4 Els
4 Els
4 Els
M514 M516
AU400
EL Software Manual
Rev. 0700
Service Group
Mishina Olympus Optical Co. Ltd.
el_tc_01.doc
M514 EL SOFTWARE MANUAL
Table of Contents
2
4. Special operation
Operation for Service Engineers........................................................4-1
5. Mode outline
Mode outline ......................................................................................5-1 to 5-2
Outline of ISE mode...........................................................................5-3
Outline of operation mode..................................................................5-4 to 5-5
8. File configuration
Program file information....................................................................8-1
(* Documents being prepared by P.D.Dept.)
9. Initialize
ANL-DPR communication outline .....................................................9-1 to 9-5
LED control .......................................................................................9-6
Abstract of initialization.....................................................................9-7
Macro action ......................................................................................9-7 to 9-8
For control line...................................................................................9-8 to 9-11
Mechanism line control ......................................................................9-12 to 9-19
ISE initialization action......................................................................9-20 to 9-25
el_tc_01.doc
M514 EL SOFTWARE MANUAL
Table of Contents
3
0400 0300
es_tc_01.doc
M514 EL SOFTWARE MANUAL
Table of Contents
4
0400
es_tc_01.doc
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-1
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-2
for the use of this manual
n The abbreviations and symbols used for the electric parts (sensor, motor , etc.) described in this manual are
summarized here. Please check the relevant nomenclature here.
The symbols (S** and M**) conform to the mechanical and electrical parts and the current nomenclature
assigns different nomenclature to these different fields. Please check the lists at the end of this chapter.
2) Rack aspiration
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-3
for the use of this manual
n Rack transfer
3) Rack storage
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-4
for the use of this manual
n Sample dispense
1) Sample probe transfer (FA)
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-5
for the use of this manual
n Reagent dispense
1) Reagent probe transfer (FD)
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-6
for the use of this manual
n Reagent dispense
3) Reagent dispense line
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-7
for the use of this manual
n Mix
1) Mix rod transfer (FC)
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-8
for the use of this manual
n Mix
2) Mixing rod washing
n Wash line
1) Wash nozzle (FB)
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-9
for the use of this manual
n Wash line
2) Cuvette washing
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-10
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-11
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-12
for the use of this manual
n ISE
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-13
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-14
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-15
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 0 Caution item and method 0-16
for the use of this manual
ES00_01.DOC
M514 EL SOFTWARE MANUAL
1-1
Chapter 1 Configuration of AU400 program
1. MS-Windows NT (Workstation)
This is the basic OS for the AU400 system. The AU400 program functions as an application running under this OS.
6. Individual firmware
Each optical communication secondary station connected to the network has an exclusive program in ROM.
ES01_01.DOC
M514 EL SOFTWARE MANUAL
1-2
Chapter 1 Configuration of AU400 program
• Turn off the power source to the AU400 and then turn it on. The system automatically forwards the program to the
SRAM of the ANL CPU PCB. ( See (1) in the figure below.)
• If using the Automatic Start function, the system forwards the program as well.
If the system is turned on manually before you set the Automatic Start function, the system forwards the program as
well.
• If the normal END process is operated and then the system turns off, the ANL CPU is continuously supplied power.
So the power does not need to be turned on to forward the program again. However if the power source is turned of
by E. STOP or the breaker, or if power is cut to the facility, it is necessary to forward the program again the next
time the power is turned on. There is no need to be aware of the above condition, because the system senses the
condition when the power is cut (E.STOP, breaker) and then automatically forwards the program.
* While the system forwards the program, the download condition can be monitored by the 8 bit LED. For more
information refer to page 9-6.
(Caution !) If there is no communication between DPR and ANL (including ISE) for more than 10 minutes, the
program cannot be loaded to the ISE because communication is automatically disconnected. (It must be restarted.)
• Turn off the power to the AU400 and then turn it on. By executing [ISE PROGRAM INSTALL] in the System
Maintenance Program, the program is forwarded to the flash ROM of the ISE Control PCB through the ANL PCB.
( See (2) in the figure below.)
• Dissimilar to the ANL Control Program, program forwarding is not executed without directions. Exercise care,
because even though the new ISE Control Program is installed in the DPR, it is not updated.
* For more information about program loading, please refer to “ System Maintenance ”.
Install Program
(1) (2)
ANL CPU
MV2074
(2)
ISE Controller
MV2190
ES01_01.DOC
M514 EL SOFTWARE MANUAL
1-3
Chapter 1 Configuration of AU400 program
2. MPU (which can run Windows-NT and conforms to the following requirement)
• Pentium 166MHz or better
3. RAM (which can run Windows-NT and conforms to the following requirement)
• 48MB or more (64 MB is recommended)
4. FD
• 3.5inch, 1 drive minimum
5. CD
• 1 drive minimum
7. Mouse (which can run Windows-NT and conforms to the following requirement)
• PS/2 Mouse which can be recognized Windows-NT
8. Sound board (which can run Windows-NT and conforms to the following requirements)
• Sound board which can be recognized by Windows-NT
• With adjustable volume speaker
9. CRT (which can run Windows-NT and conforms to the following requirements)
• CRT which can be recognized by Windows-NT
• 15 or 17 inch color
• Resolution 1024 x 768 pixels and 65,536 colors
10. Touch Panel (option) (functions with Windows-NT standard and conforms following requirements)
• Touch Panel which can be recognized by Windows-NT
• Resolution can be set to 1024 x 768
• Windows-NT can recognize it as a mouse
• SIO standard and can be connected to extension board 1 or extension board 2
11. Modem (option) (which can run Windows-NT and conforms to the following requirements)
• Modem which can be recognized by Windows-NT
• SIO standard and can be connected to extension board 1 and extension board 2
ES01_01.DOC
M514 EL SOFTWARE MANUAL
1-4
Chapter 1 Configuration of AU400 program 0200 0100
1) OAI
Port Name Type Use Port Address IRQ
COM 1 Standard ANL Communication 0x3F8~0x3FF 4
COM 2 Standard Online Communication 0x2F8~0x2FF 3
COM 3 A Modem with a built-in Modem 0x3E8~0x3EF 10
2) ODH
Port Name Type Use Port Address IRQ
COM 1 Standard ANL Communication 0x3F8~0x3FF 4
COM 2 Standard Online Communication 0x2F8~0x2FF 3
COM 3 A Modem with a built-in Modem 0x3E8~0x3EF 7
COM 4 Extension B. 1 Serial 1 Touch Panel 0x2E8~0x2EF 11
3) ASIA
Port Name Type Use Port Address IRQ
COM 1 Standard ANL Communication 0x3F8~0x3FF 4
COM 2 Standard Online Communication 0x2F8~0x2FF 3
COM 3
COM 4
13. Parallel Port (functions with Windows-NT standard and conforms to the following requirements)
Port Name Type Use Port Address IRQ
LPT 1 Standard Standard Printer 0x378~0x37F 7
LPT 2 Extension B. Optional Printer 0x278~0x27F (5)
ES01_01.DOC
M514 EL SOFTWARE MANUAL
1-5
Chapter 1 Configuration of AU400 program 0400 0300 0200
ES01_01.DOC
M514 EL SOFTWARE MANUAL
1-6
Chapter 1 Configuration of AU400 program 0600 0500 0400 0300
ES01_01.DOC
M514 EL SOFTWARE MANUAL
3-1
Chapter 3 Sample Work Sheet Management
n Operation explanation
It’s described the menu operation inquired many times in the operation menu.
The contents is added on demand.
n Contents of explanation
Menu Contents Page
Sample worksheet management Inquiry flow of test requisition 3-2 to 3-19
Inquiry list for each analysis type
& sample type & analysis mode
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-2
Chapter 3 Sample Work Sheet Management
n Inquiry flow
1. Inquiry Flow
(1) Normal sample
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-3
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-4
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-5
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-6
Chapter 3 Sample Work Sheet Management
n Inquiry list
The item inquiries for each analysis and sample type will be explained.
Confirm the analysis type, the sample type, and the analysis mode to which reference is to be made from the
following table, find the corresponding number (C-1-1, I-1-1, etc.), and confirm the detailed contents via the
corresponding text.
Analysis type Sample type Analysis mode
Sequential Rack No. Bar code
Normal analysis General sample C-1-1 (p.3-7) C-1-2 (p.3-9) C-2-1 (p.3-13)
Emergency sample C-1-1 (p.3-7) C-1-2 (p.3-9) C-2-1 (p.3-13)
STAT sample C-1-3 (p.3-11) C-1-3 (p.3-11) C-2-3 (p.3-15)
Standard reanalysis General sample i-1-1 (p.3-17) i-1-1 (p.3-17) i-2-1(p.3-20)
Emergency sample i-1-1 (p.3-17) i-1-1 (p.3-17) i-2-1(p.3-20)
STAT sample i-1-2 (p.3-19) i-1-2 (p.3-19) i-2-2 (p.3-21)
Reanalysis with STAT table use i-1-2 (p.3-19) i-1-2 (p.3-19) i-2-2 (p.3-21)
Automatic reanalysis General sample J-1-1 (p.3-22) J-1-1 (p.3-22) J-2-1 (p.3-23)
Emergency sample J-1-1 (p.3-22) J-1-1 (p.3-22) J-2-1 (p.3-23)
STAT sample − − −
Reanalysis with STAT table use − − −
RACK Reagent blank K-1-1 (p.3-24) K-1-1 (p.3-24) K-1-1 (p.3-24)
Calibration K-1-1 (p.3-24) K-1-1 (p.3-24) K-2-1 (p.3-25)
Accuracy management sample K-1-1 (p.3-24) K-1-1 (p.3-24) K-2-1 (p.3-25)
STAT manual Reagent blank L-1-1 L-1-1 L-1-1
Calibration L-1-1 L-1-1 L-2-1
Accuracy management sample L-1-1 L-1-1 L-2-1
STAT automatic Reagent blank − − −
Calibration − − −
Accuracy management sample − − −
• The mixture type (serum, urine, other) is discriminated by comparing the rack No. information set by the
system conditions with the rack No. sent at the time of an inquiry from ANL.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-7
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1 Already received processing 1
Cup present Allotted S. No. received, Analyzed, In analysis
l Inspection reception (batch input)
For received and analyzed samples, the item selection information is created, and response is made with “cup
present”.
Internal flag selection is possible for the following processing.
l Real-time on-line inquiry
1) An inquiry to the host is made with the larger one of start S No. and max. received S No., the item
selection information from the host is recorded in the work list, and a response is made.
3-1-2 Already received processing 2
Cup present Allotted S. No. received, Not analyzed
l Inspection reception (batch input)
For received and not analyzed samples, the item selection information is created, and response is made with
“cup present”.
l Real-time on-line inquiry
For each sample type, internal flag selection is possible for the following processing.
1) An inquiry is made to the host, the item selection information from the host is recorded in the work list,
and a response is made.
2) No inquiry is made to the host in the same way as inspection reception (batch input).
3-2 Already received processing 3
No cup, allotted S. No. received
l Inspection reception (batch input)
NOP response when there is no cup while an S. No. has been allotted.
An alarm is caused.
l Real-time on-line inquiry
NOP response when there is no cup while an S. No. has been allotted.
No inquiry is made to the host.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-8
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-9
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Cup present, allotted S. No. received, Analyzed, In analysis
l Inspection reception (batch input)
The item selection information is created for the received and analyzed samples, and a “cup present” response
is made.
l Real-time on-line inquiry
1) No inquiry is made to the host in the same way as inspection reception (batch input).
3-1-2
Cup present, allotted S. No. received, Not analyzed
l Inspection reception (batch input)
For received and not analyzed samples, the item selection information is created, and response is made with
“cup present”.
l Real-time on-line inquiry
1) An inquiry is made to the host, the item selection information from the host is recorded in the work list,
and a response is made.
3-2
No cup, the allotted S. No. has been received.
l Inspection reception (batch input)
NOP response when there is no cup while an S. No. has been allotted.
An alarm is caused.
l Real-time on-line inquiry
NOP response when there is no cup while an S. No. has been allotted.
An alarm is caused.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-10
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-11
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Cup present and allotted S. No. received, Analyzed, In analysis
l Inspection reception (batch input)
The item selection information is processed from the received and analyzed samples, and a “cup present”
response is made.
l Real-time on-line inquiry
1) No inquiry is made to the host in the same way as inspection reception (batch input).
3-1-2
Cup present and allotted S. No. received, Not analyzed
l Inspection reception (batch input)
The item selection information is processed from the received and not analyzed samples, and a "cup present"
response is made.
l Real-time on-line inquiry
1) No inquiry is made to the host in the same way as inspection reception (batch input).
3-2
No cup and the allotted S. No. has been received.
l Inspection reception (batch input)
• When there is no cup, there is no inquiry from ANL.
l Real-time on-line inquiry
• When there is no cup, there is no inquiry from ANL.
4-1 Processing of not received samples 1-1
Cup present and allotted S. No. not received
l Inspection reception (batch input)
Automatic generation is made with the allotted S. No. and a “cup present” response is made.
l Real-time on-line inquiry
An inquiry is made to the host, the item selection information from the host is recorded in the work list, and a
response is made.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-12
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-13
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
A sample of the mixture type and the S. ID has been received, Not analyzed (all items)
l Inspection reception (batch input)
• The sample information is retrieved from the bar code and the mixture type, and a response is made with
the retrieved S. No. and the created item information.
An alarm is issued.
l Real-time on-line inquiry
1) The sample information is retrieved from the bar code and the mixture type, and when a corresponding
sample is present, the retrieved S. No. and the item selection information are given as response.
Alarm is issued.
When there is no corresponding sample, an inquiry is made to the host with the larger one of start S No.
and max. received S No., the item selection information from the host is memorized, and a response is
made.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-14
Chapter 3 Sample Work Sheet Management
3-1-2
A sample of the mixture type and the S. ID has been received, Analyzed (there are not analyzed items), In analysis
l Inspection reception (batch input)
1) When received and not analyzed samples with the same ID are retrieved, the item selection information is
created and a response is made.
If none are found, automatic generation is made with the larger one of start S No. and max. received S No.
+ 1.
l Real-time on-line inquiry
1) When received and not analyzed samples with the same ID are retrieved, the item selection information is
created and a response is made.
When none are found, an inquiry to the host is made with the larger one of start S No. and max. received S
No., the item selection information from the host is memorized, and a response is made.
4-1 Processing of not received samples
l Inspection reception (batch input)
• S. No. allotment is made sequentially from a number set by internal parameter, automatic generation is
made from the larger one of start S No. and max. received S No. + 1, and a “cup present” response is made.
l Real-time on-line inquiry
1) An inquiry to the host is made with the larger one of start S No. and max. received S No., the item
selection information from the host is memorized, and a response is made.
5. Bar Code Read Error Processing
l Inspection reception (batch input)
Automatic generation is made with the larger one of start S No. and max. received S No. + 1, and a “cup
present” response is made.
(The error No. is inserted for the S. ID.)
l Real-time on-line inquiry
1) Automatic generation is made with the larger one of start S No. and max. received S No. + 1, and a “cup
present” response is made.
6. Real-time on-line error processing
l Real-time on-line inquiry
1) Automatic generation is made with the larger one of start S No. and max. received S No. + 1, and a “cup
present” response is made.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-15
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
A sample of the mixture type and the S. ID has been received., Not analyzed (all items)
l Inspection reception (batch input)
• The sample information is retrieved from the bar code and the mixture type, and a response is made with
the created item information.
