PM172EH Operating Manual
PM172EH Operating Manual
Powermeters
Installation and
Operation Manual
BG0361 Rev. A2
LIMITED WARRANTY
The manufacturer offers the customer a 24-month functional warranty on
the instrument in respect of faulty workmanship or parts from date of
dispatch from the distributor. In all cases, this warranty is valid for 36
months from the date of production. This warranty is on a return to factory
basis.
The manufacturer does not accept liability for any damage caused by
instrument malfunction. The manufacturer accepts no responsibility for
the suitability of the instrument to the application for which it was
purchased.
Failure to install, set up or operate the instrument according to the
instructions herein will void the warranty.
Only a duly authorized representative of the manufacturer may open your
instrument. The unit should only be opened in a fully anti-static
environment. Failure to do so may damage the electronic components
and will void the warranty.
NOTE
The greatest care has been taken to manufacture and calibrate your
instrument. However, these instructions do not cover all possible
contingencies that may arise during installation, operation or
maintenance, and all details and variations of this equipment are not
covered by these instructions.
For additional information regarding installation, operation or maintenance
of this instrument, contact the manufacturer or your local representative or
distributor.
IMPORTANT
Please read this instruction manual before installing the unit, and
take note of the following precautions:
ii
3. Do not connect the instrument to a power source if
it is damaged.
4. Do not expose the instrument to rain or moisture.
5. The secondary of an external current transformer
must never be allowed to be an open circuit when the
primary is energized. An open circuit can cause high
voltages, possibly resulting in equipment damage, fire
and even serious or fatal injury. Ensure that the current
transformer wiring is made through shorting switches
and is secured using an external strain relief to reduce
mechanical strain on the screw terminals, if necessary.
6. Setup procedures must be performed only by
qualified personnel familiar with the instrument and its
associated electrical equipment.
7. DO NOT open the instrument under any
circumstances.
iii
Table of Contents
Quick Start ...................................................................... v
Chapter 1 Introduction .................................................. 1
1.1 About this Manual ......................................................................1
1.2 About the PM172EH ..................................................................1
Chapter 2 Installation .................................................... 8
2.1 Mechanical Installation...............................................................8
2.2 Electrical Installation ................................................................11
Chapter 3 Using the Menus ........................................ 18
Chapter 4 Setup Menus............................................... 20
4.1 Basic Setup Menu....................................................................20
4.2 Communications Port Setup Menus.........................................22
4.3 Digital Inputs Setup Menu........................................................24
4.4 Pulse Counters Setup Menu ....................................................25
4.5 Analog Output Setup Menu......................................................26
4.6 Analog Expander Setup Menu .................................................29
4.7 Pulsing Output Setup Menu .....................................................30
4.8 Alarm/Event Setpoints Setup Menu .........................................31
4.9 Timers Setup Menu..................................................................38
4.10 RTC Setup Menu .....................................................................39
4.11 Display Setup Menu.................................................................41
4.12 User Selectable Options Menu ................................................42
4.13 Relay Operation Control Menu.................................................45
4.14 Access Control Menu...............................................................46
4.15 Reset Menu .............................................................................47
Chapter 5 Data Display ............................................... 49
5.1 Navigating in the Display Mode................................................49
5.2 Data Display Formats ..............................................................52
5.3 Self-Test Diagnostics Display ..................................................58
Chapter 6 Viewing Status Information....................... 59
6.1 The Status Information Menu...................................................59
6.2 Status Display Formats............................................................60
Chapter 7 Communications........................................ 62
Appendix A Technical Specifications........................ 65
Appendix B Using Setpoints ...................................... 71
Appendix C Configuring TOU..................................... 75
Appendix D Event and Data Logging ........................ 77
Appendix E Power Quality Setup............................... 81
iv
Quick Start
TYPICAL INSTALLATION
Shorting
Switches
2A
RELAY
OUTPUT
250VAC,5A
18 19 20 21 22 23 24 25 26 27 28 29
1 2 + - + - 1 2
1 2 +
STATUS INPUT ANALOG OUTPUT RELAYS 1
I1
2 - 3
V1
+ 4
I2
5 - 6
V2
+ 7
I3
8 - 9
V3 N
N/- 10
COM.1 COM.2
RS-232 RS-422/RS-485
5 1 L
11 -TX -RX +TX +RX L/+ 12
9 6 13 14 15 16 17
VN RS-422/485
5 4 3 2 1
5A
9 8 7 6
LINE AC
- +
-RX +RX N
+TX
-TX
POWER- -
METER +
RS-485
UP TO 31
DEVICES
+ -
- + - + POWER-
METER UP TO 31
PC 1 PC 2 RS-485 DEVICES
RS-485 RS-485
00-06011
v
SETUP
Setups can be performed directly on the front panel or via PAS communication
software - see Chapter 4 for full instructions. menu bASc
Performing Basic and Communications Setup option ConF
Press SELECT Î CHG Î ENTER value 4L-n
Press SELECT to activate middle window; press ▲▼ to scroll to option.
Press SELECT to activate lower window; press ▲▼ to scroll to value. menu Prt.1
Press ENTER to save selected value. option Prot
value ASCII
Basic and Communications Setup: Default Options
Code Parameter Default Description
ConF Wiring mode 4Ln3 4-wire Wye using 3 PTs
Pt PT ratio 1.0 Potential transformer primary to secondary
ratio
U.SEC PT secondary voltage 120∗ (U) Nominal secondary voltage of the phase
400∗, V potential transformer
Ct CT primary current 5A Primary rating of the phase current
transformer
d.P Power demand period 15 min Length of demand period for power
demand calculations, in minutes.
