Through the selection of the Start value and End value, the proportional output
curves can be operated with a positive or negative characteristic curve.
Positive characteristic curve:
Select the End value to be greater than the Start value.
The turn-on duration or frequency increases with an increasing measured
value (see page 131).
Negative characteristic curve:
Select the End value to be smaller than the Start value.
The turn-on duration or frequency decreases with an increasing measured
value (see page 132).
The maximum values for the pulse width or switching frequency are assigned to
the End value value and the minimum values for turn-on duration or frequency
are assigned to the Start value value.
Switching frequency f or
Pulse width v
Max.
Min.
Measured value
Start value End value
Fig. 7-6 Positive characteristic curve
Switching frequency f or
Pulse width v
Max.
Min.
Measured value
End value Start value
Fig. 7-7 Negative characteristic curve
Positive character- The proportional output range begins above the initial value. If the proportional
istic curve range is undercut or exceeded, the selected behavior comes into force.
1 Measured value 2
Fig. 7-8 Pulse width output
120
Maximum
switching frequency
60
50 % of maximum
Proportional switching frequency
band
Time
0
Minimum
1 Measured value 2 switching frequency
Fig. 7-9 Frequency output
1 Start value
Measured value with a minimum pulse width or switching frequency
2 End value
Measured value with a maximum pulse width or switching frequency
Negative charac- The proportional output range begins below the initial value. If the proportional
teristic curve range is undercut or exceeded, the selected behavior comes into force.
2 Measured value 1
Fig. 7-10 Pulse width output
100
Maximum
switching frequency
50
50 % of maximum
Proportional switching frequency
band
Time
0
Minumum
2 Measured value 1 switching frequency
Fig. 7-11 Frequency output
1 Start value
Measured value with a minimum pulse width or switching frequency
2 End value
Measured value with a maximum pulse width or switching frequency
In the overview Settings of outputs and links, field Ser. no., outputs
that are linked with sensors have the name of the linked sensor.
7 Using <>, select one of the functions listed below. Confirm with
<OK>.
Function Description
No function The relay output is not used.
System monitoring see section 7.6.2
Sensor monitoring see section 7.6.3
Limit indicator see section 7.6.4
Frequency controller see section 7.6.5
Pulse-width output see section 7.6.6
Cleaning see section 7.6.7
Sensor-controlled see section 7.6.8
Manual control see section 7.6.9
Alarm contact see section 7.6.10
8 Make the settings for the relay outputs with <> and <OK>.
The settings comprise the relay action (see section 7.6.1) and sensor
dependent settings.
9 Using <> and <OK>, highlight and confirm Save and quit.
The new settings are stored.
As soon as a function was selected for a relay output, you can select a relay
action (see section 7.6.1).
The following actions of the relay can be defined in the Action setting:
Settings Explanations
Open The relay should open on every event.
Close The relay should close on every event.
Function The System monitoring function enables the monitoring of system errors.
In order to set up the System monitoring function for a relay output, the relay
output must not be linked with a sensor (see section 7.4).
It can be used to monitor the following system errors.
Collective error In the case of a Collective error message, the relay opens if one of the following
message malfunctions occurs:
One of the sensors properly registered to the DIQ/S 28X does not provide a
valid main measured value
One of the sensors properly registered to the DIQ/S 28X does not provide a
valid secondary measured value
The monitoring output module has not received any new data DIQ/S 28X for
2 minutes.
In any case, the relay remains open for 10 seconds and only closes when the
malfunction is no longer present.
In the following cases, the relay does not open despite an invalid measured
value:
The sensor is being calibrated
The sensor is in the maintenance condition
The sensor is being cleaned with the aid of a valve module in the system
(compressed air operated cleaning system).
Function The Sensor monitoring function enables the monitoring of sensor errors and the
maintenance condition.
In order to set up the Sensor monitoring function for a relay output, the relay out-
put must be linked with a sensor (see section 7.4).
Sensor messages include errors and information that are registered by the sen-
sor.
Special The special sensor errors are sensor-dependent. Details of this are given in the
sensor errors component operating manual of the respective sensor.
General Init This can prompt a relay action for a short time, depending on
Sensor errors the starting behavior of the system
---- Invalid measured value, or defective sensor
Error Communication with sensor interrupted
OFL Measuring range undercut or exceeded (overflow)
Function The characteristic of the limit indicator is laid down in the Limit value UL, Limit
value LL, Hysteresis UL and Hysteresis LL settings. The fundamentals of the
function are described in the introductory chapter (see section 7.2.2).