(The mixture type is decided from the STAT table layout set in advance.)
l Real-time on-line inquiry
1) An inquiry is made with the larger one of start S No. and max. received S No. + 1 for each mixture type
decided from the STAT table layout.
3-1-2
A sample of the mixture type and the S. ID has been received., Analyzed (there are not analyzed items)
l Inspection reception (batch input)
1) When received and not analyzed samples with the same ID are retrieved, the item selection information is
created and a response is made. When none are found, automatic generation is made with the larger one of
start S No. and max. received S No. + 1, and a “cup present” response is made.
l Real-time on-line inquiry
The following processing can be selected by internal flag.
1) An inquiry to the host is made with the larger one of start S No. and max. received S No., the item
selection information from the host is memorized, and a response is made. Alarm is issued.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-16
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-17
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Cup present, allotted reanalysis No. received, Analyzed
l Inspection reception (batch input)
The item selection information is created from the received and analyzed samples, and a “cup present”
response is made.
l Real-time on-line inquiry
1) An inquiry is made to the host, the item selection information from the host is recorded in the work list,
and a response is made.
3-1-2
Cup present, allotted S. No. received, Not analyzed
l Inspection reception (batch input)
The item selection information is created from the received and not analyzed samples, and a “cup present”
response is made.
l Real-time on-line inquiry
1) An inquiry is made to the host, the item selection information from the host is recorded in the work list,
and a response is made.
3-2 Processing 2 for the same S. No.
No cup, the allotted reanalysis No. has been received
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
An alarm is caused.
l Real-time on-line inquiry
No cup with the allotted S. No., NOP response
No inquiry to the host
An alarm is caused.
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-18
Chapter 3 Sample Work Sheet Management
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-19
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Cup present, allotted reanalysis No. received, Analyzed, In analysis
l Inspection reception (batch input)
The item selection information is created from the received and analyzed samples, and a “cup present”
response is made.
l Real-time on-line inquiry
• In the same way as inspection reception (batch input), no inquiry is made to the host.
3-1-2
Cup present, allotted S. No. received, Not analyzed
l Inspection reception (batch input)
The item selection information is created from the received and not analyzed samples, and a “cup present”
response is made.
l Real-time on-line inquiry
• In the same way as inspection reception (batch input), no inquiry is made to the host.
3-2
No cup, allotted reanalysis No. received
l Inspection reception (batch input)
No inquiry
l Real-time on-line inquiry
• Same as inspection reception (batch input).
4-1 Processing 1-1 for not received samples
Cup present, allotted reanalysis No. not received
l Inspection reception (batch input)
Cup present, NOP response Alarm is issued. No. count up
l Real-time on-line inquiry
An inquiry is made to the host, the item selection information from the host is recorded in the work list, and a
response is made.
4-2 Processing 1-2 for not received samples
No cup, allotted reanalysis S.No. not received
l Inspection reception (batch input)
No inquiry
l Real-time on-line inquiry
• Same as inspection reception (batch input).
5. Real-time on-line error processing
l Real-time on-line inquiry
NOP response to ANL
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-20
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Sample of mixture type and S. ID received, Not analyzed (all items)
l Inspection reception (batch input)
• The sample information is retrieved from the bar code and the mixture type, and a response is made with
the retrieved S. No. and the created item information.
l Real-time on-line inquiry
The sample information is retrieved from the bar code and the mixture type, and when there is a corresponding
sample, a response is made with the retrieved S. No. and the item selection information.
(No inquiry is made to the host.)
3-1-2
(Same S. ID processing)
Sample of mixture type and S. ID received, Analyzed (there are not analyzed items), In analysis
l Inspection reception (batch input)
The item selection information is created from the received and analyzed samples, and a response is made.
l Real-time on-line inquiry
The item selection information is created from the received and analyzed samples, and a response is made.
4-1 Processing of not received samples
l Inspection reception (batch input)
Cup present, NOP response Alarm is issued.
l Real-time on-line inquiry
An inquiry to the host is made with the larger one of start S No. and max. received S No., the item selection
information from the host is memorized, and a response is made.
5. Bar Code Read Error Processing
l Inspection reception (batch input)
NOP response Alarm is issued
l Real-time on-line inquiry
NOP response
6. Real-time on-line error processing
l Real-time on-line inquiry
NOP response
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-21
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Sample of mixture type and S. ID received, Not analyzed (all items)
l Inspection reception (batch input)
In regard to the sample of the “cup present” position, DPR
• retrieves the sample information from the bar code and the mixture type and makes a response with the
retrieved S. No. and the item information.
l Real-time on-line inquiry
The sample information is retrieved from the bar code and the mixture type, and when there is a corresponding
sample, a response is made with the retrieved S. No. and the item selection information.
(No inquiry is made to the host.)
3-1-2
(Same S. ID processing)
Sample of mixture type and S. ID received, Analyzed (there are not analyzed items), In analysis
l Inspection reception (batch input)
The item selection information is created from the received and analyzed samples, and a response is made.
l Real-time on-line inquiry
The item selection information is created from the received and analyzed samples, and a response is made.
4-1 Processing of not received samples
l Inspection reception (batch input)
Cup present, NOP response Alarm is issued.
l Real-time on-line inquiry
An inquiry to the host is made with the larger one of start S No. and max. received S No., the item selection
information from the host is memorized, and a response is made.
5. Bar Code Read Error Processing
l Inspection reception (batch input)
NOP response Alarm is issued.
l Real-time on-line inquiry
NOP response
6. Real-time on-line error processing
l Real-time on-line inquiry
NOP response
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-22
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Reanalysis work list registration, during reanalysis analysis, Reanalysis completed
l Inspection reception (batch input)
The reanalysis S. No. is used as the retrieved No., and NOP response is made.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
3-1-2
Reanalysis work list registered, reanalysis not analyzed
l Inspection reception (batch input)
The item information is created from the registered and not analyzed samples, and a response is made to ANL.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
3-2
No reanalysis work list registration
l Inspection reception (batch input)
NOP response
l Real-time on-line inquiry
• Same as inspection reception (batch input).
Investigation
The sample No. is taken as the registered No. + 1, the original sample No. is retrieved from the DPR
identification key, and an inquiry is made to the host.
3-3
Reanalysis work list registration, during reanalysis analysis, Reanalysis completed
l Inspection reception (batch input)
The item information is created from the registered and not analyzed samples, and a response is made to ANL.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
4. Real-time on-line error processing
l Real-time on-line inquiry
NOP response to ANL
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-23
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1-1
Reanalysis work list registration, during reanalysis analysis, Reanalysis completed
l Inspection reception (batch input)
The reanalysis S. No. is used as the retrieved No., and NOP response is made.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
3-1-2
Reanalysis work list registration, reanalysis not analyzed
l Inspection reception (batch input)
The item information is created from the registered and not analyzed samples, and a response is made to ANL.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
3-2
No reanalysis work list registration, no reanalysis analysis data
l Inspection reception (batch input)
NOP response
l Real-time on-line inquiry
• Same as inspection reception (batch input).
Investigation
The sample No. is taken as the registered No. +1, the original sample No. is retrieved from the DPR
identification key, and an inquiry is made to the host.
3-3
Reanalysis work list registration, during reanalysis analysis, Reanalysis analyzed
l Inspection reception (batch input)
The item information is created from the registered and not analyzed samples, and a response is made to ANL.
l Real-time on-line inquiry
No inquiry is made to the host, and the same processing as inspection reception (batch input) is performed.
4. Real-time on-line error processing
l Real-time on-line inquiry
NOP response to ANL
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-24
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1
When a cup is present and the sample corresponding to the allotted cup position has been received (independent of
whether analysis has been performed or not)
l Inspection reception (batch input)
The item selection information is created from the received samples, and a “cup present” response is made.
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
3-2
When no cup is present and the sample corresponding to the allotted cup position has been received (independent
of whether analysis has been performed or not)
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
An alarm is caused.
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
4-1 Processing 1-1 for not received samples
Cup present, and the sample corresponding to the cup position has not been received
l Inspection reception (batch input)
One of the following responses with the allotted S. No.
Selection is performed by internal flag.
Cup present, NOP response
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
4-2 Processing 1-2 for not received samples
No cup, allotted S. No. not received
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-25
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1
When a cup is present and the sample corresponding to the allotted cup position has been received (independent of
whether analysis has been performed or not)
l Inspection reception (batch input)
The item selection information is created from the received samples, and a “cup present” response is made.
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
3-2
When no cup is present and the sample corresponding to the allotted cup position has been received (independent
of whether analysis has been performed or not)
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
An alarm is caused.
(is calculated from the rack No. and the cup position, and the item information is determined.)
4-1 Processing 1-1 for not received samples
Cup present, and the sample corresponding to the cup position has not been received
l Inspection reception (batch input)
One of the following responses with the allotted S. No.
Selection is performed by internal flag.
Cup present, NOP response
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
4-2 Processing 1-2 for not received samples
No cup, allotted S. No. not received
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
5. Bar code read error
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
ES03_01.DOC
M514 EL SOFTWARE MANUAL
3-26
Chapter 3 Sample Work Sheet Management
Inquiry/Response Processing
3-1
When a cup is present and the sample corresponding to the allotted cup position has been received (independent of
whether analysis has been performed or not)
l Inspection reception (batch input)
The item selection information is created from the received samples, and a “cup present” response is made.
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
3-2
When no cup is present and the sample corresponding to the allotted cup position has been received (independent
of whether analysis has been performed or not)
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
An alarm is caused.
(is calculated from the rack No. and the cup position, and the item information is determined.)
4-1 Processing 1-1 for not received samples
Cup present, and the sample corresponding to the cup position has not been received
l Inspection reception (batch input)
One of the following responses with the allotted S. No.
Selection is performed by internal flag.
Cup present, NOP response
(The kind No. is calculated from the rack No. and the cup position, and the item information is determined.)
4-2 Processing 1-2 for not received samples
No cup, allotted S. No. not received
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
5. Bar code read error
l Inspection reception (batch input)
No cup with the allotted S. No., NOP response
ES03_01.DOC
M514 EL SOFTWARE MANUAL
4-1
Chapter 4 Special operation
• Password
The setting below is reserved by the system.
ES04_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 5 Mode outline
5-1
n Mode outline
• Types
Below are the main analyzer modes.
Modes General Contents
POWER OFF If no electricity is supplied to the CPU Power Supply for ANL. (Software control is not
possible.)
POWER ON IPL, etc., are under control. (No mode will be displayed.)
INITIALIZE Ready to run.
STOP No action is taken.
MEASURE Running.
END End (or preparing to end).
• Initialization processing
Initialization will be performed as follows.
1. Initialization of internal peripheral, relocation, and register.
2. Initialization of internal peripheral selective register.
3. Initialization of timer and clock selective register.
4. Initialization of timer / counter and unit.
5. Initialization of interrupt control and unit.
6. Initialization of DPR communication.
7. Initialization of universal IO.
8. Others (necessary software initialization)
• Setting (0)... ‘check = YES’ in the ISE Overflow Sensor Check Counter
(ISE Overflow Sensor Check Counter will countdown if other setting than ‘check = YES’).
• Inside condition
The STOP & MEASURE mode can be divided into the following modes.
1) STOP
ES05_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 5 Mode outline
5-2
2) MEASURE
® Action condition
Name Description
During sample The rack or STAT sample is transferred.(The condition is managed individually.)
transfer
During reagent The reagent (R1, R2) is dispensing. (The condition is managed individually.)
dispense
During reaction Sample and reagent dispensing has ended and available cuvette data being sent.
photometry
During wash Available cuvette data sent and washing used cuvette.
During STOP All processing ends and stops.
ES05_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 5 Mode outline
5-3
2) Mode shift
*1 During STOP initialization is performed by 1) shifting the mode to STANDBY (ANL) or 2) pressing
the ISE READY key.
If initialization ends normally, the mode shifts to READY.
Initialization with no connections executes control line initialization and mechanism line initialization of
ISE.
Initialization if a connection exists executes mechanism line initialization.
*2 The mode shifts to STOP if 1) the mode shifts to STOP (ANL), or if 2) the ISE STOP key is pressed, or
if 3) the action ends abnormally (including communication error).
*3 The instruction (including MEASURE, W1, manual operations for ISE) resulted in the BUSY mode
being shifted to.
*4 If the action ends normally (including MEASURE, W1) the mode READY is shifted to.
ES05_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 5 Mode outline
5-4
ES05_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 5 Mode outline
5-5
Continued ...
Mode Contents
PAUSE Suspending analysis due to a PAUSE request or generation of RACK FULL. ( Although analysis
is being performed, rack feeding is suspended.)
Analysis can be restarted after resolving the cause.
STOP Stopping analysis forcibly due to RACK JAM, CUVETTE WHEEL ERROR (still can be
rotated), CUVETTE ROTATION IMPOSSIBLE, or WATER SUPPLY (or EVACUATION)
ERROR errors.
W1 Performing W1 operation (washing all cuvettes and bleeding air bubbles of sample and reagent
probes) by pressing W1 key.
W2 Performing W2 operation (washing all cuvettes and probes with detergent) by pressing W2 key.
PHOTOCAL Performing PHOTOCAL operation (measuring cuvette blank) by pressing PHOTOCAL key.
MAINTENANCE Maintenance operation mode for customer and for instrument check by the service engineer. The
following menus are available in this mode.
Maintenance Shift to the maintenance mode from the WARM UP, STANDBY, or STOP
mode.
There are two types: one is for data processing and the another is for analysis.
Manufacturer The maintenance mode can be shifted to while in the STOP mode.
Maintenance There are five types: Program Version, Analyzer Diagnostics, ISE
Diagnostics, Alarm Log, and File.
END End process of the system.
ES05_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 6 Mode Shift Drawing
6-1
* The above describes the shifts that take place due to user operation (SW, Key, and Menu operation).
Other shifts depend on the specifications of each condition.
ES06_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 6 Mode Shift Drawing
6-2
*1 The mode will not shift to the STOP mode during MAINTE. The mode will shift to STOP
after completing MAINTE.
If the mode shifts to MAINTE while in the STOP mode, the mode will shift to STOP after completing
MAINTE.
ES06_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 6 Mode Shift Drawing
6-3
*1 The mode will not shift to the STOP mode during MAINTE. The mode will shift to STOP after
completing MAINTE. If the mode shifts to MAINTE while in the STOP mode, the mode will shift
to STOP after completing MAINTE.
ES06_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 6 Mode Shift Drawing
6-4
(3) During MEASURE (except for AUTO REPEAT run of normal/emergency samples)
Note 1) No mode shifting will take place even if ISE errors are generated during analysis.
Note 2) The mode will shift to STOP if mechanical errors are detected when performing a
reagent check while in the PAUSE mode.
Note 3) The mode will shift to STOP if mechanical errors in the S-Probe are detected when
performing ISE manual operation while in the PAUSE mode.
ES06_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 6 Mode Shift Drawing
6-5
Note 1) No mode shifting will take place even if ISE errors are generated during analysis.
Note 2) The mode will shift to STOP if mechanical errors are detected when performing
reagent check while in the PAUSE mode.
Note 3) The mode will shift to STOP if mechanical errors in the S-Probe are detected when
performing ISE manual operation while in the PAUSE mode.