E = external synchronization
n.dp Number of power 1 No. of demand periods to be averaged for
demand periods sliding window demands. 1 = block interval
A.dP Ampere/Volt demand 900 s Length of demand period for volt/ampere
period demand calculations, in seconds
0 = measuring peak current
buF Buffer size 8 Number of measurements for RMS
averaging
Freq Nominal frequency 50/60 Hz Nominal power utility frequency
LoAd Maximum demand load 0 Maximum demand load current used in
current TDD calculations: 0 = CT primary current
rSt Reset enabled En Enabled (disable to protect all reset
functions)
Prot Communications ASCII ASCII protocol
protocol
rS Interface standard 485 RS-485 interface
Addr Address 1 Device address
bAud Baud rate 9600 bps Baud rate
dAtA Data format 8n Data format (8 bits, no parity)
H.Sh Flow control nonE No flow control
(handshaking)
rtS RTS control nonE RTS is not used
CPtb ASCII compatibility diS Disabled
mode
vi
Chapter 1 Introduction
1. Chapter 1 Introduction
Chapter 1 Introduction 1
Features
Local Display: The front panel features bright LED displays (three
windows, up to 134 pages, 6-digit energy counters) with adjustable
update time, a bar graph showing percent load with respect to user-
definable nominal load current, two communications receive/transmit
LEDs, and an energy pulsing LED.
Display auto scroll is available on the main screen with a programmable
scroll interval of 2 to 15 seconds. Automatic return to the main screen is
available after 30 seconds of uninterrupted use.
AC Inputs: 3 voltage and 3 current galvanically isolated inputs for direct
connection to power line or via potential and current transformers.
Harmonic Measurements:
• Individual voltage and current harmonics with harmonic
angles up to order 40
• Harmonic power direction (through PAS software)
• Synthesized voltage and current waveforms (through
PAS software)
• Total harmonic powers and energies (active
import/export and apparent)
Setup is menu driven, with optional password protection.
Two communication ports are available for RS-232/RS-485 and RS-
422/RS-485 standards, with ASCII, Modbus RTU and DNP3.0 protocols.
In ASCII and Modbus protocols, 120 assignable registers allow the user
to re-map any register address accessible in the instrument to the user
assignable register area. Changing setups and resetting accumulated
data through communications can be secured by the password.
Two relays are provided for energy pulsing (KYZ) or alarm and remote
control.
Two optically isolated status inputs are provided for status monitoring
with timestamp, for counting pulses and for external demand and time
synchronization.
Two optically isolated analog outputs with an internal power supply are
provided for remote monitoring or control. Current loop options are
0-20mA, 4-20mA, 0-1mA, ± 1mA.
Real Time Clock is provided for date and time stamp log and demand
interval synchronization. Standard or Daylight Savings Time (DST) with
automatic time adjustment is available. DST switch dates can be
configured for the use in different time zones.
The TOU (Time of Use) system:
2 Chapter 1 Introduction
• 7 configurable TOU energy registers for accumulating kWh
(import and export), kvarh (import and export), kVAh and energy
from two external meters through 2 pulsing inputs
• 3 configurable TOU Maximum Demand registers for recording
maximum kW (import and export), and kVA demands
• 16 tariffs
• 16 daily profiles (e.g., 4 seasons, 4 daily profiles per season)
• 8 daily tariff start times (tariff changes)
• 2 calendars
• automatically configurable daily and monthly TOU profile log
for each allocated energy and maximum demand register using
season energy tariffs
A 0.5 Megabyte onboard non-volatile memory is provided for data, event
and waveform recording.
• Event recorder provides logging configuration
changes, resets, device diagnostics and setpoint
operation events.
The event-driven Data recorder is provided for recording user-selectable
data in up to 8 data log files on a periodic basis or on any internal and
external trigger. It can be used for load profile and TOU energy profile
recording, or to log fault data on alarm conditions or digital input changes.
For more detailed information on event and data logging, see Appendix D.
Fast Waveform Recorder is provided for waveform capture and logging
on different events. Along with the disturbance monitor it allows for
capture and recording of various types of disturbances with a duration
from one millisecond and up to tens seconds - transients, outages, sags,
surges and deviations in voltage level.
User-selectable options are provided (see Section 4.11):
• Two power calculation modes
• Programmable energy rollover value
• Phase energy calculations option
• Harmonic power/energy calculation option
Chapter 1 Introduction 3
PM172EH Dimensions
4.5
" 0
87 ( 11 0"
0.7 4.0
0 . 0) )
(2
.
AX
" M AX.
00
4.500
M
5.0 7. 0 )
(114.0 )
(12
5.0
(1 2 00"
7.0
)
5. 000"
(127.0)
SW
ITC
HB
OA
RD
PA
NE
L
99-06020
4 Chapter 1 Introduction
Measured Parameters
NOTE: Real-time values are measured over 1 cycle of fundamental frequency; Average values are
of 8, 16 or 32 Real-time values. # = setup via panel, $ = setup via PC
Chapter 1 Introduction 5