In order to set up the Limit indicator function for a relay output, the relay output
must be linked with a sensor (see section 7.4).
Function The characteristic of the frequency output is laid down in the Start value, End
value, Frequency (f) min. and Frequency (f) max. settings. The fundamentals of
the function are described in the introductory chapter (see section 7.2.3).
In order to set up the Frequency controller function for a relay output, the relay
output must be linked with a sensor (see section 7.4).
Characteristic If a value for End value that is greater than the Start value is entered, the output
curve has a positive characteristic curve.
In order to obtain a negative characteristic curve, a value for End value must be
entered that is smaller than the value for Start value.
Function The characteristic of the pulse width output is laid down in the Start value, End
value, Pulse width (v) min. and Pulse width (v) max. settings. The fundamentals
of the function are described in the introductory chapter (see section 7.2.3).
In order to set up the Pulse-width output function for a relay output, the relay
output must be linked with a sensor (see section 7.4).
Characteristic You can specify the minimum and maximum pulse width (v). This determines
curve the steepness of the characteristic curve of the output.
7.6.7 Cleaning
Function The Cleaning function enables the time controlled automatic start of the sensor
cleaning function with the aid of a relay of the combination output module.
The relay controls the MIQ/CHV valve module and switches the compressed air
for the CH sensor cleaning head on or off.
In order to set up the Cleaning function for a relay output, the relay output must
be linked with a sensor (see section 7.4).
The relay of the combination output module assigned always works as a closer.
The cleaning cycle consists of Cleaning duration and Adjustment time.
During the cleaning cycle the Clean display flashes. The outputs linked with this
sensor are frozen. The maintenance condition (see section 7.10) is active.
After the Cleaning duration, the relay is opened. During the following Adjustment
time the outputs remain blocked.
The outputs linked with this sensor are only released when the cleaning cycle is
complete. The Clean display disappears. The maintenance condition is finished.
Testing the You can test the operativeness of the cleaning system as follows: Manually
operativeness open or close the relay with the Manual control function (see section 7.6.9) and,
while doing so, check the behavior of the cleaning system.
Alternatively, you can test the operativeness of the cleaning system by checking
the performance of the function at the start time set up (reference time ±
interval). To perform a test immediately, the reference time can be set so that
the next cleaning cycle will start in a few minutes (settings: see following table).
* With short cleaning intervals, the adjustable values for the Cleaning duration
and Adjustment time are limited. The following values apply:
With this, the cleaning times are fixed. They only change when the
reference time is changed.
The reference time and all further cleaning times relate to the date
and time of the system clock. How to set the system clock is
described in the system operating manual.
2
1
t1a t1b
relay 4 4
condition 1
t1 t1 t1
closed
open
t2 t2 t2 t2
1 Reference time
Begin of a cleaning cycle (t1)
Begin of the specified Cleaning duration (t1a)
2 End of the specified Cleaning duration (t1a)
Begin of the specified Adjustment time (t1b)
3 End of the specified Adjustment time (t1b)
End of the cleaning cycle (t1)
4 Reference time ± Cleaning interval (t2)
Begin of a cleaning cycle
Each time the cleaning cycle is canceled, the relay opens immediately.
If the cleaning cycle is canceled automatically, the outputs linked to the sensor
are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the maintenance
condition. The linked outputs are only released after the maintenance condition
was terminated manually.
The next cleaning cycle will be performed at the time set up.
7.6.8 Sensor-controlled
Set up the cleaning process in the menu Settings of sensors and diff.
sensors of the relevant sensor.
Function The Manual control function can be used to test the operativeness of an instru-
ment that is connected to the relay. To do so, close or open the relay manually
and, while doing so, check the behavior of the connected instrument.
The settings for other functions in the Relay function menu as, for
example, Frequency controller and Pulse-width output are retained
while the Manual control is carried out.
Function The Alarm contact function triggers a relay action (opening or closing) if a
defined alarm event takes place. The Alarm contact function is only available for
relays that are not linked with a sensor. If necessary, an existing link must be
erased.
5 Using <>, highlight a line for a current output (Cx) in the column
Feature. Confirm with <OK>.
The Settings of outputs and links display opens.
6 Using <>, highlight the menu item, Current output, and confirm
with <OK>.
Function Settings
No function The current output is not used.