ES06_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-1
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-2
Continued,
ANL DIAG ANL DIAG command using MAINTE or manufacturer’s maintenance menu
ISE DIAG ISE DIAG command using MAINTE or manufacturer’s maintenance menu
STANDBY System termination command
W1 STOP STOP request (STOP/STDBY key) or abnormal ending
WARM UP W1 ended normally from WARM UP mode
STANDBY W1 ended normally from STANDBY mode
W2 STOP STOP request (STOP/STDBY key) or abnormal ending
WARM UP W2 ended normally from WARM UP mode
STANDBY W2 ended normally from STANDBY mode
PHOTOCAL STOP STOP request (STOP/STDBY key) or abnormal ending
WARM UP Photometry ended
STANDBY Photometry ended
MAINTE STOP STOP request (STOP/STDBY key) or errors during MAINTE
WARM UP MAINTE display ended
STANDBY MAINTE display ended
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-3
(1) START (2) FEEDER STOP (3)PAUSE (4) END (5) STOP/ STANDBY (6) FORCED STANDBY (7) W1
(8) W2 (9) PHOTOCAL (10) REAGENT CHECK, STAT CHECK (11) ISE MANUAL (12) DIAG
(13)MAINTENANCE (14) STAT
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-4
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-5
n Initial
1. Shifting condition
1) Shifts to the first mode after the power is turned on.
2) Shifts by the STANDBY command from the stop mode.
2. Initialization process
1) If the ISE mode is READY, set it to BUSY.
2) Initialize the control line.
‹ Initialize the photometry slave.
› Initialize the LAN master.
fi Initialize ISE/ Barcode communication.
fl If ISE designation has been made, the power has been turned on, and the ISE mode is not connected,
initialize the ISE.
If initalization results in an error, leave in not connect state and execute following process.
Initialize R. and ID barcode reader.
– Initialize the mechanical control system.
3) If shifting occurs after the power has been turned on, check for power failure.
4) Initialize all mechanisms.
5) Execute the neccessary procedure to transfer racks from the inside to the outside.
6) Check vacuum pressure.
3. Normal process
1) If there is no detergent solution, and if no “no DI water" or "no detergent" message has been issued, make
the solution.
If an error results, do not execute the process described under 2).
2) If a STOP command from DPR, process 5 issues a command to shop the operation. Stop other processes at
the appropriate step of each process.
*Only execute the processess decribed above.
4. Completion condition
Shift to the STOP mode after nornal/ abnormal completion of the inital process, or by issuing a STOP
command from DPR.
5. Completion process
If ISE initalization has been normally completed, set the ISE mode to READY for connection.
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-6
n Stop mode
1. Shifting condition
1) Completion of the initial mode.
2) Completion of the MEASURE mode.
2. Initial process
1) Drain waste liquid from the vacuum tank. (Do not wait until completion.)
2) Determine the internal condition by referring to the shifting condition.
3) Set the magnetization to off at the STAT of rotation.
3. Normal process
1) Environmental change processing
Item Processing
Temparature steady Shift the mode to STAND-BY, if the internal mode is WARM UP.
(Requires 37°C for 20 minutes.) ( If started from E.STOP, 90 minutes are
required.)
No detergent solution Make the detergent solution if not notified to be "no DI water" or "no
detergent".
Reagent compartment Set the magnetization of the reagent conpertment lid to off.
lid open
STAT switch pushed Notify DPR, if the internal mode is STANDBY or PAUSE.
STAT ROT SW Make the STAT rotate if the internal mode is WARM UP, STANDBY or
(DIAG SW) pushed PAUSE.
If the DIAG process is designated with a trigger, execute the selected process.
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-7
*1: If performing STAT manually, turn LED off, then at completion turn the LED on.
*2: Maintain the status when performing analysis.
4. Completion condition
• With the STANDBY instruction from DPR, shift to the initial mode.
• With the START instruction from DPR, shift to the MEASURE mode.
• With the System End instruction from DPR, shift the mode to END.
5. Completion process
1) Stop the vacuum tank waste pump operation.
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-8
n MEASURE mode
1. Shifting condition
1) Shifted by the condition commands (START) from the STOP condition.
2. Initialization process
1) Start check
(1)Check the water supply line.
If original detergent A or detergent B is inadequate, an error is issued.
If DI water, detergent water A, or detergent water B is inadequate, an error is issued.
(2)Check the waste water line.
If the vacuum tank is full, an error is issued.
If an option exists when the concentrated waste becomes full, an error is issued.
If an option exists when the non-concentrated waste becomes full, an error is issued.
2) Initialize the Barcode reader, STAT, and ID.
3) Initialize the photometric constant.
4) Execute the 0% Photometry operation.
5) Execute Start Initialization.
6) Confirm the samples for analysis.
(1) If it concerns a Rack start command, execute rack confirmation operation.
If there is no rack, an error is issued.
(2) If it concerns a STAT start command, perform STAT cup confirmation.
If no cup, an error is issued.
7) The procedure for setting a negative pressure is performed.
8) Turn the bellows of the pump on.
DI water (P1)
Degasser (P3)
Detergent solution A (P4)
Detergent solution B (P5)
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-9
3. Normal process
• If it is during washing after analysis, stop it, then perform washing before
analysis.
• If the above operations are completed normally, instruct the analysis
• If it is during washing after analysis, stop it, then perform washing before
analysis.
• If the above operations are completed normally, instruct the analysis
Status command If the status after shifting is not defined, command the analysis operation to
(FEEDER STOP) stop the rack supply.
Status command If it is during R1 reagent dispensing and the status after shifting is not defined
(PAUSE) or is TO STANDBY, instruct the analysis operation to stop reagent
dispensing.
Status command If ISE is performing measurement, wait until it has been completed.
(ISE STOP) After measurement has been completed, shift the ISE mode to STOP, don’t
perform further measurement.
ES07_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 7 Mode shift triggers
7-10
4. Completion condition
1) Shift to the STOP status initialization it abnormally completed during initialization processing.
• The internal mode of the STOP status is defined by the STOP status specification.
(If there are no mechanical abnormalities, it returns to the former internal status before shifting.)
2) It shifts to the STOP status by DPR instruction (STOP) or if the analysis operation is completed
abnormally.
3) It shifts to the STOP status after all analysis operations have been completed (after analysis wash).
5. Completion process
1) The procedure for negative pressure cancellation is performed.
2) The pump turned on in the initialization process is turned off.
n End mode
1. Shifting condition
1) Shifts from the STOP status by status instruction (END).
2. Initialization operation
1) If automatic power on is not set, execute the following operations.
‹ Fill in the POWER-OFF information in the address assigned.
• Detection of EM. STOP
• Normal conpletion of cuvettes (= empty)
› Interruption not permitted
fi Turn the mode of POWER OFF of ANL control PCB into ON.
fl Wait untill the mode of POWER ON of ANL control PCB turns OFF.
Turn the ANL POWER OFF mode control PCB off.
– Set it as permanent loop.
3. Completion process
Shift to the turned off status by executing a hardware reset.
ES07_01.DOC
M514 EL SOFTWARE MANUAL
8-1
Chapter 8 File configuration
ES08_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-1
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-2
(on-line )
• Memorize information requisition
• Sample preparation
Tube, rack
Start requisition • Memorize information requisition
Index, round, start no. • Switch input (STAT start)
• Analysis start notice STAT request
START key • Start analysis
analysis parameter (1,6)
(if completed abnormally) status indication(START)
analysis parameter (others)
• Start check
• Starting initialization
(completed abnormally)
alarm request
• Mode indication status notice (STOP/STANDBY)
analysis parameter (end)
• Disregard analysis parameter
START key
(If completed normally) • Start analysis
analysis parameter(1,6)
status indication(START)
(STAT, rack order) • Start check
analysis parameter (others) • Start initialization (or STAT check during
(completed normally) STAT)
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-3
• Interruption of analysis
PAUSE key
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-4
(automatic operation)
• Analysis data calculation analysis data
repeat run judgment ANALY
repeat run inquiry -SIS
on time after data output (10 samples)
repeat run information • Rack retrieve, rack collection
sample inquiry (rack I.D., position,
S.ID , rerun identification)
sample condition
when rerun stopped, the mode does not
change
• Auto QC, auto ACAL
manual instruction (forced) analysis condition if forced, it is started as an
(only during start) status instruction (START) ordinary start, and
the starting cause must be notified
• Check the starting cause
(during sampling only) • Analysis operation
(squeeze into samples)
same as ordinary analysis
• Data edition, correction
menu operation memorize data edited
• Error occurrence as (TO STOP)
• Mode indication status notice (TO STOP)
• Stop after analysis of the current
processing cuvettes
• Stop analysis
STOP key (STOP key is effective in the all mode)
mode instruction or if error occurrence required STOP
• Stop within 1 cycle
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-5
POWER
OFF
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-6
n LED control
If there is a communication error, the following indication is made (Controls the 8-bit LED as fllows).
llllllll
ERROR CODE
• ERROR CODE (first 4-bit)
l l l O : received not STX, if waiting for STX NOTE: O : ON/ lit
l l O l : SIO error, if receiving information l : OFF/ not lit
l l O O : SIO error, if sending information
l O l l : received character, if sending timing
l O l O : BCC error
l O O l : CODE error
l O O O : LEN (Text length ) error
O l l l : ENQ acknowledgment is not ACK
O l l l : acknowledgment is not ACK
• If forwarding program to ANL when the power is turned on (PHASE indication) (last 4-bit)
( ERROR ) ( PHASE )
MSB LSB
¬ If checking if loading exists. llll lOOl
- If starting sum check. llll lOOO
® If starting to send loading request. llll Olll
¯ If starting to receive ANL program. llll OllO
° If making notification of loading complete. llll OlOl
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-7
n Abstract of initialization
• Initialization type
There are three types of initialization.
1. Macro action
These are grouped in each function.
2. Control line
Initialization of the function for control of each device (control driver, etc.).
3. Mechanism line control
Initialization of the function for the mechanism for each device (action driver, etc.).
n1 Macro action
If an error occurs, the next action isn't executed.
Before the action, magnetization ON is output to all mechanisms operating.
(Ex.: If the reagent line is reset, reset the exitation for the fridge after reagent probe rotation and up/down results are
on.)
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-8
* The initialization position is the position managed by the driver and not the position detected by the sensor.
• Timer/counter unit
1) Counter 0
2) Counter 1
3) Counter 2
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-9
• DPR communication
1) Setting the clock (DPR communication baud rate setting / send and receive clock 9600 x 64)
2) Serial communication IC initialization
OUTPUT (DPR communication command) = 4Fh
• Baud rate 64 times clock
• Character length 8 bit
• STOP 1 bit
• Parity None
• Photometry control
1) Photometry sub - CPU unit initialization
2) Arcnet initialization
• LAN master
1. Control instruction " Reset " is output.
2. Waiting for "READY" for 1st station status.
If it doesn't become "READY" in 2 seconds, an error is displayed.
If the status is "disconnection,” “Send error,” “Receive error," and "board error” , an error is displayed.
3. If the status of the second station is "READY," the instruction for address table conversion is output.
If it is not "READY," an error is displayed.
If the status is "ERROR," an error is displayed.
* The setting for the second station for the physical address and logical address is the same.
4. Flag set instruction is output to each second station.
The condition is the same as ??.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-10
• Barcode reader
1) Sample/ Rack ID
A command for tentative change of setting is sent.
During sample ID
Par0 84h SecurityLevel(NO)SecurityZone(6)
Par1 ??h CodeTYPE select code set in sample ID
bit1 Code39 *All bits on for multi.
bit2 ITF(INT2of5)
bit3 code128(ISBT128)
bit4 Codabar/Monac(NW-7)
bit6 DTF(STD2of5)
Par2 80h Triggermode (Host), Condence Code 39 (Disable)
Par3 08h Code ID (Symbol ID), NR (Disable), Code 128 Function (Disable)
Scan Direction (Disable)
Par4 00h UPC/EAN AUTO-D (Disable), 5Digit (Disable), 2Digit (Disable)
ITF MOD10 (Disable), DTF MOD10 (Disable)
Par5 21h Decode Attempt Time (time out when reading) (1.0sec)
Inter Decode Time (1.0sec)
Par6 06h Power Mode (Continuous), Baud Rate (4800), Parity (Non)
Beeper(Disable)
Par7 00h ITF Length 1 (0)
Par8 00h ITF Length 2 (0)
Par9 00h DTF Length 1 (0)
Par10 00h DTF Length 2 (0)
Par11 80h Preamble (Not send)
Par12 0Dh Postamble-1 (CR)
Par13 0Ah Postamble-2 (LF)
b) On rack ID
(message contents)
Par1 ??h CodeTYPE select code set in rack ID
Par5 22h DecodeAttemptTime (time out when reading) (1.5sec)
InterDecodeTime(1.0sec)
2) STAT ID
A command for tentative change of setting is sent.
(message)
Same as " Sample ID. "
3) R. ID
A command for tentative change of setting is sent.
(message)
Par1 ??h Code TYPE Only the code set in R.ID is set.
Except for “Part 1” it is the same as sample ID.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-11
n ISE
• If unit configuration information is " ISE exist " and the power of the ISE is on, the following items are
executed.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-12
Caution
1) After moving to the origin, and if the depth input action is executed continuously, the sensor position isn’t
checked in the after-action check.
2) This statement describes the assignment of a OPMC abnormality check but the check has to be executed.
If the same motor operates continuously, a minimum of 200 ms is allowed for step out prevention.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-13
2) If the rack detection sensor (S7) is off, the rack is supplied one pitch * pulse.
•If the rack detection sensor (S7) is on, it an error results and no action takes place.
•If BUSY isn't canceled for the motor within one second, an error results.
•If BUSY isn't canceled for the motor within 500 ms, an error results.
•If BUSY isn't canceled for the motor within 2 seconds, an error results.
•If the ID mirror origin sensor (S9) stays off after an action has ended, an error results.
•If BUSY isn’t canceled for the motor within 2 seconds, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If BUSY isn’t canceled for the motor within 2 seconds, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If BUSY isn’t canceled for the motor within 2 seconds, an error results.
2) If the repeat run rack detection sensor (S20) is off, the retrieve action is performed for the pulses for 1 pitch *
pulses.
•If the repeat run rack detection sensor (S20) is on, an error results.
•If BUSY isn’t canceled for the motor within 1 second, an error results.
3) A little storage action is done the pulses for depth input * pulses.
•If the repeat run rack receive pitch sensor (S18) remains off after 500 ms, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-14
results.
•If BUSY isn’t canceled for the motor within 3 seconds, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
results.
•If BUSY isn’t canceled for the motor within 3 seconds, an error results.
•If the push out position sensor (S25) remains off after 2 seconds, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If the stop position sensor (S30) remains on after 500 ms, an error results.
3) Left movement to reset the pulses is performed until at the reset origin.
If the sample probe is in the STAT position and the upper point sensor is off, an error results and no action
takes place.
If the sample probe is in the STAT position and the up/down motor is BUSY an error results and no action
takes place.
•If BUSY isn’t canceled for the motor within 8 seconds, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If the specified point sensor (S30) remains off after 500 ms, an error results.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-15
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If BUSY isn’t canceled for the motor after an action, an error results.
•If the Cuvette position sensor (S34) remains on after an action, an error results.
•If the position check sensor (S35) remains on after an action, an error results.
•If the position check sensor (S35) remains off after 500 ms, an error results.
•If the upper point sensor (S36) remains on after 500 ms, an error results.
•If BUSY isn’t canceled for the motor after 500 ms, an error results.
•If the crash detection sensor remains on after 500 ms, an error results.
2) The S probe up/down motor (M11) is moved upward to reset for X pulses at reset upper point.
•If BUSY isn’t canceled for the motor after 3 seconds, an error results.
•If BUSY isn’t canceled for the motor after 500 ms, an error results.
•If BUSY isn’t canceled for the motor after 500 ms, an error results.