Parameter Display Comm. Analog Pulse Alarm
Total Energy
Total Active Energy Import & Export 3 3 #$
Total Reactive Energy Import & Export 3 3 #$
Total Reactive Energy Net 3
Total Apparent Energy 3 3 #$
TOU Registers
7 Energy registers 3 3
3 Maximum demand registers (selectable kW 3
import & export, kvar import & export, kVA
demands)
16 Tariffs for each TOU register 3
Harmonic Measurements
Voltage THD per phase 3 3 #$ $
Current THD per phase 3 3 #$ $
Current TDD per phase 3 3 #$ $
K-factor per phase 3 3 #$ $
Voltage harmonics per phase up to order 40 3 3
Current harmonics per phase up to order 40 3 3
Voltage harmonic angles up to order 40 3
Current harmonic angles up to order 40 3
Total Harmonic Powers
Total Harmonic Active Power 3 3 $
Total Harmonic Apparent Power 3 3 $
Total Harmonic Energy
Total Harmonic Active Energy Import & Export 3 3
Total Harmonic Apparent Energy 3 3
Fundamental Values
Voltage and Current per phase 3
kW, PF per phase 3 3
kvar, KVA per phase 3
Total kW, PF 3 3
Total kvar, KVA 3
Real-time (RT) Values
RT RMS Voltage per phase 3 #$ $
RT RMS Current per phase 3 #$ $
RT Active Power per phase 3 $
RT Reactive Power per phase 3 $
RT Apparent Power per phase 3 $
RT Power Factor per phase 3 $
RT Total Active Power 3 #$ $
RT Total Reactive Power 3 #$ $
RT Total Apparent Power 3 #$ $
RT Frequency 3 #$ $
RT Neutral Current 3 #$ $
RT Total Power Factor 3 #$ $
RT Voltage & Current unbalance 3 $
Min/Max Logging
Min/Max A, V, total kW, kvar, kVA, PF 3 3
Min/Max Frequency, Neutral current 3 3
6 Chapter 1 Introduction
Parameter Display Comm. Analog Pulse Alarm
Min/Max THD, TDD, K-Factor per phase 3
Voltage Disturbance $
Phase Rotation 3 $
Day and Time 3 3 $
Pulse Counters 3 3 $
Remote Relay Control 3
Inputs & Outputs Status
Digital Inputs Status 3 3 $
Alarm Relay Status 3 3 $
Alarm Trigger/Setpoint Status 3 $
Self-diagnostic Tests 3 3
Chapter 1 Introduction 7
Chapter 2 Installation
2. Chapter 2 Installation
D
IS
P
L
A
Y
HEX NUT #8-32
LOC K WASHER #8
4 .0
"
STANDARD
4" ROUND
SW
I TC
HB
OA
RD
PA
NE
L
99-05013
Figure 2-1 STEP 1 (4” round cut-out): Mount the Display Module
on cut-out
8 Chapter 2 Installation
D
IS
P
L
A
Y
HEX NUT #8-32 92
.0 m
m
LOCK WASHER #8
92.0mm
FLAT WASHER #8 (WIDE)
SW
I TC DIN43700
HB
O 92 x92mm
AR
D PA
NE
L
99-05014
Figure 2-2 STEP 1 (92x92mm Square DIN cut-out): Mount the Display
Module on cut-out
LONG STUD
SW
IT C
HB
OA
RD
PA
NE
L
99-05015
Chapter 2 Installation 9
N
T IO
EC
DI R
NG 99-05016
IDI
SL
Figure 2-4 STEP 3: Slide and Position the Powermeter on Locating Studs
LOCK WASHER M4
THUMB NUT M4
99-05017
Figure 2-5 STEP 4: Affix the Powermeter Using the Thumb Nuts
10 Chapter 2 Installation
2.2 Electrical Installation
) Before installation ensure that all incoming power
sources are shut OFF. Failure to observe this practice
can result in serious or even fatal injury and damage to
equipment.
Connections to the PM172EH are made via terminals located on the rear
panel of the instrument as shown in Figure 2-6.
18 19 20 26 27 28 29
18 19 20 21 22 23 24 25 26 27 28 29
1 2 + - + - 1 2
1 2 +
STATUS INPUT ANALOG OUTPUT RELAYS 1
R
I1
2 - 3
POWER SUPPLY NOM. VOLT. MEASUREMENT
90-264VAC
V1 50/60Hz
STANDARD
690V
13 14 15 16 17
00-06007
Figure 2-6 Terminals - Rear View
Chapter 2 Installation 11
2.2.2 Current Inputs
To ensure accurate readings, the input current should not exceed 1.2A
RMS and 1.76A amplitude for the 1A CT secondary, or 6A RMS and 8.8A
amplitude for the 5A CT secondary.
Copper wiring 2.5 – 3.5 mm2 (12 AWG) should be used.
2.2.3 Ground
Connect the chassis ground of the PM172EH to the switchgear earth
ground using dedicated wire greater than 2 mm2/14 AWG.
12 Chapter 2 Installation
Figure 2-7 Three Wire Direct Connection Using 2 CTs (2-element)
Wiring Mode = 3dir2
K L
LINE 1 (A)
K L
+ -
LINE 2 (B)
K L LOAD
+ -
LINE 3 (C) + -
N
Shorting
Switches
Voltages Currents
1
+
V1 -
3
4 +
V2 6 -
7 +
V3 9
-
10
VN 12
c99-05030
Figure 2-8 Four Wire WYE Direct Connection Using 3 CTs (3-
element) Wiring Mode = 4LL3 or 4Ln3
Chapter 2 Installation 13
Figure 2-9 Four Wire WYE Connection Using 3 PTs, 3 CTs (3-
element) Wiring Mode = 4LL3 or 4Ln3
14 Chapter 2 Installation
Figure 2-11 Four Wire Wye Connection Using 2 PTs, 3 CTs
(2½-element) Wiring Mode = 3LL3 or 3Ln3
This configuration will provide accurate power measurements only if the
voltages are balanced.
Figure 2-12 Three Wire Open Delta Connection Using 2 PTs, 3 CTs
(2½-element) Wiring Mode = 3OP3
Chapter 2 Installation 15
* *
L1 L 1 (A) K L
LINE 1(A) + -
240VAC * *
L2 L 2 (B) K L
LINE 2(B) 120VAC
-
+
240VAC
240VAC * * LOAD
208VAC
L3 L 3 (C) K L
LINE 3(C)
N 120VAC N + -
N
Shorting
Switches
Grounded delta connection
Voltages Currents
* 1 +
L 2 (B) -
2 V1 3
4
5 V2 6
7 -
208V
8 V3 9 +
* 10
120V 120V
*
L 1 (A) L 3 (C) 11 VN 12
N
VOLTAGES DISPLAYED:
1. Line to neutral voltages: 120V; 208V; 120V.
2. Line to line voltages: 240V; 240V; 240V.
c99-05031
2.2.5 Relays
Two relays are provided for energy
pulsing, alarms or remote control.
16 Chapter 2 Installation
2.2.7 Analog Output
The PM172EH provides two optically isolated analog outputs with an
internal power supply and current output options of 0-20 mA and 4-20 mA
(current loop load of up to 500 Ohm), 0-1 mA and ±1 mA (current loop
load of 10 kOhm).