Recorder see section 7.7.1
PID controller see section 7.7.2
Fixed current value see section 7.7.3
8 Make the settings for the current output with <> and <OK>.
9 Using <> and <OK>, highlight and confirm Save and quit.
The new settings are stored.
7.7.1 Recorder
Function The measured values of the linked sensor at the current output are set up as cur-
rent intensity in the Recorder application. The output of measured values is laid
down in the Recorder type, Start value and End value settings.
Function The PID controller function can use an output as a controller output. The regu-
lator can be configured as a Proportional regulator with switchable Integral and
Differential regulator parts (PID Controller).
The control response of the PID controller is described by the following formula:
1 dx
I Controller = I 0 + K xe + xe dt + Td e
Ti dt
with:
I max − I min
K=
Xp
xe = xnom − xactual
The adjustable control parameters are xset, I0, Xp, Imin, Imax, ti and td (see setting
table on page 153).
By activating or deactivating the Integral (ti) and Differential (td) controller part,
the following controller types can be configured:
Characteristic For a pure P controller the correlation between the measured value and current
curve of the I on the controller output results in the following characteristic curve:
proportional
controller
Current I [mA]
Xp
Imax 14
Io 12
Imin 8
0
2 4 8 Measured value [mg/l]
Nominal value
Fig. 7-18 Characteristic curve for application example
6 mA mA
K = =1
6 mg/l mg/l
Within the proportional range an increase of the concentration by 1 mg/l
causes a reduction of the output current by 1 mA. If the measured con-
centration is 5 mg/l, for example, 11 mA is output:
mA
I Controller = 12 mA + 1 ⋅ (4 mg/l − 5 mg/l)
mg/l
mA
I Regler = 12 mA + 1 ⋅ (− 1 mg/l) = 11 mA
mg/l
The highest concentration at which the controller still works in the propor-
tional range is 8 mg/l (corresponding to Imin = 8 mA) and the lowest is
2 mg/l (corresponding to Imax = 14 mA).
Function The Fixed current value function can be used to test the operativeness of the
instruments connected to the outputs: output different current values to the out-
put and, while doing so, check the behavior of the connected instrument.
The settings for other functions in the Current output menu as, for
example, PID controller and Recorder are retained while the Fixed
current value is carried out.
For linked relay outputs or current outputs, you can determine the behavior of
the outputs in case of errors.
Depending on the use of the output, the behavior in case of errors is set in the
following menus:
Output Menu
Frequency controller Frequency with error (see section 7.6.5)
Pulse-width output Error pulse width (see section 7.6.6)
Recorder Current with error (see section 7.7.1)
Error events The specified behavior occurs with the following events or conditions:
The linked sensor does not provide any valid measured value
(display of Init, Error, "-----", or OFL)
Communication with the DIQ/S 28X is interrupted for more than 2 minutes.
The supply voltage for the DIQ/S 28X is too low.
In the function Recorder, the measured value of the linked sensor is outside
the range between Start value and End value.
Freezing the Irrespective of the behavior in case of error that was defined, the following situ-
conditions of the ations cause the conditions of the outputs to be frozen:
outputs The linked sensor is in the maintenance condition
(display of Cal, Clean, or flashing measured value).
Communication with the DIQ/S 28X is temporarily interrupted. After an inter-
ruption of 2 minutes the output changes to the behavior defined for error.
Restoring the The relay or current output automatically returns to its normal condition as soon
normal functioning as all errors have been eliminated and all conditions that caused the freezing of
the outputs were eliminated.
If you want to clean, service or repair (remove and exchange) a sensor, switch
on the maintenance condition manually (see section 7.10.1).
When you have finished cleaning, servicing or repairing the sensor, switch off
the maintenance condition manually (see section 7.10.2).
8.2 Cleaning
IQ SENSORNET, Clean components mounted in the open of gross contamination as necessary.
DIQ modules, We recommend cleaning the worst of the dirt on the enclosure and the area
MIQ modules directly around it each time before opening in order to prevent contamination
from entering the open enclosure.
To clean the module, wipe the enclosure surfaces with a damp, lint-free cloth. If
compressed air is available on site, blow off the worst of the dirt beforehand.
Keep the enclosure closed while doing so.
NOTE
Do not use high-pressure water blasters for cleaning (danger of water penetra-
tion!). Also do not use aggressive cleaning agents such as alcohol, organic sol-
vents or chemical detergents. These types of cleaning agent can attack the
surface of the enclosure.
The enclosure and the window of the display are made of plastic.