2) The dispenser discharge for X pulses at low speed at reset upper point.
•BUSY turns off for 5 seconds (timeout time).
•If the upper point sensor (S38) does not turn on at the end of an action, an error results.
•If the lower point sensor (S39) remains on after an action ends, an error results.
•When the motor BUSY isn’t canceled after 500 ms, it becomes error.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-16
• If BUSY is not canceled for the motor within 500 ms, set the error.
2) If the cover open sensor (S51) is shut, perform reset right motion at the reset cuvette position for * pulses.
If the cover open sensor (S51) is open, treat this as an error, and do not let it move.
If the upper point sensor (S46) is off, treat this as an error, and do not let it move.
• Wait 4 seconds for the Busy motor timeout.
• After completing the motion, and cuvette position sensor (S43) do not become on, set the error.
• After completing the motion, and when the position check sensor (S44) is on, set the error.
• After completing the motion and when the reagent position sensor (S45) is on, set the error.
3) Waits 200 ms
4) Perform a left movement for the washing position.
If the upper point sensor (S46) is off, treat this as an error, and do not let it move.
• After 500 ms, when the position check sensor (S44) has been set to off, set the error.
• After 500 ms, and if BUSY for the motor has not been canceled, set the error.
• After 500 ms and if BUSY is not canceled for the motor, set the error.
• After completing the motion, and if the upper point sensor (S46) does not turn on, set the error.
• After completing the motion, and when BUSY is not canceled for the motor, set the error.
• After 500 ms, and if BUSY is not canceled for the motor, set the error.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-17
• If BUSY isn’t canceled for the motor within 500 ms, an error results.
• If the outer specifiied point sensor (S49) remains on after 500 ms, an error results.
• If the inside specified point sensor (S50) remains on after 500 ms, an error results.
5) Waits 1000 ms
6) The motor is moved leftward at a constant speed for X pulses to enter the depth.
•If the origin sensor (S48) remains off after 500 ms, an error results.
•If BUSY isn’t canceled for the motor within 500 ms, an error results.
•If the outer specified point sensor (S49) remains of after 500 ms, it is set to off.
•If the inside specified point sensor (S50) remains off after 500 ms, it is set to off.
• If BUSY isn’t canceled for the motor within 500 ms, an error results.
2) The probe dispense at low speed for X pulses at the upper reset position
•BUSY for the motor turns off for 5 seconds (timeout).
•If the upper point sensor (S53) is not on after 500 an action, error results.
•If the lower point sensor (S54) is on after 500 an action, error results.
• If BUSY for the motor is not canceled within 500 ms, an error results.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-18
• If BUSY isn’t canceled for the motor within 500 ms, an error results.
• If BUSY isn’t canceled for the motor within 500 ms, an error results.
2) The mix rod is moved upward to reset for X pulses at reset upper point.
• If BUSY isn’t canceled for the motor within 13 seconds, an error results.
• If BUSY isn’t canceled for the motor within 500 ms, an error results.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-19
2) The wash nozzle is moved upward to reset for X pulses at upper reset point.
• The motor 3 seconds to switch from BUSY to OFF (timeout time).
• If the upper point sensor (S61) is off after the action has finished, an error results.
• If BUSY isn’t canceled for the motor after action has finished, an error results.
3) The wash nozzle is moved upward at a constant speed for X pulses for entering depth.
• If the upper point sensor (S61) remains of after 500 ms, an error results.
• If BUSY isn’t canceled within 500 ms, an error results.
19. ISE
1) If an ISE connection exists, the mechanism reset command is issued.
• If mechanism reset isn’t completed within 25 seconds, an error results.
• If the status is abnormal after the action has ended, an error results.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-20
n Initialization action
• ISE individual ON process
This mode is used if the connection to the ANL has been disconnected. Periodically wash and watch for execution of ANL
power. If the ISE is connected to the ANL temporarily, connection processing is performed. The processing flow is shown
below.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-21
After connection processing ends normally, the mode is shifted to the READY mode.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-22
n Power on sequence
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-23
(Caution !)
On the ISE, mechanical actions are performed until they are finished even if the connection (between the ISE and ANL) is
disconnected in mid-process.
If during this time the ANL receives a ISE instruction from the DPR, and ISE communication restarts, an error occurs
when executing reset operations while mechanical operations are executing, the connection can’t be established. In order to
prevent this, if the ISE is (temporarily) disconnected from the ANL, the connection processing isn't performed (until all
actions have ended and the condition is READY) even if it is a tentative connection.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-24
(Caution !)
If the ANL is aware of the ISE communication error without being aware of the ISE POWER signal, the problem below
occurs. Therefore it's control of the ISE mode is necessary while watching for ISE power signal and communication errors.
If ISE power is switched off immediately after ISE transmits the message to the ANL, the communication task of the ISE
may operate before ENQ retry time-out of the ANL. If the task is performed, the ISE waits for the communication check
message because the ISE wants to perform a series of POWER ON sequences, but the ANL determines that ISE
communication continues normally.
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 9 Initialize
9-25
1. Action instruction
Each type of prime action
Calibration
Each type of cleaning
Each type of washing action Shifted to each action process
Reset
Sample aspiration
Sample mix rotation
Stop Nothing is done (response only).
ES09_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-1
Abstract
n Explanation of actions
• Action type
• Manual actions are instructions given at any time by the user. Diagnostic and maintenance actions are excluded.
The types of manual actions are given below.
2) STAT check
ID for RB, control (QC), and calibrator (ACAL) set on the STAT table are automatically checked. This is required
if the user sets auto/forced ACAL or QC.
7) STAT rotation
The STAT table is rotated to the specified angle. This is required in cup setting.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-2
n Reagent check
1. Type
Reagent check action include the following types.
1) Reset
2) Remainder check
3) R. ID check
4) Remainder + R. ID check
2. Reset
1) Reagent line initialization is performed.
2) The check action for the reagent surface is performed.
3. Others
1) If the exitation of the reagent fridge is off or the position of the fridge is unstable, initialization of fridge
rotation is performed.
2) The position designation (in the action instruction) determines the action.
2. If R. ID check is performed
Stop position: The stop position of R. ID
Action contents: R. ID check is performed.
3) The stop position required in item 2), the action (according to the stop position number) is done in turn from
the position whose number is smaller than the current stop position number of fridge.
1. Remainder check
2. R. ID check
3. Remainder + R. ID check
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-3
1. Remainder
If the action is normal, the remainder is determined by (remainder step number up/down action + descent
step numbers after liquid level detection).
If the action is abnormal and the bottle is absent, "0" is displayed.
2. R. ID
If the action is normal, read data is checked for reagent ID. If it is okay, read data is specified as R. ID.
If the action is abnormal and if a read error and check error have also occurred, this is specified as "????"
(18 characters).
If the bottle is absent, "0" is displayed.
1. Remainder check
Obtain the remainder at the stop position.
2. R. ID check
R. ID of designation position for obtained stop position is transmitted.
3. Remainder + R. ID check
Data is accumulated from procedures 1) and 2).
The position data (where data is prepared) is transmitted.
7) The number of the stop position is greater than <(Possible reagent numbers to set) + 1> and measurement of
the remainder at the pre-dilution position is performed.
Data processing is in accordance with 4 and 5.
n Explanations of actions
Refer to the time chart on the next page.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-4
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-5
< Explanations of actions > 9) Degassed supply pump (P11) is turned off.
10) Inner wash back valve (V9) is turned off.
1) DI water pump (P1) and degasser (P3) are turned on. 11) If the upper point sensor (S46) is turned on, the R
2) If the position check sensor (S44) is turned on, R rotation rotation motor (M14) is moved left from the pre-dilution
motor (M14) is moved to the right from the wash position to position to the wash position.
pre-dilution position. If the upper position is turned off, an error results.
If the reagent check sensor (S44) is turned off, an 12) If the position check sensor (S44) is turned on, the R
error results. up/down motor (M15) is lowered to the wash position.
3) R liq level detection is made effective. If the position check sensor is turned off, an error
4) If the position check sensor (S44) is turned on and the results.
upper point sensor (S46) is on, R up/down motor (M15) is 13) SR outer wash valve (V10) is turned on.
lowered to the pre-dilution position. 14) R inner wash valve (V6) is turned on.
5) R liquid level detection is made ineffective. 15) R up/down motor (M15) is raised 2 mm.
6) R up/down motor (M15) is raised 5 mm. 16) R up/down motor (M15) is raised to the wash position.
7) R up/down motor (M15) is raised to the upper position at 17) R inner wash valve (V6) is turned off.
the pre-dilution position. 18) Degassed supply pump (P11) and SR outer wash valve
8) Inner wash back valve (V9) is turned on. (V10) are turned off.
19) DI water pump (P1) and degasser (P3) are turned off.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-6
< Explanations of actions > If the upper point sensor is turned off, an error
results.
1. If conditions 1 and 2 are satisfied, the fridge rotation If the inside or outside fixed point sensor is turned
motor (M6) is moved left/right to the designated stop off, an error results.
position. If the BUSY signal is not turned off within 2.5
seconds, an error results.
1) Upper point sensor (S46) of R up/down motor (M15) is If the inside or outside fixed point sensor is turned
turned on. off, corresponded to the stop position after the action is end,
2) Inside fixed point sensor (S50) is turned on if M6 is in an error results.
the even position, and the outside fixed point sensor (S49) is 2. If the bottle presence sensor (S52) is turned on, R. ID is
turned on if M6 is in the odd position. issued.
If the bottle presence sensor (S52) is turned off, the
action is ended.
3) Fridge rotation motor (M16) is moved left at low speed.
4) Fridge rotation motor (M16) is moved left to a fixed
point.
(If it ends normally, the position information is updated.)
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action Reagent check
10-7
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-8
STAT check
n STAT check
•1 STAT check action
1. If the exitation of the STAT rotation motor (M9) is off and the position of motor is unstable, initialize the STAT
rotation table.
If an error has occurred, an alarm is output and the actions below is not performed.
2. Initialization of the barcode reader at the STAT ID position is performed using the S. ID conditions.
3. The information for the stop position and the action contents are generated from the designated position.
Stop position: Stop position of STAT. ID + Stop position for cup detection
Stop position of STAT. ID : Designated position - 4 (STAT.ID position offset)
Stop position for cup detection : Designated position - 5 (Cup detection position offset)
However if the stop position exceeds 22 (possible number to set), the position is determined by (stop
position number + possible number to set).
Contents of action : Combination of cup detection and STAT.ID check
The stop position is calculated from the STAT number of the access position of the sample probe.
4. At the stop position required in item 3, a STAT check is performed in turn from the position whose number is
larger than current stat stop position.
5. The data obtained from the action is processed according to the procedure below.
Cup presence : Whether the cup exists or nothing is in the designated position is determined by the
sensor.
STAT.ID : If the action is normal, and if a read data check performed by the sample ID is okay,
read data is determined as STAT. ID.
If an abnormal action, read error, or check error has occurred, it becomes ??? (22characters).
6. The data of the position (that is prepared cup presence data and STAT. ID data both) is output. If there is no
cup, "0" is displayed for STAT. ID
2. If the upper point sensor (S36) of the S probe up/down is on, the STAT rotation motor (M9) is rotated leftward.
If the BUSY signal isn't turned off within 2 seconds, an error results.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-9
STAT check
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-10
W1 action
n W1 action
1. Action for start check
This process is performed as follows. If an abnormal end and an action stop has occurred, jump to step 4.
In each item action stop is possible by outer instruction.
1) The action for cuvette preparation is performed in 4.5 second-cycles while moving the cuvette rotation motor
(M18) rightward by one pitch.
<Cuvette>
In one pitch action the action stop is possible by outer instruction.
1) The cuvette wash action is performed in a 4.5 second-cycle while moving the cuvette rotation motor (M18) one
pitch rightward.
In the beginning, the contents of all cuvettes, except the cuvette in position is 80 has not been determined.
<ISE>
If there is a stop instruction, the ISE is stopped.
1) All prime of ISE is performed.
4. End action
1) Next pump is turned off.
DI water pump (P1), Degasser pump (P3), Detergent water pump (P4), Detergent water pump (P5)
2) Negative pressure cancel action is performed.
3) If the ISE mode is "BUSY", "READY" is set.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-11
W2 action
n W2 action
1. Action for start check
The process is performed as follows. If an abnormal end and action stop has occurred, jump to step 4.
In each item action stop is possible by outer instruction.
3) Initialization is performed.
The process is performed as follows. If an abnormal end and action stop has occurred, jump to step 4.
In one pitch actions the action stop is possible by outer instruction.
The following action is performed simultaneously.
(If an error has occurred in the action, except in steps 7 to 11, the action is stopped at the end of unit action
(item)).
<Cuvette>
1. Cuvette preparation action is performed in 4.5 second-cycles while moving the cuvette rotation motor (M18)
one pitch rightward.
2. The cuvette rotation motor (M18) is rotated and moves 1 pitch to the right in 4.5 second-cycles while washing
the cuvette until the head cuvette (in the 84th position after the preparation action) is in next position (in the 79th
position) to dry.
<Mix rod>
1. Action for preparation of mix rod is performed.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-12
W2 action
n W2 action
3. Action for wash
The process is performed as follows. If an abnormal end and action stop has occurred, jump to step 4.
Action stop is possible by outer instruction at determined position in each action.
4. End action
1) Following pumps are turned off.
DI water pump (P1), Degasser pump (P3), Detergent water pump (P4), Detergent water pump (P5)
2) Negative pressure cancel action is performed.
3) If the ISE mode is "BUSY", "READY" is set.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-13
PHOTOCAL
1) The action for cuvette preparation is performed in 4.5 second-cycles while moving the cuvette rotation motor
(M18) rightward by one pitch.
2) Photometry check action is performed.
3) The cuvette rotation motor (M18) is rotated and moves 1pitch to the right in 4.5 second-cycles while washing
the cuvette until the head cuvette (in the 84th position after the preparation action) is in the R1 dispense position
(in the 51st position).
1. The following action is performed in 6.8 second-cycles while moving the cuvette rotation motor (M18)
rightward by one pitch. (with Photocal measurement)
1) Degassed water dispensation is performed for all cuvettes (88 pcs.) from the head cuvette.
2) Wash action (detergent A, detergent B, water) for the cuvettes is performed in the action for cuvette
rotation motor (M18).
3) The photocal photometry command is output until all the data has been collected (20 times) after the
degassed water dispense action has ended.
4) After item 3 has ended, data processing for the photometry command is performed.
2. The cuvette wash action is performed in 4.5 second-cycles while moving the cuvette rotation motor (M18)
rightward by one pitch.
4. End action
Following pumps are turned off.
DI water pump (P1), Degasser pump (P3), Detergent water pump (P4), Detergent water pump (P5)
Negative pressure cancel action is performed.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-14
ISE action
n ISE action
1. Type
The manual action of the ISE has 2 starting methods.
1) Individual action : It is selected and started from the DPR menu.
2) Prime action : It is selected from the DPR menu and started with the switch on the ANL side.
The action is described below.
2. Individual action
If an abnormal end or action stop has occurred, the procedure is stopped at the end of each unit action.
1) Calibration
The time-out for each action is set, and the end information or time-out results in an error.(Action has
ended.)
There are three types of calibration.
Serum only : 1,7,8,9,10,11,12 are executed.
Urine only: 1,2,3,4,5,6,12 are executed.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 10 Explanation for action
10-15
ISE action
n ISE action
2) Selectivity check
The time-out of each action is set and the end information or time-out results in an error.