Chapter 2 Installation 17
Chapter 3 Using the Menus
3. Chapter 3 Using the Menus
Press and release SELECT to enter the setup mode. The primary menus
will appear:
Press SELECT again to activate the window of the desired primary menu.
Press ENTER
Select CHG to initialize or modify the instrument setup, or to clear the accumulated
values stored in the instrument. Entry to this menu can be protected by a
password.
SELECT Î CHG Î ENTER
Select StA to view extended status information which may be useful during
installation and in certain applications.
SELECT Î StA Î ENTER
Select OPS for viewing (not editing) the instrument setup options.
SELECT Î OPS Î ENTER
After selecting either OPS or CHG, the list of setup menus is displayed in
the upper window. Figure 3-1 presents a complete menu list. Depending
on the model of your instrument, some menus may not appear.
Password
The Setup Change Menu can be secured by a user-defined password
comprised of 4 digits. The instrument is shipped with password protection
disabled. To enable password protection, go to the Access Control Menu
(see Section 4.14).
The Password Menu appears if password protection is enabled.
To enter a password: PASS
9 Set the first digit using the up and down arrow keys. 0000
9 Press SELECT to advance to the next digit.
9 Set the other password digits in the same manner.
9 Press ENTER to continue setup. If your password is incorrect,
you will return to the Primary Selection Menu.
Status Information
StA
Setup Options ENTER
OPS
Setup Change Password Ï Ð Ï Ð
CHG PASS
Reset Functions
rSt Phase Rotation
Counters Cnt.2
Cnt Counter #3
Analog Outputs Cnt.3
(if ordered) Aout Counter #4
This menu contains the basic configuration options which define the
general operating characteristics of your instrument, such as wiring mode,
input scales, the size of the RMS averaging buffer, etc. Table 4-1 lists the
basic setup options, their code names and applicable ranges.
Activate the middle window to scroll through the list of available options,
and then activate the lower window to set the option value.
bASc To select and view a setup option:
ConF 9 Press SELECT to activate the middle window
9 Use the up/down arrow keys to scroll to the desired
4L-n option. The current value for this option appears in the
lower window.
To change the value of the selected option:
9 Press SELECT to make the lower window active.
9 Press the up/down arrow keys to scroll to the desired value.
9 Press ENTER to store the selected value, or press ESC
to quit the menu.
NOTES
1. The maximum value for CT PRIMARY CURRENT × PT RATIO is 10,000,000. If this
product is greater, power related values will be zeroed.
2. Always specify WIRING MODE, PT RATIO and CT PRIMARY CURRENT prior to
setting up alarm setpoints and analog output channels, otherwise the alarm/event
setpoints and analog outputs, which use these parameters, will automatically be
disabled.
These menus allow you to access the communications port options for
communication with a master computer. Table 4-2 lists the communications
options and their code names.
From the main menu, select the menu for the port you want to configure.
This menu is used to set up the two digital inputs provided by the
PM172EH. Each digital input can be allocated as:
- a status input to monitor external contact status, or
- a pulse input to sense pulses provided by an external
source. One of the inputs can be configured to receive
an external synchronization pulse indicating the
beginning of a new demand interval for power demand
measurements.
A pulse input can also be configured to receive time synchronization
pulses to provide synchronization of the instrument clock with a precise
external time source. Time synchronization pulses can follow in intervals
of one minute multiples aligned at 00 seconds. Receipt of the external
pulse adjusts the RTC to the nearest round minute. Whenever a precise
external demand synchronization source is used, the same input that is
allocated for this pulse can be configured as a time synchronization input.
An input allocated for the external synchronization pulse will be
automatically configured as a pulse input. Status inputs do not have to be
explicitly allocated in your instrument. All digital inputs except those you
have allocated as pulse inputs are automatically configured as status
inputs.
Pulse inputs External demand Time synchronization
synchronization input input
NOTE
Digital inputs configured as status inputs can be monitored via the Status
Information Menu (see Chapter 6) and communications. The pulses being received
via pulse inputs can be directed to one of the four pulse counters (see Section 4.4)
and, at the same time, to any of the TOU energy registers.
This menu allows you to set up an output value and its zero and full
scales for either of the two internal analog output channels. Table 4-3
explains the analog output setup options, and Table 4-4 lists all
measurement parameters that can be directed to analog output.
Output parameter Zero-scale output Full-scale output
An. 1 An. 1 An. 1
Outp Lo Hi
rt U1 0 828
NOTES
1. Except for the signed power factor, the output scale is linear within the value
range. The scale range will be inverted if the full scale specified is less than the
zero scale.
2. The output scale for the signed power factor is symmetrical with regard to
±1.000 and is linear from -0 to -1.000, and from 1.000 to +0 (note that -1.000 ≡
+1.000). Negative power factor is output as [-1.000 minus measured value], and
non-negative power factor is output as [+1.000 minus measured value]. To define
NOTES
This menu allows you to program either of the two relays provided by your
instrument to output energy pulses. Available pulsing parameters are
listed in Table 4-5.
NOTES
1. You will not be able to store your setup in the instrument if you assigned a
parameter to relay output with a zero number of unit-hours per pulse, or if the
parameter you selected has just been assigned to another relay output.
2. If a relay you allocated for pulsing has been manually operated or released, it
reverts automatically to normal operation.
Operate delay
SP 1
On d
The delays before operation (On d)
5
and release (OFFd) are set at 5
SP 1 Release delay seconds and 10 seconds respectively.
OFFd
10
NOTES
1. When you enter the setpoints setup menu at the protected level, monitoring of
setpoints is temporarily suspended until you return to the main setup menu.
2. Each time you select a new trigger parameter, the operate and release limits are
set by default to zero.
3. You will not be able to store your setpoint setup to the instrument if a setpoint
action is directed to a relay allocated for pulsing.
4. The setpoint action directed to a relay output can be overridden using
commands sent via communications. A relay can be operated or released
manually. When the relay reverts to normal operation, it returns automatically to
setpoint control.