Thus, avoid contact with acetone and similar detergents.
Remove any splashes immediately.
IQ sensors The cleaning of the IQ sensors depends greatly on the respective application.
Instructions for this are given in the respective component operating manual.
A valve module for the compressed air-operated cleaning of sensors is available
as an accessory.
9 What to do if ...
9.1 Information on errors
Log book The IQ SENSORNET system performs a comprehensive cyclical self test during
operation. While doing so, the system identifies all states that deviate from
normal operation and enters corresponding messages in the log book (informa-
tion or error message).
With the aid of the log book, you can call up instructions on how to clear the error
directly on the Universal Transmitter. The log book is described in detail in
section 4.5 MESSAGES AND LOG BOOK.
Passive components include all components that the Universal Transmitter can-
not recognize.
These include:
MIQ/PS (power supply module)
DIQ/JB (branching module)
DIQ/CHV (module for automatic compressed air cleaning)
Cables (SNCIQ, SACIQ).
WARNING
If the MIQ/PS power supply module is opened during op-
eration, there is a danger to life due to possible hazard of
electric shock. The MIQ/PS may only be opened when the
line voltage is switched off. Secure the line voltage
against being switched on again.
Only ever replace components when the IQ SENSORNET
system is switched off. Defective components are re-
moved in the reverse order to the installation (see
chapter 3 INSTALLATION).
Inactive datasets If an IQ sensor is removed from the system, its settings remain stored in the Uni-
of IQ sensors versal Transmitter. A "?" appears in the left column of the Edit list of sensors
menu. This indicates an "inactive dataset". A dataset contains the following
information:
Serial number of the IQ sensor (and, with it, the type of sensor)
Display position
All sensor settings
All characteristics of the link with an output.
Case 1:
The serial number of the IQ sensor The connected IQ sensor is automat-
is identical with the serial number of ically assigned to the inactive dataset
an inactive dataset. and starts to operate again.
Case 2:
The type of sensor is identical with Operator intervention is required here.
the type of sensor in an inactive The connected IQ sensor can:
dataset (or several inactive datas- – be assigned to the inactive dataset
ets), but the serial numbers differ. (or one of the inactive datasets).
Example: Replacing an IQ sensor.
Make sure that the substitute sen-
sor has at least the software status
of the active IQ sensor.
– be included as a new module in the
list of the sensors.
The operating sequence to do this is
described below.
Operating
1 Connect a new IQ sensor.
sequence in case 2
2 Switch to the measured value display with <M>. The component data-
base is updated. The following display appears (example):
3 Select the required option with <> and confirm with <OK>.
– If Add new sensor was selected, the system changes directly to the
measured value display. As soon as the IQ sensor is ready for oper-
ation, it delivers a measured value.
– If Assign sensor as a substitute was selected, a list of matching inac-
tive datasets appears:
4 Select the required inactive dataset with <> and confirm with
<OK>. The system changes to the measured value display. The
IQ sensor takes over all settings of the inactive dataset. As soon as the
IQ sensor is ready for operation, it delivers a measured value.
WARNING
There is also a danger to life from voltages that may pres-
ent a hazard of electric shock on the relay contacts of
open DIQ or MIQ output modules (e.g. DIQ/CR3 or MIQ/
CR3). Output modules with relays may only be opened if
all external voltages have previously been switched off.
Secure all external voltages against being switched on
again for the duration of the work.
When replacing DIQ or MIQ modules make sure that the substitute
module has at least the software status of the active module.
Only replace any DIQ or MIQ output modules (all DIQ or MIQ modules with relay
contacts and/or electrical signal outputs) while the DIQ/S 28X system is
switched off. Defective DIQ or MIQ output modules are removed in the reverse
order to that of the installation. The installation is described in the INSTALLATION
chapter of the respective component operating manual.
Inactive datasets If an DIQ or MIQ output module is removed from the system, its settings remain
in MIQ output stored in the Universal Transmitter. A "?" appears in the left column (= inactive
modules dataset) of the Edit list of outputs menu. The dataset contains the following infor-
mation:
All settings of the relays
All settings of the current outputs.
The maximum number of data sets (active and inactive datasets) for
DIQ or MIQ output modules is limited to 2 in the IQ SENSORNET sys-
tem DIQ/S 28X. When this number is reached, no further DIQ or
MIQ output module can be installed. If necessary, an inactive data-
set has to be erased to make an extension possible.