3) Cleaning
The cleaning of ISE washing is performed.
4) Reset
The mechanism line initialization action of ISE is performed.
3. Prime action
The selected action is performed by pressing DIAG SW until the action end instruction is performed.
Each action is stopped when abnormal end or action stop instruction is received.
The timeout is set in each action. It is ended by completion information or time out time.
1) Electrode replacement
First time of action selection is executed No.1 and from 2nd time, executed No.2.
1. PRIME_3 (waste the water via flow cell) is executed.
2. PRIME_1 (MID/REF prime) is executed.
2) General prime
ISE wash of all prime is performed.
3) Buffer prime
1. PRIME_4 (waste the water via bypass) is executed.
2. PRIME_6 (buffer liquid individual prime) is executed 9 times.
6) P tube replacing
1. PRIME_4 (waste the tube via bypass) is executed.
2. PRIME_7 (MID liquid individual prime) is executed 2 times.
3. PRIME_1 (MID/ REF prime) is executed 2 times.
ES10_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-1
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-2
2. Processing flow
Processing outline
Photometry data processing
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-3
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-4
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-5
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-6
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-7
( When n = 10 )
Slope Judgment formula
+ (P10 - P11) / 3 > d
- (P11 - P10) / 3 > d
“ d ” is the capable reverse OD value from the 9 second-unit.
“ d ” is included in the maker fixed parameter as an inside value.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-8
*2 Lineality check
Available points are divided in the front half and rear half and ∆E is calculated (for each).
If the following judgment formula isn’t satisfied, the error mark “ * ”is appended as linearity check error.
Judgment formula
∆E1- ∆E2
∆E1 + ∆E2 x 100 < linearity limitation
2 (Set by parameter menu)
or
∆E1 - ∆E2 < C
C: Set by parameter menu
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-9
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-10
Others
Process the value of the reaction OD as follows.
< 1. If either the value of the reaction OD for the emit item or the blank item cannot be calculated. >
The value of the reaction OD as the sample blank item is handled as if there is no data, and the abnormal value
mark is attached as OR type for the emit item and blank item.
< 2. If either the value of the reaction OD for the emit item or blank item can't be analyzed by reagent error or
sample error. >
The value of the reaction OD for the sample blank is calculated for a data item that exists as a sample blank
item, and the abnormal value mark is the same as the above.
< 3. If either the value of the reaction OD for the emit item or blank item has an abnormal value mark. >
The abnormal value mark is attached to the value of the reaction OD for the sample blank.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-11
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-12
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-13
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-14
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-15
5) Cross-check items
(1) Check timing
The operated items designated in the check condition between items are all entered after final concentration
calculation, or after the calculation of one piece of sample data has been performed.
(2) Check contents
The check is performed for the setting range, operation formula, and item combination set by the check
condition between items for concentration value.
If the operation results are outside the setting range, it’s attached the mark on all operated items.
(3) Data mark
If the range is outside the lower limit to upper limit range. [T]
6) Calculated Test
(1) Check timing
After performing the necessary tests are output in CONC, or after calculating the results of one sample.
(2) Check contents
Checking will be performed against the CONC value along with the conditions that are designated by
CALCULATED TEST CONDITION, such as Test Matching, Formula and Allowed Range, then if the results
are out of the designation, appropriate marks will be attached to the results.
The same process will be performed as the Normal Range and POS/NEG.
(3) Marks
If ‘Level Check = NO’
Exceeds the Upper Limit: [H]
Below the Lower Limit: [L]
If ‘Level Check = YES’
Exceeds the Upper Limit: [P]
Below the Lower Limit: [N]
7) QC Check
< Single Check (per One Test) >
Tests to be checked
On the QC condition, per one test, six types of Control Limits can be set in maximum. Each Control Limit
Value will be used for checking results that are designated as SINGLE in the parameter. There are two types of
Control Limit Values which can be designated by the QC condition.
a) Pre-set Mode
b) Accumulated Value
c) OFF mode (Checking will not be done.)
(1) Check timing
On the QC samples, after checking the Correlation Correction, checking will be performed for one test per
type.
(2) Check contents
Checking will be performed based on the Control Limit Value Range designated by the QC condition, then if
the result is out of range, a mark will be attached to the result.
(3) Flags
Out of Control Limit Range: [1]
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-16
Explanation of QC logic
QC ABNORMAL
Indicates that the result was out of the control limit.
NO QC ABNORMAL
Indicates that the result was within the control limit.
12s
Indicates that the QC results exceeded the control limit of ‘MEAN ± 2SD’.
13s
Indicates that the QC results exceeded the control limit of ‘MEAN ± 3SD’.
22s
Indicates that two consecutive QC results exceeded the control limit of ‘MEAN ± 2SD’ in the same
direction.
R4s
Either of two QC results, are consecutive in H/L, exceeded ‘MEAN + 2SD’, and the other one exceeded
‘MEAN -2SD’, or exceeded the range of ‘4SD’.
41s
Checking if the four consecutive results exceeded the control limit of ‘MEAN + 1SD’, or ‘MEAN -1SD’.
10x
Seven to ten consecutive QC results exceeded the ‘+’ side of ‘-’ side of control limit.
Others
For the mark, ‘X’, when QC analysis (with a twin of H/L) was performed, if one was marked (including QC
marks) and another one was not marked, ‘X’ will be marked for the QC results of one to remove the QC
result from QC control.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-17
NOTE: With the one wave method, the above process is performed only for one wave length.
NOTE: With the one wave method, the above process is performed only for one wave length.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-18
Data correction
(Advance dilution correction, Pre-dilution correction, dispensation
correction)
Correlation correction
Correlation correction
correction formula C = A * Cd + B
C: corrected concentration
A: correlation correction coefficient "A"
B: correlation correction coefficient "B"
Cd: concentration before correction
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-19
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-20
Judgment flow
Refer to the next page
5) Level judgment and type for indication and print
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-21
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-22
Output data
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-23
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-24
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-25
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-26
WB correction
by cuvette data of RB sample
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-27
1. Processing flow
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-28
Others
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-29
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-30
The calibration curve is generated using the smallest squares method and using each standard and
measured OD value, OD value = 0, and Concentration value = 0.
• If the value of concentration “ 0 ” exists in the standard.
The calibration curve is generated using the smallest squares method and using each standard and
measured OD value.
(7) Spline
If the concentration value "0" is not in the standard :
The calibration curve is generated using the smallest squares method and using each standard and the value of
the measured OD, OD value = 0, and concentration value = 0.
If the concentration value "0" is in the standard :
The calibration curve is generated using the smallest squares method and each standard and the measured OD
value.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-31
If the concentration “ 0 ” as a standard exists, and the data can be corrected using two points (concentration “ 0 ”
and other point ).
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-32
3) Other processing
• Processing if the index is changed
If the index is changed and if the QC sample data exists in the start condition menu and it's also for the calculation
for the value of daily statistics, the middle value of statistics calculation in each sample type, item and level.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-33
•LIH
Analysis process and output of analysis data isn't performed.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 20 Photometry data processing
20-34
2 Contents of processing
1) Photometry point data (from ANL after WB correction) adds the value of OD and the value of concentration for
the cuvette, and records it on the hard disk. The recording unit is the cuvette. And the data about ISE item and
LIH.
ES20_01.DOC
M514 EL SOFTWARE MANUAL
22-1
Chapter 22 Abnormal data mark 0300 0200 0100
n Data mark
1. Abstract
1) Abnormal mark (concerning hardware) attached to ANL.
2) Abnormal mark is attached to OD value level
3) Abnormal mark is attached to concentration value level
4) The mark indicated the data condition.
ES22_01.DOC
M514 EL SOFTWARE MANUAL
22-2
Chapter 22 Abnormal data mark 0300
ES22_01.DOC
M514 EL SOFTWARE MANUAL
22-3
Chapter 22 Abnormal data mark 0400
ES22_01.DOC
M514 EL SOFTWARE MANUAL
23-1
Chapter 23 Alarm list
n Outline of alarms
The alarm messages of the AU400 are shown in Chapter 24.
If an alarm is occurs, see “alarm HELP” which explains countermeasures.
Because of a shortage of information, system errors caused by the BIOS built into the OS of the personal
computer, cannot be explained. If a system error isoccurs, consult the OS maker or the manufacturer of
the PC.
• Outline of errors
1) Timing : ANL alarms issued every 4.5 seconds for a maximum of 5 points.
2) Output contents :
Error indication
Alarm No. Message Time Cause Action table Action phase
A. Alarm occurrence time : If the mechanism action error occurs, the alarm occurrence time (100 ms count
value) is set.
Caution ) If the time indication is not required, it’s set to “ FFFFH ”.
B. Alarm details : Alarm details (Max. 20 bytes) are set from the beginning, and “ 00H ” is set in the
empty space.
1. Error cause : 1 to 255, or bit information (8 bit)
2. Action table : Mechanism action table number (1 to 65535) in mechanism action
3. Action phase : Mechanism action relative time (1 to 65535) in mechanism action
es23_01.doc
M514 EL SOFTWARE MANUAL
23-2
Chapter 23 Alarm list 0600 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-3
Chapter 23 Alarm list 0500 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-4
Chapter 23 Alarm list 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-5
Chapter 23 Alarm list 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-6
Chapter 23 Alarm list 0600 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-7
Chapter 23 Alarm list 0600 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-8
Chapter 23 Alarm list 0600 0500 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-9
Chapter 23 Alarm list 0400
es23_01.doc
M514 EL SOFTWARE MANUAL
23-10
Chapter 23 Alarm list 0400
9462 The disk media is not recognized. It may not be formatted 24-129
9463 The media is write protected 24-129
9471 File genric Error [%s] 24-129
9472 File not found [%s] 24-130
9473 Bad pathname [%s] 24-130
9474 Too many open files [%s] 24-130
9475 File access denied [%s] 24-130
9476 Invalid File handle [%s] 24-131
9477 Can’t remove current directory [%s] 24-131
9478 Full directory [%s] 24-131
9479 Bad seek pointer [%s] 24-131
9480 File Hardware I/O Error [%s] 24-132
9481 File sharing violation [%s] 24-132
9482 File lock violation [%s] 24-132
9483 File disk full [%s] 24-132
9484 File end of file [%s] 24-133
es23_01.doc
M514 EL SOFTWARE MANUAL
24-1
Chapter 24 Alarm list 0400
1001 INITIALIZE
System Status
(1) The system has started initialization of it's mechanical components. This is an advisory alarm.
1002 WARMUP
System Status
(1) The temperature and optical systems are stabilizing. This is an advisory alarm.
User Action
(1) None. The mode automatically changes to WARM-UP.
1003 STANDBY
System Status
(1) The system has switched to the Stand-By mode. This is an advisory alarm.
User Action
(1) None. The mode automatically changes to STAND-BY.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-2
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-3
Chapter 24 Alarm list 0400
1012 PAUSE
System Status
(1) The mode changes to PAUSE. An advisory alarm is given.
User Action
(1) Pressing the Start key can continue system operation.
1013 STOP
System Status
(1) The mode changes to STOP. An advisory alarm is given.
User Action
(1) This alarm is usually triggered by a mechanical or computer error.
The cause must be determined and corrected before restarting.
1015 W1 START
System Status
(1) W1 operation starts. An advisory alarm is given.
1016 W1 END
System Status
(1) W1 operation ends. An advisory alarm is given.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-4
Chapter 24 Alarm list 0400
1017 W2 START
System Status
(1) An advisory alarm is given. W2 operation starts.
1018 W2 END
System Status
(1) W2 operation ends. An advisory alarm is given.
1051 POWER ON
es24_01.doc
M514 EL SOFTWARE MANUAL
24-5
Chapter 24 Alarm list 0400
aaaa : Sample ID
aaaaa : Sample ID
es24_01.doc
M514 EL SOFTWARE MANUAL
24-6
Chapter 24 Alarm list 0400
2008 Ordering
System Status
(1) System does not change modes.
User Action
(1) Restart after analysis has been completed.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-6-1
Chapter 24 Alarm list 0600 0400
System Status
User Action
(1). Analyze the reagent blank of the item. After confirmation of the result, perform the
analysis of the item.
Cause
Error detail.
aaaaa : Item name
b : Reagent sequence No.(Displayed in case of the item set to Advanced
Calibration.)
c : Sample type(SPACE:Serum U:Urine X:Others)
System Status
User Action
(1). Analyze the calibration of the item. After confirmation of the result, perform the
analysis of the item.
Cause
Error detail.
aaaaa : Item name
b : Reagent sequence No.(Displayed in case of the item set to Advanced
Calibration.)
c : Sample type(SPACE:Serum U:Urine X:Others)
es24-6-1.doc
M514 EL SOFTWARE MANUAL
24-6-2
Chapter 24 Alarm list 0600 0400
es24-6-1.doc
M514 EL SOFTWARE MANUAL
24-7
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-8
Chapter 24 Alarm list 0400
aa : Reagent position : 1 to 76
es24_01.doc
M514 EL SOFTWARE MANUAL
24-8-1
Chapter 24 Alarm list 0400
System Status
User Action
(1). Open the edit menu in Reagent status and set the cleaner.
Cause
(1). Though the cleaner is set in Contamination parameters, it is not set in Reagent status.
aa : Reagent position : 1 to 76
es24-8-1.doc
M514 EL SOFTWARE MANUAL
24-9
Chapter 24 Alarm list 0400
aa : Reagent position : 1 to 76
aa : Reagent Position : 1 to 76
es24_01.doc
M514 EL SOFTWARE MANUAL
24-10
Chapter 24 Alarm list 0400
aa : Message No.
bb : First byte of the message
es24_01.doc
M514 EL SOFTWARE MANUAL
24-11
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-12
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-13
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-14
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-15
Chapter 24 Alarm list 0600 0400
es24-15.doc
M514 EL SOFTWARE MANUAL
24-16
Chapter 24 Alarm list 0600 0400
Cause
(1) In the Online Parameter menu, submenu SetUp, the T.R.I. receive is not designated as "YES" for the
necessary parameter.
(2) The display in the [ ] and its meaning are as follows.
es24-16.doc
M514 EL SOFTWARE MANUAL
24-17
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for samples following the alarm.
Cause
(1). When receiving the test requisition information text using online, the text classification
code is out of specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dd : Received Text Code
eeeeeeeee : Transferred sample ID
es24-17.doc
M514 EL SOFTWARE MANUAL
24-18
Chapter 24 Alarm list 0600 0400
System Status
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "STOP", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). System will not receive text for samples following the sample generting the alarm.
3). In a case when the test requisition is received in real-time, the "STOP" is ignored
and the test requisition text will be received with real-time again during the next
analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "Continue", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). The test requisition information text for the next sample will be received.
Additional Information
(1). When receiving the requistion information text using online, the block identification No.
is out of specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dd : Previously received block identification No.
ee : Last received block identification No.
fffffffff : Transferred sample ID
es24-18.doc
M514 EL SOFTWARE MANUAL
24-19
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for other samples within the run.
Cause
(1). When receiving the test requisition information text using online, sample No. of the
sample generating the alarm does not meet the specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
es24-19.doc
M514 EL SOFTWARE MANUAL
24-20
Chapter 24 Alarm list 0600 0400
2138 ONLINE ILLEGAL RACK No. (aaaaaa - bbbbb - cc : ddddd - ee) fffffffff
System Status
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "STOP", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). System will not receive text for samples following the sample generating the alarm.
3). In a case when the test requisition is received in real-time, the "STOP" is ignored
and the test requisition text will be received with real-time again during the next
analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "Continue", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). The test requisition information text for the next sample will be received.