5. By using logical operators [OR/AND], trigger parameters can be mixed in the
setpoint in an arbitrary sequence. Logical operations on triggers have no specific
priority or precedence, so that the result of the entire logical expression is defined
only by the location of the logical operator in the expression. The entire logical
expression is evaluated in the direction from the left to right; any logical operator
affects all the conditions evaluated before it. To avoid confusion, it is
recommended not to alternate different logical operators in one setpoint. Instead,
bring all triggers using the same logical operator to one side of the expression, and
the others - to the opposite side.
Code Action
none No action 1
rEL.1 Operate relay #1 2
rEL.2 Operate relay #2 2
Inc.Cn.1 Increment counter #1
Inc.Cn.2 Increment counter #2
Inc.Cn.3 Increment counter #3
Inc.Cn.4 Increment counter #4
Clr.Cn.1 Clear counter #1
Clr.Cn.2 Clear counter #2
Clr.Cn.3 Clear counter #3
Clr.Cn.4 Clear counter #4
Clr.Cnt Clear all counters
ELoG Event log 3
dLoG.1 Data log #1
dLoG.2 Data log #2
dLoG.3 Data log #3
dLoG.4 Data log #4
dLoG.5 Data log #5
dLoG.6 Data log #6
dLoG.7 Data log #7
dLoG.8 Data log #8
32.LoG Waveform log #1 (32 samples/cycle rate)
128.LoG Waveform log #2 (128 samples/cycle rate)
Clr.Enr Reset total energy
Clr.dnd Reset all total maximum demands
Clr.P.dn Reset power maximum demands
Clr.A.dn Reset volt/ampere maximum demands
Clr.tEn Reset TOU energy
Clr.tdn Reset TOU maximum demands
Clr.LHi Clear Min/Max registers
FG1.On Set event flag #1
FG2.On Set event flag #2
FG3.On Set event flag #3
FG4.On Set event flag #4
FG1.OFF Clear event flag #1
FG2.OFF Clear event flag #2
FG3.OFF Clear event flag #3
FG4.OFF Clear event flag #4
1 When a setpoint is operated, its status is always stored to the alarm status
register even if no action is assigned to the setpoint. The alarm status register can
be polled and cleared through communications.
This menu allows you to access the two interval timers provided by the
PM172EH which can trigger setpoints on a user-defined time interval
basis. This is useful for continuous data logging at specified time intervals
in order to produce trend and load profile graphs.
Each timer has a time interval range up to 9999 seconds at a one-second
resolution and runs independently. The timer accuracy is about ±0.05 sec.
To use a timer as the trigger for a setpoint, simply select one of the timers
as a trigger when defining the setpoint, and then specify for the selected
timer a non-zero time interval at which you want the periodic action (for
example, a data log) to be made. To stop a timer, set the time interval to
zero. Each timer can be used to trigger multiple setpoints, for example, if
you need multiple data logs at the same time.
Timer setup To select a timer:
t-r 9 Press SELECT to activate the middle window,
t-r.1 and then use the up/down arrow keys to scroll
to the desired timer. The time interval
1 associated with the timer is displayed in the
lower window.
To change the timer interval:
9 Press SELECT to activate the lower window.
9 Use the up/down arrow keys to set the desired
time interval. The available range is 0-9999
(seconds). Setting the interval to zero stops
the timer.
9 Press ENTER to store your new setting.
9 Press ESC to leave the timer setup unchanged.
To quit the timer setup menu:
9 Press ESC .
This menu allows you to view and set the time, date and day of week in
the onboard Real Time Clock (RTC), and to modify the Daylight Savings
Time (DST) settings for your time zone.
dSt.S This entry specifies the DST start date when Daylight
Savings Time begins. Press SELECT to select the date
APr parameter you wish to change. By default, DST starts at
1St.Sun 2:00 AM on the first Sunday in April of each year.
dSt.E This entry specifies the DST end date when Daylight
Savings Time ends. Press SELECT to select the date
Oct parameter you wish to change. By default, DST ends at
LSt.Sun 2:00 AM on the last Sunday in October of each year.
This menu allows you to view and change display properties. Table 4-9
lists available options with their code names and applicable ranges.
diSP UPdt Display update 0.1 - 10.0 s Defines interval between display
time (0.5)* updates
UPdt
0.5
diSP dAtE Date format n.d.Y* Defines the date format in the RTC
d.n.Y display: d=day, n=month, Y=year
dAtE Y.n.d Each date format character is set
n.d.Y separately.
This menu allows you to change options which relate to the instrument
features and functionality. Table 4-10 lists all available options with their
code names and applicable ranges.
To select an option:
OPtS
9 Press SELECT to activate the middle window,
P.cAL and then use the up/down arrow keys to scroll to
rEAc the desired option.
To change the selected option:
9 Press SELECT to activate the lower window.
9 Use the up/down arrow keys to set the desired
value.
9 Press ENTER to store your new setting or ESC
to leave the previous setting unchanged.
To quit the display setup menu:
9 From the middle window, press ESC or ENTER .
N = S2 − P2
Mode 1 is recommended for electrical networks with low harmonic distortion
(voltage THD < 5%, current THD < 10%); Mode 2 is recommended for all other
cases.
The roll value may be changed in accordance with the average load of the power
line. For example, if average power is 400 kW and the counter must be reset every
3 months (2160 hours), then energy during this period equals 864000 kWh (6
digits) and the roll value = 10.E6.
3 Energy pulse test mode is used for testing the device’s energy measurement
accuracy. When the meter is put into the test mode, the energy pulse LED flashes
at a rate of 0.1 Wh/pulse (10,000 pulses/kWh) in secondary readings as the meter
measures imported (consumed) active or inductive reactive energy. When the
meter is in the test mode, the common energy accumulators are not affected.
This menu allows you to set the relay operation mode: non-failsafe or
failsafe. Failsafe relay operation is the opposite of normal operation where
relay contacts are closed when a relay is operated (activated), and are
open when a relay is released (de-activated). In failsafe mode, an alarm is
activated by a non-energized relay which will open in all cases when an
alarm condition is present or an alarm setpoint is not operational either
due to a loss of control power or due to corruption of the setpoint setup
configuration. A failsafe relay is closed only if it is under setpoint control
and no alarm conditions exist, or if it is manually operated via
communications.