If a DIQ or MIQ output module is connected to the system when an inactive data-
set is present, the following cases are possible:
Case 1:
The serial number of the DIQ or The connected DIQ or MIQ output
MIQ output module is identical with module is automatically assigned to
the serial number in an inactive the inactive dataset and start operat-
dataset. ing again.
Case 2:
The type of the DIQ or MIQ output Operator intervention is required
module is identical with the type of here. The connected DIQ or
an inactive dataset (or several inac- MIQ output module can:
tive datasets), but the serial num- – be assigned to the inactive dataset
bers differ. (or one of the inactive datasets).
Example: Replacing a DIQ or
MIQ output module.
– be included as a new module in
the list of outputs.
The operating sequence to do this is
described below.
Operating
1 Remove the (defective) DIQ or MIQ output module. The output module
sequence in case 2
is removed in the reverse order to the installation. The installation is
described in the INSTALLATION chapter of the respective component
operating manual.
2 Install the new DIQ or MIQ output module (INSTALLATION chapter of the
respective component operating manual).
3 Switch to the measured value display with <M>. The component data-
base is updated. The following display appears (example):
4 Select the required option with <> and confirm with <OK>.
– If Add new output module was selected, the system changes directly
to the measured value display.
– If Assign output module as a substitute was selected, a list of the
matching inactive datasets appears:
5 Select the required inactive dataset with <> and confirm with
<OK>. The system changes to the measured value display. The
MIQ output module takes over all settings of the inactive dataset.
10 Technical data
10.1 DIQ/S 282, DIQ/S 284
144,0
M C S
USB
11,0
Rear view: 16,5
115.0
70.0
45.0
72.0
137.0
Stack mounting:
48 48
137.0
144,0
M C S
USB
21,0
11,0
72,0
137,0
Stack mounting
48 48
137,0
Fig. 10-2 Dimension drawing DIQ/S 28X-MOD, DIQ/S 28X-PR (dimensions in mm)
144,0
M C S
USB
28,0
11,0
72,0
137,0
Stack mounting
48 48
137,0
Type of protection IP 67
(not suitable for direct conduit connection).
Ambient Temperature
conditions
Mounting/installation/ + 5 °C ... + 40 °C (+ 41 ... +104 °F)
maintenance
Operation - 20 °C ... + 55 °C (- 4 ... + 131 °F)
Storage - 25 °C ... + 65 °C (- 13 ... + 149 °F)
Relative humidity
Mounting/installation/ ≤ 80 %
maintenance
Yearly average ≤ 90 %
Dew formation Possible
Electrical data Power supply Nominal voltage: 100 ... 240 VAC ± 10 %
DIQ/S 28X
(240 V AC/DC line Frequency: 50/60 Hz
power version) according to DIN IEC 60038
Overvoltage II
category
Power consumption max. approx. 20 W
Connection:2 pin
Switch-on current:
1.5 A AC/DC (100 ms)
Power consumption max. approx. 20 W
EMC product and EN 61326 EMC requirements for electrical resources for con-
system character- trol technology and laboratory use
istics
– Resources for industrial areas, intended for indis-
pensable operation
– Interference emission limits for resources of class
A
System lightning pro- Extended protective characteristics as opposed to
tection EN 61326
FCC, class A
Ethernet RJ45 socket (can be retrofitted as a climate resistant RJ45 socket for operation
connection at an open-air test site section 11).
(DIQ/S 28X-E[F])
* Connection via 9-pin D-SUB socket on the underside of the housing, compatible
with Phoenix connector (IP67).
144,0
115.0 11.0
70.0
16.5
Rear view:
45.0
72.0
10.3 DIQ/JB
Dimensions 94 57
DIQ/JB
94
ca. 25
1 2 3 4 5 6 7
Electrical (7 passive, potential free terminals for
connections line extension or branching)
1 2 3 4 5 6 7
10.4 DIQ/CHV
Dimensions 94 57
DIQ/CHV
94
28
(HILFLSKONTAKTE) VENTIL
(AUXILIARY) VALVE
X6 X5 X4 X3 X2 X1
NOTE
The valve may only be operated with the auxiliary voltage of the DIQ/S 28X Uni-
versal Transmitter.