Additional Information
(1). When receiving the requistion information text using online, the rack No. is out of
specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
es24-20.doc
M514 EL SOFTWARE MANUAL
24-21
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text using online, sample kind No. of
the sample generating the alarm does not meet the specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dd : Received Sample type
-------------------------------------------------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
es24-20.doc
M514 EL SOFTWARE MANUAL
24-22
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for other samples within the run.
Cause
(1). When receiving the requistion information text using online, the sex of this text is not
within the specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
d : Sex Gender Received
eeeeeeeee : Transferred sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL
24-23
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for other samples within the run.
Cause
(1). When receiving the requistion information text using online, age or month age (age of
baby before one year) of this text is out of the specification.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
ddd : Received age
ee : Received month age
fffffffff : Transferred sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL
24-24
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for other samples within the run.
Cause
(1). When receiving the test requisition information text using online, sample type of this
text does not match the required sample type.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dd : Transferred Sample type
-------------------------------------------------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other stat sample
-------------------------------------------------------------------------------------------------------
ee : Received sample type
-------------------------------------------------------------------------------------------------------
__ : Normal sample
es24-20.doc
M514 EL SOFTWARE MANUAL
24-25
Chapter 24 Alarm list 0600 0400
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other stat sample
-------------------------------------------------------------------------------------------------------
fffffffff : Transferred sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL
24-26
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text using online, sample type of this
does not match the required sample.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dddddd : Received sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
es24-20.doc
M514 EL SOFTWARE MANUAL
24-27
Chapter 24 Alarm list 0600 0400
2144 ONLINE RACK No. MISMATCH (aaaaaa : bbbbb - cc<> ddddd - ee)
fffffffff
System Status
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "STOP", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). System will not receive text for samples following the sample generating the alarm.
3). In a case when the test requisition is received in real-time, the "STOP" is ignored
and the test requisition text will be received with real-time again during the next
analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is
designated as "Continue", the following events occur :
1). Disregard the test requisition information text for the sample generating the alarm.
2). The test requisition information text for the next sample will be received.
Additional Information
(1). When receiving the test requisition information text using online, rack No. of this does
not match the required sample.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other that sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
ddddd : Received Rack No.
ee : Received Cup position (1 to 10)
fffffffff : Transferred sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL
24-28
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text uisng online, sample ID of this text
does not match the required ones.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
ddddddddd : Transferred sample ID
eeeeeeeee : Received sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL 24-28-1
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text using online, the test requisition
cannot be stored.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
_ _ nnnn : Normal sample
_E nnnn : Emergency sample
_P nnnn : Stat sample
U_ nnnn : Urine normal sample
UE nnnn : Urine emergency sample
UP nnnn : Urine stat sample
X_ nnnn : Other normal sample
XE nnnn : Other emergency sample
XP nnnn : Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
ddddddddd : Transferred sample ID
es24-20.doc
M514 EL SOFTWARE MANUAL 24-28-2
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the repeat run test requisition information for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text for repeat run using online, repeat
run test requisition cannot be stored with for the following reasons.
1). Original sample that is designated in the repeat run text is not stored in the normal
run/sample test requisition information.
2). Sample No. is not set to the original sample that is designated in the repeat run text.
3). Original sample that is designated in the repeat run text has already been stored as a
repeat run sample for another sample.
(2). The display of [ ] and its meaning are as follows.
aaaaaa : Transferred sample type and sample No.
-------------------------------------------------------------------------------------------------------
H_ _ nnnn : repeat run,Normal sample
H_E nnnn : repeat run,Emergency sample
H_P nnnn : repeat run,Stat sample
HU_ nnnn : repeat run,Urine normal sample
HUE nnnn : repeat run,Urine emergency sample
HUP nnnn : repeat run,Urine stat sample
HX_ nnnn : repeat run,Other normal sample
HXE nnnn : repeat run,Other emergency sample
HXP nnnn : repeat run,Other stat sample
-------------------------------------------------------------------------------------------------------
bbbbb : Transferred Rack No.
cc : Transferred Cup position (1 to 10)
dddddd : sample type and original sample No.
-------------------------------------------------------------------------------------------------------
_ _nnnn : Normal sample
_ _Ennn : Emergency sample
_ _Pnnn : Stat sample
_U_nnn : Urine normal sample
_UEnnn : Urine emergency sample
_UPnnn : Urine stat sample
_Xnnnn : Other normal sample
es24-20.doc
M514 EL SOFTWARE MANUAL 24-28-3
Chapter 24 Alarm list 0600 0400
System Status
(1). Disregard the test requisition information text for the sample generating the alarm.
(2). Continue to receive the test requisition information text for the other samples within the
run.
Cause
(1). When receiving the test requisition information text using online, length of this text is out of the specification.
es24-20.doc
M514 EL SOFTWARE MANUAL
24-29
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-30
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-31
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
2208 ANL Communication Not Receive STX First Time Error (aa,bb,c)
Cause
(1) DPR received except STX with the top of text when receiving from ANL. However, ANL was retried
and communication completed without trouble.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-32
Chapter 24 Alarm list 0400
2209 ANL Communication Receive Text length not equal First Time
Error (aa,bb,c)
Cause
(1) DPR received the illegal length of the text when receiving from ANL. However, ANL was retried and
communication completed without trouble.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-33
Chapter 24 Alarm list 0400
2213 ANL Communication Text Sending Char Receive First Time Error
(aa,bb,c)
Cause
(1) DPR received text under transmitting to ANL. However, ANL was retried and communication
completed without trouble.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-34
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-35
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-36
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-37
Chapter 24 Alarm list 0400
Error details
aa : Message No.
bb : Error No.
c : Retry No.
Error details
aa : Message No.
bb : Error No.
c : Retry No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-38
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-39
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-40
Chapter 24 Alarm list 0400
Error details
es24_01.doc
M514 EL SOFTWARE MANUAL
24-41
Chapter 24 Alarm list 0400
(2) Check connection of the optical communication connectors (U39 and U40) on the optical
communication master ROM (MV1030) and ANL control board (MV1019), then restart the system. If a
ROM error recurs, replace the MV1030.
(3) If an S-RAM or DP-RAM error occurs, replace the board MV1019.
If the line is disconnected or a communication error occurs, check connection of all connectors on the
secondary error station board and restart the system.
(4) If an error reoccurs, replace the secondary error station board.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-42
Chapter 24 Alarm list 0400
Additional Information
(1) Check connectors 370 and 379 on the ANL control board (MV1019) and all connectors on the
photometry boards (MV1005 and MV1006) then restart the system. If an error reoccurs, replace the
MV1019, MV1005 and / or MV1006 boards.
(2) Reset the system. If an error reoccurs in resetting, check the sub-CPU using ANL DIAG. If an error still
recurs, take the same measures as taken for an error in initialization.
(3) The ANL receives an invalid message code from the photometry sub-CPU.
Error details
aa : Received message code (Hex.)
bb : Data in the ARCK NET status register (Hex.)
cc : Data in the ARCK NET status configuration register (Hex.)
dd : Data in the ARCK NET DIAG register (Hex.)
Error details
aa : Message code to be received (Hex.)
bb : Data in the ARCK NET status register (Hex.)
cc : Data in the ARCK NET status configuration register (Hex.)
dd : Data in the ARCK NET DIAG register (Hex.)
es24_01.doc
M514 EL SOFTWARE MANUAL
24-43
Chapter 24 Alarm list 0400
Error details
es24_01.doc
M514 EL SOFTWARE MANUAL
24-44
Chapter 24 Alarm list 0400
Error details
aa : Send command (Hex.)
b : System status
1 : System busy
2 : Wheel rotation disable
3 : Parameter error
4 : Unknown command
5 : Double receive command
6 : Initial position error after INITIALIZE
7 : Stop position error after photometry
es24_01.doc
M514 EL SOFTWARE MANUAL
24-45
Chapter 24 Alarm list 0400
Additional Information
(1) In initialization
Check the optical communication connectors (U39 and U40) on the ANL control board (MV1019),
then restart the system.
If an error recurs, check connection of all connectors on the secondary error station board, then restart
the system.
If an error recurs, replace the secondary error station board.
(2) In Operation (2) or (3)
Reset the system. If an error recurs in resetting, check the sub-CPU using ANL DIAG. If an error still
recurs, take the same measures as taken for an error in initialization.
Error Details
aa : Error type (Hex.)
10 : DMA error when writing primary station command
11 : Cannot write command to primary station
20 : Semaphore error when writing parameter to secondary station t
40 : DMA / Transfer / receive error when writing command to non-selected secondary station
41 : Cannot write command when writing command to non-selected secondary station
Case 2 (aa=20)
01H : Reagent container
02H : R probe
03H : Sample probe
04H : Rack unloading
05H : Rack recovery
06H : Dispenser
07H : Mixer
es24_01.doc
M514 EL SOFTWARE MANUAL
24-46
Chapter 24 Alarm list 0400
Error Details
aa : Error Secondary station
bb : Time table No.
cc : Phase No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-47
Chapter 24 Alarm list 0400
Additional Information
(1) In Initialization
Check connectors (375 and 385) on the ANL control board (MV1019), ISE I / F ROM (MV2310), all
connectors on the ISE control board (MV2190) and ISE control ROM (MV2191), then restart the
system. If an error recurs, replace MV2310, MV1019, MV2191 and MV2190.
(2) In case of Operation (2) or (3)
Reset the system. If an error recurs in resetting, take the same measures as taken for an error in
initialization.
Error details
aa : (Hex.)
01H : Rack ID / S.ID
03H : STAT
04H : Reagent
bb : (Hex.)
bit 1 : No acknowledgment is received in time
bit 2 : NAK is received
bit 3 : One-character transmission is not completed in time
es24_01.doc
M514 EL SOFTWARE MANUAL
24-48
Chapter 24 Alarm list 0400
Error details
aa : (Hex.)
01H : Rack ID / S.ID
03H : STAT
04H : Reagent
bb : (Hex.)
bit 0 : Message length error
bit 1 : Failure to complete from STX to LF
bit 2 :
bit 3 :
bit 4 : Over-run error
bit 5 : Framing error
bit 6 :
bit 7 : Overwriting the message
es24_01.doc
M514 EL SOFTWARE MANUAL
24-49
Chapter 24 Alarm list 0400
Error details
aa : Message number (Hex.)
bb : (Hex.)
bit 0 : STX cannot be received within time
bit 1 : Received except STX
bit 2 : Failure to complete from STX to BCC
bit 3 : BCC error
bit 4 : No EXT
bit 5 : Framing error or over-run error
Error details
aa : Message No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-50
Chapter 24 Alarm list 0400
Error details
aa : Message No. (Hex.)
bb : (Hex.)
bit 0 : An acknowledge to ENQ is not received in time
bit 1 : ENQ is returned in response to ENQ
bit 2 : An acknowledge to transmission is not received in time
bit 3 : NAK is returned in response to transmission
bit 4 : Non ACK / NAK is returned in response to transmission
bit 5 : Text is received in transmission
bit 6 : Time-out is in transmission
Error details
aa : (Hex.)
bit 0 : Reading is not completed in time
bit 1 : Check character check error
bit 2 : Wrong number of digits
es24_01.doc
M514 EL SOFTWARE MANUAL
24-51
Chapter 24 Alarm list 0400
Error details
aa : Rack type (Hex.)
1 : General
2 : Emergency
3 : Repeat run
4 : ACAL
5 : QC
bb : (Hex.)
bit 0 : Reading is not completed in time
bit 1 : Check character check error
bit 2 : Wrong number of digits
es24_01.doc
M514 EL SOFTWARE MANUAL
24-52
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-53
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-54
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-55
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-56
Chapter 24 Alarm list 0400
aa : (Decimal)
1 : Lamp brightness
2 : 0% data error
3 : Gain error
4 : Offset error
es24_01.doc
M514 EL SOFTWARE MANUAL
24-57
Chapter 24 Alarm list 0400
aa : (Hex.)
bit 0 : Unloading backward motion prohibition sensor(S7) turns on
bit 1 : Unloading rack arrival sensor(S17) turns on
bit 2 : Cup detection sensor(S8) turns on
bit 4 : Error of Rack step at dispensing sample position (S16)
es24_01.doc
M514 EL SOFTWARE MANUAL
24-58
Chapter 24 Alarm list 0400
3150 NO RACK
System Status
(1) The mode is no change.
Cause
(1) No rack is found on the rack loading mechanism during MEASURE start.
User Action
(1) Set a rack on the rack feed caterpillar. If there is a rack, call Olympus Technical Support to check
operation of the sensor.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-59
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-60
Chapter 24 Alarm list 0400
Additional Information
(1) Received sample data is inconsistent with the S.No. inquiry from the ANL to the DPR.
Error details
aa (Hex) :
bit 0 : sequential No.
bit 1 : Kind of Rack
bit 2 : Rack No.
bit 3 : Cup position
cc : 1 : General
2 : Emergency
3 : Calibration
4 : QC control
5 : Reagent blank
7 : Repeat of standard
8 : Stat
3161 QC INCOMPLETE
System Status
(1) The system proceeds to the next QC sample.
Cause
(1) The last number of QC rack are not set or the rack feeder was stopped during QC.
User Action
(1) Check if the cups and racks are set according to the control position programming. If they are set
properly, call Olympus Technical Support to check for a faulty cup sensor.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-61
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-62
Chapter 24 Alarm list 0400
Error details
T : Time table No.
aa : (Decimal)
1 : Except Upper position where sample probe is Rack position.
2 : Up/Down motor is busy where sample probe is Rack position.
Error code
T : Time Sequence No.
aa : (Decimal)
1 : Motor busy
2 : The unloading-side disable motion sensor(S7) turns on when operation starts
3 : The rack pass sensor(s27) turns on when operation starts
4 : The backward prohibition sensors(S28) turns on when backward motion starts
5 : The unloading-side stop position sensor(S1) turns off when forward motion is completed
6 : The recovery-side stop position sensors(S2) turns off when backward motion is
completed
bb : Time Table No.
cc : Phase No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-63
Chapter 24 Alarm list 0400
Error codes
T : Time Sequence No.
aa : (Decimal)
1 : Motor busy
2 : The unloading disable sensor turn on when operation starts (S20)
3 : The recovery rack detection sensor(S18) turns on when motion to the recovery position
starts
4 : The unloading-side stop position sensor(S19) turns off when motion to the unloading
position is completed
5 : The recovery-side stop position sensor(S17) turns off when motion to the recovery
position is completed
bb : Time Table No.
cc : Phase No.
Error details
T : Time Sequence No.
aa : (Decimal)
1 : Motor busy
2 : The origin sensor(S11) turns off after the pawls move to the origin
3 : The origin sensor turns(S11) off after the pawls move to the stop or open position
bbbb : (Decimal)
Time table No.
cccc : (Decimal)
Phase No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-64
Chapter 24 Alarm list 0400
Error details
T : Time Sequence No.
aa : (Decimal)
1 : Motor busy
2 : The origin sensor(S14) turns off after the pawls move to the origin
3 : The origin sensor(S14) turns off after the pawls move to the stop or open position
bbbb : (Decimal)
Time table No.
cccc : (Decimal)
Phase No.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-65
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-66
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-67
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-68
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-69
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-70
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-71
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-72
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-73
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-74
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-75
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-76
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-77
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-78
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-79
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-80
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-81
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-82
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-83
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-84
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-85
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-86
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-86-1
Chapter 24 Alarm list 0600 0400
System Status
User Action
(1) If there are not another filled up reagent bottle, once the system returns to Stand-By,
replenish the reagent bottle.