To select a relay:
rELo 9 Press SELECT to activate the middle
rEL.1 window, and then use the up/down arrow
keys to scroll to the desired relay.
nor
To change the relay operation mode:
9 Press SELECT to activate the lower
window.
9 Use the up/down arrow keys to set the
desired option. Select nor for normal
(non-failsafe) relay operation, or select
FSAFE for failsafe relay operation.
9 Press ENTER to store your new setting
or pressESC to leave your previous
setting unchanged.
To quit the setup menu:
9 From the middle window, press ESC or
ENTER .
NOTES
1. You will not be able to change the relay operation mode if a relay has been
allocated for pulsing.
2. When a failsafe relay is allocated for pulsing, it automatically reverts to normal
operation.
This menu can only be accessed via the Setup Change Menu (CHG). It is
used in order to:
• change the user password
• enable or disable password check from the front
panel keypad
• enable or disable password protection for
downloading setups and resetting data through
communications
To view an option setting:
9 Press SELECT to activate the middle window.
9 Use the up/down arrow keys to scroll to the desired option
(PASS, CtrL, Port).
Password Setting Password Protection Password
Protection
for the keypad for communications
This menu allows you to reset to zero the accumulators and Min/Max
registers in your instrument. The menu can only be accessed via the
Setup Change Menu (CHG). If the reset is disabled from the Basic Setup
Menu (see Section 4.1), you will not be able to enter this menu.
The following designations are used in the menu to specify a data location
to be reset:
Lo.Hi Resets Min/Max registers (does not affect maximum demands)
A.dnd Resets volt/ampere maximum demands
P.dnd Resets total power maximum demands
dnd Resets all total maximum demands
EnrG Resets total and phase energies
tOU.E Resets the TOU energy registers
tOU.d Resets the TOU maximum demand registers
Cnt Resets all pulse counters
Cnt.1 Resets counter # 1
Cnt.2 Resets counter # 2
Cnt.3 Resets counter # 3
Cnt.4 Resets counter # 4
The up/down arrow keys are used as follows in the Display Mode:
Ð Scrolls through the pages downward (forward)
Ï Scrolls through the pages upward (backward)
Ï Ð Returns to the first page within current measurement group
Small LEDs -
indicate current
Load bar graph -
measurement
indicates % of
group and
the current load
parameter
The front panel display is updated approximately twice per second; you
03-12008
can adjust the display update rate via the Display Setup Menu (see
Section 4.11).
NOTES
1. The common measurements display (main screen) does not have a
designated indicator LED. If no arrow LED is lit up below the display, this
means that the common measurement parameters are being displayed at
this time. To return to the common measurements from another group, just
press the same key that you used to display this group (the key pointed to
by an illuminated arrow LED) until the illuminated LED goes out.
2. When you move to another measurement group, the instrument stores
your last location; when you return to the previous group, the instrument
restores the last page. At power up, the instrument always returns to the
common measurements group and shows you the last page that was
displayed prior to loss of power.
NOTE
The PM172EH provides a self-check alarm register accessible through
communications that indicates possible problems with instrument hardware or
setup configuration. The hardware problems are indicated by the appropriate bits,
which are set whenever the instrument fails self-test diagnostics, or in the event of
loss of power. The setup configuration problems are indicated by the dedicated bit
which is set when either configuration register is corrupted. In this event, your
instrument will use the default configuration. For more information on the self-
check alarm register, refer to the communications reference guides shipped with
your instrument.
Chassis Ground
DB9
RS-422
(MALE)
+T 1
5 -R
9 +R 2
4 -T TO SATEC
CONVERTER
8 -T 3
TO PM172 +T
3
COM1
7 -R 4
+R
2
6 5
1
RS-422 CABLE
Chapter 7 Communications 63
Computer Connections COM2 - RS-422/RS-485
c98-10007/4
64 Chapter 7 Communications
Appendix A Technical
Specifications
8. Appendix A Technical Specifications
Display
Indication 3 high-brightness seven-segment digital LEDs
Environmental Conditions
Operating temperature -20°C to +60°C (-4°F to +140°F)
Storage temperature -25°C to +80°C (-13°F to +176°F)
Humidity 0 to 95% non-condensing
Construction
Instrument body Case enclosure: Plastic PC/ABS blend
Dimensions: 127 x 127 x 127 mm ( 5.0 x 5.0 x 5.0 “)
Mounting: 4-inch round or 92x92mm square panel mount
Instrument weight 1.23 kg (2.7 lb.)