Compressed air Required air quality Dry, free of dust and oil
Operating pressure Max. 7x105 Pa (7 bar) absolute
Connections on the DIQ/CHV 6 mm hose nozzles
Space required
195
290
for screwdriver
48
340
Fig. 10-7 Space required for wall and top hat rail mounting: (dimensions in mm)
Panel mounting 48 5
Panel mounting
(PMS/IQ)
215
185
Space required
for screwdriver 105
Fig. 10-8 Space required for panel mounting (PMS/IQ) (dimensions in mm)
12 Messages
12.1 Explanation of the message codes
The log book contains a list with all the messages from all modules. Each
message consists of message code, date and time. You can obtain more
detailed information by opening the full message text (see section 4.5).
The full message text comes from the component that has triggered the
message. Therefore, these texts are only available from components which are
connected to the system and ready for operation.
If a message text is not available, because the component is not connected to
the system, you can look up the texts of the messages in the operating manual
of the respective component.
The following lists contain the message codes and the related message texts
that are shown on the display. Error messages and info messages are listed
separately.
General explanations on the subjects of messages, message codes and log
book are given in section 4.5 of this system operating manual.
13 Index
A External voltages ......................................164
Access control ............................................ 81
Air pressure .............................................. 104 F
Ambient conditions ................................... 171 Fieldbus ....................................................120
Arrow keys 57, 125, 128, 134, 137, 139, 143, Function code ...........................................108
144, .......................... 146, 147, 154, 157, 158
Auxiliary voltage
I
Terminal strip ....................................... 53
Usage ................................................... 46 Inactive dataset
Output module ..............................96, 165
Sensors .........................................88, 161
B Information ..................................................67
Bus interfaces .......................................... 108 Initialization .................................................48
IQ Web Connect .......................................116
C
Calibration data .......................................... 72 K
Calibration history ...................................... 72 Keys ............................................................57
Cleaning
of module enclosures ......................... 159
L
Current outputs
Terminal strip ....................................... 53 Language ....................................................80
Lightning protection
External protective measures ...............19
D Installation guidelines ...........................19
Daily load diagram of selected sensor ....... 64 Internal protective measures ................19
Data transmission ...................................... 66 Line power connection
Dimensions 24 V version ..........................................41
DIQ/S 182 .................................. 169, 170 Line power version ................................38
DIQ/S 28X .......................... 168, 169, 170 Log book .............................................66, 160
DIQCHV ............................................. 178 Calibration entry ....................................72
DIQJB ................................................ 177 Structure ...............................................68
MIQ modules ...................................... 175
Display ....................................................... 55
Display of measured values ....................... 62
M
Display position .......................................... 88 Maintenance condition ................................74
Measured value display ........................48, 56
Measured value recording ........................108
E Displaying recorded measured values ..64
Effects of the weather ................................ 29 Recording duration .............................108
Electrical data Recording interval ...............................108
DIQ/S (24 V version) .......................... 172 Setting the recording duration .............109
DIQ/S (line power version) ................. 171 Setting the recording interval ..............109
Email ........................................................ 106 Settings ...............................................109
Email settings ........................................... 106 Messages ...................................................66
EMC characteristics ................................. 173 Code .....................................................68
Entering texts ............................................. 59 Texts .....................................................70
Error ........................................................... 66 Types ....................................................66
Ethernet ................................................... 111 Monthly load diagram of selected sensor ...64
R
Relay
Terminal strip ........................................ 53
Retrofitting components ............................. 11
S
Self-test ...................................................... 48
SENSOR connection .................................. 28
SENSOR connection (terminal strip) .......... 53
Sensor name .............................................. 87
Sensor-controlled cleaning ....................... 146
Sensor-sensor link ..................................... 91
Setting the date ........................................ 103
Setting the time ........................................ 103
Settings ...................................................... 80
Air pressure ........................................ 104
Alarms .................................................. 97
Date .................................................... 103
Sensors ................................................ 91
Site altitude ......................................... 104
Terminal ............................................... 80
Time ................................................... 103
Settings of bus interfaces ......................... 108
Settings TCP/IP ........................................ 105
14 Contact Information
14.1 Ordering & Technical Support
Email: [email protected]
Internet: www.ysi.com
When returning a product for service, include the Product Return form with
cleaning certification. The form must be completely filled out for an YSI Service
Center to accept the instrument for service. The Product Return form may be
downloaded at www.ysi.com and clicking on the ‘Support‘ tab.
YSI
1725 Brannum Lane
Yellow Springs, OH 45387
Tel: +1 937-767-7241; 800-765-4974
Fax: +1 937-767-1058
Email: [email protected]
Web: www.ysi.com
©Xylem Inc