Cause
Additional Information
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-87
Chapter 24 Alarm list 0400
3301 NO ISE
Cause
(1) There has been a complete loss of communications with the ISE.
User Action
(1) Contact Olympus Technical Support.
Additional Information
(1) The ISE power is turned off.
(2) Check the source voltage of the ISE and connector (106) between the ANL control board (MV1019)
and ISE control board (MV2190).
Error details
aa : bit 0 : The check sum of the IPL block of the ISE FROM is not correct
bit 1 : The check sum of the control block of the ISE FROM is not correct
bit 2 : Data cannot be read or written from / into the ISE RAM
bit 3 : An error occurs in ISE's pulse motor control 1
bit 4 : An error occurs in ISE's pulse motor control 2
bit 6 : The ISE does not get ready
bit 7 : The ISE cannot transmit or receive an ANL message properly in ISE communication
check
(It considers it is not A55A.)
es24_01.doc
M514 EL SOFTWARE MANUAL
24-88
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-89
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-90
Chapter 24 Alarm list 0400
aa : Message number
Error details
bb : bit 0 : Message length error
bit 1 : Communication is not completed within time from STX to BCC
bit 2 : BCC error
bit 3 : No ETX
bit 4 : Framing error
bit 5 : Over-run error
bit 6 : Undefined message
bit 7 : Message overwriting
es24_01.doc
M514 EL SOFTWARE MANUAL
24-91
Chapter 24 Alarm list 0400
aa : Message number
bb : Detail of operation command ("0" if non-received message is not an operation response.)
aa : Message number
Error details
bb : bit 0 : An ENQ response is not received in time
bit 2 : A transmission response is not received in time
bit 4 : One character is not transmitted in time
es24_01.doc
M514 EL SOFTWARE MANUAL
24-92
Chapter 24 Alarm list 0400
aa : Message number
Error details
bb : bit 0 : Message length error
bit 1 : Communication is not completed within time from STX to BCC
bit 2 : BCC error
bit 3 : No ETX
bit 4 : Framing error
bit 5 : Over-run error
bit 6 : Undefined message
bit 7 : Message overwriting
aa : Message number
bb : Details of operation command ("0" if non-received message is not an operation response.)
es24_01.doc
M514 EL SOFTWARE MANUAL
24-93
Chapter 24 Alarm list 0400
Error details
aa : ISE command 1
bb : ISE command 2
es24_01.doc
M514 EL SOFTWARE MANUAL
24-94
Chapter 24 Alarm list 0400
Error details
T : Time sequence No.
aa : 1 : OPMC busy
2 : Motor busy
3 : The upper position sensor turns on after discharging
4 : The lower position sensor turns off after aspiration
5 : The upper position sensor turns off after moving to the upper position
6 : The lower position sensor turns on after moving to the upper position
7 : The upper position sensor turns on after getting out of the upper position
8 : The lower position sensor position turns off after moving to the lower position
es24_01.doc
M514 EL SOFTWARE MANUAL
24-95
Chapter 24 Alarm list 0400
Error details
T : Time sequence No.
aa : 1 : OPMC busy
2 : Motor busy
es24_01.doc
M514 EL SOFTWARE MANUAL
24-96
Chapter 24 Alarm list 0600 0400
Error details
T : Time sequence No.
aa : 1 : Overflow
2 : The overflow sensor turns off on checking
aa : Message number
Error details
bb : bit 2 : A communication response was not received in time
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-96-1
Chapter 24 Alarm list 0600 0400
System Status
(1). No sample cup has been detected for the sample ordered.
User Action
(1). Verify the test assays selected on the test requisition menu.
Cause
(1). No sample cup is detected though there is a test requsition ordered for the sample
System Status
(1). Default panel is requested for the sample generating the alarm.
User Action
(1). Verify that test items have been selected for the sample.
Cause
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-97
Chapter 24 Alarm list 0600 0400
System Status
(1). Preset default panel of tests is requested for the sample generating the alarm.
Cause
Error detail
aaaaa : Sample No.
bbbbb : Rack No. or STAT table No.
cc : Cup position or STAT position
ddddddddd: Sample ID
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-98
Chapter 24 Alarm list 0400
Error detail
aaaaa : Requested sample No. from ANL
bbbbb : Responded sample No. from DPR
Error detail
aaaaa : Sample No.
bbbbb : Rack No. or STAT table No.
cc : Cup position or STAT position
Error detail
aaaaa : Sample No.
bbbbb : Rack No. or STAT table No.
cc : Cup position or STAT position
es24_01.doc
M514 EL SOFTWARE MANUAL
24-99
Chapter 24 Alarm list 0600 0400
Error detail
aaaaa : space
bbbbb : Rack No. or STAT table No.
cc : Cup position (1 to 10) or STAT position
aaaaa : Space
bbbbb : Rack No. or Stat table No.
cc : Cup position or stat position
System Status
Cause
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-100
Chapter 24 Alarm list 0400
Error detail
aaaaa : Item name
c : Sample type (serum: space urine:U other:X)
Error detail
aaaaa : Item name
b : Sample type (serum: space urine:U other:X)
Error detail
aaaaa : Item name
b : calibration point (1 to 7)
c : Sample type (serum: space urine:U other:X)
es24_01.doc
M514 EL SOFTWARE MANUAL
24-101
Chapter 24 Alarm list 0600 0400
Error detail
aaaaa : Item name
b : calibration point (1 to 7)
c : Sample type (serum: space urine:U other:X)
Error detail
aaaaa : Item name
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-101-1
Chapter 24 Alarm list 0600 0400
System Status
User Action
Cause
Error detail
System Status
User Action
Cause
System Status
User Action
Cause
Error detail
aaaaa : Sample No.
bb : QC control No. (1 to 60)
ccccc : Item name
d : Sample type (serum: space urine:U other:X)
eeeeeeeee : Control ID
System Status
Description
Error detail
aaaaa : Sample No.
bbbbb : Rack ID or Stat No.
cc : Cup position (1 to 10 ) 1 to 22 for stat sample
dd : Item name
eeeeeeeee : Sample ID
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-103
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-104
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-105
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-106
Chapter 24 Alarm list 0600 0400
System Status
Cause
Error detail
aaaaa : Sample No.
bbbbb : Rack No.
cc : Cup position
ddddd : Color Test Name
eeeee : Blank Test Name
fffffffff : Sample ID
Error detail
aaaaa : Sample No.
bbbbb : Rack No.
cc : Cup position
ddddd : Color Test Name
eeeee : Blank Test Name
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-107
Chapter 24 Alarm list 0600 0400
Error detail
aaaaa : Original Sample No.
bbbbb : Rack No.
cc : Cup position
Error detail
aaaaa : Repeat run sample ID
System Status
User Action
Cause
(1). The number of samples reaches the maximum in the M1 mode, or the sample number
reaches the upper limit.
Error detail
aaaaa : S.No. having reached the upper limit
bbbbbbbbb: Sample ID
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-107-1
Chapter 24 Alarm list 0600 0400
User Action
(1). Confirm the rack ID and Rack No. Limit in [Parameter][System][ System] menu. If you
find that sample No. can not calculate because of wrong rack ID or wrong Rack No.
Limit, correct them.
Cause
(1). In the case of Rack ID analysis, sample No. can not be calculated based on Rack No.
Error detail
aaaaa : Rack No.
bb : Cup position (1 to 10)
System Status
(1). System moves to the next sample without analyzing the sample.
User Action
(1). After requesting tests or confirm tests being requisitioned in HOST, re-run the sample
according to the necessity.
(2). In the case of Sequential Requisition, set the start S.No. in the Start Condition menu for
the system to process the sample as the correct S.No.
Cause
(1). There are no tests requisitioned for the sample generating the alarm even after receiving
the test requisition information text using online or being requested according to default
panel.
Error detail
aaaaa : Sample No.
bbbbb : Rack No.
cc : Cup position
ddddddddd: Sample ID
es24-86-1.doc
M514 EL SOFTWARE MANUAL
24-108
Chapter 24 Alarm list 0400
Error detail
aa : Calibrator No. (1-80)
bb : Position No. of STAT table (1-22)
Cup position is displayed in checking whether there are the cups in measure start with STAT
switch in Calibrator Barcode Operation.
Error detail
aa : QC No. (1-60)
bb : Position No. of STAT table (1-22)
Cup position is displayed in checking whether there are the cups in measure start with STAT
switch in Barcode QC Operation.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-109
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-110
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-111
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-112
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-113
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-114
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-115
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-116
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-117
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-118
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-119
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-120
Chapter 24 Alarm list 0400
9426 A hard disk controller reset was needed, but even that failed
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-121
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-122
Chapter 24 Alarm list 0400
9437 The user account has time restrictions and cannot be logged
onto at now
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-123
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-124
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-125
Chapter 24 Alarm list 0400
9446 An attempt was made to move the file pointer before the
beginning of the file
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-126
Chapter 24 Alarm list 0400
9450 The I/O operation was aborted due to either thread exit or
application request
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-127
Chapter 24 Alarm list 0400
9457 The file pointer cannot be set on the specified device or file
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-128
Chapter 24 Alarm list 0400
9460 The process cannot access the file because it is being used by
another
System Status
(1) Proceed to the next process or Unknown.
User Action
(1) Verify that the data is correct.
(2) Contact Technical Support Department.
Cause
(1) When running programs in the data processing unit, error occurs due to the hardware or software of the
computer.
es24_01.doc
M514 EL SOFTWARE MANUAL
24-129
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-130
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-131
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-132
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
24-133
Chapter 24 Alarm list 0400
es24_01.doc
M514 EL SOFTWARE MANUAL
26-1
Chapter 26 Maintenance diagnosis (User Maintenance)
* Activation method
Select using the DPR menu and start the operation with the DIAG switch.
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-2
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-3
Chapter 26 Maintenance diagnosis (User Maintenance)
Probe & Syringe replacing 8)SR outer wash valve (V10) & DI water pump (P10) &
Degasser (P3) Degassed supply pump(P11) is turned OFF.
< Explanations of action >
1)Power ON of DI water pump (P1) & degasser (P3) 1)DI water pump (P1) & Degasser (P3) is turned ON.
2)Turn ON inner wash back valve (V9). 2)Inner wash back valve (V9) is turned ON.
3)Turn ON degassed supply pump (P11). 3)Degassed supply pump (P11) is turned ON.
4)Turn OFF inner wash back valve (V9). 4)Inner wash back valve (V9) is turned OFF.
5)If the non-concentrated waste water isn’t full (when 5)If the non-concentrated waste water isn’t full (when
option exists), SR outer wash valve (V10) is turned ON. option exist), SR outer wash valve is turned ON. In the
I case non-concentrated waste water is full, error results.
6)If the non-concentrated waste water isn’t full (when 6)If the weak waste water isn’t full (when option exist), R
option exists), S inner wash valve (V5) is turned ON. In inner wash valve (V6) is turned ON. In the case non-
the case non-concentrated waste water is full, error concentrated waste water is full, it becomes error.
results. 7)R inner wash valve (V6) is turned OFF.
7)S inner wash valve (V5) is turned OFF. 8)SR outer wash valve (V10) & DI water pump (P1) &
Degasser (P3) & Degassed supply pump (P11) is turned
OFF.
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-4
Probe & Syringe replacing 10) SR outer wash valve (V10) & DI water pump (P10) &
Degasser (P3) Degassed supply pump(P11) is turned OFF.
11)If the upper sensor (S53) is OFF and lower point sensor (S54)
< The explanation for action >
is OFF, R dispense motor (M17) is moved for specified amount
1) Power ON of DI water pump (P1) & degasser (P3) at high speed for discharge. If the upper point sensor or
2) Turn ON inner wash back valve (V9). lower point sensor is ON, error results.
3) Turn ON degassed supply pump (P11). 12)Check the action end of R dispense motor (M17).
4) Turn OFF inner wash back valve (V9). If the mode is BUSY, error results.
5) If the non-concentrated waste water isn’t full (when option If the upper point sensor or lower point sensor
exist), SR outer wash valve (V10) is turned ON. is ON, error results.
6) If the upper sensor (S53) is OFF and lower point sensor (S54) is
OFF, R dispense motor (M17) is moved for specified amount for
high speed aspiration. If the upper point sensor or lower
point sensor is ON, error results.
7) If the non-concentrated waste water isn’t full (when option
exists), R inner wash valve (V6) is turned ON. In the
case non-concentrated waste water is full, error results
8) Check the action end of R dispense motor (M17).
If the mode is BUSY, error results.
If the upper point sensor or lower point sensor
is ON, error results.
9) R inner wash valve (V6) is turned OFF.
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-5
Chapter 26 Maintenance diagnosis (User Maintenance)
2. If the S probe rotation motor (M10) is in the wash position, the probe is moved left to the cuvette aspiration
position.
If the motor is in the cuvette aspiration position, the probe is moved right to the wash position.
In the case the action doesn’t end within 4 seconds, error results (when the corresponded sensor is OFF).
If the R rotation motor (M14) is in the wash position, the probe is moved right to the cuvette position.
If the motor is in the cuvette position, the probe is moved left to the wash position.
In the case the action doesn’t end within 4 seconds, error results (when the corresponded sensor is OFF).
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-6
Chapter 26 Maintenance diagnosis (User Maintenance)
n Wash nozzle
1. Contents
1) Photometry line is initialized.
If an error occurs, abnormal end results and the following action isn’t performed.
* If this action is selected continuously and ended normally, this action isn’t performed.
2) Negative pressure setting action is performed.
If an error occurs, abnormal end results and shifts to No.5.
3) If the cuvette position 80 isn’t empty, low speed aspiration action is performed.
If an error occurs, abnormal end results and shifts to No.5.
4) Wash action for one cuvette by detergent A or B or water is performed.
If an error occurs, abnormal end results and shifts to No.5.
5) Negative pressure cancel action is performed.
ES26_01.DOC
M514 EL SOFTWARE MANUAL
26-7
Chapter 26 Maintenance diagnosis (User Maintenance)
ES26_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-1
(manufacturer's maintenance/DIAG)
n DIAG outline
1. Outline
• Here, the general contents of the operations for maintenance operation, diagnosis operation, etc. will be
treated.
2. Operation Selection
• Operation selection is performed from the DPR menu.
• The selected operation is held until the next operation is selected.
• To deselect, operation selection cancellation (deselection) must be performed.
3. Operation Start
• Operation is started by DPR menu operation or with the ANL DIAG switch.
1) Operation for start by DPR menu operation
• Sub-CPU check
• Program version
• 0 - 100% (not continuous, continuous)
• Rotation measuring (not continuous, continuous)
2) Operation for start using the ANL DIAG switch
• All operations other than the above
4. Operation Completion
• Operation is completed after the previously arranged processing has been completed.
• The selected operation is held even after operation completion.
• For ANL, when operations require time for preparations etc., the preparations are performed with the first
operation, and control for return to the initial status is performed with the last operation.
1) Operations requiring explicit completion (deselection) at the time of operation completion
• Maintenance operation: Washing nozzle exchange
• Maintenance operation: Washing tube exchange
• Diagnosis operation: Washing sequence
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-2
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-3
(manufacturer's maintenance/DIAG)
l Sub-CPU check
• The following processing is performed, and each CPU responds as normal or abnormal.
For each processing, abnormal completion is made in case of an error. (Each CPU shall be independent.)
<Check of the photometry control sub-cpu>
1) Perform initialization of photometry control.
2) Perform initialization of the photometry constants.