Display body Display body: plastic PC/ABS blend
Front Panel: plastic PC
Dimensions: 114.0x114.0x20.0 mm (4.5x4.5x.787”)
Standard Compliance
Standards UL File # E129258 (PENDING)
Directive Complied With: EMC: 89/336/EEC as amended by
92/31/EEC and 93/68/EEC
LVD: 72/23/EEC as amended by 93/68/EEC and 93/465/EEC
Harmonized Standards to which Conformity is Declared:
EN55011:1991; EN50082-1:1992; EN61010-1:1993; A2/1995
ANSI C37.90.1 1989 Surge Withstand Capability (SWC)
EN50081-2 Generic Emission Standard - Industrial Environment
EN50082-2 Generic Immunity Standard - Industrial Environment
EN55022: 1994 Class A
EN61000-4-2
ENV50140: 1983
ENV50204: 1995 (900MHz)
ENV50141: 1993
EN61000-4-4:1995
EN61000-4-8: 1993
Voltage 120V×PT @ 120V 0.2 0.1 10% to 120% FS 0 to 999,000 V Direct wiring (PT=1):
400V×PT @ 690V 0.1 V @ 0.1V to 999.9 V
Wiring via PTs (PT>1):
0.001 kV @ 0.001kV to 9.999 kV
≤0.1% @ 10.00 kV to 999.0 kV
Starting voltage 1.5% FS
Line current CT PRIMARY CURRENT 0.3 0.015 1% to 120% FS 0 to 9999 A 0.01 A @ 0.01A to 99.99 A
≤0.1% @ 100.0 A to 9999 A
Starting current 0.1% FS
Active power 0.36×PT×CT @ 120V input 0.4 0.02 PF = 1 -2,000,000 to Direct wiring (PT=1):
1.2×PT×CT @ 690V input 0.5 0.02 |PF| ≥ 0.5 1 +2,000,000 kW 0.001 kW @ 0.001kW to 9.999 kW
Wiring via PTs (PT>1):
0.001 MW @ 0.001MW to 9.999 MW
≤0.1% @ 10.00 MW to 2000 MW
Reactive power 0.36×PT×CT @ 120V input 0.6 0.02 |PF| ≤ 0.9 1 -2,000,000 to Direct wiring (PT=1):
1.2×PT×CT @ 690V input +2,000,000 kvar 0.001 kvar @ 0.001kvar to 9.999 kvar
Wiring via PTs (PT>1):
0.001 Mvar @ 0.001Mvar to 9.999
Mvar
≤0.1% @ 10.00 Mvar to 2000 Mvar
Apparent power 0.36×PT×CT @ 120V input 0.5 0.02 1 0 to Direct wiring (PT=1):
1.2×PT×CT @ 690V input 2,000,000 kVA 0.001 kVA @ 0.001kVA to 9.999 kVA
Wiring via PTs (PT>1):
0.001 MVA @ 0.001MVA to 9.999 MVA
≤0.1% @ 10.00 MVA to 2000 MVA
Power factor 1 0.6 |PF| ≥ 0.5, -0.999 to +1.000 0.001
I ≥ 10% FSI
Frequency 0.02 45.00 to 65.00 0.01 Hz
Hz
Neutral CT PRIMARY CURRENT 0.5 2% to 120% FS 0 to 9999 A 0.01 A @ 0.01A to 99.99 A
(unbalanced) ≤0.1% @ 100.0 A to 9999 A
current Starting current 0.5% FS
Ampere demand Same as for current
KW demand (block Same as for kW
& sliding)
KVA demand Same as for kVA
(block & sliding)
Total harmonic 999.9 1.5 0.2 THD ≥ 2% FS, 0 to 999.9 0.1
Distortion U (I) ≥ 10% FSU
THD U (I), (FSI)
%U1(I1)
Total Demand 100 1.5 TDD ≥ 2% FS, 0 to 100 0.1
Distortion TDD, % I ≥ 10% FSI
Voltage & Current 100 1.5 1.0 HD ≥ 2% FS, 0 to 100 0.01
individual Harmonic U (I) ≥ 10% FSU
Distortion HD, % (FSI)
Harmonic active 0.36×PT×CT @ 120V input 1.5 1.0 PF = 1 -2,000,000 to Direct wiring (PT=1):
power 1.2×PT×CT @ 690V input 2.0 1.0 |PF| ≥ 0.5 3 +2,000,000 kW 0.001 kW @ 0.001kW to 9.999 kW
Wiring via PTs (PT>1):
0.001 MW @ 0.001MW to 9.999 MW
≤0.1% @ 10.00 MW to 2000 MW
Harmonic apparent 0.36×PT×CT @ 120V input 1.5 1.0 3 0 to Direct wiring (PT=1):
power 1.2×PT×CT @ 690V input 2,000,000 kVA 0.001 kVA @ 0.001kVA to 9.999 kVA
Wiring via PTs (PT>1):
0.001 MVA @ 0.001MVA to 9.999 MVA
≤0.1% @ 10.00 MVA to 2000 MVA
Voltage & Current 360° 0.6° + 0.5° × 3 -180° to 180° N/A
harmonic angles, harmonic order
degree
Active energy Class 0.5S (IEC 687-1992-6) 0 to 1 kWh @ 1 to 999,999 kWh
Import & Export 999,999 MWh 10 kWh @ 1000 to 9,999.99 MWh
100 kWh @ 10,000 to 99,999.9 MWh
1MWh @ 100,000 to 999,999 MWh
Reactive energy Under conditions of Class 1.0 0 to 1 kvarh @ 1 to 999,999 kvarh
Import & Export (IEC 1036) 999,999 Mvarh 10 kvarh @ 1000 to 9,999.99 Mvarh
|PF| ≤ 0.9 1 100 kvarh @ 10,000 to 99,999.9 Mvarh
1Mvarh @ 100,000 to 999,999 Mvarh
Additional Notes
1. Accuracy is expressed as ± (percentage of reading + percentage of full scale) ± 1 digit. This does not include inaccuracies
introduced by the user's potential and current transformers.
2. Specifications assume: voltage and current waveforms with THD ≤ 5% for kvar, kVA and PF; reference operating temperature:
20-26°C.
3. Ordinary measurement error is considerably less than the specified accuracy which indicates maximum error.
4. Harmonic real time values assume errors averaged over 10-20 readings.
Memory is allocated for each file statically when you set up the files and
will not change unless you re-organize your files.
The following table shows how to calculate a file size for different log files.
File Record Size, Bytes File Size, Bytes
Event Log 14 Record size x Number of events
Data Log 8 + 4 x Number of Record size x Number of events
parameters in a record
TOU Profile Data Log 8 + 4 x Number of season Record size x Number of TOU
(#7, #8) tariffs registers x Number of events
Waveform Log #1 1040 x 6 channels = 6240 Record size x Number of cycles
(32 samples per cycle) (16 cycles per record) per event/16 x Number of events
Waveform Log #2 1040 x 6 channels = 6240 Record size x Number of events
(128 samples per cycle) (4 cycles per record)
Event Recording
By default, the meter stores to the Event Log file all events related to
configuration changes, resets, and device diagnostics.
In addition, the meter can record to the Event Log events related to
setpoint operations. To record setpoint operations to the Event Log, add
the “Event log” action to the setpoint actions list. When a setpoint event
happens, the meter records all setpoint conditions that caused the event
and all setpoint actions performed in response to the event.
