0% found this document useful (0 votes)
102 views8 pages

Mark V Advanced Settings Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
102 views8 pages

Mark V Advanced Settings Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

MARK V Alcovisor®

Advanced Settings Manual


PAS Systems International, Inc.
P O Box 330
Fredericksburg, Virginia 22404

www.pasintl.com
(800) 660-SNIF · (540) 372-3431
Copyright© 2010 by PAS Systems International Inc., All Rights Reserved
Introduction
This manual contains the passwords and instructions on operating the advanced
settings of the MARK V along with the proper procedures to be used for instrument
calibration.

Passwords
Default Passwords:

Calibration: 119119
Advanced Settings: 112112

The default passwords for advanced settings and calibration can be changed. To
change the passwords, please refer to the “Password” section under Advanced
Settings in this manual.

WARNING: If you change the password, write it down. The new


password must be sent in with your unit if it should need service.
Advanced Settings
To access the Advanced Settings menu:
1. Choose the Settings Icon
2. Select “Advanced”
3. Enter the password 112112

Advanced Settings:

1. Blow: This icon under advanced settings allows the user to change the blow
time (in seconds) and blow pressure (1 to 6, with 6 being the hardest). These
are set to optimal levels already and it is not recommended to change them.

2. Default: Select “yes” under this icon to return instrument to Default settings.
Note: Settings have been optimized; it is not recommended to return to Default
settings.

3. Unit: Set the measurement unit and BAC/BrAC ratio. If the unit is changed, the
instrument does not need to be recalibrated.

4. Print: Select Auto Print for a result to be printed automatically (if the optional
printer is available) and input the number of copies to be printed. Select “Cal.”
to have the Last Calibration Date print.

5. Delete: This option allows for the deletion of ALL saved test records stored in
the MARK V. Once you select “Yes”, the test records cannot be restored.

6. Calibration: The password to access the calibration option is 119119. Please


refer to the calibration instructions included in this manual for proper calibration
procedures.

7. Password: The default passwords for advanced settings and calibration can be
changed. Select “Cal” or “Advanced Setting” to change the password. Enter
the new password and select to confirm. Screen will display “Password
Setting Succeed”.

NOTE: Instrument requires a 6-digit password. If less than six


digits are used, password will not store properly.

Warning: If you change the password, write it down. The new password
must be sent in with your unit if it should need service.
8. Calibration Period:
a. Select the number of days in the first column and the number of tests in the
second column that may be allowed to occur before calibration is needed.
b. Select “LOCK” to lock the device if calibration is needed.
c. Default is 360 days, 700 tests and unlocked.

Calibration
PAS recommends all MARK V units be calibrated annually or whenever an accuracy
test shows that BAC readings are no longer within tolerance limits. Because precise
calibration is crucial, it can be performed only by PAS Systems International or by
persons who have been properly trained for alcohol detector calibration. (Calibration
of the MARK V must be performed using a NHTSA-approved wet bath simulator or
certified dry gas standard.) Successful completion of calibration should be recorded
in a calibration log.

NOTE: When using dry gas (PAS recommends 0.080% dry gas concentration) for
calibration, the alcohol concentration printed on the label of the dry gas cylinder must
be corrected to account for altitude when conducting tests at or above an altitude of
250 feet above sea-level. Please refer to the Operating Manual for specific
instructions for altitude corrections.

CALIBRATION: (Using an approved Wet Bath Simulator or Certified Dry Gas


with proper Regulator)

1. Select the Settings Icon and select “Cal.”. Calibration can also be
accessed under Advanced Settings.
2. Enter the password 119119.
3. Enter the concentration of the calibration solution/cylinder. Accuracy should be
to one decimal place. The preset unit is mg/100mL. For example, enter 80.0
for 0.080% or enter 100.0 for a 0.100% solution.
More examples are in the following table:

Mg/100mL % ‰ & g/L


20.0 0.02 0.20
50.0 0.05 0.50
80.0 0.080 0.80
100.0 0.10 1.00
150.0 0.15 1.50

4. Select to confirm.
5. The next screen to appear is the calibration screen. The first line shown is the
concentration of the calibration solution/cylinder. The second line is the
temperature (22 to 28 C). The third line is the calibrated value (to be tested).
6. Place a new mouthpiece in the MARK V and Connect the MARK V to an
approved wet bath simulator/cylinder. Use calibration adaptors between the
mouthpiece and simulator (or cylinder) for an airtight fit, if necessary. Highlight
“Cal” on the screen using the left (red) button. Allow simulator/cylinder to run for
approximately 5 sec (simulator)/ 8-10 sec (dry-gas cylinder) and press the right
(green) button to select “cal” to acquire a sample. Remove from simulator. The
screen will briefly flash and the calibrated value in the third line will be changed
(this value is a mixture of letters and numbers).
7. Select to finish calibration.
8. Verify the calibration results by performing an accuracy check test.
Safety, Maintenance, & Warranty
Safety
For correct and effective use of the MARK V, it is essential to read and strictly follow
the instructions contained in this document. The MARK V is to be used only for the
purposes specified herein.

Maintenance
Repairs of the MARK V may only be performed by PAS Systems International, Inc. or
an authorized service technician. Only original MARK V parts may be used.

To keep the instrument clean, periodically use a mild disinfectant and a soft cloth on
the outside of the case. DO NOT use alcohol to clean the unit!

Warranty
The MARK V comes with a one (1) year limited parts and labor warranty, effective on the date of
purchase. The warranty does not cover batteries or accessories. The warranty does not cover
freight to the service facility, misuse or abuse of the product. Warranty is void if unit is found to have
been tampered with.

DATIA Member US DOT Approved CE Approved


MARK V Alcovisor®

PAS Systems International, Inc.


PO Box 330
Fredericksburg, VA 22404-0330

http://www.pasintl.com
Tel: (540) 372-3431
800-660-SNIF

Printed 04/13

You might also like