<Check of the LAN master cpu>
1) Perform LAN master initialization.
2) Perform initialization of the mechanical control system.
l Sample counter
1) Output of 1 to the sample counter.
2) Waiting for 100 msec.
3) Output of 0 to the sample counter.
l Program version
• The version of the program being used is given as response.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-4
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-5
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-6
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-7
(manufacturer's maintenance/DIAG)
n Operation details
l Overall - reset
(1) Full mechanical initialization is performed.
l Overall - excitation OFF
• Initially, (1) is performed, and in case of consecutive selection, operation is performed in the order of (2), (1),
(2), ... .
(1) Excitation is switched OFF for rack supply (M1), ID rotation (M3), detection stepping (M4), aspiration
stepping (M5), rack recovery (M6), rack storage (M8), STAT rotation (M9), S rotation (M10), S vertical
(M11), S dispensing (M12), R rotation (M14), R vertical (M15), refrigerator rotation (M16), R dispensing
(M17), C rotation (M18), M rotation (M19), M vertical (M20), M mixing (M21), washing vertical (M22).
(2) Excitation is switched ON for the motors of (1).
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-8
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-9
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-10
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-11
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-12
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-13
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-14
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-15
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-16
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-17
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-18
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-19
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-20
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-21
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-22
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-23
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-24
(manufacturer's maintenance/DIAG)
l Reaction photometry
(1) Reset
1) Photometry system initialization is performed.
(2) Specified position rotation
• Stop position and rotation direction are specified from DPR.
1)
1) When the present position is not established or when the top point (S36) of S vertical (M11) is OFF
or when the top point (S46) of R vertical (M15) is OFF or when the top point (S60) of M vertical
(M20) is OFF or when the top point (S61) of washing vertical (M22) is OFF, photometry system
initialization is performed.
2) When the top point (S36) of S vertical (M11), the top point (S46) of R vertical (M15), the top point
(S60) of M vertical (M20), the top point (S61) of washing vertical (M22), and the top point (S57) of
C rotation (M18) all are ON, C rotation (M18) is performed from the present position to the specified
position as specified position left movement (CCW movement) or specified position right movement
(CW movement).
• An error is caused when the above sensors are OFF.
• An error is caused when the operation is not completed within 5 sec.
• An error is caused when the fixed point is OFF after operation completion.
(3) Pulse
• 1) is performed when the operation instruction is CW, and 2) is performed when the operation instruction
is CCW.
1) Pulse right movement is performed for the specified number of pulses.
2) Pulse left movement is performed for the specified number of pulses.
(4) Excitation ON/OFF
• 1) is performed for the first operation, and in case of following consecutive selection, the operation will be
2), 1), 2), ... .
1) The C rotation (M18) excitation is switched OFF.
2) The C rotation (M18) excitation is switched ON.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-25
(manufacturer's maintenance/DIAG)
l Reaction photometry
(5) 0 - 100%
• The following operations are performed according to the specification of single/continuous or cuvette.
• 1) is performed for the first operation, and in case of following consecutive selection after normal
completion, the operation will be from 2) on.
1) Photometry system initialization is performed.
• In case of an error, abnormal completion is made.
2) The photometry constants are initialized.
• In case of an error, abnormal completion is made.
3) 0% photometry is performed.
• The measured data are transmitted.
• In case of an error, abnormal completion is made.
4) C rotation (M18) initialization is performed.
• In case of an error, abnormal completion is made.
5) When S probe transfer is (not at the cuvette position (S34) or at the top point (S36)), R probe transfer is
(not at the cuvette position (S43) or at the top point (S46)), mixing rod transfer is at the top point (S60),
washing nozzle transfer is at the top point (S61), and the fixed point (S57) of C rotation (M18) is ON,
C rotation to the right is performed to the specified position.
• An error is caused when the operation is not completed within 5 sec or when it is completed
abnormally.
6) Static photometry is performed when the fixed point (S57) of C rotation (M18) is ON.
• An error is caused when the fixed point of C rotation is OFF.
• An error is caused when the operation is not completed within 1 sec or when it is completed
abnormally.
• The photometry data are transmitted.
* In case of continuous operation, 6) is performed 20 times.
7) C rotation (M18) initialization is performed.
• In case of an error, abnormal completion is made.
(6) Rotation photometry
• The following operations are performed according to the specification of single/continuous or cuvette.
• 1) is performed for the first operation, and in case of following consecutive selection after normal
completion, the operation will be from 2) on.
1) Photometry system initialization is performed.
• In case of an error, abnormal completion is made.
2) The photometry constants are initialized.
• In case of an error, abnormal completion is made.
3) When S probe transfer is (not at the cuvette position (S34) or at the top point (S36)), R probe transfer is
(not at the cuvette position (S43) or at the top point (S46)), mixing rod transfer is at the top point (S60),
washing nozzle transfer is at the top point (S61), and the fixed point (S57) of C rotation (M18) is ON,
C rotation to the right is performed to the specified position.
• An error is caused when the operation is not completed within 5 sec or when it is completed
abnormally.
4) DIAG photometry is performed when the fixed point (S57) of C rotation (M18) is ON.
• An error is caused when the fixed point of C rotation is OFF.
• An error is caused when the operation is not completed within 7 sec or when it is completed
abnormally.
• The photometry data are transmitted.
* In case of continuous operation, 6) is performed 20 times.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-26
(manufacturer's maintenance/DIAG)
l Mixing
(1) Reset
1) Mixing system initialization is performed.
(2) Vertical step
• The time-out for each operation shall be 2 sec.
• 1) is performed for the first operation, and in case of following operation selection, the operation will be
performed in the sequence of 2), 1), 2), ... .
1)
a) When the present position in regard to rotation and vertical direction is not established, mixing
system initialization is performed.
• In case of an error, abnormal completion is made and the following operations are not performed.
b) When the top point (S60) of M vertical (M20) is ON, downward movement is performed.
• An error is caused when the top point is OFF.
2) When the top point is OFF, upward movement is performed.
• An error is caused when the top point is ON.
• An error is caused when the top point is OFF after operation completion.
(3) Rotation step
• The time-out for completion of each operation shall be 2 sec.
1)
1) When the present position in regard to rotation and vertical direction is not established, mixing
system initialization is performed.
• In case of an error, abnormal completion is made and the following operations are not performed.
2) When the top point (S59) of M rotation (M19) is ON, right movement is performed.
• An error is caused when the fixed point is OFF.
• An error is caused when the fixed point is OFF after operation completion.
(4) Vertical pulse
• 1) is performed when the operation instruction is CW, and 2) is performed when the operation instruction
is CCW.
1) Pulse upward movement is performed for the specified number of pulses.
2) Pulse downward movement is performed for the specified number of pulses.
(5) Rotation pulse
• 1) is performed when the operation instruction is CW, and 2) is performed when the operation instruction
is CCW.
1) Pulse right movement is performed for the specified number of pulses.
2) Pulse left movement is performed for the specified number of pulses.
(6) Mixing rod rotation
1) The mixing rod (M21) performs mixing.
• An error is caused when the operation is not completed within 2 sec.
(7) Sequence
• At the first selection of this operation, all operations are performed from 1) on.
When continuous operation is selected afterwards and the operation has been completed normally, 1) is
not performed.
1) Mixing system initialization is performed.
C rotation (M18) initialization is performed.
• In case of an error, abnormal completion is made and the following operations are not performed.
2) The detergent washing of “2.4.6. Mixing Rod Washing” is performed.
• In case of an error, abnormal completion is made and the following operations are not performed.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-27
(manufacturer's maintenance/DIAG)
l Washing
(1) Reset
1) Washing vertical (M22) initialization is performed.
(2) Vertical step
• The time-out for completion of each operation shall be 2 sec.
• 1) is performed for the first operation, and in case of following operation selection, the operation will be
performed in the sequence of 2), 1), 2), ... .
1)
a) Washing vertical (M22) initialization is performed.
• In case of an error, abnormal completion is made and the following operations are not performed.
b) When the top point (S61) of washing vertical (M22) is ON, downward movement is performed.
• An error is caused when the top point is OFF.
• An error is caused when the bottom point (S62) is OFF after operation completion.
2) When the bottom point (S62) is ON, upward movement is performed.
• An error is caused when the bottom point is ON.
• An error is caused when the top point (S61) is OFF after operation completion.
(3) Vertical pulse
• 1) is performed when the operation instruction is CW, and 2) is performed when the operation instruction
is CCW.
1) Pulse upward movement is performed for the specified number of pulses.
2) Pulse downward movement is performed for the specified number of pulses.
(4) Sequence
• At the first selection of this operation, 1) is performed, and then 2) is performed.
When continuous operation is selected afterwards and the operation has been completed normally, 1) is
not performed. 3) is performed at the time of operation completion.
1) Washing vertical (M22) initialization is performed.
C rotation (M18) initialization is performed.
Negative pressure setting is performed.
• In case of an error, abnormal completion is made, and the following operations are not performed.
2) Low-speed aspiration operation is performed when the cuvette position 80 is not empty.
When it is empty, the sequence for one cycle at the time of analysis operation is performed.
• In case of an error, abnormal completion is made, 3) is executed, and the operation is ended.
3) Negative pressure cancellation operation is performed.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-28
(manufacturer's maintenance/DIAG)
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-29
(manufacturer's maintenance/DIAG)
n ISE DIAG
l Types
ISE diagnosis operations are divided roughly into mechanism-related, data-related, and priming measuring
operation. Division according to the operation start method is made into the following two types.
Priming : Selection from the DPR menu and start by a switch on the ANL side.
Others : Selection/start from the DPR menu.
The operation details are described in the following.
l Mechanism-related
• A time-out until completion is set for each operation, and the completion information is gathered or error
completion is performed.
(1) Reset
• ISE of “2.2.3. Mechanical System Initial Operation” is performed.
(2) Sample mixing
1) Mixing operation (MIX) is performed.
(3) Overflow sensor check
1) Start check 2 (STARTCHECK2) is performed.
(4) Thermistor check
1) Start check 1 (STARTCHECK1) is performed.
l Data-related
• A time-out until completion is set for each operation, and the completion information is gathered or error
completion is performed.
(1) Calibration (serum)
1) “Serum CAL” is performed.
• The same sync signal as for washing before analysis (FLUSH_1) is put out.
• From the S-L position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 20 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• From the S-H position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 20 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• Washing after analysis (FLUSH_2) is performed.
2) The measuring data are put out.
(2) Calibration (urine)
1) “Urine CAL” is performed.
• The sync signal for washing before analysis (FLUSH_1) is put out.
• From the U-L position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 25 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• From the U-H position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 25 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• Washing after analysis (FLUSH_2) is performed.
2) The measuring data are put out.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-30
(manufacturer's maintenance/DIAG)
n ISE DIAG
(3) Calibration (serum + urine)
1) “Serum + urine CAL” is performed.
• The sync signal for washing before analysis (FLUSH_1) is put out.
• From the U-L position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 25 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• From the U-H position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 25 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• Washing after analysis (FLUSH_2) is performed.
• The sync signal for washing before analysis (FLUSH_1) is put out.
• From the S-L position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 20 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• From the S-H position at the STAT inner circumference, mixture is dispensed with a liquid quantity
of 20 µL, a diluent quantity of 10 µL, and 1 µL of air, and the sync signal is put out. (This is done
four times.)
• Washing after analysis (FLUSH_2) is performed.
2) The measuring data are put out.
(4) Selectivity check
• The selectivity check of “2.5.7. ISE” is performed.
l Priming
• A time-out until completion is set for each operation, and the completion information is gathered or error
completion is performed.
(1) Full priming
• Full priming of “2.4.5. ISE” Washing is performed.
(2) Buffer priming
1) Buffer liquid priming (PRIME_2) is performed.
(3) Internal standard solution priming
1) MID/REF priming (PRIME_1) is performed.
(4) Fluid discharge (electrode)
1) Fluid discharge (via the flow cell) (PRIME_3) is performed.
(5) Fluid discharge (bypass)
• Fluid discharge (bypass) of “2.5.7. ISE” is performed.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 27 Maintenance and Diagnosis 27-31
(manufacturer's maintenance/DIAG)
n ISE DIAG
l Measuring operation
• A time-out until completion is set for each operation, and the completion information is gathered or error
completion is performed.
(1) One-sample measuring (rack)
1) When not at the initial position, sample system aspiration stepping (M5) initialization operation is
performed.
2) With sample aspiration as the rack position, sample dispensing and ISE measuring are performed in the
same way as for analysis operation, and analysis of the urine or serum is executed.
3) The measuring data are put out.
(2) One-sample measuring (STAT)
1) When not at the initial position, sample system initialization operation is performed.
2) With sample aspiration as the STAT outer position, the same operations as 2) and 3) of (1) are
performed.
(3) Continuous measuring 1 (rack)
1) When not at the initial position, sample system aspiration stepping (M5) initialization operation is
performed.
2) Washing before analysis (FLUSH_1) is performed.
3) With sample aspiration as the rack position, sample dispensing and ISE measuring are performed in the
same way as for analysis operation, and urine (M1_U) or serum (M1_S) analysis is executed.
4) 3) is performed nine times (for a total of 10 times).
5) Washing after analysis (FLUSH_2) is performed.
6) The measuring data are put out.
(4) Continuous measuring 1 (STAT)
1) When not at the initial position, sample system initialization operation is performed.
2) With sample aspiration as the STAT outer position, the same operations as 2) to 6) of (3) are performed.
(5) Continuous measuring 2
1) Washing before analysis (FLUSH_1) is performed.
2) Internal standard solution measuring (M_2) is performed.
3) 2) is executed nine times (for a total of 10 times).
4) Washing after analysis (FLUSH_2) is performed.
5) The measuring data are put out.
(6) Continuous measuring 3
1) Washing before analysis (FLUSH_1) is performed.
2) Static measuring (M_3) is performed.
3) 2) is executed nine times (for a total of 10 times).
4) Washing after analysis (FLUSH_2) is performed.
5) The measuring data are put out.
ES27_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-1
Note : In case of urine + serum (with PRE-check), ANL performs urine CAL (with PRE-check)
and then serum CAL (with PRE-check).
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-2
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-3
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-4
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-5
The data from (1) to (8) out of the voltage data obtained with the above calibration procedure are used.
The third and forth voltage mean values are calculated for STD-H, for STD-L, and for STD-M (internal
standard solution).
EH : Mean voltage of (1) and (3) (STD-H voltage value)
EMH : Mean voltage of (2) and (4) (STD-M voltage value)
EL : Mean voltage of (5) and (7) (STD-L voltage value)
EML : Mean voltage of (6) and (8) (STD-M voltage value)
The following formula is used to calculate the slope. S = 0 is used when the stability check is NG.
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-6
The factor with the internal standard solution density CM given by the analysis conditions message is
calculated.
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-7
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-8
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-9
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-10
3) FLUSH-2
FLUSH-2 is performed.
H. Serum CAL (without stability check/H-L STD PRE-check)
The following operations are performed, and CAL data are calculated in the same way as for urine CAL
without PRE-check.
1) Serum HIGH-STD measuring
The serum HIGH-STD is measured four times as follows in the same way as for serum CAL without
PRE-check.
2) Serum LOW-STD measuring
The serum LOW-STD is measured four times as follows in the same way as for serum CAL without
PRE-check.
3) FLUSH-2
FLUSH-2 is performed.
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-11
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-12
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-13
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-14
ES28_01.DOC
M514 EL SOFTWARE MANUAL
Chapter 28 ISE Maintenance, Diagnosis Outline
28-15
ES28_01.DOC