Data Recording
The data recorder is event-driven and can be triggered by any event
through the programmable setpoints.
Waveform Recording
The waveform recorder is event-driven and can be triggered by any event
through the programmable setpoints.
Two waveform log files are organized as multi-section files that store data
for each recording channel in a separate section.
Waveform log #1 is normally intended for recording disturbances at a rate
of 32 samples per cycle. It stores 6 AC channels simultaneously - three
voltages and three currents. The number of pre-event cycles (1 to 8) and
the number of cycles per one disturbance event can be programmed
through the Waveform Recorder setup. The maximum number of cycles
that the meter can record per one event is 1280. To configure the
Waveform Log #1 parameters, select Memory/Log from the Meter Setup
menu, click on the Waveform Recorder tab, and then select the desired
number of pre-event cycles and the number of cycles per one waveform
event series.
Waveform log #2 is intended for recording harmonic disturbances at a
rate of 128 samples per cycle. It records 6 channels - three voltages and
three currents, where channels related to different phases are not
synchronized. Waveform log #2 always stores four cycles of the sampled
waveforms preceding the trigger event.
Each power quality event will force the meter to record the time and fault
magnitude of the power quality parameter to the Event Log both at the
start and the end of the event, and to record all power quality parameters
together at the time of the event to Data Log #3. The following table
shows an example for selecting parameters for Data Log #3. See
Appendix D for information on how to allocate memory for your power
quality Data Log.
Data Log #3
No. Group Parameter No. Group Parameter
1 RT PHASE V1 9
2 RT PHASE V2 10
3 RT PHASE V3 11
4 RT PHASE V1 THD 12
5 RT PHASE I2 THD 13
6 RT PHASE I3 THD 14
7 RT AUX FREQ 15
8 16
81
Fast transient voltages will trigger the Waveform Recorder to log voltage
and current waveforms at the time of the event. Note that the recorded
voltage waveforms will represent line-to-neutral voltages in all wiring
modes except of Open Delta where they will be line-to-line voltages. The
voltage disturbance threshold is defined in percent of the line-to-neutral or
line-to-line nominal voltage respectively.
Take into consideration that in the 4LN3 and 3LN3 wiring modes, the high
and low voltage triggers will represent line-to-neutral voltages; in all other
wiring modes they will be line-to-line voltages (see Note 6 to Table 4-7 in
Section 4.8).
The characteristics of the power quality events, such as a category,
phase, magnitude and duration, can be obtained by analyzing the Event
Log and Data Log files.
The duration of a power quality event can be calculated by taking the
difference between time stamps of the pair of records in the Event Log
recorded when a setpoint corresponding to the power quality trigger was
operated at the beginning of the event and was released at the end of the
event.
The phase, category and magnitude of a voltage and harmonic fault can
be obtained from the power quality Data Log record corresponding to the
event, i.e., having the same setpoint ID and the time stamp close to that
was recorded to the Event Log at the beginning of the event. The phase
and category of a voltage fault can be identified by comparing phase
magnitudes with thresholds recognized for different categories of voltage
faults. Use the following table to classify your voltage faults.
Voltage Fault Category Magnitude Typical Duration
Voltage interruption < 0.01 Vn on all phases
Voltage dip (sag) < 0.9 Vn 0.5 cycles – 1 min
Undervoltage < 0.9 Vn > 1 min
Voltage swell > 1.1 Vn 0.5 cycles – 1 min
Overvoltage > 1.1 Vn > 1 min
82 Index
INDEX
accuracy, 43, 73 54, 56, 57, 58, 60, 61, 70, 71,
active energy, 72 72, 73
active power, 48, 70, 71, 72 power demand, 5, 26, 29, 40
ampere demand, 5, 26, 39, 40, 58 power factor, 71
analog output, 1, 7, 27, 31, 32, 34, power source, 1, 16
48 PT, 5, 17, 26, 27, 33, 60, 68, 70,
analog outputs, 27 71, 72, 73
basic setup, 25 PT RATIO, 27
burden, 68, 69 pulsing, 7, 35, 37, 50
communications, 73 reactive energy, 73
connections, 6 reactive power, 48, 70, 83
CT, 5, 17, 26, 27, 33, 46, 54, 68, relays, 7, 35, 42
70, 71, 72, 73 reset, 5, 26, 45, 49, 52, 54, 55, 56,
CT PRIMARY CURRENT, 27, 70, 71, 61, 62
73 self tests, 61
current inputs, 68 setpoints, 27, 30, 36, 37, 43
current transformers, 73 setup menus, 23, 25
demand, 5, 7, 8, 11, 26, 29, 33, 39, setup mode, 23
40, 52, 55, 56, 58, 71 standards, 7, 68
demand period, 5, 26 status, 6, 7, 8, 23, 29, 30, 42, 51,
diagnostic, 12 62, 63, 64
digital inputs, 8, 29, 30 terminals, 2, 16, 17, 68
display, 6, 46, 47, 50, 52, 53, 54, voltage inputs, 68
55, 56, 60, 62, 63 wiring configuration, 17
display mode, 46, 52, 54, 56, 62 Wye, 5, 26
energy, 7, 8, 30, 35, 42, 48, 49, 52,
55, 56, 60, 72, 73, 83
external synchronization, 5, 7, 26,
29
frequency, 5, 10, 26, 33, 39, 41, 57,
71
harmonic, 6
inputs, 12, 29, 30, 63, 68
Load bar graph, 54
measurement, 73
menu, 23, 25, 27, 29, 30, 31, 34,
35, 36, 37, 43, 44, 45, 46, 47,
50, 51, 52, 60, 62
Modbus, 2, 7, 28, 65
mounting, 13, 14
open delta, 17, 26
options, 22
overload withstand, 68
PAS, 25, 78, 80
password, 7, 23, 51, 52, 54, 62
, 25
PAS, 5
power, 1, 5, 6, 7, 16, 26, 27, 29,
31, 32, 40, 42, 48, 49, 50, 52,
83
NOTES
84 Index